ID |
Date |
Author |
Type |
Category |
Subject |
5313
|
Sat Aug 27 20:38:17 2011 |
Suresh | Update | IOO | Light is back on WFS | [Valera, Suresh]
We wanted to continue the work with WFS servo loops. As the current optical paths on the AP table do not send any light to the WFS, I changed a mirror to a 98% window and a window to a mirror to send about 0.25mW of light towards the WFS. The MC locking is unaffected by this change. The autolocker works fine.
When the power to the MC is increased, these will have to be replaced or else the WFS will burn. |
5312
|
Sat Aug 27 15:47:59 2011 |
rana | Update | CDS | OSEM noise / nullstream and what does it mean for satellites | In the previous elog of mine, I looked at the nullstream (aka butterfly mode) to find out if the intrinsic OSEM noise is limiting the displacement noise of the interferometer or possibly the Wiener FF performance.
The conclusion was that its not above ~0.2 Hz. Due to the fortuitous breaking of the ITMX magnet, we also have a chance to check the 'bright noise': what the noise is with no magnet to occlude the LED beam.
As expected, the noise spectra with no magnets is less than the calculated nullstream. The attached plot shows the comparison of the LL OSEM (all the bright spectra look basically alike) with the damped
optic spectra from 1 month week ago.
From 0.1 - 10 Hz, the motion is cleanly larger than the noise. Below ~0.2 Hz, its possible that the common mode rejection of the short cavity lengths are ruined by this. We should try to see if the low frequency
noise in the PRC/SRC is explainable with our current knowledge of seismicity and the 2-dimensional 2-poiint correllation functions of the ground.
So, the question is, "Should we try to upgrade the satellite boxes to improve the OSEM sensing noise?" |
Attachment 1: Untitled.png
|
|
5311
|
Sat Aug 27 14:33:04 2011 |
Jenne | Update | SUS | ITMX magnet status | As I feared, since I couldn't see the magnet-to-dumbbell joint from all angles, they ended up being off by ~1/3 of a magnet diameter.
Because I don't want to deal with finding another failed glue joint tomorrow, I removed the magnet and dumbbell from the optic, and broke the manget off of the dumbbell. As with yesterday, I kept track of which end of the magnet had been glued to the dumbbell.
I got a new dumbbell, removed all the glue from the magnet, and reglued them together, in the fixture that ensures they are well aligned.
Tomorrow I will come in and glue the magnet dumbbell assembly to the ITM. |
5310
|
Fri Aug 26 16:17:31 2011 |
Jenne | Update | General | drawer cabinet moves in |
Quote: |
Jamie and Shuresh moved in Jenne's 11 drawers cabinet and relocated old note book boxes on the inside of the vac tube.
|
Barring other chores for next Wednesday, we're going to spend Wednesday afternoon populating the new cabinet with all of the optics hardware: posts, forks, dogs, everything! It's going to be so organized and awesome!! |
5309
|
Fri Aug 26 15:59:33 2011 |
steve | Update | General | drawer cabinet moves in | manufacturer info
Jamie and Shuresh moved in Jenne's 11 drawers cabinet and relocated old note book boxes on the inside of the vac tube. |
Attachment 1: P1080192.JPG
|
|
5308
|
Fri Aug 26 15:30:36 2011 |
Jenne | Update | SUS | ITMX magnet reglued | The ITMX UL magnet has been reglued.
I *very carefully* using the corner of a cleaned razor blade dropped single drops of acetone onto the top of the dumbbell, and scratched off the residual glue. I didn't want to get even a sprinkle of acetone on the dumbbell-glass junction, and I managed to avoid it. Also, the dumbbell never broke off of the glass (something I've never been able to achieve before), so all I had to do was glue the magnet back onto the dumbbell.
I also scratched the glue from the magnet, after soaking in acetone. I made sure to keep track of which way the magnet had been glued by putting it in the pickle picker that I received from Betsy before getting rid of the glue. I specifically did not compare the polarity of this magnet to the others still glued, because I have seen that in the past break magnets from dumbbells. They can't really handle sideways forces. But since it's glued the same way that it was, it should be fine.
I then aligned the optic in the gluing fixture. I test-fit the pickle picker with magnet, to ensure that the axes of the dumbbell and magnet were aligned as closely as possible. I adjusted the optic to make this axial alignment as perfect as I could see with my eye. Unfortunately the fixture doesn't allow a whole lot of viewing angles of the magnet-dumbbell joint, so we'll see how well I did after I remove it from the fixture.
I put a little dab of epoxy on the end of the magnet, spread it around so it coated the whole surface, and glued it on.
I'll come in tomorrow (Saturday) to check on it, and take it out of the fixture. If it's going to break coming out of the fixture, which I hope won't happen, but has happened before, then I want to be able to fix it again asap. |
5307
|
Fri Aug 26 10:45:17 2011 |
steve | Update | General | crane at ETMY is fixed |
Quote: |
The horizontal trolley drive stopped working at the east end this morning. It is working intermittently. In the worst case we can take the door off with the manual -Genie- lift.
I'm working with Konecrane to solve the wormgear drive problem.
|
New gear box installed and tested by Fred KoneCranes. |
Attachment 1: P1080190.JPG
|
|
Attachment 2: P1080188.JPG
|
|
5306
|
Fri Aug 26 07:53:59 2011 |
steve | Update | SUS | Broken UL magnet on ITMX |
Quote: |
Dmass just reminded me that the usual procedure is to bake the optics after the last gluing, before putting them into the chambers. Does anyone have opinions on this?
On the one hand, it's probably safer to do a vacuum bake, just to be sure. On the other hand, even if we could use one of the ovens immediately, it's a 48 hour bake, plus cool down time. But they're working on aLIGO cables, and might not have an oven for us for a while. Thoughts?
|
Follow full procedure for full strength, minimum risk |
5305
|
Thu Aug 25 17:57:35 2011 |
Suresh | Update | SUS | Broken UL magnet on ITMX |
Quote: |
Dmass just reminded me that the usual procedure is to bake the optics after the last gluing, before putting them into the chambers. Does anyone have opinions on this?
