40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 215 of 349  Not logged in ELOG logo
ID Date Author Type Categoryup Subject
  9806   Mon Apr 14 11:19:55 2014 JenneUpdateLSCMC WFS found off

I'm not sure why, but the WFS were turned off when I came in this morning.  The MC was not staying locked, and even during brief locks, the FSS FAST out was railed at 10. 

Aligning the MC mirrors to maximize the transmission, and then engaging the WFS seems to have made things better.

  9807   Mon Apr 14 13:20:45 2014 JenneUpdateLSCIFO Configure screen updated, CARM / DARM scripts added

I have compressed the IFO Configure screen.  All PRMI things (sideband lock and carrier lock) are in the PRMI button, all arm things (both RF and ALS) are in the respective arm buttons.

I have also made a new set of scripts for CARM and DARM lock acquisition with ALS. 

I hope that each button's purpose is clear, but take a second to look at them before you next use the IFO Configure screen.

  9809   Mon Apr 14 19:02:09 2014 JenneUpdateLSCMICH gets noisy as CARM or DARM offset reduced

This afternoon, I was toying around with reducing either the CARM or DARM offsets (so, put in a CARM offset, leave DARM zero, lock the PRMI, then reduce CARM offset to zero.  Or, put in a DARM offset, leaving CARM offset zero, lock the PRMI, then reduce the DARM offset to zero).

When looking at the data, I see that the MICH error signal gets fuzzier when the arms get close to resonance. (Note here that because I forgot to zero the carm offset before finding the resonances, -3 is my zero point for this plot and the next.)

MICH_fuzzy_when_offsets_small_longerData.png

Here is a zoom of the last piece of this time series, but with both TRX and TRY plotted (along with POPDC, CARM_ERR and DARM_ERR), where you can see that I had a momentary power buildup of > 100 transmission counts, which is about 20% of our final expected power.

TRX_TRY_100cts.png

Here is a different time series, showing a reduction of the DARM offset, and you can see that as the offset approaches zero, the MICH error signal gets noticeably more fuzzy.  Somewhere near the 240 second mark, I lose PRMI lock.

MICH_fuzzy_when_offsets_small.png

  9810   Tue Apr 15 02:19:54 2014 JenneUpdateLSCAnalog phasing of REFL11 and REFL55

[Jenne, EricQ]

I told Koji that I wanted to play with the common mode servo this evening, and he pointed out that we only get the signals after the digital demod phase angle in the digital system (obviously).  So, if I want to use either REFL11 or REFL55 for my CARM signal, I want to do something in analog-land so that my digital demod phase is close to 0 or 90. 

While we had the PRFPMI locked (with CARM offset of 2 or 3 nm), we set the demod phases of REFL11 and REFL55 to minimize a CARM line in the Q-phase.  This gave us -34 degrees for REFL11, and -75 degrees for REFL55. 

We calculated that about 1 degree of phase shift is about 1/(2 * pi * freq), or about 1.4e-8 seconds of delay for 11MHz.  We took the speed of light in the cables to be about 2/3*c, so 1.4e-8 * 2e8 = 2.9 meters per degree for 11MHz.  Since REFL11 was 34 degrees from 0, we estimate that we need to add about 98 meters of cable to the REFL11 signal path.  The same calculation for 55 MHz, but with a 15 degree shift required, gives 8.8 meters of cable to be added to the REFL55 signal path. 

I connected up some long BNC cables, and inserted them between the heliax breakout board on the LSC rack, and the respective PD inputs of the REFL11 and REFL55 demod boards.  I used (45 meters + 45 meters + a little bit) for REFL11, and used about 9 meters for REFL55. 

When we relocked the PRFPMI, and redid the phasing, we were very close to zero for both REFL11 and REFL55!  REFL11's digital demod phase is now +1 degree, and REFL55's digital demod phase is -5 degrees.

We changed the input of the CM servo board from POY (which Den and Koji had been using in December - see elog 9500) to REFL11 I MON. 


Q locked the FPMI (separate reply elog), and then we tried engaging the CM analog servo.  We were not successful. 

 

These settings were mostly copied from elog 9500, so they are almost surely not correct. 

CM servo screen:  In1 gain = 31dB, switch on, offset = -2.7V, boost off, super boosts off, option=disable, 79:1.6k switch disabled, polarity minus, option disable, AO gain=8dB, limiter enable.

For the slow path, CM_SLOW -> MC LSC servo had a +1 in the input matrix. 

CM filters in the AUX_ERR screen:  FM1 (unwhite) on, all others off, gain = 2.6. 

MC servo filters:  FM7, FM10 on, all others off (no triggered filter modules).  Gain = 0 initially.

MC servo board AO path disabled initially, G=-32dB initially.

 

Once Q had the FPMI locked, I tried increasing just the CM analog gain (by enabling the AO path on the MC board, and increasing the gain).  Doing this, I lost lock at -3 dB. 

I then tried again, this time alternating increasing the analog gain, and increasing the MC LSC servo gain.  I got up to 3e-3 for the MC digital gain, and -7 dB for the analog gain before we lost lock again.

 

We have determined that we should probably try just locking one of the arms with POX or POY, as Den and Koji did, to get a feel for how the system works.

 

 

  9811   Tue Apr 15 02:26:45 2014 ericqUpdateLSCAnalog phasing of REFL11 and REFL55

For future reference:

As we were poking around with the common mode servo in an FPMI configuration, we locked CARM/DARM with ALS as in recent ELOGs.

MICH was locked on ASDC: ASDC -> MICH = 10.0 in the DCPD DoF Matrix (I couldn't easily get AS55Q working, ASDC worked quickly and good enough)

MICH gain +25, FM4 FM5 On, FM2 switched on once locked. Offset was manually adjusted to get closer to dark fringe.

Actuated on BS: MICH->BS = 0.5 in Output Matrix.

  9812   Tue Apr 15 08:55:57 2014 KojiUpdateLSCAnalog phasing of REFL11 and REFL55

I have never used such a long cable for RF phase adjustment. The speed of the signal is 2e8 m/s and the frequency is ~10e6 Hz.
This means that the wavelength is only about 20m. How could you end up with ~100meters?
The convenient way to remember the cable delay is "1m, 1MHz, 2deg". This gives us ~1.5m for 11MHz and 34deg.

In fact, 1 degree of phase shift is not 1/(2 pi freq) second of delay, but f/360.

For such a precise phase adjustment, it is better to calibrate the delay with the network analyzer.

Quote:

We calculated that about 1 degree of phase shift is about 1/(2 * pi * freq), or about 1.4e-8 seconds of delay for 11MHz.  We took the speed of light in the cables to be about 2/3*c, so 1.4e-8 * 2e8 = 2.9 meters per degree for 11MHz.  Since REFL11 was 34 degrees from 0, we estimate that we need to add about 98 meters of cable to the REFL11 signal path.  The same calculation for 55 MHz, but with a 15 degree shift required, gives 8.8 meters of cable to be added to the REFL55 signal path.   

 

  9813   Tue Apr 15 09:32:19 2014 GabrieleUpdateLSCMICH gets noisy as CARM or DARM offset reduced

I guess this is normal. DARM has (almost) the same effect of MICH on the corner interferometer signals, just increased in amplitude by the arm cavity amplification. When the arm is out of resonance, DARM is almost completely depressed and it is not affecting MICH at all. On the other hand, when the cavities are exactly at resonance, DARM signal is amplified w.r.t. MICH by the cavity gain (2F/pi).

Since DARM is still controlled with ALS, it is probably noisy. The closer to resonance you move the cavities, the more ALS noise in DARM will affect MICH.


Quote:

When looking at the data, I see that the MICH error signal gets fuzzier when the arms get close to resonance. (Note here that because I forgot to zero the carm offset before finding the resonances, -3 is my zero point for this plot and the next.) 

  9816   Wed Apr 16 01:51:16 2014 JenneUpdateLSCScripts written for ALS acquisition, CARM and DARM transitions

[Jenne, EricQ]

This evening, as part of locking activities, we threw together some handy scripts.


The first one, "Lock_ALS_CARM_and_DARM.py" (no judging of my naming style!!), lives in .../scripts/ALS/ . 

It acquires ALS lock in CARM and DARM mode, so we don't have to do it by hand anymore.

The first thing that it does is ask you to acknowledge that your beatnotes are in place, and they follow our new (newer than the last elog about conventions) beatnote convention.  You are reminded in the terminal window what that convention is:  When the temperature sliders for either arm is INCREASED, the beatnote frequency should INCREASE. 

