40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 20 of 339  Not logged in ELOG logo
ID Date Authorup Type Category Subject
  14571   Thu Apr 25 03:32:25 2019 AnjaliUpdateFrequency noise measurementMZ interferometer ---> DAQ
  • Attachment #1 shows the time domain output from this measurement. The contrast between the maximum and minimum is better in this case compared to the previous trials.
  • We also tried to extract the frequency noise of the laser from this measurement. Attachment #2 shows the frequency noise spectrum. The experimental result is compared with the theoretical value of frequency noise. Above 10 Hz, the trend is comparable to the expected 1/f characteristics, but there are other peak also appearing. Similarly, below 10 Hz, the experimentally observed value is higher compared to the theory.
  • One of the uncertainties in this result is because of the length fluctuation of the fiber. The phase fluctuation in the system could be either because of the frequency noise of the laser or because of the length fluctuation of the fiber.  So,one of the reasons for the discrepancy between the experimental result and theory could be because of  fiber length fluctuation. Also, there were no locking method been applied to operate the MZI in the linear range.
  • The next step would be to do a heterodyne measurement. Attachment #3 shows the schematic for the heterodyne measurement. A free space AOM can be inserted in one of the arms to do the frequency shift. At the output of photodiode, a RF heterodyne method as shown in attachment #3 can be applied to separate the inphase and quadrature component. These components need to be saved with a deep memory system. Then the phase and thus the frequency noise can be extracted.
  • Attachment #4 shows the noise budget prepared for the heterodyne setup. The length of the fiber considered is 60 m and the photodiode is PDA255. I also have to add the frequency noise of the RF driver and the intensity noise of the laser in the noise budget.
Quote:
  1. Delay fiber was replaced with 5m (~30 nsec delay)
    • The fringing of the MZ was way too large even with the free running NPRO (~3 fringes / sec)
    • Since the V/Hz is proportional to the delay, I borrowed a 5m patch cable from Andrew/ATF lab, wrapped it around a spool, and hooked it up to the setup
    • Much more satisfactory fringing rate (~1 wrap every 20 sec) was observed with no control to the NPRO
  2. MZ readout PDs hooked up to ALS channels
    • To facilitate further quantitative study, I hooked up the two PDs monitoring the two ports of the MZ to the channels normally used for ALS X.
    • ZHL3-A amps inputs were disconnected and were turned off. Then cables to their outputs were highjacked to pipe the DC PD signals to the 1Y3 rack
    • Unfortunately there isn't a DQ-ed fast version of this data (would require a model restart of c1lsc which can be tricky), but we can already infer the low freq fringing rate from overnight EPICS data and also use short segments of 16k data downloaded "live" for the frequency noise measurement.
    • Channels are C1:ALS-BEATX_FINE_I_IN1 and C1:ALS-BEATX_FINE_Q_IN1 for 16k data, and C1:ALS-BEATX_FINE_I_INMON and C1:ALS-BEATX_FINE_I_INMON for 16 Hz.

At some point I'd like to reclaim this setup for ALS, but meantime, Anjali can work on characterization/noise budgeting. Since we have some CDS signals, we can even think of temperature control of the NPRO using pythonPID to keep the fringe in the linear regime for an extended period of time.

Attachment 1: Time_domain_output.pdf
Time_domain_output.pdf
Attachment 2: Frequency_noise.pdf
Frequency_noise.pdf
Attachment 3: schematic_heterodyne_setup.png
schematic_heterodyne_setup.png
Attachment 4: Noise_budget_1_micron_in_Hz_per_rtHz.pdf
Noise_budget_1_micron_in_Hz_per_rtHz.pdf
  14576   Thu Apr 25 15:47:54 2019 AnjaliUpdateFrequency noise measurementHomodyne v Heterodyne

My understanding is that the main advantage in going to the heterodyne scheme is that we can extract the frequecy noise information without worrying about locking to the linear region of MZI. Arctan of the ratio of the inphase and quadrature component will give us phase as a function of time, with a frequency offset. We need to to correct for this frequency offset. Then the frequency noise can be deduced. But still the frequency noise value extracted would have the contribution from both the frequency noise of the laser as well as from fiber length fluctuation. I have not understood the method of giving temperature feedback to the NPRO.I would like to discuss the same.

The functional form used for the curve labeled as theory is 5x104/f. The power spectral density (V2/Hz) of the the data in attachment #1 is found using the pwelch function in Matlab and square root of the same gives y axis in V/rtHz. From the experimental data, we get the value of Vmax and Vmin. To ride from Vmax to Vmin , the corrsponding phase change is pi. From this information, V/rad can be calculated. This value is then multiplied with 2*pi*time dealy to get the quantity in V/Hz. Dividing V/rtHz value with V/Hz value gives  y axis in Hz/rtHz. The calculated value of shot noise and dark current noise are way below (of the order of 10-4 Hz/rtHz) in this frequency range. 

I forgor to take the picture of the setup at that time. Now Andrew has taken the fiber beam splitter back for his experiment. Attachment #1 shows the current view of the setup. The data from the previous trial is saved in /users/anjali/MZ/MZdata_20190417.hdf5

 

Quote:

If I understand correctly, the Mach-Zehnder readout port power is only a function of the differential phase accumulated between the two interfering light beams. In the homodyne setup, this phase difference can come about because of either fiber length change OR laser frequency change. We cannot directly separate the two effects. Can you help me understand what advantage, if any, the heterodyne setup offers in this regard? Or is the point of going to heterodyne mainly for the feedback control, as there is presumably some easy way to combine the I and Q outputs of the heterodyne measurement to always produce an error signal that is a linear function of the differential phase, as opposed to the sin^2 in the free-running homodyne setup? What is the scheme for doing this operation in a high bandwidth way (i.e. what is supposed to happen to the demodulated outputs in Attachment #3 of your elog)? What is the advantage of the heterodyne scheme over applying temperature feedback to the NPRO with 0.5 Hz tracking bandwidth so that we always stay in the linear regime of the homodyne readout?

