ID |
Date |
Author |
Type |
Category |
Subject |
3342
|
Sat Jul 31 17:37:36 2010 |
josephb | Update | CDS | Cables needed for CDS test | Last Thursday, Kiwamu and I went through the cabling necessary for a full damping test of the vertex optics controled by the sus subsytem, i.e. BS, ITMX, ITMY, PRM, SRM. The sus IO chassis is sitting in the middle of the 1X4 rack. The c1sus computer is the top 1U computer in that rack.
ADCs:
The hardest part is placing the 2x D-sub connectors to scsi on the lemo break out boxes connected to the 110Bs. The breakout boxes can be seen at the very top of the picture Kiwamu took here. These will require a minor modification to the back panel to allow the scsi cable to get out. There are two of these boxes in the new 1X5 rack. These would be connected by scsi to the ADC adapters in the back of the sus IO chassis in 1X4. The connectors are currently behind the new 1X5 rack (along with some spare ADCs/DACs/BOs.
There are 3 cables going from 40 IDC to 37 D-sub (the last 3 wires are not used and do not need to be connected, i.e. 38-40). These plug into the blue and gold ADC adapter box, the top one shown here. There is one spare connection which will remain unused for the moment. The 40 IPC ends plug into the Optical Lever PD boxes in the upper right of the new 1X4 rack (as seen in the top picture here - the boards on the right). At the back of the blue and gold adapter box is a scsi adapter which goes to the back of the IO chassis and plugs into an ADC.
In the back of the IO chassis is a 4th ADC which can be left unconnected at this point. It will eventually be plugged into the BNC breakout box for PEM signals over in the new 1X7 rack, but is unneeded for a sus test.
DACs:
There are 5 cables going from 3 SOS dewhite/anti-image boards and 2 LSC anti-image boards into 3 blue and gold DAC adapter boxes. Currently they plug into the Pentek DACs at the bottom of the new 1X4 rack. Ideally we should be able to simply unplug these from the Pentek DACs and plug them directly into the blue and gold adapter boxes. However at the time we checked, it was unclear if they would reach. So its possible new cables may need to be made (or 40 pin IDC extenders made). These boxes are then connected to the back of the IO chassis by SCSI cables. One of the DAC outputs will be left unconnected for now.
Binary Output:
The Binary output adapter boxes are plugged into the IO chassis BO cards via D-sub 37 cables. Note one has to go past the ADC/DAC adapter board in the back of IO chassis and plug directly into the Binary Output cards in the middle of the chassis. The 50 pin IDC cables should be unplugged from XY220s and plugged into the BO adapter boxes. It is unclear if these will reach.
Timing:
We have a short fiber cable (sitting on the top shelf of the new 1X3 rack) which we can plug into the master timing distribution (blue box located in the new 1X6 rack) and into the front of the SUS IO chassis. It doesn't quite make it going through all the holes at the top of the racks and through the cabling trays, so I generally only plug it in for actual tests.
The IO chassis is already plugged into the c1sus chassis with an Infiniband cable.
So in Summary to plug everything in for a SUS test requires:
- 6x SCSI cables (3 ADC, 3 DAC) (several near bottom of new 1X3 rack)
- 4x 37 D-sub to 37 D-sub connector (end connectors can be found behind new 1X5/1X6 area with the IO chassis stuff - Need to be made) (4 BO)
- 3x 40 IDC to 37 D-sub connectors (end connectors can be found behind new 1X5/1X6 area - Need to be made)(ADC)
- 5x 64 pin ribbon to 40 IDC cable (already exist, unclear if they will reach) (DAC)
- 8x 50 pin IDC ribbon (already exist, unclear if they will reach) (BO)
- 1x Double fiber from timing master to timing card
- 1x Infiniband cable (already plugged in)
Tomorrow, I will finish up a channel numbering plan I started with Kiwamu on Thursday and place it in the wiki and elog. This is for knowing which ADC/DAC/BO channel numbers correspond to which signals. Which ADCs/DACs/BOs the cables plug into matter for the actual control model, otherwise you'll be sending signals to the wrong destinations.
WARNING: The channel numbers on the front Binary Output blue and gold adapter boxes are labeled incorrectly. Channels 1-16 are really in the middle, and 17-32 are on the left when looking at the front of the box. The "To Binary IO Module" is correct. |
3343
|
Sat Jul 31 22:35:01 2010 |
Koji | Update | VAC | Vac-P1 still 1.2mtorr | I resumed the pumping from 19:00.
Now the valve RV1 is full open. But the pumping is really slow as we are using only one RP.
After 3hrs of pumping, P1 reached 1.2mmtorr but still we need 2hrs of pumping...
I stopped pumping at 22:30. |
3344
|
Sun Aug 1 18:50:35 2010 |
Koji | Update | VAC | P1 = 0.48 torr | I resumed the pumping today. Now the pressure is 0.48 torr. The pump was stopped.
Quote: |
I resumed the pumping from 19:00.
Now the valve RV1 is full open. But the pumping is really slow as we are using only one RP.
After 3hrs of pumping, P1 reached 1.2mmtorr but still we need 2hrs of pumping...
I stopped pumping at 22:30.
|
|
3348
|
Mon Aug 2 17:12:28 2010 |
Koji | Update | General | Lab schedule for the week of Aug. 2 | Aug
2 Mon - 5 Thu WFS work (Nancy)
2 Mon - 4 Wed
Jenne: Seismometer fix / Seismic measurements on the PSL table
TT characterization (with Koji)
Preparations ETM suspensions (optional: may be in later weeks)
Kiwamu: CDS test for SUS (may be involving Koji)
Alberto: RF system prep.
All: For 5th and 6th: PSL cabling works Koji
5 Thu PSL Table prep
6 Fri PSL Table prep / Likely to shut down the PSL
9 Mon PSL Table prep / shutting down of the PSL (optional)
10 Tue PSL box Frame lifting
12 Thu PSL table tapping
16 Mon - 17 Tue concrete pouring preparation
19 Thu - 23 Fri Tripod placement
24 Tue - 26 Thu concrete pouring |
Attachment 1: PSL_work_schedule.pdf
|
|
3349
|
Mon Aug 2 18:02:46 2010 |
Jenne | Update | PEM | Seismometer Update: Still not good, but perhpas getting better | Today's seismometer diagnosis activities are still underway, this is just an update (since I did some reboots):
Problem 1: X and Z channels on both seismometers were flipped. I unplugged an X cable (East/West on the cable labels) and the Z channel (vert) would go to floating ADC zero. Rana suggested that the ADCs sometimes have random channel hopping, and that a reboot of the c0dcu1 computer which handles the PEM ADCU should fix this problem. I keyed the c0dcu1 / c0daqawg crate, those computers came back just fine, and the channels were no longer flipped. This is a good thing. Although now it's actually the Z channels that were / are bad on both seismometers, not the X's.
While rebooting those computers, c1iovme, c1sosvme, c1susvme1 and c1susvme2 crashed. I rebooted them, although for the first few power cycles c1susvme1&2 couldn't mount /cvs/cds/caltech. Eventually they did, and life is good again. Except that the seismometers are still funny. |
3350
|
Mon Aug 2 21:52:57 2010 |
Koji | Update | IOO | MC is running at the full power | [Nancy and Koji]
We restored the full power operation of the MC.