On the one hand, it's probably safer to do a vacuum bake, just to be sure. On the other hand, even if we could use one of the ovens immediately, it's a 48 hour bake, plus cool down time. But they're working on aLIGO cables, and might not have an oven for us for a while. Thoughts?
|
I think we should follow the established procedure in full, even though it will cost us a few more days. I dont think we should consider the vacuum bake as something "optional". If the glue has any volatile components they could be deposited on the optic resulting in a change in the coating and consequently optical loss in the arm cavity.
|
5304
|
Thu Aug 25 17:40:07 2011 |
Dmass | Update | Computer Scripts / Programs | elog broke, fixed | elog died b/c someone somewhere did something which may or may not have been innocuous. I ran the script in /cvs/cds/caltech/elog to restart the elog (thrice).
I have now banned Warren from clicking on the elog from home |
5303
|
Thu Aug 25 17:14:49 2011 |
steve | Update | VAC | clean room fashion changes | Jamie is modeling our next generation in-vac & clean room bonny suit that Jenne and myself already tested.
It is quite bearable with our traditional cleanroom beret-bouffant cap. Please use these in the future.
This will help to avoid the farther degradation of somewhat dusty 40m vac envelope.
It is the required dress code to enter the clean assembly room in the 40m.
We have small, med, large and x-large in stock. I'm getting larger sizes.
It will not allow certain people to climb inside the vacuum chamber in dirty pants. |
Attachment 1: P1080185.JPG
|
|
Attachment 2: P1080183.JPG
|
|
5302
|
Thu Aug 25 15:20:03 2011 |
Jenne | Update | SUS | Broken UL magnet on ITMX | Dmass just reminded me that the usual procedure is to bake the optics after the last gluing, before putting them into the chambers. Does anyone have opinions on this?
On the one hand, it's probably safer to do a vacuum bake, just to be sure. On the other hand, even if we could use one of the ovens immediately, it's a 48 hour bake, plus cool down time. But they're working on aLIGO cables, and might not have an oven for us for a while. Thoughts? |
5301
|
Thu Aug 25 13:10:42 2011 |
Jenne | Update | SUS | Drag wiping | As we have seen in the past, both of the ITMs were more dusty than the ETMs, presumably because we have the vertex open much more often than the ends. Kiwamu and I wiped all of the optics until we could no longer see any dust particles within a ~1.5 inch diameter area around the center.
Since we have ITMX out for magnet gluing, I'll probably drag wipe both front and back surfaces before putting it back in the suspension cage. All of the optics have clear dust on the AR surfaces, but we can't get to that surface while the optics are suspended. For the ETMs this isn't too big of a deal, but it does concern me a bit for the ITMs and other transmissive optics we have. I don't think it's bad enough yet though to warrant removing optics from suspensions just to wipe them. |
5300
|
Thu Aug 25 08:12:09 2011 |
steve | Update | SUS | ETMY & ITMY wiped, replaced, vac door on |
Quote: |
We've closed up ETMX:
- the optic was drag wiped
- the suspension tower was put back in place
- earthquake stops were backed off the appropriate number of turns, and de-ionized
- chamber door was put on
|
jamie, jenne, kiwamu, suresh, steve
ETMY and ITMY were treated the same way as ETMX. The BS chamber was closed with heavy vac door yesterday also. The IOO access connector's inner jamnuts are torqued to 45 ft/lbs as all vac door bolts.
The vac envelope is ready for pumpdown condition, except ITMX chamber with light atm door cover.
Jenne will summeries the condition of dust on the TMs before and after the drag wipes. |
5299
|
Wed Aug 24 17:05:11 2011 |
Jenne | Update | SUS | Broken UL magnet on ITMX |
Quote: |
The ITMX tower was shipped into the Bob's clean room to put the magnet back on.
|
Repair work is delayed. I need the "pickle pickers" that hold the magnet+dumbbell in the gluing fixture, for gluing them to the optic. Here at the 40m we have a full set of SOS gluing supplies, except for pickle pickers. We had borrowed Betsy's from Hanford for about a year, but a few months ago I returned all of the supplies we had borrowed. Betsy said she would find them in her lab, and overnight them to us. Since the problem occurred so late in the day, they won't get shipped until tomorrow (Thursday), and won't arrive until Friday.
I also can't find our magnet-to-dumbbell gluing fixture, so I asked her to send us her one of those, as well.
I have 2 options for fixing ITMX. I'll write down the pros and cons for each, and we can make a decision over the next ~36 hours.
OPTIONS:
(#1) Remove dumbbell from optic. Reglue magnet to dumbbell. Reglue magnet+dumbbell to optic.
(#2) Carefully clean dumbbell and magnet, without breaking dumbbell off of optic. Glue magnet to dumbbell.
PROS:
(#1) Guarantee that magnet and dumbbell are axially aligned.
(#2) Takes only 1 day of glue curing time.
CONS:
(#1) Takes 2 days of glue curing time. (one for magnet to dumbbell, one for set to optic.)
(#2) Could have slight mismatch in axis of dumbbell and magnet. Could accidentally drop a bit of acetone onto dumbbell-to-optic glue, which forces us into option 1, since this might destroy the integrity of the glue joint (this would take only the 2 days already required for option 1, it wouldn't force us to take 2+1=3 days). |
5298
|
Wed Aug 24 16:13:36 2011 |
kiwamu | Update | SUS | broke UL magnet on ITMX | I broke the UL magnet on ITMX
The ITMX tower was shipped into the Bob's clean room to put the magnet back on.
Since we found that all the magnets were relatively high (#5296) in the shadow sensors, we decided to slide the OSEM holder bar upward.
