  2189   Fri Nov 6 00:40:29 2009 AlbertoUpdateLSCEverything put back as it was

I disconnected the setup for the arm cavity TF measurement. I opened the scitation switch on the ISS medm screen. I reconnected the OMC ISS EXC cable to the breakout box on the floor.

The photodiode on the Y end is stilll connected.

Also the RFAm (whish is not disbaled anymore) still has a 50% beam splitter before it.

I'm also running the Align Full IFO script.

  2217   Mon Nov 9 15:11:02 2009 AlbertoUpdateLSCEverything put back as it was

Quote:

I disconnected the setup for the arm cavity TF measurement. I opened the scitation switch on the ISS medm screen. I reconnected the OMC ISS EXC cable to the breakout box on the floor.

The photodiode on the Y end is stilll connected.

Also the RFAm (whish is not disbaled anymore) still has a 50% beam splitter before it.

I'm also running the Align Full IFO script.

 

I removed the beam splitter and the PDA 255.

the beam path to the RFAM photodiode is clear again.

  4187   Sat Jan 22 01:56:04 2011 SureshUpdateGreen LockingExamining the stability of VCO PLL at low frequencies

[Kiwamu, Suresh, Rana]

Our goal:

        We wished to determine the performance of the VCO PLL at low frequencies,. 

The procedure we followed:

        The scheme is to use the Marconi (locked to Rb Clock) as an 80MHz reference and lock to it using the PLL. 

        We set up the VCO PLL as in the diagram shown in the attachment and obtained the spectra shown below.

Results:

          We need to figure out the PLL servo gain profile in order to build the Inv PLL filter....

 

   

 

 

VCO_PLL_stability.png

 

 

  4188   Sat Jan 22 02:03:55 2011 KojiUpdateGreen LockingExamining the stability of VCO PLL at low frequencies

Damn. If this figure is true, we were looking at wrong signal. We should look at the feedback signal to the VCO.

  3090   Sat Jun 19 17:31:48 2010 ranaDAQCDSExcess Noise in C1:IOO-MC_DRUM1 fixed by reboot

I was getting an excess noise in the C1:IOO-MC_DRUM1 channel - it was a flat spectrum of 10 cts/rHz (corresponding to 600 uV/rHz).

I tried a few things, but eventually had to power cycle the crate with c1iovme in order to recover the standard ADC noise level of 3x10^-3 cts/rHz with a 1/sqrt(f) knee at 10 Hz.

I checked the gain of the channel by injecting a 2 Vpp sine wave at 137.035 Hz. 2Vpp as measured on a scope gives 31919 cts instead of the expected 32768, giving a 2.5% error from what we would have naively calculated.

Even so, the noise in this channel is very surprisingly good: 0.003 / (31919 / 2) = 187 nV /rHz.  The best noise I have previously seen from an ICS-110B channel is 800 nV/rHz. What's going on here?

  6007   Fri Nov 25 18:45:31 2011 Den, RanaSummarySUSExcess Noise in Digital Filtering

We are now trying to understand why the coherence between SUS-X_SUSPOS_IN1 and SUS-X_SUSPOS_OUT is lost below 1 Hz for X = MC1, MC2, MC3, BS, ITMX, ITMY, ETMX, ETMY, SRM. It is OKEY only for PRM but the different filteres are used there. For PRM - 30:0.0 and Bounce Roll, for all others - 30:0.0 and Cheby. The transfer functions between these two signals plotted in foton and fft tools are also not the same.

If we switch off all the filters between these 2 signals, than the coherence is one. If one of the filters is switched on, everything is also fine. But if there are several present, than they filter the signal in unexpected way.

Moreover it seems that the coherence is dependent on the input signal. The coherence is better with local dumping than without.

Attachment 1: FiltNoise.png
FiltNoise.png
  6008   Fri Nov 25 19:45:36 2011 Mirco, DenSummarySUSExcess Noise in Digital Filtering

Quote:

We are now trying to understand why the coherence between SUS-X_SUSPOS_IN1 and SUS-X_SUSPOS_OUT is lost below 1 Hz for X = MC1, MC2, MC3, BS, ITMX, ITMY, ETMX, ETMY, SRM. It is OKEY only for PRM but the different filteres are used there. For PRM - 30:0.0 and Bounce Roll, for all others - 30:0.0 and Cheby. The transfer functions between these two signals plotted in foton and fft tools are also not the same.

If we switch off all the filters between these 2 signals, than the coherence is one. If one of the filters is switched on, everything is also fine. But if there are several present, than they filter the signal in unexpected way.

Moreover it seems that the coherence is dependent on the input signal. The coherence is better with local dumping than without.

 We have done the following on the c1sus and c1lsc computers : provided excitation, let the signal pass through filters 30:0.0 and Cheby and plotted the coherence between in and out signals.

c1sus computer - coherence is corrupted

c1lsc computer - coherence is not corrupted

Attachment 1: sus_coh-crop.pdf
sus_coh-crop.pdf
Attachment 2: lsc_coh-crop.pdf
lsc_coh-crop.pdf
  15641   Fri Oct 23 16:41:06 2020 KojiUpdateIOOExcess laser freq noise investigation

[Koji, Rana]

We wanted to track down the excess noise seen in MC_F and other places (see the previous report by Gautam)


Setup1: The IMC was locked and MC_F signal between 500 and 1500Hz was observed. The DTT template was saved as /users/Templates/MC/MCF_noise_201023.xml

- Suspected mech resonance/jitter coupled with clipping or any other imperfections. Poked the various optics and optomechanics on the table. Basically no change. If we tap the laser chassis and the optics close to the laser source, we occasionally unlocked the IMC

- When we touched (lifted) the Innolight controller box from the shelf, for the first time we saw a significant change in the shape of the noise spectrum. The peak around the 700Hz shited towards lower frequency by a few %. Other peaks have no obvious change in the shapes and the heights.

- While observing the MC_F signal on the laptop, we went to the back of the laser controller. Placing a hand close to the fan clearly changes the peak frequency lower. By temporarily disconnecting the fan from the power supply for a short moment, the 700Hz peak could be eliminated. We also tried to see the noise level with the slow thermal servo and diagnosis DB cable disconnected, but we didn't see any significant change of the noise level.


