ID |
Date |
Author |
Type |
Category |
Subject |
7999
|
Tue Feb 5 09:08:11 2013 |
Steve | Update | PEM | exterier doors to be painted |
Quote: |
The wood exteier walls, gutters and doors were painted at CES-Annex building #69
|
The east and south end of the 40m emergency exit doors are sealed- tapped off temporarily. They will be painted on the out side only. This job will be done by tomorrow noon
Do not open chamber if you smell the paint ! |
Attachment 1: IMG_0055.JPG
|
|
8000
|
Tue Feb 5 10:09:08 2013 |
yuta | Summary | General | rough analysis of aligned PRM-PR2 mode scan | stdev of [0.764, 0.751] is 0.007, but what we need is the error of the averaged number. Statistical error of the averaged number is stdev/sqrt(n).
Quote: |
0.764 and 0.751 do not give us the stdev of 0.005.
|
|
8001
|
Tue Feb 5 10:18:54 2013 |
Steve | Update | PEM | high particle count ALART |
Quote: |
Quote: |
The BS camber is open only. We should close ASAP
Outside air quality is 1.7- 2.2 million particles / cf min of 0.5 micron
|
Air is still bad and the chambers are closed. Before lunch Jamie repointed the PRM oplev. Manasa and I reset oplevs: BS and ITMX.
ETMX and ETMY are fine.
SRM and ITMY oplevs needs more work.
|
The bad outside air quality is pushing up the inside counts.
The outside air is 5 million counts / cf min for 0.3 micron and 2 million counts / cf min for 0.5 micron particles
Do not open chamber over 10,000 counts / cf min of 0.5 micron |
Attachment 1: badair.png
|
|
Attachment 2: 100d_airq.png
|
|
8002
|
Tue Feb 5 11:30:19 2013 |
Koji | Summary | General | rough analysis of aligned PRM-PR2 mode scan | Makes sense. I mixed up n and n-1
Probability function: X = (x1 + x2 + ... + xn)/n, where xi = xavg +/- dx
Xavg = xavg*n/n = xavg
dXavg^2 = n*dx^2/n^2
=> dXavg = dx/sqrt(n)
Xavg +/- dXavg = xavg +/- dx/sqrt(n) |
8003
|
Tue Feb 5 12:08:43 2013 |
Max Horton | Update | Summary Pages | Updating summary pages | Getting started: Worked on understanding the functionality of summary_page.py. The problem with the code is that it was written in one 8000 line python script, with sparse documentation. This makes it difficult to understand and tedious to edit, because it's hard to tell what the precise order of execution is without tracing through the code line by line. In other words, it's difficult to get an overview of what the code generally does, without literally reading all of it. I commented several functions / added docstrings to improve clarity and start fixing this problem.
Crontab: I believe I may have discovered the cause of the 6PM stop on data processing. I am told that the script that runs the summary_pages.py is called every 6 hours. I believe that at midnight, the script is processing the next day's data (which is essentially empty) and thus not updating the data from 6PM to midnight for any of the days.
Git: Finally, created git repository called public_html/__max_40m-summary_testing to use for testing the functionality of my changes to the code (without risking crashing the summary_pages). |
8004
|
Tue Feb 5 15:31:03 2013 |
Steve | Update | General | clean assembly room benches cleaned up | Manasa, Jamie and Steve,
Tip-Tilts and parts moved into the most north " 40m " cabinet in the assembly room.
Green-black glass and related components were moved to the 40m E0 cabinet in plastic boxes.
The north flow bench has a few items that belong to us: HE/Ne laser, qpd on translation stages, an iris and one red mirror. These were moved to the north edge of this bench.
However this leveled table is still full with other people's stuff |
Attachment 1: IMG_0057.JPG
|
|
Attachment 2: IMG_0061.JPG
|
|
8005
|
Tue Feb 5 19:16:22 2013 |
Jamie | Summary | General | arbcav of PRC with +600 RoC PR2/3 | This is just a simple rerun of arbcav from #7995 but with the PR2/3 RoCs set to 600, instead of -600. Overall g-factor = 0.922, and the modes are well separated:
This doesn't take into account the effect of traveling through the substrates (still working on it). It assumes the PR2/3 have been moved such that the cavity fold lengths remain the same.
This is something that we need to keep in mind: we will need to adjust the position of the PR2/3 to keep the fold lengths the same. |
8006
|
Tue Feb 5 19:32:47 2013 |
yuta | Summary | General | PR2/PR3 flipping and PRC stability | We are considering of flipping PR2 and/or PR3 to make PRMI stable because PR2/PR3 seems to be convex.
I calculated dependency of the PRC stability on the PR2/PR3 curvature when PR2/PR3 flipped and not flipped.
Flipping looks OK, from the stability point of view.
Assumption:
PRM-PR2 distance = 1.91 m
PR2-PR3 distance = 2.33 m
PR3-ITM distance = 2.54 m
PRM RoC = +122.1 m
ITM RoC = Inf
theta_inc PRM = 0 deg
theta_inc PR2 = 1.5 deg
theta_inc PR3 = 41 deg (all numbers from elog #7989)
Here, RoC means RoC measured from HR side. RoC measured from AR side will be -n_sub*RoC, assuming flat AR surface.
I also assumed mirror thickness to be negligible.
Method:
1. I used Zach's arbcav and modified it so that it only tells you your cavity is stable or not.
(It lives in /users/yuta/scripts/mode_density_PRC/stableornot.m )
2. Swept PR2/PR3 RoC (1/RoC from -0.005 to 0.005 1/m) to see the stability condition.
Results:
1. Stability condition of the PRMI when PR2 and PR3 is not flipped is depicted in the graph below. Black region is the unstable region. We all know that current PRMI is unstable, so we are in the black region.

2. Stability conditions of PRMI with one of the PR2/PR3 flipped are depicted in the graphs below. If we flip one of them, PRMI will likely to be stable, but if the flipped one is close to flat and the RoC of the other one is >~ -250 m (more convex than -250 m), PRMI will remain unstable.
