ID |
Date |
Author |
Type |
Category |
Subject |
7936
|
Thu Jan 24 08:28:35 2013 |
Steve | Update | VAC | when to pumpdown ? |
Reminder: I will be on vacation next week. We would have to put the access connector in tonight if I'm pumping down Friday, tomorrow. |
7937
|
Thu Jan 24 11:31:45 2013 |
Steve | Update | SUS | ETMX damping restored |
ETMX sus damping restored. It is still noisy |
7938
|
Thu Jan 24 11:40:55 2013 |
Jenne | Update | SUS | ETMX damping restored |
Quote: |
ETMX sus damping restored. It is still noisy
|
I should have elogged, but I turned off the watchdog to remind myself that iscex computer is still crashed. "Turning on" the damping doesn't do anything since there aren't any signals going to the coils from the computer. |
7939
|
Thu Jan 24 14:40:09 2013 |
Steve | Update | VAC | RGA backgroung |
RGA background with VM2 open to Maglev at day 37
Note: The PAN gauge of the annulos is at atm. Please do not vent this 200 ft long annulos line when you venting the annulos of a chamber. The chamber annulos should be closed off to this long 2" OD. pipe before you vent the annulos of a chamber.
|
Attachment 1: Screenshot.png
|
|
7940
|
Thu Jan 24 15:16:50 2013 |
Manasa | Update | General | Laseroptik mirror - SN6 |
I repeated the transmittance measurements of Laseroptik SN6 at 1064nm. The rotation stage could only resolve 2 deg rotation (We should consider buying a better rotation stage).
s-polarization
Percentage transmittance
0.177% 42 deg
0.806% 44 deg
0.57% 46 deg
54.8% 0 deg
p-polarization
Percentage transmittance
1.039% 42deg
1.155% 44 deg
1.159% 46 deg
65.6% 0 deg
|
7941
|
Thu Jan 24 16:23:24 2013 |
Koji | Update | General | Laseroptik mirror - SN6 |
The mirror T is completely out of spec. We should find or request the data sheet of the mirror.
> We should consider buying a better rotation stage
I'm already on it |
7942
|
Thu Jan 24 16:31:46 2013 |
Steve | Update | PEM | building exterier wall painted |
The wood exteier walls, gutters and doors were painted at CES-Annex building #69 |
Attachment 1: IMG_1880.JPG
|
|
7943
|
Thu Jan 24 16:34:56 2013 |
rana | Update | General | Laseroptik mirror - SN6 |
I have two questions:
1) Are we sure that the T measurement is not being compromised by some systematic? i.e. some leakage is making the apparent T appear too high.
2) IF the T is really so high, how should we decide whether or not to use this one rather than the G&H? Is the 532 nm property more important than the high recycling gain? |
7944
|
Fri Jan 25 08:10:42 2013 |
Steve | Update | General | Laseroptik mirrors |
Here are the German plots. Unfortunately they are not logarithmic.
Ed: Proprietary data removed. Use wiki (Koji)
Steve uploaded data to the 40m wiki / Aux_Optics on 02-07-2013
|
7945
|
Mon Jan 28 17:01:19 2013 |
Den | Update | Locking | Video of PRM-flat test cavity |
What mode will you get if lock the cavity PRM - ITMY/ITMX/TEST MIRROR without PR2, PR3 and BS?
Is it possible to skip MC1, MC3 and lock the laser to this test cavity to make sure that this is not actuator/electronics noise? |
7946
|
Mon Jan 28 17:59:02 2013 |
Riju | Update | | Photodiode transimpedance |
Summary: Measurement and plot of shot-noise-intercept-current for PDA255.
Motivation:It is to measure the shot noise intercept current for PDA255 - the MC transmission RF photodiode to get an idea for the noise current for the detector
Result: The final plot is attached here. The plot suggests that the value of shot-noise-intercept current is 3.06mA
Discussion:
The plot is for the measured data of Noise voltage (V/sqrt(Hz)) vs DCcurrent(A). The fitted plot to this measured data follows the noise equation
Vnoise = gdet* sqrt[ 2e (iDC+idet)] , where gdet= transimpedance of the PD in RF region as described in manual of PDA255 (i.e. 5e3 when it is not in High-impedance region).
On the other hand for DCcurrent calculation we must use the high-impedance value for the transimpedance i.e. 1e4 Ohm. idet is the shot noise intercept current.
For the rough calculation of the noise level we may use the following formulae:
Vnoise = gdet*sqrt[2e (iDC+idet)] = gdet*sqrt(2e in), when in=iDC+idet;
For say, in1=1mA; Vnoise1=gdet*sqrt(2e *in1)
and sqrt(2e *in1)~18pA/sqrt(Hz)
In current case dark noise is ~1.5e-7 V/sqrt(Hz)
Therefore dark current(in2) ~dark noise voltage/RF transimpedance = 30pA/sqrt(Hz)
i.e. sqrt(2e *in2)=30pA/sqrt(Hz)
i.e. sqrt(in2/in1)=30/18
therefore, in2~3mA (since in1=1mA)
For, iDC=0, in=idet.