After you acknowledge that the beatnotes are good, it sets the CARM and DARM servo gains to zero, enables the outputs, sets the input matrix elements, clears the phase tracker histories, and starts ramping up the gains (with +1,+1 for DARM, the darm servo gain is +positive.  with -1*ALSX,+1*ALSY for CARM, the carm servo gain is -negative).  At a gain of 3, it engages the integrators and the resonant gains.  At the final gain of 6, it engages the boosts.

We have used this script ~10 times tonight, and it's been great every time.


The next two scripts are for making the transition from ALS to IR signals.  They both live in ..../scripts/PRFPMI/

"Transition_CARM_ALS_to_TransSqrtInv.py" (again - no judging!) slowly blends the input matrix elements to swap CARM control from the ALS signals to the 1/sqrt(trans) signals.  It takes a few steps, and asks for a keyboard input between steps.  This is because if our 1/sqrt(trans) offsets aren't perfect, we can start to lose transmission power.  To mitigate this, we decrease the offset in the CARM servo filter bank to get more power back.  This script requires an input, which is what you want the final sqrtinv matrix elements to be.  It will fail without this.  For a CARM offset, both of the final sqrtinv matrix elements will have the same sign.

"Transition_DARM_ALS_to_AS55.py" (I can telepathically hear you judging me right now.)  does the same blending, except to swap DARM control from ALS signals to AS55Q.  For the same reason of imperfect offset-setting, it takes several steps, to allow you to adjust the CARM offset if needed. Although, after typing this, I realized that perhaps we should have been tweaking the DARM offset.  Either way, this transition required much less tweaking of offsets than the CARM transition did.  Again, the script requires an input, which is your final desired AS55Q->DARM matrix element value.

Both of these scripts should be run at a digital CARM offset of about 2 counts, although with the offset tweaking during the CARM transition, I usually end at about 1.5 counts. 

*  To determine the final gain value for the CARM sqrtinv matrix elements, we have been using a spare filter bank (ex. XARM), and having the input to that be the sum of the sqrtinv channels.  We then put in a CARM line, and look at the transfer function between the temporary filter bank's input, and the CARM_IN1. 

*  To determine the final gain value for the DARM AS55 matrix element, we have been doing a similar thing, looking at the transfer function between DARM_IN1 and AS55Q with a DARM line on.  We have been putting this DC gain into the static PD normalization (4th block from the left on the big LSC screen), although with the new script, it will be easier to just put that value into the matrix element.

  9817   Wed Apr 16 02:11:40 2014 JenneUpdateLSCCARM and DARM on IR signals, boosts engaged

[Jenne, EricQ]

Tonight, we transitioned CARM and DARM to IR signals, took loop transfer functions, and determined that we could engage the LSC boosts (FM4 in the CARM and DARM servos, which are the same as the XARM and YARM servos). 

Q is preparing spectra to post, and I will dig out time series.  Look for these tomorrow, if they aren't posted tonight.

For the time series data fetching, I have taken notes on what we were doing when, so that I can actually find the data.


11:09pm:  CARM's LSC boost on for the first time

11:14pm:  DARM transferred to AS55Q

11:21pm:  DARM's LSC boost on for the first time

(lockloss)

11:53pm:  CARM transition

12:02am:  DARM transition done, both LSC boosts on

12:04am:  lockloss after reducing CARM digital offset to 0.4

12:45am: PRMI + 2 arms flashing, with no CARM or DARM offsets (arms still on ALS) because we forgot to put in the CARM offset before restoring PRM alignment.  PRMI may have been actually locked, or we may just have been flashing....need to look through the data to see what our recycling looked like.

(lockloss)

1:05am:  pretty smooth transition completed (both CARM and DARM), but we lost lock while reducing the CARM offset.

1:19am: lockloss - why?? We were just sitting at a CARM offset of about 1.3nm (1.3 counts), holding on IR signals.  We were not touching any IFO things while looking at some plots, and just lost lock.  Want to see if we can understand why.

1:27am:  another nice smooth transition for both CARM and DARM to IR signals, but almost immediate lockloss when reducing the CARM offset.


Using the new ALS lock acquisition scripts (elog 9816) and our transition scripts, getting back to PRFPMI lock is pretty smooth and procedural.

* Align arms using ASS (ifo configure screen, restore xarm and yarm, run both arms' ass scripts).

* Align PRMI, no arms (ifo configure screen, restore prmi sideband)

* Find ALS beatnotes, with arm lasers on opposite sides of the PSL.  For both, when increasing the value of the temperature slider, the beatnote should increase in frequency.  (ifo configure screen, restore CARM and DARM als)

* Run ...../scripts/ALS/Lock_ALS_CARM_and_DARM.py

* Run "Find resonance" scripts from ALS screen for each arm.

* Put in a 3 count offset to CARM loop.

* Restore PRM alignment.  (PRMI should acquire lock immediately, although PRM may need some small alignment tweaking).  Enable PRCL and MICH outputs, PRM and BS actuation outputs.

* Reduce CARM offset to 2 counts. 

* Set offsets of 1/sqrt(TRX) and 1/sqrt(TRY) filter banks in the AUXERR section of the LSC screen.  The outputs of both should equal 2 counts (to match the 2 count offset in the CARM loop). 

* Run .../scripts/PRFPMI/Transition_CARM_ALS_to_TransSqrtInv.py , making sure to reduce the CARM digital offset if needed, to keep the arm transmissions at about 0.1 counts.

* Engage FM4 of the CARM filter bank, which is the LSC boost.

* Run .../scripts/PRFPMI/Transition_DARM_ALS_to_AS55.py , making sure to reduce the CARM (or should be DARM?) digital offset if needed, to keep the arm transmissions at about 0.1 counts.

* Engage FM4 of the DARM filter bank, which is the LSC boost.


Notes for going forward:

When we have small-ish digital CARM offsets, such that both of our arm transmitted powers are about 0.1 or higher, we see clear coherence between our CARM_IN1 (the 1/sqrt(trans) signals) and a normalized REFL11_I (again using a spare filter bank like XARM to get REFL11 normalized by (TRX+TRY) ).  We have not yet tried transitioning the CARM digital error signal to this normalized REFL11.

Even though we see that the IFO is much less noisy (as measured by significantly reduced RIN in TRX and TRY as visible by eye on Dataveiwer), we are still losing lock when we reduce the CARM offset.  I have noted above several times, for when we had locklosses, so that I can see if I see anything elucidating in the time series data.

  9818   Wed Apr 16 02:29:30 2014 ericqUpdateLSCCARM and DARM on IR signals, boosts engaged

 As Jenne mentioned, we took OLTF transfer functions, and determined that we had more than enough phase margin to switch on the LSC boosts in FM4. This improved the error signal noise spectra quite a lot, and noticeably reduced the TRX/TRY fluctuations, and actuation output. 

Here's the CARM OLTF (FM4 boost on in red, boost off in black)

carmOLTF.pdf

 

Here's what happened to the CARM and DARM spectra when we turned on the boosts. (ALS only in black, initial IR signal transitions in mid-color, boosted IR signals in bright color)

boostPlot.png 

  9819   Thu Apr 17 00:49:06 2014 JenneUpdateLSCCARM and DARM on IR signals, boosts engaged

I looked at 2 of the locklosses from last night, (1:19am and 1:27am), and saw that for both, the DARM loop started to oscillate just before we lost lock.  In the trials tonight, we were more watchful of gain peaking.

Here is the 1:19am lockloss

Lockloss_DARMgainTooHigh_119am.png

And here is the 1:27am lockloss

Lockloss_DARMgainTooHigh_127am.png


 So you can see what we were doing, and what the effect was, here is a few minutes of data just before the 1:27am lockloss. The times I note below are rough, but should give you an idea of what happened at which time.

0 sec:  Arms are held on resonance with ALS.

10 sec:  CARM offset of 3nm added.

20 sec:  PRM restored, one flash, then PRMI acquires lock.

30 sec:  CARM offset reduced to 2nm, transmitted powers are about 0.1

50 sec:  Transition CARM to 1/sqrt(trans) signals.  Note that we are using the high gain Thorlabs PD here, so the noise is better than last Thursday.

60-110 sec:  CARM offset reduction to about 1nm.

110 sec:  CARM's LSC low frequency boost engaged.

120 sec:  DARM transitioned to AS55Q.

170 sec:  DARM's LSC low frequency boost engaged.

SmoothCARMandDARMtransitions_LSCboosts.png

  9820   Thu Apr 17 01:01:02 2014 JenneUpdateLSCLSC model modifications

Last night, EricQ and I were concerned that we might need some CARM UGF servoing, so I added a UGF servo block, copied from the aLIGO LSC model, to our LSC model.  The block is inline with the CARM servo, after the output triggering, just before the output matrix.  Q put together some screens, which are accessible from the main LSC screen. 

