ID Date Author Type Categorydown Subject Text Attachments
  5759   Fri Oct 28 18:33:59 2011 steveUpdateVACleaking nitrogen line fixedI was lucky to notice that the nitrogen
supply line to the vacuum valves was leaking.
Closed ALL valves. Open supply line to atm.
 allvalvesclosed.png 
  5767   Mon Oct 31 08:55:19 2011 steveUpdateVACpressure plot at day 53


   
        
            
 pd71-m-d53.png 
  5887   Mon Nov 14 15:46:14 2011 steveUpdateVACTP2's forepump changedThe foreline pressure of TP2 was 1.4 Torr
this morning. This drypump worked well for
ten months.
 drypump.png 
  5950   Fri Nov 18 16:37:14 2011 steveUpdateVACpreparing for ac power interruptionThe vacuum is ready for no AC power for
1 hr on Sunday morning at 10am
 
 planednopower.png 
  5962   Sun Nov 20 13:51:54 2011 steveUpdateVACvacuum is back up after ac interruptionState condition: Vac Normal,         
CC1 ~8e-6 Torr
The IFO pressure peaked at 
 notpumping.png 
  6790   Fri Jun 8 15:24:08 2012 steveUpdateVACRGA scan at day 276 and pressure plots

    

       
            
 40m71pdM276d.pngpd71m276d.pngpd71d276.png 
  6894   Fri Jun 29 11:02:00 2012 steveUpdateVACthe vent is at 500 torr

    

       
            
 sussum.pngpzt.png 
  6895   Fri Jun 29 13:11:31 2012 steveUpdateVAC40m IFO at atm The 4 hrs vent plot at 3 torr/min
rate.
Nitrogen was used from 1e-6 torr
 vent06292012.png40m@ATM.png 
  6897   Fri Jun 29 22:56:35 2012 JamieUpdateVACinput telescope beam clipping on Faraday[Yuta, Koji, Jamie]
We went into ITMX chamber to inspect
the situation there.  We looked for
  
  6906   Tue Jul 3 17:23:50 2012 steveUpdateVACpumpdown completedVacuum Normal State is reached in 9 hours. 
CC1 =  2e-5 Torr
Aux dry pump #3 is still running. 
 pd72.pngpd72vacnormal.pngpd72at9hrs.png 
  6929   Fri Jul 6 09:27:40 2012 steveUpdateVAC4 days at atm I'm looking for some movement indicators
of the vent-pump down events.
 
 vent_moves.png 
  6930   Fri Jul 6 09:52:35 2012 steveUpdateVACpump down hick up I can not understand what really
happened here with CC1 gauge or there was
really a pressure glitch.
 d3wFireworks.pngpdhickup.png 
  6970   Thu Jul 12 16:27:04 2012 steveUpdateVACdrypump changedTP2  dry_pump was changed at intake
pressure 982 mTorr at 84,889 hrs  This
seal hold up for one year.
  
  7044   Fri Jul 27 14:28:36 2012 steveUpdateVACcold cathode gaugesOur cold
cathode 423 gauges are 10 years old.
They get insulated by age and show no ionization
 cc_gauges.png 
  7051   Mon Jul 30 15:49:16 2012 steveUpdateVACcc1 switched

    

       
            
 cc1changedfVtH.png 
  7056   Tue Jul 31 08:01:22 2012 steveUpdateVACcc1 switched

    

       
            
 cc1h1d.png 
  7089   Mon Aug 6 10:53:32 2012 steveUpdateVACstatus: vauum normal 
 The power outage did not have
any effect on the vacuum system.  I
 CCgauges.png 
  7272   Fri Aug 24 16:03:39 2012 SteveUpdateVACVacuum related work at atmVacuum related work at atmosphere:
Atm1,  Check all chamber dog
clamps tightness with torque wrench,
 IMG_1563.JPGIMG_1565.JPGIMG_1566.JPG 
  7278   Sun Aug 26 20:58:21 2012 JenneUpdateVACDon't vent!!!![Koji, Jenne]
Steve, do not vent tomorrow morning! 
We are still not prepared, and will not finish
  
  7293   Tue Aug 28 09:37:33 2012 SteveUpdateVACRGA scan at day 56RGA scan,  Maglev pumping speed at
day 56
CC1 is dying. CC4 is real.
 pd72m56d.pngpd72m56dpr.png 
  7300   Tue Aug 28 17:59:03 2012 JenneUpdateVACVENT: GO!I have turned of the high voltage supplies
for PZT1 and PZT2.  The OMC PZT high
voltage supplies were already off, since
  
  7306   Wed Aug 29 11:47:21 2012 ericqUpdateVACVenting [Steve, Eric]
I've been helping Steve vent this
morning. The following things were done (from
  
  7307   Wed Aug 29 12:46:49 2012 SteveUpdateVACVent completed in 4 hrs 
 vent72.pngbeforeVentOpl.pngbeforeVent.png 
  7321   Thu Aug 30 19:08:08 2012 JenneUpdateVACDogs on BS, ITM chambers checkedI tightened as many of the dog clamps on
the bottom of the BS, ITMX and ITMY chambers
as I could find.  I used a torque wrench
  
  7327   Fri Aug 31 10:24:36 2012 SteveUpdateVACchamber dog clamps checked

    

       
            
  
  7342   Tue Sep 4 20:25:22 2012 jamieOmnistructureVACbetter in-air "lite" access connector neededWe really need something better to replace
the access connector when we're at air. 
This tin foil tunnel crap is dumb. 
  
