  15211   Thu Feb 13 21:30:55 2020 shrutiUpdateALSALS OOL noise with arms locked

[Meenakshi, Gautam, Shruti]

Summary:

- We initially aligned the arm cavities to get the green lasers locked to them. For the X arm cavity, we tweaked the ITMX and ETMX pitch and yaw and toggled the X green shutter until we saw something like a TEM00 mode on the monitor and a reasonable transmitted power.

- With the LSC servo enabled, the IR light also became resonant with the cavities.

- Then we measured the noise in different configurations. Attachment 1 shows the the ALS OOL (in the IR beat signal) noise with the arms locked inidividually via PDH.


The script for plotting the ALS beat frequency noise is:

users/Templates/ALS/ALS_outOfLoop_Ref.xml
  15213   Fri Feb 14 14:02:13 2020 shrutiUpdateALSALS OOL noise with arms locked

[Meenakshi, Shruti]

Even though we were not able to lock the the IR beat (by enabling LSC) during the day possibly because of increased seismic activity, we tried to the measure the ALS beat frequency noise by changing the PDH side-band frequency to different values.

I tried choosing values that corresponded to the peaks in the PM/AM as found in elog:15206 but for some reason unknown to us the cavity did not lock between 700-800 kHz.

The three attachments have data for different sideband frequencies:

Attachment 1: 819.472 kHz (peak in PM/AM, measured around noon)

Attachment 2: 225.642 kHz (peak in PM/AM, measured earlier in the morning)

Attachment 3: 693.500 kHz (not a peak in PM/AM)

We don't think these plots mean much and will do the measurement at some quieter time more systematically.

 

While doing the experiment, the ITMY pitch actuation was changed from -2.302 to -2.3172V because it locked better.

The ITMX, ETMX alignment was also tweaked to try to lock with different sideband frequencies and then restored to the values that were found earlier in the morning.

  15216   Tue Feb 18 18:14:59 2020 shrutiUpdateALSALS OOL noise with arms locked

We proceeded with the trying to measure the ALS out-of-loop noise of the X arm when the X arm cavity is green-locked using different PDH sideband frequencies.

Before doing the experiment, Koji helped us with getting the arm cavities locked in IR using LSC (length) and ASC (angular).

With the arms locked in IR and green, we repeated the same measurements as before at different sideband frequencies (Refer Attachment 1 - label in Hz). We did not optimize the phase nor did we look at the PDH error signal today which is possibvly why we did not see an improvement in the noise. We will look into this possibly tomorrow.

  15218   Fri Feb 21 10:59:08 2020 shrutiUpdateALSPDH error signals?
Here are a few PDH error signals measured without changing the servo gain or phase from that optimized for 231.25 kHz. This was done by keeping the X arm cavity and laser unlocked but keeping the shutter for green open; so I did not force a frequency sweep but saw the unhindered motion of cavity wrt the laser using the PDH servo error monitor channel from the box (not sure if this is the best way to do it?).
 
Koji mentioned that there is a low pass filter with a cutoff frequency probably lower than 700 kHz which at the moment would hinder the efficacy of the locking at higher frequencies. The transfer function on the wiki suggests the same, although we are yet to investigate the circuit.
 
I measured the maximum range in the linear region of the signal, and here are the results:
  • Attachment 1: 231.25 kHz (current PDH sideband mod freq): 1.7 V
  • Attachment 2: 225.642 kHz: 1.2 V
  • Attachment 3: 100 kHz: 900 mV
  • Attachment 4: 763.673 kHz: 220 mV
Right now we have only inverted the phase to try locking at different frequencies (no finer adjustments were performed so elog 15216 cannot be an accurate representation of the true performance)
 
Ideas from the 40m meeting for adjusting the phase:
  1. Delay line for adding extra phase (would require over 40m of cable for 90 deg phase shift)
  2. Using two function generators for generating the sideband, clocked to each other, so that one can be sent to the PZT and the other to the mixer for demodulation.
  3. Use a different LPF (does not seem very useful for investigating multiple possible frequencies)

Once we adjust the phase we may be able to increase the servo gain for optimal locking. Unless it may be a good idea to increase the gain without optimizing the phase?

