ID |
Date |
Author |
Type |
Category |
Subject |
2759
|
Sat Apr 3 11:35:47 2010 |
rana | Configuration | PSL | Reference Cavity PD Noise Spectrum |
The units on this plot are completely bogus - we know that the thermal noise from the resonant part of the circuit is just V = sqrt(4*k*T*Z) ~ 3nV/rHz. Then the gain of the MAX4107 stage is 10. The output resistor is 50 Ohms, which forms a divide by 2 with the input impedance of the spectrum analyzer and so the bump in the dark noise should only be 15 nV/rHz and not microVolts.
Quote: |
[Rana, Alberto]
This evening we measured the noise spectrum of the reference cavity PD used in the FSS loop. From that we estimated the transimpedance and found that the PD is shot-noise limited. We also found a big AM oscillation in correspondence of the FSS modulation sideband which we later attenuated at least in part.
This plot shows the spectrum noise from the RF output of the photodetector.
|
|
2760
|
Sat Apr 3 16:07:40 2010 |
Alberto | Configuration | PSL | Reference Cavity PD Noise Spectrum |
I was aware of a problem on those units since I acquired the data. Then it wasn't totally clear to me which were the units of the data as downloaded from the Agilent 4395A, and, in part, still isn't.
It's clear that the data was in units of spectrum, an not spectral density: in between the two there is a division by the bandwidth (100KHz, in this case). Correcting for that, one gets the following plot for the FSS PD:

Although the reason why I was hesitating to elog this other plot is that it looks like there's still a discrepancy of about 0.5dBm between what one reads on the display of the spectrum analyzer and the data values downloaded from it.
However I well know that, I should have just posted it, including my reserves about that possible offset (as I'm doing now).
Quote:
|
The units on this plot are completely bogus - we know that the thermal noise from the resonant part of the circuit is just V = sqrt(4*k*T*Z) ~ 3nV/rHz. Then the gain of the MAX4107 stage is 10. The output resistor is 50 Ohms, which forms a divide by 2 with the input impedance of the spectrum analyzer and so the bump in the dark noise should only be 15 nV/rHz and not microVolts.
Quote: |
[Rana, Alberto]
This evening we measured the noise spectrum of the reference cavity PD used in the FSS loop. From that we estimated the transimpedance and found that the PD is shot-noise limited. We also found a big AM oscillation in correspondence of the FSS modulation sideband which we later attenuated at least in part.
This plot shows the spectrum noise from the RF output of the photodetector.
|
|
|
2789
|
Mon Apr 12 16:20:05 2010 |
Alberto | Configuration | 40m Upgrading | REFL55 improved |
During the commissioning of the AS55 PD, I learned how to get a much better rejection of the 11MHz modulation.
So I went back to REFL55 and I modified it using the same strategy. (Basically I added another notch to the circuit).
After a few days of continuous back and forth between modeling, measuring, soldering, tuning I got a much better transfer function.
All the details and data will be included in the wiki page (and so also the results for AS55). Here I just show the comparison of the transfer functions that I measured and that I modeled.
I applied an approximate calibration to the data so that all the measurements would refer to the transfer function of Vout / PD Photocurrent. Here's how they look like. (also the calibration will be explained in the wiki)
.
The ratio between the amplitude of the 55Mhz modulation over the 11MHz is ~ 90dB
The electronics TF doesn't provide a faithful reproduction of the optical response. |
2805
|
Mon Apr 19 05:54:50 2010 |
rana | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
In order to measure the transfer function of the RC cavity's foam, I've turned off the servo so that the room temperature noise can excite it.
The attached plot shows a step response test from 2 weeks ago. Servo is nominally still working fine. |
Attachment 1: Untitled.png
|
|
2808
|
Mon Apr 19 13:23:03 2010 |
josephb | Configuration | Computers | yum update fixed on control room machines |
I went to Ottavia, and tried running yum update. It was having dependancy issues with mjpegtools, which was a rpmforge provided package. In order to get it to update, I moved the rpmforge priority above (a lower number) that of epel ( epel -> 20 from 10, rpmforge -> 10 to 20). This resolved the problem and the updates proceeded (all 434 of them). yum update on Ottavia now reports nothing needs to be done.
