    Reply  Tue Jan 5 16:33:45 2016, ericq, Update, LSC, AUX X Freq Noise attempt 
Turning on the MCL path (in addition to the MCL FF we always have on) let me lock the PLL for multiple seconds, but low frequency excursions still break
it in the end. I was able to briefly observe a level of ~50Hz/rtHz at 1kHz, which may or may not be real. Tomorrow we'll send the PLL control
signal to MC2, which should lock it up just fine and give us time to twiddle laser diode current, measure the PLL loop shape, etc. 
    Reply  Thu Jan 7 04:28:39 2016, ericq, Update, LSC, AUX X Freq Noise measured PLLspec.pdfPLL_OLG.pdf
[ericq, Gautam]

Brief summary of tonights work:


Locked Marconi to AUX X vs PSL beat at around 320MHz, PSL shutter closed (i.e. both lasers free running)
Measured control
    Reply  Thu Jan 7 16:52:32 2016, ericq, Update, LSC, AUX X Freq Noise measured AUXfreqnoise.pdfauxXmeasurements.zip
Here is some of the promised data. As mentioned, changing diode current and crystal temperature didn't have much effect on the frequency noise spectrum;
but the spectrum itself does seem too high for our needs. 

At each temperature, we started measuring the spectrum at 1.8A, and stepped the current up, hoping to reach 2.0 A.
    Reply  Thu Jan 7 19:04:25 2016, Koji, Update, LSC, AUX X Freq Noise measured 
The next step is to compare this data with the same measurement with the PSL and the AUX laser on the PSL table (or the end Y laser). If these show a
lot lower noise level, we can say 1) the x-end laser is malfunctioning and 2) the y-end and AUX laser on the PSL are well low noise.
    Reply  Thu Sep 21 01:32:00 2017, gautam, Update, ALS, AUX X Innolight AM measurement running Innolight_AM.pdfInnolight_AM.tar.gzIMG_7599.JPG20170921_203741_TFAG4395A_21-09-2017_115547_FourSquare.pdf
[rana,gautam]

We set up a measurement of the AUX X laser AM today. Some notes:


PDA 55 that was installed as a power monitor for the AUX X laser has been moved into the main green beam path - it is just upstream of
Entry  Thu Aug 27 02:09:14 2015, ericq, Update, LSC, AUX X Laser Current Changed 
I spent some time tonight chasing down the cause of huge RIN in the X green PDH transmitted light, which I had started seeing on Monday. This was preventing
robust locking, since the ALS sensing noise was ~10x worse above 50Hz, thus making the AO transition much flakier (though, impressively, not impossible!)

I went down to the X end, and found that turning the laser diode current down by 0.1A (from 2.0 to 1.9) smoothed things out completely. Unfortunately,
    Reply  Fri Sep 18 11:59:08 2015, ericq, Update, LSC, AUX X Laser Current Reverted 
Once again, the transmitted X green beam was showing enormous intensity fluctuations (50x higher than normal). Last month, I reduced the AUX X laser
current from 2.0A to 1.9A, which I thought had fixed it somehow.

However, when I sent to the end to check it out today, I found the SR560 which is there to amplify the green PDH error signal before being
    Reply  Mon Dec 21 02:29:49 2015, ericq, Update, LSC, AUX X RIN measurements auxXRIN.pdf
I'll finish up the beat / frequency noise parts of the diagnosis tomorrow later, but I've done some investigation of the AUX X laser RIN. 

