40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 140 of 344  Not logged in ELOG logo
ID Date Authordown Type Category Subject Text Attachments
  15852   Mon Mar 1 12:36:38 2021 gautamSummaryIMCgetting familiar with IMC controlsPretty minor thing - but PMCT and PMCR
were switched on Quad 2 for whatever reason.
I switched them back because I prefer the
  
  15853   Mon Mar 1 16:27:17 2021 gautamUpdateLSCPRM violin filter excessive?The PRM violin filter seems very suboptimal
- the gain peaking shows up in the MICH OLTF,
presumably due to the MICH-->PRM LSC output
 violins.pdfPRviolin.pdf 
  15855   Tue Mar 2 19:52:46 2021 gautamUpdateLSCREFL55 demod board reworkThere were multiple problems with the REFL55
demod board. I fixed them and re-installed
the board. The TFs and noise measured on
  
  15859   Wed Mar 3 22:13:05 2021 gautamUpdateLSCREFL55 demod board reworkAfter this work, I measured that the orthogonality
was poor. I confirmed on the bench that the
PQW-2-90
  
  15860   Wed Mar 3 23:23:58 2021 gautamUpdateALSArm cavity scanI see no evidence of anything radically
different from my PSL table optical
characterization in the IMC transmitted
 armScan.pdf 
  15867   Fri Mar 5 13:53:57 2021 gautamUpdateLSCREFL55 demod board rework0 dBm ~ 0.63 Vpp. I guess there is ~4dB
total loss (3dB from splitter
and 1dB from total excess loss above theoretical
  
  15868   Fri Mar 5 15:03:28 2021 gautamSummaryElectronicsA bunch of electronics receivedThe PCBs for the D1002593
BIO I/F (5pcs ea of D1001050
and D1001266)
  
  15871   Fri Mar 5 16:24:24 2021 gautamUpdateLSCREFL55 demod board re-installed in 1Y2I don't have a good explanation why,
but I too measured similar numbers to what
Koji measured. The overall conversion gain
 REFL55.pdfdemodNoise.pdf 
  15873   Fri Mar 5 22:25:13 2021 gautamUpdateLSCPRMI 1f SB locking recoveredNow that the REFL55 signal chain is capable
of providing balanced, orthogonal readout
of the two quadratures, I was able to recover
 PRMI1f_noArmssensMat.pdf 
  15874   Sat Mar 6 12:34:18 2021 gautamUpdateLSCSensing matrix settings messed withTo my dismay, I found today that somebody
had changed the oscillator frequencies for
the sensing matrix infrastructure we have.
 sensMat.png 
  15875   Sun Mar 7 15:26:10 2021 gautamUpdateLSCHousekeeping + more PRMI
Beam pointing into PMC was tweaked
to improve transmission.
AS110
photodiode was re-installed on the AS table
 PRMI_SBres.pngMICH_act_calib.pdf 
  15878   Mon Mar 8 12:40:35 2021 gautamSummarytrainingInvestigate how-to XARM lockingFor the arm locking, the "Restore
Xarm (XARM POX)" script from the "IFO_CONFIGURE"
MEDM screen should get you there (I just checked
  
  15879   Mon Mar 8 12:54:54 2021 gautamUpdateEquipment loan40m-->Cryo
Busby box
SR554 transformer
preamplifier
  
  15880   Mon Mar 8 17:09:29 2021 gautamUpdateSUSPRM coil actuators heavily imbalancedI realized I hadn't checked the PRM
actuator as thoroughly as I had the others.
I used the Oplev as a sensor to check the
 PRMact.pdf 
  15883   Mon Mar 8 22:01:26 2021 gautamUpdateLSCMore PRMIThere are still many mysteries remaining
- e.g. the MICH-->PRCL contribution still
can't be nulled. But for now, I have
 PRMI1f_noArmssensMat.pdfPRMI3f_noArmssensMat.pdf 
  15887   Tue Mar 9 14:37:26 2021 gautamSummarySUSPRM suspensionThe PRM got tripped ~5AM this morning.
The cause is unclear - the seismometer reports
elevated activity ~10 minutes before the
 PRMtrip.png 
  15892   Wed Mar 10 00:32:03 2021 gautamUpdateLSCPRFPMiThe interferometer can nearly be locked
again. I was unable to fully hand off control
from ALS-->RF, I suspect I may be using
 PRMI3f_ALSsensMat.pdf 
  15894   Wed Mar 10 11:55:22 2021 gautamUpdateSUSPRM suspension suspectThe procedure is that the optic is kicked
to excite it, and allowed to ring down for
~1ksec, with damping turned off. The procedure
 PRM_sensorSpectra.pdfPRM_pkFitNaive.pdfPRM_diagComp.pdf 
  15895   Wed Mar 10 15:00:16 2021 gautamSummaryIMCIMC free swinging prepDid you fix this issue? It is helpful to
post a screenshot of the offending MEDM screen
in addition to witticisms. The elog says
  
