    Reply  Tue Aug 8 17:04:26 2017, Steve, Update, VAC, unintended pump down pumping_down.png
IFO pressure 2 Torr,    PSL  shutter closed.  I'm pumping down with 2 roughing pumps with ion pump gate valves open and annulosses
at atm.

The vacuum envelope was vented to 17 Torr while I was replacing the USP battery stack. More about this later.....
    Reply  Tue Aug 8 17:44:11 2017, Steve, Update, VAC, unintended pump down stopped_pumping.png
Pumpdown stopped for over night  at ~  1 Torr

The roughing line disconnected.  Valves condition indicator  "moving " means that it is closed and it's cable disconnected
so it can not move.
    Reply  Wed Aug 9 11:59:51 2017, Steve, Update, VAC, unintended pump down at vacuum normal pdc.pngpd80b@vacnormal.png
pd80b has reached Vac Normal. IFO pressure 0.5 mTorr

We need our vauum channels back in dataviewer.




Entry  Wed Aug 9 16:34:46 2017, rana, Update, VAC, Vacuum Document recovered 
Steve and I found the previous draft of the 40m Vacuum Document. Someone in 2015 had browsed into the Docs history and then saved the old 2013 version
as the current one.

We restored the version from 2014 which has all of Steve's edits. I have put that version (which is now the working copy) into the DCC:  
    Reply  Thu Aug 10 15:45:34 2017, Steve, Update, VAC, accidental vent to 17 Torr Mag_UPS.jpgUPSbatteryPack.jpgclosing_V1.jpg
Finally we got the cold cathode gauge working. IFO pressure 7e-6 Torr-it, vacuum normal valve configuration with all 4 ion pump gate valves closed at
~ 9e-6 Torr.  The cryo pump was also pumped yesterday to remove the accumulated outgassing build
up.
    Reply  Mon Aug 21 14:53:32 2017, gautam, Update, VAC, RGA reset 
[gautam, steve]

In the aftermath of the accidental vent, it looks like the RGA was shutdown.

We followed the instructions in this elog to restart the RGA.
    Reply  Mon Aug 21 16:35:48 2017, gautam, Update, VAC, UPS checkup 
[steve, gautam]

At Rolf/Rich Abbott's request, we performed a check of the UPS today.

Steve believed that the UPS was functioning as it should, and the recent accidental vent was because the UPS batteries were insufficiently charged
    Reply  Wed Aug 23 10:11:46 2017, Steve, Update, VAC, vacuum valve specifications 
The V1 gate valve specs installed at 40m wiki page. VAT model number 10846-UE44-0007
Our main volume pumping goes through this 8" id gate valve V1 to Maglev turbo 
or Cryo pump to  VC1
    Reply  Mon Aug 28 10:28:07 2017, Steve, Update, VAC, RGA scan 20 day after 17 Torr 20d_after_17_torr.png
The RGA was turned on 7 days ago.  It's 46 C now.   The X-arm room tem ~20 C

IFO pressure 6.5e-6 Torr at IT-Hornet gauge. Valve configuration vacuum normal.
    Reply  Tue Aug 29 15:35:19 2017, Steve, Update, VAC, vacuum pump specifications & manuals 1759489_995814a9-55c1-4fa1-b333-f446d272c1cc.jpg
RP1 and RP3 roughing pump manual of Leybold D30A oily rotory
pump

Fore pump of
    Reply  Tue Sep 19 11:05:26 2017, Steve, Update, VAC, RGA scan at day 334  RGAscan334d.png
 

 
    Reply  Fri Oct 27 10:14:11 2017, Steve, Update, VAC, RGA scan at day 372 RGA_scan_d372.png
Valve configuration: Vacuum normal

Note: Tp2 running at 75Krpm 0.25A 26C has a  load high pitch sound today. It's fore line pressure 78 mTorr. Room temp 20C

 
    Reply  Mon Oct 30 10:09:41 2017, Steve, Update, VAC, TP2 failed tp2failed.pngifo_1.0E-5_Torrit.pngtp2failed2dago.png4days.png
 IFO pressure 1.2e-5 Torr at 9:30am




Quote:



