  2076   Fri Oct 9 16:36:13 2009 rob UpdateIOOfrequency noise problem

I used the XARM as a reference to measure the frequency noise after the MC.  It's huge around 4kHz--hundreds of times larger than the frequency noise the MC servo is actually squashing.  This presents a real problem for our noise performance.

An elog search reveals that this noise has been present (although not calibrated till now) for years.  We're not sure what's causing it, but suspicion falls on the piezojena input PZTs. 

I didn't bother too much about it before because we previously had enough common mode servo oomph to squash it below other DARM noises, and I didn't worry too much about stuff at 4kHz..  Now that we have a weaker FSS and thus much weaker CM servo, we can't squash it, and the most interesting feature of our IFO is at 4kHz. 

I'll measure the actual voltage noise going to the PZTs.  I remember doing this before and concluding it was ok, but can't find an elog entry.  So this time maybe I'll  do it right.

Attachment 1: freqnoiseaftermc.png
freqnoiseaftermc.png
  2077   Fri Oct 9 17:37:21 2009 JenneUpdateIOOMC2 trans beam is dumped

Last night while noise hunting, Rana found that the MC2 trans beam has been left for an unknown length of time just hitting the side of the black enclosure box.  Today I put a brand-spankin'-new razor dump on the MC2 table, to dump the beam.

  2102   Fri Oct 16 10:15:02 2009 steveUpdateIOOIP ang & pos recentered

Pointing stability of 4 days. Initial pointing does not go through suspended optics. It is launched  right after the Piezo Jena steering mirrors in the BS chamber.

IP-ANG on epics screen is  C1:ASC-IBQPD_X and Y in dataviewer  were recentered. This beam is clipping a bit in ETMX chamber  pick off mirror.

IP-POS pick  off is in the BS chamber and it's qpd on the BS_ISCT This beam is also clipping just a little bit. This is easy to fix. We'll have to remove an iris from the BS optical levers table.

note: arms were not locked when I recentered

Attachment 1: ip4d.jpg
ip4d.jpg
  2123   Tue Oct 20 02:14:29 2009 robUpdateIOOMC2 alignment bias changed

the mode cleaner was having trouble locking in a 00 mode, needing several tries.  I changed the MC2 coil biases, and it seems better for now.

  2124   Tue Oct 20 10:46:18 2009 steveUpdateIOOclipping IP-ANG beam at ETMX chamber

Initial pointing beam is clearly clipping on 2" pick off mirror in  ETMX vacuum chamber.

Atm. 1  The pick off mirror is just north west of the ETMX test mass

Atm. 2 The camera is looking in from the north view port of ETMX chamber. The back side of pick off mirror is visible now with the face view of the "IP-ANG-OUT" mirror.

 

Attachment 1: PA200175.JPG
PA200175.JPG
Attachment 2: PA200177.JPG
PA200177.JPG
  2132   Thu Oct 22 08:45:58 2009 steveUpdateIOOIP ang & pos recentered

Quote:

Pointing stability of 4 days. Initial pointing does not go through suspended optics. It is launched  right after the Piezo Jena steering mirrors in the BS chamber.

IP-ANG on epics screen is  C1:ASC-IBQPD_X and Y in dataviewer  were recentered. This beam is clipping a bit in ETMX chamber  pick off mirror.

IP-POS pick  off is in the BS chamber and it's qpd on the BS_ISCT This beam is also clipping just a little bit. This is easy to fix. We'll have to remove an iris from the BS optical levers table.

note: arms were not locked when I recentered

 IP-ANG clipping can be traced back to our last vent of Aug. 18, 2008  See elog entry #845

This was an after earth quake - sus repair vent

Attachment 1: pointing1000.jpg
pointing1000.jpg
  2144   Mon Oct 26 18:15:57 2009 robUpdateIOOMC OLG

I measured the mode cleaner open loop gain.  It's around 60kHz with 29 degs of phase margin.

  2168   Mon Nov 2 13:00:55 2009 KojiUpdateIOOPMC aligned, MC WFS aligned

The beam to PMC aligned. The beam to MC WFS cameras aligned.
PMC Trans increased from 2.73 to 2.75 (+1%).
MC Trans increased from 7.80 to 7.87 (+1%).

  2172   Tue Nov 3 03:45:04 2009 rob UpdateIOOfrequency noise problem

Quote:

I used the XARM as a reference to measure the frequency noise after the MC.  It's huge around 4kHz--hundreds of times larger than the frequency noise the MC servo is actually squashing.  This presents a real problem for our noise performance.

An elog search reveals that this noise has been present (although not calibrated till now) for years.  We're not sure what's causing it, but suspicion falls on the piezojena input PZTs. 

I didn't bother too much about it before because we previously had enough common mode servo oomph to squash it below other DARM noises, and I didn't worry too much about stuff at 4kHz..  Now that we have a weaker FSS and thus much weaker CM servo, we can't squash it, and the most interesting feature of our IFO is at 4kHz. 

I'll measure the actual voltage noise going to the PZTs.  I remember doing this before and concluding it was ok, but can't find an elog entry.  So this time maybe I'll  do it right.

 

This level of frequency noise has not changed, but we now have increased common mode servo gain and so it's not as huge of a deal, although we should still probably do something about it. 

 

Attached is a plot of the piezojena noise measurement, estimated into Hz, along with another measurement of frequency noise as described above. 

To get the piezojena voltage noise into Hz, I estimated the PZTs within have a flat 2 micron/V response (based on a rough knowledge of their geometry and assuming a 10 milliradian / 150V steering range).  This is the voltage noise with the PZTs operating in closed loop mode, which is how we normally run them.  This plot also ignores the transfer function of the Pomona box, as we are mainly looking at noise in the kHz band.  I think this plot shows that these PZTs are a good candidate for creating this frequency noise, especially near their mechanical resonances (the manual says the unloaded resonances are in the 3-4kHz range).   

I've been operating one DOF of the piezojenas in open loop mode for a couple of weeks now, and the feared drift has not been a problem at all.  If we plan to keep using these after the upgrade, we should definitely put some big resistors in series at the outputs and operate them in open loop mode.

Also attached is a plot of RF DARM noise, with a frequency noise spectrum.  That spectrum is a REFL 2I spectrum put into DARM units using a measured TF (driving MC_L and measuring REFL 2I and DARM_ERR), and then put into meters using the same DARM calibration as used for the DARM curve.