I placed a PDA255 at one of the rejected beams from the PBS on the downstream side of the IR faraday, making sure the power didn't saturate
the PD. I measured the RIN on a SR785, and simultaneously looked at the signal on a 100MHz scope. 
    Reply  Tue Feb 10 16:01:21 2015, manasa, Update, General, AUX X fiber coupled 72% AUXxTelescope.png
Plan C finally worked. We have 1.454mW of AUX X light at the PSL table (2mW
incident on the fiber coupler). 
    Reply  Tue Jan 19 17:27:58 2016, gautam, Update, Green Locking, AUX X power investigations X_END_POWER_BUDGET.JPG
Last week, Eric and I noticed that the green transmission levels at the PSL table seem much lower now than they did a month or two ago. To investigate
this, I attempted to reproduce a power budget for the X endtable setup - see the attached figure (IR powers measured with calorimeter, green powers measured
with Ophir power meter). A summary of my observations:
Entry  Fri Jan 9 14:01:17 2015, manasa, Update, General, AUX Y + PSL beat note at 1064nm 
I worked around the PSL table today.

The Y+PSL output from the optical fiber module for FOL was fed to the input of the Thorlabs FPD310. 
    Reply  Tue Jan 13 10:28:02 2015, manasa, Update, General, AUX Y + PSL beat note at 1064nm 



Quote:


 
I plan to use the
Minicircuits ZFL-1000ln that is on the IOO rack but not being used (This was used for green beatnote
    Reply  Tue Jan 13 13:22:54 2015, Jenne, Update, General, AUX Y + PSL beat note at 1064nm: needs work 
I'm super excited about this new frequency readback, but I'm not sure that it's reliable yet.  Without touching any settings, the readback is currently saying 78.6MHz, and is changing slightly (as is the FSS Slow
    Reply  Tue Jan 13 14:01:20 2015, manasa, Update, General, AUX Y + PSL beat note at 1064nm: needs work 



Quote:



I'm super excited
about this new frequency readback, but I'm not sure that it's reliable yet.  Without touching any settings, the readback is currently saying
    Reply  Fri Jan 8 14:47:33 2016, ericq, Update, LSC, AUX Y Freq Noise measured AUX_freqnoises.pdfAUXY_Jan8.zip
Here are some results from measuring the PSL / AUX Y beat. 

With the Y end laser, I was able to lock the PLL with a lower actuation range (1.6MHz/V), and with the PSL in both the free-running and MCL locked
configurations. (In the latter, I had to do a bit of human-turning-knob servo to keep the control signal from running away). I also took a spectrum with
    Reply  Fri Jan 8 20:02:49 2016, rana, Update, LSC, AUX Y Freq Noise measured 
Unless this is the limit from the way you guys set up the PLL, it seems like there's no difference between the two lasers that's of any import.
So then the locking problem has been something else all along - perhaps its noise in the X-PDF lock somehow? PDH box oscillations?
    Reply  Sat Jan 9 00:39:15 2016, gautam, Update, LSC, AUX Y Freq Noise re-measured AuxPLL.pdf



Quote:



With the Y end laser,
I was able to lock the PLL with a lower actuation range (1.6MHz/V), and with the PSL in both the free-running and MCL locked configurations.
Entry  Wed Jul 30 18:09:08 2014, Harry, Update, General, AUX and PSL Coupling Telescopes PSLTelescope.pngAUXTelescope.pngtelescopes.zipAUXTelescopePlot.pngPSLTelescopePlot.png
 Purpose
These telescopes will be used to mode match//couple the dumped SHG light from both PSL and AUX (Y-Arm) lasers into PM fibers for use in FOL.
Methods
    Reply  Thu Jun 14 22:45:05 2018, johannes, Update, General, AUX beam SRC alignment 
[Jon, Gautam, Johannes]

Jon spent some time trying to align the AUX beam to the SRC today, I got to the game kind of late so maybe others can add more detail.

The AUX beam that is reflected by the SRM looks terribly misshapen - it is quite elongated in vertical direction. Unfortunately I didn't
Entry  Tue Jun 19 18:56:55 2018, Jon, Update, General, AUX beam alignment issues 
Not much progress today with the AUX cavity scans. I've determined there still are some alignment issues.

