40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 205 of 341  Not logged in ELOG logo
ID Date Author Type Categoryup Subject
  9295   Fri Oct 25 21:36:51 2013 MasayukiUpdateLSC'scope and spectrum analyser for REFL165

Quote:

As Jenne's Elog we want to see Spectrum and time series of REFL 165 (our PRMI LSC locking PD) to see if the signal is saturated while bring the arms into resonance.
I started to connect the spectrum analyser and the 'scope to REFL165 output.

Directional coupler (Mini=-circuits ZMDC-10-2 ZMDC-20-3) was connected just before the dimod boad input. The main output of coupler is plugged into demod board's input.The other output of the coupler is connected to AG4395A using BNC cable.

The spectrum analyser output can be read using netgpibdata in control room. The IP address is 192.168.113.108 and the GPIB address is 17. For this I dissconected the network hub from another AG4395A, which is at the front of 1X2 lack.

I didn't connected the 300 MHz 'scope right now, but tomorrow it will be connected using power splitter and also be able to get data by internet. For connect 'scope to network, I disconected the network hub from SR785.

[Jenne, Masayuki]

We changed the Directional coupler from ZMDC-20-3 to ZMDC-20-5-S+ because that coupler seemed to introduce some high frequency noise.

  9304   Mon Oct 28 14:24:01 2013 MasayukiUpdateLSC'scope and spectrum analyser for REFL165

 

 I connected the 'scope between REFL165 output and demod board input. I split the signal from coupler using the splitter (Mini-Circuits ZFSC-2-5). One signal is going to 'scope CH1 and the other is going to spectrum analyzer. I connected the 'scope to 40MARS. The IP adress is 192.168.113.25. I connected that by cabling from 1X2.

 

  9305   Mon Oct 28 18:57:27 2013 MasayukiHowToLSCread 'scope and spectrum analyser datas

 

The command to get the data from spectrum analyzer right now

From command line, put ./netgpibdata -i 192.168.113.108 -d AG4395A -a 17 -f meas01

(EDIT JCD:  You must first be in the correct folder:  /opt/rtcds/caltech/c1/scripts/general/netgpibdata/)

(EDIT JCD again: "meas01" in the command line instruction will be the name of the filename.  Also, the output file meas01.dat has a comment in the first line that must be deleted before you can plot the data.  This sucks, and we should write a script to strip that line, then make nice plots.)

Please take notice that although IP address of AG4395A is same as written in the help of netgpibdata, the GPIB address is not same. It's 17.

 

How to use  'scope from control room.

Open the browser. Put the IP adress of 'scope (192.168.113.25) into adrress bar of the browser. If it's on the network, below screen will open.

You can control 'scope, get the data, and so on from control room.

Please take notice that Google Chrome cannot connect the 'scope. So you have to use the Firefox or other browser.

  9312   Wed Oct 30 00:02:25 2013 JenneUpdateLSCLSC demod boards need some thought

As we are meditating on things to look at for PRMI + 2 arms, Rana brought up the question of the demod board situation. 

We then found this table on the wiki (LSC demod boards) that indicates that all of the demod boards were originally given lowpass filters, no matter the demodulation frequency.  Back in September, I switched out the low pass filter for a bandpass filter in POP110, and put in the same bandpass when putting together AS110 (elog 9100).  So, the 11MHz diodes are probably okay with lowpasses, and the 110 diodes are okay, but we need to think about all the other ones. 

We should probably do a first guess by putting in a bandpass filter, but then simulate and measure to figure out what our requirements are for attenuation at the non-demodulation frequencies for each board.

The SXBPs from Minicircuits look pretty good, but there are lots of options on their website.

 

For tonight, Rana has put a coax 100 MHz highpass filter on the input to the REFL165 demod board.

  9313   Wed Oct 30 01:22:56 2013 JenneUpdateLSCREFL 165 demod phase adjusted

Quote:

For tonight, Rana has put a coax 100 MHz highpass filter on the input to the REFL165 demod board.

 This of course changes our demod phase.  Rana plotted a 4th order elliptic filter in Matlab, and from the plot determined that we should expect around 60 degrees of difference in our phase. 

To actually set the phase, I locked PRMI on AS55Q and REFL33I (MICH gain = -8.0, PRCL gain = +0.05, with 1's in the matrix elements).  I then turned on the PRCL oscillation notch (564 Hz), and turned on the sensing matrix's drive at that frequency, and looked at the spectrum of REFL165. 

The previous REFL165 demod phase was 96 degrees, so I was looking around either 36 degrees or 156 degrees.  The phase that minimized the peak in the Q signal while driving PRCL was 37.5 degrees.  Good work Matlab/Rana.

I then looked at the transfer functions between REFL33 and AS55 and REFL165, to see if there were any sign flips that happened.  There were not.  As expected, it was just a little extra phase delay.

I was able to lock PRMI with REFL 165 again after this phasing, and I am now taking transfer functions of the MICH and PRCL loops to make sure that we have the gains about right.

  9314   Wed Oct 30 01:44:13 2013 JenneUpdateLSCMICH and PRCL gains adjusted (Config file saved)

Quote:

I am now taking transfer functions of the MICH and PRCL loops to make sure that we have the gains about right.

 I have set the PRCL UGF to be about 180Hz, and the MICH UGF to be about 70 Hz. 

This is with locking on REFL165 I&Q, with MICH gain of -2.0 and PRCL gain of 0.70 . 

The PRCL loop only has about 30 degrees of phase margin, and is not near the top of its phase bubble.  During the day, I need to look at why we don't have more phase near 200 Hz.

  9316   Wed Oct 30 03:33:17 2013 RanaUpdateLSCLSC demod boards need some thought

 

 0309.png

I worked on the script SPAG4395A.py tonight with Masayuki's help. This sets up the parameters on the Agilent 4395A and then acquires the spectrum data. It had a couple of bugs before: no matter what channel you requested, you always got channel R. It also would disobey any requests to reduce the attenuation and left the Auto Atten ON. The version now in the SVN allows you to choose the channel and the attenuation.

It then makes this plot using matplotlib. The attached image is from the REFL165 pickoff at a time tonight when the arm powers were ~5-10. I have converted the spectrum from RF electrical Watts into Volts (V = 50*sqrt(W)). To go from the analyzer input to the demod board input we should scale this spectrum by a factor of ~15 (to account for the 20 dB from the coupler and the 3 dB of the splitter and a little more for losses). On the oscilloscope we see Vpp ~5 mV, so that's ~75 mVpp at the output of the BBPD which we're using for REFL165. Perhaps we can handle another factor of ~2-3 ? I'm not sure what we have in terms of linearity measurements on this thing.

EDIT: Evan is right, its V = sqrt(50*W), not V = 50*sqrt(W). ignore y-axis above

  9317   Wed Oct 30 03:36:51 2013 JenneUpdateLSCMICH and PRCL UGFs change with ALS enabled

Masayuki was able to hold both arms off-resonance with ALS long enough for me to lock the PRMI (arms still held off resonance), and take a set of transfer functions.

MICH gain is still -2.0, PRCL gain is still 0.070, which, with the ETMs misaligned, gave me UGFs of 70 for MICH and 180 for PRCL.

Now, however, with the ETMs aligned, but arms held off resonance with ALS, the UGFs have been lowered by a factor of 2 in frequency!  What is doing this??  MICH is now 40 Hz, and PRCL is now 80 Hz.

We measured the MICH and PRCL loops for several arm powers, and there was no change, at least until the arms were both resonating with powers of ~4 . 

After misaligning the ETMs, I remeasured the loops, and the UGFs went back up to where they started.

  9326   Fri Nov 1 17:01:46 2013 GabrieleSummaryLSCSimulation of REFL_3f signal when the arms come in

 I simulated how the 3f signal is affected by the resonance condition of the arms.

To keep it simple, I only simulated a double cavity. The attached plot shows the result. In x there is the arm cavity detuning from resonance (in log scale to show what happens close to the 0 value). In the y axis there is the PRC detuning. So every vertical slice of the upper plot gives a PDH signal for a given arm detuning. The bottom plot shows the power build up inside the arm, which is dominated by the carrier.

refl_3f_I_vs_cavity_tuning_noresf2.png

The 3f signal is not perturbed in any significant way by the arm resonance condition. This is good and what we expected.

