ID Date Authorup Type Category Subject Text Attachments
  12900   Wed Mar 22 16:58:25 2017 gautamUpdateIMCWFS sensing matrix measurementsI've taken a bunch of transfer function
measurements from the MC ASC PIT and YAW
channels to the WFS error signals using the
  
  12901   Thu Mar 23 01:44:53 2017 gautamUpdateIMCWFS sensing matrix measurementsThanks to Koji's nice MATLAB script
using DttData functions, I was able to quickly
analyze the TF data. Essentially, this measurement
 IMC_WFS_segment_TF.pdfIMC_WFS_channels_TF.pdfTFsummary.pdfIMC_WFS_170322.xlsx.zip 
  12903   Thu Mar 23 23:38:58 2017 gautamUpdateIMCMC SUS damping gains stepped downI've reduced the gains of the damping
on all 3 MC SUS by a factor of 3 for overnight
observation as part of the ongoing feedforward
  
  12904   Fri Mar 24 11:26:57 2017 gautamUpdateIMCMC SUS damping gains restoredI've restored the damping loop gains
to their nominal values. Analysis of the
coherence between MCL and seismometer channels
  
  12905   Fri Mar 24 12:21:27 2017 gautamSummaryIOOMCL / MCF / CalibrationI repeated this measurement as follows:


Added a filter in the MC_F
filterbank (FM9) to account for the Pomona
 6x 
  12906   Fri Mar 24 19:04:18 2017 gautamUpdateIMCSeismic feedforward and WFS[valera, gautam]

On Wednesday at the meeting, we
were discussing why we aren't able to
 MCL_comparison.pdfseis_comparison.pdf 
  12916   Wed Mar 29 11:41:19 2017 gautamUpdatePSLPMC DAQ assay for feed-forward integrationThe C1IOO frontend machine that resides
in 1X1/1X2 has 2 ADCs, ADC0 and ADC1. The
latter has 28 out of 32 channels unused at
 C1ALS_PSLnamespace.pngPMCcheckout.pdfD980352-A-40m_151119.pdf 
  12918   Thu Mar 30 00:16:09 2017 gautamUpdatePSLPSL NPRO PZT calibrationAs part of the ongoing effort to try and
calibrate the PMC DAQ channels into physical
units, I tried to get a calibration for the
  
  12924   Mon Apr 3 17:09:47 2017 gautamUpdateCDSC1PSL burt-restoredWhen I came in this morning, Steve had
re-locked the PMC and IMC - but I could see
a ~1Hz intensity fluctuation on the PMC REFL video
  
  12925   Mon Apr 3 17:25:13 2017 gautamUpdatePSLPSL NPRO PZT calibrationSummary:

By sweeping the laser frequency
and looking at the PMC PDH error signal,
 PDHerr.pdfNPROcalib.pdf 
  12926   Mon Apr 3 23:07:09 2017 gautamUpdatePSLPMC DAQ assay for feed-forward integrationI made some changes to the DAQ path on
the PMC servo board, as per the plan posted
earlier in this thread. Summary of changes:
  
  12929   Wed Apr 5 16:05:47 2017 gautamUpdateGeneralNB code checkout[evan, gautam]

We spent some time trying to get
the noise-budgeting code running today. I
  
  12936   Mon Apr 10 15:37:11 2017 gautamUpdateCOCRC folding mirrors - v3 of specs uploaded Koji and I have been going over these calculations
again before we send a list of revised requirements
to Ramin. I've uploaded v3 of the specs
 8x 
  12939   Tue Apr 11 00:38:37 2017 gautamUpdatePSLPMC demod moved off servo boardAs discussed at the Wednesday meeting last
week, I tried moving the demodulation of
the PMC error signal off the PMC servo board,
 PMC_Ctrl_spec.pdf 
  12940   Wed Apr 12 00:36:53 2017 gautamUpdatePSLPMC demod moved off servo boardHere is a more detailed comparison of the
spectra of the signals at the front panel
DAQ LEMO output, measured with the Agilent
 PMC_digitalSpec.pdfPMC_DAQ_spectra.pdf 
  12944   Tue Apr 18 01:01:03 2017 gautamUpdatePSLPMC OLTF measured, DAQ channels calibratedQuick entry, details to follow in the AM
tomorrow.


