Wed Jul 2 19:15:52 2008 |
Alberto | Update | General | Status of the alignment of the NPRO beam for the Absolute Length Measurement |
Today I've tried to bring the frequency of the NPRO laser close enough to that of the IFO beam so that the beat between the two beams can be at a detectable frequency for the photodiode. The way I've been changing the frequency is by the NPRO's temperature control on its driver.
Looking at the signal from the AS OSA should enable us to monitor the direction in which the frequency is changing. Every time the resonances of the IFO beam and of the NPRO beam overlap, we know that the frequencies of the two beams are some FSR of the OSA away from each other. At the overlapping of the resonances, if the difference of frequency is within the detectable range of the photodiode, we should see a peak in the network/spectrum analyzer.
This way turned out not very easy in practice because from the AS OSA one can hardly distinguish the resonances of the primary beam from those of the secondary beam. The cause is mainly the flashing of the IFO beam at the AS port which produces a pattern of resonances of different amplitude. Also for some reason, triggering the output signal from the OSA at the oscilloscope doesn't work very well.
However, even if we didn't have these problems, I think that the two beams are not very well aligned, at least not anymore. I'm attaching some pictures from the AS port. The bright spot on the left is the NPRO beam and the one in the center which flashes is the IFO beam. We probably need some more work in the alignment of the NRPO beam. |
Attachment 1: DSC_0156.JPG
|
|
Attachment 2: DSC_0158.JPG
|
|
628
|
Thu Jul 3 11:53:30 2008 |
Koji | Update | General | Status of the alignment of the NPRO beam for the Absolute Length Measurement |
The method itself looked fine.
Use of the one arm configuration will make the work easier as constant power at the AS port is obtained.
How much is the FSR of the OSA?
Apparently the alignment is not good any more as Alberto pointed. Everytime you touch the flipper you'd better to adjust it. Then, if necessary, adjust the injection steerings.
If the PSL beam is blocked, only the injection beam appears at the optical ports. The spot is obtained at the AS port and the SY port (REFL) at the same time. I recommend to confirm the transmittion to the SY port too by the CDD, the card, and whatever. Note that this may be difficult because this will have the beam power of below 1 mW.
Quote: | Alberto> Today I've tried to bring the frequency of the NPRO laser ... |
|
630
|
Thu Jul 3 13:12:32 2008 |
Rob, Yoichi, John | Update | Locking | More oscillations |
Bounce/ roll filters were added to the short degrees of freedom to reduce the effect of the 24Hz line seen on Tuesday night.
However last night saw the arrival of a new oscillation at ~34Hz. This may be the second harmonic of the MOS roll mode. Reducing the arm offset would cause this oscillation to ring up and break the lock (first plot). This effect was repeatable.
No signal was seen in the oplevs or osems which leads us to rule out alignment problems, at least for now.
Although one can clearly see DARM_ERR increasing as arm power increases adding a resonant gain in the DARM loop had no effect.
We also noticed that x arm transmission was significantly more noisy than the Y (second plot). And showed greater coherence with the increase in DARM noise. Investigations showed that the PD was not the source of the difference.
Turning up the MC gain seems to help a little.
We're now looking at POX as a candidate for RF_CARM (third plot). |
Attachment 1: LOL.png
|
|
Attachment 2: NoisyX080702.png
|
|
Attachment 3: POXforCARM080702.png
|
|
634
|
Thu Jul 3 18:48:09 2008 |
Alberto | Update | General | Beats of the two lasers in the absolute length measurement observed |
I adjusted the alignment of the flipper mirror as suggested by Koji making the two beam spots match. I also aligned all the IFO mirrors (ITMs, PRM, SRM, ETMs) to have more power for the IFO signal at the AS port. When I did that I could see the beats at the AS OSA. Then I explored the range of temperature of the NPRO from 35deg (C) to 51.2807deg and at that point I could observe a peak corresponding to the beat at about 10MHz on the network analyzer. The peak tends to drift because the laser takes probably a longer time to actually thermalize and it moves very rapidly changing the temperature of the laser. |
635
|
Thu Jul 3 22:54:45 2008 |
Koji | Update | General | Beats of the two lasers in the absolute length measurement observed |
Great! Conguraturation! I wish if I could see it! It's nice if you can put the photo or anything of the RF spectrum analyzer.