However, in this simulation I had to ensure that the 1f sidebands are not perfectly anti-resonant inside the arms. They are indeed quite far away from resonance. If the modulation frequency is chosen in order to make the 1f sidebands exactly ant-resonant, the 2f will be resonant. This screws up the signal: REFL_3f is made of two contributions of equal amplitude, one on the PRC sidebands resonance and the other on the PRC carrier resonance. When the arm tuning goes to zero, these two cancels out and there is no more PDH...

refl_3f_I_vs_cavity_tuning.png

However, this is a limit case, since the frequency show match perfectly. If the modulation frequency is few arm line widths away from perfect anti-resonance, we have no problem.

  9327   Fri Nov 1 17:44:06 2013 KojiSummaryLSCSimulation of REFL_3f signal when the arms come in

Yes, the resonance of the 2nd-order sidebands to the IFO screws up the 3f scheme.

2f (~22MHz) and 10f (~110MHz) are at x 5.6 and x 27.9 FSR from the carrier, so that's not the case.

Could we also see how much gain fluctuation of the 3f signals we would experience when the arm comes into the resonance?

  9337   Mon Nov 4 14:11:23 2013 GabrieleSummaryLSCSimulation of REFL_3f signal when the arms come in

Quote:

Yes, the resonance of the 2nd-order sidebands to the IFO screws up the 3f scheme.

2f (~22MHz) and 10f (~110MHz) are at x 5.6 and x 27.9 FSR from the carrier, so that's not the case.

Could we also see how much gain fluctuation of the 3f signals we would experience when the arm comes into the resonance?

 From the simulation there is no visible change in the gain.

  9338   Mon Nov 4 15:46:17 2013 JenneUpdateLSCThoughts and Conclusions from last week's PRMI+2arms attempt

5:31pm - This is still a work in progress, but I'm going to submit so that I save my writing so far. I think I'm done writing now.


First, a transcription of some of the notes that I took last Tuesday night, then a few looks at the data, and finally some thoughts on things to investigate.


MICH and PRCL Transfer Functions while arms brought in to resonance (both arms locked to ALS beatnotes):

This is summarized in elog 9317, which I made as we were finishing up Tuesday night.  Here's the full story though.  Note that I didn't save the data for these, I just took notes (and screenshots for the 1st TF).

POP22I was ~140 counts, POP110I was ~100 counts.

MICH gain = -2.0, PRCL gain = 0.070. 

First TF (used as reference for 2-10), PRMI locked on REFL165, Xarm transmission = 0.03, Yarm transmission = 0.05 (both arms off resonance).  MICH UGF~40Hz, PRCL UGF~80Hz.

MICH_40Hz.pngPRCL_80Hz.png

2: X=off-res (xarm not moved), Y=0.13, no change in TF

3: X=off-res (xarm not moved), Y=0.35, no change in TF

4: X=off-res (xarm not moved), Y=0.60, MICH high freq gain went up a little, otherwise no change (no change in either UGF)

5:  X=off-res (xarm not moved), Y=0.95, same as TF#4.

6: X=0.20, Y=1.10 (yarm not moved), same as TF#4

7:  X=0.40, Y=1.30 (yarm not moved), same as TF#4

8:  X=0.70, Y=1.55 (yarm not moved), same as TF#4

9: X=1.40, Y=2.20 (yarm not moved), same as TF#4

10: X=4.0, Y=4.0 (yarm not moved), PRCL UGF is 10Hz higher than TF#4, MICH UGF is 20Hz lower than TF#4.

11: (No TF taken), Xarm and Yarm transmission both around 20!  To get this, MICH FMs that were triggered, are no longer triggered to turn on.  Also, MICH gain was lowered to -0.15 and PRCL gain was increased to 0.1

12: (No TF taken), Xarm and Yarm transmissions both around 40!  The peaks could be higher, but we don't have the QPD ready yet.

After that, we started moving away from resonance, but we didn't take any more transfer functions.


OpLev spectra for different arm resonance values:

We were concerned that the ETMs and ITMs might be moving more, when the arms are resonating high power, due to some optical spring / radiation pressure effects, so I took spectra of oplevs at various arm transmissions.

I titled the first file "no lock", and unfortunately I don't remember what wasn't locked.  I think, however, that nothing at all was locked.  No PRMI, no arm ALS, no nothing.  Anyhow, here's the spectrum:

ALS_noLock.pdf

I have a measurement when the Yarm's transmission was 1, and the Xarm's transmission was 1.75.  This was a PRMI lock, with ALS holding the arms partially on resonance:

ALS_X1pt75Y1.pdf

Next up, I have a measurement when Yarm was 0.8, Xarm was 2.  Again, PRMI with the arms held by ALS:

ALS_X2Y0pt8.pdf

And finally, a measurement when Xarm was 5, Yarm was 4:

ALS_X5Y4.pdf

Just so we have a "real" reference, I have just now taken a set of oplev spectra, with the ITMs, ETMs and PRM restored, but I shut the PSL shutter, so there was no light flashing around pushing on things.  I noticed, when taking this data, that if the PSL shutter was open, so the PRFPMI is flashing (but LSC is off), the PRM oplev looks much like the original "no Lock" spectra, but when I closed the shutter, the oplev looks like the others.  So, perhaps when we're getting to really high powers, the PRM is getting pushed around a bit?

ALS_noLock_noLaser.pdf

Conclusions from OpLev Spectra:  At least up to these resonances (which is, admittedly, not that much), I do not see any difference in the oplev spectra at the different buildup power levels.  What I need to do is make sure to take oplev spectra next time we do the PRMI+2arms test when the arms are resonating a lot. 


Time series while bringing arms into resonance:

PRMI_2arms_29Oct2013_POPrin.png

I had wondered if, since the POP 22 and 110 values looked so shakey, we were increasing the PRCL RIN while we brought the arms into resonance.  You can see in the above time series that that's not true.  The left side of the plot is PRMI locked, arms held out of resonance using ALS.  First the Yarm is brought close to resonance, then the Xarm follows.  The RIN of the arms is maybe increasing a little bit as we get closer to resonance, but not by that much.  But there seems to be no correlation between arm power and RIN of the power recycling cavity.

Alternatively, here is some time series when the arm powers got pretty high:

PRMI_2arms_29Oct2013_POPrin_highArmPowers.png


Possible Saturation of Signals:

One possibility for our locklosses of PRMI is that some signal somewhere is saturating, so here are some plots showing that that's not true for the error and control signals for the PRMI:

PRMI_2arms_29Oct2013_LSCcontrolSignals.png

Here, for the exact same time, is a set of time series for every optic except the SRM.  We can see that none of the signals are saturating, and I don't see any big differences for the ITMs or ETMs in the times that the PRMI is locked with high arm powers (center of the x-axis on the plot) and times that the PRMI is not locked, so we don't have high arm powers (edges of the plot - first half second, and last full second).  You can definitely see that the PRM moves much more when the PRMI is locked though, in both pitch and yaw. 

PRMI_2arms_29Oct2013_OpLevs_highArmPowers.png

DCPD signals at the same time:

PRMI_2arms_29Oct2013_DCPDs.png

NB:  These latest 3 plots were created with the getdata script, with arguments "-s 1067163405 -d 7".  It may be a good idea to take some spectra starting at, say 1067163406, 1 second in, and going for ~2 seconds. (It turns out that this is kind of a pain, and I can't convince DTT to give me a sensible spectrum of very short duration....we'll just need to do this live next time around).


Things to think about and investigate:

Why are we losing lock? 

On paper, is the (will the) optical spring a problem once we get high resonance in the arms? 

Spectra of oplevs when we're resonating high arm power.

What is the coupling between 110MHz and 165MHz on the REFL165 PD?  Do we need a stronger bandpass? 

Why are things so shakey when the arm power builds up?

Why do PRCL and MICH have different UGFs when the arms are controlled by ALS vs. ETMs misaligned?

Does QPD for arm transmissions switching work?  Can we then start using TRX and TRY for control?