Valve configuration:
    Reply  Mon Nov 6 09:09:43 2017, Steve, Update, VAC, setting up new TP2 turbo 
Our new Agilent Technology TwisTorr
84FS AG rack controller ( English Manual pages 195-297 )  RS232/485, product number X3508-64001, sn IT1737C383

This controller, turbo and
    Reply  Fri Nov 10 08:52:21 2017, Steve, Update, VAC, TP3 drypump replaced 
PSL shutter closed at 6e-6 Torr-it    

   The foreline pressure of the drypump is 850 mTorr at 8,446 hrs of seal life

V1 will be closed for ~20 minutes for drypump replacement..........
    Reply  Sun Nov 26 19:38:40 2017, Steve, Update, VAC, TP3 drypump replaced again after_replacement.png
The TP3 foreline pressure was 4.8 Torr, 50K rpm 0.54A and 31C........Maglev rotation normal 560 Hz.......    IFO pressure 7.2e- 6 Torrit was
not effected

V1 closed ......replaced drypump.........V1 opened
    Reply  Thu Nov 30 10:38:39 2017, Steve, Update, VAC, annuloses are not pumped AnnulosesNotPumped.png
Annuloses are not pumped for 30 days, since TP2 failed.  IFO pressure 7e-6 Torr it,  Rga 2.6e-6 Torr

Valve configuration: Vacuum Normal as TP3 is the forepump of Maglev, annuloses are not puped at 1.1 Torr

TP3 50K rpm,  0.15A  24C, foreline pressure 16.1 mTorr
    Reply  Tue Jan 2 15:43:35 2018, Steve, Update, VAC, pumpdown after power outage pumpdown_from_44_mTorr.png
 




Quote:



There was a power outage.
    Reply  Tue Jan 30 10:22:20 2018, Steve, Update, VAC, RGA scan at day 175 RGApd80-b-468-175d.pngRgaOn20hrs.png
pd80b rga scan at 175 day.  IFO pressure 7.3e-6 Torr-it

Condition: vacuum normal, annuloses not pumped. Rga turned on yesterday.

The rga was not on since last poweroutage Jan 2, 2018 It is warming up and outgassing
    Reply  Wed Feb 7 15:50:42 2018, Steve, Update, VAC, IFO pressure monitoring CC_analog.png
Hornet cold cathode gauge analoge output  [ DSub9 pin 3 and 7 ] are wired to go ETMX Acromag. It was reading 4.9V at 7.8e-6 Torr [ 3,110 V 
8.35e-5A ] at the end of a 24ft BNC cable. Now it has to be hook up to an Acromag channel.

This will replace the not functioning C1: Vac-CC1_pressure
    Reply  Fri Feb 16 15:57:58 2018, Steve, Update, VAC, TP3 drypump replaced 
The forline pressure of TP3 was 399 mTorr

It was replaced this morning at  TP3 controller 134,638hrs with the "failed TP2 station" drypump. The foreline pressure now at
TP3 is 100 mTorr at 6 hrs of operation.[ at day 3  63 mT ]
Entry  Wed Feb 21 17:20:32 2018, johannes, Update, VAC, Hornet gauge connected to DAQ. VacLog.png
I wired the six available BNC connectors on the front panel of the new XEND slow DAQ to physical Acromag channels. There were two unused ADC channels
and eight DAC channels, of which I connected four. The following entries were added to /cvs/cds/caltech/target/c1auxex2/ETMXAUX2.db /caltech/target/c1auxex2/ETMXaux2.db




Connector
Acromag Channel
EPICS
    Reply  Fri Feb 23 07:47:54 2018, Steve, Update, VAC, CC1 Hornet InstruTech_Hornet_CC1.pngCC1_readback_updated.png
We have the IFO pressure logged again! Thanks Johannes and Gautam

This InstruTech cold cathode ionization vacuum gauge " Hornet " was installed
2016 Sep 14
    Reply  Mon Mar 5 10:13:21 2018, gautam, Update, VAC, vacuum health VacuumStatus.png
In Steve's absence, I've tried to keep an eye on the health of the vacuum system. From Attachment #1, the pressure of the main volume seems stable,
no red flags there. I also don't here any anomalously loud sounds near the vacuum pumps. I've changed the N2 cylinders that keep V1 open twice,
on Wednesday and Sunday of last week. So
    Reply  Fri Mar 16 09:36:56 2018, Steve, Update, VAC, RGA scan at day 511,218d RgaScan511d.png08.png
Pumpdown 80 at 511 days and pd80b at 218 days

