  10021   Tue Jun 10 19:11:27 2014 EvanConfigurationWikiDokuWikis are back up

As of this writing, the DokuWiki appears to be working.

As you and I suspected, it looks like this was a clusterwhoops with the permissions for the NFS mount. Let's recap what happened in the past 24 hours:

  1. Yesterday, 8 PM: I restart the Apache server, thereby resurrecting the SVN (now conveniently located at /export/home/svn). The DokuWikis remain borked.
  2. Yesterday, 7 to 11 PM: Zach, Rana, and Jenne perform deep magic to get the front-end machines up and running again. This should be irrelevant for this Apache/SVN/DokuWiki witchcraft.
  3. Today, morning: the townsfolk happily resume their svn up and svn ci festival.
  4. Today, ca. 3 PM: Zach runs stopapache.sh to bring down Apache, thinking he can bring it back up momentarily with startapache.sh. But NFS is a jealous and vengeful god, and Apache now complains that it doesn't have permission to write to its logfiles, and therefore can't start httpd.
  5. Today, 5 PM: "How can this be?", Zach and I ask. Apache had no problem starting up yesterday night, and to our knowledge nobody futzed with chiara's NFS mount of /home/cds. This change remains mysterious to us.

Despite the aforementioned mystery, Zach and I pressed on and tried to diagnose the permissions issue. We found that even if you are logged into nodus or pianosa or rossa or whatever as the controls user, the NFS mount saw us as the user nobody (in the group nogroup). If we created a file on the NFS mount, it was owned by nobody/nogroup. If we tried to modify a file on the NFS mount that was owned by controls/controls or controls/staff, we got a "permission denied" error, even if we tried with superuser privileges.

It turns out this has to do with the vagaries of NFS (scroll down to gotcha #4). We have all_squash enabled in /etc/exports , which means that no matter your username or group on nodus, rossa, pianosa, or harpischordsa, NFS coerces your UID/GID to chiara's nobody/nogroup. Anyway, the fix was to go into chiara's /etc/exports and change this

/home/cds 192.168.113.0/255.255.255.0(rw,sync,no_subtree_check,all_squash,insecure)

to this

/home/cds 192.168.113.0/255.255.255.0(rw,sync,no_subtree_check,all_squash,anonuid=1001,anongid=1001,insecure)

where 1001/1001 are the UID/GID for chiara's controls/controls (as opposed to 65534/65534 for chiara's nobody/nogroup). That way, the NFS mount sees you as chiara's controls/controls.
 
In order to make chiara's NFS daemon aware of the new /etc/export settings, I ran sudo exportfs -r based on the answer to this StackOverflow question. As with all the best StackOverflow questions, the moderators closed it for being "off-topic".
 
[Edit, 2014-06-11, 11 AM: I've repeated the above anonuid/anongid change for the /home/cds/caltech/home/controls mount, so that nodus's /home/controls is writeable as well. I've also added a .screenrc for nodus in order to maintain sanity.]
  10024   Wed Jun 11 10:15:15 2014 EvanConfigurationWikiDokuWikis are back up

Quote:

It looks like auth is broken on the AIC wiki (though working fine on ATF and Cryo). I did some poking around but can't see how anything we did could have broken it.

Quote:

As of this writing, the DokuWiki appears to be working.

 

I went into local.php and changed $conf['useacl'] = 1; to $conf['useacl'] = 0; and it looks like the auth issue goes away (I've changed it back). This isn't a fix (we want to use access control), but it gives us a clue as to where the problem is.

  10235   Fri Jul 18 14:59:07 2014 EvanUpdateIOOMC servo TFs

[Rana, Evan]

This morning we took several TFs of the MC servo board using the HP4395A.

The 4395 source was teed, with one output of the tee going to 4395 R and the other output going to the board's IN1. We then took TFs of (4395 A) / (4395 R), where 4395 A was one of the following four points on the servo board:

  • OUT2
  • A TEST1
  • B TEST1
  • SERVO

For each of these points, we took a TF at two gain settings: IN1 and VCO gains both at 0 dB, and then IN1 and VCO gains both at 20 dB.

Before doing these measurements, we calibrated out the cable delay. Additionally, SERVO was always loaded with 50 Ω—either from the 4395 or from a terminator.

The attached png shows the servo board settings when these TFs were taken with the 0 dB gain settings. The settings for the 20 dB measurements are identical, except for the higher IN1 and VCO gains.

Attachment 1: mcServoTFSettings.png
mcServoTFSettings.png
Attachment 2: MCtfs.pdf
MCtfs.pdf
  10243   Sun Jul 20 09:26:27 2014 EvanUpdateElectronicsMC servo card modifications in DCC

Quote:

[Rana, Jenne]

We have decided to keep better track (using new-fangled digital "computers") of our modifications to electronics boards. 

The idea will be to create a new DCC document for every electronics board (when we pull a board and modify it, it should receive this treatment) that we have, and that document will become a history of the board's life.  Version 1 will be a copy of the original drawing.  Version 2 should be a modified version of that drawing with the current situation.  All future versions should be modified from the most recent version, to reflect any changes.  Notes for each updated version should include an elog reference to the work, so that we know why we did things, and have a place to find photos of the actual modifications.  Elogs should also include a link to the DCC version.  DCC titles should include the phrase "40m Revisions" for ease of searching.

Patient Zero for this new system will be the PMC servo card.  The DCC number is D1400221.  As of this moment, this just has the V1 original drawing with no modifications.

This has been included in the 40m's DCC document tree that Jamie started back in November 2012.

Patient One for this new system will be the MC servo card. The DCC number is D1400242. Currently, v1 is just the original drawing with no modifications. I've updated the DCC document tree at E1400326 accordingly.

It looks like we can use Jenne's information in 40m:9892 to deduce the modifications that have been made (alternatively, someone can just pull the board and examine it on the bench).

  10250   Tue Jul 22 08:24:42 2014 EvanUpdateElectronicsMC servo card: modified schematic

Quote:

Patient One for this new system will be the MC servo card. The DCC number is D1400242. Currently, v1 is just the original drawing with no modifications. I've updated the DCC document tree at E1400326 accordingly.

It looks like we can use Jenne's information in 40m:9892 to deduce the modifications that have been made (alternatively, someone can just pull the board and examine it on the bench).

The attached zip file has a modified schematic of the MC servo card (011/MC), as deduced from Jenne's photos. Someone should go through and verify that the schematic is correct. Then it can go on the DCC as D1400242-v2.

To modify the schematic, I used Inkscape (the svg files for each sheet are included in the zip file). Then to generate the pdf, I ran

for i in sheet*.svg; do inkscape -A "${i/svg/pdf}" "$i"; done

pdftk sheet*.pdf cat output D1400242

Attachment 1: D1400242.zip
  10251   Tue Jul 22 08:36:08 2014 EvanUpdateIOOMC servo TFs

Quote:

[Rana, Evan]

This morning we took several TFs of the MC servo board using the HP4395A.

The 4395 source was teed, with one output of the tee going to 4395 R and the other output going to the board's IN1. We then took TFs of (4395 A) / (4395 R), where 4395 A was one of the following four points on the servo board:

  • OUT2
  • A TEST1
  • B TEST1
  • SERVO

For each of these points, we took a TF at two gain settings: IN1 and VCO gains both at 0 dB, and then IN1 and VCO gains both at 20 dB.

Before doing these measurements, we calibrated out the cable delay. Additionally, SERVO was always loaded with 50 Ω—either from the 4395 or from a terminator.

