ID |
Date |
Author |
Type |
Category |
Subject |
266
|
Fri Jan 25 11:38:16 2008 |
josephb | Configuration | Cameras | Working GiGE video on Linux - sort of |
1)I have been able to compile the SampleViewer program which can stream the video from the Prosilica 750C camera. This was accomplished on my 64-bit laptop running Ubuntu, after about 3 hours of explicitly converting strings to wxStrings and back again within the C++ code. (There was probably an easier way to simply overload the functions that were being called, but I wasn't sure how to go about doing so). By connecting it to the CDS network, I was able to immediately detect the camera and display the images.
Unfortunately, I have not yet been able to get it to compile on Mafalda with the x86 architecture. This may be do the fact that it has wxWidgets version 2.8.7 while my laptop has 2.8.4. Certainly the failure at compile time looks different from the errors earlier, and seem to be within the wxWidget code rather than the SampleViewer code. I may simply need to uninstall 2.8.7 and install 2.8.4 of wxWidgets.
The modified code that will compile on my machine has been copied to /cvs/cds/caltech/target/Prosilica/examples/SampleViewer2b.
2)The Snap program (under /cvs/cds/caltech/target/Prosilica/examples/Snap) also will now take full resolution images even on Mafalda. This was achieved by reducing the packet size to 1000 and also increasing the wait until timeout time up to 400 ms, which originally was at 100. Apparently, it takes on the order of 1 ms per packet as far as I can tell. So full resolution at 752x480 required something of order 360 packets.
To Do:
1) Get sample viewer to compile on Mafalda, and then statically compile it so it can be run from any Linux based machine.
2) Get a user friendly version of Snap up and running, statically compiled, with options for a continuous loop every X seconds and also to set desired parameters (such as height, width, file name to save to, save format, etc).
3) Figure out data analysis with the images in Matlab and an after the fact image viewer.
Attached is an example .tiff image from the Snap program. |
Attachment 1: snap.tiff
|
267
|
Fri Jan 25 13:36:13 2008 |
josephb | Configuration | Cameras | Working GiGE video on Mafalda |
Finally got the GiGE camera sample viewer video running on Mafalda by updating to the latest API (version 1.16 from Dec 16, 2007) from Prosilica and then using the modified Sample Viewer code I had written. The API version previously in cvs was 1.14.
It can currently be run by ssh -X into Mafalda and going to /cvs/cds/caltech/target/Prosilica/bin-pc/x86 and running the SampleViewer executable found there. |
269
|
Fri Jan 25 17:11:07 2008 |
Max , Andrey | Configuration | General | NEW_FETCH_SHOUROV and GET_DATA do not work |
The problem which started yesterday after Andrey's framebuilder restart still persists.
It is still impossible to read data in the past from the channels using "get_data" which in turn uses "new_fetch_shourov".
Max was trying to read data from the channel
"C1:LSC-DARM_CTRL",
and he got the same error messages as Andrey.
Andrey tried earlier today to read data from "C1:SUS-ITMS_SUS" or "C1:SUS-ETMX_SUS" with the error meassge
Error in ==> new_fetch_shourov at 22
at (start_time+duration) > stops(end)
So, it seems that Robert Ward fixed just one problem out of two problems.
Robert revived the realtime signals in Dataviewer,
but did not revive the memory of channels for new_fetch_shourov.
To be more precise, channels have memory (it is possible to see the "Playback" curves in Dataviewer"),
but "get_data" and "new_fetch_shourov" do not see the data from those channels. The problem appeared immediately after Andrey's clicking on blue buttons to restart the framebuilder.
Andrey again apologizes. |
281
|
Mon Jan 28 17:16:54 2008 |
Andrey | Configuration | Computers | Matlab libraries DO NOT WORK properly sometimes |
Working in Matlab, I encountered at two different times today the license distribution problem:
??? License checkout failed.
License Manager Error -4
Maximum number of users for Curve_Fitting_Toolbox reached.
Try again later.
To see a list of current users use the lmstat utility or contact your License Administrator.
Troubleshoot this issue by visiting:
http://www.mathworks.com/support/lme4a |
288
|
Thu Jan 31 12:39:14 2008 |
John | Configuration | General | Y arm test mass cameras |
I've adjusted the test mass cameras on the y arm to make the beam injected through ETMY more visible. |
289
|
Thu Jan 31 16:53:41 2008 |
josephb | Configuration | Cameras | Improving camera user interface |
There's a new and improved version of Snap program at the moment people are free to play with.
Located in /cvs/cds/caltech/target/Prosilica/40mCode/
The program Snap now has a -h or --help option which describes some basic command line arguments. The height (in pixels), width (in pixels), exposure time (in micro seconds), file name to be saved to (in .tiff format), and packet size can all be set. The format type (i.e. pixel format such as Mono8 or Mono16) doesn't work at the moment.
At the moment, it only runs on mafalda.
Currently in the process of adding a loop option which will take images every X seconds, saving them to a given file name and then appending the time of capture to the file name.
