40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 154 of 344  Not logged in ELOG logo
ID Date Author Type Category Subject
  9585   Wed Jan 29 16:36:37 2014 KojiSummaryGeneralHigh power beam blasting of the aLIGO RFPD

[Rich, Jay, Koji]

We blasted the aLIGO RF PD with a 1W IR beam. We did not find any obvious damage.
Rich and Jay brought the PD back to Downs to find any deterioration of the performance with careful tests.

The power modulation setup is at the rejection side of the PBS in front of the laser source.
I checked the beams are nicely damped.
As they may come back here tomorrow, a power supply and a scope is still at the MC side of the PSL enclosure.

  9584   Tue Jan 28 23:32:12 2014 KojiUpdateGeneralX/Y arm locked with the IR beam

[Koji EricQ]

The both arms have been locked with IR and aligned by ASS.

The IFO was left with ITMX/Y, ETMX/Y, BS, and PRM aligned, and the PSL shutter closed.


YARM
SIGNAL PATH:
POY11I(+45dB)->YARM(G=+1.0)->ETMY
NORM: TRYx10
TRIG: TRY 0.01up/0.005down
FM TRIG: FM2/3/6/7/8/9 0.01up/0.05down, 0.5 sec delay

XARM
SIGNAL PATH:
POX11I(+45dB)->XARM(G=+4.0)->ETMX
NORM: TRXx10
TRIG: TRX 0.01up/0.005down
FM TRIG: FM2/3/6/7/8/9 0.01up/0.05down, 0.5 sec delay


For decent locks, it was necessary that the offset of the error signals are trimmed at the input filters
even after running LSCoffset.py script.

Once the cavities were aligned for the IR, we could see the green beams are also flashing.
The Y arm was actually locked with the green with a TEM00 mode

Attachment 1: good_alignment.png
good_alignment.png
  9583   Tue Jan 28 22:24:46 2014 ericq UpdateGeneralFurther Alignment

[Masasa, ericq]

Having no luck doing things remotely, we went into the ITMX chamber and roughly aligned the IR beam. Using the little sliding alignment target, we moved the BS to get the IR beam centered on ITMX, then moved ITMX to get good michelson fringes with ITMY. Using an IR card, found the retroflection and moved ETMX to make it overlap with the beam transmitted through the ITM. With the PRM flashing, X-arm cavity flashes could be seen. So, at that point, both the y-arm and x-arm were flashing low order modes. 

  9582   Tue Jan 28 16:26:40 2014 SteveUpdateVAC vacuum computers are back without safety

Quote:

[Steve Koji]

We pushed the reset button of c1vac1 and c1vac2 and the vacuum screen is back.

First, we pushed the reset button of c1vac1 and pushed the one on c1vac2.
This did not bring c1vac2 up. We pushed the reset of c1vac2 again and now everything of the vacuum screen is back.

 1, Reset works

 2, Default values are lost. They actually reversed into open and turn on everything when power recycle c1vac1, c1vac2 or 24V dc power supply. This can vent the IFO in an event of power failure!

      It's may be the time to go back to an isolated, vacuum controller only computer.

 

The present valve configuration at Atm1

a, all annuloses are vented

b, valve cables disconnected at: VC1 and 4 IP gate valves

c, the RGA is off and it is pumped by the Maglev through VM2

d, cryo pump is being roughed with TP3  It's outgassing rate was 25 mTorr / min

e, Roughing hose is disconnected

Attachment 1: c1vac1_vac2areback.png
c1vac1_vac2areback.png
  9581   Tue Jan 28 11:13:50 2014 KojiUpdateVAC vacuum monitor is still blank

[Steve Koji]

We pushed the reset button of c1vac1 and c1vac2 and the vacuum screen is back.

First, we pushed the reset button of c1vac1 and pushed the one on c1vac2.
This did not bring c1vac2 up. We pushed the reset of c1vac2 again and now everything of the vacuum screen is back.

  9580   Tue Jan 28 09:51:47 2014 SteveUpdateIOOlow power pointing

 

PSL output is stable.

Attachment 1: IOO4dlowpr.png
IOO4dlowpr.png
  9579   Mon Jan 27 21:36:35 2014 ericq UpdateGeneralVent so far

After turning the slow FSS threshold down, the mode cleaner stays locked enough to do other things. We were able to align the tip tilts to the y-arm such that we were able to get some flashes in what looks like a TM00-ish mode. (It was necessary to align the PRM such that there was some extra power circulating in the PRC to be able to see the IR flashes on the ITMY face camera) This is enough to convince us that we are at least near a reasonable alignment, even though we couldn't lock to the mode. 

The x-arm was in a hairier situation; since the green beam wouldn't flash into any modes, we don't even know that a good cavity axis exists. So, I used the green input PZTs to shine the green beam directly on the earthquake stops on the ITMX cage, and then inferred the PZT coordinates that would place the green beam roughly on the center of ITMX. I moved the ETMX face camera such that it points at the ETMX baffle. I tried looking for the retroreflected green spot to no avail. Hopefully tomorrow, we can get ourselves to a reasonably aligned state, so we can begin measuring the macroscopic PRC length. 

  9578   Mon Jan 27 17:49:46 2014 ranaHowToComputer Scripts / Programssendmail started on nodus: fixing SwiftMail on Dokuwiki

Since the recent filesystem fracas, the new accounts could not be created on nodus / dokuwiki (for the controls workshop, for example).

I started sendmail on nodus using the command:   sudo /etc/init.d/sendmail start

and the SwiftMail plugin on there is now sending out the confirmation emails again. This will happen each time we reboot nodus, so let's replace it.