I went to Rosalba and found rpmfusion repositories enabled. The only one of the 3 repositories in each file enabled was the first one.
I then added priority listing to all the repositories on Rosalba. I set CentOS-Base and related to priority=1. I set CentOS-Media.repo priority to 1 (although it is disabled - just to head off future problems). I set all epel related to priorities to 20. I set all rpmforge related priorities to 10. I set all rpmfusion related priorities to 30, and left the first repo in rpmfusion-free-updates and rpmfusion-nonfree-updates were enabled. All other rpmfusion testing repositories were disabled by me.
I then had to by hand downgrade expat to expat-1.95.8-8.3.el5_4.2.x86_64 (the rpmforge version). I also removed and reinstalled x264.x86_64. Lastly I removed and reinstalled lyx. yum update was then run and completed successfully.
I installed yum-priorities on Allegra and made all CentOS-Base repositories priority 1. I similarly made the still disabled CentOS-Media priority 1. I made all epel related repos priority 20. I made all lscsoft repos priority=50 (not sure why its on Allegra and none of the other ones). I made all rpmforge priorities 10. I then ran "yum update" which updated 416 packages.
So basically all the Centos control room machines are now using the following order for repositories:
CentOS-Base > rpmforge > epel > (rpmfusion - rosalba only) > lscsoft (allegra only)
I'm not sure if rpmfusion and lscsoft are necessary, but I've left them for now. This should mean "yum update" will have far fewer problems in the future.
|
2849
|
Tue Apr 27 11:16:13 2010 |
josephb | Configuration | CDS | Wiki page with CDS .mdl names, shared memory allocation |
I've added a new page in the wiki which describes the current naming scheme for the .mdl model files used for the real time code generator. Note, that these model names do not necessarily have to be the names of the channels contained within. Its still possible to make all suspension related channels start with C1:SUS- for example. I'm also allocating 1024 8 byte channels for shared memory address space for each controller and each simulated plant.
The wiki page is here
Name suggestions, other front end models that are needed long term (HEPI is listed for example, even though we don't have it here, since in the long run we'd like to port the simulated plant work to the sites) are all welcome. |
2873
|
Mon May 3 17:49:41 2010 |
rana | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
Quote: |
In order to measure the transfer function of the RC cavity's foam, I've turned off the servo so that the room temperature noise can excite it.
The attached plot shows a step response test from 2 weeks ago. Servo is nominally still working fine.
|
I've just now re-enabled the temperature control of the reference cavity can. Trend of the last 8 days is attached. |
Attachment 1: rct.png
|
|
2875
|
Tue May 4 02:28:38 2010 |
rana | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
My attempt to passively measure the transfer function of the foam failed fantastically.
As it turns out, the room temperature fluctuations inside the PSL box reach the 1 mK/rHz noise floor of the AD590 (or maybe the ADC) at ~1-2 mHz. Everything at higher frequencies is noise.
So to see what the foam is doing we will have to do something smarter - we need a volunteer to disable the RC temperature servo from the EPICS screen and then cycle the PSL table lights every hour in the morning.
We'll then use our knowledge of the Laplace transform to get the TF from the step responses. |
2876
|
Tue May 4 06:32:58 2010 |
alberto | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
Quote: |
My attempt to passively measure the transfer function of the foam failed fantastically.
As it turns out, the room temperature fluctuations inside the PSL box reach the 1 mK/rHz noise floor of the AD590 (or maybe the ADC) at ~1-2 mHz. Everything at higher frequencies is noise.
So to see what the foam is doing we will have to do something smarter - we need a volunteer to disable the RC temperature servo from the EPICS screen and then cycle the PSL table lights every hour in the morning.
We'll then use our knowledge of the Laplace transform to get the TF from the step responses.
|
more detailed instructions needed.... |
2881
|
Wed May 5 02:37:55 2010 |
rana | Configuration | PSL | RC Temperature Servo Turned OFF temporarily |
Quote: |
more detailed instructions needed....
|
I showed Kiwamu and Alberto how to turn the lights on and off in the PSL. This is why Caltech is such a fine institution: most schools would have TAs delivering this kind of optics instruction.