Restoration of the suspensions
- Found the suspension watch dogs are left turned off.
- Found c1susvme1/2 were not running.
- Launched the realtime processes on c1susvme1/2 and c1iscey
- Restored the watch dogs. The suspensions looked fine.
Preparations for the high power
- Put an ND2.0 before the MCT CCD. Confirmed the ND reflection is damped.
MCT QPD is not necessary to be touched.
The high power operation of the MC / post lock adjustment
- Locked the MC under the autolocker being disabled.
- Adjusted the aperture on the MC2 face camera
- Adjusted the spot positions on the WFS QPDs
- Reverted the scripts to the high power ones
(mcup / mcdown / autolockMCmain40m)
- Logged in to op340m and restarted autolockMCmain40m
The autolocker seems working correctly. |
3351
|
Tue Aug 3 01:51:01 2010 |
Jenne | Update | PEM | Seismometer Update: Still not good, but perhaps getting better |
Quote: |
Today's seismometer diagnosis activities are still underway, this is just an update (since I did some reboots):
Problem 1: X and Z channels on both seismometers were flipped. I unplugged an X cable (East/West on the cable labels) and the Z channel (vert) would go to floating ADC zero. Rana suggested that the ADCs sometimes have random channel hopping, and that a reboot of the c0dcu1 computer which handles the PEM ADCU should fix this problem. I keyed the c0dcu1 / c0daqawg crate, those computers came back just fine, and the channels were no longer flipped. This is a good thing. Although now it's actually the Z channels that were / are bad on both seismometers, not the X's.
While rebooting those computers, c1iovme, c1sosvme, c1susvme1 and c1susvme2 crashed. I rebooted them, although for the first few power cycles c1susvme1&2 couldn't mount /cvs/cds/caltech. Eventually they did, and life is good again. Except that the seismometers are still funny.
|
Some more progress, but still not complete:
Jan and I looked at all of the Gur channels on a 'scope (battery, so as not to be grounded), and 5 of the 6 looked good. We were looking at the BNCs just as they go into the ADC. The one which still looks bad is Gur1Z. The 'scope just doesn't see any signal on that channel.
In addition, the ADC's BNC input #4 (which normally has Gur2Z) looks totally shot. When it's floating, the signal on dataviewer definitely doesn't look floating. I'm probably going to have to move over to another channel, and just give that one up (this ADC already has several channels which have been declared bad, so maybe it's not a surprise that this can happen?)
Since one of the Gur signals looks bad (Gur1Z) and one of the ADC channels looks bad (usual Gur2Z), I switched the Z channels on the ADC board, so the channel being saved as Gur1Z is in fact Gur2Z. This is valid as of ~1:15am until further elog notice.
During my investigations into why Gur1Z is funny, I also looked at the signal on the BNC octopus cable coming straight from the output of the Guralp Breakout Box (this is the cable which goes from "ADC Out" on the back of the box which is a 37 pin D-sub to 9 differential BNCs), and sometimes I saw zero on the 'scope, but sometimes there was a signal which would coincide with jumping tests. Whenever there was a signal however, it was always a way lower amplitude (at least by a factor of 10?) than the other channels.
All of this craziness led to me pulling the Guralp box to investigate.
Upon opening the box, I recalled that the channels go in order: Vert, NS, EW. The Gur1Z channel is that first vert channel, and it's the one which always had a blue input capacitor rather than a surface mount one. Being suspicious of Frank and Alastair, since they seemed unhappy with my capacitor choices, I wondered if they had wiggled the blue cap, and tore something loose. Just in case, and to make things seem more uniform, I replaced the blue cap with a surface mount 1uF cap. (Actually its 0.909uF, replacing the 0.905uF blue cap, according to the black DMM that measures capacitance.) While I was in there, since it had been a problem in the past (elog 2811), I relflowed the solder on some of the resistors, especially near the output op amp.
Anyhow, none of that may have been necessary. All 6 of the Gur channels were examined on a 'scope, using clip doodles to measure the various Test Points on the circuit. I looked at all of the TPs in Gur1Z, and I didn't find that any particular stage was any noisier than the others. Also, all 6 of the Gur channels seemed totally fine in terms of sending a good signal to the output of the box, including Gur1Z which is currently under investigation. All of the channels passed the "output looks ~20x the input" test, and for approximately equal thumping on the ground all 6 channels seemed to have similar amplitude outputs. The Z channels on both channels one and channels two were a little bigger than the X's or Y's, but the 2 Z channels were about the same. This test was done using Guralp2 and the Gur2 cable on both channels 1 and 2, and then checked with Guralp 1, using the Gur2 cable on channels 1. The Vert1 channel always seemed good.
I now am suspicious of one or more of the cables: either the Gur1 cable from the seismometer to the box, or the Vert1 channel of the octopus cable. I'm satisfied that the BNC cables running through the cable tray are okay (although it might not hurt to check that they all successfully send a sine wave...) I opened up the backshell of the Gur1 cable, on the end that connects to the breakout box. Nothing seemed amiss. I still need to Beep the cable to check its connections, and look at the octopus cable.
Recap / Current Status: Breakout Box is reinstalled, both seismometers hooked up. The Z channels on the seismometers are swapped at the ADC input. The dataviewer channels Gur1_X, Gur1_Y, Gur1_Z (which is actually Z of Gur2 seismometer) and Gur2_X, Gur2_Y are all good. Nancy is going to leave the MC in a happy place, and note the time when she's done. Tomorrow I'll check out the cables for the Gur1Z seismometer channel. |
3352
|
Tue Aug 3 03:15:06 2010 |
nancy | Update | IOO | MC back to locked mode | I turned the WFS gain to 0.02 back, and the MC is locked, the data for the seismic motion might be meaningful nowforth. |
3353
|
Tue Aug 3 11:17:10 2010 |
kiwamu | Update | CDS | Diagrams for Cables needed for CDS test | Current Wiring Setup for the Suspension Controls

New Wiring Plan for the Suspension Controls with the New CDS

Missing Stuff for the CDS test
Ideally we can reuse the existing cables, but some of them may not be long enough for the new wiring.
The diagram below shows extremely non-ideal case.

Some more information will be summarized on the wiki later.
|
3354
|
Tue Aug 3 11:50:16 2010 |
Jenne | Update | PEM | Seismometer Problem Tracked down | After some cable swapping this morning, I have determined which cable is bad. It's the Gur1 cable between the seismometer and the breakout box. This is a milspec -> 37pin d-sub cable. I'll pull out the cable and have a look at it after lunch. |
3356
|
Wed Aug 4 03:01:56 2010 |
Jenne | Update | PEM | Catastrophic Cable Failures |
Quote: |
After some cable swapping this morning, I have determined which cable is bad. It's the Gur1 cable between the seismometer and the breakout box. This is a milspec -> 37pin d-sub cable. I'll pull out the cable and have a look at it after lunch.
|
So, I was wrong about which cable it was, probably in my rush to get some lunch. The actual culprit was the octopus cable that Bob made waaay back in the day (~2 years ago) to go from the "ADC out" of the breakout box (37pin Male Dsub) to 9 BNCs. As it turns out, the Gur1Z channel of that cable was broken on both ends!!!
