ID |
Date |
Author |
Type |
Category |
Subject |
2939
|
Mon May 17 10:57:16 2010 |
steve | Configuration | SUS | damping restored to ETMYs |
ETMY-south sus damping was restored |
2943
|
Tue May 18 11:54:29 2010 |
steve | Configuration | SAFETY | psl mechanical shutter is not working |
As we learned yesterday, the PSL laser power out put mechanical shutter is not working in the remote mode. It only works in local manual mode.
Do not rely on the MEDM screen monitor readout! The position is only changing on the monitor. The main beam must be blocked before the output periscope. |
2947
|
Tue May 18 15:09:02 2010 |
steve | Configuration | SAFETY | PSL output shutter in order again |
Quote: |
As we learned yesterday, the PSL laser power out put mechanical shutter is not working in the remote mode. It only works in local manual mode.
Do not rely on the MEDM screen monitor readout! The position is only changing on the monitor. The main beam must be blocked before the output periscope.
|
Ben found the Sorenson 5V ps off. It was turned off since our last scheduled power outage. I wonder what else is running on 5V in the PSL? This power supply should be on the
"alarm handler" list to avoid future repeat of this condition. However a real safety switch would have confirming position sensors of the shutter open or closed. Is there such thing at the sites? |
2957
|
Thu May 20 12:34:46 2010 |
kiwamu | Configuration | 40m Upgrading | optical breadboards with legs |
Yesterday Steve and I revived two legs to mount some optical breadboards outside of the end table.
These legs had been used as oplev's mounts many years ago, but now they are served for 40m upgrading. These are really nice.
By putting them on the side of the end table, a mirror mounted on the top of the leg can reflect the beam outside of the end table.
Once we pick off the green beam from the end table to its outside, the green beam can propagate through the 40m walkway along the Y-arm.
So that we can measure the beam profile as it propagates.
These legs are also going to be used during mode matching of the vacuum optics. |
Attachment 1: leg1_small.png
|
|
Attachment 2: leg2_small.png
|
|
2965
|
Fri May 21 08:28:29 2010 |
steve | Configuration | VAC | TP3 turned off |
Quote: |
[Jenne, Kiwamu, and Steve via phone]
Around 9:30pm, Kiwamu and I came back from dinner, and were getting ready to begin the beam scan measurements. I noticed that one of the vacuum pumps was being very loud. Kiwamu noted that it is the fore pump for TP3's turbo, which he and Steve replaced in January (elog 2538). We had not noticed these noises before leaving for dinner, around 8pm.
We called Steve at home, and he could hear the noise through the phone. He said that even though it was really loud, since it was reading 3.3mTorr (on the display of the controller, in the vacuum rack just above head-height) which is close to the nominal value, it should be fine to leave. He will check it out in the morning. If it had been reading at or above ~1Torr, that's indicative of it being really bad, and we would have needed to shut it off.
For future reference, in case we need to turn it off, Steve said to use the following procedure:
1. Close VM3, to isolate the RGA, which is what this pump is currently (while we're at atmosphere) pumping on. I don't know if there are other things which would need to be shut at this stage, if we were at vacuum nominal.
2. Close VM5, which is right in front of TP3, so TP3's pump is just pumping on itself.
3. Push the "Stop" button on the Turbo controller for TP3, in the vacuum rack, about waist level. Turning off the turbo will also turn off the fore pump.
UPDATE, 1am: The controller in the rack is reading 3.1mTorr, so the pump, while still noisy, still seems to be working.
|
The foreline pressure of TP3 is 2.9 mTorr The drypump is loosing it's bearing and it is very noisy.
V3, V5 closed and TP3 small turbo controller off. This turned off the noisy forepump that has to be replaced.
RGA is running at cc4 2e-6 Torr
The RGA was turned off at cc4 1e-5 Torr |
2970
|
Fri May 21 16:37:35 2010 |
steve | Configuration | VAC | tp3's drypump replaced |
Bob replaced the tipseal in an other drypump and I swapped it in. TP3 turbo is running again, it's foreline pressure is 40 mTorr. The RGA is still off |
2972
|
Mon May 24 07:53:57 2010 |
steve | Configuration | PEM | air cond. just turned ON |
IFO room temp 27.5C , Please remember to turn AC back on ! |
Attachment 1: acoff.jpg
|
|
2980
|
Tue May 25 09:12:46 2010 |
kiwamu | Configuration | Green Locking | effect from air conditioner |
We should completely turn off the air conditioner when working on green locking.