The model is compiled and running.  We didn't get very far in testing it though before Koji pointed out that it is a slow solution, and not a fast one like we were searching for.  We were hoping to deal with the momentary power buildup, and thus optical gain change, as the arms flash close to resonance.  The UGF servo will not work nearly that fast though.  We may want it for slow UGF servo-ing, but it's not the solution to what Q and I were thinking about yesterday.  Regular ol' dynamic normalization is closer to the right answer for this.

In tonight's activities, Koji and I found that we probably want a CESAR block for DARM as well as CARM, so that we can independently normalize AS55Q. 

To solve the DARM oscillation issue from last night (that I discovered this evening when I finally looked at the time series data), we may want to implement a DARM UGF servo.  For tonight, as we reduced the CARM offset and started seeing gain peaking in the DARM spectra, I hand-reduced the DARM gain.

 

  9821   Thu Apr 17 01:18:34 2014 JenneUpdateLSCAttempt at handing CARM to REFL11

[Jenne, Koji]

This evening, Koji and I followed the procedure from last night (elog 9817), with the exceptions that as we saw gain peaking in the DARM spectrum, we lowered the DARM servo gain.  We also left the DARM FM4 boost off, and added (TRX+TRY) power normalization to AS55 (although we still had to hand-reduce the gain).    These changes enabled us to reduce the CARM offset much further.  We were able to get the transmitted powers to hold steady at about 1 count while on the IR signals, which is a new record for us.  (We had in the past held the arms with ALS at several counts, but the power fluctuations were huge, and now they are nice and small).

After that, we started looking at whether we could transition CARM over to REFL11I.  We tried a few times, but never made it all the way.


Here are some times for data-foraging tomorrow:

8:27pm, nice transition, CARM offset reduction to 0.6 before lockloss.

9:19pm, turned on power normalization for AS55Q, then reduced CARM offset to 0.5

9:40pm, Lockloss after reducing CARM offset to -0.24, arm transmitted powers around 0.9.

gps 1081748419: First trial trying to transition CARM to REFL11I normalized by (TRX+TRY).

gps 1081749965:  Tried to transition CARM to (REFL11 + REFL33)/(TRX+TRY).  Got about 1/3 of the way through the transition (in terms of matrix element value steps) before lockloss.

11:56pm, Tried to add in REFL11I to CARM error signal (without reducing 1/sqrt(trans) matrix elements).  We increased the REFL11 matrix element until we saw gain peaking, and then tried reducing the 1/sqrt(trans) contribution, and lost lock.  We were only at an offset of 0.3, so we probably weren't close enough to the resonance yet.  We were able to add in REFL11 information, but this was probably not too hard, since there wasn't much actual information in it.


Thoughts:

* It's a little weird that once we are on IR signals, the 0 CARM offset point that we find with ALS is not the true CARM offset point.  Although, this may be because we're just going to an averaged no CARM offset place with ALS, but since ALS is noisy, we won't ever really be holding on the zero offset point.  Anyhow, when we were using the 1/sqrt(trans) signals for CARM, and the CARM digital offset was -0.24, the ALSX and ALSY outputs were both about 0.5 in magnitude.

* We're getting there! 

  9827   Thu Apr 17 17:27:32 2014 ericqUpdateLSCSome reference Plots

Jenne made some suggestions for some plots that would be useful on our CARM offset reduction adventures, so I made some with my MIST model. 

First, here's a plot showing the transfer function of CARM to TRX, with logarithmically spaced offsets out to 3nm. While not a control signal, it shows us where the optical plant resonance stuff is happening. The peaks in this TF correspond to peaks in REFL11, REFL55, AS11, etc., as in the close-to-resonance TFs in ELOG 9785

carm2TRX.pdf

[more to come, had a MATLAB issue]

 

Attachment 2: carmSignalLevels.pdf
carmSignalLevels.pdf
  9829   Fri Apr 18 12:53:54 2014 JenneUpdateLSCAttempt at handing CARM to REFL11: Time series

Some time series data from Wednesday night. 

Here is the first time we've gotten the arm transmissions to about 1 count, while holding CARM and DARM on IR signals, so the transmission, as well as POPDC, were relatively quiet.

CARMin1_POPDC_TRX_1000sec_TRXwas1ct.png

As we were attempting to transition CARM to REFL11I, at least 2 of the times, we were hitting a CARM oscillation:

lockloss_CARMoscillating.png

lockloss_CARMoscillating_2.png

  9832   Fri Apr 18 20:17:17 2014 JenneUpdateLSCALS noisy

Last night, as well as tonight, the ALS seems not quite as robust as it was earlier in the week.

I have just taken noise spectra, and ALS is definitely more noisy than usual. 

These plots are with the arms held in CARM and DARM mode, with servo gains of 8. I was seeing the beginnings of gain peaking at a gain of 10, so I turned it back to 8.  Our ALS in-loop RMS is usually something like a few hundred Hz, but I'm seeing over 1kHz, so I have a factor of 4 or 5 too much noise.  Why?!?!?

ALS_1kHzRMSnoise.pdf

ALS_oolNoise.pdf

  9836   Mon Apr 21 22:53:16 2014 manasaUpdateLSCALS noise

Quote:

Last night, as well as tonight, the ALS seems not quite as robust as it was earlier in the week.

I have just taken noise spectra, and ALS is definitely more noisy than usual. 

These plots are with the arms held in CARM and DARM mode, with servo gains of 8. I was seeing the beginnings of gain peaking at a gain of 10, so I turned it back to 8.  Our ALS in-loop RMS is usually something like a few hundred Hz, but I'm seeing over 1kHz, so I have a factor of 4 or 5 too much noise.  Why?!?!?

I have noticed that ALS noise has been at 1KHz rms since LSC arm lock servos have been used to lock arms using ALS error signals. May be this has not been given much attention.

But looking more closely at the ALS noise (better dtt resolution for noise power spectrum) , there seems to be too much noise suppression at <1Hz and not much happening at around 10Hz.

Attachment 1 (data files at /users/manasa/data/140421/)

 

So I made a bunch of transfer function measurements for ALS and phase tracker servo. Koji will be using these and redesigning the servo filters so that we can get more suppression at 10Hz.

Other than this I also found that the Y arm showed more high frequency noise as compared to the X arm. (Edit by manasa: Thinking back now, this could be related to the onset of 60Hz noise at the Y end elog 9838. But still has to be looked at after fixing TRY)

Attachment 2

Tip: Once the arms are ALS locked, enabling the SLOW_SERVO helps hold the lock stably. 

P.S. I realigned the Y green to the arm and brought GTRY to 0.93

To do:

Find out what makes Y arm in-loop noise at high frequency higher than X arm.

Attachment 1: ALSX_FreeInLoop.jpg
ALSX_FreeInLoop.jpg
Attachment 2: ALSXY_inLoop.jpg
ALSXY_inLoop.jpg
  9838   Tue Apr 22 01:11:42 2014 JenneUpdateLSCTRY 60Hz noise

Quote:

 

P.S. I realigned the Y green to the arm and brought GTRY to 0.93

This evening, I was not able to successfully transition CARM from ALS to 1/sqrt(trans) signals.  The TRY time series looked odd, so I took a spectra, and we have huge 60Hz noise in TRY. 

I found a lock stretch from around 6:30pm that did not show the 60Hz noise, and then there was a lock stretch around 8pm that did have the noise.  So, something happened at the Yend between 6:30 and 8pm tonight.

Asking around, this was the time frame in which Manasa was down at the Yend to realign the green beam, and to check cabling for the PZT_OUT and ERR_MON signals to the ADC.

Looking at the spectra, Rana noted that we have even as well as odd harmonics of the 60Hz line, which is unusual.

TRY_60Hz_noise_21Apr2014.pdf

To try to diagnose the problem, Rana and I tried to make sure no cables' connectors were touching, and that no equipment was plugged in that shouldn't be.  We noticed that none of: the shutter, the Thorlabs TRY PD, or the QPD TRY are isolated from the table.  To see if perhaps the shutter was the problem, I turned off the power to the Yend green shutter, and unplugged the cable.  The cable is laying on the table, with the connector sitting on a piece of plastic to isolate it.  Removing the shutter from the system did not change anything. 

We don't see the 60Hz noise in the Xarm, so it's not on the laser light itself.  Also, we don't see the 60Hz lines in the Yarm feedback signal, so we're not putting the lines onto the mirror, and thus onto the Yarm's light. 

Manasa, can you please take a look, and see if you can figure out what is going on?  We need TRY so that we can transition to 1/sqrt(trans) signals for CARM.  Thanks!!

