ID |
Date |
Author |
Type |
Category |
Subject |
16628
|
Thu Jan 27 18:03:36 2022 |
Paco | Summary | BHD | Part III of BHR upgrade - Install PR2, balance, and attempt OSEM tuning |
[Paco, Anchal, Tega]
After installing the short OSEMs into PR2, we moved it into ITMX Chamber. While Tega loaded some of the damping filters and other settings, we took time to balance the heavily tilted ITMX chamber. After running out of counterweigths, Anchal had to go into the cleanroom and bring the SOS stands, two of which had to be stacked near the edge of the breadoard. Finally, we connected the OSEMs following the canonical order
LL -> UR-> UL
LR -> SD
But found that UR was reading -14000 counts. So, we did a quick swap of the UR and UL sensors and verified that the OSEM itself is working, just in a different channel... So it's time to debug the electronics (probably PR2 Sat Amp?)...
PR2 Sat Amp preliminary investigation:
- The UR channel (CH1 or CH4, on PR2 Sat Amp) gives negative value as soon as an OSEM is connected.
- Without anything connected, the channel reports the usual 0V output.
- With the PDA inputs of PR2 Sat Amp shorted, we again see 0V output.
- So it seems like when non-zero photodiode current goes, the circuit sees a reversal of gain polarity and the gain is roughly half in magnitude as the correct one.
|
16627
|
Thu Jan 27 17:57:35 2022 |
Anchal | Summary | BHD | Part III of BHR upgrade - Replaced small suspended PR3 with new SOS PR3 and OSEM tuning |
[Anchal, Paco]
We removed the old PR3 housed in a tip-tilt style suspension and put it on the North flow bench in the cleanroom. I put PR3 in an accessible position near the North West edge of the BS chamber and balanced the table again with many weights. The OSEM tuning was very uneventful and easy. Following are the full brightness ADC counts for the OSEMs:
UL 25693. -> 12846
UR 24905. -> 12452
LL 23298. -> 11649
LR 24991. -> 12495
SD 26357. -> 13178
I was able to damp the optic easily after the OSEM installation with no issues.
Photos: https://photos.app.goo.gl/jcAqwFJoboeUuR7F9 |
16626
|
Thu Jan 27 16:40:57 2022 |
Tega | Summary | Computer Scripts / Programs | SUS Plant Plan for New Optics |
[Ian, Paco, Tega]
Last night we set up the four main matrices that handle the conversion between the degrees of freedom bases and the sensor bases. We also wrote a bash script to automatically set up the system. The script sets the four change of bases matrices and activates the filters that control the plant. this script should fully set up the plant to its most basic form. The script also turns off all of the built-in noise generators.
After this, we tried damping the optic. The easiest part of the system to damp is the side or y motion of the optic because it is separate from the other degrees of freedom in both of the bases. We were able to damp that easily. in attachment 1 you can see that the last graph in the ndscope screen the side motion of the optic is damped. Today we decided to revisit the problem.
Anyways, looking at the problem with fresh eyes today, I noticed the in pit2pit coupling has the largest swing of all the plant filters and thought this might be the reason why the inputs (UL,UR,LR,LL) to the controller was hitting the rails for pit DoF. I reduce the gain of the pit2pit filter then slowly increased it back to one. I also reduced the gain in the OSEM input filter from 1 to 1/100. The attached image (Attachment2) is the output from this trial. This did not solve the problem. The output when all OSEM input filter gain set to one is shown in Attachment2.
We will try to continue to tweak the coefficients. We are probably going to ask Anchal and Paco to sit down with us and really hone in on the right coefficients. They have more experience and should be able to really get the right values. |
Attachment 1: simplant_control_1.png
|
|
Attachment 2: simplant_control_0.png
|
|
16625
|
Thu Jan 27 08:27:34 2022 |
Paco | Summary | BHD | Part IX of BHR upgrade - AS1 inspection and correction |
[Paco]
This morning, I went into ITMY chamber to inspect AS1 after the free swinging test failed. Indeed, as forecasted by Anchal, the top front EQ stop was slightly touching, which means AS1 was not properly installed before. I proceeded by removing it well behind any chance of touching the optic, and did the same for all the other stops, of which most were already recessed significantly. Finally, the OSEMs changed accordingly to produced a PITCHed optic (top front EQ was slightly biasing the pitch angle), so I did a reinstallation until the levels were around the 14000 count region. After damping AS1 relatively quickly, I closed the ITMY chamber.
Quote: |
For some reasonf the free swing test showed only one resonance peak (see attachment 1). This probably happened because one of the earthquake stops is touching the optic. Maybe after the table balancing, the table moved a little over its long relazation time and by the time the free swing test was performed at 3 am, one of the earthquake stops was touching the optic. We need to check this when we open the chamber next.
|
|
16624
|
Tue Jan 25 18:37:12 2022 |
TYehonathan | Update | BHD | PR2 Suspension |
PR2's side magnet height was adjusted and its roll was balanced (attachment 1,2). I verified that the OpLev beam is still aligned. The pitch was balanced: First, using an iris for rough adjustment. Then, with the QPD. I locked the counterweight setscrew.
I turned off the HEPAs, damped PR2, and measured the QPD spectra (attachment 3). Major peaks are at 690mHz, 953mHz, and 1.05Hz. I screwed back the lower OSEM plate. The wires were clamped to the suspension block and were cut. Winch adapter plate removed. I wanted to push OSEMs into the OSEM plates but the wiki is down so I can't tell what was the plan. This will have to wait for tomorrow. Also here like with AS1 we need to apply glue to the counterweights. |
Attachment 1: PR2_magnet_height.png
|
|
Attachment 2: PR2_roll_balance.png
|
|
Attachment 3: FreeSwingingSpectra_div_50mV.pdf
|
|
16623
|
Tue Jan 25 16:42:03 2022 |
Anchal | Summary | BHD | Reduced filter gains in all damped new SOS |
I noticed that our current suspension damping loops for the new SOS were railing the DAC outputs. The reason being that cts2um module has not been updated for most optics and thus teh OSEM signal (with the new Sat Amps) is about 30 times stronger. That means our usual intuition of damping gains is too high without applying correct conversion cts2um filter module. I reduced all these gains today and nothing is overflowing the c1su2 chassis now. I also added two options in the "!" (command running drop down menu) in the sus_single medm screens for opening ndscope for monitoring coil outputs or OSEM inputs of the optic whose sus screen is used.