  9577   Mon Jan 27 12:26:00 2014 KojiUpdateIOOIOO Slow Actuator Servo threshold changed

In order to activate the slow actuator servo for the MC locking,
the threshold level for this servo (C1:PSL-FSS_LOCKEDLEVEL) was changed from 10000 to 700.

Now the servo started to move the PZT fast out to be controlled to 5V.

  9576   Mon Jan 27 09:08:00 2014 SteveUpdateVAC4 days at atmosphere

 

 

Attachment 1: day4atm.png
day4atm.png
Attachment 2: vent76bgTp3d4.png
vent76bgTp3d4.png
  9575   Sat Jan 25 21:09:16 2014 gabrieleHowToLSCProcedure to measure PRC length

I know the drawing is wrong. I put random distances, not realistic ones, and I did not try to get something close to reality. Once we put the measured distances, the drawing should (hopefully) be correct.

  9574   Fri Jan 24 13:10:12 2014 JamieHowToLSCProcedure to measure PRC length

Quote:

I wrote a MATLAB script that takes as input the measured distances and produce the optical path lengths. The script also produce a drawing of the setup as reconstructed, showing the measurement points, the mirrors, the reference base plates,  and the beam path. Here is an example output, that can be used to understand which are the five distances to be measured. I used dummy measured distances to produce it.

map.pdf

This path does not look correct to me.  Maybe it's because this is supposed to represent "optical path lengths" as opposed to actual physical location of optics, but I think locations should be checked.  For instance, PRM looks like it's floating in mid-air between the BS and ITMX chambers, and PR2 is not located behind ITMX.  Actually, come to think of it, it might just be that ITMX (or the ITMs in general) is in the wrong place?

Here is a similar diagram I produced when building a Finesse model of the 40m, based on the CAD drawing that Manasa is maintaining:

path.pdf

  9573   Fri Jan 24 12:44:25 2014 GabrieleHowToLSCProcedure to measure PRC length

Here is how to measure the PRC length with a set of distance measurements in the optical setup. 

We need to take distance measurements between reference points on each mirror suspension. For the large ones (SOS) that are used for BS, PRM and ITMs, the reference points are the corners of the second rectangular base: not the one directly in contact with the optical bench (since the chamfers make difficult to define a clear corner), but the rectangular one just above it. For the small suspensions (TT) the points are directly the corners of the base plates.

From the mechanical drawings of the two kind of suspensions I got the distances between the mirror centers and the reference corners. The mirror is not centered in the base, so it is a good idea to cross check if the numbers are correct with some measurements on the dummy suspensions.

I assumed the dimensions of the mirrors, as well as the beam incidence angles are known and we don't need to measure them again. Small errors in the angles should have small impact on the results.

I wrote a MATLAB script that takes as input the measured distances and produce the optical path lengths. The script also produce a drawing of the setup as reconstructed, showing the measurement points, the mirrors, the reference base plates,  and the beam path. Here is an example output, that can be used to understand which are the five distances to be measured. I used dummy measured distances to produce it.

map.pdf

In red the beam path in vacuum and in magenta the beam path in the substrate. The mirrors are the blue rectangles inside the reference bases which are in black. The thick lines are the HR faces. The green points are the measurement points and the green lines the distances to be measured. The names on the measurement lines are those used in the MATLAB script. 

The MATLAB scripts are attached to this elog. The main file is survey_v2.m, which contains all the parameters and the measured values. Update it with the real numbers and run it to get the results, including the graphic output. The other files are auxiliary functions to create the graphics. I checked many times the code and the computations, but I can't be sure that there are no errors, since there's no way to check if the output is correct... The plot is produced in a way which is somehow independent from the computations, so if it makes sense this gives at least a self consistency test. 

Attachment 2: survey_v2.m
global sos_lx sos_ly sos_cx sos_cy tt_lx tt_ly tt_cx tt_cy

%% Survey of the PRC length

%% measured distances
d_MB2_MY  = 2000.0;
d_MB3_MX  = 2000.0;
d_MB1_M31 = 400.0;
d_M32_M21 = 3000.0;
d_M22_MP  = 2000.0;
... 210 more lines ...
Attachment 3: distance.m
function d = distance(c1, c2)
    d = sqrt(sum((c1-c2).^2));
end
Attachment 4: draw_beam.m
function draw_beam(c1, c2, color)
    plot( [c1(1), c2(1)], [c1(2), c2(2)], color, 'LineWidth', 2)
end
Attachment 5: draw_measurement.m
function draw_measurement(c1, c2, color, name)
    plot( [c1(1), c2(1)], [c1(2), c2(2)], color)
    text( (c1(1)+c2(1))/2, (c1(2)+c2(2))/2 + 20, name, ...
        'FontSize', 5, 'HorizontalAlignment', 'center', ...
         'VerticalAlignment', 'middle')
end
Attachment 6: draw_point.m
function draw_point(c)
    plot(c(1), c(2), 'go', 'LineWidth', 2, 'MarkerSize', 3);
end
Attachment 7: draw_sos.m
function draw_sos(C, angle)
    global sos_lx sos_ly sos_cx sos_cy tt_lx tt_ly tt_cx tt_cy

    c(:,1) = [-sos_lx/2, -sos_ly/2 + sos_cy-sos_ly/2]';
    c(:,2) = [-sos_lx/2, sos_ly/2 + sos_cy-sos_ly/2]';
    c(:,3) = [sos_lx/2, sos_ly/2 + sos_cy-sos_ly/2]';
    c(:,4) = [sos_lx/2, -sos_ly/2 + sos_cy-sos_ly/2]';
    c(:,5) = [-sos_lx/2, -sos_ly/2 + sos_cy-sos_ly/2]';
    