At the start of today a large AUX/PSL beat note was visible on the AS110 sensor, at a similar power as where we left off last night (-60 dBm).
Proceeding from there, I attempted to reproduce Johannes' measurement of the cavity
    Reply  Wed Jun 20 00:57:04 2018, johannes, Update, General, AUX beam alignment issues as110_rf_30MHz.pdfyend_rf_30MHz.pdf
We did swap a lens as discussed in elog 13968, but they both had f=100mm specified,
the difference being one was AR-coated for 1064 and bi-convex, while the other one was plano-convex and had a different coating. The reason for the large
beam spot was something else: The fiber wasn't sitting in the coupler properly. When reconnecting the fiber after taking it out make sure to align
    Reply  Sun Jun 24 23:13:46 2018, johannes, Update, General, AUX beam alignment issues AS_AUX_SETUP.pdf
At some point we want to change the AUX injection on the AS table to interfere less with the normal interferometer path, and avoid 10/90 beamsplitters
which produce a fair amount of ghosting. The plan is to replace the 99/1 BS whose reflection goes to AS110 and AS55, while the transmission goes to the
AS camera, with a 90/10 BS as shown in the attachment. This results in ~10% less light on the PDs compared to the pre-AUX era. Between this BS and the
Entry  Sun Jul 8 22:27:25 2018, keerthana, Update, , AUX diagram PLL-single_demodulation.pdfPLL-double_demodulation.pdf
(Analisa, Keerthana, Sandrine)

So far we tried four different techniques to scan the AUX laser. They are,

1. Scanning the marconi frequency to sweep the central frequency of the AUX laser.
Entry  Fri May 11 18:17:07 2018, gautam, Update, DetChar, AUX laser PLL setup 
[koji, gautam]

As discussed at the meeting earlier this week, we will use some old *MOPA* channels for interfacing with the PLL system Jon is setting up. He
is going to put a sketch+photos up here shortly, but in the meantime, Koji helped me identify a channel that can be used to tune the temperature of the
Entry  Fri Mar 29 15:47:58 2019, gautam, Update, AUX, AUX laser fiber moved from AS table to PSL table 
[anjali, gautam]

To facilitate the 1um MZ frequency stabilization project, I decided that the AUX laser was a better candidate than any of the other 3 active
NPROs in the lab as (i) it is already coupled into a ~60m long fiber, (ii) the PSL table has the most room available to set up the readout optics for the
    Reply  Sun Mar 31 18:39:45 2019, Anjali, Update, AUX, AUX laser fiber moved from AS table to PSL table Schematic_of_experimental_setup_for_frequency_stabilisation_of_1_micron_source.png1_micron_setup_for_frequency_noise_measurement.JPG1_micron_setup_for_frequency_noise_measurement_power_levels.png


Attachment #1 shows the schematic of the experimental setup for the frequency noise measurement of 1 um laser source.


AUX laser will be used as the seed source and it is already coupled to a 60 m fiber (PM980). The other end of the fiber was at
the AS table and we have now removed it and placed in the PSL table.
    Reply  Tue Jun 12 03:22:25 2018, gautam, Update, LSC, AUX laser shuttered 
I worked a bit on recovering the DRMI locking again tonight. I decided to shutter the AUX laser on the PSL table at least until I figured out the correct
locking settings. As has become customary now, there was a cable in the AS beampath (leading from the AS55 DC monitor to nothing, through the enclosure
side panel, it is visible in Attachment #3 in this elog) which I only found after 30mins of
Entry  Thu May 31 02:04:55 2018, johannes, Update, PSL, AUX laser state of mind IMG_20180530_220058190.jpg
The AUX laser is down to 5.4 mW output power 

What's worse, because we wanted those fast switching times by the AOM for ringdowns, I made the beam really small, which
    Reply  Thu Jul 19 01:56:25 2018, gautam, Summary, VAC, AUX pump shutdown 
[koji, gautam]

Per Steve's instructions, we did the following:


TP3fl pressure reading was 65 torr.
TP3 controller reported pumping current of ~0.18A, temperature of 24C.
We throttled
Entry  Fri May 18 19:59:55 2018, Jon Richardson, Configuration, Electronics, AUX-PSL PLL Characterization Measurements SR560_OL.pdfSR560_CL.pdf
Below is analysis of measurements I had taken of the AUX-PSL PLL using an SR560 as the servo controller (1 Hz single-pole low-pass, gain varied
100-500). The resulting transfer function is in good agreement with that found by Gautam and Koji (#13848).
The optimal gain is found to be 200, which places the UGF at 15 kHz with a 45 deg phase margin.
Entry  Fri May 4 13:24:56 2018, Jon Richardson, Configuration, Electronics, AUX-PSL PLL Implementation & Characterization PLL_Schematic.pdfPLL_AUX-PSL_40m.pdf
Attached are final details of the phase-locked loop (PLL) implementation we'll use for slaving the AUX 700 mW NPRO laser to the PSL.

The first image is a schematic of the electronics used to create the analog loop. They are curently housed on an analyzer cart beside the
PSL table. If this setup is made permanent, we will move them to a location inside the PSL table enclosure.
    Reply  Fri May 4 19:06:28 2018, rana, Configuration, Electronics, AUX-PSL PLL Implementation & Characterization 
this doesn't make much sense to me; the phase to frequency conversion (mixer-demod to PZT ) should give us a 1/f loop as Johannes mentioned in the
meeting. That doesn't agree with your loop shape.

How about give us some more details of the setup including photos and signal/power levels? And maybe measure the LB1005 TF by itself to find
    Reply  Mon Jan 22 20:54:58 2018, Koji, Summary, General, AUX-PSL beat setup IMG_3051.JPG
The beat setup has been made on the PSL table. The BS and the PD were setup. The beat was found at 29.42degC and 50.58degC for the PSL
and AUX crystal temperatures, respectively.
We are ready for the EOM test. I have instruments stacked around the PSL table. Please leave them as they are
Entry  Fri Jan 22 11:33:20 2016, gautam, Update, Green Locking, AUX-X AM/PM investigations AUX_X_AM.pdfAUX_X_RIN.pdfNE_Mephisto.png
I was trying to characterize the AM/PM response of the X end laser. I tried to measure the AM response first, as follows:


I used the Thorlabs PDA 55, whose datasheet
says it has 10MHz bandwidth - I chose it because it has a larger active area than the PDA 255, but has sufficient bandwidth for this measurement. 
My
    Reply  Fri Jan 22 13:33:37 2016, ericq, Update, Green Locking, AUX-X AM/PM investigations 



Quote:



Attachment #1 shows the measured AM response. It differs qualitatively in shape from the earlier measurements reported in this
    Reply  Fri Jan 22 17:22:06 2016, gautam, Update, Green Locking, AUX-X AM/PM investigations AM_response.pdfNE_investigations.pdf
There were a number of directories in /users/OLD/mott/PZT/2NPRO, I've used the data in Innolight_AM_New. Also, I am unsure as to what their "calibration"
factor is to convert the measured data into RIN, so I've just used a value of 0.8, with which I got the plot to match up as close as possible to the
plot in this elog. I also redid the measurement today, given that the laser parameters have
    Reply  Fri Jan 22 18:46:03 2016, rana, Update, Green Locking, AUX-X AM/PM investigations 
The PDA photodetectors are DC coupled, so you cannot use them to go directly into the analyzer. Must use the DC block so that you can reduce the input
attenuation on the B channel and then lower the drive amplitude.

Good policy for TF measurements: drive as softly as you can and still measure in a reasonable amount of time, but no softer than that.
    Reply  Mon Oct 3 13:04:12 2011, Jenne, Update, SUS, AUXEX, AUXEY rebooted 


    
        
            Quote:
        
        
            
              + I found that burtrestore for the ETMX DC coil forces were not functional.
    Reply  Fri Nov 15 13:26:00 2013, Jenne, Update, CDS, AUXEY is back 


    
        
            Quote:
        
        
            

           
            
                
           
