  8472   Mon Apr 22 17:43:09 2013 SteveUpdate40m Upgrading ETMY optical table & enclosure is ready for optics

Quote:

Quote:

 Enclosure is at the east end. It has it's bottom o-ring in place. It will be ready for optics tomorrow around 5pm

I have to shim out the enclosure, finish leveling the table and cut surgical tubing O-ring for the top.

 

 Glued surgical latex tubing with super glue into O-ring shape. The existing in place tubing K-100, OD 0.125" (actual size 0.140"), wall 0.031", ID 0.062".

I have just found out that tolerances on tubing OD are + - 0.026" by the manufacturer. I'm getting larger tubing for better fit.

The table is ready for optics.

Things left to do:

1, finalize o-ring size  2, finish cable feedthrough  3, finalize window connection 4, IR-Thermashield strips for bridge sides 5, replace bridge support post with solid one

 

 I'm working on to improve the quality of the enclosure.

The short comings are: more cable feedthroughs needed, latches to anchor top covers air tight and posts to support bending bridges.

Red triangles are compression latches at 10 places to hold the top air tight on surgical tubing

Green lines represent 4  posts of Al 1" OD to support the covers and maximize their eigenfrequencies.

Black crosses are 4 spring loaded push-bottom quick release pins to anchor the top covers to the bridges. This connection will  not be air tight.

(quarter-turn wing head fastener have the same problem) I'm thinking of some solution to minimize the leak.    

Violet _ steel plate (1" wide, 15" long, 0.125" thick)  between the two posts will anchor the quick release pins and make bridge rigid.

Blue rectangle is an other cable feedthrough exiting on the chamber side.

Planning to substitute window with soft - air tight ( Aluminized thin wall hose )  connection to vacuum view port where white circle is representing the Al adaptor ring.

Updated after Wednesday meeting 4-24-2013

 

  12226   Wed Jun 29 08:20:04 2016 SteveUpdateSUS First Contact shelf life
We'll follow LIGO policy:
Our policy is to use first contact within 1 year of purchase for use in the interferometers.  For inspection use I am comfortable with out-of-date use.

GaryLinn offered their indate First Contact for use.

 

  525   Fri Jun 6 16:47:04 2008 josephbConfigurationCameras GC650 scatter images of 1064nm light
Took images similar to the scattered light images from earlier, except with the CCD GC650 camera. The first three attached plots are an average of all 200 images, an average of the first 100 and then an average of the last 100 images.

They show no definite structure. The big red blob which changes with time may be a brighter reflection, although it virtually the same type of setup as the GC750 images.

To do this properly, I should grab a short focal length lens and simply blow up the beam to a size greater than the detector area and simply fix both cameras looking into.

The last set of plots are mean and standard deviation plots from a previous set of runs on 5/29/08 with the GC750 and GC650 running at the same time. The GC650 was receiving approximately 33% of the total power and GC750 was receiving 66% (in otherwords a factor of 2 more).
  8834   Fri Jul 12 09:00:10 2013 SteveUpdateGeneral Halogen illumination cables disconnected

Alex and Steve,

Old halogen chamber illuminator cabling disconnected and potenciometer board removed at 1Y1 in order to give room for pd calibration fibre set up.

  8835   Fri Jul 12 12:30:47 2013 manasaUpdateGeneral Halogen illumination cables disconnected

Quote:

Alex and Steve,

Old halogen chamber illuminator cabling disconnected and potenciometer board removed at 1Y1 in order to give room for pd calibration fibre set up.

 During the process, they had also removed the power cable to the ITMY camera. Steve and I fixed this...so the camera is back.

  2585   Wed Feb 10 16:27:47 2010 steveConfigurationGeneral IFO beam heights

IN VACUUM beam heights are ALL 5.5"  This is measured from the top of the optical table to the center of all TMs, mirrors and other optical components. This beam is ~36" above the floor.

PSL (inside of enclosure) main-output  beam: PMC, MZ, RC and ISS  are at 3" heights. IOO Angle & Position, MC-Trans and RFAM qpds are at 4"

ALL OTHER beam heights at atmosphere and  different ISCT (interferrometer sensing, control optical table)s are at 4"

 

 

  10428   Mon Aug 25 09:56:21 2014 SteveUpdateIMC IMC WFSs upgrade

 

 The Napa earth quake magnitude 6 did not have any effect on the suspensions.