 
3. Stability condition of PRMI with both PR2 and PR3 flipped is depicted in the graph below. If we flip both, PRMI will be stable.

Discussion:
1. Flipping one of PR2/PR3 seems OK, but I cannot guarantee. TMS measurement insists RoC of PR2 to be ~ -190 m, if we believe PRM RoC = +122.1 m (elog #7997). We need more precise measurement if we need to be sure before flipping. I prefer PR2 flipping because PR3 flipping gives us longer path in the substrate and larger astigmatism. Also, PR3 RoC is phase-map-measured to be ~ -600 m and PR2 RoC seems to be more convex than -600 m from the TMS measurement.
2. Flipping both is good from stability point of view. We need calculation of the loss in the PRC (and mode-mismatch to the arms). Are there any requirements?
3. If we are going to flip PR3, are there any possibilities of clipping the beam at PR3? We need to check.
4. I need to calculate whether mirror thickness and AR surface curvature are negligible or not.
Conclusion:
I want to flip only PR2 and lock PRMI.
By the way:
I don't like matlab plots. |
8007
|
Wed Feb 6 11:59:12 2013 |
Jenne | Update | Locking | PRC cavity gains | EDIT: These numbers are for a perfect, non-lossy arm cavity. So, a half real, half imaginary world.
Carrier uses arm cavity reflectivity for perfectly resonant case.
PRC carrier gain, flipped PR2, PR3 = 61
PRC carrier gain, regular PR2, PR3 = 68 (same value, within errors, for no folding at all).
Carrier gain loss = (68-61)/68 = 10%
SB uses arm cavity reflectivity for perfectly anti-resonant case.
PRC SB gain, flipped PR2, PR3 = 21
PRC SB gain, regular PR2, PR3 = 22 (same value, within errors, for no folding at all). <--- yes, this this "regular PR2, PR3 = 22..."
SB % gain loss = (22-21)/22 = 4.5%
I claim that we will be fine, recycling gain-wise, if we flip the folding mirrors. If we do as Yuta suggests and flip only one folding mirror, we'll fall somewhere in the middle. |
8008
|
Wed Feb 6 14:51:25 2013 |
Jenne | Update | Electronics | 1 power supply replaced! |
Quote: |
Currently, DC power for amplifiers ZHL-1000LN+ is supplied by Aligent E3620A.
I tried to use power supply from the side of 1X1 rack, but fuse plug(Phoenix Contact ST-SI-UK-4) showed red LED, so I couldn't use it.
|
Yuta, Jenne
We fixed things so that we are now using regular fused rack power for these amplifiers. The fuse no longer had a red LED, but it measured open when we checked the resistance. Although, somehow (magic?) 13.73V were getting to the other side of the fuse.
Anyhow, replacing the fuse with a new one fixed the problem right up. |
8009
|
Wed Feb 6 15:05:18 2013 |
Steve | Update | SAFETY | cameras must be anchord | Cameras must be immediately anchord to avoid a possible collusion with the view port ! |
Attachment 1: IMG_0070_1.JPG
|
|
Attachment 2: IMG_0069_1.JPG
|
|
8010
|
Wed Feb 6 15:10:22 2013 |
Steve | Update | PEM | new safety signs on exterier doors |
Quote: |
Quote: |
The wood exteier walls, gutters and doors were painted at CES-Annex building #69
|
The east and south end of the 40m emergency exit doors are sealed- tapped off temporarily. They will be painted on the out side only. This job will be done by tomorrow noon
Do not open chamber if you smell the paint !
|
The east and south end of the 40m emergency exit doors received new safety signs. |
Attachment 1: IMG_0068.JPG
|
|
8011
|
Wed Feb 6 15:11:21 2013 |
Jenne | Update | Electronics | "Temporary" power supply situation | [Jenne, Yuta, Rana, Steve, Manasa]
We have taken stock of the lab "temporary" power supply situation, and things look much better.
This morning, I removed 2 unused power supplies and a function generator from the PSL table - these had been used for MC ringdown things.
I also removed the non-connected cables that had been used for the RAMMON setup, and the EOM temperature controller circuit.
This afternoon, Yuta removed the 2 HV power supplies that were used to keep PZT2 working near the end of its life. Since we now have the active TTs, these have been turned off for a while, and just needed to be removed.
Manasa removed the power supply under the POP/POX table that was powering the amplifier for POP22. If we are going to continue using a Thorlabs PD for POP22, then we need to make a twisted pair of wires (~20 feet) to get power from 1X1. If we are going to (finally) install a gold RF PD, then that will not be necessary.
I removed the power supply sitting near the bottom of the LSC rack, for another amplifier for POP22 (with minicircuits filters attached). Again, if we get a gold RF PD, we can remove the filters and amplifier. If not, we can use the existing twisted pair of wires, and plug them into the rack rather than a power supply.
The power supply under the NE corner of the PSL table was no longer in use. It was most recently used for amplifiers for the green beat PDs, as Yuta mentioned in elog 6862, those were moved over to 1X2. In elog 8008 I mentioned that Yuta and I moved those amplifiers over to rack power.
The HV supply, the function generator and the OSA controllers that were on top of the short OMC rack next to the AS table have all been removed. We need to come up with a better place for the OSAs, since we need to re-install them. The power supply and the function generator (which was used just for a voltage offset) were formerly used for the output steering PZTs, but lately we have just been using those mirrors as fixed mirrors, since we don't need to steer into the OMC. Some day, we will replace those mirrors with the output steering active tip tilts, and re-commission the OMC....someday.
The power supply for the amplifier set (that goes with the set of minicircuits filters) for the RAMMON PD (which took light from the IPPOS path) has been removed. If we determine that we need RAMMON back, we will have to make a twisted pair to power those amplifiers.
SUMMARY:
* If we don't install a gold RF PD for POP22, we need a 20ft twisted pair for +15/GND.
* Also, if we don't install a gold RF PD for POP22, we need to plug the amplifier at the LSC rack into the rack power (twisted pair already exists).
* If we need RAMMON back, we will need a twisted pair to power those amplifiers.