I calibrated the PMC DAQ
 PMCspectra_calibrated.pdfPMC_MEDMs.png 
  12945   Tue Apr 18 16:10:00 2017 gautamUpdatePSLPMC OLTF measured, DAQ channels calibratedHere are the details:


PMC OLTF:


the
procedure used was identical to what Koji
 PMC_OLTF_170418.pdf 
  12947   Wed Apr 19 15:13:30 2017 gautamUpdatePSLPMC/MCL multicoherenceI used a one hour stretch of data from
last night to look at coherence between the
PMC control signal and MCL, to see if the
 PMC_MCL_multicoherence.pdf 
  12948   Wed Apr 19 15:46:24 2017 gautamUpdateGeneral1611/1811 inventory checkI looked through the lab area to do a fast
photodiode inventory check, as we may need
to buy some for the higher order mode spectroscopy
  
  12950   Tue Apr 25 19:35:41 2017 gautamUpdateGeneralIPCS -qDataviewer wouldn't launch on pianosa
- it seemed to work fine on Donatella though.
Rana suggested using the ipcs -q command.
  
  12951   Wed Apr 26 01:00:23 2017 gautamUpdateGeneralDRMI locking Since we'd like to get back to DRSE
locking, I tried locking the DRMI tonight.
I did the following:
 DRMI_1f_26April2017.pngIMG_7373.JPGIMG_7372.JPG 
  12954   Fri Apr 28 02:04:36 2017 gautamUpdateGeneralDRMI locking I got a couple of ~30min long DRMI lock
stretches today. The settings I used are
essentially the same as what I had back in
 20170428_DRMI.pdf 
  12957   Fri Apr 28 19:32:06 2017 gautamUpdateGeneralDRMI locking - PRCL angular FFI took a closer look at the POP QPD/ PRC
angular feedforward situation yesterday.
I thought it would be useful to have a 
 PRC_angularFF.pdfPRC_TFs.pdf 
  12960   Mon May 1 16:29:51 2017 gautamUpdateGeneralDRMI locking For the traces I posted, I had not turned
on the whitening for the SRCL sensing PD
(REFL55). However, I took a spectrum on a
 LSC_sensingNoise.pdfIMG_7378.JPG 
  12963   Wed May 3 16:00:00 2017 gautamSummaryGeneralNetwork Topology Check[johannes, gautam]

I forgot we had done this last
year already, but we updated the control
 Network_topology_3May2017.pdf 
  12972   Thu May 4 19:03:15 2017 gautamUpdateGeneralDRMI locking - preliminary MICH NBSummary:

 DRMI_noArms_April30.pdfMICH_OLTF.pdfC1NB_disp_40m_MICH_NB_30_April_2017.pdf40m_REFL_RFPDs_efficiency.pdf 
  12975   Fri May 5 12:10:53 2017 gautamUpdateGeneralMICH NB questions[quote]Is suspension thermal noise missing?
I take it "Thermal" refers just
to thermal things going on in the optic,
 C1NB_disp_40m_MICH_NB_30_April_2017.pdf 
  12979   Wed May 10 01:56:06 2017 gautamUpdateGeneralMICH NB - OL couplingLast night, I tried to estimate the contribution
of OL feedback signal to the MICH length
error signal.

In order to do so,
 OL_BS_ITM_comp.pdfC1NB_disp_40m_MICH_NB_8_May_2017.pdfC1NB_disp_40m_MICH_NB_10_May_2017.pdf 
  12980   Wed May 10 12:37:41 2017 gautamUpdateCDSMCautolocker deadThe MCautolocker had stalled - there were
no additional lines to the logfile after
12:17pm (~20mins ago). Normally, it suffices
  
  12983   Wed May 10 17:17:05 2017 gautamUpdateGeneralDAC / Coil Driver noiseSuspension
Actuator noise:

There
are 3 main sources of electronics noise which
 actuation.jpg 
  12984   Wed May 10 17:46:44 2017 gautamUpdateGeneralDAC / Coil Driver noise - SRM coil driver + dewhite board removedI've
removed the SOS coil driver (D010001-B,
  