Attachment 1: noise.png
noise.png
Attachment 2: spectra.pdf
spectra.pdf
  2227   Tue Nov 10 17:01:33 2009 AlbertoConfigurationIOOc1ioovme and c1iool0 rebooted

This afternoon, while I was trying to add the StochMon channels to the frames, I rebooted the c1ioovme and c1iool0.

I had to do it twice because of a mispelling in the C1IOO.INI file that the first time prevented the computer to restart properly.

Eventually I restored the old .ini file, as it was before the changes.

After rebooting I also burtrestored.

During the process the mode cleaner got unlocked. Later on the autoclokcer couldn't engage. I had to run the MC_down and MC_up scripts.

  2234   Wed Nov 11 09:48:04 2009 steveUpdateIOOwhere is IOO-RFAMPD_DCMON ?

RFAMPD_DCMON disappered on Nov 5, 2009

Attachment 1: rfampddc.jpg
rfampddc.jpg
  2318   Mon Nov 23 21:36:38 2009 KojiUpdateIOOAligned PMC/RC

I aligned the beam goes to PMC. It increased the MC Trans from 8.25 to 8.30.

I also aligned the beam goes to RC.
When I touched the FSS box (wrong: this was the VCO driver) that was close to one of the steering mirror, suddenly the RC trans increased.
It is now 9.8. I am afraid that it gets saturated. I could not reproduce the phenomenon. This could be caused by a bad contact?
Note that I didn't see there is any loose optic.

Attachment 1: 091123_PSL.png
091123_PSL.png
  2357   Sat Dec 5 17:34:30 2009 robUpdateIOOfrequency noise problem

There's a large broadband increase in the MC_F spectrum.  I'm not totally sure it's real--it could be some weird bit-swapping thing.  I've tried soft reboots of c1susvme2 and c1iovme, which haven't helped.  In any case, it seems like this is preventing any locking success today.  Last night it was fine.

Attachment 1: mcf.png
mcf.png
  2359   Sat Dec 5 22:31:52 2009 robUpdateIOOfrequency noise problem

Quote:

There's a large broadband increase in the MC_F spectrum.  I'm not totally sure it's real--it could be some weird bit-swapping thing.  I've tried soft reboots of c1susvme2 and c1iovme, which haven't helped.  In any case, it seems like this is preventing any locking success today.  Last night it was fine.

 Rebooting c1iovme (by keying off the crate, waiting 30 seconds, and then keying it back on and restarting) has resolved this.  The frequency noise is back to the 'usual' trace.

  2406   Sun Dec 13 20:50:45 2009 ranaSummaryIOOMach Zender Calibration

I ramped the MZ PZT (with the loop disabled on the input switch) to calibrate it. Since the transmission has been blocked, I used the so-called "REFL" port of the MZ to do this.

The dark-to-dark distance for the MZ corresponds to 2 consecutive destructive interferences. Therefore, that's 2 pi in phase or 1 full wavelength of length change in the arm with the moving mirror.

Eyeballing it on the DTT plot (after lowpassing at 0.1 Hz) and using its cursors, I find that the dark-to-dark distance corresponds to 47.4 +/- 5 seconds.

So the calibration of the MZ PZT is 88 +/- 8 Volts/micron.

Inversely, that's a mean of 12 nm / V.

why am I calibrating the MZ? Maybe because Rob may want it later, but mainly because Koji won't let me lock the IFO.

Apparently, we haven't had a fast channel for any of the MZ board. So I have temporarily hooked it up to MC_DRUM at 21:13 and also turned down the HEPA. Now, let's see how stable the MZ and PMC really are overnight.

EDIT: it railed the +/- 2V ADCwe have so I put in a 1:4 attenuator via Pomona box. The calibration of MC_DRUM in terms of MZ_PZT volts is 31.8 cts/V.

So the calibration of MC_DRUM1 in meters is: 0.38 nm / count


Attachment 1: Untitled.png
Untitled.png
  2407   Sun Dec 13 23:18:09 2009 ranaSummaryIOODisplacement noise on the PSL table

For the Laser Gyro, I wondered how much mechanical noise we might get with a non-suspended cavity. My guess is that the PMC is better than we could do with a large ring and that the MZ is much worse than we could do.

Below 5 Hz, I think the MZ is "wind noise" limited. Above 10 Hz, its just ADC noise in the readout of the PZT voltage.

Attachment 1: mz.pdf
mz.pdf
  2431   Fri Dec 18 15:40:33 2009 KojiUpdateIOOMC2 spot centered / MCT QPD issue

This afternoon I felt like saying hello to the input mode cleaner. So I decided to center the spot on MC2.

Motivation

MC has 6 alignment dofs. 4 of them are controlled by the WFSs. Remaining 2 appears at the spot position on MC2.
If the spot on the MC2 is fixed, the beam hits the same places of three mirrors. If the mirrors are completely fixed
in terms of the incident beam, I suppose the reflected beam is also fixed. This makes the WFS spots more stable.
Then I feel better.

Today's goal is to confirm the behaviour of MC such as dithering amplitude, response of the couplings to the alignment,
behavior of the WFS, and the transmitted power.


Method

1) Turned off MC auto locker. Turned off MC WFS as the WFS servos disturbs my work.
2) Dithered MC2 in Pitch and Yaw using DTT. There looks elliptic filter (fc=28Hz) in the ASC path, I used 20Hz-ish excitations.
- C1:SUS-MC2_ASCPIT_EXC 100cnt_pk@19Hz
- C1:SUS-MC2_ASCYAW_EXC 100cnt_pk@22Hz
3) Looked at C1:SUS-MC2_MCL_OUT to find the peaks at 19Hz and 22Hz. These are caused by alignment-length coupling.
If they are minimized I assume the spot is somehow centered on MC2.
Note: This may not be the true center. The suspension response should be investigated. But this is a certain reporoducible spot position.
Note: I should use ezcademod in order to obtain the phase information of the dither result.

4) Move MC2 Pitch for certain amount (0.01cnt) by the alignment slider. Align MC1/MC3 to have max transmittion.
5) If the Pitch peak got lower, the direction of 4) was right. Go further.
5') If the Pitch peak got higher, the direction of 4) was wrong. Go the other direction.

6) Repeat 4)&5) for Yaw.