Therefore the shot-noise-intercept current will be ~3mA
Then Vdc = in2*1e4 = 30V
According to the experiment and also from the PDA255 manual the DC voltage level never goes beyond ~10V. Therefore following the photodiode characteristics(we work in reverse bias) we may infer that it can never become shot noise limited.
Also, from PDA255 manual, at 1650nm the dark noise is 30pW/sqrt(Hz) and the responsivity is 0.9A/W. Therefore the noise current level will be = noise power* responsivity ~27pA/sqrt(Hz). The value matches well with our expectation.
|
Attachment 1: shotnoiseinterceptpda255.pdf
|
|
7947
|
Mon Jan 28 19:07:45 2013 |
Manasa | Update | General | SN6 Laseroptik mirror - Tranmittance measurements |
I repeated the measurements using NPRO instead of Crystalaser. I am attaching optical layouts for these measurements for future reference.
Lesson learnt : Do not use Crystalaser for transmission measurements and always separate the transmitted main beam from other beams that result from the wedged surface of the mirror.

Measurements match specs provided by Laseroptik
p-polarization
T percentage = 0.10% 42 deg
0.092% 44 deg
0.086% 46 deg
Minimum transmittance = 0.081% 52deg
s-polarization
T percentage = 0.048% 42 deg
0.047% 44 deg
0.047% 46 deg
Minimum transmittance = 0.047% 46 deg

|
7948
|
Mon Jan 28 19:15:14 2013 |
Manasa | Update | Scattering | Scattering setup |
[Jan, Manasa]
We are trying to get some scattering measurements in the Y-arm cavity. We have removed one of the viewport windows window covers of ETMY chamber and have installed cameras on a ring that clamps to the window. The window along with the ring attachment is covered with aluminium foil when not in use. |
7949
|
Mon Jan 28 21:32:38 2013 |
jamie | Update | Alignment | tweaking of alignment into half PRC |
[Koji, Jamie]
We tweaked up the alignment of the half PRC a bit. Koji started by looking at the REFL and POP DC powers as a function of TT2 and PRM alignment.
He found that the reflected beam for good PRC transmission was not well overlapped at REFL. When the beam was well overlapped at REFL, there was clipping in the REFL path on the AS table.
We started by getting good overlap at REFL, and then went to the AS table to tweak up all the beams on the REFL pds and cameras.
This made the unlocked REFL DC about 40 count. This was about 10mV (=0.2mA) at the REFL55 PD.
This amazed Koji since we found the REFL DC (of the day) of 160 as the maximum of the day for a particular combination of the PRM Pitch and TT2 Pitch. So something wrong could be somewhere.
We then moved to the ITMX table where we cleaned up the POP path. We noticed that the lens in the POP path is a little slow, so the beam is too big on the POP PD and on the POP camera (and on the camera pick-off mirror as well).
We moved the currently unused POP55 and POP22/110 RFPDs out of the way so we could move the POP RF PD and camera back closer to the focus. Things are better, but we still need to get a better focus, particularly on the POP PD.
We found two irides on the oplev path. They are too big and one of these is too close to the POP beam. Since it does not make sense too to have two irides in vicinity, we pulled out that one from the post.
Other things we noticed:
- The POP beam is definitely clipping in the vacuum, looks like on two sides.
- We can probably get better layout on the POP table, so we're not hitting mirrors at oblique angles and can get beams on grid paths.
After the alignment work on the tables, we started locking the cavity. We already saw the improvement of the POPDC power from 1000 cnt to 2500 cnt without any realignment.
Once PRM is tweaked a little (0.01ish for pitch and yaw), the maximum POPDC of 6000 was achieved. But still the POP camera shows non-gaussian shape of the beam and the Faraday camera shows bright
scattering of the beam. It seems that the scattering at the Faraday is not from the main beam but the halo leaking from the cavity (i.e. unlocking of the cavity made the scattering disappeared)
Tomorrow Jenne and I will go into BS to tweak the alignment of the TEMP PRC flat mirror, and into ITMX to see if we can clean up the POP path. |
7951
|
Tue Jan 29 10:50:02 2013 |
Jenne | Update | Locking | Video of PRM-flat test cavity |
I think Den accidentally edited and overwrote my entry, rather than replying, so I'm going to recreate it from memory:
I aligned the PRM-flat test cavity (although not as well as Jamie and Koji did later in the evening) and took some videos. Note that these may not be as relevant any more, since Jamie and Koji improved things after I left.
Also, before doing anything with the cavity, I tuned up the PMC since the pitch input alignment wasn't perfect (we were getting ~0.7 transmission), and also tuned up the MC alignment and remeasured the MC spot positions, to maintain a record. |
7953
|
Tue Jan 29 14:20:02 2013 |
Koji | Update | General | Finer rotation stage for optics characterization |
A rotation stage has been ordered.
