Entry  Tue Jul 8 12:25:54 2008, John, Summary, PSL, ISS gain set to 2dB 
 
Entry  Tue Jul 8 21:46:38 2008, Yoichi, Configuration, PSL, GC650M moved to the PMC transmission 
I moved a GC650M, which was monitoring the light coming out of the PSL, to the transmission port of the PMC to see the transmitted mode shape.
It will stay there unless someone find other use of it.
Entry  Tue Jul 8 21:58:22 2008, alberto, Update, General, Secondaty beam aligned to the IFO beam again 
Yesterday the alignment of the secondary beam to the IFO was completely lost and today I had to realign all the optics before I was able to match the two
spots again. I had to reset the height of the irises and I had also to replace mirror M1 with one with a larger angular motion. Eventually I obtained the
beat again. Working on the optics table I inadvertently misaligned the OSA but I didn't make in time to bring it back before the night shift people came.
Entry  Wed Jul 9 12:42:14 2008, John, Update, Locking, Hand off to RF CARM LongLock080709.png
Rob, Yoichi, John

Last night we were able to reduce the CARM offset to around 80. This was achieved by increasing the DARM gain and
Entry  Wed Jul 9 15:04:22 2008, steve, Metaphysics, Photos, SURFs helping hands surfjob.png
Surf students are helping out with baffle cleaning.
Entry  Wed Jul 9 17:58:19 2008, John, Summary, General, Illuminator alarms 
This morning some time was wasted on alignment due to the illuminators.

I added the illuminators to the alarm handler. They will give a [COLOR=red]RED[/COLOR] alarm whenever
Entry  Thu Jul 10 13:47:12 2008, Yoichi, HowTo, Computers, svn access via https 
Now you can access to the svn repository on nodus by https.
To perform a checkout, you can use the following command
Entry  Thu Jul 10 19:12:07 2008, Alberto, Update, General, abs cavity length measurement experiment 
Yesterday morning, when I started, I found the IFO beam on a different position and the beam spot at the AS port looked very deformed. The overlapping with
the secondary beam was not good enough to observe the beats anymore. Restoring the alignments of the interferometers did not work because, as John found
out later, some of the photodiodes had offsets and gain which made the restore script ineffective. After resetting the parameters, we had to align every
Entry  Thu Jul 10 23:27:57 2008, John, Metaphysics, Cameras, Secret handshakes 
Rob and I have joined the ranks of the illuminati and exercised our power.

[quote] Osamu showed me the secret way to change the video labels for the quads and
Entry  Fri Jul 11 00:30:24 2008, rob, Metaphysics, Computers, strange SUS controllers 

[I]rob, johnnie[B]M[/B][/I]
Entry  Fri Jul 11 09:29:02 2008, steve, Update, PSL, the PMC still hangs up pmc4d.jpg
Morning alarms:

PMC high voltage is railing, it's auto locker is frosen
Entry  Fri Jul 11 20:16:01 2008, Eric, DAQ, Cameras, Taking data from the GC 750 Camera 
Mafalda has been set up with a background process to constantly take data from the GC 750 camera (at the end of the x-arm) for the weekend.  This camera
will otherwise be inoperable until then.
Entry  Thu Jul 10 14:59:01 2008, rob, Update, Locking, RF common mode at zero offset lock_control_sequence_worked.pngtrendpowerbuild.png
[I]rob, john, yoichi[/I]

[SIZE=3][FONT=Comic Sans MS]Last night we succeeded in reducing the CARM offset to zero[/FONT][/SIZE].  
    Reply  Fri Jul 11 23:55:25 2008, alan, Update, Locking, RF common mode at zero offset 
[quote][I]rob, john, yoichi[/I]

[SIZE=3][FONT=Comic Sans MS]Last night we succeeded in reducing the CARM offset to zero[/FONT][/SIZE].  
Entry  Sat Jul 12 23:28:31 2008, rana, Update, PSL, PMC needs help Untitled.png
As everyone has noticed recently, the PMC seems to have a PZT problem. It often zooms 
of to one of its rails after locking as if the PZT range has decreased dramatically.
Entry  Sun Jul 13 17:19:29 2008, rana, Summary, PSL, MOPA SLOWM Calibration SLOWDCcalibration.png
John, Rana

