6873
|
Tue Jun 26 00:52:18 2012 |
yuta | Update | Computer Scripts / Programs | PMC locker |
Quote: |
I thought we rewrite auto lockers once per year, but this time it took us only a month. I wrote it for PMC on May 24. Is it not working?
|
I know.
I just wanted to use pyNDS for this kind of scanning & locking situation.
c1ioo was down for the weekend and I couldn't test my script for ALS, so I used it for PMC.
But I think PMClocker.py can relock PMC faster because it can sweep C1:PSL-PMC_RAMP continuously and can get continuous data of C1:PSL-PMC_PMCTRANSPD. |
9141
|
Thu Sep 19 18:48:24 2013 |
rana | Update | Computer Scripts / Programs | PMC locker |
In May of 2013 Den wrote a PMC Autolocker because he ignored / didn't want to read anyone else's code. Later that year Yuta also wrote another one from scratch for the same reasons.
I tried to use both today, but neither one runs. Yuta's one doesn't run because he was using a bunch of private yuta library stuff in the yuta directory. That kind of programming style is pretty useless for us since it never works after some time.
So I re-activated and tested the PMCAutolock bash script (it is actually a symbolic link called "PMCAutolock" which points to AutoLock.sh). These scripts are all basically the same:
They turn off the loop (or turn down the gain) and then scan the PZT, look for a resonance, and then activate the loop.
One problem with the logic has been that turning off the loop makes the gain so low that the peak flashes by too fast. But leaving the loop ON and just sweeping with the gain turned down to -10 dB is also not good. That only reduces the UGF from 1 kHz to ~100 Hz. What we want is more like a 10 Hz UGF while scanning the length. SO, I edited the script to turn down the modulation depth on the EOM by that factor. After acquiring lock, it returns all settings to the nominal levels as defined on the PSL_SETTINGS screen.
I've tested it a few times and it seems to work OK. You can run it from the yellow shabang button on the PMC screen.
I also changed the .bashrc aliases for the MEDM command so that if you type medm_good at the command line you get MEDM screens with scalable fonts. So you can stretch the screens. |
Attachment 1: pmc.png
|
|
9142
|
Thu Sep 19 21:15:44 2013 |
rana | Update | Computer Scripts / Programs | PMC locker |
I used a script (~PSL/PMC/testAutoLocker.sh) to unlock the PMC and run autlocker ~100 times to see how robust the new autlocker is.
It failed to grab it 2 out of 137 times. During those times it just went on trying to ramp the PZT even after it had gone to a rail. Once someone resurrects Rob's 'trianglewave' script we should be OK. Even so, I think this is good enough. Please try this out via the yellow button next time the PMC needs to be locked.
It usually takes 10-30 seconds to lock, depending upon where the fringe is compared to the upper voltage rail. Good enough. |
Attachment 1: Untitled.pdf
|
|
13457
|
Wed Nov 29 15:33:16 2017 |
rana | Update | PSL | PMC locking |
PMC wasn't locking. Had to power down c1psl. Did burt restore. Still not great.
I think many of the readbacks on the PMC MEDM screen are now bogus and misleading since the PMC RF upgrade that Gautam did awhile ago. We ought to fix the screen and clearly label which readbacks and actuators are no longer valid.
Also, the locking procedure is not so nice. The output V adjust doesn't work anymore with BLANK enabled. Would be good to make an autolocker script if we find a visitor wanting to do something fun. |
6229
|
Thu Jan 26 19:28:02 2012 |
kiwamu | Update | IOO | PMC low transmission |
After I recovered the lock of PMC, I found that the PMC transmission was quite low. It was about 0.26 in the EPICS display.
I zeroed the PSL temperature feedback value which had been -2.3 and then the PMC transmission went back to a normal value of 0.83.