The attached png shows the servo board settings when these TFs were taken with the 0 dB gain settings. The settings for the 20 dB measurements are identical, except for the higher IN1 and VCO gains.

Using the modified schematic (40m:10250), I've made a plot of the TFs I expect for GIN1 = GVCO = 0 dB, taking into account our 50 Ω loading of the board.

Evidently I'm somehow missing a factor of 2 in the gain of the overall TF, but the shapes of the expected vs. measured magnitudes agree quite well.

At 1 MHz, I expect we should have accumulated about 80 degrees of phase going through the servo board. In reality, we appear to have lost more like 105 degrees.

Attachment 1: MCtfExpectations.pdf
MCtfExpectations.pdf
  7836   Fri Dec 14 17:12:19 2012 Evan HallUpdatePSLPMC yaw tune-up (from Wednesday night)

Wednesday night, there was ~0.4 V on the PMC transmission PD. I adjusted the steering mirrors into the PMC and got the transmission up to 0.81 V.

  10013   Mon Jun 9 19:02:34 2014 Evan, EricUpdateComputer Scripts / ProgramsSVN is back

The SVN Apache server was not happy trying to read from /cvs/cds/caltech/svn/; it complains "Value too large for defined data type" when trying to modify certain files.

To remedy this, Eric ran an rsync job to copy over the svn directory to /export/home/svn/, which is directly on nodus rather than on the NFS mount.

Accordingly, I edited the httpd-ssl.conf file in /cvs/cds/caltech/apache/etc/ so that SVNPath points to /export/home/svn. The original config file is preserved as httpd-ssl.conf.old_20140609.

Then I started the Apache server using the instructions on the 40 m wiki (search "Apache"). The SVN now appears to be working fine; you can svn up and svn ci as necessary.

However, this means that we now need to start backing up /export/home/svn/, rather than the NFS-mounted directory.

  11405   Mon Jul 13 18:27:27 2015 EveConfigurationGeneralHow to set up your own summary page environment on the LDG cluster

I'd like to build off of Koji's instructions with a few useful tips I discovered while setting up my own summary page environment.

To only make a specified selection of tabs for the summary pages, copy only the corresponding .ini files into /home/albert.einstein/summary/config and run the gw_daily_summary_custom following Koji's instructions below. When asked for nodus's password either hit "enter" three times without providing the password or comment out this section of the code to stop the summary page creation process from taking current data files from nodus. This is especially helpful when the 40m is down (like it is now).

After running the summary page code, the pages can be viewed at https://ldas-jobs.ligo.caltech.edu/~albert.einstein/summary/day/YYYYMMDD/ and corresponding error logs can be found at ~/public_html/summary/logs/gw_summary_pipe_local-687496-0.err.

  11411   Tue Jul 14 16:47:18 2015 EveUpdateSummary PagesSummary page updates continue during upgrade

I've continued to make changes to the summary pages on my own environment, which I plan on implementing on the main summary pages when they are back online.

 

Motivation:

I created my own summary page environment and manipulated data from June 30 to make additional plots and change already existing plots. The main summary pages (https://nodus.ligo.caltech.edu:30889/detcharsummary/ or https://ldas-jobs.ligo.caltech.edu/~max.isi/summary/) are currently down due to the CDS upgrade, so my own summary page environment acts as a temporary playground to continue working on my SURF project. My summary pages can be found here (https://ldas-jobs.ligo.caltech.edu/~eve.chase/summary/day/20150630/); they contian identical plots to the main summary pages, except for the Summary tab. I'm open to suggestions, so I can make the summary pages as useful as possible.

 

What I did:

  • SUS OpLev: For every already existing optical lever timeseries, I created a corresponding spectrum, showing all channels present in the original timeseries. The spectra are now placed to the right of their corresponding timeseries. I'm still playing with the axes to make sure I set the best ranges.
  • SUSdrift: I added two new timeseries, DRMI SUS Pitch and DRMI SUS Yaw, to add to the four already-existing timeseries in this tab. These plots represent channels not previously displayed on the summary pages
  • Minor changes
    • Added axis labels on IOO plot 6
    • Changes axis ranges of IOO: MC2 Trans QPD and IOO: IMC REFL RFPD DC
    • Changes axis label on PSL plot 6

 

Results:

So far, all of these changes have been properly implemented into my personal summary page environment. I would like some feedback as to how I can improve the summary pages.

 

 

  11414   Tue Jul 14 17:14:23 2015 EveSummarySummary PagesFuture summary pages improvements

Here is a list of suggested improvements to the summary pages. Let me know if there's something you'd like for me to add to this list!

  • A lot of plots are missing axis labels and titles, and I often don't know what to call these labels. I could use some help with this.
  • Check the weather and vacuum tabs to make sure that we're getting the expected output. Set the axis labels accordingly. 
  • Investigate past periods of missing data on DataViewer to see if the problem was with the data requisition process, the summary page production process, or something else.
  • Based on trends in data over the past three months, set axis ranges accordingly to encapsulate the full data range.
  • Create a CDS tab to store statistics of our digital systems. We will use the CDS signals to determine when the digital system is running and when the minute trend is missing. This will allow us to exclude irrelevant parts of the data.
  • Provide duty ratio statistics for the IMC.
  • Set triggers for certain plots. For example, for channels C1:LSC-XARM OUT DQ and page 4 LIGO-T1500123–v1 C1:LSC-YARM OUT DQ to be plotted in the Arm LSC Control signals figures, C1:LSCTRX OUT DQ and C1:LSC-TRY OUT DQ must be higher than 0.5, thus acting as triggers.
  • Include some flag or other marking indicating when data is not being represented at a certain time for specific plots.
  • Maybe include some cool features like interactive plots.
  11437   Wed Jul 22 22:06:42 2015 EveSummarySummary PagesFuture summary pages improvements

- CDS Tab

We want to monitor the status of the digital control system.

1st plot
Title: EPICS DAQ Status
I wonder we can plot the binary numbers as statuses of the data acquisition for the realtime codes.
We want to use the status indicators. Like this:
https://ldas-jobs.ligo-wa.caltech.edu/~detchar/summary/day/20150722/plots/H1-MULTI_A8CE50_SEGMENTS-1121558417-86400.png

channels:
C1:DAQ-DC0_C1X04_STATUS
C1:DAQ-DC0_C1LSC_STATUS
C1:DAQ-DC0_C1ASS_STATUS
C1:DAQ-DC0_C1OAF_STATUS
C1:DAQ-DC0_C1CAL_STATUS

C1:DAQ-DC0_C1X02_STATUS
C1:DAQ-DC0_C1SUS_STATUS
C1:DAQ-DC0_C1MCS_STATUS
C1:DAQ-DC0_C1RFM_STATUS
C1:DAQ-DC0_C1PEM_STATUS

C1:DAQ-DC0_C1X03_STATUS
C1:DAQ-DC0_C1IOO_STATUS
C1:DAQ-DC0_C1ALS_STATUS

C1:DAQ-DC0_C1X01_STATUS
C1:DAQ-DC0_C1SCX_STATUS
C1:DAQ-DC0_C1ASX_STATUS

C1:DAQ-DC0_C1X05_STATUS
C1:DAQ-DC0_C1SCY_STATUS
C1:DAQ-DC0_C1TST_STATUS

1st plot
Title: IOP Fast Channel DAQ Status
These have two bits each. How can we handle it?
If we need to shrink it to a single bit take "AND" of them.
C1:FEC-40_FB_NET_STATUS (legend: c1x04, if a legend placable)
C1:FEC-20_FB_NET_STATUS (legend: c1x02)
C1:FEC-33_FB_NET_STATUS (legend: c1x03)
C1:FEC-19_FB_NET_STATUS (legend: c1x01)
C1:FEC-46_FB_NET_STATUS (legend: c1x05)

