Entry  Fri Sep 19 08:09:55 2008, steve, Update, PSL, psl HEPAs turned on 
I have just turned on the PSL HEPA filters at 60% operational speed.
Entry  Mon Sep 22 18:54:54 2008, Jenne, Update, PSL, PMC transfer functions with various brick-on-top configurations PMC_OLG_100Hz_to_100kHz.pngPMC_OLG_5kHz_to_30kHz.png
Attached below is a graphical summary of different things that I have tried putting on the PMC to reduce the noise in the loop.  The motivation behind these
measurements is the current inability here at the 40m to increase the UGF of the PMC.  This is part of a broader ISS loop/gain/noise problem that we are
having, which is causing Rob's locking efforts to have trouble.  (The ISS is next on the to-do list, after we find the best configuration for the PMC,
Entry  Mon Sep 22 21:30:06 2008, rana, Configuration, PSL, bad FSS DAQ.png
The MC refl power was going up and the FSS PC drive was so large that I had to turn up the FSS
common gain from 1.5 dB to 10.5 dB to get it to be better. Attached are the before (REF) and
after plots of frequency noise. Is the FSS gain really supposed to be 1.5 dB?? How did we gain
    Reply  Mon Sep 22 22:24:19 2008, rana, Configuration, PSL, bad FSS 
[quote]The MC refl power was going up and the FSS PC drive was so large that I had to turn up the FSS...[/quote]

Looks like I bumped the PS for the 21.5 MHz test setup and changed the supply voltage of the amplifier
Entry  Tue Sep 23 11:17:59 2008, steve, Update, PSL, PMC scattering spot rainbow.pngpmcclip.png
The PMC output side has a new madly scattering spot at chamfer 2 o'clock position
Entry  Tue Sep 23 15:28:06 2008, pete, Configuration, PSL, new 21.5 MHz FSS reference installed fss_ref_001.jpgfss_ref_013.jpg
The new 21.5 MHz FSS reference is now installed in the rack with the 7 Sorensen PS.  Both outputs give 18.7 dBm.  The MC seems happy.

Bob did the +24 V and +15 V hookups for the amp and the Wenzel oscillator respectively, off of the din strips on the right of the rack. 
Entry  Thu Sep 25 02:35:21 2008, rana, Summary, PSL, FAST is moving alot Untitled.pngUntitled.png
It looks like the FAST signal has started moving a lot - this is partly what inspired us to tune the SLOW loop.

Some of the spiking events happen when people go on the table or the MC loses lock. But at other times it just
Entry  Thu Sep 25 10:48:29 2008, Yoichi, Update, PSL, FSS calibration again PDTrans.pngPDHsignal.png
I did a calibration of the FSS error signal again with a different method.
This time, I swept the laser frequency with the NPRO PZT around a resonance.
The attached figures show the transmitted light power and the PDH error signal vs the applied voltage to the PZT.
Entry  Fri Sep 26 00:19:54 2008, Jenne, Update, PSL, Filter-action with the PMC 
Written, but not posted on 24Sept2008:

 
Entry  Fri Sep 26 18:35:17 2008, Jenne, Update, PSL, PMC filter is out for tuning 
The PMC's new Pomona Box filter is out for tuning.  I'd like to get the notch right on the 18.3kHz, rather than being off in 21.7kHz land.
Entry  Fri Sep 26 19:08:43 2008, rana, Configuration, PSL, Refcav Trans: PD + Camera + Dumps 
I went out to improve the Refcav trans path.

I removed all ND filters to get rid of the fringing. 
Entry  Mon Sep 29 01:19:40 2008, rana, Summary, PSL, Laser chiller running a little hot 
I looked at it some last night and my suspicion was the ISS. Whenever the ISS switch came on the FAST got a kick.

We should try to disable the MC locking and ISS and see if the FSS/PMC/MZ are stable this way. If so this may be
    Reply  Mon Sep 29 13:23:40 2008, rob, Summary, PSL, Laser chiller running a little hot 
[quote]I looked at it some last night and my suspicion was the ISS. Whenever the ISS switch came on the FAST got a kick.