Also, what is the functional form of the curve labelled "Theory" in Attachment #2? How did you convert from voltage units in Attachment #1 to frequency units in Attachment #2? Does it make sense that you're apparently measuring laser frequency noise above 10 Hz? i.e. where do the "Dark Current Noise" and "Shot Noise" traces for the experiment lie relative to the blue curve in Attachment #2? Can you point to where the data is stored, and also add a photo of the setup?

 

Attachment 1: Experimental_setup.JPG
Experimental_setup.JPG
  14578   Thu Apr 25 18:14:42 2019 AnjaliUpdatePSLDoor broken

It is noticed that one of the doors (door # 2 ) of the PSL table is broken. Attachement #1 shows the image

Attachment 1: IMG_6069.JPG
IMG_6069.JPG
  14579   Fri Apr 26 12:10:08 2019 AnjaliUpdateFrequency noise measurementFrequency noise measurement of 1 micron source

From the earlier results with homodyne measurement,the Vmax and Vmin values observed were comparable with the expected results . So in the time interval between these two points, the MZI is assumed to be in the linear region and I tried to find the frequency noise based  on data available in this region.This results is not significantly different from that we got before when we took the complete time series to calculate the frequency noise. Attachment #1 shows the time domain data considered and attachment #2 shows the frequecy noise extracted from that. 

As discussed, we will be trying the heterodyne method next. Initialy, we will be trying to save the data with two channel ADC with 16 kHz sampling rate. With this setup, we can get the information only upto 8 kHz. 

Attachment 1: Time_domain_data.pdf
Time_domain_data.pdf
Attachment 2: Frequency_noise_from_data_in_linear_region.pdf
Frequency_noise_from_data_in_linear_region.pdf
  14586   Tue Apr 30 17:27:35 2019 AnjaliUpdateFrequency noise measurementFrequency noise measurement of 1 micron source

We repeated the homodyne measurement to check whether we are measuring the actual frequency noise of the laser. The idea was to repeat the experiment when the laser is not locked and when the laser is locked to IMC.The frequency noise of the laser is expected to be reduced at higher frequency  (the expected value is about 0.1 Hz/rtHz at 100 Hz ) when it is locked to IMC . In this measurement, the fiber beam splitter used is Non PM. Following are the observations

1. Time domain output_laser unlocked.pdf : Time domain output when the laser is not locked. The frequency noise is estimated from data corresponds to the linear regime. Following time intervals are considered to calculate the frequency noise (a) 104-116 s (b) 164-167 s (c) 285-289 s

2. Frequency_noise_laser_unlocked.pdf: Frequency noise when the laser is not locked. The model used has the functional form of 5x104/f as we did before. Compared to our previous results, the closeness of the experimental results to the model is less from this measurement. In both the cases, we have the uncertainty because of the fiber length fluctuation. Moreover, this measurement could have effect of polarisation fluctuation as well.

3.Time domain output_laser locked.pdf :Time domain output when the laser is locked. Following time intervals are considered to calculate the frequency noise (a) 70-73 s (b) 142-145 s (c) 266-269 s. 

4. Frequency_noise_laser_locked.pdf : Frequency noise when the laser is locked

5. Frequency noise_comparison.pdf : Comparison of frequency noise in two cases. The two values are not significantly different above 10 Hz. We would expect reduction in frequency noise at higher frequency once the laser is locked to IMC. But this result may indicate that we are not really measuring the actual frequency noise of the laser.

Attachment 1: Homodyne_repeated_measurement.zip
  14616   Fri May 17 10:12:07 2019 AnjaliSummaryEquipment loanBorrowed component

I borrowed one Marconi (2023 B) from 40 m lab to QIL lab.

  8183   Wed Feb 27 14:39:59 2013 AnnalisaUpdateLSCFibre laid for RFPD audio

 [Annalisa, Jenne, Rana, Steve]

We installed the fibres on cable trays the 1Y2 and the Control Room.

Still to do: find a power supply for the Fiboxes and plug everything in.

  8190   Wed Feb 27 19:27:29 2013 AnnalisaHowToCOMSOL TipsMirror support Eigenfrequency

 I studied the eigenfrequencies of a mirror support using COMSOL.

 

Attachment 1: IronSupport.png
IronSupport.png
Attachment 2: IronSupportEigenfreq.png
IronSupportEigenfreq.png
  8226   Mon Mar 4 20:03:42 2013 AnnalisaHowToCOMSOL TipsStudy of mirror mount eigenfrequencies

 I studied the eigenfrequencies of a mirror mount designed with COMSOL.

I imposed fixed constraints for the base screws and for the screw connecting the base with the pedestal. Note that the central screw is connected to the base only for a small thickness, and the pedestal touches the base only with a thin annulus. This is in way to make a better model of the actual stress.

Shown in fig. 2 is the lowest eigenfrequency of the mount.

I' going to change the base and study the way the eigenfrequency vary, in way to find the configuration which minimizes the lowest eigenfrequency.

 

Attachment 1: MirrorSupport1.png
MirrorSupport1.png
Attachment 2: MirrorSupportEig1.png
MirrorSupportEig1.png
Attachment 3: pedestal.png
pedestal.png
Attachment 4: Base2.png
Base2.png
  8244   Wed Mar 6 18:51:07 2013 AnnalisaUpdateAlignmentAuxiliary laser installed for FSR and TMS measurement of the PRC

We want to measure the g-factor of the PRC using the beat note of the main laser with an auxiliary NPRO laser.

