ID |
Date |
Author |
Type |
Category |
Subject |
16929
|
Fri Jun 17 16:22:21 2022 |
yuta | Update | LSC | Actuator calibration of BS. ITMX, ITMY, updated MICH displacement spectra from c1cal |
Following what we have done in 2013 (40m/8242), actuator calibration was done using MICH.
AS55_Q in MICH : 9.74e8 counts/m
BS : 26.08e-9 /f^2 m/counts
ITMX : 5.29e-9 /f^2 m/counts
ITMY : 4.74e-9 /f^2 m/counts
Optical gain is 25% lower than the measurement in June 6 (40m/16892), probably because our estimate was too rough then and also we now have ~15% lower IMC transmission.
Actuator gains are 2-30% higher than the measurement in 2013.
MICH error signal calibration:
C1:LSC-AS55_Q_ERR was calibrated by taking data with C1:LSC-ASDC_OUT, when Michelson was aligned and free swinging (Attachment #1).
AS55_Q and ASDC were X-Y plotted and fitted with ellipse to get an amplitude of AS55_Q to be 82.51 counts (Attachment #2).
4*pi*A/lambda gives you 9.74e8 counts/m, where meters are in terms of difference between BS to ITMX length and BS to ITMY length.
Jupyter notebook: https://git.ligo.org/40m/scripts/-/blob/main/CAL/MICH/MICHOpticalGainCalibration.ipynb
Openloop transfer function for actuator calibration:
C1:LSC-MICH_GAIN was lowered to -1 (instead of -6), and some of filters are turned off to make the MICH UGF to be ~10.
Also, ellip("LowPass",4,1,40,50) was added to C1:LSC-MICH_A filter bank to cut the feedback above 50 Hz, so that the loop does not suppress the measurement.
The configuration is in Attachment #3.
Actuator calibration of BS, ITMX, ITMY:
With this MICH OLG, transfer functions from C1:LSC-BS,ITMX,ITMY_EXC to C1:LSC-AS55_Q_ERR were measured.
AS55_Q was calibrated to meters using the calibration factor above, and fitted the transfer function with 1/f^2 in 70-150 Hz range to get the actuator efficiency mentioned above (Attachement #4).
Thus, meters in this calibration is in terms of ITM POS motion (not in BS POS motion).
Jupyter notebook: https://git.ligo.org/40m/scripts/-/blob/main/CAL/MICH/MICHActuatorCalibration.ipynb
MICH displacement noise:
Measured values were added to c1cal model as follows.
C1:CAL-MICH_CINV FM2: 1/9.74e8 = 1.03e-9
C1:CAL-MICH_A FM2: 2.608e-8 (it was 2.07e-8 from 2013!)
C1:CAL-MICH_A_GAIN = 0.5 to take into account of C1:LSC-OUTPUT_MTRX_8_2=0.5 in the LSC output matrix for BS
Spectrum of C1:CAL-MICH_W_OUT (now calibrated in nm) with configuration in Attachment #5 was taken.
Attachement #6 is the result. I also took the spectrum with PSL shutter off to measure the sensing noise. The sensing noise limits our sensitivity above ~40 Hz at 5e-11 m/rtHz. |
Attachment 1: MICHOpticalGainCalibrationFig1.png
|
|
Attachment 2: MICHOpticalGainCalibrationFig2.png
|
|
Attachment 3: Screenshot_2022-06-17_14-23-04_MICHOLTF_ActuatorCalibration.png
|
|
Attachment 4: MICHActuatorCalibration.png
|
|
Attachment 5: Screenshot_2022-06-17_15-54-41_MICHCalibrationFilters.png
|
|
Attachment 6: Screenshot_2022-06-17_15-53-41_MICHDisplacement.png
|
|
16977
|
Thu Jul 7 18:18:19 2022 |
yuta | Update | LSC | Actuator calibration of ETMX and ETMX |
(This is a complete restore of elog 40m/16970 from July 5, 2022 at 14:34)
ETMX and ETMY actuators were calibrated using single arm lock by taking the actuation efficiency ratio between ITMs. Below is the result.