  15220   Fri Feb 21 20:44:18 2020 shrutiUpdateALSALS OOL noise and PDH

[Meenakshi, Shruti]

In order to adjust the relative phase for PDH locking, we used the Siglent SDG 1032X function generator which has two outputs whose relative phase can be adjusted.

This Siglent function generator was borrowed from Yehonathan's setup near the PSL table and can be found at the X end disconnected from our setup after our use.

Initially, we used the Siglent at 231.250 kHz and 5 Vpp from each output with zero relative phase to lock the green arm cavity. By moving the phase at intervals of 5deg and looking at the PDH error signals when the cavity was unlocked we concluded that 0deg probably looked like it had the largest linear region (~1.9 V on the yaxis. Refer elog 15218 for more information) as expected.

Then we tried the same for 225.642 kHz, 5 Vpp, and found the optimal demod phase to be -55deg, with linear region of ~3 V (Ref. Attachment 2). A 'bad' frequency 180 kHz was optimized to 10deg and linear region of ~1.5 V.

The error signals at higher frequencies appeared to be quite low (not sure why at the moment) and tuning the phase did not seem to help this much.

For the noise measurement, the IFO arms were locked to IR and green, but even after optimizing the transmission with dither, we couldn't achieve best locking (green transmission was around ~0.2). Further, the IMC went out of lock during the experiment after which Koji helped us by adjusting the gains a locking point of the PMC servo. Attachment 1 contains some noise curves for the 3 frequencies with a reference from an earlier 'good' time.

  12342   Wed Jul 27 15:03:02 2016 sisyphusMetaphysicsSUSBroken wire

While tightening the bolts on the ETMX wire clamp, the wire broke. All four face magnets broke off. 

Fortunately, no pieces were lost.

  4958   Fri Jul 8 20:50:49 2011 sonaliUpdateGreen LockingPower of the AUX laser increased.

The ETMY laser was operating at 1.5 A current and 197 mW power.

For the efficient frequency doubling of the AUX  laser beam at the ETMY table, a higher power is required.

Steve and I changed the current level of the laser from 1.5 A to 2.1 A in steps of 0.1 A and noted the corresponding power output . The graph is attached here.

The laser has been set to current 1.8 Amperes. At this current, the power of the output beam just near the laser output is measured to be 390 mW.

The power of the beam which is being coupled into the optical fibre is measured to be between 159 mW to 164 mW (The power meter was showing fluctuating readings).

The power out of the beam coming out of the fibre far-end at the PSL table is measured to be 72 mW. Here, I have attached a picture of the beam paths of the ETMY table with the beams labelled with their respective powers.

Next we are going to adjust the green alignment on the ETMY and then measure the power of the beam.

At the output end of the fibre on the PSL, a power meter has been put to dump the beam for now as well as to help with the alignment at the ETMY table.

  4965   Thu Jul 14 02:32:11 2011 sonaliUpdateGreen LockingPower of the AUX laser increased.

Quote:

The power of the beam which is being coupled into the optical fibre is measured to be between 159 mW to 164 mW (The power meter was showing fluctuating readings).

The power out of the beam coming out of the fibre far-end at the PSL table is measured to be 72 mW. Here, I have attached a picture of the beam paths of the ETMY table with the beams labelled with their respective powers.

 For the phase locking or beat note measuring we only need ~1 mW. Its a bad idea to send so much power into the fiber because of SBS and safety. The power should be lowered until the output at the PSL is < 2 mW. In terms of SNR, there's no advantage to use such high powers.

  4973   Fri Jul 15 13:48:56 2011 sonaliUpdateGreen LockingPower of the AUX laser increased.

Quote:

Quote:

The power of the beam which is being coupled into the optical fibre is measured to be between 159 mW to 164 mW (The power meter was showing fluctuating readings).