* All other power supplies have been removed, and put away. We currently have 0 "temporary" power supplies in use in the lab! |
8012
|
Wed Feb 6 15:20:55 2013 |
yuta | Summary | General | FWHM was wrong | I have to blame Jamie for putting extra 2 randomly.
Measured PRM-PR2 cavity finesse was actually 108 +/- 3 (even if you use digital system to get data).
Lorentzian fit:
Lorentzian function is;
f(x;x0,gamma,A) = A * gamma**2/((x-x0)**2+gamma**2)
where x0 is the location of the peak, gamma is HWHM, and A is the peak height.
Lorentzian fitting function in my original code (/users/yuta/scripts/modescanresults/analyzemodescan.py) was
fitFunc = lambda p,x,m: (m-p[2])*p[0]**4/(4*(x-p[1])**2+p[0]**4)+p[2]
In this function, p[0] is sqrt(FWHM), not sqrt(HWHM). I doubled gamma to make it FWHM and squared it because they should be positive.
During Jamie's modification of my code, he doubled p[0]**2 to get FWHM, which is wrong (/users/jrollins/modescan/modescan.py).
I should have commented that p[0] is sqrt(FWHM).
Redoing the analysis:
1. I pulled 2 out, and modified Jamie's modescan.py so that you can name each peak with peakdistinguish=True option. I also modified fitpeak function so that it throws away "peaks" which don't look like a peak.
2. If you run /users/yuta/PRCmodescan/run.py and name each peak, you will get peaks.csv which includes peak position, FWHM, and the type of the peak;
0.065017,0.001458,l
0.070446,0.001463,3
0.075940,0.001509,2
0.081552,0.001526,1
0.087273,0.001565,0
0.112027,0.001911,u
0.278660,0.002211,u
0.306486,0.001658,0
0.312480,0.001576,1
0.313626,2.507910,
0.318486,0.001626,2
0.319730,2.633097,
0.324801,0.001739,3
0.331848,0.001922,l
0.527509,0.001603,l
0.533231,0.001445,3
0.538648,0.001488,2
0.544081,0.001455,1
0.549517,0.001498,0
0.551725,2.422759,
0.570972,0.001346,u
3. /users/yuta/PRCmodescan/calcmodescanresults.py reads peaks.csv and tells you the results;
Time between TEM00 and sideband 0.0239435 pm 0.00115999887452 sec
Calibration factor is 462.167602898 pm 22.3907907867 MHz/sec
FSR is 78.4797010471 MHz
FWHM is 0.729828720682 pm 0.0174145743828 MHz
TMS is 2.64718671684 pm 0.0538858477824 MHz
Finesse is 107.53166986 pm 2.5658325169
Cavity g-factor is 0.994390582331 pm 0.000228155661075
Cavity g-factor is 0.988812630228 pm 0.000453751681357 (Edited by YM; see elog #8056)
RoC of PR2 is -187.384503001 pm 4.26100999578 m (assuming PRM RoC= 122.1 m)
RoC of PRM is 217.915890722 pm 5.65451518991 m (assuming PR2 RoC= -600 m) |
8013
|
Wed Feb 6 15:39:19 2013 |
Steve | Update | Electronics | DC power supplies in cabinets | East arm cabinet E9 and E10
|
Attachment 1: IMG_0066_1.JPG
|
|
8014
|
Wed Feb 6 18:39:08 2013 |
Jenne | Update | Locking | PRC cavity gains | [Yuta, Jenne]
We have both calculated, and agree on the numbers for, the PRC gain for carrier and sideband.
We are using the measured arm cavity (power) loss of 150ppm....see elog 5359.
We get a PRC gain for the CARRIER (non-flipped folding) of 21, and PRC gain (flipped folding) of 20. This is a 4.7% loss of carrier buildup.
We get a PRC gain for the SIDEBANDS (non-flipped folding) of 69, and PRC gain (flipped folding) of 62. This is an 8.8% loss of sideband buildup.
The only difference between the "flipped" and "non-flipped" cases are the L_PR# values - for "non-flipped", I assume no loss of PR2 or PR3, but for the "flipped" case, I assume 1500ppm, as in Rana's email. Also, all of these cases assume perfect mode matching. We should see what the effect of poor mode matching is, once Jamie finishes up his calculation.
Why, one might ask, are we getting cavity buildup of ~20, when Kiwamu always quoted ~40? Good question! The answer seems, as far as Yuta and I can tell, to be that Kiwamu was always using the reflectivity of the ITM, not the reflectivity of the arm cavity. The other alternative that makes the math work out is that he's assuming a loss of 25ppm, which we have never measured our arms to be so good.
For those interested in making sure we haven't done anything dumb:
ppm = 1e-6;
% || | | || ||
% PRM PR2 PR3 ITM ETM
T_PRM = 0.05637;
t_PRM = sqrt(T_PRM);
L_PRM = 0 *ppm;
R_PRM = 1 - T_PRM - L_PRM;
r_PRM = sqrt(R_PRM);
T_PR2 = 20 *ppm;
t_PR2 = sqrt(T_PR2);
L_PR2 = 1500 *ppm;
R_PR2 = 1 - T_PR2 - L_PR2;
r_PR2 = sqrt(R_PR2);
T_PR3 = 47 *ppm;
t_PR3 = sqrt(T_PR3);
L_PR3 = 1500 *ppm;
R_PR3 = 1 - T_PR3 - L_PR3;
r_PR3 = sqrt(R_PR3);
T_ITM = 0.01384;
t_ITM = sqrt(T_ITM);
L_ITM = 0;%100 *ppm;
R_ITM = 1 - T_ITM - L_ITM;
r_ITM = sqrt(R_ITM);
T_ETM = 15 *ppm;
t_ETM = sqrt(T_ETM);
L_ETM = 0 *ppm;
R_ETM = 1 - T_ETM - L_ETM;
r_ETM = sqrt(R_ETM);
rtl = 150*ppm; % measured POWER round trip loss of arm cavities.
rtl = rtl/2; % because we need the sqrt of the exp() for ampl loss....see Siegman pg414.