We first unlocked the FSS and ramped the SLOW actuator. With the PMC locked we observed the PMC PZT voltage
Entry  Sun Jul 13 22:39:16 2008, John, Summary, General, Edited medm screens  
I've edited the FSS and PMC screens so that red boxes are shown around the appropriate slider if a gain or offset is not within the limits defined in C1PSL_SETTINGS_SET.adl

With the current setting of 0 V the FSS input offset is red. According to the settings screen the nominal value is 0.3 +/- 0.050. Are there any objections
Entry  Mon Jul 14 00:36:19 2008, John, Summary, PSL, Slow sweep of laser temp - PMC PZT response SlowTsweep.png
John, Rana

Follow up to # 663
Entry  Mon Jul 14 10:57:00 2008, Koji, Frogs, Environment, Someone at 40M sent LHO water of life 
Someone at the 40m sent Mike@LHO a pound of peets coffee with the name of Koji Arai.
It was a good surprise! Thanks, we will enjoy it!
I will return to Pasadena next week. See you then.
Entry  Mon Jul 14 12:43:07 2008, John, Summary, Computers, Restarted fb40m, tpman and c1ass 
 
Entry  Mon Jul 14 19:15:43 2008, Alberto, Update, General, abs cavity length measurement experiment 
Lately I've been dealing with the alignment of the interferometer to have a good beam spot at the AS port. Today the alignment script kept failing  because
of computer problems (failure of the frame builder) and also because the IFO was probably too far from the range where the automatic alignment works.
    Reply  Mon Jul 14 21:34:10 2008, Alberto, Update, General, abs cavity length measurement experiment 
[quote]Lately I've been dealing with the alignment of the interferometer to have a good beam spot at the AS port. Today the alignment script kept failing
 because of computer problems (failure of the frame builder) and also because the IFO was probably too far from the range where the automatic alignment
works.
Entry  Tue Jul 15 09:47:09 2008, steve, Update, PEM, lab temps and particles temps24d.jpg
All air condtion units were serviced last Friday.
AC filters are trying to control our particle counts but they have no capacity to match bad Pasadena conditions.
IFO room filters at CES were really clean.
Entry  Tue Jul 15 10:09:42 2008, Eric, DAQ, Cameras, Did anyone kill the picture taking process on Mafalda? 
Did anyone kill the process on Mafalda that was taking pictures of the end mirror of the x-arm last Friday?  I need to know whether or not it crashed of
its own accord.
Entry  Tue Jul 15 10:24:57 2008, steve, Update, PSL, PMC temp & pzt voltage pmctemp.jpg
The PMC pzt HV was happy with no HEPA temp stability.
Can we thermally insulate the pmc ?
Entry  Tue Jul 15 11:47:56 2008, Jenne, DAQ, PEM, Accelerometer channels in ASS Adapt MEDM screen 
Jenne, Sharon

We have traced which accelerometers correspond to which channels in the C1ASS_TOP MEDM screen.  
Entry  Tue Jul 15 12:23:22 2008, Jenne, Update, General, MC2 Watchdog tripped 
Alberto, Jenne

Mode Cleaner was unlocked.  We checked, and found that MC2's watchdog was tripped.  It didn't look like anything bad was going on, so we turned the optic
Entry  Tue Jul 15 12:44:08 2008, John, Summary, Locking, DRFPMI with DC readout 
Rob, John

Last night, despite suspect alignment, we were again able to reduce the CARM offset to zero using 
Entry  Tue Jul 15 19:15:57 2008, rana, Summary, SUS, ETMX Dewhitening characterization etmx-dw-transient.pdf
Since the boys found that the ETM dewhitening transient was kicking the IFO out of lock we
decided to investigate. 
Entry  Wed Jul 16 09:27:17 2008, steve, Update, ALARM, PSL-FSS_RMTEMP alarm is false frmtemp2y.jpg
Morning alarm sound is good for people who does not drink coffee.
Our 40m alarm is on every morning.
Those whom are not here in the morning thinks that this beeping sound is inspirational.
Entry  Wed Jul 16 10:50:55 2008, Eric, Summary, Cameras, Weekly Summary AttemptedPeakFit3.tiff
Finished unwrapping, cleaning, baking, wrapping, wrapping again, packing, and shipping the baffles.  