We are going to phase lock the NPRO to the main laser (taking it from POY) and then we will inject the NPRO  through the AS edge of the ITMY.

Today Sendhil and I installed the auxiliary laser on the ITMY table moving it from the AS table.

We also installed the beam steering optics, except the BS which will launch the beam through the AR edge of the ITMY.

To do: install the BS, take the POY beam and mix it with the auxiliary laser on a photodiode to phase lock the two lasers, do better calculations for the mode matching optics to be used for the auxiliary laser beam.

Attachment 1: IMG_3-6-13.JPG
IMG_3-6-13.JPG
  8257   Fri Mar 8 12:57:57 2013 AnnalisaUpdateABSLBS installed on ITMY table

 Sendhil and I installed the S polarized BS on the ITMY table to steer the NPRO beam through the AR wedge and align it to the POY beam. 

We took a shutter from the BSPRM table (which was not used) and a beam dump from the AS table (which was used by the auxiliary laser already removed and installed on the ITMY).

To do: do better alignment of the NPRO beam, maybe installing some iris after the BS and before the AS wedge, phase lock the two beams. 

  8303   Mon Mar 18 12:02:12 2013 AnnalisaConfigurationABSLABSL setup for g-factor measurement of PRC
Motivations
The ABSL technique has been already used in the past to measure the absolute length of the interferometer's optical cavities by means of an auxiliary laser source, as described in LIGO-P1200048-v3 and in Alberto Stochino thesis work.
Using the same technique it is possible to measure the g-factor of the power recycling cavity by measuring the cavity Transverse Mode Spacing.
 
Plan for experimental setup
The auxiliary laser is set on the POY table and is injected through the ITMY window in way to follow the same path of the POY beam. It hits the AR wedge of ITMY and is reflected back to the BS and the PRM.
 
Since the main beam is P-polarized, all the optics in the central IFO are P-polarization dependent, so it is useful to P-polarize the auxiliary beam before it enters the IFO.  
I made a mode matching calculation with a la mode script, in order to mode match the auxiliary beam waist to the waist of the main laser.
However, before ordering and installing steering optics and mode maching lenses, I'm waiting to know whether someone has an NPRO laser to install on the END table in place of the broken one, otherwise the one I'm using could be taken.
In this case a possibility could be to take the auxiliary beam from the end table with an optical fiber, but it means to use the auxiliary laser alternately to lock the arm or make a measurement of TMS. If so, a new calculation for the mode matching needs to be done.
Anyway, I hope that another laser will be found!
 
In order to phase lock the auxiliary beam with the main beam, the latter will be taken from the PSL table after the PMC through a single mode fiber, which will be brought up to the POY table. This solution results to be more reliable then taking the POY beam to phase lock the two laser, because POY is related to the locking. 
 
The signal with the beat note between the two lasers can be detected by the transmission from PR2 (POP). 
 
 
 
  8305   Mon Mar 18 12:35:29 2013 AnnalisaBureaucracyAuxiliary lockingYend table upgrade - go fetch NPRO from ATF

Quote:

Zach has just replied, and said that we should feel free to take the laser from his iodine setup in the West Bridge subbasement, in the ATF lab. 

Annalisa, please ask Koji or Tara to show you where it is, and help you bring it to the 40m.  You should install it (temporarily) on the PSL table, measure the waist, and find the beat in IR.  Elog 3755 and elog 3759 have some of the details on how it has been done in the past.

 Ok, I'm going to contact Koji.

  8308   Mon Mar 18 20:13:18 2013 AnnalisaBureaucracyAuxiliary lockingYend table upgrade - go fetch NPRO from ATF

Quote:

1) Annalisa is going to start  working on mode profiling and beat note search for the old MOPA NPRO.

2) In the meantime, Manasa is working on the end table items. This will be reviewed by KA in the afternoon.

The laser at ATF is moved to the 40m when the status of 1) and 2) is determined by KA to be reasonable.

We also make the beat note measurement for the ATF laser too.
 

Today I installed mirrors to steer the pick-off from the main laser beam in a more free part of the PSL table and make the beat note measurement between it and the NPRO.

At the beginning I took the beam from the harmonic separator after the doubling crystal, and I was going to bring it in a less full part of the table . At the end I realized that there was already a beam steered up to a more free part of the table, and the beam is taken from the transmission of the PMC.

Tomorrow I'm going to use that beam to find the beat note with the NPRO.

I also removed almost all the steering  optics that I used on the ITMY table to send the auxiliary beam for ABSL through the window parallel to the POY beam. The most important thing is that I removed the BS, which was on the same path of the POY beam (see elog 8257).

 

  8313   Tue Mar 19 20:24:56 2013 AnnalisaUpdateAuxiliary lockingAuxiliary laser on PSL table

 I moved the auxiliary laser from the ITMY table to the PSL table and installed all the optics (mirrors and lenses) to steer the beam up to a PDA55 photodiode, where also the pick-off of the PSL is sent.

Tomorrow I'm going to measure the beat note between the two.

  8322   Thu Mar 21 09:53:46 2013 AnnalisaUpdateLockingFinding the beat note

 Yesterday I tried to find the beat note between the main PSL and the auxiliary NPRO, but I didn't :( 

Today I will do a better alignment of the two beams in the PD and try again.

  8327   Thu Mar 21 13:11:42 2013 AnnalisaUpdateLockingFinding the beat note

Quote:

Give us more info on the elog:
What PD are you using? How much power the beams on the recombining BS are? What kind of BS is it?
How are you looking for the beat note? (on the scope? or spectrum analyzer?)
What was the scanned temp range?