I believe it was because the PSL was running with two different oscillation modes due to the big temperature offset. |
3560
|
Sun Sep 12 23:02:53 2010 |
valera | Update | | PMC mode matching |
Kiwamu and I found that the first lens in the PMC mode matching telescope was mislabeled. It is supposed to be PLCX-25.4-77.3-C and was labeled as such but in fact it was PLCX-25.4-103.0-C. This is why the PMC mode matching was bad. We swapped the lens for the correct one and got the PMC visibility of 82%. The attached plot shows the beam scans before and after the PMC. The data were taken with the wrong lens. The ABCD model shown in the plot uses the lens that was there at the time - PLCX-25.4-103.0-C. The model for the PMC is just the waist of 0.371 mm at the nominal location. The snap shot of the ABCD file is attached. The calculation includes the KTP for FI and LiNb for EOM with 4 cm length. The distances are as measured on the table. |
Attachment 1: pmc.pdf
|
|
Attachment 2: pmc-abcd.tiff
|
3899
|
Thu Nov 11 18:05:55 2010 |
valera | Update | PSL | PMC mode matching at full laser power |
The PMC mode matching was initially done at low power ~150 mW. It was expected and found that at full power ~2 W (injection current 2.1 A) the mode matching got much worse:
the visibility degraded from 80% to 50% (1 - refl locked/refl unlocked) . The thermal lensing could be in the laser, EOM, or FI.
The first attached plot shows the scan of the beam after the EOM at low and full laser power. At full power the waist position is 10 mm after the turning mirror after the EOM and the waist size is 310 um.
The second plot shows the ABCD calculation for the mode matching solution. 
I removed the MM lens PLCX-25.4-77.3-C and placed the PLCX-25.4-180.3-UV about 20 mm after the first PMC periscope mirror (the second mirror after the EOM).
The PMC visibility improved to 94% and the power through the PMC, as measured by the PMC transmission PD, went up by a factor of 2. |
Attachment 1: scan.pdf
|
|
Attachment 2: pmc2-abcd.png
|
|
3856
|
Wed Nov 3 19:14:00 2010 |
rana | Update | PSL | PMC mode matching update |
I moved the lens just before the PMC to check the mode matching landscape. The PMC trans went up from ~6.5 to ~6.8. That's 5% with ~1 hour of work.
As per the micrometer, this took ~7-8 mm of travel. Since there's so much power left in the HOMs, we we we will have to do a proper mode scan and re-calculate the solution.
The measured transmission is now ~610 mW. The power reflected from the PMC with it unlocked is ~1400 mW.  |
15139
|
Wed Jan 22 11:22:39 2020 |
gautam | Update | PSL | PMC modulation depth measurement |
Summary:
I estimate the PMC servo modulation depth to be approximately 50 mrad. This is only 15% lower than what was measured in Jan 2018, and cannot explain the ~x50 reduction of optical gain measured earlier in this thread. Later in the day, I also confirmed that the LO input to the ZAD-6 mixer is +7 dBm. So the crystal is not to blame.
Details:
- PSL frequency is locked to the IMC length.
- Arm lengths are locked to the PSL frequency using POX/POY.
- EX green laser locked to the X arm length using end PDH servo. GTRX was ~0.4 in this measurement, which is the nominal value.
- The 20dB coupled port of the beat between the EX and PSL lasers was monitored using the AG4395A in "Spectrum" units.
- The beat was set at ~90 MHz, and a spectrum was taken for ~100 MHz span centered at the beat frequency.
- The modulation depth is estimated by considering the ratio of power at the beat frequency relative to that 35.5 MHz away. See Attachment #1.
Assuming a finesse of 700 for the PMC, we expect an optical gain of 2*Pin*J0(50e-3)*J1(50e-3)/fp ~ 1.2e-7 W/Hz (=0.089 GW/m). I can't find a measurement of the PMC RFPD transimpedance to map this onto a V/Hz value. |
Attachment 1: modDepth.pdf
|
|
9421
|
Thu Nov 21 16:32:20 2013 |
Steve | Update | IOO | PMC needs a touch of love |
The PMC power degrading on this 3 days plot. MC2 -T = 14,200 counts. C1:IOO-MC_TRANS_SUM can not be ploted in dataviewer. The MEDM screen has a valid number.
Initial pointing is not so bad (what does "not so bad" mean ???)