Result

After the adjustment, the couplings got lower about 10 times. (Sorry! The explanation is not so scientific.)
Next time I (or someone) should make a script to do it and evaluate the coupling by the estimated distance of the spot from the center of the mirror (the center of the rotation).
I have not seen visible change in the spectrum of C1:SUS-MC2_MLC_OUT.


MCT QPD issue

By the spot centering, I could expected to see some improvement of the transmittion. But in reality, there was no change.
In fact, the transmittion power was getting down for those weeks.

I checked WFS and MCT paths. Eventually I found that a couple of possible problems:
1) MCT Total output varies more than 10% depending on the spot position on MCT QPD.
2) Just before the QPD, there is a ND1 filter.
This may suggest that:
a) Four elemtns of the MCT QPD have different responses.
b) The ND filter is causing a fringe.

So far I aligned the ND filter to face the beam. The reflection from the filter was blocked at a farther place.
Still the output varies on the spot position. Probably I have to look at the QPD someday.

So far the spot on the QPD was defined so that I get the maximum output from the QPD. This is about 8.8.
As I touched the steering mirrors, the X and Y outputs of the QPD are no longer any reference.

For now, I closed the PSL table. The full IFO was aligned.

  2435   Sun Dec 20 23:42:44 2009 JenneUpdateIOONew Input Mode Matching Telescope

I've got most of the new Mode Matching Telescope figured out.  The scripts and an example result are at: MMT09 wiki  (Rather, the scripts are in the svn: MMT svn)

Issues still to be resolved: 

* We're getting pretty iffy 'angles' between tilt and translation when using the mode matching mirrors for steering.

* I haven't taken into account the astigmatism which occurs when you tilt the mode matching mirrors. 

The nifty thing about these scripts is that they take a look at the mode matching overlap:  For each possible mode matching solution it adds noise to all of the distances and radii of curvature during ~10,000 iterations and plots a histogram of the overlap so that we can see which solutions have a better chance of giving us the optimal overlap, even if we place the optics in slightly the wrong place.

 

I'd like to update the overlap part of the script with the astigmatism business:  do we lose goodness of overlap if we tilt the mirrors by a bit?  I think this will require redoing the overlap part with the X and Y directions separate.  Koji has done this in the past.  My current code assumes that the beam is always symmetric in X and Y. 

  2444   Tue Dec 22 11:23:51 2009 kiwamu, SteveUpdateIOOMC relocked

In this morning I found MC unlocked.

Steve restored the watchdogs before I found that.

Then I relocked MC and now MC is locked and working well.

The reflected DC power is ~0.38, which is usual number.

 

  2448   Wed Dec 23 16:34:25 2009 KojiUpdateIOOMCT QPD/MC REFL QPD disabled

For a certain investigation of the sum/diff module for MCT QPD/MC REFL QPD, I removed it from the system.

 

  2451   Thu Dec 24 19:13:29 2009 KojiUpdateIOOMCT QPD investigation

I found that MCT QPD has a dependence of the total output on the position of the spot. Since the QPD needs the supply and bias voltages from the sum/diff amp, I could not separate the problems of the QPD itself and the sum/diff amplifier by the investigation on Tuesday. On Wednesday, I investigated a generic quad photodiode interface module D990692.

...I was so disappointed. This circuit was left uninvestigated and used so long time with the following sorrowful conditions.
- This circuit has 4 unbuffered inputs with input impedance of 300~400 Ohm. It's way too low!
- Moreover, those channels have different input impedances. Ahhhh.
- Even worse, the QPD circuit D990272 has output impedance of 50 Ohm.
- The PCB of this circuit has four layers. It is quite difficult to make modifications of the signal route.
- It is a headache: this circuit is "generic" and used in many places.

D990692 has 4 channel inputs that are not buffered. Each channel has two high impedance buffers but they are used only for the monitors. The signal paths have no buffer.

The differential amplifier is formed by R=1k Ohm. The inverted side of the input has 1kOhm impedance. The non-inverted side has 1.5kOhm impedance.

CH1: 10K // 1.5k // 1.5k // 1k = 411 Ohm
CH2: 10K // 1.5k // 1k // 1k = 361 Ohm
CH3: 10K // 1k // 1k // 1k = 323 Ohm
CH4: 10K // 1k // 1.5k // 1k = 361 Ohm

Considering the output impedance of 50Ohm for the QPD, those too low input impedances result in the following effect:
- Because of the voltage division, we suffer absolute errors of 10.8~13.4%. This is huge.
- Because of the input impedance differences, we suffer a relative error of 1.5%~3%. This is also huge.

Unfortunately, the circuit has no room to modify; the signal paths are embedded in the internal layer.

I decided to replace the resistors of the sum/diff amps from 1k to 10k. Also the input impedance of the buffer was removed as the input is terminated by the sum/diff amps in any case.This changes the input inpedance to the followings:

CH1: 15k // 15k // 10k = 4286 Ohm
CH2: 15k // 10k // 10k = 3750 Ohm
CH3: 10k // 10k // 10k = 3333 Ohm
CH4: 10K // 15k // 10k = 3750 Ohm

These yield the absolute error of 1.2-1.5%. The relative error is now 0.3%. I can accept these numbers, but later I should put additional terminating resistors to compensate the differencies.

So far I have modified the resistors for the MCT as the modification for a QPD needs 17 10k resistors.
Next thing I have to check is the dependence of the QPD outputs on the spot positions.

-----------------------------------------------

Edit: Feb 11, 2010

I talked with Frank and he pointed out that the impedances are not the matter but the gains of the each channels are the matters (after considering the output impedance of the QPD channels).
If we assume the ideal voltage sources at the QPD and the symmetric output impedances of 50Ohm, the gain of the each circuit are affected but the change should be symmetric.

He found that several things:
- The analog switch (MAX333) used in the QPD unit adds more output impedance (somewhat randomly!).
- The resistance of the sum/diff circuits may vary each other unless we use 0.1% resistors.

 

Attachment 1: D990692.png
D990692.png
  2473   Mon Jan 4 17:21:30 2010 JenneConfigurationIOOElusive Mode Matching Solution found!

I think I have finally found a Mode Matching solution for our new Input Mode Matching Telescope!  And after looking at the layout diagram with Koji and Raffaele, it seems like all of the optics will fit into the chambers / onto the tables (not true as of last week). 