We've turned off the RC temperature stabilization and the lights will supply the quasi-random heat input to the table and the cavity. Alberto and Kiwamu will be turning the lights on and off at random times.
The attached plot is the spectrum of temperature fluctuations of the room and the vacuum can with no stabilization from this weekend. I think the rolloff above 10 mHz is kind of fake - I had the .SMOO parameter set to 0.99 for both of these channels. I've just now set the .SMOO to 0 for both channels, so we should now see the true ADC or sensor noise level. It should be ~1 mK/rHz. |
Attachment 1: Picture_7.png
|
|
2901
|
Sun May 9 20:02:23 2010 |
rana | Configuration | SUS | SUS filters deleted again to reduce CPU load on c1susvme2 again |
On Friday, I deleted a bunch of filters from the c1susvme2 optics' screens (MC1,2,3 + SRM) so as to reduce the CPU load and keep it from going bonkers.
This first plot shows the CPU trend over the last 40 days and 40 nights. As you can see the CPU_LOAD has dropped by 1 us since I did the deleting.
 
In the second plot (on the right) you can see the same trend but over 400 days and nights. Of course, we hope that we throw this away soon, but until then it will be nice to have the suspensions be working more often. |
2922
|
Wed May 12 12:32:04 2010 |
josephb | Configuration | CDS | Modified /etc/rc.d/rc.local on megatron |
I modified the /etc/rc.d/rc.local file on megatron removing a bunch of the old test module names and added the new lsc and lsp modules, as well as a couple planned suspension models and plants, to shared memory so that they'll work. Basically I'm trying to move forward into the era of working on the actual model we're going to use in the long term as opposed to continually tweaking "test" models.
The last line in the file is now: /usr/bin/setup_shmem.rtl lsc lsp spy scy spx scx sus sup&
I removed mdp mdc mon mem grc grp aaa tst tmt. |
2923
|
Wed May 12 12:58:26 2010 |
josephb | Configuration | CDS | Setup fb to handle lsc, lsp models on megatron |
I modified /cvs/cds/caltech/target/fb and changed the line "set controller_dcu=10" to "set controller_dcu=13" (where 13 is the lsc dcu_id number).
I also changed the set gds_server line from having 10 and 11 to 13 and 14 (lsc and lsp).
The file /cvs/cds/caltech/fb/master was modified to use C1LSC.ini and C1LSP.ini, as well as tpchn_C2.par (LSC) and tpchn_C3.par (LSP)
testpoint.par in /cvs/cds/caltech/target/gds/param was modified to use C-node1 and C-node2 (1 less then the gds_node_id for lsc and lsp respectively).
Note all the values of gds_node_id, dcu_id, and so forth are recorded at http://lhocds.ligo-wa.caltech.edu:8000/40m/Electronics/Existing_RCG_DCUID_and_gds_ids |
2938
|
Mon May 17 02:10:10 2010 |
Koji | Configuration | IOO | How to lock / align the MC |
Let me remind you how to lock and align the IMC
To lock
1. Open the doors for the IMC/OMC chambers. Open the manual shutter of the PSL just in front of the optical window
2. Run scripts/MC/mcloopson
3. Set the MC length path gain 0.3 / Set the MC total gain "+20"
4. If you want to avoid excitation of the mirrors by air turbulence, put a big plastic film and put three posters on the top and both the sides on the floor to block the wind go into the chamber.
To shut down
1. Run scripts/MC/mcloopsoff
2. Close the manual shutter, Remove the wind blockers, and the light door of the chambers
To align the MC
1. Tweak MC2 and MC3 to get maximum transmittion and/or minimum reflection. |
2939
|
Mon May 17 10:57:16 2010 |
steve | Configuration | SUS | damping restored to ETMYs |
ETMY-south sus damping was restored |
2943
|
Tue May 18 11:54:29 2010 |
steve | Configuration | SAFETY | psl mechanical shutter is not working |
As we learned yesterday, the PSL laser power out put mechanical shutter is not working in the remote mode. It only works in local manual mode.
