Entry  Tue Oct 21 16:17:45 2008, Yoichi, Update, PSL, AD797 Oscillation in the FSS board 
I checked each op-amp's output in the FSS board to see if any indication of slew-rate saturation can be found.
PA85, which was the most suspicious one, actually has a very large slew rate limit (1000V/usec).
Its output swing was about 5V/usec. So PA85 was ok in terms of slew rate.
Entry  Fri Apr 13 16:09:43 2012, Suresh, Update, Computer Scripts / Programs, ACAD 2002 installed on C21530 
I have installed ACAD 2002 on one of the Windows machines in the Control Room.    It is on the machine which
has Solid Works (called C21530).  
Entry  Fri Nov 7 08:08:00 2008, steve, Update, PEM, AC turned off in IFO room 
All [COLOR=red]tree[/COLOR] AC units in IFO room 104 switched off at 7:45am
I'm looking for the periodic thump signal in the Vertex area
noted in entry 1113 of PEM-ACC_MC1
Entry  Fri Feb 10 15:54:27 2012, steve, Update, PEM, AC turned back ON 
The air cond was off for 2 hrs.  I just switched it back on at 15:51
    Reply  Wed Dec 19 11:53:23 2012, Steve, Update, VAC, AC removed workingatatm.png
 Koji, Manasa, Jenne, Jamie, Bob and Steve
Access connector removed this morning and work has began in the IOO chamber.  BE WARE OF ANTS !
    Reply  Mon Jul 29 16:04:40 2013, Steve, Update, VAC, AC removed acout.pngAlAcACa.jpgAlAcACb.jpg
 Manasa, Eric, Evan, Koji and Steve,
Access connector removed in order to complete alignment. Light aluminum with acetate windows AC installed.
 
    Reply  Mon Jul 29 16:15:13 2013, Koji, Update, VAC, AC removed 
Prior to the access connector removal, Manasa and I aligned the IFO mirrors. 
The arms were locked and aligned by ASS.
Entry  Wed Mar 21 10:31:28 2012, steve, Update, SAFETY, AC power to ETMY turned off 
ETMY sus damping was disabled. Green locking laser and associated electronics turned off. Computers and power supplies turned off at rack 1Y4
The electricians picking up ac power from 1Y4 manual disconnect box and installing conduit line to ISCT-ETMY east end optical table.
There will be no more daisy chaining this way. 
    Reply  Wed Mar 21 11:32:47 2012, steve, Update, SAFETY, AC power to ETMX turned off 


    
        
            Quote:
        
        
            
            ETMY sus damping was disabled. Green locking laser and associated electronics turned off. Computers and power supplies
    Reply  Mon Jun 25 08:21:40 2012, steve, Update, General, AC power disturbance on Sat powerdisturbed.png


    
        
            Quote:
        
        
            

           I was notified by CIT Utilities that there was a power surge or short power outage this after noon.
    Reply  Wed Mar 21 19:12:27 2012, steve, Update, Computers, AC power back on both ends 


    
        
            Quote:
        
        
            

           
            
                
 
    Reply  Mon Apr 26 11:13:28 2010, steve, Omnistructure, PEM, AC on/off is working 


    
        
            Quote:
        
        
            
            Koji and I wanted to turn off the IFO-room AC so the wind would not blow on MC1-3. We could not. The switches were
Entry  Fri Nov 7 10:52:57 2008, steve, Update, PEM, AC is back on acf.jpgacc_mc1.jpgtest.jpg
The 20 minutes and 6 sec thumps are not related to the 40m ac units
This period is precisely right on so it is not likely mechanical as a source.
The time and frequency domain of this signal at attachment 2&3
    Reply  Fri Nov 7 15:58:10 2008, rana, Update, PEM, AC is back on mc-acc.pdfmc-acc-quad.pdf
I'll bet Steve a dollar that it is mechanical. The attached PDF compares all of the accelerometers from right now.