3. RoCMMT1 is -5m
   RoCMMT2 is 8m,
   with the MMTs 1.89m apart.
   This is a 1.6x telescope.
   MMT2 is 2.2641m from the PRM
   MMT1 is 2m from MC3.
   The Condition Number for this optical chain is 89219047.5781.

This layout is very similar to the one that Koji posted on the wiki yesterday:  Upgrade09/Optical Layout.  The difference is that I want to move MMT1 ~20cm closer to the MC13 table, so just on the other side of the main red beam that goes directly to PRM.  There is plenty of space there, so it should be all good.  The tricky bit is that the flat steering mirrors fit into things now while they are piezos, but they will be trickier to fit if we make them into Tip Tilts.  But I have full faith in Koji's amazing optical table layout skills, that he can make it happen. 

Unless there are major objections, I think this is the MMT that we're going to go with. (So speak now or forever hold your peace.)  The angle between tilt and translation isn't quite what we'd like it to be (at ~18deg), but it's not too terrible.  And we still have 99.5% overlap which is very important.

Attachment 1: Awesome_MM_Solution.png
Awesome_MM_Solution.png
  2481   Wed Jan 6 03:44:41 2010 KojiConfigurationIOOElusive Mode Matching Solution found!

I am in the way to get a reasonable optical layout.
Please calculate the final results with the following conditions.

"Result" =
- mode overlapping with astigmatism
- alignment matrix (m/rad, rad/rad) for Pitch and Yaw
- alignment orthogonality
- sensitivity of the mode overlapping to the perturbations
  * histgram
  * individual scan of the optic positions

Optics chain: MC3 - SM1(flat) - MMT1(f=-5m) - MMT2(f=+8m) - SM2(flat) - PRM

Incident angles: SM1 24deg, MMT1 3deg, MMT2 1deg, SM2 44.5deg

Distances:
MC3 HR - SM1: 884mm
SM1 - MMT1: 1058.2mm
MMT1 - MMT2: 1890mm
MMT2 - SM2: 2007.9mm
SM2 - PRM HR: 495.6mm

It has ~200mm deviation from the solution. I can move only MMT1 for final optimization.
Give us the numbers if it can improve the performance.
Note that this move changes SM1-MMT1 and MMT1-MMT2 simultaneously.

Quote:

I think I have finally found a Mode Matching solution for our new Input Mode Matching Telescope!  And after looking at the layout diagram with Koji and Raffaele, it seems like all of the optics will fit into the chambers / onto the tables (not true as of last week). 

3. RoCMMT1 is -5m
   RoCMMT2 is 8m,
   with the MMTs 1.89m apart.
   This is a 1.6x telescope.
   MMT2 is 2.2641m from the PRM
   MMT1 is 2m from MC3.
   The Condition Number for this optical chain is 89219047.5781.

This layout is very similar to the one that Koji posted on the wiki yesterday:  Upgrade09/Optical Layout.  The difference is that I want to move MMT1 ~20cm closer to the MC13 table, so just on the other side of the main red beam that goes directly to PRM.  There is plenty of space there, so it should be all good.  The tricky bit is that the flat steering mirrors fit into things now while they are piezos, but they will be trickier to fit if we make them into Tip Tilts.  But I have full faith in Koji's amazing optical table layout skills, that he can make it happen. 

Unless there are major objections, I think this is the MMT that we're going to go with. (So speak now or forever hold your peace.)  The angle between tilt and translation isn't quite what we'd like it to be (at ~18deg), but it's not too terrible.  And we still have 99.5% overlap which is very important.

 

  2535   Thu Jan 21 10:09:27 2010 KojiSummaryIOOPhotos of the optical tables

I made a wiki page dedicated for the photos of the optical tables.
The current layouts were uploaded.

http://lhocds.ligo-wa.caltech.edu:8000/40m/Optical_Tables

  2559   Tue Feb 2 13:14:09 2010 KojiHowToIOOAnatomy of New Focus Resonant EOM

Joe let me use the resonant EOM for GigE phase camera for a while.
Then, I immediately started to open it :)

it uses the MiniCIrcuits T5-1T transformer and a TOKO RCL variable inductor.

The photos are on the Picasa 40m album.

http://lhocds.ligo-wa.caltech.edu:8000/40m/40m_Pictures

  2584   Tue Feb 9 17:51:48 2010 JenneSummaryIOOInput Mode Matching Telescope design is complete

The upgrade's input mode matching telescope design is complete.  A summary document is on the MMT wiki page, as are the final distances between the optics in the chain from the mode cleaner to the ITMs.  Unless we all failed kindergarden and can't use rulers, we should be able to get very good mode matching overlap.  We seem to be able (in Matlab simulation land) to achieve better than 99.9% overlap even if we are wrong on the optics' placement by ~5mm.  Everything is checked in to the svn, and is ready for output mode matching when we get there.

  2698   Tue Mar 23 00:31:51 2010 KojiUpdateIOOMC realigned

This is the first touch to the MC mirrors after the earthquake on 16th.

  • I made an aluminum access connector so that we can work on the MC even the door is open. We still can be able to open the aluminum tube. The photos are attached. Steve, could you please look it at a glance whether the seal is enough or not.
  • MC resonances were flashing. Align MC2 and MC3 so that we have many TEM00s.
  • Found c1vmesus2 gone mad. Restarted remotely according to the wiki entry. 
  • Reset the MC coil output matrix to 1. (Previously, balance was adjusted so that A2L was minimized.)
  • Excite MC2 Pitch/Yaw at 8 and 9 Hz, looking at the peaks in the MC-MCL output. Move MC2 Pitch/Yaw so that the peak
    is reduced. (*)
  • MC1/MC3 were aligned so that we get the maximum transmission (or minimum reflection). (**)
  • Repeat (*) and (**)

So far, I have aligned in Yaw such that the yaw peak is minimized.

Attachment 1: IMG_2346.jpg
IMG_2346.jpg
Attachment 2: IMG_2347.jpg
IMG_2347.jpg
  2699   Tue Mar 23 09:37:36 2010 steveUpdateIOOvac envelope has to be sealed as antproof for overnight

Quote:

This is the first touch to the MC mirrors after the earthquake on 16th.