3rd plot
Title C1LSC CPU Meters
channels:
C1:FEC-40_CPU_METER (legend: c1x04)
C1:FEC-42_CPU_METER (legend: c1lsc)
C1:FEC-48_CPU_METER (legend: c1ass)
C1:FEC-22_CPU_METER (legend: c1oaf)
C1:FEC-50_CPU_METER (legend: c1cal)
The range is from 0 to 75 except for c1oaf that could go to 500.
Can we plot c1oaf with the value being devided by 8? (Then the legend should be c1oaf /8)

4th plot
Title C1SUS CPU Meters
channels:
C1:FEC-20_CPU_METER (legend: c1x02)
C1:FEC-21_CPU_METER (legend: c1sus)
C1:FEC-36_CPU_METER (legend: c1mcs)
C1:FEC-38_CPU_METER (legend: c1rfm)
C1:FEC-39_CPU_METER (legend: c1pem)
The range is be from 0 to 75 except for c1pem that could go to 500.
Can we plot c1pem with the value being devided by 8? (Then the legend should be c1pem /8)

5th plot
Title C1IOO CPU Meters
channels:
C1:FEC-33_CPU_METER (legend: c1x03)
C1:FEC-34_CPU_METER (legend: c1ioo)
C1:FEC-28_CPU_METER (legend: c1als)
The range is be from 0 to 75.

6th plot
Title C1ISCEX CPU Meters
channels:
C1:FEC-19_CPU_METER (legend: c1x01)
C1:FEC-45_CPU_METER (legend: c1scx)
C1:FEC-44_CPU_METER (legend: c1asx)
The range is be from 0 to 75.

7th plot
Title C1ISCEY CPU Meters
channels:
C1:FEC-46_CPU_METER (legend: c1x05)
C1:FEC-47_CPU_METER (legend: c1scy)
C1:FEC-91_CPU_METER (legend: c1tst)
The range is be from 0 to 75.

=====================

IOO

We want a duty ratio plot for the IMC. C1:IOO-MC_TRANS_SUM >1e4 is the good period.

Duty ratio plot looks like the right plot of the following link
https://ldas-jobs.ligo-wa.caltech.edu/~detchar/summary/day/20150722/lock/segments/

=====================

SUS: OPLEV

OL_PIT_INMON and OL_YAW_INMON are good for the slow drift monitor.
But their sampling rate is too slow for the PSDs.
Can you use
C1:SUS-ETM_OPLEV_PERROR
C1:SUS-ETM_OPLEV_YERROR
etc...
For the PSDs? They are 2kHz sampling DQ channels. You would be able to plot
it up to ~1kHz. In fact, we want to monitor the PSD from 100mHz to 1kHz.
How can you set up the resolution (=FFT length)?

=====================

LSC / ASC / ALS tabs

Let's make new tabs LSC, ASC, and ALS

LSC:

We should have a plot for
C1:LSC-TRX_OUT_DQ
C1:LSC-TRY_OUT_DQ
C1:LSC-POPDC_OUT_DQ
It's OK to use the minute trend for now.
You can check the range using dataviewer.

ASC:

Let's use
C1:SUS_MC1_ASCPIT_OUT16 (legend: IMC WFS)
C1:ASS-XARM_ITM_YAW_OSC_CLKGAIN (legend: XARM ASS)
C1:ASS-YARM_ITM_YAW_OSC_CLKGAIN (legend: YARM ASS)
C1:ASX-XARM_M1_PIT_OSC_CLKGAIN (legend: XARM Green ASS)
as the status indicators. There is no YARM Green ASS yet.

ALS:

Title: ALS Green transmission
We want a time series of
ALS-TRX_OUT16
ALS-TRY_OUT16

Title: ALS Green beatnote
Another time series
ALS-BEATX_FINE_Q_MON
ALS-BEATY_FINE_Q_MON

Title: Frequency monitor
We have frequency counter outputs, but I have to talk to Eric to know the channel names

  11448   Mon Jul 27 17:51:06 2015 EveUpdateSummary PagesNew summary page tabs and other improvements

The summary pages can still be found at https://nodus.ligo.caltech.edu:30889/detcharsummary/ (EDIT: in an older version of this post I listed an incorrect url). They are operational and include data from some channels for intermittent periods of time.

 

Motivation: to make the summary pages more informative and useful for all

 

What I did:

I have added tabs for ALS, ASC, and LSC subsystems. While there is currently no data on the plots, I plan on checking all channels with DataViewer to set appropriate axis ranges so that we can actually see the data.

I altered which channels are used to represent spectra for OpLev systems to more appropriately provide PSDs.

I've changed the check code status page to include "warning" labels. Previously, when the summary pages ran, resulting in a warning message, the check code status page would list this as an "error", implying that the summary pages were not properly produced.

 

Results:

All features were implemented, but I need to investigate some of these channels to understand why we aren't seeing many channels in the plots. I am working on some other changes to the summary pages, including providing a Locked status which will only show data in a timeseries for a selected period of time.

  11467   Thu Jul 30 14:27:18 2015 EveUpdateSummary PagesALS, ASC, LSC Summary Pages

I've switches the ALS, ASC, and LSC plots on the summary pages from plotting raw frames, to plotting minute trends, instead. Now, the plots contain information, instead of being completely blank, but data is not recorded on the plots after 12UTC.

Typically, I make changes to the summary pages on my own version of the pages, found at https://ldas-jobs.ligo.caltech.edu/~eve.chase/summary/day/, where I change the summary pages for June 30 and then import such changes into the main summary pages. 

 

  11474   Sat Aug 1 17:04:29 2015 EveUpdateSummary PagesStates and Triggers in SPs

I've added states to the summary pages to only show data for times at which one certain channel is above a specified threshold. So far, I've incorporated states for the IOO tab to show when the mode cleaner is locked.

You can see these changes implemented in the IOO tab of my personal summary pages for June 30: https://ldas-jobs.ligo.caltech.edu/~eve.chase/summary/day/20150630/ioo/.

I've written a description of how to add states to summary pages here: https://wiki-40m.ligo.caltech.edu/DailySummaryHelp#How_to_Define_and_Implement_States.

  11480   Wed Aug 5 17:15:08 2015 EveUpdateSummary PagesFixed ASC Tab

I've fixed the ASC tab on the summary pages to populate the graphs with data without causing an error.

Motivation: The ASC tab was showing no data. It resulted in a name error when generated.

What I did:

A name error indicates a bad channel name in the plot definition. I identified two errors in the code:

  1. I said C1:SUS_MC1_ASCPIT_OUT16.mean instead of C1:SUS-MC1_ASCPIT_OUT16.mean (underscore should be dash)
  2. The channel C1:ASX-XARM_M1_PUT_OSC_CLKGAIN was resulting in a name error. I removed it.

Results:

The plots are not processing without error. However, no titles or axis labels are present on the plots- I'll work on adding these.

 

  11494   Tue Aug 11 16:13:28 2015 EveUpdateGeneralGaussianity tests

I’m working on a code to determine the Gaussianity of a PSD.