|
16622
|
Tue Jan 25 11:02:54 2022 |
Anchal | Summary | BHD | Part IX of BHR upgrade - AS1 free swing test failed |
For some reasonf the free swing test showed only one resonance peak (see attachment 1). This probably happened because one of the earthquake stops is touching the optic. Maybe after the table balancing, the table moved a little over its long relazation time and by the time the free swing test was performed at 3 am, one of the earthquake stops was touching the optic. We need to check this when we open the chamber next. |
Attachment 1: AS1_freeSwingTestFailed.pdf
|
|
16621
|
Tue Jan 25 10:55:02 2022 |
Anchal | Summary | BHD | Part VIII of BHR upgrade - LO2 input matrix diagonalization performed. |
The free swinging test was successful. I ran the input matrix diagonalization code (/opt/rtcds/caltech/c1/Git/40m/scripts/SUS/InMAtCalc/sus_diagonalization.py) on the LO2 free-swinging data collected last night. The logfile and results are stored in /opt/rtcds/caltech/c1/Git/40m/scripts/SUS/InMAtCalc/LO2 directory. Attachment 1 shows the power spectral density of the DOF basis data (POS, PIT, YAW, SIDE) before and after the diagonalization. Attachment 2 shows the fitted peaks.
Free Swinging Resonances Peak Fits
|
Resonant Frequency [Hz] |
Q |
A |
POS |
0.981 |
297 |
3967 |
PIT |
0.677 |
202 |
1465 |
YAW |
0.775 |
2434 |
1057 |
SIDE |
1.001 |
244 |
4304 |
LO2 New Input Matrix
|
UL |
UR |
LR |
LL |
SIDE |
POS |
0.46
|
1.237
|
1.094
|
0.318
|
0.98
|
PIT |
1.091
|
0.252
|
-1.512
|
-0.672
|
-0.088
|
YAW |
0.722
|
-1.014
|
-0.217
|
1.519
|
0.314
|
SIDE |
-0.747
|
1.523
|
1.737
|
-0.534
|
3.134
|
The new matrix was loaded on LO2 input matrix and this resulted in no control loop oscillations at least. I'll compare the performance of the loops in future soon. |
16620
|
Mon Jan 24 21:18:40 2022 |
Paco | Summary | BHD | Part IX of BHR upgrade - AS1 placed and OSEM tuned |
[Paco]
AS1 was installed in the ITMY chamber today. For this I moved AS4 to its nominal final placement and clamped it down with a single dog clamp. Then, I placed AS1 near the center of the table, and quickly checked AS4 could still be damped. After this, I leveled the table using a heavier/lighter counterweight pair.
Once things were leveled, I proceeded to install AS1 OSEMs. The LL, UL, UR OSEMs had a bright level of 27000 counts, while SD and LR were at 29500, and 29900 respectively. After a while, I managed to damp all degrees of freedom around the 50% thousand count levels, and decided to stop.
UL 27000. -> 16000
UR 27000. -> 13800
LL 27000 -> 14600
LR 29900 -> 14900
SD 29500 -> 12900
Free swinging test set to trigger
AS1 is set to go through a free swinging test at 3 am this evening. We have used this script (Git/40m/scripts/SUS/InMatCalc/freeSwing.py) reliably in the past so we expect no issues, it has a error catching block to restore all changes at the end of the test or if something goes wrong.
To access the test, on allegra, type:
tmux a -t AS1
Then you can kill the script if required by Ctrl-C, it will restore all changes while exiting.
SUSPENSION STATUS UPDATED HERE |
16619
|
Mon Jan 24 20:48:48 2022 |
Anchal | Update | BHD | AS4 Input Matrix Diagonalization performed. |
I agree. That's an interesting idea. But does that mean that there is an always working inverse matrix solution or that any solution will be vulnerable to the alignment biases.
I think we can also calculate the matrix rotation required as a function of dc biases and do that rotation in the simulimk model.
Quote: |
I think our suspension input matrix diagonalization is not so robust usually because we only choose a inverting matrix which gives the best separation for a single suspension alignment.
i.e. we have seen in the past that adjusting the bias for the alignment makes the matrix inversion not work well. Sometime people turn OFF the alignment bias before making the ringdown and that makes the whole measurement invalid.
This is because the sensitivity of the OSEMs to longitudinal and/or transverse motion is significantly different for different alignment.
I wonder if there's a way we can choose a better matrix by putting in random gain errors on the shadow sensor signals and then finding the matrix which gives the best diag under an ensemble of gain errors.
|
|
16618
|
Mon Jan 24 18:53:09 2022 |
Anchal | Summary | BHD | Part VIII of BHR upgrade - LO2 placed and OSEMs tuned |
I placed LO2 in its planned position in BS chamber, inserted the OSEMs, and tuned their position to halfway brightness. At the end of the work, I was able to damp the optic successfully. The full open (full brightness) OSEM ADC counts are:
UL 25743. -> 12876
UR 27384. -> 13692
LL 25550. -> 12775
LR 27395 -> 13697
SD 28947 -> 14473
Today's OSEM tuning was relatively unhappening. I have only following two remarks:
- BS table was 3 SOS near the East end and PRM is parked in the center, thus the table is very unevenly balanced. I had to use all available counter weights to make it flat near the LO2 suspension.
- The side OSEM for LO2 is not exactly centered (probably due to table imbalance). I was able to balance the table to a point though that the side OSEM was responsive to full range and I was able to damp the optic.
Free swinging test set to trigger
LO2 is set to go through a free swinging test at 10 pm tonight. We have used this script (Git/40m/scripts/SUS/InMatCalc/freeSwing.py) reliably in the past so we expect no issues, it has a error catching block to restore all changes at the end of the test or if something goes wrong.
To access the test, on allegra, type:
tmux a -t LO2
Then you can kill the script if required by Ctrl-C, it will restore all changes while exiting.
