  6097   Fri Dec 9 17:08:41 2011 JenneUpdateRF SystemLots of current used in Rich's demod box

I checked the power regulators on the Rich demod box (according to the schematic, D1000217).  The positive one is LM2941CT, and the negative one is LM2991T.  Both accept input voltage up to +26V or -26V respectively.  So my use of +\- 24V to be regulated down to +\- 15V isn't too crazy.  It's a little crazy, but not too crazy.  They recommend having only a 3V difference between the input and output voltages.  We don't have any 18V or 20V power supplies in the regular LSC power supply rack, so Rana suggested using the 24's. 

When I plug in and turn on the Rich box, the current on the +24V power supply goes up by about 0.8A, and the -24V supply goes up by about 0.3A.  That seems like kind of a lot.  Is that too much?  Do I need to find a better plan that involves +\- 18V?  Thoughts?

For now, the Rich box is off, just in case.

  6099   Fri Dec 9 17:44:45 2011 KojiUpdateRF SystemLots of current used in Rich's demod box

Those asymmetric currents are same as what I saw with the table top +/-18V supply. If you really don't like it, there could be an option to disconnect CH3/4 in the box.

In any case, this could be a good long-run test of the demod box, couldn't it?

Quote:

I checked the power regulators on the Rich demod box (according to the schematic, D1000217).  The positive one is LM2941CT, and the negative one is LM2991T.  Both accept input voltage up to +26V or -26V respectively.  So my use of +\- 24V to be regulated down to +\- 15V isn't too crazy.  It's a little crazy, but not too crazy.  They recommend having only a 3V difference between the input and output voltages.  We don't have any 18V or 20V power supplies in the regular LSC power supply rack, so Rana suggested using the 24's. 

When I plug in and turn on the Rich box, the current on the +24V power supply goes up by about 0.8A, and the -24V supply goes up by about 0.3A.  That seems like kind of a lot.  Is that too much?  Do I need to find a better plan that involves +\- 18V?  Thoughts?

For now, the Rich box is off, just in case.

 

  6101   Fri Dec 9 20:03:57 2011 ZachUpdateRF SystemLots of current used in Rich's demod box

D0902745-v5 (probably the AP1053's):

Screen_shot_2011-12-09_at_8.00.54_PM.png

Quote:

Those asymmetric currents are same as what I saw with the table top +/-18V supply. If you really don't like it, there could be an option to disconnect CH3/4 in the box.

In any case, this could be a good long-run test of the demod box, couldn't it?

Quote:

I checked the power regulators on the Rich demod box (according to the schematic, D1000217).  The positive one is LM2941CT, and the negative one is LM2991T.  Both accept input voltage up to +26V or -26V respectively.  So my use of +\- 24V to be regulated down to +\- 15V isn't too crazy.  It's a little crazy, but not too crazy.  They recommend having only a 3V difference between the input and output voltages.  We don't have any 18V or 20V power supplies in the regular LSC power supply rack, so Rana suggested using the 24's. 

When I plug in and turn on the Rich box, the current on the +24V power supply goes up by about 0.8A, and the -24V supply goes up by about 0.3A.  That seems like kind of a lot.  Is that too much?  Do I need to find a better plan that involves +\- 18V?  Thoughts?

For now, the Rich box is off, just in case.

 

 

  6104   Mon Dec 12 11:16:02 2011 JenneUpdateRF SystemFoam house on EOM

Foam house installed on EOM a few min ago.  We'll leave it until ~tomorrow, then try out the heater loop.

  6109   Mon Dec 12 16:57:38 2011 JenneUpdateRF SystemRAMmon, 4 day trend

EOM was aligned to minimize the 11 and 55 MHz peaks in the RAMmon PD the other day (elog 6089), and was left with just the temperature sensor attached, no heater, no foam box.

Here is a 4 day trend:

EOM_tempSense_noFoam_noHeater_RAMmon_4days.png

I don't have a whole lot to say about this, other than there's a lot of stuff going on.  The craziness at the end is me realigning the PMC and MC since, as you can see, MC trans was way down.  The foam box was put on earlier today (elog 6104), so we'll see how that changes things over night.

  6111   Tue Dec 13 11:34:32 2011 JenneUpdateRF SystemRAMmon, 5 day trend

Now we've got another day of data, with the foam box on for the last 24hrs.

First plot is a 5 day trend so you can see that the RAM has gotten a little bit smaller, as has the temperature drift, but not by a whole lot.

EOM_tempSense_RAMmon_5days_last_day_withFoam_noHeater.png

Second plot is the last 19 hours (so excluding much of the time while I was realigning beams on the PSL table yesterday), to zoom in on just the time when the foam box was installed.

EOM_tempSense_withFoam_noHeater_RAMmon_1day.png

After lunch Zach and I are going to engage the heater to temperature stabilize the system, to see how that affects things.

In other news, the MC looks like it was fine for a good long time, and ~3 hours ago it went bad.  The mode that's flashing is really bad in both pitch and yaw.  I don't know what happened, but something is not so awesome. Edit: Steve said that he opened the PSL table at some point this morning to look around but not touch, and also it's Janitor Day, and Kevin comes in around 8ish. That doesn't mean I know the actual cause, but those are the only things that happened in the IFO room this morning that anyone is aware of.

  6116   Wed Dec 14 12:18:11 2011 ZachUpdateRF Systemheater reengaged

I reengaged the heater this morning, to compare it with the free-wafting and passive box-covered data. In order to make the loop stable, I had to reduce the gain of the AD620 by 10. I have increased the TEMP_MON preamp gain by 10, so the calibration should still be ~3.5 V/K into the ADC (and in DV).

