40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m elog, Page 223 of 357  Not logged in ELOG logo
ID Date Author Type Category Subjectup
  5593   Sat Oct 1 22:53:49 2011 kiwamuSummaryGeneralRecovery from the power shutdown : lockable

Found the Marconi for the 11 MHz source had been reset to its default.

 => reset the parameters. f = 11.065910 MHz (see #5530) amp = 13 dBm.

Interferometer became lockable. I checked the X/Y arm lock, and MICH lock, they all are fine.

  5599   Mon Oct 3 08:38:21 2011 steveSummaryVACRecovery from the power shutdown is completed

 

I failed to start Rana's favorit anciant desktop computer at the vac rack. He has to baby this old beast just a little bit more.

Vacuum status: Vac Normal was reached through Farfalla: Rga was switched back to IFO and and Annuloses are beeing pumped now.

V1 was closed for about a day and the pressure reached  ~2.8 mTorr in the IFO.  This leakrate plus outgassing is normal

The ref cavity 5000V was turned on.

The only thing that has to be done is to restart the RGA. I forgat to turn it off on Friday.

 

  5662   Thu Oct 13 21:40:59 2011 ranaSummaryVACRecovery from the power shutdown is completed

 As it turns out Steve is not crazy in this particular instance: the vacuum computer, linux3 , has some issues. I can login just fine, but trying to open a terminal makes the CPU rail and the RAM max out and eventually the machine freezes.

Under KDE, I can open a terminal OK as root, but if I then try a 'su controls', the same issue happens. Let's wait for Jamie to fix this.

  5596   Sun Oct 2 13:45:13 2011 ranaSummaryGeneralRecovery from the power shutdown: apache / svn

Restarted Apache on nodus using Yoichi's wiki instructions. SVN is back up.

  4432   Wed Mar 23 12:40:22 2011 Chief Recycling OfficerHowToEnvironmentRecycle stuff!

The following is a message from the LIGO 40m Chief Recycling Officer:

Please get up off your (Alignment Stabilization Servo)es and recycle your bottles and cans!  There is a recycling bin in the control room.  Recent weeks have seen an increase in number of bottles/cans thrown away in the regular garbage.  This is not cool. 

Thank you,

---L4mCRO

  10449   Thu Sep 4 01:28:32 2014 ericqUpdateLSCRecycling cavity lengths

 Going off some discussion we had at lunch today, here is my current knowledge of the state of cavity lengths. 

Acknowledging that Koji changed the sideband modulation frequency recently, the ideal cavity lengths are (to the nearest mm):

  • Lprc = c / ( 4 * fmod) = 6.773 m
  • Lsrc = c / ( 5 * fmod) = 5.418 m

We when last hand measured distances, after moving PR2, we found:

  • Lprc  = 6.752 m = 2.1 cm short
  • Lsrc  = 5.474 m = 5.6 cm long. 

However, when I looked at the sideband splitting interferometrically, I found:

  • Lprc = 6.759m = 1.4 cm short

This is only 5mm from the hand measured value, so we can believe that the SRC length is between 5 and 6 cm too long. I'm building a MIST model to try and see what this may entail. 

  10451   Thu Sep 4 10:10:23 2014 KojiUpdateLSCRecycling cavity lengths

Com'on. This is just a 60ppm change of the mod frequency from the nominal. How can it change the recycling cav length by more than a cm?

https://wiki-40m.ligo.caltech.edu/IFO_Modeling/RC_lengths

This describes how the desirable recycling cavity lengths are affected by the phase of the sidebands at non-resonant reflection of the arms.

If we believe these numbers, L_PRC = 6.7538 [m] and L_SRC = 5.39915 [m].

Compare them with the measured numbers

  • Lprc = 6.752 m
  • Lsrc  = 5.474 m

You should definitely run MIST to see what is the optimal length of the RCs, and what is the effect of the given length deviations.

  10453   Thu Sep 4 18:16:20 2014 ericqUpdateLSCRecycling cavity lengths

Koji correctly points out that I naïvely overlooked various factors. With a similar analysis to the wiki page, I get:

  • Ideal arm length of 37.795 m
  • Ideal PRC length of 6.753 m
  • Ideal SRC length of 5.399 m

This means that:

  • The PRC, measured at 6.759m, is 6mm long. 
  • The SRC, measured at 5.474m, is 7.5 cm long

Next step is to see how this may affect our ability to sense, and thereby control, the SRC when the arms are going. 

MIST simulations and plots are in the attached zip. 

  2071   Thu Oct 8 21:32:59 2009 KojiSummaryGeneralRecycling cavity loss

I looked at the data of the day before yesterday (Oct 06) to know how much is the recycling gain.

X arm: (TRX_PRecycled) / (TRX_PRMmisaligned) * T_PRM = 83.1/0.943*0.07 = 6.17
Y arm: (TRX_PRecycled) / (TRX_PRMmisaligned) * T_PRM = 99.2/1.017*0.07 = 6.83

==> G_PR = 6.5 +/- 0.5     (oh...this estimation is so bad...)

From the recycling gain and the arm cavity reflectance, one can get the loss in the recycling cavity.

G_PR = T_PRM  / (1-Sqrt(R_PRM * (1-L_PRC)*R_cav))^2

==> loss in the recycling cavity L_PRC: 0.009+/-0.009
       (About 1% loss is likely in the recycling cavity)

Quote:

<<X arm>>
Measured arm reflectivity R_cav: 0.875 +/- 0.005
Estimated round trip loss L_RT: 157ppm +/- 8ppm
Estimated finesse F: 1213+/-2

<<Y arm>>
Measured arm reflectivity R_cav:
0.869 +/- 0.006
Estimated round trip loss L_RT: 166ppm +/- 8ppm
Estimated finesse F: 1211+/-2

 

  1900   Fri Aug 14 02:57:46 2009 ClaraUpdatePEMRedo of the Huddle Test

I put all three seismometers and all six accelerometers together in the foam box with peanuts. Three of the accelerometers are facing in the x-direction and three are in the y-direction. Both Guralps are aligned on the NS axis and the Ranger is pointing vertically.

**EDIT: The accelerometers are in the x and z directions, not x and y. Sorry, I was sleepy when I wrote this.**

One of the accelerometers was refusing to show anything, and after a few hours of checking connections and swapping cables, I discovered that someone had unplugged the cable from the ADC. A quick glance in the dataviewer shows that the channel has been unplugged since about 3 in the afternoon on August 8th (Saturday). So... obviously all the accelerometer measurements made with that channel since then did not actually get recorded. Yay.

Anyway, as of 2:45, everything is working and taking data. Clearly we're not getting a full night's worth... hopefully that's okay.

  16623   Tue Jan 25 16:42:03 2022 AnchalSummaryBHDReduced filter gains in all damped new SOS

I noticed that our current suspension damping loops for the new SOS were railing the DAC outputs. The reason being that cts2um module has not been updated for most optics and thus teh OSEM signal (with the new Sat Amps) is about 30 times stronger. That means our usual intuition of damping gains is too high without applying correct conversion cts2um filter module. I reduced all these gains today and nothing is overflowing the c1su2 chassis now. I also added two options in the "!" (command running drop down menu) in the sus_single medm screens for opening ndscope for monitoring coil outputs or OSEM inputs of the optic whose sus screen is used.

 

  17805   Wed Aug 23 16:52:52 2023 Paco, Radhika, MurtazaUpdateASSReducing XARM-ASS Errors

We're trying to reduce the demodulated error signals after running the ASS script for the XARM.

