  11785   Wed Nov 18 22:26:33 2015 yutaroUpdateOptical LeversCalibration of oplevs for ITMY/ETMYBased on elog
1403, I calibrated the oplevs for ITMY/ETMY.

Summary of this method is following:
 calib_etmypit2.pdfcalib_etmyyaw2.pdfcalib_itmypit2.pdfcalib_itmyyaw2.pdfcalib_oplev.zip 
  11786   Wed Nov 18 23:18:07 2015 ericqUpdateComputer Scripts / Programsnodus /boot cleared upThe /boot partition was filling up with
old kernels. Nodus has automatic security
updates turned on, so new kernels roll in
  
  11787   Wed Nov 18 23:40:01 2015 ranaUpdateOptical LeversCalibration of oplevs for ITMY/ETMYOMG. Please try to use larger fonts and
PDF so that we can read the plots.

  
  11788   Thu Nov 19 14:50:34 2015 Max IsiUpdateGeneralNew 2D histogram plot for summary pagesA new type of plot is now available for
use in the summary pages, based on EricQ's 2D
histogram plots (elog
 C1-ALL_AB3834_HISTOGRAM2D-1131926417-86400.png 
  11789   Thu Nov 19 15:16:24 2015 ericqUpdateCDSc1iscey IO chassis now has brackets[Steve, ericq]

Brackets for the c1iscey IO
chassis cards have been installed. Now, I
  
  11790   Thu Nov 19 16:06:54 2015 yutaroUpdateOptical LeversCalibration of oplevs for ITMY/ETMYI'm sorry. I will be careful about
that. And I updated the plots in elog
11785.
  
  11794   Sat Nov 21 00:45:30 2015 KojiUpdateIOOIMC fixBased on the observation of the PMC error
signal, I started measuring the IMC OLTF.
Immediately, it
 PDRFsignal_cable.JPGQsignal_cable1.JPGQsignal_cable2.JPGQsignal_cable3.JPG 
  11795   Sat Nov 21 00:46:33 2015 KojiUpdateIOOIMC OLTFHere is the comparison before and after
the fix.

Before the work, the UGF was ~40kHz.
 IMC_OLTF.pdfPMC_noise_comparison.pdf 
  11796   Sun Nov 22 07:09:01 2015 ranaUpdateIOOIMC fixOn the demod board there is a 10 dB attenuator
(AT1), which lowers the level to -10 dBm
before the ERA-5. Then it should be 10 dBm
  
  11797   Sun Nov 22 12:07:09 2015 KojiUpdateIOOIMC fixTO DO

Done (Nov 23, 2015) -
Check if the attenuator is still there in
  
  11798   Sun Nov 22 12:12:17 2015 KojiUpdateIOOIMC OLTFWell. I thought a bit more and now I think
it is likely that this is just the servo
bump as you can see in the closed-loop TF.
  
  11799   Mon Nov 23 14:45:39 2015 ericqUpdateComputer Scripts / ProgramsNew softwareCOMSOL 5.1 has been installed at: /cvs/cds/caltech/apps/linux64/comsol51/bin/comsol

MATLAB 2015b has been installed
  
  11800   Mon Nov 23 20:32:43 2015 KojiUpdateLSCFrequency source fixed, IMC LO level adjustedThe frequency source was fixed. The IMC
LO level was adjusted.

IMC is locked => OLTF measured
  
  11801   Mon Nov 23 21:48:49 2015 KojiUpdateLSCFrequency source fixed, IMC LO level adjustedThe trouble we had: the 29.5 MHz
source had an output of 6 dBm instead
of 13 dBm.
 IMG_2136.JPGIMG_2129.JPGIMG_2133.JPG 
  11802   Mon Nov 23 22:12:10 2015 KojiUpdateIOOLO level check for the IMC demod boardIn order to check the proper LO level,
the IMC demod board was checked. As
a short summary, -8dBm is the proper input
 IMG_2137.JPGIMG_2138.JPG 
  11803   Mon Nov 23 23:42:56 2015 ericqUpdateLSCALSY recovered[ericq, gautam]

