ID Date Author Type Categoryup Subject Text Attachments
  15408   Thu Jun 18 14:13:03 2020 JonUpdateVACQuestions/comments on vacuumI agree there
were MEDM fields, but I can't find any
  
  15409   Thu Jun 18 15:25:08 2020 JordanUpdateVACTP2 and TP3 Forepump removalI removed the backing pumps for TP2 and
TP3 today to test ultimate pressure and determine
if they need a tip seal replacement. This
  
  15410   Thu Jun 18 15:46:34 2020 gautamUpdateVACQuestions/comments on vacuumSo why not just have a special mode for
the interlock code during pumpdown and venting,
and during normal operation we expect the
  
  15411   Thu Jun 18 16:56:34 2020 JordanUpdateVACTP2 and TP3 Forepump removal


Quote:


  
  15412   Thu Jun 18 22:33:57 2020 JonOmnistructureVACVac hardware purchase listReplacement Hardware Purchase List

I've
created a purchase
list of hardware needed to restore the aging
  
  15413   Fri Jun 19 07:40:49 2020 JonUpdateVACQuestions/comments on vacuumI think we should discuss interlock possibilities at
a 40m meeting. I'm reluctant to make
the system more complicated, but perhaps
  
  15415   Fri Jun 19 09:57:35 2020 gautamUpdateVACQuestions/comments on vacuumFor this particular email service, ideally
the email should be sent out as soon as the
interlock is tripped, so this would require
  
  15417   Fri Jun 19 14:03:50 2020 JordanUpdateVACForepump Tip Seal ReplacementTip Seals were replaced on the forepumps
for TP2 and TP3, and both are ready to be
installed back onto the forelines.
  
  15421   Mon Jun 22 10:43:25 2020 JonConfigurationVACVac maintenance at 11 amThe vac system is going down at 11 am today
for planned maintenance:


Re-install the repaired
  
  15424   Mon Jun 22 20:06:06 2020 JonConfigurationVACVac maintenance completeThis work is finally complete. The dry
pump replacement was finished quickly but
the controls updates required some substantial
 Pumpdown-6-22-20.png 
  15425   Tue Jun 23 17:54:56 2020 ranaConfigurationVACVac maintenance completeI propose we go for all CAPS for all channel
names. The lower case names is just a holdover
from Steve/Alan from the 90's. All other
  
  15446   Wed Jul 1 18:03:04 2020 JonConfigurationVACUPS replacements​I looked into how the new UPS devices
suggested by Chub would communicate with
the vac interlocks. There are several possible
  
  15465   Thu Jul 9 18:00:35 2020 JonConfigurationVACUPS replacementsChub has placed the order for two new UPS
units (115V for TP2/3 and a 220V version
for TP1).
  
  15499   Thu Jul 23 15:58:24 2020 JonSummaryVACVacuum controls refurbishment planThis year we've struggled with vacuum
controls unreliability (e.g., spurious interlock
triggers) caused by decaying hardware. Here
  
  15500   Fri Jul 24 15:40:59 2020 JordanUpdateVACInstallation of two new UPS unitsI installed the Tripp
Lite SMX1000RT2U and Tripp
Lite Smart1000LCD at the
 UPS.jpg 
  15501   Mon Jul 27 15:48:36 2020 JonSummaryVACVacuum parts orderedTo carry out the next steps of the vac
refurbishment plan [ELOG
15499], I've ordered parts necessary
  
  15502   Tue Jul 28 12:22:40 2020 JonUpdateVACVac interlock test today 1:30 pmThis afternoon Jordan is going to carry
out a test of the V4 and V5 hardware
interlocks. To inform the interlock improvement
  
  15504   Tue Jul 28 14:11:14 2020 JonUpdateVACVac interlock test today 1:30 pmThis test has been completed. The IFO configuration
has been reverted to nominal.

For future reference: yes,
  
  15526   Fri Aug 14 10:10:56 2020 JonConfigurationVACVacuum repairs todayThe vac system is going down now for planned
repairs [ELOG
15499]. It will likely take most
  
  15527   Sat Aug 15 02:02:13 2020 JonConfigurationVACVacuum repairs todayVacuum work is completed. The TP2 and TP3
interlocks have been overhauled as proposed
in ELOG
  
  15528   Sat Aug 15 15:12:22 2020 JonConfigurationVACOverhaul of small turbo pump interlocksSummary

Yesterday I completed
the switchover of small turbo pump interlocks
as proposed in ELOG
 small_tp_signal_routing.pngsmall_tp_interlock_signal.pngsmall_tp_signal_routing.pngmedm_screen.png 
  15537   Mon Aug 24 08:13:56 2020 JonUpdateVACUPS installationI'm in the lab this morning to interface
the two new UPS units with the digital controls
system. Will be out by lunchtime. The disruptions
  
  15538   Mon Aug 24 11:25:07 2020 JonUpdateVACUPS installationI'm leaving the lab shortly. We're
not ready to switch over the vac equipment
to the new UPS units yet.
  
  15541   Wed Aug 26 15:48:31 2020 gautamUpdateVACControl screen left open on vacuum workstationI found that the control MEDM screen was
left open on the c1vac workstation. This
should be closed every time you leave the
  
  15556   Fri Sep 4 15:26:55 2020 JonUpdateVACVac system UPS installationThe vac controls are going down now to
pull and test software changes. Will advise
when the work is completed.
  
