40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 1 of 346  Not logged in ELOG logo
IDdown Date Author Type Category Subject Text Attachments
  17432   Fri Jan 27 21:22:57 2023 yutaSummaryBHDV beam dump installed for BH44 RF PDOne of these V beam dumps was installed
for BH44 RF PD.
The rest is now stored in the box
 V.JPG 
  17431   Fri Jan 27 19:35:31 2023 yutaSummaryBHD60 Hz noise investigations around IMCSo far different measurements are consistent
with the hypothesis that the 60 Hz noise
is from PSL frequency noise (40m/17423).
 6x 
  17429   Fri Jan 27 11:26:31 2023 ranaUpdatePEMHEPA Monitor setup using WFS BLRMSthis is very exciting! Its the beginning
of the task of reducing vast amounts of PEM
data into human-useful (bite sized) info.
  
  17428   Thu Jan 26 23:33:19 2023 KojiUpdateLSCEP30-2 Epoxy bonding for Yuta@OMC Lab
EP30-2 Epoxy bonding of V beam dumps
for LSC PDs.
 PXL_20230127_053113856.jpgPXL_20230127_053145432.jpg 
  17427   Thu Jan 26 18:20:43 2023 AnchalUpdatePEMHEPA Monitor setup using WFS BLRMSI implemented the BLRMS on WFS1/2_I_PIT/YAW
outputs and using there value, a HEPA state
can be defined. I've currently set it
 HEPA_MONITOR.png 
  17426   Thu Jan 26 16:07:05 2023 yutaSummaryPSLPMC aligned, now PMC transmission is 0.7PMC aligned (Attachment #1).
Over the past month, PMC transmission
is actually slowly growing from 0.68 to 0.70
 Screenshot_2023-01-26_16-10-20.pngScreenshot_2023-01-26_16-10-35.png 
  17425   Thu Jan 26 15:56:30 2023 AnchalUpdateASC1X1 -5V sorenson tripped[Yuta, Anchal]




  
  17424   Thu Jan 26 00:18:54 2023 KojiUpdateCamerasRecording CCD camerasConnect any video signal supported by the
adapter. Use Windows / Mac or any other OS.
If it keeps complaining, contact Magwell
  
  17423   Wed Jan 25 18:01:21 2023 yutaSummaryBHDBH44 setup isolated, but 60 Hz noise stays the same[Paco, Anchal, Yuta]

Isolating
BH44 setup from the rest didn't help
 BH44BH55Diagram_Disconnected.pdf20230125_FPMIBHDPDs_Spectra_BothArmsLocked.pdf20230125_FPMIBHDPDs_Spectra_BothArmsLockedBH44Removed.pdf20230125_MCF.pdf 
  17422   Wed Jan 25 16:58:19 2023 AlexUpdateCamerasRecording CCD camerasThus far, the software needed for the Magewell
video encoder has been successfully installed
on Donatella. OBS studio has also been installed
  
  17421   Wed Jan 25 14:33:07 2023 ranaSummaryBHDREFL55 visually inspected, BH44 Kapton tapedKapton tape is a good insulator, so its
a good block for 60 Hz. But it is mostly
useless for RF since the capacitance between
  
  17420   Wed Jan 25 12:49:14 2023 RadhikaUpdateALSXARM green laser lock debugging   XEND_PDHservo_boost_on_gain2.pdfXEND_PDHservo_boost_on_gain4.pdfXEND_PDHservo_boost_on_gain6.pdfXEND_PDHservo_boost_on_gain8.pdfXEND_PDHservo_boost_on_gain10.pdf 
  17419   Wed Jan 25 11:35:20 2023 yutaSummaryBHDREFL55 visually inspected, BH44 Kapton tapedFollowing was done to investigate 60 Hz
noise issue, but no significant change in
the FPMI noise observed.
 REFL55.JPGBH44Kapton.JPG20230125_FPMIBHDPDs_Spectra_FPMILockedBH55Locked.pdf20230125_FPMIBHDPDs_Spectra_FPMILockedBH55LockedNoWhitening.pdf20230125_FPMIBHDPDs_Spectra_FPMILockedBH55LockedResG60InCARM.pdf 
  17418   Wed Jan 25 10:02:47 2023 yehonathanUpdate Earthquake, MC3 watchdog trippedWe came this morning and noticed the FSS_FAST
channel was moving very rapidly. Short inspection
showed that MC3 watchdog got tripped. After
  
  17417   Tue Jan 24 21:29:31 2023 yutaSummaryBHDElectronics diagram around BH44 and BH551) Turning the whitening filter before
the ADC on/off didn't changed the relative
height of 60 Hz peak compared with the
 BH44BH55Diagram.pdf 
  17416   Tue Jan 24 21:04:59 2023 AnchalUpdateASCAS WFS path beam profiledI completed the mode matching calculation
today and found good solution with 360.6
mm ROC PLCX lens at -1.2 m from z=0 point.
 ASBeamFocusingLens.png 
  17415   Tue Jan 24 16:12:07 2023 ranaSummaryBHDFPMI displacement noise with 60 Hz harmonics side lobes1) do a comparison with the whtiening before
the ADC on/off. This will tell us if it is
pickup before the whtiening filter or not.
  