Three points to be checked:

- Polarization

- Alignment

- Temperature

Experimental Setup

I'm using a 1611 New Focus PD (1 GHZ, with maximum input power 1mW), and the total power hitting on the PD is of about 0.650 mW.

The current of the NPRO laser is set to 1.38 A, so that the input power is 19 mW. The beam is initially damped by a 10% reflection BS and then it hits a 33% reflection BS (where it recombines with the PSL pick-off beam) with 2 mW power.

After this second BS the power is reduced to 0.592 mW.

The PSL pick-off hits on the 33% reflection BS with 65.5 uW power, and it exit with a 47 uW power.

I connected a power supply to apply a Voltage to the slow frequency BNC, in way to tune the laser frequency.

I'm using the AGILENT 4395A Spectrum analyzer to make the measurement. I tried to use the HEWELETT PACKARD 8591E spectrum analyzer, but the monitor didn't turn on.

The temperature spanned until now in only of about 10 deg C, because I realized that I needed a better alignment, so I added a lens in front of the PD and I did a better alignment. 

Moreover, the current of the laser is too low, so I need to increase it and add more beam splitters in the beam path to dump the beam, in way to don't reach the PD threshold.

I knew that both the beams are s-polarized, but maybe I can check it again.

Attachment 1: Beat_note_setup.jpg
Beat_note_setup.jpg
  8329   Thu Mar 21 14:02:45 2013 AnnalisaUpdateLockingincident angles

Quote:

Is there a reason to use non-45 degree incident angles on the steering mirrors between the laser and the PD?  I would always use 45 degree incident angles unless there is a really good reason not to.

 Actually not, it is a mistake! It is one of the things I'm going to modify, in addition to add more BS to reduce the power. 

  8333   Fri Mar 22 23:23:38 2013 AnnalisaUpdateLockingBeat note still missing

[Annalisa, Manasa, Koji]

I updated the setup for the beat note. The main reason is that I needed to keep the ADJ to 0 in way to operate at the nominal laser power.

Now the input power of the laser is increased (about 315 mW) and needs to be dumped so as not to exceed the PD threshold of 1mW. 

Moreover, a lens has been added to match the two beams size.

A BS has been removed from the PSL pick-off beam path, so the PSL power hitting the BS is now about 100 uW, and the total power on the PD is 0.7mW.

I also verified that both the beams are S polarized.

To find the beat note, the laser temperature has been varied  through the laser controller and not adding a Voltage with the power supply.

A range of temperature of 30 degC has been spanned, but we suppose there should be some calibration problem with the controller, since set temperature is not the same as Laser temperature on the display. 

Anyway, no beat note has been found up to now.

 

An external monitor has to be added to check the real temperature of the crystal.

The next possible plan is to vary the PSL temperature and try to find the beat note. 

 

P.S.: The HEWELETT PACKARD 8591E spectrum analyzer works! The monitor only took some time to turn on!

 

Attachment 1: Beat_note_-_new_setup.jpg
Beat_note_-_new_setup.jpg
  8345   Mon Mar 25 23:20:57 2013 AnnalisaSummaryAuxiliary lockingBeat note found!

[Annalisa, Manasa]

The beat note between the main PSL and the auxiliarly NPRO has been found!

The setup didn't change with respect to the one described on the previous note on the elog. A multimeter has been connected to the laser controller diagnostic pin to read out the voltage that indicated the laser crystal temperature.

The connector has been taken from the Yend table laser controller.

The voltage on the multimeter gave the same temperature shown by "Laser temperature" on the display of the controller, while "set temperature" was wrong.

The temperature has been varied using the laser controller with reference to the voltage read on the multimeter display.

Starting from 35.2 °C, the temperature has been first lowered until 20 °C and no beat note has been found, then temperature has been increased up to 35.2 °C and the first beat note has been found at 38.0 °C.

It has been detected at a frequency of about 80 MHz with an RF power of -27 dBm and a frequency fluctuation of about  +/- 4 MHz.

To do:

I made more measurements slowly varying the laser temperature, to see how the beat note frequency changes with it. I'll make the plot and post it as soon.

  8361   Wed Mar 27 21:53:21 2013 AnnalisaUpdateABSLBeat note of ATF auxiliary laser

After measuring the beat note, the "Alberto" NPRO auxiliary laser has been moved from the PSL table to the POY table. Its beam profile is going to be measured. It's going to be used as green laser on the END table, in place of the broken one.

The auxiliary laser borrowed form ATF lab (which will be used for the ABSL measurement) has been set on the PSL table to make a measurement of the beat note between it and the main laser.

The setup is mostly the same of the previous beat note measurement . In this case, laser input power is 326 mW, so I needed to replace one of the mirrors of the steering optics with a BS 50% reflecting in order to have less than 1 mW on the PD.

Now, the total power on the PD is less than 0.5 mW.

I didn't measure the beat note yet to leave the PSL table as quite as possible for the locking procedures.

To do:

Measure the beat note, fiber coupling the NPRO laser to bring it to the POY table.

 

  8368   Thu Mar 28 19:32:22 2013 AnnalisaSummaryAuxiliary lockingBeat note found!

 

 I plot the variation of the beat note frequency as a function of "Alberto" NPRO laser's temperature.

After some discussion, now I'm going to vary the PSL temperature and find the auxiliary NPRO temperature matching to have the beat note between the two.

Attachment 1: BeatFreq.jpg
BeatFreq.jpg
  8369   Thu Mar 28 23:00:30 2013 AnnalisaUpdateABSLBeat note of ATF auxiliary laser found

 

The beat note for the ATF lab laser has been found. 

The measurement has been carried out in the same way as described in elog 8368.