ETMX : 2.65e-9 /f^2 m/counts (0.5007 times ITMX)
ETMY : 10.91e-9 /f^2 m/counts (2.3017 times ITMY)
Motivation:
- ETMX and ETMY actuators seemed to be unbalanced when locking DARM (see 40m/16968)
What we did:
- Reverted to C1:LSC-ETMX_GAIN = 1
- XARM was locked using POX11_I_ERR (42dB whitening gain, 132.95 deg for demod phase) with ETMX and C1:LSC-XARM_GAIN=0.06
- YARM was locked using POY11_I_ERR (18dB whitening gain, -66.00 deg for demod phase) with ETMX and C1:LSC-YARM_GAIN=0.02
- OLTFs for each was measured to be Attachment #1; UGF was ~180 Hz for XARM, ~200 Hz for YARM.
- Measured TF from C1:LSC-(E|I)TM(X|Y)_EXC to C1:LSC-(X|Y)ARM_IN1 (see Attachment #2)
- Took the ratio between ITM actuation and ETM actuation to calculate ETM actuation. For ITM actuation, we used the value measured using MICH (see 40m/16929). The average of the ratio in the frequency range 70-150 Hz was used.
Files:
- Measurement files live in https://git.ligo.org/40m/measurements/-/tree/main/LSC/XARM and YARM
- Script for calculation lives at https://git.ligo.org/40m/scripts/-/blob/main/CAL/ARM/ETMActuatorCalibration.ipynb
Discussion:
- ETMX actuation is 4.12 times less compared with ETMY. This is more or less consistent with what we measured in 40m/16968, but we didn't do loop-correction at that time.
- We should check if this imbalance is as expected or not.
Summary of actuation calibration so far:
BS : 26.08e-9 /f^2 m/counts (see 40m/16929)
ITMX : 5.29e-9 /f^2 m/counts (see 40m/16929)
ITMY : 4.74e-9 /f^2 m/counts (see 40m/16929)
ETMX : 2.65e-9 /f^2 m/counts (0.5007 times ITMX)
ETMY : 10.91e-9 /f^2 m/counts (2.3017 times ITMY)
|
Attachment 1: Screenshot_2022-07-05_14-52-01_OLTF.png
|
|
Attachment 2: Screenshot_2022-07-05_14-54-03_TF.png
|
|
Attachment 3: Screenshot_2022-07-05_14-56-41_Ratio.png
|
|
16978
|
Thu Jul 7 18:22:12 2022 |
yuta | Update | LSC | Actuator calibration of MC2 using Yarm |
(This is also a restore of elog 40m/16971 from Jul 5, 2022 at 17:36)
MC2 actuator calibration was also done using Yarm in the same way as we did in 40m/16970 (now 40m/16977).
The result is the following;
MC2 : -14.17e-9 /f^2 m/counts in arm length (-2.9905 times ITMY)
MC2 : 5.06e-9 /f^2 m/counts in IMC length
MC2 : 1.06e+05 /f^2 Hz/counts in IR laser frequency
What we did:
- Measured TF from C1:LSC-MC2_EXC to C1:LSC-YARM_IN1 during YARM lock using ETMY (see Attachment #1). Note that the sign of MC2 actuation and ITMY actuation is flipped.
- Took the ratio between ITM actuation and MC2 actuation to calculate MC2 actuation. For ITM actuation, we used the value measured using MICH (see 40m/16929). The average of the ratio in the frequency range 70-150 Hz was used (see Attachment #2).
- The actuation efficiency in meters in arm length was converted into meters in IMC length by multiplying it by IMCLength/ArmLength, where IMCLength=13.5 m is half of IMC round-trip length, ArmLength=37.79 m is the arm length.
- The actuation efficiency in meters in arm length was converted into Hz in IR laser frequency by multiplying it by LaserFreq/ArmLength, where LaserFreq=1064 nm / c is the laser frequency.
