40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 155 of 350  Not logged in ELOG logo
ID Date Authorup Type Category Subject
  8813   Tue Jul 9 17:03:06 2013 SteveUpdateGreen Lockingfiber layed for Y arm

Alex, Gautam and Steve,

Single mode fiber 50m long is layed out into cable tray that is attached to the beam tube of the Y arm.

It goes from ETMY to PSL enclosure. It is protected at both ends with " clear- pvc, slit corrugated loom tubing " 1.5" ID

The fiber is not protected between 1Y1 and 1Y4

Attachment 1: fromETMYtowardPSL.jpg
fromETMYtowardPSL.jpg
Attachment 2: fibreETMYtoPSL50m.jpg
fibreETMYtoPSL50m.jpg
Attachment 3: PSLfiberfromETMY.jpg
PSLfiberfromETMY.jpg
  8822   Wed Jul 10 14:49:00 2013 SteveUpdateGreen Locking more fiber protective tubing layed

Quote:

Craig, Gautam and Steve,

Single mode fiber 50m long is layed out into cable tray that is attached to the beam tube of the Y arm.

It goes from ETMY to PSL enclosure. It is protected at both ends with " clear- nylon slit corrugated loom tubing " 1.5" ID

The fiber is not protected between 1Y1 and 1Y4

 Installed 0.5" ID 10 ft long protective tubing at the PSL end of the  ETMY fiber this morning. Here I had to cable tie a bunch of cables at the east side of the PSL enclosure.

They were hanging off the table blocking space were the sliding doors move.

 At the ETMX end of the X-arm fiber received the same protective tubing.

Attachment 1: fibreETMXprotectedendOnly.jpg
fibreETMXprotectedendOnly.jpg
  8827   Thu Jul 11 09:15:10 2013 SteveUpdateendtable upgradeETMY optable grounded

ETMY optical table top was grounded to the ETMY chamber through 1 Mohms this morning. I  also strain releifed relieved a few cables that were pulling on components directly.

  8828   Thu Jul 11 09:24:03 2013 SteveUpdateGeneralthere should be nothing left on optable tops

There are 4 oscilloscopes left on the AP optical table top.... It's only 25 lbs... Do not leave anything on the optical table tops!

  8834   Fri Jul 12 09:00:10 2013 SteveUpdateGeneral Halogen illumination cables disconnected

Alex and Steve,

Old halogen chamber illuminator cabling disconnected and potenciometer board removed at 1Y1 in order to give room for pd calibration fibre set up.

Attachment 1: 1Y1oldilluminatersDisconnected.jpg
1Y1oldilluminatersDisconnected.jpg
  8847   Mon Jul 15 14:12:15 2013 SteveUpdateVACRGA scan at day 144

 Pumpdown 75, vacuum normal condition at day 144

 

Attachment 1: pd75mRGA144d.png
pd75mRGA144d.png
Attachment 2: pd75m144d.png
pd75m144d.png
  8856   Tue Jul 16 13:48:26 2013 SteveUpdatePEMcranes cleaned

Keven and Steve,

The 3 cranes tested and  wiped off as preparation for upcoming vent.

  8904   Tue Jul 23 08:21:31 2013 SteveUpdateVACVent has started

 

Thanks for the good preparation.  IFO pressure P1=20 Torr

Attachment 1: beforVent.png
beforVent.png
Attachment 2: startVent.png
startVent.png
  8907   Tue Jul 23 14:00:03 2013 SteveUpdateVACthe vent is completed

 The 40m IFO has reached  atmospher in 5 hours. It is ready to open chamber condition. The RGA is pumped with the maglev.

P1 pirani gauge is contact dependent as you see it on the linear plot It will be replaced during this vent.

The venting speed was 2-4 Torr / min

Atm2 shows how the BS is sensing the venting air cylinder changes. 

The 4th cylinder of instrument  grade air  bump is overlapping with our janitor working at the BS chamber.