Do not rely on the MEDM screen monitor readout! The position is only changing on the monitor. The main beam must be blocked before the output periscope. |
2947
|
Tue May 18 15:09:02 2010 |
steve | Configuration | SAFETY | PSL output shutter in order again |
Quote: |
As we learned yesterday, the PSL laser power out put mechanical shutter is not working in the remote mode. It only works in local manual mode.
Do not rely on the MEDM screen monitor readout! The position is only changing on the monitor. The main beam must be blocked before the output periscope.
|
Ben found the Sorenson 5V ps off. It was turned off since our last scheduled power outage. I wonder what else is running on 5V in the PSL? This power supply should be on the
"alarm handler" list to avoid future repeat of this condition. However a real safety switch would have confirming position sensors of the shutter open or closed. Is there such thing at the sites? |
2957
|
Thu May 20 12:34:46 2010 |
kiwamu | Configuration | 40m Upgrading | optical breadboards with legs |
Yesterday Steve and I revived two legs to mount some optical breadboards outside of the end table.
These legs had been used as oplev's mounts many years ago, but now they are served for 40m upgrading. These are really nice.
By putting them on the side of the end table, a mirror mounted on the top of the leg can reflect the beam outside of the end table.
Once we pick off the green beam from the end table to its outside, the green beam can propagate through the 40m walkway along the Y-arm.
So that we can measure the beam profile as it propagates.
These legs are also going to be used during mode matching of the vacuum optics. |
Attachment 1: leg1_small.png
|
|
Attachment 2: leg2_small.png
|
|
2965
|
Fri May 21 08:28:29 2010 |
steve | Configuration | VAC | TP3 turned off |
Quote: |
[Jenne, Kiwamu, and Steve via phone]
Around 9:30pm, Kiwamu and I came back from dinner, and were getting ready to begin the beam scan measurements. I noticed that one of the vacuum pumps was being very loud. Kiwamu noted that it is the fore pump for TP3's turbo, which he and Steve replaced in January (elog 2538). We had not noticed these noises before leaving for dinner, around 8pm.
We called Steve at home, and he could hear the noise through the phone. He said that even though it was really loud, since it was reading 3.3mTorr (on the display of the controller, in the vacuum rack just above head-height) which is close to the nominal value, it should be fine to leave. He will check it out in the morning. If it had been reading at or above ~1Torr, that's indicative of it being really bad, and we would have needed to shut it off.
For future reference, in case we need to turn it off, Steve said to use the following procedure:
1. Close VM3, to isolate the RGA, which is what this pump is currently (while we're at atmosphere) pumping on. I don't know if there are other things which would need to be shut at this stage, if we were at vacuum nominal.
2. Close VM5, which is right in front of TP3, so TP3's pump is just pumping on itself.
3. Push the "Stop" button on the Turbo controller for TP3, in the vacuum rack, about waist level. Turning off the turbo will also turn off the fore pump.
UPDATE, 1am: The controller in the rack is reading 3.1mTorr, so the pump, while still noisy, still seems to be working.
|
The foreline pressure of TP3 is 2.9 mTorr The drypump is loosing it's bearing and it is very noisy.
V3, V5 closed and TP3 small turbo controller off. This turned off the noisy forepump that has to be replaced.
RGA is running at cc4 2e-6 Torr
The RGA was turned off at cc4 1e-5 Torr |
2970
|
Fri May 21 16:37:35 2010 |
steve | Configuration | VAC | tp3's drypump replaced |
Bob replaced the tipseal in an other drypump and I swapped it in. TP3 turbo is running again, it's foreline pressure is 40 mTorr. The RGA is still off |
2972
|
Mon May 24 07:53:57 2010 |
steve | Configuration | PEM | air cond. just turned ON |
IFO room temp 27.5C , Please remember to turn AC back on ! |
Attachment 1: acoff.jpg
|
|
2980
|
Tue May 25 09:12:46 2010 |
kiwamu | Configuration | Green Locking | effect from air conditioner |
We should completely turn off the air conditioner when working on green locking.
Even if green beams propagates inside of chambers, the air conditioner does affect the spatial jitter of the beam.
The attached picture was taken when Steve and I were seeing how the green beam jittered.