eIkx_r = exp(-1i*2*pi);
r_cav_res = -r_ITM + (t_ITM^2 * r_ETM * eIkx_r * exp(-rtl)) / (1 - r_ITM*r_ETM * eIkx_r * exp(-rtl) );
eIkx_ar = exp(-1i*pi);
r_cav_antires = -r_ITM + (t_ITM^2 * r_ETM * eIkx_ar * exp(-rtl)) / (1 - r_ITM*r_ETM * eIkx_ar * exp(-rtl) );
%% PRC buildup gain
g_antires = t_PRM*eIkx_ar / (1-r_PRM*r_PR2*r_PR3*r_cav_antires*eIkx_ar);
G_ar = g_antires^2;
G_ar = abs(G_ar) % Just to get rid of the imag part that matlab is keeping around.
g_res = t_PRM*eIkx_r / (1-r_PRM*r_PR2*r_PR3*r_cav_res*eIkx_r);
G_r = g_res^2;
G_r = abs(G_r)
|
8015
|
Wed Feb 6 19:59:35 2013 |
rana | Update | Locking | PRC cavity gains | Getting closer, but need to use the real measured AR reflectivity values, not the 1500 ppm guess. These should be measured at the correct angles and pol, using an NPRO. |
8016
|
Wed Feb 6 20:00:06 2013 |
Manasa | Update | Electronics | BNC cables piled up at every corner | [Yuta, Steve, Manasa]
There are cables piled up around the access connector area which have been victims of stampedes all the time. I have heard these cables were somehow Den's responsibility.
Now that he is not around here:
I found piled up bnc's open at one end and with no labels lying on the floor near the access connector and PSL area. Yuta, Steve and I tried to trace them and found them connected to data channels. We could not totally get rid of the pile even after almost an hour of struggle, but we tied them together and put them away on the other side of the arm where we rarely walk.
There are more piles around the access connector...we should have a next cleanup session and get rid of these orphaned cables or atleast move them to where they will not be walked on.

|
8017
|
Wed Feb 6 20:03:50 2013 |
Manasa | Update | Locking | PRC cavity gains |
Quote: |
Getting closer, but need to use the real measured AR reflectivity values, not the 1500 ppm guess. These should be measured at the correct angles and pol, using an NPRO.
|
I'm still on that! |
8018
|
Wed Feb 6 20:19:52 2013 |
Manasa | Update | Optics | G&H and LaserOptik mirrors | [Koji, Manasa]
We measured the wedge angle of the G&H and LaserOptik mirrors at the OMC lab using an autocollimator and rotation stage.
The wedge angles:
G&H : 18 arc seconds (rough measurement)
LaserOptik : 1.887 deg |
8019
|
Wed Feb 6 22:39:23 2013 |
Jamie | Update | General | PRC/arm mode matching with flipped PR2/PR3: coming soon | I intended to post a long analysis of the PRC/arm mode matching for the various TT situations using Nic's a la mode mode matching program, but I seem to have encountered what I think might be a bug. I'll talk to Nic about it first thing in the AM. Once the issue is resolved I should be able to post the full analysis fairly quickly. Sorry about the delay. |
8020
|
Thu Feb 7 09:03:54 2013 |
Manasa | Update | General | Store optics in respective cabinets |
@Yuta
The ITMX table has been left open since yesterday. I am disconnecting your oscilloscope and closing the table.
To whomsoever it may concern...
I found about half a dozen new cvi optics (beam splitters, waveplates and lenses) lying around on the SP table.
Please store optics back in their respective cabinets if you are not using them immediately. Somebody might be looking around to use them.
|
8021
|
Thu Feb 7 10:35:35 2013 |
yuta | Update | General | Store optics in respective cabinets | I'm not the one who opened the ITMX table yesterday, but thanks for reminding me.
I put POP DC oscilloscope and its cables back.
Also, I relocked PMC and MC. It was unlocked since last night. |
8022
|
Thu Feb 7 12:56:18 2013 |
Jamie | Summary | General | PRC/arm mode matching calculations | NOTE: There was a small bug in my initial calculation. The plots and numbers have been updated with the fixed values. The conclusion remains the same.
Using Nic's a la mode mode matching program, I've calculated the PRC mode and g-parameter for various PR2/3 scenarios. I then looked at the overlap of the resultant PRC eigenmodes with the ARM eigenmode. Here are the results:
NOTE: each optical element below (PR2, ITM, etc.) is represented by a compound M matrix. The z axis in these plots is actually just the free space propagation between the elements, not the overall optical path length.
ARM
This is the ARM mode I used for all comparisons:
 
|
tangential |
sagittal |
gouy shift, one-way |
55.63 |
55.63 |
g (from gouy) |
0.303 |
0.303 |
g (product of individual mirror g) |
0.303 |
0.303 |
PRC, nominal design (flat PR2/3)
This is the nominal "as designed" PRC, with flat PR2/3 folding mirrors.
 
|
tangential |
sagittal |
gouy shift, one-way |
14.05 |
14.05 |
g (from gouy) |
0.941 |
0.941 |
g (product of individual mirror g) |
0.942 |
0.942 |
ARM mode matching: 0.9998
PRC, both PR2/3 flipped
This assumes both PR2 and PR3 have a RoC of -600 when not flipped, and includes the affect of propagation through the substrates.
 
|
tangential |
sagittal |
gouy shift, one-way |
19.76 |
18.45 |
g (from gouy) |
0.886 |
0.900 |
g (product of individual mirror g) |
0.888 |
0.902 |
ARM mode matching: 0.9806
PRC, only PR2 flipped
In this case we only flip PR2 and leave PR3 with it's bad -600 RoC:
 
|
tangential |
sagittal |
gouy shift, one-way |
18.37 |
18.31 |
g (from gouy) |
0.901 |
0.901 |
g (product of individual mirror g) |
0.903 |
0.903 |
ARM mode matching: 0.9859
Discussion
I left out the current situation (PR2/3 with -600 RoC) and the case where only PR3 is flipped, since those are both unstable according to a la mode.