After that need to add the ability to identify and choose the camera you want (as opposed to the first one it finds).
Lastly, I've been finding on occassion that the frame fails to save. However if you try again a few seconds later with the exact same parameters, it generally does save the second time. Not sure whats causing this, whether on the camera or network side of things.
I've attached two images, the first at default exposure time (15,000 microseconds) and the second at 1/5th that time (3,000 microseconds).
The command line used was "./Snap -E 3000 -F 'Camera_exp_3000.tiff' " |
Attachment 1: Camera_exp_15000.tiff
|
Attachment 2: Camera_exp_3000.tiff
|
292
|
Fri Feb 1 15:04:54 2008 |
josephb | Configuration | Cameras | Snap with looping functionality available |
New GiGE camera code is available in /cvs/cds/caltech/target/Prosilica/40mCode/. Currently only runs on Mafalda.
Snap has expanded functionality to continuously loop infinitely or for a maximum number of images set by the user. File names generated with the loop option have the current Unix time and .tiff appended to them. So -f './test' will produce tiff files with format "test1234567.tiff". The -l option sets the number of seconds between images.
"./Snap -l 5 -i -f './test' " will cause the program to infinitely loop, saving images every 5 seconds. Using "-m 10" instead of "-i" will take a series of 10 images every 5 seconds (so taking a total of 50 seconds to run).
It also now defaults to 16-bit (in reality only 10 bit) output instead of 8 bit output. You can select between the two with -F 'Mono8' or -F 'Mono16'.
Use --help for a full list of options.
Note that if you ctrl-c out of the loop, you may need to run ./ResetCamera 131.215.113.104 (or whatever the IP is - use ./ListCameras to determine IP if necessary) in order to reset the camera because it doesn't close out elegantly at the moment. |
297
|
Tue Feb 5 15:32:29 2008 |
josephb | Configuration | Cameras | PMC and the GigE Camera |
The PMC transmission video camera has been removed and replaced with the GigE GC750 camera for the moment.
A ND4.0 filter has been added in the path to that camera to reduce saturation for the moment.
The old camera has been placed on the elevated section inside the enclosure, and the cable for it is still on the table proper.
The Gige camera is currently running the Snap code on Linux3 with the following command line:
./Snap -E 2000 -l 60 -m 1440 -f './pmc_trans/pmc_trans'
So its going to be taking tiff images every minute for the next 24 hours into the cvs/cds/caltech/target/Prosilica/40mCode/pmc_trans/ directory.
Attached is an example image with exposure set to 2000, loaded into matlab and plotted with the surf command. 2500 microseconds looked like it was still saturating, but this seems to be a good level (with a max of 58560 out of 65535). |
Attachment 1: pmc_trans.jpg
|
|
298
|
Tue Feb 5 17:39:05 2008 |
jweiner | Configuration | PEM | PEM-AS_MIC taken down from AS table, will put in PSL enclosure soon |
I took down the microphone that Andrey hung above the AS table his first week in lab. I want to hang the microphone above the PMC to check the effect of acoustic noise on the PMC. The cables were a little more tangled than I thought so I've only taken the microphone down and haven't hung it back up yet, but on Thursday I'll have enough time to carefully put it up inside the PSL and see what I can find out about acoustic noise inside the PSL. I think the microphone should be sensitive enough for the frequencies we're interested in, and I'll hopefully find out if it's sufficient once I put it up in the PSL. The microphone cable and microphone are on top of the PSL for now. |
300
|
Wed Feb 6 16:50:47 2008 |
josephb | Configuration | Cameras | Regions of Interest and max frame rate |
The Snap code has once again been modified such that setting the -l option to 0 will take images as fast as possible. Also, the -H and -W options set the height and width, while in principle the -Y and -X options set the position in pixels of the top edge and left edge of the image. It also seems possible to set these values such that the saved image wraps around. I'll be adding some command checking so that the user can't do this in the near future.
Doing some timed runs, using a -H 350 and -W 350 (as opposed to the full 752x480), 100 images can be saved in roughly 8 seconds, and 1000 images took about 73 seconds. This corresponds to a frame rate of about 12-13 frames per second (or a 12-13 Hz display). The size of this area was sufficient to cover the current PMC transmission beam.
The command line I used was
time ./Snap -l 0 -m 1000 -f 'test' -W 350 -H 350 -Y 50 -X 350 -E 2000
Interestingly enough, there would be bursts of failed frame saves if I executed commands in another terminal (such as using ls on the directory where the files were being stored).
As always, this code is available in /cvs/cds/caltech/target/Prosilica/40mCode/. |
301
|
Wed Feb 6 19:39:11 2008 |
rana | Configuration | Cameras | Regions of Interest and max frame rate |
We really need to look into making the 40m CDS network have an all GigE backbone so that we can have cooler cameras as well as collect multiple datastreams... |
303
|
Fri Feb 8 17:55:53 2008 |
josh | Configuration | PEM | PEM-AS_MIC now in PSL enclosure |
I have moved the microphone to the PSL enclosure, hanging near the south (Y) side from a support rod for the overhanging storage area so that it's reasonably close to the PMC. I've fastened it in many places using cable ties to make sure that it won't fall.