Valve configuration:  special vacuum normal, annuloses are not pumped at 3 Torr, IFO pressure 7.4e-6 Torr at vac envelope temp 22 +- 1C
degrees
    Reply  Mon Mar 26 09:25:18 2018, Steve, Update, VAC, VM1 opened CC4VM1.png
CC1 old MKS cold cathode gauge randomly turns on- off. This makes software interlock close VM1 to protect RGA  So the closed off RGA region
pressure goes up and the result is distorted RGA scan.

CC1 MKS gauge is disconnected and VM1 opened. This reminds me that we should connect our interlocks to CC1 Hornet
    Reply  Tue Mar 27 08:58:21 2018, Steve, Update, VAC, VM1 opened.......scan fine rga2018march27.png
 




Quote:



CC1 old MKS cold cathode gauge randomly turns on- off. This makes software interlock close VM1 to protect
    Reply  Wed Jun 6 10:02:52 2018, Steve, Update, VAC, RGA scan RGA302d.pngannuloses_NOT_pumped.pngtemp_vac.png
 
    Reply  Wed Jun 6 10:44:52 2018, gautam, Update, VAC, Annulus pressure channels added to frames AnsPressureLogged.png
[steve, gautam]

We added the following channels to C0EDCU.ini and restarted the daqd processes. Channels seem to have been added successfully, we will check
trend writing later today. Motivation is to have a long term record of annulus pressure (even though we are not currently pumping on the annulus).
    Reply  Thu Jul 12 16:18:22 2018, Steve, Update, VAC, Vent preparations before_Vent.pngbefore_Vent_cond.png
We are getting ready to vent.

 
Entry  Fri Jul 13 10:54:55 2018, aaron, Update, VAC, Vent 80 vent80wtiptilts.png
[aaron, steve]

Steve gave me a venting tutorial. I'll record this in probably a bit more detail than is strictly necessary, so I can keep track of some
of the minor details for future reference.
    Reply  Fri Jul 13 16:26:52 2018, Steve, Update, VAC, Vent 80 is completing... @atm.pngvent80_7h.pngventregN2&Air_c.jpg
Steve and Aaron,

6 hrs vent is reaching equlibrium to room air. It took 3 and a half instrument grade air cilynders [ AI UZ300 as labelled ] at 10 psi pressure.
Average vent speed ~ 2 Torr/min
Entry  Mon Jul 16 15:07:19 2018, Koji, Summary, VAC, Oven C vent P_20180716_141512.jpgP_20180716_141601.jpgP_20180716_141610.jpgP_20180716_141827.jpgP_20180716_143901.jpg
[Steve Koji]

- Attachment1: Removed the thermal cap. Checked the temperature of the oven. It was totally cold.

- Attachment2: Confirmed the RGA section was isolated. The pumps for the RGA was left running.
Entry  Mon Jul 16 18:12:00 2018, Koji, Update, VAC, Adding a manual gate valve between TP1 and V1/VM2 P_20180716_155413.jpgP_20180716_155645.jpgP_20180716_155738.jpgP_20180716_162307.jpgP_20180716_172000.jpg
[Steve Koji]

We are in the process of adding a manual gate valve between TP1 (Osaka Maglev) and the other gate valves (I suppose V1 and VM2).
The work is still on going
    Reply  Tue Jul 17 18:16:38 2018, Steve, Summary, VAC, pumpdown 81 at 6 hrs 6x
Precondition:  4 days at atm.   Atm5

HEPA tent used during the vent at ETMY  It reduced partical count 10 fold of 0.5 and 0.3 micron particals. Atm6

New items in vacuum:  Clean manual gate valve [Cetec made] from John Worden with 6" id....as it came from Hanford... [ Throttle able
    Reply  Tue Jul 17 22:25:41 2018, Koji, Summary, VAC, pumpdown 81 at 6 hrs 
10:20PM


Opened VM2 to pump down the RGA section with TP1
Stopped rotary roughing pumps

Manually closed RV1
Closed
V3
Stopped RP1 and RP3
Vented the RP hose




The P1 pressure is 380mTorr. I allowed Gautam to use
    Reply  Wed Jul 18 12:49:08 2018, Steve, Summary, VAC, pumpdown 81 at 6 +4.5hrs manlGateScan.pnghanding_over_Mag.pngTGVw2auxfans_.jpg
The manual gate valve scan was clean. Atm1     TP1 was pumping on it overnight.

