40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 255 of 355  Not logged in ELOG logo
ID Date Author Type Categoryup Subject
  2663   Tue Mar 9 09:04:20 2010 steveUpdatePEMkeep vacuum chamber closed

They are sandblasting at CES: our particle counts are very high. DO NOT OPEN CHAMBER!

Attachment 1: sandblasting.jpg
sandblasting.jpg
  2665   Tue Mar 9 12:06:53 2010 rana, JenneUpdatePEMStyrofoam Cooler on the Seismos

Looks like the GUR2_X signal is bad. Jenne says that we need to center it mechanically before the signals will become useful again. Maybe Steve will do this - instructions are in the manuel ?

  2666   Wed Mar 10 15:02:58 2010 ranaUpdatePEMStyrofoam Cooler on the Seismos

This is the spectra and coherence from a quiet time last night. I've lowered the Guralp cal by a factor of 2 to account for the fact that the gain in the breakout box is actually 20 and not 10 as I previously said.

The AD620 stage in the front part has a gain of 10 and then there's a single-to-differential stage in the output which gives us a gain of 2. The DTT cal in counts is now 3.8e-9 (m/s)/count.

The second plot shows the Guralp and Ranger signals at the ADC input (converted from counts to Volts for usefulness). The thick grey line is the expected noise of the Guralp breakout box

(mainly the AD620) propagated to the ADC (via multiplication by 2). It looks like the preamp board should not be a problem as long as we can reach the AD620 limit.

So the excess noise in the Guralp is not the fault of the preamp, but more likely the mounting and insulation of the seismometers.

Attachment 1: Seismic-Ref_100310.png
Seismic-Ref_100310.png
Attachment 2: Untitled.gif
Untitled.gif
  2671   Sat Mar 13 21:30:22 2010 ranaUpdatePEMStyrofoam Cooler on the Seismos

Untitled.png

 

  2673   Mon Mar 15 09:43:47 2010 steveUpdatePEMmore sandblasting today

Do not open IFO vacuum envelope today! They are sandblasting again at CES

  2674   Mon Mar 15 16:39:36 2010 steveUpdatePEMGuralp2 centered

Quote:

Untitled.png

 

Guralp 2 centered.The mass position offsets are: E-W 0.05V, N-S 0V, Z 0.4V

Guralp 1: E-W -0.1V, N-S -0.25V, Z 0V measured, not adjusted

The GUR2_X channel has an offset. See plot below when seismometers are disconnected. This offset has to be removed.

NOTE: this huddle is on bad-soft ground-lenoleum tile from prehistoric Flintstone age

 

Attachment 1: grlp2ffst.jpg
grlp2ffst.jpg
  2678   Thu Mar 18 08:49:51 2010 steveUpdatePEMGuralp2 centered again ?

 

Someone adjusted the Guralp2 mass position last night??

NO

Attachment 1: grlp2xadj.jpg
grlp2xadj.jpg
  2685   Fri Mar 19 18:00:14 2010 jenneUpdatePEMGuralp2 centered again

[Jenne, Sanjit]

It looks like Steve used a GND-12V supply to power the Guralp through the little breakout box (the box is for checking the centering of the mass).  This is BAD.  The Guralps want +/- 12V.

We centered all of the channels on Gur2, and checked the channels on Gur1, so we'll see how they're feeling after a while.

  2686   Fri Mar 19 21:15:33 2010 ranaUpdatePEMGuralp2 centered again

This trend of the last 200 days shows that GUR2 has been bad forever...until now anyways.

Attachment 1: Untitled.png
Untitled.png
  2687   Fri Mar 19 23:03:41 2010 ranaUpdatePEMGuralp2 centered again

I went and double-checked and aligned the styrofoam cooler at ~5:00 UTC. It was fine, but we really need a better huddling box. Where's that granite anyway?

Here's the new Huddle Test output. This time I show the X-axis since there's some coherence now below 0.1 Hz.

You'll also notice that the Wiener filter is now beating the FD subtraction. This happened when I increased the # of taps to 8000. Looks like the noise keeps getting lower as I increase the number of taps, but this is really a kind of cheat if you think about it carefully.

Attachment 1: huddlez.png
huddlez.png
  2689   Sun Mar 21 19:25:29 2010 ranaUpdatePEMGuralp2 centered again

From this morning, now in calibrated units, and with the Güralp self noise spec from the Güralp manual.

Attachment 1: huddlez.png
huddlez.png
  2694   Mon Mar 22 11:37:09 2010 steveUpdatePEMjackhammering

There was more jackhammering this morning just about 20 ft north-west of the beamsplitter chamber, outside.

