  12473   Tue Sep 6 20:30:56 2016 ranaUpdateSUSITMY_UL is sickIn the morning, Steve will start opening
the north BS door so that we can enter to
inspect the PRM LR OSEM.
  
  12474   Tue Sep 6 20:45:14 2016 gautamUpdateSUSX arm test masses back in chamber[Teng, Johannes,
Lydia, gautam]


The
 IMG_3137.JPGIMG_3143.JPGIMG_3142.JPGIMG_3148.JPG 
  12475   Tue Sep 6 20:52:42 2016 gautamUpdateSUSOSEM issues



The
 P9060254.JPG 
  12476   Wed Sep 7 17:23:26 2016 gautamUpdateSUSPRM LR fixed for now[johannes, gautam]


We took the heavy door
off in the morning with Steve's help
The
 IMG_3208.JPGIMG_3211.JPG 
  12477   Wed Sep 7 18:00:47 2016 LydiaUpdateGeneralAcromag Progress[Teng, Lydia]

We
would like to establish a system for setting
  
  12478   Thu Sep 8 22:12:36 2016 gautamUpdateSUSX arm in place, locked to green, IR flashes visibleDetailed elog to follow but summary of
todays activities:


ITMX and ETMX are back
  
  12479   Fri Sep 9 11:24:15 2016 SteveUpdatesafetysafety trainingVisiting
graduate student Teng Zhang from Glasglow
  
  12480   Fri Sep 9 17:50:02 2016 gautamUpdateSUSHeavy doors on BS-PRM, ETMY chambers[steve, teng, johannes, lydia, gautam]


we set about doing some
final checks on the Y arm while Johannes
  
  12481   Sat Sep 10 14:26:01 2016 ranaUpdateGeneralPRM SRM alignmentFor the PRM, I aligned it until the arm
flashes were maximized and the REFL camera
showed a centered spot with dips happening
 Osamu.JPG 
  12482   Mon Sep 12 17:15:22 2016 JohannesUpdateGeneralPRM SRM alignment[Gautam, Steve, Johannes]

We put on the remaining heavy doors
on the chambers (ITMY, ITMX,ETMX, in this
 IMG_3273.JPGIMG_3275.JPG 
  12483   Mon Sep 12 17:54:24 2016 SteveUpdateVACpumpdown 79 We are pumping down. The annuloses
are below 10 mTorr

The vacuum gauges are not communicating
  
  12484   Mon Sep 12 20:15:22 2016 LydiaUpdateSUSDiagonalization in air[Lydia, Teng]

We ran the scripts to diagonalize
the damping matrices using the free swinging
 ITMY_fit.jpgBS_diag.jpg 
  12485   Mon Sep 12 20:19:25 2016 LydiaUpdateGeneralMC REFL beam splitter not replacedThe
beam splitter that directs light into the
MC REFL photodiode has not been replaced;
  
  12486   Tue Sep 13 11:00:59 2016 LydiaUpdateSUSETMY UL glitch returned 




  
  12488   Tue Sep 13 17:47:04 2016 SteveUpdateVACpumpdown 79 completedP1 IFO pressure is 1 mTorr, valve configuration: vacuum normal,
annulosses are pumped, RGA is off, not pumped.
  
  12489   Tue Sep 13 19:02:56 2016 TengUpdateGeneralITMX sensor[Lydia,Teng]

Something strange happened to the
ITMX osem reading around 4.pm. PDT as shown
 47.png34.png 
  12490   Tue Sep 13 19:18:43 2016 LydiaUpdateSUSDiagonalization in air[Lydia, Teng]

We continued to work on the diagonalization
scripts today and devised a way of choosing
 ITMY_diag.jpg38.png 
  12491   Wed Sep 14 09:33:54 2016 SteveUpdateGeneralITMX magnets are stuckedI believe that the UR and LR magnets are
stuck. There was no earth quake at 16:18
yesterday. Something had to kick it into
 ITMXmagStuck.png 
  12492   Wed Sep 14 09:44:09 2016 SteveUpdateVACpumpdown 79 completedThe pumpdown started at
9-12-2016

The IFO is at 5e-5 Torr vacuum
 pd79.pngpd79d2.jpgCCcg.jpg 
  12493   Wed Sep 14 19:41:23 2016 JohannesUpdateGeneralPSL back to high powerToday's summary:


Replaced mirror in MC REFL
path with R=10% BS and aligned beam on PD
  
  12494   Wed Sep 14 20:05:32 2016 LydiaUpdateGeneralITMX magnets are stuckedWhen I restarted c1susaux yesterday, I
didn't know that I needed to disable
the coil outputs first. So when it came back
  
  12495   Wed Sep 14 20:27:03 2016 LydiaUpdateSUSDiagonalizationToday the main optics were free swinging
for several hours, so I attempted diagonalization
in vacuum.
 ETMY_diag.png 
  12496   Thu Sep 15 10:36:32 2016 SteveUpdateVACpumpdown 79 day 3IFO pressure 3.7e-5 Torr at new cold cathode
InstruTech - Hornet

New
 vent78for73d.png 
  12497   Thu Sep 15 18:37:20 2016 LydiaUpdateSUSDiagonalization[Teng, Lydia]


We fixed the 60Hz filter
on ITMY. This improved the phase problems
 ETMY_osemspec.pngITMY_osemspec.png 
  12498   Fri Sep 16 14:15:28 2016 steveUpdatePEMpartical countsSouth end flow bench and both clean room
assembly flow benches measured zero counts
for 0.3 and 0.5 micron size particales.
 80days.png 
  12499   Fri Sep 16 19:14:27 2016 LydiaUpdateSUSDiagonalization[Lydia, Teng]

