  40m Log, Page 24 of 341  Not logged in ELOG logo
ID Dateup Author Type Category Subject Text Attachments
  1165   Mon Dec 1 15:09:27 2008 robUpdatePEMhalf-micron particle count is alarming    
  1166   Tue Dec 2 17:56:56 2008 Alberto, RanaConfigurationPSLMC AlignmentIn the attempt to maximize the Mode Cleaner
transmission and minimize the reflection
from the steering mirrors of the MC periscope,
  
  1167   Tue Dec 2 19:18:10 2008 ranaSummaryPEMRanger SS-1In entry http://dziban.ligo.caltech.edu:40/40m/881
and a follow up from Jenne I put in the Ranger
calibration.
  
  1168   Tue Dec 2 19:51:32 2008 ranaUpdatePEMhalf-micron particle count is alarmingThe 0.5 micron dust monitor count is now pretty
high (36000). I wandered around the lab to
see if there was anything
 Untitled.pngdust.png 
  1169   Wed Dec 3 11:58:10 2008 AlbertoUpdatePSLMC AlignmentRana, Alberto,

more details on the MC alignment we did yesterday.
  
  1170   Wed Dec 3 12:49:11 2008 jenneUpdateComputerssomething sketchy with NDS ... or something[SIZE=2][COLOR=green]Never mind...I had forgotten
that you have to run mdv_config every time
you open matlab, not just every time you
  
  1171   Wed Dec 3 19:21:09 2008 ranaConfigurationPEMRanger moveI looked at the Ranger signals. [B]Somehow
it has a relative transfer function of 'f'
between it and the Guralp.[/B]
 c.pdfwsnb.pdf 
  1172   Wed Dec 3 20:10:09 2008 Jenne, RanaUpdatePEMComparing Wiener subtraction with different seismometersAttached is a plot of MC_L, and then the residual
MC_L after static Wiener filtering, using
different combinations of our accelerometers
 Dec032008_c1wino_seisCombos.png 
  1173   Wed Dec 3 20:36:07 2008 Jenne, RanaUpdatePEMComparing Wiener subtraction with different seismometersThe Ranger has now been moved over to sit
underneath the MC2 tank (it was previously
close to the PSL rack). It
 d.pdfsco.pngfly.pdf 
  1174   Thu Dec 4 13:49:39 2008 JenneUpdatePEMMore of: Comparing Wiener subtraction using different sensorsHere is another version of the same type plot
I put in the elog yesterday.  This plot is
looking at the 7200 seconds after 04Dec2008
 Dec042008_c1wino_seisCombos.png 
  1175   Thu Dec 4 16:29:20 2008 josephbConfigurationComputersError message on Frame Builder Raid ArrayThe Fibrenetix FX-606-U4 RAID connected to
the frame builder in 1Y7 is showing the following
error message: IDE Channel #4  Error Reading
  
  1176   Thu Dec 4 17:42:23 2008 carynUpdateIOOdrum modes observable without excitationSo, the mode cleaner was evidently aligned
better and now the drum modes are observable
using DTT. 
 drum_modes.pdfdrum_modes2.pdf 
  1177   Fri Dec 5 01:41:33 2008 YoichiConfigurationComputersMEDM screen snapshot now works on linux machinesAs a part of my "make everything work on linux"
project, I modified 'updatesnap' script so
that linux machines can update MEDM screen
  
  1178   Fri Dec 5 01:58:58 2008 YoichiConfigurationASCtdscntr.pl now works at 40mTobin gave me the perl version of tdscntr
some time ago.
Pinkesh and I modified and tested it at LHO.
  