 The Goy phase upgrade was done nicely. The IOO pointing did not change. Credit owned to Nick and Andres.

  IFO is locked right on.

  7466   Wed Oct 3 07:41:23 2012 SteveUpdateSUS MC2 sus damping restored

This is the third morning in a row that the MC2 was tripped.  Would you look at it Koji?

  7473   Wed Oct 3 23:24:05 2012 KojiUpdateSUS MC2 sus damping restored

Quote:

This is the third morning in a row that the MC2 was tripped.  Would you look at it Koji?

This may be cause by the impact of crazy WFS signal after the lock loss.
The auto locker is not fast enough to shut the WFS down before the mirrors are kicked.

Jenne and I discussed the issue and agreed that this can be solved by implementing
the same triggering algorithm as the LSC triggers.

Give us a bit more time to work on this.

  7496   Sun Oct 7 15:05:42 2012 DenUpdateSUS MC2 sus damping restored

Quote:

This is the third morning in a row that the MC2 was tripped.

 MC2 was tripped again. I think the answer is that watchdog's critical value was too small C1:SUS-MC2_PD_MAX_VAR = 10, so seismic could trip MC2. I've changed the value to 100.

mc2.png

  7460   Tue Oct 2 07:45:49 2012 SteveUpdateSUS MC2 & PRM sus damping restored
  5139   Mon Aug 8 13:24:00 2011 steveMetaphysicsTreasure Monday morning

Quote:

Just a reminder that a film crew will be here Monday morning, filming Christian Ott for some Discovery channel show.

They are slated to be here from 8am to 12:30pm or so.  They will take a couple of shots inside the lab, and the rest of the filming should be of Christian in the control room (which they will "clean up" and fit with "sexy lighting").  I will try to be here the whole time to oversee everything.

Konecrane Fred was early this morning. He diagnosed the ETMY crane horizontal drive gear box dead and left just before the film crew showed up.

New gear box should be here by the end of this week for installation.

The lab air quality is high ~20,000 counts of particles of 0.5 micron.  Keep an eye on this before you open the chamber.

  17908   Thu Oct 19 07:14:44 2023 JCUpdateIOO Morning Alignment

Basic Arm Alignment has been done

  • WFS seemed to have railed overnight.
    • The first action I took this mornign was to turn wfs off. Once I saw that I holding well, I ran the 60s WFS Relief script.
  • I align the single arms, C1:LSC-TRY_OUT_DQ transmission to ~1.0 and C1:LSC-TRX_OUT_DQ ~ .8

I'm going to continue working on the BHD platform now.

The first assemblies I am working on are D2200406, D2200154,and D2200392.

  7568   Wed Oct 17 11:15:05 2012 DenUpdateSUS PITCH damping needed

Quote:

 

 Pitch damping solution needed! It should be in the machine shop already.

I think we can put ø2mm × 10mm long magnetic material inside 4 holes with actuation magnets. Then magnetic field on the other side of the mirror will be close to one produced by actuation magnet. Magnetic cylinder center of inertia will be in the vertical plane where mirror's center of inertia is. So this should not change alignment significantly. Eddy current dumping will be applied to the end of the magnetic cylinder opposite to the magnet using aluminium disks, we have them in the clean room.

pitch_damping.png

  7571   Thu Oct 18 02:42:38 2012 DenUpdateSUS PITCH damping needed

Quote:

 

I think we can put ø2mm × 10mm long magnetic material inside 4 holes with actuation magnets. Then magnetic field on the other side of the mirror will be close to one produced by actuation magnet. Magnetic cylinder center of inertia will be in the vertical plane where mirror's center of inertia is. So this should not change alignment significantly. Eddy current dumping will be applied to the end of the magnetic cylinder opposite to the magnet using aluminium disks, we have them in the clean room.

 I've tested this approach. As we do not have required cylinders with high magnetic permittivity, I replaced them with magnets simular to actuator magnets ø2mm × 3mm long. Using them and aluminium disks from other TT I've made a "pitch dumping" construction.

DSC_4759.JPG

Pitch Q reduced but not that much as I could expect. I did a ringdown test. 