Files:
- Measurement files live in https://git.ligo.org/40m/measurements/-/tree/main/LSC/YARM
- Script for calculation lives at https://git.ligo.org/40m/scripts/-/blob/main/CAL/ARM/ETMActuatorCalibration.ipynb
Summary of actuation calibration so far:
BS : 26.08e-9 /f^2 m/counts (see 40m/16929)
ITMX : 5.29e-9 /f^2 m/counts (see 40m/16929)
ITMY : 4.74e-9 /f^2 m/counts (see 40m/16929)
ETMX : 2.65e-9 /f^2 m/counts (0.5007 times ITMX)
ETMY : 10.91e-9 /f^2 m/counts (2.3017 times ITMY)
MC2 : -14.17e-9 /f^2 m/counts in arm length (-2.9905 times ITMY)
MC2 : 5.06e-9 /f^2 m/counts in IMC length
NOTE ADDED by YM on July 7, 2022
To account for the gain imbalance in ETMX, ETMY, MC2, LSC violin filter gains were set to:
C1:LSC-ETMX_GAIN = 4.12
C1:LSC-MC2_GAIN = -0.77
This is a temporary solution to make ETMX and MC2 actuation efficiencies from LSC in terms of arm length to be the same as ETMY 10.91e-9 /f^2 m/counts.
I think it is better to make C1:LSC-ETMX_GAIN = 1, and put 4.12 in C1:SUS-ETMX_TO_COIL gains. We need to adjust local damping gains and XARM ASS afterwards.
As for MC2, it is better to put -0.77 in LSC output matrix, since this balancing depends on LSC topology.
|
Attachment 1: TF.png
|
|
Attachment 2: MC2.png
|
|
16981
|
Fri Jul 8 16:18:35 2022 |
rana | Update | LSC | Actuator calibration of MC2 using Yarm |
although I know that Yuta knows this, I will just put this here to be clear: the NNN/f^2 calibration is only accurate abouve the pendulum POS eiegenfrequency, so when we estimate the DC part (in diaggui, for example), we have to assume that we have a pendulum with f = 1 Hz and Q ~5, to get the value of DC gain to put into the diaggui Gain field in the calibration tab. |
6163
|
Tue Jan 3 20:42:05 2012 |
Leo Singer | Update | General | Actuators for Stewart platform |
I checked on the two single-axis shakers that are present at the 40m that Steve pointed out:
- Brüel & Kjær type 4809, rated for 45 N peak, and
- Brüel & Kjær type 4810, rated for 10 N peak.
Neither of these meet the force requirement of 2.04 kN peak. |
6165
|
Wed Jan 4 02:43:40 2012 |
rana | Update | General | Actuators for Stewart platform |
Quote: |
I checked on the two single-axis shakers that are present at the 40m that Steve pointed out:
- Brüel & Kjær type 4809, rated for 45 N peak, and
- Brüel & Kjær type 4810, rated for 10 N peak.
Neither of these meet the force requirement of 2.04 kN peak.
|
Time to lower your expectations!
Do we really need 40 microns at 500 Hz? Or perhaps should there be a frequency dependent displacement requirement? |
5811
|
Fri Nov 4 15:24:13 2011 |
Mirko | Update | Adaptive Filtering | Adaptive FF on the MC doesn't make sense |
[Den, Jenne, Mirko]

Here is the story:
1. High gain
The control loop has a high gain at the interesting frequencies. The error point (EP) before the servo is approx. zero and the information how much the mirror would move is in the feedback point (FB) behind the servo. The mirror doesn’t actually move because of the high gain. This is the case of the grav. wave detectors and medium frequencies (> approx. 50Hz, <<1kHz). Adding feed-forward (FF) to this doesn’t actually keep the mirror quieter. In fact if you look into the FB and subtract the seismic you make the mirror move more. Yes this is the case we have for the mode cleaner, doesn’t make sense.
In a real GW detector FF however isn’t totally useless. The FB tells you how much the mirror moves, due to GWs, seismic etc. When you record the FB and subtract (offline) the seismic you get closer to the real GW signal.