During the work, I haven't made the OSEMs far enough from the magnets.
So the magnets and OSEMs touched as I moved the holder.
Then the UL magnets were broken off and fell into the UL coil.
|
5297
|
Wed Aug 24 12:08:56 2011 |
jamie | Update | SUS | ITMX, ETMX, ETMY free swinging | ITMX: 998245556
ETMX, ETMY: 998248032 |
5296
|
Wed Aug 24 11:40:21 2011 |
jamie, jenne, kiwamu, suresh, steve | Update | SUS | problem with ITMX | ITMX was drag wiped, and the suspension was put back into place. However, after removing all of the earthquake stops we found that the suspension was hanging in a very strange way.
The optic appears to heavily pitched forward in the suspension. All of the rear face magnets are high in their OSEMs, while the SIDE OSEM appears fine. When first inspected, some of the magnets appeared to be stuck to their top OSEM plates, which was definitely causing it to pitch forward severely. After gently touching the top of the optic I could get the magnets to sit in a more reasonable position in the OSEMs. However, they still seem to be sitting a little high. All of the PDMon values are also too low:
|
nominal |
now |
UL |
1.045 |
0.767 |
UR |
0.855 |
0.718 |
LR |
0.745
|
0.420
|
LL |
0.780
|
0.415 |
SD |
0.840
|
0.752 |
Taking a free swing measurement now. |
5295
|
Wed Aug 24 11:30:27 2011 |
jamie, jenne, kiwamu, suresh, steve | Update | SUS | ETMX wiped, replaced, door on | We've closed up ETMX:
- the optic was drag wiped
- the suspension tower was put back in place
- earthquake stops were backed off the appropriate number of turns, and de-ionized
- chamber door was put on
|
5294
|
Wed Aug 24 09:11:19 2011 |
jamie | Update | SUS | ETMY SUS update: looks good. WE'RE READY TO CLOSE | We ran one more free swing test on ETMY last night, after the last bit of tweaking on the SIDE OSEM. It now looks pretty good:
ETMY |
 |
pit yaw pos side butt
UL -0.323 1.274 1.459 -0.019 0.932
UR 1.013 -0.726 1.410 -0.050 -1.099
LR -0.664 -1.353 0.541 -0.036 0.750
LL -2.000 0.647 0.590 -0.004 -1.219
SD 0.021 -0.035 1.174 1.000 0.137 |
4.23371 |
So I declare: WE'RE NOW READY TO CLOSE UP. |
5293
|
Tue Aug 23 18:25:56 2011 |
jamie | Update | SUS | SRM diagnalization OK | By looking at a longer data stretch for the SRM (6 hours instead of just one), we were able to get enough extra resolution to make fits to the very close POS and SIDE peaks. This allowed us to do the matrix inversion. The result is that SRM looks pretty good, and agrees with what was measured previously:
SRM |
 |
pit yaw pos side butt
UL 0.869 0.975 1.140 -0.253 1.085
UR 1.028 -1.025 1.083 -0.128 -1.063
LR -0.972 -0.993 0.860 -0.080 0.834
LL -1.131 1.007 0.917 -0.205 -1.018
SD 0.106 0.064 3.188 1.000 -0.011 |
4.24889 |
|
5292
|
Tue Aug 23 17:51:37 2011 |
Keiko | Update | LSC | Tolerance of PRC, SRC, MICH length = 2 mm ? | Keiko, Kiwamu
We noticed that we have used wrong code for MICH degree of freedom for both of the ELOG entries on this topic (cavity lengths tolerance search). It will be modified and posted soon. |
5291
|
Tue Aug 23 17:45:22 2011 |
jamie | Update | SUS | ITMX, ITMY, ETMX clamped and moved to edge of tables | In preparation for tomorrow's drag wiping and door closing, I have clamped ITMX, ITMY, and ETMX with their earthquake stops and moved the suspension cages to the door-edge of their respective tables. They will remain clamped through drag wiping.
ETMY was left free-swinging, so we will clamp and move it directly prior to drag wiping tomorrow morning. |
5290
|
Tue Aug 23 17:21:45 2011 |
kiwamu | Update | SUS | free swinging test for ETMY |
Excited ETMY
Tue Aug 23 17:20:45 PDT 2011
998180460
|
5289
|
Tue Aug 23 16:23:33 2011 |
Jenne | Update | VAC | Access connector in place | [Steve, Bob, Jamie, Kiwamu, Valera, Jenne]
The access connector is now in place, in preparation for pump-down. Tomorrow (hopefully) we will do all the other doors.
|
5288
|
Tue Aug 23 14:49:14 2011 |
jamie, jenne, kiwamu, suresh, keiko | Update | SUS | Adjustment of ETMY, issue with ITMY whitening | Before lunch we took a closer look at two of the suspensions that were most problematic: ITMY and ETMY. Over lunch we took new free swinging data. Results below:
- For ITMY we discovered that the whitening on the UL sensor was not switching. This was causing the UL sensor to have a different response, with a steeper roll of, which was causing all of the transfer function estimates to the other sensors to have large imaginary components. We took new free swing data with all of the whitening turned OFF. The result is a much improved matrix and diagnalization. The input matrix elements are mostly the same, but the coupling is basically gone. We'll fix the whitening after the pump down.
