    Reply  Thu Jan 7 21:16:22 2021, gautam, Update, ALS, Noisy ALS 
I thought about it, but wouldn't that show up at the AUX PDH error point? Or because the loop gain is so high there we wouldn't see a small excess?
I suppose there could be some clipping on the Faraday or something like that. But the GTRX level and the green REFL DC level when locked are nominal.




Quote:


    Reply  Thu Jan 7 23:25:19 2021, Koji, Update, ALS, Noisy ALS 
If the sensing noise level of the end PDH degraded for some reason, it'd make the out of loop stability worse without making the end pdh error level
degraded.
It's just speculation.
    Reply  Fri Jan 8 20:01:11 2021, gautam, Update, ALS, Noisy ALS ALSX_IR_green.pdf
I did this test today. The excess noise around 100 Hz doesn't show up in the green beat.

See Attachment #1. The setup
was as usual:
Entry  Wed Mar 3 23:23:58 2021, gautam, Update, ALS, Arm cavity scan armScan.pdf
I see no evidence of anything radically different from my PSL table optical characterization
in the IMC transmitted beam, see Attachment #1.
The lines are just a quick indicator of what's what and no sophisticated peak fitting has been done yet (so the apparent offset between the transmission
Entry  Tue May 25 17:42:11 2021, Anchal, Paco, Summary, ALS, ALS Single Arm Noise Budget ALS_IR_b.svgALS_Single_Arm_IR.pdf
Here is our first attempt at a single-arm noise budget for ALS.

Attachment 1 shows the loop diagram we used to calculate the contribution of different noises.

Attachment 2 shows the measured noise at C1:ALS-BEATX_PHASE_FINE_OUT_HZ when XARM was locked to the main laser and Xend Green
    Reply  Thu May 27 11:03:15 2021, Anchal, Paco, Summary, ALS, ALS Single Arm Noise Budget ALS_Single_X_Arm_IR.pdfALS_OOL_with_Ref.pdf
Here's an updated X ARM ALS noise budget.

Things to remember:


Major mistake we were making earlier was that we were missing the step of clicking 
'Set Phase UGF' before taking the measurement.
Click the clear phase history just before taking measure.
Make sure
    Reply  Fri May 28 17:32:48 2021, Anchal, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat SingleArmActCalwithIRALSBeat.pdfstateSpaceModel.zip
I attempted a single arm actuation calibration using IR beatnote (in the directions of soCal idea for DARM calibration)


Measurement and Inferences:


I sent 4 excitation signals at C1:SUS-ITM_LSC_EXC wit 30cts at 31Hz,
200cts at 197Hz, 600cts at 619Hz and 1000cts at 1069 Hz.
These were sent simultaneously using compose function in python awg.
The
    Reply  Tue Jun 1 16:55:32 2021, Anchal, Paco, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat SingleArmActCalwithIRALSBeat-1306624785.pdf
Rana suggested in today's meeting to put in a notch filter in the XARM IR PDH loop to avoid suppressing the excitation line. We tried this today
first with just one notch at 1069 Hz and then with an additional notch at 619 Hz and sent two simultaneous excitations.


Measurement and Analysis:


We added notch filters with Q=10, depth=50dB, freq=619 Hz and 1069 Hz
    Reply  Tue Jun 8 11:40:53 2021, Anchal, Paco, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat ITMX_Cal_Noise_Spectrum_1307143423.pdf
We attempted to simulate "oscillator based realtime calibration noise monitoring" in offline analysis with python. This helped us in finding
about a factor of sqrt(2) that we were missing earlier in 16171. we measured C1:ALS-BEATX_FINE_PHASE_OUT_HZ_DQ
when X-ARM was locked to main laser and Xend green laser was locked to XARM. An excitation signal of amplitude 600 was setn at 619 hz at C1:ITMX_LSC_EXC.
    Reply  Wed Jun 9 18:29:13 2021, Anchal, Paco, Summary, ALS, Check for saturation in ITMX SOS Driver 
We did a quick check to make sure there is no saturation in the C1:SUS-ITMX_LSC_EXC analog path. For this, we looked at the SOS driver
output monitors from the 1X4 chassis near MC2 on a scope. We found that even with 600 x 10 = 6000 counts
of our 619 Hz excitation these outputs in particular are not saturating (highest mon signal was UL coil with
    Reply  Fri Jul 9 15:39:08 2021, Anchal, Summary, ALS, Single Arm Actuation Calibration with IR ALS Beat [Correction] ITMX_calibration_With_ALS_Beat.pdf
I did this analysis again by just doing demodulation go 5s time segments of the 60s excitation signal. The major difference is that I was not summing
up the sine-cosine multiplied signals, so the error associated was a lot more. If I simply multpy the whole beatnote signal with digital LO created at
excitation frequency, divide it up in 12 segments of 5 s each, sum them up individually, then take the mean and standard deviation, I get the answer as:
Entry  Wed Sep 15 23:38:32 2021, Koji, Update, ALS, ALS ASX PZT HV was off -> restored P_20210915_151333.jpg
It was known that the Y end ALS PZTs are not working. But Anchal reported in the meeting that the X end PZTs are not working too.