Plots:

yaw ringdown using original construction     |  yaw ringdown with added pitch damping

------------------------------------------------------------------------------------------------------------------------

pitch ringdown using original construction   |  pitch ringdown with added pitch damping

    yaw_nodamp_20.JPG   yaw_damped_13.JPG

pitch_nodamp_30.JPG     pitch_damped_10.JPG

 

 From this data I've estimated Q factor for yaw (135 vs 88) and pitch (192 vs 77) (original vs added pitch damping). Thess results diverges with the ones obtained by designes. They measured Q~40-50 for original construction. Pitch and yaw have 2 close resonances so this time domain method can not be very precise. I've measured the same with SR785.

yaw.pngpitch.png

In these comparison plots excitation was not the same as coils are not plugged in yet, but resonance Q factors can be compared.

 

  7583   Fri Oct 19 18:30:07 2012 DenUpdateSUS PITCH damping needed

CQuote:

 

 From this data I've estimated Q factor for yaw (135 vs 88) and pitch (192 vs 77) (original vs added pitch damping).

 I've made a more precise measurement of pitch damping using spectrum analyzer.

damp.png   pitch.png

Measurements confirm that damping using small actuation magnets reduces pitch Q by a factor of 4 and is not enough.

  7585   Sat Oct 20 01:23:56 2012 DenUpdateSUS PITCH damping needed

Quote:

Measurements confirm that damping using small actuation magnets reduces pitch Q by a factor of 4 and is not enough.

 I've tested the idea to use coils as eddy current dampers. I terminated them with a wire and measured Q factor during the ringdown test. Sadly, I did not see any significant damping and Q was ~150. We need stronger magnets if we want eddy current dumping down to Q~1.

P1010059.jpg      P1010060.jpg    P1010058.jpg

  7586   Sat Oct 20 20:37:55 2012 DenUpdateSUS PITCH damping needed

Quote:

 

 We need stronger magnets if we want eddy current dumping down to Q~1.

 I've inserted 10mm * 10mm magnets to the 4 corner holes on the front side of the mirror frame according to actuation magnets polarity. I realigned TT and measured Q factor for pitch and yaw, it was 5-10.

DSC_4778.JPG     pitch.JPG   yaw.JPG

 

I was able to do it for 1 TT only, because others have smaller (~0.1 mm) hole diameter and magnets can't go inside. I tried to warm holes up to 850 F but still was not able to insert a magnet.

DSC_4776.JPG

  7588   Mon Oct 22 11:26:12 2012 ranaUpdateSUS PITCH damping needed

 

Too bad - I thought it would at least give a little damping. Since we want the viscous-like energy loss to be ~49x larger, we need to have the field modulation in the damper (not dumper) increase by ~7.

  7593   Tue Oct 23 01:46:53 2012 DenUpdateSUS PITCH damping needed

Quote:

 

Too bad - I thought it would at least give a little damping. Since we want the viscous-like energy loss to be ~49x larger, we need to have the field modulation in the damper (not dumper) increase by ~7.

 I've made SolidWorks models of damping bracket and eddy current disk. They will me manufactured and used instead of old ones. New bracket will be mounted in exactly the same place where the old one was. Drawings might not be complete but all dimensions are in the models so we can fix drawing tomorrow before going to machine shop.

I think we can use ring magnets for passive damping. Then we won't have the vent problem. I've found some at K&J Magnetics, we can get them any time. Magnets are Ni-Cu-Ni (fine for vacuum?) Diameter is 3/8'' with advertised tolerence 0.004'', so they should fit the holes.

  7598   Tue Oct 23 17:12:30 2012 DenUpdateSUS PITCH damping needed

 Koji and Steve pointed out that previous design  of a damping bracket was a bit complicated to manufacture. So I made it simpler and also added a tap hole for original yaw damping. We'll give drawing to Mike in the machine shop tomorrow morning.

I've purchased K&J magnets for eddy current damping, they should be here in 2 days. 

  7602   Tue Oct 23 18:18:29 2012 JenneUpdateSUS PITCH damping needed

Quote:

Quote:

 

Too bad - I thought it would at least give a little damping. Since we want the viscous-like energy loss to be ~49x larger, we need to have the field modulation in the damper (not dumper) increase by ~7.

 I've made SolidWorks models of damping bracket and eddy current disk. They will me manufactured and used instead of old ones. New bracket will be mounted in exactly the same place where the old one was. Drawings might not be complete but all dimensions are in the models so we can fix drawing tomorrow before going to machine shop.

I think we can use ring magnets for passive damping. Then we won't have the vent problem. I've found some at K&J Magnetics, we can get them any time. Magnets are Ni-Cu-Ni (fine for vacuum?) Diameter is 3/8'' with advertised tolerence 0.004'', so they should fit the holes.