 

 

Attachment 1: Atm.png
Atm.png
Attachment 2: 5cyl.png
5cyl.png
Attachment 3: ventSum1d.png
ventSum1d.png
  8915   Wed Jul 24 10:35:41 2013 SteveUpdateVACBS, ITMY doors are removed

Quote:

We will open the BS and ITMY doors first thing tomorrow morning. I plan to try to be in around 9 am. The first order of business will be to flip the folding mirrors that are not currently flipped (SR2, SR3, PR3).

 Jenne, Annalisa & Steve

Attachment 1: beforeDoorsOff.png
beforeDoorsOff.png
Attachment 2: particlecount10d.png
particlecount10d.png
  8933   Mon Jul 29 16:04:40 2013 SteveUpdateVACAC removed

 Manasa, Eric, Evan, Koji and Steve,

Access connector removed in order to complete alignment. Light aluminum with acetate windows AC installed.

 

 

Attachment 1: acout.png
acout.png
Attachment 2: AlAcACa.jpg
AlAcACa.jpg
Attachment 3: AlAcACb.jpg
AlAcACb.jpg
  8945   Wed Jul 31 09:51:43 2013 SteveUpdatesafetySujan receives safety training

Sujan got 40m specific basic safety training this morning.

Attachment 1: IMG_0031.JPG
IMG_0031.JPG
  8946   Wed Jul 31 10:27:00 2013 SteveUpdateVACinstrument grade air is discontinoued

We have been using Air Liquide Instrument Grade Air for venting the 40m vacuum envelope. It is no longer available.

The replacement is Alphagas 1, total hydrocarbon <0.1% ppm

LINK TO Alphagaz

 

A- A+ Imprimer Partager

ALPHAGAZ™ Specifications

ALPHAGAZ™ range of Pure Gases for Analysis: two purity levels tailored to cover all your analytical needs.

Because each analytical application is sensitive to critical impurities that can affect your results, tests in our research center and analyzer manufacturers' studies confirm that the main impurities are moisture, oxygen and hydrocarbons.

Based on these critical impurities, Air Liquide has developed ALPHAGAZ™ brand with a Maximum Impurity Level (H2O, O2, CnHm).

  • The Maximum Impurity Level is the same across the whole product range. This constitutes a guarantee of quality with regard to major critical impurities and contamination tracers. ALPHAGAZ™ 2 offers an even greater guarantee in terms of purity (CO, CO2, H2).
  • The Minimum Total Purity is a prerequisite for product quality, but we do not stop there: Air Liquide selects and controls its gas sources by defining specifications regarding impurities that are critical for your analysis or application.

 

ALPHAGAZ™ 1 ALPHAGAZ™ 1 ALPHAGAZ™ 2 ALPHAGAZ™ 2 ALPHAGAZ™ AUTO IV ALPHAGAZ™ CO2 SFC
He, N2, H2,
Ar, O2
Air He, N2, H2,
Ar, O2
Air Air*, N2, O2 CO2
Specifications (ppm vol):
H2O < 3 ppm
O2 < 2 ppm
CnHm < 0.5 ppm
H2O< 3 ppm
CnHm< 0.1 ppm
CO2 < 1 ppm
CO < 1 ppm
H2O < 0.5 ppm
O2 < 0.1 ppm
CO2 < 0.1 ppm
CO < 0.1 ppm
CnHm < 0.1 ppm
H2 < 0.1 ppm
H2O < 0,5 ppm
CnHm < 50 ppb
CO2 < 0,1 ppm
CO < 0,1 ppm
NOx < 10 ppb *
SO2 < 10 ppb*
CnHm < 0.2 ppm
CO2 < 0.5 ppm
CO < 0.5 ppm
NO/NOx < 0.1 ppm
H2O < 5 ppm
O2 < 2 ppm
CnHm < 2 ppm
NVOC* < 2 ppm
CO < 5 ppm
NO + NO2 < 2 ppm
S Total < 1 ppm
H2 < 0.5 ppm
  20 ± 1 % O2 (Europe)
21 ± 1% O2 (North America)
  20,9 ± 1 % O2
*: Detection limit
*20 ± 1 % O2 *NVOC: Non Volatile Organic Compounds
Minimum Total Purity

 

 

  8953   Thu Aug 1 16:03:52 2013 SteveUpdateGeneralreopen BS & IOO

Quote:

[Koji, Manasa]

We missed to check that we had the green transmitted to the PSL after flipping the SRC and PRC folding mirrors.
There is no green transmission reaching the PSL even after locking the arms to green.