  12986   Thu May 11 18:59:22 2017 gautamUpdateGeneralSRM coil driver + dewhite board initial surveyI've added marked-up schematics + high-res
photographs of the SRM coil driver board
and dewhitening board to the 40m DCC Document
 D010001-B_40m.pdfD000183-C8_40m.pdf 
  12987   Fri May 12 01:36:04 2017 gautamUpdateGeneralSRM coil driver + dewhite board LISO modelingI've
made the LISO models for the dewhitening
board and coil driver boards I pulled out.
 SRM_bypass_plus_CoilDriver.pdfliso.zip 
  12988   Fri May 12 12:34:55 2017 gautamUpdateGeneralITM and BS coil driver + dewhite board pulled outI first set the bias sliders to 0 on the
MEDM screen (after checking that the nominal
values were stored), then shut down the watchdogs,
  
  12990   Fri May 12 18:50:08 2017 gautamUpdateGeneralITM and BS coil driver + dewhite board pulled outI've uploaded high-res photos + marked
up schematics to the same DCC page linked
in the previous page. I've noted the
  
  13003   Mon May 22 13:37:01 2017 gautamUpdateGeneralDAC noise estimateSummary:

I've
 DAC_noise_model.pdfC1NB_disp_40m_MICH_NB_22_May_2017.pdf 
  13010   Tue May 23 22:58:23 2017 gautamUpdateGeneralDe-Whitening board noisesSummary:

I
 ITMX_deWhite_ch1_noise.pdf 
  13012   Thu May 25 12:22:59 2017 gautamUpdateCDSslow machine bootfestAfter ~3months without any problems on
the slow machine front, I had to reboot 
  
  13015   Thu May 25 19:27:29 2017 gautamUpdateGeneralCoil driver board noises [Koji, Gautam]

Summary: 


Attachment
#1 shows the measured/modeled noise
 coilDriverNoises.pdf 
  13017   Mon May 29 16:47:38 2017 gautamUpdateGeneralCoil driver boards reinstalledYesterday,
I reinstalled the de-whitening boards + coil
driver boards into their respective Eurocrate
 deWhitening_consolidated.pdfdeWhitening_measurements.zipETMY_OL.png 
  13019   Tue May 30 16:02:59 2017 gautamUpdateGeneralCoil driver boards reinstalledI think the reason I am unable to engage
the de-whitening is that the OL loop is injecting
a ton of control noise - see Attachment #1.
 OL_noiseInjection.pdf 
  13026   Thu Jun 1 00:10:15 2017 gautamUpdateGeneralCoil driver boards reinstalled[Koji, Gautam]

We tried to debug the mysterious
sudden failure of ASS - here is a summary
  
  13028   Thu Jun 1 15:37:01 2017 gautamUpdateCDSslow machine bootfestSteve alerted me that the IMC wouldn't
lock. Reboots for
  
  13033   Fri Jun 2 01:22:50 2017 gautamUpdateASSASS restoration workI started by checking if shaking an optic
in pitch really moves it in pitch - i.e.
how much PIT to YAW coupling is there. The
 ASS_Y_recovery.pngASS_Repairs.zipOLs.png 
  13034   Fri Jun 2 12:32:16 2017 gautamUpdateGeneralPower glitchLooks like there was a power glitch at
around 10am today.

All frontends, FB, Megatron, Optimus
  
  13035   Fri Jun 2 16:02:34 2017 gautamUpdateGeneralPower glitchToday's
recovery seems to be a lot more complicated
than usual.


 IMG_7399.JPG 
  13036   Fri Jun 2 22:01:52 2017 gautamUpdateGeneralPower glitch - recovery[Koji, Rana, Gautam]

Attachment #1 - CDS
status at the end of todays efforts. There
 power_glitch_recovery.pngIMG_7406.JPGIMG_7407.JPGIMG_7400.JPG 
  13038   Sun Jun 4 15:59:50 2017 gautamUpdateGeneralPower glitch - recoveryI
think the CDS status is back to normal.


Bit
 powerGlitchRecovery.png 
  13048   Wed Jun 7 14:11:49 2017 gautamUpdateASSY-arm coil driver electronics investigationRana suggested taking a look at the Y-arm
test mass actuator TFs (measured by driving
the coils one at a time, with only local
 CoilTFs.pdfY_OL_calib_check.png 
  13051   Wed Jun 7 17:45:11 2017 gautamUpdateASSY-arm coil driver electronics investigationI repeated the test of driving C1:SUS-<Optic>_<coil>_EXC
individually and measuring
 CoilTFs.pdf 