On one end, we have the 37 pin Dsub. The cable used was so thick (way too thick for this application) that it made a super rigid connection between the wires and the connector, and any bending of the cable stressed this connection, despite the strain-relief of the backshell. The Gur1Z connection snapped off when I was gently wiggling the connections to check them out. Also, since the wires were all so thick, they didn't really fit into the hole in the backshell, so 2 or 3 of them were squished.....straight through the insulation so that several channels were shorted together / potentially shorted to ground. This may explain some of the nasty behavior that Rana and I had seen (although I might have forgotten to elog? My bad.) that even with the inputs of the breakout box all terminated, there was high coherence between the channels. Terminated inputs should give random noise, so this was fishy.
On the other end of the cable we have the 9 BNCs. I had finished redoing the 37pin end of the cable, and was 'beeping' it to check it out, when to my dismay I found that the Gur1Z channels (the inside and the outer shield of the BNC connector) were shorted together. I removed these 2 wires from the Dsub connector to confirm that the BNC was at fault. Koji looked at the BNC with me after I chopped it off of the cable. Bad news strikes again. To get the wires to fit in the inner pin of the BNC connector, the cable-maker had cut off several strands of the wire to make it skinnier. It appears that over the years these cut-off strands wiggled their way to touching the outer shield. This appears to be a danger for all of the BNCs on this cable: a little bit of torque (which one might expect during plugging and unplugging a BNC) and the 2 sides of the differential measurement will be shorted together.
I then decided to start afresh and make my own cable. I found some AWG26 8-twisted-pair cable laying around underneath the Yarm (since this was all I could find, I was just going to do the Gur1 and Gur2 channels, and leave out the Gur3's). The 37 Dsub side was easy, but I seem not able to connect such skinny wire to the BNC connectors in a robust way. Since this bad cable has so far cost me ~2.5 16-hour days of grief, I don't want my new version of it to also be bad. At this point, I await the advice of one wiser than I. I think BNC connectors are designed for something a little closer to ~20AWG, but I could be wrong. Also, they are obviously optimized for coax cable. So what I have now is never going to be great. Maybe tomorrow I can go to the Electronics Shop / Store and buy BNC connectors that are meant to be soldered-to. That would be awesome.
Since I currently have no functional cable to go from Breakout box to ADC, the Guralps are unplugged for tonight.
Conclusions for the day / evening: Frank, Alastair and Jenna are mostly absolved of blame, although the traveling to Bridge and opening and closing the box (which usually involves more plugging and unplugging of cables) probably didn't help this cable out too much. Also, Bob definitely owes me a Sugar Napoleon or something.
In other news, since the Gur2Z ADC channel is totally wacked, I have taken over (but not renamed) the Ranger channel for Gur2Z for now. Jan still has the Ranger hostage over in Bridge, so this is okay for now. |
3357
|
Wed Aug 4 11:10:28 2010 |
nancy | Update | IOO | Mode Cleaner WFS | Yesterday, I started twiddling with the Mode Cleaner at about 2 pm.
So the seismic data should be all good before that.
I was using it till about 3.30 am, and then left for the night with locking it and swithcing on back the WFS control
Today morning, I have started twiddling with it again, at about 10.30 am.
About my work with the mode cleaner :
I am primarily exciting the mirrors in pitch and yaw, and trying to measure the response of the WFS and the MC2 OPLEV wrt the excitation.
This thus involves switching off the WFS control while measurement.
After two more of those measurements today, I will get to finding new values for the Output Matrix of the WFS for controlling MC1 & 3, and also, try giving in control to MC2 alignment using OPLEV signals.
|
3358
|
Wed Aug 4 12:49:42 2010 |
nancy | Update | IOO | Mode Cleaner WFS |
Quote: |
Yesterday, I started twiddling with the Mode Cleaner at about 2 pm.
So the seismic data should be all good before that.
I was using it till about 3.30 am, and then left for the night with locking it and swithcing on back the WFS control
Today morning, I have started twiddling with it again, at about 10.30 am.
About my work with the mode cleaner :
I am primarily exciting the mirrors in pitch and yaw, and trying to measure the response of the WFS and the MC2 OPLEV wrt the excitation.
This thus involves switching off the WFS control while measurement.
After two more of those measurements today, I will get to finding new values for the Output Matrix of the WFS for controlling MC1 & 3, and also, try giving in control to MC2 alignment using OPLEV signals.
|
TFs after the measurement -
In the order - MC1 , MC2 , MC3 -pitch and yaw.
These plots let us know about how do the wavefront sensor signals actually respond to the mis-alignments in the mirrors.
For legibility, legend has been includded in only one plot in each pdf., its typically the same for all 3 plots.
the actual xml files for this measurement are in the directory /cvs/cds/caltech/users/nancy/Align_Matrix/highpower/spot_center
It was made sure before each measurement that the MC is best aligned, the WFS are turned off, and the spots on all 3 QPDs are centered.
|
Attachment 1: pit.pdf
|
|
Attachment 2: yaw.pdf
|
|
Attachment 3: pit.pdf
|
|
Attachment 4: yaw.pdf
|
|
Attachment 5: pit.pdf
|
|
Attachment 6: yaw.pdf
|
|
3360
|
Wed Aug 4 16:52:59 2010 |
Razib, Aidan | Update | Phase Camera | Sideband power measurement (updated) | Aidan and I made some attempt to measure the power of the sidebands so that we can calculate our expected signal strength.
Our setup looks like the following:

As light from the laser is split into two at BS1, the transmitted beam has higher power as our BS1 is only coated for 1064nm. We get two reflected beams from BS1, one reflected of the front surface and the other from the back surface. We took the stronger back reflected beam to the EOM driven at 40 MHz (also at 25 MHz at a later time). The AOM produced a reference beam with 40 .000 005 MHz offset which we recombined with the sidebands obtained from the EOM. The beat produced is sent off to PDA 10CF connected to 4395A spectrum analyzer.
The plots for 40MHz sidebands and 25 MHz sidebands looks like this:

From the above spectra, at 40 MHz sideband regime:
Power of the carrier @ 40 MHz = -39.72 dBm
Power of the sideband @ 80 MHz = -60.39 dBm

At 25 MHz sideband regime,
Power of the carrier @ 40 MHz = -40.22 dBm
Power of the upper sideband @ 65 MHz = -61.72 dBm
Power of the lower sideband @ 15 MHz = -60.99 dBm
Power Measurement:
We made some necessary power measurement using a PD connected to a voltmeter after the EOM and the AOM when the EOM is driven at 40 MHz:
___________________________________________________________
Dark : 0.025 V
AOM on: 4.10 V (EOM blocked)
EOM : 2.425 V (AOM blocked)
___________________________________________________________
From the earlier calculation (ref: Elog entry July 28) the power that we expect to see at the PD is,
P= A_c ^2 + A_r^2 + A_(-sb)^2+ A_sb ^2 +2* A_r* A_sb * cos ( w_(r,sb) t ) , where A_c= carrier; A_r= reference beam; A_sb=Upper sideband; A_(-sb)= Lower sideband, w_(r,sb) = w_r - w_sb
P = A_c ^2 + A_r^2 + A_(-sb)^2+ A_sb ^2 +2* A_r* A_sb , letting cos (w_(r,sb) go to 1) is order to approximate the maximum signal
So the signal that we expect to see relative to the DC ( i.e A_c ^2 + A_r^2 + A_(-sb)^2+ A_sb ^2, the first four terms of the power equation) is,
Sig = 2* A_r* A_sb / { A_c ^2 + A_r^2 + A_(-sb)^2+ A_sb ^2 },
Since the modulation index is small, the power in the sideband is very small compared to carrier and the reference beam. So we can ignore the sideband power for the signal expression.