 

Motivation:

It can be difficult to distinguish between GW events and non-Gaussian noise, especially in burst searches. By characterizing noise Gaussianity, we can better recognize noise patterns and distinguish between GW events and noise.

 

What I did:

I analyzed an hour of S5 L1 data. First, I plotted a timeseries, just to see what I was working with. Then, I produced a PSD (technically, an ASD) for the timeseries using Welch’s method in Python.

 

I split the data segment into smaller time-chunks and then produced a PSD for each chunk. All PSDs were superimposed in one plot. Here’s a plot for 201 time-chunks of equal length:

For a specific frequency, I can view the spread in PSD value through the production of a histogram.

 

Results:

I’ve made histograms displaying varying PSD values for the 201 PSD plot at 100 Hz, 500 Hz, and 1kHz.

For Gaussian noise, an exponential decay plot is expected. I will continue this analysis by following the statistical method in Ando et al. 2003 to calculate specific values indicative of the Gaussianity of various distributions. I’ll then look at different periods of time in the S5 L1 data to find periods of time suggesting non-Gaussian behavior.

  11511   Sun Aug 16 23:26:40 2015 EveUpdateGeneralGaussianity tests

I've continued to work on my Gaussianity tests for S5 L1 data. 

Following the statistical measure in Ando et al. (2003), I've calculated the Laguerre coefficient, c2, for all frequencies present in my S5 L1 PSD as a metric of Gaussianity. When c2 is zero, the distribution is Gaussian. A positive c2 corresponds to glitchy noise, while a negative c2 suggests stationary noise.

Below is a plot displaying variation in c2 for this PSD:

By observing the c2 value and histogram of distribution of various PSD values at a given frequency, we can elucidate statistical differences in the Gaussian nature of noise at that frequency which are unclear in the standard PSD.

Attachment 1: Gaussianity_noc1.png
Gaussianity_noc1.png
  11385   Tue Jun 30 20:26:24 2015 Eve UpdateGeneralMinor Summary Page Changes

I made several small, nit-picky changes to the summary pages.

 

Motivation:

I'm still working on getting used to editing the summary pages. I also wanted to change some of the easy-to-alter cosmetics of the pages.

 

What I did:

I changed axis ranges, axis labels, and typos throughout the summary pages. Read below for an excrutiating list of the minor details of my alterations, if you wish:

  • Changed axes on LSC control signals plots on the Summary tab (but will probably change these back to their original state)
  • Moved an OpLev plot from the Sandbox tab to "Eve" tab
  • Increased the y axis range on IOO MC2 Trans QPD and IMC REFLY RFPD DC plots (which may change when I better incorporate triggers into these plots)
  • Fixed title on IOO Whitened Spectrogram and Rayleigh Spectrogram
  • Fixed degree sign on Weather: Temperature and PSL Table Temperature
  • Fixed percent sign on Weather: Humidity
  •  

Results:

So far, everything looks good. I'll continue to make more changes later this week and hope to soon get on to more substatial changes.

  11380   Fri Jun 26 23:18:52 2015 Eve ChaseUpdateCDSSummary Page Updates

Motivation:

My SURF project largely focuses on updating and improving the 40m summary pages. I began to explore and experiment with the already existing summary page code to better understand the required code and eventually lead to tangible improvements of the summary pages.

 

What I did:

I practiced moving from one tab of the summary pages to another. Specifically, I copied the ETM Optical Levers plot from SUS: OpLev to Sandbox, without removing it from its original location. Additionally, I created my own tab to further test the summary pages, titled “Eve”.

KA ed: The configuration files are located at /cvs/cds/caltech/chans/GWsummaries. It is under svn control.

Result:

All changed appeared on the summary pages without much hassle and without any errors.

  2891   Thu May 6 19:23:54 2010 FrankSummaryComputerssvn problems

i tried to commit something this afternoon and got the following error message:

Command: Commit 
Adding: C:\Caltech\Documents\40m-svn\nodus\frank 
Error: Commit failed (details follow): 
Error: Server sent unexpected return value (405 Method Not Allowed) in response to  
Error: MKCOL request for '/svn/!svn/wrk/d2523f8e-eda2-d847-b8e5-59c020170cec/trunk/frank' 
Finished!:  

anyone had this before? what's wrong?

  3117   Thu Jun 24 18:47:26 2010 FrankDAQIOOVME crate rebooted

we had to reboot the IOO VME crate right before lunch as the DAQ wasn't working correct meaning showing no real signals anymore, only strange noise. The framebuilder and everything else was working fine at that time.

  • The channel used for the phase noise measurement stopped showing any useful signal right after midnight, so all the other IOO-MC signals.
  • The data taken with those channels showed something like a 140 counts or so of steady offset with something which looked like the last bit fluctuating.
  • Whatever signal we connected to the input it didn't change at all, floating/shorted input, sine wave etc.
  • the other channels for the MC which we checked showed the same strange behaviour

As the other channels showed the same effect we decided to reboot the crate and everything was fine afterwards.

  3875   Sat Nov 6 01:54:15 2010 FrankSummaryComputersvirus definition file update on laptop for dinocam

i took some pictures with the dinocam this afternoon. I used the laptop computer next to it using wireless lan connection to the caltech network to send the pictures to me.

The installed anti virus software was bitching about the old database and wanted to update that. As the installed virus definition database was from mid last year i agreed and started downloading the update. As the file was huge (~100MB) it wasn't finished when i left. computer is still running and probably waiting for instructions.

Will come back on the weekend to finalize the new virus definition file database installation.

  4017   Mon Dec 6 22:43:19 2010 FrankConfigurationelogelog restarted

I restarted the elog because i changed the configuration for the cryo-elog.

Used the "start-elog.csh" script in /cvs/cds/caltech/elog/

  8376   Fri Mar 29 19:56:02 2013 GabrieleMetaphysicsLSCLock of PRMI on sidebands

I finally managed to get long stretches of PRMI lock, up to many minutes. The lock is not yest very stable, it seems to me that we are limited by some yaw oscillation that I could not trace down. The oscillation is very well visible on POP.

Presently, PRCL is controlled with REFL55_I, while MICH is controlled with AS55_Q. This configuration is maybe not optimal from the point of view of phase noise couplings, but at least it works quite well. I believe that the limit on the length of locks is given by the angular oscillation. I attach to this entry few plots showing some of the lock stretches. The alignment is not optimal, as visible from a quite large TEM01 mode at the dark port.

Here are the parameters I used:

MICH gain -10   PRCL gain -0.1

Normalization of both error signal on POP22_I with factor 0.004

Triggering on POP22: in at 100, out at 20 for both MICH and PRCL.

POP55 demodulation phase -9

MICH and PRCL control signal limits at 2000 counts

 

There is a high frequency (628 Hz) oscillation going on when locked (very annoying on the speakers...), but reducing the gain made the lock less stable. I could go down to MICH=-1.5 and PRCL=-0.02, still being able to acquire the lock. But the oscillation was still there. I suspect that it is not due to the loops, but maybe some resonance of the suspension or payload (violin mode?). There is still some room for fine tuning...

Lock is acquired without problems and maintained for minutes.

Have a nice week-end!

