  13197   Fri Aug 11 18:53:35 2017 gautamUpdateCDSSlow EPICS channels -> Frames re-enabled


  
  13198   Fri Aug 11 19:34:49 2017 JamieUpdateCDSCDS final bits status updateSo it appears we now have full frames and
second, minute, and minute_raw trends.

We are still not able to raise
  
  13199   Sat Aug 12 14:09:36 2017 gautamUpdateSUSGlitches stay on MC1Even
in the switched state, the glitches stayed
on MC1. 

The coil
 MC1_glitch.pngMC_12hr_trend.pngMC1_glitches_intermittent.png 
  13200   Sat Aug 12 22:35:22 2017 ranaUpdateSUSGlitches stay on MC1To add to Gautam's entry: we swapped
the cables at the coil driver side (these
are the ones that go from coil driver to
  
  13201   Sun Aug 13 01:35:09 2017 KojiUpdateSUSGlitches stay on MC1We used to have similar suspension excursion
at ETMX. This was the motivation to replace
the stand-offs from Al ones to ruby ones.
  
  13202   Mon Aug 14 09:49:18 2017 KiraUpdatePEMtemperature sensorDecided to try adding in an OP amp just
to see if it would work. Added LT1012
and a 100k resistor to the circuit (I originally
 IMG_20170814_092452.jpgIMG_20170814_092513.jpg 
  13203   Mon Aug 14 12:52:33 2017 KiraUpdatePEMtemperature sensorI didn't realize that the LT1012 needed
an additional input to function. I added
in +15V and -15V to pins 7 and 4, respectively and
 IMG_20170814_121131.jpgIMG_20170814_121139.jpgIMG_20170814_121758~2.jpg 
  13204   Mon Aug 14 16:24:09 2017 gautamUpdateALSFiber ALSToday, I borrowed the fiber microscope
from Johannes and took a look at the fibers
coupled to the PDs. The PD labelled "
 IMG_7471.JPGIMG_7472.JPGIMG_7473.JPGIMG_7474.JPG 
  13205   Mon Aug 14 19:41:46 2017 JamieUpdateCDSfront-end/DAQ network down for kernel upgrade, and timing errorsI'm upgrading the linux kernel for
all the front ends to one that is supposedly
more stable and won't freeze when we
  
  13206   Mon Aug 14 20:01:38 2017 gautamUpdateSUSGlitches stay on MC1I don't think we can say for sure. I
was just talking to EricQ about this, he
said the glitches were often seen when changing
  
  13207   Mon Aug 14 20:12:09 2017 Jamie, GautumUpdateCDSWeird problem with GPS receiverToday we saw a weird issue with the GPS
receiver (EndRun Technologies Tempus LX). 
GPS timing on fb1 (which is handled via IRIG-B
  
  13208   Tue Aug 15 08:22:16 2017 SteveUpdateGeneralprojector light bulb replacedBL-FS300C-PH-LE was
replaced after 2,904 hrs  It did
not explode this time.  The 4 monts
  
  13210   Tue Aug 15 13:32:38 2017 KiraUpdatePEMtemperature sensorTested to make sure that even when only
the AD586 was heated that there was no change
in the reading. I did so by placing the AD586
  