Below is a screenshot showing that the RAMmon signals are pushed to some (nonzero) value, and it appears that they stay there despite the changing PSL table temperature as measured by FSS_RMTEMP. My post from last week shows that without the heater servo the temperature of the EOM can follows RMTEMP almost exactly. So, it seems like the heater is working well at low frequencies, modulo sensor noise, which ought to be low for the thermistor. Since several things (MC, etc.) have changed since out baseline data, it migth be prudent to let this sit for a little while and then disconnect the heater to see what happens.

heater_reengaged_12_14_11.png

  6119   Wed Dec 14 14:30:43 2011 JenneUpdateRF SystemLO for new demod box

The Rich demod box wants 10dBm for the local oscillator inputs, so I measured the values that we have coming out of the distribution box.  I'm using the "Spare 55MHz" and the "POP11" outputs, both of which had terminators so were not in use. 

The 55MHz had ~600mV peak, so between 5 and 6 dBm. 

The 11MHz had ~800mV peak, so about 8 dBm.

This is not enough dBm for either.  Going in search of RF amplifiers now...

  6121   Wed Dec 14 16:19:46 2011 ZachUpdateRF SystemLO for new demod box

I'm not sure I agree with your conversions, BUT:

The IQ boards use a PE4140, fancy MOSFET array as the mixer, and according to Peregrine (manufacturer), they can be operated with 0-20 dBm LO drive. I'm not recommending we drive them at 0 dBm, but perhaps the numbers you mentioned are OK.

Quote:

The Rich demod box wants 10dBm for the local oscillator inputs, so I measured the values that we have coming out of the distribution box.  I'm using the "Spare 55MHz" and the "POP11" outputs, both of which had terminators so were not in use. 

The 55MHz had ~600mV peak, so between 5 and 6 dBm. 

The 11MHz had ~800mV peak, so about 8 dBm.

This is not enough dBm for either.  Going in search of RF amplifiers now...

 

  6122   Wed Dec 14 18:06:39 2011 ZachUpdateRF SystemLO for new demod box

Actually, the LO inputs to the IQ boards have AP1053 (Cougar) amps on them. These are 10 dB amps and so putting 10 dBm in puts us on the very maximum of the LO range at 20 dBm.

I think the distribution box levels are fine. 

Quote:

I'm not sure I agree with your conversions, BUT:

The IQ boards use a PE4140, fancy MOSFET array as the mixer, and according to Peregrine (manufacturer), they can be operated with 0-20 dBm LO drive. I'm not recommending we drive them at 0 dBm, but perhaps the numbers you mentioned are OK.

Quote:

The Rich demod box wants 10dBm for the local oscillator inputs, so I measured the values that we have coming out of the distribution box.  I'm using the "Spare 55MHz" and the "POP11" outputs, both of which had terminators so were not in use. 

The 55MHz had ~600mV peak, so between 5 and 6 dBm. 

The 11MHz had ~800mV peak, so about 8 dBm.

This is not enough dBm for either.  Going in search of RF amplifiers now...

 

 

  6123   Wed Dec 14 19:59:12 2011 JenneUpdateRF SystemLO for new demod box

Quote:

Actually, the LO inputs to the IQ boards have AP1053 (Cougar) amps on them. These are 10 dB amps and so putting 10 dBm in puts us on the very maximum of the LO range at 20 dBm.

I think the distribution box levels are fine. 

Quote:

I'm not sure I agree with your conversions, BUT:

The IQ boards use a PE4140, fancy MOSFET array as the mixer, and according to Peregrine (manufacturer), they can be operated with 0-20 dBm LO drive. I'm not recommending we drive them at 0 dBm, but perhaps the numbers you mentioned are OK.

Quote:

The Rich demod box wants 10dBm for the local oscillator inputs, so I measured the values that we have coming out of the distribution box.  I'm using the "Spare 55MHz" and the "POP11" outputs, both of which had terminators so were not in use. 

The 55MHz had ~600mV peak, so between 5 and 6 dBm. 

The 11MHz had ~800mV peak, so about 8 dBm.

This is not enough dBm for either.  Going in search of RF amplifiers now...

 

 

 Yeah, I looked and saw that it's a semiconductor mixer, so it doesn't have to be as perfect. 

Everything is plugged in now to the new demod board.  More details soonly...

The I & Q outs are plugged into whitening filter #3, channels 5-8.  11MHz I = chan 5, 11MHz Q = chan 6, 55MHz I = chan 7, 55MHz Q = chan 8.  These channels are probably already recorded, but I haven't checked yet.  Hopefully I'll have time tonight after I pack for my trip.  But Zach, can you look into it tomorrow just to check??  Backup plan is to just go back to using the AS11 and POP55 boards and channels if the new board isn't doing what it's supposed to.

I disconnected the 3rd and 4th channels of the demod box since they were drawing unnecessary current, and making the box hot.  Now the box is just warmish.

  6138   Mon Dec 19 23:50:23 2011 ZachUpdateRF SystemRAMmon

I have been looking at the swings in the RAMmon channels since the heater was reengaged, to compare them to the data from beforehand (with and without the foam box). With the large grains of salt that I will list after, it appears that the EOM temperature controller does in fact reduce the amplitude of the swings by a measurable factor.

Salt:

  1. The reason I have not included any plots here is because the data suck. What we should ideally have is a continuous stream of RAMmon signals split into three chunks: 1) no foam, no heat, 2) foam, no heat, and 3) foam and heat. Instead, we have pieces of each kind of data on different days, before and after the MC has been realigned, some in old channels and some in new so that the calibration is different, etc. This piecemeal shit will not do.
  2. I realized that the LF boost was not engaged on the heater when I turned it on most recently. For this reason, the EOM temperature has not been stabilized as well as it might have been on diurnal timescales, and so with the boost it could be that the noise reduction is greater. For posterity, the DC suppression level is ~20x without the boost.

It seems impractical to try and rope off essentially 3 straight days where nothign major can be done to the IFO just to take RAM data. Instead, I think we should figure out a way to mimic the diurnal temperature swings on ~hour timescales. The EOM can temperature follows PSL-FSS_RMTEMP almost exactly and with a very short delay, so we can probably even accomplish this by stepping the lab A/C temperature. If this won't work, we can use an incandescent lamp or something similar to heat up the area around the EOM by a noticeable amount.

I'll try to come up with a good way to do this so that we can get some reliable data...