Attachment 1: lock_prmi5.pdf
lock_prmi5.pdf
Attachment 2: lock_prmi6.pdf
lock_prmi6.pdf
Attachment 3: lock_prmi7.pdf
lock_prmi7.pdf
Attachment 4: oscillation.pdf
oscillation.pdf
Attachment 5: lock_prmi8.pdf
lock_prmi8.pdf
  8401   Wed Apr 3 14:46:17 2013 GabrieleSummaryLSCError signal simulation in PRMI

Here is a summary of a simulation of the error signal behavior in the PRMI configuration. The main parameters are:
L_PRC = 6.7538 m
Schnup = 0.0342 m
fmod1 = 11.065399e6 Hz
fmod2 = 5 * fmod1

prmi_prclsweep_pop22.pngprmi_prclsweep_pop110.png

These two plots shows the response of the POP22 and POP110 signals (in almost arbitrary units) to a PRCL sweep around the resonance. The splitting of the 55 sideband peaks is well visible in the second plot. It is due to the fact that the 55MHz sidebands are not perfectly matched to the PRC length

prmi_michsweep_pop22.pngprmi_michsweep_pop110.png

The same thing when sweeping MICH. The peaks are wider and it is not possible to see the splitting.

prmi_prclsweep_errsig_not_tunedphase.pngprmi_michsweep_errsig_not_tunedphase.png

These are the error signals (REFL11_I/Q and REFL_55_I/Q) as a function of the PRCL (left) and MICH (right) sweep. Here the demodulation phases are not properly tuned. This is just to show that when the phase is wrong, you can get multiple zero crossings (in this case only in the Q signals, but in general also in I) close to resonance.

prmi_prclsweep_errsig_tunedphase.pngprmi_michsweep_errsig_tunedphase.png

If the phases are tuned in order to maximize the slope of the I signals with respect to PRCL, one gets these "optimized phase" responses. It is that the phase does not correspond to the one that makes the PRCL peak to peak signal small in Q. The Q signals are indeed flat around resonance for a PRCL motion, but they deviate quite a lot from zero when moving more far from resonance. Moreover, both the REFL_55 error signals (I for PRCL and Q for MICH) are crossing again zero at two additional positions, but those are quite far from the resonance point.

prmi_prclsweep_triggering.pngprmi_michsweep_triggering.png

These plots just show the PRCL and MICH error signals together with the POP22 and POP110 signals, to give an idea of the level of triggering that might be needed to be inside the linear range. It seems that if we trigger on POP22 when using the REFL55 signal we loose a bit of linear range, but not that much.

prmi_prclsweep_linearized_signals.pngprmi_michsweep_linearized_signals.png

If you reached this point it means you're really interested in this topic, or maybe you have nothing better to do... However, this plot shows the effect of linearization of the error signal, obtained dividing them by the proper POP22/110 signal. The linear range is increased, but unfortunately for the 55 signals, the additional zero crossing I was mentioning before creates two sharp features. Those are however quite outside the triggering region, so they should not be harmful.

Attachment 1: prmi_michsweep_pop22.png
prmi_michsweep_pop22.png
Attachment 2: prmi_michsweep_pop110.png
prmi_michsweep_pop110.png
  8411   Thu Apr 4 10:12:55 2013 GabrieleSummaryLSCPOP22 and POP110

I had a look at the POP110 signal, with the PRMI flashing.

1) The LSCoffset script does not zero any more POP22_I_ERR offset. I did it by hand

2) The gain of POP22 is changed a lot, as well as the sign: now sidebands are resonant when POP22_I is negative

3) POP110 seems to deliver good signals. The plot attached shows that when we cross the sideband resonance, there is a clear splitting of the peak. If we rely on the simulations I posted in entry 8401, the full width at half height of the POP_22 peak is of the order of 5 nm. Using this as a calibration, we find a splitting of the order of 7 nm, which is not far from the simulated one (5 nm)

Attachment 1: pop110_2.pdf
pop110_2.pdf
  8412   Thu Apr 4 10:32:42 2013 GabrieleSummaryLSCREFL55 error signals

The attached plot shows that also the behaviour of the REFL11 and 55 signals is qualitatively equal to the simulation outcome.

Attachment 1: prmi_refl_signals.pdf
prmi_refl_signals.pdf
  9326   Fri Nov 1 17:01:46 2013 GabrieleSummaryLSCSimulation of REFL_3f signal when the arms come in

 I simulated how the 3f signal is affected by the resonance condition of the arms.

To keep it simple, I only simulated a double cavity. The attached plot shows the result. In x there is the arm cavity detuning from resonance (in log scale to show what happens close to the 0 value). In the y axis there is the PRC detuning. So every vertical slice of the upper plot gives a PDH signal for a given arm detuning. The bottom plot shows the power build up inside the arm, which is dominated by the carrier.

refl_3f_I_vs_cavity_tuning_noresf2.png

The 3f signal is not perturbed in any significant way by the arm resonance condition. This is good and what we expected.

However, in this simulation I had to ensure that the 1f sidebands are not perfectly anti-resonant inside the arms. They are indeed quite far away from resonance. If the modulation frequency is chosen in order to make the 1f sidebands exactly ant-resonant, the 2f will be resonant. This screws up the signal: REFL_3f is made of two contributions of equal amplitude, one on the PRC sidebands resonance and the other on the PRC carrier resonance. When the arm tuning goes to zero, these two cancels out and there is no more PDH...

refl_3f_I_vs_cavity_tuning.png

However, this is a limit case, since the frequency show match perfectly. If the modulation frequency is few arm line widths away from perfect anti-resonance, we have no problem.

  9337   Mon Nov 4 14:11:23 2013 GabrieleSummaryLSCSimulation of REFL_3f signal when the arms come in

Quote:

Yes, the resonance of the 2nd-order sidebands to the IFO screws up the 3f scheme.

2f (~22MHz) and 10f (~110MHz) are at x 5.6 and x 27.9 FSR from the carrier, so that's not the case.

Could we also see how much gain fluctuation of the 3f signals we would experience when the arm comes into the resonance?

 From the simulation there is no visible change in the gain.

  9344   Tue Nov 5 16:39:54 2013 GabrieleUpdateLSCThoughts on Transition to IR

Quote:

Gabriele and I talked for a while on Wednesday afternoon about ideas for transitioning to IR control, from ALS. 

I think one of the baseline ideas was to use the sqrt(transmission) as an error signal.  Gabriele pointed out to me that to have a linear signal, really what we need is sqrt( [max transmission] - [current transmission] ), and this requires good knowledge of the maximum transmission that we expect.  However, we can't really measure this max transmission, since we aren't yet able to hold the arms that close to resonance.  If we get this number wrong, the error signal close to the resonance won't be very good.

Gabriele suggested maybe using just the raw transmission signal.  When we're near the half-resonance point, the transmission gives us an approximately linear signal, although it becomes totally non-linear as we get close to resonance.  Using this technique, however, requires lowering the finesse of PRCL by putting in a medium-large MICH offset, so that the PRC is lossy.  This lowering of the PRC finesse prevents the coupled-cavity linewidth of the arm to get too tiny.  Apparently this trick was very handy for Virgo when locking the PRFPMI, but it's not so clear that it will work for the DRFPMI, because the signal recycling cavity complicates things.

I need to look at, and meditate over, some Optickle simulations before I say much else about this stuff.

 The idea of introducing a large MICH offset to reduce the PRC finesse might help us to get rid of the transmitted power signal. We might be able to increase enough the line width of the double cavity to make it larger than the ASL length fluctuations. Then we can switch from ASL to the IR demodulated signal without transitioning through the power signal.