 Den mentioned that the disks will have threaded holes, and that he has made a note to that effect on the paper copy of the drawing that he will bring to Mike at the shop.  Also, all threaded holes in the new plate are marked on the paper copy.

  7604   Wed Oct 24 01:02:10 2012 KojiUpdateSUS PITCH damping needed

Wow... This is even more complicated than the original "Y" design...

  7640   Mon Oct 29 18:14:55 2012 DenUpdateSUS PITCH damping needed

 

 We've received all parts that we need for eddy current damping. I've made an estimate of Q with dirty tip-tilt. It looks fine (Q~1)

We need to check ring magnets for vacuum compatibility. Bob start baking on Friday.

DSC_4787.JPG     DSC_4791.JPG

  12844   Wed Feb 22 08:54:17 2017 steveUpdatePEM PSL enclousure particle count

ETMX sus damping recovered.  PSL enclousure is dusty at 20V rotation speed. Rainy days as outside condition.

Quote:

The MET#1 particle counter was moved from CES wall at ITMX to PSL enclousure south west corner at 11am.

The HEPA filter speed at the Variac was turned down to 20V from 40

This counter pumps air for 1 minute in every 20 minutes. Soft foam in bags used to minimize this shaking as it is clamped.

 

 

  13549   Tue Jan 16 11:05:51 2018 gautamUpdatePSL PSL shelf - AOM power connection interrupted

While moving the RefCav to facilitate the PSL shelf install, I bumped the power cable to the AOM driver. I will re-solder it in the evening after the shelf installation. PMC and IMC have been re-locked. Judging by the PMC refl camera image, I may also have bumped the camera as the REFL spot is now a little shifted. The fact that the IMC re-locked readily suggests that the input pointing can't have changed significantly because of the RefCav move.

 

  13551   Tue Jan 16 21:46:02 2018 gautamUpdatePSL PSL shelf - AOM power connection interrupted

Johannes informed me that he touched up the PMC REFL camera alignment. I am holding off on re-soldering the AOM driver power as I could use another pair of hands getting the power cable disentangled and removed from the 1X2 rack rails, so that I can bring it out to the lab and solder it back on.

Is anyone aware of a more robust connector solution for the type of power pins we have on the AOM driver?

Quote:

While moving the RefCav to facilitate the PSL shelf install, I bumped the power cable to the AOM driver. I will re-solder it in the evening after the shelf installation. PMC and IMC have been re-locked. Judging by the PMC refl camera image, I may also have bumped the camera as the REFL spot is now a little shifted. The fact that the IMC re-locked readily suggests that the input pointing can't have changed significantly because of the RefCav move.

 

 

  13538   Fri Jan 12 10:26:24 2018 SteveUpdatePSL PSL shelf work schedule

Measurements for good fit were made. The new shelf will be installed on next Tuesday at 2pm

The reference cavity ion pump is in the way  so the cavity will be moved 5" westward. The shelf height space will be 10"  Under shelf working height 18" to optical table.

Quote:

I have just received the scheduling of the PSL self work for tomorrow. Gautam and I agreed that if it is needed I will shut the laser off and cover the hole table with plastic.

 

  7566   Wed Oct 17 08:30:49 2012 SteveUpdateSUS PTICH damping needed

Quote:

Quote:

  [Koji, Steve, Den]

TT alignment is fine, yaw damping is satisfactory, pitch damping is slow. We might want to add magnets to the mirror and attach blades to the frame for pitch edge current damping.

We are moving towards electronics testing.

Atm1,  TT 1.5" high adaptor base will be back from the shop in 10 days.

Atm2,  There is no PITCH damping, YAW edie current damping works well at 0.5 mm gap

Atm3,  Adjustable Al -disc that contains a small magnet is purely designed.

We have to come up with a solution to have damping in PITCH

 Pitch damping solution needed! It should be in the machine shop already.

  352   Mon Mar 3 13:58:10 2008 steveUpdateComputers RFM Network are down
The CODAQ_RFNETWORK are down, except C1SUSVME & AWG
  353   Mon Mar 3 19:34:40 2008 ranaUpdateComputers RFM Network are down

Quote:
The CODAQ_RFNETWORK are down, except C1SUSVME & AWG

All of the FE machines were found to be down this afternoon. I called Alex and he suggested several
things which didn't work (restart EPICS tasks, power cycle RFM switch, etc.).