C1iscey comes and goes again.
|
Attachment 1: PMCplus.png
|
|
662
|
Sat Jul 12 23:28:31 2008 |
rana | Update | PSL | PMC needs help |
As everyone has noticed recently, the PMC seems to have a PZT problem. It often zooms
of to one of its rails after locking as if the PZT range has decreased dramatically.
WE should check this on Monday by disabling the FSS and applying a slow triangle wave to
the NPRO frequency. The PMC will track this frequency change and this will allow us to
diagnose its problems.
If it has real problems, I have a spare in W. Bridge which we can swap in temporarily.
The attached plot shows 3 years of trend; looks like it went bad in summer of last year.
This also seems to be the cause of our ISS saturation problems: lowering the gain of the PMC
even slightly increased the intensity noise enough to cause saturation. Increasing the gain
even slightly increased the intensity noise enough to cause saturation due to PMC servo oscillation. |
Attachment 1: Untitled.png
|
|
4798
|
Thu Jun 9 10:38:30 2011 |
steve | Update | PSL | PMC needs help |
The PMC is losing power. |
Attachment 1: pmcnothp.jpg
|
|
8542
|
Tue May 7 18:42:20 2013 |
Jamie | Update | PSL | PMC not locking |
I'm just now realizing that the PMC has also not been locked since noon today, and doesn't seem to be responding to anything right now.
wtf is going on here? |
3817
|
Fri Oct 29 04:24:34 2010 |
Koji | Update | IOO | PMC output increased: need attention |
[Kevin Koji]
- The PBS alignment increased the transmitted power
- The first faraday and the PMC EOM were realigned.
- The transmission of the PMC increased from ~5.4V to ~6.5V.
Thus we need to pay attention to the incident beam power on to the MC
so that it does not exceed the power of 20-40mW.
Kevin will give us the detail of the work. |
8077
|
Wed Feb 13 16:31:08 2013 |
Jenne | Update | PSL | PMC pitch input tuned, MC yaw input tuned |
[Jenne, Yuta]
I looked at PMCR camera on the MC1 tv, and tweaked up the beam going into the PMC - it only needed a little bit of pitch.
Yuta and I measured the MC spots, determined (consistent with my measurements this morning) that they were only off in yaw. We touched the 2nd steering mirror in the zigzag on the PSL table in yaw a small amount (top of knob away from me), realigned the MC, and things were good. The plot is zoomed in to show only measurements taken today. 2 in the morning, before anything in the IFO room was touched. 1 this afternoon after tweaking PMC. 1st attempt at PSL beam tweaking was successful, 2nd measurement confirms it wasn't a fluke.

|
116
|
Tue Nov 20 10:11:33 2007 |
John | Summary | PSL | PMC pole measurements |
We measured the PMC pole in the following way.
1. Reduced laser power by rotating lambda/2 plate at laser output. Thermal effects in the PZT distort resonance peaks. Reducing power too much leads to problems with digitisation error.
2. Sweep NPRO PZT (C1: PSL-FSS_INOFFSET) using trianglewave. Record ramp, PMC transmission and reference cavity transmission ('C1: PSL-FSS_FAST','C1: PSL-ISS_INMONPD_F','C1: PSL-FSS_RCTRANSPD_F).
3. Since the PZT cannot sweep a full FSR in the PMC we looked at the sideband resonances within the reference cavity to calibrate the actuator.
Result: 7.35 +/- 0.22 MHz/V
4. Use #3 to calibrate the x axis of the PMC transmission.
5. Fit PMC resoances to an Airy function to get finesse. Take an average, weighted according to the resnorm. Calculate cavity pole frequency.
Result: 380kHz +/- 59kHz. This corresponds to a finesse of ~936. According to this plot the nominal pole is at 488kHz and the finesse is 732.
This is by no means a definitive measurement due to the misshapen resonance peaks recorded. |
Attachment 1: FittedPMCPeak.jpg
|
|
2155
|
Wed Oct 28 09:12:18 2009 |
steve | Update | PSL | PMC power on the rise? |
The PMC power is seems to be on the rise, ( MOPA_AMPMON is dropping ?) but I do not think it is real. We have Santa Anna wind condition, when the relative humidity drops and ......