After running the ASS script, we initially tried to play around with the with the EXC Gain and brought all of them down to 300. It didn't make a huge difference on the error signals or the transmission signal. We then tried tweaking the XARM_OUT_MTRX by flipping the signs/changing the magnitude but it mostly just made things worse. We then changed that matrix to closely resemble the YARM_OUT_MTRX (structurally). At an XARM GAIN of about 0.02, with the EXC Gains at 300 and the XARM_SEN_MTRX having 1.00 on the diagonal terms, the error signals slowly started converging to 0. However, X_ARM_ETM_PIT_L_DEMOD_I_OUT16 kept oscillating which wasn't good.

We later tried looking at the spectrum for the demodulated signals to see if there were any peaks at frequencies outside of the delmodulating frequencies. Most of them looked consistent with peaks at demodulation frequencies (and modes) and signal input frequencies (60Hz and modes). We compared the spectrum with the YARM where everything was optimized, there were no noticable differences.

Later in the day, both XARM and YARM lost lock a couple of times for reasons unknown. We restored to an earlier point in the day (12:00) suspecting there was misalignment with the input optics.

THE XARM ASS MYSTERY REMAINS.

  17820   Fri Sep 1 18:06:35 2023 Paco, Radhika, MurtazaUpdateASSReducing XARM-ASS Errors

[Radhika, Murtaza]

XARM ASS

We resumed playing around with ASS for XARM. We approached each error signal one at a time to try to determine the sign of actuation and ensure the error was reduced.

Steps:

1. Turned on dithering and reduced all excitation amplitudes to 300 cts.
2. Cleared current output matrix to start from scratch.
3. Made all servo control gains positive, for consistency (all YARM ASS gains are +1).
3. Started with the "fast" loop, using transmission error signals to align the cavity.
    a. Used ETM transmission error signal to feedback to ETM - this worked great! Configured as in Attachment 1.
4. Tried to apply same logic to the ITM, but transmission dropped with both choices of sign in output matrix.

Next steps:

1. Resume nailing down the "fast" control by using ITM transmission error signal to feedback to ITM

2. Add in "slow" pointing control by feeding back ETM LSC error (centering) to BS.

     *NOTE* We tried to do this before feeding back any ITM error signal, but this immediately caused transmission to drop because ITM had no way to adjust to new input pointing.

 

  17825   Tue Sep 5 11:04:33 2023 ranaUpdateASSReducing XARM-ASS Errors

I recommend usinng the DC offset method that Koji and I used for measuring the IMC WFS sensing matrix (not the AC method that Anchal used). With a sensing matrix, you should be able to do some partial inversion.

Without any sensing matrix inversion, we would have to rely on a gain hierarchy for getting the loops to work.

With some approximate matrix inversion, the loops are more indepedent of each other. Also if you look at the spectrum of the error signals, it should be clear that the sensing noise is pretty large, and so that sets a natural upper limit to the UGFs. We only want integrator (1/f) loops, but the LPFs cause some extra phase lag.

Quote:

[Radhika, Murtaza]

XARM ASS

  17828   Wed Sep 6 12:53:11 2023 Paco, Radhika, MurtazaUpdateASSReducing XARM-ASS Errors

[Radhika, Murtaza]

To create the sensing matrix, we tried the DC offset method by giving offsets in the Pitch and Yaw DOFs for ITMX, ETMX and the BS respectively. The signals we looked at were the demodulated ETMX_L, ETMX_T and ITMX_T. We wrote a quick notebook that does the following things for each DOF:

1. Calculate the mean error signal (over 10s)
2. Give an offset of 3 steps in each DOF corresponding to their step size serially with some buffer time (restoring the offset after each DOF)
3. Calculate the new mean error signal (over 10s)
4. Find the difference in error signals and divide by their respective step sizes to get each sensor's sensitivity to the offset.
5. Invert to obtain the sensing matrix.

Sensing Matrix for Pitch:

\begin{bmatrix} ETM P \\ ITMP \\ BSP \end{bmatrix} = \begin{bmatrix} 1.63590936e+00 && -9.67386830e+00 && 2.65620052e+00 \\ 7.58125093e-03 && -2.23977732e-02 && 1.02654009e-01 \\ 2.03270795e-02 && -4.60362844e-02 && 3.66501511e-02 \end{bmatrix} \begin{bmatrix} ETM P L err\\ ETMP T err\\ ITM P Terr \end{bmatrix}

Sensing Matrix for Yaw:

\begin{bmatrix} ETM Y \\ ITM Y \\ BS Y \end{bmatrix} = \begin{bmatrix} -4.69197238 && 19.11902408 && -13.96830153 \\ -0.04886671 && 0.103685 && -0.06974976 \\ 0.03923102 && -0.12202067 && 0.10306225 \\ \end{bmatrix}\begin{bmatrix} ETM YLerr \\ ITM YTerr \\ BS YTerr \end{bmatrix}

*NEED TO TRY THESE OUT*

%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%

The Output Matrix for the from intuition was set to *Attachment 2* which improved the net average transmission (see Attachment 1), but wasn't really stable after the improvement.

  17832   Thu Sep 7 18:42:02 2023 Paco, Radhika, MurtazaUpdateASSReducing XARM-ASS Errors

[Radhika, Murtaza]

We recalculated the sensing matrix for XARM ASS by collecting each sensor's step response to an offset in each DOF. This produced the following dense output matrix A (see Attachment 1 for rows/cols):

      [[-0.02047695,  0.        , -0.10262752,  0.        , -0.0157128 , 0.        ],
       [ 0.        ,  0.16908344,  0.        , -0.00929291,  0.        ,-0.35916455],
A =    [-0.28050764,  0.        ,  0.26982002,  0.        , -0.55100297, 0.        ],
       [ 0.        ,  0.85501491,  0.        ,  0.0606197 ,  0.        , 0.27568672],
       [-0.95963335,  0.        ,  0.95742611,  0.        ,  0.83435534, 0.        ],
       [ 0.        , -0.49026554,  0.        , -0.99811768,  0.        , 0.89162641]]

Turning the XASS gain up slowly to ~0.15, we observed that several error signals diverged and transmission started to drop. Debugging this matrix proved difficult since there were many nonzero elements to consider. So we reverted to build the matrix from our intuition, considering the centering and input pointing loops, and using the YASS output matrix as a reference.

The YARM ASS servo gains are all +1. The YASS output matrix has the following length (centering) signal mapping:

ITM PIT/YAW L ----> ETM feedback
(ETM PIT/YAW L - ITM PIT/YAW L) ----> ITM feedback

We mirrored this in the XASS output matrix. Note that previously the ITM L error signals were not used for XASS. To simplify the process, we decided to just work out the beam centering first and ignore the input pointing coming from the beam splitter (setting BS PIT/YAW matrix elements to 0). We also set all the XARM ASS servo gains to +1. See the output matrix below:

We cleared the outputs and turned on the XARM GAIN slowly (0.1) and immediately noticed the YAW signals in ETM start to diverge (C1:ASS-XARM_ETM_YAW_T_DEMOD_I_OUT16, C1:ASS-XARM_ETM_YAW_L_DEMOD_I_OUT16). We turned down the XARM gain and flipped the sign for the signal going to ETM YAW. (suspect a difference in sign convention).

To check the stability, we sequentially gave offsets in PIT/YAW for ETM and ITM. We saw the signal (C1:ASS-XARM_ETM_PIT_L_DEMOD_I_OUT16) oscillate wildly at a frequency of ~(1/15 Hz). We suspected the ASS loop was driving these oscillations so we turned down the gain going to ETM PIT to 0.25 which worked really well and the transient oscillation of further checks was gone.