  9373   Wed Nov 13 09:31:15 2013 GabrieleUpdateLSCPRM motion causing trouble?

Interesting results. When you compute the effect of ETM motion, you maybe should also consider that moving around the arm cavity axis changes the matching of the input beam with the cavity, and thus the coupling between PRC and arms. But I believe this effect is of the same order of the one you computed, so maybe there is only one or two factors of two to add. This do not change significantly the conclusion.

Instead, the numbers you're giving for PRM motion are interesting. Since I almost never believe computations before I see that an experiment agrees with them, I suggest that you try to prove experimentally your statement. The simplest way is to use a scatter plot as I suggested the past week: you plot the carrier arm power vs PRM optical lever signals in a scatter plot. If there is no correlation between the two motions, you should see a round fuzzy ball in the plot. Otherwise, you will se some non trivial shape. Here is an example: https://tds.ego-gw.it/itf/osl_virgo/index.php?callRep=18918

 

  9464   Fri Dec 13 11:47:11 2013 GabrieleUpdateGreen LockingBetter Xarm OLTF

I'm not as good as a fit, but it seems that there is a loop delay of about 30 microseconds, looking at the high frequency phase. This might explain the limitation on the BW. Eric should be able to get the delay out of the fit with some care.

  9490   Wed Dec 18 17:28:50 2013 GabrieleSummaryLSCEstimate of the PRC length error

Measurement 

Looking back at what I did in april (see log #8411) I realized that it is possible to get an estimate of how much the PRC length is wrong looking at the splitting of the sideband resonant peak as visible in the POP_110_I signal. With the help of Jenne the PRMI was aligned and left swinging. The first plot shows a typical example of the peak splitting of 55MHz sidebands. This is much larger than what was observed in April.

When the sidebands resonate inside the PRC they get a differential dephasing given by 

dPhi = 4*pi*f_mod/c * dL

where dL is the cavity length error with respect to the one that makes the sidebands perfectly resonant when the arms are not there. This is not exactly the error we are interested in, since we should take into account the shift from anti-resonance of the SBs in the arm cavities.

Nevertheless, I can measure the splitting of the peak in units of the peaks full width at half maximum (FWHM). I did this fitting few peaks with the sum of two Airy peaks. Here is an example of the result

data_fit2.png

The average splitting is 1.8 times the FWHM. Knowing the PRC finesse, one can determine the length error:

dL = c / (4 * f_mod * Finesse) * (dPhi / FWHM)

Assuming a finesse of 60, I get a length error of 4 cm.

To get another estimate, we kicked the PRM in order to get some almost linear sweeps of the PRC length. Here is one of the best results:

data_kick.png

The distance between consecutive peaks is the free spectral range (FSR) of the PRC cavity. Again, I can measure the peak splitting in units of the FSR and determine the length error:

dL = c / (4 * f_mod) * (dPhi / FSR)

The result is again a length error of 4 cm.

Simulation

An error of 4 cm seems pretty big. Therefore I set up a quick simulation with MIST to check if this makes sense. Indeed, if I simulate a PRMI with the 40m parameters and move the PRC length from the optimal one, I get the following result for POP_110_I, which is consistent with the measurement.

pop110_vs_dL.png

 

 Therefore, we can quite confidently assume that the PRC is off by 4 cm with respect to the position that would make the 55 MHz sideband resonant without arms. Unfortunately, it is not possible with this technique to infere the direction of the error.

 

 

 

 

 

 

Attachment 3: pop110_vs_dL.png
pop110_vs_dL.png
  9495   Thu Dec 19 18:33:25 2013 GabrieleSummaryLSCEstimate of the sign of the PRC length error

Maybe I'm getting confused, but I still believe there is no way to decide the direction from yesterday's measurement.

Let's say for example that the arm sideband detuning from antiresonance is equivalent to a PRC length change of +1cm away from the position of ideal resonance of the sidebands without arms. Then we can get a measured separation of the sidebands, without arms, corresponding to 5cm both if the PRC is off by +4cm or by -6cm...

  9537   Wed Jan 8 13:01:48 2014 GabrieleSummaryLSCEffect of PRC length mismatch on error signals

I ran a simulation of a double cavity with a PRC length mismatched w.r.t. the modulation frequency. I summarized the results in the attached PDF. I think it would be important to have a cross check of the results.

In brief:

A mismatch between PRC length and modulation frequency do have an effect on error signals

Multiple zeros appear in REFL_3f/PRCL that can be removed by careful tuning of the demodulation phase (however, the shape of the signal makes difficult to understand which phase is good…)

No visible effect on REFL_1f/CARM

But a large PRCL signal appears in REFL_1f_I, which is used to control CARM. This is not good.

A mismatch of the order of 0.5 cm has a small effect.

 

 

 

 

 

Attachment 1: REFL_vs_PRClength.pdf
REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf REFL_vs_PRClength.pdf
  9541   Wed Jan 8 19:05:30 2014 GabrieleSummaryLSCEffect of PRC length mismatch on error signals

 [Gabriele, EricQ]


Actually it is difficult to see any laser frequency line in the dark fringe signal, since the Schnupp asymmetry is small. It is much better to use a differential MICH excitation which gives a better signal at the dark port.

We repeated the simulation explained before. We can use both the AS55 or the AS11 signals, bout the first one has a limited linear range and the expected 4cm value is very close to saturation.

as11.pngas55.png

  9548   Sun Jan 12 09:57:24 2014 GabrieleSummaryLSCEffect of PRC length mismatch on error signals

Quote:

 

 Its very doubtful that the MC yaw drift matters for the IFO. That's just a qualitative correlation; the numbers don't hang together.

 Then there must be something else slowly drifting. It was very clear that the good alignment of the IFO was every time lost after few minutes...

  9551   Mon Jan 13 19:31:04 2014 GabrieleSummaryLSCPRMI locking

[Gabriele, EricQ]

We wanted to try the PRC length measurement,but we ended up spending all the afternoon to lock the PRMI on sidebands. Here are some results

  • Lock of PRM on carrier is easy with MICH on AS55_Q and PRCL on REFL55_I. The optimal gains to avoid correction gain peaking are MICH=-5 and PRCL=0.012. The alignment today was must more stable over time than on Friday
  • We wanted to move MICH on REFL55_Q. After a few trials we discovered that REFL55_Q is not seeing any MICH signal at all. This is quite strange and we don't understand this.
  • Locking PRMI on carrier using REFL165_I and Q was difficult. We thought that was due to a RF amplifier installed in December. We tried to remove it, but since this did not help, we put it back
  • We could lock PRMI on carrier and sidebands using REFL11 signals. The optimal demodulation phase is 155. Lock on carrier was achieved with gains MICH=5, PRCL=0.7. Lock on sidebands with MICH gain=-10 and PRCL gain=-0.3. The MICH correction is very high and exciting a tone at few hundreds Hz. Maybe a violin mode of the ITMs?
  • Finally we could tune the REFL165 phase to 118.5 and lock on carrier. The lock was not very stable. Gain are: lock on carrier MICH=-0.5, PRCL=0.05. The error signal has a very big offset: to get a dark fringe we had to add a MICH offset of 500 counts. We also had to engage the CLP400 filter to avoid saturating ITMs corrections with MICH.
  9554   Tue Jan 14 19:05:48 2014 GabrieleSummaryLSCPRMI locked on sideband

[Gabriele, EricQ] 

Finally, we managed to lock PRMI on sidebands:

  • MICH locked on AS55_Q with gain -10. Demod phase of AS55=18
  • PRCL locked on REFL55_I with gain -0.04. Demod phase of REFL55=88
  • Triggering on POP110_I at 50/10
  • Filter "1:5" of MICH engaged, this improved a lot the stability

 

  9555   Tue Jan 14 19:10:51 2014 GabrieleSummaryLSCPRC length measurement

 [EricQ, Gabriele]

We could carry out the measurement of PRC length. The AS110 photodiode was plugged into REFL11. So REFL11 is giving us the AS11 signal. Here is the procedure.