ITMY |
 |
pit yaw pos side butt
UL 0.157 1.311 1.213 -0.090 0.956
UR 1.749 -0.490 0.886 -0.038 -1.042
LR -0.251 -2.000 0.787 -0.007 1.066
LL -1.843 -0.199 1.114 -0.059 -0.936
SD -0.973 -0.205 1.428 1.000 0.239 |
4.34779 |
- ETMY has a very problematic SIDE OSEM. The magnet does not line up with the OSEM axis, and since there is no lateral adjustment in the side OSEMs, there's not much we can do about this. We're using aluminum foil to wedge the OSEM over as far as possible, but it's not quite enough. With the OSEM plates horizontal there is a lot of POS->SIDE coupling. With the OSEM plates vertical, the magnetic sits a little too close to the rear face, which can cause the magnet to get stuck to the LED plate. We're trying to decide where to leave it now, but the new diagnalization with the OSEM plates vertical is definitely better:
ETMX |
 |
pit yaw pos side butt
UL -0.138 1.224 1.463 -0.086 0.944
UR 0.867 -0.776 1.501 -0.072 -1.051
LR -0.995 -0.896 0.537 -0.045 0.754
LL -2.000 1.104 0.499 -0.059 -1.251
SD 0.011 0.220 1.917 1.000 0.224 |
4.42482 |
|
5287
|
Tue Aug 23 11:57:22 2011 |
kiwamu | Update | SUS | free swinging test during lunch time | excited all the optics. (with ITMY WTF OFF)
Tue Aug 23 11:52:52 PDT 2011
998160788 |
5286
|
Tue Aug 23 10:38:27 2011 |
jamie | Update | SUS | SUS update | SUS update before closing up:
- MC1, MC2, ITMX look good
- MC3, PRM look ok
- SRM pos and side peaks are too close together to distinguish, so the matrix is not diagnalizable. I think with more data it should be ok, though.
- all ITMY elements have imaginary components
- ITMY, ETMX, ETMY appear to have modest that swapped position:
- ITMY: pit/yaw
- ETMX: yaw/side
- ETMY: pos/side
- MC3, ETMX, ETMY have some very large/small elements
Not particularly good. We're going to work on ETMY at least, since that one is clearly bad.
OPTIC |
|
M |
cond(B) |
MC1 |
 |
pit yaw pos side butt
UL 0.733 1.198 1.168 0.050 1.057
UR 1.165 -0.802 0.896 0.015 -0.925
LR -0.835 -1.278 0.832 -0.002 0.954
LL -1.267 0.722 1.104 0.032 -1.064
SD 0.115 0.153 -0.436 1.000 -0.044 |
4.02107 |
MC2 |
 |
pit yaw pos side butt
UL 1.051 0.765 1.027 0.128 0.952
UR 0.641 -1.235 1.089 -0.089 -0.942
LR -1.359 -0.677 0.973 -0.097 1.011
LL -0.949 1.323 0.911 0.121 -1.096
SD -0.091 -0.147 -0.792 1.000 -0.066 |
4.02254 |
MC3 |
 |
pit yaw pos side butt
UL 1.589 0.353 1.148 0.170 1.099
UR 0.039 -1.647 1.145 0.207 -1.010
LR -1.961 -0.000 0.852 0.113 0.896
LL -0.411 2.000 0.855 0.076 -0.994
SD -0.418 0.396 -1.624 1.000 0.019 |
3.60876 |
PRM |
 |
pit yaw pos side butt
UL 0.532 1.424 1.808 -0.334 0.839
UR 1.355 -0.576 0.546 -0.052 -0.890
LR -0.645 -0.979 0.192 0.015 0.881
LL -1.468 1.021 1.454 -0.267 -1.391
SD 0.679 -0.546 -0.674 1.000 0.590 |
5.54281 |
BS |
 |
pit yaw pos side butt
UL 1.596 0.666 0.416 0.277 1.037
UR 0.201 -1.334 1.679 -0.047 -0.934
LR -1.799 -0.203 1.584 -0.077 0.952
LL -0.404 1.797 0.321 0.247 -1.077
SD 0.711 0.301 -3.397 1.000 0.034 |
5.46234 |
SRM |
NA |
NA |
NA |
ITMX |
 |
pit yaw pos side butt
UL 0.458 1.025 1.060 -0.065 0.753
UR 0.849 -0.975 1.152 -0.199 -0.978
LR -1.151 -1.245 0.940 -0.243 1.217
LL -1.542 0.755 0.848 -0.109 -1.052
SD -0.501 -0.719 2.278 1.000 -0.153 |
4.4212 |
ITMY |
 |
pit yaw pos side butt
UL 0.164 1.320 1.218 -0.086 0.963
UR 1.748 -0.497 0.889 -0.034 -1.043
LR -0.252 -2.000 0.782 -0.005 1.066
LL -1.836 -0.183 1.111 -0.058 -0.929
SD -0.961 -0.194 1.385 1.000 0.239 |
4.33051 |
ETMX |
 |
pit yaw pos side butt
UL 0.623 1.552 1.596 -0.033 1.027
UR 0.194 -0.448 1.841 0.491 -1.170
LR -1.806 -0.478 0.404 0.520 0.943
LL -1.377 1.522 0.159 -0.005 -0.860
SD 1.425 3.638 -0.762 1.000 -0.132 |
4.89418 |
ETMY |
 |
pit yaw pos side butt
UL 0.856 0.007 1.799 0.241 1.005
UR -0.082 -1.914 -0.201 -0.352 -1.128
LR -2.000 0.079 -0.104 -0.162 0.748
LL -1.063 2.000 1.896 0.432 -1.119
SD -0.491 -1.546 2.926 1.000 0.169 |
9.11516 |
|
5285
|
Tue Aug 23 09:40:37 2011 |
kiwamu | Update | General | Re: AS clipping fixed | Indeed it was suspenseful.
We tried finding where the clipping happened, but we couldn't find any obvious clippings.
So we checked centering of the beams on all the optics associated with the AS path, starting from BS, SR3,... to the AS optical bench.
And during the work some of them were recentered.
At the end we found no clipping. To make sure we tested the available range (no clipping range) by exciting the angular motion of BS with AWG (f ~ 1Hz, a ~ 1000).
The beam looked successfully coming out at the most of the angular oscillation point.
Quote from #5284 |
Where was the AS clipping?! Ah, the suspense...
|
|
5284
|
Tue Aug 23 06:49:24 2011 |
Anamaria | Update | General | more in-vac work : AS clipping fixed and OSEM/oplev adjustment | Where was the AS clipping?! Ah, the suspense...