  17414   Tue Jan 24 11:21:59 2023 yutaSummaryBHDFPMI displacement noise with 60 Hz harmonics side lobesJust to show how bad 60 Hz noise is, I
compared FPMI displacement noise with pre-BH44
era (measured on Jan 13, 40m/17400).
 FPMI_calibrated_noise_20230124_60Hz.pdfScreenshot_2023-01-24_11-27-22_FPMIndscope.png 
  17413   Mon Jan 23 22:51:17 2023 yutaSummaryBHD60 Hz harmonics side lobe investigations[Paco, Yehonathan, Yuta]

Since we have installed BH44, we
are seeing side lobes of 60 Hz + harmonics
 Screenshot_2023-01-24_11-43-40_POXPOYDark.png 
  17412   Mon Jan 23 20:50:58 2023 AnchalUpdateASCAS WFS path beam profiledI measured the expected beam profile by
WFS photodiodes by measuring the beam when
mode cleaner was unlocked from the point
 WFSPathBeamProfile.pdfASFocPathBeamProfile.pdfASPathBeamProfile.pdf 
  17411   Mon Jan 23 16:31:23 2023 ranaUpdateALSDFD and Phase tracker AM couplingBoth the TF measurement and the noise measurements
are useful, but the nosie measurement is
much more meaningful. Since we expect the
  
  17410   Mon Jan 23 11:20:44 2023 JCSummaryBHDBH44 RFPD optical path and LO/AS cameraHere's the beam path of BH44.  BH44.jpg 
  17409   Sat Jan 21 18:01:06 2023 AnchalUpdateALSDFD and Phase tracker AM couplingI took transfer function measurement
of DFD AM coupling using noise excitation.


Noise excitation
 BeatYRFAmplitudeNoiseASDwithSimulation.pdfAWGguiSettingsForSimulatedRFAMNoise.pngDFD_AM_to_Hz_Coupling_upto_0p1Hz.pdfDFD_Phase_Tracker_Noise_Due_To_Amplitude_Noise.pdf 
  17408   Sat Jan 21 15:32:40 2023 AnchalUpdateASCAS WFS path nominally setI've completed the beam redirection
path for AS beam to WFS heads in a nominal
way. By that I mean that all mirrors (M1,
 PXL_20230121_231740878.NIGHT.jpg 
  17407   Fri Jan 20 20:13:20 2023 AnchalUpdateASCInstalled 2 flipper mirrors for handingl MC reflection beam to cameraAfter discussions with Yuta, I figured
that a better optical layout is possible
which does not interfere with the existing
 PXL_20230121_035908170.NIGHT.jpg 
  17406   Thu Jan 19 20:35:54 2023 AnchalUpdateASCInstalled 2 flipper mirrors for handingl MC reflection beam to cameraToday I installed two flipper mirrors M3
and M4 (see attached photo) to create alternate
route for MC reflection camera beam. Both
 PXL_20230120_041313615.jpg 
  17405   Thu Jan 19 18:15:48 2023 yutaSummaryBHDBH44 RFPD optical path and LO/AS camera[JC, Paco, Anchal, Yuta]

- We installed the new BH44 beam path
 PXL_20230120_000603659.jpgScreenshot_2023-01-19_18-42-03_LO-ITMX_BH44Lock.pngScreenshot_2023-01-19_18-29-24_60Hzandfriends_FPMIBHD.png 
  17404   Thu Jan 19 14:58:40 2023 ranaSummaryBHDIQ demod orthogonalthe
problems with these circle plots:
  
  17403   Thu Jan 19 12:12:09 2023 AnchalSummaryBHDIQ demod orthogonalBy sending two frequencies offset from
eachother to LO input and RF input, we measured
the remaining phase difference between I
 BH44_IQDemodMeasuredDiff_1358192471.pdf 
  17402   Wed Jan 18 20:57:53 2023 PacoSummaryBHDIQ demod orthogonalI tested a spare IQ demod board labeled
33.3 MHz (closer to 44 MHz than the 165 MHz
we had started using) and using the Moku
  
  17401   Tue Jan 17 20:03:19 2023 yutaSummaryBHDBH44 installations: IQ demodulator is not orthogonal[Anchal, Paco, JC, Yuta]

We have started hardware installations
 6x 
  17400   Fri Jan 13 18:54:59 2023 yutaSummaryLSCInvestigations of LO phase locking in FPMI BHD[Paco, Yuta]

After several hours of unattended
 Screenshot_2023-01-13_18-41-48_SensMat_FPMIBHD.pngBHD_DARM_Sensors_20230113.pdfFPMI_calibrated_noise_20230113_LO1vsAS4.pdf 
  17399   Fri Jan 13 14:20:34 2023 yutaSummaryLSCCalibration friendly FPMI BHD[Paco, Yuta]