Even if green beams propagates inside of chambers, the air conditioner does affect the spatial jitter of the beam.
The attached picture was taken when Steve and I were seeing how the green beam jittered.
At that time the beam was injected from the end table and going through inside of the ETM, the ITM and the BS camber.
Eventually it came out from the camber and hit the wall outside of the chamber. It was obvious, we could see the jittering when the air cond. was ON. |
Attachment 1: green_spot.png
|
|
2996
|
Wed May 26 22:22:03 2010 |
Alberto | Configuration | 40m Upgrading | Arm cavity length |
The second sideband is resonant in the arms for a cavity length of 37.9299m.
The nearest antiresonant arm lengths for f2 (55MHz) are 36.5753m and 39.2845m.
If we don't touch the ITMs, and we use the room we still have now on the end tables, we can get to 37.5m.
This is how the power spectrum at REFL would look like for perfect antiresonance:

And this is how it looks like for 37.5m:

Or, god forbid, we change the modulation frequencies... |
3015
|
Sun May 30 15:33:21 2010 |
Alberto | Configuration | IOO | mode cleaner and air conditioning |
The mode cleaner is locked and the air conditioning is full on. So the the air conditioning doesn't seem to be so important for the lock to hold. |
3016
|
Sun May 30 15:36:22 2010 |
Alberto | Configuration | PSL | IMC periscope shutter |
Two days ago I opened the PSL shutter by switching the switch on the shutter driver. That caused the shutter's switch on the medm screen to work in reversed mode: open meant closed and closed meant open.
I fixed that. Now the medm screen switch state is correct. |
3023
|
Tue Jun 1 06:30:38 2010 |
Koji | Configuration | 40m Upgrading | effect of the arm length |
I checked the effect of the arm length to the reflectance of the f2(=5*f1) sidebands.
Conclusion: If we choose L_arm = 38.4 [m], it looks sufficiently being away from the resonance
We may want to incorporate small change of the recycling cavity lengths so that we can compensate the phase deviation from -180deg.
f1 of 11.065399MHz is assumed. The carrier is assumed to be locked at the resonance.
Attachment 1: (Left) Amplitude reflectance of the arm cavity at f2 a a function of L_arm. (Right) Phase
Horizontal axis: Arm length in meter, Vertical Magnitude and Phase of the reflectance
At L=37.93 [m], f2 sidebands become resonant to the arm cavity. Otherwise, the beam will not be resonant.
Attachment 2: close-up at around 5 f1 frequency.
The phase deviation from the true anti resonance is ~0.7deg. This can be compensated by both PRC and SRC lengths. |
Attachment 1: arm_cavity1.png
|
|
Attachment 2: arm_cavity2.png
|
|
Attachment 3: arm_cavity.nb.zip
|
3081
|
Wed Jun 16 18:12:16 2010 |
nancy | Configuration | Computers | 40MARS |
i added my laptop's mac address to teh martian at port 13 today.
|
3083
|
Wed Jun 16 18:44:07 2010 |
Alberto | Configuration | Computers | 40MARS |
Quote: |
i added my laptop's mac address to teh martian at port 13 today.
|
No personal laptop is allowed to the martian network. Only access to the General Computing Side is permitted.
Please disconnect it. |
3109
|
Wed Jun 23 18:05:00 2010 |
Koji | Configuration | PSL | FSS SLOWDC should be ~-4.0 |
FSS SLOWDC slider is at around 0.
Please someone relock this at ~-4.0 to exploit some last juice of the fruit.
See this entry for the details of the operating point.
|
Attachment 1: C1PSL_FSS.png
|
|
3110
|
Wed Jun 23 23:08:30 2010 |
rana | Configuration | PSL | FSS SLOWDC should be ~-4.0 |
|
3133
|
Tue Jun 29 11:48:17 2010 |
Jenne | Configuration | SAFETY | Back in LASER HAZARD mode. |
[Steve, Kiwamu, Jenne]
The 40m is now back in Laser Hazard mode. Safety glasses are required for entry into the LVEA / IFO room. |
3135
|
Tue Jun 29 14:16:35 2010 |
Koji | Configuration | SAFETY | Back in LASER HAZARD mode. |
The insects and the laser trouble... Strange coincidences with LHO surprised me, but now I have been relieved.