Attempted to set up the Snap software so that it could talk directly to EPICS channels.  This is not currently working due to a series of very strange
Entry  Wed Jul 16 11:00:15 2008, Masha, Update, Auxiliary locking, improving ADC input for the mach zehnder setup and completely unrelated happenings 
For most of last week, the SURFs + Jenne were helping Mike and Ken with "stray light control for
Enhanced LIGO", i.e. cleaning and baking many many baffles which will catch scattered light in the
interferometer. 
Entry  Wed Jul 16 11:26:47 2008, Max Jones, Update, , This Week 
Baffles. 

I got a battery for the magnetometer today which is slightly too large (~2 mm) in one dimension. Not sure what I'm going to do. 
Entry  Wed Jul 16 15:59:04 2008, josephb, Eric, Configuration, Cameras, PMC trans camera path 
In order to reduce saturation, we placed a Y1 plate (spare from the SP table) in transmission just before the GC650 camera looking at the PMC transmision.
 The reflection (most of the light) was dumped to a convient razor blade dump.  We also removed the 0.3 and 0.5 ND filters and placed them in the 24 hour
loan ND filter box.
Entry  Wed Jul 16 16:28:14 2008, josephb, Configuration, Computers, Fixed IP address on Switch 
Realized today that the change I made back on June 30th to the switch was to the wrong switch.  I had disabled the DHCP setting and mislabeled the switch
in the control room (which seems to not have affected anything).
Entry  Wed Jul 16 16:59:07 2008, Alberto, Update, General, Aligment 
I think the two beams are aligned again - they both pass the Faraday, they match at the irises and all along the optical path on the AP table. Although
the NPRO beam does not show up at the AS port.
Entry  Wed Jul 16 17:36:51 2008, John, DAQ, PSL, FSS input offset 
I changed the nominal FSS input offset to 0 from 0.3. Tolerance remains unchanged at +/-0.05.
Entry  Wed Jul 16 17:51:58 2008, Masha, Update, Auxiliary locking, long measurement 
I'm taking a measurement on the SR785 spectrum analyzer at low frequencies, so I'm going to leave it by the symmetric port table for a while. Please don't
move it!
    Reply  Wed Jul 16 22:29:05 2008, Masha, Update, Auxiliary locking, long measurement 
[quote]I'm taking a measurement on the SR785 spectrum analyzer at low frequencies, so I'm going to leave it by the symmetric port table for a while. Please
don't move it![/quote]
Entry  Thu Jul 17 00:59:18 2008, Jenne, Summary, General, Funny signal coming out of VCO VCOout_time.PNGVCOout_freq.PNG
While working on calibrating the MC_F signal, Rana and I noticed a funny signal coming out of the VCO.  We expect the output to be a nice sine wave at about
80MHz.  What we see is the 80MHz signal plus higher harmonics.  The reason behind the craziness is to be determined.  For now, here's what the signal looks
like, in both time and frequency domains.
Entry  Thu Jul 17 08:30:15 2008, steve, Update, PSL, PMC relocked manually pmc4d.jpg
The PMC pzt HV and the servo gain adj. are railing at max this morning