2. Low gain
When you, for technical reasons, can’t have a high gain in your control loop the EP contains information of how the mirror actually moves. You can then feed this into the adaptive filter and add its output to the FB. This will minimize the EP reducing the actual mirror motion. This is the case we will have for most or all other degrees of freedom in the 40m.
The reason we have so much gain in the mode cleaner length control is that we don’t actually move mirrors around. We change the frequency of the incoming laser light. You can do that crazy fast with a big amplitude. This gives us a high UGF and lots of gain in the 1Hz range we are interested in.
We now change the adaptive filter to look at the EP for all DOFs except for the MC. We calculate the effect of the FF on the MC length signal without ever applying the FF to the MC length control. |
384
|
Mon Mar 17 18:30:48 2008 |
mevans | Configuration | PEM | Adaptive Filtering |
It seems that adaptive filtering can achieve results similar to those of the MISO FIR Wiener (entry 369). The adaptive code simulates real-time operation, but uses the same data used by Rana for the Wiener filter. I ran the adaptive filter over the data 100 times to ensure that it was well trained... maybe too well. |
Attachment 1: mcacc_adaptive.png
|
|
387
|
Thu Mar 20 17:45:36 2008 |
rana | Summary | ASS | Adaptive Filtering in the ASS system |
Over the past couple weeks we (Matt, Alex, Rob, me) have worked on getting an adaptive filter
system working. We wanted to load this system into c1ass to use this processor. The dither alignment
system hasn't been employed here for awhile and so we have just used this box.
The signals are acquired in the PEM ADCU. Alex modified the code there to send the signals over to
the new system. We also get the SUS-LSC_OUT signals from each of the suspensions so that we can
try to minimize them.
The outputs of the adaptive filter go into the unused SUS-MCL inputs of all the suspensions (except
for MC2). In the current setup, we have 6 accelerometers and 1 seismometer around the MC to be used
to demonstrate the principle of the whole thing.
Much more documentation and description of everything is necessary. We'll try to get Matt, Rob, and Alex
to use the elog. |
563
|
Wed Jun 25 09:46:45 2008 |
Sharon | Update | | Adaptive Filters |
I have been learning about different methods for applying adaptive filters to improve the Mode Cleaner lock in specific, and other LIGO systems in general.
Finding the exact number of coeffs we would like to apply for our FIR adaptive filter is very important to the efficiency of the filter. Getting this number higher might improve the accuracy of the filter, but costs time we do not have. Another important number to find is the step size. The step size is the variable that controls how far back we want to look into our data for finding the new coeffs. To understand more about the step size it is necessary to learn about the standard deviation of our input and output signals. By getting the step size too big, we are considering long term behavior, but might be missing out on a short term one.
In the near future I will be learning about the meanings of these variables and their contribution to the over all accuracy of our filters.
Results will be posted. |
704
|
Mon Jul 21 09:52:05 2008 |
Sharon | Update | | Adaptive code changes |
Thanks to Alex, we now save the coefficients of the adaptive filter every cycle. When we choose ENABLE: OFF on the MEDM screen, suppressing the signal to the MC1, we stop and save the last coefficients. When enabling it again, it starts from the last coefficients saved. I will take some measurements today to check how this contributes to the adaptation rate. If you change the number of taps or the number of AUX channels, the coefficients are again set to zero. |
5738
|
Tue Oct 25 20:04:40 2011 |
Mirko | Update | Adaptive Filtering | Adaptive filter witness and EP SNR |
We currently have the code running for all DOFs using all witness channels. By default nothing is applied. C-Code parameters can be changed via the respective EPICS variables. Sanity checks in the C-Code make sure the code doesn't crash when nothing / zeros are fed to the code. Let's look into applying FF to one DOF only as a starting point. We start with MCL.
Remember there are two possible signals to look into MC-F and MC-Servo. See page 5695 http://nodus.ligo.caltech.edu:8080/40m/5695
Dark noise: MC-F over MC-Servo which is unconnected in this measurement:

=> At least 20dB SNR. ADC noise should not be an issue. Of course more is always better.