Setup 2: Using the ALS phase tracker, we can observe the relative freq noise of the PSL laser and the ETMY AUX laser without any servo involved. This way we can freely disconnect any cables from the lasers. The measurement template for DTT was saved as /users/Templates/ALS/Y_ALS_FINE_PHASE_OUT_102320.xml

- Noise spectrum before disconnecting the cable (REF0, RMS REF1)

- The Fast PZT input to the PSL was disconnected => This made all the peaks (including the 700Hz) disappeared (REF2, RMS REF3)

- The Fast PZT input was restored as before, then the chain was disconnected at the input of the HV PZT driver (Thorlabs) => Again, this made the peaks disappeared (REF4, RMS REF5)

- The chain was disconnected at the input of the TTFSS box => Again, this made the peaks disappeared (REF6, RMS REF7)

- Disconnected the demod input and the AO cables from the IMC servo board => This made the peaks came back (REF8)

- Disconnected all the input/peripheral cables from the IMC servo board except for the connection to the TTFSS box => Still the excess noise was observed  (REF9)

- In addition to the above, the cable to the FSS box was disconnected but the ground was still touching the MC servo board => This made the peaks disappeared (REF10)

The conclusion is that the noise is injected from the main circuit of the IMC servo board.


Next time we will check if the backplane connection is doing something wrong. Also, we'll test if the presence of the RF signals does something bad to the IMC board via EMI and RFI.

We have reverted the connection and tested if we lock the IMC and Y arm. ==> We saw at least they were locked for a short period. The things are still stabilizing, but left them turned on so they keep trying to lock automatically for the night.

Attachment 1: plot.pdf
plot.pdf
  15643   Mon Oct 26 13:35:58 2020 KojiUpdateIOOExcess laser freq noise investigation

In fact, the problem was the grounding issue (presumably on the IOO racks).
A temporary differential receiver at the TTFSS side was built using an SR560 and a few ponoma cables. This removed the structures ~850Hz.


The MC Servo Output was disconnected from the TTFSS box and monitored with SR785. The 850Hz structure was kept visible no matter what cables, including all the acromag DB cables, were removed. This made me suspicious about the measurement setup. The SR785 was connected to an AC power strip under the SP table and this was too far from the IOO rack.

The SR785 was connected to the AC power strip on 1X2, and now the difference becomes clear. No matter if the acromag cables are connected or not, the connection (particularly ground connection) between the MC servo module and the TTFSS box causes the MC servo output contaminated. (Comparison between Blue and Orange of Attachment #1). During the measurement, the EPICS switch for the fast path was disengaged (=no signal) and the VCO gain (...so called. It's just the MC Servo Gain) was set to be 0dB.

To test if the differential receiving of the MC Servo Output at the PSL helps to reduce this noise, I've built a simple (hacky) differential receiver using an SR560. (Attachment #2)
This kept the noise level same as the disconnected case (Comparison between Green and Orange of Attachment #1, I don't think the difference between them is not significant), while the IMC is locked as before.
Note that we can see that the 36kHz line was significantly reduced. Did we remove this annoying noise?

After talking with Gautam, we decided to leave this configuration while the SE-Diff cable was replaced with a more robust one. (See Attachment #3)


The PSL laser frequency performance was evakluated in the following two ways as we did last week:
1) Use the beat frequency of the free running PSL and the Y-end laser (Attachment #4). The PSL shutter was closed and thus the IMC was not locked.
2) Use the IMC MCF while the IMC was locked. (Attachment #5)

For both cases, the improvement was confirmed.


I also tried to check the reported issue by Gautam on this elog. He used 1Hz BW, but I cheated with 16Hz BW and 10x12.8kHz span PSDs. (Attachment #6)

For the measurement, IN1 GAIN of the IMC Servo was set to be 0dB and the OUT2 was switched to monitor the IN1 noise, while IN1 was terminated by a 50Ohm.

As I mentioned above, the AC power of SR785 was taken from a 1X2 power strip. Is this the reason for the power line forest look less severe compared to the previous case???
Anyway, I tried to use the same differential receiving technique (but with gain of x100) to see if this helps. The differential receiver helped to reduce the structure above 50kHz. The floor noise level was observed to be higher. I didn't pursue this any further, but the forest of the power line looked like a part of the measurement noise. This is indicative that the grounding condition on 1X2 is really not great and we need to review the configuration of the acromag grounding.

Attachment 1: MC_Servo_Output.pdf
MC_Servo_Output.pdf
Attachment 2: 20201026135735_IMG_0175.jpg
20201026135735_IMG_0175.jpg
Attachment 3: 20201026153435_IMG_0176.jpg
20201026153435_IMG_0176.jpg
Attachment 4: Screen_Shot_2020-10-26_at_1.15.54_PM.png
Screen_Shot_2020-10-26_at_1.15.54_PM.png
Attachment 5: Screen_Shot_2020-10-26_at_1.35.19_PM.png
Screen_Shot_2020-10-26_at_1.35.19_PM.png
Attachment 6: MC_Servo_Error_Mon.pdf
MC_Servo_Error_Mon.pdf
  15644   Mon Oct 26 17:26:26 2020 gautamUpdateIOOExcess laser freq noise investigation

Apart from the questionable wiring on the Acromags, one other important difference is in the way the connections were made between the old VME crates to the Eurocrate backplanes, and how we do it now. The thick cables had their sheilds connected to the eurocrate ground (or at least, there was a dedicated ground lug on those cables which we screwed on to the ground terminals on the Eurocrate backplanes). However, in our current configuration, we interface the Acromag ADCs and DACs to the backplane via these adaptor boards. The shields of the DSUB cables are presumably NOT connected to the Eurocrate grounds. This should also be investigated as one potential cause of the grounding issue - while on some of the Eurocrate modules, the P1/P2 connectors may have either the "A" or "C" row of connectors shorted to ground, some may not, and the TTFSS may suffer from such an issue?

Note that we have this problem in all of the slow machines that were upgraded to Acromag (if this turns out to be the issue). 

Quote:

In fact, the problem was the grounding issue (presumably on the IOO racks).

  15259   Fri Mar 6 01:19:19 2020 gautamUpdateIOOExcess laser frequency noise

Summary:

Sometime between 1PM and 6PM on Tuesday, excess laser frequency noise shows up in MCF at around 800 Hz, as shown in Attachment #1. Sigh.

Details:

While I show the MCF spectrum here, I confirmed that this noise is not injected by the IMC loop (with the PSL shutter closed, and the IMC servo board disconnected from the feedback path to the NPRO, the PMC error and control points still show the elevated noise, see Attachment #2). I don't think the problem is from the PMC loop - see Attachment #3 which is the ALS beat out-of-loop noise with the PMC unlocked (the PSL beam doesn't see the cavity before it gets to the ALS setup, and we only actuate on the cavity length for that loop, so this wasn't even really necessary).