What is the meaning of the similar features in both transmission signals, and the power recycling cavity?  Power fluctuation in the PRC due to PRM motion? 

  9339   Mon Nov 4 17:08:23 2013 JenneUpdateLSCThoughts on Transition to IR

Gabriele and I talked for a while on Wednesday afternoon about ideas for transitioning to IR control, from ALS. 

I think one of the baseline ideas was to use the sqrt(transmission) as an error signal.  Gabriele pointed out to me that to have a linear signal, really what we need is sqrt( [max transmission] - [current transmission] ), and this requires good knowledge of the maximum transmission that we expect.  However, we can't really measure this max transmission, since we aren't yet able to hold the arms that close to resonance.  If we get this number wrong, the error signal close to the resonance won't be very good.

Gabriele suggested maybe using just the raw transmission signal.  When we're near the half-resonance point, the transmission gives us an approximately linear signal, although it becomes totally non-linear as we get close to resonance.  Using this technique, however, requires lowering the finesse of PRCL by putting in a medium-large MICH offset, so that the PRC is lossy.  This lowering of the PRC finesse prevents the coupled-cavity linewidth of the arm to get too tiny.  Apparently this trick was very handy for Virgo when locking the PRFPMI, but it's not so clear that it will work for the DRFPMI, because the signal recycling cavity complicates things.

I need to look at, and meditate over, some Optickle simulations before I say much else about this stuff.

  9340   Mon Nov 4 18:24:15 2013 KojiUpdateLSCThoughts on Transition to IR

 You have the data. Why don't you just calculate 1/SQRT(TRX)?

...yeah, you can calculate it but of course you don't have no any reference for the true displacement...

  9341   Mon Nov 4 23:11:00 2013 JenneUpdateLSCSmall updates to LSC screen

I made some small edits to the LSC screen. 

* When I added columns for the new AS110 PD, I had forgotten to make the Trigger matrix and Power Normalization matrix icons on the screen bigger, so we weren't seeing the last 2 columns in the overview screen.

* I added "show if not zero" oscillator icons to the Sensing Matrix part of the LSC overview screen, so that it's easier at a glance to see that there is an oscillator on.

  9344   Tue Nov 5 16:39:54 2013 GabrieleUpdateLSCThoughts on Transition to IR

Quote:

Gabriele and I talked for a while on Wednesday afternoon about ideas for transitioning to IR control, from ALS. 

I think one of the baseline ideas was to use the sqrt(transmission) as an error signal.  Gabriele pointed out to me that to have a linear signal, really what we need is sqrt( [max transmission] - [current transmission] ), and this requires good knowledge of the maximum transmission that we expect.  However, we can't really measure this max transmission, since we aren't yet able to hold the arms that close to resonance.  If we get this number wrong, the error signal close to the resonance won't be very good.

Gabriele suggested maybe using just the raw transmission signal.  When we're near the half-resonance point, the transmission gives us an approximately linear signal, although it becomes totally non-linear as we get close to resonance.  Using this technique, however, requires lowering the finesse of PRCL by putting in a medium-large MICH offset, so that the PRC is lossy.  This lowering of the PRC finesse prevents the coupled-cavity linewidth of the arm to get too tiny.  Apparently this trick was very handy for Virgo when locking the PRFPMI, but it's not so clear that it will work for the DRFPMI, because the signal recycling cavity complicates things.

I need to look at, and meditate over, some Optickle simulations before I say much else about this stuff.

 The idea of introducing a large MICH offset to reduce the PRC finesse might help us to get rid of the transmitted power signal. We might be able to increase enough the line width of the double cavity to make it larger than the ASL length fluctuations. Then we can switch from ASL to the IR demodulated signal without transitioning through the power signal.

  9345   Tue Nov 5 16:47:09 2013 JenneUpdateLSCEnd transmission QPDs

I think Steve is trying to align the end transmission QPDs, since the arms are locked nicely right now.  I noticed that the QPDX pitch and yaw signals were digital zeros.  A quick look determined that the QPDX matrix to go from 4 quadrants to 3 degrees of freedom had been filled in for the POS row, but not pitch and yaw.  So, I copied the QPDY matrix over to QPDX (so the ordering of the rows and columns is assumed to be the same). 

Hopefully this will get us close to centered, but I suppose we ought to check really which quadrant is which, by shining a laser pointer at each quad at each end.

  9346   Tue Nov 5 16:47:19 2013 ranaFrogsLSCillegal power supply about to expire

 Is this your illegally installed HP bench power supply?

20131023_222351.jpg

If so, or if not but you care about the signal that passes through these amplifiers, I suggest you remove this temporary power supply and wire the power from the rack power supplies through the fuse blocks and possibly use a voltage regulator.

In 24 hours, that power supply will be disconnected and the wires snipped if they are still there.

  9347   Tue Nov 5 17:12:34 2013 SteveUpdateLSCcentered qpds

 Full list tomorrow: IP-Ang & Pos, ETMY-T, ETMY-Oplev, ETMX-T, IOO-Ang & Pos

 RA: No one in the control room this evening can understand what this ELOG means. Please use more words.

Attachment 1: recentered.png
recentered.png
  9348   Tue Nov 5 17:12:48 2013 JenneFrogsLSCillegal power supply about to expire

Quote:

 Is this your illegally installed HP bench power supply?

 Steve has promised to add another row of fuses to the LSC rack first thing in the morning.  Then, during Wednesday Chores, we can move the wires from the power supply to the fused power.

STEVE:  NEVER MIND about doing this in the morning.  Let's chat at the lunch meeting about what needs to be done to power things down, then back up again, in a nice order, and we can do it after lunch.  

So, please do not do anything to the LSC rack tomorrow!  Thank you.

  9349   Tue Nov 5 19:39:27 2013 JenneUpdateLSCOpLev time series

[Rana, Jenne]

We looked at the time series for all the oplevs except the BS, from last Tuesday night, during a time when we were building up the power in the arms.  We conclude from a 400 second stretch of data that there is not discernible difference in the amount of motion of any optic, when the cavities are at medium power, and when they're at low power.  Note however, that we don't have such a nice stretch of data for the really high powers, so the maximum arm power in these plots is around 5.  Both the TRX and TRY signals look fairly stationary up to powers of 1 or 2, but once you get to 4 or 5, the power fluctuations are much more significant.  So, since this isn't caused by any optic moving more, perhaps it's just that we're more sensitive to optic motion when we're closer to resonance in the arms.

However, from this plot, it looks like the ETMY is moving much more than any other optic.  On the other hand, ETMY has not ever been calibrated (there's an arbitrary 300 in there for the calibration numbers on the ETMY oplev screen).  So, perhaps it's not actually moving any more than other optics.  We should calibrate the ETM oplevs nicely, so we have some real numbers in there.  ETMX also only is roughly calibrated, relative to the OSEMs.  We should either do the move-the-QPD calibration, or a Kakeru-style pitch and yaw some mirrors and look at transmitted power.

Traces on this xml file have been filtered with DTT, using zpk([0],[0.03],1,"n").

OpLevs_during_PRMI_2arms.pdf

 

  9353   Wed Nov 6 14:47:41 2013 SteveFrogsLSCDin connectors added at 1Y2

 The north side of the LSC rack is full. I installed more DIN connectors with fuses on the south side of the rack 1Y2

The access to this may be a little bit awkward. You just remove the connector, wire it and put it back in.

 

Attachment 1: 1Y2moreDinFuses.jpg
1Y2moreDinFuses.jpg
  9355   Wed Nov 6 15:57:22 2013 JenneUpdateLSCPower Supply solution

We have decided that, rather than replacing the power source for the amplifiers that are on the rack, and leaving the Thorlabs PD as POP22/110, we will remove all of the temporary elements, and put in something more permanent.

So, I have taken the broadband PDs from Zach's Gyro experiment in the ATF.  We will figure out what needs to be done to modify these to notch out unwanted frequencies, and amplify the signal nicely.  We will also create a pair of cables - one for power from the LSC rack, and one for signal back to the LSC rack.  Then we'll swap out the currently installed Thorlabs PD and replace it with a broadband PD.