  6197   Fri Jan 13 17:40:38 2012 ZachUpdateRF Systemfoam box and temp controller taken off of PSL table

I stole the foam EOM box and the temperature controller circuit from the PSL table so that I could continue with the RAM measurements here at the ATF.

That is all.

  6390   Fri Mar 9 10:44:57 2012 steveUpdateRF SystemOSA

Optical spectrum analyzers like the Attachment made by Coherent , Meles Griot- CVI and Spectral Product are all discontinued.

The 40m have Coherent models C240 analyzer with controller C251 Their Finesse measured in 2004: sn205408  F302,  sn205409  F396,

Jenne borrowed Jan's Meles Griot model 13SAE006, Peter King has the same model. FSR 300 MHZ, finnees 200 minimum

Attachment 1: OSA.pdf
OSA.pdf
  6402   Mon Mar 12 22:14:56 2012 SureshUpdateRF SystemCalibration of Demod Board Efficiency.

I have completed the calibration of the demod board efficiencies.  Here is the schematic of the set-up.

 Calibration_Schematic.png

The data is given below and the data-file is attached in several different formats.

 Demod_calib.png

 

Attachment 3: Demod_calib.txt
								
	Measurements			 After corrections			Efficiency= out/in	
Demod Board	mV_ampl	mV_pk-pk	mV_pk-pk	mV_ampl	mV_ampl	mV_ampl	Vout/Vin	Vout/Vin
	PD in	Q out	I out	PD in	Q out	I out	Q out	I out
REFL33	10.6	10.0	10.0	9.4	5.0	5.0	0.53	0.53
AS11	24.0	10.0	11.0	21.3	5.0	5.5	0.23	0.26
REFL11	22.5	240.0	255.0	20.0	120.0	127.5	6.00	6.38
POX11	24.0	9.2	8.5	21.3	4.6	4.3	0.22	0.20
POY11	22.4	10.5	9.0	19.9	5.3	4.5	0.26	0.23
AS55	17.6	268.0	268.0	15.6	134.0	134.0	8.57	8.57
REFL55	19.7	15.8	15.5	17.5	7.9	7.8	0.45	0.44
POP55	18.8	278.0	274.0	16.7	139.0	137.0	8.32	8.20
REFL165	21.2	16.0	16.4	18.8	8.0	8.2	0.42	0.44
POY110	23.4	14.7	14.4	20.8	7.4	7.2	0.35	0.35
POY22	17.5	11.9	9.3	15.6	6.0	4.6	0.38	0.30
Attachment 4: Demod_calib.xlsx
  7353   Thu Sep 6 18:49:30 2012 JenneUpdateRF SystemAS 55 may be broken

I was going to lock MICH, but I don't see anything on dataviewer for either AS55Q or ASDC.  I went out onto the table, and there is beam on the diode, but no mV out on a voltmeter connected to the DC monitor point.  I shine a flashlight, and still I see 0.0mV.  So, something is up with AS55, but since the michelson is aligned right now, I'm not going to mess with the PD.  I won't lock MICH, I'll just move on.  Koji is taking a look at the diode, but if he doesn't get it figured out tonight, we can take a closer look after we pump down.

  7355   Thu Sep 6 19:36:19 2012 JenneUpdateRF SystemAS 55 is fine

Quote:

I was going to lock MICH, but I don't see anything on dataviewer for either AS55Q or ASDC.  I went out onto the table, and there is beam on the diode, but no mV out on a voltmeter connected to the DC monitor point.  I shine a flashlight, and still I see 0.0mV.  So, something is up with AS55, but since the michelson is aligned right now, I'm not going to mess with the PD.  I won't lock MICH, I'll just move on.  Koji is taking a look at the diode, but if he doesn't get it figured out tonight, we can take a closer look after we pump down.

 Never mind.  I was using an LED flashlight, which doesn't emit light that the PD is sensitive to.  A regular flashlight gives plenty of signal on the DC out. 

Using an SR560 with 30Hz low pass and gain of 100, it was pretty easy to align the light on the PD. 

Koji calculates in his head that there is about 6 microwatts of light incident on the PD, which is not a lot of light. Our SNR may be kind of lame for locking right now.

  7974   Thu Jan 31 14:46:05 2013 JenneUpdateRF SystemPhotodiode transimpedance

Quote:

Today I collected the data for shot noise intercept current for MC REFL PD. I didn't get many data points at higher DC voltage of the photodiode, cause the incandescent bulbs get burnt at that level; two bulbs I have burnt today. I will process the data and report.

 This work was done in-situ, so no optics on the AS table were moved.  The PSL shutter was blocked since the IR beam was not necessary, and would scatter off the bulb Riju put in front of the PD. 

  7976   Thu Jan 31 15:34:22 2013 RijuUpdateRF SystemPhotodiode transimpedance

Quote:

Quote:

Today I collected the data for shot noise intercept current for MC REFL PD. I didn't get many data points at higher DC voltage of the photodiode, cause the incandescent bulbs get burnt at that level; two bulbs I have burnt today. I will process the data and report.

 This work was done in-situ, so no optics on the AS table were moved.  The PSL shutter was blocked since the IR beam was not necessary, and would scatter off the bulb Riju put in front of the PD. 

 Thanks Jenne.

  8038   Fri Feb 8 17:15:56 2013 JenneUpdateRF SystemMC REFL Photodiode transimpedance

This measurement was done already about a week ago, in elog 7984.  Can you please describe why the numbers for the last measurement were not believable, and what was done differently this time?

  8044   Fri Feb 8 20:27:56 2013 KojiUpdateRF SystemMC REFL Photodiode transimpedance

The comment itself was added by me.
The difference between the previous and new measurements should be described by Riju.

In the entry 7984, the description has several PDs mixed up. The measurement was done with the MCREFL PD.
But the DC transimpedance of the thorlabs PD (5e3) was used, according to the text.
I first wonder if this is only a mistake not in the calculation but only in the elog due to a sloppy copy-and-paste.
But the resulting shot-noise-intercept current was 50uA, which is way too small
compared with a realistic value of 0.1~1mA. I have never seen such a good value with
C30642 at the resonant freq ~30MHz. That's why I said "hard to believe". I guessed this wrong
DC transimpedance was actually used for the calculation. 