  • I made an aluminum access connector so that we can work on the MC even the door is open. We still can be able to open the aluminum tube. The photos are attached. Steve, could you please look it at a glance whether the seal is enough or not.
  • MC resonances were flashing. Align MC2 and MC3 so that we have many TEM00s.
  • Found c1vmesus2 gone mad. Restarted remotely according to the wiki entry. 
  • Reset the MC coil output matrix to 1. (Previously, balance was adjusted so that A2L was minimized.)
  • Excite MC2 Pitch/Yaw at 8 and 9 Hz, looking at the peaks in the MC-MCL output. Move MC2 Pitch/Yaw so that the peak
    is reduced. (*)
  • MC1/MC3 were aligned so that we get the maximum transmission (or minimum reflection). (**)
  • Repeat (*) and (**)

So far, I have aligned in Yaw such that the yaw peak is minimized.

 This seal is good for daily use- operation only. The IFO has to be sealed  with light metal doors every night so ants and other bugs can not find their way in.

Our janitor Kevin is mopping the hole IFO room floor area with 5%  ant killing solution in water in order to discourage bugs getting close to our openings of the vented chamber.

You may be sensitive to this chemical too.  Do not open chamber till after lunch.

Attachment 1: pc3.JPG
pc3.JPG
Attachment 2: pc4.JPG
pc4.JPG
  2700   Tue Mar 23 09:55:20 2010 KojiUpdateIOOvac envelope has to be sealed as antproof for overnight

Roger.

Quote:

 This seal is good for daily use- operation only. The IFO has to be sealed  with light metal doors every night so ants and other bugs can not find their way in.

 

  2705   Wed Mar 24 02:06:24 2010 KojiUpdateIOOvac envelope has to be sealed as antproof for overnight

Matt and Koji:

We closed the light doors of the chambers.

Quote:

Roger.

Quote:

 This seal is good for daily use- operation only. The IFO has to be sealed  with light metal doors every night so ants and other bugs can not find their way in.

 

 

  2827   Wed Apr 21 21:46:53 2010 KojiUpdateIOOMC spot diagnosis

Zach and Koji

We measured uncalibrated angle-to-length coupling using tdssine and tdsdmd.
We made a simple shell script to measure the a2l coupling.

Details:

- Opened the IMC/OMC light door.

- Saw the large misalignment mostly in pitch. Aligned using MC2 and MC3.

- Locked the MC in the low power mode. (script/MC/mcloopson AND MC length gain 0.3->1.0)

- Further aligned MC2/3. We got the transmission of 0.16, reflection of 0.2

- Tried to detect angle-to-length coupling so that we get the diagnosis of the spot positions.

- Tried to use ezcademod. Failed. They seems excite the mirror  but returned NaN.

- We used tdssine and tdsdmd instead. Succeeded.

- We made simple shell script to measure the a2l coupling. It is so far located users/koji/100421/MCspot

- We blocked the beam on the PSL table. We closed the chamber and left.

 

  2852   Tue Apr 27 22:28:58 2010 ZachUpdateIOOMC alignment

Beginning last week, I have been helping Koji with some of the IO work that must be done for the 40m upgrade. The first thing he asked me to do is to help with the alignment of the MC.

As I understand, it became apparent that the IFO beam was not centered on all (or any) of the MC mirrors, which is disadvantageous for obvious reasons. We are trying to correct this, using the following strategy:

  1. Adjust the MC mirrors into rough alignment, isolate a strong TEM00, and lock the cavity
  2. Fine-tune the alignment by minimizing the REFL power when locked (in these first two steps, we adjusted only MC2 & MC3, assuming that the REFL beam was centered on the PD, and wanting to keep it that way). At this point, the cavity is resonating some asymmetric mode, looking something like (not to scale---for illustration only):MC_misaligned.png
  3. Shake all three mirrors (in succession) in pitch and yaw, each time demodulating the error signal at the frequency used for the excitation and recording the magnitude and phase of the response.
  4. Move one mirror's DC orientation, repeat step 3, and then restore the mirror to its original position
  5. Repeat step 4 for both angular degrees of freedom of each mirror

Using the results of these measurements, it is possible to evaluate the components of a block-diagonal matrix M which relates the tilt-to-displacement coupling of each DOF to each mirror's misalignment in that degree, i.e.,

a = M x

with a a 6-dimensional vector containing the coupling of each degree of freedom to the length of the cavity and x a 6-dimensional vector containing the angular misalignments of each. Due to orthogonality of pitch and yaw, M will take the form of a 6x6 matrix with two non-zero 3x3 blocks along the diagonal and zero matrices on the off-diagonal blocks.

The idea is to isolate components of M by moving one mirror at a time, solve for them, then find the inverse M-1 that should give us the required angular adjustments to obtain the beam-centered ideal cavity mode.

In theory, this need only be done once; in practice, our measurement error will compound and M will not be accurate enough to get the beams exactly centered, so we will have to iterate.

NOTE: The fact that we are adjusting the three cavity mirrors to obtain the ideal mode means that we will necessarily tarnish our coupling into the cavity. Once we have adjusted the mirrors once, we will need to re-steer the input beam and center it on the REFL diode. 

Status: This process has been completed once through step 5. I am in the process of trying to construct the matrix for the first adjustment.

 

  2855   Wed Apr 28 12:05:44 2010 ZachUpdateIOOMC alignment

I have worked out the first set of adjustments to make on the MC mirrors (all angle figures are in units of the increments on the control screen)

Using the method described in the previous post, I obtained the following matrix relating the angle-to-length coupling and the angular deviations. In the following matrix, Mij corresponds to the contribution of the jth degree of freedom to the ith A-to-L coupling, with the state vector defined as xi = (MC1P, MC2P, MC3P, MC1Y, MC2Y, MC3Y), where each element is understood as the angular deviation of the specific mirror in the specific direction from the ideal position, such that x = 0 when the cavity eigenmode is the correct one and the beams are centered on the mirrors (thus giving no A-to-L coupling regardless of the components of M).

 

M =

   1.0e+03 *

   -0.2843   -0.4279   -0.1254         0         0         0

   -0.8903   -0.4820   -0.6623         0         0         0

    0.5024    0.0484   -0.0099         0         0         0

         0         0         0         0.1145   -0.1941   -0.3407

         0         0         0         0.0265    1.5601    0.2115

         0         0         0         0.1015    0.1805   -0.0103,

giving an inverse

M-1 =  

    0.0003   -0.0001    0.0020         0         0         0

   -0.0031    0.0006   -0.0007         0         0         0

    0.0018   -0.0018   -0.0022         0         0         0

         0         0         0        -0.0013   -0.0015    0.0117

         0         0         0         0.0005    0.0008   -0.0008

         0         0         0        -0.0037   -0.0010    0.0044

The initial coupling vector is then acted on with this inverse matrix to give an approximate state vector x containing the angular misalignments of each mirror in pitch and yaw. The results are below:

x

   1P:  0.0223

   2P: -0.0733

   3P:  0.3010

  1Y:  -0.1372

  2Y:   0.0194

  3Y:  -0.0681

 

  2856   Wed Apr 28 14:15:58 2010 AlbertoUpdateIOOMC alignment

 

 That's interesting.