    m_lx = 25.4*2;
... 18 more lines ...
Attachment 8: draw_tt.m
function draw_tt(C, angle)
    global sos_lx sos_ly sos_cx sos_cy tt_lx tt_ly tt_cx tt_cy

    c(:,1) = [-tt_lx/2, -tt_ly/2 + tt_cy-tt_ly/2]';
    c(:,2) = [-tt_lx/2, tt_ly/2 + tt_cy-tt_ly/2]';
    c(:,3) = [tt_lx/2, tt_ly/2 + tt_cy-tt_ly/2]';
    c(:,4) = [tt_lx/2, -tt_ly/2 + tt_cy-tt_ly/2]';
    c(:,5) = [-tt_lx/2, -tt_ly/2 + tt_cy-tt_ly/2]';
    
    m_lx = 25.4;
... 18 more lines ...
  9572   Thu Jan 23 23:10:19 2014 ericq UpdateGeneralVent so far

[ericq, Manasa, Jenne]

Summary: We opened up the BS and both ITM chambers today, and put the light doors on. //Edit : Manasa  Post-vent the MC was very much misaligned in yaw. Both the ITMs moved in pitch as inferred from the oplev; but there is still light on the oplev PDs//. We toiled with the PMC and mode cleaner for a while to get reasonable transmission and stability (at least for a period of time). We then tried to lock IR to the y-arm, to no avail. 

Locking the PMC doesn't seem very robust with the low power level we have; adjusting the gain at all when it's locked throws it right out. The mode cleaner spot was visibly moving around on MC2 as well. We'll continue tomorrow. 

Details about alignment efforts: Manasa and I tried for a while to try and align the y-arm for IR. Straight out of venting the green TM00 would lock to the y-arm with about .45, as compared to .8 before venting, so it didn't seem to drift too far. The x-arm would even flash any modes, however. For a while, IR was no where to be seen after the mode cleaner. Eventually, we used the tip tilts to bring the AS beam onto the camera, which exhibited fringes, so we knew we were hitting the ITMs somewhere. We wandered around with the ETM to see if any retroflection was happening, and saw the IR beam scatter off of the earthquake stop. We moved it to the side to see it hitting the OSEM holder, and moved down to the bottom OSEM holder to get an idea of where to put pitch to get roughly the center of the ITM, then undid the yaw motion.

There, we would see very infrequent, weak flashes. We weren't able to distinguish the mode shape though; however, the flashes were coincident with where the green would lock to a very yaw-misaligned fishbone mode, to the lower right of the optic's center. We figured that if we gradually fixed the green alignment with the mode shapes we could see and actually lock on, we could use the tip tilts to adjust the IR pointing and keep it coincident and eventually resonate more. However, this didn't really work out. The flashes were very infrequent, and at this point the PMC/MC were getting very touchy, and would cease to stay locked for more than a minute or two. At this point, we stopped for the day. 

 

  9571   Thu Jan 23 13:23:10 2014 SteveUpdateVAC40m IFO is at atmosphere

Quote:

 

There is BLANK VacControl_BAK.adl screen only. 

I can move a valve by disconnecting it's  solenoid power if it's position is normally open.

I will close V1 and check computer cable connections and move on with manual - hand disconnect ea valve to be moved into the right position for vent. Valve positions will be confirmed by looking manual indicators on valves.

 The 40m vacuum envelope vent is completed with instrument grade air.

Valve configuration: chamber open, RGA is pumped through VM3 by TP3,

  9570   Thu Jan 23 08:04:04 2014 SteveUpdateVACvacuum control screen is blank

 

There is BLANK VacControl_BAK.adl screen only. 

I can move a valve by disconnecting it's  solenoid power if it's position is normally open.

I will close V1 and check computer cable connections and move on with manual - hand disconnect ea valve to be moved into the right position for vent. Valve positions will be confirmed by looking manual indicators on valves.

Attachment 1: pd76m170d2dRgaOn.png
pd76m170d2dRgaOn.png
  9568   Wed Jan 22 20:00:41 2014 JenneUpdateGeneralVENT GO!

Steve, please begin the vent!!

[EricQ, Jenne]

We have followed the pre-vent checklist, and done everything except check the jam nuts (which Steve can do in the morning).

We are ready to vent, so Steve, please begin bringing us up to atmosphere first thing in the morning.

Here is a copy of the list, from the wiki:

 

 

  • Center all oplevs/IPPOS/IPANG
  • Align the arm cavities for IR and align the green lasers to the arms. (Green powers were both ~0.8.  We only touched the Xend PZTs remotely, did not touch Yend).
  • Make a record of the MC pointing
  • Align the beam at the PSL angle and position QPDs (Did not need doing, left QPDs as-is so we keep our long-term trend.)
  • Reduce input power by touching wave plate on the PSL table BEFORE THE PMC.  (HWP was at 269degrees, now at 3 hundred something to make power just before PSL shutter 90mW)
  • Replace 10% BS before MC REFL PD with Y1 mirror and lock MC at low power.
  • Close shutter of PSL-IR and green shutters at the ends
  • Make sure the jam nuts are protecting bellows
  •  

     

    Attachment 1: IFOstatus_lowPower_preVent.png
    IFOstatus_lowPower_preVent.png
      9567   Wed Jan 22 18:17:46 2014 JenneUpdateCDSfb timing was off

    Since this morning, the fb's timing has been off.  Steve pointed it out to me earlier today, but I didn't have a chance to look at it until now. 

    This was different from the more common problem of the mx stream needing to be restarted - that causes 3 red blocks per core, on all cores on a computer, but it doesn't have to be every computer.  This was only one red block per core in the CDS FE status screen, but it was on every core on every computer. 