We should fix this tomorrow. The BS heavy door should come off.

Steve! Do not start pump down tomorrow !

 We closed the chambers last night with heavy doors and reopened it today.

Attachment 1: IOOnorth.jpg
IOOnorth.jpg
  8963   Mon Aug 5 10:50:48 2013 SteveUpdateVACRGA background

 RGA background at day 12 of this vent . The maglev is pumping on the rga through VM2

 

Attachment 1: v75bgMd12.png
v75bgMd12.png
  8970   Tue Aug 6 10:06:29 2013 SteveUpdateVACPumpdown is at 500Torr

 The pumpdown has started at 8:38am

Manasa was here to confirm good alignment of the IFO

Inner jam nuts of AC bellow were torqued to 45 ft/lbs and   door nuts were check on opened chambers.

Annulos were roughed down to 500 mTorr

Oplev servos turned off, PSL and green shutters closed before pumpdown started

 

Attachment 1: befPd.png
befPd.png
  8973   Tue Aug 6 16:45:31 2013 SteveUpdateVACPumpdown is completed

 IFO  P1=1mT PSL shutter is opened

 

Attachment 1: 8hrPd76.png
8hrPd76.png
Attachment 2: 1mT.png
1mT.png
  8976   Wed Aug 7 10:51:49 2013 SteveUpdateVACPumpdown at 26hrs

 Cold cathode gauge CC1 -h (horizontal) just coming on 9.2e-5 Torr

P2 is the fore line pressure of the maglev. One can see the 4 Torr load during switching over to turbo pumping.

CC4  5e-9 Torr is the performance of the maglev pumping on the RGA only.

The annuloses are not pumped  now.  They are closed off to see how much outgassing plus leak they have.

Configuration: vacuum normal, annuloses not  pumped

Condition: normal

Precondition: 14 days at atm, IOO chamber north door was taken off as a new entrance, the ETMX chamber was not opened.

What is new in the vacuum system: new P1 pirani gauge, gold plated clean allen wrench and ..........what else was dropped?

Note: the wireless laptop did not fail once all day yesterday. I want to give credit to the person who is responsible for this.

 

Attachment 1: 26hr.png
26hr.png
Attachment 2: 14datm.png
14datm.png
  8978   Wed Aug 7 15:36:29 2013 SteveUpdateVACTP2 drypump replaced

 

 TP2's fore line - dry pump replaced at performance level 600 mTorr after 10,377 hrs of continuous operation.

Where are the foreline pressure gauges? These values are not on the vac.medm screen.

The new tip seal dry pump lowered the small turbo foreline pressure 10x

TP2fl after 2 day of pumping 65mTorr

Attachment 1: forelinesPs.jpg
forelinesPs.jpg
  8984   Thu Aug 8 07:57:23 2013 SteveUpdateVACpumpdown at 48 hrs

Quote:

 

 TP2's fore line - dry pump replaced at performance level 600 mTorr after 10,377 hrs of continuous operation.

 

Attachment 1: 48hrsPd76.png
48hrsPd76.png
  8985   Thu Aug 8 10:31:28 2013 SteveUpdateVACfew reminders of this vent

 1, Vacuum envelope grounds must be connected all times!  After door removal reconnect both cables immediately.

 2, The crane folding had a new issue of getting cut as picture shows.

 3, Too much oplev light is scattered. This picture was taken just before we put on the heavy door.

 4, We were unprepared to hold the smaller side chamber door 29" od of the IOC

 5, Silicon bronze 1/2-13 nuts for chamber doors will be replaced. They are not smooth turning.