  15898   Wed Mar 10 17:35:47 2021 gautamUpdateSUSSpooky action at a distanceAs I am sitting in the control room, the
PRM suspension watchdog tripped again. This
time, there is clearly no
 PRMtrip.pngPRMtrip_zoom.png 
  15899   Wed Mar 10 19:58:27 2021 gautamUpdateLSCSR785 hooked up to CM boardIn preparation for later today evening.
The TT alignment wasn't visibly disturbed.
  
  15900   Thu Mar 11 01:45:42 2021 gautamUpdateLSCPRFPMi
PRM satellite box indeed seems
to have been the culprit - shortly after
I swapped it to the SRM, its shadow sensors
  
  15903   Thu Mar 11 14:03:02 2021 gautamUpdateLSCAO pathThere is some evidence of weird saturation
but the gain balancing (0.8dB) and orthogonality
(~89 deg) for the daughter board on the REFL11
 IMG_9127.HEIC 
  15904   Thu Mar 11 14:27:56 2021 gautamUpdateCDStimesync issue?I have recently been running into hitting
the 4MB/s data rate limit on testpoints -
basically, I can't run DTT TF and spectrum
 timesync.pngc1lscMods.png 
  15905   Thu Mar 11 18:46:06 2021 gautamUpdateCDScds rebootSince Koji was in the lab I decided to
bite the bullet and do the reboot. I've
modified the reboot script - now, it prompts
 CDSoverview.png 
  15906   Thu Mar 11 20:18:00 2021 gautamUpdateLSCHigh bandwidth POYI repeated the high bandwidth POY locking
experiment.


The "Q" demod
 OLGevolution.pdfinLoopSpec.pdf 
  15911   Fri Mar 12 11:02:38 2021 gautamUpdateCDScds rebootI looked into this a bit today morning.
I forgot exactly what time we restarted the
machines, but looking at the timesyncd logs,
 timeDrift.png 
  15913   Fri Mar 12 12:32:54 2021 gautamSummarySUSCoil Rs & Ls for PRM/BS/SRMFor consistency, today, I measured both
the BS and PRM actuator balancing using the
same technique and don't find as serious
 BS_actuator.pdfPRMact.pdf 
  15915   Fri Mar 12 13:48:53 2021 gautamSummarySUSCoil Rs & Ls for PRM/BS/SRMI didn't repeat Koji's measurement,
but he reports the expected ~3.2mH per coil
on all the BS and PRM coils.
  
  15917   Fri Mar 12 19:44:31 2021 gautamUpdateLSCDelay lineI may want to use the delay line phase
shifter in 1Y2 to allow remote actuation
of the REFL11 demod phase (for the AO path,
  
  15918   Fri Mar 12 21:15:19 2021 gautamUpdateLSCcoronaversary PRFPMiAttachment
#1 - proof that the
 PRFPMI.pngCARM_OLTF.pdf 
  15920   Mon Mar 15 20:22:01 2021 gautamUpdateASCc1rfm model restartedOn Friday, I felt that the ASC performance
when the PRFPMI was locked was not as good
as it used to be, so I looked into the situation
 RFM.pngCDSoverview.pngRFMchans.pngASCloops.png 
  15925   Tue Mar 16 19:04:20 2021 gautamUpdateCDSFront-end testingNow that I think about it, I may only have
backed up the root file system of chiara,
and not
  