I guess the main take away is that we get slightly better PRC stability and mode matching to the arms by only flipping PR2. |
8023
|
Thu Feb 7 14:10:25 2013 |
Manasa | Update | Optics | LaserOptik - AR Reflectivity - Bad data | Reflectivity of AR surface of LaserOptik (SN6)

The first step measurements of R for AR surface. I am not convinced with the data....because the power meter is a lame detector for this measurement.
I'm repeating the measurements again with PDs. But below is the log R plot for AR surface.
R percentage
6000ppm @ 42 deg
3560ppm @ 44 deg
7880ppm @ 46 deg
4690ppm @ 48 deg

|
8024
|
Thu Feb 7 15:46:42 2013 |
Jenne | Update | Locking | PR2 flipped | More correctly, a different G&H mirror (which we have a phase map for) was put into the PR2 TT, backwards.
Order of operations:
* Retrieve flat test G&H from BS chamber. Put 4th dog clamp back on BS optic's base.
* Remove flat G&H from the DLC mount, put the original BS that was in that mount back. Notes: That BS had been stored in the G&H's clean optic box. The DLC mount is engraved with the info for that BS, so it makes sense to put it back. The DLC mount with BS is now back in a clean storage box.
* Remove PR2 TT from ITMX chamber.
* Remove suspension mounting block from TT frame, lay out flat, magnets up, on lint-free cloth on top of foil.
* Remove former PR2 G&H optic.
* Put what was the flat G&H test optic into the PR2 optic holder, with AR surface at the front.
* Put PR2 suspension block back onto TT frame.
* Put PR2 assembly back in the chamber, solidly against the placement reference blocks that Evan put in last Thursday.
* Close up, clean up, put labels on all the boxes so we know what optic is where.
Why the switcho-changeo? We have a phase map for the G&H that is the new PR2, and a measured RoC of -706m, surface rms of 8.7nm. Now, we can measure the former PR2 and see how it compares to our estimate of the RoC from the cavity measurements we've taken recently. |
8025
|
Thu Feb 7 17:10:11 2013 |
Koji | Summary | General | PRC/arm mode matching calculations |
Quote: |
I left out the current situation (PR2/3 with -600 RoC) and the case where only PR3 is flipped, since those are both unstable according to a la mode.
|
This surprises me. I am curious to know the reason why we can't make the cavity stable by flipping the PR3 as PR3 was supposed to have more lensing effect than PR2 according to my statement. |
8026
|
Thu Feb 7 17:24:13 2013 |
Steve | Update | SAFETY | fire extinguishers checked | The fire department weighted and pressure checked our units today. Surprisingly they found one powder filled can. We can only use HALON gas in the lab.
|
8027
|
Thu Feb 7 19:24:57 2013 |
Riju | Update | | Photodiode transimpedance | Summary: Measurement and plot of shot-noise-intercept-current for MC REFL PD.
Motivation:It is to measure the shot noise intercept current for MC REFL PD.
Result: The final plot is attached here. The plot suggests that the value of shot-noise-intercept current is 1.9mA
Discussion:
The plot is for the measured data of Noise voltage (V/sqrt(Hz)) vs DCcurrent(A). The fitted plot to this measured data follows the noise equation
Vnoise = gdet* sqrt[ 2e (iDC+idet)] , where gdet= transimpedance of the PD in RF region ~600
To get an approximate idea of the shot noise intercept current, we may follow the same procedure described in 7946
In the present case minimum noise value is 1.46e-08 V/sqrt(Hz)
Therefore dark current(in2) ~dark noise voltage/RF transimpedance ~25pA/sqrt(Hz)
Therefore the approximate shot noise intercept current value is (25/18)^2 ~ 1.92mA, which matches well to the fitted value.
|
Attachment 1: reflshotnoise.pdf
|
|
8028
|
Thu Feb 7 19:25:22 2013 |
yuta | Update | CDS | C1ALS filters reloaded | Filters for C1ALS were all gone. So, I copied /opt/rtcds/caltech/c1/chans/C1GCV.txt and renamed it as C1ALS.txt.
I also fixed links in the medm screens; C1ALS.adl and C1ALS_COMPACT.adl.
I'm not sure what happened to C1SC{X,Y} screens.
Quote: |
I decided to rename the c1gcv model to be c1als. This is in an ongoing effort to rename all the ALS stuff as ALS, and get rid of the various GC{V,X,Y} named stuff.
(...snip...)
The above has been done. Still todo:
- FIX SCRIPTS! There are almost certainly scripts that point to GC{V,X,Y} channels. Those will have to be fixed as we come across them.
- Fix the c1sc{x,y}/master/C1SC{X,Y}_GC{X,Y}_SLOW.adl screens. I need to figure out a more consistent place for those screens.
- Fix the C1ALS_COMPACT screen
- ???
|
|
8029
|
Fri Feb 8 00:23:33 2013 |
rana | Summary | General | PRC/arm mode matching calculations |
I would guess that either flipping PR2 or PR3 would give nearly the same effect (g = 0.9) and that flipping both makes it even more stable (smaller g). But what we really need is to see the cavity scan / HOM resonance plot to compare the cases.
The difference of 0.5% in mode-matching is not a strong motivation to make a choice, but sensitivity to accidental HOM resonance of either the carrier or f1 or f2 sidebands would be. Should also check for 2*f2 and 2*f1 resonances since our modulation depth may be as high as 0.3. Accidental 2f resonance may disturb the 3f error signals. |
8030
|
Fri Feb 8 02:12:14 2013 |
Jenne | Update | Locking | PRMI work | [Jenne, Yuta]
Lots of work, no solid conclusions yet. In-vac, we aligned MICH and the PRM. Out of vac, we got beam on AS and REFL paths. We can lock MICH, but we're not as happy with PRCL.
In-vac alignment:
To get the beam centered on TT2 in yaw, Koji helped us out and moved TT1 with the sliders a little bit. Then to get the beam centered on PRM and PR2, Koji moved the TT2 sliders a little bit.