Was there some work on the PSL table on Tuesday afternoon that can explain this

Attachment 1: MCF.pdf
MCF.pdf
Attachment 2: ExcessFreqNoise.pdf
ExcessFreqNoise.pdf
Attachment 3: ALSnoise.pdf
ALSnoise.pdf
  15260   Fri Mar 6 16:33:11 2020 gautamUpdateIOOExcess laser frequency noise

I did some preliminary debugging of this, and have localized the problem to the output path (after MC slow) on the IMC Servo card. Basically, I monitored the spectrum of the ALS beat frequency fluctuations under a few different conditions: 

  • With the BNC to the NPRO PZT disconnected, there is no noise. So the laser and the FSS phase correction EOM (which the ALS beat pickoff sees) are not responsible.
  • With the input to the Koji summing box disconnected, still no excess - so the summing box + Thorlabs HV driver are not responsible.
  • With the TTFSS output connected to the summing box, but with the input switch to the TTFSS box disabled (isolating the on-PSL table parts of the FSS system), still no excess.
  • With the input to the TTFSS enabled, and the BNC output of the IMC Servo card disconnected at 1X2, still no excess.
  • Finally, when I connect the BNC cable, the excess starts to show up.

Toggling C1:IOO-MC_FASTSW, which supposedly isolates the post-MC slow (a.k.a. MCL) part of the servo, I see no difference. I am also reasonably confident this switch itself works, because I can break the IMC lock by toggling it. So pending a more detailed investigation, I am forced to conclude that the problem originates in the part of the IMC servo board after the MCL pickoff. Some cabling was removed at 1X2 on Tuesday between the times when there was no excess and when it showed up, but it's hard to imagine how this could have created this particular problem.

  16329   Tue Sep 14 17:19:38 2021 PacoSummaryPEMExcess seismic noise in 0.1 - 0.3 Hz band

For the past couple of days the 0.1 to 0.3 Hz RMS seismic noise along BS-X has increased. Attachment 1 shows the hour trend in the last ~ 10 days. We'll keep monitoring it, but one thing to note is how uncorrelated it seems to be from other frequency bands. The vertical axis in the plot is in um / s

Attachment 1: SEIS_2021-09-14_17-33-12.png
SEIS_2021-09-14_17-33-12.png
  16331   Tue Sep 14 19:12:03 2021 KojiSummaryPEMExcess seismic noise in 0.1 - 0.3 Hz band

Looks like this increase is correlated for BS/EX/EY. So it is likely to be real.

Comparison between 9/15 (UTC) (Attachment 1) and 9/10 (UTC) (Attachment 2)

Attachment 1: C1-ALL_393F21_SPECTRUM-1315699218-86400.png
C1-ALL_393F21_SPECTRUM-1315699218-86400.png
Attachment 2: C1-ALL_393F21_SPECTRUM-1315267218-86400.png
C1-ALL_393F21_SPECTRUM-1315267218-86400.png
  5623   Wed Oct 5 18:31:02 2011 KatrinUpdateGreen LockingExchanged mirror on YARM table

On the Green YARM end table the second mirror behind the laser has been exchanged.

Reason: The light is impinging on the mirror under an angle of  about 10 degrees, but the old mirror was coated for angle of incidence (aoi) of 45°.

Thus it was more like a beam splitter. The new mirror is a 1" Goock & Housego mirror which has a better performance for an aoi of 10 degree.

Realignment through Faraday Isolator and SHG cristall as well as 532nm isolator is almost finished.

  8900   Tue Jul 23 04:07:48 2013 gautamUpdateCDSExcitation points set up on c1scx

 

 In light of recent events and the decision to test the piezo tip-tilts for green beam steering on the X-end table, I have set up 8 excitation points to channels 8 through 15 of the DAC on c1scx (as was done earlier for the DAC at 1Y4 with Jenne's help) in order to verify that the pin-outs of the DAC interface board. I have not yet compiled the model or restarted the computer, and will do these tomorrow, after which I will do the test. The channels are named YYY_CHAN9 etc. 

 

 

 

 

  8909   Tue Jul 23 16:47:01 2013 gautamUpdateCDSExcitation points set up on c1scx

 I just compiled and installed the model with the excitation points on c1scx and then restarted framebuilder. The channels I set up are now showing up in the awggui dropdown menu. I will do the tests on the DAC channels shortly.

Just to keep things on record, these are the steps I followed:

  • opened the model c1scx (path: /opt/rtcds/userapps/release/sus/c1/models) with MATLAB
  • Added 8 excitation points and saved the model. A copy has been saved as c1scx.mdl.r2010b because of the recent upgrade to r2013a. 
  • ssh to c1iscex (computer running the model c1scx). 
  • Entered the following sequence of commands in terminal: rtcds make c1scx ,  rtcds install c1scx , rtcds start c1scx 
  • ssh to framebuilder, and restarted the framebuilder by entering telnet fb 8088   and then   shutdown.
  1708   Sat Jun 27 03:16:16 2009 ClaraUpdatePEMExciting microphone things!

So, I'm double-posting, but I figured the last post was long enough as it was, and this is about something different. After double and triple checking the XLR cables, I hooked up the microphone setup (mic---preamp---output) to the oscilloscope to figure out what kind of voltage would register with loud noises. So, I clapped and shouted and forgot to warn the other people in the lab what I was doing (sorry guys) and discovered that, even on the lowest gain setting, my loud noises were generation 2-3 times as much voltage as the ADC can handle (2V). And, since our XLR cables are so freaking long, we probably want to go for a higher gain, which puts us at something like 20 times too much voltage. I doubt this is really necessary, but it's late (early) and I got camera-happy, so I'm going to share anyway:

mic_voltage_out.png

So, to deal with this issue, I made some nifty voltage dividers. Hopefully they are small enough to fit side-by-side in the ports without needing extra cableage. Anyway, they should prevent the voltage from getting larger than 2V at the output even if the mic setup is producing 50V. Seeing as my screaming as loud as I could about 2mm away from the mic at full gain could only produce 45V, I think this should be pretty safe. I put the ADC in parallel with a 25.5 kOhm resistor, which should have a noise like 10^-8 V/rHz. This is a lot smaller than 1 uV/rHz (the noise in the ADC, if I understood Rana's explanation correctly), so the voltage dividers should pose a noise issue. Now for pictures.

voltage_dividers.png

I opened one so you can see its innards.

voltage_divider_sketch.png

In case the diagram on the box was too small to decipher...