We built matrices for ITMY
and ETMY by driving one degree of freedom
 ETMY_yawdrivedefault.pngETMY_yawdrivenew.pngETMY_yawdriveold.png57.png 
  12500   Fri Sep 16 19:48:52 2016 LydiaUpdateGeneralAlignment statusToday the Y arm was locking fine. The alignment
had drifted somewhat so I ran the dither
and TRY returned to ~0.8. However, the mode
  
  12501   Sat Sep 17 02:00:23 2016 ranaUpdateSUSAlignment statusAll is not lost. I've stuck and unstuck
optics around a half dozen times. Can you
please post the zoomed in time series (not
  
  12502   Sat Sep 17 16:51:01 2016 LydiaUpdateSUSAlignment statusHere's the timeseries plots. I've
zoomed in to right after the problem- did
you want before? We pretty much know
 Screenshot_from_2016-09-17_16-45-00.png 
  12503   Sun Sep 18 16:18:05 2016 ranaUpdateSUSAlignment statussusaux is responsible for turning on/off
the inputs to the coil driver, but not the
actual damping loops. So rebooting susaux
  
  12504   Mon Sep 19 11:11:43 2016 ericqUpdateSUSAlignment status[ericq, Steve]

ITMX is free, OSEM signals all
rougly centered. 
  
  12505   Mon Sep 19 13:25:03 2016 TengUpdateElectronicsSatellite Amplifier 

In order to figure out the difference
betweent simulated result and measurement,
 Ele_measurement.pngEle_noise.png 
  12506   Mon Sep 19 13:57:21 2016 ranaUpdateGeneralnever post EPS files in the ELOG. Ever.http://tex.stackexchange.com/questions/2092/which-figure-type-to-use-pdf-or-eps   
  12507   Mon Sep 19 22:03:10 2016 ericqUpdateGeneralFurther recovery progress[ericq, Lydia, Teng]

Brief summary of this afternoon's
activities:
  
  12508   Tue Sep 20 10:45:06 2016 ranaUpdateGeneralFurther recovery progressRana suspicious. We had arms locked before
pumpdown with beams on Transmon PDs. If they're
off now, must be beams are far off on the
  
  12509   Tue Sep 20 17:04:46 2016 SteveUpdateElectronicsREF33REF33 was removed for taking picture of
the bare C30362 InGaAs photodiode per Rana's
request. All other rf photodiodes have their
 A005_-_20160920_135529_-_Shortcut.lnk.bmp 
  12510   Wed Sep 21 01:08:02 2016 ericqUpdateGeneralFurther recovery progressThe misalignment wasn't as bad as I
had intially feared; the spot was indeed
pretty high on ETMX at first. Both transmon
  
  12511   Wed Sep 21 09:04:57 2016 SteveUpdateGeneral8 hours recovery progressGood 8 hours




 8hrs.png 
  12512   Thu Sep 22 15:44:21 2016 SteveUpdateGeneralITMX magnets are stucked again 

 
 StuckAgain.png 
  12513   Thu Sep 22 20:01:47 2016 LydiaUpdateGeneralITMX freed, all optics kickedRana came by and freed ITMX again. I think
it shouldn't be a problem for me to free
it if it happens again. 
  
  12514   Thu Sep 22 20:18:27 2016 LydiaUpdateGeneralAcromag ProgressWe moved the Acromag and its power supply
to the X end, where we connected it to the
diagnostic output of the NPRO controller.
  
  12515   Thu Sep 22 22:52:08 2016 ericqUpdateGeneralDamping found to be onJust a heads up, it looks like the damping
came on at around 8:30pm. Not sure why. 
  
  12516   Fri Sep 23 01:09:04 2016 ericqUpdateSUSETMX hysteresis testI had hoped to do some ALS work, but I
realized too late that we loaned our HP analyzer
to Andrew. I decided instead to do some ETMX
  
  12517   Sat Sep 24 11:04:00 2016 ericqUpdateSUSETMX hysteresis testSeems like the angular position was fairly
stable, though there is some change in the
ETMX pitch that could be hysterisis or normal
 hysterisis_test.png 
  12518   Mon Sep 26 19:48:09 2016 LydiaUpdateSUSITMX stuck again, ITMY whitening issueThis afternoon around 2:45, ITMX started
ringing up at ~.9Hz for about a minute and
then
 44.png26.png 
  12519   Tue Sep 27 08:49:47 2016 SteveUpdateSUSseismic activity is upThe earth quake shook ITMX free for a 
short while.

 
 4.3mSaltonSee.pngITMXstuck.png 
  12520   Tue Sep 27 18:04:50 2016 LydiaUpdateSUSITMX slow channels down, ITMY diagonalization update[Teng, Lydia]

When we plugged in the back cables
yesterday on the whitening boxes after switching
 ITMY_diagsuccess.pdf 
  12521   Wed Sep 28 04:27:33 2016 ericqUpdateGeneralmucking aboutPMC was terribly misaligned. The PMCR camera
seems to have drifted somewhat off target
too, but I didn't touch it.
 alscheck.pdfweird_olpit.pdf 
  12522   Thu Sep 29 09:49:53 2016 ranaUpdateGeneralmucking aboutWith the WFS and OL, we never have figured
out a good way to separate pit and yaw. Need
to figure out a reference for up/down and
  
  12523   Thu Sep 29 16:19:29 2016 LydiaUpdateSUSFree swing eigenmodes[Lydia, Teng]

Motivated by the strange pitch/yaw
coupling behavior we ran into while doing
 05.pngSUS_eigenmodes.png 