    The error message, when you click into the details of a single core, was 0x4000.  I elog searched for that, and found elog 6920, which says that this is a timing issue with the frame builder.  Since Jamie had already set things on nodus' config correctly, all I did was reconnect the fb to the ntp: 

    fb$ sudo /etc/init.d/ntp-client restart

    As in elog 6920, the daqd stopped, then restarted itself, and cleared the error message. It looks like everything is good again.

    I suspect (without proof) that this may have to do with the campus network being down this morning, so the computers couldn't sync up with the outside world.

      9566   Wed Jan 22 16:36:45 2014 ericqUpdateElectronicsRF distribution box power button fail

    Quote:

    Rana, Gabriele and I are trying to measure the FSR of the PRC (elog about that later), and we turned off the power to the RF generation box so that we could switch cables at the EOM combiner.  However, as in elog 9101, the power button won't latch when we try to turn the power back on.  All 3 of us tried, to no avail.  For our measurement, poor Gabriele is standing holding the button pushed in, so that we can have some RF sidebands. 

    Tomorrow, we'll have to pull the RF generation box, and put in a better switch.

    I replaced the stupid broken fancy button with a simple sturdy switch. I had to file out the hole in the chassis a bit, but the switch is pressed in tightly and securely. I put the box back in the rack, but the power cable was coming directly from the power supplies with no fuses. The box was drawing ~.9 and 1.5 Amps from two supplies, so I put 2A fuses on both. Plugged everything back in, and the mode cleaner locks, so it looks like all is well.

    RXA: When its so close, I prefer to size it up by 1 step. Please change to 5A fuses. Otherwise, we may blow them from power glitches.

    Q: 5A fuses have been swapped in

      9565   Wed Jan 22 15:24:11 2014 SteveSummaryVACRga scan after reboot

    Quote:

    [Jenne, Steve]

    Steve noticed that the RGA was not logging data and that not all the correct connection lights were on, and he wasn't able to run the "RGAset.py" script (in ...../scripts/RGA/) that sets up the proper parameters. 

    I looked, and the computer was not mounting the file system.  I did a remote shutdown, then Steve went in and pushed the power button to turn the machine back on.  After it booted up, it was able to talk to the file system, so I started ..../scripts/RGA/RGAset.py .  The first 2 times I ran the script, it reported errors, but the 3rd time, it reported no communication errors.  So, now that the computer can again talk to the file system, it should be able to run the cronjob, which is set to take data at 4am every day.  Steve will check in the morning to confirm that the data is there.  (The last data that's logged is 22Dec2013, 4am, which is right around when Koji reported and then fixed the file system).

     

     

     We are venting tomorrow. This give us an opportunity to fix sick vacuum controller computer. Jamie volunteered.

    Remember to rough down cryo and ion pump volumes. Their pressure can be at 1 Torr range after years of accumulated outgassing.  Without total valve controls it is dangerous to have these air pockets. Some of their  gate valves can be leaking and that would explane the slower pumpdown speed. 

     

    Attachment 1: Rgascan169dwarmingup.png
    Rgascan169dwarmingup.png
    Attachment 2: BlankMedmMustBeFixed.png
    BlankMedmMustBeFixed.png
      9564   Wed Jan 22 09:05:42 2014 GabrieleConfigurationLSCREFL11 back

     Yesterday night I plugged back the REFL11 RF cable into the corresponding demodulation board.

      9563   Tue Jan 21 19:41:59 2014 JenneUpdateElectronicsRF distribution box power button fail

    Rana, Gabriele and I are trying to measure the FSR of the PRC (elog about that later), and we turned off the power to the RF generation box so that we could switch cables at the EOM combiner.  However, as in elog 9101, the power button won't latch when we try to turn the power back on.  All 3 of us tried, to no avail.  For our measurement, poor Gabriele is standing holding the button pushed in, so that we can have some RF sidebands. 

    Tomorrow, we'll have to pull the RF generation box, and put in a better switch.

      9562   Tue Jan 21 17:26:59 2014 JenneSummaryVACRebooted RGA computer and reset RGA settings

    [Jenne, Steve]

    Steve noticed that the RGA was not logging data and that not all the correct connection lights were on, and he wasn't able to run the "RGAset.py" script (in ...../scripts/RGA/) that sets up the proper parameters. 

    I looked, and the computer was not mounting the file system.  I did a remote shutdown, then Steve went in and pushed the power button to turn the machine back on.  After it booted up, it was able to talk to the file system, so I started ..../scripts/RGA/RGAset.py .  The first 2 times I ran the script, it reported errors, but the 3rd time, it reported no communication errors.  So, now that the computer can again talk to the file system, it should be able to run the cronjob, which is set to take data at 4am every day.  Steve will check in the morning to confirm that the data is there.  (The last data that's logged is 22Dec2013, 4am, which is right around when Koji reported and then fixed the file system).

     

     

      9561   Fri Jan 17 11:44:25 2014 GabrieleSummaryLSCMore length measurements, more confusion

     I analyzed the data taken yesterday. 

    The AS11 data in PRMI configuration is very bad, while the AS55 seems good enough:

    results_as11.pdfresults_as55.pdf

    The phase differences are 

    AS11 = 21 +- 18 degrees (almost useless due to the large error)

    AS55 = 11.0 +- 0.4 degrees

    The AS55 phase difference is not the same measured in the last trial, but about half of it. The new length estimates are:

    AS11 = 3.2 +- 2.8 cm

    AS55 = 0.47 +- 0.01 cm

    We can probably forget about the AS11 measurement, but the AS55 result is different from the previous estimate... Maybe this is due to the fact that Eric adjusted the PRCL offset, but then we're going in the wrong direction....