Quote: |
+ fixed the AS clipping issue
Quote from #5275 |
We need to check/fix the AS beam clipping and once it's done we will readjust the OSEM mid range and the oplevs.
|
|
|
5283
|
Tue Aug 23 02:03:04 2011 |
Suresh | Update | IOO | MC realigned and spot positions recentered | After the MC1 and MC3 OSEMs were repositioned MC had to be realigned and the beam spots had to be recentered on the actuation nodes.
To do that I had to change the input beam direction into the MC and the coil offsets.
I also measured the resultant spot positions
spot positions in mm (MC1,2,3 pit MC1,2,3 yaw):
0.1354 -0.2522 -0.1383 -1.0893 0.7122 -1.5587

The MC1 and MC3 yaw can be improved further after the chambers are closed and evacuated. The PZT adjustments needed to realign the input beam pointing are quite small and should not pose a problem.
|
5282
|
Tue Aug 23 01:09:44 2011 |
kiwamu | Update | SUS | free swinging test | excited all the optics ---
Tue Aug 23 01:08:00 PDT 2011
998122096 |
5281
|
Tue Aug 23 01:05:40 2011 |
Jenne | Update | Treasure | All Hands on Deck, 9am! | We will begin drag wiping and putting on doors at 9am tomorrow (Tuesday).
We need to get started on time so that we can finish at least the 4 test masses before lunch (if possible).
We will have a ~2 hour break for LIGOX + Valera's talk.
I propose the following teams:
(Team 1: 2 people, one clean, one dirty) Open light doors, clamp EQ stops, move optic close to door. ETMX, ITMX, ITMY, ETMY
(Team 2: K&J) Drag wipe optic, and put back against rails. Follow Team 1 around.
(Team 3 = Team 1, redux: 2 people, one clean, one dirty) Put earthquake stops at correct 2mm distance. Follow Team 2 around.
(Team 4: 3 people, Steve + 2) Close doors. Follow Team 3 around.
Later, we'll do BS door and Access Connector. BS, SRM, PRM already have the EQ stops at proper distances.
|
5280
|
Tue Aug 23 00:55:13 2011 |
Jenne | Update | General | IFO ready for doors | [Kiwamu, Jenne]
After the IFO was aligned in air one final time, we tapped on a few OSEMs until we were happy with all of the centering of all of the optics' OSEMs. All are within 0.05 of their halfway values, with the exception of one each on MC1 and MC3, one of which is within 0.06, and the other 0.08. Because of the realignment pain of dealing with MC OSEMs, we elected to leave these alone. Also, since we obviously didn't open the MC2 tank, we don't know how they are, although the numbers look reasonable.
Also, we took photos (to be posted on Picasa in a day or two) of all the main IFO magnet-in-OSEM centering, as best we could. SRM, BS, PRM all caused trouble, due to their tight optical layouts. We got what we could. Various people have been looking at these for the past 2 weeks, and I think they're all fine, even if we didn't get stellar photos.
We are now prepared for pumping. For real this time.
|
5279
|
Mon Aug 22 21:32:10 2011 |
kiwamu | Update | General | more in-vac work : AS clipping fixed and OSEM/oplev adjustment | [Keiko / Jenne / Jamie / Kiwamu]
We did the following things today :
+ fixed the AS clipping issue
+ realigned all the oplevs
+ checked and adjusted the all OSEM DC values, including PRM, SRM, BS, ITMs, ETMs, MC1 and MC3
Since we touched the OSEMs the alignment has changed somewhat.
Right now Jenne, Suresh and I are working on the "confirmation alignment".
Once we find the alignment is still good (steerable by the PZTs and the DC coil bias), tomorrow we will do the drag&wipe and door closing.
Quote from #5275 |
We need to check/fix the AS beam clipping and once it's done we will readjust the OSEM mid range and the oplevs.
|
|
5278
|
Mon Aug 22 20:37:43 2011 |
Anamaria | Configuration | RF System | Plan for install of 3f PDs | I made a quick sketch of how to include two more RF PDs on the REFL beam, given the space we have on the table. We want to install REFL33 and REFL165, 3f signals for the the two modulation frequencies we are using. The point is to make the distance from first beam splitter the same to all PDs so that we can use only one lens before this BS to make the beam the right size. Currently there are 2 PDs on the refl beam, REFL11 and REFL55, predictably. So the drawing shows 4 PDs. Drawing is to scale but is a bit coarse. Hopefully we'll take pictures once we're done.
Reference from current BS splitting beam to the existing PDs. |
Attachment 1: 40m3fReflPdLayout.pdf
|
|
5277
|
Mon Aug 22 16:28:44 2011 |
Suresh | Update | General | SRM tower shifted on Friday | [Kiwamu, Suresh] This is a belated elog entry from last Friday night + Saturday morning!
We shifted the SRM tower across the beam and away from the door by about 5mm.
After the input beam from MC was aligned to the Y-arm, Kiwamu noticed that the AS beam was being clipped and that the correction had to start from SRM onwards as the beam had become offcentered on the SRM. So we shifted the SRM tower by about 5mm away from the door and transverse to the beam and rotated it by a few degrees to center the OSEM offsets. After this we aligned all optics along the AS beam path to extract the AS beam from the chamber. We then aligned each optic in the vertex so that their beams overlap at the AS port with the reflection from ITMY. Then we aligned BS to center the beam on ETMX and then looked for flashes from the Y arm.