Photos: https://photos.app.goo.gl/Ff3yGBprj9xgPbnLA
SUSPENSION STATUS UPDATED HERE |
16617
|
Mon Jan 24 17:58:21 2022 |
Yehonathan | Update | BHD | PR2 Suspension |
I picked up the PR2 mirrors (labeled M1, M2) from Anchel's table and took them to the cleanroom. By inspection, I spotted some dust particles on M1. I wasn't able to remove them with clean air so I decided to use M2 which looked much cleaner. I wasn't able to discern any wedge angle on the optic. I inserted the optic into a thin optic adapter. The optic is thicker than I expected so I use long screws for the mirror clamping. I expect that the pitch balance will shift towards the front of the mirror so I assembled only 1 counterweight for now. The side blocks with wires in them were installed.
I engraved the SOS and installed the winches on it. Paco came in and helped me to hang the optic. Looking at the wire hanging angle I realize that 1 counterweight at the front is not enough. I install a second counterweight at the back and observe that I can cross the balancing point.
I locked the EQ stops. Suspension work continues tomorrow... |
16616
|
Mon Jan 24 17:12:27 2022 |
rana | Update | BHD | AS4 Input Matrix Diagonalization performed. |
I think our suspension input matrix diagonalization is not so robust usually because we only choose a inverting matrix which gives the best separation for a single suspension alignment.
i.e. we have seen in the past that adjusting the bias for the alignment makes the matrix inversion not work well. Sometime people turn OFF the alignment bias before making the ringdown and that makes the whole measurement invalid.
This is because the sensitivity of the OSEMs to longitudinal and/or transverse motion is significantly different for different alignment.
I wonder if there's a way we can choose a better matrix by putting in random gain errors on the shadow sensor signals and then finding the matrix which gives the best diag under an ensemble of gain errors. |
16615
|
Mon Jan 24 17:10:25 2022 |
Tega | Summary | Computer Scripts / Programs | SUS Plant Plan for New Optics |
[Ian, Tega]
Connected the New SUS screens to the controller for the simplant model. Because of hard-coded links in the medm screen links, it was necessary to create the following path in the c1sim computer, where the new medm screen files are located:
/opt/rtcds/userapps/trunk/sus/c1/medm/templates/NEW_SUS_SCREENS
We noticed a few problems:
1. Some of the medm files still had C1 hard coded, so we need to replace them with $IFO instead, in order for the custom damping filter screen to be useful.
2. The "Load coefficient" button was initially blank on the new sus screen, but we were able to figure out that the problem came from setting the top-level DCU_ID to 63.
medm -x -macro "IFO=X1,OPTIC=OPT_CTRL,DCU_ID=63" SUS_SINGLE_OVERVIEW.adl
[TODO]
Get the data showing the controller damping the pendulum. This will involve tweaking some gains and such to fine-tune the settings in the controller medm screen. Then we will be able to post some data of the working controller.
[Useful aside]
We should have a single place with all the instructions that are currently spread over multiple elogs so that we can better navigate the simplant computer. |
Attachment 1: Screen_Shot_2022-01-24_at_5.33.15_PM.png
|
|
16614
|
Mon Jan 24 12:33:41 2022 |
rana | Configuration | Wiki | AIC Wiki: txz files allowed |
I updated the mime.local.conf file for the AIC Wiki so as to allow attachments with the .txz format. THis should be persistent over upgrades, since its a local file. |
16613
|
Fri Jan 21 16:40:10 2022 |
Anchal | Update | BHD | AS4 Input Matrix Diagonalization performed. |
The free swinging test was successful. I ran the input matrix diagonalization code (/opt/rtcds/caltech/c1/Git/40m/scripts/SUS/InMAtCalc/sus_diagonalization.py) on the AS4 free-swinging data collected last night. The logfile and results are stored in /opt/rtcds/caltech/c1/Git/40m/scripts/SUS/InMAtCalc/AS4 directory. Attachment 1 shows the power spectral density of the DOF basis data (POS, PIT, YAW, SIDE) before and after the diagonalization. Attachment 2 shows the fitted peaks.
Free Swinging Resonances Peak Fits
|
Resonant Frequency [Hz] |
Q |
A |
POS |
1.025 |
337 |
3647 |
PIT |
0.792 |
112 |
3637 |
YAW |
0.682 |
227 |
1228 |
SIDE |
0.993 |
164 |
3094 |
AS4 New Input Matrix
|
UL |
UR |
LR |
LL |
SIDE |
POS |
0.844
|
0.707
|
0.115
|
0.253
|
-1.646
|
PIT |
0.122
|
0.262
|
-1.319
|
-1.459
|
0.214
|
YAW |
1.087
|
-0.901
|
-0.874
|
1.114
|
0.016
|
SIDE |
0.622
|
0.934
|
0.357
|
0.045
|
3.822
|
The new matrix was loaded on AS4 input matrix and this resulted in no control loop oscillations at least. I'll compare the performance of the loops in future soon.
|
Attachment 1: AS4_SUS_InpMat_Diagnolization_20220121.pdf
|
|
Attachment 2: AS4_FreeSwingData_PeakFitting_20220121.pdf
|
|
16612
|
Fri Jan 21 14:51:00 2022 |
Koji | Update | BHD | V6-704/705 Mirror now @Downs |
Camille@Downs measured the surface of these M1 and M2 using Zygo.
Result of the ROC measurements:M1: ROC=2076m (convex)M2: ROC=2118m (convex)
Here are screenshots. One file shows the entire surface and the other shows the central 30mm. |
Attachment 1: M1.PNG
|
|
Attachment 2: M1_30mm.PNG
|
|
Attachment 3: M2.PNG
|
|
Attachment 4: M2_30mm.PNG
|
|
16611
|
Fri Jan 21 12:46:31 2022 |
Tega | Update | CDS | SUS screen debugging |
All done (almost)! I still have not sorted the issue of pitch and yaw gains growing together when modified using ramping time. Image of custom ADC and DAC panel is attached.
Quote: |
Seen. Thanks.
Quote: |
Indicated by the red arrow:
Even when the side damping servo is off, the number appears at the input of the output matrix
Indicated by the green arrows:
The face magnets and the side magnets use different ADCs. How about opening a custom ADC panel that accommodates all ADCs at once? Same for the DAC.
Indicated by the blue arrows:
This button opens a custom FM window. When the pitch gain was modified with a ramping time, the pitch and yaw gain grows at the same time even though only the pitch gain was modified.