      9560   Thu Jan 16 21:38:13 2014 ericqUpdateLSCRepeat of PRC length measurement

    [ericq,Jenne]

    Since we don't have agreement between the measurements we made the other day and the earlier estimations, I wanted to repeat the demodulation angle measurement. We had to do a few things to keep the PRMI locked, since in the last few days, it hasn't been stable enough.

    The mode cleaner had been very fussy lately; the WFS were pushing in a way that caused fast oscillations of the transmission and reflection powers. I turned off the servos, manually aligned the mode cleaner to transmission of about 15k and refl of about .4, centered the beams on the WFS QPDs, and turned the loops back on. Things were much stable after that. Also, Jenne noticed that the PMC loop had walked the laser PZT temperature to a bad place, and fixed it.

    After aligning the carrier locked PRMI, the last piece needed to get things stable enough for sideband locking was turning off the angular damping on the PRM suspension screen (this was turned back on when we were done). Waiting until evening noise levels probably helped too. We used a 1000 count MICH excitation in the PRMI case, and recorded data for about a minute in one degree steps around the demodulation phase that looked to put the excitation entirely within the Q of the PD. Also, we notched out the excitation frequency in the MICH servo bank for today's measurement; I think it's outside of the loop bandwidth anyways, but it's good to be sure. 

    Jenne and I pondered a bit whether changing the AS55 demodulation phase while it (AS55 Q) is being used as the MICH control signal introduces subtleties that we haven't anticipated, but couldn't come up with anything concrete. Changing the angle from the what maximizes the Q just looks like a slight change in MICH gain, and shouldn't affect the phase of the excitation signal on the PD...

    In any case, the data have been recorded, and the results will follow soon. 

      9559   Thu Jan 16 08:19:29 2014 SteveUpdatePEM4.4 and 3.8M local earth quakes


    It looks like there was a 4.4 magnitude earthquake near Fontana, CA around 1:30am today.  This tripped all of the suspension watchdogs, which Q has just now re-enabled. 

       Earth quake shake down yesterday Atm1

    Atm2, today's shake

    Attachment 1: local4.4eq.png
    local4.4eq.png
    Attachment 2: local3.8eq.png
    local3.8eq.png
      9558   Wed Jan 15 18:42:57 2014 JenneUpdateASCPOP ASC QPD offline for a few hours this afternoon

    I was in the lab, near the south end of the ITMX oplev table, looking for something, and I bumped the POP ASC QPD's power supply.  I thought that it was fine, but did not adequately check it.  When EricQ asked me just now about why the PRC is so wobbly today, I checked, and the power for the QPD wasn't properly connected (it's kind of a crappy connector, that if you nudge, contacts or loses contact).  Anyhow, I restored power to the QPD, and the PRC looks a little more stable now.  My fault for not checking more carefully, and my apologies to Q and Gabriele for their frustrations this afternoon.

      9557   Wed Jan 15 18:18:15 2014 GabrieleSummaryLSCPRC length measurement analysis

     I analyzed the data we took yesterday, both using AS11 and AS55. For each value of the phase I estimated the Q/P ratio using a demodulation code. Then I used a linear regression fit to estimate the zero crossing point.

    Here are the plots of the data points with the fits:

    experimental_data_as11.pdfexperimental_data_as55.pdf

    The measurements a re more noisy in the PRMI configuration, as expected since we had a lot of angular motion. Also, the AS11 data is more noisy. However, the estimated phase differences between PRMI and MICH configurations are:

    • AS11 = -10.9 +- 1.0 degrees
    • AS55 = -21.1 +- 0.4 degrees

    The simulation already described in slogs 9539 and 9541 provides the calibration in terms of PRC length. Here are the curves

    simulation_phase_difference.pdf

    The corresponding length errors are

    • AS11 = 1.44 +- 0.13 cm
    • AS55 = 0.59 +- 0.01 cm

    The two results are not consistent one with the other and they are both not consistent with the previous estimate of 4 cm based on the 55MHz sideband peak splitting.

    I don't know the reason for this incongruence. I checked the simulation, repeating it with Optickle and I got the same results. So I'm confident that the simulation is not completely wrong.

    I also tried to understand which parameters of the IFO can affect the result. The following ones have no impact

    • Beam matching
    • ITM curvatures
    • Schnupp asymmetry
    • Distance PR-BS
    • ITM and PRM misalignments

    The only parameters that could affect the curves are offsets in MICH and PRCL locking point. We should check if this is happening. A first quick look (with EricQ) seems to indicate that we indeed have an offset in PRCL. However, tonight the PRMI is not locking stably on the sidebands. 

    If possibile, we will repeat the measurement later on tonight, checking first the PRCL offset.

     

      9556   Wed Jan 15 13:05:30 2014 JenneUpdatePEM4.4 EQ in Fontana, CA

    It looks like there was a 4.4 magnitude earthquake near Fontana, CA around 1:30am today.  This tripped all of the suspension watchdogs, which Q has just now re-enabled. 

      9555   Tue Jan 14 19:10:51 2014 GabrieleSummaryLSCPRC length measurement

     [EricQ, Gabriele]

    We could carry out the measurement of PRC length. The AS110 photodiode was plugged into REFL11. So REFL11 is giving us the AS11 signal. Here is the procedure.

    1. Lock MICH.
    2. Add a line in MICH (amplitude 20000 counts)
    3. Tune AS11 demod phase to have the line in I.
    4. Change the demod phase by steps of 1 degree around the rough optimum, taking one minute of data at each point
    5. Lock PRMI on sidebands
    6. Add a line in MICH (amplitude 500 counts)
    7. Tune AS11 demod phase to have the line in I.
    8. Change the demod phase by steps of 1 degree around the rough optimum, taking one minute of data at each point

    We repeated the same measurement also using AS55, with the same procedure.