And finally, I came up with a name scheme for the mics and pre-amps. We now have two Bluebird (bacteriophage) mics named Bonnie and Butch Cassidy. Their preamps are, naturally, Clyde and The Sundance Kid. Sadly, no photos. I know it's disappointing. Also, before anyone gives me crap for putting the labels on the mics upside-down, they are meant to be hung or mounted from high things, and the location (and stiffness) of the cable prevents us from simply standing them up. So they will more than likely be in some kind of upsidedownish position.

  15763   Thu Jan 14 11:46:20 2021 gautamUpdateCDSExpansion chassis from LHO

I picked the boxes up this morning. The inventory per Fil's email looks accurate. Some comments:

  1. They shipped the chassis and mounting parts (we should still get rails to mount these on, they're pretty heavy to just be supported on 4 rack nuts on the front). idk if we still need the two empty chassis that were requested from Rich.
  2. Regarding the fibers - one of the fibers is pre-2012. These are known to fail (according to Rolf). One of the two that LHO shipped is from 2012 (judging by S/N, I can't find an online lookup for the serial number), the other is 2011. IIRC, Rolf offered us some fibers so we may want to take him up on that. We may also be able to use copper cables if the distances b/w server and expansion chassis are short.
  3. The units are fitted with a +24V DC input power connector and not the AC power supplies that we have on all the rest of the chassis. This is probably just gonna be a matter of convenience, whether we want to stick to this scheme or revert to the AC input adaptor we have on all the other units. idk what the current draw will be from the Sorensen - I tested that the boards get power, and with noi ADCs/DACs/BIOs, the chassis draws ~1A (read off from DCPS display, not measured with a DMM). ~Half of this is for the cooling fans It seems like KT @ LLO has offered to ship AC power supplies so maybe we want to take them up on that offer.
  4. Without the host side OSSI PCIe card, timing interface board, and supermicro servers that actually have enough PCIe slots, we still can't actually run any meaningful test. I ran just a basic diagnostic that the chassis can be powered on, and the indicator LEDs and cooling fans run.
  5. Some photos of the contents are here. The units are stored along the east arm pending installation.

    >     Koji,
    >
    >     Barebones on this order.
    >
    >        1. Main PCIe board
    >        2. Backplane (Interface board)
    >        3. Power Board
    >        4. Fiber (One Stop) Interface Card, chassis side only
    >        5. Two One Stop Fibers
    >        6. No Timing Interface
    >        7. No Binary Cards.
    >        8. No ADC or DAC cards
    >
    >     Fil Clara
    >       
  15764   Thu Jan 14 12:19:43 2021 JonUpdateCDSExpansion chassis from LHO

That's fine, we didn't actually request those. We bought and already have in hand new PCIe x4 cables for the chassis-host connection. They're 3 m copper cables, which was based on the assumption of the time that host and chassis would be installed in the same rack.

Quote:
  1. Regarding the fibers - one of the fibers is pre-2012. These are known to fail (according to Rolf). One of the two that LHO shipped is from 2012 (judging by S/N, I can't find an online lookup for the serial number), the other is 2011. IIRC, Rolf offered us some fibers so we may want to take him up on that. We may also be able to use copper cables if the distances b/w server and expansion chassis are short.
  15766   Fri Jan 15 15:06:42 2021 JonUpdateCDSExpansion chassis from LHO

Koji asked me assemble a detailed breakdown of the parts received from LHO, which I do based on the high-res photos that Gautam posted of the shipment.

Parts in hand:

Qty Part Note(s)
2 Chassis body  
2 Power board and cooling fans As noted in 15763, these have the standard LIGO +24V input connector which we may want to change
2 IO interface backplane  
2 PCIe backplane  
2 Chassis-side OSS PCIe x4 card  
2 CX4 fiber cables These were not requested and are not needed

Parts still needed:

Qty Part Note(s)
2 Host-side OSS PCIe x4 card These were requested but missing from the LHO shipment 
2 Timing slave These were not originally requested, but we have recently learned they will be replaced at LHO soon

Issue with PCIe slots in new FEs

Also, I looked into the mix-up regarding the number of PCIe slots in the new Supermicro servers. The motherboard actually has six PCIe slots and is on the CDS list of boards known to be compatible. The mistake (mine) was in selecting a low-profile (1U) chassis that only exposes one of these slots. But at least it's not a fundamental limitation.

One option is to install an external PCIe expansion chassis that would be rack-mounted right above the FE. It is automatically configured by the system BIOS, so doesn't require any special drivers. It also supports hot-swapping of PCIe cards. There are also cheap ribbon-cable riser cards that would allow more cards to be connected for testing, although this is not as great for permanent mounting.

It may still be better to use the machines offered by Keith Thorne from LLO, as they're more powerful anyway. But if there is going to be an extended delay before those can be received, we should be able to use the machines we already have in conjunction with one of these PCIe expansion options.

  15767   Fri Jan 15 16:54:57 2021 gautamUpdateCDSExpansion chassis from LHO

Can you please provide a link to this "list of boards"? The only document I can find is T1800302. In that, under "Basic Requirements" (before considering specific motherboards), it is specified that the processor should be clocked @ >3GHz. The 3 new supermicros we have are clocked at 1.7 GHz. X10SRi-F boards are used according to that doc, but the processor is clocked at 3.6 or 3.2 GHz.

Please also confirm that there are no conflicts w.r.t. the generation of PCIe slots, and the interfaces (Dolphin, OSSI) we are planning to use - the new machines we have are "PCIe 2.0" (though i have no idea if this is the same as Gen 2). 

Quote:

The motherboard actually has six PCIe slots and is on the CDS list of boards known to be compatible.

As for the CX4 cable - I still think it's good to have these on hand. Not good to be in a situation later where FE and expansion chassis have to be in different racks, and the copper cable can't be used.

Attachment 1: Screenshot_2021-01-15_17-00-06.png
Screenshot_2021-01-15_17-00-06.png
  15770   Tue Jan 19 13:19:24 2021 JonUpdateCDSExpansion chassis from LHO

Indeed T1800302 is the document I was alluding to, but I completely missed the statement about >3 GHz speed. There is an option for 3.4 GHz processors on the X10SRi-F board, but back in 2019 I chose against it because it would double the cost of the systems. At the time I thought I had saved us $5k. Hopefully we can get the LLO machines in the near term---but if not, I wonder if it's worth testing one of these to see whether the performance is tolerable.