The only difference is that in this case I started from a temperature of 35.2 degC, and I reduced it until the minimum which was 30.71 degC. No beat note in this range.

Then I rised on the temperature and I found the first beat note at 41.46 degC. It has been detected at a frequency of about 120 MHz with an RF power of -53 dBm and a frequency fluctuation of about  +/- 5 MHz. 

I tried to improve the alignment to have a stronger beat, but it was the maximum I could reach. Maybe I could increase the power hitting the photodiode, which was 0.453 mW. 

 

 

  8370   Thu Mar 28 23:06:48 2013 AnnalisaUpdateAuxiliary locking"Alberto" NPRO laser again on PSL table

 "Alberto"NPRO laser has been moved again on PSL table in order to make a measurement of the beat note varying also the PSL temperature.

It is useful because if the PSL temperature would drift  we have to know which is the NPRO temperature that returns the beat.

I'm going to measure it tomorrow.

 

 

  8386   Mon Apr 1 23:22:17 2013 AnnalisaUpdateAuxiliary lockingBeat note between "Alberto" NPRO laser and PSL laser

I measured the beat note between the "Alberto" NPRO laser and the PSL varying the PSL temperature and find the matching NPRO temperature that gave the beat.

I first switched off the FSS loop for the PSL, then I varied its temperature and switched on the loop back.

PSL temperature has been varied starting from 31.88 °C (its starting temperature) down to 23.88 by 1°C step, and then from 31.88 °C up to 36.92 °C, always with a 1°C step.

For each PSL temperature, the NPRO temperature was varied as well, in way to find the temperature to have a beat note between the two.

The trend of the NPRO laser temperature reminds the frequency change of the laser as a function of the crystal temperature continuous tuning.

I made measurements only for the first temperature of the NPRO laser which gave me the beat note. Tomorrow I'm going to find the beat note also for higher frequencies of the NPRO laser.

 

Attachment 1: Beat_Note.jpg
Beat_Note.jpg
  8396   Tue Apr 2 22:39:17 2013 AnnalisaUpdateAuxiliary lockingBeat note between "Alberto" NPRO laser and PSL laser

 

 The beat note between the PSL laser and the "Alberto" NPRO laser has been measured. In particular, for each PSL temperature, more than one Aux laser frequency has been found.

The second of the three curves seems to be more stable than the other two, even if a "step" trend can be found in all of them (maybe due to the frequency change of the NPRO laser as a function of the crystal temperature continuous tuning, as mentioned in the previous elog). This is the reason why the points are not perfectly aligned, and the errors on the fit parameters are so big.

 

 

Attachment 1: Beat_note_3col.jpg
Beat_note_3col.jpg
  8437   Wed Apr 10 15:49:22 2013 AnnalisaConfigurationCOMSOL TipsYend table eigenfrequency simulation with COMSOL

 I made a Simulation with COMSOL for the Yend table. Mainly, I tried to see how the lower eigenmode changes with the number and the size of the posts inside.

The lateral frame is just sitting on the table, it is fixed by its weight. I also put a couple of screws to fix it better, but the resulting eigenfrequency didn't change so much (less than 1 Hz). 

In Fig. 1 I didn't put any post. Of course, the lowest eigenfrequency is very low (around 80 Hz).

Then I added 2 posts, one per side (Fig. 2 and Fig. 3), with different diameter.

In some cases posts don't have a base, but they are fixed to the table only by a screw. It is just a condition to keep them fixed to the table

Eventually I put 4 posts, 2 per side. 

The lowest eigenfrequency is always increasing.

At the end I also put a simulation for 4 1.6 inch diameter posts without base, and the eigenfrequency is slightly higher. I want to check it again, because I would expect that the configuration shown in Fig.5a could be more stable.

P.S.: All the post are stainless steel.

 

Attachment 1: Pics_end_table.pdf
Pics_end_table.pdf Pics_end_table.pdf Pics_end_table.pdf
  8495   Fri Apr 26 10:50:07 2013 AnnalisaUpdateABSLATF laser on PSL

The ATF NPRO auxiliary laser has been moved on the PSL table. All the optics for beat note measurement are in place and alignment has been done.

The setup for this measurement is the same as described in elog 8333.

  8561   Fri May 10 20:05:21 2013 AnnalisaUpdate40m UpgradingEndtable upgrade for auxiliary green laser : progress

I rotated some mounts along the green beam path, and I started aligning the beam again.

The beam is aligned up to the waveplate just before the doubler crystal, even if I couldn't reach more than 88% transmission for the Faraday. Next week I will finish the alignment and I'll put the lenses that Manasa already ordered.

 

  8565   Mon May 13 21:35:55 2013 AnnalisaUpdate40m UpgradingEndtable upgrade for auxiliary green laser : progress

Yend table - Current status

OPLEV

Today the 2m focal length lens along the oplev path (just after the laser) has been added. In Manasa's layout it allows to have a beam waist of 3.8mm on the OPLEV QPD, even if it seems to be smaller.

The laser is closer to the box wall than the layout shows (it's on the line n.1 instead of line n.9), so maybe it has to be moved in the position shown in the layout, as Steve suggests, to leave empty space just before the window.

Rana suggests a 2mm diameter beam on the QPD, so a new calculation has to be done to add a second lens.

GREEN

The beam has been aligned until the doubler, but after the crystal it it has a small tilt, so a better alignment has to be done.

Moreover, the beam waist has to be measured after the Faraday for the green, in way to choose the focal length of the lenses necessary for the mode matching.

Then the three steering mirrors to send the beam into the arm have to be put.

TRANSMON PATH

A lens which has to be put on the Transmon path (already ordered) has to be added, and the beam alignment on the QPD-y and on the PDA520 has to be done.