  1. Lock MICH.
  2. Add a line in MICH (amplitude 20000 counts)
  3. Tune AS11 demod phase to have the line in I.
  4. Change the demod phase by steps of 1 degree around the rough optimum, taking one minute of data at each point
  5. Lock PRMI on sidebands
  6. Add a line in MICH (amplitude 500 counts)
  7. Tune AS11 demod phase to have the line in I.
  8. Change the demod phase by steps of 1 degree around the rough optimum, taking one minute of data at each point

We repeated the same measurement also using AS55, with the same procedure.

Roughly, the phase difference for AS11 was 11 degrees and for AS55 it was 23 degrees. A more detailed analysis and a calibration in terms of PRC length will follow.

  9557   Wed Jan 15 18:18:15 2014 GabrieleSummaryLSCPRC length measurement analysis

 I analyzed the data we took yesterday, both using AS11 and AS55. For each value of the phase I estimated the Q/P ratio using a demodulation code. Then I used a linear regression fit to estimate the zero crossing point.

Here are the plots of the data points with the fits:

experimental_data_as11.pdfexperimental_data_as55.pdf

The measurements a re more noisy in the PRMI configuration, as expected since we had a lot of angular motion. Also, the AS11 data is more noisy. However, the estimated phase differences between PRMI and MICH configurations are:

  • AS11 = -10.9 +- 1.0 degrees
  • AS55 = -21.1 +- 0.4 degrees

The simulation already described in slogs 9539 and 9541 provides the calibration in terms of PRC length. Here are the curves

simulation_phase_difference.pdf

The corresponding length errors are

  • AS11 = 1.44 +- 0.13 cm
  • AS55 = 0.59 +- 0.01 cm

The two results are not consistent one with the other and they are both not consistent with the previous estimate of 4 cm based on the 55MHz sideband peak splitting.

I don't know the reason for this incongruence. I checked the simulation, repeating it with Optickle and I got the same results. So I'm confident that the simulation is not completely wrong.

I also tried to understand which parameters of the IFO can affect the result. The following ones have no impact

  • Beam matching
  • ITM curvatures
  • Schnupp asymmetry
  • Distance PR-BS
  • ITM and PRM misalignments

The only parameters that could affect the curves are offsets in MICH and PRCL locking point. We should check if this is happening. A first quick look (with EricQ) seems to indicate that we indeed have an offset in PRCL. However, tonight the PRMI is not locking stably on the sidebands. 

If possibile, we will repeat the measurement later on tonight, checking first the PRCL offset.

 

  9561   Fri Jan 17 11:44:25 2014 GabrieleSummaryLSCMore length measurements, more confusion

 I analyzed the data taken yesterday. 

The AS11 data in PRMI configuration is very bad, while the AS55 seems good enough:

results_as11.pdfresults_as55.pdf

The phase differences are 

AS11 = 21 +- 18 degrees (almost useless due to the large error)

AS55 = 11.0 +- 0.4 degrees

The AS55 phase difference is not the same measured in the last trial, but about half of it. The new length estimates are:

AS11 = 3.2 +- 2.8 cm

AS55 = 0.47 +- 0.01 cm

We can probably forget about the AS11 measurement, but the AS55 result is different from the previous estimate... Maybe this is due to the fact that Eric adjusted the PRCL offset, but then we're going in the wrong direction....

  9564   Wed Jan 22 09:05:42 2014 GabrieleConfigurationLSCREFL11 back

 Yesterday night I plugged back the REFL11 RF cable into the corresponding demodulation board.

  9573   Fri Jan 24 12:44:25 2014 GabrieleHowToLSCProcedure to measure PRC length

Here is how to measure the PRC length with a set of distance measurements in the optical setup. 

We need to take distance measurements between reference points on each mirror suspension. For the large ones (SOS) that are used for BS, PRM and ITMs, the reference points are the corners of the second rectangular base: not the one directly in contact with the optical bench (since the chamfers make difficult to define a clear corner), but the rectangular one just above it. For the small suspensions (TT) the points are directly the corners of the base plates.

From the mechanical drawings of the two kind of suspensions I got the distances between the mirror centers and the reference corners. The mirror is not centered in the base, so it is a good idea to cross check if the numbers are correct with some measurements on the dummy suspensions.

I assumed the dimensions of the mirrors, as well as the beam incidence angles are known and we don't need to measure them again. Small errors in the angles should have small impact on the results.

I wrote a MATLAB script that takes as input the measured distances and produce the optical path lengths. The script also produce a drawing of the setup as reconstructed, showing the measurement points, the mirrors, the reference base plates,  and the beam path. Here is an example output, that can be used to understand which are the five distances to be measured. I used dummy measured distances to produce it.

map.pdf

In red the beam path in vacuum and in magenta the beam path in the substrate. The mirrors are the blue rectangles inside the reference bases which are in black. The thick lines are the HR faces. The green points are the measurement points and the green lines the distances to be measured. The names on the measurement lines are those used in the MATLAB script. 

The MATLAB scripts are attached to this elog. The main file is survey_v2.m, which contains all the parameters and the measured values. Update it with the real numbers and run it to get the results, including the graphic output. The other files are auxiliary functions to create the graphics. I checked many times the code and the computations, but I can't be sure that there are no errors, since there's no way to check if the output is correct... The plot is produced in a way which is somehow independent from the computations, so if it makes sense this gives at least a self consistency test. 

Attachment 2: survey_v2.m
global sos_lx sos_ly sos_cx sos_cy tt_lx tt_ly tt_cx tt_cy

%% Survey of the PRC length

%% measured distances
d_MB2_MY  = 2000.0;
d_MB3_MX  = 2000.0;
d_MB1_M31 = 400.0;
d_M32_M21 = 3000.0;
d_M22_MP  = 2000.0;
... 210 more lines ...
Attachment 3: distance.m
function d = distance(c1, c2)
    d = sqrt(sum((c1-c2).^2));
end
Attachment 4: draw_beam.m
function draw_beam(c1, c2, color)
    plot( [c1(1), c2(1)], [c1(2), c2(2)], color, 'LineWidth', 2)
end
Attachment 5: draw_measurement.m
function draw_measurement(c1, c2, color, name)
    plot( [c1(1), c2(1)], [c1(2), c2(2)], color)
    text( (c1(1)+c2(1))/2, (c1(2)+c2(2))/2 + 20, name, ...
        'FontSize', 5, 'HorizontalAlignment', 'center', ...
         'VerticalAlignment', 'middle')
end
Attachment 6: draw_point.m
function draw_point(c)
    plot(c(1), c(2), 'go', 'LineWidth', 2, 'MarkerSize', 3);
end
Attachment 7: draw_sos.m
function draw_sos(C, angle)
    global sos_lx sos_ly sos_cx sos_cy tt_lx tt_ly tt_cx tt_cy