Quote: |
[Steve, Kiwamu, Jenne]
The 40m is now back in Laser Hazard mode. Safety glasses are required for entry into the LVEA / IFO room.
|
|
3151
|
Wed Jun 30 23:03:46 2010 |
rana | Configuration | IOO | Power into MC restored to max |
Kiwamu, Nancy, and I restored the power into the MC today:
- Changed the 2" dia. mirror ahead of the MC REFL RFPD back to the old R=10% mirror.
- Since the MC axis has changed, we had to redo the alignment of the optics in that area. Nearly all optics had to move by 1-2 cm.
- 2 of the main mounts there had the wrong handedness (e.g. the U100-A-LH instead of RH). We rotated them to some level of reasonableness.
- Tuned the penultimate waveplate on the PSL (ahead of the PBS) to maximize the transmission to the MC and to minimize the power in the PBS reject beam.
- MC_REFL DC =1.8 V.
- Beams aligned on WFS.
- MC mirrors alignment tweaked to maximize transmission. IN the morning we will check the whole A2L centering again. If its OK, fine. Otherwise, we'll restore the bias values and align the PSL beam to the MC via the persicope.
- waveplates and PBS in the PSL were NOT removed.
- MC TRANS camera and QPD have to be recentered after we are happy with the MC axis.
- MC REFL camera has to be restored.
- WFS measurements will commence after the SURF reports are submitted.
We found many dis-assembled Allen Key sets. Do not do this! Return tools to their proper places or else you are just wasting everyone's time!
|
3152
|
Thu Jul 1 10:26:34 2010 |
steve | Configuration | VAC | pd68 at vacuum normal state |
Slow pump down _pd68 has reached the vacuum normal state. CC4_Rga region is pumped now. The RGA is still off.
Precondition to this pump down: 129 days at atm, ITMs replaced. MMT, oplev and other components were removed from BSC, ITMCs. New MMT mirrors are in. IOO_access_connector was out. The end chambers were not opened.
|
Attachment 1: pd68slow2d.jpg
|
|
Attachment 2: pem134d.jpg
|
|
3170
|
Wed Jul 7 17:18:57 2010 |
Alberto | Configuration | Electronics | Stochmon and LSC AM Stabilizer Decomissioned |
Today I disconnected and removed the Stochmon box from the 1Y2 rack.
I also removed the amplifiers that were sitting on the PSL table, next to the RF AM PD, that were connected to the Stochmon. I pulled back the RG cable and the power cables that went from the PSL table to the 1Y1 and 1Y2 racks.
The power cable, all rolled up, is now sitting on the floor, inside the 1Y1 rack and one of its end is still connected to the power of the rack. We'd like to turn off the entire rack in order to safely remove it. But since the laser driver is there too, we should do it the first time we have to turn off the rack for some other reason.
I also removed two of the AM stabilizers from the 1Y2 rack. The other one, which is currently running th MC modulations, is still in the rack, and there it is going to remain together with its distribution box.
I stored both AM stabilizers and the Stochmon box inside the RF cabinet down the East arm. |
3173
|
Wed Jul 7 22:52:38 2010 |
rana, nancy | Configuration | IOO | bad length control offset for the MC |
Rana found out that a connection was bad in the shown place, due to which the MEDM screen was showing bad offset for length control.
Basically, the offset slider value would not go into the system because of that bad connection, and was locking the mode cleaner at the wrong location. |
Attachment 1: P7070251.jpg
|
|
Attachment 2: Screenshot.png
|
|
3178
|
Thu Jul 8 15:19:27 2010 |
josephb, koji | Configuration | Computers | Added Zonet camera to IP table on linux1 |
We gave the Zonet camera the IP 192.168.113.26 and the name Zonet1.
We did this by modifying the /var/named/chroot/var/named/113.168.192.in-addr.arpa.zone and martian.zone files on linux1 as root. |
3191
|
Mon Jul 12 02:21:01 2010 |
Koji | Configuration | ASC | Resurrection of MC WFS |
I have resurrected the MC WFS on Friday night.