We went down to the X arm in the afternoon and checked the status. The HV (KEPCO) was off from the mechanical switch. I don't know this KEPCO
has the function to shutdown the switch at the power glitch or not.
    Reply  Wed Jun 1 11:56:28 2022, yuta, Update, ALS, Shutter driver for GRY replaced Screenshot_2022-06-01_12-01-53.png
[JC, Yuta]

We replaced a shutter driver for GRY since it stopped working this morning.
Entry  Fri Jun 24 17:16:59 2022, Paco, Update, ALS, XAUX cable in control room 
[JC, Paco]

We took the long BNC cable that ran from ETMX to ETMY and ran it from ETMX into the control room instead. This way Cici and Deeksha can send
small voltage signals to the AUX PZT and read back using the beatnote pickoff that is usually connected to the spectrum analyzer.
Entry  Tue Jun 28 16:59:35 2022, Paco, Summary, ALS, ALS beat allan deviation (XARM) xbeat_1340469968.pdffrac_avar_xbeat.pdf
[Paco]

I took ~ 7 minutes of XALS beatnote data with the XAUX laser locked to the XARM cavity, and the XARM locked to PSL to develop an allan deviation
estimator. The resulting timeseries for the channel C1:ALS-BEATX_FINE_PHASE_OUT_HZ_DQ (decimated
    Reply  Tue Jun 28 18:53:16 2022, rana, Summary, ALS, ALS beat allan deviation (XARM) 
what's the reasoning behind using df/f_beat instead of df/f_laser ?




Quote:



[Paco]
    Reply  Wed Jun 29 14:28:06 2022, Paco, Summary, ALS, ALS beat allan deviation (XARM) frac_adev_xbeat.pdf
I guess it didn't make sense since f_beat can be arbitrarily moved, but the beat is taken around the PSL freq ~ 281.73 THz. Attachment #1
shows the overlapping tau allan deviation for the exact same dataset but using the python package allantools,
where this time I used the PSL freq as the base frequency. This time, I can see the minimum fractional deviation of 1.33e-13 happening
Entry  Thu Jun 30 18:06:22 2022, Paco, Update, ALS, Optimum ALS recovery - part I als_ool_06_2022.png
[Paco]

In the morning I took some time to align the AUX beams in the XEND table. Later in the afternoon, I did the same on the YEND table. I then locked
the AUX beams to the arm cavities while they were stabilized using POX/POY and turned off the PSL hepa off temporarily (this
Entry  Tue Jul 12 19:46:54 2022, Paco, Summary, ALS, How (not) to take NPRO PZT transfer function 
[Paco, Deeksha, rana]

Quick elog for this evening:


Rana disabled MC servo .
Slow loop also got disengaged.
AUX PSL beatnote is best taken with *free running lasers*
Entry  Wed Aug 31 12:57:07 2022, rana, Summary, ALS, control of ALS beat freq from command line -easy Untitled.png
The PZT sweeps that we've been making to characterize the ALS-X laser should probably be discarded - the DFD was not setup correctly for this during
the past few months.