  1179   Fri Dec 5 09:29:59 2008 ranaUpdateIOOdrum modes observable without excitationNot sure what the y-scale is since there aren't
any y axis labels in the plot, but it seems
like we
  
  1180   Fri Dec 5 14:13:41 2008 ranaSummaryIOOMC trend for the last 4 days[COLOR=pink][SIZE=2][FONT=Comic Sans MS]The
MC has stayed locked for ~3 days! I just
broke it to reset the MZ.[/FONT][/SIZE][/COLOR]
 g.png 
  1181   Fri Dec 5 20:40:38 2008 YoichiHowToComputersElog multi-keyword searchThe current Elog search allows you to look
for only one keyword in the text.
You cannot search for two keywords by simply
  
  1182   Fri Dec 5 21:31:11 2008 YoichiUpdateIOOdrum modes observable without excitationThe calibration of the MC_F feedback is posted
in elog:1032.
I'm not sure where Caryn took MC signal,
  
  1183   Sun Dec 7 16:02:46 2008 ranaUpdateGeneralMag 5.1 EQ halfway to VegasThere was a [URL=http://earthquake.usgs.gov/eqcenter/recenteqsww/Quakes/ci14408052.php#details]Mag
5.1 EQ[/URL] Friday night at 8:18 PM.
 g.pngUntitled.png 
  1184   Sun Dec 7 16:12:53 2008 ranaUpdateDAQbooted awg[SIZE=3]because it was [COLOR=red]red[/COLOR][/SIZE]   
  1185   Mon Dec 8 00:10:42 2008 carynSummaryGeneralcalibrating the jenne laserI apologize in advance for the long list of
numbers in the attachment. I can't seem to
make them hide for some reason. 
 IvsDCmax.pngdata.tar.gzjcal_dip_2_norm_log.pngjcal_flat_3_norm_log.png 
  1186   Mon Dec 8 11:41:27 2008 YoichiSummaryVACThe rough pump for the TP2 replacedBob, Yoichi

The foreline pressure of the TP2 (the foreline
  
  1187   Mon Dec 8 11:54:27 2008 YoichiUpdateGeneralIFO mirrors alignedThis morning, I re-aligned the IFO mirrors
to see if they were badly moved by the earthquake.
The both arms locked just by the restoring
  
  1188   Mon Dec 8 17:50:21 2008 YoichiUpdateSUSITMY driftThe suspension drift monitor shows that the
ITMY alignment was shifted after the earthquake.
Looks like only the UL sensor had a step
 ITMY-OSEMs.pngITMY-UL.pngITMY-08-18.pdfITMY-12-08.pdf 
  1189   Tue Dec 9 10:48:17 2008 CarynSummaryGeneralcalibrating the jenne laser: impedance mismatch?
We sent RFout of network analyzer to a splitter,
with one side going back to the network analyzer
 reflection.png 
  1190   Fri Dec 12 22:51:23 2008 YoichiUpdatePSLReference cavity ring down measurement againBob made new HV-cables with HV compatible
coaxes. The coax cable is rated for 2kV,
which was as high as Bob
 RC-Ringdown.png 
  1191   Tue Dec 16 19:06:01 2008 YoichiUpdatePSLReference cavity ring down repeated many timesToday, I repeated the reference cavity ring
down measurement many times to see how much
the results vary.
 RC_Ringdown_Estimates.pngCavity_Pole_Histogram.png 
  1192   Thu Dec 18 12:52:00 2008 AlbertoConfigurationSUSMode Cleaner Cavity AlignmentThis morning I found the MC locked to the
10 mode. When I locked it on the 00 mode,
it was unstable and eventually it always
  
  1193   Thu Dec 18 19:15:54 2008 Alberto, YoichiConfigurationSUSMode Cleaner Cavity Alignment[quote]This morning I found the MC locked
to the 10 mode. When I locked it on the 00
mode, it was unstable and eventually it always
  
  1194   Fri Dec 19 11:18:52 2008 AlbertoConfigurationGeneralMode Cleaner Temperature MonitorI reduced from 10 to 5 the gain of the SR560
that Caryn has set up after the lock-in amplifier
nest to the PSL rack because the overload
  
  1195   Fri Dec 19 11:29:16 2008 Alberto, YoichiConfigurationMZMZ Trans PDLately, it seems that the matching of the
input beam to the Mode Cleaner has changed.
Also, it is drifting such that it has become
  