Why is it on the decreasing side of FSS_RTTEMP slope?
Entry  Thu Jul 17 12:15:21 2008, Eric, Update, PSL, Swept PMC PZT voltage range 
I unlocked the PMC and swept over C1:PSL-PMC_RAMP's full range a couple of times this morning.  The PMC should now be relocked and returned 
to normal.
Entry  Thu Jul 17 13:08:37 2008, John, Summary, LSC, HOM resonances in the arms HOMresonances.png
On Tuesday night we attempted to lock the full DRFPMI in the optical spring configuration with the +f2 sideband resonant in the SRC.
Despite having no problems locking on the +f2 in a DRM we couldn't lock the full IFO.
Entry  Thu Jul 17 16:39:58 2008, Max Jones, Update, DAQ, Magnetometer Installed 
Today I installed the magnetometer near the beam splitter chamber. It is located on the BSC chamber at head height on the inner part of the interferometer
(meaning I had to crawl under the arms to install it). I don't think I disturbed anything during installation but I think that's it's probably prudent
to tell everyone that I was back there just in case. I plan to run 3 BNC cables (one for each axis) from the magnetometer to the DAQ input either tonight
Entry  Thu Jul 17 20:13:34 2008, Yoichi, Update, PSL, PMC alignment/mode matching effort 
I'm working to improve the mode matching of PMC.
Because I noticed that the beam was hitting the aperture of the EOM for PMC, I moved the EOM a little bit to maximize the transmission.
This did not change the alignment to the reference cavity but changed the alignment of the PMC a lot.
Entry  Fri Jul 18 12:24:15 2008, josephb, Eric, Configuration, Cameras, Changed Lenses on GC750 at ETMX 
We removed the giant TV zoom lens and replaced it with a much smaller fixed zoom lens.  Currently it views the entire optic.  We have another (also small)
zoom lens which focuses much better on the spot itself.  With how far back the camera is currently placed, neither of these fixed zoom lenses can touch
or hit the view port or the chamber while still attached to camera and mount, even using all of the mount's motion range.  So this should be less of a
Entry  Fri Jul 18 16:57:37 2008, Jenne, Update, IOO, Calibrated MC_F MC_Fcalib.png20080717MC_F-MC_I.pdf
I have calibrated MC_F.  The conversion factor is 137.49 MHz/count.

The calibration data taken is attached, along with a calibrated power spectrum.
Entry  Fri Jul 18 17:06:20 2008, Jenne, Update, Computers, Computers down for most of the day, but back up now 
[Sharon, Alex, Rob, Alberto, Jenne]

Sharon and I have been having trouble with the C1ASS computer the past couple of days.  She has been corresponding with Alex, who has been rebooting the
Entry  Fri Jul 18 17:12:35 2008, Jenne, Update, IOO, Checking out the MC Servo Board 
[Jenne, Max]

One of the things that Rana thinks that might be causing my MC_F calibration to be off is that the MC Servo Board's filters don't match those on the schematics.
    Reply  Fri Jul 18 19:15:15 2008, Jenne, Update, IOO, Checking out the MC Servo Board 
[quote][Jenne, Max]

I haven't checked to ensure that the MC still locks, because Yoichi is doing stuff on the PSL table, but I didn't change anything on the board, and hooked
Entry  Fri Jul 18 19:30:20 2008, Masha, Update, Auxiliary locking, moving from 40m 
I will be working in the basement of Bridge probably starting next week; I moved the NPRO laser and some of the optics from my mach zehnder setup on the
SP table to Bridge. Thanks for your help!
Entry  Fri Jul 18 19:43:55 2008, Yoichi, DAQ, Computers, PSL fast channels cannot be read by dataviewer 
At this moment only the PSL fast channels have trouble.
Rob restarted fb40m, c1IOVME, but no effect.
Entry  Fri Jul 18 19:41:09 2008, Yoichi, Update, PSL, PMC PZT investigation Hysteresis.pngStepResponse.png
I measured the HV coming to the PMC PZT by plugging it off from the PZT and hooking it up to a DVM.
The reading of DVM is pretty much consistent with the reading on EPICS. I got 287V on the DVM when the EPICS says 290V.
    Reply  Fri Jul 18 23:24:24 2008, rob, Update, PSL, PMC PZT investigation 
[quote]I measured the HV coming to the PMC PZT by plugging it off from the PZT and hooking it up to a DVM.
The reading of DVM is pretty much consistent with the reading on EPICS. I got 287V on the DVM when the EPICS says 290V.
       Reply  Sat Jul 19 19:39:44 2008, rob, Update, PSL, PMC PZT investigation 
[quote][quote] The 64k is the series resistor after the PA85, not the modeled resistance of the PZT itself.[/quote]
 Yes. What I meant was that because the measured voltage across the PZT was a half of the open voltage of the HV amp, the DC impedance of the PZT is expected
to be similar to the output impedance of the HV amp. Of course, I don't think the DC impedance of a normal PZT should be such low.
Entry  Sat Jul 19 19:41:56 2008, Yoichi, AoG, PSL, The author of the entry 702 is Yoichi not Rob 
I made a mistake.
ELOG V3.1.3-