Coherence of seismometers to MCL:
STS1 is located at the vertex. x-axis along the x arm.
GUR1 is located at the IMC MC2 mirror. Same orientation.

=> 1. Only the x-direction has good coherence (to be expected)
2. Only good coherence at 1.5-4Hz (huh?)
So probably other noise sources are dominating. Let's look into noise projections. Remember IMC autoalignment is off.
A quick adaptive filter run with only the GUR1 and STS1 witnesses applied only to MCL didn't really do anything. Some more thought needs to be invested into the AA and shaping filters. |
5740
|
Tue Oct 25 21:49:13 2011 |
Den | Update | Adaptive Filtering | Adaptive filter witness and EP SNR |
Quote |
Coherence of seismometers to MCL:
STS1 is located at the vertex. x-axis along the x arm.
GUR1 is located at the IMC MC2 mirror. Same orientation.

=> 1. Only the x-direction has good coherence (to be expected)
2. Only good coherence at 1.5-4Hz (huh?)
So probably other noise sources are dominating. Let's look into noise projections. Remember IMC autoalignment is off.
A quick adaptive filter run with only the GUR1 and STS1 witnesses applied only to MCL didn't really do anything. Some more thought needs to be invested into the AA and shaping filters.
|
Indeed, only GUR1_X is reasonable. Static Wiener filtering (length = 2500) of MCL with witness channels GUR_1_X, GUR_1_Y, GUR_1_Z proves your measurements.
We need to callibrate seimometers. I think that now we see velocity, not displacement. It might be useful to amplify the seimometer singal before ADC to make sure that our signal is not ADC noise. |
Attachment 1: gur1_x.jpg
|
|
Attachment 2: gur1_y.jpg
|
|
Attachment 3: gur1_z.jpg
|
|
5777
|
Tue Nov 1 18:16:50 2011 |
Den | Update | Adaptive Filtering | Adaptive filter witness and EP SNR |
Quote: |
Coherence of seismometers to MCL:
STS1 is located at the vertex. x-axis along the x arm.
GUR1 is located at the IMC MC2 mirror. Same orientation.

=> 1. Only the x-direction has good coherence (to be expected)
2. Only good coherence at 1.5-4Hz (huh?)
So probably other noise sources are dominating. Let's look into noise projections. Remember IMC autoalignment is off.
A quick adaptive filter run with only the GUR1 and STS1 witnesses applied only to MCL didn't really do anything. Some more thought needs to be invested into the AA and shaping filters.
|
The possible explanation to this effect is the following:
Seismic noise mainly consists of the Love and Rayleigh surface waves. In the simulations we've taken 2 perpendicular Love waves and 2 perpendicular Rayleigh waves that interfere under the test mirrors. This interference produces both translational and tilt movements. Then we can see the coherence between translational motion and cavity length.

1. The coherence at big frequencies is small due to the passive isolation.
2. The coherence at 1 Hz is 0 due to the wire resonance.
3. The coherence between 1 and 10 Hz is reasonable. At the real 40m's measurements we can see only good coherence for gur1_x and sts1_x but this is the matter of adjusting seismic waves amplitude and direction. In the simulation we've assumed that all waves are of the same amplitude. The really interesting thing is that
4. The coherence below 0.8 Hz began to grow. We don't see this in real measurements.
But let's simulate the seismometer measurements. It measures not only translational motion but also tilt and with amplitude proportional to g / omega^2. On the Figure below the spectrum of translation motion, tilt and tilt as seen by seismometer is presented. We can see that at low frequencies tilt begins to dominate over the translational motion. We assumed the speed of waves in the region 30 - 60 m/sec.

Let's now plot the coherence between the cavity length and seismometer signal.

We can see that the coherence between seismic signal from measured by seismometer and cavity length is gone below 1 Hz where tilt becomes important.