At that time the beam was injected from the end table and going through inside of the ETM, the ITM and the BS camber.
Eventually it came out from the camber and hit the wall outside of the chamber. It was obvious, we could see the jittering when the air cond. was ON. |
Attachment 1: green_spot.png
|
|
2996
|
Wed May 26 22:22:03 2010 |
Alberto | Configuration | 40m Upgrading | Arm cavity length |
The second sideband is resonant in the arms for a cavity length of 37.9299m.
The nearest antiresonant arm lengths for f2 (55MHz) are 36.5753m and 39.2845m.
If we don't touch the ITMs, and we use the room we still have now on the end tables, we can get to 37.5m.
This is how the power spectrum at REFL would look like for perfect antiresonance:

And this is how it looks like for 37.5m:

Or, god forbid, we change the modulation frequencies... |
3015
|
Sun May 30 15:33:21 2010 |
Alberto | Configuration | IOO | mode cleaner and air conditioning |
The mode cleaner is locked and the air conditioning is full on. So the the air conditioning doesn't seem to be so important for the lock to hold. |
3016
|
Sun May 30 15:36:22 2010 |
Alberto | Configuration | PSL | IMC periscope shutter |
Two days ago I opened the PSL shutter by switching the switch on the shutter driver. That caused the shutter's switch on the medm screen to work in reversed mode: open meant closed and closed meant open.
I fixed that. Now the medm screen switch state is correct. |
3023
|
Tue Jun 1 06:30:38 2010 |
Koji | Configuration | 40m Upgrading | effect of the arm length |
I checked the effect of the arm length to the reflectance of the f2(=5*f1) sidebands.
Conclusion: If we choose L_arm = 38.4 [m], it looks sufficiently being away from the resonance
We may want to incorporate small change of the recycling cavity lengths so that we can compensate the phase deviation from -180deg.
f1 of 11.065399MHz is assumed. The carrier is assumed to be locked at the resonance.
Attachment 1: (Left) Amplitude reflectance of the arm cavity at f2 a a function of L_arm. (Right) Phase
Horizontal axis: Arm length in meter, Vertical Magnitude and Phase of the reflectance
At L=37.93 [m], f2 sidebands become resonant to the arm cavity. Otherwise, the beam will not be resonant.
Attachment 2: close-up at around 5 f1 frequency.
The phase deviation from the true anti resonance is ~0.7deg. This can be compensated by both PRC and SRC lengths. |
Attachment 1: arm_cavity1.png
|
|
Attachment 2: arm_cavity2.png
|
|
Attachment 3: arm_cavity.nb.zip
|
3081
|
Wed Jun 16 18:12:16 2010 |
nancy | Configuration | Computers | 40MARS |
i added my laptop's mac address to teh martian at port 13 today.
|
3083
|
Wed Jun 16 18:44:07 2010 |
Alberto | Configuration | Computers | 40MARS |
Quote: |
i added my laptop's mac address to teh martian at port 13 today.
|
No personal laptop is allowed to the martian network. Only access to the General Computing Side is permitted.
Please disconnect it. |
3109
|
Wed Jun 23 18:05:00 2010 |
Koji | Configuration | PSL | FSS SLOWDC should be ~-4.0 |
FSS SLOWDC slider is at around 0.
Please someone relock this at ~-4.0 to exploit some last juice of the fruit.
See this entry for the details of the operating point.
|
Attachment 1: C1PSL_FSS.png
|
|
3110
|
Wed Jun 23 23:08:30 2010 |
rana | Configuration | PSL | FSS SLOWDC should be ~-4.0 |
|
3133
|
Tue Jun 29 11:48:17 2010 |
Jenne | Configuration | SAFETY | Back in LASER HAZARD mode. |
[Steve, Kiwamu, Jenne]
The 40m is now back in Laser Hazard mode. Safety glasses are required for entry into the LVEA / IFO room. |
3135
|
Tue Jun 29 14:16:35 2010 |
Koji | Configuration | SAFETY | Back in LASER HAZARD mode. |
The insects and the laser trouble... Strange coincidences with LHO surprised me, but now I have been relieved.