Can you please provide a link to this "list of boards"? The only document I can find is T1800302....

I confirm that PCIe 2.0 motherboards are backwards compatible with PCIe 1.x cards, so there's no hardware issue. My main concern is whether the obsolete Dolphin drivers (requiring linux kernel <=3.x) will work on a new system, albeit one running Debian 8. The OSS PCIe card is automatically configured by the BIOS, so no external drivers are required for that one.

Please also confirm that there are no conflicts w.r.t. the generation of PCIe slots, and the interfaces (Dolphin, OSSI) we are planning to use - the new machines we have are "PCIe 2.0" (though i have no idea if this is the same as Gen 2). 

  7905   Wed Jan 16 18:08:06 2013 JenneUpdateLockingExpected PRC gains

I was calculating the power recycling gains we expect for different versions of the PRC, and I am a little concerned that we aren't going to have much gain with the new LaserOptik mirrors.

I'm using

                       t_PRM^2

G =  -------------------------------------------

       (1 - r_PRM * r_PR2 * r_PR3 * r_end)^2

 

from eqn 11.20 in Siegman.

r_end is either the ITM (for a symmetric Michelson) or the flat mirror that we'll put in (for the PR-flat test case).

r = sqrt( R ) = sqrt( 1 - T ) for mirrors whose power transmission is the quoted value.

 

Some values: 

t_PRM^2 = T_PRM = 0.055   --------->   r_PRM = sqrt( 1 - 0.055 )

T_G&H = 20e-6   ---->   r_G&H = sqrt( 1 - 20e-6 )

T_LaserOptic = 0.015 (see elog 7624 where Raji measured this...1.5% was the best that she measured for P polarization.  Elog 7644 has more data, with 3.1% for 40deg AoI) -------> r_LasOpt = sqrt( 1 - 0.015 ) or sqrt( 1 - 0.031)

T_ITM = 0.014 -----------> r_ITM = sqrt( 1 - 0.014 )

 

Some calculations with 1.5% LaserOptik transmission:

G_PRC_2G&H = 45

G_PRC_G&H_LasOpt = 31

G_PRM_flatG&H = 51

With the 3% LaserOptik transmission:

G_PRC_G&H_LasOpt = 22

G_PRM_flatG&H = 30

More ideal case of just PRM, flat mirror (either ITM or G&H), ignoring the folding mirrors:

G_PRM_ITM = 45

G_PRM_flatG&H = 70

 

Punchline:

If the LaserOptik mirror has 1.5% transmission at ~45 degrees, the regular PRC expected gain goes down to 31, from 45 with both folding mirrors as G&Hs.

  7909   Wed Jan 16 20:27:16 2013 ranaUpdateLockingExpected PRC gains

Why would we use such a bad optic in our recycling cavity? Is 1.5% the spec for these mirrors? Is this the requirement that Kiwamu calculated somehow? Did anyone confirm this measurement?

I can't believe that we'll have low noise performance in a RC where we dump so much power.

  7910   Thu Jan 17 00:17:31 2013 JenneUpdateLockingExpected PRC gains

Quote:

Why would we use such a bad optic in our recycling cavity? Is 1.5% the spec for these mirrors? Is this the requirement that Kiwamu calculated somehow? Did anyone confirm this measurement?

I can't believe that we'll have low noise performance in a RC where we dump so much power.

 Yeah, Koji mentioned in response to Raji's measurements several months ago that the LaserOptic mirros were pretty far out of spec. We should probably redo the measurement to confirm.

  3224   Wed Jul 14 19:36:17 2010 GopalUpdateOptic StacksExperimental Confirmation of COMSOL Analysis

I experimentally determined the spring constant of a single Vitol spring in order to obtain a rough estimate for the natural frequency of single-stack oscillation.

The procedure basically involved stacking metal bars of known mass onto the Vitol and using a caliper to measure deviations from the equilibrium length.

The plot below shows that, for small compressions, the response is linear to an R-squared of 0.98.

Untitled.png

The experimental spring constant came out to be about 270 lb/ft, or 3900 N/m.

Previous documents have listed that the top stack takes on a load of approximately 43 kg. per individual spring. A bit of calculation yields an experimental resonant frequency of 9.5 Hz.

Compared with the theoretical COMSOL first harmonic of about 7.5 Hz, there is a reasonable amount of error. Of course, I used this calculation as a simple ballpark estimate: errors from misplacement onto the Viton were minimized through use of a level, but were still inevitable on the mm scale. Since the two methods yield answers with the same order of magnitude, we are ready to move forward and build the remaining layers of the stack.

  4208   Wed Jan 26 12:04:31 2011 josephbUpdateCDSExplanation of why c1sus and c1lsc models crash when the other one goes down

So apparently with the current Dolphin drivers, when one of the nodes goes down (say c1lsc), it causes all the other nodes to freeze for up to 20 seconds.

This 20 seconds can force a model to go over the 60 microseconds limit and is sufficiently long enough to force the FE to time out.  Alex and Rolf have been working with the vendors to get this problem fixed, as having all your front ends go down because you rebooted a single computer is bad.

[40184.120912] c1rfm: sync error my=0x3a6b2d5d00000000 remote=0x0
[40184.120914] c1rfm: sync error my=0x3a6b2d5d00000000 remote=0x0
[44472.627831] c1pem: ADC TIMEOUT 0 7718 38 7782
[44472.627835] c1mcs: ADC TIMEOUT 0 7718 38 7782
[44472.627849] c1sus: ADC TIMEOUT 0 7718 38 7782
[44472.644677] c1rfm: cycle 1945 time 17872; adcWait 15; write1 0; write2 0; longest write2 0
[44472.644682] c1x02: cycle 7782 time 17849; adcWait 12; write1 0; write2 0; longest write2 0
[44472.646898] c1rfm: ADC TIMEOUT 0 8133 5 7941

The solution for the moment is to start the computers at exactly the same time, so the dolphin is up before the front ends, or start the models by hand after the computer is up and dolphin running, but after they have timed out.  This is done by:

sudo rmmod c1SYSfe

sudo insmod /opt/rtcds/caltech/c1/target/c1SYS/bin/c1SYSfe.ko

 

Alex and Rolf have been working with the vendors to get this fixed, and we may simply need to update our Dolphin drivers.  I'm trying to get in contact with them and see if this is the case.