Newport Rotation Stage, 360° Coarse, 5° Fine Rotation, Micrometer |
Newport |
481-A |
Newport Solid Insert for RSP-1T Rotation Stage |
Newport |
RSA-1TI |
Newport Universal Mounting Plate, 2.56 in. x 2.56 in. x 0.5 in., 1/4-20 Thread |
Newport |
UP-1A |
Specification: Newport 481-A
- Sensitivity: 15 arcsec
- Graduations: 1 deg
- Vernier: 5 arcmin
- Fine travel range: 5 deg
- With Micrometer
|
7954
|
Tue Jan 29 14:34:42 2013 |
Jenne | Update | PEM | Secret Seis filters |
The BLRMS have been bad again, since the computer crash of last week. Finally getting around to looking into it, I discovered that there are filter banks that have the microns/sec calibration filters, which are not accessible from the sitemap. I have added links to them for GUR1 and GUR2. We need to make the PEM/BLRMS screens macro-expansion-y, so that I don't have to change each screen individually.
Anyhow, the BLRMS are back.
|
7956
|
Tue Jan 29 18:40:20 2013 |
Riju | Update | | Photodiode transimpedance |
Today I have taken data for shot noise intercept current for PDA10CF. I will process the data and report.
Note: GPIB address changed, new command for AG4395A network/spectrum analyzer: ./netgpibdata.py -i 192.168.113.108 -d AG4395A -a 10 -f filename |
7957
|
Tue Jan 29 19:50:49 2013 |
Jenne | Update | Locking | Better POP layout, no extra PRM motion with locked cavity |
[Jenne, Jamie, Manasa]
Today's activities focused on getting the POP layout improved, so that we could get clean data for the mode scan measurement.
As Jamie and Koji pointed out yesterday, the beam was still a little too big on the POP DC PD, and was falling off the diode when the beam moved a small amount. We have fixed things so that the PD is now at the focus of the lens, and the camera is at a place where the beam takes up most of the area on the TVs. The beam no longer falls off the PD with cavity fluctuations. A key point of this work was also to use an extra 2" optic to steer the beam down the length of the POP table, and then do the 50/50 beam splitting later with a 1" optic. The 1" BS that we had been using (including with the "real" POP beam) is too small. We could not find a 2" 50/50 BS, so we opted to do the splitting closer to the focal point. Also, the BS that was splitting the beam between the PD and the camera was a 33% reflector, but now is a 50/50 BS. When we put back the 'real' POP path, we need to consider using larger optics, or a faster lens. The POP path is now good, hopefully for the duration of the half cavity test.
After getting the POP path taken care of, and tweaking up the cavity alignment a little bit, the transmitted power on POP DC is ~22,000 counts, with occasional fluctuations as high as 25,000 counts.
Jamie looked at the REFL path, and things look sensible there. The unlocked REFL power is ~36 counts, and the locked power is ~20 counts. I'm not sure what the 160 counts that Koji mentioned in his edits to elog 7949 is about.
I looked at the PRM oplev with the cavity locked and unlocked, and with today's alignment, there seems to be no difference in the amount of PRM motion when the cavity is locked vs unlocked.

It still looks like we might be seeing some clipping in the in-vac POP steering mirrors - we haven't gotten to them yet.
Jamie is currently modifying Yuta's mode scan analysis script to look at the data that we have of the cavity.
We need more 2" optics. There are no mounted 2" spares in the various optic "graveyards" (which, PS, we should consolidate all into the cabinet with doors near the optics bench), and the options for boxes in the drawers is slim pickin's. We have some S-pol stuff, but no Y1s or BS-50s for P-pol. Since POP, POX, POY, IPANG, TRX and TRY all come out of the vacuum with large beams, we should have some options for these laying around for this kind occasional temporary thing. We also need to choose, then purchase better 2" lenses for the pickoffs. |
Attachment 1: HalfPRCL_PRM-flatMirror_RefsAreLocked_OthersUnlocked.pdf
|
|
7958
|
Tue Jan 29 20:28:11 2013 |
ericq | Update | General | Early work on Mirror Mounts |
[Q, Chloe]
Chloe has been to the lab twice to start up her investigations in acoustic noise coupling to mirrors. The general idea for the setup is a HeNe laser bouncing off a mirror and onto a QPD, whose signal provides a measure of beam displacement noise. The mirror will be mounted and excited in various ways to make quantitative conclusions about the quality of different mounting schemes.
We have set up the laser+mirror+QPD on the SP table, and collected data via SR560s->SR785, with the main aim of evaluating the suitability of this setup. The data we collected is not calibrated to any meaningful units (yet). For now, we are just using QPD volts.
Chloe collected data of vertical displacement noise for the following schemes: Terminated SR785 input, Terminated SR560 inputs, Laser centered directly onto the QPD, Laser shining on mirror centered on QPD, laser/mirror/qpd with some small desktop speakers producing white noise from http://www.simplynoise.com. Data shown below.