At this point Kiwamu checked and found that the input beam from the MC had shifted. It was landing on the ITMY about 5mm below the center. And (inexplicably) it was still centered on ETMY! The Y- green which traced the cavity axis (since this was still flashing) was not coincident with the IR beam. So all the work we did in aligning the AS beam and the vertex optics work was lost..... and had to be done again.
|
5276
|
Mon Aug 22 11:40:08 2011 |
steve | Update | VAC | cranes checked | Cranes are checked and they are ready for lifting. At the east end will use the manual Genei-lift to put door on. |
5275
|
Mon Aug 22 02:43:28 2011 |
kiwamu | Update | General | in-vac work | [Kieko / Kiwamu]
The AS beam shows a little bit of clipping and right now this is the only concern for the alignment of the interferometer.
Other than that everything is okay including :
+ Beam centering on all the suspended optics
+ Arm cavity alignments. The Fabry-Perot fringes from both arm cavity were found on the AS camera.
+ POX/POY/POP, they are still successfully coming out from the chambers
+ IPPOS/ANG
+ Alignment of the green beams and the associated optics. Both green are reaching to the PSL table
We need to check/fix the AS beam clipping and once it's done we will readjust the OSEM mid range and the oplevs.
Then it will be ready for the drag/wipe and door closing. |
5274
|
Sat Aug 20 23:01:39 2011 |
Jenny | Update | PSL | Taking temperature step-response data: successes and tribulations | After finishing my last elog entry, I monitored the digital loop's error signal (the control signal for the fast loop) and the output to the laser heater remotely, (from West Bridge), using dataviewer. The ref cav surrounding can temperature was set to 36 degrees C.
With the loop closed and a gain of 0.008, after seeing the output voltage to the laser heater (TMP_OUTPUT) remain fairly constant and the error signal (TMP_INMON) stay close to zero for ~3 hours, I tried to step the temperature. (This was at 2am last night). I was working remotely from West Bridge. To step the temperature I used the following command:
ezcawrite C1:PSL-FSS_RCPID_SETPOINT 35.5
Rather than change the can temperature to 35.5 C, it outputted:
C1:PSL-FSS_RCPID_SETPOINT=0.
It had set the setpoint to 0 degrees C, which was essentially turning the heater off. I tried resetting it back to 36 and had no luck. I tried changing the syntax slightly.: ezcawrite C1:PSL-FSS_RCPID_SETPOINT=36 and ezcawrite C1:PSL-FSS_RCPID_SETPOINT (36). No success.
I ran over to the 40m and changed the temperature back to 36 manually. The in-loop temp sensor had decreased to 31.5 degrees C before I was able to step the setpoint back up. The system seems to have recovered from this large impulse though, and the laser has remained locked.


(5 hours of minute-trend data)
From left to right:
Top: Out-of-loop can temp sensor; Voltage sent to heat can
Middle: signal sent to heat the laser (TMP_OUTPUT); room temp
Bottom: Error signal for slow loop (sampled control signal from fast loop); In-loop can temp sensor
At 9:30 this morning (7 and a half hours after accidentally setting the setpoint to zero), I came in to the 40m. TMP_OUTPUT was still decreasing but was slowing somewhat, so I decided to step the can temperature up half a decree to 36.5 C.
TMP_OUTPUT responded to the step, but it is also oscillating slowly with room-temperature changes, and these oscillations are on the same order as the step response. The oscillations look like the room-temp oscilations, but inverted. (TMP_OUTPUT reaches maxima when RMTEMP reaches minima). Oddly, there does not appear to be much of a time delay between the room temperature and TMP_OUTPUT signals. I would expect a time delay since there's a time constant for a room-temperature change to propagate into the cavity. Perhaps the laser itself is susceptible to room-temperature changes and those propagate into the laser cavity on a much faster time scale. I don't know the thermal coupling of ambient temperature changes into the laser.


(24-hours of second-trend data)
Options are:
--If the system can handle it, do a larger temperature step (3 degrees, say), so that I can more clearly distinguish the oscillations with room temp from the step response.
--Insulate the cavity with foam (will in principle make the temperature over the can surrounding the ref cav more uniform and less affected by room temperature changes).
--Insulate the laser? Is this possible?
--Leave the system as is and, as a first approximation, fit the room-temp data to a sine wave and subtract it off somehow from my data to just see the step response.
--Don't bother with steps and just try to get the transfer function from out-of-loop temperature (RCTEMP, which is affected by temperature noise from the room) to TMP_OUTPUT via taking the Fourier transforms of both signals.
I'm flying out tomorrow morning, so I'll either need to figure out how to step the temperature set point of the can remotely, successfully, or I'll need someone to manually enter in the temperature steps for me in the control room. |
5273
|
Sat Aug 20 00:42:22 2011 |
Keiko | Update | LSC | Tolerance of PRC, SRC, MICH length = 2 mm ? |
Keiko, Kiwamu
I have run Kiwamu's length tolerance code (in CVS iscmodeling, ArmTolerance.m & analyseArmTorelance.m ) for the vertex ifo.
In his previous post, he monte-carlo-ed the arm lengths and saw the histogram of the sensing matrix and the demodulation phase between POP55 MICH and POP55 SRCL. From these plots, he roughly estimated that the tolerance is about 1 cm (sigma of the rondom gaussian) and in that case POP55 MICH and SRCL is separated by the demodulation phase 60-150 degrees.
This time I put the length displacements of random gaussian on PRC, SRC, MICH lengths at the same time (Fig.1).

Fig. 1. History of random walk in PRC, SRC, MICH lengths parameter space. Same as Kiwamu's previous post, The position of the three degrees are randomly chosen with a Gaussian distribution function in every simulaton. This example was generated when \sigma = 1 cm for all the three lengths, where \sigma is the standard deviation of the Gaussian function. The number of simulation is 1000 times.
When the sigma is 1 cm, we found that the sensing matrix is quite bad if you look at Fig. 2. In Fig.2 row POP55, although the desired degrees of freedoms are MICH and SRCL, they have quite a bit of variety. Their separation in the demodulation phase is plotted in Fig.3. The separation in the demodulation phase varies from 40 degrees to 140 degrees, and around 270 degrees. It is not good as ideally we want it to be 90.