I have uncommented the WFS part of the MC autolocker.
The WFS total gain was empirically set to 0.1 such that the loops have no instability.
The loops somewhat worked through the weekend although they seemed to have the drift of the operating points
in accordance with the WFS spot. |
3206
|
Tue Jul 13 13:56:29 2010 |
Gopal | Configuration | Optic Stacks | Vitol Material Properties |
Viton flouroelastomers have somewhat variable material properties. The following parameters are being used for eigenfrequency analysis.
Young's Modulus: 72,500-87,000 psi (cite: http://www.row-inc.com/pfa.html) *Accurate for PFAs
Poisson's Ratio: 0.48-0.50 (cite: http://www.engineeringtoolbox.com/poissons-ratio-d_1224.html) *Accurate for rubber
Density: 1800 kg/m^3 (cite: http://physics.nist.gov/cgi-bin/Star/compos.pl?matno=275) |
3222
|
Wed Jul 14 19:00:56 2010 |
Gopal | Configuration | Optic Stacks | Vitol Material Properties, REVISED |
Quote: |
Viton flouroelastomers have somewhat variable material properties. The following parameters are being used for eigenfrequency analysis.
Young's Modulus: 72,500-87,000 psi (cite: http://www.row-inc.com/pfa.html) *Accurate for PFAs
Poisson's Ratio: 0.48-0.50 (cite: http://www.engineeringtoolbox.com/poissons-ratio-d_1224.html) *Accurate for rubber
Density: 1800 kg/m^3 (cite: http://physics.nist.gov/cgi-bin/Star/compos.pl?matno=275)
|
The Young's Modulus for PFA turned out to be orders of magnitude greater than for Viton. The revised values produced much more accurate results:
Young's Modulus for Viton-75: 1950 psi or 6.6 MPa
(Courtesy Row, Inc. and Steve Vass)
This website contains other details: http://www.row-inc.com/viton.html |
3227
|
Thu Jul 15 12:21:08 2010 |
Alberto | Configuration | LSC | PRC and SRC length adjustements |
Lately I've been trying to calculate the corrections to the recycling cavity lengths that would compensate for the phase that the sidebands will pick up from the arms in the upgraded interferometer.
To do that calculation , I tried two quite different ways, although equivalent in principle. They both use the optickle model of the 40m, but the calculation is made differently.
In the first way, I looked directly at the phases of the field: phase of [input field] / [reflected field], phase of [input field at PRM] / [transmitted field at SRM].
In the second way I looked at the demodulation phases of the LSC signals.
The first way is much simpler, especially from a computational point of view. It is the first I tried several weeks ago, but then I had abandoned because back then I thought it wasn't the correct way.
Anyway, both ways gave me the same results for the PRC length.
For the SRC length, the first way has given me a clear outcome. On the other hand, the second way has produced a less clear result.
According to these results, these would be the proposed adjustements to the cavity lengths:
dl(PRC) = -0.0266 m; dl(SRC) = 0.0612 m
I) 1st Way
a) case of arms ideal length (33.86 m)

b) case arm length = 38.40 m
PRC
zoom -> 
SRC
zoom -> 
II) 2nd Way
a) case of arms ideal length (33.86 m)

b) case arm length = 38.40 m

|
3228
|
Thu Jul 15 15:57:10 2010 |
Koji | Configuration | LSC | PRC and SRC length adjustements |
Tell me whether it is correct or not. Otherwise I won't be able to sleep tonight.
Quote: |
According to these results, these would be the proposed adjustements to the cavity lengths:
dl(PRC) = -0.0266 m; dl(SRC) = 0.612 m
|
|
3229
|
Thu Jul 15 16:16:51 2010 |
Alberto | Configuration | LSC | PRC and SRC length adjustements |
Quote: |
Tell me whether it is correct or not. Otherwise I won't be able to sleep tonight.