There is an other funky think. The room temp became rock solid. The PSL HEPAs running at 20% and IFO-room ACs are also in normal operational mode. |
Attachment 1: pmcprising.jpg
|
|
645
|
Tue Jul 8 08:16:56 2008 |
steve | Update | PSL | PMC problem |
The PMC is unhappy. PMC auto locker is not working.
DC output slider adjust has to be moved from rail to rail before it locks.
MZ is working great. |
Attachment 1: pmc.jpg
|
|
8545
|
Tue May 7 20:09:10 2013 |
Jamie, Rana | Update | PSL | PMC problem was FSS slow actuator |
Rana showed up and diagnosed the problem as a railed FSS SLOW output. The SLOW Monitor about was showing ~6V, which is apparently a bad mode-hoppy place for the NPRO. Reducing the SLOW output brought things back into a good range which allowed the PMC to lock again.
In attempting to diagnose the problem I noticed that there is -100 mV DC coming out of the PMC RFPD RF output. This is not good, probably indicating a problem, and was what I thought was the PMC lock issue for a while. Need to look at the PMC RFPD RF output. |
6995
|
Fri Jul 20 12:12:25 2012 |
rana | Update | PSL | PMC problems examined |
Jenne, Den, Rana
The PMC transmission has been varying a lot and the MC seems unstable. Superstitious people might blame this on the El-nino or the alignment with Sagitarius, but we are ostensibly scientists.
WE found that the PMC EPICS values had not been toggled since the reboot and so the RF phase and Amplitude were totally wrong (we should replace this with a fixed oscillator box as we did with FSS).
Also, the NPRO SLOW slider was at -2 V which made the mode going into the PMC funny (although the mode was OK this morning before I started playing with the PMC sliders).
Before adjustment, there was a strong correlation between the seismic motions and the PMC reflection. This means that the PMC gain was low and it couldn't stay locked. Now, after fixing the RF and upping the gain slider it looks more stable. Let's watch it for a few days to see if there's an improvement in the trends.
The 10-minute trend of the lat 400 days shows that nothing has changed much this year; its been equally bad for a long while. |
Attachment 1: Untitled.png
|
|
6998
|
Sat Jul 21 14:05:21 2012 |
Den | Update | PSL | PMC problems examined |
Quote: |
WE found that the PMC EPICS values had not been toggled since the reboot and so the RF phase and Amplitude were totally wrong (we should replace this with a fixed oscillator box as we did with FSS).
Also, the NPRO SLOW slider was at -2 V which made the mode going into the PMC funny (although the mode was OK this morning before I started playing with the PMC sliders).
|
PMC transmission is oscillating in the range 0.5 - 0.85. PMC PZT voltage is 1-2 V.
FSS slow controls was -2.5 V. I adjusted it to 0 and PMC stabilized. PMC PZT voltage is 128, transmission is 0.845.
But most probably, slow control will drift again.

|
9155
|
Tue Sep 24 10:55:45 2013 |
rana | Update | PSL | PMC re-aligned |
After relocking the PMC at a good voltage, Steve and I re-aligned the beam into the PMC by walking the last two steering mirrors. After maximizing the power, we also aligned the reflected beam by maximizing the PMC_REFL_DC with the unlocked beam.
Transmission is back to 0.84 V. We need Valera mode matching maintenance to get higher I guess. Maybe we can get a little toaster to keep the PMC PZT more in the middle of its range? |
Attachment 1: psl-trend.png
|
|
15004
|
Thu Oct 31 10:44:40 2019 |
gautam | Update | PSL | PMC re-locked |
PMC got unlocked at ~4am. I re-locked it. Also tweaked the input pointing into the cavity. The misalignment was mostly in pitch.
There was also a loud buzzing in the control room due to the audio cable being improperly seated in the mixer. I re-seated it. |
15202
|
Mon Feb 10 10:07:20 2020 |
gautam | Update | PSL | PMC re-locked |
I found the PMC unlocked this morning. It was re-locked using the usual procedure. I feel like this has been happening more frequently in the last month than before. In the past, the cause seems to have been the PZT voltage drifting too close to one of the rails - however, in this case, it looks like an IMC unlock event is what triggered the PMC lockloss (admittedly the PZT voltage was somewhat close to the rail). It would be good if someone can re-connect the PMC Transmission photodiode, it was a useful diagnostic channel we had working fine before the ringdowns started.