  7343   Wed Sep 5 09:50:25 2012 SteveOmnistructureVACbetter in-air "lite" access connector needed

    

       
            
 IMG_1612.JPG 
  7344   Wed Sep 5 10:50:15 2012 jamieOmnistructureVACbetter in-air "lite" access connector needed

    

       
            
  
  7345   Wed Sep 5 13:11:43 2012 jamieOmnistructureVACbetter in-air "lite" access connector needed

    

       
            
  
  7432   Mon Sep 24 16:59:31 2012 JenneUpdateVACDoors on, ready to pump[The 40m Family]
The access connector and all heavy
doors are back on. 
  
  7434   Mon Sep 24 17:47:05 2012 SteveUpdateVAC ready to pump  The IFO will be at atmosphere 
overnight. The annuloses are pumped
down. The access connector jam nuts are tightened
  
  7438   Tue Sep 25 17:21:01 2012 SteveUpdateVACpumpdown completed IFO P1= 1mTorr,   CC1 =
5e-4 Torr.  The IFO is ready for work.
Be free to open the shutter and turn on HV
 pd73.pngpd73b.png 
  7441   Wed Sep 26 11:03:04 2012 SteveUpdateVACpumpdown completed The new cold cathode gauge CC1 is
in place. We were at atmosphere for 28 days
......more later
 40mpd73md1.png28datm.png 
  7454   Mon Oct 1 09:02:17 2012 SteveUpdateVACpumpdown reached vacuum normal

    

       
            
 pd73md6.pngpd73md6b.png 
  7461   Tue Oct 2 09:13:16 2012 SteveUpdateVACRGA scan at day 7

    

       
            
 pd73md7.png 
  7491   Fri Oct 5 14:40:55 2012 SteveUpdateVACRGA scan after power outage

    

       
            
 afterpoweroutaged10.pngpd73m10d.png 
  7545   Mon Oct 15 08:12:45 2012 SteveUpdateVACblank vac monitorVME
crate  blew its fuse and the fuse
housing is hard to access.
 blankvac.png 
  7547   Mon Oct 15 13:07:51 2012 jamieUpdateVACvacuum VME crate broken, replaced, minor vacuum mayhem ensuesSteve and I managed to access the fuse
in the vacuum VME crate, but replacing it
did not bring it back up.  We decided
  
  7549   Mon Oct 15 17:08:01 2012 SteveUpdateVACvacuum is back with some disabled valves

    

       
            
 VMEcrateout.pngIMG_1725.JPGwarning.pngshortcomings.jpg 
  7564   Wed Oct 17 08:04:31 2012 SteveUpdateVACRoughing pumps on for oil changePR1, PR2 and RP3 turned on for warming
up for oil change. Oil changed with 3.2L
of  MVT-19
  
  7582   Fri Oct 19 16:45:39 2012 SteveUpdateVACvent has started

    

       
            
  
  7584   Fri Oct 19 19:26:42 2012 JenneUpdateVACvent has started

    

       
            
  
  7587   Mon Oct 22 09:10:07 2012 SteveUpdateVAC@750Torr

    

       
            
 ventOct222012.png 
  7594   Tue Oct 23 08:45:00 2012 SteveUpdateVACvent day 3

    

       
            
 ventd4.pngIMG_1743.JPG 
  7597   Tue Oct 23 16:18:03 2012 SteveUpdateVACRoughing pumps on for oil change

    

       
            
 oilchange.jpg 
  7605   Wed Oct 24 09:15:12 2012 SteveUpdateVACvacuum rack 24V ps was current limited

    

       
            
 currentlimited24V.jpg 
  7669   Mon Nov 5 10:34:48 2012 SteveUpdateVACexisting vacuum documents

    

       
            
  
  7696   Mon Nov 12 09:21:10 2012 SteveUpdateVACpumpdown has startedI'm looking for IOO steering mirror power
supplies. I found two so far: one at 1Y3
as Kepco  + 75V and one at OMC_North
 beforepd.pngIMG_1800.JPGIMG_1801.JPG 
  7697   Mon Nov 12 16:52:03 2012 SteveUpdateVACpumpdown has completed The pressure is CC1 3e-4 Torr at
8 hours of pumping.  Status: vacuum
normal
 pd74.pngpd74at8h.png 
  7703   Tue Nov 13 08:34:11 2012 SteveSummaryVACpumpdown after 24days at atmosphere 
 
 pd74d1.png24datm.pngpdd1.png 