  13211   Tue Aug 15 16:32:42 2017 Jamie, GautumUpdateCDSGPS receiver apparently set to correct mode as "UTC"


  
  13212   Wed Aug 16 14:54:13 2017 gautamUpdateCDSPSL monitoring Acromag EPICS server restarted[johannes,
gautam, jamie]


Made a directory 
  
  13213   Wed Aug 16 14:57:01 2017 SteveUpdatePSLref Cavity heating blanket power supplyThe last entry I found relating to ref
cavity was
2011 Aug 19
 refCavPSheat.jpgrefCavTempSens.jpgrefTempSen_1X1.jpg 
  13214   Wed Aug 16 16:05:53 2017 KiraUpdatePEMtemp sensor PCBTried taking the circuit from the breadboard
to the PCB. I attached all the components
to adapters that would allow them to be connected
 IMG_20170816_154514.jpgIMG_20170816_154541.jpg 
  13215   Wed Aug 16 17:05:53 2017 JamieUpdateCDSfront-end/DAQ network down for kernel upgrade, and timing errorsThe CDS system has now been up moved to
a supposedly more stable real-time-patched
linux kernel (3.2.88-csp) and RCG r4447 (roughly
 2017-08-16-163725_1366x495_scrot.png 
  13216   Wed Aug 16 17:14:02 2017 KojiUpdateCDSfront-end/DAQ network down for kernel upgrade, and timing errorsWhat's the current backup situation?   
  13217   Wed Aug 16 18:01:28 2017 JamieUpdateCDSfront-end/DAQ network down for kernel upgrade, and timing errors


  
  13218   Wed Aug 16 18:06:01 2017 KojiUpdateCDSfront-end/DAQ network down for kernel upgrade, and timing errorsWe also need to copy chiara's root.
What is the best way to get the full image
of the root FS?
  
  13219   Wed Aug 16 18:50:58 2017 JamieUpdateCDSfront-end/DAQ network down for kernel upgrade, and timing errors


 2017-08-16-184910_1394x488_scrot.png 
  13220   Wed Aug 16 19:50:17 2017 gautamUpdateSUSMC1 <--> MC3 switched backNow that all the CDS overview lights are
green, I decided to switch back the coil
driver outputs to their original state so
  
  13221   Wed Aug 16 20:01:03 2017 gautamUpdateGeneralSUS model ASC input weirdnessI'm not sure if this has something
to do with the model restarts / new RCG,
but while I was re-enabling the MC watchdogs,
 ITMX_ASC.png 
  13222   Wed Aug 16 20:24:23 2017 gautamUpdateALSFiber ALSToday, with Johannes' help, I cleaned
the fiber tips of the photodiodes. The effect
of the cleaning was dramatic - see Attachments
 IMG_7476.JPGIMG_7477.JPGIMG_7478.JPGIMG_7479.JPG 
  13223   Thu Aug 17 08:42:27 2017 SteveUpdatePSLPSL HEPAThe PSL HEPA was running noisy at 100V  
The bearing is wearing out. I turned it down
to 30V It is quiet there.
  
  13224   Thu Aug 17 10:41:58 2017 KiraUpdatePEMtemp sensor PCBGot it to work. One of the connections
was faulty. I decided to check the temperature
measured against a thermometer. The sensor
 IMG_20170817_095917.jpg 
  13225   Thu Aug 17 11:17:49 2017 gautamUpdateSUSMC1 <--> MC3 switched backSeems like this modification didn't
really work. There were several large MC1
glitches, and one of them misaligned MC1
 MC1_misaligned.pngMC1_glitch.png 
  13226   Thu Aug 17 17:33:01 2017 gautamUpdateSUSMC1 <--> MC3 switched backthat's why the Autolocker clears the
outputs; we don't want to be holding
the offsets from the last ms of lock when
  
  13227   Thu Aug 17 22:54:49 2017 ericqUpdateComputersTrying to access JetStor RAID filesThe JetStor RAID unit that we had been
using for frame writing before the fb meltdown
has some archived frames from DRFPMI locks
  
  13228   Fri Aug 18 21:58:35 2017 gautamUpdateGeneralSUS model ASC input weirdnessI spent some time today trying to debug
this issue.