Indicated by the orange circle:
The numbers are not indicated here, but they are input-related numbers (for watchdogging) rather than output-related numbers. It is confusing to place them here.
|
|
|
Attachment 1: Custom_ADC_DAC_monitors.png
|
|
16610
|
Fri Jan 21 11:24:42 2022 |
Anchal | Summary | BHD | SR2 Input Matrix Diagonalization performed. |
The free swinging test was successful. I ran the input matrix diagonalization code (/opt/rtcds/caltech/c1/Git/40m/scripts/SUS/InMAtCalc/sus_diagonalization.py) on theSR2 free-swinging data collected last night. The logfile and results are stored in /opt/rtcds/caltech/c1/Git/40m/scripts/SUS/InMAtCalc/SR2 directory. Attachment 1 shows the power spectral density of the DOF basis data (POS, PIT, YAW, SIDE) before and after the diagonalization. Attachment 2 shows the fitted peaks.
Free Swinging Resonances Peak Fits
|
Resonant Frequency [Hz] |
Q |
A |
POS |
0.982 |
340 |
3584 |
PIT |
0.727 |
186 |
1522 |
YAW |
0.798 |
252 |
912 |
SIDE |
1.005 |
134 |
3365 |
SR2 New Input Matrix
|
UL |
UR |
LR |
LL |
SIDE |
POS |
1.09
|
0.914
|
0.622
|
0.798
|
-0.977
|
PIT |
1.249
|
0.143
|
-1.465
|
-0.359
|
0.378
|
YAW |
0.552
|
-1.314
|
-0.605
|
1.261
|
0.118
|
SIDE |
0.72
|
0.403
|
0.217
|
0.534
|
3.871
|
The new matrix was loaded on SR2 input matrix and this resulted in no control loop oscillations at least. I'll compare the performance of the loops in future soon.
|
Attachment 1: SR2_SUS_InpMat_Diagnolization_20220121.pdf
|
|
Attachment 2: SR2_FreeSwingData_PeakFitting_20220121.pdf
|
|
16609
|
Thu Jan 20 18:41:55 2022 |
Anchal | Summary | BHD | SR2 set to trigger free swing test |
SR2 is set to go through a free swinging test at 3 am tonight. We have used this script (Git/40m/scripts/SUS/InMatCalc/freeSwing.py) reliably in the past so we expect no issues, it has a error catching block to restore all changes at the end of the test or if something goes wrong.
To access the test, on allegra, type:
tmux a -t SR2
Then you can kill the script if required by Ctrl-C, it will restore all changes while exiting.
|
16608
|
Thu Jan 20 18:16:29 2022 |
Anchal | Update | BHD | AS4 set to trigger free swing test |
AS4 is set to go through a free swinging test at 10 pm tonight. We have used this script (Git/40m/scripts/SUS/InMatCalc/freeSwing.py) reliably in the past so we expect no issues, it has a error catching block to restore all changes at the end of the test or if something goes wrong.
To access the test, on allegra, type:
tmux a -t AS4
Then you can kill the script if required by Ctrl-C, it will restore all changes while exiting. |
16607
|
Thu Jan 20 17:34:07 2022 |
Koji | Update | BHD | V6-704/705 Mirror now @Downs |
The PR2 candidate V6-704/705 mirrors (Qty2) are now @Downs. Camille picked them up for the measurements.
To identify the mirrors, I labeled them (on the box) as M1 and M2. Also the HR side was checked to be the side pointed by an arrow mark on the barrel. e.g. Attachment 1 shows the HR side up |
Attachment 1: PXL_20220120_225248265_2.jpg
|
|
Attachment 2: PXL_20220120_225309361_2.jpg
|
|
16606
|
Thu Jan 20 17:21:21 2022 |
Tega | Update | SUS | Temporary watchdog |
Temp software watchdog now operational for LO1 and the remaining optics!
Koji helped me understand how to write to switches and we tried for a while to only turnoff the output switch of the filters instead of the writing a zero that resets everything in the filter.
Eventually, I was able to move this effort foward by realising that I can pass the control trigger along multiple records using the forwarding option 'FLNK'. When I added this field to the trigger block, record(dfanout,"C1:SUS-LO1_PUSH_ALL"), and subsequent calculation blocks, record(calcout,"C1:SUS-LO1_COILSWa") to record(calcout,"C1:SUS-LO1_COILSWd"), everything started working right.
Quote: |
After some work on the reference database file, we now have a template for temporary watchdog implementation for LO1 located here "/cvs/cds/caltech/target/c1susaux/C1_SUS-AUX_LO1.db".
Basically, what I have done is swap the EPICS asyn analog input readout for the COIL and OSEM to accessible medm channels, then write out watchdog enable/disable to coil filter SW2 switch. Everything else in the file remains the same. I am worried about some of the conversions but the only way to know more is to see the output on the medm screen.
To test, I restarted c1su2 but this did not make the LO1 database available, so I am guessing that we also need to restart the c1sus, which can be done tomorrow.
|
|
16605
|
Thu Jan 20 17:03:36 2022 |
Anchal | Summary | BHD | Part IV of BHR upgrade - SR2 OSEM tuned. |
The main issue with SR2 OSEMs, now that I think of it, was that the BS table was very inclined due to the multiple things we removed (including counterweights). Today the first I did was level the BS table by placing some counterweights in the correct positions. I placed the level in two directions right next to SR2 (clamped in its planned place), and made the bubble center.
While doing do, at one point, I was trying to reach the far South-West end of the table with the 3x heavy 6" cylindrical counterweight in my hand. The counterweight slipped off my hand and fell from the table. See the photo in attachment 1. It went to the bottommost place and is resting on its curved surface.
This counterweight needs to be removed but one can not reach it from over the table. So to remove it, we'll have to open one of the blank flanges on the South-west end of BS chamber and remove the counterweight from there. We'll ask Chub to help us on this. I'm sorry for the mistake, I'll be more careful with counterweights in the future.
Moving on, I tuned all the SR2 OSEMs. It was fairly simple today since the table was leveled. I closed the chamber with the optic free to move and damped in all degrees of freedom.