Gautam couldn't observe a Y
green beatnote earlier, so we checked
  
  11804   Tue Nov 24 01:14:23 2015 KojiUpdateLSCALSY recoveredSorry, I completely forgot to turn the
Marconi on...
  
  11805   Tue Nov 24 11:18:47 2015 yutaroUpdateOptical LeversBeam centering for the oplev of ITMY I made the beam spot   
  11806   Tue Nov 24 14:58:40 2015 yutaroUpdateLSCITMX misalignedI misaligned ITMX. The oplev servo for
ITMX is now turned off. You can restore ITMX
alignment by running "restore".
  
  11807   Wed Nov 25 04:24:21 2015 ranaUpdateIOOLO level check for the IMC demod boardHmmm. Very non-standard demod. From the
photo, looks like someone did some surgery
with the attenuators (AT1, AT2, AT3) in the
  
  11808   Wed Nov 25 10:07:10 2015 KojiUpdateIOOLO level check for the IMC demod boardI didn't finish making the DCC entry
for this module yet.

But the attenuators are
  
  11809   Wed Nov 25 14:46:53 2015 gautamUpdateCDSc1lsc models restartedI noticed that all the models running
on C1LSC had crashed when I came in earlier
today. I restarted all of them by ssh-ing
  
  11810   Wed Nov 25 16:40:32 2015 yutaroUpdateLSCround trip loss of Y armI measured round trip loss of Y arm. The
alignment of relevant mirrors was set ideal
with dithering (no offset).
  
  11811   Wed Nov 25 16:46:10 2015 KojiUpdateIOOLO level check for the IMC demod boardAwwww. I found that the demod board has
the power splitter (PSCJ-2-1) with one output
unterminated.
 D990511-40m_151123.pdf 
  11812   Wed Nov 25 20:07:35 2015 ranaUpdateIOOLO level check for the IMC demod boardPerhaps we can replace T1 with a mini-circuits
hybrid 0-90 deg splitter and then remove
the trim caps. (JSPQ-80, JYPQ-30, SCPQ-50)
  
  11813   Wed Nov 25 22:37:12 2015 yutaroUpdateLSCremoval of Gautam's cable in 1Y2 and restoration of POYDC[yutaro, Koji]

We disconnected the cable that
was connected to CH5 of the whitening filter in
  
  11814   Wed Nov 25 22:59:42 2015 yutaroUpdateLSCAS table optics realignmentI slightly changed the orientation
of a few mirrors on AS table that
are used to make the AS light get into PDs,
  
  11815   Wed Nov 25 23:17:34 2015 yutaroUpdateLSCstrange behavior of ASDC[yutaro, Koji]

I noticed that ASDC level changes
depending on the angle of ITMY when trying
 14.png 
  11816   Wed Nov 25 23:34:52 2015 yutaroUpdateLSCround trip loss of Y arm[yutaro, Koji]

Due to the strange behavior (elog
 14.png 
  11817   Thu Nov 26 19:39:27 2015 KojiUpdateLSCCurrent state of the frequency source, and possible improvementUploaded on T1000461
too.
 RF_Frequency_Source.pdf 
  11819   Fri Nov 27 22:20:24 2015 yutaroUpdateLSCround trip loss of Y armHere, I upload data I took last night,
including the power of reflected power (locked/misaligned)
and transmitted power for each point (attachement
 log.txt.zip17.png 
  11820   Sat Nov 28 11:46:40 2015 yutaroUpdateLSCpossible error source of loss map measurementI found that TRY level degraded and the
beam shape seen with CCD camera at AS port
was splitted when the beam spot on ETMY was
  
  11821   Sun Nov 29 05:23:57 2015 ranaUpdateLSCCurrent state of the frequency source, and possible improvementI need some more hints to understand the
improvement, although its generally good
to re-build it considering the sad state
  
  11822   Sun Nov 29 12:32:26 2015 KojiUpdateLSCCurrent state of the frequency source, and possible improvementI'm not claiming we need to modify
the frequency source immediately as we are
not limited by the oscillator amplitude or
  
  11823   Mon Nov 30 10:41:45 2015 yutaroUpdateLSCDoes a baffle in front of ETMY have effect on loss map measurement?It might have, so I think
I need to estimate shift of beam spot
more preciely.
  