You may wonder why this 50uA is unreasonable number.
Basically this is just my feeling and probably is same as Rana's feeling.
But "my feeling" can't be a scientific explanation. Here is some estimation.

Looking at my note in 2010:
https://wiki-40m.ligo.caltech.edu/40m_Library (Comparisons of the PD circuits by KA)

The expected shot noise intercept current (idc) is

idc = 2 kB T / (e Rres),

where Rres is the impedance of the resonant circuit at the resonant freq.

This Rres is expressed as

Rres = 1/(4 pi^2 fres^2 Cd^2 Rs),

where Cd and Rs are the capacitance and series resistance of the diode.

If we input realistic numbers,

Cd = 100pF
Rs = 10 Ohm
fres = 30MHz

We obtain, Rres ~ 300Ohm, and idc = 0.2mA

In other words, Rs needs to be 2~3Ohm in order to have idc = 50uA.
This is too small from the previous measurements.
Test Results for C30642 LSC Diode Elements by Rich Abbott

  8094   Sat Feb 16 18:32:01 2013 yutaSummaryRF Systemphase tracker: OLTF

I measured openloop transfer function of the phase tracking loop for the first characterization of phase tracker.

What is phase tracker:

  See elog #6832.
  For ALS, we use delay-line frequency discriminator, but it has trade-off between sensitivity and linear range. We solved this trade-off by tacking the phase of I/Q signals.
  Figure below is the current diagram of the frequency discriminator using phase tracker.
phasetracker.png

OLTF of phase tracking loop:
  Below. UGF at 1.2 kHz, phase margin 63 deg for both BEATX and BEATY. Phase delay can be clearly explained by 61 usec delay. This delay is 1 step in 16 KHz system.
  Note that UGF depends on the amplitude of the RF input. I think this should be fixed by calculating the amplitude from I/Q signals.
  BEAT(X|Y)_PHASE_GAIN were set to 300, and I put -3dBm 100 MHz RF signal to the beatbox during the measurement.
BEATX:BEATXFINEPHASE_OLTF.png BEATY:BEATYFINEPHASE_OLTF.png

Other measurements needed:

 - Linear range: By sweeping the RF input frequency and see sensitivity dependence.
 - Bandwidth: By measuring transfer function from the modulation frequency of the RF input to phase tracker output.
 - Maximum sensitivity: Sensitivity dependence on delay-line length (see PSL_Lab #825).
 - Noise: Lock oscillator frequency with phase tracker and measure out-of-loop frequency noise with phase tracker.
 - Sensitivity to amplitude fluctuation: Modulate RF input amplitude and measure the sensitivity.

  8199   Thu Feb 28 05:54:54 2013 yutaUpdateRF Systemphase tracker: calibration

I swept the frequency of RF input to the beatbox to calibrate and check linearity range of phase tracker.
Calibration factors are;
  C1:ALS-BEATX_FINE_PHASE_OUT    52.1643 +/- 0.0003 deg/MHz
  C1:ALS-BEATY_FINE_PHASE_OUT    51.4788 +/- 0.0003 deg/MHz


There was systematic error to the linearity check, but at least, calibration factor changes less than 50 % in the frequency range of 10 MHz to more than 500 MHz.


What I did:
  Used network analyzer(Aligent 4395A) to sweep the frequency RF input to the beatbox and getdata of phase tracker signal. I swept from 10 Hz to 500 MHz with 501 points in 50 sec. This sweep is slow enough considering we could lock the 40m arms (typical speed of a mirror is 1 um/s, so bandwidth of the phase tracker should be more than 1 um/sec / 40 m * 3e14 Hz = 75 MHz/s).
  RF amplitude was set to be -3 dBm and splitted into BEATX and BEATY.


Result:
  Plots for BEATX and BEATY are below;
ALS-BEATX_FINE_PHASE_OUT.pngALS-BEATY_FINE_PHASE_OUT.png


Discussion:
  - Considering delay line length is ~30m, expected calibration factor is;

    2*pi*l/v = 2*pi * 30 m / (2e8 m/s) = 0.94 rad/MHz = 54 deg/MHz

so, this calibration is reasonable.

  - Since frequency sweep of network analyzer is not continuous, phase tracker output is like steps with some ringdown. This makes some systematic error for checking linearity. I'm planning to do slower sweep or continuous sweep. Also, the phase tracker seems like he can exceed 500 MHz.

  8200   Thu Feb 28 06:51:17 2013 yutaUpdateRF Systemphase tracker: noise level

I measured noise level of the phase tracker by inputting constant frequency RF signal from marconi.
Measured frequency noise was ~2 Hz/rtHz @ 100 Hz. It's not so good.

What I did:
  1. Unplugged 11MHz marconi and put RF signal to the beatbox from this. Frequency and amplitude I put are 100 MHz and -3 dBm.
  2. Measured spectra of phase tracker outputs, C1:ALS-BEATX_FINE_PHASE_OUT, C1:ALS-BEATY_FINE_PHASE_OUT.
  3. Calibrated using the factor I measured (elog #8199).
  4. Put marconi back to orignal settings.

Result:
frequencynoise.png

Discussion:
  - According to Schilt et al., this noise level is not so good.
  - By changing the delay-line cable length or optimizing whitening filter etc., we can improve this.
 

  8214   Sat Mar 2 20:09:12 2013 yutaUpdateRF Systemphase tracker: DAQ noise limited

[Koji, Yuta]

We found that our phase tracker noise is currently limited by the noise introduced in DAQ.
We confirmed that the frequency noise was improved from 2 Hz/rtHz to 0.4 Hz/rtHz by increasing the gain of the whitening filter.
The whitening filters should definitely be refined.

What we did:
  1. Put constant frequency RF input to the beatbox from Marconi and measured noise spectrum of the beatbox output(BEATX I) after the whitening filter with a spectrum analyzer. Noise floor level was ~0.2 Hz/rtHz at carrier frequency range of 15-100 MHz. Calibration factor of the beatbox output was ~380 mV/MHz.

