40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 198 of 341  Not logged in ELOG logo
IDup Date Author Type Category Subject Text Attachments
  9905   Fri May 2 14:31:26 2014 KojiUpdateLSCALS Y beat setup alignedPlease check the X&Y ALS out-of-loop
stability. Use fine resolution (BW0.01).
Calibrate the POX/POY in Hz.
  
  9906   Fri May 2 19:03:13 2014 JenneUpdateLSCALS out of loop spectraI have taken out of loop spectra for both
arms, by looking at POX/POY while the arms
were controlled with ALS.
 ALS_OutOfLoop2_2May2014.pdf 
  9907   Sun May 4 14:20:04 2014 ericqUpdateIOOPMC relockedThe PMC has been unlocked for ~23 hours.
FSS slow was at ~-1.5 V. I zeroed it, and
relocked the PMC, transmission is ~0.81V.
  
  9908   Sun May 4 22:28:54 2014 ericqUpdateLSCfarther into CM [Rana, ericq]
Today, we got a ~2kHz bandwidth
lock of the YARM with the AO path. We weren't
 OLTF.png 
  9909   Mon May 5 19:26:43 2014 JenneUpdateLSCOverride ability for whitening triggering Today I finally implemented a feature
in the whitening triggering that I should
have a long time ago:  an override button.
  
  9910   Mon May 5 19:34:54 2014 jamieUpdateCDSc1ioo/c1ioo control output IPCs changed to PCIE DolphinNow the c1ioo in on the Dolphin network,
I changed the c1ioo MC{1,2,3}_{PIT,YAW} and
MC{L,F} outputs to go out over the Dolphin
 ioo-ipc.png 
  9911   Mon May 5 19:51:56 2014 jamieUpdateCDSc1oaf model broken because of broken BLRMS blockI finally tracked down the problem with
the c1oaf model to the BLRMS part:
/opt/rtcds/userapps/release/cds/common/models/BLRMS.mdl
 blrms-hot-mess.pngsddefault.jpg 
  9912   Tue May 6 02:48:50 2014 JenneUpdateLSCAO path engaged with AS55 as error signal for Yarm locking[Rana, Jenne]
This evening, we were able to
lock the Yarm through the common mode board,
 TF3_5May2014_BoostON_UGF5kHz.pngzipped.tgzCM_TP2A_140506_boostON_realloc.pngout.pdf 
  9913   Tue May 6 03:17:15 2014 ranaUpdateLSCfarther into CMYes,
we still need to do these things, day team.
  
  9914   Tue May 6 09:46:50 2014 steveUpdatePEMcleaning day Keven and Steve,
We cleaned around the Vertex chambers,
PSL and MC2 floor areas.
 120d.png 
  9915   Tue May 6 10:22:28 2014 steveUpdateCDSc1ioo dolphin fiber

    

       
            
 dolphin1x2.png 
  9916   Tue May 6 10:31:58 2014 jamieUpdateCDSc1ioo dolphin fiber

    

       
            
  
  9917   Tue May 6 17:58:44 2014 ericqUpdateLSCfarther into CM I took a look at the MC OLTF and
AO path TFs with the fast agilent analyzer. 
I played with the relative gain of
 AOTFs.pdf 
  9918   Tue May 6 18:32:14 2014 steveUpdateLSCTRY 60Hz noise huntThis
is an effort to get rid of our ground loops
 ETMY-ISCT_EISOL.jpg 
  9919   Tue May 6 19:38:13 2014 JenneUpdateLSCSet up for PRFPMI CM lockingTo get ready for the PRFPMI CM trials tonight,
I put AS55's cables back to their nominal
state, and now have REFL11 I going to IN1
  
  9920   Wed May 7 04:01:44 2014 rana, jenneUpdateLSCPRFPMI: Common Mode servo using REFL_DC ON, CARM offset still non-zero
    With REFL_DC coupled into
the CM board through an SR560 (with an offset
subtractor), we were able to transition to
  
  9921   Wed May 7 14:01:36 2014 steveUpdateLSCTRY 60Hz noise hunt

    

       
            
  
  9922   Wed May 7 16:31:12 2014 jamieUpdateCDScdsutils updated to version 226
controls@pianosa:~ 0$ cd /opt/rtcds/cdsutils/trunk/
controls@pianosa:/opt/rtcds/cdsutils/trunk
0$ svn update
...
At revision 226.
controls@pianosa:/opt/rtcds/cdsutils/trunk
0$ make
echo "__version__ = '226'" >lib/cdsutils/_version.py
echo
  
  9923   Wed May 7 17:10:59 2014 ranaUpdateCDScdsutils updated to version 226 This upgrade from Jamie has given
us the new apps (avg, servo, and trigservo).
We should figure out if there's a way to
  
  9924   Wed May 7 22:47:33 2014 ranaUpdateCDScdsutils updated to version 226: not working on pianosa or rossa controls@rossa:~ 0$ cdsutils
read C1:LSC-DARM_GAIN
Traceback (most recent call last):
  
  9925   Wed May 7 23:09:06 2014 rana, jamieSummaryComputer Scripts / ProgramsOttavia back on networkAfter Jamie fixed the third party repo
issue with Ottavia, he was able to upgrade
it to Ubuntu 12.04 Precise Pangolin. But
  
  9926   Wed May 7 23:30:21 2014 jamieUpdateCDScdsutils should be working nowShould be fixed now.  There were python2.6
compatibility issues, which only show up
on these old distros (e.g. ubuntu 10.04).
  