Would it be possible to write about the technique on a wiki page as you get measurements and results?

  2858   Wed Apr 28 14:42:55 2010 ZachUpdateIOOMC alignment

Sure. I figured I would put up a How-To if it works. 

Quote:

 

 That's interesting.

Would it be possible to write about the technique on a wiki page as you get measurements and results?

 

  2864   Sun May 2 15:28:25 2010 KojiUpdateIOOSpot Positions of MC1/MC3

Summary

The spot positions on the MC mirrors were measured with coil balance gains.
The estimated spot positions from the center of the MC1 and MC3 are as followings:

MC1H = +0.29 mm
MC1V = -0.43 mm
MC3H = +1.16 mm
MC3V = -0.68 mm

The cordinates are described in the figure

Method

As far as the cavity mirrors are aligned to the incident beam, spots on the MC1 and MC3 tell us the geometry of the incident beam.
Note that spot position on the MC2 is determined by the alignment of the MC1 and MC3, so it does not a big issue now.
The calibration between the coil balance and the spot position are described in the previous entry.

  1. Lock the MC. Align it with MC2/MC3
  2. Run A2L scripts. script/A2L/A2L_MC1 and so on.
    • The scripts run only on the solaris machines. They require "expect" in stalled some specific place which does not exist on the linux machines.
    • Excitation amplitude, excitation freq, readback channels were modified

Result

Beam powers
MC Trans: 0.18
MC Refl: 0.12-0.13

Alignment biases
MC1P 3.2531
MC1Y -1.0827
MC2P 3.4534
MC2Y -1.1747
MC3P -0.9054
MC3Y -3.1393

Coil balances
MC1H 1.02682
MC1V 0.959605
MC3H 0.936519
MC3V 1.10755

(subtract 1, then multiply 10.8mm => spot position.)

Attachment 1: spot_position.png
spot_position.png
  2880   Wed May 5 01:19:05 2010 KojiUpdateIOOMC spot centering cont"d

Koji and Zach

We improved the beam axis rotaion on the MC. We still have 3mrad to be corrected.
So far we lost the MC Trans spot on CCD as the beam is now hitting the flange of the window. We need to move the steering mirror.

To do next:

- MC2 spot is too much off. Adjust it.

- Rotate axis for 3mrad more.

- MC2 spot is too much off. Adjust it.

- Adjust Vertical spot position as a final touch.


Monday

- Incident beam had 7mrad rotation.

- Tried to rotate in-vac steering mirror (IM1) in CCW

- After the long struggle the beam from PSL table started to hit north-east side of IM1 mount.

- Moved the IM1. All of the beam (input beam, MC Trans, MC Refl) got moved. Chaotic.

- Recovered TEM00 resonance. MC Trans CCD image missing. The beam axis rotation was 8.5mrad.
  Even worse. Disappointed.

Tuesday

- We made a strategic plan after some deliberation.

- We returned to the initial alignment of Saturday only for yaw.
  Not at once, such that we don't miss the resonance.

- Adjusted SM2Y and IM1Y to get reasonable resonance. Then adjusted MC2/3 to have TEM00 lock.

- Measured the spot positions. The axis rotation was 4.8mrad.

- Moved the spot on IM1 by 7mm by rotating SM2Y in CCW - ((A) in the figure)

- Compensated the misalignment by IM1Y CCW. ((B) in the figure)
  Used a large sensor card with puch holes to see the spot distribution between the MC1 and MC3.

- Fine alignment by MC2/MC3. Lock to TEM00. The beam axis rotation was 3mrad.The beam axis translation was 3mm.

- This 3mm can be Compensated by IM1Y. But this can easily let the resonance lost.
  Put the sensor card between MC1/MC3 and compensated the misalignment by MC3 and MC1.

Note: You match the returned spot from the MC2 to the incident beam by moving the spot deviation by MC3,
the spot returns to the good position on MC1. But the angle of the returned beam is totally bad.
This angle deviation can be adjusted by MC1.

Note2: This step should be done for max 2mm (2mrad) at once. As 2mrad deviation induces the spot move on the MC2 by an inch.

- After all, what we get is

MC1H = -0.15 mm
MC1V = -0.33 mm
MC3H = +0.97 mm
MC3V = -0.33 mm

This corresponds to the axis rotation of 3mrad and the beam axis translation of 0.8mm (to north).

Attachment 1: steering.png
steering.png
  2884   Thu May 6 01:06:16 2010 KojiUpdateIOOMC spot centering cont'd (Triumph)

Zach and Koji,

We finally aligned the incident beam enough close to the center of the all MC mirrors! Uraaaaah!

MC1H = -0.12mm
MC1V =
-0.13mm
MC2H = -0.15mm

MC2V = +0.14mm
MC3H = -0.14mm
MC3V = -0.11mm

The aperture right before the vacuum window has been adjusted to the beam position. This will  ensure that any misalignment on the PSL table can have the correct angle to the mode cleaner as far as it does resonate to the cavity. (This is effectively true as the small angle change produces the large displacement on the PSL table.)

If we put an aperture at the reflection, it will be perfect.

Now we can remove the MZ setup and realign the beam to the mode cleaner!


 

Method:

- The beam axis rotation has been adjusted by the method that was used yesterday.

Differential: SM2Y and IM1Y

Common: SM2Y only

- We developped scripts to shift the MC2 spot without degrading the alignment.

/cvs/cds/caltech/users/zach/MCalign/MC2_spot/MC2_spot_up
/cvs/cds/caltech/users/zach/MCalign/MC2_spot/MC2_spot_down
/cvs/cds/caltech/users/zach/MCalign/MC2_spot/MC2_spot_left
/cvs/cds/caltech/users/zach/MCalign/MC2_spot/MC2_spot_right

These scripts must be upgraded to the slow servo by the SURF students.