  15503   Tue Jul 28 13:55:11 2020 HangUpdateBHDExploring bilinear SRCL->DARM coupling

We explore bilinear SRCL to DARM noise coupling mechanisms, and show two cases that by doing BHD readout the noise performance can be improved. In the first case, the bilinear piece is due to residual DHARD motion (see also LHO:45823), and it matters mostly for the low-frequency (<100 Hz) part, and in the second piece the bilinear piece is due to residual SRCL fluctuation and it matters mostly for the a few x 100 Hz part. Details are below:

=================================================

General Model:

We can write the SRCL to DARM transfer function as (Evan Hall's thesis, eq. 2.29)

Z_s2d(f) = C_lf(f) * F^2 * x_D + C_hf(f) * F * dphi_S * x_D    ---- (1)

where

C_lf ~ 1/f^2 and C_hf ~ f are constants at each frequency unless there are major upgrades to the IFO,

F is the finesse of the arm cavity which depends on the alignment, spot position on the TMs, etc., 

dphi_S is the SRCL detuning (wrt the nominal 90 deg value), 

x_D is the DC DARM offset. 

The linear part of this can be removed with feedforward subtractions and it is the bilinear piece that matters, which reads

dZ_s2d = C_lf * <F>^2 * dx_D + C_hf * <F> * <dphi_S> * dx_D

             + 2C_lf * <F> * <x_D>  * dF + C_hf * <dphi_S> * <x_D> * dF

             + C_hf  * <F> * <x_D> * d(dphi_S).     ---- (2)

The first term in (2) is due to residual DARM motion dx_D. This term does not depends on the DC value of DARM offset <x_D> and thus does not depend on doing BHD or DC readout. On the other hand, the typical residual DARM motion is 1 fm << 1 pm of DARM offset. Since the current feedforward reduction factor is about 10 (see both Den Martynov's thesis and Evan Hall's thesis), clearly we are not limited by the residual DARM motion. 

The second term is due to the change in the arm finesse, which can be affected by, e.g., the alignment fluctuation (both increasing the loss due to scattering into 01/10 modes and affecting the spot positon and hence changing the losses), and is likely to be the reason why we see the effect being modulated by DHARD. 

The last term in (2) is due to the residual SRCL fluctuation and is important for the ~ a few x 100 Hz band.

=================================================

DHARD effects. 

As argued above, the DHARD affects the SRCL -> DARM coupling as it changes the finesse in the arm cavity (through scattering into 01/10 modes; in finesse we cannot directly simulate the effects due to spot hitting a rougher location). 

Since in the second term of eq. (2) the LF part depends on the DARM DC offset <x_D>, this effect can be improved by going from DC readout to BHD. 

To simulate it in finesse, at a fixed DARM DC offset, we compute the SRCL->DARM transfer functions at different DHARD offsets, and then numerically compute the derivative \partial Z_s2d / \partial \theta_{DH}. Then multiplying this derivative with the rms value of DHARD fluctuation \theta_{DH} we then know the expected bilinear coupling piece. 

The result is shown in the first attached plot. Here we have assumed a flat SRCL noise of 5e-16 m/rtHz for simplicity (see PRD 93, 112004, 2016). We do not account for the loop effects which further reduces the high frequency components for now. The residual DHARD RMS is assumed to be 1 nrad. 

In the first plot, from top to bottom we show the SRCL noise projection at different DARM DC offsets of (0.1, 1, 10) pm. Since the DHARD alignment only affects the arm finesse starting at quadratic order, it thus matters what DC offset in DHARD we assume. In each pannel, the blue trace is for no DC offset in DHARD and the orange one for a 5 nrad DC offset. As a reference, the A+ sensitivity is shown in grey trace in each plot as a reference. 

We can see if there is a large DC offset in DHARD (a few nrad) and we still do DC readout with a few pm of DARM offset, then the bilinear piece of SRCL can still contaminate the sensitivity in the 10-100 Hz band (bottom panel; orange trace). On the other hand, if we do BHD, then the SRCL noise should be down by ~ x100  even compared to with the top panel. 

(A 5 nrad of DC offset in DHARD coupled with 1 nrad RMS would cause about 0.5% RIN in the arms. This is somewhat greater than the typically measured RIN which is more like <~ 0.2%. See the second plot). 

=================================================

SRCL effect. 

Similarly we can consider the SRCL->DARM coupling due to residual SRCL rms. The approach is very similar to what we did above for DHARD. I.e., we compute Z_s2d at fixed DARM offset and for different SRCL offsets, then we numerically evaluate \partial Z_s2d / \partial dphi_S. A residual SRCL rms of 0.1 nm is then used to generate the projection shown in the third figure. 

Unlike the DHARD effect, the bilinear SRCL piece does not depend on the DC SRCL detuning (for the 50-500 Hz part). It does still depends on the DARM DC offset and therefore could be improved by BHD.

Since we do not include the LP of the SRCL loop in this plot, the HF noise at 1 kHz is artifical as it can be easily filtered out. However, the LP will not be very strong around 100-300 Hz for a SRCL UGF ~ 30 Hz, and thus doing BHD could still have some small improvements for this effect. 

Attachment 1: SRCL_bilin_DHARD.pdf
SRCL_bilin_DHARD.pdf
Attachment 2: ARM_RIN.pdf
ARM_RIN.pdf
Attachment 3: SRCL_bilin_SRCL.pdf
SRCL_bilin_SRCL.pdf
  16100   Thu Apr 29 17:43:48 2021 AnchalUpdateCDSF2A Filters double check

I double checked today and the F2A filters in the output matrices of MC1, MC2 and MC3 in the POS column are ON. I do not get what SDF means? Did we need to add these filters elsewhere?

Quote:

The IMC suspension team should double check their filters are on again. I am not familiar with the settings and I don't think they've been added to the SDF.

Attachment 1: F2AFiltersON.png
F2AFiltersON.png
  16105   Fri Apr 30 00:20:30 2021 gautamUpdateCDSF2A Filters double check

The SDF system is supposed to help with restoring the correct settings, complementary to burt. My personal opinion is that there is no need to commit these filters to SDF until we're convinced that they help with the locking / noise performance.

Quote:

I double checked today and the F2A filters in the output matrices of MC1, MC2 and MC3 in the POS column are ON. I do not get what SDF means? Did we need to add these filters elsewhere

  6004   Thu Nov 24 20:22:42 2011 MirkoUpdateIOOF2A filter for MC

I calculated the F2A filters for the input mode cleaner optics as described in T010140-01-D eq (4). On Ranas recommendation I added an s/ ( w_0 * Q ) term to the numerator.