Alberto helped me solder together a female BNC-female 3.5 mm stereo adapter so that I can use the DAQ to output through BNC to PC speakers. My plan is do sweep sine output through PC speakers to find the transfer function of sound from outside the enclosure to inside the enclosure and by moving the microphone more centrally over the PSL table, check if there are any strong resonances. Hopefully I can use this technique at other places around the interferometer or measure the effect of installing acoustic foam. |
307
|
Sun Feb 10 21:43:16 2008 |
rob | Configuration | IOO | MC alignment tweaked |
I adjusted the alignment of the free hanging mode cleaner to best transmit the PSL beam. |
309
|
Mon Feb 11 22:44:29 2008 |
rob | Configuration | DAQ | Change in channel trending on C1SUS2 |
I removed the MC2 optical lever related channels from trends, and the SRM POUT and YOUT (as these are redundant if we're also trending PERROR and YERROR). I did this because the c1susvme2 processor was having bursts of un-syncy lateness every ~15 seconds or so, and I suspected this might interfere with locking activities. This behaviour appears to have been happening for a month or so, and has been getting steadily worse. Rebooting did not fix the issue, but it appears that removing some trends has actually helped. Attached is a 50 day trend of the c1susvme2 sync-fe monitor. |
Attachment 1: srm_sync.png
|
|
311
|
Tue Feb 12 16:18:29 2008 |
rob | Configuration | Computer Scripts / Programs | autoburt cron moved to op340m |
I moved the autoburt cron job from op440m to op340m. For some reason, burtrb requires gcc to run (I gather it uses the C-preprocessor to parse input files), so I had to install that on op340m to get it to work properly.
There are no more cron jobs running on op440m now. |
345
|
Thu Feb 28 16:19:37 2008 |
josephb | Configuration | Computers | Mafalda rewired and multiple cameras running |
1) Mafalda is now connected via an orange Cat5E ethernet cord to the gigabit ethernet switch in rack in the office space. It has been labeled at both ends with "mafalda".
2) Both the GC650M camera (from MIT) and the GC750M are working. I can run the sampleviewer code and get images simultaneously. Unforutnately, the fps on both cameras seems to drop roughly in half (not an exact measurement) when displaying both simultaneously at full resolution.
3)Discovered the Gigabit ethernet card in Mafalda doesn't support jumbo packets (packets of up to 9k bytes), which is what they recommend for optimum speed.
4)However, connecting the cameras through only gigabit switches to Mafalda did seem to increase the data rate anyways, roughly by a factor of 2. (Used to take about 80 seconds to get 1000 frames saved, now takes roughly 40 seconds).
5)Need to determine the bottleneck on the cameras. It may be the ethernet card, although its possible to connect multiple gigabit cards to single computer (depending on the number of PCI slots it has). Given the ethernet cards are cheap ($300 for 20) compared to even a single camera (~$800-1500), it might be worth while outfitting a computer with multiple. |
346
|
Thu Feb 28 19:37:41 2008 |
rob | Configuration | Computers | multiple cameras running and seisBLRMS |
Quote: | 1) Mafalda is now connected via an orange Cat5E ethernet cord to the gigabit ethernet switch in rack in the office space. It has been labeled at both ends with "mafalda".
2) Both the GC650M camera (from MIT) and the GC750M are working. I can run the sampleviewer code and get images simultaneously. Unforutnately, the fps on both cameras seems to drop roughly in half (not an exact measurement) when displaying both simultaneously at full resolution.
3)Discovered the Gigabit ethernet card in Mafalda doesn't support jumbo packets (packets of up to 9k bytes), which is what they recommend for optimum speed.
4)However, connecting the cameras through only gigabit switches to Mafalda did seem to increase the data rate anyways, roughly by a factor of 2. (Used to take about 80 seconds to get 1000 frames saved, now takes roughly 40 seconds).
5)Need to determine the bottleneck on the cameras. It may be the ethernet card, although its possible to connect multiple gigabit cards to single computer (depending on the number of PCI slots it has). Given the ethernet cards are cheap ($300 for 20) compared to even a single camera (~$800-1500), it might be worth while outfitting a computer with multiple. |
I found the SampleViewer running and displaying images from the two cameras. This kept mafalda's network so busy that the seisBLRMS program fell behind by a half-hour from its nominal delay (so 45 minutes instead of 12), and was probably getting steadily further behind. I killed the SampleViewer display on linux2, and seisBLRMS is catching up. |
356
|
Tue Mar 4 19:14:09 2008 |
rana | Configuration | CDS | TDS & SVN |
Matt, Rob, Rana
Today we added the TDS software to the 40m SVN repo.
First we rationalized things by deleting all the old TDS directories and taking
the tds_mevans dir and making it be the main one (apps/linux/tds).