We should try to disable the MC locking and ISS and see if the FSS/PMC/MZ are stable this way. If so this may be
Entry  Mon Sep 29 15:09:36 2008, steve, Update, PSL, almost 4 yrs plot of power & temps 3.5y.jpg
The water chiller is normally running 1.5 C warmer than the laser head temp.
When control room temp is stable and PEM-count_temp is stable we can expect the head temp to be stable 20.0 C
Entry  Mon Sep 29 17:53:33 2008, Yoichi, Update, PSL, ISS update IMG_1671.JPG
ISS has been saturating easily.
Today I opened the PSL enclosure to inspect the ISS box. Then I found that the sensor PD was disconnected from the box.
I don't know for how long it has been like this, but it is clearly bad.
Entry  Tue Sep 30 19:50:27 2008, Jenne, Update, PSL, Quicky Summary - more details later mcf.png
Quicky summary for now, more details later tonight / tomorrow morning:

PMC notch:  It's tuned up, but it is out, and it [COLOR=green][SIZE=4]is staying out[/SIZE][/COLOR].  It looks like the 18.3kHz junk isn't being helped
Entry  Wed Oct 1 02:47:53 2008, rana, Update, PSL, PSL ERR & LODET: Too much offset Untitled.png
Looks like there is an anomolous mixer offset correlated with the increase in the LO level. This may be leading to crazy offset locking in the FSS and too
much coupling from ISS to FSS.
Entry  Wed Oct 1 23:05:14 2008, Yoichi, Update, PSL, ISS RIN spectra RIN.pngOPLTF.png
Stefan, Yoichi

We took relative intensity noise (RIN) spectra of the ISS error point and the monitor PD (attm1).
Entry  Wed Oct 1 23:21:03 2008, Yoichi, Configuration, PSL, Reference cavity reflection camera 
I re-centered the reference cavity reflection camera, which has been mis-aligned for a while.
I also tweaked an input steering mirror to make the alignment better. This resulted in the increase of the transmission PD voltage
from 8V to 9V.
Entry  Fri Oct 3 15:09:58 2008, rob, Update, PSL, FAST/SLOW FASTSLOW.png

Last night during locking, for no apparent reason (no common mode), the PSL FAST/SLOW loop starting going just a little
nutz.  Attached is a two day plot.  The noisy period started around 11-ish last night.
Entry  Wed Oct 8 19:17:55 2008, Yoichi, Configuration, PSL, Laser power is slowly recovering 
This afternoon we (rich, steve, yoichi) shutdown the laser for the DC-DC converter installation. 
(we decided not to do so. Detail will be posted soon.)
After we turned on the laser again,the laser power has been recovering but very slowly.
Entry  Wed Oct 8 21:26:20 2008, Yoichi, Update, PSL, Attempt to replace the DC-DC converter (aborted) DCDC.JPGNPRO.JPG
Rich, Steve, Yoichi

We opened the MOPA box and inspected our NPRO.
Entry  Wed Oct 8 23:18:23 2008, Yoichi, Update, PSL, Correlation between the Sorensen switching noise and the FSS error signal PowerLineSpe.pngCoherence.png
I took some spectra and coherence function of the FSS error signal and the +24V Sorensen power line.
The first plot shows spectra of the two signals. Looks like Sorensen is not responsible for most of the lines
in the FSS error signal.
Entry  Mon Oct 13 11:32:50 2008, Yoixhi, Update, PSL, MOPA is in trouble now powerdrop.png
Steve, Alberto, Yoichi

A quick update.
Entry  Mon Oct 13 13:51:49 2008, pete, Configuration, PSL, attempt to measure FSS ref phase 
On Friday I began a measurement of the FSS reference phase.  The setup involves the following:
  + turn off the 166 MHz LO (top signal generator on 1Y2 rack)
  + bring FSS LO 21.5 MHz to the 166 MHz delay line phase shifter, and back out the phase shifter with a second length of cable
    Reply  Mon Oct 13 13:56:03 2008, Yoichi, Update, PSL, MOPA is not that much in trouble now 
The problem was found to be all to do with the shutter.
The shutter started to work again, after a while, apparently for no clear reason.
The alignment to the PA was actually not screwed, and the MOPA output is now slowly increasing.
Entry  Mon Oct 13 18:59:39 2008, Yoichi, Update, PSL, NPRO EMI and FSS error signal correlation IMG_1692.JPGSpectrum.pngSpectrumFar.pngCoherence.png
I made a simple loop antenna to measure the electro-magnetic inteference (EMI) around the master oscillator NPRO.