The used values are:

w_0 = 2pi / s
h= 0.0009
D= 2.46957E-2
Q=10

UpperCoils.pdf

LowerCoils.pdf

I put theses filters into C1:SUS-MC1_TO_COIL_1_1 to _4_1 . For convenience split in Z and P. Well it doesn't work. After a few seconds the optic begins to swing wildly.

  6006   Fri Nov 25 17:52:28 2011 ranaUpdateIOOF2A filter for MC

Woo. Pretty crazy. The numerators should only be ~10% larger than the denominator below 1 Hz. Let's try again.

  6012   Fri Nov 25 23:25:24 2011 MirkoUpdateIOOF2A filter for MC

Quote:

Woo. Pretty crazy. The numerators should only be ~10% larger than the denominator below 1 Hz. Let's try again.

 [Rana, Mirko]

I redid this calculation. The idea behind it is to get rid on any pitch that is introduced by applying longitudinal feedback to the mirrors. This coupling happens because the center of percussion for pitch , which is identical with the point where the wires lift off of the mirror, is above the center of mass.

With the same values as before, just less faulty math and Q = 2 instead of 10 we end up with the following filters:

For the lower coils (red), compared to corresponding preexisting BS filters (black):

F2aForMCcomparedToBS.pdf

The upper coils' TF is just mirrored at the 0dB magnitude axis, and have a corresponding frequency response.

I switched the F2a filters on for all MC mirrors. For convenience they are split into F2aZeros and F2aPoles. Everything seems fine. The F2a filters seem to be off for ( all ?) other mirrors.

  6021   Mon Nov 28 10:54:40 2011 ranaUpdateSUSF2A filter for MC

Our approach to making the F2A or F2P filters for the MC is to use the measured resonant frequencies and then calculating the appropriate mechanical dimensions of each suspension. This is basically because we don't have optical levers with normal incidence on these optics, but the method should be fine.

To find the formulas, I asked Gaby for her old cheat sheet: Its now in the DCC. Its only for Large optics, but you should be able to reconstruct the right ones for SOS by just changing the parameters.

  989   Thu Sep 25 02:35:21 2008 ranaSummaryPSLFAST is moving alot
It looks like the FAST signal has started moving a lot - this is partly what inspired us to tune the SLOW loop.

Some of the spiking events happen when people go on the table or the MC loses lock. But at other times it just
spikes for no apparent reason. You can also see from the first plot (9 day 10-minute trend) that there is no
great change in DTEC so we shouldn't be worried about clogging in the NPRO head.

The second plot is a 1 day minute-trend.
Attachment 1: Untitled.png
Untitled.png
Attachment 2: Untitled.png
Untitled.png
  1023   Fri Oct 3 15:09:58 2008 robUpdatePSLFAST/SLOW

Last night during locking, for no apparent reason (no common mode), the PSL FAST/SLOW loop starting going just a little
nutz. Attached is a two day plot. The noisy period started around 11-ish last night.
Attachment 1: FASTSLOW.png
FASTSLOW.png
  6639   Thu May 10 22:05:21 2012 DenUpdateCDSFB

Already for the second time today all computers loose connection to the framebuilder. When I ssh to framebuilder DAQD process was not running. I started it

controls@fb ~ 130$ sudo /sbin/init q

But I do not know what causes this problem. May be this is a memory issue. For FB

Mem:   7678472k total,  7598368k used,    80104k free

Practically all memory is used. If more is needed and swap is off, DAQD process may die.

  6640   Fri May 11 08:07:30 2012 JamieUpdateCDSFB

Quote:

Already for the second time today all computers loose connection to the framebuilder. When I ssh to framebuilder DAQD process was not running. I started it

controls@fb ~ 130$ sudo /sbin/init q

Just to be clear, "init q" does not start the framebuilder.  It just tells the init process to reparse the /etc/inittab.  And since init is supposed to be configured to restart daqd when it dies, it restarted it after the reloading of /etc/inittab.  You and Alex must have forgot to do that after you modified the inittab when you're were trying to fix daqd last week.

daqd is known to crash without reason.  It usually just goes unnoticed because init always restarts it automatically.  But we've known about this problem for a while.

Quote:

But I do not know what causes this problem. May be this is a memory issue. For FB

Mem:   7678472k total,  7598368k used,    80104k free

Practically all memory is used. If more is needed and swap is off, DAQD process may die.

This doesn't really mean anything, since the computer always ends up using all available memory.  It doesn't indicate a lack of memory.  If the machine is really running out of memory you would see lots of ugly messages in dmesg.

  13152   Mon Jul 31 15:13:24 2017 gautamUpdateCDSFB ---> FB1

[jamie, gautam]

In order to test the new daqd config that Jamie has been working on, we felt it would be most convenient for the host name "fb" (martian network IP 192.168.113.202) to point to the physical machine "fb1" (martian network IP 192.168.113.201).

I made this change in /var/lib/bind/martian.hosts on chiara, and then ran sudo service bind9 restart. It seems to have done the job. So as things stand, both hostnames "fb" and "fb1" point to 192.168.113.201.

Now, when starting up DTT or dataviewer, the NDS server is automatically found.

More details to follow.

  11076   Thu Feb 26 13:17:31 2015 ericqUpdateComputer Scripts / ProgramsFB IO load

Over the past few days, I've occasionally been peeking at the framebuilder IO load to see If I could correlate anything with it, but it's usually been low when I looked. I.e. with daqd and all models running, the %wa time was in the few percents at most.

Just now, I was seeing some EPICS sluggishness, and sure enough, the %wa was in the 50-60 range. I used iostat -xmh 5 on the framebuilder to see that /dev/sda, the /frames drive, was at 100% utilization, which means it was reading and writing as fast as it possibliy could. 

I ssh'd over to nodus, and with iotop found that an rsync job was running (rsync -am --exclude .*.gwf full 131.215.114.19::40m/full), and its IO rates corresponded very closely to the data read rates on the framebuilder from /frames. 

I killed the rsync process on nodus, and the %wa time on the framebuilder dropped to near zero. The ASS striptools, where I had noticed the sluggishness, immediately started updating faster.

While rsync is supposed to play nice with a system's IO demands, maybe it only knows about nodus's IO usage, not fb which is the underlying NFS server where the frames live. I think it would be good to throttle the bandwidth of these jobs to a specific bandwidth. 50MB/s seemed like too much, so maybe 10MB/s is ok?