Photos: https://photos.app.goo.gl/CQ6VouSB1HX2DPqW6
SUSPENSION STATUS UPDATED HERE |
Attachment 1: DJI_0144.JPG
|
|
16604
|
Thu Jan 20 16:42:55 2022 |
Paco | Update | BHD | AS4 OSEMs installation - part 2 |
[Paco]
Turns out, the shifting was likely due to the table level. Because I didn't take care the first time to "zero" the level of the table as I tuned the OSEMs, the installation was b o g u s. So today I took time to,
a) Shift AS4 close to the center of the table.
b) Use the clean level tool to pick a plane of reference. To do this, I iteratively placed two counterweights (from the ETMX flow bench) in two locations in the breadboard such that I nominally balanced the table under this configuration to zome reference plane z0. The counterweight placement is of course temporary, and as soon as we make further changes such as final placement of AS4 SOS, or installation of AS1, their positions will need to change to recover z=z0.
c) Install OSEMs until I was happy with the damping. ** Here, I noticed the new suspension screens had been misconfigured (probably c1sus2 rebooted and we don't have any BURT), so quickly restored the input and output matrices.
SUSPENSION STATUS UPDATED HERE |
16603
|
Thu Jan 20 12:10:51 2022 |
Yehonathan | Update | BHD | AS1 resurrection |
I was wondering whether I should take AS1 down to redo the wire clamping on the side blocks. I decided to take the OpLev spectrum again to be more certain. Attachments 1,2,3 show 3 spectra taken at different times.
They all show the same peaks 744mHz, 810mHz, 1Hz. So I think something went wrong with yesterday's measurement. I will not take AS1 down for now. We still need to apply some glue to the counterweight. |
Attachment 1: FreeSwingingSpectra.pdf
|
|
Attachment 2: FreeSwingingSpectra_div_20mV.pdf
|
|
Attachment 3: FreeSwingingSpectra_div_50mV.pdf
|
|
16602
|
Thu Jan 20 01:48:02 2022 |
Koji | Update | BHD | PR2 transmission calculation updated |
IMC is not such lossy. IMC output is supposed to be ~1W.
The critical coupling condition is G_PRC = 1/T_PRM = 17.7. If we really have L_arm = 50ppm, we will be very close to the critical coupling. Maybe we are OK if we have such condition as our testing time would be much longer in PRMI than PRFPMI at the first phase. If the arm loss turned out to be higher, we'll be saved by falling to undercoupling.
When the PRC is close to the critical coupling (like 50ppm case), we roughly have Tprc x 2 and Tarm to be almost equal. So each beam will have 1/3 of the input power i.e. ~300mW. That's probably too much even for the two OMCs (i.e. 4 DCPDs). That's OK. We can reduce the input power by 3~5.
Quote: |
LO Power
LO power when PRFPMI is locked is calculated by assuming 1 W of input power to IMC. IMC is assumed to let pass 10% of the power ( ).
|
|
16601
|
Thu Jan 20 00:26:50 2022 |
Koji | Update | SUS | Temporary watchdog |
As the new db is made for c1susaux, 1) it needs to be configured to be read by c1susaux 2) it requires restarting c1susaux 3) it needs to be recorded by FB 4) and restartinbg FB.
(^-Maybe not super exact procedure but conceptually like this)
cf.
https://wiki-40m.ligo.caltech.edu/How_To/Add_or_rename_a_daq_channel
|
16600
|
Wed Jan 19 21:39:22 2022 |
Tega | Update | SUS | Temporary watchdog |
After some work on the reference database file, we now have a template for temporary watchdog implementation for LO1 located here "/cvs/cds/caltech/target/c1susaux/C1_SUS-AUX_LO1.db".
Basically, what I have done is swap the EPICS asyn analog input readout for the COIL and OSEM to accessible medm channels, then write out watchdog enable/disable to coil filter SW2 switch. Everything else in the file remains the same. I am worried about some of the conversions but the only way to know more is to see the output on the medm screen.
To test, I restarted c1su2 but this did not make the LO1 database available, so I am guessing that we also need to restart the c1sus, which can be done tomorrow. |
16599
|
Wed Jan 19 18:15:34 2022 |
Yehonathan | Update | BHD | AS1 resurrection |
Today I suspended AS1. Anchal helped me with the initial hanging of the optics. Attachments 1,2 show the roll balance and side magnet height. Attachment 3 shows the motion spectra.
The major peaks are at 668mHz, 821mHz, 985mHz.
For some reason, I was not able to balance the pitch with 2 counterweights as I did with the rest of the thin optics (and AS1 before). Inserting the weights all the way was not enough to bring the reflection up to the iris aperture that was used for preliminary balancing. I was able to do so with a single counterweight (attachment 4). I'm afraid something is wrong here but couldn't find anything obvious. It is also worth noting that the yaw resonance 668mHz is different from the 755mHz we got in all the other optics. Maybe one or more of the wires are not clamped correctly on the side blocks?
The OSEMs were pushed into the OSEM plate and the plates were adjusted such that the magnets are at the center of the face OSEMs. The wires were clamped and cut from the winches. The SOS is ready for installation.
Also, I added a link to the OSEM assignments spreadsheet to the suspension wiki.
I uploaded some pictures of the PEEK EQ stops, both on the thick and thin optics, to the Google Photos account. |
Attachment 1: AS1_roll_balance2.png
|
|
Attachment 2: AS1_magnet_heigh2.png
|
|
Attachment 3: FreeSwingingSpectra.pdf
|
|
Attachment 4: IMG_6324.JPG
|
|
16598
|
Wed Jan 19 16:22:48 2022 |
Anchal | Update | BHD | PR2 transmission calculation updated |
I have further updated my calculation. Please find the results in the attached pdf.
Following is the description of calculations done:
Arm cavity reflection:
Reflection fro arm cavity is calculated as simple FP cavity reflection formula while absorbing all round trip cavity scattering losses (between 50 ppm to 200 ppm) into the ETM transmission loss.
So effective reflection of ETM is calculated as


The magnitude and phase of this reflection is plotted in page 1 with respect to different round trip loss and deviation of cavity length from resonance. Note that the arm round trip loss does not affect the sign of the reflection from cavity, at least in the range of values taken here.