Quote: |
According to these results, these would be the proposed adjustements to the cavity lengths:
dl(PRC) = -0.0266 m; dl(SRC) = 0.612 m
|
|
Sorry. I was in a rush to go to the LIGO "all hands" meetings when I posted that elog entry, that I forgot a zero in the SRC length value. The correct values are:
dl(PRC) = -0.0266 m; dl(SRC) = 0.0612 m
The cavity absolute lengths are then:
L(PRC) = 0.5/2/f1*c - 0.0266 = 6.7466 m
L(SRC) = c/f2 + 0.0612 = 5.4798 m
where c is the speed of light; f1 = 11065399 Hz; f2 = 55326995 Hz |
3239
|
Fri Jul 16 16:12:31 2010 |
Alberto | Configuration | Computers | c1susvme1/2 rebooted |
Today I noticed that the FE SYNC counters of c1susvme1/2 on the RFM network screen were stuck at 16384. I tried to reboot the machines to fix the problem but it didn't work.
The BS watchdog tripped off when I did that, because I had forgotten to disable it. I had to wait for a few minutes before it settled down again.
Later I also re-locked the mode cleaner. But before I could do it, Rana had to reduce the MC_L offset for me. |
3242
|
Sun Jul 18 20:50:03 2010 |
rana | Configuration | IOO | MC TRANS optics changed |
To make the beam on the MC trans camera bigger, I removed the lens + ND filter that was in that path.
The camera was getting the transmission through a BS1-33 (33% reflector). The reflection went to the TRANS QPD. I changed
the R=33% into an HR mirror (Y1-45P) so now the camera has a nice beam. The QPD was now saturating so I put a ND06 into that path
so now the TRANS_SUM is ~4.5-5 V when the MC is aligned.
The MC was also misaligned and failing to lock all weekend (why??) so I aligned the MC mirrors to get it to acquire again. Since we want to
collect MC seismic data, please make sure the MC is locked and running after finishing with your various MC or PSL work (this means YOU). |
3248
|
Tue Jul 20 08:04:31 2010 |
steve | Configuration | VAC | slow vent has started |
The PSL shutter was closed. The beam path blocked two places. High voltage power supplies to IOO and OMC PZT were checked to be off. Oplevs are off.
The south arm green cavity was misaligned yesterday
We would like to keep the vent speed at 1 torr / min. I'm venting with N2 now up to 25 PSI. We have 3 cylinder of instrument grade air inside the lab. Additional supply will arrive later. It can be as late as 1pm |
3252
|
Tue Jul 20 17:38:16 2010 |
Gopal | Configuration | Optic Stacks | Stack Type Clarifications |
Some clarification is warranted regarding the different shapes of stacks. Corrections are appreciated:
1) The single-leg stack that I just completed should function as a working model for the IO, OO, and MC1/3. Rana commented, however, that the current dimensions are slightly off for MC1/3 (which makes sense since I could only find drawings for the IOC). If anyone knows the whereabouts of similar drawings for MC1/3, I'd much appreciate it.
2) A triple-leg stack can model the BS, ITMX, and ITMY chambers. I don't have exact dimensions for these, but I can make decent approximations from to-scale stack drawings. I'll probably work on a model for this style next, since at least I have some information regarding this version.
3) The MC2 chamber has its own stack model, about which I haven't found any drawings in the binders. I can't start on MC2C at all until I find such drawings. |
3261
|
Wed Jul 21 17:41:17 2010 |
Gopal | Configuration | Optic Stacks | Pictures of Stacks |
Now that venting is complete, this is a request for anyone who opens any chamber:
1) Please notify me immediately so I can take pictures of the stacks in that chamber.
2) If I'm not around, please take a few pictures for me. I'm most interested in the shape, number of layers, size, and damper arrangements of each stack.
This is most important for the MC1/MC3 chamber, MC2 chamber, and BS/ITMX/ITMY chambers.
Thanks! |
3331
|
Fri Jul 30 09:55:34 2010 |
steve | Configuration | VAC | slow pumpdown has started |
Bob and Steve closed BS chamber with the help of the manual Genie lift and the pump down started. The PSL shutter was closed and manual block was placed in the beam path. High voltage power supplies were checked to be off.
Pumping speed ~ 1 Torr/min was achieved at 1/8 of a turn opened roughing valve RV1 |
Attachment 1: slowpd5hr.jpg
|
|
3332
|
Fri Jul 30 12:46:38 2010 |
Koji | Configuration | 40m Upgrading | BS chamber status |
|
Attachment 1: BS_layout_Jul_29_2010.png
|
|
3335
|
Fri Jul 30 17:52:12 2010 |
Koji | Configuration | General | Jacket |
Jenne and I need our jackets. We removed them from the Rb clock.