Next step:
o You can try to maximize the beat amplitude by the tuning of the Injection steering mirrors.
o At the south end of the SP table, I prepared a frequency mixer. You can put the beat signal into the RF input, and an oscillator (which you can bring from somewhere) to the LO input in order to obtain the error signal of the PLL. Put the IF output of the mixer in a SR560, and please try to lock it by a simple 6db/oct (1st order) LPF of the SR560. For the actuator you can use the fast-pzt input of the NPRO.
Quote: | Then I explored the range of temperature of the NPRO from 35deg (C) to 51.2807deg and at that point I could observe a peak corresponding to the beat at about 10MHz on the network analyzer. |
|
637
|
Mon Jul 7 11:22:02 2008 |
Alberto | Update | General | Beats of the two lasers in the absolute length measurement observed |
I didn't post a screenshot from the RF SA because I had troubles with the interface with the computer (unfortunately the network SA cannot export the data either).
There is problem with the PLL circuit. The signal, beside the beat, also contains peaks at 33, 66 and 99 MHz, so we should think about filtering those out.
Quote: | Great! Conguraturation! I wish if I could see it! It's nice if you can put the photo or anything of the RF spectrum analyzer.
Next step:
o You can try to maximize the beat amplitude by the tuning of the Injection steering mirrors.
o At the south end of the SP table, I prepared a frequency mixer. You can put the beat signal into the RF input, and an oscillator (which you can bring from somewhere) to the LO input in order to obtain the error signal of the PLL. Put the IF output of the mixer in a SR560, and please try to lock it by a simple 6db/oct (1st order) LPF of the SR560. For the actuator you can use the fast-pzt input of the NPRO.
Quote: | Then I explored the range of temperature of the NPRO from 35deg (C) to 51.2807deg and at that point I could observe a peak corresponding to the beat at about 10MHz on the network analyzer. |
|
|
638
|
Mon Jul 7 13:06:38 2008 |
Koji | Update | General | Beats of the two lasers in the absolute length measurement observed |
One may need an RF filter after the mixer. I expect the SR560 does work for this purpose.
If it does not, a passive LPF can be used.
Quote: | I didn't post a screenshot from the RF SA because I had troubles with the interface with the computer (unfortunately the network SA cannot export the data).
There's is problem with the PLL circuit. The signal, beside the beat, also contains peaks at 33, 66 and 99 MHz, so we should think about filtering those out, correct?
|
|
641
|
Mon Jul 7 14:02:05 2008 |
Yoichi | Update | Computers | SVN conversion progress |
So far /cvs/cds/caltech/medm, /cvs/cds/caltech/chans and /cvs/cds/caltech/scripts have been converted to svn working copies.
Now /cvs/cds/caltech/target is being converted. |
642
|
Mon Jul 7 16:30:08 2008 |
steve | Update | PSL | PSL-PEM 16 days trend |
This morning the laser head temp was up to 20.3C because the laser chiller was overflowing.
I removed 700 cc water.
The PSL-FSS_RMTEMP became much more stable during the holidays as the psl enclosure was closed for 4 days
The high particle counts can be explaned by construction activity today.
The PMC & MZ PZT high voltages were out of range this morning. |
Attachment 1: pempsl16d.jpg
|
|
643
|
Mon Jul 7 19:15:38 2008 |
Alberto | Update | General | Optics alignement on the ABS length experiment |
Today I started setting up the PLL instruments to lock the frequency of the NPRO beam to the IFO beam. with no need of a new alignment after the weekend I was able to see the beat again, although this time I found at a different temperature of the NPRO laser of about 54 degrees (vs 51 of the last time).
I've got the Marconi as local oscillator (LO), the mixer Koji suggetsed, the SR560 and a 5 MHz low pass filter to cut the 33, 66 and 99 MHz present in the output signal from the PD. The filter worked well and I was able to single out only the beat resonance from the power spectrum.
In the attempt to enhance the amplitude of the beat, as Koji suggested, I tried to work on the alignment of the steering mirrors. While I was doing that, for some reason the pre-modecleaner lost the alignment and I had to ask John to help me lock it again. during the process I lost the old alignment but at the end I got a new one, apparently (from the camera) even better than the other. Although after that the beats were gone. Actually after the lock-in of the PMC the IFO beam didn't look as good as before, so it might be also for that reason.