Yuta and I then moved PR2 forward a few mm, to keep the optical path length of the PRC approximately (within ~1mm, hopefully) the same as always. After my PR2 optic swapping earlier, the pitch alignment was no longer good. I loosened the screws holding the wire clamp to the optic holder, and tapped it back and forth until the alignment was good. Of course, the screw-tightening / pitch-checking is a stochastic process, but eventually we got it. A small amount of yaw adjustment by twisting the PR2 TT was also done, but not much was needed.
Beam was a little off in pitch at ITMY, so Yuta poked the top of PR3, and that one single poke was perfect, and the beam was very nicely centered on the ITMY target. Beam was getting through BS target just fine. We checked at ITMX, and the beam looked pretty centered, although we didn't put in a target. We didn't do anything to BS while we were in-vac, since it was already good.
We aligned the ITMs so that their beams were retroreflecting back to the BS. After this, we saw nice MICH fringes.
We aligned the PRM so that its beam was retroreflecting.
We checked that we were getting REFL and AS beams out of the vacuum, which we were (a small amount of adjustment was done to AS path steering mirrors).
AS table alignment:
We did a bit of tweaking of the REFL path, and lots of small stuff to the AS path.
The AS beam was coming out of vac at a slightly different place in yaw, so we moved the first out of vac AS steering mirror so the beam hit the center, rather than ~1/3 of the way to the edge. We then aligned the beam through the lens, to the camera, and to AS55. Most significantly, we removed the BS that was just before AS55. This was sending beam to a dump, but it is in place to send beam over to AS110, once we get back to real locking. We measured ~30 microwatts of power going to the AS55 PD, while MICH fringes were fringing.
The REFL path didn't need much, although we had never been going through the center of the HWP and PBS that are used to reduce the power before going to the PDs, so we translated them a millimeter or two.
We see signal on dataviewer for all of the channels that we're interested in....AS55 I&Q, ASDC, REFL11 I&Q, REFLDC (which comes from REFL55).
Locking:
Locking MICH was very easy, after we rotated the phase of AS55 to get all the good MICH signal in the Q phase. Part of the criteria for this was that the AS55_Q_ERR signal should cross zero when ASDC went to 0. This was done very coarsely, so we need to do it properly, but it was enough to get us locked. We changed the phase from 24.5 to 90 deg.
PRCL has been more of a challenge, although we're still working on it.
On the back face of the Faraday, we see the michelson fringes, but they are not getting through the Faraday's aperture. This implies that we have a poorly aligned michelson, in that the interference between the returning beams from the ITMs is happening at a different place than the original beam splitting. Yuta is working on getting a better MICH right now. EDIT, 10 minutes later.... This seems to be fixed, and the MICH fringes enter the back aperture of the FI, but there is still the PRM refl problem (next paragraph).
Also, when we get the most bright REFL beam, we see that there is some very obvious clipping in the back of the Faraday aperture, and this is matched by a clipped-looking REFL beam on the AS table. We must understand what we have done wrong, such that when the beam is actually going through the Faraday, we see a much dimmer beam. It's possible that there is some clipping happening at that time with the in-vac REFL path....we need to check this. It's not a clipping problem on the AS table - I checked, and the beams are still reasonably well centered on all of the mirrors.
We think that the MICH / REFL beam problems may be that the input pointing is close, but not perfect. We have not confirmed today that the beam is centered on ETMY. We should do this as part of our final alignment procedure before putting on doors.
Plans for tomorrow:
Get POP aligned, especially the camera, so we can see what our intracavity mode really looks like in the PRC. This is probably (in part, at least) due to our having moved PR2 around, so the transmitted beams aren't in exactly the same place.
We think that it's more useful in the short term to check out the PRC, and since the clipping problem with the REFL beam is likely an imperfect input pointing, we want to use the other measured G&H mirror, and do another half-PRC test, with the test mirror in front of the BS. This requires much less perfection in the input pointing, so it should be very quick to set up.
Confirm that PRM oplev is still aligned (turn laser back on first).
Plans for next week:
Perfect the input pointing, by checking the beam position at ETMY. Recheck all corner alignment.
Try again locking PRMI in air. First, confirm ITM and BS oplevs are all aligned. |
8031
|
Fri Feb 8 02:38:04 2013 |
Koji | Update | Locking | PRMI work | I feel it's too hasty to use the PRMI.
I support the idea of the half-PRC test, to make an apple-to-apple comparison.
Make haste slowly. |
8032
|
Fri Feb 8 11:01:18 2013 |
Jamie | Update | Locking | PRMI work | I completely agree with Koji. We definitely should have locked the half PRC first. We were all set up for that. Why go through all this work to align MICH when we haven't confirmed with the half PRC that the flipping is helping us? The first rule of debugging is to only make one change at a time. We have measurements from the half PRC, so we could have made a direct comparison with those to see how things have changed. If we jump the gun we're going to end up wasting more time when we have to back-track.
Also, we never talked about moving PR2 to adjust optical path length, although I can understand why we would think that should be done. My calculations were all done assuming the free-space separation between PRM/PR2 and PR2/PR3 were unchanged. It's possible changing the position is better, but again, it's more work and it changes multiple things at one. I can redo my calculations for this new scenario, but we need to update our drawings with this new configuration. Please note precisely where PR2 has moved to.
We should have just flipped PR2 and that's it. Then we could have run the exact same measurements we had previously. Only then, once we understood this new simple cavity, should we have done further adjustments. |
8033
|
Fri Feb 8 11:07:07 2013 |
Jamie | Summary | General | PRC/arm mode matching calculations |
Quote: |
I would guess that either flipping PR2 or PR3 would give nearly the same effect (g = 0.9) and that flipping both makes it even more stable (smaller g). But what we really need is to see the cavity scan / HOM resonance plot to compare the cases.