The power out of the beam coming out of the fibre far-end at the PSL table is measured to be 72 mW. Here, I have attached a picture of the beam paths of the ETMY table with the beams labelled with their respective powers.

 For the phase locking or beat note measuring we only need ~1 mW. Its a bad idea to send so much power into the fiber because of SBS and safety. The power should be lowered until the output at the PSL is < 2 mW. In terms of SNR, there's no advantage to use such high powers.

 

Well,the plan is to put in  a neutral density filter in the beam path before it enters the fibre. But before I could do that, I set up the camera on the PSL table to look at the fiber output . I will need it while I realign the  beam after putting in the Neutral Density Filter. I have attached the ETMY layout with the Neutral Density filter in place herewith.

  10   Tue Oct 23 11:08:20 2007 steveOtherGeneralbrush fires
There are big brush fires around LA
40 days plot show no effect in the 40m lab
  12   Wed Oct 24 08:58:09 2007 steveOtherPSLlaser headtemp is up
C1:PSL-126MOPA_HTEMP is 19.3C

Half of the chiller's air intake was covered by loose paper
  17   Fri Oct 26 09:10:17 2007 steveRoutinePEMPEM &PSL trend
The fires are out, lab particle counts are up.
Psl HEPAs are at 100% and mobel HEPAs are just turned on
20 days plot and 5 hrs plot below
  23   Mon Oct 29 09:16:31 2007 steveRoutineVACthe rga is back
We had no filament current since last power glitch of Oct. 8, 2007
First I thought that the filament was lost, but it was only bad contact.
The rga head pins were oxidized. Rga was turned back on last Friday.
It's temp is 55.3C normal
  24   Mon Oct 29 09:46:50 2007 steveRoutineVACvac & pem trend
Pumpdown 64 pumped by maglev for 125 days
pd64-m-d125

Rob, can you tell me, when did the fire start on this plot?
  44   Thu Nov 1 09:17:27 2007 steveRoutineVACvent 64
Yesterday before vent I could not lock MC, therfore I could not measure the
transmitted power at MC2
The vent went well. We had lots of help.

We could not find the Nikon D40
PLEASE BORROW THINGS when taking them away
and bring them back promtly.

The laser was turned off for better visibility.

I see clean room frorks laying around here and there.
Please put them away so we do not carry excess particles into the chamber.
  57   Fri Nov 2 08:59:30 2007 steveBureaucracySAFETYthe laser is ON
The psl laser is back on !
  65   Tue Nov 6 09:14:37 2007 steveSummaryVACpump down 65
8 hr plot,
precondition: 5 days at atm,
vent objective: drag wiping mc1, mc2 & mc3 accomplished,
hardware changes: IOO access connector, mc2 chamber door south & west
were removed and reinstalled
pump down mode: slow to avoid steering up dust
One roughing pump was used with closed down valve position in the first 4 hrs