We saw similar wild oscillatory signals in ITM PIT (C1:ASS-XARM_ITM_PIT_T_DEMOD_I_OUT16, C1:ASS-XARM_ITM_PIT_L_DEMOD_I_OUT16) on applying offsets so we reduced the gain going to ITM PIT to 0.3. (0.25 and 0.3 are arbitary relatively smaller weights, can be fine tuned).

We checked the stability of this setup as a whole by giving a few offsets to ITMX and ETMX, with a servo gain of 0.15 it did a great job! (0.25 made it diverge once again). See final state for centering in Attachment 1, and error signal suppression in Attachment 2. (Ignore XAUX transmission in grey - we were toggling the shutter.) Note that the Length error signals were successfully suppressed, but the dark green/brown Transmission error signals were not fed back and thus remain nonzero.

WE SHALL INVESTIGATE THE INPUT POINTING NEXT FEEDING BACK TO THE BS and ITMX. We will give an update shortly about whether restoring XARM ASS is feasible by Monday.

  17835   Sat Sep 9 16:25:07 2023 RadhikaUpdateASSReducing XARM-ASS Errors

[Radhika, Murtaza]

This post summarizes XARM ASS efforts from Friday 9/8 and Saturday 9/9.

FRIDAY

On Friday, we continued with our previous output matrix that used the length error signals (ITM/ETM PIT/YAW L) to feed back to ITMX and ETMX (see the previous ELOG). In that state we did not use the transmission error signals and had no feedback going to the BS. We then tried to use the transmission error signals ITM PIT/YAW L as a proxy for BS input pointing and feed them back to the BS. For both PIT and YAW, both signs of feedback resulted in diverging T error signals and a decrease in transmission.

SATURDAY

On Saturday, we used the transmission error signals (ITM/ETM PIT/YAW T) in the sensing matrix to build the output matrix. We got it to a state where we could get the controlled error signals to converge by just feeding back to the ITMX and ETMX (Attachments 1,2). Once we had this working, we tried to feed back a combination of (ETM PIT/YAW L and ITM PIT/YAW T) to correct BS pointing. However, any combinations and signs to the BS dropped transmission and led to diverging error signals.
We then attempted to use the latest working XASS output matrix (before the acromags were pulled out) and see the effect of flipping signs in there (one optic+DOF at a time) We then tried to use the sign logic from the previously working ETM/ITM feedback we got partially working; however the error signals did not converge with any combination.

SUMMARY:

- We are able to successfully feed back to ITMX and ETMX, using either length or transmission error signals. It is when we try to add BS feedback that ASS fails. This can be due to the fact that we need to consider the relative servo gains when treating these loops separately, like Koji mentioned.

- The sensing matrix approach might be the only way to simultaneously optimize feedback for all optics, avoiding the need to tune servo gains. We will revisit this approach on Monday.

     - Koji pointed out that we are reading out the low-passed error signals in order to calculate each step response - we will need to consider our sampling rate and duration of averaging accordingly.

     - It will be harder to iteratively flip signs of each matrix element for this dense matrix, and we will have to be clever about which sign combinations we try for actuation.

  17836   Mon Sep 11 19:35:27 2023 RadhikaUpdateASSReducing XARM-ASS Errors

[Radhika, Murtaza, Paco]

Today we decided to take a closer look at the demod phases of the T and L error signals for XARM ASS. By eye we tuned the phases to minimize the signal in Q. Here are the new demod phases:
(THE DEMODULATION PHASE VALUE DO NOT RESTORE BACK TO THE ORIGINAL VALUES WHEN DITHER IS TURNED ON.)

C1:ASS-XARM_ETM_PIT_L_DEMOD_PHASE:  15 -> 35
C1:ASS-XARM_ETM_YAW_L_DEMOD_PHASE: 176 -> 180
C1:ASS-XARM_ITM_PIT_L_DEMOD_PHASE:   0 -> -5
C1:ASS-XARM_ITM_YAW_L_DEMOD_PHASE:  10 -> -10
C1:ASS-XARM_ETM_PIT_T_DEMOD_PHASE:  10 -> -3.5
C1:ASS-XARM_ETM_YAW_T_DEMOD_PHASE: -10 -> -5
C1:ASS-XARM_ITM_PIT_T_DEMOD_PHASE:   0 -> -15
C1:ASS-XARM_ITM_YAW_T_DEMOD_PHASE:  -5 -> 30

We also noticed that MEDM indicator for dithering on (white --> green LO symbol) for ETM_YAW_L_OSC was tied to the wrong excitation gain channel (C1:ASS-XARM_ITM_YAW_OSC_CLKGAIN instead of C1:ASS-XARM_ETM_YAW_OSC_CLKGAIN). We went ahead and changed this in [insert medm file location]. So now the right green LO symbol appears when the appropriate excitation is turned on.

  17903   Tue Oct 17 19:35:31 2023 RadhikaUpdateASSReducing XARM-ASS Errors

I resumed work to restore XARM ASS

I manually aligned XARM and calculated the output matrix using scripts/ASS/getASSOutMat.py. This script gives an offset to the ETM/ITM/BS in pitch and yaw and records all error signals to construct a sensing matrix (8 err signals x 6 actuators). Then the ITMX PIT/YAW L error signal rows are removed (as per XASS historical practice) and the resulting square matrix is inverted.

A few times I loaded the output matrix and turned up the ASS servo gain, but this caused error signals to diverge and transmission to drop. I realigned XARM in between each time and recalcuated the output matrix until one looked promising [Attachment 1].

- From a decently aligned state, the servo maintained transmission and controlled error signals began to converge to 0.

- However when I gave an offset to some optic, transmission started to recover but then dropped even as all controlled error signals were minimized [Attachment 2]. A zoomed version of this is in Attachment 3.

     - ITMX PIT L (light blue) / ITMX YAW L (light pink) are uncontrolled

     - ETMX PIT L (dark blue) has a ~0.2 Hz oscillation, but its mean error is roughly 0. This behavior has been present since XASS stopped working, but it is yet to be understood.

The output matrix is successfully reducing the error signals it controls, but transmission is not maximized. I wonder if it's stuck in a local minimum, but I haven't convinced myself that these error signals can be 0 when the beam spots aren't centeredand cavity/beam axes aren't matched. Maybe we need to use the ITMX PIT/YAW L error signals.

Things to try:

1. Assume pitch and yaw are decoupled and calculate a checkerboard output matrix, and debug each angular DOF at a time. I briefly did this today and got the pitch loop to increase transmission, but not the yaw loop.

2. Altering the sensing matrix to include ITMX PIT/YAW L error signals. Then I'll use the same script to compute the output matrix.

  5096   Tue Aug 2 17:40:04 2011 JennyUpdatePSLReducing beam intensity incident on photodiode

I am using a PDA255 photodiode to measure the power outputted by the NPRO beam on the PSL table. (I'm going to then use a network analyzer to measure the amplitude response of the PZT to being driven at a range of frequencies. I'll detect the variation in in response to changing the driving frequency using this PDA255.)

The PDA255 has an active area of 0.8mm^2 and a maximum intensity for which the response is linear of 10mW/cm^2. This means that a beam I focus on the PD must have a power less than 0.08 mW (and even less if the spot size is smaller than the window size).

I used a power meter to measure the beam power and found it was 0.381 mW.

The second polarizing beam splitter in the setup transmits most of the beam power, but reflects 0.04 mW (according to the power meter). I'm going to place the photodiode there in the path of the reflected beam.