  8576   Tue May 14 21:03:15 2013 AnnalisaUpdate40m UpgradingEndtable upgrade for auxiliary green laser : progress

 

GREEN

The new lenses arrived, and I put the right 250mm before the doubler. I'm still not so confident with the alignment, because I cannot get more than 11-12 uW out from the "green" Faraday, with more than 200uW going in.

TRANSMON

I replaced the Y1 mirror with an HR1064-HT532. The alignment has to be done. Today the 50cm focal length lens arrived, and I'm going to put in tomorrow.

 

  8584   Wed May 15 21:27:39 2013 AnnalisaUpdate40m UpgradingEndtable upgrade for auxiliary green laser : progress

 

GREEN

I still have problems in maximizing the power out from the doubler. I realized that the real green power I obtain is about 30 uW, and it is the power which really enters the Faraday.

Before I was measuring it just after the Harmonic separator, and there was some residual IR beam which increased the power on the power meter, that's why I obtained about 200 uW.

I also tried to slightly vary the position of the mode matching lens, but I was not able to get more than 30 uW on the power meter.

TRANSMON PATH

The 50 cm focal length lens has been added in the position shown on Manasa's layout, and the beam has been focused on the PD.

 

 

  8594   Fri May 17 00:32:32 2013 AnnalisaUpdate40m upgradingETMY - progress

[Rana, Annalisa] 

 GREEN

 The alignment for the green has been improved, so that we have much more green power.

The first lens position along the IR path has been changed in way to have the beam waist at the center of the first Faraday. In this way we had about 91% of the input power out from it.

The two cylindrical lenses which were used to correct the ellipticity of the beam have been replaced by a single lens. Its focal length is intermediate between the focal lengths of the two cylindrical. 

Moving the position of the lens before the doubler crystal and improving the alignment we got about 1mW of green light (0.35% of the incoming IR beam).

TO DO

 

After aligning the green beam through the second Faraday, the beam waist of the outgoing beam has to be measured and the mode matching calculation has to be done to choose the two MM lenses. Then the steering mirrors will be placed to send the beam into the arm.

Attachment 1: IMG_0536.JPG
IMG_0536.JPG
  8597   Fri May 17 18:24:04 2013 AnnalisaUpdate40m upgradingETMY - progress

I aligned the green beam into the Faraday. I needed an HWP to have the right polarization for the light entering the Faraday itself.

I tried to dump as much beams as possible with razor dumps, but eventually I had to use some "temporary solutions" for higher beams, because I didn't find the right mounts for razor dumps.

I measured the beam waist after the Faraday with the beam scan. Analysis and MM calculation to follow.

  8622   Thu May 23 00:16:32 2013 AnnalisaUpdate40m upgradingETMY - progress

 [Annalisa, Koji] 

 GREEN

I aligned back the beam (we lost part of the alignment after we put back the box and after the posts were installed). The green beam out from the crystal is still low, but anyway I get about 1.2 mW of green out from the Faraday. 

TO DO 

Mode Matching calculation (tomorrow)

Fix the dumping situation

Replace some of the mounts with more solid ones (in the future)

TRANSMON PATH

 QPD, PD and Camera have been rotated as Rana suggested last Wednesday. A 1m focal length lens is on the main beam transmitted path (before the harmonic separator), and the beam diameter on the QPD is about 5mm. We put another lens with a shorter focal length to put the PD very close to the beam waist and in way to have a reasonable beam size on the camera. Tomorrow I will write down all the correct sizes of the beams.

OPLEV 

(for Steve) I marked a possible beam path for the Oplev (the laser is not in the right place in the picture, but I left it in the correct place on the table). I also put the QPD for the IP-ANG, so we know in which part of the table the beam can be steered.

The space in the red rectangle (right corner) has to be left empty to put a PD for the rejected beam from the green Faraday.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Attachment 1: TransMonAndOplev.jpg
TransMonAndOplev.jpg
  8637   Fri May 24 02:12:50 2013 AnnalisaUpdate40m upgradingETMY - Mode Matching for green

 

 Mode Matching calculation for green beam - Yarm

After measuring the beam radius out from the Faraday for the green, I made the calculation to match the green beam mode with the IR mode inside the arm.

The beam waist after the Faraday is elliptical, and I found the following value for the waist:

w0x = 3.55e-5 m @ z0x = -0.042 m

w0y = 2.44e-5 m @ z0y = -0.036 m

(the origin of the z axis is the output of the Faraday, so the waist is inside the Faraday itself)

I did the calculation using a la mode, using as beam waist and its position the following values:

w0 = sqrt(w0x*w0y) = 2.943e-5 m @ z0 = (z0x+z0y)/2 = -0.039 m

The results are shown in the attached plots.

                      Focal length (m)             position (m)

lens1            0.125                                0.1416

lens2            0.100                                0.5225

L                    1.000                                1.5748 (fixed lens used to focus transmitted beam)

 

As the first plot shows, the green beam size on the ETMY is about 6mm. My concern is that it could be too big.

The third plot shows the X and Y section of the beam. It is strongly elliptical, but nevertheless the coupling factor calculated with Koji's formula  gives C=0.936 for the astigmatic beam, and C=0.985 for the non astigmatic beam, so it seems to be still ok.

 

 

Attachment 1: ModeMatchingGreen.jpg
ModeMatchingGreen.jpg
Attachment 2: ModeMatchingGreenZoom.jpg
ModeMatchingGreenZoom.jpg
Attachment 3: XYpath.jpg
XYpath.jpg
  8639   Fri May 24 12:50:25 2013 AnnalisaUpdate40m upgradingETMY - Mode Matching for green

Quote:

I got confused. Is the mode calculation in the cavity correct?
Are you sure the wavelength in the code is 532nm?