- These are the record of the alignment and the actuator balances

C1:SUS-MC1_PIT_COMM   =  2.4005
C1:SUS-MC1_YAW_COMM   = -4.6246
C1:SUS-MC2_PIT_COMM   =  3.4603
C1:SUS-MC2_YAW_COMM   = -1.302
C1:SUS-MC3_PIT_COMM   = -0.8094
C1:SUS-MC3_YAW_COMM   = -6.7545
C1:SUS-MC1_ULPIT_GAIN =  0.989187
C1:SUS-MC1_ULYAW_GAIN =  0.987766
C1:SUS-MC2_ULPIT_GAIN =  0.985762
C1:SUS-MC2_ULYAW_GAIN =  1.01311
C1:SUS-MC3_ULPIT_GAIN =  0.986771
C1:SUS-MC3_ULYAW_GAIN =  0.990253

  2899   Sat May 8 02:38:08 2010 KojiSummaryIOOMC incident power

As per Steve's request I checked the MC incident power as a function of time.

The output is negative: the lower voltage, the higher power.

Before I put the attenuator the incident power was 1.1W. It appear as -5V.

Now the output is -0.1V. This corresponds to 22mW.

 

Attachment 1: MC_input.png
MC_input.png
  2900   Sat May 8 03:09:15 2010 KojiUpdateIOOSteering around MC

After the MZ-removal work:

- I found that the input steering (IM1) was right handed. This was different from the CAD layout. This was the main reason why the MC trans was kicked by the mount.
- Removed the mount from the post and converted it to a keft handed.
- Align IM1 so that we can get TEM00 lock. Align IM1 further.

- After the IM1 was optimized for the TEM00, move the periscope mirrors to have best alignment.

- Checked the beam spot positions. They looks quite good (MC2 is not the matter now).

C1:SUS-MC1_ULPIT_GAIN = 0.998053
C1:SUS-MC1_ULYAW_GAIN = 0.992942
C1:SUS-MC2_ULPIT_GAIN = 1.00856
C1:SUS-MC2_ULYAW_GAIN = 1.04443
C1:SUS-MC3_ULPIT_GAIN = 0.99868
C1:SUS-MC3_ULYAW_GAIN = 1.00041

  2918   Wed May 12 03:56:54 2010 KojiUpdateIOOFaraday aligned

Zach and Koji

The old small MMT was removed and wrapped by Al foils.

The steering mirror IM2-IM4 were displaced and aligned.

The Faraday isolator block is moved and aligned.

The MC is realigned and resonatng TEM-00.

Now the MC has slightly miscentered beam on the mirrors owing to change of the stack leveling.
OSEMs are also in a strange state. We should check this later.

  2928   Thu May 13 23:59:46 2010 ZachSummaryIOOMC table leveled

 After the recent removal of the old IMMT and the relocation of the Faraday isolator, the MC table was tilted a bit (southward and slightly westward---as of when I opened the chamber this afternoon). I re-leveled it by putting an extra two rectangular ballast blocks on the stack that was already hanging off the NNE edge of the table (there are a total of 4 in the stack now). I also screwed down the circular block that Koji and I put between the Faraday and SM1 on Tuesday, and re-mounted the two wire harness towers onto the table.

Needless to say, this threw the MC way out of alignment. I spent the rest of the afternoon reacquiring alignment and getting it to lock robustly. Here is a summary:

  • I adjusted MC3 until I got the 2nd, 3rd+ pass beams to overlap with the input beam between MC1&3, then I adjusted MC2&1 semi-methodically until I got something flashing at the transmitted end. This took some time.
  • I went back into the control room, engaged the loops and acquired lock on the TEM00 mode, whereupon I found that the beam spot was WAY off center on MC2 (due to my meddling with all the mirrors to get resonance flashes). I began using the MC2_spot_up (etc) scripts we wrote the other day to re-center it.
  • After a few iterations, the lock became weak, and eventually gave out. This is because the REFL beam was falling off the RFPD (and being clipped by the iris on the AP table), so I moved the iris and re-centered the beam on the diode.
  • With that, I was able to get the MC2 spot more or less centered, but then I noticed that---though the lock was clearly strong as evidenced both by the REFL power dip and visually via the camera on MC2---it looked like crap on the CCD. It seemed like there was some higher order mode structure sloshing around on top of the 00 spot, which didn't make any sense, until I realized that it was just a diffraction pattern from the TRANS beam getting clipped somewhere on the way out of the vacuum system.
  • I went back to the AP table, where I noticed that the TRANS beam was hitting near the edges of several of the mirrors on the way back to the PSL table, including the first one out of the viewport, so I turned IM4 to center the beam on this mirror, then proceeded to center the beam on each mirror downstream and then onto the CCD.
  • After getting a clear picture of the transmission on the CCD, centering the spot even better on MC2, then fine-tuning MC2&3 to strengthen the lock, I went back to the MC table to check that the transmitted beam was still passing through the center of the Faraday, which, by none other than an act of God, it was.
  • Having done the necessary work in the tank, I ran the A2L_MC2 script to fine-tune the centering of the spot on MC2. It needed a couple steps up and to the side, but after that the actuator gains for pitch and yaw were both balanced again to within ~2%, which is only slightly above the measurement error. We will probably need to adjust this continually, especially during the upgrade, so I didn't bother with getting it better than that.

After that, I shut off the loops, blocked the beam, and put the light doors back on the tanks. Then I went to the parking lot, then I got in my car, etc, etc, etc.

  2929   Fri May 14 03:30:45 2010 KojiSummaryIOOMC table leveled

Thanks Zach.This was a great job.

It was not mentioned but: was the Faraday clamped down on the table?

 

  2931   Fri May 14 10:33:01 2010 ZachSummaryIOOMC table leveled

Ah... no, I didn't. That explains why there were loose dogclamps on the table. I wrapped them in foil and put them on the clean cart. Can this wait until the next time we open the tank (i.e. to measure the beam profile), or should I go over there and clamp it down today?

 

  2935   Sat May 15 04:13:33 2010 KojiSummaryIOOMC table leveled

Fixing at the next time is absolutely OK.

Quote:

Ah... no, I didn't. That explains why there were loose dogclamps on the table. I wrapped them in foil and put them on the clean cart. Can this wait until the next time we open the tank (i.e. to measure the beam profile), or should I go over there and clamp it down today? 