  741   Fri Jul 25 19:57:18 2008 JenneUpdatePSLRef Cav & PMC
"PMC is in, but is still being worked on. Leave it alone." ---Rana

Ref. Cavity is locked again. Still a work in progress. I think we're ready to mode match on Monday. ---Jenne
  1847   Thu Aug 6 18:26:26 2009 JenneUpdatePSLRef Cav and PMC aligned

[Alberto, Jenne]

We aligned both the reference cavity and the PMC, each by looking at their Trans PD on Davaviewer, and adjusting the two steering mirrors to maximize the transmission power.  We got a pretty good amount of improvement for the ref cav, but since the PMC hasn't decayed a whole lot, we got a much smaller amount of improvement.

  1850   Thu Aug 6 23:29:47 2009 JenneUpdatePSLRef cav reflection PD is funky

After Alberto and I worked on aligning the reference cavity, Rob asked the important and useful question: what is the visibility of the reference cavity.  This helps tell us if we're optimally aligned or not even close.

I did a scan of the ref cav temperature, using /scripts/PSL/FSS/SLOWscan, but there seems to be no real signal is C1:PSL-FSS_RFPDDC.  As shown in Alberto's 200-day plot, it does change sometimes, but if you zoom in on the flat parts, it seems like it's not really reading anything meaningful.  I did a cursory check-out of it, but I'm not 100% sure where to go from here:  There are (as with all of these gold-box PDs) 3 outputs:  a ribbon cable (for ADC purposes I think), an SMA for the RF signal, and a BNC for the DC signal.  The photodiode is clearly working, since if you stick the Lollypop in front of the PD, the cavity unlocks.  I plugged a 'scope into the DC BNC, and it also behaves as expected: block the beam and the signal goes down; unblock the beam and the signal goes up.  Something of note is that this readout gives a positive voltage, which decreases when the beam is blocked.  However, looking at the dataviewer channel, nothing at all seems to happen when the beam is blocked/unblocked.  So the problem lies somewhere in the get-signal-to-DAQ path.  I unplugged and replugged in the ribbon cable, and the value at which the channel has been stuck changed.  Many days ago, the value was -0.5, for the last few days it's been -1.5, and after my unplug/replug, it's now back to ~ -0.5 . The other day Alberto mentioned, and made the point again today that it's a little weird that the PD reads out a negative voltage.  Hmm.

 

Do we have a tester-cable, so that instead of the ribbon cable, I can plug that connector (or pins thereof) into a 'scope?

  1851   Fri Aug 7 00:10:14 2009 ranaUpdatePSLRef cav reflection PD is funky

we have a tester cable, but you don't want it. Instead the problem is probably at the cross-connect. The D-cable goes to a cross-connect and you can probe there with a voltmeter. If the signal is good there, trace it to the ADC. Also trend for several years to see when this happened - Yoichi may know the history better.

Also, we still need to complete the FSS RFPD task list from last year.

 

  1860   Fri Aug 7 17:05:34 2009 JenneUpdatePSLRef cav reflection PD is funky

Quote:

we have a tester cable, but you don't want it. Instead the problem is probably at the cross-connect. The D-cable goes to a cross-connect and you can probe there with a voltmeter. If the signal is good there, trace it to the ADC. Also trend for several years to see when this happened - Yoichi may know the history better.

Also, we still need to complete the FSS RFPD task list from last year.

 

 [Jenne, Ben]

I called in the reinforcements today.  Ben came over and we looked all around at all of the cross-connects and cables relating to the FSS.  Everything looks pretty much okey-dokey, except that we still weren't getting signal in the DataViewer channels.  Finally we looked at the psl.db file, which indicates that the C1:PSL-FSS_RFPDDC channel looks at channel 21 of the ADC cross connect thing.  We followed the cable which was plugged into this, and it led to a cable which was disconnected, but laying right next to the Ref Cav refl PD.  We plugged this into the DC out SMA connection of the photodiode (which had not been connected to anything), and suddenly everything was mostly golden again in dataviewer land.  RFPDDC_F now has a signal, but RFPDDC is still flat. 

 

Even though this seems to be working now, it's still not perfect.  Rob suggested that instead of having this SMA cable going from the photodiode's DC out, we should take the signal from the ribbon cable.  So I'm going to figure out which pin of the D-connector is the DC out, and take that from the cross connect to the ADC cross connect.  This will help avoid some persnickity ground loops. 

  5072   Sat Jul 30 20:41:50 2011 ranaUpdatePSLRefCav Stabilization back on

 Untitled.png

I turned the RefCav heater and servo back on a couple days ago. At first it was stabilizing again at a low setpoint, but in reality the right temperature (~40 C). After fixing the in-loop signal offsets, the setpoint now correctly reflects the actual temperature.

Jenny is going to calibrate the sensors using some kind of dunking cannister next week.

  12573   Wed Oct 19 18:32:25 2016 rana, yinziUpdatePSLRefCav thermal control: heater is dead

We wanted to re-activate the Heater for the reference cavity today to use it as a testbed for PID autotuning and the new heater driver circuit that Andrew is working on for the coating thermal noise experiment.

Unfortunately, it seems that the large power supply which is used for the heater is dead.sad Or maybe I don't remember how to use it?

The AC power cord was plugged in to a power strip which seems to work for IO chassis. We also tried swapping power strip ports.

We checked the front panel fuses. The power one was 3 Ohms and the 'bias' one was 55 Ohms. We also checked that the EPICS slider did, in fact, make voltage changes at the bias control input.

Non of the front panel lights come onfrown, but I also don't remember if that is normal.

Have those lights been dead a long time? We also reconnected the heater cable at the reference cavity side.

  1001   Fri Sep 26 19:08:43 2008 ranaConfigurationPSLRefcav Trans: PD + Camera + Dumps
I went out to improve the Refcav trans path.

I removed all ND filters to get rid of the fringing.

I removed the anodized Al dump that was there. Black anodized Aluminum dumps are forbidden for use as
dumps in any low phase noise setup (such as our frequency stabilization cavity). The scatter was going
directly back into the cavity and making noise. For now its undumped, but Steve will find the
reflections and dump them on unblemished razor blade dumps mounted stiffly.

I will post a photo of the new setup later - the new setup is sketched on the control room markerboard.

The transPD level is now 8 V, up from its previous 3-4 V. We will probably have to also put a lens
in front of it to get the beam size down.
  1923   Tue Aug 18 14:24:43 2009 YoichiSummaryPSLReference Cavity Inspection
Rana, Koji, Yoichi

To see why the reflected beam from the RC is distorted, we took out
the periscope and the iris in front of the RC. The periscope mirrors
had some gunk and dusts on them. We blew nitrogen air onto them to
remove the dust. Since the gunk did not come off with the air, we
wrapped a Q-tip with lens cleaning paper soaked in Methanol, and wiped
the surface of the mirrors. We did this because it was hard to remove
the mirrors from the periscope (they were in a spring loaded mirror
holders. The springs were too strong to safely remove them without
damaging the mirrors).

Looking into the RC from the front mirror revealed nothing obstructing
in the path.

After the cleaning, we put the periscope back and observed the direct
reflection from the cavity (not locked). It was still distorted
exactly like before.

So we did some tests.
First we injected He-Ne to the RC. It turned out that multiple
reflections from the optical window (not AR coated for He-Ne) made it
almost impossible to investigate anything with He-Ne. But this
observation made us to suspect maybe one side of the window is not AR
coated.

We placed the periscope about 50cm away from the RC and injected the
beam from an angle, so that we can observe the direct reflection.