The first plot says "the waist radius at ITMY is 2.15mm". This number is already very close to
the waist size of the cavity mode (2.1mm@ITM, 3.7mm@ETM), but the spot radius at ETMY is 6mm.
They are inconsistent.

 

 Jenne and I just realized that a la mode has 1064e-9 m as default value. I'll change it and make the calculation again.

  8645   Sat May 25 02:03:48 2013 AnnalisaUpdate40m upgradingETMY - Mode Matching for green - new calculation

 Mode matching calculation for green - Yarm

I did again the mode matching calculation. The previous one was using 1064nm as wavelength, so it was wrong.

The seed beam waist and its position are the same as in elog 8637. The new results are shown in the attached graphs.

I got the following values for focal lengths and positions of the two Mode Matching lenses:

 

                    Focal length (m)             Distance from the Faraday output (m)

lens1            0.125                                                  0.1829

lens2           -0.200                                                  0.4398

L                   1.000                                                  1.4986 (fixed)

The position of the lens L has changed because the path lengh has been slightly  reduced. 

The Coupling factor for he astigmatic beam is C = 0.959 (it is C = 0.9974 if we consider the beam as non astigmatic).

I put the lenses and aligned the beam up to the shutter, which has been moved from its initial position because the beam size on it was too large. 

TO DO

The green beam needs to be aligned and sent into the arm cavity. 

Polarization has to be checked.

Many beams still have to be dumped, both in IR and Green paths. 

 

 

 

 

Attachment 1: ModeMatchingGreenNEW.jpg
ModeMatchingGreenNEW.jpg
Attachment 2: ModeMatchingGreenZoomNEW.jpg
ModeMatchingGreenZoomNEW.jpg
Attachment 3: XYpathNEW.jpg
XYpathNEW.jpg
Attachment 4: photo1.JPG
photo1.JPG
Attachment 5: photo2.JPG
photo2.JPG
  8646   Mon May 27 21:38:53 2013 AnnalisaUpdate40m upgradingETMY - Beam Dumps

 

I put many razor dumps along the IR/green path. The rejected beam from the IR Faraday needs to be dumped (about 1.5 mW). I used all the new razor blade I had, so I need one more for that beam.

The IR reflection of the Harmonic separator right after the doubler needs to be dumped in a better way. At the moment there is a black screen, but we need something suitable to dump more than 300 mW.

After the second steering mirror along the green beam path there is a very small transmission (about 6 uW), which is difficult to dump because there is no space enough. Can it be dumped with a black screen?

 

The Oplev has a lot of reflection hitting the central BS (The BS for the transmitted beam). It is very difficult to dump them without intercepting the main beam path. Maybe we have to slightly change the Oplev beam angle to avoid so many reflections.

 

  8652   Tue May 28 22:11:23 2013 AnnalisaUpdate40m upgradingETMY - Yarm shutter cable

 

 For some strange reason the Yarm shutter cable runs up to the POY table, where it is connected to another cable going to the rack. It has to be put off from the table, at least. It would be better to have only one cable going directly to the rack.

I roughly aligned the green into the Yarm and I've seen the green beam flashing on the PSL table, but the mode matching is not so good and I get an higher order mode, so I'm going to fix the mode matching tomorrow.

  8660   Thu May 30 20:45:53 2013 AnnalisaUpdateendtable upgradeETMY - oplev

Quote:

Quote:

 Temporary oplev in place. The spot on the qpd is still big. My two lens solution did not work.

I will finalize optical component position of the oplev after the the arm transmitted and green beam optics in place. They have priority.

 

 Oplev spot size on qpd ~ 1 mm 

PS: I realized it later that the returning beam is going through a lens for TRY. This is a nono.

      This beam path will be relayed again as the TRY, green beam and IP-ang get there place.

 Oplev is disabled. I removed one of the steering mirrors because it was on the green beam  path.

  8662   Fri May 31 11:15:01 2013 AnnalisaUpdateendtable upgradeETMY - Mode Matching for green - new calculation

 Since the beam waist after the Faraday had changed since the last time I measured it (maybe alignment changed a bit), I made a new mode matching calculation for green. I attached the results.

I'm going to align the beam into the Yarm.

RXA: JPG images deleted - replace with PDF please.

  8671   Tue Jun 4 16:04:34 2013 AnnalisaUpdateLSCALS-TRY_OUT needs anti-whitening

Quote:

 After working some more on the EY table, we are getting some TEM00 flashes for the Y arm green. We have had to raise the height of one of the MM lenses to prevent clipping.

We used a function generator to apply a ~300 mV 10 Hz triangle wave to scan the laser frequency while aligning.

We tried to use the C1:ALS-TRY_OUT channel to help us in our alignment but there are a couple problems:

1) It seems that there is an uncompensated whitening filter before the ADC - Annalisa is making a compensation filter now.

2) The data delay is too much to use this for fast alignment. We might need to get a coax cable down there or mount a wired ethernet computer on the wall.

3) We need to make DQ channels for the TRY and TRX OUT. We need long term data of these, not just test points.

 I made the anti-whitening filter for the C1:ALS-TRY_OUT channel. But then I forgot to make an ELOG because I am bad.

  8672   Tue Jun 4 17:28:09 2013 AnnalisaUpdateLSCC1:ALS-TRY_OUT filter and green progress

 [Annalisa, Gautam, Rana]

I made the anti-whitening filter for the C1:ALS-TRY_OUT channel.

zpk [[150],[15],1] Hz

Now we can look at the picks of this signal to align the green into the cavity.

We already had some 00 flash, but a better alignment has to be done.