We also deleted all of the TDS directories in the project area. It is now very
likely that several scripts will not work. We're going to have the teething
problems of repointing everything to the nominal paths (in the apps areas).
Finally we did:svn import tds https://40m.ligo.caltech.edu/svn/40m/tds --username rana
to stick it in. To check it out do:
svn checkout https://40m.ligo.caltech.edu/svn/40m/tds --username rana
We'll get a couple of the O'Reilly SVN books as well to supplement our verion control knowledge.
Unitl then you can use the SVN cheat sheets available at:http://www.digilife.be/quickreferences/quickrefs.htm |
357
|
Tue Mar 4 20:14:02 2008 |
rana | Configuration | IOO | MC Alignment |
The MC alignment was pretty far off. We were getting TEM01 mode locks only.
Rather than inspect what happened I just aligned the MC suspensions to get
the transmission higher. Now Matt should be able to lock the X arm and collect
adaptive filter data. |
363
|
Fri Mar 7 00:47:54 2008 |
rana | Configuration | PEM | Ranger SS-1 |
Yesterday evening around 7:30 PM, I changed the Ranger seismometer from a
vertical to a horizontal seismometer. To do this I followed the instructions
in the manual.
1) Lock it down.
2) Turn it sideways. Use the leveling screws to center the bubble level.
3) Carefully loosen the hanger rod and release slowly the tension to allow
the mass to recenter.
4) Look through the little viewhole next to the rod to make the white lines
line up. This means the mass is centered.
5) Look at the output on a scope. It should be freely moving with a ~1 sec.
period.
The attached plot shows the before and after spectra. |
Attachment 1: ss1.pdf
|
|
367
|
Mon Mar 10 20:46:41 2008 |
John | Configuration | LSC | ETMY Trans PD & QPD |
I've placed a 10% reflector in the path from ETMY to the trans and quadrant photodiodes. |
368
|
Tue Mar 11 23:14:01 2008 |
rana | Configuration | PEM | Accelerometer and Seismometer movements |
Steve and Matt moved the accelerometers and seismometers today.
The accelerometers are now placed around the MC and the seismometer is in-between MC1 & MC2.
We have changed the names of the acc channels to reflect whether they are close to MC1/MC3
or MC2. We tested the accelerometer to channel name mapping by switching gains at the wilcoxon
breakout box and also by tapping. It seems now that the previous setup near the ITMX/ETMX had
some few channels mislabeled which would have given some confusing results.
Alex, Jay, and Rolf came over today and installed, then de-installed some of the hardware for
sending the PEM channels over to the C1ASS machine where the adaptive filter front end will go.
Everything should be back to the way it was...hopefully, the guys will modify the ADCU PEM
code to send the signals to the new FE over the reflective memory net and then send them to the
MCL inputs of the suspensions. So the first incarnation should use the accelerometers and seismometer
to drive MC1 and/or MC3. |
Attachment 1: Acc.pdf
|
|
369
|
Wed Mar 12 00:36:52 2008 |
rana | Configuration | PEM | Accelerometer and Seismometer movements |
I used the MISO FIR Wiener matlab code to see how well we might do in principle.
The attached 3 page PDF file shows the MC_L control signal (force on MC2) and the residual
after subtracting off the accelerometer and seismometer using a 32 Hz sample rate and
512 taps (page 1), 1024 taps (page 2), and 2048 taps (page 3). As Matt smarmily points out,
there's not a lot to win by going beyond 512; maybe a factor of sqrt(2) for a factor of 4
tap number. |
Attachment 1: finished.pdf
|
|
370
|
Wed Mar 12 00:40:35 2008 |
rana | Configuration | PEM | Accelerometer and Seismometer movements |
Same as above but with 2048 taps and a 128 Hz sample rate. Does much better at the 16 Hz bounce mode. |
Attachment 1: mc2048-128.pdf
|
|
371
|
Wed Mar 12 00:47:26 2008 |
rana | Configuration | PEM | Accelerometer and Seismometer movements |
And this is a cool snapshot showing how this operation used 16 cores on menkar ! |
Attachment 1: Screenshot.png
|
|
373
|
Thu Mar 13 02:52:06 2008 |
Lisa | Configuration | LSC | Locking with 3f |
Today we have tried to use the reflected signal demodulated at 3*f1 ~ 99 MHz (REFL31) for length control.
This signal is cool because it is generated by the beating of sidebands, so it is not very sensitive to what the carrier does inside the IFO.
In particular, its gain and the demodulation phase shouldn't change much while changing the CARM offset during the locking sequence.
The idea is therefore to use REFL31_I and REFL31_Q for controlling MICH and PRCL, with the goal of making the lock acquisition sequence more robust.
We minimized hardware changes by using the 199MHz demodulation board, changing the local oscillator to 99.586317 MHz, with an amplitude of +10 dbm (the 3f signals are therefore acquired as LSC-PD6_I and LSC-PD6-Q).