Attachment 1: jackhammering2.jpg
jackhammering2.jpg
  2701   Tue Mar 23 10:11:13 2010 steveUpdatePEM small air cond unit failed yesterday

Old control room air condition failed yesterday around noon. It was blowing 80-85F hot air for about 2-3 hours at racks 1Y4-7 and the entry room 103

Attachment 1: 20datm.jpg
20datm.jpg
  2713   Thu Mar 25 09:07:08 2010 steveUpdatePEMGuralp2 x problem is back

Quote:

I went and double-checked and aligned the styrofoam cooler at ~5:00 UTC. It was fine, but we really need a better huddling box. Where's that granite anyway?

Here's the new Huddle Test output. This time I show the X-axis since there's some coherence now below 0.1 Hz.

You'll also notice that the Wiener filter is now beating the FD subtraction. This happened when I increased the # of taps to 8000. Looks like the noise keeps getting lower as I increase the number of taps, but this is really a kind of cheat if you think about it carefully.

 The same thing happening again.  The intermittent offset upstream of the seismometer that never got fixed.

The granite plate and ball bearings are in. I will place seismometers on it.

Attachment 1: grlp2xproblm.jpg
grlp2xproblm.jpg
  2743   Wed Mar 31 16:31:44 2010 steveUpdatePEMGuralp interface box turned off

Quote:

Quote:

I went and double-checked and aligned the styrofoam cooler at ~5:00 UTC. It was fine, but we really need a better huddling box. Where's that granite anyway?

Here's the new Huddle Test output. This time I show the X-axis since there's some coherence now below 0.1 Hz.

You'll also notice that the Wiener filter is now beating the FD subtraction. This happened when I increased the # of taps to 8000. Looks like the noise keeps getting lower as I increase the number of taps, but this is really a kind of cheat if you think about it carefully.

 The same thing happening again.  The intermittent offset upstream of the seismometer that never got fixed.

The granite plate and ball bearings are in. I will place seismometers on it.

 

Attachment 1: grlpntrfcbxoff.jpg
grlpntrfcbxoff.jpg
  2764   Mon Apr 5 01:02:07 2010 ranaUpdatePEMGuralp interface box turned off

I was checking into the Guralp situation today. I put the rubber balls underneath the granite block (the Q is too high), but found unfortunately that Jenne's styrofoam box is too short to cover the Guralps on top of the granite. If the box was skinny enough to fit on the block or taller by ~6 inches, it would be perfect. We need some new Seismo boxes.

 

Here's the story of the Gur2 noise so far. We need to pull out and repair the breakout box.

1) At some point we noticed that the Guralp2 X channel was behaving badly.

2) Steve tried recentering with just a +12V supply - this didn't work. Jenne then centered it using the +/- 12V supply. This was OK.

3) Around noon on March 24, the channel 'goes bad' again.

4) On the afternoon of the 25th, most of the channels go to zero, but the GUR2X channel stays bad. There's NO ENTRY in the elog about this. This is UNACCEPTABLE. Apparently, the seismometers were disconnected without shutting off the power to the box. You MUST elog everything - otherwise, go home and sit on your hands.

5) On the evening of the 31st, Steve turns off the Guralp breakout box. From the trend, you can see that the signals all go to zero at that time.

6) From then until today, there is no noise in the GUR2X channel. From these tests we can guess that the problem is in the GUR2X channel of the breakout box, but not in the AA Chassis or the ADC, since those showed no excess noise with the box turned OFF. Its hard to be sure without elog entries, but I assume that 3/25-3/31 was a 'seismometer disconnected', but 'box on' state.

Attachment 1: Untitled.png
Untitled.png
  2765   Mon Apr 5 08:43:48 2010 steveUpdatePEMearthquake mag 7.2

Large earthquake shakes Baja California, Mexico and 6 over Magnitude 5 aftershakes follow.  The frontend computers are still down since Friday.

Attachment 1: eq7.2.jpg
eq7.2.jpg
  2778   Wed Apr 7 09:00:01 2010 steveHowToPEMprepare to open chamber

In order to minimize the diffusion of more dust particles into the vented IFO vacuum envelope

BEFORE opening chamber:

-Have a  known plan,

-Heavy 1" thick door requires 3 persons- of  one experienced and one certified crane operator and steel tow safety shoes

-Block IFO beams, be ware of experimental set up of other hazards: 1064,  visible or new-special installation

- Look at the particle counter, do not open above 6,000 particles of 0.5 micron. Construction activities are winding down. See  plot of 35 days since we  vented.