 

Attachment 1: GROUND!.jpg
GROUND!.jpg
Attachment 2: bad_Folding.jpg
bad_Folding.jpg
Attachment 3: toomuchred.jpg
toomuchred.jpg
Attachment 4: BETTERholder.jpg
BETTERholder.jpg
  9002   Tue Aug 13 07:40:53 2013 SteveUpdateCDSc1iscex needs help

 

 Sorrensen ps ouput of +15V at rack 1X9 was current limited to 10.3V @ 2A

Increased threshold to 2.1A  and the voltage is up to 14.7V

Attachment 1: c1iscexSick.png
c1iscexSick.png
  9003   Tue Aug 13 11:04:44 2013 SteveUpdatePEMfluorecent lights

Our fluorecent lights became obsolete.  We'll have change fixtures over to some more energy efficient one. Do you have any recommendation regarding to less noise performer unit?

We may go this direction of LED fluorecent lamps ?

  9011   Wed Aug 14 08:24:20 2013 SteveUpdateVACpumpdown at day 8

 

 

Attachment 1: pd76md8.png
pd76md8.png
  9013   Thu Aug 15 09:34:12 2013 SteveUpdateGeneralWilcoxon cables rescued

Eric and Steve,

 We removed Wilcoxon Accelerometer PS and Amplifier unit under the BS optical tabel yesterday. The six cabels going to DAQ  were labeled and left in place. Gain setting were 100, except channel 3 was 10.

The ~ 40 m long 2 sets of 3 cables were very happy to get their kinks out. Especially the set going just south of ITMX optical  table.

We have to take better care of these cables! Your data will be useless this way.

Attachment 1: rescuedGraycables.jpg
rescuedGraycables.jpg
Attachment 2: wilconoxOut.jpg
wilconoxOut.jpg
Attachment 3: chanGains.jpg
chanGains.jpg
  9017   Fri Aug 16 09:35:18 2013 SteveUpdateVACVacuum Normal state recognition is back

Quote:

Quote:

Quote:

Quote:

Apparently all of the ION pump valves (VIPEE, VIPEV, VIPSV, VIPSE) opened, which vented the main volume up to 62 mTorr.  All of the annulus valves (VAVSE, VAVSV, VAVBS, VAVEV, VAVEE) also appeared to be open.  One of the roughing pumps was also turned on.  Other stuff we didn't notice?  Bad. 

 Several of the suspensions were kicked pretty hard (600+ mV on some sensors) as a result of this quick vent wind.  All of the suspensions are damped now, so it doesn't look like we suffered any damage to suspensions.

CLOSE CALL on the vacuum system:

Jamie and I disabled V1, VM2 and VM3 gate valves by disconnecting their 120V solenoid actuator before the swap of the VME crate.

The vacuum controller unexpectedly lost control over the swap as Jamie described it. We were lucky not to do any damage! The ion pumps were cold and clean. We have not used them for years so their outgassing possibly  accumulated to reach ~10-50 Torr

I disconnected_ immobilized and labelled the following 6 valves:  the 4 large ion pump gate valves and VC1,  VC2  of the cryo pump. Note: the valves on the cryo pump stayed closed. It is crucial that a warm cry pump is kept closed!

This will not allow the same thing to happen again and protect the IFO from warm cryo contamination.

The down side of this that the computer can not identify vacuum states any longer.

This vacuum system badly needs an upgrade. I will make a list.

 While I was doing the oil change of the roughing pumps I accidentally touched the 24 V adjustment knob on the power supply.

All valve closed to default condition. I realized that the current indicator was red at 0.2A  and the voltage fluctuated from 3-13V

Increased current limiter to 0.4A and set voltage to 24V     I think this was the reason for the caos of valve switching during the VME swap.

 

 Based on the facts above I reconnected VC1 and VC2 valves.  State recognition is working.  Ion pumps are turned off and their gate valves are disabled. 

We learned that even with closed off gate valves while at atmosphere  ion pumps outgass hydrocarbons at 1e-6 Torr level.  We have not used them for this reason in the passed 9 rears.

 

I need help with implementing V1 interlock triggered by Maglev failure signal  and-or P2 pressure.

MEDM screen agrees with vacuum rack signs.