Then he suggested that I go around and power cycle every crate!!! And that sometimes the order of this
matters!!! I think he was just recording this conversation so that he could have a laugh by playing it on Youtube.

However, power cycling all of the FE crates seemed to work. Alex's theory is that 'something goes bad in the
RFM cards sometimes'.

Its all green now.
  8686   Thu Jun 6 15:46:10 2013 Max HortonSummaryGeneral Smart UPS 2200 Batteries Replaced

Replaced the batteries successfully in the control room.  We just had to switch the clips from the old batteries to the new one, which we didn't know was possible until now.

  12043   Wed Mar 23 11:55:47 2016 SteveUpdateGeneral Smart UPS 2200 Battery Replaced

Batteries replaced in control room UPS after 3 years from replaceUPSbattery.com

 

 

  12952   Thu Apr 27 16:41:13 2017 Eric GustafsonUpdateLSC Status of the 40 m PD Frequency Response Fiber System
There two reports in the DCC describing the state of the system as of October 2014 including: (1)  Alex Cole’s “T1300618 Automated photodiode Frequency Response Measurement System” and a Wiki  created by Alex Cole where there are some instructions on the Master Script at https://wiki.ligo.caltech.edu/ajw?AlexanderCole    

And (2)  P140021 “Final Report: Automated Photodiode Frequency Response Measurement System for 40m Lab” by Nichin Sreekantaswamy and also as part of Nichin’s report by there is an archive of data at   https://wiki-40m.ligo.caltech.edu/Electronics/PDFR%20system   

I made a visual inspection of the system and saw that the following fibers collimators are still mounted in alignment mounts and the fiber is attached and pointed at a photodetector but possibly not aligned. 

ASP Table

Photodetector Label                             Fiber Label

REFL11                                              REFL55 Fiber on mount        

REFL33                                              REFL33 Fiber on mount

REFL55                                              REFL11 Fiber on mount

REFL165                                            No Fiber

AS55                                                   AS55 Fiber on mount

MCREFPD                                         MCREFPD Fiber on mount

No PD                                                 Loose unlabeled Fiber No mount

 

ITMX Optics Table

Photodetector Label                             Fiber Label

POX11                                                POX11 on mount

Unlabeled PD                                      POP22/POP110 on mount

NO PD                                                 POP55 loose fiber No mount

 

The RF switch seems to be hooked up and there is a fiber running from the Diode Laser module to the fiber splitter module. So REFL 11 and REFL545 seem to be illuminated by the wrong fiber. I’ll try and run the software on Monday and check to see if I need to move the fibers or just relabel them.

  8432   Tue Apr 9 21:27:48 2013 JenneUpdate40m Upgrading TRY temporarily in place

Quote:

I've used a Y1 mirror to steer the Y transmission to an R98% BS. The reflected beam falls on PDA520 and the transmitted beam is steered to the camera. The earlier normalization of TRY is no more valid as the power distribution at the PD has changed.

 To take this into account, last night, I reduced the TRY gain by a factor of 2.  This is not exactly correct - when the layout is finalized we need to figure out what the pickoff situation used to be (we think, based on the Xend, that it could have been 0.5*0.9), and do the correct normalization.

  12788   Thu Feb 2 12:17:48 2017 SteveUpdateGeneral USB microscope

This AmScope microscope would have 3.5x-180x magnification, calibratable measurement function, 5MP picture and good working distance to work on printed circuit boards.

 

  12813   Thu Feb 9 08:03:08 2017 SteveUpdateGeneral USB microscope ordered

http://www.amscope.com/3-5x-180x-boom-stand-trinocular-zoom-stereo-microscope-with-144-led-ring-light-and-10mp-camera.html will be ordered today.

The actual unit we are getting has lockable zoom for better repeatability after calibration: SM-3NTPZZ-144

Quote: CWQ6-020817

 

 

  12848   Thu Feb 23 14:50:26 2017 SteveUpdateGeneral USB microscope returned

The microscope shipped back to the vendor for credit yesterday.

Quote:

http://www.amscope.com/3-5x-180x-boom-stand-trinocular-zoom-stereo-microscope-with-144-led-ring-light-and-10mp-camera.html will be ordered today.