You can see that the RMS in MC2 is way bigger than MC1.
Entry  Wed Nov 23 14:47:14 2011, Jenne, Configuration, Environment, AC in the IFO room was off 
I turned it back on, maybe around 11am?  Definitely a little while before the 12:30 meeting.
EDIT by KI:
 Sorry, it's me. I was checking if AC was doing something bad on the ALS noise.
    Reply  Tue Apr 20 10:52:49 2021, Anchal, Paco, Update, SUS, AC gain coil output balancing for IMC POS2PYuncoupled.pdfPIT2PYuncoupled.pdfMC1YAWexc.pdfMC1PITexc.pdf
[Paco, Anchal]


We adopted the following procedure to balance the coil output gains using a high-frequency (> 10 Hz) excitation on "C1:SUS-MCX_ASCPIT_EXC",
"C1:SUS-MCX_ASCYAW_EXC", and "C1:SUS-MCX_LSC_EXC", where X is one of {1, 2, 3} for the three IMC optics, and the cavity sensors (MC_F,
Entry  Wed Sep 17 19:13:47 2008, Alberto, Update, General, ABSL: status 
I installed the setup for measuring TEM01/10 on the X end table. 
I'm leaving. I shut the laser, flipped down the flipper mirror, disconnected the pzt drive signal from the laser.
[B]For Jenne. The power cable for the Guralps' board is now connected to the PDH box on my instruments cart but you can take it.[/B]
Entry  Mon Sep 22 20:00:35 2008, Alberto, Update, General, ABSL: running measurement 
I'm leaving the X arm locked on the TEm01 mode while a measurement is running. Just please wait for 40 minute if you need the interferometer tonight.
Entry  Wed Sep 17 13:58:36 2008, Alberto, Update, General, ABSL: results from the X arm Sept17_XarmFSRmeasurement_report.ps
Today I repeated the measurement of the FSR on the X arm cavity. The noise in the transmitted power that made the measures fluctuate was much reduced after
last night Rob worked on the interferometer. The X arm cavity length is now: (38.4580+/-0.0003)m. I'm attaching a summary of the data I've taken.
Entry  Sun Sep 14 18:30:32 2008, Alberto, Update, General, ABSL: measured X arm 
Today I measured the X arm FSR.
Hi moved the fast PD (Thor Labs PDA255) from the Y end table to the X end table. I had to use a beam splitter to pick out the transmitted beam from the
cavity beam and send it to the PD. I did not want to interpose the BS before the TRANS X PD, so I had to move the ETMXT camera to an other place in the
    Reply  Sun Sep 14 19:29:07 2008, Alberto, Update, General, ABSL: measured X arm 
[quote]Today I measured the X arm FSR.
Hi moved the fast PD (Thor Labs PDA255) from the Y end table to the X end table. I had to use a beam splitter to pick out the transmitted beam from the
cavity beam and send it to the PD. I did not want to interpose the BS before the TRANS X PD, so I had to move the ETMXT camera to an other place in the
Entry  Wed Sep 24 17:57:04 2008, Alberto, Update, General, ABSL: FSS Slow Actuator Control 
Rana, Alberto

Today when I started working with the PLL that I use to control the secondary laser on the ABSL experiment, I found that the beat between the two lasers
Entry  Mon Mar 18 12:02:12 2013, Annalisa, Configuration, ABSL, ABSL setup for g-factor measurement of PRC 
Motivations
The ABSL technique has been already used in the past to measure the absolute length of the interferometer's optical cavities
by means of  an auxiliary laser source, as described in LIGO-P1200048-v3 and in Alberto Stochino
Entry  Tue Jul 5 17:38:46 2011, kiwamu, Update, ABSL, ABSL laser frequency-locked RFsetup.png
In this past weekend the ABSL laser was successfully frequency-locked to the
PSL laser with a frequency offset of about 100 MHz.
In the current setup a mixer-based frequency discriminator is used for
Entry  Thu Aug 4 09:42:19 2011, Koji, Update, ABSL, ABSL Laser shutter closed 
The shutter of the ABSL laser is closed for the vent work.
Entry  Sat Jul 30 00:33:33 2011, Koji, Update, PSL, ABSL Laser crystal temp left largely excited & left unattended for more than 3hours 
I found that the ref cav trans CCD view was blinking with 30-50 fringe amplitudes. This meant the
laser freq was swinging ~50GHz.
I checked the ABSL laser and the SG out of a lock-in amplifier was connected to the slow input.
    Reply  Sat Jul 30 02:47:43 2011, rana, Update, PSL, ABSL Laser crystal temp left largely excited & left unattended for more than 3hours 
 Hmm. Should have only been +/- 1 GHz. Some setting got changed apparently...
This is a part of the RefCav temperature measurement setup. You'll get an elog from Jenny very soon.
Entry  Wed Jul 20 06:52:01 2011, Suresh, Update, LSC, ABSL - PSL beat lock 
The ABSL locking setup to the PSL is down. 
According the plan, I started to use the IR beam dumped after the doubling crystal for the
IR beat lock (Sonali's project).  The beat lock was disturbed when I shifted some clamps to make way for a few mirrors.  So I set about fixing
the beat lock.  I reobtained the lock but noticed that the net beam power reaching the Newfocus 1611 detector was around 15mW.  10mW from
    Reply  Wed May 15 17:38:49 2013, Jamie, Summary, CDS, AA/AI requirements 