The first plot shows the comparison of the FSS error signal with the EMI measured when the antenna was put next to the NPRO (the MOPA box was opened).
Entry  Tue Oct 14 14:19:36 2008, pete, Configuration, PSL, FSS ref phase 
Today I made several measurements which should yield the optimized phase for the FSS 21.5 MHz reference.  I made two sets of measurements, using the 166
MHz phase delay shifter.  For each phase value I made 5 measurements of a 500 kHz injection into test2 at 1 Vpp,  with the 4195 spectrum analyzer on in1
with the high impedance probe, 51 points, a 10 kHz range.  It was surprisingly noisy.  I will make plots using matlab to find the maximum, and hope for
Entry  Tue Oct 14 19:24:34 2008, Yoichi, Configuration, PSL, FSS light power reduced 
Rana, Yoichi

To change the gain distribution in the FSS, Rana reduced the VCO power for the AOM to reduce the light incident to the reference cavity.
Entry  Wed Oct 15 17:40:50 2008, rana, Update, PSL, PMC Offset adjusted 
I set the PMC servo input offset: closed the MOPA shutter, zeroed the mixer output with the offset slider, 
relocked everything, and set the nominal to the new value of -6 V.
Entry  Wed Oct 15 22:07:52 2008, pete, Configuration, PSL, FSS ref phase measurements phasedelay.png
Optimizing the FSS LO/RF phase at 500 kHz, above the servo band, proved to be noisy and those measurements were useless.  Today I repeated
the measurement at 35 kHz and got good signal to noise.  I've attached a plot of the 35 kHz peak in dBm as measured at IN2 by SR785, with 
an injection into TEST2 at 35 kHz with 0.2 Vpp, as a function of delay in ns given by the delay phase shifter normally used by the 166 MHz.  
Entry  Thu Oct 16 09:44:49 2008, Yoichi, Update, PSL, Bad cable for FSS 
Yesterday arount 1:30PM, we lost the LO signal for the FSS.
I found it was caused by a bad cable connecting from the peter's RF oscillator box to the LO of the FSS.
I temporarily replaced it with a BNC cable of comparable length. 
    Reply  Thu Oct 16 09:47:49 2008, Yoichi, Configuration, PSL, FSS ref phase measurements 
[quote] I fit the bottom (quadratic) portion of this curve, and found an optimum delay of 25.8 ns, which can be implemented as 25.81 ns on the phase shift
box (25 + 1/2 + 1/4 + 1/16). [/quote]
Entry  Thu Oct 16 13:12:58 2008, pete, Configuration, PSL, phase between FSS reference outputs 
I verified the phase between the FSS reference outputs (used for LO and RF) using matched BNC cables.  I measured 0.95 degree (average of 12 scope measurements).
Entry  Thu Oct 16 16:26:26 2008, pete, Configuration, PSL, FSS phase matching cable installed 
RG 405 cable has a solid teflon dielectric, and a velocity factor of 0.69.  To get the 8.2 degrees of additional phase on the LO output at 21.5 MHz then
requires 22 cm of cable.  I made a cable that ended up being 21 cm long after I'd gained some experience putting on the connector.  It gives a phase difference
between LO and RF of about 10 degrees.  It is currently installed.
Entry  Mon Oct 20 20:50:09 2008, Yoichi, Configuration, PSL, FSS board chip replacement 
A quick update.