So,
Sig = 2* A_r* A_sb / ( A_c ^2 + A_r^2 )
So if we want to maximize this signal w.r.t the reference then,
d (sig)/ d(A_r) = 2 { ( A_c ^2 - A_r^2) *A_sb } / {( A_c^2 + A_r^2)} ^2
Thus, the signal is maximized when,
A_r^2 = A_c^2
We adjusted the AOM to be driven at + 7.7 dBM so that the new power at the AOM matched the EOM power, which is 2.397 in the voltmeter.
So the power at both the AOM and the EOM are:
P_AOM = ( V_AOM - V_dark) / (PD responsitivity * Transimpedance gain)
= ( 2.397 - 0.025 ) / ( 0.45 * 1.5 x 10 ^5 )
= 3.51 x 10 ^ - 5 W
P_EOM = (V_EOM - V _dark) / (PD responsitivity * Transimpedance gain)
= ( 2. 425 - .0.025) / ( 0.45 * 1.5 x 10 ^5 )
= 3.55 x 10^ - 5 W
From the spectra of the 40 MHz sideband above, the ratio of the carrier and the sideband amplitude is: A_c / A_sb = 10.8 .
P_EOM = A_c ^2 + 2 A_sb ^2
Therefore, A_sb = sqrt ( P_EOM / 118.64) = 5.47 x 10^ - 4 V/m
Thus, A_c = 5.908 x 10^ -3 V/m
and A_r = sqrt ( P_AOM) = 5.92 x 10 -3 V/m.
This measurement can be used to calculate the signal to contrast ratio (SCR) that we expect to see:
SCR = 2 A_r * A_sb / ( A_c^2 + A_r^2 ) = 0.09
Our next step is to measure the actual signal to constrast ratio as seen by the camera. Details of that will be posted soon. |
3362
|
Wed Aug 4 20:15:07 2010 |
Jenne | Update | PEM | Yay! Guralps work again! | After much hassle, the Guralp cable from the ADC Out of the breakout box to the ADC is fixed, and everything is plugged in and working again. The seismometers are back in their regular positions at the ends of the MC, ready for some excellent seis/MC combo data.
I solidified the change of putting the Gur2Z channel into a different BNC input on the ADC. The C1ADCU_PEM.ini file has been changed so that what used to be the Ranger's channel is now recognized as Gur2Z.
Also, I changed the same .ini file to reflect Koji's move of ACC_MC1_Z to the old AUDIO_MIC2 channel, so now all 6 Accelerometer channels have the same calibrations again.
Another big change is the change from old-left-handed convention to new-right-handed convention. The seismometers are aligned the same way they always have been (with the North-South markers aligned with the MC), but now the North-South output is plugged into the BNC on the ADC that is associated with Gur*_X, and the East-West output is plugged into the ADC channel associated with Gur*_Y. This is true for both Guralp Seismometers.
So, now we have:
Gur1_X = Gur1_NS = ADC#10
Gur1_Y = Gur1_EW = ADC#11
Gur1_Z = Gur1_Vert = ADC#12
Gur2_X = Gur2_NS = ADC#2
Gur2_Y = Gur2_EW = ADC#3
Gur2_Z = Gur2_Vert = ADC#24
SEIS_Ranger_Y = no longer in the .ini file |
3363
|
Wed Aug 4 20:58:22 2010 |
Gopal | Update | WIKI-40M Update | 7.28.10 - 8.4.10 Weekly Update | Summary of this week's activities:
7/28: Finished Y-Translational 4-Stack Analysis
"Tapered Cantilever" COMSOL tutorial
Tried (and failed) isolating gravity from oscillation
7/29: Developed tilt/rotation load combinations for torsional inputs and showed these to work in the model
Tried using Normal Vector mode on top plate to obtain output tilts; worked for the rectangular bar, but not for the full stack 
Talked to Jan about a 1st-order alternative to gravity - requires Weak Form (only found in COMSOL 3.5 right now)
Began Z-Translational 4-Stack Analysis -- Ran Overnight
7/30: Progress Report 1st Draft
Completed Z-Translational 4-Stack Analysis
8/1: Progress Report 2nd Draft
8/2: Progress Report 3rd Draft
Submitted Progress Report
8/3: Finalized Eigenfrequency Analysis for MC1/MC3 Stack 
24 Physical Eigenmodes plotted and recorded, as expected
Should be good enough for the final report --> focus on transfer function analysis for the remainder of the SURF
8/4: Prescribed Displacement Tests on Simple Rectangular Block --> shown to better produce displacement-displacement transfer functions
X-to-X Transfer Function seems much better when plotted
Should now be able to do the Displacement portion of Transfer Function Analysis on MC1/MC3 for Translational Modes
(I apologize that this update is a little late) |
3364
|
Thu Aug 5 00:17:41 2010 |
Koji | Update | PSL | PSL preparation work | We start the work on the cables at around the PSL table.
Aug 5th 10am-4pm?: (Kiwamu, Alberto, Koji)
- Removal of the unused cables around the PSL table and the control room
- Removal of the cable ties on the PSL frame
- Removal of the big nuts at the side of the PSL table
Aug 6th 10am-4pm?: (Kiwamu, Alberto, Koji, Jenne (~noon) )
- Labeling of the cables
- Planning of the disconnection
Aug 9th 9am-5pm: (Steve, Jenne, Alberto, Koji)
- Shutting down of the PSL
- Disconnection of the cables
- Draining of the cooling water
- Removal of the accelerometers
- Removal of the PSL chamber
- Sealing of the table with the plastic sheets |
3365
|
Thu Aug 5 01:29:39 2010 |
nancy | Update | IOO | Mode Cleaner WFS |
I calculated the MC1&3 Vs WFS1&2 Output Matrix today from the above measurements with koji's help.
the matrix can be generated from the m file at /cvs/cds/caltech/users/nancy/Align_Matrix/matrix.m
these values were put in, and the direction of control is sort of confirmed. I tried twiddling with the gains in the loop to get a 4*4 stable control, but could not succeed.
the mode cleaner is back locked now, and WFS matrix as well as gains are reverted to the old values. (1.30 am)
The output Matrices are
Pitch
0.724
|
0.197
|
0
|
0
|
-1.448
|
-0.758
|
Yaw
0.919
|
0.139
|
0
|
0
|
-0.0106
|
-0.1245
|
|
3366
|
Thu Aug 5 11:48:52 2010 |
nancy | Update | IOO | Mode Cleaner WFS |
Quote: |
I calculated the MC1&3 Vs WFS1&2 Output Matrix today from the above measurements with koji's help.
the matrix can be generated from the m file at /cvs/cds/caltech/users/nancy/Align_Matrix/matrix.m
these values were put in, and the direction of control is sort of confirmed. I tried twiddling with the gains in the loop to get a 4*4 stable control, but could not succeed.
the mode cleaner is back locked now, and WFS matrix as well as gains are reverted to the old values. (1.30 am)
The output Matrices are
Pitch
0.724
|
0.197
|
0
|
0
|
-1.448
|
-0.758
|
Yaw
0.919
|
0.139
|
0
|
0
|
-0.0106
|
-0.1245
|
|
I realised today morning that there was a flaw in my calculations for the yaw matrix.