PRC Gain
The Michelson in PRFPMI is assumed to be perfectly aligned so that one end of PRC cavity is taken as the arm cavity reflection calculated above at resonance. The other end of the cavity is calculated as a single mirror of effective transmission that of PRM, 2 times PR2 and 2 times PR3. Then effective reflectivity of PRM is calculated as:


Note, that field transmission of PRM is calculated with original PRM power transmission value, so that the PR2, PR3 transmission losses do not increase field transmission of PRM in our calculations. Then the field gain is calculated inside the PRC using the following:

From this, the power recycling cavity gain is calculated as:

The variation of PRC Gain is showed on page 2 wrt arm cavity round trip losses and PR2 transmission. Note that gain value of 40 is calculated for any PR2 transmission below 1000 ppm. The black verticle lines show the optics whose transmission was measured. If V6-704 is used, PRC Gain would vary between 15 and 10 depending on the arm cavity losses. With pre-2010 ITM, PRC Gain would vary between 30 and 15.
LO Power
LO power when PRFPMI is locked is calculated by assuming 1 W of input power to IMC. IMC is assumed to let pass 10% of the power ( ). This power is then multiplied by PRC Gain and transmitted through the PR2 to calculate the LO power.

Page 3 shows the result of this calculation. Note for V6-704, LO power would be between 35mW and 15 mW, for pre-2010 ITM, it would be between 15 mW and 5 mW depending on the arm cavity losses.
The power available during alignment is simply given by:


If we remove PRM from the input path, we would have sufficient light to work with for both relevant optics.
I have attached the notebook used to do these calculations. Please let me know if you find any mistake in this calculation. |
Attachment 1: PR2transmissionSelectionAnalysis.pdf
|
|
Attachment 2: PR2_Trans_Calc.ipynb.zip
|
16597
|
Wed Jan 19 14:41:23 2022 |
Koji | Update | BHD | Suspension Status |
Is this the correct status? Please directly update this entry.
LO1 [Glued] [Suspended] [Balanced] [Placed In Vacuum] [OSEM Tuned] [Damped]
LO2 [Glued] [Suspended] [Balanced] [Placed In Vacuum] [OSEM Tuned] [Damped]
AS1 [Glued] [Suspended] [Balanced] [Placed In Vacuum] [OSEM Tuned] [Damped]
AS4 [Glued] [Suspended] [Balanced] [Placed In Vacuum] [OSEM Tuned] [Damped]
PR2 [Glued] [Suspended] [Balanced] [Placed In Vacuum] [OSEM Tuned] [Damped]
PR3 [Glued] [Suspended] [Balanced] [Placed In Vacuum] [OSEM Tuned] [Damped]
SR2 [Glued] [Suspended] [Balanced] [Placed In Vacuum] [OSEM Tuned] [Damped]
Last updated: Fri Jan 28 10:34:19 2022 |
16596
|
Wed Jan 19 12:56:52 2022 |
Paco | Summary | BHD | AS4 OSEMs installation |
[Paco, Tega, Anchal]
Today, we started work on AS4 SOS by checking the OSEM and cable. Swapping the connection preserved the failure (no counts) so we swapped the long OSEM for a short one that we knew was working instead, and this solved the issue. We proceeded to swap in a "yellow label", long OSEM in place and then noticed the top plate had issues with the OSEM threads. We took out the bolt and inspected its thread, and even borrowed the screw from PR2 plate but saw the same effect. Even using a silver plated setscrew such as the SD OSEM one resulted in trouble... Then, we decided to not keep trying weird things, and took our sweet time to remove the UL, UR OSEMs, top earthquake stops, and top plate carefully in-situ. Then, we continued the surgery by installing a new top plate which we borrowed from the clean room (the only difference is the OSEM aperture barrels are teflon (?) rather than stainless. The operation was a success, and we moved on to OSEM installation.
After reaching a good place with the OSEM installation, where most sensors were at 50% brightness level and we were happy with the damping action (!!), we fixed all EQ stops and proceeded to push the SOS to its nominal placement. Then upong releasing the EQ stops, we found out that the sensor readings were shifted. |
16595
|
Wed Jan 19 12:50:10 2022 |
Anchal | Summary | BHD | Part IV of BHR upgrade - SR2 OSEM tuning progress. |
It was indeed the issue of the top OSEM plate not being in the right place horizontally. But the issue was more non-trivial. I believe because of the wedge in thick optics, there is a YAW offset in the optic in the free hanging position. I had to readjust the OSEM plate 4 times to be able to get full dark to bright range in both upper OSEMs. After doing that, I tuned the four OSEMs somewhat near the halfway point and once I was sure I'm inside the sensitive region in all face OSEMs, I switched on POS, PIT, and YAW damping. Then I was able to finely tune the positions of both upper OSEMs.
However, on reaching to lower right OSEM, I found again the same issue. I had to stop to go to the 40m meeting, I'll continue this work in the afternoon. But OSEM plate adjustment in the horizontal direction, particularly for thick optics is required to be done before transporting them. I achieved the best position by turning the OSEM 90 degrees and using the OSEM LED/PD plates to determine the position. This was the final successful trial I did in adjusting the plate position horizontally.
|
16594
|
Tue Jan 18 18:19:22 2022 |
Koji | Summary | BHD | AS4 placed in ITMY Chamber, OSEMs connected |
AS4 satellite amplifier D1002818 / D080276 troubleshoot
I dug into the circuit to see what/where things were wrong.
- UL saturation issue: The open light voltage at the TIA output (I-V out) was 10.4V. It seemed that the photocurrent of 86uA was simply a little too much for the transimpedance gain of 121kOhm. So the R18 was replaced to 100kOhm. This made the I-V out to be 8.6V and the ADC input count to be 28200 (Attachment 1). This modification was done on the unit S2100742 CH1 (LEFT CH1)
- Non responding LL issue: Now moved on to LL (LEFT CH2). The basic circuit test didn't reveal any problem. So the DSUB25 cables were swapped at the vacuum feedthru flange. The result is shown in Attachment 2. LL OSEM issue was moved to the 2nd ch of the right channel of the sat amp (CH6). This is means that the problem is somewhere in the vacuum chamber (including the vacuum feedthru). We need to check the in-vacuum cable and the OSEM. We can test the OSEM by swapping the position of the OSEM connector between LL and UL (for example).
|
Attachment 1: Screen_Shot_2022-01-18_at_17.48.16.png
|
|
Attachment 2: cable_swap.png
|
|
16593
|
Tue Jan 18 18:16:28 2022 |
Anchal | Summary | BHD | Part IV of BHR upgrade - SR2 Sat Amp Box inspection |
I tested the monitor ports on the SR2 Sat Amp Box but found that all LED Mon and PD Mon are giving expected values. I disconnected the cable to OSEM and checked the PD monitors and found no offsets in case of no PD current. I realised that PD transimpedance offset should be checked with PD inputs shorted instead. So I created a male DB 25 connector with pinds 2-3, 50-6, 8-9 and 11-12 shorted. This on connecting to the OSEM cable at the back of sat amp boxes should short the PD inputs. On using this plug, I found no offsets in any of the Sat Amp PD output channels.