Thanks for making them warm. Probably a Scotland sweater would fit. |
3337
|
Fri Jul 30 19:20:15 2010 |
Alastair | Configuration | General | Jacket |
Quote: |
Jenne and I need our jackets. We removed them from the Rb clock.
Thanks for making them warm. Probably a Scotland sweater would fit.
|
Thanks for the loan guys. I do have a lot of warm weather clothing at home that is not so necessary in the California climate. I will find some suitable attire for the Rb clocks there. |
3347
|
Mon Aug 2 10:26:15 2010 |
steve | Configuration | VAC | very slow pumpdown completed |
Quote: |
Bob and Steve closed BS chamber with the help of the manual Genie lift and the pump down started. The PSL shutter was closed and manual block was placed in the beam path. High voltage power supplies were checked to be off.
Pumping speed ~ 1 Torr/min was achieved at 1/8 of a turn opened roughing valve RV1
|
3 days pump down #69 completed. Thanks to Koji and Kiwamu who did more roughing on Saturday and Sunday. |
Attachment 1: veryslowpd.jpg
|
|
3355
|
Wed Aug 4 01:44:56 2010 |
Koji | Configuration | PEM | Accelerometer located on and below the PSL table |
[Jenne and Koji]
In order to characterize the seismic vibration of the PSL table, we put the accelerometers on and below the PSL table.
On the PSL: MC2 Accelerometers (X, Y, Z) - being connected to CH1-3 of the preamps
East side of the PSL table. X, Y, and Z is directed to North, East, and Up.
On the ground: MC1 Accelerometers (X, Y, Z) - being connected to CH4-6 of the preamps
Beneath the West side of the PSL table. X, Y, and Z is co-aligned to the MC2 ACC.
I found that the C1:PEM-ACC_MC1_Z has large noise in the low freq (~1Hz) region. I tracked down the noise source
and found the noise is still present in the down stream even when the CH17 (C1:PEM-ACC_MC1_Z ) of the ADC IF BNC
(@1Y7) was terminated.
I consulted with Jenne and decided to connect this channel to CH14, which is vacant and has name
C1:PEM-AUDIO_MIC2 (16k).
(for the details of the channel configurations, see /cvs/cds/caltech/chans/daq/C1ADCU_PEM.ini )
I first tried magnetometer channels to steal, but they didn't seem reacting (and the connected to the wrong channels).
I am feeling that we should once entirely check the I/F box.
Note that there looks the difference of the gain by x10 between C1:PEM-ACC_MC* channels and C1:PEM-AUDIO_MIC2 .
[From Jenne: The gain difference is because the C1:PEM-ACC_MC* channels have gain=10 in the .ini file, while C1:PEM-AUDIO_MIC2 uses the default gain=1. ] |
Attachment 1: IMG_2675.jpg
|
|
Attachment 2: IMG_2678.jpg
|
|
Attachment 3: IMG_2682.jpg
|
|
3361
|
Wed Aug 4 19:50:58 2010 |
rana | Configuration | Electronics | Rubidium clocks too hot: hut removed |
Alastair found that the foam hut that he and Jan put on top of the Rb clocks to temperature stabilize them was too good of an insulator. The Rb boxes had gotten very hot and became internally unlocked as seen on the front panel.
After we let them cool down with the box off, I turned them back on. After several minutes the 'Locked' light came back on. Some minutes after that the '1PPS Sync' light also came on, indicating that the two had become somewhat synchronized. It really means that the frequencies are kind of close: I think its roughly that f1-f2 < 2 mHz.
I put the yellow box back on and have left it with a small gap on the bottom so that the hot air can get out. Hopefully, this will protect the clocks from the wind, but not cause them to overheat.
The signal going to the DAQ right now is DC-coupled, with a gain of 1. The peak-peak beat signal in this situation is 6300 counts.