-Have clean door stand for heavy door, covered with merostate at the right location and dry-clean screws for light covers,

-Prepare lint free wipers for o-rings,(no solvent on o-ring!) Kimwipes for outside of chamber and metal covers, methanol and powder free gloves

-Wipe with wet Kimwipe-tissue of methanol around the door, chamber of interest and o-ring cover ring

-Cut door covering merostate and tape it into position,..if in place...check  folded-merostate position, if dusty... replace it

-Is your cleanroom garment clean?.......if in doubt ....replace it

-Keep surrounding area free and clean

-Make sure that HEPAs are running: PSL-enclosure, two mobile units and south end flow banch

-Check the tools: are they really clean? wipe it with wet Kimwipe, do you see anything on the Kimwipe?

 

-You are responsible to close chamber ASAP with light door or doors as you finished for the day.

Merostate cover down is appropriate during daily brakes.

Attachment 1: 0.5micron.jpg
0.5micron.jpg
  2802   Fri Apr 16 17:26:23 2010 JenneUpdatePEMGuralp Breakout Box pulled

I pulled the Guralp breakout box from the rack, and it's sitting on the EE bench here.  The game plan is to check out the Gur2X channel.  

Rana and Steve have been investigating, and found that the X channel has been funky (which has been known for ~a month or two) when the seismometer has been plugged in, and also when the seismometers have been unplugged, but the box is left on.  The funkyness goes away when the box is turned off.  Since it's not there when the box is off, it seems that it's not a problem with the cable from the box to the ADC, or in the ADC channel.  Since it is there when the box is on, but the seismometer is unplugged, it's clear that it's probably in the box itself.  

Preliminarily, I've connected a set of BNC clipdoodles to the input testpoints, and another set to the output.  They're both connected to a 'scope (which is on it's battery so it's not connected to any Ground), and when I tap on the circuit board the input trace is totally unchanged, but the output trace goes kind of crazy, and gets more fuzzy, and picks up a DC offset.  Koji is concerned that some of the big capacitors may have an iffy connection to the board. 

Investigations will continue Monday morning. 

  2811   Tue Apr 20 00:32:30 2010 JenneUpdatePEMGuralp Breakout Box put back

Quote:

I pulled the Guralp breakout box from the rack, and it's sitting on the EE bench here.  The game plan is to check out the Gur2X channel.  

Rana and Steve have been investigating, and found that the X channel has been funky (which has been known for ~a month or two) when the seismometer has been plugged in, and also when the seismometers have been unplugged, but the box is left on.  The funkyness goes away when the box is turned off.  Since it's not there when the box is off, it seems that it's not a problem with the cable from the box to the ADC, or in the ADC channel.  Since it is there when the box is on, but the seismometer is unplugged, it's clear that it's probably in the box itself.  

Preliminarily, I've connected a set of BNC clipdoodles to the input testpoints, and another set to the output.  They're both connected to a 'scope (which is on it's battery so it's not connected to any Ground), and when I tap on the circuit board the input trace is totally unchanged, but the output trace goes kind of crazy, and gets more fuzzy, and picks up a DC offset.  Koji is concerned that some of the big capacitors may have an iffy connection to the board. 

Investigations will continue Monday morning. 

 The Guralp Box appears to be back in working order.  It's reinstalled with the 2 seismometers plugged in.

In order:

* Koji suggested retouching the through-board solder joints on the broken channel (EW2 = Gur2X) with a bit of solder to ensure the connections were good.  Check.

* "C7", one of the giant 1uF capacitors on each channel is totally bypassed, and since that was one of the original suspects, Rana removed the (possibly) offending capacitor from EW2.

* Rana and I isolated the craziness to the final differential output stage.  We tried each of the testpoints after the individual gain / filter stages, and found that the signals were all fine, until after the output stage.

* I started to remove the resistors in the output stage (with the plan to go through the resistors, capacitors, and even the amplifier chip if neccessary), and noticed that 2 of the 1k resistors came off too easily, as if they were just barely connected in the first place.  After replacing only the 4 1k resistors, the craziness seemed to be gone.  I poked and gently bent the board, but the output wouldn't go crazy.  I declared victory.

* I reinstalled the box in its normal spot, and put Gur2 (which had been out by the bench for use as a test signal) back next to the other seismometers.  We are in nominal condition, and should be able to do a huddle test this week.

 

I looked at the time traces of all the seismometer channels, and they all looked good.  I'll put a spectra in in the morning....I'm too impatient to wait around for the low frequency FFTs.

