  11870   Thu Dec 10 12:33:04 2015 yutaroUpdateLSCstrange behavior of ASDC

I did additional tests for the strange behavior of ASCD. ETMY, ETMX and ITMY were misaligned so that only light reflected by ITMX went into AS port. I had done similar measurement before with ITMY YAW varied.

Attachment 1 shows how ASDC level changed when ITMX PIT varied.

Attachment 2 shows how ASDC level changed when ITMX YAW varied.

Attachment 3 shows how the power of light measured by a power meter just after the AS view port varied when ITMX YAW varied.

 

Comparing 1 & 2, we can say that this behavior is not unique to YAW direction.

From 2 & 3, we can say something strange is happening inside the chamber.   

 

Attachment 1: 07.png
07.png
Attachment 2: 28.png
28.png
Attachment 3: ASDC.png
ASDC.png
  11871   Thu Dec 10 19:53:22 2015 yutaroUpdateLSCstrange behavior of ASDC

To check if the strange behavior of ASDC is caused by SR2/SR3 or not, I did the following measurement:

ASDC measures the power of the light reflected by ITMX. POXDC measures the power of the light reflected by ITMX and SRM successively. Then I varied the angle of ITMX in YAW direction and compared the behaviors of ASDC and POXDC.

The results are shown in Attachments 1-3.

As you can see in these figures, the strange up-and-down behavior appeared ONLY in ASDC. Therefore, the cause of this behavior exists between AS table and SRM (I had confirmed that the angle of SRM did not affect ASDC).

And this behavior is fringe-like, as can be seen in the figures (there seems to be 3 "peaks" and 2 "valleys"), so the cause could be interference between main path and not good AR reflection at a mirror after SRM before AS table (I suspect a mirror is flipped mistakenly).   

Attachment 1: 30.png
30.png
Attachment 2: 11.png
11.png
Attachment 3: 49.png
49.png
  11872   Fri Dec 11 09:35:44 2015 yutaroUpdateLSCPower recycling gain estimation from arm loss measurement

I took PR3 AR reflectivity and calculated PRG (PR3 is flipped and so AR surface is inside PRC).

As shown in attached figure, which shows AR specification of the LaserOptik mirror (PR3 is this mirror), AR reflectivity of PR3 is ~0.5 %. Since resonant light in PRC goes through AR surface of PR3 4 times per round trip, round trip loss due to this is ~2 %. Then I got

PRG = 7.8.    

 

Attachment 1: LaserOptikAR.png
LaserOptikAR.png
  11874   Fri Dec 11 15:37:50 2015 yutaroUpdateLSCPower recycling gain estimation from arm loss measurement

Attached is the plot of relation between the average arm round trip loss and power recycling gain. 2 % loss due to PR3 AR reflection is taken into account.

Attachment 1: PRG_plot.png
PRG_plot.png
  11875   Fri Dec 11 16:16:36 2015 yutaroUpdateOptical LeversCalibration of oplevs for ITMX/ETMX

Based on calibration measurement I have done (elog 11785, 11831), I updated calibration factors of oplevs on medm screen as follows. Not to change loop gain oplev servo, I also changed oplev servo gain.

C1:SUS-ETMX_OL_PIT_CALIB, C1:SUS-ETMX_OL_PIT_GAIN

(45.1,16) => (200,3.5)

C1:SUS-ETMX_OL_YAW_CALIB, C1:SUS-ETMX_OL_YAW_GAIN

(85.6,8) => (222,3.0) 

C1:SUS-ETMY_OL_PIT_CALIB, C1:SUS-ETMY_OL_PIT_GAIN

(26,-16) => (140,-3.0) 

C1:SUS-ETMY_OL_YAW_CALIB, C1:SUS-ETMY_OL_YAW_GAIN

(31,-21) => (143,-4.5) 

C1:SUS-ITMX_OL_PIT_CALIB, C1:SUS-ITMX_OL_PIT_GAIN

(110,8) => (122,7.2) 

C1:SUS-ITMX_OL_YAW_CALIB, C1:SUS-ITMX_OL_YAW_GAIN

(81,-11) => (147,-6) 

C1:SUS-ITMY_OL_PIT_CALIB, C1:SUS-ITMY_OL_PIT_GAIN

(159,15) => (239,10) 

C1:SUS-ITMY_OL_YAW_CALIB, C1:SUS-ITMY_OL_YAW_GAIN

(174,-21) => (226,-16) 

 