Quote: |
[Steve, Kiwamu, Jenne]
The 40m is now back in Laser Hazard mode. Safety glasses are required for entry into the LVEA / IFO room.
|
|
3151
|
Wed Jun 30 23:03:46 2010 |
rana | Configuration | IOO | Power into MC restored to max |
Kiwamu, Nancy, and I restored the power into the MC today:
- Changed the 2" dia. mirror ahead of the MC REFL RFPD back to the old R=10% mirror.
- Since the MC axis has changed, we had to redo the alignment of the optics in that area. Nearly all optics had to move by 1-2 cm.
- 2 of the main mounts there had the wrong handedness (e.g. the U100-A-LH instead of RH). We rotated them to some level of reasonableness.
- Tuned the penultimate waveplate on the PSL (ahead of the PBS) to maximize the transmission to the MC and to minimize the power in the PBS reject beam.
- MC_REFL DC =1.8 V.
- Beams aligned on WFS.
- MC mirrors alignment tweaked to maximize transmission. IN the morning we will check the whole A2L centering again. If its OK, fine. Otherwise, we'll restore the bias values and align the PSL beam to the MC via the persicope.
- waveplates and PBS in the PSL were NOT removed.
- MC TRANS camera and QPD have to be recentered after we are happy with the MC axis.
- MC REFL camera has to be restored.
- WFS measurements will commence after the SURF reports are submitted.
We found many dis-assembled Allen Key sets. Do not do this! Return tools to their proper places or else you are just wasting everyone's time!
|
3152
|
Thu Jul 1 10:26:34 2010 |
steve | Configuration | VAC | pd68 at vacuum normal state |
Slow pump down _pd68 has reached the vacuum normal state. CC4_Rga region is pumped now. The RGA is still off.
Precondition to this pump down: 129 days at atm, ITMs replaced. MMT, oplev and other components were removed from BSC, ITMCs. New MMT mirrors are in. IOO_access_connector was out. The end chambers were not opened.
|
Attachment 1: pd68slow2d.jpg
|
|
Attachment 2: pem134d.jpg
|
|
3170
|
Wed Jul 7 17:18:57 2010 |
Alberto | Configuration | Electronics | Stochmon and LSC AM Stabilizer Decomissioned |
Today I disconnected and removed the Stochmon box from the 1Y2 rack.
I also removed the amplifiers that were sitting on the PSL table, next to the RF AM PD, that were connected to the Stochmon. I pulled back the RG cable and the power cables that went from the PSL table to the 1Y1 and 1Y2 racks.
The power cable, all rolled up, is now sitting on the floor, inside the 1Y1 rack and one of its end is still connected to the power of the rack. We'd like to turn off the entire rack in order to safely remove it. But since the laser driver is there too, we should do it the first time we have to turn off the rack for some other reason.
I also removed two of the AM stabilizers from the 1Y2 rack. The other one, which is currently running th MC modulations, is still in the rack, and there it is going to remain together with its distribution box.
I stored both AM stabilizers and the Stochmon box inside the RF cabinet down the East arm. |
3173
|
Wed Jul 7 22:52:38 2010 |
rana, nancy | Configuration | IOO | bad length control offset for the MC |
Rana found out that a connection was bad in the shown place, due to which the MEDM screen was showing bad offset for length control.
Basically, the offset slider value would not go into the system because of that bad connection, and was locking the mode cleaner at the wrong location. |
Attachment 1: P7070251.jpg
|
|
Attachment 2: Screenshot.png
|
|
3178
|
Thu Jul 8 15:19:27 2010 |
josephb, koji | Configuration | Computers | Added Zonet camera to IP table on linux1 |
We gave the Zonet camera the IP 192.168.113.26 and the name Zonet1.
We did this by modifying the /var/named/chroot/var/named/113.168.192.in-addr.arpa.zone and martian.zone files on linux1 as root. |
3191
|
Mon Jul 12 02:21:01 2010 |
Koji | Configuration | ASC | Resurrection of MC WFS |
I have resurrected the MC WFS on Friday night.
I have uncommented the WFS part of the MC autolocker.
The WFS total gain was empirically set to 0.1 such that the loops have no instability.