The difference of 0.5% in mode-matching is not a strong motivation to make a choice, but sensitivity to accidental HOM resonance of either the carrier or f1 or f2 sidebands would be. Should also check for 2*f2 and 2*f1 resonances since our modulation depth may be as high as 0.3. Accidental 2f resonance may disturb the 3f error signals.
|
You would guess, and I would have guessed too, but the calculations tell the story. Flipping both does not increase the stability. The main issue is that flipping PR3 induces considerable astigmatism. This is why flipping PR3 alone does not make the cavity stable. I will do some simple calculations today that will demonstrate this effect.
But again, we should only change one thing at a time and understand that before moving on. Given that the calculations show that flipping only PR2 should alone have a positive affect, we should do just that first, and verify that we understand what's going on, before we move on to making more changes.
I will try to make some more arbcav runs as well, for just the flipped PR2. |
8034
|
Fri Feb 8 12:39:32 2013 |
yuta | Update | Locking | PRMI work | Half-PRC at this time already have two changes from the previous half-PRC; PR2 replaced/flipped and different TM before BS.
PRMI has only one change from the previous PRMI; PR2 replaced/flipped.
This is why I wanted to try PRMI first. But we now recognized that MI alignment (including REFL and AS alignment) is tough without using the arms, I agree that we should try half-PRC first.
I don't exactly know what the situation in the Jamie's calculation, but to make the optical path length the same before and after flipping, PR2 holder have to move about n*t, where n is the substrate refractive index and t is the thickness of the mirror, towards PRM/PR3.
Quote: |
The first rule of debugging is to only make one change at a time.
Also, we never talked about moving PR2 to adjust optical path length,
|
|
8035
|
Fri Feb 8 12:42:45 2013 |
nicolas | Summary | General | PRC/arm mode matching calculations |
Quote: |
The main issue is that flipping PR3 induces considerable astigmatism.
|
Yes, at 45degrees PR3 will only have a curvature of about 850m for the vertical mode of the beam, apparently not enough to stabilize the cavity. |
8036
|
Fri Feb 8 12:43:26 2013 |
yuta | Update | Computers | videocapture.py now supports movie capturing | I updated /opt/rtcds/caltech/c1/scripts/general/videoscripts.py so that it supports movie capturing. It saves captured images (bmp) and movies (mp4) in /users/sensoray/SensorayCaptures/ directory.
I also updated /opt/rtcds/caltech/c1/scripts/pylibs/pyndslib.py because /usr/bin/lalapps_tconvert is not working and now /usr/bin/tconvert works.
However, tconvert doesn't run on ottavia, so I need Jamie to fix it.
videocapture.py -h:
Usage:
videocapture.py [cameraname] [options]
Example usage:
videocapture.py MC2F -s 320x240 -t off
(Camptures image of MC2F with the size of 320x240, without timestamp on the image. MUST RUN ON PIANOSA!)
videocapture.py AS -m 10
(Camptures 10 sec movie of AS with the size of 720x480. MUST RUN ON PIANOSA!)
Options:
-h, --help show this help message and exit
-s SIZE specify image size [default: 720x480]
-t TIMESTAMP_ONOFF timestamp on or off [default: on]
-m MOVLENGTH specity movie length (in sec; takes movie if specified) [default: 0] |
8037
|
Fri Feb 8 15:53:48 2013 |
Jenne | Update | Locking | PRMI work |
Quote: |
I completely agree with Koji. We definitely should have locked the half PRC first. We were all set up for that.
|
I reminded Jamie this morning that we were not, in fact, set up yesterday for a half PRC. I had extracted what was the flat test mirror, to put in as PR2. The test mirror was the better of the 2 G&Hs that we had measurements for, so I had used it as the flat test mirror, but then also wanted it to be the more permanent PR2. After doing the PR2 flip, the IFO was naturally all aligned for PRMI, which is part of why we just did that.
Anyhow, Jamie used his tallness to put the other measured G&H mirror into the mount, and put that in front of the BS. He aligned things such that he saw fringes in the half PRC.
I then aligned POP onto the camera, and onto the PD. Yuta is confirming that we're maximally on the REFL PDs.
We're starting locking in 5 min. |
8038
|
Fri Feb 8 17:15:56 2013 |
Jenne | Update | RF System | MC REFL Photodiode transimpedance | This measurement was done already about a week ago, in elog 7984. Can you please describe why the numbers for the last measurement were not believable, and what was done differently this time? |
8039
|
Fri Feb 8 17:41:34 2013 |
Jenne | Update | Locking | PRMI work |
[Yuta, Jenne]
After much tweaking of the alignment using TT1, TT2 and PRM sliders, we were able to get a TEM00 mode locked with the half PRC!
PRCL gain is -0.010
FM4, 5 are always on. FM2,3,6 (boosts and stack res-gains) are triggered to come on after the cavity is locked.
We see a little clipping of POPDC, even though there are 2 BSs in the beam path, to dump 50% and then 67% of the beam. But it's not so much that we can't align.
REFLDC goes from 28.5 to 24.5, so we don't have great visibility.
Please watch our awesome video of the cavity, where we demonstrate that the half cavity is stable:
The cavity is flashing for the 1st 15 sec, then locks. Upper right is REFL, Lower right is POP, Upper left is back of the Faraday, Lower left is MC2F. Note that we definitely see some not so beautiful modes flashing, but most of that is due to the half cavity length and thus greater degeneracy of modes. Jamie is posting a HOM plot presently.
BEAM MOTION:
The beam is moving way more than it should be. Right now the PRM oplev is not coming out of the vacuum, since the flat test mirror mount is obstructing it. However, as we saw with other half-cavity tests, turning on the PRM oplev helps, but does not completely eliminate the beam motion. We should consider putting oplevs on one of the passive TTs, at least temporarily, so we know what kind of motion is coming from where. |
8040
|
Fri Feb 8 18:23:32 2013 |
Jamie | Summary | General | arbcav of half PRC with flipped PR2 | Arbcav with half PRC (flat temporary mirror in front of BS), PR2 RoC = 600m, PR3 RoC = -600m:
 
NOTE: this does NOT include the affect of the PR2 substrate in the cavity. Arbcav does not handle that. It would have to be modified to accept arbitrary ABCD matrices.