  2. Measured noise spectrum of C1:ALS-BEATX_FINE_I_OUTPUT(figure below). The noise floor didn't change when there was RF input of 100 MHz from Marconi(blue) and DAQ input was terminated (green). Also, C1:ALS-BEATX_FINE_I_IN1(which is before unwhitening filter) showed a flat spectrum. These show our spectrum is limited by DAQ noise, which is introduced after the whitening filter.

  3. We increased the gain of whitening filter by x20 to show frequency noise performance can be improved by better whitening filter(red). But we can not use this setup as the other quadrature will be saturated by a too much gain at DC. Thus we need to carefully consider the signal level and the gain distribution of the whitening filters.
frequencynoisewhitening.png


Next:
  - Better whitening filters. The current one consists of zero 1 Hz and pole 10 Hz with DC gain of 5 using SR560.
  - Better beatbox. We can increase the RF input power to the mixer and unify the preamplifier and the whitening filter in the box.

  8485   Wed Apr 24 14:36:06 2013 JenneUpdateRF SystemPD frequency response

I think you have the splitter that splits the RF signal from the network analyzer in the wrong place. 

Usually you split the signal immediately after the RF Out, so that half of the signal goes to the A-input of the Analyzer, and the other half goes to your controller (here, the laser diode controller).  Then you would take the output of your controller and go straight to the actual laser diode, with no splitting in this path.

  8488   Thu Apr 25 00:59:37 2013 RijuUpdateRF SystemPD frequency response

Quote:

I think you have the splitter that splits the RF signal from the network analyzer in the wrong place. 

Usually you split the signal immediately after the RF Out, so that half of the signal goes to the A-input of the Analyzer, and the other half goes to your controller (here, the laser diode controller).  Then you would take the output of your controller and go straight to the actual laser diode, with no splitting in this path.

 Here our device under test is the photodiode. So for the reference I wanted to retain the response of the laser diode controller. Otherwise I have to consider the transfer function of that LDC too. I may check both the options at the time of experiment.

Thanks

  8499   Fri Apr 26 21:38:06 2013 JenneConfigurationRF SystemPD frequency response

I was sad to see that there wasn't a photo of the POX situation after the fiber work was done on Thursday.

Also, I was out looking at something else, and noticed that the fibers aren't in a very good/safe place from the POX table over to your splitter.  Getting to the POX table is certainly more tricky than the AP table, since the fiber splitter is right next to the AP table, but we should go back and try to make sure the fibers to the more distant tables are laid in a nice, safe way.

Is there a reason that we're not using the clear plastic tubing that Eric bought to put the fibers into?  It seems like that would help a lot in keeping the fibers safe.

I took a few photos of the things that I'm sad about:

1. We should not be keeping fibers on the floor in an area where they can be stepped on.  This will be fixed (I hope) as part of putting the extra coiled length over by the splitter.

IMG_0498.JPG

2. Again, in an area where we semi-regularly walk, the fibers should not be a tripping hazard.  Behind the table legs (rather than under the middle of the table) is safer, and will help tuck them out of the way.

IMG_0499.JPG

3.  It's not obvious when we're pumped down, but we remove the access connector (top right side of this photo), and need to walk in this area.  I can pretty much guarantee that within 1 day of the next time we vent, these fibers will be stepped on, tripped over, and broken if they are not moved to a different location.  I'm not yet sure what the best way to route these fibers is, but this is not it. 

IMG_0500.JPG

Riju, since Eric will be away next week, please let one of us "40m Regulars" know when you plan to come over (at least a few hours ahead of time), and we can give you a hand in protecting these fibers a little bit better.  Thanks!

  8506   Mon Apr 29 17:26:22 2013 RijuConfigurationRF SystemPD frequency response

 Today I have rerouted the fibers on AP table to remove the fiber rolls out of the AP table.  I removed the fibers one by one from both ends - from the 1x16 splitter and from the AP table - keeping the fiber roll intact, and then connected it in reverse way, i.e. the fiber end which was on AP table now is connected to the splitter (since length of the outside the roll is shorter that side) and the fiber end connected to splitter is now rerouted on AP table.

We need to keep the fibers in such a fashion so that no sharp bending occurs anywhere, and also it does not get strained due to its weight, particularly near the 1x16 splitter. Jenne suggested to use a plastic box over the splitter rack to keep the fiber rolls for time-being. We discussed a lot how this can be done nicely; in future we may use array of hooks, Koji suggested to use cable hangers and to tie the rolls using more than one hanging point, Jenne suggested to use the bottom shelf of the rack or to use one plastic box with holes. We tried to make holes on the plastic box using drill, but it developed crack on the box. So ultimately I used the opened box only and put it over the rack.

The corresponding photographs are attached herewith.

Tomorrow we will reroute the fibers for POX table.  

Attachment 1: IMG_0504.JPG
IMG_0504.JPG
Attachment 2: IMG_0503.JPG
IMG_0503.JPG
Attachment 3: IMG_0505.JPG
IMG_0505.JPG
Attachment 4: IMG_0506.JPG
IMG_0506.JPG
Attachment 5: IMG_0508.JPG
IMG_0508.JPG
Attachment 6: IMG_0509.JPG
IMG_0509.JPG
Attachment 7: IMG_0510.JPG
IMG_0510.JPG
  8512   Tue Apr 30 19:39:14 2013 RijuConfigurationRF SystemPD frequency response

Today I have rerouted the fibers on POX table. The aim was to lay it overhead through the plastic pipe. A pipe ~50ft (~15.5m) long was taken for this purpose. I disconnected the two 25m long fibers for POP55 and POX11 PD (those had been already routed) from both of their ends - i.e. from the POX table and also from 1x16 splitter. Jenne and Koji suggested that we may have another two PDs ( POP22 and POP110) on POX table in future. So we used another 25m long fiber for these two (POP22/POP110). We could not use two fibers for these two since we have only four 25m long fibers and one of them we need for POY11 PD on POY table. Jenne and me put the three fibers inside the pipe using a copper tube. The tube then was put on the overhead rack, Manasa helped me to do it. The fiber ends were finally laid on the POX table at one end and connected to the 1x16 splitter at the other end.