We locked both the PRM and the DRM in a stable way using the REFL31_I and REFL31_Q, after tuning the demodulation phase (50) and removing their offsets.
On the other hand, we weren't able to acquire the lock in the DRM configuration directly by using the 3f signals. We needed instead to use the f signals first, and switch to the 3f signals once the lock was already acquired, otherwise ending up locking DRM at a different working point.
One explanation for that might be the fact that the beam impinging upon the 3f diode is too big compared with the diode size (only 1 mm, half of the size of the f1 diode).
For these reason, in presence of misalignments, some of the reflected light goes in high order modes, which can be partially (or all) off the diode, thereby generating multi-zero crossing in the demodulated error signal.
The next step before making the test with the whole IFO is therefore to modify the telescope in front of the 3f diode in order to reduce the beam size and repeat the tests we did tonight in DRM configuration.
P.S.: We made a test by changing the frequency of the local oscillator by a little bit and then coming back to the original value. We observed that the phase of the signal can change, so every time this frequency is moved the 3f demod phase need to be retuned.
John, Rob, Rana, Lisa |
378
|
Fri Mar 14 12:06:29 2008 |
josephb | Configuration | Cameras | GC750 looking at ETMX while locked |
The GC750 (CMOS) is currently looking at the front of ETMX. Unfortunately, its being routed through a 10Mbit connection (which I will be purchasing a replacement for today), so getting it to send images to Mafalda/Linux 2 or 3 isn't working well, but by using a local gigabit switch and a laptop I can get sufficient speed for full images with the sample viewer.
The attached image is from a full 752x480 reslution with 10,000 microsecond exposure with the X-arm locked. Although it looks like I still need to work on the focusing. Will be switching the GC750 with the GC 650 (CCD) later today and comparing the resulting images. |
Attachment 1: ETMX_zoom_exp_10000_750.tiff
|
379
|
Fri Mar 14 14:59:51 2008 |
josephb | Configuration | Cameras | Comparison between GC650 (CCD) and GC750 (CMOS) looking at ETMX |
Attached are images taken of ETMX while locked.
The first two are 300,000 microsecond exposure time, with approximately the same focusing/zoom. (The 750 is slightly more zoomed in than the 650 in these images). The second are 30,000 microsecond exposures. The la
The CMOS appears to be more sensitive to the 1064 nm reflected light (resulting in bright images for the same exposure time). This may make a difference in applications where images are desired to be taken quickly and repeatedly.
Both seem to be resolving individual specks on the optic reasonably well.
Next test is to place both camera on a Gaussian beam (in a couple different modes say 00, 11, and so forth), probably using the PMC. |
Attachment 1: ETMX_z2_exp_300000_650.tiff
|
Attachment 2: ETMX_z2_exp_300000_750.tiff
|
Attachment 3: ETMX_z2_exp_30000_650.tiff
|
Attachment 4: ETMX_z2_exp_30000_750.tiff
|
381
|
Fri Mar 14 15:52:07 2008 |
rob | Configuration | LSC | LSC code change |
I've edited the LSC code to send different signals to the ASS box. Now, instead of the previously selected error signals deemed to be acceptable for the Alignment Sensing and Stabalization system, it sends the LSC control signals for each suspension to the ASS box (in its new incarnation as the Adaptive Susurration Subtraction system). These are the signals after the output matrix, and also after the LSC-[SUS] filter modules. |
383
|
Sun Mar 16 17:03:32 2008 |
rob | Configuration | CDS | ASS code change |
I've updated the ass.mdl file in the directory:
/cvs/cds/caltech/users/alex/cds/advLigo/src/epics/simLink/
to get us started in the adaptive PEM noise subtraction.
After several iterations of remote help from Alex, the code compiles and runs, receives signals from the LSC, PEM, and MC2, and communicates with the suspension controllers. I've also adapted the .par file from the code generator, but haven't got the testpoints working with the new ASS code. There are no MEDM screens yet, and Matt's adaptive filter code has not been installed (there's a matrix as a placeholder).
Putting in the adaptive code should be simple, building the MEDM screens tedious, and getting the testpoints working uncertain. I noticed that the new testpoint.par file starts at a different channel number than the previous (working) version, which is strange. I probably have a script somewhere to change all these numbers by a constant offset, but I don't know if that's the actual problem--maybe stuff just needs to be rebooted.
The code receives as input the first 24 channels from the PEM ADCU, the eight suspension control signals from the LSC, and the output of the MCL filter from MC2. It outputs to the MCL filter input of each suspension (except MC2). |
384
|
Mon Mar 17 18:30:48 2008 |
mevans | Configuration | PEM | Adaptive Filtering |
It seems that adaptive filtering can achieve results similar to those of the MISO FIR Wiener (entry 369). The adaptive code simulates real-time operation, but uses the same data used by Rana for the Wiener filter. I ran the adaptive filter over the data 100 times to ensure that it was well trained... maybe too well. |
Attachment 1: mcacc_adaptive.png
|
|
386
|
Thu Mar 20 16:06:27 2008 |
rob | Configuration | LSC | LSC code change |
I changed the LSC code again. I noticed that when turning off the LSC (e.g., going from LA to OFF), the cpu time would jump from ~50 to ~80, and irrevocably de-sync all the SUS controllers. This was because turning off the LSC would suddenly zero the inputs to the decimation filters that send information to the ASS box, which for some reason greatly increases the computation time of the iir filter function call. I changed the code so that these inputs are never zeroed. The ASS receives inputs from the LSC all the time now.