    Roughly, the phase difference for AS11 was 11 degrees and for AS55 it was 23 degrees. A more detailed analysis and a calibration in terms of PRC length will follow.

      9554   Tue Jan 14 19:05:48 2014 GabrieleSummaryLSCPRMI locked on sideband

    [Gabriele, EricQ] 

    Finally, we managed to lock PRMI on sidebands:

    • MICH locked on AS55_Q with gain -10. Demod phase of AS55=18
    • PRCL locked on REFL55_I with gain -0.04. Demod phase of REFL55=88
    • Triggering on POP110_I at 50/10
    • Filter "1:5" of MICH engaged, this improved a lot the stability

     

      9553   Tue Jan 14 10:34:57 2014 SteveUpdatePSLgreen transmission measurment

    GariLyn is using our green light on the west side of the PSL table. The green PDA36As were moved and the HEPA turned up to 60V

    Attachment 1: greenPickUp.jpg
    greenPickUp.jpg
      9552   Tue Jan 14 10:12:12 2014 SteveUpdatePSLlaser drift monitor set up idea

    Quote:

    this locationQuote:

    Quote:

    Quote:

    I wonder what's drifting between the laser and the PMC? And why is it getting worse lately?

     The PMC refl is bad in pitch today, and the transmission is only 0.76, rather than our usual 0.83ish.

    I did a quick, rough tweak-up of the alignment, and now we're at 0.825 in transmission.

     The PMC transmission continuously degrades. In order to see what is really drifting the laser output after PBS was sampled as shown.

     IOO pointing is drifting in pitch. I'd like to use a QPD instead of the paper target to see if the Innolite output is stable. The idea is to move temporarily IOO-QPD_POS to  this location

     I do like to move IOO-QPD_POS temporarily to see that the feedback has anything to do with with the pointing.

    Attachment 1: bad4thday.png
    bad4thday.png
      9551   Mon Jan 13 19:31:04 2014 GabrieleSummaryLSCPRMI locking

    [Gabriele, EricQ]

    We wanted to try the PRC length measurement,but we ended up spending all the afternoon to lock the PRMI on sidebands. Here are some results

    • Lock of PRM on carrier is easy with MICH on AS55_Q and PRCL on REFL55_I. The optimal gains to avoid correction gain peaking are MICH=-5 and PRCL=0.012. The alignment today was must more stable over time than on Friday
    • We wanted to move MICH on REFL55_Q. After a few trials we discovered that REFL55_Q is not seeing any MICH signal at all. This is quite strange and we don't understand this.
    • Locking PRMI on carrier using REFL165_I and Q was difficult. We thought that was due to a RF amplifier installed in December. We tried to remove it, but since this did not help, we put it back
    • We could lock PRMI on carrier and sidebands using REFL11 signals. The optimal demodulation phase is 155. Lock on carrier was achieved with gains MICH=5, PRCL=0.7. Lock on sidebands with MICH gain=-10 and PRCL gain=-0.3. The MICH correction is very high and exciting a tone at few hundreds Hz. Maybe a violin mode of the ITMs?
    • Finally we could tune the REFL165 phase to 118.5 and lock on carrier. The lock was not very stable. Gain are: lock on carrier MICH=-0.5, PRCL=0.05. The error signal has a very big offset: to get a dark fringe we had to add a MICH offset of 500 counts. We also had to engage the CLP400 filter to avoid saturating ITMs corrections with MICH.
      9550   Mon Jan 13 16:50:55 2014 SteveUpdateVACMaglev controller needs service

    Quote:

     The date is correct on this monitor.

    Last stored RGA scan data Dec 20, 2013

    IFO pressure at CC1 5.8e-6 Torr

    Valve configuration: Vacuum Normal, confirmable only by manual checking of position indicators and pressure gauge controllers  readouts

     

     The Osaka TG390MCAB maglev turbo pump's controller TC010M has passed the 40,000 hrs of operation. This triggered the " alarm" LED  warning light to come on. 

    It is normal maintenance.  Maglev TP-1 is running perfectly.  Osaka will send us a loaner-controller that we can use while they do the std maintenance.

    I'm thinking of ~ February to do this.

      9549   Mon Jan 13 11:08:48 2014 SteveUpdatePSL3 good days of IOO pointing

     Three good days of IOO pointing: Friday, Sat and Sun    What was changed?  May be the clamping on Friday?

    IOO vertical changes recovering as tempeture. IP is clipping at plastic enclosure of ETMY

     

    NOTE: ANTS at the PSL optical table.  I will mop with chemicals tomorrow if we see more.

     

    Attachment 1: 3gdPSLpointing.png
    3gdPSLpointing.png
      9548   Sun Jan 12 09:57:24 2014 GabrieleSummaryLSCEffect of PRC length mismatch on error signals

    Quote:

     

     Its very doubtful that the MC yaw drift matters for the IFO. That's just a qualitative correlation; the numbers don't hang together.

     Then there must be something else slowly drifting. It was very clear that the good alignment of the IFO was every time lost after few minutes...

      9547   Fri Jan 10 15:33:02 2014 SteveUpdatePSLlaser drift monitor set up idea

    this locationQuote:

    Quote:

    Quote:

    I wonder what's drifting between the laser and the PMC? And why is it getting worse lately?

     The PMC refl is bad in pitch today, and the transmission is only 0.76, rather than our usual 0.83ish.

    I did a quick, rough tweak-up of the alignment, and now we're at 0.825 in transmission.

     The PMC transmission continuously degrades. In order to see what is really drifting the laser output after PBS was sampled as shown.