 

Attached are the before and after pictures of the output stage of EW2 / Gur2X.  The "before" is the one with the OUT+ and OUT- words upsidedown.  The "after" picture has them right side up.

Attachment 1: Gur2X_before_20Apr2010.jpg
Gur2X_before_20Apr2010.jpg
Attachment 2: Gur2X_after_replace_100k_resistors_20Apr2010.jpg
Gur2X_after_replace_100k_resistors_20Apr2010.jpg
  2825   Wed Apr 21 15:53:57 2010 JenneUpdatePEMSeismometers now on the granite slab

The 3 seismometers are now on the granite slab.  The Ranger is now aligned with the Xarm (perpendicular to the Mode Cleaner) since that's the only way all 3 would fit on the slab.

Attachment 1: P4210048.JPG
P4210048.JPG
  2831   Thu Apr 22 09:03:54 2010 steveOmnistructurePEMaircondition can not be turned off

Koji and I wanted to turn off the IFO-room AC so the wind would not blow on MC1-3. We could not. The switches were probably bypassed when the power transformer was replaced at the last scheduled power outage.

  2833   Thu Apr 22 20:28:40 2010 JenneUpdatePEMSeismometer Noise Characteristics

Quote from elog 2237, 11 Nov 2009:

The attached plot shows the spectra of the 3 Z axes of the 3 seismometers we have (this data is from ~20Aug2009, when the Ranger was in the Z orientation) in Magenta, Cyan and Green, and the noise of each of the sensors in Red, Blue and Black.  The noise curves were extracted from the spectra using the Huddle Test / 3 Corner Hat method.  The Blue and Black traces which are just a few points are estimates of the noise from other spectra.  The Blue points come from the Guralp Spec Sheet, and the Black comes from the noise test that Rana and I did the other day with the Ranger (elog 2223).  

I'm not really happy with the black spectra - it looks way too high.  I'm still investigating to see if this is a problem with my calibration/method....

 So, as it turns out (surprise), I'm a spaz and forgot a 2*pi when calibrating the Guralp noise spectra from the spec sheet.  I noticed this when redoing the Huddle Test, and comparing my Spec Sheet Guralp noise with Rana's, which he shows in elog 2689.  When going from m/s^2, the units in the spec sheet, I just tilted the line by a factor of frequency.  Koji pointed out that I needed a factor of 2*pi*f.  That moves the Guralp spec line in the plot in elog 2237 (to which this entry is a reply) down by ~6, so that my measured noise is not, in fact, below the spec.  This makes things much more right with the world.

In other news, I redid the Huddle analysis of the 2 Guralp seismometers, ala Rana's elog 2689. The difference is now we are on the granite slab, with soft rubber feet between the floor and the granite.  We have not yet cut holes in the linoleum (which we'll do so that we're sitting directly on the 40m's slab).

Rana> this seems horrible. Its like there's a monster in there at 6-7 Hz! Either the seismos are not centered or the rubber balls are bad or Steve is dancing on the granite slab again.

 

Attachment 1: Gur1_Gur2_noise.png
Gur1_Gur2_noise.png
  2842   Mon Apr 26 11:13:28 2010 steveOmnistructurePEMAC on/off is working

Quote:

Koji and I wanted to turn off the IFO-room AC so the wind would not blow on MC1-3. We could not. The switches were probably bypassed when the power transformer was replaced at the last scheduled power outage.

 There is one three position manual/off/auto switch next to the filter for each unit at CES. They have to be in AUTO position when we want to turn AC on/off from the lab.

  2889   Thu May 6 18:25:20 2010 JenneUpdatePEMAcc power supply turned on

The accelerometer power supply / preamp board has been OFF because of exciting new accelerometer measurements.  It's now on, so watch out and make sure to turn it back off before plugging / unplugging accelerometers.

  2890   Thu May 6 18:43:58 2010 ranaUpdatePEMNoise floor of the Ranger Seismometer

I added a noise model of the SR560 to the LISO opamp.lib. This assumes you're using it in G=100, low-noise mode. The voltage noise is correct, but I had to guess on the current noise because I didn't measure it before. Lame.

This can be compared with the noise that we measure when locking it down...

Attachment 1: ranger.pdf
ranger.pdf
  2892   Thu May 6 19:51:22 2010 JenneUpdatePEMNew 'Seismic Spectrum of the 40m'

For reasons unknown, the seismic spectra posted above Rosalba has been wrong since ~January when it was first posted.  The noise that we were claiming was waaaay lower than is really possible.