I also tweaked the input pointing into the PMC and ran the WFS DC offset relief script. |
Attachment 1: PMCunlock.png
|
|
15662
|
Fri Nov 6 14:08:44 2020 |
gautam | Update | PSL | PMC re-locked |
The PMC servo railed and so I re-locked it at ~half range. I've been noticing that the diurnal drift of the PZT control voltage has been larger than usual - not sure if it's entirely correlated with temperature on the PSL table. Anyway the cavity is locked again so all is good. |
15570
|
Tue Sep 15 10:57:30 2020 |
gautam | Update | PSL | PMC re-locked, RGA re-enabled |
The PMC has been unlocked since September 11 sometime (summary pages are flaky again). I re-locked it just now. I didn't mess with the HEPA settings for now as I'm not using the IFO at the moment, so everything should be running in the configuration reported here. The particulate count numbers (both 0.3um and 0.5um) reported is ~x5-8 of what was reported on Thursday, September 10, after the HEPA filters were turned on. We don't have logging of this information in any automated way so it's hard to correlate things with the conditions in Pasadena. We also don't have a working gauge of the pressure of the vacuum envelope.
The RGA scanning was NOT enabled for whatever reason after the vacuum work. I re-enabled it, and opened VM1 to expose the RGA to the main volume. The unit may still be warming up but this initial scan doesn't look completely crazy compared to the reference trace which is supposedly from a normal time based on my elog scanning (the timestamp is inherited from the c0rga machine whose clock is a bit off).
Update 1500: I checked the particle count on the PSL table and it barely registers on the unit (between 0-20 between multiple trials) so I don't know if we need a better particle coutner or if there is negligible danger of damage to optics due to particulate matter. |
Attachment 1: RGAscan.pdf
|
|
9960
|
Fri May 16 00:25:53 2014 |
rana | Update | PSL | PMC realign |
Tonight I noticed that the drop in PMC transmission was ~1V, more than the usual of ~0.5V from the daily drift.
While re-aligning on the table, I noticed that the misalignment was not from either of the steering mirrors; i.e. I has to walk them both to get the alignment back. This implies that the misalignment is generated far upstream. Maybe the the laser itself is moving. We need some updates from Steve's laser misalignment tracker. |
6153
|
Tue Dec 27 23:03:56 2011 |
kiwamu | Update | PSL | PMC realigned |
I have realigned the steering mirrors for PMC because the transmitted light had been at ~ 0.741
After the alignment it went back to ~ 0.850. |
6350
|
Mon Mar 5 03:22:54 2012 |
kiwamu | Update | PSL | PMC realigned |
I realigned the steering mirrors for the PMC. The trans value went up from 0.79 to 0.83.
The misalignment was largely in the pitch direction. |
10885
|
Fri Jan 9 19:18:51 2015 |
Jenne | Update | PSL | PMC realigned |
A few hours ago I tweaked up the alignment to the PMC. It was really bad in pitch, and the transmission was down to about 0.711. |
10909
|
Thu Jan 15 19:01:30 2015 |
ericq | Update | PSL | PMC realigned |
PMC realigned again... The transmission was down to 0.70, and the MC was having a hard time trying to autolock. |
6217
|
Mon Jan 23 15:43:47 2012 |
Jenne | Update | IOO | PMC realignment |
Quote: |
I realigned the incident beam to PMC at 23:30. The transmitted light went up from 0.78 to 0.83.
|
Do we have PSL pos and ang QPD trends? We should start watching them, because the PMC drifted back down to 0.76 transmission, ~3.5 days after Kiwamu realigned it (his elog is from last Thurs). Not so awesome.