|
7959
|
Tue Jan 29 21:07:48 2013 |
Jenne | Update | Locking | PRM coils need diagonalizing |
[Jenne, Jamie]
We tried actuating on PRM so that we go through fringes in a known, linear way. We used C1:SUS-PRM_LSC_EXC and awggui. It seems that we get a lot of angular motion when we actuate....we need to look into this tomorrow.
EDIT/UPDATE: Last night we tried several combinations of frequency and amplitude, but just for an idea, we were using 2Hz, 1000cts. Using Kiwamu's calibration in elog 5583 for the PRM actuator of 2e-8/f^2 m/cts, this means that we were pushing ~5nm. But when we pushed much harder (larger amplitude) than that, we saw angular fringing. |
7960
|
Wed Jan 30 03:01:55 2013 |
Koji | Update | General | Early work on Mirror Mounts |
I can't believe that SR785 can have such a low input noise level (<1nV/rtHz). Review your calibration again.
It is also described in the manual that SR560 typically has the input noise level of 4nV/rtHz, although this number depends on which gain you use. |
7961
|
Wed Jan 30 11:16:32 2013 |
Jenne | Update | Locking | Mode spacing calc |
[Jenne, Jamie]
We did a few pen and paper calculations yesterday to confirm for ourselves that the half PRC should have nicely separated modes. The half cavity is L=4.34m long, assuming flat mirror is 3.5 inches in front of BS. That 3.5" is a guess, not a measurement.
Finesse
F = ( pi * sqrt(r1 * r2) ) / (1 - r1*r2) = 111.
Full width at half max
FWHM = c / (2 * L * F) = 311 kHz
FWHM in meters = FWHM * L/f = L*1064nm/c = 4.8 nm
Free spectral range
nu_fsr = F * FWHM = 34.5 MHz
Mode Spacing (eq 19.23 from Siegman)
omega = (n + m) * arccos(\pm sqrt(g1*g2)) / pi * (2*pi*c)/(2L)
For our half cavity, g1*g2 = 0.96
For the 01 or 10 modes, n+m = 1
omega = 13.7e6 rad/sec
mode spacing between 00 and 01 = 2.2 MHz
Thus, the modes should be well separated
=> spacing is 2.2 MHz while FWHM is 0.311 MHz (cavity fsr = 34.5 MHz)
EDIT JCD 31Jan2013: Fixed mode spacing eqn to be diff between TEM00 mode and HOM, not plane wave and HOM. Then fixed the factor of 2 error in the mode spacing numbers. |
7962
|
Wed Jan 30 11:18:31 2013 |
Manasa | Update | Scattering | Scattering setup |
Quote: |
[Jan, Manasa]
We are trying to get some scattering measurements in the Y-arm cavity. We have removed one of the viewport windows window covers of ETMY chamber and have installed cameras on a ring that clamps to the window. The window along with the ring attachment is covered with aluminium foil when not in use.
|
[Jan, Manasa]
To align the camera to see small angle scattering from the ITMY, we tried shooting a green laser pointer at the pickoff mirror that was installed in the ETMY chamber such that we hit the face of ITMY. But we concluded that to be a very bad way to align the camera because we have no means to reconfirm that the camera was exactly looking at the scattering from ITMY.
Since we are in air, we came up with a plan B. The plan is to temporarily install a mirror in the ITMY chamber to steer the beam from the laser pointer (installed on the POY table) through ITMY to the pickoff mirror at the ETMY end. This way, we can install the camera at the ETMY window and be sure we are looking at ITMY scattered light. |
7963
|
Wed Jan 30 13:50:27 2013 |
Jenne | Update | Computers | c1iscex still down |
[Koji, Jenne]
We noticed that the iscex computer is still down, but the IOP is (was) running. When we sat down to look at it, c1x01 was 'breathing', had a non-zero CPU_METER time, and the error was 0x4000, which I've never seen before. The fb connection was still red though. Also, it is claiming that its sync source is 1pps, not TDS like it usually is.
Since things were different, Koji restarted the 2 other models running on iscex, with no resulting change. We then did a 'rtcds restart all', and the IOP is no longer breathing, and the error message has changed to 0xbad. The sync source is still 1pps.
Moral of the story: c1iscex is still down, but temporarily showed signs of life that we wanted to record. |
7964
|
Wed Jan 30 14:00:02 2013 |
Charles | Update | ISS | ISS Design and Prototyping |
Attached are both the circuit diagram and the liso formatted *.fil for the main branch of the ISS, as well as the resulting transfer function when analyzed. Unfortunately, as noted in the file, not all of the elements are possible to analyze in liso, such as any type of op-amp with more than two inputs and one output (AD602 used in this design has 16 pins with two distinct amplifiers contained within).