Rana and I checked the calibrations, and the numbers in DTT for the Ranger and the Guralp are correct (it's unknown what was being used at the time of the bad plot) - Cal for the Guralp is 3.8e-9 m/s, and for the Ranger is 1.77e-9 m/s.

Something is funny with the accelerometer calibration.  Hopefully Kevin's investigation will sort it out.  Their calibration used to be 1.2e-7 m/s^2 , but it was changed to be 7e-7 m/s^2 to match the noise level of the accelerometers with the seismometers at ~10Hz. We need to go through the calibration carefully and figure out why this is!

Posted above Rosalba for easy reference, and attached below, is the new seismic spectra.  The black trace is when the Ranger's mass is locked down, and the teal circle markers indicate the Guralp Spec-Sheet Noise Floor.

** Rana says> the y-axis in Jenne's plot is (m/s)/sqrt(Hz). The Guralp has a velocity readout bandwidth of 0.03-40 Hz, so we would have to modify the calibration to make it right in those frequencies. I believe the Ranger cal has the correct poles in it. The huge rise at low frequencies is because of the 1/f noise of the SR560.

Attachment 1: SeisRef_6May2010_AccelCalFudged.png
SeisRef_6May2010_AccelCalFudged.png
  2934   Fri May 14 16:19:22 2010 JenneUpdatePEMGuts of a Guralp

[Jenne, Rana]

We took apart and examined one of the Guralp seismometers this afternoon.  For the most part we think we understand how it works. The horizontal sensors are a little more confusing, since we didn't end up finding the moving masses.  The vertical sensor is a flat rectangle, hinged at one edge.  There are capacitive sensors above and below the rectangle.  The hinged end is connected to a leaf spring. 

The PCBs are packed full of old-school 80's components.  We probably need an actual schematic to figure out where the preamp circuit is, which is what we'd want to think about fitzing with, if we were to try to improve the noise of the seismometer.  For now, we put it all back together, and back out on the granite slab. 

There was a wee bit of confusion when putting the N/S marker-spikes back on as to where they should go.  The solution is that the handle of the seismometer is aligned with the North/South axis, so the spikes should be aligned with the handle.  The lid of the seismometer is uniquely aligned to the stuff inside by the ribbon cable connector, as well as the holes in the lid for accessing the centering potentiometers.  So, align the lid to the pots, and then align the spikes to the handle.

Photos are on Picasa.

  2972   Mon May 24 07:53:57 2010 steveConfigurationPEMair cond. just turned ON

IFO room temp 27.5C , Please remember to turn AC back on !

Attachment 1: acoff.jpg
acoff.jpg
  3002   Thu May 27 23:59:54 2010 ranaUpdatePEMNew Foam Box installed

Valera and I put the 2 Guralps and the Ranger onto the big granite slab and then put the new big yellow foam box on top of it.

There is a problem with the setup. I believe that the lead balls under the slab are not sitting right. We need to cut out the tile so the thing sits directly on some steel inserts.

You can see from the dataviewer trend that the horizontal directions got a lot noisier as soon as we put the things on the slab.

Attachment 1: Untitled.png
Untitled.png
  3003   Fri May 28 00:40:53 2010 ranaUpdatePEMDAQ down

 Although trends are available, I am unable to get any full data from in the past (using DTT or DV). I started the FB's daqd process a few times, but no luck. 

I blame Joe's SimPlant monkeying from earlier today for lack of a better candidate. I checked and the frames are actually on the FB disk, so its something else.

  3005   Fri May 28 10:44:47 2010 josephbUpdatePEMDAQ down

Quote:

 Although trends are available, I am unable to get any full data from in the past (using DTT or DV). I started the FB's daqd process a few times, but no luck. 

I blame Joe's SimPlant monkeying from earlier today for lack of a better candidate. I checked and the frames are actually on the FB disk, so its something else.

 I tried running dataviewer and dtt this morning.  Dataviewer seemed to be working.  I was able to get trends, full data on a 2k channel (seismic channels) and full data on a 16k channel (C1:PEM-AUDIO_MIC1)  This was tried for a period 24 hours a go for a 10 minute stretch.

I also tried dtt and was able to get 2k and 16k channel data, for example C1:PEM-AUDIO_MIC1.  Was this problem fixed by someone last night or did time somehow fix it?

  3006   Fri May 28 11:26:35 2010 JenneUpdatePEMNew Foam Box installed

Quote:

Valera and I put the 2 Guralps and the Ranger onto the big granite slab and then put the new big yellow foam box on top of it.

There is a problem with the setup. I believe that the lead balls under the slab are not sitting right. We need to cut out the tile so the thing sits directly on some steel inserts.