I also noticed that the ASS machine was running in ~2400 usec. Yes, 2,400 microseconds. I don't know how long it's been doing that, but I restarted it. Immediately after restart, it ran at 1700 microseconds. After using the "RESET" field in the adaptOnline code, that dropped to ~100 usec. Now it's not doing any adaptive filtering, as I don't know what the good settings are and no-one has been elogging their IFO work the last few days. |
390
|
Fri Mar 21 17:01:21 2008 |
rana | Configuration | DMF | Locale change on Mafalda & seisBLRMS restart |
Ever since we moved the accelerometers to be around the MC and changed their names, the seisBLRMS
has not been working. I tried to restart it today after fixing the channel names in the par file
but I ran into a PERL / UBUNTU bug.
perl: warning: Setting locale failed.
perl: warning: Please check that your locale settings:
LANGUAGE = (unset),
LC_ALL = (unset),
LC_TIME = "en_US.ISO8859-1",
LC_MONETARY = "en_US.ISO8859-1",
LC_CTYPE = "en_US.ISO8859-1",
LC_COLLATE = "en_US.ISO8859-1",
LC_MESSAGES = "C",
LC_NUMERIC = "en_US.ISO8859-1",
LANG = "en_US.UTF-8"
are supported and installed on your system.
perl: warning: Falling back to the standard locale ("C").
I don't know how this crept up or when it started. There were a bunch of fixes on the Ubuntu
forums which didn't work.
In the end I just set the 'unset' environment variables via our cshrc.40m file and this seemed
to make ligotools/perl happy. Lets hope this lasts...I love Linux. |
391
|
Fri Mar 21 23:15:11 2008 |
rana | Configuration | PEM | Ranger SS-1: New Setup |
The Ranger seismometer has been in a bad state. Its output had been sent into a SR560 without any termination.
The seismometer is, internally, just a mass on a flexure with a magnet and a pickup coil for readout.
The damping of the system depends on the resistor hooked up across the coil. With the SR560 this is
the 1 Meg input impedance of it and so the mass is undamped.
I installed a 4300 Ohm resistor in there which seems to nearly critically damp it. However, this will not
allow us to reach the ultimate quantum noise limited performance. We will have to analyze the thermal, voltage,
and current noise to get that.
I then also increased the gain from 10 to 100 on the SR560. This should now make the front end noise of the
seismometer/SR560 close to equal to the noise of the PEM ADC. |
392
|
Fri Mar 21 23:17:47 2008 |
rana | Configuration | DMF | seisBLRMS restarted |
I updated the seisBLRMS par file with the new channel names of the accelerometers and the seismometer and then
recompiled the code and restarted it according to Rob's elog entry. It went fine and the seisBLRMS is now back in
action. |
412
|
Thu Apr 3 18:46:04 2008 |
Andrey | Configuration | Computers | "Network switch board" and "c1pem1 crate" were touched |
While working with the weather station, I did two things that potentially (with a very small probability) might influence the smooth work of other processors/computers.
I did the following on Wednesday, April 2nd, in times between 1PM and 3PM.
(1) I turned off for several seconds and returned into the initial position the switch-key on the rack with computer (processor) 'c1pem1' in order to reboot processor 'c1pem1'. The turning off/on of that key-switch was repeated several times.
(2) I pulled gently the whole "Network-Switch Board" towards me in order to replace an ethernet cat 5 cable going into the board form the processor 'c1pem1'. Some other connections of other ethernet cables might be flimsy, and then other people in 40-meter might have problems with computers other than 'c1pem1'. It should not happen, but in case of extraordinary behaviour of any other computer in our lab, people should check the connectors on the network-switch board. It is located near the middle of Y-arm. See picture. |
Attachment 1: Computer_Rack.JPG
|
|
418
|
Tue Apr 8 09:08:54 2008 |
steve | Configuration | PEM | weather station disconnected |
We can not leave cables in the walkways and go on vacation.
I disconnected the weather station from the DAQ
Every Tuesday is janitor day in the 40m.
We have to give him free space for doing a good job.
Burned toast award goes to Andrey. |
Attachment 1: toast.jpg
|
|
419
|
Tue Apr 15 18:44:25 2008 |
rana | Configuration | Computers | Rosalba |
There is a new computer in the control room -- its called Rosalba,
in keeping with our naming convention. Its a quad-core machine that
Dmass found for cheap somewhere; we've installed the CentOS on it
that Alex recommended.