Attachment 1: VacuumNormal.png
VacuumNormal.png
Attachment 2: vacValvesDisabled.jpg
vacValvesDisabled.jpg
  9024   Mon Aug 19 07:53:48 2013 SteveUpdateSUSETMX damping restored

ETMX sus damping restored

  9026   Mon Aug 19 09:54:13 2013 SteveUpdatesafetyMasayuki receives safety training

Masayuki Nakano, a student of Seiji's from ICRR / U Tokyo, is visiting us here at the 40m lab for the next couple months.

He received 40m specific basic safety training this morning.

  9027   Mon Aug 19 10:03:17 2013 SteveHowToGeneralhow not to leave a cable

We can not leave cables connected like this. This is a burned toast award.

Attachment 1: NONOcable.jpg
NONOcable.jpg
  9029   Mon Aug 19 11:12:54 2013 SteveUpdateVACRGA scan at day 13

 

 

Attachment 1: rgaScan13d.png
rgaScan13d.png
  9034   Mon Aug 19 17:40:32 2013 SteveUpdateGreen LockingXend green layout corrections

 Shutter moved, no more clipping.

Pick-off mirror 2" replaced by 1" one. Laseroptik HR 532nm, incident angle 30-45 degrees, AR 532 nm

Green REFL PD moved to 4" close to pick-off mirror. Pd being close to pick-off does not separate multiple reflections on it. I'll replace Laseroptic mirror with Al one. It is not easy to find.

 Hole cut into side wall for doubler oven cable to exit.

 

 

Attachment 1: beforeC.jpg
beforeC.jpg
Attachment 2: nowC.jpg
nowC.jpg
Attachment 3: stillMultiple.jpg
stillMultiple.jpg
  9039   Tue Aug 20 10:59:15 2013 SteveUpdateGreen LockingXend green layout corrections

Quote:

 Shutter moved, no more clipping.

Pick-off mirror 2" replaced by 1" one. Laseroptik HR 532nm, incident angle 30-45 degrees, AR 532 nm

Green REFL PD moved to 4" close to pick-off mirror. Pd being close to pick-off does not separate multiple reflections on it. I'll replace Laseroptic mirror with Al one. It is not easy to find.

 Hole cut into side wall for doubler oven cable to exit.

 

 

 Beam trap for Pd refl is in place. Cabeling is ti·died up.

 Laseroptic 1" mirror is replaced by Al 1" mirror. Problem remains the same. This diffraction patter has to be coming from the Faraday.

  Atm1, good separation when Pd is far 

  Atm2, bad separation when Pd is close 

Attachment 1: faraway.jpg
faraway.jpg
Attachment 2: closer.jpg
closer.jpg
  9083   Wed Aug 28 11:15:02 2013 SteveUpdateGreen LockingXend green layout corrections

Quote:

Quote:

 Shutter moved, no more clipping.

Pick-off mirror 2" replaced by 1" one. Laseroptik HR 532nm, incident angle 30-45 degrees, AR 532 nm

Green REFL PD moved to 4" close to pick-off mirror. Pd being close to pick-off does not separate multiple reflections on it. I'll replace Laseroptic mirror with Al one. It is not easy to find.

 Hole cut into side wall for doubler oven cable to exit.

 

 

 Beam trap for Pd refl is in place. Cabeling is ti·died up.

 Laseroptic 1" mirror is replaced by Al 1" mirror. Problem remains the same. This diffraction patter has to be coming from the Faraday.

  Atm1, good separation when Pd is far 

  Atm2, bad separation when Pd is close 

 The extra high post 3.375"  for PZT is ready. We also have 2 more 2" green Laseroptik mirrors. I'm ready to swap them in.

The  75 mm  focal length  lens  was placed in front of the green REFL PD yesterday.

  9084   Wed Aug 28 11:23:41 2013 SteveUpdateVACcrane repair sheduled

Quote:

 1, Vacuum envelope grounds must be connected all times!  After door removal reconnect both cables immediately.

 2, The crane folding had a new issue of getting cut as picture shows.