Jamie
 ITMXP.pngASC_model_outmatrix.pngASC_medm.pngASC_outMat.png 
  13229   Fri Aug 18 23:59:53 2017 gautamUpdateALSX Arm ALS lock[ericq, gautam]


I was just getting the
IFO aligned, and single arm lock going, when
  
  13230   Sat Aug 19 01:35:08 2017 ericqUpdateALSX Arm ALS lockMy motivation tonight was to get an up-to-date
spectrum of a calibrated measurement
of the out-of-loop displacement of an arm
 ALS_comparison.pdf 
  13231   Mon Aug 21 09:08:54 2017 SteveUpdateSUStiny glitchesThey are synchronised tiny glitches. They
are not mechanical.
 
 glitches.png 
  13232   Mon Aug 21 13:07:08 2017 KiraUpdatePEMtemp sensor PCBOn Friday, I cleaned up the circuit so
that there are only three connections needed
(+15V, -15V, GND) and a BNC connector for
 IMG_20170821_124121.jpgIMG_20170821_124429~2.jpgIMG_20170821_124108.jpg 
  13233   Mon Aug 21 14:53:32 2017 gautamUpdateVACRGA reset[gautam, steve]

In the aftermath of the accidental
vent, it looks like the RGA was shutdown.
  
  13234   Mon Aug 21 16:35:48 2017 gautamUpdateVACUPS checkup[steve, gautam]

At Rolf/Rich Abbott's request,
we performed a check of the UPS today.
  
  13237   Mon Aug 21 23:38:55 2017 gautamUpdateALSALS out-of-loop noiseI worked a little bit on the Y arm ALS
today. 


Started by locking the
 ALS_21082017.pdf 
  13238   Tue Aug 22 02:19:11 2017 gautamUpdateALSALS OLTFsAttachment #1 shows the
results of my measurements tonight (SR785
data in Attachment #2).
 ALS_OLTFs.pdfALS_OLTF_Aug2017.zip 
  13239   Tue Aug 22 15:17:19 2017 ericqUpdateComputersOld frames accessible againIt turns out the problem was just a bent
pin on the SCSI cable, likely from having
to stretch things a bit to reach optimus
  
  13240   Tue Aug 22 15:40:06 2017 gautamUpdateComputersOld frames accessible again [jamie, gautam]

I had some trouble getting the
daqd processes up and running again using
  
  13242   Tue Aug 22 17:11:15 2017 gautamUpdateComputersc1iscex model restarts[jamie, gautam]

We tried to implement the fix
that Rolf suggested in order to solve (perhaps
  
  13243   Tue Aug 22 18:36:46 2017 gautamUpdateComputersAll FE models compiled against RCG3.4After getting the go ahead from Jamie,
I recompiled all the FE models against the
same version of RCG that we tested on the
 CDS_Aug22.png 
  13244   Tue Aug 22 23:27:14 2017 ranaUpdateALSALS OLTFsDidn't someone look at what the OLG
req. should be for these servos at some point?
I wonder if we can make a parallel digital
  
  13245   Wed Aug 23 10:11:46 2017 SteveUpdateVACvacuum valve specificationsThe V1
gate valve specs installed at 40m wiki
page. VAT model number 10846-UE44-0007 Our
  
  13246   Wed Aug 23 17:22:36 2017 gautamUpdateALSFiber ALS - reinstalledI completed the revamp of the box, and
re-installed the box on the PSL table today.
I think it would be ideal to install this
  
  13248   Thu Aug 24 00:39:47 2017 gautamUpdateLSCDRMI locking attemptSince the single arm locking and dither
alignment seemed to work alright after the
CDS overhaul, I decided to try some recycling
  
  13249   Thu Aug 24 17:36:11 2017 gautamUpdateCDSFSS Slow Python maintenanceA couple of weeks ago, I was trying to
modernize the python version of the FSS Slow
temperature control loops, when I accidentally
  
  13252   Fri Aug 25 01:20:52 2017 gautamUpdateLSCDRMI locking attemptI tried some DRMI   
  13253   Fri Aug 25 11:11:26 2017 gautamUpdateGeneralMC1 kicked againLooks like MC1 got another big kick just
under 4 hours ago. None of the other optics
show any evidence of a glitch so it seems
 MC1_glitch.png 
ELOG V3.1.3-