You can see from the dataviewer trend that the horizontal directions got a lot noisier as soon as we put the things on the slab.

 You'll have to ask Steve how deep he cut, but the tile is cut around the lead balls, so they are not sitting on the linoleum.  They might just be sitting on the concrete slab, or whatever Steve found underneath the tile, instead of fancy steel inserts, but at least they're not on the tile.  I don't know why things got noisier though...

  3017   Sun May 30 17:51:04 2010 kiwamuHowToPEMAllegra dataviewer

I found the dataviewer didn't work only on Allegra. This thing sometimes happened as described in the past entry.

I rebooted Allegra, then the problem was fixed.

 

  3018   Sun May 30 22:18:49 2010 ranaUpdatePEMGranite slab w/ lead balls is so far a flop

The seismometers showed an increased noise in the Y-direction when put on top of the granite slab. By tapping the slab, you can tell that its really a mechanical resonance of the lead balls + granite system at ~15-20 Hz.

I tried new balls, flipping the slab upside down, and sitting on the slab for awhile. None of this changed the qualitative behavior, although each of the actions changed the resonance frequencies by several Hz.

I have removed the granite/balls and put the seismometers back on the linoleum floor. The excess noise is gone. I have put the new big box back on top of them and we'll see how the data looks overnight.

 

I expect that we should remove the linoleum in a wider area and put the seismometers directly on the floor.

  3022   Mon May 31 22:52:57 2010 ranaUpdatePEMGranite slab w/ lead balls is so far a flop

This plot shows the noise with the box on, but no granite. We're still pretty far off from the Guralp data sheet.

Untitled.png

I implemented software rotation in the huddle subtraction as Valera suggested and it works much better. The two plots below show the before and after. So far this is just 2 deg. of rotation around the z-axis. I'm assuming that aligning the seismometers vertically via bubble level is good enough for the z-axis, but I haven't calibrated the bubble yet.

huddlez.pnghuddlez.png

The residual slope is now suspiciously smooth. I somehow suspect that our readout electronics can still be responsible. We need to hook up a 9V battery to the input terminals to check it out. Its a little steeper than 1/f and I thought that we had exonerated the Guralp breakout box in the past, but now I'm not so sure. I'll let Jenne comment on that.

I also noticed that we have not yet divided by sqrt(2) to account for the fact that we are subtracting 2 seismometers. In principle, an unbiased estimate of the single seismometer noise will be lower by sqrt(2) than the green curve.

  3024   Tue Jun 1 11:47:14 2010 steveUpdatePEMlead balls on concrete

Quote:

Valera and I put the 2 Guralps and the Ranger onto the big granite slab and then put the new big yellow foam box on top of it.

There is a problem with the setup. I believe that the lead balls under the slab are not sitting right. We need to cut out the tile so the thing sits directly on some steel inserts.

You can see from the dataviewer trend that the horizontal directions got a lot noisier as soon as we put the things on the slab.

 The tiles were cut out in 1.5" ID circle to insure that the 7/16" OD lead balls would not touch the tiles on Wednesday, May 26, 2010

Granite surface plate specifications: grade B, 18" x 24" x 3" , 139 lbs

These balls and granite plate were removed by  Rana in entry log #3018 at 5-31-2010

Attachment 1: P1060269.JPG
P1060269.JPG
  3025   Tue Jun 1 15:51:42 2010 steveUpdatePEMfoam box for good thermal stability

This box was made  to provide good thermal stability for seismometer calibration. There is an inner solid shell of 0.064" Al box that is covered by 2" insulation

inside and outside. The polystyrene foam is "CertiFoam 25 SE " from McMasterCarr #9255K3.

Attachment 1: albx2.PDF
albx2.PDF
  3037   Wed Jun 2 18:09:32 2010 steveUpdatePEMseismometers off of linoleum floor

Steve for Nancy,

Seismometer interface box ac power was turned off, Guralps disconnected and moved. Ranger locked, moved and released. Nancy will describe the rest soon.

 

  3039   Wed Jun 2 21:21:43 2010 steveUpdatePEMseismometers off of linoleum floor

Quote:

Steve for Nancy,

Seismometer interface box ac power was turned off, Guralps disconnected and moved. Ranger locked, moved and released. Nancy will describe the rest soon.

The flattened lead balls were checked for their heights by the calliper, and were all in the range of 9.50 to 9.70 mm.

The rechecking was done by using these balls between two aluminium plates and checking their levelling. When confirmed this, we proceeded to install the balls(no more balls :P ) in their place.