  9840   Tue Apr 22 02:14:55 2014 ericqUpdateLSClock acquisition path for the CM servo

In an effort to familiarize myself with the analog CM servo, I've begun to replicate Koji and Den's work from the ELOG post that this is a reply to.

I hooked POX11Q into the IN1 of the CM board. (POX is rotated by ~86 degrees in the CDS, meaning analog Q is almost perfect.)

While there, I took out the too-long delay cables Jenne introduced for REFLs 11 and 55. (Also note: when we do cable-based analog phasing in the future, we should do it on the LO side, instead of the PD input side.) I also heard a dangerously crinkly sound from the short SMA cable for REFL11, so I replaced it with a beefy looking new one I found on the SP table.

I messed with the gain and offset in the CM_SLOW input filter to get it to look just like POX11_I_ERR, and was able to lock the arm on it without an issue. I then put the SR560 between the CM and MC (30k pole, but also AC coupled, because I figure the digital loop should be doing the work down there, and don't want to kick the AO with an offset), and was able to turn on the AO path with a gain of 8dB on the CM board and 10dB on the MC board, as detailed in Koji's procedures.

I wasn't able to increase the AO gain to 9dB without breaking lock, but maybe this is ok, because by judging by the LSC filter gains, POX11 might be about 3 times bigger than POY, so maybe 8dB AO gain on POX ~ =18dB AO gain on POY? I was able to put the CM servo offset at 0, but turning on boosts promptly kicked the MC out of lock.

I'm stopping for the night; but tomorrow I'll bust out a spectrum analyzer to see if I actually have won some bandwidth with the CM servo, and check out the situation with the offsets and boosts.

  9841   Tue Apr 22 21:54:50 2014 manasaUpdateLSCTRY 60Hz noise

Quote:

Quote:

 

P.S. I realigned the Y green to the arm and brought GTRY to 0.93

This evening, I was not able to successfully transition CARM from ALS to 1/sqrt(trans) signals.  The TRY time series looked odd, so I took a spectra, and we have huge 60Hz noise in TRY. 

Manasa, can you please take a look, and see if you can figure out what is going on?  We need TRY so that we can transition to 1/sqrt(trans) signals for CARM.  Thanks!!

 I went to the Y end to look at the TRY 60Hz noise situation this morning. While looking at TRY noise on dtt, I found that just lifting the cable away from the cable bunch that runs out of the table suppressed the noise drastically. 

Attachment 1

I removed the unwanted bnc connector in the path of the already long TRY cable running from the PD to the 1Y4 rack and isolated it from the bunch. TRY became less noisy.

But the noise was back again earlier in the evening and it looks like the noise is very much related to the TRY cable. TRY cable might have moved from its sweet spot while I was around checking cable connections yesterday.

I couldn't find a spare to replace it right away today (We need a BNC to 4 pin lemo).

Attachment 1: 60HzTRY.jpg
60HzTRY.jpg
  9842   Tue Apr 22 22:49:10 2014 ranaUpdateLSCTRY 60Hz noise

 

 The detectors and electronics on this table are not properly isolated. To reduce the 60 Hz and ground loops, photodiodes and shutter must be isolated by using plastic spacers as we usually do elsewhere - this table just seems to have a few oversights.

Steve can start assembling all of the pieces to do this in the morning and then we can start the swapping after the meeting.

The high gain Transmon cable should be a regular BNC. There's no need for 4-pin LEMO in this usage, so the best move is to modify the board and replace the 4-pin LEMO connector with an isolated panel mount BNC female.

The AC adapter for this diode (and all of the detectors on the table) should get their power from a power strip which gets plugged into the rack with the whitening boards. The SHG oven, the Uniblitz shutter, and any cameras can get their power from another power strip if needed/wanted.

  9843   Wed Apr 23 19:58:00 2014 manasaUpdateLSCTRY 60Hz noise

[Steve, Manasa]

To find noise source

1. Swapped the power cable of the PD and checked that it is connected to the right power source.

2. Changed the aluminium base of the post holding the diode so that the diode is floating

3. Grounded the table and the rack

4. Routed the cable on the other side of the beam tube to isolate it from other cables.

After all the above, we still found that shaking the cable was making TRY noisy.

I pulled out the PD whitening board to replace the 4 pin lemo connector with a bnc connector so that we can swap the cable with a new one. So there is no TRY right now.

 

  9844   Wed Apr 23 23:48:30 2014 manasaUpdateLSCY end whitening board

The MON outputs of the Y end QPD whitening board were hot earlier today while pulling it out of the crate. After swapping the 4 pin lemo connector with an isolated panel mount bnc connector, I stuck the board back into the crate and this immediately kicked the ETMY suspension. Jenne and I went to the Y end to look at what was going on. We removed the board from the crate after smelling something burning. The MON output ports of the whitening board were super hot this time. There is no sign of any components melting on the board (comparing the board with its pictures that were taken earlier) and a tester board stuck into the crate lights up just fine.

So the back panel is still ok. We need to troubleshoot or replace the whitening board.

Edit, JCD:  The attached photos are from right after I replaced the "Rgain" resistor, elog 9823.  What they show is that it looks like some of the melting / burning may have already been happening before I pulled the board, and I just never noticed :(  In particular, look at the resistors on the main board above the blue "G" sticker.  There isn't a difference that I can tell between this photo from last week, and today's situation. 

 

 IMG_1378.JPG

Attachment 1: IMG_1378.JPG
IMG_1378.JPG
Attachment 2: IMG_1379.JPG
IMG_1379.JPG
  9845   Thu Apr 24 00:11:35 2014 JenneUpdateLSCYend shutter back.

Quote:

To see if perhaps the shutter was the problem, I turned off the power to the Yend green shutter, and unplugged the cable.  The cable is laying on the table, with the connector sitting on a piece of plastic to isolate it.  Removing the shutter from the system did not change anything.

 I re-plugged in the Yend shutter, and turned it on.

  9846   Thu Apr 24 02:12:05 2014 JenneUpdateLSCLocking without TRY

I tried some locking anyway tonight, even though we don't have TRY. 

The biggest conclusion is that I miss the auto-resonance-finding.  I've been roughly scanning the Y-ALS offset to find the POY zero crossing when I see the resonance on the test mass face cameras. 

The next-biggest conclusion, is that I can hold the PRFPMI close to resonance, using ALS for CARM and DARM.  I was trying to transition DARM to AS55, but I couldn't get the last bit of the way.  That is, I couldn't turn off the ALS control.  So, I think that AS55 wasn't actually holding DARM, until maybe the last moment or so.

Anyhow, here are some time series.  My average TRX value is around 40 counts, and POPDC is maybe 250 counts (just PRMI, POPDC is about 75 counts).  Obviously this is noisy as hell, but I'm not using any IR signals for the arms.  Near the end of this first time series, I am trying to switch to AS55 for DARM.

TRX_avg_40cts_POPDC_avg_200cts.png

Zooming in, my real lockloss is due to PRCL oscillating at ~350 Hz:

Lockloss_PRCL_350Hz.png

However, I also saw ~25Hz peaks in CARM and DARM on the spectra starting to show up, and I see a ~25 Hz oscillation in DARM a few moments after the PRCL lockloss.  (Plot #2 is a zoom of the ~1.1 second mark on Plot #3.)

Lockloss_DARM_20Hz.png


The locking parameters:

CARM:

Input:  Using the new CESAR matrix, -1*ALSX, +1*ALSY.  Beatnotes both move up in freq if temp sliders move up.

Servo: gain = 6, FMs 1, 2, 3, 5, 6, 7, 9 on.  Offset = 0 counts. 

Output = -1*MC2

DARM:

Input:  +1*ALSX, +1*ALSY

Servo:  gain = 4.  FMs 1, 2, 3, 5, 6, 7, 9 on.  Offset = 0 counts.

Output = -1*ETMX, +1*ETMY

PRCL:

Input:  +1*REFL33_I, Norm = +0.01*POPDC, sqrt engaged.

Servo:  acquisition easier with -0.04 or -0.06, less gain peaking at -0.02   FMs 4, 5 on; 2, 3, 6, 9 triggered with 0.5 sec delay.  Servo trigger = POPDC, up 100, down 10.  FM trigger = POPDC, up 300, down 20.

Output = +1*PRM

PRCL ASC off, PRM oplev on.

MICH:

Input:  +1*REFL33_Q, Norm = +0.01*POPDC, sqrt engaged.

Servo:  gain = 2, FMs 4, 5 on; 2, 3 triggered with 0.2 sec delay.  Servo trigger = POPDC, up 100, down 10.  FM trigger = POPDC, up 300, down 20.

Output = +0.5*BS, -0.2625*PRM

PDs:

REFL33 analog gain set to 30 dB for both I&Q.

