  8651   Tue May 28 19:35:17 2013 ManasaUpdateGreen LockingStill searching for the X green beat noteProcedure
1. Aligned X-arm to IR.
2. Aligned green to the X-arm.
  
  1353   Wed Mar 4 03:50:17 2009 YoichiUpdateLockingStill won't go above arm power 10Yoichi, Kentaro

The IFO still loses lock at around arm power
 lockLoss1.pnglockLoss2.pnglockLoss3.png 
  5682   Mon Oct 17 23:28:32 2011 ranaUpdateElectronicsStochMonTo get to the bottom of the RFAM mystery,
we've got to resurrect the StochMon to trend
the RFAM after the IMC.
  
  5706   Wed Oct 19 18:18:03 2011 SureshUpdateElectronicsStochMon : Filters installed
 PA190146.JPGPA190143.JPG 
  1093   Mon Oct 27 11:16:23 2008 AlbertoConfigurationIOOStochMon CalibrationI implemented the calibration for the four
channels of the StochMon in the ioo EPICS
database. Now the output of those channels,
  
  2253   Thu Nov 12 12:50:35 2009 AlbertoUpdateComputersStochMon calibrated channels added to the data trendI added the StochMon calibrated channels
to the data trend by including the following
channel names in the C0EDCU.ini file:
  
  554   Mon Jun 23 19:48:28 2008 rana,albertoSummaryIOOStochMon trends (80 days)Here's a StochMon plot showing the RFAM after
the MC. Remember that in these units, 2V
means no RFAM
 e.png 
  3170   Wed Jul 7 17:18:57 2010 AlbertoConfigurationElectronicsStochmon and LSC AM Stabilizer DecomissionedToday I disconnected and removed the
Stochmon box from the 1Y2 rack.
I
also removed the amplifiers that were sitting
  
  5996   Thu Nov 24 05:47:16 2011 KojiSummaryIOOStochmon runningNow stochmon for 11MHz and 55MHz is running.
The calibration / noise measurement are going
to be post later...
  
  6028   Mon Nov 28 18:19:53 2011 kiwamuUpdateIOOStochmon seems workingHere is a 48 hours trend of the RFAM
monitor (a.k.a StochMon):
 RFAM_48hours.png 
  5941   Fri Nov 18 01:51:37 2011 KojiUpdateIOOStochmon updateUpdate
of the stochmon status
 stochmon.pdfstochmon_calib.pdfRFAM_PD_noise.pdf 
  6009   Fri Nov 25 20:03:05 2011 KojiUpdateIOOStochmon update New RFAM mon calibration  stochmon_calib.pdf 
  5921   Thu Nov 17 11:04:02 2011 JenneUpdateRF SystemStochmon?Is there an update on Stochmon?  Are
the signals acquired somewhere already? 
What's the current deal-io?  The new
  
  5923   Thu Nov 17 11:35:27 2011 KojiUpdateRF SystemStochmon?The  Stochmon channels for 11&55MHz
have been reasonably working since last night.
The output is not yet calibrated
  
  5926   Thu Nov 17 14:38:16 2011 ZachUpdateRF SystemStochmon?It turns out that we don't have all the
parts I would need to do a full prototype
  
  15379   Sat Jun 6 14:07:30 2020 JonUpdateBHDStock-Part Mode-Matching Telescope OptionSummary

For the initial phase of
BHD testing, we recently discussed whether
the mode-matching telescopes could be built
 ASpathStock.pdfLOpathStock.pdf 
  2339   Wed Nov 25 20:28:17 2009 AlbertoUpdateABSLStopped working on the AbsLI closed the shutter of the NPRO for the
night.
  
  8020   Thu Feb 7 09:03:54 2013 ManasaUpdateGeneralStore optics in respective cabinets
@Yuta
The ITMX table has been left open since
  
  8021   Thu Feb 7 10:35:35 2013 yutaUpdateGeneralStore optics in respective cabinetsI'm not the one who opened the ITMX table
yesterday, but thanks for reminding me.
I put POP DC oscilloscope and its
  
  4132   Tue Jan 11 11:19:13 2011 josephbSummaryCDSStoring FE harddrives down Y armLacking
a better place, I've chosen the cabinet down
the Y arm which had ethernet cables and various
  
  10830   Mon Dec 22 16:21:15 2014 ericqUpdateelogStrange ELOG searchIn order to fix ELOG search, I have started
running ELOG v2.9.2 on Nodus.
Sadly,
  
  10839   Tue Dec 23 16:49:32 2014 ericqUpdateelogStrange ELOG search So, despite having registered users,
it turns out that the "Author"
field is still open for editing when making
  
  10827   Mon Dec 22 13:34:34 2014 KojiUpdateelogStrange ELOG serachI tried to find my own entry and faced
with a strange behavior of the elog.
The search button invoked the following
  
  5241   Mon Aug 15 17:36:10 2011 jamieUpdateSUSStrangeness with ETMY (was: Monday SUS update)For some reason ETMY has changed a lot. 
Not only does it now have the worst "badness"
(B matrix condition number) at ~10, but the
  
  6694   Sun May 27 17:19:27 2012 ranaSummaryloreStrawberries We have placed some sweet giant strawberries
in the fridge; free for eating for anyone
working in the lab today or tomorrow:
 20120527_145826.jpg 
  2082   Mon Oct 12 17:27:20 2009 KojiConfigurationSAFETYStray beam blockingSteve, Kiwamu, and Koji
We went through the PSL table to
make sure any strong beam did not hit the
 PSL.png 
  2084   Mon Oct 12 18:38:07 2009 JenneConfigurationSAFETYStray beam blocking


   
        
            
  
  2085   Mon Oct 12 19:53:44 2009 KojiConfigurationSAFETYStray beam blockingI aligned the EOM and the beam to the PMC.
The beam is still hitting the bottom of the
EOM aperture, 
  
  2087   Mon Oct 12 20:01:13 2009 KojiConfigurationSAFETYStray beam blockingOK! I saw the optics are redundant and
some of the components are not in a right
place.
  