  9360   Thu Nov 7 14:39:49 2013 JenneUpdateLSCLock Acquisition Game Plan

Between the 40m meeting, and chatting with Gabriele, there was lots of talking yesterday about our 40m Lock Acquisition game plan. 

From those talks, here is my current understanding of the plan, in a Ward-style cartoon:

 LockAcquisition_7Nov2013.pdf(This is a 2 page document - description of steps is on 2nd page)

If you look closely, you will notice that there are several places that I have used "?" rather than numbers, to indicate what RFPD signal we should be using.  To fill these in, I need to look at some more simulations, and think more carefully about what signals exist at what ports, and what SNR we have at each of those ports.

Also, while the overall scale of the arm power plot is correct, the power level at each step is totally arbitrary right now, and should just be taken to mean places (in time) where the CARM offset is reduced a little more.


There are several things at this point that we know we need to look into:

* POP 22/110 PD and filtering electronics should be switched to a broadband PD, rather than the Thorlabs PD + Miniciruits filters. (Hardware)

* Whitening for the transmission QPDs needs to be thought about more carefully. (Calculation, then hardware)

* Chose a good SNR REFL DC signal, which may or may not be from the PD we are currently using (I think it's the DC of REFL11, but I'll have to check). (Calculation)

* For DRMI locking, what is the size of the SRCL error signal at AS55, AS165, and the REFL ports?  Do we need to lock with AS port, and then switch over to a REFL 3f port, to make acquisition easier?  (Simulation)

* Similarly, I want to make the equivalent of Figure 3 of T1000294, with our 40m parameters. (Simulation)

* To set the phase of AS110, simulate the demod phase of AS110 in both DRMI and SRMI cases.  If no (significant) change, maybe we can set the phase in the real system by misaligning the PRM, and watching the SRMI flash.  (Simulation)

* Simulate an arm sweep, up to many orders of the sidebands, to see how close to the carrier resonance any sideband resonances might be.  If something like the 4th order sideband resonates, and then beats with a 1st order sideband, is that signal big enough to disturb our 3f locking of the PRMI / DRMI?  We want to be holding the arms off resonance with ALS closer to the carrier than any "important" sideband resonances (where the definition of "important" is still undetermined).  (Simulation)

* Check if we can hand DARM from the DC transmission signals to the final RF signal while we still have a large CARM offset. Is there a point where the CARM offset is too large, and we must be still using the DC signals? (Simulation)

* At what arm power level can we transition from ALS to IR DC transmission signals for the individual arms? (Simulation)

* Still need to finish calculating what could be causing our big arm power fluctuations (Test mass angular motion?  PRM angular motion? ALS noise?) (Calculation)


Replys, and comments are welcome, particularly to help me understand where I may have (likely did) go wrong in drawing my cartoon. 

  9361   Fri Nov 8 17:19:27 2013 JenneUpdateLSCNew Broadband PD for POP 22/110

Here is a photo of the board inside the broadband photodiode (one of them) that I took from the Gyro experiment:

IMG_1255.JPG

This PD is Serial Number S1200271.

We need to have a look at the schematic, figure out what's in here now, and then modify this to be useful (appropriate resonances / notches, as well as amplification) for POP 22/110.

  9362   Fri Nov 8 18:12:21 2013 JenneUpdateLSCPRFPMI: Not crossing any resonances

Quote:

There are several things at this point that we know we need to look into:

* Simulate an arm sweep, up to many orders of the sidebands, to see how close to the carrier resonance any sideband resonances might be.  If something like the 4th order sideband resonates, and then beats with a 1st order sideband, is that signal big enough to disturb our 3f locking of the PRMI / DRMI?  We want to be holding the arms off resonance with ALS closer to the carrier than any "important" sideband resonances (where the definition of "important" is still undetermined).  (Simulation)

 I have done a sweep of CARM, while looking at the fields inside of one arm (I've chosen the Xarm), to see where any resonances might be, that could be causing us trouble in keeping the PRMI locked as we bring the arms into resonance. 

2f_resonances_Xarm_CARMsweep.png

Since Gabriele pointed out to me that we're using the 3x55MHz signal for locking, we should be most concerned about resonances of the higher orders of 55, and not of 11.  So, on this plot, I have up to the 6th order 55 MHz sidebands, which are 332 MHz.  Although the Matlab default color chart has wrapped around, it's clear that the carrier is the carrier, and the +4f2, which is the same blue, is not the giant central peak.  So, it's kind of clear which trace is which, even though the legend colors are degenerate.  Also, the main point that I want to show here is that there is nothing going on near the carrier, with any relevant amplitude.  The nearest things are the plus and minus 55 MHz sidebands themselves, and they're more than 50 nm away from the carrier. 

Recalling from elog 9122, the PRFPMI and DRFPMI linewidths are about 40pm.  50pm away from the resonant point is ~1/10 the power, and 100pm away from the resonant point is ~1/100 the power.  So, 50 nm is a looooong ways away. 

Just for kicks, here is a plot of all the resonances of the 1f and 2f modulation frequencies, up to 30*f1, which is the same 6*f2:

AllModFreq_resonances_Xarm_CARMsweep.png

The resonances which are "close" to the carrier are the 9th order 11 MHz sidebands, and they're 280pm from the carrier, so twice as far as we need to be, to get our arm powers to ~1/100 of the maximum, and, they're a factor of ~1e4 smaller than the carrier.

  9363   Sat Nov 9 10:25:43 2013 KojiUpdateLSCNew Broadband PD for POP 22/110

General Remarks on the BBPD

- To form the LC network: Use fixed SMD inductors from Coilcraft. SMD tunable capacitors are found in the shelf right next to Steve's desk.
  If the tuning is too coarse, combine an appropriate fixed ceramic SMC C and the tunable C (in parallel, of course)

- L1/C1a/C1b pads are specifically designed for an additional notch

- Another notch at the diode stage can be formed between the middle PD pin (just left of the marking "C3b") to the large GND pad (between C1a/C1b to C3a).
  You have to scratch off the green resin with a small flat screw driver (or anything similar)

- A notch at the amplifier stage can be formed between the output of MAR-6SM ("+" marking)  and one of the GND pads (left side of the "U1" marking)

- The original design of the PD is broadband. So additional notches on the diode stage provides notches and resonances.
  Check if the resonances do not hit the signal frequencies.

- One would think the PD can have resonant feature to reduce the coupling of the undesired signals.
  In some sense it is possible but it will be different from the usual resonant tank circuit in the following two points.

  * Just adding a parallel L between the cathode and ground does not work. As this DC current should be directed to the DC path,
    L&C combo should be added. In fact this actually give a notch-resonance pair. This C should be big enough so that you can ignore it
    at the target resonant frequency. Supply complimentary small C if necessary to keep low impedance of the Cs at the target frequency.
    (i.e. Check SRF - self-resonant frequency of the big C)

  * Since the input impedance of MAR-6SM is 50Ohm, the top of the resonant curve will be cut at 50Ohm. So the resultant shape looks
    like a bandpass rather than a resonance.

- So in total, simulation of the circuit is very important to shape the transimpedance. And, consider the circuit can not be formed as simulated
  because of many practical imperfections like stray Ls and Cs.

 

  9367   Tue Nov 12 16:49:22 2013 JenneUpdateLSCXend QPD and Whitening board pulled

Quote:

* Whitening for the transmission QPDs needs to be thought about more carefully. (Calculation, then hardware)

 I have the X end transmission QPD, as well as the whitening board, out on the electronics bench.  Since the Thorlabs high-gain TRX PD also goes through this whitening board, we have no transmission signal for the Xarm at this time. The whitening board was in the left-most slot, of the top crate in the Xend rack.  The only cables that exist for it (like the Yend), are the ribbon from the QPD, the 4-pin lemo from the Thorlabs PD, and the ribbon going to the ADC.

I have taken photos, and want to make sure that I know what is going on on the circuits, before I put them back in. 

The QPD:

IMG_1262.JPG

IMG_1260.JPG

The whitening board:

IMG_1264.JPG

IMG_1276.JPG

  9371   Wed Nov 13 01:35:40 2013 JenneUpdateLSCPRM motion causing trouble?

Quote:

* Still need to finish calculating what could be causing our big arm power fluctuations (Test mass angular motion?  PRM angular motion? ALS noise?) (Calculation)

 I think that our problem of seeing significant arm power fluctuations while we bring the arms into resonance during PRMI+arms tests is coming from PRM motion. I've done 3 calculations, so I will describe below why I think the first two are not the culprit, and then why I think the PRM motion is our dominant problem.