Since the DFD only had a peak-peak range of ~5 MHz, whenever the beat frequency drifts out of the linear range (~2-3 MHz), the data would have
Entry  Fri Sep 2 15:40:25 2022, Anchal, Summary, ALS, DFD cable measurements ICableOpenEnd_2022-09-02_3_08_34_PM.pngICableShortedEnd_2022-09-02_3_04_43_PM.pngQCableOpenEnd_2022-09-02_3_09_49_PM.pngQCableShortedEnd_2022-09-02_3_00_55_PM.pngDFD_Delay_Measurement.pdf
[Anchal, Yehonathan]

I laid down another temporary cable from Xend to 1Y2 (LSC rack) for also measuring the Q output of the DFD box. Then to get a quick measurement
of these long cable delays, we used Moku:GO in oscillator mode, sent 100 ns pulses at a 100 kHz rate from one end, and measured the difference between
Entry  Wed Jul 16 09:27:17 2008, steve, Update, ALARM, PSL-FSS_RMTEMP alarm is false frmtemp2y.jpg
Morning alarm sound is good for people who does not drink coffee.
Our 40m alarm is on every morning.
Those whom are not here in the morning thinks that this beeping sound is inspirational.
Entry  Thu May 21 12:58:32 2009, rob, Configuration, ALARM, PEM count-half disabled 
I've disabled the alarm for PEM_count_half, using the mask in the 40m.alhConfig file.  We can't do anything about it, and it's just annoying.
Entry  Tue Jan 26 20:18:32 2010, rana, Configuration, ALARM, op540m: alarms and BLRMS and StripTool restored 
I turned the StripTool and ALARMS and BLRMS back on on op540m. Looks like it has been rebooted 5 days ago and no one turned these back on. Also, there
was a bunch of junk strewn around its keyboard which I restrained myself from throwing in the trash.
The BLRMS trends should be active now.
Entry  Thu Apr 23 23:07:23 2015, Dugolini, Frogs, ALARM, laptops warning popeye-and-spinach.jpg
Please!



Don't put laptops on the ISC Tables!
Entry  Sun Dec 20 23:23:54 2015, ericq, Update, ALARM, Rats. 
A small rat / large mouse just ran through the control room. Ugh.
Entry  Wed Dec 23 16:15:47 2015, rana, Omnistructure, ALARM, fire alarm 
Fire alarm went off several minutes ago. Talked to security and they said there was no fire. It beeped twice again just now. No one has been working
on the IFO today.
    Reply  Mon Jun 27 15:26:03 2016, Steve, Omnistructure, ALARM, fire alarm test fireAlarmTest.png
The fire alarm came on around 15:05  for about 2-3 minutes. We all  left the lab and counted heads.  I called Paul Mackel x2646 (cell
626/ 890- 3259) at Fire Protection Services. He said that this alarm test was planned and we should of got an email notice. Perhaps I missed that notes.




Entry  Wed Nov 14 14:35:38 2018, Steve, Update, ALARM, emergency calling list for 40m Lab  
It is posted at the 40m wiki with Gautam' help. Printed copies posted around doors
also.
Entry  Fri Sep 6 16:38:24 2019, aaron, Update, ALARM, Alarm noise from smart-ups machine under workstation? 
There was an alarm sound from the Smart-UPS 2200 sitting under the workstation. I see that the 'replace battery' light is red, and this
elog tells me that these batteries are replaced every ~1-4 years; the last replacement was march 2016. Holding down the 'test' button for 2-3
seconds results in the alarm sound and does not clear the replace battery indicator.
Entry  Mon Oct 5 14:37:29 2009, Alberto, Update, ABSL, Absolute Length Meaasurement NPRO is on 
In the revival of the experiement length measurement for the recycling cavities, I turned the auxiliary NPRO back on. The shutter is closed.
I also recollected all the equipment of the experiment after that during the summer it had been scattered around the lab to be used for other purposes
(Joe and Zach's cameras and Stephanie and Koji's work with the new EOM).
Entry  Mon Nov 9 11:14:57 2009, Alberto, Update, ABSL, Started working on the PSL table 
I'm working on the PSL table to set up the PLL setup for the AbsL experiment.
Entry  Mon Nov 9 13:17:07 2009, Alberto, Configuration, ABSL, NPRO on 
I turned the auxialiary NPRO for the AbsL Experiment on. Its shutter stays closed.
Entry  Tue Nov 10 19:19:57 2009, Alberto, Update, ABSL, Rotated polarizer on the PSL table, along the MC input pick off beam  
Aligning the beam for the PLL of the AbsL Experiement I rotated the polarizer along the path of the MC Input pick off beam (= the pick off coming from
the MC periscope).
Entry  Tue Nov 10 19:21:53 2009, Alberto, Update, ABSL, PLL Alignment 
I've been trying to lock the PLL for the AbsL Experiment but I can't see the beat (between the auxiliary NPRO and the PSL).
I believe the alignment of the PLL is not good. The Farady Isolator is definitely not perfectly aligned (you can see it from the beam spot after
it) but still it should be enough to see something at the PLL PD.
Entry  Wed Nov 11 15:04:52 2009, Alberto, Update, ABSL, Working on the AP table 
I've opened the AP table and I'm working on it.
    Reply  Wed Nov 11 16:18:57 2009, Alberto, Update, ABSL, Working on the AP table 