Correcting the values, and also making teh tables more readable.
I will test these values once our computers are back to working condition.
PITCH
|
WFS1
|
WFS2
|
MC1
|
0.724
|
0.1964
|
MC2
|
0
|
0
|
MC3
|
-1.4436
|
-0.756
|
YAW
|
WFS1
|
WFS2
|
MC1
|
0.0710
|
0.1074
|
MC2
|
0
|
0
|
MC3
|
0.0082
|
-0.962
|
|
3367
|
Thu Aug 5 13:05:53 2010 |
Koji | Update | IOO | Mode Cleaner WFS | Upon Nancy's request, I checked the status of the suspensions.
I found that the power strip of the 1Y4 rack was turned off.
Since it has a over current breaker, I don't know whether it happened by someone or over current.
Anyway, I restarted the sus computers, and now the suspensions are damping as usual.
The MC has been aligned, the auto locker is also working.
Incidentally, I found that the WFS servos are not working. Actually since the last night
It repeated losing lock and unlock.
Probably some values of the matrix or the gain is wrong.
I left the WFS as it is because Nancy will put new values this afternoon.
I will ask her to confirm that the old values work at the end of her work. |
3368
|
Thu Aug 5 16:18:37 2010 |
Alastair | Update | General | clocks still not locked fully - hut removed | The rubidium clocks are still not quite locked together, though it is clear that the beat frequency has dropped a lot since yesterday.
I checked on the clocks and the 1pps sync light is on. The clocks are really hot again though despite the gap left at the bottom of the igloo. The side of the clocks were hot enough to not be touchable.
I made the executive decision that I would remove the hut just now. We can let the clocks lock together and then put the hut back on just before measuring. This way the hut will isolate from temperature fluctuations during the measurement but it won't be running at the hotter equilibrium temperature. I hope that the temperature won't change too much during the measurement if we put the hut back on.
RA: Attachment deleted because it was in Postscript format. Also not allowed here are the Stone Tablet and Cave Painting formats. |
3369
|
Thu Aug 5 17:59:23 2010 |
Koji | Update | PSL | Cable removal from the control room | [Alberto, Kiwamu, and Koji]
We removed the BNC cables from the control room.
The work was as hard as the one I had when I swept a 300m tunnel with a vacuum...
If we could remove the video cables, that would be a real epoch.
We found that the cabling behind the AP table is still quite ugly....grurrrh |
Attachment 1: IMG_2684.jpg
|
|
3370
|
Thu Aug 5 22:36:11 2010 |
Koji | Update | PSL | PSL preparation work | PSL preparation work
Aug 5th 10am-4pm?: (Kiwamu, Alberto, Koji)
Removing the unused cables around the PSL table and the control room
Aug 6th 10am-4pm?: (Kiwamu (ex. noon-2pm), Alberto, Koji, Jenne ('till noon) )
- Labeling the cables to be disconnected / making the records ==> All
- Removals
- the big nuts at the side of the PSL table ==> Steve
- the cable ties on the PSL frame ==> Easy
- Innolight 2W ==> Kiwamu
- the green pickoff optics at the edge, if necessary ==> Kiwamu talking with Steve
- Optics on the shelf ==> Jenne / Koji
- Oscilloscopes on the shelf ==> Jenne / Koji
- CCD camera connections (optional, as far as not critical for the operation)
- Put poles on the table (for the plastic sheet) ==> Alberto / Koji
Aug 9th 9am-5pm: (Steve, Jenne, Alberto, Koji)
- Disconnecting the cables ==> All
- Shutting down the PSL ==> Steve/Koji
- Draining the cooling water ==> Steve
- Removals
- The accelerometers ==> Jenne
- the PSL chamber ==> Steve
- Periscopes ==> Alberto
- Sealing of the table with the plastic sheets
- The chiller is planned to go to MIT
|
3371
|
Fri Aug 6 08:09:15 2010 |
steve | Update | PSL | PSL preparation work |
Quote: |
PSL preparation work
Aug 5th 10am-4pm?: (Kiwamu, Alberto, Koji)
Removing the unused cables around the PSL table and the control room
Aug 6th 10am-4pm?: (Kiwamu (ex. noon-2pm), Alberto, Koji, Jenne ('till noon) )
- Labeling the cables to be disconnected / making the records ==> All
- Removals
- the big nuts at the side of the PSL table ==> Steve
- the cable ties on the PSL frame ==> Easy
- Innolight 2W ==> Kiwamu
- the green pickoff optics at the edge, if necessary ==> Kiwamu talking with Steve
- Optics on the shelf ==> Jenne / Koji
- Oscilloscopes on the shelf ==> Jenne / Koji
- CCD camera connections (optional, as far as not critical for the operation)
- Put poles on the table (for the plastic sheet) ==> Alberto / Koji
Aug 9th 9am-5pm: (Steve, Jenne, Alberto, Koji)
- Disconnecting the cables ==> All
- Shutting down the PSL ==> Steve/Koji
- Draining the cooling water ==> Steve
- Removals
- The accelerometers ==> Jenne
- the PSL chamber ==> Steve
- Periscopes ==> Alberto
- Sealing of the table with the plastic sheets
- The chiller is planned to go to MIT
|
Monday, August 9
We should move the reference cavity too. Will this cavity be pumped while relocated?
Check and insure that attached and cut-free cables of PSL have enough room to tolerate the raising of the enclosure by 6"
I had second thoughts about the power line to the OMC. Koji was right, we should disconnect them from the power supplies.
The PSL enclosure doors on the north side will have to be removed some times to move exiting and entering ports.
|
3372
|
Fri Aug 6 10:44:55 2010 |
nancy, koji | Update | IOO | Mode Cleaner WFS |
Quote: |
Upon Nancy's request, I checked the status of the suspensions.
I found that the power strip of the 1Y4 rack was turned off.
Since it has a over current breaker, I don't know whether it happened by someone or over current.
Anyway, I restarted the sus computers, and now the suspensions are damping as usual.
The MC has been aligned, the auto locker is also working.
Incidentally, I found that the WFS servos are not working. Actually since the last night
It repeated losing lock and unlock.
Probably some values of the matrix or the gain is wrong.
I left the WFS as it is because Nancy will put new values this afternoon.
I will ask her to confirm that the old values work at the end of her work.
|
Yesterday , I put in the Output Matrix, and changed the gain sliders for the 4 WFS loops.
It worked and was keeping the lock for the MC.
I then tested whether the MC1 and 3 were following any change in MC2 alignment. It was indeed workinng,
Next we stepped to putting in the gains for the MC2 oplev servo.
the signs are decided on the basis of convergence, and the magnitude is kept very low, to have a very slow control for MC2.
This complete 6 * 6 model does work, and was able to keep the transmission held.
I also tried poking each mirror in pitcg and yaw, and the cavity comes back to high resonance after some time.
This time is indeed large if the poking is made for MC2, and the transmission comes back to normal after big oscillations.