  9927   Thu May 8 00:40:39 2014 ericqUpdateLSCBNC vs. 2pin LEMO for AO I've checked that the 2pin lemo connector
that was run some time ago from the LSC rack
to the MC board does indeed transmit signals.
 AOcablesWide.pdf 
  9928   Thu May 8 01:33:21 2014 ericqUpdateCDSpython issuesOn pianosa: The ezca.Ezca class somehow
initializes with its prefix set to "C1:",
even though the docstring says the default
  
  9929   Thu May 8 02:03:51 2014 ranaUpdateISSISS: fuse was blown, repaired, loop back onBack in November, Nic and Evan turned on
an SR560 based ISS. It uses the PMC TRANS
PD as the error signal and makes an AC coupled
 ISS.pdf 
  9930   Thu May 8 14:37:02 2014 JenneUpdateASCPOP ASC QPD powerI was thinking about POP today, and wanted
to know if there was something to be done
to allow us to use the PRCL ASC for at least
  
  9931   Thu May 8 15:55:43 2014 jamieUpdateCDSpython issues

    

       
            
  
  9932   Thu May 8 17:00:56 2014 rana, QSummaryLSCREFL_DC handoff didn't work last nightLast
night after checking cabling and turning
on ISS, we tried several times to handoff
  
  9933   Thu May 8 17:25:17 2014 steveUpdateLSCTRY 60Hz noise hunt I
worked at the ETMY-ISCT this morning and
  
  9934   Fri May 9 01:36:28 2014 ranaSummarySUSOptical Lever QPD Sum trends: they're almost all too weak We want there to be ~16000 cts of
signal per quadrant on the optical levers.
I think that most of the QPDs have been modified
 OLtrend.png 
  9935   Fri May 9 04:09:39 2014 JenneUpdateLSCCM board boost turn-on checkoutAs part of checking the common mode board
before we get too carried away with using
it, I looked at the time series of the
  
  9936   Fri May 9 04:51:13 2014 ericqSummaryLSCREFL_DC handoff didn't work last night

    

       
            
  
  9937   Fri May 9 11:23:11 2014 steveUpdateGreen Lockingdecreased X green light power Green light power decreased from 3 mW to
1 mW at the ETMX-ISCT shutter. More later.
  
  9938   Fri May 9 14:01:24 2014 JenneUpdateLSCCM board boost turn-on checkoutNote:  I thought I was editing elog
9935, but somehow this became a whole
 17x 
  9939   Fri May 9 21:18:51 2014 KojiUpdateGreen LockingReverted X green light power It is actually very tricky to measure the
green power at the output of the doubling
crystal as the IR often leaks into the measurement.
  
  9940   Mon May 12 10:42:01 2014 ranaUpdateSUSsome Arm maintenanceI ran the ASS/ADS for the arms because
the X-arm was way out. There was also some
problem with its locking due to bad ramps
  
  9941   Mon May 12 14:42:25 2014 steveUpdateendtable upgradeoptical table enclosure wall proposal and table at ETMY

    

       
            
 Hilite70.jpgETMY-ISCT.jpgRoscoThermashield.pdf 
  9942   Mon May 12 22:42:19 2014 ranaSummarySUSOptical Lever QPD Sum trends: they're almost all too weakFor some reason or another, I decided that
we should see if the optical lever servos
were injecting too much noise into the test
 OL-FB.pngarms_140512.pdf 
  9943   Mon May 12 22:52:59 2014 JenneSummarySUSOptical Lever Filters are all differentWe need to go back and have a look at all
of our optical lever control filters, and
make sure they make sense. 
 OpLevFilters.pdf 
  9944   Tue May 13 00:46:58 2014 ranaHowToPSLPMC relockingThe PMC runs out of range sometimes due
to the daily temperature swing. The voltage
swings up after sunset and then starts to
  
  9945   Tue May 13 03:30:10 2014 JenneUpdateLSCLocking activities - no progress[Jenne, Rana, EricQ]
We tried a few times to engage
the AO path while holding CARM on sqrtTrans
 ALS_outofloop.pdf 
  9946   Tue May 13 13:27:58 2014 SteveSummarySUSOptical Lever Servos setting table

    

       
            
 OLsums.png 
  9947   Tue May 13 17:03:05 2014 SteveSummarySUSETMX oplev laser replaced

    

       
            
  
  9948   Tue May 13 17:31:32 2014 JenneSummarySUSETMX oplev laser replaced: New oplev gains setI took loop measurements of ETMX pit and
yaw, and set the upper UGF to be ~6Hz for
both.  This required a pitch gain of
 OLerr_13May2014.pdfOLfb_13May2014.pdf 
  9949   Tue May 13 17:45:21 2014 ranaUpdateCDS/frames space cleared up, daqd stabilized 
 Late last night we were getting
some problems with DAQD again. Turned out
  
  9950   Tue May 13 22:55:57 2014 JenneSummarySUSETMX oplev: cleanupI believe that the Xend aux laser was turned
off earlier today, for Steve's work swapping
out the oplev.  When I went down there,
  
  9951   Wed May 14 02:37:58 2014 JenneUpdateLSCNo big locking newsTonight was mostly cleaning up some scripts,
including the re-writing the restore and
align scripts for the optics. 
  
  9952   Wed May 14 10:04:06 2014 SteveSummarySUS oplev laser and temperature ETMX oplev qpd gain has to be increased.
 
 TempRules.png 
  9953   Wed May 14 14:39:31 2014 JenneUpdateSUSNew buttons on IFO_ALIGN screenIt's starting to get a little crowded,
but I modified the IFO_ALIGN screen to have
new buttons to show the aligned / misaligned
 Screenshot-Untitled_Window.png 
  9954   Wed May 14 17:36:32 2014 ericqUpdateCDSNew netgpib scripts for SR785 I have redone the SPSR785 (spectrum measurement)   
ELOG V3.1.3-