     IOO pointing is drifting in pitch. I'd like to use a QPD instead of the paper target to see if the Innolite output is stable. The idea is to move temporarily IOO-QPD_POS to  this location

    Attachment 1: 2daysDrift.png
    2daysDrift.png
      9546   Fri Jan 10 15:31:07 2014 ranaSummaryLSCEffect of PRC length mismatch on error signals

     

     Its very doubtful that the MC yaw drift matters for the IFO. That's just a qualitative correlation; the numbers don't hang together.

      9545   Fri Jan 10 10:28:03 2014 SteveUpdatePSLPSL pointing changes

     

    I looked at IOO QPDs again. QPD_POS was clamped by one screw. Dog clamp was added on the unclamped side.

    QPD_ANG chassis has no isolation to optical table..._POS has.  QPD_ANG  base was tightened also.

    Both QPDs moved a little bit but I did not centered them.  The spot sizes are 2-3 mm  They should be smaller.

    How ever, we still can not explane the pitch movement of the IOO beam

     

    Razor beam dumps were labeled at the AP table.

     

    The 40m roof was cleaned from leafs this morning.

     

     

    Attachment 1: clamped.png
    clamped.png
      9544   Thu Jan 9 17:58:31 2014 ericqSummaryLSCEffect of PRC length mismatch on error signals

    [ericq, Gabriele, Manasa]

     We wanted to perform the PRC length measurement today with an AS11 signal, but such a signal didn't exist. So, we have temporarily connected the AS110 PD signal (which is some Thorlabs PD, and not a resonant one) into the REFL11 demod board. 

    We then proceeded with the goal of locking the PRC with REFL165. A few parameters that were changed along the way as we aligned and locked things:

    • the XARM gain was increased from 0.4 to 0.5 to help it acquire lock
    • the MICH gain was decreased from -10 to -5 since there was some gain peaking in its servo output
    • the REFL165 demodulation phase was changed from 155 to 122, to place a PRCL excitation entirely within I (we did this while locked on the carrier)

    Sadly, in the end, we couldn't lock the PRC on a sideband in a stable manner. The alignment would drift faster than we could optimize the alignment and gains for the PRC. I.e. we would lock the PRC on the carrier, align PRM (and maybe touch ITMX) to maximize POPDC, switch to sideband locking, try to lock, and things would start looking misaligned. Switching back to carrier locking, the beam spots on REFL (for example) would have moved.

    Manasa noted the MC_TRANS_Y has been substantially drifting along with small drift in MC_TRANS_P as well. So we need to fix the source of the mode cleaner beam drifting if we want to make this measurement. 

      9543   Thu Jan 9 17:21:45 2014 ranaUpdateGeneralIFO plan, IPANG telescope

    Quote:

    For the IPANG telescope design, we are in the 'beyond the Rayleigh range' regime. So using a single lens to make the beam small is not a great idea. I

    Can you please explain this? I don't understand what exactly is the issue or 'great idea'.

    I think we should be OK with just a single lens in the vacuum. But what we need is the ray tracing analysis to show what the effect will be on the IPANG readout.

      9542   Thu Jan 9 10:34:58 2014 SteveUpdatePSLPSL pointing changes in pitch

      IOO QPDs tested in dark, lighted and open PSL enclosure. The created temperature change 0.03 C has  effect on monitoring  in pitch.

     

     Atm1,  all lights off 10 min, PSL enclosure lights on  10 min, all lights off 15 min, open  door # 11 at north east corner of enclosure ( HEPA filters are running at 30V ) for 10 min, closed-dark enclosure 15 min

                  dark 10, lighted 10, dark 15, open-dark 10 and closed-dark 15 minutes

     

    Atm2, Pitch drift of 24 hours does not recover

    Attachment 1: Lfnfdnc.png
    Lfnfdnc.png
    Attachment 2: 24hPSLpointing.png
    24hPSLpointing.png
      9541   Wed Jan 8 19:05:30 2014 GabrieleSummaryLSCEffect of PRC length mismatch on error signals

     [Gabriele, EricQ]


    Actually it is difficult to see any laser frequency line in the dark fringe signal, since the Schnupp asymmetry is small. It is much better to use a differential MICH excitation which gives a better signal at the dark port.

    We repeated the simulation explained before. We can use both the AS55 or the AS11 signals, bout the first one has a limited linear range and the expected 4cm value is very close to saturation.

    as11.pngas55.png

      9540   Wed Jan 8 17:53:26 2014 manasaUpdateGeneralIFO plan, IPANG telescope

    For the IPANG telescope design, we are in the 'beyond the Rayleigh range' regime. So using a single lens to make the beam small is not a great idea. I have put down a solution where we use a pair of lenses; one of which will be mounted in-vacuum in the ETMY chamber and the other on the endtable.
    This way we will also allow have some freedom to configure the layout out-of vacuum in case the need arises. The layout will look something like in the cartoon:
    IPANG_layout.png

    I also made a choice of using longer focal length lenses (CVI 2" lenses f =1 m). Below is the beam path summary for IPANG telescope. I have used the waist diameter at the ITM for propagation. The endtable is roughly at 41.2m. The QPD will be placed in front of the waist (w0=47um).
    IPANG.png 

      9539   Wed Jan 8 16:08:52 2014 ericqSummaryLSCEffect of PRC length mismatch on error signals

     [ericq,Gabriele]

    So, we want an relatively quick measurement of the PRC length error (with sign!) at the order of .5 centimeter or so. Rana suggested the "demodulation phase method," i.e. lock the simple Michelson, measure what demodulation phase brings the 1F signal entirely within the phase quadrature, then lock the PRMI and measure the demodulation phase again. This tells you something about the length of the PRC. 