The loops somewhat worked through the weekend although they seemed to have the drift of the operating points
in accordance with the WFS spot. |
3206
|
Tue Jul 13 13:56:29 2010 |
Gopal | Configuration | Optic Stacks | Vitol Material Properties |
Viton flouroelastomers have somewhat variable material properties. The following parameters are being used for eigenfrequency analysis.
Young's Modulus: 72,500-87,000 psi (cite: http://www.row-inc.com/pfa.html) *Accurate for PFAs
Poisson's Ratio: 0.48-0.50 (cite: http://www.engineeringtoolbox.com/poissons-ratio-d_1224.html) *Accurate for rubber
Density: 1800 kg/m^3 (cite: http://physics.nist.gov/cgi-bin/Star/compos.pl?matno=275) |
3222
|
Wed Jul 14 19:00:56 2010 |
Gopal | Configuration | Optic Stacks | Vitol Material Properties, REVISED |
Quote: |
Viton flouroelastomers have somewhat variable material properties. The following parameters are being used for eigenfrequency analysis.
Young's Modulus: 72,500-87,000 psi (cite: http://www.row-inc.com/pfa.html) *Accurate for PFAs
Poisson's Ratio: 0.48-0.50 (cite: http://www.engineeringtoolbox.com/poissons-ratio-d_1224.html) *Accurate for rubber
Density: 1800 kg/m^3 (cite: http://physics.nist.gov/cgi-bin/Star/compos.pl?matno=275)
|
The Young's Modulus for PFA turned out to be orders of magnitude greater than for Viton. The revised values produced much more accurate results:
Young's Modulus for Viton-75: 1950 psi or 6.6 MPa
(Courtesy Row, Inc. and Steve Vass)
This website contains other details: http://www.row-inc.com/viton.html |
3227
|
Thu Jul 15 12:21:08 2010 |
Alberto | Configuration | LSC | PRC and SRC length adjustements |
Lately I've been trying to calculate the corrections to the recycling cavity lengths that would compensate for the phase that the sidebands will pick up from the arms in the upgraded interferometer.
To do that calculation , I tried two quite different ways, although equivalent in principle. They both use the optickle model of the 40m, but the calculation is made differently.
In the first way, I looked directly at the phases of the field: phase of [input field] / [reflected field], phase of [input field at PRM] / [transmitted field at SRM].
In the second way I looked at the demodulation phases of the LSC signals.
The first way is much simpler, especially from a computational point of view. It is the first I tried several weeks ago, but then I had abandoned because back then I thought it wasn't the correct way.
Anyway, both ways gave me the same results for the PRC length.
For the SRC length, the first way has given me a clear outcome. On the other hand, the second way has produced a less clear result.
According to these results, these would be the proposed adjustements to the cavity lengths:
dl(PRC) = -0.0266 m; dl(SRC) = 0.0612 m
I) 1st Way
a) case of arms ideal length (33.86 m)

b) case arm length = 38.40 m
PRC
zoom -> 
SRC
zoom -> 
II) 2nd Way
a) case of arms ideal length (33.86 m)

b) case arm length = 38.40 m

|
3228
|
Thu Jul 15 15:57:10 2010 |
Koji | Configuration | LSC | PRC and SRC length adjustements |
Tell me whether it is correct or not. Otherwise I won't be able to sleep tonight.
Quote: |
According to these results, these would be the proposed adjustements to the cavity lengths:
dl(PRC) = -0.0266 m; dl(SRC) = 0.612 m
|
|
3229
|
Thu Jul 15 16:16:51 2010 |
Alberto | Configuration | LSC | PRC and SRC length adjustements |
Quote: |
Tell me whether it is correct or not. Otherwise I won't be able to sleep tonight.