First, we checked the shape of the beam leaving the periscope. It was good.
We then observed the reflected beam from the RC. It was also good, no distortion.
We made sure that it was really the reflection from the mirror, not from the window
as follows.
We measured the separation between the in coming beam to the cavity and the reflected beam
at two locations. From this, we can guess where the two beams intersect (the reflection point).
The estimated reflection point was far inside the RC enclosure, indicating that it was really
reflection from the front mirror of the RC.
Since we did not see any other reflection beam, we concluded that the AR coating of the window
is good.

We checked the direct reflection beam shape with several different incident angles, but the
beam shape was always good.

We put back the periscope to the original position. This time, we put a high reflectivity mirror
after the output mirror of the periscope. The beam coming out of the circulator (PBS) had a good
circular shape. But still if we let the beam reflected by the cavity, the beam shape is distorted.
Something must be happening in the RC. Unfortunately, we could not figure out what it is.

We put everything back to the original configuration, except for the iris, and the RC alignment
was already good (surprise). After Koji's final tweak, the FSS is now doing fine, but still
the reflected beam is ugly.
  3240   Fri Jul 16 20:25:52 2010 MeganUpdatePSLReference Cavity Insulation

Rana and I

1) took the temperature sensors off the reference cavity;

2) wrapped copper foil around the cavity (during which I learned it is REALLY easy to cut hands with the foil);

3) wrapped electrical tape around the power terminals of the temperature sensors (color-coded, too! Red for the out of loop sensor, Blue for the first one, Brown for the second, Gray for the third, and Violet for the fourth. Yes, we went with an alphabetical coding system, excluding the out of loop sensor);

4) re-wrapped the thermal blanket heater;

5) covered the ends of the cavities with copper, ensuring that the beam can enter and exit;

6) took pretty pictures for your enjoyment!

We will see if this helps the temperature stabilization of the reference cavity.

 

DSC_2271.JPG

The end of the reference cavity, with a lovely square around the beam.

 

DSC_2266.JPG

The entire, well-wrapped reference cavity!

  3241   Fri Jul 16 23:53:27 2010 RanaUpdatePSLReference Cavity Insulation

From the trend, it seems that the Reference Cavity's temperature servo is working fine with the new copper foil. I was unable to find the insulating foam anywhere, but that's OK. We'll just get Frank to make us a new insulation with his special yellow stuff.

The copper foil that Steve got is just the right thickness for making it easy to form around the vacuum can, but we just have to have the patience to wrap ~5-10 more layers on there. We also have to get a new heater jacket; this one barely fits around the outside of the copper wrap. The one we have now seems to have a good heating wire pattern, but I don't know where we can buy these.

I also turned the HEPA's Variac back down to the nominal value of 20. Please remember to turn it back up to 100 before working on the PSL.

  3280   Fri Jul 23 16:02:16 2010 RanaUpdatePSLReference Cavity Insulation

This is the trend so far with the copper foil wrapping. According to Megan's calculation, we need ~1 mm of foil and the thickness of each layer is 0.002" (1/20th of a mm), so we need ~20 layers. We have ~5 layers so far.

As you can see the out-of-loop temperature sensor (RCTEMP) is much better than before. We need another week to tell how well the frequency is doing -

the recent spate of power cycles / reboots of the PSL have interrupted the trend smoothness so far.

  3282   Fri Jul 23 21:14:29 2010 RanaUpdatePSLReference Cavity Insulation

I wrapped another ~3 layers onto there. It occurs to me now that we could just get some 2mm thick copper plates made to fit over the stainless steel can.

They don't have to completely cover it, just mostly. I also took the copper circles that Steve had made and marked them with the correct beam height

and put them on Steve's desk. We need a 1" dia. hole cut into these on Monday.

To compensate for the cooling during my work, I've set the heater for max heating for 1 hour and then to engage the temperature servo.

I also noticed the HEPA VARIAC on the PSL was set to 100. Please set it back to 20 after completing your PSL work so that it doesn't disturb the RC temperature..

  2732   Mon Mar 29 21:43:27 2010 AlbertoConfigurationPSLReference Cavity PD Noise Spectrum

[Rana, Alberto]

This evening we measured the noise spectrum of the reference cavity PD used in the FSS loop. From that we estimated the transimpedance and found that the PD is shot-noise limited. We also found a big AM oscillation in correspondence of the FSS modulation sideband which we later attenuated at least in part.

This plot shows the spectrum noise from the RF output of the photodetector.
 
 (here you should be able to see an attached figure, if not it's probably becasue imagemagic has having problems with displaying png files)
2010-03-29_FSS_PD_shotnoise_and_darknoise.png
 
The tall peak at 21.5 MHz is the AM modulation introduced by the EOM. It seems to be caused by a misalignment of the EOM which might be somehow modulating the polarization.
The mount in which the EOM sits is not very solid. We should change it with something similar to that of the other two EOMs in the Mach Zehnder.
By tightening down the plastic screws of the mount Rana reduced the amplitude of the AM modulation by 20dB.
 
The bump in both the dark and shot noise are in corrispondence of the resonance of the PD's electronics. As it appears, the electronics is not well tuned: the bump should coincide with the AM peak.
 
In the case of the dark noise spectrum, the bump is due to the thermal noise of the electronics. It's a good sign: it means that the electronics is good enough to be sensitive to it.
 
Transimpedance Estimate
As a "sanity check" we made an approximate estimate of the transimpedance to make sure that the PD is dominated by shot noise rather than other noises, ie electronic's noise.
 
  1. Supposing that the laser beam hitting the PD was shot noise limited, we measured 1.1V at the DC output. That let us estimate the photocurrent at DC of 20mA, for a 50Ohm output impedance.
  2. The shot noise for 20mA is 80 pA/rtHz
  3. From the nosie spectrum, we measured 3e-7 v/rtHz at 21.5MHz
  4. The impedance at RF is then Z_rf = 3e-7 V/rtHz / 80e-12 pA ~ 4000 Ohm
  5. Since the RF path inside the PD has a gain of 10, the transimpedance is ~400Ohm, which is about as we (ie Rana) remembered it to be.
  6. The PD seems to be working fine.
  2733   Tue Mar 30 06:37:32 2010 ranaConfigurationPSLReference Cavity PD Noise Spectrum

Some more words about the RFAM: I noticed that there was an excess RFAM by unlocking the RC and just looking at the RF out with the 50 Ohm input of the scope. It was ~100 mVp-p! In the end our method to minimize the AM was not so sensible - we aligned the waveplate before the EOM so as to minimize the p-pol light transmitted by the PBS cube just ahead of the AOM. At first, this did not minimize the RFAM. But after I got angry at the bad plastic mounting of the EOM and re-aligned it, the AM seemed to be small with the polarization aligned to the cube. It was too small to measure on the scope and on the spectrum analyzer, the peak was hopping around by ~10-20 dB on a few second timescale. Further reduction would require some kind of active temperature stabilization of the EOM housing (maybe a good SURF project!).

For the EOM mount we (meaning Steve) should replace the lame 2-post system that's in there with one of the mounts of the type that is used in the Mach-Zucker EOMs. I think we have spare in the cabinet next to one of the arms.

After the RFAM monkeying, I aligned the beam to the RC using the standard, 2-mirror, beam-walking approach. You can see from the attached plot that the transmission went up by ~20% ! And the reflection went down by ~30%. I doubt that I have developed any new alignment technique beyond what Yoichi and I already did last time. Most likely there was some beam shape corruption in the EOM, or the RFAM was causing us to lock far off the fringe. Now the reflected beam from the reference cavity is a nice donut shape and we could even make it better by doing some mode matching! This finally solves the eternal mystery of the bad REFL beam (or at least sweeps it under the rug).