 

  2938   Mon May 17 02:10:10 2010 KojiConfigurationIOOHow to lock / align the MC

Let me remind you how to lock and align the IMC

To lock

1. Open the doors for the IMC/OMC chambers. Open the manual shutter of the PSL just in front of the optical window

2. Run scripts/MC/mcloopson

3. Set the MC length path gain 0.3 / Set the MC total gain "+20"

4. If you want to avoid excitation of the mirrors by air turbulence, put a big plastic film and put three posters on the top and both the sides on the floor to block the wind go into the chamber.

To shut down

1. Run scripts/MC/mcloopsoff

2. Close the manual shutter, Remove the wind blockers, and the light door of the chambers

To align the MC

1. Tweak MC2 and MC3 to get maximum transmittion and/or minimum reflection.

  2941   Mon May 17 19:42:11 2010 JenneUpdateIOOFirst steps toward MC mode measuring

[Jenne, Kevin, Steve]

We made some progress toward getting the MC's beam profile measured.  In the end, no changes were made to anything today, but we're more prepared to go for tomorrow.

What we did:

* Grabbed the scanning slit beam scan from the PSL lab.  It's the same kind as we had here at the 40m, so Kevin was able to hook it up to the computer, and confirmed that it works.

* Opened the IOO and OMC chamber doors, and locked the MC.  Unfortunately the MC mode was awful in Yaw.  Awful like TEM(0,10+). But it still locked.  

* Confirmed that the beam went through the Faraday.  I looked at the beam before and after the Faraday on a card, and it was the same nasty beam both before and after.  So it looks like Zach did a good job aligning the Faraday and everything else.  I was going to clamp the Faraday, but I didn't yet, since I wanted to see the nice happy TEM00 mode go through without clipping before risking moving the Faraday during clamping (I don't know how heavy it is, so I'm not sure how much it might potentially move during clamping.)

* Noticed that there is a whole lot of crap on both the OMC and BS tables that's going to have to move.  In particular, one of the weights leveling the OMC table is right where I need to put MMT2.  Steve suggested putting the optic there, in its approximate place, before doing too much other stuff, since it could potentially affect the leveling of the table, and thus the input pointing to the MC.  Unfortunately, to do that I'll need to move the weight, which is definitely going to change things.  Sad face.  Moving the weight will likely be one of the first things I do tomorrow, so that all 3 profile measurements have the same configuration. 

* Before closing up, I tried to align the MC, to get back to TEM00, to no avail.  I got as far as achieving TEM11 flashing, along with a bunch of other crappy modes, but didn't get 00.  That's also on the to-do list.

What we're going to do:

* Open the chambers, and align the MC to TEM00 (using the sliders on the MC align screen).

* Check with an IR card that the beam goes through the Faraday.

* Clamp the Faraday, reconfirm.

* Remove the weight on the OMC table.

* Place MMT2 on the OMC table in it's approximate final location.

* Realign the MC, and make sure the beam goes through the Faraday.  If this doesn't happen smoothly, I may need more instruction since I've never dealt with aligning the Faraday before.  What are the appropriate mirrors to adjust? 

* Move the PZT flat steering mirror from the BS table to the IOO table.  (Thoughts on this?  This will change the table leveling, and also includes the trickiness of needing to move the connectors for the PZT.)

* Place a flat mirror on the BS table to route the MC beam out to the BS/PRM/SRM oplev table. 

* Measure the mode using the beam scan: on the BS oplev table, on the POX table, and then perhaps by shooting the beam through the beamtube on the ETMY (new convention) table.

* Place MMT1 on the BS table, use flat mirrors to get it out of the chambers, repeat measurements.

* Place MMT2 in the correct position, use flat mirrors to get it out of the chambers, repeat measurements.

All of this may require some serious cleaning-up of the BS table, which is going to be ugly, but it has to happen sometime. Hopefully I can get away with only moving a minimal number of things, in order to get these measurements done.

 

Another note: Don't trust the PSL shutter and the switch on the MEDM screens! Always use a manual block in addition!!! We discovered upon closeup that hitting the "Closed" button, while it reads back as if the shutter is closed (with the red box around the buttons), does not in fact close the shutter.  The shutter is still wide open.  This must be fixed.

  2942   Tue May 18 01:40:56 2010 KojiUpdateIOOFirst steps toward MC mode measuring

OK. Don't worry. This is just an initial confusion which we also had for the suspensions a while ago.

The faraday must be clamped. It shakes the table terribly but it is fine. The leveling may change a bit but should be small enough. Otherwise, just tweak the weights. In fact, the faraday has enough large apertures and we hope we don't need to move it again, as far as the MC incident beam is not moved. But if necessary, we don't move the mirrors but move the faraday itself.

Usually the alignment of the MC is taken by MC2/MC3 such that we don't  move the refl. But if you think what have moved is the MC1/MC3 (i.e. activity in the IMC chamber), take the alignment of the MC1/MC3.

It is just a matter of time to get TEM00. If you get TEM11, it is already close. If you align for TEM11, it is enough aligned to lock TEM10 or TEM01. Once you got better mode, align for it again. Eventually you will get TEM00.

The leveling may change by moving the optics and the weight again. But once the leveling is recovered by arranging the weights somewhere else,
the pointing must be fine again. If necessary, You can remove two optics for squeezing injection (strange motorized rotating mirror and a mount sticking out from the table to south.)

Yes, we need to move the PZT mirror. For the connection, only Steve can give us the right way to do it. If it is too much hussle, just move only the mirror and ignore the wiring for now.

I will update how the mirrors should be migrated from the table to the table.

 

  2945   Tue May 18 12:04:13 2010 robUpdateIOOFirst steps toward MC mode measuring

Quote:

Another note: Don't trust the PSL shutter and the switch on the MEDM screens! Always use a manual block in addition!!! We discovered upon closeup that hitting the "Closed" button, while it reads back as if the shutter is closed (with the red box around the buttons), does not in fact close the shutter.  The shutter is still wide open.  This must be fixed.

 Has anyone tried pushing the "reset" button on the Uniblitz driver?

  2949   Tue May 18 16:44:35 2010 KojiUpdateIOOFirst steps toward MC mode measuring

Here is the upadted list http://lhocds.ligo-wa.caltech.edu:8000/40m/Upgrade_09/Optics

Quote:

I will update how the mirrors should be migrated from the table to the table. 

 

ELOG V3.1.3-