I have begun prototyping this circuit on a breadboard. |
Attachment 1: ISS.fil
|
## ISS Main Branch
##
## All circuit elements are named according to the circuit diagram
## "D020241-D2.pdf" by R. Abbott.
# Stages are separated by empty lines and elements between stages are
# also separated by empty lines for easy file navigation
# Before the first stage there is a 'fully differentiable' op-amp
# that I believe serves to isolate the device from the power supply
# However, liso does not have the capability to analyze such an op-amp,
... 79 more lines ...
|
Attachment 2: ISS_Transfer_Function.png
|
|
Attachment 3: D020241-D2.pdf
|
|
7965
|
Wed Jan 30 14:37:01 2013 |
Zach | Update | ISS | ISS Design and Prototyping |
Quote: |
Unfortunately, as noted in the file, not all of the elements are possible to analyze in liso, such as any type of op-amp with more than two inputs and one output (AD602 used in this design has 16 pins with two distinct amplifiers contained within).
|
Typically, you can still find a way to model the important parts of the stages that are not as simply added. In the case of the differential input stage, in particular, it is important to include it because it will usually set the input noise level of the circuit. In this case, the noise is the same as the second stage (U5) and it has a gain of 1, so there is essentially no difference (up to factors of sqrt(2) or 2).
You can edit the opamp.lib file and add in custom components. For the input stage, you can just pretend it is a simple non-inverting amplifier with the specified noise characteristics from the datasheet: un = 1.3n, uc = 50 Hz (see below).
For dual op amps, you can usually just model each part separately. For example, the OPA2604 is a dual op amp that is included in the opamp.lib and can be treated as a single one in a model.

|
7966
|
Wed Jan 30 15:45:09 2013 |
Zach | Update | Locking | Mode spacing calc |
Quote: |
Thus, the modes should be well separated
=> spacing is 4.3 MHz while FWHM is 0.311 MHz (cavity fsr = 34.5 MHz)
|
Something looks fishy. I calculate a transverse mode spacing of 2.21 MHz---is there a factor of two missing somewhere in your analytical calculation?
delta_f = (1/2/pi) * w01 - w00 = (1/2/pi) * acos(±sqrt(0.96)) /pi *2 * pi * c /2 /L = 2.21 MHz
I guess that's still OK, but if you are using 11-MHz sidebands, there is a n+m=5 mode within one linewidth of resonance. Can you use 55?
-------------
May I suggest my arbcav() tool for things like this? I think it's pretty handy for just this sort of calculations. I'm actually hoping to revamp the I/O to make it much cleaner and more intuitive.
>> T = [0.055 20e-6];
>> L = [4.34 4.34];
>> RoC = [115.5 1e10];
>> theta = [0 0];
>> fmod = 11e6;
>> lambda = 1064e-9;
>> num_pts = 1000;
>> loss = 50e-6;
>> [fin,coefs,df] = arbcav(T,L,RoC,theta,fmod,loss,lambda,num_pts);
>> fmod = 55e6;
>> [fin,coefs,df] = arbcav(T,L,RoC,theta,fmod,loss,lambda,num_pts);

|
7968
|
Wed Jan 30 19:30:21 2013 |
Jenne | Update | Locking | POP in-vac improved |
Quote: |
It still looks like we might be seeing some clipping in the in-vac POP steering mirrors - we haven't gotten to them yet.
|
[Jenne, Jamie]
We fixed up, as best we can, the in-vac POP alignment. We are entirely limited in yaw by the aperture size of the 2" 45deg mirror launching the beam out of the vacuum. The main centroid of the beam is well centered, but the inflated weird part of the beam is totally clipped. There's nothing we can do about it except use a much larger mirror, install a fast lens inside the chamber, or just fix the damn PRC. I vote for the third option there.
How did we work our magic?
We put a green laser pointer where the POP DC PD was, and injected it into the vacuum, just like we normally do. However, this time, we made sure the green laser was centered on all of the out of vacuum mirrors, so that there was no real work to do once we turned off the laser pointer. We locked the cavity, and confirmed that we are well centered on all of the in and out of vacuum mirrors, and discovered our aperture problem with the last in-vac mirror.
Here is a snapshot of the POP camera:

|
7969
|
Wed Jan 30 19:34:17 2013 |
Jenne | Update | Locking | PRM coils need diagonalizing |
[Koji, Jamie, Jenne]
Koji did this, while we actuated on PRM in pos, and watched the oplev. Empirically, he found the following values for the POS column of the output matrix:
UL = 1.020
UR = 0.990
LL = 1.000
LR = 0.970
SD = 0.000
(The nominal values are all +1, except for Side, which is 0).
Actuation of PRM was through C1:SUS-PRM_LSC_EXC, f=0.1Hz, A=100 counts.
Ed by KA:
This means UL and UR are increased by 2% and UR and LR are decreased by 3%. More precisely UR should be 1.02*0.97.
This is just a quick hack which works only for the DC. |
7970
|
Thu Jan 31 10:23:39 2013 |
Jamie | Update | Computers | c1iscex still down |
Quote: |
[Koji, Jenne]
We noticed that the iscex computer is still down, but the IOP is (was) running. When we sat down to look at it, c1x01 was 'breathing', had a non-zero CPU_METER time, and the error was 0x4000, which I've never seen before. The fb connection was still red though. Also, it is claiming that its sync source is 1pps, not TDS like it usually is.