===============================================================

ALS length fluctuations

Arm length fluctuations seem not to be a huge problem for us right now, in terms of what is causing our arm power fluctuations.

What I have done is to calculate the derivative of the power in the arm cavity, using the power buildup that optickle gives me. The interferometer configuration I'm using is PRFPMI, and I'm doing a CARM sweep. Then, I look at the power in one arm cavity. The derivative gives me Watts buildup per meter CARM motion, at various CARM offsets. Then, I multiply the derivative by 60 nm, which is my memory of the latest good rms motion of the ALS system here at the 40m. I finally divide by the carrier buildup in the arm at each offset, to give me an approximation of the RIN at any CARM offset.

I don't know exactly what the calibration is for our ALS offset counts, but since we are not seeing maximum arm cavity buildup yet, we aren't very close to zero CARM offset. 

From this plot, I conclude that we have to be quite close to zero offset for arm length fluctuations to explain the large arm power fluctuations we have been seeing.

 xarmRIN_vs_CARM.png

=======================================================================

AS port contrast defect from ETM motion

For this calculation, I considered how much AS port contrast defect we might expect to see given some ETM motion. From that, I considered what the effect would be on the power recycling buildup. 

Rather than doing the integrals out, I ended up doing a numerical analysis. I created 2 Gaussian beams, subtracted the fields, then calculated the total power left. I did this for several separations of the beams to get a plot of contrast defect vs. separation. My simulated Gaussian beams have a FWHM of 1 unit, so the x-axis of the plot below is in units of spot motion normalized by spot size. 

Unfortunately, my normalization isn't perfect, so 2 perfectly constructively interfering beams have a total power of 0.3, so my y-axis should all be divided by 0.3. 

The actual beam separation that we might expect at the AS port from some ETM motion (of order 1e-6 radians) causing some beam axis shift is of the order 1e-5 meters, while the beam spot size is of the order 1e-3 meters. So, in normalized units, that's about 1e-2. I probably should change the x-axis to log as well, but you can see that the contrast defect for that size beam separation is very small. To make a significant difference in the power recycling cavity gain, the contrast defect, which is the Michelson transmission, should be close to the transmission of the PRM. Since that's not true, I conclude that ETM angular motion leading to PRC losses is not an issue. 

I still haven't calculated the effect of ITM motion, nor have I calculated either test mass' angular effect directly on arm cavity power loss, so those are yet to be done, although I suspect that they aren't our problem either. 

ASportOverlap_moveGaussianBeams.png

========================================================================

PRM motion

 

I think that the PRM moving around, thus causing a loss in recycling gain, is our major problem. 

First, how do I conclude that, then some thoughts on why the PRM is moving at all. 

=========

theta = 12e-6 radians (ref: oplev plot from elog 9338 last week)

L = 6.781 meters

g = 0.94

a = theta * L /(1-g) = 0.0014 meters axis displacement

w0 = 3e-3 meters = spot size at ITM

a^2/w0^2 = 0.204 ==>> 20% power loss into higher order modes due to PRM motion.

That means 20% less power circulating, hitting the ITMs, so less power going into the arm cavities, so less power buildup. This isn't 50%, but it is fairly substantial, using angular fluctuation numbers that we saw during our PRMI+arms test last week. If you look at the oplev plot from that test, you will notice that when the arm power is high (as is POP), the PRM moves significantly more than when the carrier buildup in the cavities was low. The rms motion is not 12 urad, but the peak-to-peak motion can occasionally be that large. 

So, why is that? Rana and I had a look, and it is clear that there is a difference in PRM motion when the IFO is aligned and flashing, versus aligned, but PSL shutter is closed. Written the cavities flash, the PRM gets a kick. Our current theory is that some scattered light in the PRC or the BS chamber is getting into the PRM's OSEMs, causing a spike in their error signal, and this causes the damping loops to push on the optic. 

We should think a little more on why the PRM is moving so much more that any other optic while the power is building up, and if there is anything we can do about the situation without venting. If we have to, we should consider putting aluminum foil beam blocks to protect the PRM's OSEMs. 

  9373   Wed Nov 13 09:31:15 2013 GabrieleUpdateLSCPRM motion causing trouble?

Interesting results. When you compute the effect of ETM motion, you maybe should also consider that moving around the arm cavity axis changes the matching of the input beam with the cavity, and thus the coupling between PRC and arms. But I believe this effect is of the same order of the one you computed, so maybe there is only one or two factors of two to add. This do not change significantly the conclusion.

Instead, the numbers you're giving for PRM motion are interesting. Since I almost never believe computations before I see that an experiment agrees with them, I suggest that you try to prove experimentally your statement. The simplest way is to use a scatter plot as I suggested the past week: you plot the carrier arm power vs PRM optical lever signals in a scatter plot. If there is no correlation between the two motions, you should see a round fuzzy ball in the plot. Otherwise, you will se some non trivial shape. Here is an example: https://tds.ego-gw.it/itf/osl_virgo/index.php?callRep=18918

 

  9378   Wed Nov 13 19:22:58 2013 JenneUpdateLSCPRM motion correlated to intracavity power

[Gabriele, Jenne]

Nic and Evan put the ISS together (elog 9376), and we used an injection into the error point (?) to modulate the laser power before the PMC (The AOM had a bias offset, but there is no loop).  This gives us some RIN, that we can try to correlate with the PRM OSEM sensors. 

We injected several lines, around 100, 200, 500 and 800 Hz.  For 100, 200 and 800 Hz lines, we see a ratio between POPDC and the OSEM sensors of 1e-4, but at 500 Hz, the ratio was more like 1e-3. We're not sure why this ratio difference exists, but it does.  These ratios were true for the 4 face OSEMs.  The side OSEM saw a slightly smaller signal.  

For these measurements, the PRMI was sideband locked, and we were driving the AOM with an amplitude of 10,000 counts (I don't know what the calibration is between counts and actual drive, which is why we're looking at the POPDC to sensor *ratio*).  

To get a more precise number, we may want to consider locking the PRMI on carrier, so we have more power in the cavity, and so more signal in the OSEMs.  

These ratios look, by eye, similar to the ratios we see from the time back on 30 Oct when we were doing the PRMI+2arms test, and the arms were resonating about 50 units.  So, that is nice to see some consistency.

AOMmodulated_POPDC_OSEM.pdf

This time series is from 1067163395 + 27 seconds, from 30 Oct 2013 when we did the PRMI+2arms.

POPDC_sensors_30oct2013.png

 


Ideas to go forward:

We should think about chopping the OSEM LEDs, and demodulating the PD sensors. 

We should also take a look in the chamber with a camera from the viewport on the north side of the BS chamber, to see if we see any flashes in the chamber that could be going into the OSEMs, to see where we should maybe put aluminum foil shields.

  9382   Thu Nov 14 02:50:43 2013 JenneUpdateLSCPRM oplev measured and modeled TF

In the process of figuring out what we can do to fix our PRM motion problem, I am looking at the PRM oplev. 

Eventually (as in, tomorrow), I'd like to be able to simulate some optic motion as a result of an impulse, and see what the oplev loops do to that motion.  (For starters, I'll take the impulse response of the OSEM loop as my time series that the oplev loop sees).

One thing that I have done is look at the oplev model that Rana put together, which is now in the noisebudget svn: /ligo/svncommon/NbSVN/aligonoisebudget/trunk/OpLev/C1

This script plots the open loop gain of the modeled oplev:

PRM_OL_TF_model.png

This should be compared to the pitch and yaw measured transfer functions:

 PRM_OLPIT_TF.pdf

PRM_OLYAW_TF.pdf

In the YAW plot, there are 2 transfer functions.  The first time around, the UGF was ~2.5Hz, which is too low, so I increased the gain in the C1:SUS-PRM_OLYAW filter bank from -3 to -9. 