The corresponding photos are attached herewith.

Attachment 1: IMG_0511.JPG
IMG_0511.JPG
Attachment 2: IMG_0512.JPG
IMG_0512.JPG
Attachment 3: IMG_0513.JPG
IMG_0513.JPG
Attachment 4: IMG_0514.JPG
IMG_0514.JPG
Attachment 5: IMG_0515.JPG
IMG_0515.JPG
  8513   Tue Apr 30 21:24:15 2013 JenneConfigurationRF SystemPOX fiber laying

Nice work.  That was a lot of effort, but having done it so nicely will definitely pay off, since it is now much harder to break the fibers.

2 small issues:  In your attachment 3, I see a coil of fiber just outside the POX table.  I thought Koji had asked that all spare coiled-up length of fiber always be at the splitter side.  Also, in securing the plastic tubing as it comes down near the PSL table, you have zip-tied the tubing to the PSL table.  Since that is a space that we need to access to align the Xgreen beatnote stuff, please disconnect that zip tie, and secure the tubing on the north side somewhere, underneath the AP table, rather than the PSL table (when you look closer, you may notice that no cables in that bundle are attached to the PSL side at the bottom, for this same reason). 

  8515   Tue Apr 30 23:04:23 2013 JenneConfigurationRF SystemOnly 4 25m cables ordered

I have found in the depths of the elog the (original?) list of fibers and lengths that were decided upon:  elog 6535.

In Suresh's elog, we were assuming that POP22 & POP110 would be served by a single PD.  This is still the nominal plan, although we (Rana is maybe still thinking about this in the back of his head?) think that it might not be feasible.  Riju and I were hoping to put a 4th fiber in the tubing so that we wouldn't have to add it later if POP22 & POP110 are eventually 2 separate PDs.  Anyhow, for now, all we have available are 3 fibers for the POX table, so that is what was installed this afternoon.

  8520   Wed May 1 17:36:26 2013 RijuConfigurationRF SystemPD frequency response

 Today I routed fiber from 1x16 splitter to POY table. Manasa helped me doing that. The fiber(25m) was laid on overhead rack through plastic pipe of length ~76ft. We put the fiber inside the pipe using one copper tube, and then tied the plastic pipe on the overhead rack. Finally one end of the fiber was laid on POY table and the other end was connected to the 1x16 splitter. The photographs corresponding are attached. There is no picture of splitter end, cause it was dark that time.

Attachment 1: IMG_0517.JPG
IMG_0517.JPG
Attachment 2: IMG_0518.JPG
IMG_0518.JPG
Attachment 3: IMG_0519.JPG
IMG_0519.JPG
  8538   Tue May 7 17:13:30 2013 JenneUpdateRF SystemIdeal PRMI RF frequency

Koji asked me to look at what the ideal RF modulation frequency is, for just the PRMI case (no arms).  If we had a perfect interferometer, with the sidebands exactly antiresonant in the arms when the arms resonate with the carrier, this wouldn't be an issue.  However, due to vacuum envelope constraints, we do not have perfect antiresonance of the sidebands in the arm cavities.  Rather, the sideband frequencies (and arm lengths) were chosen such that they pick up a minimum amount of extra phase on reflection from the arms.  But, when the arms are off resonance (ex, the ETMs are misaligned), the sideband frequencies see a different amount of phase.   

We want to know what a rough guess (since we don't have a precise number for the length of the PRC since our last vent) is for the ideal RF modulation frequency in just the PRMI. 

If I take (from Manasa's kind measurements from the CAD drawing yesterday) the relevant distances to be:

L_P[meters] = 1.9045 + 2.1155 + 0.4067

L_X[meters] = 2.3070 + 0.0254*n

L_Y[meters] = 2.2372 + 0.0359*n + 0.0254*n

L_PRCL = L_P + (L_X + L_Y)/2 = 6.7616 meters.

The *n factors (n=1.44963) are due to travel through glass of the BS, and the substrate of the ITMs. 

I find the FSR of the PRC to be 22.1686 MHz. For the sidebands to be antiresonant, we want them to be 11.0843 MHz. This would correspond to a mode cleaner length of 27.0466 meters.  Our current modulation frequency of 11.066134 MHz corresponds to a MC length of 27.091 meters.  So, if we want to use this 'ideal' modulation frequency for the PRC, we need to shorten the mode cleaner by 4.4cm!  That's kind of a lot.

  8539   Tue May 7 17:30:28 2013 KojiUpdateRF SystemIdeal PRMI RF frequency

To change the MC length is not the point.

If we can improve the length sensing by the intentional shift of the modulation frequency from the MC FSR, that's worth to try, I thought.

But that is tough as the freq difference is 18kHz that is ~x4 of the line width of the MC.
Not only the 55MHz sidebands, but also the 11MHz sidebands will just be rejected.

Nevertheless: Is there any possibility that we can improve anything by shifting the modulation frequency by ~1kHz?

  8573   Tue May 14 19:52:58 2013 RijuConfigurationRF SystemPD frequency response

 [Eric, Riju, Annalisa]

Today we have cleared up the fiber spool near AP table. We have put the 1x16 fiber splitter and a box (we made two openings on it) for fiber spool on a different part of the rack. Also put a plastic tubing or the fibers coming out of AP table. Now the fibers coming out from AP table and also from POX table first enter the box through one opening and the end of the fibers come out of the other opening to get connected to to splitter. Photographs of the work are attached. I don't think enough fiber is there to make a similar loop for fiber coming from POY table.

 

 

Attachment 1: IMG_0520.JPG
IMG_0520.JPG
Attachment 2: IMG_0521.JPG
IMG_0521.JPG
  8578   Wed May 15 08:29:28 2013 SteveConfigurationRF Systemfiber protection at splitter box area

 I positioned the fiber loaded protecting tubing and anchored them so they can do their job.

However, the area needs a good clean up.