At the end, I also fixed the alignment of the RFPD. It should be set so the incident angle of the beam is ~20-40 deg, but it was instead set to be near normal incidence ?! Its also on flimsy plastic legs. Steve, can you please replace this with the new brass ones?

  2759   Sat Apr 3 11:35:47 2010 ranaConfigurationPSLReference Cavity PD Noise Spectrum

The units on this plot are completely bogus - we know that the thermal noise from the resonant part of the circuit is just V = sqrt(4*k*T*Z) ~ 3nV/rHz. Then the gain of the MAX4107 stage is 10. The output resistor is 50 Ohms, which forms a divide by 2 with the input impedance of the spectrum analyzer and so the bump in the dark noise should only be 15 nV/rHz and not microVolts.

Quote:

[Rana, Alberto]

This evening we measured the noise spectrum of the reference cavity PD used in the FSS loop. From that we estimated the transimpedance and found that the PD is shot-noise limited. We also found a big AM oscillation in correspondence of the FSS modulation sideband which we later attenuated at least in part.

This plot shows the spectrum noise from the RF output of the photodetector.

  2760   Sat Apr 3 16:07:40 2010 AlbertoConfigurationPSLReference Cavity PD Noise Spectrum

 I was aware of a problem on those units since I acquired the data. Then it wasn't totally clear to me which were the units of the data as downloaded from the Agilent 4395A, and, in part, still isn't.

It's clear that the data was in units of spectrum, an not spectral density: in between the two there is a division by the bandwidth (100KHz, in this case). Correcting for that, one gets the following plot for the FSS PD:

2010-03-29_FSS_PD_shotnoise_and_darknoise.png

Although the reason why I was hesitating to elog this other plot is that it looks like there's still a discrepancy of about 0.5dBm between what one reads on the display of the spectrum analyzer and the data values downloaded from it.

However I well know that, I should have just posted it, including my reserves about that possible offset (as I'm doing now).

Quote:

The units on this plot are completely bogus - we know that the thermal noise from the resonant part of the circuit is just V = sqrt(4*k*T*Z) ~ 3nV/rHz. Then the gain of the MAX4107 stage is 10. The output resistor is 50 Ohms, which forms a divide by 2 with the input impedance of the spectrum analyzer and so the bump in the dark noise should only be 15 nV/rHz and not microVolts.

Quote:

[Rana, Alberto]

This evening we measured the noise spectrum of the reference cavity PD used in the FSS loop. From that we estimated the transimpedance and found that the PD is shot-noise limited. We also found a big AM oscillation in correspondence of the FSS modulation sideband which we later attenuated at least in part.

This plot shows the spectrum noise from the RF output of the photodetector.

  2742   Wed Mar 31 15:31:53 2010 steveUpdatePSLReference Cavity RF PD base upgraded

Quote:

Some more words about the RFAM: I noticed that there was an excess RFAM by unlocking the RC and just looking at the RF out with the 50 Ohm input of the scope. It was ~100 mVp-p! In the end our method to minimize the AM was not so sensible - we aligned the waveplate before the EOM so as to minimize the p-pol light transmitted by the PBS cube just ahead of the AOM. At first, this did not minimize the RFAM. But after I got angry at the bad plastic mounting of the EOM and re-aligned it, the AM seemed to be small with the polarization aligned to the cube. It was too small to measure on the scope and on the spectrum analyzer, the peak was hopping around by ~10-20 dB on a few second timescale. Further reduction would require some kind of active temperature stabilization of the EOM housing (maybe a good SURF project!).

For the EOM mount we (meaning Steve) should replace the lame 2-post system that's in there with one of the mounts of the type that is used in the Mach-Zucker EOMs. I think we have spare in the cabinet next to one of the arms.

After the RFAM monkeying, I aligned the beam to the RC using the standard, 2-mirror, beam-walking approach. You can see from the attached plot that the transmission went up by ~20% ! And the reflection went down by ~30%. I doubt that I have developed any new alignment technique beyond what Yoichi and I already did last time. Most likely there was some beam shape corruption in the EOM, or the RFAM was causing us to lock far off the fringe. Now the reflected beam from the reference cavity is a nice donut shape and we could even make it better by doing some mode matching! This finally solves the eternal mystery of the bad REFL beam (or at least sweeps it under the rug).

At the end, I also fixed the alignment of the RFPD. It should be set so the incident angle of the beam is ~20-40 deg, but it was instead set to be near normal incidence ?! Its also on flimsy plastic legs. Steve, can you please replace this with the new brass ones?

 Teflon feet removed and heavy brass-delrin pd base installed. Ref-cavity reflected light remains to be beautiful doughnut shape on camera.

  1954   Wed Aug 26 19:58:14 2009 Rana, AlbertoUpdatePSLReference Cavity Temperature Control: MINCO PID removed

Summary: This afternoon we managed to get the temperature control of the reference cavity working again.

We bypassed the MINCO PID by connecting the temperature box error signal directly into EPICS.

We couldn't configure the PID so that it worked with the modified temperature box so we decided to just avoid using it.

Now the temperature control is done by a software servo by using the channel C1:PSL-FSS_MINCOMEAS as error signal and driving C1:PSL-FSS_TIDALSET (which we have clip-doodle wired directly to the heater input).

 

We 'successfully' used ezcaservo to stabilize the temperature:

ezcaservo -r C1:PSL-FSS_MINCOMEAS -s 26.6 -g -0.00003 C1:PSL-FSS_TIDALSET

 

We also recalibrated the channels:

C1:PSL-FSS_RMTEMP

C1:PSL-FSS_RCTEMP

C1:PSL-FSS_MINCOMEAS

with Peter King on the phone by using ezcawrite (EGUF and EGUL) but we didn't change the database yet. So please do not reboot the PSL computer until we update the database.

 

More details will follow.

  1956   Thu Aug 27 13:42:08 2009 ranaSummaryPSLReference Cavity Temperature Control: psl.db changes

I made the changes to the psl.db to handle the new Temperature box hardware. The calibrations (EGUF/EGUL) are just copied directly from the LHO .db file (I have rsync'd their entire target area to here).