    
        
            Quote:
        
        
            

           What this means:
Entry  Fri Dec 4 19:23:40 2020, Koji, Update, Electronics, AA/AI board testing ongoing 
I have the setup built for the AA/AI board testing around the PD testing area. Please let me leave it like that for a week or so.

12/4 TF Tested 5 PCBs
12/6 TF Tested 19 PCBs (12min/PCB) - found 1 failure (S2001479 CH1) -> Fixed 12/11
Entry  Wed Apr 16 21:11:12 2008, rana, Summary, DAQ, AA/AI Filters for the DAQ & FE systems DAQ_filters_080416.pdf
I used Foton to make up some new filters which will be used all over the project in order to downsample/upsample.

There will be 2 flavors:
Entry  Fri Nov 12 20:40:51 2010, josephb, yuta, valera, Configuration, Electronics, AA voltage range 
We changed the range of the two SUS AA boards in the corner from +/-2 V to +/-10 V by changing the supply voltage from +/-5 V to +/-15 V. The change
was made by switching the AA power feed wires on  the cross connect. The max supply according to the spec of DRV134/INA134 is +/-18 V.
We checked the new range by applying the voltage to the input of AA and measuring the output going to the ADCs. The local damping MC1,2,3 appears
Entry  Sun Jul 24 20:36:03 2011, haixing, Summary, Electronics, AA filter tolerance analysis 6x
Koji and Haixing,
We did a tolerance analysis to specify the conner frequency for passive low-pass filtering in the AA filter of Cymac. The 
link to the wiki page for the AA filter goes as follows (one can have a look at the simple schematics): 
    Reply  Sun Jul 24 20:47:21 2011, rana, Summary, Electronics, AA filter tolerance analysis 
This is sort of OK, except the capacitor connects across the (+) terminals of the two input opamps, and does not connect to ground.
Also, we don't care about the CMRR at 64 kHz. We care about it at up to 10 kHz, but not above. The sample frequency of the ADC is 64 kHz, but all
of the models run at 16 kHz or less, so the Nyquist frequency is 8 kHz.
    Reply  Sun Jul 24 22:19:19 2011, haixing, Summary, Electronics, AA filter tolerance analysis low_pass_new.png
 
>> This sort of OK, except the capacitor connects across the (+) terminals of the two input opamps, and does not connect to ground:
    Reply  Tue Jul 26 21:11:40 2011, haixing, Summary, Electronics, AA filter tolerance analysis 
 
Entry  Thu May 19 12:09:41 2011, josephb, Update, CDS, AA filter box pulled from 1X5, optic suspensions currently off 
[Steve, Joe]
Steve pulled the top AA filter box from 1X5 which handled some of the suspensions channels.  We turned off all the watchdogs before pulling
it out, as well as recorded which cables were connected to which inputs.
    Reply  Thu May 19 17:53:03 2011, steve, Update, CDS, AA filter box modified at 1X5 P1070727.JPG


    
        
            Quote:
        
        
            