Discussion with Koji
It is possible that the issue is there because the magnet is missing the LED-PD path way because it is offset sideways. In fact, in my limited memory, I do not recall seeing the UL OSEM signal ever going to complete darkness either. Tomorrow, we should take a photo of the OSEMs from the back and see if any sideways adjustment of the top OSEM plate is required. If any adjustment is required, we must take the OSEMs out and then do the adjustment. Do not attempt to adjust OSEM plate with OSEMs inserted in-situ. That will most probably knock off the magnets. |
16592
|
Tue Jan 18 15:19:49 2022 |
Paco | Summary | BHD | Part IV of BHR upgrade - Replaced old SR3 with SOS SR2. OSEM tuning attemtped. |
[Tega, Anchal, Paco]
We started working on SR2 installation. Preliminary work involved
- Removing SR3 from BS chamber. For whatever reason this optic was still installed. It got relocated in the west corner of the ETMX flow bench.
That was pretty much it. After identifying the cabling situation, we proceeded to bring SR2 from the cleanroom. The magnets and wires remained well through their travel.
Connected OSEM one-by-one. Starting from top right to left (PIn1)
1st connector: LL -> UR -> UL
2nd connector: LR -> SD** (we had some trouble here where the first time we made a connection we didn't see any signal, after a brief review of cables, sat amp unit, cables again with Koji, and sat amp again, we found out a connection was not done in the front of the SR2 SatAmp box, after which we saw the sensor signals).
Loosening all OSEMs and taking them out and noting full bright readings:
- SD: 29600 -> 14980
- LR: 21840 -> 10920
- UR: 25900 -> 12950
- LL: 28020 -> 14010
- UL: 25750 -> 12875
After finishing the initial SD osem tuning, we moved onto UL, and then to UR, but we noticed that the UR was not able to drop to its target value of ~13000 counts, even when the OSEM face was < 1 mm from the adapter (see Attachments #1-2). Apart from becoming harder to push in, it became apparent that the dark level (full shadow) is not consistent with ~ 0 counts; is there an offset coming from SatAmp? We quickly checked the OSEM by replacing it in-situ with another working one from the cleanroom batch, but the issue persisted. We decided to stop here, as we suspect the SatAmp box might have some issue.
Photos: https://photos.app.goo.gl/HXzG6mj4oxKTksjZ7 |
Attachment 1: PXL_20220119_012954365.jpg
|
|
Attachment 2: PXL_20220119_013003522.jpg
|
|
16591
|
Tue Jan 18 14:26:09 2022 |
Paco | Summary | BHD | Part IX of BHR upgrade - Placed remaining filters SR2, PR3, PR2 |
[Anchal, Paco]
Added input filters, input matrix, damping filters, output matrix, coil filters, and copy the state over from LO1 into SR2, PR2, PR3 screens in anticipation for damping. |
16590
|
Fri Jan 14 18:12:47 2022 |
Anchal | Summary | BHD | AS4 placed in ITMY Chamber, OSEMs connected |
AS4 was succesfully suspended and trasported to ITMY chamber (40m/16589). We placed it near the door to make it easy to tune the OSEMs. We connected the OSEMs and found that the LL OSEM is not responding. Even though the the OSEMs are completely out right now, there was no signal on this OSEM. This could be an issue in either at the LED driver circuit or the PD circuit in AS4 Sat Amp box, or it could be the OSEM that is bad. We'll investigate further next day. For now, we recorded the full brightness reading for the OSEMs:
- UL: 32767 -> 16383
- UR: 29420 -> 14710
- LR: 30100 -> 15050
- SD: 29222 -> 14611
Another thing to note is that UL value above is not changing at all. I checked the CDS screen and the the ADC input is overflowing in complete bright position of the OSEM. |
16589
|
Fri Jan 14 17:33:10 2022 |
Yehonathan | Update | BHD | AS4 resurrection |
{Yehonathan, Anchal}
Came this morning, the gluing of the magnets was 100% successful. Side blocks, counterweights were assembled. We suspend AS4 and adjust the roll balance and the magnet height (attachments 1,2). OpLev was slightly realigned.
The pitch was balanced. We had to compensate for the pitch shift due to the locking of the counterweights. Once we got good pitch balance, the motion spectrum was taken (attachment 3). Major peaks are at 755mHz, 953mHz, 1040mHz.
Previous peaks were 755mHz, 964mHz, and 1.062Hz so not much has changed. We pushed back the OSEMs, adjusted OSEM plate and locked it tightly. We lock the EQ stops and transfer AS4 to the vacuum chamber in foil. We open the foil inside the chamber. No magnets were broken. Everything seems to be intact. We connect the OSEMs to CDS. |
Attachment 1: AS4_roll_balance2.png
|
|
Attachment 2: AS4_magnet_height2.png
|
|
Attachment 3: FreeSwingingSpectra.pdf
|
|
16588
|
Fri Jan 14 14:04:51 2022 |
Paco | Update | Electronics | RFSoC 2x2 board arrived |
The Xilinx RFSoC 2x2 board arrived right before the winter break, so this is kind of an overdue elog. I unboxed it, it came with two ~15 cm SMA M-M cables, an SD card preloaded with the ARM processor and a few overlay jupyter notebooks, a two-piece AC/DC adapter (kind of like a laptop charger), and a USB 3.0 cable. I got a 1U box, lid, and assembled a prototype box to hold this board, but this need not be a permanent solution (see Attachment #1). I drilled 4 thru holes on the bottom of the box to hold the board in place. A large component exceeds the 1U height, but is thin enough to clear one of the thin slits at the top (I believe this is a fuse of some sort). Then, I found a brand new front panel, and drilled 4x 13/32 thru holes in the front for SMA F-F connectors.
