40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 2 of 341  Not logged in ELOG logo
ID Dateup Author Type Category Subject Text Attachments
  53   Thu Nov 1 19:55:03 2007 Andrey RodionovBureaucracyPhotosAndrey's photo   DSC_0055.JPG 
  54   Thu Nov 1 19:55:59 2007 Andrey RodionovBureaucracyPhotosAndrey, Tobin, Robert - photo   DSC_0092.JPG 
  55   Thu Nov 1 19:58:07 2007 Andrey RodionovBureaucracyPhotosSteve and Tobin's picture   DSC_0023.JPG 
  56   Thu Nov 1 20:03:00 2007 Andrey RodionovSummaryPhotosProcedure "Drop and Drag" in pictures   DSC_0072.JPGDSC_0083.JPGDSC_0099.JPGDSC_0100.JPG 
  57   Fri Nov 2 08:59:30 2007 steveBureaucracySAFETYthe laser is ONThe psl laser is back on !   
  58   Fri Nov 2 12:18:47 2007 waldmanSummaryOMCLocked OMC with DCPD[Rich, Sam]

We locked the OMC and look at the signal
 071102_OMC_LockedDCPD.gif071102_OMC_LockedDCPD.pdf 
  59   Sat Nov 3 16:20:43 2007 waldmanSummaryOMCA good day's work
I followed up yesterday's test of the PZT
with a whole mess of characterizations of
 6x 
  60   Sun Nov 4 23:22:50 2007 waldmanUpdateOMCOMC PZT and driver response functionsI wrote a big long elog and then my browser
hung up,  so you get a less detailed entry.
 I used Pinkesh's calibration of the PZT
 7x 
  61   Sun Nov 4 23:55:24 2007 ranaUpdateIOOFriday's In-Vac workOn Friday morning when closing up we noticed
that we could not get the MC to flash any
modes.
  
  62   Mon Nov 5 07:29:35 2007 ranaUpdateIOOFriday's In-Vac workLiyuan recently did some of his pencil beam
scatterometer measurements measuring not
the
  
  63   Mon Nov 5 14:44:39 2007 waldmanUpdateOMCPZT response functions and De-whiteningThe PZT has two control paths:  a DC coupled
path with gain of 20,  range of 0 to 300
V,  and a pair of 1:10 whitening filters,
 07110_DithertoVmonAC_sweep2-0.png071105_LSCtoVmonDC_sweep4-0.png07110_DithertoVmonAC_sweep2.pdf071105_LSCtoVmonDC_sweep4.pdf 
  64   Mon Nov 5 22:24:38 2007 Andrey, SteveOmnistructureVACPumping down goes smoothly
We (Steve and Andrey) started pumping down
at 3.25PM today. At 9 PM we turned off the
  
  65   Tue Nov 6 09:14:37 2007 steveSummaryVACpump down 658 hr plot,
precondition: 5 days at atm,
vent objective: drag wiping mc1, mc2 & mc3
 pd65.jpg 
  66   Tue Nov 6 09:45:22 2007 steveSummarySUSvent sus trend The mc optics dragwippings were done by locking
optics by eq stops and rotating-moving
cages so access were good. This technic worked
 ventsustrend.jpg 
  67   Tue Nov 6 10:42:01 2007 robConfigurationIOOmode cleaner lockedIncreased the power exiting the PSL by turning
the half-wave plate after the MOPA, opened
the PSL shutter, and aligned the mode cleaner
  
  68   Tue Nov 6 14:51:03 2007 tobin, robUpdateIOOMode cleaner lengthUsing the Ward-Fricke variant* of the [url=http://ilog.ligo-wa.caltech.edu/ilog/pub/ilog.cgi?group=detector&date_to_view=02/14/2004&anchor_to_scroll_to=2004:02:15:00:06:50-daniel]Sigg-Frolov
method[/url], we found the length of the
mode cleaner to be [b]27.0934020183 meters[/b],
  
  69   Tue Nov 6 15:36:03 2007 robUpdateLSCXARM lockedEasily, after resetting the PSL Uniblitz shutters.
 There's no entry from David or Andrey about
the recovery from last week's power outage,
  
  70   Tue Nov 6 15:37:34 2007 robConfigurationSUSrampdown script/cvs/cds/caltech/scripts/SUS/rampdown.pl is
now in the crontab for op340m, running every
half-hour at 15&45.  It checks the suspension
  
  71   Tue Nov 6 16:48:54 2007 tobinConfigurationComputersscopes on the netI configured our two 100 MHz Tektronix 3014B
scopes with IP addresses: 131.215.113.24
([url=http://scope0/]scope0[/url]) and 113.215.113.25
  
  72   Tue Nov 6 18:18:15 2007 tobinConfigurationComputersI broke (and fixed) conloggerIt turns out that not only restart_conlogger,
but also conlogger itself checks to see that
it is running on the right machine.  I had
  
  73   Tue Nov 6 23:45:38 2007 tobinConfigurationComputerstektronix scripts!I cooked up a little script to fetch the data
from the networked Tektronix scope.  Example
usage:
  
  74   Wed Nov 7 00:51:33 2007 andrey, rob, tobinConfigurationIOOMC ringdownsWe completed several ringdown measurements
this afternoon; Andrey is currently processing
the data.
  