Gains in DARM are corrected to make
 Screenshot_2023-01-13_14-55-34.pngFPMI_calibrated_noise_20230113.pdf 
  17398   Fri Jan 13 13:34:12 2023 AnchalSummaryBHDBH44 tuned and transimpedance measuredI've tuned one gold box RFPD to be
resonant at 44.26 MHz and I left the notch
to be near 66 MHz, however, it is only effective
 BH44_Transimpedance_From_Test_Port.pdf 
  17397   Thu Jan 12 18:51:17 2023 PacoUpdateALSDFD and Phase tracker AM coupling[Anchal, Paco]

We measured DFD AM coupling; it
seems to be minimum at higher RF input and
 DFD_AM_coupling.pdf 
  17396   Thu Jan 12 15:31:27 2023 RadhikaUpdateALSXARM green laser lock debugging[Radhika, Anchal, Paco]

AUX
PDH Loop Stability
 IMG_4166.jpg 
  17395   Thu Jan 12 12:00:09 2023 PacoSummaryLSCFPMI BHD sensitivity curve with higher resolution upto 6.9 kHzHere's the same sensitivity plot from
yesterday (40m/17392),
but with higher frequency resolution, upto
 FPMI_calibrated_noise_20230112.pdfFPMI_calibrated_noise_20230112_HF.pdf 
  17394   Thu Jan 12 10:06:10 2023 PacoUpdateALSDFD discriminant calibration[Paco, Anchal] Log from yesterday
work around 1Y2 rack; note that while this
work was ongoing, TT2 position drifted slowly
 xbeat_dfd.png 
  17393   Wed Jan 11 17:05:55 2023 AnchalUpdateASCWFS demodulation board 112B - Working as expectedThe other modified board 112B has been
fixed and tested now. See the results attached.
The issue was in some malfunctioning OP284
 WFS_Board_112B_Test.pdf 
  17392   Wed Jan 11 16:56:57 2023 yutaSummaryLSCFPMI BHD recovered, LO phase noise not limiting the sensitivity[Paco, Yuta]

We recovered FPMI BHD, and sensitivity
 6x 
  17391   Tue Jan 10 20:24:29 2023 AnchalUpdateASCWFS demodulation board 111B - Working as expectedI've completed the modifications on
two WFS demod boards. This required replacing
all 8 mixer ICs on each board. I also tuned
 WFS_Board_111B_Test.pdf 
  17390   Tue Jan 10 16:06:58 2023 yutaSummaryPSLPMC transmission dropped to 0.68[JC, Paco, Yuta]

It seems like PMC transmission
(C1:PSL-PMC_PMCTRANSPD) dropped to ~0.68
 Screenshot_2023-01-10_14-26-46.pngIMG_1994.JPG 
  17389   Tue Jan 10 15:07:33 2023 JCUpdateGeneralClean Large Optical TableI have been working on cleaning the large
optical table next to the PSL table. I have
placed the known optics along in the optical
 IMG_6465.JPG 
  17388   Mon Jan 9 19:41:01 2023 AnchalUpdateSUSNull stream (butterfly/pringle) row added and DQedI updated the suspension model (/cvs/cds/rtcds/userapps/trunk/sus/c1/models/lib/sus_single_control_new.dml)
to add a 5th row in the input matrix so that
we can put in the calculated NULL stream
 SUS_Model_Changes.pngMC1_INMATRIX.png 
  17387   Thu Jan 5 14:30:32 2023 PacoHowToDAQnds2 server restartAfter being unable to fetch data offsite
using the nds40 server, I found enlightment
here.
  
  17386   Wed Jan 4 17:15:41 2023 KojiConfigurationCDSunknown dhcp request to fb1The dhcpd error on the log file stopped
when the yellow (DAQ) ethernet cable was
removed. With Chris's permission I left
 PXL_20230104_233926284.MP.jpg 
  17385   Wed Jan 4 15:10:19 2023 KojiConfigurationCDSunknown dhcp request to fb1Jamie reported that:


The logs (/var/log/daemon.log)
on fb1 are filling with this line:
 PXL_20230104_233907953.jpg 
  17384   Wed Jan 4 12:12:57 2023 PacoSummaryCalibrationALS calibration error from DFD[Paco, Anchal]

One of the crucial and currently
limiting calibration errors is in the ALS
 als_rf_power_screenshot.png 
  17383   Wed Jan 4 08:52:11 2023 JCUpdateOPLEV TablesETMX laser has diedKoji mentioned to me that this laser has
a degraded output. I removed this replacement
laser and put in a new laser from the Y Arm
 DACBE2CD-35EB-4314-92FD-71C5270A6961.jpeg45D7AF8F-5955-4036-B8AE-D761C2838408.jpeg 
  17382   Tue Jan 3 17:41:55 2023 KojiUpdateCDSChiara local backupChris modified the fstab so that the disk
is automatically mounted. The key was that
these two disks have an identical UUID and
  
ELOG V3.1.3-