Entry  Wed Apr 1 15:47:48 2009, Yoichi, Update, Locking, 3.8kHz peak looks like a real optical response of the interferometer CARM-ETM-EXC.pngAOpath-TFs.png
Yoichi, Peter

To see where the 3.8kHz peak comes from, we locked the interferometer with the CARM fed back only to ETM and increased the arm power to 4.
Entry  Wed Apr 1 16:14:36 2009, Yoichi, Update, Locking, 3.8kHz peak does not change with SRC offset 
Yoichi, Peter

We suspected that maybe the 3.8kHz peak is the DARM RSE somehow coupled to the CARM.
Entry  Thu Mar 26 04:27:26 2009, Yoichi, Update, Locking, 3.8kHz peak as a function of the arm power CARM-PODC.pdf
During the power ramp-up, I actuated CARM using ETMs and measured the transfer functions to the PO_DC at several arm powers.
The peak grows rapidly with the power. It also seems like the frequency shifts slightly as the power goes up, but not much.
Entry  Thu Jun 26 22:08:33 2008, Jenne, Update, Electronics, 3.7MHz bandstop filter in MC Servo  MCLoopGainBoth.pngTP1ASpectrumBoth.pngQmonWithFilt.pngMCOut2WithFilt.png
The 3.7MHz elliptical bandstop filter that I made during my SURF summer is now installed in the MC servo loop to reduce the noise at 3.7MHz. 

I have taken transfer functions with and without the filter between TP1A and TP2A, with EXCA at -20dBm, using the HP4195A Network Analyzer.  I have also
Entry  Fri May 3 08:55:55 2013, Steve, Update, PEM, 3.2 M earthquake 3.2eqChannelIland.png3.2eq.png
 
Entry  Wed Mar 17 10:52:48 2021, Jordan, Update, SUS, 3" Ring Adpater for SOS  Side_View.pngFront_View.pngRing_with_Modifed_Bracket.pngBack_of_ring.pngFront_of_Ring.png
I have added a .1", 45deg chamfer to the bottom of the adapter ring. This was added for a new placement of the eq stops, since the barrel screws
are hard to access/adjust.

This also required a modification to the eq stop bracket, D960008-v2, with 1/4-20 screws angled at 45 deg to line up with the chamfer.
    Reply  Mon Jan 13 11:08:48 2014, Steve, Update, PSL, 3 good days of IOO pointing 3gdPSLpointing.png
 Three good days of IOO pointing: Friday, Sat and Sun    What was changed?  May be the clamping on Friday?
IOO vertical changes recovering as tempeture. IP is clipping at plastic enclosure of ETMY
 
    Reply  Mon Aug 29 15:15:46 2022, Tega, Update, Computers, 3 FEs from LLO got delivered today IMG_20220829_145533452.jpgIMG_20220829_144801365.jpg
[JC, Tega]

We got the 3 front-ends from LLO today. The contents of each box are:


FE machine
OSS adapter card for connecting to I/O chassis
PCI riser cards (x2)
Timing Card and cable
Power
    Reply  Tue Aug 30 15:21:27 2022, Tega, Update, Computers, 3 FEs from LHO got delivered today IMG_20220830_144925325.jpgIMG_20220830_142307495.jpgIMG_20220830_143059443.jpg
[Tega, JC]

We received the remaining 3 front-ends from LHO today. They each have a timing card and an OSS host adapter card installed. We also receive 3
dolphin DX cards. As with the previous packages from LLO, each box contains a rack mounting kit for the supermicro machine.
    Reply  Mon Nov 21 16:08:04 2011, kiwamu, Update, Green Locking, 2nd trial of Y arm ALS noise budget : broad band noise gone Yarm_ALS_2011Nov19.png


    
        
            Quote from #5930
        
    
   
            
            Right now the fluctuation
Entry  Mon Aug 10 14:58:01 2009, Jenne, Update, PEM, 2nd set of Guralp channels plugged into ADCU 
The second set of Guralp channels is now plugged into the PEM ADCU, into channels which are confirmed to be working.  (Method: 1Vpp sine wave into
channel, check with DataViewer).
 