Entry  Fri Jun 6 14:56:09 2014, ericq, Update, elog, Aaaaaand we're back! 
ELOG is back up and running after last Friday's disk-crash-a-thon. SVN is still a work in progress. Jenne and I are now restarting computers and such.
Entry  Thu Nov 12 10:45:02 2009, Alberto, Update, PSL, Abandoned Frequency Generator 
This morning I found a frequency generator connected to something on the PSL table sitting on the blue step next to the sliding doors.
Is anyone using it? Has it been forgotten there? If that's the case, can the interested person please take care of removing it?
    Reply  Thu Nov 12 11:19:10 2009, Koji, Update, PSL, Abandoned Frequency Generator 
Last night there was an activity for a calibratuon work, which I helped. I can take care of the FG.


    
        
            Quote:
        
        
            
    Reply  Sat Dec 17 03:59:32 2011, kiwamu, Update, SUS, Aborted Hysteresis test 


    
        
            Quote from #6128
        
    
   
            
            To test it, we are shaking all of the suspension biases
    Reply  Sat Dec 17 11:53:46 2011, Zach, Update, SUS, Aborted Hysteresis test 
Do you guys have timestamps for when you started/ended the test? I have been trying to take some long-term RAM data but keep getting foiled by stuff
(this test, RTS upgrade, switching of RAMmon channels, etc.)


    
        
            
    Reply  Sat Dec 17 12:41:46 2011, Koji, Update, SUS, Aborted Hysteresis test 
The test was from: 2011-12-17 09:48 to 11:49 (UTC).
This corresponds to the period from  2011-12-17 01:48 to 3:49 (PST).
ZK: Thanks
Entry  Thu Jun 9 18:54:40 2016, Aakash, Update, General, About Acromag | SURF 2016 
Today I tried to setup Acromag Busworks card. I was able to calibrate and test it over USB but I couldn't test it over ethernet. I'll utilize
a few hours tomorrow to test it over ethernet and see if I can make it work. I have also found a few RTDs which I want to use for temperature sensing via
four probe method. So, tomorrow I'll get these RTD details revived by Gautam and Steve.
Entry  Fri Jul 29 03:44:04 2016, Aakash, Summary, , About Acromag | SURF 2016 
I tried to recompile the modbusApp binary for linux-arm acrhitecture since I suspected someting wrong with it. But still the problem persists; I can
connect to acromag but cannot access the channels. I have also reconfigured new acromag bus works terminal XT 1221-000 and I want to test if I could access
its channels. My target is to complete this acromag setup work before sunday morning so that I can focus towards having some useful results for my presentation.
Entry  Thu Oct 23 15:27:19 2008, Alberto, Update, General, Abs length TRX_aplot_03_04_05_together.png
Here are the measurements I've got yesterday. The plot shows the transmitted power after the X arm while sweeping the frequency of the beat between the
two lasers. That frequency is changed by scanning the frequency of the local oscillator of the PLL (that is the Marconi).
The X arm cavity has been locked to the TEM00 of the main beam. I tilted ITMX in order to enhance the higher modes of the secondary beam with the purpose
    Reply  Thu Oct 23 18:23:47 2008, Alberto, Update, General, Abs length TRX_plot_06_07_08_together.png
[quote]Here are the measurements I've got yesterday. The plot shows the transmitted power after the X arm while sweeping the frequency of the beat between
the two lasers. That frequency is changed by scanning the frequency of the local oscillator of the PLL (that is the Marconi).
The X arm cavity has been locked to the TEM00 of the main beam. I tilted ITMX in order to enhance the higher modes of the secondary beam with the purpose
    Reply  Thu Oct 23 18:27:04 2008, Alberto, Update, General, Abs length HOM_Frequencies.m
[quote]Here are the measurements I've got yesterday. The plot shows the transmitted power after the X arm while sweeping the frequency of the beat between
the two lasers. That frequency is changed by scanning the frequency of the local oscillator of the PLL (that is the Marconi).
The X arm cavity has been locked to the TEM00 of the main beam. I tilted ITMX in order to enhance the higher modes of the secondary beam with the purpose