Its a 64-bit Linux and so that's going to cause some problems. Alex has done this
before and so we have some confidence that we can get our regular tools (DTT, Dataviewer)
to run on it.
I have made a new apps tree for all of our future 64-bit Linux machines. So far, there is
a 64-bit firefox and a 64-bit matlab in there. As we start using this machine some more, we
will be forced to install more 64-bit Linux stuff.
We also didn't have enough network cables to run to both linux3 and rosalba. Andrey has decided that we
should not ditch linux3 and so he will run another cable for it tomorrow. |
434
|
Tue Apr 22 08:34:22 2008 |
josephb | Configuration | Cameras | Current Network Diagram |
The attached network diagram has also been added to the 40m Wiki at http://lhocds.ligo-wa.caltech.edu:8000/40m/Image_Processing_with_GigE_Cameras |
Attachment 1: Network.pdf
|
|
439
|
Tue Apr 22 22:51:30 2008 |
rana | Configuration | IOO | McWFS Status |
I've been working a little on the MC WFS in the last few days. I have made many
changes to the sensing matrix script and also to the MCWFSanalyze.m script.
The output matrix, as it was, was not bad at low frequencies but was making noise in
the ~1 Hz band. Turning the gain way down made it do good things at DC and not make
things work higher.
The output matrix generating script now works after Rob fixed the XYCOM issue. Not sure
what was up there. As Caryn mentioned the SUS2.ini channels were all zero after Andrey's
PEM power cycle a few days ago. Rob booted c1susvme to get the SUS1 channels back and
today we did c1susvme2 to get the IOO-MC_L et. al. back.
Even after doing the matrix inversion there is some bad stuff in the output matrix. I
checked that the sensing matrix measurement has good coherence and I measured and set the
MC WFS RF phases (they were off by ~20-30 deg.). Still no luck.
My best guess now is that the RG filters I've used for POS damping and the movement of the
beam on the MC mirror faces has made a POS<->YAW instability at low frequencies. My next
move is to revert to velocity damping and see if things get better. Should also try redoing
the A2L on the MC1-3. |
443
|
Thu Apr 24 15:57:53 2008 |
steve | Configuration | SAFETY | Safety at AP-ISCT |
I measured the output power of the psl after the mechanical shutter.
It was 1.1 W with Ophir power meter, than unlocked the MC and measured
the power at the MC-REFL Beam Dump at the AP-ISCT 0.9 W
Power on MC-REFL photodiode 92 mW
High power metal beam shields were installed around the beam path of
MC-REFL between AP-Viewport and MC-REFL Beam Dump.
Placed HIGH POWER LASER BEAM PATH warning signs on table frame and top
covers.
Last week I placed a small monitor on the top of the OOC that
monitors the resonant spot of MC2. Please keep an eye on this monitor
when working on the AP-ISCT
AP table should NOT be left uncovered. One experienced laser operator
has to be present if the top is removed and IR-viewer scan required.
We need your full cooperation to keep this lab safe. |
Attachment 1: P1020197.JPG
|
|
Attachment 2: mcrefl3.JPG
|
|
447
|
Fri Apr 25 11:33:40 2008 |
Andrey | Configuration | Computers | Computer controlling vaccum equipment |
Old computer (located in the south-end end of the interferometer room) that was almost unable to fulfill his duties of controlling vacuum equipment has been replaced to "Linux-3". MEDM runs on "Linux-3".
We checked later that day together with Steve Vass that vacuum equipment (like vacuum valves) can be really controlled from the MEDM-screen 'VacControl.adl'.
Unused flat LCD monitor, keyboard and mouse (parts of the former LINUX-3 computer) were put on the second shelf of the computer rack in the computer room near the HP printer. |
451
|
Fri Apr 25 20:53:02 2008 |
rana | Configuration | IOO | MC WFS with more gain |
Quick update: we found that the reason for the MC WFS instability was that the digital anti-whitening was one but not the analog whitening.
We turned off the digital filters and were able to increase the gain by a factor of ~30. It is left like this, but if it hampers IFO locking then best to just turn it back down to an overall gain of 0.1 or 0.05. |
454
|
Sun Apr 27 02:11:11 2008 |
rana | Configuration | IOO | MC WFS Notes |
As noted in the elog from Friday, the WFS has been bad ever since someone switched on the digital whitening filters (FM1 & FM2)
in the MC WFS I&Q filter banks.
On Friday evening, John, Alan, and I went to the rack and verified that although the drawing shows a hookup for the whitening
filters, there is actually no such thing and so we can't have the whitening. So the anti-whitening turns on two lag filters
(2 poles at 4 Hz) and without the hardware this makes the servos unstable by adding 90 deg of phase lag at 4 Hz.
There are still several problems in this system:
- AD797 is used after the mixer. This is an unreliable, noisy part. We need to change this out
with some OP27s so that this becomes reliable and has a more reasonable noise figure.
- Hard wire the whitening filters ON. We never want these to be off. Then we can turn on the
anti-whitening. This will give us a factor of 100 better noise without filtering.