Fig. 2 Histgram of the sensing signal power in the matrix when 1 cm sigma rondom gaussian is applied on PRC, SRC, MICH lengths. x axis it the signal power in log10.

Fig.3 POP55 MICH and POP55 SRCL separation with the displacement sigma 1 cm.
Kiwamu suspected that PRC length as more strict tolerance than other two (SRC, MICH) for POP55, as 55MHz is fast and can be sensitive to the arm length change. So I ran the same monte-carlo with SRC, MICH displacements but no PRC displacements when sigma is the same, 1cm. The results were almost same as above, nothing obvious difference.
With 2mm sigma, the signal power matrix and the POP55 MICH and POP55 SRCL separation in the demodulation phase look good (Fig. 4 and Fig. 5).
Fig.4 Signal power matrix when PRC, SRC, MICH lengths fractuate with random gaussian distribution with 2mm sigma. The signal powers are shown in log10 in x axis, and they do not vary very much in this case.
Fig.5 POP55 MICH and POP55 SRCL separation with the displacement sigma 2 mm. The separation of the two signal is 60-90 degrees, much better than when sigma is 1 cm. We may need to check 60 degree separation is really ok or not.
PRC SRC MICH lengths tolerances of 2 mm in the real world will be very difficult !
Next I will check what happens on 3f signals.
Quote: |
Required arm length = 37.7974 +/- 0.02 [m]
This is a preliminary result of the estimation of the Arm length tolerance.
This number was obtained from a simulation based on Optickle.
Note that the simulation was done by considering misplacements in only the arm lengths while keeping PRCL, SRCL and MICH at the ideal lengths.
Therefore the tolerance will be somewhat tighter if misplacements in the central part are taken into account.
Next : check 3f signals, and include misplacements in PRCL, SRCL and MICH.
Figure.2 A sensing matrix of the 40-m DRFPMI while changing the position of ETMX/Y by \sigma = 2 cm.
For convenience, only REFL11, AS55, POP11 and POP55 are shown. They are the designed signal ports that
mentioned in the aLIGO LSC document ( T1000298). In all the histograms, x-axis represents the optical gain in log scale in units of [W/m].
The y-axis is the number of events. The diagonal ports are surrounded by red rectangular window.
(Results2 : demodulation phase of MICH and SRCL on POP55)
Now a special attention should be payed on the MICH and SRCL signals on POP55.
Since MICH and SRCL are designed to be taken from POP55, they must be nicely separated in their demodulation phases.
Therefore the demodulation phase of MICH and SRCL has to be carefully examined.
The plot in Figure.3 is the resultant phase difference between MICH and SRCL on POP55 when \sigma_x = \sigma_y = 2 cm.
As shown in the plot the phase are always within a range of 60 - 120 deg, which satisfies my requirement (2) mentioned in the last section.
Figure.3 Difference in the demodulation phase of MICH and SRCL on POP55.
x-axis is the difference in the demodulation phase of MICH and SRCL, and y-axis the number of events.
|
|
5272
|
Fri Aug 19 23:41:20 2011 |
Jenny | Update | PSL | Relocking NPRO to reference cavity. |
Quote: |
I am trying again to measure a temperature step response on the reference cavity on the PSL table.
I have been working to relock the NPRO to the cavity. I unblocked the laser beam, reassembled the setup described in my previous elog entry: 5202. I then did the following:
1) Monitored error signal (from LB1005 PDH servo), transmitted signal, and control signal sent to drive PZT on oscilloscope.
2) With loop open, swept through 0,0-mode resonance, saw a peak in the transmission, saw an accompanying error signal similar to the signal shown in 5202.
3) Tried to lock. Swept the gain on the LB1005 and could not find a gain that would make it lock. Tried changing the PI-corner freq. from 10 kHz to 30 kHz and back and still could not lock.
4) Noticed that the open loop error signal displayed on the scope was DC-offset from zero. Changed the offset to zero the error signal.
5) Tried to lock again and succeeded.
6) Noticed that upon closing the loop, the error signal became offset from zero again. Turning on the integrator on the LB1005 increased DC-offset.
7) Reduced the gain on the SR560 being used as a low pass filter from 5 to 1. Readjusted the open loop error signal offset on the LB1005.
8) Closed the loop and locked. Closing the loop then caused a much smaller DC change in the signal than I had seen with the larger gain (now around 3mV). RMS fluctuations in error signal are now 1 mV (well within the linear region of the error signal).
9) Noticed transmission has a strange distorted harmonic oscillation in it a 1MHz. (Modulation freq is 230kHz, so it's not that). Checked reflected signal and also saw a strange oscillation--in a sawtooth-like pattern.
I intend to
1) Post oscilloscope traces here showing transmitted and reflected signal when locked.
2) Look upstream to see if the sawtooth-like oscillation is in the laser beam before it enters the cavity:
- Sweep the temperature of the laser so that the beam is no longer resonating in the cavity.
- Compare the reflected signal off the cavity to the signal detected before being directed into the cavity (using the PDA255 that I used for measuring the AM response of the PZT) both with and and without the frequency modulation.
3) At some point, try to close the slow digital loop, perhaps readjusting the gain.
4) Try to measure a temperature step response.
|
I decided to go forward and try to close the digital loop in spite of the unexplained oscillations in the transmission.
1) Plugged the 20dB attenuator into the slow port on the laser drive. This pushed the laser out of lock and, for some reason, made the laser temperature stop responding to sweeping the set point manually with the knob.
2) Plugged the output from the digital system into the slow port (with the attenuator still in place).
3) Displayed the beam seen by the camera on a monitor in the control room
4) Stepped the laser temperature using MEDM until finding the 0,1 mode. Locked to that mode.