I'll try again tomorrow, after that probably tonight Rob is going to reset the alignments of the interferometers. |
645
|
Tue Jul 8 08:16:56 2008 |
steve | Update | PSL | PMC problem |
The PMC is unhappy. PMC auto locker is not working.
DC output slider adjust has to be moved from rail to rail before it locks.
MZ is working great. |
Attachment 1: pmc.jpg
|
|
646
|
Tue Jul 8 10:20:10 2008 |
steve | Update | PEM | HEPA turned on |
It is specially important to run the PSL-HEPA filters when inside counts is peaking at 30,000 counts
There is a small label at the hepa on/off switch:
enclosure open 100%,
low noise off,
normal 60% of Variac voltage setting on the top of the enclosure
Not running the HEPAs will lower the temp fluctuations from 1.5 to o.5 degree C
at the cost of particle counts from 0 to immidiate room counts. |
Attachment 1: hepa.jpg
|
|
647
|
Tue Jul 8 10:26:30 2008 |
Masha | Update | Auxiliary locking | setup updates |
Yesterday I changed one of the beam splitters in the Mach Zehnder to one with a more stable mount as to reduce the system's coupling to environmental noise.
With help from John, I worked out how to get the signals from the two channels of the interferometer into the digital system. I put up BNC cables along part of the Y arm to connect the output of the detectors into the digital channels. |
650
|
Tue Jul 8 21:58:22 2008 |
alberto | Update | General | Secondaty beam aligned to the IFO beam again |
Yesterday the alignment of the secondary beam to the IFO was completely lost and today I had to realign all the optics before I was able to match the two spots again. I had to reset the height of the irises and I had also to replace mirror M1 with one with a larger angular motion. Eventually I obtained the beat again. Working on the optics table I inadvertently misaligned the OSA but I didn't make in time to bring it back before the night shift people came. I'll work on that tomorrow morning. |
651
|
Wed Jul 9 12:42:14 2008 |
John | Update | Locking | Hand off to RF CARM |
Rob, Yoichi, John
Last night we were able to reduce the CARM offset to around 80. This was achieved by increasing the DARM gain and
switching to AS_I when AS_Q went bad. This is probably a temporary solution, we will probably switch to DC readout
for DARM as we bring the arms on resonance.
Having reduced the arm offset enough to get us into the linear region of the RF_CARM signal (POX_I) we worked on
analogue conditioning of this signal to allow us to hand over. Lock was maintained for over 20 minutes as we did
this work.
We were able to partially switch over both the frequency and length paths to this new signal before losing lock. |
Attachment 1: LongLock080709.png
|
|
655
|
Thu Jul 10 14:59:01 2008 |
rob | Update | Locking | RF common mode at zero offset |
rob, john, yoichi
Last night we succeeded in reducing the CARM offset to zero.
We handed off control of the common mode servo from PO-DC to POX-I.
We pushed the common mode servo bandwidth to ~19kHz. Without the boosts, it had ~80 degs of phase margin. Didn't measure it after engaging the boosts (Boost + 1 superboost). Trying to engage the second superboost stage broke the lock.
The process is fully scripted, and the script worked all the way through several times.
The DARM ugf was ~200Hz. The RSE peak could clearly be seen. No optical spring, as expected (we're locking in anti-spring mode).
Engaging test mass de-whitening filters did not work (broke the lock).
I'm attaching a lock control sequence diagram and a trend of the arm power during a scripted up-sequence. I think the script can be sped up significantly (especially the long ramp period).
Up next:
Calibrated DARM spectrum
Noise hunting (start with dewhites)
DC - Readout
Lock to the springy side. |
Attachment 1: lock_control_sequence_worked.png
|
|
Attachment 2: trendpowerbuild.png
|
|
656
|
Thu Jul 10 19:12:07 2008 |
Alberto | Update | General | abs cavity length measurement experiment |
Yesterday morning, when I started, I found the IFO beam on a different position and the beam spot at the AS port looked very deformed. The overlapping with the secondary beam was not good enough to observe the beats anymore. Restoring the alignments of the interferometers did not work because, as John found out later, some of the photodiodes had offsets and gain which made the restore script ineffective. After resetting the parameters, we had to align every mirror of the interferometers and save the configuration twice. The second times was because on the first time the alignment had been done with the illuminators on. To avoid that in the future, John wrote an alarm to warn about the status of the lights.