  15557   Fri Sep 4 21:12:51 2020 JonUpdateVACVac system UPS installationThe vac work is completed. All of the vacuum
equipment is now running on the new 120V
UPS, except for TP1. The 230V TP1 is still
  
  15558   Sat Sep 5 12:01:10 2020 JonUpdateVACVac system UPS installationSummary

Yesterday's UPS
switchover was mostly a success. The new
Tripp
 vac_medm.pngvac_medm.png 
  15577   Wed Sep 16 12:03:07 2020 JonUpdateVACReplacing pressure gaugesAssembled is the list of dead pressure
gauges. Their locations are also circled
in Attachment 1.
 vac_gauges.png 
  15582   Sat Sep 19 18:07:35 2020 KojiUpdateVACTP3 RP failureI came to the campus and Gautam notified
that he just had received the alert from
the vac watchdog.
 vac.pngScreen_Shot_2020-09-19_at_17.52.40.png 
  15586   Sat Sep 19 19:37:16 2020 not KojiUpdateVACTP3 RP failureDisconcerting because those tip seals were
just replaced [15417].
Maybe they were just defective, but if there
  
  15591   Mon Sep 21 15:57:08 2020 JordanUpdateVACTP3 Forepump Replacement and Vac resetI removed the forepump (Varian SH-110)
for TP3 today to see why it had failed over
the weekend. I tested it in the C&B lab
 Screenshot_from_2020-09-21_15-56-04.png20200921_145043.jpg20200921_145040.jpg 
  15599   Wed Sep 23 08:57:18 2020 gautamUpdateVACTP2 running HOTThe interlocks tripped at ~630am local
time. Jordan reported that TP2 was supposedly
running at 52 C (!).
 Screen_Shot_2020-09-23_at_8.55.39_AM.png 
  15600   Wed Sep 23 10:06:52 2020 KojiUpdateVACTP2 running HOTHere is the timeline. This suggests TP2
backing RP failure.

1st line: TP2 foreline pressure
 Screen_Shot_2020-09-23_at_10.00.43.png 
  15602   Wed Sep 23 15:06:54 2020 JordanUpdateVACTP2 Forepump Re-installI removed the forepump to TP2 this morning
after the vacuum failure, and tested in the
C&B lab. I pumped down on a small volume
 Screenshot_from_2020-09-23_15-15-43.pngMainVolPumpDown.png 
  15615   Tue Oct 6 14:35:16 2020 JordanUpdateVACSpare forepumpsI have placed 3 new in box, IDP 7 forepumps
along the x arm of the interferometer. These
are to be used as spares for both the 40m
  
  15668   Tue Nov 10 11:59:37 2020 gautamUpdateVACStuck RV2I've uploaded some more photos here.
I believe the problem is a worn out thread
where the main rotary handle attaches to
 damagedThread.001.jpegIFOstatus.pngP1a_leakTest.png 
  15681   Wed Nov 18 17:51:50 2020 gautamUpdateVACAgilent pressure gauge controller deliveredIt is stored along with the cables that
arrived a few weeks ago, awaiting the gauges
which are now expected next week sometime.
  
  15686   Mon Nov 23 16:33:10 2020 gautamUpdateVACMore vacuum deliveries Five Agilent pressure gauges were delivered
to the 40m. It is stored with the controller
and cables in the office area. This completes
  
  15692   Wed Dec 2 12:27:49 2020 JonUpdateVACReplacing pressure gaugesNow that the new Agilent full-range gauges
(FRGs) have been received, I'm putting
together an installation plan. Since
  
  15698   Thu Dec 3 10:33:00 2020 gautamUpdateVACTrippLite UPS deliveredThe latest greatest UPS has been delivered.
I will move it to near the vacuum rack in
its packaging for storage. It weighs >100lbs
 DFDD4F39-3F8A-439D-888D-7C0CE2E030CF.jpeg 
  15703   Thu Dec 3 14:53:58 2020 JonUpdateVACReplacing pressure gaugesUpdate to the gauge replacement plan (15692),
based on Jordan's walk-through today.
  
  15721   Wed Dec 9 20:14:49 2020 gautamUpdateVACUPS failureSummary:


The (120V) UPS at the vacuum
rack is faulty.
The drypump backing
 rackBeforenAfter.pdfIMG_0008.jpgIMG_0009.jpgvacStatus.png 
  15722   Thu Dec 10 11:07:24 2020 ChubUpdateVACUPS faultIs that a fault code that you can decipher
in the manual, or just a light telling you
nothing but your UPS is dead?
  
  15723   Thu Dec 10 11:17:50 2020 ChubUpdateVACUPS faultI can't find anything in the manual
that describes the nature of the FAULT message. 
In fact, it's not mentioned at all. 
  
  15724   Thu Dec 10 13:05:52 2020 JonUpdateVACUPS failureI've investigated the vacuum controls
failure that occurred last night. Here's
what I believe happened.
 TP2_time_history.pngslow_controls_monitors.png 
  15725   Thu Dec 10 14:29:26 2020 gautamUpdateVACUPS failureI don't buy this story - P2 only briefly
burped around GPStime 1291608000 which is
around 8pm local time, which is when I was
 vacDiag1.png 
  15748   Wed Jan 6 15:28:04 2021 gautamUpdateVACVac rack UPS batteries replaced[chub, gautam]

the replacement was done this afternoon.
The red "Replace Battery" indicator
  
  16047   Mon Apr 19 09:17:51 2021 JordanUpdateVACEmpty N2 TanksWhen I came into the lab this morning,
I noticed that both N2 tanks were empty.
I had swapped one on Friday (4-16-21) before
 N2_Pressure.PNG 
  16064   Wed Apr 21 12:56:00 2021 JordanUpdateVACEmpty N2 TanksInstalled T2 today, and leaked checked
the entire line. No issues found. It could
have been a bad valve on the tank itself.
 Screenshot_2021-04-21_12-53-26.png 
  16090   Wed Apr 28 11:31:40 2021 JonUpdateVACEmpty N2 TanksI checked out what happened on c1vac. There
are actually two independent monitoring codes
running:
  