NOTE: I added to the mode plot the frequency separation of the first HOMs from the carrier (\omega_{10/01}), in units of carrier FSR. |
8041
|
Fri Feb 8 19:29:44 2013 |
yuta | Summary | General | arbcav of half PRC with flipped PR2 | We need expected finesse and g-factor to compare with mode-scan measurement. Can you give us the g-factor of the half-PRC and what losses did you assumed to calculate the finesse?
Also, flipped PR2 should have RoC of - R_HR * n_sub (minus measured RoC of HR surface multiplied by the substrate refractive index) because of the flipping.
According to Jenne dictionary, HR curvature measured from HR side is;
PRM: -122.1 m
PR2: -706 m
PR3: - 700 m
TM in front of BS: -581 m
Please use these values to calculate expected g-factor so that we don't get confused.
Quote: |
Arbcav with half PRC (flat temporary mirror in front of BS), PR2 RoC = 600m, PR3 RoC = -600m:
|
|
8042
|
Fri Feb 8 19:39:02 2013 |
Koji | Update | Locking | PRMI work | It seems that the cavity trans looks much better than before. Cool.
At least the optical gain is ~x5 of the previous value. This means what we did was something good.
Looking forward to seeing the further analysis of the caivty... |
8043
|
Fri Feb 8 20:05:15 2013 |
Jenne | Update | Locking | PRMI work | I fixed up the POP path so that there is no clipping, so that Yuta can take a cavity mode scan. |
8044
|
Fri Feb 8 20:27:56 2013 |
Koji | Update | RF System | MC REFL Photodiode transimpedance | The comment itself was added by me.
The difference between the previous and new measurements should be described by Riju.
In the entry 7984, the description has several PDs mixed up. The measurement was done with the MCREFL PD.
But the DC transimpedance of the thorlabs PD (5e3) was used, according to the text.
I first wonder if this is only a mistake not in the calculation but only in the elog due to a sloppy copy-and-paste.
But the resulting shot-noise-intercept current was 50uA, which is way too small
compared with a realistic value of 0.1~1mA. I have never seen such a good value with
C30642 at the resonant freq ~30MHz. That's why I said "hard to believe". I guessed this wrong
DC transimpedance was actually used for the calculation.
You may wonder why this 50uA is unreasonable number.
Basically this is just my feeling and probably is same as Rana's feeling.
But "my feeling" can't be a scientific explanation. Here is some estimation.
Looking at my note in 2010:
https://wiki-40m.ligo.caltech.edu/40m_Library (Comparisons of the PD circuits by KA)
The expected shot noise intercept current (idc) is
idc = 2 kB T / (e Rres),
where Rres is the impedance of the resonant circuit at the resonant freq.
This Rres is expressed as
Rres = 1/(4 pi^2 fres^2 Cd^2 Rs),
where Cd and Rs are the capacitance and series resistance of the diode.
If we input realistic numbers,
Cd = 100pF
Rs = 10 Ohm
fres = 30MHz
We obtain, Rres ~ 300Ohm, and idc = 0.2mA
In other words, Rs needs to be 2~3Ohm in order to have idc = 50uA.
This is too small from the previous measurements.
Test Results for C30642 LSC Diode Elements by Rich Abbott
|
8045
|
Fri Feb 8 21:14:52 2013 |
Manasa | Update | Optics | G&H - AR Reflectivity | Hours of struggle and still no data 
I tried to measure the AR reflectivity and the loss due to flipping of G&H mirrors
With almost no wedge angle, separating the AR reflected beam from the HR reflected beam seems to need more tricks.

The separation between the 2 reflected rays is expected 0.8mm. After using a lens along the incident beam, this distance was still not enough to be separable by an iris.
The first trick: I could find a prism and tried to refract the beams at the edge of the prism...but the edges weren't that sharp to separate the beams (Infact I thought an axicon would do the job better..but I think we don't have any of those).
Next from the bag of tricks: I installed a camera to see if the spots can actually be resolved.
The camera image shows the 2 sets of focal spots; bright set to the left corresponding to HR reflected beam and the other from the AR surface. I expect the ghost images to arise from the 15 arcsec wedge of the mirror. I tried to mask one of the sets using a razor blade to see if I can separate them and get some data using a PD. But, it so turns out that even the blade edge is not sharp enough to separate them.
If there are any more intelligent ideas...go ahead and suggest!

|
8046
|
Fri Feb 8 22:49:31 2013 |
Koji | Update | Optics | G&H - AR Reflectivity | How about to measure the AR reflectivity at larger (but small) angles the extrapolate the function to smaller angle,
or estimate an upper limit?
The spot separation is
D = 2 d Tan(\phi) Cos(\theta), where \phi = ArcSin(Sin(\theta) * n)
D = 2 d Tan(\phi) Cos(\theta), where \phi = ArcSin(Sin(\theta) / n) (<== correction by Manasa's entry)
\theta is the angle of incidence. For a small \theta, D is propotional to \theta.
So If you double the incident angle, the beam separation will be doubled,
while the reflectivity is an even function of the incident angle (i.e. the lowest order is quadratic).
I am not sure until how much larger angle you can use the quadratic function rather than a quartic function.
But thinking about the difficulty you have, it might be worth to try. |
8047
|
Fri Feb 8 23:04:40 2013 |
Manasa | Update | Optics | G&H - AR Reflectivity |
Quote: |
D = 2 d Tan(\phi) Cos(\theta), where \phi = ArcSin(Sin(\theta) * n)
\theta is the angle of incidence. For a small \theta, D is propotional to \theta.
|
n1Sin(\theta1) = n2 Sin(\theta2)
So it should be
\phi = ArcSin(Sin(\theta) / n
I did check the reflected images for larger angles of incidence, about 20 deg and visibly (on the IR card) I did not see much change in the separation. But I will check it with the camera again to confirm on that. |
8048
|
Fri Feb 8 23:22:48 2013 |
Den | Summary | Modern Control | progress report | I wrote a small document on the application of LQG method to a Fabry-Perot cavity control. |
Attachment 1: LQG.pdf
|
|
|