Quote: |
According to these results, these would be the proposed adjustements to the cavity lengths:
dl(PRC) = -0.0266 m; dl(SRC) = 0.612 m
|
|
Sorry. I was in a rush to go to the LIGO "all hands" meetings when I posted that elog entry, that I forgot a zero in the SRC length value. The correct values are:
dl(PRC) = -0.0266 m; dl(SRC) = 0.0612 m
The cavity absolute lengths are then:
L(PRC) = 0.5/2/f1*c - 0.0266 = 6.7466 m
L(SRC) = c/f2 + 0.0612 = 5.4798 m
where c is the speed of light; f1 = 11065399 Hz; f2 = 55326995 Hz |
3239
|
Fri Jul 16 16:12:31 2010 |
Alberto | Configuration | Computers | c1susvme1/2 rebooted |
Today I noticed that the FE SYNC counters of c1susvme1/2 on the RFM network screen were stuck at 16384. I tried to reboot the machines to fix the problem but it didn't work.
The BS watchdog tripped off when I did that, because I had forgotten to disable it. I had to wait for a few minutes before it settled down again.
Later I also re-locked the mode cleaner. But before I could do it, Rana had to reduce the MC_L offset for me. |
3242
|
Sun Jul 18 20:50:03 2010 |
rana | Configuration | IOO | MC TRANS optics changed |
To make the beam on the MC trans camera bigger, I removed the lens + ND filter that was in that path.
The camera was getting the transmission through a BS1-33 (33% reflector). The reflection went to the TRANS QPD. I changed
the R=33% into an HR mirror (Y1-45P) so now the camera has a nice beam. The QPD was now saturating so I put a ND06 into that path
so now the TRANS_SUM is ~4.5-5 V when the MC is aligned.
The MC was also misaligned and failing to lock all weekend (why??) so I aligned the MC mirrors to get it to acquire again. Since we want to
collect MC seismic data, please make sure the MC is locked and running after finishing with your various MC or PSL work (this means YOU). |
3248
|
Tue Jul 20 08:04:31 2010 |
steve | Configuration | VAC | slow vent has started |
The PSL shutter was closed. The beam path blocked two places. High voltage power supplies to IOO and OMC PZT were checked to be off. Oplevs are off.
The south arm green cavity was misaligned yesterday
We would like to keep the vent speed at 1 torr / min. I'm venting with N2 now up to 25 PSI. We have 3 cylinder of instrument grade air inside the lab. Additional supply will arrive later. It can be as late as 1pm |
3252
|
Tue Jul 20 17:38:16 2010 |
Gopal | Configuration | Optic Stacks | Stack Type Clarifications |
Some clarification is warranted regarding the different shapes of stacks. Corrections are appreciated:
1) The single-leg stack that I just completed should function as a working model for the IO, OO, and MC1/3. Rana commented, however, that the current dimensions are slightly off for MC1/3 (which makes sense since I could only find drawings for the IOC). If anyone knows the whereabouts of similar drawings for MC1/3, I'd much appreciate it.
2) A triple-leg stack can model the BS, ITMX, and ITMY chambers. I don't have exact dimensions for these, but I can make decent approximations from to-scale stack drawings. I'll probably work on a model for this style next, since at least I have some information regarding this version.
3) The MC2 chamber has its own stack model, about which I haven't found any drawings in the binders. I can't start on MC2C at all until I find such drawings. |
3261
|
Wed Jul 21 17:41:17 2010 |
Gopal | Configuration | Optic Stacks | Pictures of Stacks |
Now that venting is complete, this is a request for anyone who opens any chamber:
1) Please notify me immediately so I can take pictures of the stacks in that chamber.
2) If I'm not around, please take a few pictures for me. I'm most interested in the shape, number of layers, size, and damper arrangements of each stack.
This is most important for the MC1/MC3 chamber, MC2 chamber, and BS/ITMX/ITMY chambers.
Thanks! |
3331
|
Fri Jul 30 09:55:34 2010 |
steve | Configuration | VAC | slow pumpdown has started |
Bob and Steve closed BS chamber with the help of the manual Genie lift and the pump down started. The PSL shutter was closed and manual block was placed in the beam path. High voltage power supplies were checked to be off.
Pumping speed ~ 1 Torr/min was achieved at 1/8 of a turn opened roughing valve RV1 |
Attachment 1: slowpd5hr.jpg
|
|
3332
|
Fri Jul 30 12:46:38 2010 |
Koji | Configuration | 40m Upgrading | BS chamber status |
|
Attachment 1: BS_layout_Jul_29_2010.png
|
|