The Guralps were switched off by switching off the power supply to them. The ranger mass was clamped in order to be able to move it. This can be undone by rotating the transport screw counter-clockwise.

We installed the flattened lead ballsin the space made for them. The granite was then placed on it with the help of many other people in the lab.  It was lowered by hanging it on two straps held by people , and then placed in the space marked for it.

Did we then turn on the seismometers? Did we release the locking screw on the Ranger? What happened to Bat-Boy??? Since they make a good mystery I will choose to leave them out of my elog entry.

  3047   Thu Jun 3 22:17:05 2010 ranaUpdatePEMseismometers off of linoleum floor

 At ~2350 UTC on June 2, the seismometers were turned off. After the granite slab was repositioned with the new lead, the Ranger was turned on, but not the Guralps.

Now, after ~24 hours, I have put the Guralps onto the granite and turned them on. During this off time, the input channels should be ADC noise limited (or perhaps limited by the INA134 differential receiver chips inside of the Sander Liu AA chassis). The following plot shows that this noise level is ~0.8 uV/rHz and then rising like ~1/sqrt(f) below 5 Hz:

Untitled.png

I checked the slab again by whacking it. It still rings with a Q of several, so I think we need to make the lead flatter. There should barely be any room between the granite and the linoleum.

UPDATE:

I guessed that it should be possible to make the slab-to-floor coupling better with flatter lead (Brian Lantz suggested to use lead sheets). So I removed my booties and jumped up and down on the granite several times. Because of my soft sole shoes, I was able to make an impulsive impact without shattering the granite. The effect of the stomping was pretty dramatic - the horizontal resonance frequency has gone up by a factor of 2. The red trace shows the new TF after the stomping:

a.png

 And the resulting spectrum is here too. As you can see, there is no excess between the Ranger and the Guralps until ~50 Hz where the mechanical resonance in the short direction (non-MC dir) takes over.

Untitled.png

 So, the lesson for next time is to flatten the balls a little more. I leave it to Nancy to calculate the horizontal resonant frequencies of this lead/granite combo to see if it matches with our measurements.

  3050   Fri Jun 4 23:52:57 2010 ranaUpdatePEMseismometers off of linoleum floor

Untitled.pnghuddlez.png

For the huddle test, I have updated the code to divide the residual by sqrt(2) because of the assumption of equal noise from the 2 Guralps. We would have to multiply this trace by sqrt(2) to compare with the previous results.

Now the question is, how do I add a low noise ~50 mV offset to the front of the Guralp breakout box to test for the noise of the box?

  3056   Tue Jun 8 18:39:36 2010 ranaUpdatePEMDAQ down

As before, I am unable to get data from the past. With DTT on Allegra I got data from now, but its unavailable from 1 hour ago. Same problem using mDV on mafalda. I blame Joe again - or the military industrial complex.

 

Quote:

Quote:

 Although trends are available, I am unable to get any full data from in the past (using DTT or DV). I started the FB's daqd process a few times, but no luck. 

I blame Joe's SimPlant monkeying from earlier today for lack of a better candidate. I checked and the frames are actually on the FB disk, so its something else.

 I tried running dataviewer and dtt this morning.  Dataviewer seemed to be working.  I was able to get trends, full data on a 2k channel (seismic channels) and full data on a 16k channel (C1:PEM-AUDIO_MIC1)  This was tried for a period 24 hours a go for a 10 minute stretch.

I also tried dtt and was able to get 2k and 16k channel data, for example C1:PEM-AUDIO_MIC1.  Was this problem fixed by someone last night or did time somehow fix it?

  3057   Tue Jun 8 20:52:25 2010 josephbUpdatePEMDAQ up (for the moment)

As a test, I did a remote reboot of both Megatron and c1iscex, to make sure there was no code running that might interfere with the dataviewer.  Megatron is behind a firewall, so I don't see how it could be interfering with the frame builder.  c1iscex was only running a test module from earlier today when I was testing the multi-filter matrix part.  No daqd or similar processes were running on this machine either, but it is not behind a firewall at the moment. 

Neither of these seemed to affect the lack of past data.  I note the error message from dataviewer was "read(); errno=9".

Going to the frame builder machine, I ran dmesg.  I get some disturbing messages from May 26th and June 7th. There are 6-7 of these pairs of lines for each of these days, spread over the course of about 30 minutes.