My guess is that the clocks will by synchronized by tomorrow afternoon so that we can get the measurement done. Please don't disturb the clocks or the yellow box around them. Try to minimize any activity around that area. |
Attachment 1: beat.png
|
|
3383
|
Sat Aug 7 11:07:44 2010 |
Koji | Configuration | PSL | MC kept locked / PMC control gain reduced to +13dB |
Jenne asked us to keep th MC locked and let the seismometers happy through this weekend.
Note that the work at the control room and the desks are no problem as far as you are quiet.
Nancy told Jenne and me that she finished the work and reverted the WFS to the old state at 4:30AM.
She could not make the elog as it has been crashed.
MC and old MC WFS looks working as usual.
From 6:40AM to 9:40AM the oscillation of the PMC looks present.
At 10:30AM I reduced the gain of the PMC from +15dB to +13dB. |
3384
|
Sat Aug 7 21:09:59 2010 |
Dmass | Configuration | Computer Scripts / Programs | eLog changes |
I made some changes to the elog on nodus:
- Made a backup of /cvs/cds/caltech/elog/elog-2.7.5/elogd.cfg called elogd.cfg.bk.20100407 in the same directory
- Added a folder: /cvs/cds/caltech/elog/elog-2.7.5/logbooks/EAGER_Lab
- Restarted the elog daemon via the start-elog-nodus script in the elog-2.7.5 directory
I saw that the current version of the elog seems to be in the svn, so tried to svn the changes from nodus via ssh, but got this message:
"svn: This client is too old to work with working copy '/cvs/cds/caltech/elog/elog-2.7.5'; please get a newer Subversion client."
I feel I should svn this but don't want to *&#@ the svn/elog up.
For now I will leave it alone and ask a question: Is the folder /cvs/cds/caltech/elog/elog-2.7.5/ under SVN control? Is it also under CVS control?
TL;DR: New tab added to elog.
|
3433
|
Wed Aug 18 12:02:29 2010 |
Alastair | Configuration | Computers | elog had crashed again... |
...I restarted it. |
3438
|
Wed Aug 18 20:54:23 2010 |
rana, yoichi | Configuration | Environment | Chiller's chiller turned off |
WE programmed the MOPA chiller's AC unit to turn off at 6PM each day and automatically turn on at 7 AM each morning. Its now very quiet in the control room at night.
Next, we'll leave it off during the day and see if it makes Steve go crazy or not. |
3456
|
Mon Aug 23 15:24:24 2010 |
kiwamu | Configuration | SUS | watchdogs off |
For the new CDS test, I turned off the watchdogs for PRM, SRM, BS, ITMs and MCs.
I will restore these watchdogs after several hours from now.
|
3457
|
Mon Aug 23 18:18:22 2010 |
kiwamu | Configuration | SUS | Re:watchdogs off |
Now watchdogs are back.
The suspensions are well damped. |
3496
|
Tue Aug 31 14:47:53 2010 |
steve | Configuration | PEM | toilet upgrade |
Our toilets were replaced by a green Gerber 1.6 gpf ( 6 lpf ) last week. |
Attachment 1: toil-n.JPG
|
|
3501
|
Wed Sep 1 07:52:27 2010 |
Alberto | Configuration | Electronics | PMC board unplugged, turned on Sorensen switches on 1Y1 rack |
Today I put the FSS frequency box back into the 1Y1 rack.
To power it on, I turned on the 24V and 15V Sorensen switches in the same rack.
The PMC crystal board in the same rack should not be affected (it runs with 10V), but, to make sure it was not powered, I disconnected it from its crate. Since the board was disconnected from the EOM for the PSL table's upgrade, I wanted to avoid having the RF output floating.
We just have to remember to plug it back in, when we need it again. |
3504
|
Wed Sep 1 08:40:28 2010 |
Alberto | Configuration | Electronics | PMC board unplugged, turned on Sorensen switches on 1Y1 rack |
Quote: |
Today I put the FSS frequency box back into the 1Y1 rack.
To power it on, I turned on the 24V and 15V Sorensen switches in the same rack.
The PMC crystal board in the same rack should not be affected (it runs with 10V), but, to make sure it was not powered, I disconnected it from its crate. Since the board was disconnected from the EOM for the PSL table's upgrade, I wanted to avoid having the RF output floating.
We just have to remember to plug it back in, when we need it again.
|
I just turned on the other Sorensen's too in 1Y1. |