AS55 set to 0 dB for both I&Q.  AS55 had DC normalization of 80 counts (which was the measured number for PRFPMI when TRX was about 0.1 count this evening)

 

  9847   Thu Apr 24 11:19:50 2014 KojiUpdateLSCLocking without TRY

This seems the ever best stability at the zero offset PRFPMI.

Can you look at REFLDC in this data stream too? How was it promising?

  9848   Thu Apr 24 14:00:42 2014 JenneUpdateLSCLocking without TRY

Here is 1 second of data, with REFLDC, POPDC and TRX:

REFLDC_1sec.png

Here is a zoom of the first 3 big peaks in TRX.  The weird jumps at the beginning of each TRX peak are due to the triggered switching between the Thorlabs trans PD and the QPD trans PD.  Clearly we need to work on their relative normalizations.  There are also little jumps after each peak as the triggering sends the signal back to the Thorlabs PD.

REFLDC_3peaks.png

Here is a zoom of the single big peak about halfway through the 1 second of data:

REFLDC_1peak.png

And here is a zoom of the tail of that peak.  It looks to me like we want to start thinking about using REFL DC when our transmitted powers are around 2 counts.  We could do as soon as 1 count, but 2 is a little farther into the dip.

REFLDC_1peak_zoom.png

  9849   Thu Apr 24 14:23:09 2014 not manasaUpdateLSCY end whitening board

 

 maybe the tantalum caps on the daughter board power supply lines are blown? If so, replace with 35V+ ceramic.

  9850   Thu Apr 24 16:25:31 2014 ericqUpdateLSCQuick CM servo prep

I added ~1m of cable to the LO side of the REFL11 Demodulator, which brought its PRCL demod phase to about 8 degrees. According to my simulations, PRCL and CARM have the same angle (but opposite sign) at resonance. There seems to be a severe lack of SMA cables in the lab, so I didn't tune it to be any closer. Cos(8 degrees)=.99, so I think it should be fine to use it for the CARM servo, since none of the other signals are going to be nearly as big. I plugged analog REFL 11 I back into the CARM servo IN1. 

As for IN2, I threw together a temporary setup for using REFLDC as a complementary signal. I T'd off the REFLDC signal (which is the DC signal out of REFL55), and sent it into an SR560 to subtract an offset. The offset comes from a 1Hz-passive-pomona-box-low-passed C1:IOO-TT4_LR output, since there are 8 DAC channels set up for the nonexistent tip tilts 3 and 4 actively running. The output of the SR560 is sent to the CARM servo IN2. 

I adjusted the offset by turning on only IN2 in the CARM servo MEDM screen, and looking at the CM_SLOW signal in data viewer. I adjusted gains and such to get it to look just like REFLDC with the PRC locked. There was good coherence and no appreciable phase difference from DC out to some kHz, albeit a dip in coherence to about .8-.9 from ~40 to 300Hz, for some reason. (This included turning on the unWhite FM in the REFLDC filter bank)

If this signal turns out to be useful, it will be relatively straightforward to put together a little box that does the offset subtraction nicely, but this should do for our immediate needs.

Lastly, I hung up this plot in the control room to give us information about the DC values of different signals as the CARM offset changes. This is helpful to see what our CARM offset is, based on the transmission we se, when different signals start to have length dependence, where they start/stop being linear, etc. The TRX curve is scaled to a maximum of 600, REFLDC is normalized to input power = 1, and all the rest are arbitrarily scaled to fit on the plot. I've assumed 75ppm loss on all mirrors in my simulation (PRM, BS, 2xITM, 2xETM), mostly to get some realistic profile of REFLDC. 

carmOffsetProfiles.pdf

  9852   Thu Apr 24 23:55:31 2014 KojiUpdateLSCY end whitening board

The main problem was a panel fixing bolt that caused the short circuits between power supply layers.
This burned the PCB and secondarily caused permanent short circuit between +15V/-15V/+5V layers.

Diagnosis

- The resistances between +15V, +5V, and -15V were low. The resistance between +15V and -15V is 13 Ohm.
  The one between +5V and -15V is 7Ohm. And the one between +15 and +5 is 19Ohm. So the situation is

                o -15V
                |
+15V o-(13 Ohm)-+-(9 Ohm)-o +5V

Even after removing all of the active components from the board, they remained the same.

- The tantalum caps were removed from the board and it was confirmed that they are not the cause of the issue.

- The panel was removed from the module for the component migration to a spare board (to be described in the other entry).
I found that the screw hole and the screw have burnt marks. The screw need an insulation tube to avoid short circuit.
The other screw was also bare. The spare board has the screws with the insulation tubes.

 

Attachment 1: P4245550.JPG
P4245550.JPG
  9853   Fri Apr 25 03:14:46 2014 ericqUpdateLSClocking activity

[ericq, Jenne, Zach]

We spent some time tonight trying to push our CARM locking further, to little avail. DARM/CARM loop oscillations kept sneaking up on us. We measured some MC2 motion -> REFL11 Transfer Functions to see if we could see CARM plant features; plots will come in the near future...

  9854   Fri Apr 25 10:43:57 2014 KojiUpdateLSC(Fixed) Y end whitening board

I went to WB and found the last spare module of D990399 revB. We need to thank Frank for his foresight.

The original (=broken) board had various modifications from this revB.
I had to check the schemaric diagram and the difference between the boards and migrate some of the SMD components from left to right.


Here is the deciphered features of the QPD whitening board:
- The input stage is a VGA amp (AD602). It has the internal input impedance of 100 Ohm. The series resister
  of 909 Ohm gives us 1/10 voltage division! It is more tricky as the QPD (D990272) has the output impedances of 50Ohm
  (for the both side of the differential out) and on resistance of MAX333A. So it could have been deviated by ~10% from the nominal.

- Variable gain control: The input has 1/10 voltage division. The gain is fixed at the unity. In total the gain of the variable control stage is 1/10.
  This gives us the gain range of +42dB/-22dB for +10V/-10V. The actual range is limited to be -10~30dB.

- Whitening stages. Each channel has two sets of the whitening path and the bypass path.
  They could be switched by binary control inputs but I permanently enabled the whitening by pulling the MAX333 control inputs to the ground.
  The whitening zero and pole are at 4.02Hz and 40.6Hz.

  Each bypass path has an additional cap of 220pF in parallel to 35.7kOhm (R101 and R103 for CH1), resulting in the pole at 20.2kHz.
  Each whitening paths had a 5.6nF cap (C53 and C64). This cap was replaced with 350pF, resulting in the move of the pole freq from 800Hz to 12.7kHz.

- There are two anti-aliasing stages which were designed for 2kHz sampling rate. They are identical sallen key 2nd-order LPFs with fc=766Hz and Q=0.74 (~ butterworth).
  As all of these caps were removed, they are just voltage followers now.

- The final stage (AD620) has the gain resister of 16.5k. The gain is 1+(49.4k/16.5k) = 3.99.

- The 4pin lemo connector (J8) was removed from the board. We instead installed an isolated BNC connector on the panel for the thorlabs PD serving as the high gain PD.

- There is a daughter board for the high gain PD. This seems to be the butterworth low pass filter with fc=~30kHz.
  The differential output of the daughter board is connected to pin 17 and 18 of J10 (S5 Out and Rtn).

- The input of the daughter board is differential (AD620). Therefore the LEMO connectros next to the BNC were wrapped with Kapton tapes for isolation.

Board test at the workbench.

- The test required two dual power supply as the unit requires +/-5V and +/-15V.

- The four channels were tested with the signal injection. 1kHz input yielded 20mVpp across the AD602 input. The output of the 1st whitening stage was
  60mVpp. This makes sense as the gain of the AD620 is -10dB (1/10 and 10dB). The output of the 2nd whitening stage was 600mVpp.
  Finally the output of the output stage was confirmed to be 2400mVpp. This was confirmed for four channels.

- The daughter board output was also checked. The gain is the unity and flat upto ~10kHz.

Board installation

- Jenne installed the module. This time there was no smoke.


Gain mystery

- It was not sure how the whitening gains have been given.

- The corresponding database entry was found in /cvs/cds/caltech/target/c1auxey/ETMYaux.db as

grecord(ao,"C1:ASC-QPDY_S1WhiteGain")
grecord(ao,"C1:ASC-QPDY_S2WhiteGain")
grecord(ao,"C1:ASC-QPDY_S3WhiteGain")
grecord(ao,"C1:ASC-QPDY_S4WhiteGain")

- The gains for S2-S4 were set to be 30. However, C1:ASC-QPDY_S1WhiteGain was set to be 8.62068.
And it was not writable.