5) Closed the digital loop (input to slow laser drive attenuated 20dB attenuator). Gain 0.010
6) Loop appeared stable for 30 minutes, then temperature began shooting off. I opened the loop, cleared history, reduced the gain to 0.008, and started it again. Loop appears stable after 15 minutes of watching. I'm going to leave it for a few hours, then come back to check on it and, if it's stable, step the can temperature. |
5271
|
Fri Aug 19 19:08:40 2011 |
Jenny | Update | PSL | Relocking NPRO to reference cavity. | I am trying again to measure a temperature step response on the reference cavity on the PSL table.
I have been working to relock the NPRO to the cavity. I unblocked the laser beam, reassembled the setup described in my previous elog entry: 5202. I then did the following:
1) Monitored error signal (from LB1005 PDH servo), transmitted signal, and control signal sent to drive PZT on oscilloscope.
2) With loop open, swept through 0,0-mode resonance, saw a peak in the transmission, saw an accompanying error signal similar to the signal shown in 5202.
3) Tried to lock. Swept the gain on the LB1005 and could not find a gain that would make it lock. Tried changing the PI-corner freq. from 10 kHz to 30 kHz and back and still could not lock.
4) Noticed that the open loop error signal displayed on the scope was DC-offset from zero. Changed the offset to zero the error signal.
5) Tried to lock again and succeeded.
6) Noticed that upon closing the loop, the error signal became offset from zero again. Turning on the integrator on the LB1005 increased DC-offset.
7) Reduced the gain on the SR560 being used as a low pass filter from 5 to 1. Readjusted the open loop error signal offset on the LB1005.
8) Closed the loop and locked. Closing the loop then caused a much smaller DC change in the signal than I had seen with the larger gain (now around 3mV). RMS fluctuations in error signal are now 1 mV (well within the linear region of the error signal).
9) Noticed transmission has a strange distorted harmonic oscillation in it a 1MHz. (Modulation freq is 230kHz, so it's not that). Checked reflected signal and also saw a strange oscillation--in a sawtooth-like pattern.
I intend to
1) Post oscilloscope traces here showing transmitted and reflected signal when locked.
2) Look upstream to see if the sawtooth-like oscillation is in the laser beam before it enters the cavity:
- Sweep the temperature of the laser so that the beam is no longer resonating in the cavity.
- Compare the reflected signal off the cavity to the signal detected before being directed into the cavity (using the PDA255 that I used for measuring the AM response of the PZT) both with and and without the frequency modulation.
3) At some point, try to close the slow digital loop, perhaps readjusting the gain.
4) Try to measure a temperature step response. |
5270
|
Fri Aug 19 15:31:53 2011 |
steve | Update | General | power interruption rescheduled to 10-1-2011 | UTILITY & SERVICE INTERRUPTION
**PLEASE POST**
Building: Central Engineering Services (C.E.S.)
Date: Saturday, October 1, 2011
Time: 8:00 a.m. To 9:00 a.m.
Interruption: Electricity
Contact: Mike Anchondo ext. 4999 Tom Brennan 4984
*This interruption is required for maintenance of high voltage switchgear in Campus Sub Station.
(If there is a problem with this Interruption, please notify
the Service Center X-4717 or the above Contact as soon as possible.
If no response is received we will proceed with the interruption.)
Reza Ohadi,
Director, Campus Operations & Maintenance |
5269
|
Fri Aug 19 10:26:53 2011 |
steve | Update | SUS | OSEM sensor spectra | Free swingging OSEM sensors LL at atm |
Attachment 1: 8freeSUSosemSensors.jpg
|
|
5268
|
Fri Aug 19 09:14:56 2011 |
steve | Update | IOO | degrading laser power at atm | Light into the MC is 20 mW at atm, MC_Transmitted ~10 MW = 400 count
The PMC_T is OK but something else is drifting. |
Attachment 1: power@atm.jpg
|
|
5267
|
Fri Aug 19 01:46:06 2011 |
Suresh | Update | General | IFO alignment | [Keiko, Jamie, Kiwamu, Anamaria,
We followed the procedure that we laid-out in our elog of yesterday. We completed the first six steps and we now have the y-arm well aligned to the green beam which passes through the center of of both ETMY and ITMY.
The IR beam was steered with the PZTs to coincide with the green beam. The BS was adjusted to see IR beam scatter on a target placed near the center of the ETMX. And then the AS IR beam was steered to the AS camera by adjusting several components along OM path ( we touched OM1, OM2, OM3, OM4, OM5, OMPO and OM6). We then looked for IR fringes in the AS port from the Y-arm. But no luck there. We need to realign the IR beam into the Y-arm cavity axis using the pzts.
We aligned ITMX and PRM to get power recycled Michelson fringes at the AS.
|
5266
|
Fri Aug 19 01:15:22 2011 |
Suresh | Update | SUS | FreeSwing all optics | I ran "freeswing all" at Fri Aug 19 01:09:28 PDT 2011 (997776583) and "opticshutdown" as well.
|
5265
|
Thu Aug 18 22:24:08 2011 |
jamie | Omnistructure | VIDEO | Updated 'videoswitch' script | I have updated the 'videoswitch' program that controls the video MUX. It now includes the ability to query the video mux for the channel mapping:
controls@pianosa:~ 0$ /opt/rtcds/caltech/c1/scripts/general/videoswitch -h
Usage:
videoswitch [options] [OUT] List current output/input mapping [for OUT]
videoswitch [options] OUT IN Set output OUT to be input IN
Options:
-h, --help show this help message and exit
-i, --inputs List input channels and exit
-o, --outputs List output channels and exit
-l, --list List all input and output channels and exit
-H HOST, --host=HOST IP address/Host name
controls@pianosa:~ 0$
|
5264
|
Thu Aug 18 15:54:35 2011 |
steve | Update | SUS | damped and undamped OSEMs | damped sus at atm1 and freeswingging sus at atm2
|
Attachment 1: 5susLL.jpg
|
|
Attachment 2: 8freeSUSLL.jpg
|
|
|