           [Steve, Joe]
    Reply  Tue Dec 22 22:11:37 2020, gautam, Update, CDS, AA filt repaired and reinstalled 
Koji fixed the problematic channel - the issue was a bad solder joint on the input resistors to the THS4131. The board was re-installed. I also made
a custom 2x4-pin LEMO-->DB9 cable, so we are now recording the PMC and FSS ERR/CTRL channel diagnostics again (spectra tomorrow). Note
that Ch32 is recording some sort of DuoTone signal and so is not usable. This is due to a misconfiguration - ADC0 CH31 is the one
Entry  Wed Aug 10 12:28:22 2011, Ishwita , Manuel, Update, PEM, AA board gain pw_gur1_sts1_aug10.png
We used a function generator, an oscilloscope and the Data Viewer to check the gain of the new AA board (used for the seismometers). Putting a sine wave
of 0.3V (using a function generator) to the AA board, we could see about 500 counts in the Data Viewer. The calibration of the ADC
is 214 counts/volt, so the AA board gives to the ADC an output of 0.03V. This proves that
Entry  Mon Jul 11 10:10:31 2011, Ishwita, Configuration, , AA board  P7080305.JPGseismometers.png
The AA board shown in attachment 1 will be used in the seismometer hardware setup. A cartoon of this setup is shown in attachment 2.
BNC connectors are required for the seismometer breakout boxes. So the four-pin LEMO connectors present in the AA board were removed and panel
mount BNC connectors were soldered to it. Red and blue colored wires were used to connect the BNC connectors to the board. Red wire connects the center
Entry  Sat Jul 16 18:54:05 2011, Ishwita, Manuel, Configuration, Electronics, AA board DSC_3236.JPGpic1.pngDSC_3237.JPG
We fixed the anti-aliasing board in its aluminum black box,  the box couldn't be covered entirely because of the outgoing wires of the BNC connectors,
so we drilled additional holes on the top cover to slide it backwards by 1cm and then screw it.
We had to fix the AA board box in rack 1X7, but there wasn't enough space, so we tried to move the blue chassis (ligo electro-optical fanout chassis
    Reply  Sun Oct 14 20:51:20 2012, Den, Update, PEM, AA board acc.pngDSC_4747.JPG


    
        
            Quote:
        
        
            

            
    Reply  Mon Oct 15 10:14:22 2012, Den, Update, PEM, AA board 


    
        
            Quote:
        
        
            

            
Entry  Thu May 9 00:05:12 2013, rana, Koji, Jenne, Summary, LSC, AA and AI change AAAI.pdf
We would like to increase the UGF of the PRC loop so as to allow more suppression
of the PRC signal and less pollution of the MICH signal (remember that the PRC/MICH optical gain ratio is huge).
We were already losing phase because of delay in the LSC - SUS digital link. In addition to that, a major source of delay is the analog anti-aliasing
    Reply  Thu Oct 28 04:23:38 2010, rana, Update, IOO, A2L prep 
To get the angle to length signal before the c1ioo processor gets going, we need a length signal. We can use either the error signal or the control signal.
I recommend using the control signal since its not puny. The 4-pin LEMO inputs to the OSEM ADC that Suresh has wired are differential
so we can, in principle, use either the BNC output of the SERVO plug or the 2-pin LEMO output.
Entry  Thu Jun 21 14:23:22 2018, Udit Khandelwal, Summary, General, A summary of the Tip-TIlt Mirror Holder design changes 1.png2.png3-1.png3-2.png3-3.png
Here’s a quick summary of the Tip-Tilt Design updates (all files are in the dropbox in [TipTiltSus>TT_New]) that I have been working on with
Koji and Steve's help.

1. Plate on top to hold mirror in place:
    Reply  Thu Jun 21 14:57:59 2018, Koji, Summary, General, A summary of the Tip-TIlt Mirror Holder design changes 
> 2. Weighted screw rod at the bottom for tilting the mirror-holder:

Too long. The design of the holder should be check with the entire assembly.
We should be able to make it compact if we heavier weights.
Entry  Thu Mar 19 02:45:36 2009, Yoichi, Configuration, IOO, A loose wire found for MC1  
I found a loose connection of a wire in the cross-connect between an ADC and the MC1 coil driver's UL bias input.
I tightened it.
To see if this fixes the MC1 drift problem, I will do another round of MC1 drift measurement.
    Reply  Thu Mar 19 10:45:43 2009, Yoichi, Configuration, IOO, A loose wire found for MC1  AfterWireFix-1.pdf
I attached a 6-hour trend of the MC mirror OSEM signals with the MC unlocked.
The drift of the MC1 is within 20 counts (0.6um in terms of each OSEM).
This is comparable to the other MC mirrors.