After that we fixed the IFO beam, I had to realign the optics in the table to match the secondary beam to the IFO beam. I got the two beam overlapping and, even though the NPRO spot looked distorted, I could observe again some signal of the beat. To do that it was also necessary to have all the interferometer mirrors aligned so that we had more power from the ifo beam although it also made the spot flash. Ideally, to avoid the flashing (which we would also impede the PLL to work) we should work with the interferometer locked. Since that doesn't seem actually possible, we should just keep one of the ITM aligned and improve the beam matching so that we can observe the beats even with less power.
Today I spent the day trying to improve the alignement of the optics to observe the beats with only the ITM aligned, resetting the alignment of both beams with the ireses, with the Farady and all the rest. It was a rather long and tiring process but I think I'm close to the target and maybe tomorrow. |
659
|
Fri Jul 11 09:29:02 2008 |
steve | Update | PSL | the PMC still hangs up |
Morning alarms:
PMC high voltage is railing, it's auto locker is frosen
I have to move DC Output Adjust slider by hand and it locks immediately
PSL_ISS is saturating. Saturation goes away when PMC is locked.
Laser chiller water is overflowing again: removed 450 cc water |
Attachment 1: pmc4d.jpg
|
|
661
|
Fri Jul 11 23:55:25 2008 |
alan | Update | Locking | RF common mode at zero offset |
Quote: | rob, john, yoichi
Last night we succeeded in reducing the CARM offset to zero.
|
Congratulations! Well done! I look forward to hearing the details and further progress! |
662
|
Sat Jul 12 23:28:31 2008 |
rana | Update | PSL | PMC needs help |
As everyone has noticed recently, the PMC seems to have a PZT problem. It often zooms
of to one of its rails after locking as if the PZT range has decreased dramatically.
WE should check this on Monday by disabling the FSS and applying a slow triangle wave to
the NPRO frequency. The PMC will track this frequency change and this will allow us to
diagnose its problems.
If it has real problems, I have a spare in W. Bridge which we can swap in temporarily.
The attached plot shows 3 years of trend; looks like it went bad in summer of last year.
This also seems to be the cause of our ISS saturation problems: lowering the gain of the PMC
even slightly increased the intensity noise enough to cause saturation. Increasing the gain
even slightly increased the intensity noise enough to cause saturation due to PMC servo oscillation. |
Attachment 1: Untitled.png
|
|
668
|
Mon Jul 14 19:15:43 2008 |
Alberto | Update | General | abs cavity length measurement experiment |
Lately I've been dealing with the alignment of the interferometer to have a good beam spot at the AS port. Today the alignment script kept failing because of computer problems (failure of the frame builder) and also because the IFO was probably too far from the range where the automatic alignment works.
An other problem I keep having with the alignment of the optics on the AP table is with multiple reflection beams of the NPRO beam at the Farady.
Although I believe that now the two beams are quite well aligned, I don't see any reflection of the secondary beam from the IFO anymore.
It's like the more I try to improve the alignment, the worse I get from the beam matching. I'll keep working on this. |
669
|
Mon Jul 14 21:34:10 2008 |
Alberto | Update | General | abs cavity length measurement experiment |
Quote: | Lately I've been dealing with the alignment of the interferometer to have a good beam spot at the AS port. Today the alignment script kept failing because of computer problems (failure of the frame builder) and also because the IFO was probably too far from the range where the automatic alignment works.
An other problem I keep having with the alignment of the optics on the AP table is with multiple reflection beams of the NPRO beam at the Farady.
Although I believe that now the two beams are quite well aligned, I don't see any reflection of the secondary beam from the IFO anymore.
It's like the more I try to improve the alignment, the worse I get from the beam matching. I'll keep working on this. |
Realigning the OSA I also had to move a little bit the mirror that reflects the IFO beam of at the AS port in order to raise the beam height. This had the effect of changing the position of the AS spot on the camera and on the monitors.
Tonight with John, we made sure that the AS beam was still aligned to the PD. |
670
|
Tue Jul 15 09:47:09 2008 |
steve | Update | PEM | lab temps and particles |
All air condtion units were serviced last Friday.