The shapes of the measured and modeled transfer functions look reasonably similar, but I haven't done a plot overlay.  I suspect that the reason I don't see the same height peak as in the model is just that I'm not taking a huge number of points.  However, if the other parts of the TF line up, I'll assume that that's okay.

I want to make sure that the modeled transfer function matches the measured ones, so that I know I can trust the model.  Then, I'll figure out how to use the time series data with the simulated loop.  Ideally, I'd like to see that the oplev loop can fully squish the motion from the OSEM kicks.  Once I get something that looks good (by hand-tweaking the filter shape), I'll give it a try in the actual system.  We should, as soon as I get the optimal stuff working, redo this in a more optimal way.  Both now, and after I get an optimal design, I'll look at the actual step and impulse responses of the loop, to make sure there aren't any hidden instabilities.

Other thoughts for the night:

Rana suggests increasing the gain in some of the oplev QPD heads (including PRM), so that we're getting more than a few hundred counts of power on each quadrant.  Since our ADCs go to 32,000 counts, a few hundred is very small, and keeping us close to our noise limits.

Also, just an observation, but when I watch the REFL camera along with POP and AS, it's clear that the PRM is getting kicked, and I don't have the ETMs aligned right now, so this is just PRMI flashes.  There is also a lot of glow in the BS chamber during flashes (as seen on the PRM face video camera).

  9401   Mon Nov 18 21:02:54 2013 JenneUpdateLSCPRM oplev measured and modeled TF

I have created a new filter for the PRM oplev damping loops.  The biggest change is an increase in the gain between 0.4 - 7 Hz.

Here is a plot of the old, and my new modelled open loop gain:

PRM_OLG_NewOld.png

When I look at my step and impulse response time series, the notches for the bounce and roll were causing some ringing, so for now they are turned off, both in the model and in the real time system.  Also, the "OLG orig" trace has a 4th order elliptic lowpass at 75 Hz, but the real system had a 4th order elliptic low pass at 35 Hz.  When we use 35 Hz in the model, we get lots of ringing.  So, we have moved both model and real system to 55 Hz 4th order elliptic low passes.  Also, also, we haven't been using the 3.3 Hz resonant gain, so I removed that from the modelled loop.

I have put the "boost" for the .4-7 Hz emphasis into FM 7 of the PRM oplev filters.  I also removed several old filters that are never used.  So, for now, the PRM oplevs should have engaged:  FM 1, 7, 9. Pitch gain is +5, yaw gain is -9.  We can consider re-implementing the bounce-roll notches, and the stack resgain if it looks like those are getting rung up, and causing trouble.

Here is a set of spectra, showing the improvement.  It's unclear why yaw is worse than pitch below 4Hz, and why pitch is so much worse than yaw between 4-15 Hz, however for each of pitch and yaw, the before (reference pink and cyan traces) is higher than the improved (dark red, dark blue traces) between a few tenths of a Hz up to 3ish Hz.  And, we're not causing more noise elsewhere.  We do want to monitor to make sure we're not ringing up the bounce and roll modes, but for now they seem fine.

PRM_oplev_improvement.pdf

  9404   Mon Nov 18 21:40:24 2013 KojiUpdateLSCPRM oplev measured and modeled TF

I forgot how we could turn on the PRM oplev servo and the PRM ASC servo at the same time without conflict.
It seems that this new oplev servo covers 0.04 to 8Hz. It's pretty broadband. Do we inject the ASC signal to the oplev error?

  9405   Tue Nov 19 00:07:16 2013 JenneUpdateLSCGreen status

After I aligned the IR interferometer (no ASS - we still need to figure out what's going on with that), I am trying to find the green beatnotes for each arm.

First, I locked the green lasers to each arm.

I then went out to the PSL table and aligned the Green Yarm path by overlapping the near-field and far-field of the yarm transmission and the PSL green pickoff.  I then turned on the power for the Beat PDs, since it was off (I confirmed that the outputs were plugged into the beatbox, so they are seeing 50 ohms).  I assume that the beat PDs were off since Manasa pulled the Beatbox last week, but there is no elog reference!!  Anyhow, after seeing a real signal, I maximized the DC power on the beat PD for the Yarm.  I then maximized the light on the DC transmission PD for the Yarm.

I looked at the Xarm, and the near-field alignment looks okay, but I haven't checked the far-field.

I started looking for the beatnotes from the control room:

I am changing the SLOW_SERVO2_OFFSETs by 30 counts, and then unlocking and relocking the arms, and checking to see if I see a peak on the RF spectrum analyser. 

The Y offset started at -10320, and I found a beatnote at -11230 (beatnote is about 26MHz).  The X offset started at 4500.  Going larger seemed to get me to a less bright TEM00 mode, so I switched and have been searching by going down in offset, but haven't yet found the beatnote.  I suspect that I actually need to align the X path on the PSL table.  The Y beatnote is very small, about -30dBm, so I also need to tweak the alignment by maximizing the peak value.

  9406   Tue Nov 19 00:18:30 2013 JenneUpdateLSCGreen ALS wishlist

EricQ said that he's going to start hanging out at the 40m a bit, and I was thinking about what I can have him help me with.  This lead to me writing up a wishlist for things that have to do with the ALS system and green lasers.  Some of these are very small tasks, while others are pretty big.  They are certainly not all high priority.  But, they're on my wishlist.

Calibrations

  • How many counts of SLOW_SERVO2_OFFSET is one green FSR (for each arm)?
  • Calibrate ALS OFFSETTER#_OFFSET counts to nm or Hz offset between the end lasers and the PSL.

Automation / script writing

  • Automate finding the beatnotes (requires freq counters)
  • Automate locking the ALS

Digital Acquisition

  • All 3 laser temperatures
  • Frequency counting of beatnotes

Hardware

  • Install flipper mirrors on the PSL table to switch between trans DCPDs and far-field views of beam overlap for each arm.
  • IR beatnote project - send pickoff of end lasers to PSL via fiber, set up beat detection for each arm, create PLLs.
  • Yarm PZT installation and autoalignment.
  9407   Tue Nov 19 01:11:19 2013 JenneUpdateLSCGreen status

I am able to lock the Yarm ALS, but not at the full gain that I should be.  I attribute this to my mediocre alignment of the path on the PSL table.  EDIT: Manasa pointed out that I forgot to set the PSL FSS slow adjust to ~zero, so the PSL temperature was off, so there wasn't really any hope for me last night.

However, I decided that I should write down the ALS locking procedure, as shown to me by Masayuki on 29Oct2013, that is written in one of the Control Room notebooks.  So, here it is.  I will write channel names and DTT template names for the Y arm, but the procedure is the same for both arms.

  1. Lock and align arms using IR.
  2. Lock green beams to arms.
  3. Align green beams to arms.
  4. Check beatnote alignment on PSL table.
  5. Find beatnote by changing end laser temperature (C1:ALS-Y_SLOW_SERVO2_OFFSET) in steps of ~30, watch spectrum analyser for peak.  Easier if arms are locked in IR, but disable LSC system before moving to step 6.  Beatnote should be less than ~50 MHz, and should have a peak height of about -20dBm or more.  When doing 2 arms, be careful that beatnotes of the different arms do not overlap in frequency.  Manasa reminds me that you must also remember to set the PSL FSS SLOW actuator adjust to near zero, to get the PSL back near its nominal temperature.
  6. Check UGF of phase tracker loop.  (DTT template in /users/Templates/ALS/YALS_PT_OLTF.xml) Want UGF to be ~2kHz.  Change C1:ALS0BEATY_FINE_PHASE_GAIN as necessary.
  7. Start the watch script from the ALS screen to watch for lockloss.
  8. Look at the PHASE_OUT spectrum (DTT template in /users/Templates/ALS/ALS.xml).
  9. Clear history of Phase Tracker Loop (clear hist button on C1:ALS-BEATY-FINE_PHASE screen).  Very important to do this before step 10, every time you get to step 10  (i.e. if you lose lock and are starting over)!
  10. Check sign of loop gain by using + or - 0.1 for the gain (C1:ALS-YARM_GAIN).  Beatnote should immediately stop moving if you have the sign right.  Otherwise, it'll zip around (if it does, repeat step 9, step 10).
  11. Turn gain of ALS up to ~15.  Watch the PHASE_OUT spectrum, look for the servo bump.  When you see it, back off the gain a little.  Gain of ~15 is usually about the right ballpark.
  12. Turn on FM 2, 3, 6, 7, 8 of C1:ALS-YARM.  (FM5 should already have been on).
  13. Wait for PHASE_OUT spectrum to come down.  Turn on FM10 of C1:ALS-YARM.
  14. Check UGF of ALS loop (DTT template in /users/Templates/ALS/YALS_OLTF.xml).  Want UGF to be about 150 or 170 Hz (at the peak of the phase bubble).  Adjust C1:ALS-YARM_GAIN as necessary.
  15. ALS is locked!  Use something like the "Scan Arm" script from the ALS screen to find IR resonance, or do whatever measurement you want.  Dataviewer template /users/Templates/Dataviewer_Templates/ALSdtv.xml may be useful.
  9409   Tue Nov 19 11:56:10 2013 manasa, jenne, ericQUpdateLSCPRM- OSEM side ccd camera is in place

Quote:

Quote:

Can't we somehow hook up this camera to the MUX with the movie mode?
I think both the MUX and the sensoray are compatible with the color video signal.
Only the old CRT is B/W.