Since things were different, Koji restarted the 2 other models running on iscex, with no resulting change. We then did a 'rtcds restart all', and the IOP is no longer breathing, and the error message has changed to 0xbad. The sync source is still 1pps.
Moral of the story: c1iscex is still down, but temporarily showed signs of life that we wanted to record.
|
There's definitely a timing issue with this machine. I looked at it a bit yesterday. I'll try to get to it by the end of the week. |
7971
|
Thu Jan 31 11:53:31 2013 |
Manasa | Update | Scattering | Scattering setup |
Since we are in air, we came up with a plan B. The plan is to temporarily install a mirror in the ITMY chamber to steer the beam from the laser pointer (installed on the POY table) through ITMY to the pickoff mirror at the ETMY end. This way, we can install the camera at the ETMY window and be sure we are looking at ITMY scattered light.
|
[Jan,Manasa]
We executed plan B. We installed the green laser pointer on POY table and steered the beam through ITMY to hit the pick off mirror at the ETM end by installing *temporary mirrors. The pick off mirror was adjusted in pitch and yaw to center the reflected beam on the viewport window. We have installed irides on the ring attached to the viewport window to direct the beam to the camera.
*Temporary mirrors were removed from the ITMY chamber after this alignment. |
7972
|
Thu Jan 31 12:44:42 2013 |
Riju | Update | | Photodiode transimpedance |
Today I collected the data for shot noise intercept current for MC REFL PD. I didn't get many data points at higher DC voltage of the photodiode, cause the incandescent bulbs get burnt at that level; two bulbs I have burnt today. I will process the data and report. |
7973
|
Thu Jan 31 14:30:50 2013 |
Jenne | Update | Locking | Mode spacing calc |
I have calculated (using Zach's sweet software) the expected mode content for the various possible PRCs that we can make.
Also, Zach was right about the factor of 2. I see now that I was calculating the mode spacing between a plane wave and a HOM, so the guoy phase had a factor of (n+m+1). The right thing to do is to get the spacing between the 00 mode and HOMs, so the guoy phase just has (n+m). Switching from n+m+1=2 to n+m=1, that fixes the factor of 2 problem.
I attach my results as a pdf, since I'm listing out 5 configurations. Each config has a cartoon, with a small (hard to read) HOM plot, and then at the end, each HOM plot is shown again, but larger. Also, "TM" is the "test mirror", the flat G&H that we're using as the cavity end mirror. |
Attachment 1: Half_PRC_Configs.pdf
|
|
7974
|
Thu Jan 31 14:46:05 2013 |
Jenne | Update | RF System | Photodiode transimpedance |
Quote: |
Today I collected the data for shot noise intercept current for MC REFL PD. I didn't get many data points at higher DC voltage of the photodiode, cause the incandescent bulbs get burnt at that level; two bulbs I have burnt today. I will process the data and report.
|
This work was done in-situ, so no optics on the AS table were moved. The PSL shutter was blocked since the IR beam was not necessary, and would scatter off the bulb Riju put in front of the PD. |
7975
|
Thu Jan 31 15:20:46 2013 |
Zach | Update | Locking | Mode spacing calc |
I should mention that I just found a bug in how it treats odd-mirror-number cavities. For such cavities, HG modes with odd horizontal indices should receive an extra roundtrip phase of pi/2 (due to the rotation by the cavity). Because of a numbering convention issue, arbcav actually used to apply this phase shift to even-order modes. Essentially, the only difference is that the fundamental mode was shifted to anti-resonance. Everywhere else, there are modes at both corresponding locations in frequency space, and so it does not back a big difference in terms of cavity design.
Thanks to this IMC modeling we are doing at the workshop, I caught it! It has been fixed in the SVN.
Quote: |
I have calculated (using Zach's sweet software) the expected mode content for the various possible PRCs that we can make.
Also, Zach was right about the factor of 2. I see now that I was calculating the mode spacing between a plane wave and a HOM, so the guoy phase had a factor of (n+m+1). The right thing to do is to get the spacing between the 00 mode and HOMs, so the guoy phase just has (n+m). Switching from n+m+1=2 to n+m=1, that fixes the factor of 2 problem.
I attach my results as a pdf, since I'm listing out 5 configurations. Each config has a cartoon, with a small (hard to read) HOM plot, and then at the end, each HOM plot is shown again, but larger. Also, "TM" is the "test mirror", the flat G&H that we're using as the cavity end mirror.
|
|
7976
|
Thu Jan 31 15:34:22 2013 |
Riju | Update | RF System | Photodiode transimpedance |
Quote: |
Quote: |
Today I collected the data for shot noise intercept current for MC REFL PD. I didn't get many data points at higher DC voltage of the photodiode, cause the incandescent bulbs get burnt at that level; two bulbs I have burnt today. I will process the data and report.
|
This work was done in-situ, so no optics on the AS table were moved. The PSL shutter was blocked since the IR beam was not necessary, and would scatter off the bulb Riju put in front of the PD.
|
Thanks Jenne. |
7977
|
Thu Jan 31 15:56:38 2013 |
Riju | Update | | Photodiode transimpedance |
Summary: Measurement and plot of shot-noise-intercept-current for PDA10CF.