AC filters are trying to control our particle counts but they have no capacity to match bad Pasadena conditions.
IFO room filters at CES were really clean.
Air make up filters inside and outside were dirty.
They showed the construction effect.
Control room and clean assembly units needed all filters replaced.
Note: the PSL-FSS_RCTEMP droped o.1C when enclouser HEPAs were turned back on
The RC temp controller should be better than that! |
Attachment 1: temps24d.jpg
|
|
672
|
Tue Jul 15 10:24:57 2008 |
steve | Update | PSL | PMC temp & pzt voltage |
The PMC pzt HV was happy with no HEPA temp stability.
Can we thermally insulate the pmc ? |
Attachment 1: pmctemp.jpg
|
|
674
|
Tue Jul 15 12:23:22 2008 |
Jenne | Update | General | MC2 Watchdog tripped |
Alberto, Jenne
Mode Cleaner was unlocked. We checked, and found that MC2's watchdog was tripped. It didn't look like anything bad was going on, so we turned the optic back on, and tried to relock the MC. It looks like the Mode Cleaner is now locked, but the lock bit on the LockMC screen is still red. I don't know what's up. |
677
|
Wed Jul 16 09:27:17 2008 |
steve | Update | ALARM | PSL-FSS_RMTEMP alarm is false |
Morning alarm sound is good for people who does not drink coffee.
Our 40m alarm is on every morning.
Those whom are not here in the morning thinks that this beeping sound is inspirational.
Would someone change this sound into less punishing form, like mockingbird chirp....
The C1PSL_SETTINGS.adl (40mm PSL Settings ) indicating that
C1:PSL-FSS_INOFFSET (Input Offset Adjust ) should be 0.3 +-0.05 V (red warning tag )
Alarm Handler: 40M pointing to yellow grade warning of PSL-FSS_RMTEM
This is a false alarm.
Two years trend of these channels are here: |
Attachment 1: frmtemp2y.jpg
|
|
679
|
Wed Jul 16 11:00:15 2008 |
Masha | Update | Auxiliary locking | improving ADC input for the mach zehnder setup and completely unrelated happenings |
For most of last week, the SURFs + Jenne were helping Mike and Ken with "stray light control for
Enhanced LIGO", i.e. cleaning and baking many many baffles which will catch scattered light in the
interferometer.
Otherwise, the two channels of the Mach Zehnder which will be used to measure fibre noise were
balanced, which should reduce the effect of laser amplitude noise in phase noise detection. I have
set up two digital channels to collect time series data from the two photodiodes and took some
preliminary noise measurements. I will be using Matlab to combine the signals as to directly measure
the phase noise, and I wrote some Matlab code to speed up this process: loading the files,
manipulating time series data, and converting into frequency domain. Currently I am building a
filter that will attenuate the signal at frequencies below 1Hz and amplify at higher frequencies in
order to whiten the spectra and reduce ADC noise. |
680
|
Wed Jul 16 11:26:47 2008 |
Max Jones | Update | | This Week |
Baffles.
I got a battery for the magnetometer today which is slightly too large (~2 mm) in one dimension. Not sure what I'm going to do.
I'm attempting to calibrate the magnetometer but I'm having a hard time calibrating the axis that I cannot simply put through a coil parallel to the coils length. I have attempted to use the end fields of the solenoid but the measurements from the magnetometer are significantly different from the theoretical calculations.