I walked through the control room just now and found both PMC and MC unlocked. They're both locked now, but with PMC transmission 0.76, MC transmission ~24,500. |
6218
|
Mon Jan 23 23:12:00 2012 |
kiwamu | Update | IOO | PMC realignment |
I have realigned the beam pointing to PMC. The transmitted light increased from 0.74 to 0.83.
The misalignment was mainly in pitch. |
15247
|
Wed Mar 4 11:16:37 2020 |
gautam | Update | PSL | PMC realignment |
I realigned the input pointing into the PMC this morning. Usually, the way I do this is to minimize any discernible mode structure in the PMC reflection CCD image. Today, I noticed that making the DC reflection go down also makes the DC transmission go down. Possibilities:
- we are sampling slightly different spots inside the PMC cavity which change the buildup by ~2-3%.
- we are misaligned on the transmission/reflection photodiode.
- ??
|
Attachment 1: PMCrealignment.png
|
|
6213
|
Thu Jan 19 23:34:52 2012 |
kiwamu | Update | IOO | PMC realignment and HEPA |
I realigned the incident beam to PMC at 23:30. The transmitted light went up from 0.78 to 0.83.
Also I decreased the HEPA level down to 20 % for the night time locking. |
9329
|
Fri Nov 1 19:09:01 2013 |
rana, evan | Configuration | PSL | PMC reflected beam nonsense |
While looking at the PMC REFL beam for the AOM diffracted beam, we noticed that although only one beam exists between the PMC and the first steering mirror, there are two afterwards and they both go to the PMC REFL RFPD!!! This is madness. We only want one beam on our PDH diode.
The reason that we have two beams is that that first steering mirrors is actually a (W1-PW-1025-UV-1064-45P) non-wedged window with an AR coating on only one side. So two beams come out of it. There is a terrible and floppy and illegal anodized aluminum dump close to this beam which *someone* probably intended to use as a "scraper" to get rid of one of the beams.
Black anodized aluminum is a horrible beam dump material at 1064 - its about as grey as Steve's chair. And its so soft that it scatters light back into the PMC and makes more acoustic noise. And it is mounted so poorly (only one screw) that it can easily be bumped and twist and miss the beam. Punchline: only use anodized aluminum dumps for stray light around cameras or for HeNe for OL. Its NOT allowed anywhere where we care about interferometry of NIR beams.
It was also set to dump the dimmer beam. On Monday, we should order ~5 W1 and get them with a wedge of 1-2 deg. Then we use a black glass dump for the dim beam and orient the bright one to hit the REFL camera and the PMC REFL PD.
For the weekend, I have adjusted the crappy grey aluminum flapper to catch the bright beam so that the PMC REFL image no longer shows the interference fringe of two beams. Lets see how the PMC drifts over the next 3 days. |
9094
|
Mon Sep 2 15:22:57 2013 |
Jenne | Update | PSL | PMC relocked |
The PMC was locked on an LG 10 mode (or something like it), for at least the last 8 hours. I relocked it on the regular 00 mode, and it's fine now.
Also, in CDS news, I did an mxstream restart (the RCG upgrade is supposed to make this not an issue anymore...), and did a "diag reset" afterwards, and all of the IPC errors except for one in the LSC model have gone away (OAF is still not running....on my to-do list, but not super high priority). |
9907
|
Sun May 4 14:20:04 2014 |
ericq | Update | IOO | PMC relocked |
The PMC has been unlocked for ~23 hours. FSS slow was at ~-1.5 V. I zeroed it, and relocked the PMC, transmission is ~0.81V. MC with WFS came back fine. |
9986
|
Wed May 21 22:15:37 2014 |
ericq | Update | PSL | PMC relocked |
PMC has been unlocked for ~4hrs, not sure why. It's servo gain was down at -10dB...
Relocked with transmission of .76V, MC locks fine with WFS, transmission of 15.5k. |
9993
|
Mon May 26 20:10:14 2014 |
ericq | Update | PSL | PMC relocked |
I came in and PMC transmission was at 0.5V, and ETMX was swinging around a lot, (LSC mode was on).
Turning off oplevs let ETMX calm down. I realigned the PMC to 0.82V.
MC wouldn't relock, it looked misaligned in pitch and yaw on MC camera.