  2088   Mon Oct 12 22:15:15 2009 ranaConfigurationPSLStray beam blockingYou can remove the RFAM measuring setup.
Once we upgrade, we will no longer have a
MZ or the related problems. 
  
  5531   Fri Sep 23 17:31:14 2011 KatrinUpdateGreen LockingStray light reduction (Y)I inserted several beam blocks and iris
on the Y arm Green table. There was/is lots
of stray light because a lot of the mirrors
  
  6983   Wed Jul 18 09:09:51 2012 MashaUpdatePEMStreckeisen The Streckeisen is currently plugged into
ADC channels 13, 14, and 15 (corresponding
to STS-3 in the channels). The X, Y, and
 Screenshot.pngScreenshot-1.png 
  207   Thu Dec 20 19:10:03 2007 waldmanUpdateOMCStressful reattachment of heaterPhotos may follow eventually,  but for now
here's the rundown.  I scraped the heater
clean of the thermal epoxy using a clean
  
  16314   Fri Sep 3 02:03:15 2021 TegaSummaryComputersStrip down large error filesAlso deleted the ~50GB error files from
ldas to prevent rsync from copying them
to nodus again. With the new update to GWsumm,
  
  5390   Mon Sep 12 23:45:14 2011 SureshConfigurationComputer Scripts / ProgramsStripTool does not scale properly on PianosaWhen I run StripTool on Pianosa, I get
the following message
==== StripTool Xt Warning Handler
  
  6682   Fri May 25 16:10:53 2012 KojiUpdateGeneralStripTool on Zita restoredssh -X pianosa
On pianosa
  
  6943   Mon Jul 9 10:52:48 2012 MashaUpdatePEMStripTools on WallThe RMS signals generated by the updated
filtering process are now on the wall. The
NaN issue is gone it seems, and the template
  
  859   Wed Aug 20 11:50:10 2008 JohnSummaryComputersStripTools on op540m[SIZE=5][FONT=Comic Sans MS]
To restart the striptools on op540m:
  
  15668   Tue Nov 10 11:59:37 2020 gautamUpdateVACStuck RV2I've uploaded some more photos here.
I believe the problem is a worn out thread
where the main rotary handle attaches to
 damagedThread.001.jpegIFOstatus.pngP1a_leakTest.png 
  9696   Wed Mar 5 22:32:21 2014 manasaUpdateLSCStuck at step 2

    

       
            
  
  9698   Thu Mar 6 11:15:32 2014 KojiSummaryLSCStuck at step 2You don't need to make transition from
ALS X/Y to ALS C/D. Just stabilize the arms
with ALS C/D from the beginning.
  
  14043   Sat Jul 7 19:50:38 2018 AnnalisaConfigurationThermal CompensationStudy about the Thermal projection setup and its effect on the cavityI made some simulation to study the change
that the heater setup can induce on the Radius
of Curvature of the ETM.
 6x 
  8226   Mon Mar 4 20:03:42 2013 AnnalisaHowToCOMSOL TipsStudy of mirror mount eigenfrequencies I studied the eigenfrequencies of
a mirror mount designed with COMSOL.
I imposed fixed constraints for the
 MirrorSupport1.pngMirrorSupportEig1.pngpedestal.pngBase2.png 
  4371   Wed Mar 2 22:57:57 2011 sureshSummaryGeneralStuff from LLOHere is a partial list of stuff which is
being packed at LLO to be shipped to CIT. 
The electronics ckt boards are yet to be
 eLIGO_items_from_LLO_for_Caltech.xls 
  4381   Mon Mar 7 17:58:14 2011 sureshSummaryGeneralStuff from LLO
Here is the updated list. These lists were
used as packing lists and therefore are organised
 eLIGO_items_from_LLO_for_Caltech_Sheet1.pdf 
  15968   Thu Mar 25 18:05:04 2021 gautamUpdateElectronicsStuffed HV coil drivers received from Screaming CircuitsI think the only part missing for assembly
now are 4 2U chassis. The PA95s need to be
soldered on as well (they didn't arrive
  
  2662   Mon Mar 8 18:15:27 2010 ranaUpdatePEMStyrofoam Cooler on the SeismosI put Jenne's
cooler over the seismometers
 SeismometersHuddled.jpg 
  2665   Tue Mar 9 12:06:53 2010 rana, JenneUpdatePEMStyrofoam Cooler on the SeismosLooks like the GUR2_X signal is bad. Jenne
says that we need to center it mechanically
before the signals will become useful again.
  
  2666   Wed Mar 10 15:02:58 2010 ranaUpdatePEMStyrofoam Cooler on the SeismosThis is the spectra and coherence from
a quiet time last night. I've lowered the
Guralp cal by a factor of 2 to account for
 Seismic-Ref_100310.pngUntitled.gif 
  2671   Sat Mar 13 21:30:22 2010 ranaUpdatePEMStyrofoam Cooler on the Seismos   Untitled.png 