I would appreciate suggestions. - Max. |
683
|
Wed Jul 16 16:59:07 2008 |
Alberto | Update | General | Aligment |
I think the two beams are aligned again - they both pass the Faraday, they match at the irises and all along the optical path on the AP table. Although the NPRO beam does not show up at the AS port. |
685
|
Wed Jul 16 17:51:58 2008 |
Masha | Update | Auxiliary locking | long measurement |
I'm taking a measurement on the SR785 spectrum analyzer at low frequencies, so I'm going to leave it by the symmetric port table for a while. Please don't move it! |
686
|
Wed Jul 16 22:29:05 2008 |
Masha | Update | Auxiliary locking | long measurement |
Quote: | I'm taking a measurement on the SR785 spectrum analyzer at low frequencies, so I'm going to leave it by the symmetric port table for a while. Please don't move it! |
all done thanks. |
688
|
Thu Jul 17 08:30:15 2008 |
steve | Update | PSL | PMC relocked manually |
The PMC pzt HV and the servo gain adj. are railing at max this morning
Why is it on the decreasing side of FSS_RTTEMP slope? |
Attachment 1: pmc4d.jpg
|
|
689
|
Thu Jul 17 12:15:21 2008 |
Eric | Update | PSL | Swept PMC PZT voltage range |
I unlocked the PMC and swept over C1:PSL-PMC_RAMP's full range a couple of times this morning. The PMC should now be relocked and returned
to normal. |
691
|
Thu Jul 17 16:39:58 2008 |
Max Jones | Update | DAQ | Magnetometer Installed |
Today I installed the magnetometer near the beam splitter chamber. It is located on the BSC chamber at head height on the inner part of the interferometer (meaning I had to crawl under the arms to install it). I don't think I disturbed anything during installation but I think that's it's probably prudent to tell everyone that I was back there just in case. I plan to run 3 BNC cables (one for each axis) from the magnetometer to the DAQ input either tonight or tomorrow. Suggestions are appreciated. - Max. |
692
|
Thu Jul 17 20:13:34 2008 |
Yoichi | Update | PSL | PMC alignment/mode matching effort |
I'm working to improve the mode matching of PMC.
Because I noticed that the beam was hitting the aperture of the EOM for PMC, I moved the EOM a little bit to maximize the transmission.
This did not change the alignment to the reference cavity but changed the alignment of the PMC a lot.
So I adjusted it back.
The alignment of the PMC can be easily optimized but the Hermite 02 mode still remains. This means the mode matching is bad.
Moving the lenses by a small amount (a few mm) did not change the height of 02 mode.
I'm planning to move the lenses by a large amount tomorrow. But it will destroy the alignment to the PMC.
So I installed two irises in the beam path after the lenses to remember the alignment roughly.
Right now the PMC transmission is slightly worse than before because the lens positions are not good. |
694
|
Fri Jul 18 16:57:37 2008 |
Jenne | Update | IOO | Calibrated MC_F |
I have calibrated MC_F. The conversion factor is 137.49 MHz/count.
The calibration data taken is attached, along with a calibrated power spectrum.
On the data plot, the x axis is volts from the C1:IOO-MC_FAST_MON channel, with the calibration between FAST_MON and MC_F = -788.18 volts/count.
The linear term of the fit line = -0.085MHz/volt. Error bars are +/- 1 in the last digit of what the spectrum analyzer gave me for frequency (+/- 0.01MHz).
The net conversion factor is then (-788.18)*(-0.085)*(2) = 137.49 MHz/count. The factor of 2 is because the light passes through the AOM twice.
On the power spectrum,
REF0 and REF1 = MC unlocked, HEPAs on, MC Refl gain = 22
REF2 and REF3 = MC locked, HEPAs on, MC Refl gain = 22
REF4 and REF5 = MC locked, HEPAs on, MC Refl gain = 19
REF6 and REF7 = MC locked, HEPAs off, MC Refl gain = 19 |
Attachment 1: MC_Fcalib.png
|
|
Attachment 2: 20080717MC_F-MC_I.pdf
|
|
695
|
Fri Jul 18 17:06:20 2008 |
Jenne | Update | Computers | Computers down for most of the day, but back up now |
[Sharon, Alex, Rob, Alberto, Jenne]
Sharon and I have been having trouble with the C1ASS computer the past couple of days. She has been corresponding with Alex, who has been rebooting the computers for us. At some point this afternoon, as a result of this work, or other stuff (I'm not totally sure which) about half of the computers' status lights on the MEDM screen were red. Alberto and Sharon spoke to Alex, who then fixed all of them except C1ASC. Alberto and I couldn't telnet into C1ASC to follow the restart procedures on the Wiki, so Rob helped us hook up a monitor and keyboard to the computer and restart it the old fashioned way.
It seems like C1ASC has some confusion as to what its IP address is, or some other computer is now using C1ASC's IP address.