  15927   Wed Mar 17 00:05:26 2021 gautamUpdateLSCDelay line BIO remote controlWhile Koji is working on the REFL 11 demod
board, I took the opportunity to investigate
the non-remote-controllability of the 
  
  15932   Wed Mar 17 15:02:06 2021 gautamUpdatesafetyDoor to outside from control room was unlockedI came into the lab a few mins ago and
found the back door
  
  15933   Wed Mar 17 15:04:20 2021 gautamUpdateElectronicsRibbon cable for chassisI had asked Chub to order 100ft ea of 9,
15 and 25 conductor ribbon cable. These arrived
today and are stored in the VEA alongside
 IMG_9139.jpg 
  15935   Thu Mar 18 01:12:31 2021 gautamUpdateLSCPRFPMi
Integrated >1
hour at RF only control, high circulating
powers tonight.


All
 DARM_OLTF_20210317.pdfCARMTF_20210317.pdfPRFPMI_Mar_17sensMat.pdf 
  15940   Thu Mar 18 13:12:39 2021 gautamUpdateComputer Scripts / ProgramsOmnigraffle vs draw.ioWhat is the advantage of Omnigraffle c.f.
draw.io?
The latter also has a desktop app, and for
  
  15941   Thu Mar 18 18:06:36 2021 gautamUpdateElectronicsModified Sat Amp and Coil DriverI uploaded the annotated schematics (to
be more convenient than the noise analysis
notes linked from the DCC page) for the 
  
  15944   Fri Mar 19 11:18:25 2021 gautamUpdateLSCPRMI investigations: what IS the matrix??From Finesse simulation (and also analytic
calcs), the expected PRCL optical gain
is ~1 MW/m (there is a large uncertainty,
  
  15949   Fri Mar 19 22:24:54 2021 gautamUpdateLSCPRMI investigations: what IS the matrix??I did all these checks today. 




 IMG_9140.jpgIMG_9141.jpgREFL55.png 
  15953   Mon Mar 22 16:29:17 2021 gautamUpdateASCSome prep for AS WFS commissioning
Added rough cts2mW calibration
filters to the quadrants, see 
 AS_WFS_head.pngWFSquads.pdf 
  15956   Wed Mar 24 00:51:19 2021 gautamUpdateLSCSchnupp asymmetryI used the Valera technique to measure
the Schnupp asymmetry to be 
 Lsch.pdf 
  15958   Wed Mar 24 15:24:13 2021 gautamUpdateLSCNotes on testsFor my note-taking:


Lock PRMI with ITMs as
the MICH actuator. Confirm that the MICH-->PRCL
  
  15960   Wed Mar 24 22:54:49 2021 gautamUpdateLSCNew day, new problemsI thought I'd get started on some of
the tests tonight. But I found that this
problem had resurfaced. I don't know
 PRMI3f_noArmssensMat.pdfREFL55_whtGainStepping.pngREFL55_whtGainStepping2.png 
  15963   Thu Mar 25 14:16:33 2021 gautamUpdateCDSc1auxey assemblyIt might be a good idea to configure this
box for the new suspension config - modern
Satellite Amp, HV coil driver etc. It's
  
  15965   Thu Mar 25 15:31:24 2021 gautamUpdateIOOWFS servosThe servos are almost certainly not optimal
- but we have the IFO sort of working, so
before we make any changes, let's make
  
  15966   Thu Mar 25 16:02:15 2021 gautamSummarySUSRepeated measurement of coil Rs & Ls for PRM/BSMethod

Since I am mainly concerned with
the actuator part of the OSEM, I chose to
  
  15967   Thu Mar 25 17:39:28 2021 gautamUpdateComputer Scripts / ProgramsSpot position measurement scripts "modernized"I want to measure the spot positions on
the IMC mirrors. We know that they can't
be too far off centerBasically I did the
 MCdecenter202103251735_mcmirror0.pdfMCdecenter202103251735_mcdecenter0.pdf 
  15968   Thu Mar 25 18:05:04 2021 gautamUpdateElectronicsStuffed HV coil drivers received from Screaming CircuitsI think the only part missing for assembly
now are 4 2U chassis. The PA95s need to be
soldered on as well (they didn't arrive
  
ELOG V3.1.3-