  11824   Mon Nov 30 12:19:38 2015 yutaroUpdateComputer Scripts / Programsimage captureOn VIDEO.adl, Image Capture and Video Capture
did not seem to work and gave me some errors,
so I fixed following two things:
  
  11825   Mon Nov 30 14:12:14 2015 ericqUpdateLSCLO level check for the LSC RF distribution boxI checked the RF levels at the LSC LO distribution
box, with the agilent scope and a handful
of couplers. This was all done with the Marconi
  
  11826   Mon Nov 30 15:17:57 2015 KojiUpdateLSCLO level check for the LSC RF distribution boxT1000461
tells us that the nominal LO input is 2dBm
although we don't know what's the
  
  11827   Mon Nov 30 16:33:06 2015 ericqUpdateLSCstrange behavior of ASDCOne possible explanation of this behavior is
simply poor centering of the AS beam
on AS55 (whose DC level provides ASDC, if
 AScentering.png 
  11828   Mon Nov 30 17:17:30 2015 yutaroUpdateLSCDoes a baffle in front of ETMY have effect on loss map measurement?With captured images of ETMYF, I measured
the shift of the beam spot on ETMY.

The conclusions are:
 element2-2.pngelement0-0.pngelement4-4.png 
  11829   Mon Nov 30 18:27:30 2015 KojiUpdateLSCstrange behavior of ASDCIt wasn't fully mentioned in ELOG
11814.
We checked the PD first and this behavior
  
  11830   Tue Dec 1 10:52:52 2015 SteveUpdateGeneralDataviewerDataviewer x axis end is not there.

On ( 2600 days) longer plots it
is missing 8 moths and on (100 days) shorther
 xAxisEndMissing.png 
  11831   Tue Dec 1 11:26:23 2015 yutaroUpdateOptical LeversCalibration of oplevs for ITMX/ETMXWith the same method as reported in elog
11785, I calibrated oplevs for ITMX/ETMX.
 ep_l.pdfey_l.pdfip_l.pdfiy_l.pdf 
  11832   Tue Dec 1 16:55:04 2015 SteveUpdateVACDataviewer fixedQ adjusted the Dataviewer so it is not
chopping of data any more. Thanks.

 
 10yrsCC1&4M.png 
  11833   Tue Dec 1 17:16:31 2015 gautamUpdateCDSScript for copying BLRMS filtersWe've been talking about putting in
BLRMS filters for several channels - it would
be a pain to manually copy over the correct
 makeFilterFiles.bash.zip 
  11834   Tue Dec 1 17:26:14 2015 KojiUpdateGeneralMegatron maitenenceSLOWDC servo was dead. I followed EricQ's
instruction.
  
  11835   Tue Dec 1 20:20:16 2015 KojiUpdateGeneralMegatron maitenenceMC Autolocker got stack somewhere. I had
to go to megatron and kill MC Autolocker.

init relaunched the autolocker
  
  11836   Wed Dec 2 03:34:30 2015 ericqUpdateLSCSRCL OLG weirdness[gautam, ericq]

Since ETMX seems to have been on
good behavior lately, we tried to fire the
 weirdSRCLG.pdf 
  11837   Wed Dec 2 15:08:41 2015 ericqUpdateComputer Scripts / ProgramsDonatella sudo problem resolvedSomehow, the controls
user account on donatella lost its membership
to the sudoers
  
  11838   Wed Dec 2 15:52:28 2015 yutaroUpdateLSCrestoration of POXDCI disconnected the   