Jun 7 14:05:09 fb ufs: [ID 213553 kern.notice] NOTICE: realloccg /: file system full

Jun 7 14:11:14 fb last message repeated 19 times

There's also one:

Jun 7 13:35:14 fb syslogd: /usr/controls/main_daqd.log: No space left on device

I went to /usr/controls/ and looked at the file.  I couldn't read it with less, it errored with Value too large for defined data type.  Turns out the file was 2.3 G.  And had not been updated since June 7th.  There were also a bunch of core dump files from May 25th, and a few more recent.  However the ones from May 25th were somewhat large, half a gig each or so.  I decided to delete the main_daqd.log file as well as the core files.

This seems to have fixed the data history for the moment (at least with one 16k channel I tested quickly). However, I'm now investigating why that log file seems to have filled up, and see if we can prevent this in the future.

Quote:

As before, I am unable to get data from the past. With DTT on Allegra I got data from now, but its unavailable from 1 hour ago. Same problem using mDV on mafalda. I blame Joe again - or the military industrial complex.

 

Quote:

Quote:

 Although trends are available, I am unable to get any full data from in the past (using DTT or DV). I started the FB's daqd process a few times, but no luck. 

I blame Joe's SimPlant monkeying from earlier today for lack of a better candidate. I checked and the frames are actually on the FB disk, so its something else.

 I tried running dataviewer and dtt this morning.  Dataviewer seemed to be working.  I was able to get trends, full data on a 2k channel (seismic channels) and full data on a 16k channel (C1:PEM-AUDIO_MIC1)  This was tried for a period 24 hours a go for a 10 minute stretch.

I also tried dtt and was able to get 2k and 16k channel data, for example C1:PEM-AUDIO_MIC1.  Was this problem fixed by someone last night or did time somehow fix it?

 

  3060   Wed Jun 9 19:47:08 2010 nancyUpdatePEMlead balls on concrete

Quote:

Quote:

Valera and I put the 2 Guralps and the Ranger onto the big granite slab and then put the new big yellow foam box on top of it.

There is a problem with the setup. I believe that the lead balls under the slab are not sitting right. We need to cut out the tile so the thing sits directly on some steel inserts.

You can see from the dataviewer trend that the horizontal directions got a lot noisier as soon as we put the things on the slab.

 The tiles were cut out in 1.5" ID circle to insure that the 7/16" OD lead balls would not touch the tiles on Wednesday, May 26, 2010

Granite surface plate specifications: grade B, 18" x 24" x 3" , 139 lbs

These balls and granite plate were removed by  Rana in entry log #3018 at 5-31-2010

 I tried to calculate the frequency of resonance using Rayleigh's method.  approximated the geometry of lead to be that of a perfect cylinder, and  the deformation in the lead by the deflection in a cantilever under  a shear strain.

this rough calculation gives an answer of 170Hz and depends on the dimensions of each lead, number of leads, and mass of the granite. But the flaw pointed out is that this calculation doesnot depend on the dimension of the granite slab, nor on the exact placing of the lead spheres with respect toteh COM of the slab.

I will put up the calculations details later, and also try to do a FEM analysis of the problem.

 

BTW, latex launched this new thing for writing pdfs. doesnot require any installations.  check  http://docs.latexlab.org

  3062   Thu Jun 10 07:53:14 2010 AlbertoUpdatePEMLaTeXlabs

Quote:

BTW, latex launched this new thing for writing pdfs. doesnot require any installations.  check  http://docs.latexlab.org

 so cool!

  3139   Tue Jun 29 20:47:19 2010 JenneUpdatePEMTerminator put on Guralp Box

A little D-sub terminator was put on the Gur1 input to the Guralp box, to check again the noise level of the box.

  3147   Wed Jun 30 14:38:04 2010 JenneUpdatePEMGuralp terminator is switched

I moved the Guralp box's input terminator from Gur1 to Gur2 a minute or so ago to check the other channels.

  3149   Wed Jun 30 17:30:54 2010 JenneUpdatePEMSeismometers moved for MC Wiener work

Now that the MC is back up and running, I put the Guralp seismometers at the ends of the mode cleaner.  Gur1 is near MC2, and Gur2 is near MC1 (yes, it seems backwards....that's how the cable lengths work).  Also, the set of 3 MC2 accelerometers are in place under MC2.  I can't find the black cube for the other set of accelerometers, so there aren't any around MC1/3. 

  3153   Thu Jul 1 14:03:40 2010 josephbUpdatePEMTemporary disconnect of some PEM channels for ~20 minutes

In order to identify the output adapter of the BNC patch panel used for about 20 PEM channels, I had to disconnect its power and remove the back panel.  Channels coming into that panel (seismometers and so forth) was out from 1:36 to 1:56 pm.

I did a quick check of some of the channels and it looks like its working again after putting it all back together.

ELOG V3.1.3-