Andrey was very helpful
  66   Tue Nov 6 09:45:22 2007 steveSummarySUSvent sus trend
The mc optics dragwippings were done by locking optics by eq stops and rotating-moving
cages so access were good. This technic worked well with mc1 & mc2
MC3 osems were reoriented only.
  76   Wed Nov 7 09:38:01 2007 steveUpdateVACrga scan
pd65-m-d2 at cc1 6e-6 torr
  81   Wed Nov 7 16:07:03 2007 steveUpdatePSLPSL & IOO trend
1.5 days of happy psl-ioo with litle bumps in C1:PSL-126MOPA_HTEMP
  83   Thu Nov 8 11:40:21 2007 steveUpdatePEMparticle counts are up
I turned up the psl HEPA filter to 100%
This 4 days plot shows why
  88   Fri Nov 9 09:37:55 2007 steveUpdatePSLhead temp hiccup
Just an other PSL-126MOPA_HTEMP hiccup.
The water chiller is at 20.00C
  113   Fri Nov 16 18:46:49 2007 steveBureaucracyPSL MOPA was turned off & on
The "Mohana" boys scouts and their parents visited the 40m lab today.
The laser was turned off for their safety.
It is back on !
  114   Mon Nov 19 14:19:25 2007 steveOmnistructurePEMjackhammer
The construction personal successfully jackhemmered a fence around the "Drever's parking slot"
There is no parking space available close by
  115   Mon Nov 19 14:32:10 2007 steveBureaucracySAFETYgrad student safety training
John Miller and Alberto Stochino has received the 40m safety bible.
They still have to read the laser operation manual and sign off on it.
  122   Mon Nov 26 10:17:31 2007 steveOmnistructureSUSetmy sus damping restored
20 days plot is showing etmy loosing damping 4 times.
I zoomed in with each event. Three of them could of been triggered
by garbage loading just outside. However attachment 2 plot demonstrating that small earthquake or seismic event
did not tripped etmy damping.
The fourth event was preceded by a 4-5 hrs of continous rise of the rms motion at C1:SUS-ETMY_LLPD_VAR
  129   Wed Nov 28 08:47:29 2007 steveOmnistructureVACrga is out of order
The rga is not working since Nov 10
The controller is broken.
pd65-m-d23
  200   Wed Dec 19 11:31:01 2007 steveOmnistructurePEMaircond filter maintenance
Jeff is working on all air condiontion units of the 40m lab
This we do every six months.
  203   Wed Dec 19 16:40:12 2007 steveUpdateSAFETYlaser safety glasses measured
I measured the coarse transission at 1064nm of the 40m safety glasses today.

12 pieces of UVEX # LOTG-YAG/CO2 light green, all plastic construction, ADSORBANT

3 pieces of 6KG5, Scott colored filter glass type,

individual prescription glasses: alan, bob, ben, jay and steve

7 pieces of dual waveleght glasses

These glasses showed 0.00mW transmission out of 170mW Crysta Laser 1064
  220   Thu Jan 3 08:53:55 2008 steveUpdateSUSetmy vs etmx
Rana have corrected sus gain damping setting of ETMY 8 days ago

gain settings: pos, pit, yaw & sd
etmx: 4,2,2,& -16
etmy: 4,2,2,& 50
  221   Thu Jan 3 09:12:59 2008 steveUpdatePSLMZ servo
Here is MZ trend for one year and 40 days.
Now days it runs out of range on the low side.
This is the weakest link in the psl today.
  222   Thu Jan 3 09:55:11 2008 steveUpdateSUSetmy sus damping restored
ETMY watch dog was lost at midnight
  225   Fri Jan 4 08:42:03 2008 steveUpdateSUSetmy trips again
ETMY sus damping tripped at 6am this morning
It was reset. We should put an accelerometer to the south end to see
the garbage dumping effect.
  226   Mon Jan 7 09:01:39 2008 steveUpdateSUSBS sus damping restored
The BS sus damping was lost at 8am Sunday morning.
  232   Thu Jan 10 10:38:02 2008 steveUpdateSUSetmy damping restored
The IST building onstruction has really started yesterday and continuing today with big heavy ground breaking
machinary. The MC is holding lock and the suspentions are hanging on.

ETMY does not like this.

SUS-MC2_LLPD_VAR monitor is a good indicator of seismic activity on this 12 days plot
  233   Thu Jan 10 12:08:23 2008 steveUpdateSUSwhy did the BS move?
  235   Thu Jan 10 15:04:04 2008 steveUpdateSUSilluminator light effect on BS position
The bs chamber illuminator light was turned on this morning and left on.
Earlier on Rana noticed that the bs moved.
I follwed up to see what happened. I turned off oplev servo and tried to recenter on oplev pd
by adjusting pitch and yaw biases. It did not move. I looked at suspention and realized that the
illuminator was still on. I turned it off and to my amazement the the AP spot started flashing
  237   Mon Jan 14 14:41:09 2008 steveUpdateSUSetmy sus damping restored & mz relocked
Tree days trend of MZ HV drift is typical these days.
So as the etmy sus inability to hold damping for longer then 2-3 days.
  254   Wed Jan 23 09:27:55 2008 steveUpdate etmy sus damping restored
Seismis event trips etmy
  255   Wed Jan 23 11:41:06 2008 steveUpdatePEMsulfur smell in 40m
Led - acid car batteries were overcharged in the machine shop next door
and sulfuric acid smell is coming over to the ifo room.