- After some investigation, it was found that the database was wrong. The DAC channel was changed from S100 to S0.
The corrected entry is shown here.

grecord(ao,"C1:ASC-QPDY_S1WhiteGain")
{
        field(DESC,"Whitening gain for QPDY Seg 1")
        field(DTYP,"VMIVME-4116")
        field(OUT,"#C0 S0 @")
        field(PREC,"1")
        field(EGUF,"42")
        field(EGUL,"-22")
        field(EGU,"dB")
        field(LINR,"LINEAR")
        field(DRVH,"30")
        field(DRVL,"-10")
        field(HOPR,"30")
        field(LOPR,"-10")
}

- Once c1auxey was rebooted, the S1 whitening gain became writable. Now all of the channels were set to be +30dB (max).

 

Attachment 1: D990399-B_40m.pdf
D990399-B_40m.pdf D990399-B_40m.pdf D990399-B_40m.pdf
Attachment 2: P4245552.JPG
P4245552.JPG
Attachment 3: P4245553.JPG
P4245553.JPG
Attachment 4: P4245551.JPG
P4245551.JPG
  9855   Fri Apr 25 13:18:08 2014 Dark JamieUpdateLSClocking activity

Quote:

[ericq, Jenne, Zach]

We spent some time tonight trying to push our CARM locking further, to little avail. DARM/CARM loop oscillations kept sneaking up on us. We measured some MC2 motion -> REFL11 Transfer Functions to see if we could see CARM plant features; plots will come in the near future...

 Probably things would have worked better if you would have gotten your hair done at the same place as me.

Attachment 1: m10008_f1_bg.jpg
m10008_f1_bg.jpg
  9859   Sun Apr 27 19:53:54 2014 ericqUpdateLSCPRFP YArm Locking

Inspired by a comment by Koji the other day, I spent some time yesterday and today working on locking a (very lossy) power recycled Y-arm. ITMX was misaligned, to save myself the headache of dealing with ITMY getting a sign flip and ITMX staying the same when the arm resonates. 

My main goal was to achieve high bandwidth control with the analog CARM servo. 

TL,DR: Transisitoned 90% to REFLDC through CM_SLOW at TRY = 2.1 twice. Couldn't make it all the way over. 


PRCL settings:

  • Input: REFL165 I.
  • Actuate on PRM +1
  • Control: G=-.32 (~100Hz UGF); Acq on FM 4,5; Trig 1,2,3,6,9 (I modified the +10dB in FM1 to a 1kHz ELP)
  • Trig: POP 110 I: 1.5 up, 0 down (max was around 4 counts, very weak PRC!)

The PRC was very stable in this configuration, which doesn't surprise me due to its simplicity. I was honestly a little surprised there was enough light to lock on 3f. REFL33 didn't work. 


My efforts to bring the Y-arm into lock were very similar to the CARM procedure we've been using recently. (Which is the motivation for this exercise)

At first I was actuating on ETMY, and got to the point where I wanted to start bringing in the CARM servo slow output, then realized that I didn't want to actuate both on the ETM and MC AO. (Maybe this would be doable, but in the end, not what I'm interested in learning about in terms of overlap with CARM locking)

From then on, I only actuated YARM on MC2. (Heads up, my lock-losses will show up in the trends of the MC2 Trans addition to the WFS.)

Transitioning the arm to SqrtInv TRY control was just as straightforward as it has been for CARM. However, engaging the LSCLock FM (FM4), would sometimes work beautifully, and sometimes kick the hell out of MC2. Keeping an eye on the error signal spectrum and UGF gave no indication which outcome would happen. Once FM4 could be engaged, the transmitted power was very stable. Without FM4, reducing the offset didn't get very far without losing lock. 

I tried a few times to bring in CM_slow (set to just IN2, i.e. offset adjusted REFLDC), at arbitrary arm powers, with little success. I didn't know how much arm power to expect at resonance, and thus didn't really know where on the line width I was.

I knew I was mostly outside of the linear regime of the PDH signals, since, even though I had good coherence between, say, REFL11 I and SqrtInvTry, with an ETMY excitation on; when I would turn TRY normalization on/off, I would see the sign of the TF change. 

I then realized that I could actively keep an eye on the trend of POY11, to see when I got to the PDH "hump", which is where REFLDC starts being usable, and SqrtInv is reaching its limit. 

This brought me to a YARM offset of .115, with a steady TRY of about 2.1. I adjusted the analog offset of the REFLDC input to the CARM board, and the digital gain of the CMSLOW input filter to get 1:1 correspondence between CMSLOW and the SQRTINVY channels. Their spectra were neigh identical, with CMSLOW having slightly more high frequency noise. 

I started stepping SQRTINV down by .1, and upping CMSLOW by .1. This shifted the offset around, so I opted for taking away gain before bringing it back, because I didn't want to get so close to resonance that SQRTINV would freak out. I got to .1*SQRTINVY + .9*CMSLOW, and lost lock. TRY was getting noisier as I made the transition. 

I'm not sure what exactly was the reason for failure. I'm going to go back over some of the data to try to get an idea.. Maybe I should've loosened up some of the gain/boosts during the transition. 


So, no great success story yet, but this configuration is a lot simpler than the full PRFPMI, and I feel that I should soon be able to get it fully controlled, and figure out a systematic way to make the digital to analog transition for this PRFP cavity, and thus have a much more informed basis for doing the same for CARM control. 

  9860   Sun Apr 27 20:26:19 2014 KojiUpdateLSCPhase Tracker servo characterization

The measured open loop TF of the ALS Phase Tracker loop for each arm was characterized by an empirical model on LISO.

The model for the open loop TF has pole 1m instead of the one at DC as LISO has a difficulty to model it.
Digital time delay and the sampling effect seem to be well represented by a zero at ~8kHz and delay of  ~60us.
(cf 16kHz sampling => 61us)

The XARM phase tracker has the UGF of 1.5kHz. This is too low because
1) The phase rotation at 100Hz is visible in the plot.
2) We don't much care about the closed loop bump in the phase tracker as long as the phase tracker keeps its continuity.

So I suggest to increase the gain so that we have the UGF of 3kHz. (phase margin: 24deg)

The red curve in the plot is the closed loop response calculated by CLTF =  - OLTF / (1-OLTF).

The model results are used in the ALS servo models.

Attachment 1: ALSX_PTTF.pdf
ALSX_PTTF.pdf
Attachment 2: ALSY_PTTF.pdf
ALSY_PTTF.pdf
  9861   Sun Apr 27 21:30:59 2014 KojiUpdateLSCALS servo characterization

The measured openloop TF of the ALS servo for each was characterized by a ZPK model.

The openloop TF can be modeled by:

1) Filter TF obtained from foton
2) Actuator response with appropriate assumption
3) Phase tracker closed loop TF
4) Delay caused by the digital control
5) anything else

For 1) ZPK models of the servo filter was obtained from foton. It turned out that the TF of FM5 doesn't match with the ZPK model in foton.
Therefore the TF was exported and fitted with LISO. This seems to be related to the pole frequency (3kHz) which is too close to Nyquist frequency (8kHz).

FM(:,1)  = zero1(f,5).*pole1(f,0.001)*5000;
FM(:,2)  = zero1(f,1).*pole1(f,0.001)*1000;
FM(:,3)  = zero2(f,4.5,1.4619).*pole1(f,0.001).*pole1(f,0.001)*20.2501*1e6;
FM(:,4)  = zero2(f,35,2).*pole2(f,3,3).*zero1(f,3000).*pole1(f,1).*pole2(f,3000,1/sqrt(2)).*pole1(f,700).*zero1(f,10).*zero1(f,350).*136e1;
FM(:,5)  = zero1(f,1).*pole1(f,4.010e3).*pole2(f,17.3211e3,1.242).*zero2(f,18.865e3,100e3);
FM(:,6)  = zero2(f,3.2,0.966775).*pole2(f,3.2,30.572);
FM(:,7)  = zero2(f,16.5,2.48494).*pole2(f,16.5,78.5807).*zero2(f,24.0,2.22483).*pole2(f,24.0,7.03551);
FM(:,8)  = 1;
FM(:,9)  = zero2(f,7.50359,1.07194).*pole2(f,1.43429,0.717146)*27.5653;
FM(:,10) = 1;

dc_gain = 14;

FM1/2/3/5/6/7/9 are used for the control.

For 2), a resonant freq of 0.97 with Q of 5 was assumed.

The model for 3) was obtained by the previous entry.

Now the measured TF was divided by the known part of the model 1) ~ 3) and empirically fitted in LISO.