Now let's try to filter out the seismic noise from the cavity length using both static Wiener filtering and adaptive Mfxlms algorithm. For both filters we've used AA filter before the filters and also AI filter after adaptive filter. The downsampling ratio was 4, the sample frequency 256. We can see that nothing is really subtracted due to the pollution of the seismometer signal due to tilt motion.

Assume we do the same computational experiment but with the seismometers that measure only ground translational motion and tilt do not affect on them. Then we have a reasonable subraction of seismic noise at low frequencies even with the filters of the length 100 as shown on the figure below.

Let's look through an order of magnitude analysis. Assume ground motion consists of only one wave with amplitude A and only vertical movement: z(t) = A*sin(2 pi 0.1 t). So the frequency of the wave is 0.1 Hz. If A = 10-6 m => the amplitude of the suspended mirror motion is also approximately 10-6 m, as we have no isolation at low frequencies. The tilt angle has the amplitude alpha = 2*pi*A/lambda, where lambda - wavelength of the ground wave, lambda = v/f = 40/0.1 = 400 m, v - speed of the wave, f - frequency. Then alpha = 10-8 rad. If the distance between ground and mirror suspension point is 1 m, then mirror motion amplitude due to tilt is B = 10-8 m << A.
It turns out that tilt does not effect much on the cavity length compared to the ground translational motion, but it affects a lot on the seismometer signals, that are used as witness signals in the filtering. For that reason we need tiltmeters to filter seismometer signals in order to obtain pure translational ground motion. |
11855
|
Mon Dec 7 10:40:09 2015 |
yutaro | Update | Computer Scripts / Programs | Added 1 line to UNFREEZE_DITHER.py |
I added 1 line to one of the ASS scripts, UNFREEZE_DITHER.py like this:
L29> ez.cawrite('C1:ASS-'+dof+'_GAIN', 0)
The reason why I added this is: without this line, C1:ASS-'+dof+'_GAIN become larger that 1.0, which is nomial value, if you UNFREEZE DITHER when the dither is already running or C1:ASS-'+dof+'_GAIN is not 0.0. |
11793
|
Fri Nov 20 15:44:12 2015 |
Koji | Summary | PSL | Added 17.5kHz LPF to the PMC servo |
As a final tune of the PMC servo, I've added 1nF cap at the error signal amplification stage. The diagram has been updated and uploaded to DCC. https://dcc.ligo.org/D1400221
It should be noted that this modification yielded the error signal to have 17.5kHz roll off.
The openloop TF after the modification has been measured. (Attachment 1)
With the new nominal gain of 9dB, almost the same gain margin for the 28kHz peak has been realized.
=> We have 6dB (factor of 2) more gain at low frequency. Currently, the feature at 8kHz causes the oscillation when the gain is further increased.
Here is the model function for the OLTF.
function cmpOLTFc = PMC_OLTF_model(freqOLTFc)
cmpOLTFc = -9.5e5*pole1(freqOLTFc,0.162).*zero1(freqOLTFc,491)... % from the circuit diagram
.*pole1(freqOLTFc,17.5e3)... % Newly implemented input filter => GBW pole was replaced with this
.*zero2(freqOLTFc,12.5e3,100)... % eye-fit
.*pole2(freqOLTFc,12.2e3,6)... % eye-fit
.*pole2(freqOLTFc,28.8e3, 12)... % eye-fit
.*pole1(freqOLTFc,150e3)... % Mixer LPF estimated from Circuit Lab Simulation
.*pole1(freqOLTFc,11.3)... % Output Impedance + PZT LPF
.*pole1(freqOLTFc,3e4); % Unknown
end
The free-running round-trip displacement (roundtrip) / frequency noise is shown in Attachments. There we compare the spectra with and without IMC locked.
i.e. When the IMC is not locked, we are measuring the laser frequency noise with the sensor (PMC cavity) that is noisy due to the PMC displacement.
When the IMC is locked, the laser frequency is further stabilized while the sensor (PMC) noise is not changed.
- Without IMC locked
Can we see the laser freq noise? It seems that it is visible above 100Hz.