                                         
      Pumpdown continued to hand over the pumping to TP1 maglev turbo
    Reply  Wed Jul 18 17:36:50 2018, Steve, Summary, VAC, pumpdown 81 at 6 +9 hrs completed pd81completed.pngpd81@30hrs.png
IFO P1 6e-4 Torr,  manual gate valve is fully open

The annuloses will be pumped down tomorrow.

Valve configuration: vacuum normal, RGA and annuloses are not pumped
    Reply  Thu Jul 19 01:56:25 2018, gautam, Summary, VAC, AUX pump shutdown 
[koji, gautam]

Per Steve's instructions, we did the following:


TP3fl pressure reading was 65 torr.
TP3 controller reported pumping current of ~0.18A, temperature of 24C.
We throttled
    Reply  Thu Jul 19 11:01:03 2018, Steve, Summary, VAC, pd81 @ 2e-5 Torr pd81@2days.png
Cold cathode gauge just turned on.
    Reply  Thu Jul 19 13:35:30 2018, Steve, Summary, VAC, annuloses pumped  AnsPumped.png
Roughing down the annuloses required closing V1 for 13 minutes

IFO is 2.2e-5 Torr
    Reply  Mon Jul 23 09:58:52 2018, Steve, Summary, VAC, RGA scan at day 6 pd81-560Hz-d6.png
 

 
    Reply  Mon Aug 6 09:34:02 2018, Steve, Update, VAC, RGA scan at day 20 pd81d20.png
 

 
    Reply  Thu Aug 16 07:50:28 2018, Steve, Update, VAC, pumpdown 81 at day 30 pd81d30.png
 

 
    Reply  Thu Aug 16 14:48:14 2018, Steve, Update, VAC, why do we need a root pump? 6x


Basic Pump Throughput Concepts

What is Pump Throughput?

The manufacturer of a vacuum pump supplies a chart for each pump
showing pumping speed (volume in unit time) vs pressure. The example, for a fictitious pump, shows the pumping speed is substantially
    Reply  Fri Aug 24 10:51:23 2018, Steve, Update, VAC, pumpdown 81 at day 38 d38.png
 

 
    Reply  Wed Aug 29 09:56:00 2018, Steve, Update, VAC, Maglev controller needs service 
TP-1 Osaka maglev controller  [  model TCO10M,  ser V3F04J07 ]  needs maintenance. Alarm led  on indicating  that we need
Lv2 service.

The turbo and the controller are in good working order.
Entry  Fri Sep 21 16:51:43 2018, gautam, Update, VAC, c1vac1 is unresponsive 
Steve pointed out that some of the vacuum MEDM screen fields were reporting "NO COMM". Koji confirmed that this is a c1vac1 problem, likely
the same as reported here and can be fixed using the same procedure.

However, Steve is worried that the interlock won't kick in in case of a vacuum emergency, 
    Reply  Mon Sep 24 15:06:10 2018, gautam, Update, VAC, c1vac1 reboot + TP1 controller replacement beforeReboot.pngafterReboot.pngCC1.png
[steve, gautam]

Following the procedure in this elog, we effected a reset of the vacuum slow machines.
Usually, I just turn the key on these crates to do a power cycle, but Steve pointed out that for the vacuum machines, we
    Reply  Wed Sep 26 10:07:16 2018, Steve, Update, VAC, why reboot c1vac1 NOcomm.pngReboot_&_sawp.pngc1vac1&2_.jpg
Precondition: c1vac1 & c1vac2 all LED warning lights green [ atm3 ], the only error message is in the gauge readings NO COMM, dataviewer will plot
zero [ atm1 ], valves are operational

When our vacuum gauges read " NO COMM " than our INTERLOCKS  do  NOT