 

Attachment 1: fiberprotect.jpg
fiberprotect.jpg
  7301   Tue Aug 28 18:28:21 2012 janoschMetaphysicsRingdownripples

Let's see if the ripples observed in the MC ringdown can be due to tilt motion of the mirrors.

The time it takes to produce a phase shift corresponding to N multiples of 2*pi is given by:

t = sqrt(2*N*lambda/(L*omega_T^2*(alpha_1+alpha_2)))

L is the length of the MC (something like 13m), and alpha_1, alpha_2 are the DC tilt angles of the two mirrors "shooting into the long arms of the MC" produced by the MC control with respect to the mechanical equilibrium position. omega_T is the tilt eigenfrequency of the three mirrors (assumed to be identical). lambda = 1.064e-6m;

The time it takes from N=1 to N=2 (the first observable ripple) is given by: tau1 = 0.6/omega_T*sqrt(lambda/L/(alpha_1+alpha_2))

The time it takes from N=2 to N=3 is given by: tau2 = 0.77*tau1

etc

First, we also see in the measurement that later ripples are shorter than early ripples consistent with some accelerated effect. The predicted ripple durations tau seem to be a bit too high though. The measurements show something like a first 14us and a late 8us ripple. It depends somewhat on the initial tilt angles that I don't know really.

In any case, the short ripple times could also be explained if the tilt motions start a little earlier than the ringdown, or the tilt motion starts with some small initial velocity. The next step will be to program a little ringdown simulation that includes mirror tilts and see what kind of tilt motion would produce the ripples exactly as we observe them (or maybe tilt motion cannot produce ripples as observed).

  7302   Tue Aug 28 19:06:32 2012 KojiMetaphysicsRingdownripples

Isn't it just a ringing of the intracavity power as you shifted the laser frequency abruptly?

Quote:

Let's see if the ripples observed in the MC ringdown can be due to tilt motion of the mirrors.

The time it takes to produce a phase shift corresponding to N multiples of 2*pi is given by:

t = sqrt(2*N*lambda/(L*omega_T^2*(alpha_1+alpha_2)))

L is the length of the MC (something like 13m), and alpha_1, alpha_2 are the DC tilt angles of the two mirrors "shooting into the long arms of the MC" produced by the MC control with respect to the mechanical equilibrium position. omega_T is the tilt eigenfrequency of the three mirrors (assumed to be identical). lambda = 1.064e-6m;

The time it takes from N=1 to N=2 (the first observable ripple) is given by: tau1 = 0.6/omega_T*sqrt(lambda/L/(alpha_1+alpha_2))

The time it takes from N=2 to N=3 is given by: tau2 = 0.77*tau1

etc

First, we also see in the measurement that later ripples are shorter than early ripples consistent with some accelerated effect. The predicted ripple durations tau seem to be a bit too high though. The measurements show something like a first 14us and a late 8us ripple. It depends somewhat on the initial tilt angles that I don't know really.

In any case, the short ripple times could also be explained if the tilt motions start a little earlier than the ringdown, or the tilt motion starts with some small initial velocity. The next step will be to program a little ringdown simulation that includes mirror tilts and see what kind of tilt motion would produce the ripples exactly as we observe them (or maybe tilt motion cannot produce ripples as observed).

 

  7303   Tue Aug 28 19:21:37 2012 janoschMetaphysicsRingdownripples

Hmm. I don't know what ringing really is. Ok, let's assume it has to do with the pump... I don't see how the pump laser could produce these ripples. They have large amplitudes and so I always suspected something happening to the intracavity field. Therefore I was looking for effects that would change resonance conditions of the intracavity field during ringdown. Tilt motion seemed to be one explanation to me, but it may be a bit too slow (not sure yet). Longitudinal mirror motion is certainly too slow. What else could there be?

Quote:

Isn't it just a ringing of the intracavity power as you shifted the laser frequency abruptly?

Quote:

Let's see if the ripples observed in the MC ringdown can be due to tilt motion of the mirrors.

The time it takes to produce a phase shift corresponding to N multiples of 2*pi is given by:

t = sqrt(2*N*lambda/(L*omega_T^2*(alpha_1+alpha_2)))

L is the length of the MC (something like 13m), and alpha_1, alpha_2 are the DC tilt angles of the two mirrors "shooting into the long arms of the MC" produced by the MC control with respect to the mechanical equilibrium position. omega_T is the tilt eigenfrequency of the three mirrors (assumed to be identical). lambda = 1.064e-6m;

The time it takes from N=1 to N=2 (the first observable ripple) is given by: tau1 = 0.6/omega_T*sqrt(lambda/L/(alpha_1+alpha_2))

The time it takes from N=2 to N=3 is given by: tau2 = 0.77*tau1

etc

First, we also see in the measurement that later ripples are shorter than early ripples consistent with some accelerated effect. The predicted ripple durations tau seem to be a bit too high though. The measurements show something like a first 14us and a late 8us ripple. It depends somewhat on the initial tilt angles that I don't know really.

In any case, the short ripple times could also be explained if the tilt motions start a little earlier than the ringdown, or the tilt motion starts with some small initial velocity. The next step will be to program a little ringdown simulation that includes mirror tilts and see what kind of tilt motion would produce the ripples exactly as we observe them (or maybe tilt motion cannot produce ripples as observed).

 

 

  7304   Tue Aug 28 20:23:54 2012 KojiMetaphysicsRingdownripples

Laser frequency shift = longitudinal motion of the mirrors

Ringing: http://www.opticsinfobase.org/ol/abstract.cfm?uri=ol-20-24-2463

Quote:

Hmm. I don't know what ringing really is. Ok, let's assume it has to do with the pump... I don't see how the pump laser could produce these ripples. They have large amplitudes and so I always suspected something happening to the intracavity field. Therefore I was looking for effects that would change resonance conditions of the intracavity field during ringdown. Tilt motion seemed to be one explanation to me, but it may be a bit too slow (not sure yet). Longitudinal mirror motion is certainly too slow. What else could there be?