    Gabriele and I worked through a simulation using MIST to determine how to actually do this. We simulated the case of injecting a line at 1kHz in the laser frequency via the laser's PZT and looking at the transfer function of the 1kHz signal to the I and Q at the 1F AS demodulated signal when locked. (Michelson locked on the dark fringe, PRC locked on 11MHz sideband) With the I and Q in hand, we can measure some demodulation phase angle that would bring everything into I. 

    When the PRC length is in the ideal location, the demodulation phases in the two cases are the just about the same. Sweeping the length of the PRC around the ideal length gives us a monotonic function in the difference in the demodulation phases:

    phaseVlength.pdf

    So, with this simulation, we should be able to calibrate a measured difference in demod phase into the length error of the cavity! We will proceed and report...

      9538   Wed Jan 8 13:46:39 2014 JenneUpdateGeneralIFO plan, PRM baffle

    Quote:

    While we're in there, we need to also put a baffle on the back side of the PRM cage, to protect the OSEMs from stray light.  Den and I discovered before Christmas that turning off the OSEM and OpLev damping to the PRM (while using the POP QPD for ASC) significantly reduced the power fluctuations in the PRC.  We still had arm power fluctuations, but I believe those are likely because our ALS system can't hold an arm precisely at full resonance.  So, putting a black glass baffle with ~2 inch aperture right up against the OSEMs should help a lot.  This week, I'll ask Steve to make me a quickie to-scale cardboard version of the baffles that he has had cut, so I can try securing it to the dirty suspension cage that we have out.  I will also check to make sure I have seen with my own eyes the baffles that I need, and copper wire to tie it to the cage.

    Steve may actually be onto something with the clamps that he had made a year and a half ago.  These clamps hold the glass, and then clamp to the base of the suspension cage.  Not the table, but the base of the suspension cage. The drawings are in elog 6344.  I'm not sure that the 1/4-20 holes in the clamp things are exactly where we'll want them, but we should be able to just dog it down to the base of the suspension.  I need to check this, but it may be even easier than tieing the glass to the cage.

    Also, something to think about is that the earthquake stop screws extend backwards farther than the OSEMs.  I'm not sure anymore if we have shorter 1/4-20 earthquake stops around (if we do, they should be in the cleanroom shelves), but if we can't swap those out, they'll limit how close we can get to the OSEMs. 

    Here's an overhead photo from 6 Sept 2012:

    PRMcage_6Sept2012.JPG

      9537   Wed Jan 8 13:01:48 2014 GabrieleSummaryLSCEffect of PRC length mismatch on error signals

    I ran a simulation of a double cavity with a PRC length mismatched w.r.t. the modulation frequency. I summarized the results in the attached PDF. I think it would be important to have a cross check of the results.

    In brief:

    A mismatch between PRC length and modulation frequency do have an effect on error signals

    Multiple zeros appear in REFL_3f/PRCL that can be removed by careful tuning of the demodulation phase (however, the shape of the signal makes difficult to understand which phase is good…)

    No visible effect on REFL_1f/CARM

    But a large PRCL signal appears in REFL_1f_I, which is used to control CARM. This is not good.

    A mismatch of the order of 0.5 cm has a small effect.

     

     

     

     

     

    Attachment 1: REFL_vs_PRClength.pdf
    REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf
      9536   Tue Jan 7 23:53:35 2014 JamieUpdateCDSdaqd can't connect to c1vac1, c1vac2

    dadq is logging the following error messages to it's log related to the fact that it can't connect to c1vac1 and c1vac2:

    CAC: Unable to connect because "Connection timed out"
    CA.Client.Exception...............................................
        Warning: "Virtual circuit disconnect"
        Context: "c1vac2.martian:5064"
        Source File: ../cac.cpp line 1127
        Current Time: Tue Jan 07 2014 23:50:53.355609430
    ..................................................................
    CAC: Unable to connect because "Connection timed out"
    CA.Client.Exception...............................................
        Warning: "Virtual circuit disconnect"
        Context: "c1vac1.martian:5064"
        Source File: ../cac.cpp line 1127
        Current Time: Tue Jan 07 2014 23:50:53.356568469
    ..................................................................
    

     Not sure if this is related to the full /frames issue that we've been seeing.

      9535   Tue Jan 7 23:50:27 2014 jamieUpdateCDS/frames space cleared up, daqd stabilized

    The wiper script is done and deleted a whole bunch of stuff to clean up some space:

    controls@fb ~ 0$ /opt/rtcds/caltech/c1/target/fb/wiper.pl --delete
    
    Tue Jan  7 23:09:21 PST 2014
    
    Directory disk usage:
    /frames/trend/minute_raw 385927520k
    /frames/trend/second 125729084k
    /frames/full 12552144324k
    /frames/trend/minute 2311404k
    Combined 13066112332k or 12759875m or 12460Gb
    
    /frames size 13460088620k at 97.07%
    /frames above keep value of 95.00%
    Frame area size is 12401156668k
    /frames/full size 12552144324k keep 11781098835k
    /frames/trend/second size 125729084k keep 24802313k
    /frames/trend/minute size 2311404k keep 620057k
    Deleting some full frames to free 771045488k
    - /frames/full/10685/C-R-1068567600-16.gwf
    - /frames/full/10685/C-R-1068567616-16.gwf
    ...
    controls@fb ~ 0$ df -h /frames
    Filesystem            Size  Used Avail Use% Mounted on
    /dev/sda1              13T   12T  826G  94% /frames
    controls@fb ~ 0$
    
    So it cleaned up 826G of space.  It looks like the fb is stabilized for the moment.  On site folks should confirm...
    

     

    asdfasdfsadf sadf asdf

    ELOG V3.1.3-