 3, Too much oplev light is scattered. This picture was taken just before we put on the heavy door.

 4, We were unprepared to hold the smaller side chamber door 29" od of the IOC

 5, Silicon bronze 1/2-13 nuts for chamber doors will be replaced. They are not smooth turning.

 

 Fred Goldbar of KoneCranes will come 7:30am Wednesday, September 4 to look at this issue.

It is changed to Thursday morning.

  9085   Wed Aug 28 15:41:42 2013 SteveUpdateVACRGA scan at day 23

 Valve configuration: Vacuum Normal

pd76-m-d23

 

Attachment 1: pd75md23.png
pd75md23.png
  9092   Fri Aug 30 14:31:57 2013 SteveUpdateGeneralcleaning up

I cleaned up around MC2 and 1Y1 area this morning.

ETMY NPRO moved from the side to the center of the low shelf. Thorlab catalog " as spacer " removed after lunch.

The Lightwave Controller values: LT 40.4C,  LTEC +0.1V,  T 41.041C,  Pwr 239 mW,  Adj 0,  DC 1.82A,  DPM  0.0V,  Neon OFF,  Ldon OFF,  Display 5,  DT 21.0C,  Dtec +1.0V

 

Attachment 1: MC2areaCP.jpg
MC2areaCP.jpg
Attachment 2: 1Y1areaCleanedUp.jpg
1Y1areaCleanedUp.jpg
Attachment 3: ETMY_NPROmoved.jpg
ETMY_NPROmoved.jpg
  9097   Tue Sep 3 10:54:33 2013 SteveHowToGeneralHow To Coil Cables

Quote:

 http://youtu.be/pEd7ru24Vx0

 B grade Nobel is awarded.

                                  If cables could dream?

This skill should be  mandatory for LIGOX graduates.

  9098   Tue Sep 3 14:23:21 2013 SteveUpdatePEM earthquakes

 ~Vancuver, Canada earthquakes: magnitude 6,  5  and more shaking

 

Attachment 1: EQm6&5.png
EQm6&5.png
  9110   Thu Sep 5 10:28:17 2013 SteveUpdateSUSPRM side

I think the crane repair guy accidentally  stepped on the  BSC support beam.

PRM side is not coming down.

Attachment 1: craneguywashere.png
craneguywashere.png
  9111   Thu Sep 5 10:47:58 2013 SteveUpdateVACvertex crane folding issue

 

 [Fred Goldbar, Mike Gerfen, Dennis Coyne and Steve]

We inspected the hinge, 1.25" cross pin and I-beams. It is hard to explain what is  causing the folding I-beam corner to jam against the main I-beam.

To limit the motion of the folding I-beam  cross pin bushing will be added. This will take a week to complete.

 

 

 

 

 

Attachment 1: 5mmLess.jpg
5mmLess.jpg
  9115   Fri Sep 6 09:27:10 2013 SteveUpdateVAC31 days after pumpdown

Quote:

 Valve configuration: Vacuum Normal

 

 

Attachment 1: day31vac.png
day31vac.png
  9117   Mon Sep 9 15:33:06 2013 SteveUpdateVACvertex crane repair is scheduled

Quote:

 

 [Fred Goldbar, Mike Gerfen, Dennis Coyne and Steve]

We inspected the hinge, 1.25" cross pin and I-beams. It is hard to explain what is  causing the folding I-beam corner to jam against the main I-beam.

To limit the motion of the folding I-beam  cross pin bushing will be added. This will take a week to complete.

 

 

 

 

 

 KoneCrane contact John McDaniel (562) 903 - 1371,

Wednesday, September 18,  folding I-beam will be removed.  KoneCrane will start working at 7:30am and they should be out by 12:30pm

 Friday,  September 20, reinstalling machined hinge on the I-beam. Same timing schedule as Wednesday.

  9119   Tue Sep 10 09:18:22 2013 SteveUpdateVACdry pump is getting loud

 

 TP3 foreline's dry pump is getting noisier and noisier.  Turbo TP3 is pumping on the annulos. The foreline pressure is 7.2 mTorr and it is not degrading. It was swapped in March 5, 2013

The seal is very good, but the bearing is dying.