  1196   Fri Dec 19 14:35:58 2008 Yoichi AlbertoUpdateIOOMC WFS and IOO-POS QPD re-centeringFor the past two days, the MC alignment has
kept drifting.
This morning, the MC alignment was so bad
  
  1197   Fri Dec 19 16:38:09 2008 steveUpdateLSCall optlevs centeredAll optlevs were centered after full alignment.

Qpd sums are:
  
  1198   Sat Dec 20 23:37:43 2008 robOmnistructureGeneralSaturday Night Fever after presumed power failure
Just came by to pick something up...
  
  1199   Sun Dec 21 13:00:06 2008 steveUpdateall down cond.vac and laser back onThere was a power outage sometimes early Saturday.

All things are down.
  
  1200   Sun Dec 21 14:18:04 2008 YoichiUpdateComputersRFM network bypass box's power supply is dead I restarted the front-end computers by power
cycling them one-by-one.
After issuing startup commands, most of them
  
  1201   Mon Dec 22 13:48:22 2008 YoichiUpdateComputersRFM network bypass box's power supply is dead As a temporary fix, I cut the cable of the
power supply and connected it to the Sorensen
power supply +5V on the rack.
  
  1202   Tue Dec 23 10:35:40 2008 YoichiUpdateComputersRFM network breakdown mostly fixedRana, Rolf, Alberto, Yoichi

The source of the problem was the RFM bypass
  
  1203   Wed Dec 24 10:33:24 2008 YoichiUpdateComputersSeveral fixes. Test point problem remains.Yesterday, I fixed several remaining problems
from the power failure.
  
  1204   Wed Dec 24 12:46:54 2008 YoichiUpdateComputersTest points are backRob told me how to restart the test point
manager.
It runs on fb40m and actually there is an
  
  1205   Mon Dec 29 18:01:07 2008 AidanUpdateAuxiliary lockingUpdated 40m Upgrade Document T080074-00-R
Added a paragraph to the 40m Upgrade document
describing the fiber stabilization and frequency
  
  1206   Mon Dec 29 21:38:57 2008 YoichiUpdateComputersSnapshots of MEDM screensI wrote scripts to take snapshots of MEDM
screens in the background.
These scripts work even on a computer without
  
  1207   Mon Dec 29 21:51:02 2008 YoichiConfigurationComputersWeb server on nodusThe apache on nodus has been solely serving
for the svn web access.
I changed the configuration and all files
  
  1208   Tue Dec 30 18:51:18 2008 rana,yoichiConfigurationElectronicsIlluminator Power Supply reset[COLOR=blue][B]We noticed that none of the
illuminators were working.[/B][/COLOR]
  
  1209   Wed Dec 31 22:59:40 2008 YoichiSummaryEnvironmentParticle counts going crazyYes it is a new year's eve, and a lot of crazy
people are on Colorado to secure seats for
the parade tomorrow.
  
  1210   Thu Jan 1 00:55:39 2009 YoichiUpdateASCAlignment scripts for LinuxA Happy New Year.

The dither alignment scripts did not run
  
  1211   Thu Jan 1 01:07:03 2009 YoichiSummaryEnvironmentParticle counts going crazyI increased the fan speed of the PSL HEPA
filter to the maximum.
  
  1212   Thu Jan 1 01:15:45 2009 YoichiSummaryVACN2 line leak ?I've been replacing the N2 bottles recently.
I noticed that the consumption is too high.
I had to replace them every two days.
  
  1213   Fri Jan 2 17:20:44 2009 ranaSummaryComputer Scripts / Programs40m GWINCI have made a '40m' directory in the iscmodeling
CVS tree which allows one to run a 40m version
of GWINC.
 bench40.pdf 
  1214   Fri Jan 2 18:49:54 2009 YoichiUpdateLSCLSC modulation frequencies adjustedI noticed that the IFO did not lock in the
MICH configuration.
This was because AS166Q signal was too small.
  