Motivation:It is to measure the shot noise intercept current for PDA10CF.
Result: The final plot is attached here. The plot suggests that the value of shot-noise-intercept current is 0.21mA
Discussion:
The plot is for the measured data of Noise voltage (V/sqrt(Hz)) vs DCcurrent(A). The fitted plot to this measured data follows the noise equation
Vnoise = gdet* sqrt[ 2e (iDC+idet)] , where gdet= transimpedance of the PD in RF region as described in manual of PDA255 (i.e. 5e3 when it is not in High-impedance region).
To get an approximate idea of the shot noise intercept current, we may follow the same procedure described in 7946
In the present case dark-noise is 4.3e-08 V/sqrt(Hz)
Therefore dark current(in2) ~dark noise voltage/RF transimpedance = 8.6pA/sqrt(Hz)
Therefore the approximate shot noise intercept current ~ (8.6/18)^2=0.22mA
This value matches well with the fitted data.
From PDA10CF manual, NEP=1.2e-11W/sqrt(Hz) and responsivity~0.9A/W. Therefore the noise current level will be ~10pA.
|
Attachment 1: shotnoiseinterceptpda10cf.pdf
|
|
7978
|
Thu Jan 31 20:06:22 2013 |
Evan | Update | Locking | PRM/PR2 cavity |
[Jenne, Evan]
Tonight we made a non-folded cavity between the PRM and PR2 as follows. I put down two dog clamps to constrain the original position of the PR2 mount. I then loosened the dog clamps holding the mount to the table and nudged the mount until we saw a few reasonably well-aligned bounces in the cavity. I then dogged down the mount.
We played with the PRM and TT2 steering until we saw flashes of TEM00. However, the resonance is not clean so we couldn't lock.
Since we changed the PRM alignment, we had to redo the last bit of steering for the PRM oplev into the photodiode. We also put a few ND filters on the POP camera.
|
7980
|
Thu Jan 31 23:48:45 2013 |
Koji | Update | Locking | PRM/PR2 cavity |
Wow! What's happened?
As the video showed good quality of resonances, I stopped by at the 40m on the way back home.
I looked at the error signals and found that they indicate high finesse and clear resonance of the sidebands.
The lock was immediate once the gain is set to be -0.004 (previous 0.05ish). This implies the optical gain is ~10 times larger than the previous configration.
The alignment was not easy as POPDC was saturated at ~27000. I leave this as a daytime job.
As I misaligned the PRM, I could see that the lock hopped into the next higher order. i.e .from TEM00 to TEM01, from TEM01 to TEM02, etc
This means that the modes are closely located each other, but sufficiently separated to sustain each mode.
I definitely certify that cavity scans will give us meaningful information about the cavity. |
7981
|
Fri Feb 1 09:33:11 2013 |
Jamie | Update | Locking | PRM/PR2 cavity |
I replaced the BS1 between the POPDC PD and the camera with a 98 reflector, and moved the 50 up before the BS to dump half the light. Still saturating POPDC, but hopefully the ratio between POPDC and the camera should be better. We just need to dump more of the power before we get there. I'll come back to this after C&D if no one else has already gotten to it.
I don't know why I didn't pay more attention last night, but things look way WAY better. The beams are much cleaner and the power level is much much higher. |
7983
|
Fri Feb 1 12:34:55 2013 |
Jenne | Update | PSL | Too much power injected into vacuum |
I noticed (while relocking the MC after Jamie and I zeroed the LSC offsets) that the MC refl power was 4.8 V. Usually we should be ~4.2, so I closed the PSL shutter and went in to measure the power. We were injecting ~125mW or a little more. I had adjusted the power the other day, and through yesterday, it looked fine, but overnight it looks like it drifted up. |
7984
|
Fri Feb 1 14:47:17 2013 |
Riju | Update | | Photodiode transimpedance |
Summary: Measurement and plot of shot-noise-intercept-current for MC REFL PD.
Motivation:It is to measure the shot noise intercept current for MC REFL PD.
Result: The final plot is attached here. The plot suggests that the value of shot-noise-intercept current is 0.041mA
Discussion:
The plot is for the measured data of Noise voltage (V/sqrt(Hz)) vs DCcurrent(A). The fitted plot to this measured data follows the noise equation
Vnoise = gdet* sqrt[ 2e (iDC+idet)] , where gdet= transimpedance of the PD in RF region as described in manual of PDA255 (i.e. 5e3 when it is not in High-impedance region).
To get an approximate idea of the shot noise intercept current, we may follow the same procedure described in 7946
In the present case minimum noise value is 2.03e-08 V/sqrt(Hz)
Therefore dark current(in2) ~dark noise voltage/RF transimpedance = 4.06pA/sqrt(Hz)
Therefore the approximate shot noise intercept current value is (4/18)^2 ~ 0.049mA, which is close to the fitted value.