I tried to measure the Open loop TFs for all these loops yesterday, but somehow could not find a correct excitation.
I will do it today.
Plan ahead :
1. Center the spot on MC2 and the QPD
2. Optimize the gains by looking at response to noise.
3. Measure Power Spectrum Density of each error signal.
|
3373
|
Fri Aug 6 12:22:04 2010 |
Jenne | Update | IOO | MC data taking over the weekend | Nancy has the Mode Cleaner for her work for the night, and is going to leave the MC happy, locked, autolocker on, WFS enabled, the works, and write down in the elog the time that she's finished. After that, I'm taking MC/seismic data all weekend long. During the weekend, if at all possible, please don't go into the IFO room, especially near the Mode Cleaner. If you do need to go into the IFO room, please elog the time you went in, and the time you left so I can correlate it with my data. This is actually important, so please stick a quick elog entry in if you even think about opening the doors to the IFO room. It is much appreciated. |
3374
|
Fri Aug 6 12:33:10 2010 |
Koji | Update | IOO | Mode Cleaner WFS |
Yesterday , I put in the Output Matrix, and changed the gain sliders for the 4 WFS loops.
|
From how much to how much have you chnged the gain?
Next we stepped to putting in the gains for the MC2 oplev servo.
|
I like to put the credit to Aidan for teaching Nancy how to use FOTON.
This complete 6 * 6 model does work, and was able to keep the transmission held.
|
This should be in this size:
This complete 6 * 6 model does work,
and was able to keep the transmission held.
Yeeeeah!
|
3375
|
Fri Aug 6 12:44:29 2010 |
Koji | Update | IOO | MC data taking over the weekend | Question:
Do you like to keep the WFS turned on?
This may change the transfer functions between the ground motion to the angle time by time,
and thus change the TF between the GND and the MCL.
Quote: |
Nancy has the Mode Cleaner for her work for the night, and is going to leave the MC happy, locked, autolocker on, WFS enabled, the works, and write down in the elog the time that she's finished. After that, I'm taking MC/seismic data all weekend long. During the weekend, if at all possible, please don't go into the IFO room, especially near the Mode Cleaner. If you do need to go into the IFO room, please elog the time you went in, and the time you left so I can correlate it with my data. This is actually important, so please stick a quick elog entry in if you even think about opening the doors to the IFO room. It is much appreciated.
|
|
3376
|
Fri Aug 6 15:50:29 2010 |
Gopal | Update | Optic Stacks | (Much Better Looking) Displacement-Displacment Transfer Functions | I reran the FDA in COMSOL on the MC1/MC3 Stack and produced the following Displacement-Displacement Transfer Functions:
X-Translational Drive has a blue background
Y-Translational Drive has a red background
Z-Translational Drive has a green background
Obtaining the Displacement-to-Phase part of the Transfer Function still produces difficulties -- I'm still working on the COMSOL-Matlab interface to perhaps better facilitate this.
RA: I have deleted those plots because they weren't transfer functions. Transfer functions must always be the ratio of something to something. For example: if I had a nickel for every bad plot I see, I would be a millionaire. In that example, the transfer function would have the units of nickels/plots. For the stacks, it should be meters/meter.
|
Attachment 1: MC1_MC3_XTrans.png
|
|
Attachment 2: MC1_MC3_YTrans.png
|
|
Attachment 3: MC1_MC3_ZTrans.png
|
|
3377
|
Fri Aug 6 16:20:08 2010 |
rana | Update | General | clocks still not locked fully - phase shifted | I tried to get the clocks to be closer to 90 deg for the relative phase by adding some cable length to one of the lines, but they are still wandering too much. We need to use the serial interface and up the gain in their 1PPS locking loops. |
Attachment 1: Untitled.pdf
|
|
3380
|
Fri Aug 6 19:46:59 2010 |
Gopal | Update | Optic Stacks | (Much Better Looking) Displacement-Displacment Transfer Functions |
Quote: |
I reran the FDA in COMSOL on the MC1/MC3 Stack and produced the following Displacement-Displacement Transfer Functions:
X-Translational Drive has a blue background
Y-Translational Drive has a red background
Z-Translational Drive has a green background
Obtaining the Displacement-to-Phase part of the Transfer Function still produces difficulties -- I'm still working on the COMSOL-Matlab interface to perhaps better facilitate this.
RA: I have deleted those plots because they weren't transfer functions. Transfer functions must always be the ratio of something to something. For example: if I had a nickel for every bad plot I see, I would be a millionaire. In that example, the transfer function would have the units of nickels/plots. For the stacks, it should be meters/meter.
|
My apologies for the mislabeled axes on my previous plots. They have been corrected to a ratio (in./in.), as Rana so kindly suggested in his helpful, not-at-all-condescending response.
I have chosen to stay in the English system because all of the original stack drawings are in inches as well. |
3381
|
Fri Aug 6 20:00:03 2010 |
Koji | Update | PSL | PSL preparation work | PSL preparation work report
Aug 6th 10am-5pm: (Steve, Jenne, Alberto, Kiwamu, Koji)
- We labeled the cables to be disconnected
- These will be disconnected in order to isolate the PSL table and the frame (housing) from the other part of the lab.
- Upon the labeling we made the list and the map of the cables to be removed.
- On Monday we disconnect those cables one by one accoding to the list.
- The following stuffs have been removed from the PSL table
- The big nuts at the side of the PSL table
- The cable ties on the PSL frame
- Innolight 2W
- The green pickoff optics at the edge
- The optics on the shelf
- The oscilloscopes on the shelf
- The OMC power supply cable was visited.
- The connections to the power supply were removed. There are two HV outputs.
- We put thick and long optical poles
- They are placed at the edge of the table so that we can put the plastic sheets on the table without touching the optics.
Plan on Monday
Aug 9th 9am-5pm: (Steve, Jenne, Alberto, Koji)
- Disconnecting the cables ==> All
- Shutting down the PSL ==> Steve/Koji
- Draining the cooling water ==> Steve
- Removals
- The accelerometers ==> Jenne
- The reference cavity chamber ==> Steve
- The small periscope ==> Alberto
- Sealing of the table with the plastic sheets
- The chiller is planned to go to MIT
|
Attachment 1: PSL_cable_map.pdf
|
|
3386
|
Mon Aug 9 12:46:24 2010 |
Nancy | Update | IOO | Mode Cleaner WFS |
Yesterday , I put in the Output Matrix, and changed the gain sliders for the 4 WFS loops.
|
From how much to how much have you chnged the gain?
I changed the gains from all 4 0.01 to o.27, 0.23, 0.32 and 0.11 and the main alignment gain to be 0.8
Next we stepped to putting in the gains for the MC2 oplev servo.
|
I like to put the credit to Aidan for teaching Nancy how to use FOTON.
Yes, I am sorry for not mentioning this.
Thanks Aidan
|
3387
|
Mon Aug 9 13:32:02 2010 |
nancy | Update | IOO | Mode Cleaner ASC | E-log entry for Friday - will attach more plots to this entry on wednesday after i am back to the 40.
Started working at some 1030 hrs and recording the Open Loop Tfs for all 6 loops.
The control was not so good, and I lost the lock quite a number of times while measureing
WFS did not converge when the spot was aligned to the center. But there was convergence to a non-center point. So if the control system was switched on near those points, it was converging to that point.