    Reply  Mon Aug 10 18:12:25 2009, Jenne, Update, PEM, 2nd set of Guralp channels plugged into ADCU 


    
        
            Quote:
        
        
            
            The second set of Guralp channels is now plugged into the PEM ADCU, into channels which are confirmed to be working. 
Entry  Fri Oct 8 07:41:42 2010, steve, Configuration, SAFETY, 2W laser shutter is closed 
The 2 W Innolight shutter is closed and enclosure door removed. Beam path blocked. Do not change this condition.
    Reply  Fri Oct 8 10:38:03 2010, steve, Configuration, SAFETY, 2W laser shutter is closed P1060905.JPGP1060910.JPG


    
        
            Quote:
        
        
            
            The 2 W Innolight shutter is closed and enclosure door removed. Beam path blocked. Do not change this condition.
Entry  Wed May 5 16:58:21 2010, Koji, Update, PSL, 2W hooked up to the interlock service 
Ben, Steve, and Koji
Ben came to the 40m and hooked up a cable to the main interlock service.
We have tested the interlock and confirmed it's working.
Entry  Sat Apr 24 15:05:41 2010, Kevin, Update, PSL, 2W Vertical Beam Profile vbp.jpg
The vertical beam profile of the Innolight 2W laser was measured at eight points along the axis of the laser.
These measurements were made with the laser crystal temperature at 25.04°C and the injection current at 2.091A. z is the distance from the
razor blade to the flat black face of the front of the laser.
Entry  Wed Jun 2 22:58:04 2010, Kevin, Update, PSL, 2W Second Reflected Beam Profile profile_2nd.png
[Koji, Kevin]
The profile of the Innolight 2W was previously measured by measuring the reflected beam from the front surface of a W2 window (see
entry). To investigate thermal effects, Rana suggested also measuring the profile of the beam reflected from the back surface of the W2.
Entry  Thu Oct 14 23:33:45 2010, Suresh, Update, IOO, 2W NPRO laser output power versus temperature tempscan.pngScan2010.zip
Steve measured an apparent power drop in the 2W NPRO output from 2.1W to 1.6W(elog entry no 3698)
at 2.1A of diode current in the laser (elog entry:  2822).  It was later noticed that the laser temperature was set to about 45 degC while
    Reply  Fri Oct 15 00:15:52 2010, Koji, Update, IOO, 2W NPRO laser output power versus temperature 
From the plot, you observed the reduction of the output power only by 1% between 25deg to 45deg.
This does not agree with the reduction from 2.1W to 1.6W.
Is there any cause of this discrepancy?
Entry  Fri Oct 15 21:25:23 2010, Suresh, Update, IOO, 2W NPRO laser output power drop question 
  The power meter used in the measurements of elog entries 2822, 3698 and 3709 was the Ophir PD300-3W.  This power head has several damaged
patches  and a slight movement of the laser spot changes the reading considerably.  To verify I checked the power out with another power meter
(the Vector S310) and found that there is no significant variation of the power output with the temperature of the laser.  And the power at 2.1A of
Entry  Mon Aug 30 23:21:36 2010, rana, HowTo, PSL, 2W NPRO Mount designed with emachineshop.com software LaserMount.png
To test out this website - emachineshop.com, Jenne and I are designing some of the mounts for the new beam height.
    Reply  Wed Sep 1 08:36:59 2010, Koji, HowTo, PSL, 2W NPRO Mount designed with emachineshop.com software 
1. I can not see whether the attaching surface is flat or not.
It should have ~1mm step to avoid "the legs" of the laser at the four corners.
Otherwise we will have ~0.5mm space between the block and the laser 
Entry  Thu Jun 3 00:47:17 2010, Kevin, Update, PSL, 2W Beam Profile of Second Reflected Beam profile_2nd.png
[Koji, Kevin]