I changed two AD797s on the FSS board to AD829s to mitigate the 50MHz oscillation, which I plan to report later.
Entry  Tue Oct 21 16:17:45 2008, Yoichi, Update, PSL, AD797 Oscillation in the FSS board 
I checked each op-amp's output in the FSS board to see if any indication of slew-rate saturation can be found.
PA85, which was the most suspicious one, actually has a very large slew rate limit (1000V/usec).
Its output swing was about 5V/usec. So PA85 was ok in terms of slew rate.
Entry  Tue Oct 21 17:52:30 2008, rana, Summary, PSL, FSS Photo: early October fss.jpg
This is a photo of the FSS board before Yoichi did his surgery - it was taken with the D40 in macro mode, sitting on the big Gorilla pod.
Entry  Thu Oct 23 20:47:28 2008, pete, Configuration, PSL, FSS LO calibration for MEDM fss_lo_calibration.png
Today I took a quick series of measurements to calibrate the FSS LO power measurement in the MEDM.  This was done by using the spec.an. to measure the 21.5
MHz peak in dBm at the LO input to the FSS box on the PSL table, and recording the MEDM value, for attenuations applied at the FSS REF box output ranging
from -5 dBm to -30 dBm.  
Entry  Thu Oct 23 21:52:27 2008, Yoichi, Update, PSL, FSS UGF now 450kHz FSS_OPLTF.png
I measured the open loop transfer function of the FSS, for the first time after I mitigated the oscillation.
The attached plot shows the comparison of the OPLTF before and after the oscillation was mitigated.
Blue curves are when AD797 was oscillating, and the red ones are after AD797s were replaced by AD829s.
Entry  Thu Oct 23 23:09:18 2008, Yoichi, Update, PSL, MC UGF is now 75kHz MC_Loop.pngOldAndNew.png
I measured three loop transfer functions of the MC servo.
The blue curve in the first attachment is the overall open loop gain of the servo measured using 
the sum-amp A of the MC board (it is the sum-amp in the common part).
Entry  Fri Oct 24 11:21:26 2008, pete, Configuration, PSL, FSS LO input calibrated in dBm 
Based on the measurements described in my previous elog, I created a new calc record in the file /cvs/cds/caltech/target/c1psl/psl.db
 
[CODE]grecord(calc, "C1:PSL-FSS_LOCALC")
Entry  Mon Oct 27 14:48:27 2008, Yoichi, Configuration, PSL, EO shutter installed to the reference cavity 
I'm now preparing for cavity ring down measurements of the reference cavity.
An EOM for polarization rotation is installed between the two steering mirrors for the reference cavity.
The location is before the polarized beam splitter (used to pick-up the reflected light from the cavity) and 
Entry  Wed Oct 29 12:23:04 2008, Yoichi, Configuration, PSL, MZ alignment touched and the alarm level changed 
Since the MZ reflection is alarming all the time, I tried to improve the MZ alignment by touching the folding mirror.
I locked the X-arm and monitored the transmitted light power while tweaking the mirror alignment to ensure that the output beam pointing is not changed.
I changed the alignment only a little, almost like just touching the knob.
Entry  Mon Nov 3 09:59:47 2008, steve, Update, PSL, PSL HEPAs turned on 
The psl [COLOR=chocolate]enclosure[/COLOR] HEPAs were tuned on.

Loose paper drawing was found on the psl inside [COLOR=chocolate]shelf[/COLOR].
Entry  Tue Nov 4 17:58:42 2008, Alberto, DAQ, PSL, MC temperature sensor 
I added a channel for the temperature sensor on the MC1/MC3 chamber: C1:PSL-MC_TEMP_SEN.
To do that I had to reboot the frame builder. The slow servo of the FSS had to get restarted, the reference cavity locked and so the PMC and MZ.
Entry  Fri Nov 7 18:38:19 2008, Alberto, DAQ, PSL, MC temperature sensor hooked up 
Alberto, Rana,
we found that the computer handling the signals from ICS-110B was C1IOVME so we restarted it. We changed the name of the channel to C1:PEM_TEMPS and the
number to 16349. We tracked it up to the J14 connector of the DAQ.
Entry  Wed Nov 12 11:14:59 2008, Caryn, DAQ, PSL, MC temp sensor hooked up incorrectly 
MC Temperature sensor was not hooked up correctly. It turns out that for the 4 pin LEMO connections on the DAQ like J13, J14, etc. the channels correspond
to horizontal pairs on the 4 pin LEMO. The connector we used for the temp sensor had vertical pairs connected to each BNC which resulted in both the differential
pairs on J13 being read by the channel. 
Entry  Fri Nov 14 19:20:42 2008, Yoichi, Update, PSL, Reference cavity ring down RingDown.png
Thanks to Bob making the high-voltage BNC cables for the HV pulse generator, I was able to operate the EOM in front of
the reference cavity.
    Reply  Fri Nov 14 20:35:47 2008, rana, Update, PSL, Reference cavity ring down 
To make the DEI pulser make a fast pulse on the EO shutter EOMs, we had to make sure:

1) the cable had a high voltage rated dielectric. cheap dielectrics show the 'corona'