... hard to believe these numbers. Wrong DC transimpedance? (KA) |
Attachment 1: shotnoiseinterceptmcreflpd.pdf
|
|
7985
|
Fri Feb 1 15:12:53 2013 |
Jenne | Update | Locking | PRM/PR2 cavity |
After Jamie did all the work this morning on the POP table, I was able to get the cavity to lock. It's not very stable until I engage the boost filters in the PRCL loop. After locking, I tuned up the alignment a bit more. Now we're taking mode scan data. Look for results hopefully shortly after Journal Club! |
7986
|
Fri Feb 1 19:55:33 2013 |
Jenne | Update | Locking | PRM/PR2 cavity |
[Jamie, Koji, Jenne]
We are looking at the mode scan data, and have some preliminary results! We have data from when the cavity was aligned, when it was slightly misaligned in pitch, and slightly misaligned in yaw.
Inverting the equation for transverse mode spacing, we infer (for pitch misalignment) a cavity g-factor of 0.99, and from there (assuming the G&H mirror is flat and so has a g-factor of 1), we infer a PRM radius of curvature of 168 meters which is ~50% longer than we expected.
More results to come over the weekend from Jamie. |
7987
|
Fri Feb 1 23:12:42 2013 |
Koji | Update | Locking | PRM/PR2 cavity |
During the scanning we were riddled by the fact the PDH error and the transmission peaks do not happen simultaneously.
After a little investigation, it was found that "LP100^2" filter is left on in the POPDC filter.
Moreover, it was also found that the whitening filter switches for the POPDC does not switch the analog counterpart.
These were the culprit why we never saw accidental hitting of the max transmission by the peaks when the cavity was not locked.
I know that the most of the whitening filter in the RF paths were checked before (by Keiko?), but the similar failure still exists in the POX path.
We should check for the whitening filters in the DC path as well and fix everything at once. I can offer assistance on the fixing part. |
7988
|
Fri Feb 1 23:52:59 2013 |
rana | Update | Locking | PRM/PR2 cavity |
Very exciting result, if true. I suppose we should try to reconfirm this result by doing another phase map of PRM03.
Is it possible that PR2 is not flat? How would we test to see if the tip-tilt frame screw gives it a curvature? Perhaps we can check with COMSOL. |
7993
|
Mon Feb 4 15:26:10 2013 |
Jamie | Update | Computer Scripts / Programs | New "getdata" program to pull NDS channel data, including test points |
I've added a new program called getdata (to scripts/general/getdata) that will conveniently pull arbitrary data from an NDS server, either DQ or online (ie. testpoints).
Start times and durations may be specified. If past data is requested, you must of course be requesting DQ channels. If no start time is specified, data will be pulled "online", in which case you can specify testpoints.
If an output directory is specified, the retrieved data will be stored in that directory in files named after the channels. If an output directory is not specified, no output will be
Help usage:
controls@pianosa:~ 0$ /opt/rtcds/caltech/c1/scripts/general/getdata --help
usage: getdata [-h] [-s START] [-d DURATION] [-o OUTDIR] channel [channel ...]
Pull online or DQ data from an NDS server. Use NDSSERVER environment variable
to specify host:port.
positional arguments:
channel Acquisition channel. Multiple channels may be
specified acquired at once.
optional arguments:
-h, --help show this help message and exit
-s START, --start START
GPS start time. If omitted, online data will be
fetched. When specified must also specify duration.
-d DURATION, --duration DURATION
Length of data to acquire.
-o OUTDIR, --outdir OUTDIR
Output directory. Data from each channel stored as
'.txt'. Any existing data files will be
automatically overwritten.
controls@pianosa:~ 0$
|
7999
|
Tue Feb 5 09:08:11 2013 |
Steve | Update | PEM | exterier doors to be painted |
Quote: |
The wood exteier walls, gutters and doors were painted at CES-Annex building #69
|
The east and south end of the 40m emergency exit doors are sealed- tapped off temporarily. They will be painted on the out side only. This job will be done by tomorrow noon
Do not open chamber if you smell the paint ! |
Attachment 1: IMG_0055.JPG
|
|
8001
|
Tue Feb 5 10:18:54 2013 |
Steve | Update | PEM | high particle count ALART |
Quote: |
Quote: |
The BS camber is open only. We should close ASAP
Outside air quality is 1.7- 2.2 million particles / cf min of 0.5 micron
|
Air is still bad and the chambers are closed. Before lunch Jamie repointed the PRM oplev. Manasa and I reset oplevs: BS and ITMX.
ETMX and ETMY are fine.
SRM and ITMY oplevs needs more work.
|
The bad outside air quality is pushing up the inside counts.
The outside air is 5 million counts / cf min for 0.3 micron and 2 million counts / cf min for 0.5 micron particles
Do not open chamber over 10,000 counts / cf min of 0.5 micron |
Attachment 1: badair.png
|
|
Attachment 2: 100d_airq.png
|
|