Autolocker : switches WFS control on directly, whereas the best way is to gradually increase the gain to 1. Also, the autolocker code now needs to be changed to incoporate the switing off the MC2 oplev in down and switch it on in the up script.
After Koji locked the Reference Cavity in the evening, I resumed measurements for the Open Loop TFs.
Measurement of the Open Loop Transfer Functions :
noise waveform was generated using arbitrary wf generator and injected into each loop.
An LPF was applied to have max co-relation at minimum disturabnce. (thanks to Rana)
The Transfer functions, Co-relations and Power Spectra were then measured using the DTT.
Power Spectrum of the IN1, IN2 and EXC shows clearly the suppression of the noise, and OLTF shows the phase margins.
- Courtesy Rana again for suggesting the idea of plotting power spectra of all signals in the same graph.
Later in the night , Koji worked with me and we reflected upon all TFs and changed gains whereevr required according to the phase margin considerations from the Open Loop TFs.
We used the same output matrix given in the previous e-log.
Final gains -
Alignment Gain in the WFS Master - 1.000
Loop |
Gain |
WFS1 P |
0.27 |
WFS1 Y |
0.7 |
WFS2 P |
0.15 |
WFS2 Y |
0.110 |
MC 2OPLEV P |
-0.1 |
MC2 OPLEV Y |
-0.1 |
this measurement invloved locking the MC to the correct position, with the spot centered at both the WFS and the QPD. invloved some cheating (offsets) after we tried centering w/o offsets.
demod signal was also centered while alignment.
credits to Koji for getting the correct lock position and also staying with me till late night in the lab
Important Points to be noted
1. All loops' histories have to be cleared while swtiching them on.
2. turn the loop output before the loop input so that there is no remnant history in the loop.
2. Alignment gain was gradually increased to 1. and tehn the oplevs turned on.
Later measured teh PSD of 6 error signals under 3 conditions -
New Control ON
New Control OFF
Old Control ON
Also measured the time series for the MC_trans and MC_refl for the 3 conditions.
|
Attachment 5: refl_off.pdf
|
|
3388
|
Mon Aug 9 15:54:43 2010 |
Koji | Update | IOO | Mode Cleaner ASC | The WFS and QPD servos were working. That was great.
Everything was fine except for the time series plots.
I could not get what story you are telling with the time series.
(e.g. your's are good or bad or anything) |
3389
|
Mon Aug 9 21:50:50 2010 |
nancy | Update | IOO | Mode Cleaner ASC |
Quote: |
The WFS and QPD servos were working. That was great.
Everything was fine except for the time series plots.
I could not get what story you are telling with the time series.
(e.g. your's are good or bad or anything)
|
Well, the data is kind of not enough to be analysed in time domain,
But by far from what I analyse, I think that the new control is not worse than the old one.
I donot also find any better results, except for this one being theoritically stronger. |
3390
|
Tue Aug 10 01:05:17 2010 |
Zach | Update | elog | elog restarted | Elog was down. Restarted with the script. |
3391
|
Tue Aug 10 05:56:07 2010 |
Koji | Update | PSL | PSL Table Lifting Preparation | Work on Aug 9th
Steve, Jenne, Koji, Alberto, Aidan, Jan, Sharmila, Katherine
From 9am to 6pm
- Shutting down the PSL after the 90885 hours of service
- Removals
- The accelerometers
- The reference cavity chamber
- The periscopes were left so far
- The cables between the PSL table and the outside have been disconnected
- Listed items on Friday
- Some unlisted items recorded and disconnected
- Drained the cooling water from the chiller lines
- Pulled out the chiller connections at the control room as well as the chiller control cables (temp sens & RS232C)
- Sealed the PSL table with plastic sheets
- Put antistatic films to the table (they are supported by the long optical poles)
- Used capton tapes to fix the films on to the table
- Put the white huge plastic sheet to cover the table at once
- Some spaces at the edge of the tables are left flat such that the C-clamps can be attached
- Sealed the AP and the ITMY tables by capton tapes
Some photos are attached in this entry. All of the photos found in the picasa album (click the slideshow)
|
Attachment 1: IMG_2686.jpg
|
|
Attachment 2: IMG_2691.jpg
|
|
Attachment 3: IMG_2692.jpg
|
|
Attachment 4: IMG_2711.jpg
|
|
3392
|
Tue Aug 10 15:23:35 2010 |
Jenna | Update | Electronics | Rubidium clock time constant | [Jenna & Alastair]
We changed the locking time constant on one of the Rubidium clocks using the RbMon software that came with it. We had to use the ancient Dell laptop latitudeD810 because it has a serial port built in, and we couldn't get the usb->serial adapter to work right with the clock. We tried the usb connector on more than one computer, and we had installed the right adapter and the computer seemed to recognize it fine, it just wouldn't communicate with the clock. We even tried it with the Dell latitute laptop and it still failed to work, so the only way seems to be to use the serial port directly.
The clock has a default time constant of 18.2 hours because it's designed to be locked to a GPS clock which is less stable than the Rb clock itself, so we changed it to a time constant of .57 hours. We also changed the length of the BNC cables to get the DC offset to 10mV, but then as I was typing this, we opened up data viewer to look at the real time data and saw the output suddenly leap up, and found that the offset is now -5mV mysteriously, so we went to investigate and found that the gain of the SR560 was still set to 1 from a calibration. We beat one of the clocks with a marconi for a few minutes with the gain still at this level to do another calibration, and then hooked the clocks back up together and upped the gain to 100. The DC offset is currently about 2.5mV. We're going to leave them alone for a few hours, and then check to see what the signal looks like over that period. |
3393
|
Tue Aug 10 16:55:38 2010 |
Jenna | Update | Electronics | c1iovme restarted | Alastair and I restarted the c1iovme around the time of my last elog entry (~3:20). |
3394
|
Tue Aug 10 17:21:48 2010 |
steve | Update | General | PSL optical table is at 30" height | Mike Gerfen, Koji and Steve,
Our old ( 6 x 10 x 1 ) ft Newport optical table will have the same height as the neighboring AP table when this job is finished.
Today's work was to elevate the optical table ~ 2400 lbs to 30" clearance to create a workable space below.
Atm1, enclosure frame was raised by 6"
Atm2, 2 pellet jacks and 4x4 technology in the work.
Atm3, ready for drill and tap with jig plate in place, note 4 Al legs 8" OD. 1/4" wall safely supporting the load
Atm4, showing clamps that holding 1/2" plywood with jigs
Atm5, We removed the 4 old legs: 6" OD steel, 1/4" wall and 10" high after they were unloaded. |
Attachment 1: P1060586.JPG
|
|
Attachment 2: P1060583.JPG
|
|
Attachment 3: P1060591.JPG
|
|
Attachment 4: P1060593.JPG
|
|
Attachment 5: P1060597.JPG
|
|
3396
|
Wed Aug 11 02:44:37 2010 |
Zach | Update | elog | elog restarted | You'll never guess what happened: the elog crashed!