The actual unit we are getting has lockable zoom for better repeatability after calibration: SM-3NTPZZ-144

Quote: CWQ6-020817

 

 

 

  12570   Wed Oct 19 14:43:15 2016 SteveUpdateGeneral Viewports & coating of 2001

Tilted viewports installed in horizontal position. Atm2

  14106   Thu Jul 26 15:11:18 2018 SteveUpdateGeneral Viewports & coating of 2001

New optical quality BK-7 windows in 2001 [4 substrates ] AR coated R<0.75 % for 630-1064nm " Azury BLue" broadband : TRX, TRY, ITMY-Oplev &  ITMX-Oplev viewports.

The BS-Oplev and PRM-Oplev 10" CF with 5.38" diameter view was coated the same way. The window here is Corning 7056 Borosilicate

5 more BK-7 substrates were coated R <0.1% of 1064 nm "Golden Orange" Their location: IMC-IN, IFO-REF and OMC   The next vent we have to confirm optical quality window locations.

All other conflat flange viewports are 7056 Kovar sealed .

Technical notes of 2001 40m upgrade can be seen at LIGO-T010115- 00- R  ....page 14

  12572   Wed Oct 19 17:02:34 2016 ranaUpdateGeneral Viewports & coating of 2016

These old specs are not so bad. But we now want to get replacements for the TRX and TRY and PSL viewports that are R <0.1% at 532 and 1064 nm.

I don't know of any issues with keeping BK-7 as the substrate.

  10023   Wed Jun 11 08:57:49 2014 SteveUpdatePEM air cond maintenance tomorrow morning

Quote:

  The 4th week of no wet mopping of the floor  and no wet wiping  the vacuum envelope.

 

We cleaned the intakes of the south arm IFO air condition. The bottom duct have quite a bit accumulation Atm1

See wet wiped contrast on Atm2

We found 3 holes around pipes ( coming from CES ) on the east wall that has to be sealed.

After closer examination of these holes, they are sealed off well.

 Air condition maintenance will be finished by 11:30am tomorrow.

 

Yesterday, Kelvin mopped with chemicals the whole floor area of the  lab. This was triggered by some  visiting  ants at our PSL last week. It was 6 months since we had the last fully wet mopped IFO floor.

The cleaning- mopping water became   very dirty at the end.

  5307   Fri Aug 26 10:45:17 2011 steveUpdateGeneral crane at ETMY is fixed

Quote:

The horizontal trolley drive stopped working  at the east end this morning. It is working intermittently. In the worst case we can take the door off with the manual -Genie- lift.

I'm working with Konecrane to solve  the wormgear drive problem.

 New gear box installed and tested by Fred KoneCranes.

  12868   Mon Mar 6 09:12:59 2017 SteveUpdatesafety crane inspection 2017

All 3 cranes inspected by professional Fred Goodbar of Konecranes and load tested with 450 lbs at max reach on Friday, March 3, 2017

 

 

  13630   Mon Feb 12 14:56:00 2018 SteveUpdatesafety crane inspection 2018

Our 3 cranes passed  professional  inspection. Fred Goodbar of Konacrane with 450 lbs load at full extension.

Certificates will be posted in 40m wiki as they arrive.

 

  12020   Fri Mar 4 11:07:10 2016 SteveUpdatesafety crane inspection completes 2016 safey audit

All 3 cranes inspected by professional and load tested with 450 lbs at max reach.

 

  12656   Fri Dec 2 10:57:08 2016 steveUpdatePEM door seal replaced

The south end door leaky weather seals replaced.

The aim is here to get some overpressure inside / outside so the lab partical count would not depend on outside condition.

 

  3404   Thu Aug 12 10:56:31 2010 steveUpdateEnvironment dusty floor work in progres

I powered down 1Y1 and 1Y2 instrument racks of MC and RF-systems this morning.

Both racks, SP table, clean tools-parts boxes and MC2 area are also covered with plastics.

Bertin Bros. Construction, Inc is working on the floor preparation. Their head man is Robby Bertin, cell 626. 831.2264

Today's job:

remove tiles, clean off tile glue, drill for ribbed  steel bars, rough clean- surface for good bounding. paint floor with bounding agent

  11975   Thu Feb 4 10:14:52 2016 SteveUpdateSUS earthquake 3.5M

 

 

  11932   Thu Jan 14 13:44:05 2016 SteveUpdateSUS earthquake 3.6M

No obvious sign of damage.

 

ELOG V3.1.3-