  75   Wed Nov 7 02:14:08 2007 AndreyBureaucracyIOOMore information about MC2 ringdownAs Tobin wrote two hours ago, we (Andrey,
Tobin, Robert) made a series of ringdown
measurements for MC2
 9x 
  76   Wed Nov 7 09:38:01 2007 steveUpdateVACrga scanpd65-m-d2 at cc1 6e-6 torr  pd65d2.jpg 
  77   Wed Nov 7 10:55:21 2007 ajwConfigurationComputersbackup script restartedFollowing the reboot of computers on 10/31/07,
the backup script required restart (which
unfortunately "can't" be automated because
  
  78   Wed Nov 7 13:54:44 2007 robConfigurationIOOMode Cleaner transfer functionI performed the same procedure described [url=http://dziban.ligo.caltech.edu:40/40m/40]
here[/url], and re-measured the transfer
function of the mode cleaner to see the effect
 mctf.pdf 
  79   Wed Nov 7 14:01:31 2007 waldmanOmnistructureOMCFrequency and Intensity noiseOne of the biggest problems I had using the
PZT to lock was excessive noise.  I did a
little noise hunting and found that the problem
  
  80   Wed Nov 7 14:05:59 2007 tobinConfigurationIOOMC ringdownModeling the mode cleaner as a simple cavity
with all losses lumped together, we expect
the cavity power to be
  
  81   Wed Nov 7 16:07:03 2007 steveUpdatePSLPSL & IOO trend1.5 days of happy psl-ioo with litle bumps
in C1:PSL-126MOPA_HTEMP
 psl1.5dtrend.jpg 
  82   Thu Nov 8 00:55:44 2007 pkpUpdateOMCSuspension tests[Sam , Pinkesh]

We tried to measure the transfer functions
 9x 
  83   Thu Nov 8 11:40:21 2007 steveUpdatePEMparticle counts are upI turned up the psl HEPA filter to 100%
This 4 days plot shows why
 pslhepaon.jpg 
  84   Thu Nov 8 15:57:53 2007 tobinConfigurationPSLshelf removedI removed the sheet metal shelf from the PSL
enclosure, for easier access to the ISS.
  
  85   Thu Nov 8 18:44:01 2007 tobinConfigurationPSLISSTobin, Rob

With the Sense PD blocked, I adjusted the
  
  86   Fri Nov 9 00:01:24 2007 waldmanOmnistructureOMCOMC mechanical resonances (Tap tap tappy tap)[Pinkesh, Aidan, Sam]

We did a tap-tap-tappy-tap test of the OMC
 7x 
  87   Fri Nov 9 00:23:12 2007 pkpUpdateOMCX and Z resonancesI got a couple of resonance plots going for
now. I am still having trouble getting the
Y measurement going for some reason. I will
 Xtransferfunc.pdfZtransferfunc.pdf 
  88   Fri Nov 9 09:37:55 2007 steveUpdatePSLhead temp hiccupJust an other PSL-126MOPA_HTEMP hiccup.
The water chiller is at 20.00C
 headtempup.jpg 
  89   Fri Nov 9 17:33:33 2007 robConfigurationPSLISS
The 3.7 MHz is actually on the light.  It's
the beat between the 29.5 MHz sidebands and
  
  90   Fri Nov 9 21:36:14 2007 robConfigurationPSLFSSrob, rana

We looked at the FSS a bit today.  The most
  
  91   Sun Nov 11 21:05:55 2007 ranaHowToSUSMC Touching or notI wrote a script: [I]SUS/freeswing-mc.csh[/I],
which gives the MC mirrors the appropriate
kicks
  
  92   Sun Nov 11 21:21:04 2007 ranaHowToComputersNew DVTo use the new ligoDV (previously GEO DV)
to look at 40m data, open up a matlab, set
up for mDV as usual,
 Screenshot-1.png 
  93   Mon Nov 12 10:53:58 2007 pkpUpdateOMCVertical Transfer functions[Norna Sam Pinkesh]

These plots were created by injected white
 VerticalTrans.pdf 
  94   Mon Nov 12 14:09:19 2007 robDAQGeneraltpman dead on fb40mThe testpoint manager was dead on fb40m. 
I know I re-started it sometime after the
power outage, so something must have killed
  
  95   Mon Nov 12 15:05:49 2007 robConfigurationPSLFSS
Spent a bit of time fiddling with the FSS
again today.  In a not-particularly-systematic
  
  96   Mon Nov 12 15:18:34 2007 robUpdatePSLISS
After John soldered a 3.7 MHz notch filter
onto the ISS board, I took a quick TF and
 ISS.png 
  97   Mon Nov 12 23:44:19 2007 JohnUpdatePSLISS[quote]
After John soldered a 3.7 MHz notch filter
onto the ISS board, I took a quick TF and
 ISSfilter.jpg 
  98   Tue Nov 13 14:33:40 2007 JohnUpdatePSLISS filterThe transfer function from 'In Loop Error
Point Monitor' to TP3 the filter out test
point on the ISS board.
 PB130035.JPGDSC_0165.JPG 
  99   Wed Nov 14 07:48:38 2007 nornaOmnistructureOMCOMC Cable dressing[Snipped from an email]

1) Last Friday Pinkesh and I set the OMC
 OMC-11-13-07_011.jpgVerticalTrans.pdf 
  100   Wed Nov 14 12:33:35 2007 tobinAoGEnvironmentconstructionThe construction crews are running a jack-hammer
right outside of the control room.
 DSC_0172.JPG 
  101   Wed Nov 14 12:47:19 2007 tobinUpdatePSLISSJohn, Tobin

With John's notch filter installed and the
  
  102   Wed Nov 14 16:54:54 2007 pkpUpdateOMCMuch better looking vertical transfer functions[Norna Pinkesh]

So after Chub did his wonderful mini-surgery
 VerticalTransPreampwireremovedNov142007.pdfVerticalTranswiresclampedNov142007.pdf 
ELOG V3.1.3-