As of now, all the computers are back up. |
696
|
Fri Jul 18 17:12:35 2008 |
Jenne | Update | IOO | Checking out the MC Servo Board |
[Jenne, Max]
One of the things that Rana thinks that might be causing my MC_F calibration to be off is that the MC Servo Board's filters don't match those on the schematics. Max and I pulled the MC servo board today to check resistor and capacitor values. Alberto needed the Mode Cleaner, so we put the board back before finishing checking values. We will probably pull the board again next week to finish checking the values.
I haven't checked to ensure that the MC still locks, because Yoichi is doing stuff on the PSL table, but I didn't change anything on the board, and hooked all the cables back where they were, so hopefully it's all okay. |
697
|
Fri Jul 18 19:15:15 2008 |
Jenne | Update | IOO | Checking out the MC Servo Board |
Quote: | [Jenne, Max]
I haven't checked to ensure that the MC still locks, because Yoichi is doing stuff on the PSL table, but I didn't change anything on the board, and hooked all the cables back where they were, so hopefully it's all okay. |
I put the PMC back and the MC now locks. |
698
|
Fri Jul 18 19:30:20 2008 |
Masha | Update | Auxiliary locking | moving from 40m |
I will be working in the basement of Bridge probably starting next week; I moved the NPRO laser and some of the optics from my mach zehnder setup on the SP table to Bridge. Thanks for your help! |
699
|
Fri Jul 18 19:41:09 2008 |
Yoichi | Update | PSL | PMC PZT investigation |
I measured the HV coming to the PMC PZT by plugging it off from the PZT and hooking it up to a DVM.
The reading of DVM is pretty much consistent with the reading on EPICS. I got 287V on the DVM when the EPICS says 290V.
Then I used a T to monitor the same voltage while it is connected to the PZT. I attached a plot of the actual voltage measured by the DVM vs the EPICS reading.
It shows a hysteresis.
Also the actual voltage drops by more than a half when the PZT is connected. The output impedance of the HV amp is 64k (according to the schematic). If I believe this number, the impedance of the PZT should also be 64k. The current flowing the PZT is 1.6mA at 200V EPICS reading.
The impedance of the PZT directly measured by the DVM is 1.5M ohm, which is significantly different from the value expected above. I will check the actual output impedance of the HV amp later.
The capacitance of the PZT measured by the DVM is 300nF. I don't know if I can believe the DVM's ability to measure C.
I noticed that when a high voltage is applied, the actual voltage across the PZT shows a decay.
The second plot shows the step response of the actual voltage.
The voltage coming to the PZT was T-ed and reduced by a factor of 30 using a high impedance voltage divider to be recorded by an ADC.
The PMCTRANSPD channel is temporarily used to monitor this signal.
After the voltage applied to the PZT was increased abruptly (to ~230V), the actual voltage starts to exponentially decrease.
When the HV was reduced to ~30V, the actual voltage goes up. This behavior explains the weird exponential motion of the PZT feedback signal when the PMC is locked.
The cause of the actual voltage drop is not understood yet.
From the above measurements, we can almost certainly conclude that the problem of the PMC is in the PZT, not in the HV amp nor the read back. |
Attachment 1: Hysteresis.png
|
|
Attachment 2: StepResponse.png
|
|
701
|
Fri Jul 18 23:24:24 2008 |
rob | Update | PSL | PMC PZT investigation |
Quote: | I measured the HV coming to the PMC PZT by plugging it off from the PZT and hooking it up to a DVM.
The reading of DVM is pretty much consistent with the reading on EPICS. I got 287V on the DVM when the EPICS says 290V.
Then I used a T to monitor the same voltage while it is connected to the PZT. I attached a plot of the actual voltage measured by the DVM vs the EPICS reading.
It shows a hysteresis.
Also the actual voltage drops by more than a half when the PZT is connected. The output impedance of the HV amp is 64k (according to the schematic). If I believe this number, the impedance of the PZT should also be 64k. The current flowing the PZT is 1.6mA at 200V EPICS reading.
The impedance of the PZT directly measured by the DVM is 1.5M ohm, which is significantly different from the value expected above. I will check the actual output impedance of the HV amp later.
The capacitance of the PZT measured by the DVM is 300nF. I don't know if I can believe the DVM's ability to measure C.
I noticed that when a high voltage is applied, the actual voltage across the PZT shows a decay.
The second plot shows the step response of the actual voltage.
The voltage coming to the PZT was T-ed and reduced by a factor of 30 using a high impedance voltage divider to be recorded by an ADC.