 Watek ccd with Tamron lens is hooked up to MUX

This set up close to the viewport glass! Please be careful!

 Video captures when power recycling cavity is locked (videos 1 & 2) and flashing (video 3). Arms stayed misaligned.

1. CH1 and CH2 are loooking at PRM front and back faces. CH3 and CH4 are looking at POP and REFL

 2. CH1 and CH2 are loooking at PRM front and back faces. CH3 and CH4 are looking at the ITMs

3. CH1 and CH2 are loooking at PRM front and back faces. CH3 and CH4 are looking at POP and REFL

  9410   Tue Nov 19 14:47:44 2013 JenneUpdateLSCPRM oplev measured and modeled TF

Quote:

I forgot how we could turn on the PRM oplev servo and the PRM ASC servo at the same time without conflict.
It seems that this new oplev servo covers 0.04 to 8Hz. It's pretty broadband. Do we inject the ASC signal to the oplev error?

 Right now all 3 servos that control PRM angle (OSEM damping, Oplev, and ASC) run in parallel, and they're all AC coupled. 

  9411   Tue Nov 19 14:47:59 2013 manasaUpdateLSCGreen status

Quote:

After I aligned the IR interferometer (no ASS - we still need to figure out what's going on with that), I am trying to find the green beatnotes for each arm.

First, I locked the green lasers to each arm.

I then went out to the PSL table and aligned the Green Yarm path by overlapping the near-field and far-field of the yarm transmission and the PSL green pickoff.  I then turned on the power for the Beat PDs, since it was off (I confirmed that the outputs were plugged into the beatbox, so they are seeing 50 ohms).  I assume that the beat PDs were off since Manasa pulled the Beatbox last week, but there is no elog reference!!  Anyhow, after seeing a real signal, I maximized the DC power on the beat PD for the Yarm.  I then maximized the light on the DC transmission PD for the Yarm.

I looked at the Xarm, and the near-field alignment looks okay, but I haven't checked the far-field.

I started looking for the beatnotes from the control room:

I am changing the SLOW_SERVO2_OFFSETs by 30 counts, and then unlocking and relocking the arms, and checking to see if I see a peak on the RF spectrum analyser. 

The Y offset started at -10320, and I found a beatnote at -11230 (beatnote is about 26MHz).  The X offset started at 4500.  Going larger seemed to get me to a less bright TEM00 mode, so I switched and have been searching by going down in offset, but haven't yet found the beatnote.  I suspect that I actually need to align the X path on the PSL table.  The Y beatnote is very small, about -30dBm, so I also need to tweak the alignment by maximizing the peak value.

 I found the beatnotes for both the X and Y arm ALS this morning. The beat amplitudes measured -5dBm and -18dBm respectively and occurred at SLOW SERVO2 OFFSET 4550 and -10340. I had to only tweak the Y green PSL alignment to increase the beat amplitude.

I locked both the arms using ALS and they were stably locked until MC unlocked for a moment (nearly 16 minutes).

The only thing missing in the list of things you looked into is the status of the PSL slow actuator adjust. Check if this is near zero.

  9413   Tue Nov 19 17:47:17 2013 JenneUpdateLSCPRMI+2arms attempt

So far this afternoon, I have redone the IFO alignment, locked both arms with ALS, moved both arms off resonance, locked PRMI, and started bringing one arm back to resonance. 

The alignment was really not good, which I knew yesterday, but the ASS wasn't working yesterday.  I hand-did the alignment, and tried locking, which was easier with the slightly better alignment.

I locked both arms with ALS, found the resonances, and then moved them off resonance using Masayuki's scripts.

I then restored the PRM alignment, and locked the PRMI. 

I started bringing the Yarm back, but I kept losing lock when I got to about 0.1 transmission.


After losing lock several times, I switched over to looking at the ASS. I have figured out the problem, and fixed it.  The ASS for the arms now works again.

Looking at the StripTool plots of the lockin outputs for each arm, it was clear that the "L" traces were their usual size, but the "T" traces, which are demodulated versions of the transmission DC PDs, were tiny.  I investigated in the model, and the answer is obvious:  both the LSC and the ASS get the transmission information directly from the end sus computers.  Since we recently moved the normalization gain for the transmission diodes into the SUS models from the LSC model, this means that the ASS was seeing a differently sized signal than it had in the past. 

To fix this, I put a gain into the T_DEMOD_SIG filter banks for all 8 lockins that use info from the transmission DC PDs.  I used 1/g , where g is the gain that is in the C1:SUS-ETM#_TR#_GAIN channels.  For TRX, that number is -0.003, and for TRY that number is 0.002 .  So, in the .snap file that is used when turning on the ASS, I have given the Xarm lockins a gain of -333, and the Yarm lockins a gain of 500.  I chose this place, because the only thing that has happened to the signal until this point is a bandpass, so the rest of the servo gains can remain the same. 

I tested the ASS, and it works just like it used to.  I let it run, and align all of the optics, then I misaligned by a small amount each of the ETMs, saw that the lockin output values changed, and then were servoed back to zero.  So, it seems all good. 

  9414   Tue Nov 19 21:28:05 2013 manasaUpdateLSCPRMI carrier locking

Since we have never tried to lock PRMI on carrier after the folding mirrors were flipped, I tried to lock PRCL on carrier.

I thought this might give us some idea about the PRC stability for resonance or some clue as to what happens to the PRM suspensions and PRMI stability when we have carrier resonating in the cavity.

I changed the sign of the PRCL gain and also tried increasing the gain. But this did not work and I was not able to carrier lock PRMI. May be I am missing to change some parameter that is very trivial?

  9415   Tue Nov 19 21:59:35 2013 manasaUpdateLSCPRMI carrier locking

Quote:

Since we have never tried to lock PRMI on carrier after the folding mirrors were flipped, I tried to lock PRCL on carrier.

I thought this might give us some idea about the PRC stability for resonance or some clue as to what happens to the PRM suspensions and PRMI stability when we have carrier resonating in the cavity.

I changed the sign of the PRCL gain and also tried increasing the gain. But this did not work and I was not able to carrier lock PRMI. May be I am missing to change some parameter that is very trivial?

PRMI could not be locked on carrier using 3f. The configuration from the last time when PRMI was carrier locked (elog) were used and PRMI locked on carrier with these settings.

== PRMI carrier ==
  MICH: AS55_Q_ERR, AS55_PHASE_R = -12 deg,  MICH_GAIN = -0.2, feedback to ITMX(-1),ITMY(+1)
  PRCL: REFL55_I_ERR, REFL55_PHASE_R = 70 deg, PRCL_GAIN = 1.0, feedback to PRM

Below is the video capture showing the PRM front and back face when carrier flashes with few second locks.

EDIT by JCD:

The demod phase numbers that Manasa is quoting above were correct back in March, when the elog she's quoting from was written.  They are not true now, since we've adjusted things in the last 8 months.  Also, I'm using a gain of -1.5 for MICH, and +1.5 for PRCL.  MICH has no FMs triggered, PRCL has FM 2,3,6 triggered.  Since we won't be using this configuration for full locking, but just for some tests, I'm currently using AS55 Q for MICH, and REFL 55I for PRCL, and using the ITMs to actuate on MICH for today.