Entry room 103 is specially bad.
  258   Thu Jan 24 11:52:56 2008 steveUpdatePEMthe mud is cleaned up & MOPA shutter is opened
Safety glasses are required again!

I have just opened the mopa shutter.
One janitor came to help with muddy floor.
Rack 1x1 toward ITMX chamber and the south wall of these area
were completely covered by mud. I wiped the floor of bottom of the rack
with towels. The cables were lifted and still should be wiped.

The bottom of LSC rack got less water, only on the west side.

We are ready to bring up the computers.

Thanks to ALL with the clean up, including Alan Rice
who was really helpful.

  264   Fri Jan 25 09:22:21 2008 steveUpdatePEMburned toast award goes to
DYM for collaborating with the enemy.

In order to minimize the number of ants visiting our lab we have to take out side the lab
all food left over and organic waste. If you are eating here do not expect someone else
to clean up after you.

Thanks for your cooperation.
  299   Wed Feb 6 09:17:31 2008 steveUpdatePEMIST building construction continoues
The bulldosers at work
  308   Mon Feb 11 14:24:19 2008 steveUpdatePEMmore earthquakes
ITMX and ITMY sus damping restored after Baja earthquake 5.1 mag at 10:29 this morning.

The ground preparation for The ITS building is almost finished.
Activity is winding down, however the Baja California_ Mexico earthquake zone
"Guadala Victoria" started acting up on Friday.

http://earthquake.usgs.gov/eqcenter/recenteqsus/Maps/special/California_Nevada_eqs.php
  322   Tue Feb 19 10:14:13 2008 steveUpdateVACrga logging needs help
The rga head and controller are running fine, but the data logging is not.
  325   Wed Feb 20 11:34:17 2008 steveUpdatePSLlaser head temp is up
MOPA head temp is running at 20.3C now
Nomally it is at 18.5C
  326   Wed Feb 20 16:24:37 2008 steveUpdatePSLthe laser is recovering slowly
Head temp is still 20.5C and decreasing slowly.
Power output 2.9W
NPRO power 22 mW is increasing as head is cooling down
  327   Thu Feb 21 09:56:26 2008 steveUpdatePSLthe laser is back
The laser and the psl recovered.
The water chiller temp is 19.98C and head temp 18.4C
Power 3.2W
PMC_T 3.1

c1iovme was restarted and the mc is locking now
  335   Fri Feb 22 14:45:06 2008 steveUpdateMOPAlaser power levels

At the beginning of this 1000 days plot shows the laser that was running at 22C head temp
and it was send to LLO

The laser from LHO PA#102 with NPRO#206 were installed at Nov. 29, 2005 @ 49,943 hrs
Now,almost 20,000 hrs later we have 50% less PSL-126MOPA_AMPMON power
  351   Mon Mar 3 09:25:33 2008 steveUpdateVACrga scan logging is working now

Quote:

Quote:
The rga head and controller are running fine, but the data logging is not.


It should run tonight at 1:25 AM. To get the cron job to work properly on op340m, I had to make wrapper sh script which defines the perl library before calling the actual RGAlogger script.




Thanks to Rob, it is working !

The baked, calibrated rga head
model# M206M, s/n #c128035 was reinstalled at the 40m ifo on Feb. 8, 2008
Faraday mode dwell time was increased from 2 to 16 sec
Rga head temp at top silver gaskit 45.2 C
The noise floor is at 1E-12 Torr

There is more detail in logbook VMI-14 p 107

pd65-m-d119
ELOG V3.1.3-