  11077   Thu Feb 26 13:55:59 2015 jamieUpdateComputer Scripts / ProgramsFB IO load
We should use "ionice" to throttle the rsync. Use something like "ionice -c 3 rsync ..." to set the priority such that the rsync process will only work when there is no other IO contention. See "man ionice" for other options.
  8374   Fri Mar 29 17:24:43 2013 JamieUpdateComputersFB RAID power supply replaced

Steve ordered a replacement power supply for the FB JetStor power supply that failed a couple weeks ago.  I just installed it and it looks fine.

  12024   Sun Mar 6 15:24:05 2016 gautamUpdateCDSFB down again

I came in to check the status of the nitrogen and noticed that the striptool panels in the control room were all blank.

  • PMC was unlocked but I was able to relock it using the usual procedure
  • FB seems to be down: I was unable to ssh into it (or any of the FEs for that matter). I checked the lights on the RAID array, they are all green. I am holding off on doing a hard reboot of FB in case there is some other debugging that can be done first
  • None of the watchdogs were tripped, but judging by the green spots on the mirrors, all of them are moving quite a bit. I've shutdown the watchdogs on all the optics except the MC mirrors, but the ITMs and ETMs still seem to be moving quite a bit.

I am leaving things in this state for now. It is unclear why this should have happened, it doesn't seem like there was a power glitch?

Attachment 1: 58.png
58.png
  12025   Mon Mar 7 20:40:02 2016 ericqUpdateCDSFB down again

We went and looked at the monitor plugged into FB. All kinds of messages were being spammed to the screen (maybe RAM errors), and nothing could be done to interrupt. Sadly, a hard reboot of FB was neccesary.

Video of error messages: https://youtu.be/7rea_kokhPY

After the reboot, it just took a couple of model restarts to get the CDS screen happy.

  16294   Tue Aug 24 18:44:03 2021 KojiUpdateCDSFB is writing the frames with a year old date

Dan Kozak pointed out that the new frame files of the 40m has not been written in 2021 GPS time but 2020 GPS time.

Current GPS time is 1313890914 (or something like that), but the new files are written as C-R-1282268576-16.gwf

I don't know how this can happen but this may explain why we can't have the agreement between the FB gps time and the RTS gps time.

(dataviewer seems dependent on the FB GPS time and it indicates 2020 date. DTT/diaggui does not.)


This is the way to check the gpstime on fb1. It's apparently a year off.

controls@fb1:~ 0$ cat /proc/gps
1282269402.89

Attachment 1: Screen_Shot_2021-08-24_at_18.46.24.png
Screen_Shot_2021-08-24_at_18.46.24.png
  16298   Wed Aug 25 17:31:30 2021 PacoUpdateCDSFB is writing the frames with a year old date

[paco, tega, koji]

After invaluable assistance from Jamie in fixing this yearly offset in the gps time reported by cat /proc/gps, we managed to restart the real time system correctly (while still manually synchronizing the front end machine times). After this, we recovered the mode cleaner and were able to lock the arms with not much fuss.

Nevertheless, tega and I noticed some weird noise in the C1:LSC-TRX_OUT which was not present in the YARM transmission, and that is present even in the absence of light (we unlocked the arms and just saw it on the ndscope as shown in Attachment #1). It seems to affect the XARM and in general the lock acquisition...

We took some quick spectrum with diaggui (Attachment #2) but it doesn't look normal; there seems to be broadband excess noise with a remarkable 1 kHz component. We will probably look into it in more detail.

Attachment 1: TRX_noise_2021-08-25_17-40-55.png
TRX_noise_2021-08-25_17-40-55.png
Attachment 2: TRX_TRY_power_spectra.pdf
TRX_TRY_power_spectra.pdf
  16300   Thu Aug 26 10:10:44 2021 PacoUpdateCDSFB is writing the frames with a year old date

[paco, ]

We went over the X end to check what was going on with the TRX signal. We spotted the ground terminal coming from the QPD is loosely touching the handle of one of the computers on the rack. When we detached it completely from the rack the noise was gone (attachment 1).

We taped this terminal so it doesn't touch anything accidently. We don't know if this is the best solution since it is probably needs a stable voltage reference. In the Y end those ground terminals are connected to the same point on the rack. The other ground terminals in the X end are just cut.

We also took the PSD of these channels (attachment 2). The noise seem to be gone but TRX is still a bit noisier than TRY. Maybe we should setup a proper ground for the X arm QPD?


We saw that the X end station ALS laser was off. We turned it on and also the crystal oven and reenabled the temperature controller. Green light immidiately appeared. We are now working to restore the ALS lock. After running XARM ASS we were unable to lock the green laser so we went to the XEND and moved the piezo X ALS alignment mirrors until we maximized the transmission in the right mode. We then locked the ALS beams on both arms successfully. It very well could be that the PZT offsets were reset by the power glitch. The XARM ALS still needs some tweaking, its level is ~ 25% of what it was before the power glitch.

Attachment 1: Screenshot_from_2021-08-26_10-09-50.png
Screenshot_from_2021-08-26_10-09-50.png
Attachment 2: TRXTRY_Spectra.pdf
TRXTRY_Spectra.pdf
  9021   Sun Aug 18 16:04:07 2013 ranaSummaryCDSFB lights all RED: mxstream restart

Sun Aug 18 15:52:50 2013

Found the FB lights (C1:FEC-NN_FB_NET_STATUS and C1:DAQ-DC0_C1XXX_STATUS) RED for everything on the CDS_FE_STATUS screen.

I used the (! mxstream restart) button ro restart the mxstreams. Everything is green now.

PMC was out of lock- relocked it and the IMC locked itself as did the X & Y arms on IR. X was already green locked.

Attachment 1: IFO-Trend.png
IFO-Trend.png
  9354   Wed Nov 6 15:12:01 2013 JenneUpdateCDSFB not talking to LSC?

Something funny is going on with the framebuilder's communication with the LSC machine. 

This is a different failure mode / error than I have seen before.  It's not the type of problem that is solved by restarting the mxstreams (that is indicated by also the 2 blocks on top of one another, that are green on the lsc machine right now, being red), although I did try that, before I looked closer and realized that that wasn't the problem.

ssh-ing to c1lsc, and doing a "rtcds restart all" seems to be fixing the problem.  Both c1oaf and c1cal needed another round of restarting, because they needed their BURT buttons pressed manually.  All of the models on the lsc machine are running fine now, though.

Here's a screenshot of the CDS overview screen, with the error lights:

Screenshot-Untitled_Window-1.png