The profile of the Innolight 2W was previously measured by measuring the reflected beam from the front surface of a W2 window (see
Entry  Wed Jun 2 03:24:22 2010, Kevin, Update, PSL, 2W Beam Profile profile.pngerrors.pngLayout.jpg
[Rana, Kiwamu, Kevin]
The Innolight 2W beam profile was measured with the beam scan. A W2-IF-1025-C-1064-45P window was used to reflect a small amount of the main beam.
A 5101 VIS mirror was used to direct just the beam reflected from the front surface of the W2 down the table (the beam reflected from the back surface
    Reply  Wed Jun 2 04:27:02 2010, Koji, Update, PSL, 2W Beam Profile 
This is what I already told to Kevin and Rana:
A direct output beam is one of the most difficult measurements for the mode profiling.
I worried about the thermal lensing.
Entry  Wed Jun 25 12:25:28 2008, Eric, Summary, Cameras, 2D Gaussian Fitting Code 
I initially wrote a script in MATLAB that takes pictures of the laser beam's profile and fits them to a two dimensional gaussian in order to determine the
position and width of the beam.  This code is now (mostly) ported to C so that it can be imbedded in the camera software package that Joe is writing. 
The fitting works fairly well for pictures with the beam directly incident on the camera, and less well for pictures of scatter off the end mirrors of
    Reply  Sun Feb 5 21:56:50 2017, Lydia, Update, IMC, 29.5 MHz stabilizer box replacement 
Since the "stablizer box" doesn't really need to stabilize, it just needs to amplify, I decided to replace it with an off the shelf amplifier
we already had, ZHL-2. I worked on getting it set up today, but didn't connect
anything so that people have a chance to give some feedback. 
    Reply  Tue Feb 7 12:01:10 2017, Lydia, Update, IMC, 29.5 MHz stabilizer box replacement RF_amp.pdf
I tested the amplifier with the Agilent network analyzer and measured 19.5 dB of gain between 29 and 30 mHz. The phase only changed by 1 degree
over this same 1 MHz span. Since everything seems to be in order I'll hook it up this afternoon, unless there are any objections. 
    Reply  Tue Feb 7 17:00:55 2017, Lydia, Update, IMC, 29.5 MHz stabilizer box replacement 
I set everything up and connected it as shown on the block diagram attached to the previous entry, with the exception of the DC power. This
is becuase there is no place open to connect to on the DIN rail where the DC power is distributed, so the +24V power will have to be shut off to the other
equipment in 1X1 before we can connect the amplifier. (The amplifier is in 1X2, but the DC power distribution was more accessible in 1X1.) I
    Reply  Fri Feb 10 02:14:10 2017, gautam, Update, IMC, 29.5 MHz stabilizer box replacement 
Lydia finished up installing the new RF amplifier, and will elog the details of the installation.

I wanted to try and measure the IMC OLG to compare against my Simulink model. So I went about performing a few checks. Summary of my findings:


The amplifier seems to be working fine. I checked powers at the input, output to EOM and output to distribution box (that serves the
    Reply  Fri Feb 10 11:41:43 2017, Lydia, Update, IMC, 29.5 MHz stabilizer box replacement 
To install the replacement amplifier, I did the following:


Mounted the amplifier in a 2U chassis, with a metal plate between the amplifier and the bottom of the box. The plate is separated from
the box and the amplifier with 2 sets of Nylon screws. I did it this way to make use of the holes that were already in the chassis bottom and just drill
    Reply  Fri Feb 10 13:24:28 2017, rana, Update, IMC, 29.5 MHz stabilizer box replacement 
To remind myself about how to put filter caps on the mini-circuits RF Amps, I looked at Koji's recent
elog. Its mostly about op-amps, but the idea holds for us.

We want a big (~100 uF) electrolytic with a 50V rating for the +24V RF Amp. And then a 50V ceramic capacitor of ~0.1 uF close to the pins. Remember
Entry  Wed Jan 25 15:41:29 2017, Lydia, Update, IMC, 29.5 MHz modulation depth measurement plan 
[Lydia, gautam]

To measure the modulation depth of the 29.5 MHz sideband, we plan to connect a bidirectional coupler between the EOM and the triple resonant
circuit box. This will let us measure the power going into the EOM and the power in the reflection. According to the manual for the EOM (Newport 4064),
    Reply  Wed Jan 25 17:30:03 2017, Koji, Update, IMC, 29.5 MHz modulation depth measurement plan 
I'm afraid that the bidirectional coupler, designed to be 50ohm in/out, disturbs the resonant circuit designed for the EOM which is almost purely
capacitive.

One possible way could be to measure the transfer function using the active FET probe from the triple resonant input to the output with the EOM
    Reply  Wed Jan 25 19:39:07 2017, gautam , Update, IMC, 29.5 MHz modulation depth measurement plan 
Just collecting some links from my elog searching today here for easy reference later.


EOM datasheet: Newfocus 4064 (according to this, the input
Impedance is 10pF, and can handle up to 10W max input RF power).
An elog thread with some past measurement details: elog
    Reply  Fri Jan 27 21:25:11 2017, Lydia, Update, IMC, 29.5 MHz modulation depth  
[gautam, Lydia]

We set out to measure the 29.5 MHz power going to the EOM today but decided to start by looking at the output of the RF AM stabilizer box first.
We wanted to measure the AM noise with a mixer, so we needed to know the power it was giving. We looked at the ouput that goes to the power combiner on
    Reply  Sat Jan 28 01:25:51 2017, gautam, Update, IMC, 29.5 MHz modulation depth  IMC_mod.pdfLockMCNew.png
Some more details of our investigation:


Here is a spectrum of the signal to the power combiner on the PSL table, measured on the output of the RF AM Stabilization box.
Entry  Thu Apr 15 14:47:28 2010, steve, Update, Electronics, 25MHZ oscillation of HP4195A P1060246.JPGP1060249.JPG
The 1979 vintage RF spectrum analyzer HP4195A  sn2904J01587 shipped out  for repair today to http://www.avalontest.com
It has a 25 MHZ oscillation when you go  below 150 MHZ in your sweep....atm1 with the larger amplitude shows this 25 MHZ
Atm2 is displaying  full sweep-sign scans from 1 to 500 MHZ.....here one can clearly see the three segment of the scan:
Entry  Thu Jun 17 22:19:12 2021, Koji, Update, Electronics, 25 HAM-A coil driver units delivered P_20210617_195811.jpg
25 HAM-A coil driver units were fabricated by Todd and I've transported them to the 40m.
 2 units we already have received earlier.
The last (1) unit has been completed, but Luis wants to use it for some A+ testing. So 1 more unit is coming.
    Reply  Fri Dec 2 16:40:29 2016, gautam, Update, IMC, 24V fuse pulled out 
I've pulled out the 24V fuse block which supplies power to the AOM RF driver. The way things are set up on the PSL table, this same voltage source
powers the RF amplifiers which amplify the green beatnote signals before sending them to the LSC rack. So I turned off the green beat PDs before pulling
out the fuse. I then disconnected the input to the RF driver (it was plugged into a DS345 function generator on the PSL table) and terminated it with a
Entry  Wed Jul 2 13:56:10 2008, Rob, Yoichi, John, Update, Locking, 24.5 Hz resonance locking080702.png
Work continues on trying to reduce the CARM offset using dc signals from PO_DC. Got up to arm powers of
~35 last night.
Entry  Mon Mar 5 20:31:33 2012, Koji, Update, LSC, 22/110MHz path for POP 
This a kind of self record...
We need an RF setup at POP to extract 22 and 110 MHz components separately.
I am planning to work on this in the daytime on Tuesday.
    Reply  Tue Mar 6 22:23:04 2012, Koji, Update, LSC, 22/110MHz path for POP 
 
 As par Kiwamu's request, RF filters for POP22 and POP110 were installed. They are not really nice. We need to replace it with more fancy
electronics.
    Reply  Wed Mar 7 15:56:36 2012, Koji, Update, LSC, 22/110MHz path for POP RFseparator.pngimpedance.pdftransmission.pdf
The RF separator installed comprises of the Minicircuits filters cascaded as in the figure below. 
This has one input and 4 output ports for 11, 22, 30-60, and 110MHz signal. 
As seen in this entry #6167, we have 22 and 110MHz signals together with 11, 44, 66MHz
    Reply  Wed Feb 27 20:50:41 2013, Manasa, Update, LSC, 22/110MHz path for POP POP.png
[Yuta, Manasa]

Modified POP path.
    Reply  Tue Apr 2 23:14:02 2013, rana, Update, LSC, 22/110MHz path for POP 


    
        
            Quote:
        
        
            

           We could not find a power supply slot for the amplifiers on the LSC rack. We had to put a temporary power
Entry  Wed May 12 23:31:17 2010, Alberto, Update, 40m Upgrading, 216 MHz resonance in the POY11 PD killed POX11.png2010-05-12_POY11_CalibratedOpticalResponse0-500MHz.png
It turned out that the resonance at 216 MHz in the 11MHz PD that I showed in the elog entry 2902
was casued by an instability of the of the MAX4107 opamap' feedback loop.
As the datasheet of the opamp shows, the close-loop gain has a peak
at about 200-250MHz, in presence of even small capacitive loads.
In my case, perhaps either the capacitance of the BNC cables plugged to the
    Reply  Thu May 13 05:06:43 2010, rana, Update, 40m Upgrading, 216 MHz resonance in the POY11 PD killed 
 
 This idea was tried before by Dale in the ~1998 generation of PDs. Its OK for damping a resonance, but it has the unfortunate consequence
of hurting the dynamic range of the opamp. The 100 Ohm resistor reduces the signal that can be put out to the output without saturating the 4107.
Entry  Fri Jan 12 20:35:34 2018, Udit Khandelwal, Summary, General, 2018/01/12 Summary 

40m Lab CAD


Worked
further on positioning vacuum tubes and chambers in the building.
Next step
would be to find some drawings for optical table positions and vibration isolation stack. Need help with this! 


    Reply  Wed Mar 30 16:40:24 2016, ericq, Update, LSC, 2016 vs 2010 2016_vs_2010.png
I haven't found any data files for the DARM spectrum of the previous generation of 40m, but with some GIMP-fu, I have plotted Monday's spectrum
(green) on top of one of the figures from Rob's thesis.