- The AD602 on the front of the whitening board has a 100 Ohm internal impedance and the
resistor between the demod board and the AD602 is 909 Ohms. This results in dividing the
signal by 10.
- The signal at the ADC is ~100 cts peak-peak. The full ADC range is, of course, 65000 cts. So
we could use a lot more gain. The mean quadrant signals are also ~100 cts so we could easily
up the analog DC gain by a factor of 30 on top of the whitening filter increase.
- The AD602 at the input and the AD620 on the output are both variable gain stages but because
of our lack of control are set to ambiguous gain levels. We should set the AD602 on the input
to its max gain of 30 dB. With the -20 dB from the x10 voltage division, this will give us
an overall gain of 3 for the puny demod signals.
|
455
|
Sun Apr 27 05:09:30 2008 |
rana | Configuration | IOO | MC WFS Whitening turned on |
I hardwired on the MC WFS whitening filters.
The MAX333A switches which choose between whitening and bypass on that board were in the bypass position
because the Xycom220 connections are not there. So the control switch gets +15V but there is no pull
down to set it to the whitened mode.
The least invasive (easiest) change I could do was to tie all of those inputs to ground. This pulls a few mA
through the pull-down resistors but is otherwise innocuous. All of these control lines come in on the A-row
of the P1 connector, so I was able to solder a single wire across all of them to ground them all.
The WFS2 board had a blown electrolytic capacitor on the -15 V line and so there was probably some extra noise
getting in that way. I couldn't find any extra SMD to replace it so I cut the legs off of a 22 uF polarized
tantalum and stuck it in there. Its even close to being the same color. I checked out the other caps, and they were all
close to 68 uF as spec'd. This one had luckily blown open and so didn't suck down the Sorensen and destroy everything.
Plugged everything back in switched the WFS servos back on. Looks good. Took before and after spectra.
In the plot:
GREEN: Open loop dark noise before changes
RED: Open loop bright (MC locked but MCWFS off)
BLUE: Closed loop, MC locked
BLACK: Dark noise after whitening
ORANGE:Closed loop after whitening
The cursor is at 16.25 Hz, the SOS bounce mode.
The I ran the new setMCWFSgains script which uses pzgain to set the UGFs of the 4 loops to 4.01 Hz.
We have in the past had problems with high WFS gains causing instabilities with the CM servo around 10-30 Hz. If this happens we should
just lower the gain by a factor of ~5. |
Attachment 1: mcnoise.png
|
|
463
|
Thu May 1 12:46:02 2008 |
josephb | Configuration | Computers | Nodus gateway is up |
The computer Nodus is now acting as a gateway machine between the GC network and the martian network in the 40m. It has the same passwords as the rana gateway machine.
Its name on the GC side is nodus (ip: 131.215.115.52) and on the martian side is nodus113 (ip: 131.215.113.200). Will need to update the hosts file on the control room machines so you can just use the name nodus113 rather than the full ip.
Software is still being added to the computer, and it will remain in parallel with the rana gateway machine until everything has been working properly for a week or so. |
466
|
Tue May 6 17:28:39 2008 |
rob | Configuration | LSC | AP33 -> POX33 |
I am in the process of switching the POX166 and AP33 photodetectors, so that they become POX33 and AP166. The IFO_CONFIGURE buttons won't work until I finish. |
467
|
Wed May 7 15:25:41 2008 |
rob | Configuration | LSC | AP33 -> POX33 |
Quote: |
I am in the process of switching the POX166 and AP33 photodetectors, so that they become POX33 and AP166. The IFO_CONFIGURE buttons won't work until I finish. |
Done. We're now in the 40m CDD configuration. |
471
|
Thu May 8 16:40:36 2008 |
josephb | Configuration | Cameras | Gige Camera currently on PSL table |
Andrey and myself were working on the PSL table today, using a pickoff of a pickoff of the main beam (adding a microscope slide to pickoff ~4% of the original pickoff) to the GC750 GigeCam.
At the time we left, we scanned the area with a beam scan and didn't see any new stray beams, and nothing in any useful beam paths should have changed. We also strung a Cat 6 cable from the control room switch out to the PSL table in the cable trays, and then above the PSL table.
Currently, its not as well aligned as it could be, and also requires a very low exposure setting, of -E 50 or so to avoid saturation. |
479
|
Thu May 15 12:05:49 2008 |
josephb | Configuration | PSL | Path to PSL Position QPD |
The 50/50 beamsplitter that was being used as the last turning mirror to the PSL Position QPD has been replaced with a Y1-1037-45-S plate. This turning mirror was also moved 4" farther along the beam path, so as to produce as small (few microwatts) transmission through the plate. The lensing optics were also shifted so as to maintain a focused beam on the photodiode. Lastly, the rotating ND filter was increased from 1.5 to 2.0 to reduce the incident power on the photodiode, since twice the power is now reaching it.
The small beam on transmission will be used by the digital cameras as a test beam. |