TO DO:

- put the shutter along the beam path

- check the polarization (we have a new PBS for visible)

 

Attachment 1: green.JPG
green.JPG
Attachment 2: QUAD1_1054358327.mp4
  8674   Tue Jun 4 21:50:23 2013 AnnalisaUpdateLSCC1:ALS-TRY_OUT filter and green progress

 

 [Annalisa, Gautam]

 The green beam alignment has been improved, so we see much more 00 bright flashing. We checked the polarization and the Ygreen shutter is back in place.

A mirror is already in place to steer the rejected beam from the green Faraday into a PD, tomorrow morning we'll put a lens and the PD to take the signal for PDH locking.

 

  8679   Wed Jun 5 14:43:42 2013 AnnalisaUpdateLSCALS-TRY_OUT DQ channels

Quote:

 After working some more on the EY table, we are getting some TEM00 flashes for the Y arm green. We have had to raise the height of one of the MM lenses to prevent clipping.

We used a function generator to apply a ~300 mV 10 Hz triangle wave to scan the laser frequency while aligning.

We tried to use the C1:ALS-TRY_OUT channel to help us in our alignment but there are a couple problems:

1) It seems that there is an uncompensated whitening filter before the ADC - Annalisa is making a compensation filter now.

2) The data delay is too much to use this for fast alignment. We might need to get a coax cable down there or mount a wired ethernet computer on the wall.

3) We need to make DQ channels for the TRY and TRX OUT. We need long term data of these, not just test points.

 [Jenne, Annalisa]

DQ channels have been created in the C1ALS model for TRX and TRY. They are called TRX_OUT and TRY_OUT and the sampling rate is 2048 Hz.

  8680   Wed Jun 5 15:03:42 2013 AnnalisaUpdateLSCC1:ALS-TRY_OUT filter and green progress

Quote:

The rejected beam from this Faraday comes out at a tiny, tiny angle and so its tough to pick it off without clipping the main beam.

Some care must be taken in setting this up - Steve may have some good ideas on what kind of mount can be placed so close to the beam.

Why did we ever order this terrible Faraday? Let's never get a Faraday with a tiny angle between the beams again.

The rejected beam from the Faraday is steered with a mirror into the PDA32A PD  and a 75mm fl lens is used to focus the beam into it.

The main beam is a few millimeters away from the mirror mount (maybe 2mm), and I think it should be fine as long as the main beam is not supposed to move.

 

Attachment 1: faraday_rejected_beam.JPG
faraday_rejected_beam.JPG
  8691   Sat Jun 8 00:24:11 2013 AnnalisaUpdateGreen LockingY arm locked with green but bad mode matching

[Annalisa, Nic]

After connecting the PD with the reflection from the arm  to the PDH box, theY  arm has been locked on the 01 mode.  Maximizing the alignment, we obtained a 00 mode locking, but we couldn't maximize the power.

The size of the reflected beam was different with respect to the size of the incoming beam, so probably a bad mode matching was one of the issues.

Moreover, the reflected beam is very low power. We need to figure out why it is so (bad alignment? related to mode matching?)

 

After measuring better all the distances, I did a new mode matching calculation. I put the lenses after measuring the beam waist, so the size of the beam on the lenses was the same as expected from the calculation. Nevertheless, the beam size on the beam splitter looks bigger than expected, and also in this case green flashes into the cavity at some HOM (again 01).

I also tried to lock again the cavity and maximize the alignment, but I didn't get any improvement with respect to the previous mode matching.

 

  8692   Mon Jun 10 21:39:26 2013 AnnalisaUpdateGreen LockingY arm locked with green but bad mode matching

 

Still no good locking!

After making the reflected beam size closer to the injected one, I maximized alignment. I locked again in 00 mode, but I couldn't maximize the power. 

I just realized that maybe I'm not using the correct radius of curvature for the ETMY in the simulation. Tomorrow I will start checking from that.

  8695   Wed Jun 12 01:56:58 2013 AnnalisaUpdateGreen LockingY arm locked with green but bad mode matching

 

 For the mode matching calculation I was using the ETMY focal length that I found on Kiwamu's plot on the wiki page. 

Taking into account also the substrate, the focal length turns out to be

fl = ((n-1)*(1/R1 - 1/R2 + (n-1)d/(nR1R2)))^(-1) = -125.81 m

with n = 1.46071 (refraction index of fused silica at 532nm)

R1 = 5625 m (radius of curvature of the first surface)

R2 = 57.37 m (radius of curvature of the second surface)

d = 25mm (thickness)

 

The value of the focal length is sligthly different from the one I was using before in the calculation, but maybe it is enough to change the coupling.

The mode matching solution I found is very sensitive to the lenses position. 

The beam waist position can vary up to 20m varying by 1cm the first lens position, while it is slightly less sensitive to the second lens displacement.

As shown in the picture, along the green beam path there is also a 1m focal length lens. It's position is fixed, because it is along the IR transmetted beam path also. I tried to get a better solution without it, but I found that the waist position was still strongly dependent on one of the two lenses position, so it would not solve the problem to remove this lens.

I think that the main issue of this mode matching is related to the "space contraints", because the two lenses' positions can vary in a very small space, even though the green beam path on the table is quite long.

Eventually, I put the MM lenses found from this last simulation on the table, and it seems to work, since I've seen very strong 00 flashes. Unfortunately, while trying to maximize the alignment I broke it  and I have to do it again, but I feel confident! 

Attachment 1: waistVaryingLensPos.pdf
waistVaryingLensPos.pdf
Attachment 2: green.JPG
green.JPG
Attachment 3: ModeMatchingY.pdf
ModeMatchingY.pdf
ELOG V3.1.3-