I powered the board, and quickly accessed its tutorial notebooks, including a spectrum analyzer and signal generators just to quickly check it works normally. The board has 2 fast RFADCs and 2 RFDACs exposed, 12 and 14 bit respectively, running at up to 4 GSps. |
Attachment 1: PXL_20220114_211249499.jpg
|
|
16587
|
Fri Jan 14 13:46:25 2022 |
Anchal | Update | BHD | PR2 transmission calculation updated |
I updated the arm cavity roundtrip losses due to scattering. Yehonathan told me that arm cavity looses 50ppm every roundtrip other than the transmission losses. With the updated arm cavity loss:
|
PRFPMI LO Power (mW) |
Unlocked PRC LO Power (uW) |
PRC Gain |
pre-2010 ITM |
8 |
28 |
15.2 |
V6:704 |
24 |
113 |
12 |
|
Attachment 1: LO_power_vs_PR2_transmission.pdf
|
|
Attachment 2: PRC_Gain_vs_PR2_transmission.pdf
|
|
Attachment 3: PR2_Trans_Calc.ipynb.zip
|
16586
|
Fri Jan 14 12:01:21 2022 |
Anchal | Update | Electronics | BS & ITMY feedthroughs labeled and connected to Sat Amps |
I labeled all the newly installed flanges and connected the in-air cables (40m/16530) to appropriate ports. These cables are connected to the CDS system on 1Y1/1Y0 racks through the satellite amplifiers. So all new optics now can be damped as soon as they are placed. We need to make more DB9 plugs for setting "Acquire" mode on the HAM-A coil drivers since our Binary input system is not ready yet. Right now, we only have 2 such plugs which means only one optic and be damped at a time.
|
16585
|
Fri Jan 14 11:00:29 2022 |
Anchal | Update | BHD | PR2 transmission calculation |
Yeah, I counted the loss from arm cavities as the transmission from ETMs on each bounce. I assumed Michelson to be perfectly aligned to get no light at the dark port. Should I use some other number for the round-trip loss in the arm cavity? |
16584
|
Fri Jan 14 03:07:04 2022 |
Koji | Update | BHD | PR2 transmission calculation |
I opened the notebook but I was not sure where you have the loss per bounce for the arm cavity.
PRC_RT_Loss = 2 * PR3_T + 2 * PR2_T + 2 * Arm_Cavity_Finesse * ETM_T + PRM_T
Do you count the arm reflection loss to be only 2 * 13ppm * 450 = 1.17%? |
16583
|
Thu Jan 13 17:10:55 2022 |
Anchal | Update | BHD | PR2 transmission calculation |
I corrected the calculation by adding losses by the arm cavity ends times the arm cavity finesse and also taking into account the folding of the cavity mirror. I used exact formula for finesse calculation and divided it by pi to get the PRC gain from there. Attachment 3 is the notebook for referring to the calculations I made.
Note that using V6-704 would provide 35 mW of LO power when PRFPMI is locked and 113 uW for alignment, but will bring down the PRC Gain to 17.5.
pre-2010 ITM (if it is still an option) would provide 12 mW of LO power when PRFPMI is locked and 28 uW for alignment, but will keep the PRC Gain to 24.6.
I still have to do a curvature check on the V6-704 optic. |
Attachment 1: LO_power_vs_PR2_transmission.pdf
|
|
Attachment 2: PRC_Gain_vs_PR2_transmission.pdf
|
|
Attachment 3: PR2_Trans_Calc.ipynb.zip
|
16582
|
Thu Jan 13 16:08:00 2022 |
Yehonathan | Update | BHD | gluing magnets after AS1/4 misfortune |
{Yehonathan, Anchal, Paco}
In the cleanroom, we removed AS1 and AS4 from their SOS towers. We removed the mirrors from the adapters and put them in their boxes. The broken magnets were collected from the towers and their surfaces were cleaned as well as the magnet sockets on the two adapters and on the side block from where the magnets were knocked off.
We prepared our last batch of glue (more glue was ordered three days ago) and glue 2 side magnets and 2 face magnets. We also took the chance and apply glue on the counterweights on the thick optic adapters so there is no need to look for alternatives for now.
The peek screws and nuts were assembled on the thick optics SOS towers instead of the metal screws and nuts that were used as upper back EQ stops. |
16581
|
Thu Jan 13 12:29:27 2022 |
Anchal | Summary | BHD | AS4 LR magnet broke |
After the debacle with AS1 (40m/16580), we decided the put the PEEK earthquake stop by first removing the lower OSEM plate and then doing it. So I unfastened AS4 from its position with the earthquake stops in place and moved the suspension to the center of the table. Then I carefully removed the bottom OSEM plate. But I found out that the LR magnet is broken and lying on the floor of the suspension . Given my past on the same day, it could be me breaking it again during the moving of the suspension of taking off the OSEM plate or there is a small chance that this break happened before. Regardless of fault, this meant we have to resuspend AS4 again as well. So we transported AS4 back to the clean room and the work on it's re-suspension has begun. |
16580
|
Thu Jan 13 12:24:08 2022 |
Anchal | Summary | BHD | AS1 SD and LR magnets broke |
[Anchal (vacuum work), Paco (outside)]
After the AS1 Sat Amp fix (40m/16579), we today were able to tune all OSEMs to the mid-bright level. But when we were about to call it, we were told that the new PEEK earthquake stop screw and bolts need to go on the thin suspended optics. Against better judgment, we decided to install the new back earthquake stop in-situ since we had tuned all OSEMs already. I installed the new stop but after that found that in the process I have broken off the side magnet and LR magnet from the optic adaptor and they are inside the OSEM coils now. This means we'll have to redo the AS1 suspension almost from scratch again . We have transported AS1 to the cleanroom where the work on re-suspension has begun. |
16579
|
Thu Jan 13 09:48:41 2022 |
Anchal | Summary | BHD | AS1 Sat Amp fixed |
I fixed the issue in AS1 Sat Amp (S2100741) by using a razor blade. I cut the short between the two places, cleaned up the area and covered it with electrical tape. However, later feedback from Rana was to not use electrical tape as it dries up and becomes brittle and lfaky in long run. So after the AS1 OSEM tuning is over, I'll open this box again and use something else to insulate the exposed area. See attached pictures for current status.
|
Attachment 1: signal-2022-01-13-094823_001.jpeg
|
|
Attachment 2: signal-2022-01-13-094823_002.jpeg
|
|