I restarted it with the script. Yay. |
3397
|
Wed Aug 11 11:51:45 2010 |
Gopal | Update | WIKI-40M Update | 8.5.10 - 8.11.10 Weekly Update | Summary of this Week's Activities:
Thursday, August 5:
X-Displacement Transfer Function Measurement
JPL Tour
Friday, August 6:
Y-Displacement Transfer Function Measurement
Z-Displacement Transfer Function Measurement
Monday, August 9:
Worked on COMSOL/MatLab Interface --> problems may be due to older version
Discussed with Koji options for calling our COMSOL sales representative
Jan and I decided that there is in fact something wrong with the installations on both my Mac and Kallo
Reinstalled on both machines, but the problem was not solved
Jan said we'd go see Larry tomorrow
Tuesday, August 10:
Attempted to figure out Time-Dependent Modal Analysis --> don't think it's what we need
Began reading the LiveLink for MatLab documentation --> even the directions in this produced issues
Discovered "Prescribed acceleration" option for gravity:
A test with it on the simplest stack eliminated the unwanted oscillation, which I guess is a partial success... 
Trying the same thing with Koji on a simple pendulum, however, didn't produce the expected increase in resonant frequency
(Jan was unable to see Larry today, but we're meeting on Wednesday instead).
Wednesday, August 11 (morning):
Some background research on multiple-layer stack theory
Began working on presentations
|
3398
|
Wed Aug 11 12:58:56 2010 |
Jenna | Update | Electronics | Rubidium clock phase noise | I took some measurements of the clock this morning, first without the box, then with the box, then without the box again. All the noise levels look pretty much the same. When I first put the box on, it was only propped up on one side, so I think the clocks got a bit overheated and the data looks ridiculous, which is the first plot. I took it off and let them cool down a bit, and then put the box on, this time with a generous 3 inch gap at the bottom all the way around, and it seemed to be fine after that.
The calibration for the data is pi (rad) /6415 (counts) /100.
Aidan: I edited this post to change the plots from Postscripts to PDFs. |
Attachment 1: 08_11_Rb_crazybox.pdf
|
|
Attachment 2: 08_11_Rb_comparison.pdf
|
|
3399
|
Wed Aug 11 13:28:44 2010 |
josephb | Update | CDS | New cdsIPCx parts, old part breaks models | While working on a test stand at LLO, I've discovered that there's a new update to the Real Time Code Generator that breaks virtually all of our current models.
Previously, we've been using the cdsIPCx part as a flexible link between models, and could be set to RFM (reflected memory), SHMEM (shared memory on the local computer), or PCIE (pci express or dolphin I think) depending on what was written in the IPC file (found in /cvs/cds/caltech/chans/ipc/C1.ipc).
Recently, three new parts were added called cdsIPCx_SHMEM, cdsIPCx_RFM, cdsIPCx_PCIE, and the previous cdsIPCx broken. The cdsIPCx_***** has to match the type written in the IPC file or else it errors out with a rather unhelpful error message which doesn't tell you which part/line is in conflict...
i.e.
***ERROR: IPCx type mis-match: I vs. ISHM
make: *** [g1lsc] Error 255
In anycase, when using a current checkout (updated or checked out after ~July 30th), all cdsIPCx blocks in the simulink models needs to be changed to the approriate type. I'm trying to get a new CDS_PARTS.mdl file I updated into the svn, but for now you can just open up the model file directly in the /advLigoRTS/src/epics/simLink/lib directory and copy it in from there (i.e. cdsIPCx_SHMEM.mdl) to your model file. I'm also trying to get the feCodeGen.pl file changed so the error message actually tells you what channel/part is mismatching so you don't have to guess.
An easy way to modify a file for all shared memory connections without doing a ton of cutting, pasting and renaming is:
sed -i 's/"cdsIPCx"/"cdsIPCx_SHMEM"/g' file_name_here.mdl |
3400
|
Wed Aug 11 15:27:16 2010 |
Jenna | Update | Electronics | Rubidium clock phase noise | We unsynched the clocks by unhooking the 1pps locking. I've added it to the plot of the other measurements here, and we've divided by a factor of sqrt(2) in the calibration to get the phase noise from just one clock, so the calibration is now
pi (rad) /6415 (counts) /100/sqrt(2).
I've also added the noise of the clock according to SRS to the plot.
The units of this plot are rad/rt(Hz). I've no idea why it just says magnitude. |
Attachment 1: 08_11_Rb_withspec.pdf
|
|
3401
|
Wed Aug 11 16:13:59 2010 |
Jenna | Update | elog | elog restarted | The elog crashed, so we restarted it again. |
3402
|
Wed Aug 11 16:38:02 2010 |
Jenna | Update | Electronics | Rubidium clock phase noise |
Quote: |
The units of this plot are rad/rt(Hz). I've no idea why it just says magnitude.
|
This is a known thing (at least to me and Rana), so it's not just you. When you put in some points like your PD Spec, the units disappear, and I've never figured out how to get them back while keeping the points. Thanks for putting the units in your entry though. If anyone else does know how to get the units to 'stick' where they're supposed to be, that would be helpful. |
3404
|
Thu Aug 12 10:56:31 2010 |
steve | Update | Environment | dusty floor work in progres | I powered down 1Y1 and 1Y2 instrument racks of MC and RF-systems this morning.
Both racks, SP table, clean tools-parts boxes and MC2 area are also covered with plastics.
Bertin Bros. Construction, Inc is working on the floor preparation. Their head man is Robby Bertin, cell 626. 831.2264
Today's job:
remove tiles, clean off tile glue, drill for ribbed steel bars, rough clean- surface for good bounding. paint floor with bounding agent |
3405
|
Thu Aug 12 11:19:04 2010 |
kiwamu | Update | CDS | current status |
- Cables 
All the necessary cables are in our hands, such as SCSIs, 37 pin D-sub and 3 pin power cables.
With a big help from Jenne and Koji, we made 37 pin F/M cables and 3 pin power cables on the last Tuesday.
Now all the cables are connected to the machines except for the 3 pin power cables.
To install the power cables I will switch off Sorensens at new 1X5 for a minute.
- Suspension model file 
The model file named C1SUS was made and I succeeded in compiling and building it. So it is ready for the test.
But some medm screens look like not correctly complied.
For example the channel names which are listed on C1SUS_MONITOR_ADC0.adl aren't correctly assigned.
- ADCs 
All the ADC channels are working. Also the channel assignments are correct.
I checked all the ADC channels if it was working while I ran the front end realtime code C1SUS.
By injecting a signal from a function generator directly to the SCSIs, I could clearly see the numbers hopping on medm screens.
- DACs 
None of them are working although the computer can recognize that all three DAC cards are mounted.
I think something in the IOP file and the model file may be wrong because this symptom looks similar to that of C1ISCEX we tested two weeks before ( see this entry )
I have to fix it anyhow because the DACs are very necessary part for this damping test.
- Binary Outputs
They didn't work at all. Even the computer didn't recognize the binary output cards.
I think I should put some magical components on the model files.
- Conversion of the filter coefficients
The conversion of the filter coefficients has been doen yesterday.
It looks running well because I can load and unload these filters on medm screens.
I manually copied the filter coefficients from the current suspension filter file to that of new filter file.
The current file can be found under /cvs/cds/caltech/chans/, and the new one is under /cvs/cds/rtcds/caltech/c1/chans
Suspended works
- Binary Outputs
- simlink realtime model with the new CDS PARTS ( see the notice from Joe )
To be done
- let DACs work
- damping tests
|
|