  9416   Wed Nov 20 01:10:38 2013 JenneUpdateLSCPRMI carrier locking

I have increased the gain of the MICH loop to -100, and set FMs 2,3,7 to be triggered.  I have also increased the PRCL gain to 2.  The PRCL ASC pitch and yaw gains used to be -0.004, but I have increased them both to -0.01.

Now, I'm seeing power fluctuations in POPDC of ~200 pk-pk, at an average value of 2650.  That's a RIN of 7.5% .  If I turn off all OSEM damping for the PRM (after the cavities are already locked), I get POP DC fluctuations of 100 pk-pk at the same average value, so a RIN of 4%. 

Back on October 30th (elog 9338), we had an average POPDC of 400, with fluctuations of 200 pk-pk, so a RIN of 50%. 

So, I am pleased that, with the carrier locking, I have lower power fluctuations.  And, since there is more overall power in the PRC right now than we had 3 weeks ago, I'm hopeful that a PRMI+arms test will have lower power fluctuation.

Also, a note, when my MICH gain was still low, I had lots of power fluctuation at the AS port, which was coherent with my POPDC power fluctuations (which makes sense).  At that time, my overall RIN was higher than it is now (although I neglected to write down the numbers), but more significantly, I saw occasional 'kicks', where the ASDC and POPDC powers would ring for 1 or 2 seconds, with power fluctuations of order 40%.  I have not seen any of those kicks since increasing the MICH gain.

  9417   Wed Nov 20 16:05:52 2013 JenneUpdateLSCPRMI carrier locking, OpLev Check

[Jenne, EricQ]

We locked the PRMI on carrier again today, after lunch.  Following a suggestion from the 40m meeting, we wanted to compare the PRMI carrier fluctuations with the new vs. old OpLev servo for the PRM. 

To do change between the servo shapes, I put in an elliptic lowpass at 35Hz, since I overwrote that with the 55Hz lowpass the other day.  The only other change between shapes is turning on and off my boost / emphasis filter. 

So, the scenarios were:

(1) New OpLev servo

(2) Old OpLev servo (no boost, but 3.2Hz res gain and bounce roll notches on), with 55Hz lowpass

(3) Old OpLev servo with 35Hz lowpass

For scenario (1), like last night, there were small power fluctuations.  For scenario (2), most of the time there were small power fluctuations, but occasionally there would be a kick somewhere, and the power would dip down by ~50%, and the fluctuations would continue like a ringdown for a few seconds, and then we'd be back to small fluctuations until the next kick.  For scenario (3), even with trying different LSC servo gains, we could not get the PRMI to lock on carrier for more than a few tenths of a second.  During that time, the power fluctuations were very large. 

So, the old oplev servo was kind of okay, but the lowpass at 35 Hz was bad, bad, bad.  It seems that the new OpLev servo is doing good things for us.

  9418   Wed Nov 20 17:05:15 2013 JenneUpdateLSCXend QPD and Whitening board replaced

[EricQ, Jenne]

We have put the Xend QPD back in place, and centered it.  The whitening board was replaced by me a few days ago.

We also went down to the Yend and centered the Yend QPD.

I used the offset.py script that Masayuki wrote to zero the offsets of the individual quadrants when the PSL shutter was closed, and then I averaged the output of the SUM filter banks, and made the gains 1/AvgSum, so that both the Thorlabs PD and the QPD are normalized to 1 at single-arm resonance, for each arm.

I don't know what the gain is of the QPD head off the top of my head, relative to the Thorlabs PD, but eventually we want them to be the same, so that 1=1 and 700=700 on each PD.

  9430   Wed Nov 27 18:31:26 2013 KojiSummaryLSCAdittion of the ALS error signals to the LSC input matrix

The Phase tracker outputs (= ALS X/Y error signals) are now conveyed to the LSC model.

Their entry points at the LSC model are C1:LSC-ALSX_IN1 and C1:LSC-ALSY_IN1.
They are connected to the signal matrix (28th and 29th signals) via signal conditioning filters (C1:LSC-ALSX and C1:LSC-ALSY).

The main LSC screen has not been updated. The conventional ALS servos are still remains as they were.

This renovation required the recompilation of c1als, c1rfm, and c1lsc. Two PCIe-RFM bridge paths were added resulting in
increase of the c1rfm timing budget from 38 to 44.

  9439   Wed Dec 4 14:16:42 2013 JenneUpdateLSCPRFPMI flashes on transmission QPDs

2 weeks ago I took some data, and remembered today at the 40m meeting that I hadn't posted it.  Bad grad student.

All I'm trying to show here is that we see flashes in the arms that are larger than the ~50 units that we see saturate the Thorlabs transmission PDs. For arm power values below ~50, the QPD sum and Thorlabs PDs give approximately the same values.  So, 1 unit on the Thorlabs PDs is equivalent to 1 unit on the QPD sum, and 50 units on the Thorlabs diode is equivalent to 50 units on the QPD sum.

The situation was arms held on resonance with ALS, and the PRMI was flashing.

ArmsResonatingUsingALS_PRMIflashing_TRX_TRY.pdf

Arm powers of ~140 imply a power recycling gain of ~7.

  9440   Wed Dec 4 15:43:13 2013 JenneSummaryLSCPut LSC DAQ channels back

Last week, Koji cleaned up the LSC model to make it much more readable, while he was working on piping the ALS signals to the LSC model.  However, somehow the DAQ Channels block got deleted before the model was committed to the svn.  Since there were 2 months between svn checkins for c1lsc.mdl, it's possible that someone had the model open just to look at, and the block got deleted, and that's the version that Koji started with. 

Anyhow, thankfully we have the svn, so Koji and I found that the DAQ Channels block was (as expected) in the previously checked-in version of the LSC model.  I put a copy of the old model onto my desktop, opened it up, copied the DAQ Channels block, and then pasted it into the new cleaned-up version of the model.  (Jamie - is there a way to conveniently download a previous version through the web interface?)

I have checked it in, compiled and restarted the lsc model.  The _DQ channels are back now.

  9449   Fri Dec 6 21:38:27 2013 KojiUpdateLSCCDS related activities for LSC

I worked on the CDS related stuffs for LSC yesterday and today.


1. Slow machines:

I checked the database files for c1iscaux and c1iscaux2 (slow machines). They are mainly
used for the control of LSC whitening filters. The channel names were totally random as we
reconfigured the RF PDs while the channel names had been unchanged.

- Now the database was modified so that the PD name and the channels are related.
- saverestore.req and autoBurt.req were also changed accordingly.

- PD interface channels are completely random. Don't use them.
- I found the whitening of DCPDs are not effective.

- We need to clean up /cvs/cds/caltech/target directory. The autoBurt requests in the old targets
are making unnecessary burt files.

2. LSC screens

- The channel names on the LSC OVERVIEW screen was modified. (Attachment 1)
- A new LSC Whitening screen was made. (Attachment 2)

3. LSC screen generator

To touch the main LSC screen is very tough. The screen was split in to several sub screens
and combined with a command.

/opt/rtcds/caltech/c1/medm/c1lsc/master/generateLSCscreen/generateLSCscreen.py

This command combines the multiple adl files into a single file with x&y offsets.
This way, you can work with the each section of the screen.
Also, moving the blocks are just easy.

4. LSC Code Bug?

During the screen making, I found that a couple of the whitening switches are not
working properly.
e.g. When AS165 (either I or Q) FM1 is activated throught the whitening trigger,
the MSB bit (bit15) of the binary I/O (C1:LSC-BIO_0_0) does not .

SImilarly ASDC FM1 does not toggle bit15 of C1:LSC-BIO_0_1.

The other channels seems OK.

At first, I thought this is a bug of "Bit2Word" block. But an individual test of the block showed that
the block is not guilty. So why is only Bit15 malfunctioning???

 

Attachment 1: LSC1.png
LSC1.png
Attachment 2: LSC2.png
LSC2.png
ELOG V3.1.3-