    c(:,1) = [-sos_lx/2, -sos_ly/2 + sos_cy-sos_ly/2]';
    c(:,2) = [-sos_lx/2, sos_ly/2 + sos_cy-sos_ly/2]';
    c(:,3) = [sos_lx/2, sos_ly/2 + sos_cy-sos_ly/2]';
    c(:,4) = [sos_lx/2, -sos_ly/2 + sos_cy-sos_ly/2]';
    c(:,5) = [-sos_lx/2, -sos_ly/2 + sos_cy-sos_ly/2]';
    
    m_lx = 25.4*2;
... 18 more lines ...
Attachment 8: draw_tt.m
function draw_tt(C, angle)
    global sos_lx sos_ly sos_cx sos_cy tt_lx tt_ly tt_cx tt_cy

    c(:,1) = [-tt_lx/2, -tt_ly/2 + tt_cy-tt_ly/2]';
    c(:,2) = [-tt_lx/2, tt_ly/2 + tt_cy-tt_ly/2]';
    c(:,3) = [tt_lx/2, tt_ly/2 + tt_cy-tt_ly/2]';
    c(:,4) = [tt_lx/2, -tt_ly/2 + tt_cy-tt_ly/2]';
    c(:,5) = [-tt_lx/2, -tt_ly/2 + tt_cy-tt_ly/2]';
    
    m_lx = 25.4;
... 18 more lines ...
  9586   Wed Jan 29 21:01:03 2014 GabrieleSummaryLSCPRC length measurement analysis

 [Manasa, EricQ, Gabriele]

We managed to measure the PRC length using a procedure close to the one described in slog 9573.

We had to modify a bit the reference points, since some of them were not accessible. The distances between points into the BS chamber were measured using a ruler. The distances between points on different chambers were measured using the Leica measurement tool. In total we measured five distances, shown in green in the attached map.

We also measured three additional distances that we used to cross check the results. These are shown in the map in magenta.

The values of the optical lengths we measured are:

LX    = 6828.96 mm
LY    = 6791.74 mm
LPRC  = 6810.35 mm
LX-LY = 37.2196 mm  

The three reference distances are computed by the script and they match well the measured one, within half centimeter:

M32_MP1  = 117.929 mm   (measured = 119 mm)
MP2_MB3  = 242.221 mm   (measured = 249 mm)
M23_MX1p = 220.442 mm   (measured = 226 mm)

See the attached map to see what the names correspond to.

The nominal PRC length (the one that makes SB resonant without arms) can be computed from the IMC length and it is 6777 mm. So, the power recycling cavity is 33 mm too long w.r.t. the nominal length. This is in good agreement with the estimate we got with the SB splitting method (4cm).

According to the simulation in the wiki page the length we want to have the SB resonate when the arms are there is 6753 mm. So the cavity is 57 mm too long.

Attached the new version of the script used for the computation.

Attachment 1: map.pdf
map.pdf
Attachment 2: script.zip
  9588   Thu Jan 30 19:00:25 2014 GabrieleSummaryLSCPRC length changed

 [Manasa, EricQ, Gabriele]

Today we changed the PRC length translating PR2 by 27 mm in the direction of the corner. After this movement we had to realign the PRC cavity to get the beam centered on PRM, PR2, PR3, BS (with apertures) and ITMY (with aperture). To realign we had to move a bit both PR2 and PR3. We could also see some flashes back from the ETMY . //Edit by Manasa : We could see the ETMY reflection close to the center of the ITMY but the arm is not aligned or flashing as yet//. 

After the realignment we measured again the PRC length with the same method of yesterday. We only had to change one of the length to measure, because it was no more accessible today. The new map is attached as well as the new script (the script contains also the SRC length estimation, with random numbers in it).

The new PRC length is 6753 mm, which is exactly our target!

 

The consistency checks are within 5 mm, which is not bad.

We also measured some distances to estimate the SRC length, but right now I'm a bit confused looking at the notes and it seems there is one missing distance (number 1 in the notes). We'll have to check it again tomorrow.

 

 

 

Attachment 1: map_jan30.pdf
map_jan30.pdf
Attachment 2: survey_v3_jan30.m
clear all


global sos_lx sos_ly sos_cx sos_cy tt_lx ...
       tt_ly tt_cx tt_cy sos_sx sos_sy sos_dy

%% Survey of the PRC+SRC lengths %%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%%

%% measured distances
d_MB2_MY  = 2114 + 27 + 9;
... 446 more lines ...
  9590   Fri Jan 31 19:29:36 2014 GabrieleSummaryLSCPRC and SRC lengths

 Today we measured the missing distance to reconstruct SRC length.

I also changed the way the mirror positions are reconstructed. In total for PRC and SRC we took 13 measurements between different points. The script runs a global fit to these distances based on eight distances and four incidence angles on PR2, PR2, SR2 and SR3. The optimal values are those that minimize the maximum error of the 13 measurements with respect to the ones reconstructed on the base of the parameters. The new script is attached (sorry, the code is not the cleanest one I ever wrote...)

The reconstructed distances are:

Reconstructed lengths [mm]:
LX    = 6771
LY    = 6734
LPRC  = 6752
LX-LY = 37
LSX   = 5493
LSY   = 5456
LSRC  = 5474

The angles of incidence of the beam on the mirrors are very close to those coming from the CAD drawing (within 0.15 degrees):

Reconstructed angles [deg]:
aoi PR3 = 41.11 (CAD 41)
aoi PR2 = 1.48  (CAD 1.5)
aoi SR3 = 43.90 (CAD 44)
aoi SR2 = 5.64  (CAD 5.5)

The errors in the measured distances w.r.t. the reconstructed one are all smaller than 1.5 mm. This seems a good check of the global consistency of the measurement and of the reconstruction method.

NOTES: in the reconstruction, the BS is assumed to be exactly at 45 degrees; wedges are not considered.

 

 

Attachment 1: map_jan31st.pdf
map_jan31st.pdf
Attachment 2: survey_v3.zip
  9611   Fri Feb 7 13:30:22 2014 GabrieleUpdateGeneralIn-Vac Alignment

Quote:

Quote:

I then used the same settings as in ELOG 9554, except I used -1s instead of +1s for the POP110I trigger matrix elements. (I'm not sure why this is different, but I noticed that the PRC would lock on carrier with positive entries here, so I figured we wanted the peaks with opposite sign).

 Nice work!!  As with all the other RF PDs, POP110's phase likely needs tuning.  You want POP110 (and POP22) I-quadratures to be maximally positive when you're locked on sidebands, and maximally negative when locked on carrier.  What you can do to get close is lock PRC on carrier, then rotate the POP phases until you get maximally negative numbers.  Then, when locked on sideband, you can tweak the phases a little, if need be.

 Very good news! We should have a look at the POP110 sideband peak splitting, to see if we really got the right PRC length... 

  9768   Mon Mar 31 21:23:30 2014 GabrieleSummaryLSCMICH sensing oddities in REFL 3F

I'm not sure what may cause this. To back up this measurement/interpretation, I tried to take measurements of these transfer functions across different excitation frequencies via swept sine DTT, but seismic activity kept me from staying locked long enough...

I guess that you get an ellipse when the transfer functions to I and Q have a different phase. One mechanism could be that when driving MICH we make some residual PRCL and this couples with a different transfer function to both I and Q. However, I would expect no phase lag in the PRMI configuration, since there is no enough optical delay in the system to give significant dephasing at few hundreds Hz. This effect might come from mechanical resonances.

It is worth measuring the optical transfer functions from both PRCL and MICH to REFL signals at all frequencies, to see if we have strange features in the TFs.