allegra:c1psl>diff psl.db~ psl.db
341,353d340
< grecord(ai,"C1:PSL-FSS_TIDALOUT")
< {
<       field(DESC,"TIDALOUT- drive to the reference cavity heater")
<       field(DISV,"1")
<         field(SCAN,".5 second")
<       field(DTYP,"VMIVME-3113")
<       field(INP,"#C0 S28 @")
<       field(EGUF,"10")
<       field(EGUL,"-10")
<       field(EGU,"volts")
<       field(LOPR,"-10")
<       field(AOFF,"0")
< }
493,494c480,481
<         field(EGUF,"285.675")
<         field(EGUL,"-214.325")
---
>         field(EGUF,"67.02")
>         field(EGUL,"7.96")
508,509c495,496
<         field(EGUF,"726.85")
<         field(EGUL,"-1273.15")
---
>         field(EGUF,"75.57")
>         field(EGUL,"12.31")
531,532c518,519
<         field(EGUF,"726.85")
<         field(EGUL,"-1273.15")
---
>         field(EGUF,"75.57")
>         field(EGUL,"12.31")
605,617d591
< grecord(ai,"C1:PSL-FSS_TIDALINPUT")
< {
<       field(DESC,"TIDALINPUT- tidal actuator input")
<       field(DISV,"1")
<         field(SCAN,".5 second")
<       field(DTYP,"VMIVME-3123")
<       field(INP,"#C0 S3 @")
<       field(EGUF,"10")
<       field(EGUL,"-10")
<       field(EGU,"volts")
<       field(LOPR,"-10")
<       field(AOFF,"0")
< }
1130a1105,1130
> grecord(ai,"C1:PSL-FSS_TIDALINPUT")
> {
>       field(DESC,"TIDALINPUT- tidal actuator input")
>       field(DISV,"1")
>         field(SCAN,".5 second")
>       field(DTYP,"VMIVME-3123")
>       field(INP,"#C0 S3 @")
>       field(EGUF,"10")
>       field(EGUL,"-10")
>       field(EGU,"volts")
>       field(LOPR,"-10")
>       field(AOFF,"0")
> }
> grecord(ai,"C1:PSL-FSS_TIDALOUT")
> {
>       field(DESC,"TIDALOUT- drive to the reference cavity heater")
>       field(DISV,"1")
>         field(SCAN,".5 second")
>       field(DTYP,"VMIVME-3113")
>       field(INP,"#C0 S28 @")
>       field(EGUF,"10")
>       field(EGUL,"-10")
>       field(EGU,"volts")
>       field(LOPR,"-10")
>       field(AOFF,"0")
> }
1143,1144c1143,1144
<         field(HOPR,"0.010")
<         field(LOPR,"-0.010")
---
>         field(HOPR,"2")
>         field(LOPR,"0")

  1018   Wed Oct 1 23:21:03 2008 YoichiConfigurationPSLReference cavity reflection camera
I re-centered the reference cavity reflection camera, which has been mis-aligned for a while.
I also tweaked an input steering mirror to make the alignment better. This resulted in the increase of the transmission PD voltage
from 8V to 9V.
  1915   Mon Aug 17 02:05:49 2009 Yoichi,ranaUpdatePSLReference cavity reflection looks bad
Rana, Yoichi

It has been a well known fact that the reference cavity reflection beam looks ugly.

We measured the visibility of the RC by locking and unlocking it.
Comparing the reflected beam powers, we got the visibility of 0.46,
which is pretty bad.

The beam going into the RC looks fine (circular on a sensor card).
However, the beam reflected back from the RC is distorted into a
horizontal ellipse, even when the RC is not locked.

We took a picture of the reflected beam hitting a white paper with the
infrared camera (see the attachment). It looks like two overlapping
circles horizontally separated. Could it be a badly coated optics
producing a secondary reflection ?

We looked into the RC's front mirror with an inspection mirror, but we
could not identify any obstructing object.

Rana is now touching the RC alignment.

We plan to remove the periscope before the RC to have a better look
into the cavity for inspection.


Late breaking update:
- We also moved the Refcav reflection camera to look at the leakage through a reflection steering mirror so that there's less chance of distortion. There was previously a W1 window in there as a pickofff. Also changed the camera to autogain so that we can see something.

- Re-aligned onto the refl pd.

- Tweaked alignment into RC. Mainly in yaw. Transmission went from 5V to 7V. In your face, Aso!
  1917   Mon Aug 17 04:16:13 2009 YoichiUpdatePSLReference cavity reflection looks bad

Quote:
Rana, Yoichi
- We also moved the Refcav reflection camera to look at the leakage through a reflection steering mirror so that there's less chance of distortion. There was previously a W1 window in there as a pickofff. Also changed the camera to autogain so that we can see something.

- Re-aligned onto the refl pd.

- Tweaked alignment into RC. Mainly in yaw. Transmission went from 5V to 7V. In your face, Aso!


After our removal of the pick off window and Rana's re-alignment of the beam into the RC, the RC optical gain increased.
FSS was complaining about it by driving the PC feedback crazy.
I reduced the nominal common gain from 12.5dB to 11dB.
  1136   Fri Nov 14 19:20:42 2008 YoichiUpdatePSLReference cavity ring down
Thanks to Bob making the high-voltage BNC cables for the HV pulse generator, I was able to operate the EOM in front of
the reference cavity.

The conceptual setup is the following:
[HV pulse] ----+           +-->-- [PD2]
               V           |
->--[HWP]->-- [EOM] -->-- [PBS] --<->-- [QWP] --<->-- [Reference Cavity] -->-- [PD1]
                           |
                [PD3] --<--+

The high voltage pulse rotates the polarization of the light after the EOM. When the HV is applied, the PBS reflects most of the light
into PD2 (Thorlabs PDA255), shutting down the incident light into the cavity.
The transmitted light power of the reference cavity is monitored by PD1 (PDA255). The reflected light from the reference cavity
is monitored by the DC output of the RF PD (PD3). PD3 is low-passed so the response is not fast.
Thorlabs says PDA255 has 50MHz bandwidth.

The attached plot shows the time series of the above PD signals when the HV was applied.
Input Pulse (blue curve) is the input to the HV pulse generator. When it is high, the HV is applied.
"PBS reflection" (red) is PD2. "Reflection" (green) is PD3. "Transmission" (light blue) is PD1.

The red curve shows huge ringing. At first I thought this was caused by the bad response of the PD.
However, the same ringing can be seen in the PD3 and the peaks match very well.
When red curve goes down the green curve goes up, which is consistent with the energy conservation.
So it looks like the light power is actually exhibiting this ringing.
May be the HV pulse is distorted and the voltage across the EOM is showing this ringing.
I will check the input voltage shape to the EOM using a high impedance probe, if possible.

The green curve shows a slow decay because it has a long time constant. It is not an actual
trend of the reflected light power.

The RC transmission power shows some peaks, probably due to the ringing in the input power.
So just fitting with an exponential would not give a good estimate of the cavity pole.
Even though, we should be able to de-convolute the frequency response of the reference cavity
from the input (red curve) and output (light blue curve) signals.
  1137   Fri Nov 14 20:35:47 2008 ranaUpdatePSLReference cavity ring down
To make the DEI pulser make a fast pulse on the EO shutter EOMs, we had to make sure:

1) the cable had a high voltage rated dielectric. cheap dielectrics show the 'corona'
effect, especially when there is a bend in the cable.

2) the EO has to have a resistor on it to prevent ringing due to the impedance mismatch.

3) We needed ~3.5 kV to get the EO shutter crystal to flip the light by 90 deg.
  1138   Fri Nov 14 22:40:51 2008 YoichiUpdatePSLReference cavity ring down

Quote:

To make the DEI pulser make a fast pulse on the EO shutter EOMs, we had to make sure:

1) the cable had a high voltage rated dielectric. cheap dielectrics show the 'corona'
effect, especially when there is a bend in the cable.


I'll check it with Bob.


Quote:

2) the EO has to have a resistor on it to prevent ringing due to the impedance mismatch.


Did you use a shunt or series resistor ?
If shunt, I guess it has to have a huge heat sink.
Actually, DEI says the pulser does not require any external shunt/series resistors or impedance-matching network.
Looks like it is not true ...


Quote:

3) We needed ~3.5 kV to get the EO shutter crystal to flip the light by 90 deg.


Yes, I adjusted the voltage to maximize the power change and it was about 3.5kV.
  1140   Mon Nov 17 15:07:06 2008 YoichiUpdatePSLReference cavity ring down
I used MATLAB's system identification tool box to estimate the response of the reference cavity, i.e. cavity pole.
What I did was basically to estimate a model of the RC using the time series of the measured input and output power.