### XARM ###
pole 392.5021429051 698.1992431753m
zero 42.3128869460k 31.0954443799m
pole 589.2716424428 2.8325268375
factor 8.3430140244
delay 34.7536691023p

### YARM ###
pole 416.2463334253 743.2196174175m
zero 97.9161062704M 114.6703921876m
pole 626.0463515310 2.7671041771

factor 9.0045911761
delay 34.0945727358p

These compensation TF have weird TF. Probably the frequency response of the delay and the analog AA/AI filters without the high frequency data
led the LISO make up this. I'm requesting Masayuki to provide the AA/AI data to make the estimation more reasonable.
For the servo modeling, this is sufficient and we'll go a head.

The results of the OLTF modeling are attached.

Attachment 1: ALSX_OLTF.pdf
ALSX_OLTF.pdf
Attachment 2: ALSY_OLTF.pdf
ALSY_OLTF.pdf
  9862   Mon Apr 28 10:24:10 2014 KojiUpdateLSCerror signal characterization

As we now have the loop model, we can characterize the error signals.

We have the following data:

1) Free-running ALS error signals (i.e. phase tracker output) calibrated in Hz (for 532nm) (blue)
2) Controlled ALS error signals calibrated in Hz (for 532nm) (red)
3) ALS error signals measured with X and Y arm locked with the IR PDH. (black)
    This is likely to represent the sensing noise of the beatnote detection

from 2) we can derive the similar quantity as 1)
4) Estimated free-running ALS error signals from the controlled signals (green)

Remarks:

- From 1) and 4) we can see that the phase tracker is not perfectly linear. It seems that fast fringing of the phase tracker is causing upconversion.

- From 2) and 3) the servo loops don't have enough gain between 3Hz and 20Hz. On the other hand they have too much gain bekow 3Hz.

Attachment 1: ALSX_SPE.pdf
ALSX_SPE.pdf
Attachment 2: ALSY_SPE.pdf
ALSY_SPE.pdf
  9863   Mon Apr 28 10:34:51 2014 KojiUpdateLSCnew ALS servo design

Based on the evaluation of the error signals, the new servo was designed.

Concept:
- Don't touch the locking filters. (i.e. FM5)
- Sacrifice some phase at 150Hz to increase the gain between 3-20Hz.
- As resonant gains costs the phase without increasing the LF gains, replace them with a poles for the integrators.


FM(:,1) = zero2(f,.5,.7).*pole2(f,0.001,.7)*(0.5/0.001)^2;
FM(:,2) = zero2(f,5,2).*pole2(f,3,3).*pole1(f,1).*zero1(f,5)*5*(5/3)^2;
FM(:,3) = zero2(f,25,.7).*pole2(f,3.2,10)*(25/3.2)^2; % Zero crossing
FM(:,4) = zero2(f,35,2).*pole2(f,3,3).*zero1(f,3000).*pole1(f,1).*pole2(f,3000,1/sqrt(2)).*pole1(f,700).*zero1(f,10).*zero1(f,350).*136e1;
FM(:,5) = zero1(f,1).*pole1(f,4010).*pole2(f,17.3211e3,1.242).*zero2(f,18.865e3,100e3);
FM(:,6) = zero2(f,5,2).*pole2(f,10,2).*pole2(f,16.5,30).*zero2(f,30,2);
FM(:,7) = 1;
FM(:,8) = 1;
FM(:,9) = 1;
FM(:,10) = 1;
dc_gain = 14;

FM1/2/3/5/6 are expected to be used for the control.


FM1: Boost below 0.5Hz. This does not cost the phase margin.
FM2: Increase the gain below 5Hz. This hardly cost the phase margin.
FM3: Boost below 25Hz. This is the main phase cost at UGF. This has a complex pole pair at 3Hz with Q=10 to supress the stack motion.
FM6: zero-pole-pole-zero combination to boost the gain between 5 to 30Hz. This eats the phase margin a little.

Note that the phase tracker gain for the X arm was increased by factor of 2.5.

Attachment 1: ALSX_OLTF_new2.pdf
ALSX_OLTF_new2.pdf
Attachment 2: ALSY_OLTF_new2.pdf
ALSY_OLTF_new2.pdf
  9864   Mon Apr 28 10:48:48 2014 KojiUpdateLSCnew ALS servo design: comparison

Comparison of the new and old servo OLTF
The new servo has the same UGF, slightly less phase margin, and more gain between 1.5 and 25Hz.

Attachment 1: ALSX_OLTF_new.pdf
ALSX_OLTF_new.pdf
Attachment 2: ALSY_OLTF_new.pdf
ALSY_OLTF_new.pdf
  9865   Mon Apr 28 10:59:54 2014 KojiUpdateLSCNew ALS servo design: expected error signals

The expected error signals derived from the estimated free running error signals of the ALS.

1) Previously estimated free-running noise (blue)
2) Previous in-loop ALS error signal (red)
3) Estimated error signal with the new servo (green)
4) Out-of-loop noise of the ALS with the arm controlled with the IR PDH (black)

Now the error signal (green) is expected to be very white.
The suppressed noise between 3 to 20Hz are below the sensing noise level.
There seems a little excess at 24.5Hz and 28Hz. If it is limiting the RMS, we need to take care of them.

Attachment 1: ALSX_SPE_new.pdf
ALSX_SPE_new.pdf
Attachment 2: ALSY_SPE_new.pdf
ALSY_SPE_new.pdf
  9866   Mon Apr 28 11:03:57 2014 KojiConfigurationLSCNew ALS servo implemented for the X arm

The new ALS/LSC servo was implemented for the X arm.

I'll upload more data later but here I make quick remarks:

- We need to give the gain of 12 to have correct UGF with the ALS.

- With this servo, the Xarm PDH lock oscillates with the gain of 0.02. We need to lower the gain to 0.015.
  Also FM trigger should be changed not to trigger unused FMs (FM7/8)

  9867   Mon Apr 28 11:08:11 2014 KojiUpdateLSCNew ALS servo design: expected error signals

Here are the MATLAB scripts and LISO codes for all of these servo analyses

Attachment 1: 140421_ALS_servo.zip
  9868   Mon Apr 28 13:18:18 2014 JenneUpdateLSCLSC offsets script modified

Quote:

The weird jumps at the beginning of each TRX peak are due to the triggered switching between the Thorlabs trans PD and the QPD trans PD.  Clearly we need to work on their relative normalizations.  There are also little jumps after each peak as the triggering sends the signal back to the Thorlabs PD.

 I was unhappy with the discontinuities between the Thorlabs and QPD versions of our transmitted light powers.  I realized that in the olden days, we just used the Thorlabs PD, and we set the no-light offset in the LSC version of the TR[x,y] filter banks.  However, now that we have brought the QPDs back, we are setting the dark offsets in the end suspension models, so that the signal chosen by the trigger already has its offset taken care of before we send it to the LSC model. 

Anyhow, having the offsets script try to put a value in the C1:LSC-TR[x,y]_OFFSET was giving us an extra offset and then when we did the normalizations, the numbers came out all wrong.  So.  I have removed the C1:LSC-TR[x,y] filter banks from the offset list, since they were made redundant. 

I have redone the normalizations for both arms (after running the ASS scripts).  I checked by watching the _OUT16 versions of the Thorlabs and QPD diodes before the triggering happens, and as I put offsets into the LSC servos to change the transmitted power, the diodes both change in the same way.  So, we'll have to see if this holds true for more than just values 0-1 the next time we lock, but hopefully it won't need changing for a while.

  9872   Mon Apr 28 23:05:03 2014 JenneUpdateLSCALS CARM and DARM settings

[Jenne, Koji]

The IFO is being uncooperative tonight, and I have an early morning meeting, so I'm calling it a night. 

Koji's filter module changes have been propagated from the Xarm to the Yarm, to CARM and to DARM.  (Actually, Q overwrote the changes to Xarm on Sunday accidentally, so first he reverted those for us, and then we propagated the changes). 

Today, with careful measuring, we find that for X and Y arms individually locked with the ALS, we want the gains to be +17 for the Yarm, and -17 for the Xarm (with the beatnote up-is-up convention).  This puts the UGFs at 150 Hz. 

We then switched over to CARM and DARM locking.  We guessed that the gains should be a factor of 2 lower since we're pushing on both ETMs for DARM, and the MC2 actuator is roughly the same strength as the sum of the ETMs.  In the end, after measuring the CARM and DARM loops, we find that the gains should be +7.5 for CARM, and +8.0 for DARM to set the UGFs at 150 Hz.  The servo is a little bit delicate, so having too low of gain is not okay. 

For some reason, we seem to be utilizing more actuator range with the new setup, so the limiters in the filter banks have been set to 11,000 (previously were 8,000), and the ALS watch script (ALSdown.py) threshold has been increased to 10,000 (previously 7,000). 