    
        
            Quote:
        
        
            
            I've opened the AP table and I'm working on it.
    Reply  Wed Nov 11 17:10:51 2009, Jenne, Update, ABSL, Working on the AP table 


    
        
            Quote:
        
        
            
            
            
    
    Reply  Wed Nov 11 17:33:54 2009, Koji, Update, ABSL, Working on the AP table 
Yes it did.
For long time, the crystal temperature C1:PSL-126MOPA_LTMP was 43~46deg. Now it is 34deg. Try
~10deg lower temperature.
Entry  Thu Nov 12 10:45:36 2009, Alberto, Update, ABSL, Working on the AP table 
I've opened the AP table and I'm working on it.
Also auxiliary NPRO turned on and mechanical shutter opened.
    Reply  Thu Nov 12 11:34:38 2009, Alberto, Update, ABSL, Working on the AP table 


    
        
            Quote:
        
        
            
            Yes it did.
    Reply  Thu Nov 12 12:51:45 2009, Alberto, Update, ABSL, Working on the AP table 


    
        
            Quote:
        
        
            
            I've opened the AP table and I'm working on it.
    Reply  Thu Nov 12 16:53:59 2009, Alberto, Update, ABSL, Working on the AP table 


    
        
            Quote:
        
        
            
            
            
    
Entry  Thu Nov 12 18:10:27 2009, Alberto, Update, ABSL, PLL Locked 
I locked the PLL and made some first measuremtns of the spectrum of the error signal. I'll post them later.
I closed the shutter of the NPRO.
Entry  Tue Nov 24 14:33:22 2009, Alberto, Update, ABSL, working on the AP table 
I'm working on the AP table. I also opened the auxiliary NPRO shutter. The auxiliary beam is on its path on the AP table and PSL table.
    Reply  Tue Nov 24 19:16:02 2009, Alberto, Update, ABSL, working on the AP table 


    
        
            Quote:
        
        
            
            I'm working on the AP table. I also opened the auxiliary NPRO shutter. The auxiliary beam is on its path on the AP
Entry  Wed Nov 25 08:43:08 2009, Alberto, Update, ABSL, Working on the AP table 
I'm working on the AP table. I also opened the auxiliary NPRO shutter. The auxiliary beam is on its path on the AP table and PSL table.
Entry  Wed Nov 25 10:20:47 2009, Alberto, Update, ABSL, AbsL PLL not able to lock 
Last night something happened on the beat between the PSL beam and the auxiliary NPRO beam, that spoiled the quality of the beating I had before.
As a result the PLL has become unable to lock the two lasers.
The amplitude of the beat at the spectrum analyzer has gone down to -40 dBm from -10 that it was earlier. The frequency has also become more unstable
    Reply  Wed Nov 25 11:02:54 2009, Alberto, Update, ABSL, AbsL PLL not able to lock 


    
        
            Quote:
        
        
            
            Last night something happened on the beat between the PSL beam and the auxiliary NPRO beam, that spoiled the quality
    Reply  Wed Nov 25 15:42:27 2009, Alberto, Update, ABSL, Working on the AP table 


    
        
            Quote:
        
        
            
            I'm working on the AP table. I also opened the auxiliary NPRO shutter. The auxiliary beam is on its path on the AP