I've touched the alignment, and gotten the reflection below 0.5, but it unlocks periodically, spot positions aren't great, and turning on WFS throws it out of alignment. ughhhhh |
11127
|
Tue Mar 10 14:47:05 2015 |
manasa | Update | PSL | PMC relocked |
PMC was locked in a bad state. FSS slow actuator adjust was at ~ -0.7 and PZT voltage at ~45.
So I set these right by moving the appropriate sliders and relocked it. FSS slow actuator adjust brought back to zero and PZT voltage ~115. PMC trans after relock is 0.789.
|
12032
|
Sat Mar 12 22:23:37 2016 |
rana | Summary | IOO | PMC relocked |
Found it locked on TEM01 mode.
Sweets in the fridge for non-PhD holders, courtesy of the highest levels of Caltech. |
688
|
Thu Jul 17 08:30:15 2008 |
steve | Update | PSL | PMC relocked manually |
The PMC pzt HV and the servo gain adj. are railing at max this morning
Why is it on the decreasing side of FSS_RTTEMP slope? |
Attachment 1: pmc4d.jpg
|
|
9944
|
Tue May 13 00:46:58 2014 |
rana | HowTo | PSL | PMC relocking |
The PMC runs out of range sometimes due to the daily temperature swing. The voltage swings up after sunset and then starts to swing down before sunrise. So when you relock the PMC at the beginning of the locking night, the mnemonic from the PMC is:
Sun Go Low, Lock Me Voltage Low. |
10800
|
Mon Dec 15 22:40:09 2014 |
rana | Summary | PSL | PMC restored |
Found that the PMC gain has been set to 5.3 dB instead of 10 dB since 9 AM this morning, with no elog entry.

I also re-aligned the beam into the PMC to minimize the reflection. It was almost all in pitch. |
10140
|
Mon Jul 7 16:39:09 2014 |
manasa | Update | PSL | PMC ringdown setup |
I moved stuff on the PSL table to accommodate the PMC ringdown setup.
I used the beam that leaks from the steering mirror at the PMC transmission that was dumped to a razor blade dump. I installed a Y1 to steer the beam to the ringdown PD. Power in the beam 75mW.
Results are in here elog |
Attachment 1: PMC_ring.png
|
|
10149
|
Mon Jul 7 23:19:55 2014 |
rana | Update | PSL | PMC ringdown setup |
Quote: |
I moved stuff on the PSL table to accommodate the PMC ringdown setup.
I used the beam that leaks from the steering mirror at the PMC transmission that was dumped to a razor blade dump. I installed a Y1 to steer the beam to the ringdown PD. Power in the beam 75mW.
|
I am guessing that 75 mW will burn / destroy any Thorlabs PD. I hope that mW is supposed to be uW. |
10164
|
Wed Jul 9 16:33:05 2014 |
manasa | Update | PSL | PMC ringdown setup |
Quote: |
Quote: |
I moved stuff on the PSL table to accommodate the PMC ringdown setup.
I used the beam that leaks from the steering mirror at the PMC transmission that was dumped to a razor blade dump. I installed a Y1 to steer the beam to the ringdown PD. Power in the beam 75mW.
|
I am guessing that 75 mW will burn / destroy any Thorlabs PD. I hope that mW is supposed to be uW.
|
It was ~7.5mW and measured ~2V at the PD output (given its range 0-5V ) on the oscilloscope . So PD is safe ! |
767
|
Wed Jul 30 13:09:40 2008 |
josephb, Eric | Configuration | PSL | PMC scan experiment |
We turned the PSL power down by a factor of 4, blocked one half of the Mach Zehnder and scanned the PMC by applying a ramp signal to PMC PZT. Eric will adding plots later today of those results.
We returned the power to close to original level and removed the block on the Mach Zehnder, and then relocked the PMC. |
984
|
Tue Sep 23 11:17:59 2008 |
steve | Update | PSL | PMC scattering spot |
The PMC output side has a new madly scattering spot at chamfer 2 o'clock position |
Attachment 1: rainbow.png
|
|
Attachment 2: pmcclip.png
|
|