 

  7305   Wed Aug 29 09:35:03 2012 janoschMetaphysicsRingdownripples 2

Ok, so the whole idea that mirror motion can explain the ripples is nonsense. At least, when you think off the ringdown with "pump off". The phase shifts that I tried to estimate from longitudinal and tilt mirror motion are defined against a non-existing reference. So I guess that I have to click on the link that Koji posted...

Just to mention, for the tilt phase shift (yes, there is one, but the exact expression has two more factors in the equation I posted), it does not matter, which mirror tilts. So even for a lower bound on the ripple time, my equation was incorrect. It should have the sum over all three initial tilt angles not only the two "shooting into the long arms" of the MC.

Quote:

Laser frequency shift = longitudinal motion of the mirrors

Ringing: http://www.opticsinfobase.org/ol/abstract.cfm?uri=ol-20-24-2463

Quote:

Hmm. I don't know what ringing really is. Ok, let's assume it has to do with the pump... I don't see how the pump laser could produce these ripples. They have large amplitudes and so I always suspected something happening to the intracavity field. Therefore I was looking for effects that would change resonance conditions of the intracavity field during ringdown. Tilt motion seemed to be one explanation to me, but it may be a bit too slow (not sure yet). Longitudinal mirror motion is certainly too slow. What else could there be?

 

 

  7330   Fri Aug 31 17:44:21 2012 ManasaUpdateRingdownData

Quote:

Ok, so the whole idea that mirror motion can explain the ripples is nonsense. At least, when you think off the ringdown with "pump off". The phase shifts that I tried to estimate from longitudinal and tilt mirror motion are defined against a non-existing reference. So I guess that I have to click on the link that Koji posted...

Just to mention, for the tilt phase shift (yes, there is one, but the exact expression has two more factors in the equation I posted), it does not matter, which mirror tilts. So even for a lower bound on the ripple time, my equation was incorrect. It should have the sum over all three initial tilt angles not only the two "shooting into the long arms" of the MC.

Quote:

Laser frequency shift = longitudinal motion of the mirrors

Ringing: http://www.opticsinfobase.org/ol/abstract.cfm?uri=ol-20-24-2463

Quote:

Hmm. I don't know what ringing really is. Ok, let's assume it has to do with the pump... I don't see how the pump laser could produce these ripples. They have large amplitudes and so I always suspected something happening to the intracavity field. Therefore I was looking for effects that would change resonance conditions of the intracavity field during ringdown. Tilt motion seemed to be one explanation to me, but it may be a bit too slow (not sure yet). Longitudinal mirror motion is certainly too slow. What else could there be?

 

 

It is essential we take a look at the ringdown data for all measurements made so far to figure out what must be done to track the source of these notorious ripples. I've attached the plot for the same showing the decay time to be the same in all cases. About the ripples; it seems unlikely to both Jan and me that the ripples are some electronic noise because the ripples do not follow any common pattern or time constant. We have discussed with Koji about monitoring the frequency shift, the input power to the MC and also try other methods of shutting down the pump to track their source as the next steps.

 

cum_plot.png 

  57   Fri Nov 2 08:59:30 2007 steveBureaucracySAFETYthe laser is ON
The psl laser is back on !
  115   Mon Nov 19 14:32:10 2007 steveBureaucracySAFETYgrad student safety training
John Miller and Alberto Stochino has received the 40m safety bible.
They still have to read the laser operation manual and sign off on it.
  196   Tue Dec 18 16:50:35 2007 tobinUpdateSAFETYuvex laser safety glasses defective
A few days ago we noticed what appeared to be a blotched, speckled fracturing of the coating of the "UVEX" laser safety glasses. These are the glasses with "transparent" (reflective to 1064nm) lenses and white frames that we keep in a box on top of a filing cabinet in the control room. Today Steve measured the transmission of these glasses and found 80% transmission of 1064nm in several cases.

Do not use the white, transparent "uvex" laser safety glasses until further notice. Steve has hidden them away so that you won't be tempted.

Below is attached a photo of a bad lens.
Attachment 1: bad-glasses.jpg
bad-glasses.jpg
Attachment 2: bad-glasses-zoom.jpg
bad-glasses-zoom.jpg
  203   Wed Dec 19 16:40:12 2007 steveUpdateSAFETYlaser safety glasses measured
I measured the coarse transission at 1064nm of the 40m safety glasses today.

12 pieces of UVEX # LOTG-YAG/CO2 light green, all plastic construction, ADSORBANT

3 pieces of 6KG5, Scott colored filter glass type,

individual prescription glasses: alan, bob, ben, jay and steve

7 pieces of dual waveleght glasses

These glasses showed 0.00mW transmission out of 170mW Crysta Laser 1064
  443   Thu Apr 24 15:57:53 2008 steveConfigurationSAFETYSafety at AP-ISCT
I measured the output power of the psl after the mechanical shutter.

It was 1.1 W with Ophir power meter, than unlocked the MC and measured
the power at the MC-REFL Beam Dump at the AP-ISCT 0.9 W
Power on MC-REFL photodiode 92 mW

High power metal beam shields were installed around the beam path of
MC-REFL between AP-Viewport and MC-REFL Beam Dump.
Placed HIGH POWER LASER BEAM PATH warning signs on table frame and top
covers.

Last week I placed a small monitor on the top of the OOC that
monitors the resonant spot of MC2. Please keep an eye on this monitor
when working on the AP-ISCT

AP table should NOT be left uncovered. One experienced laser operator
has to be present if the top is removed and IR-viewer scan required.
We need your full cooperation to keep this lab safe.
Attachment 1: P1020197.JPG
P1020197.JPG
Attachment 2: mcrefl3.JPG
mcrefl3.JPG
  488   Tue May 20 09:28:42 2008 steveBureaucracySAFETYsafety traning for 40m
Tara Chalernsongsak, new gradstudent for K. Libbrecht was introduced to the basics of the 40m operations.
  491   Thu May 22 11:21:45 2008 steveBureaucracySAFETYearly surf sudent
Caltech under grad Eric Mintun received the 40m safety training.
Now he has to read and sign SOP of laser and ifo.
He'll be working with GigE cameras with Joe