 

 

 

  9120   Tue Sep 10 15:43:01 2013 SteveUpdateVACTP3' dry pump is replaced

Quote:

 

 TP3 foreline's dry pump is getting noisier and noisier.  Turbo TP3 is pumping on the annulos. The foreline pressure is 7.2 mTorr and it is not degrading. It was swapped in March 5, 2013

The seal is very good, but the bearing is dying.

 

 

 

 The drypump is replaced at 95,781 hrs on TP3 controller time. The foreline pressure is 30 mTorr and dropping.

 It is 13 mTorr after 17 hours of pumping.

  9129   Mon Sep 16 10:57:11 2013 SteveUpdateVACRGA scan at day 41

Quote:

 Valve configuration: Vacuum Normal

 

 Finally we reached 1.0e-5 Torr, cc1 (h) at day 41. There must be a leak or cold cathode gauge is not reading correctly. The gauge is new. It should not take this long.

All chamber annulos pressures are normal.

Attachment 1: pd76m41dRGA.png
pd76m41dRGA.png
Attachment 2: pd76m41d.png
pd76m41d.png
  9139   Wed Sep 18 14:57:12 2013 SteveUpdateVACvertex crane folding arm removed

 The folding beam removed as shown. Two man supporting it while I hammering it out. Pin was dry and it gulled into supporting hinges.

The rotating hinge will be machined and bushing will be added with Zerk fitting or similar. This will allow lubrication in the future.

 see elog #9111

Attachment 1: manPower.jpg
manPower.jpg
Attachment 2: rotatingHinge.jpg
rotatingHinge.jpg
Attachment 3: supportingHinge.jpg
supportingHinge.jpg
Attachment 4: gullingPin.jpg
gullingPin.jpg
  9140   Thu Sep 19 10:24:58 2013 SteveUpdatePEMPRM damping restored after earth quake

Local m3.8 eq shakes PRM lose.

Attachment 1: eq3.8mLaVerne.png
eq3.8mLaVerne.png
  9144   Fri Sep 20 08:15:30 2013 SteveUpdateComputer Scripts / ProgramsOttavia, Rossa and Pianosa

Ottavia, Rossa  and Pianosa are running out of storage  space.

Attachment 1: Ottavia.png
Ottavia.png
Attachment 2: Rossa.png
Rossa.png
Attachment 3: Pianosa.png
Pianosa.png
  9145   Fri Sep 20 09:49:06 2013 SteveUpdateSUSOptical Lever Trend for 180 days: bad ETMY & PRM

I'm working on it.

Attachment 1: ETMYoplevPRM.png
ETMYoplevPRM.png
  9146   Fri Sep 20 16:37:30 2013 SteveUpdateVACvertex crane folding arm is back on

Quote:

 The folding beam removed as shown. Two man supporting it while I hammering it out. Pin was dry and it gulled into supporting hinges.

The rotating hinge will be machined and bushing will be added with Zerk fitting or similar. This will allow lubrication in the future.

 see elog #9111

Atm1,  The folding arm is back on with 0.1" misalignment at no load in the trolly's way. The other side of the I beam is 0.02" higher than the main beam.. New bushing and pin were greased up  with Krytox before installation.

The axial Zerk 1/8" pipe in the pin upper end can not take any fitting. There is no room. It is taped off.

This gap comes down to ~ 1/16" at fully extended arm with 225 lbs load at the end of it.

The present plan is to grind down the  the misalignment of 0.1"  for a slow-loaded trolly.

Steve Baker of Konacranes will be back to grind down this ridge and load test at 500 lbs on Tuesday,  OCT  1, 2013

 

 

 

Attachment 1: bushingBottom.jpg
bushingBottom.jpg
Attachment 2: foldingI.jpg
foldingI.jpg
Attachment 3: noLoadGap.jpg
noLoadGap.jpg
Attachment 4: trollyway.jpg
trollyway.jpg
Attachment 5: trollySee.jpg
trollySee.jpg
ELOG V3.1.3-