The PMCTRANSPD channel is temporarily used to monitor this signal.
After the voltage applied to the PZT was increased abruptly (to ~230V), the actual voltage starts to exponentially decrease.
When the HV was reduced to ~30V, the actual voltage goes up. This behavior explains the weird exponential motion of the PZT feedback signal when the PMC is locked.
The cause of the actual voltage drop is not understood yet.
From the above measurements, we can almost certainly conclude that the problem of the PMC is in the PZT, not in the HV amp nor the read back. |
I'd believe the Fluke's measurement of capacitance. Here's some info from PK about the PZT:
Quote: |
But the PMC ones were something like
0.750 in. thick x 0.287 in. thick. 2 microns per 200 V displacement,
resonant frequency greater than 65 kHz. Typical capacitance is around 0.66
uF.
|
If the PZT capacitance has dropped by a factor of two, that seems like a bad sign. I don't know what to expect for a resistance value of the PZT, but I wouldn't be surprised if it's non-Ohmic. The 64k is the series resistor after the PA85, not the modeled resistance of the PZT itself. |
702
|
Sat Jul 19 19:39:44 2008 |
rob | Update | PSL | PMC PZT investigation |
Quote: |
Quote: | The 64k is the series resistor after the PA85, not the modeled resistance of the PZT itself. |
Yes. What I meant was that because the measured voltage across the PZT was a half of the open voltage of the HV amp, the DC impedance of the PZT is expected to be similar to the output impedance of the HV amp. Of course, I don't think the DC impedance of a normal PZT should be such low.
I'm puzzled by the discrepancy between this expected DC impedance and the directly measured impedance by the Fluke DVM (1.5M Ohm).
One possibility is that the PZT leaks current only when a high voltage is applied. |
704
|
Mon Jul 21 09:52:05 2008 |
Sharon | Update | | Adaptive code changes | Thanks to Alex, we now save the coefficients of the adaptive filter every cycle. When we choose ENABLE: OFF on the MEDM screen, suppressing the signal to the MC1, we stop and save the last coefficients. When enabling it again, it starts from the last coefficients saved. I will take some measurements today to check how this contributes to the adaptation rate. If you change the number of taps or the number of AUX channels, the coefficients are again set to zero. |
705
|
Mon Jul 21 10:00:42 2008 |
steve | Update | PEM | ifo room temp | Five years of temp trend with 5 C fluctuation. The campus chilled water supply was upgraded ~ 1 year ago.
I requested our thermostats to be calibrated. |
Attachment 1: temp5y.jpg
|
|
706
|
Mon Jul 21 11:54:00 2008 |
Jenne | Update | General | MC Servo Board | I pulled the MC Servo Board again, to check the components that are on the board, and compare them with the schematics. The filters that I'm interested in on the Fast Path haven't been changed. The high pass filters on the Fast Path have been changed.
Component Schematic Actual
--------- --------- ------
C140 10u open
C144 10u open
C149 open a gray Cap. value unknown
C141 10u open
C145 10u open
R97 1.58K 0
R99 open 1130
R103 open 1130
R100 open 0
R104 100 1130
R98 1.58K open
R109 367 365
Board is back in, and MC locks. |
709
|
Mon Jul 21 19:48:57 2008 |
Sharon | Update | | how tp restart C1ASS | How to restart C1ASS:
1. reboot
2. as root: caltech/target/c1ass:> ./startass
3. no need for root: burtgooey
that's it... |
711
|
Tue Jul 22 03:03:22 2008 |
John, Rob | Update | PSL | FSS open loop transfer function | With the common gain slider maxed out the unity gain frequency is 58kHz.
The reference cavity refl diode appears to be okay. RF OUT/ TEST IN transfer function was normal.
There is a ~220mV offset in the RF out. We removed this using a coupler - no change. We also checked the
diode->FSS cable.
Tomorrow I'll take a closer look at the board. |
712
|
Tue Jul 22 09:24:17 2008 |
steve | Update | PSL | laser power | Laser power reality of 120 days |
Attachment 1: power120d.jpg
|
|
713
|
Tue Jul 22 11:55:22 2008 |
rana | Update | PSL | Note from R. Abbott re: the PMC | an email from Rich:Your PZT is broken.
R |
|