When finding the IR resonances with the new scheme, we are having trouble holding lock throughout the scan.  I have set the tramp for the coarse part of the scan to be 0.05 seconds (previously 0.01 seconds), which is an increase of a factor of 5 in the ramp time.  This helps, but may still not be enough, since we don't always hold lock until both IR resonances are found.

Probably the most annoying thing from tonight is the fact that ETMY keeps drifting off, particularly in yaw, when locked.  I don't have an explanation of why this is happening, but you can watch it happen sometimes, and the lock will be lost shortly thereafter.  Definitely when we lose lock and the ETM gets kicked, it is far enough away in yaw alignment that I have to completely redo the Yarm alignment.  This happens whether or not the ETMY oplevs are on.

To summarize, 3 scripts have been modified:

(1) ALSdown - threshold increased  (Modification from last week - turns off the slow temp servos for the end lasers, clears histories)

(2) ALSfindIRresonance - increase ramp time

(3) Lock_ALS_CARMandDARM - final gain values set to 7.5 for CARM and 8 for DARM, no filters come on until gains all the way up, turns on new set of Koji filters. (Modification from last week - turns on the slow temperature servos for the end lasers)

  9874   Tue Apr 29 01:10:16 2014 KojiConfigurationLSCNew ALS servo implemented for the X arm

New ALS servo performance

Attachment 1:

Comparison between the old (orange) and new (red). The new error signal (red) is suppressed like a white noise as expected.

Comparison between the out-of-loop evaluation (black) and the in-loop signal (red). Below 50Hz the out-of-loop is limited by the sensor-noise like something.
This out-of-loop stability was measured with the ALS stayed at the top of the resonance and calibrated the POX11 error signal.

Attachment 2:

New ALS servo with the LSC PDH signal. When the PDH signal is used for the control, FM4 is additionally used.
In this condition, the error signal was measured and calibrated into frequncy noise (Hz/sqrtHz).

By comparing the POX (with the new servo) and POY (with the old servo) signals, one can see that the new servo has better supression below 30Hz with almost no cost at ~100Hz.

Attachment 1: ALSX_SPE.pdf
ALSX_SPE.pdf
Attachment 2: ALSX_PDH_SPE.pdf
ALSX_PDH_SPE.pdf
  9877   Wed Apr 30 00:40:55 2014 manasaConfigurationLSCY arm IR lock troubleshooting

[Koji, Manasa]

The Y arm locks stably for IR PDH now.  

The reason for ETMY getting kicked during lock acquisition was finally found to be related to the limiter value set in the Y arm servo. We reduced the limiter value unintentionally and found that the lock acquisition stayed smooth. The limiter value was stepped in 1000s from 7000 and eventually found that the ETMY suspension was kicked when we try to acquire lock with the limiter value was set at 11000. 

 

The limiter for X arm at 11000 is not causing any trouble at the moment.

In the process, we did a bunch of things through the evening to troubleshoot IR locking of the Y arm.

Earlier today running the IFO configure script did not restore the arms to lock and both the ETMs needed to be aligned to lock the arms. The arms stayed locked for 15 minutes and the Y arm lost lock eventually leaving the ETMY in a misaligned state. 

The state of the Y arm was similar to what Jenne has explained in ELOG where the ETMY was kicked during lock acquisition and would move to a misaligned state.

To trouble shoot, there were several things that were done. A few of them might not have any direct correlation to the locking issue but could just be a coincidence.  

1.  The trigger time for the filters in the arm filter modules were set such that they switch on after the SUS violin filters. Arm FM trigger time = 3 s (previously set at 0.1s) and SUS violin trigger time = 1s. This reduced the number of lock loss events.

2. There was some drop in transmission when the bounce filter of Y arm (FM6) turned ON. This was fixed by changing its ramp time (initially set at 1s). The filter has been modified to turn on immediately upon arm lock acquisition before the other triggered filters in the filter module turn on.

3. The QPD and SUS signal cables running to the rack were checked to be intact. Koji found some of them to be loose. But this had no evident correlation with the arm locking problem.

4.The oplev and PD alignment was checked at the Y end. The high gain trans PD for Y arm was checked for good alignment by looking at TRY. It was found that the EXIT light at the Y end is injecting some noise to the transmission PD. 

5. The ETMY was given offsets in PIT, YAW and POS and the OSEM sensor values were checked to see if the suspension is behaving well. It was behaving well.

  9880   Wed Apr 30 16:07:59 2014 manasaUpdateLSCALS X noise post servo modification

I. The Y arm stayed stable through last night and I have saved the arm lock settings to IFOconfigure.

II. ALS X arm noise measurements

I looked at the before and after noise of ALSX.

Settings:
Phase tracker gain = 85
Xarm servo gain = -17

The rms in loop noise has dropped from 3KHz to 500 Hz.

Attachment 1: Phase tracker OLTF

Attachment 2: Free running noise and in loop noise

Attachment 3: Out of loop noise (measured with arms locked using PDH for IR)

Attachment 4: ALS arm servo OLTF

xml data files can be found in /users/manasa/data/140430/

Attachment 1: ALSX_PToltf.jpg
ALSX_PToltf.jpg
Attachment 2: ALSX_FreeInloop.jpg
ALSX_FreeInloop.jpg
Attachment 3: ALSX_ool.jpg
ALSX_ool.jpg
Attachment 4: ALSX_OLTF.jpg
ALSX_OLTF.jpg
  9884   Wed Apr 30 21:16:42 2014 ranaUpdateLSCMC2 Strad

bettsreplica.jpg

I found the YARM LSC feedback going to MC2 and the MC2 violin mode (at 644.69 Hz) rung up. The existing notch was just a second order Twin-T style notch (so not a good idea) and also not turned on, since it was in the FM4 spot of LSC-MC2 and the vio triggers are ganged between all mirrors and don't touch FM4.

I copied the PRM vio bandstop into FM2 of this bank, deleted the old notch, and tuned the bandstop frequencies a little to get the violin peak into one of the zeros of the elliptic bandstop. Attached are the Y-arm / MCF spectrum with the mode rung up as well as the new filter's TF compared with the old notch.

P.S. I installed http://en.wikipedia.org/wiki/Midnight_Commander on pianosa.

Attachment 2: MC_Y_vio.pdf
MC_Y_vio.pdf
Attachment 3: MC2_vio.pdf
MC2_vio.pdf
  9887   Thu May 1 00:13:21 2014 KojiUpdateLSCALS X beat setup aligned

I saw big misalignment on the GTRX camera, I went to the PSL table and aligned the beat beams.

I disconnected the RF out of the X beat PD and  connect an oscilloscope.
The beat amplitude was 15mVpp at the beginning and is 60mVpp right now.
I checked the alignment on this RF PD and the DC PD as well as the spot on the CCD.

The RF cable was connected again.

Jenne and I ran the ALS and scanned the arm cavity. We had the impression that the noise level of the ALS improved,
but I don't have correctly calibrated measurement. Let's do it tomorrow in the day time.

The Yarm beat alignment look awful. We should align this too.

  9888   Thu May 1 03:15:03 2014 JenneUpdateLSCYarm locking with CM board

[Rana, EricQ, Jenne]

We locked the Yarm by using the CM board this evening. 

POY is going from its demod board to the CM board, and then the slow output of that is going to the POY channel of the whitening, and then on to the ADC.  So, with no AO path engaged, this is basically like regular Yarm locking. 

First of all, Den and Koji back in December were concerned that they were seeing some EOM saturation in the fast path, but we don't think that's an issue.  We looked at the FSS PCDRIVE while we increased the AO gain.  In fact, it looks like the offset is coming from the MC board's IN2 slider.  Even with no input on that slider, increasing its value puts an offset into the MC.  To fix this, I am going to put a 6.8uF cap in series with R30 in the MC board, which is part of the crossbar switch where the IN1 and IN2 get summed.  This should AC-couple the output of the IN2 slider before the summing node.

We aren't sure which sign to use for the AO path of the CM board...Eric is doing some modelling to see if he can figure it out.  He's going to try to see which spectra (below) his model matches.

For the spectra, we have a reference trace with no AO path, a trace with "Plus" polarity on the CM board which started to show a peak when the value of the MC IN2 slider was at about -6 dB, and a trace with "Minus" polarity, which started to show a peak when the value of the MC IN2 slider was at about -16 dB. 

Yarm_CMlocking_spectra_30Apr2014_copy.pdf

We took loop measurements for each of the Plus and Minus cases. Something that seems a little weird is how shallow of a slope we have in both cases near our UGF.

Yarm_CMlocking_TFs_30Apr2014_copy.pdf

 

ELOG V3.1.3-