First, I prepared the input and output time series for model estimation.
The input is the input power to the RC, which I produced by inverting the PBS reflected light power and adding an offset
so that the signal is zero at t=0. Offset removal was necessary to make sure that the input time series does not give an
unintentional step at t=0.
The output time series is the transmission power of the RC. I also added an offset to make it zero at t=0.
Then I commanded MATLAB to compute the response of a first order low-pass filter to the input and try to fit
the computed response to the measured output by iteratively changing the gain and the cut-off frequency.
("pem" is the name of the command to use if you are interested in).

The result is shown in the attachment.
Blue curve is the input signal (I added a vertical offset to show it separately from the output).
The green curve is the measured output (RC transmission). The red curve is the response of the estimated model.
The estimated cut-off frequency was about 45kHz.

You can see that the red curve deviates a lot from the green curve after t=15usec.
By looking at this, I realized that the bandwidth of the RC cavity servo was too high.
The time scale we are looking at is about 50kHz whereas the FSS bandwidth is about 400kHz.
So when the input light was cut off, the error signal of the FSS becomes meaning less and the
input laser frequency was quickly moved away from the resonance. This is why the green curve does not
respond to the large peaks in the blue curve (input). The cavity was already off-resonance when the input power
showed bumps.

Since the red curve matches nicely with the green curve at the very beginning of the ring down, the estimated 45kHz
cavity pole is probably not that a bad estimate.

To make a better measurement, I will try to reduce the bandwidth of the RC servo by using only the PZT actuator.
If there were no ringing in the input light power, we wouldn't have to worry about the bandwidth of the servo because our
feedback is all made to the laser, not the cavity length.
In order to reduce the ringing in the input power, I asked Bob to make new HV cables using HV grade coax cables.
  1190   Fri Dec 12 22:51:23 2008 YoichiUpdatePSLReference cavity ring down measurement again
Bob made new HV-cables with HV compatible coaxes. The coax cable is rated for 2kV, which was as high as Bob
could found. I used it with 3kV hoping it was ok.
I also put a series resistor to the pockels cell to tame down the ripples I saw in elog:1136.

Despite those efforts, I still observed large ringings.
I tried several resistor values (2.5k, 1k, 330ohm), and found that 330ohm gives a slightly better result.
(When the resistance is larger, the edge of the PBS Refl. becomes dull).
Since the shape of the ringing does not change at all even when the pulse voltage is lowered to less than 1kV,
I'm now suspicious of the DEI pulser.

Anyway, I estimated the cavity pole using the MATLAB's system identification toolbox again.
This time, I locked the reference cavity using only the PZT feedback, which makes the UGF about a few kHz.
So, within the time scale shown in the plot below, the servo does not have enough time to respond, thus the laser
frequency stays tuned with the cavity. This was necessary to avoid non-linear behavior of the transmitted power
caused by the servo disturbing the laser frequency. With this treatment, I was able to approximate the response of
the cavity with a simple linear model (one pole low-pass filter).

MATLAB estimated the cavity pole to be 47.5kHz.
The blue curve in the plot is the measured RC transmitted power.
The incident power to the cavity can be inferred from the inverse of the red curve (the PBS reflection power).
The brown curve is the response of the first order low-pass filter with fc=47.5kHz to the input power variation.
The blue and brown curves match well for the first 10usec. Even after that the phases match well.
So the estimated 47.5kHz is probably a reasonable number. I don't know yet how to estimate the error of this measurement.

According to http://www.ligo.caltech.edu/~ajw/PSLFRC.png the designed transmission of the reference cavity mirrors is 300ppm (i.e.
the round trip loss (RTL) is 600ppm).
This number yields fc=35kHz. In the same picture, it was stated that fc=38.74kHz (I guess this is a measured number at some point).
The current fc=47.5kHz means, the RTL has increased by 200ppm from the design and 150ppm from the time fc=38.74kHz was measured.
  1191   Tue Dec 16 19:06:01 2008 YoichiUpdatePSLReference cavity ring down repeated many times
Today, I repeated the reference cavity ring down measurement many times to see how much the results vary.

I repeated the ring down for 20 times and the first attachment shows the comparison of the measured and estimated cavity transmission power.
The blue curve is the measured one, and the red curve is the estimated one. There are only 10 plots because I made a mistake when transferring data
from the oscilloscope to the PC, and one measurement data was lost.

The second attachment shows the histogram of the histogram of the estimated cavity pole frequencies.
I admit that there are not enough samples to treat it statistically.
Anyway, the mean and the standard deviation of the estimated frequencies are 47.6kHz and 2.4kHz.
Assuming a Gaussian distribution and zero systematic error, both of which are bold assumptions though, the result is 47.6(+/-0.6)kHz.

Now I removed the Pockels Cell from the RC input beam path.
I maximized the transmission by tweaking the steering mirrors and rotating the HWP.
Since the transmission PD was saturated without an ND filter on it, I reduced the VCO RF power slider to 2.85.
Accordingly, I changed the nominal common gain of the FSS servo to 10.5dB.
  2915   Wed May 12 02:35:13 2010 Koji, Rana, KiwamuUpdateGreen LockingReflection from ETM and ITM !

We succeeded in getting the reflected green beam from both ITMY and ETMY.

After we did several things on the end table, we eventually could observe these reflections.

Now the spot size of the reflection from ITMY is still big ( more than 1 cm ), so tomorrow modematching to the 40m cavity is going to be improved by putting mode matching telescopes on right positions.

An important thing we found is that, the beam height of optics which directly guides the beam to the cavity should be 4.5 inch on the end table.

 


(what we did)

* Aidan, Kevin and Kiwamu set the beam to be linearly polarized by rotating a QWP in front of the Innolight. This was done by monitoring the power of the transmitted light from the polarizer attached on the input of the Faraday of 1064 nm. Note that the angle for QWP is 326.4 deg.

* We put some beam damps against the rejected beam from the Faraday

* To get a good isolation with the Faraday we at first rotated the polarization of the incident beam so to have a minimum transmission. And then we rotated the output polarizer until the transmission reaches a minimum. Eventually we got the transmission of less than 1mW, so now the Faraday should be working regardless of the polarization angle of the incident beam. As we predicted, the output polaerizer seems to be rotated 45 deg from that of the input.

* Rana, Koji and Kiwamu aligned the PPKTP crystal to maximize the power of 532 nm.  Now the incident power of 1064 nm is adjusted to 250mW and the output power for 532 nm is 0.77mW. Actually we can increase the laser power by rotating a HWP in front of the Faraday.

* We injected the green beam to the chamber and aligned the beam axis to the ETMY without the modematching lenses, while exciting the horizontal motion of the ETM with f=1Hz from awg. This excitation was very helpful because we could figure out which spot was the reflection from the ETM.

* Once we made the reflected beam going close to the path of the incident beam, we then put the modematching lenses and aligned the steering mirrors and lenses. At this time we could see the reflected beam was successfully kicked away by the Faraday of 532 nm.

* Koji went to ITMY chamber with a walkie-talkie and looked at the spot position. Then he told Rana and Kiwamu to go a right direction with the steering mirrors. At last we could see a green beam from ITM illuminating the ETM cage.

* We excited the ITMY with f=2Hz vertically and aligned the ITM from medm. Also we recovered a video monitor which was abandoned around ETMY chamber so that we could see the spot on the ETM via the monitor. Seeing that monitor we aligned the ITM and we obtained the reclection from the ITM at the end table.

* We also tried to match the mode by moving a lens with f=400mm, but we couldn't obtain a good spot size.

 

ELOG V3.1.3-