The red curve is the measured noise level. The NPRO (although it is LWE NPRO) noise level from S. Nagano's thesis (see our wiki) is shown there.
- With IMC locked
When the IC is locked, we see the increase of the noise between 1~4Hz. It means that the IMC is not only noisier than the laser, but also noisier than the PMC cavity.
Sounds reasonable. And the PMC is capable to handle this motion.
The reduction of the frequency noise is seen from 100Hz to 30kHz.
The interesting point is that we can see the noise increase above 30kHz when the IMC is locked.
I believe that the phase correction EOM is shared with the PMC modulation. i.e. PMC sees the corrected laser frequency.
We expect that the frequency noise is reduced at this frequency. But in reality not.
In addition, there is a sharp peak at ~35kHz. I wonder If this is caused by the IMC servo. It is worse to investigate. |
Attachment 1: PMC_OLTF.pdf
|
|
Attachment 2: PMC_noise_comparison.pdf
|
|
16957
|
Tue Jun 28 17:07:47 2022 |
Anchal | Update | Calibration | Added Beatnote channels in demodulation of c1cal |
I added today demodulation of C1:LSC-BEATX/Y_FINE_I/Q in the c1cal demodulation where different degrees of freedom can be dithered. For McCal (formerly soCal), we'll dither the arm cavity for which we can use any of the DOFs (like DARM) to send the dither to ETMX/ETMY. Then with green laser locked as well, we'll get the calibration signal from the beatnotes in the demodulaed channels. We can also read right after the mixing in c1cal model and try differnt poles for integration .
I've also added medm screens in the sensing matrix part of LSC screen. These let you see demodulation of beatnote frequency signals. |
11421
|
Thu Jul 16 16:33:56 2015 |
Jessica | Update | General | Added Bode Plots of Bandpass Filter |
I updated the bandpass filter I was using, finding that having different stopband attenuations before and after the passband better emphasized the area from 3 Hz to 20 Hz. I chose a low passband ripple but high stopband attenuation to do this. My passband ripple was 2 dB, the first stopband was 25 dB, and the second stopband attenuation was 40 dB. As can be seen in the filter Magnitude plot, this resulted in a fairly smooth passband and a fairly step dropoff to the stopband, which will better emphasize the region I am trying to isolate. My goal was to emphasize the 3-20 Hz region 10-30 times more than the outside regions. I think I accomplished this by looking at the Bode plot, but I may have chosen the second stopband attenuation to be slightly too high for this. |
Attachment 1: acc1_update.png
|
|
Attachment 2: acc2_update.png
|
|
Attachment 3: acc3_update.png
|
|
Attachment 4: bp_BodeMag.png
|
|
Attachment 5: bp_BodePhase.png
|
|
1782
|
Thu Jul 23 07:34:45 2009 |
Aidan | Update | CDS | Added C2 MEDM screens to 40m SVN. |
See Adhikari eLOG entry: http://nodus.ligo.caltech.edu:8080/AdhikariLab/194 |
707
|
Mon Jul 21 14:26:11 2008 |
Max | Summary | PEM | Added Channels |
The following channels have been added.
Channel Name DAQ port
C1 : PEM-MAG_BSC_X 27
C1 : PEM-MAG_BSC_Y 28
C1 : PEM-MAG_BSC_Z 29
Jenne and I ran the wires from near the beam splitter chamber (as described in a previous elog) to the rack Y7 and plugged the labeled BNC's into ports 27-29. The computer was c0dcu1. John then restarted the frame builder and Alberto and I restarted the front end of c0dcu1 as per the wiki's instructions. The channels seem to be working. - Max. |
2275
|
Mon Nov 16 15:58:02 2009 |
josephb | Configuration | General | Added Gige camera to AP table, added some screens |
I placed a GC750 gige camera looking at a pickoff of the AS port, basically next to the analog camera, on the AP table.
I've modified the main sitemap to include a CAM button, for the digital cameras. There's a half done screen associated with it. At the moment, it reports on the X and Y center of mass calculation, the exposure setting, and displays a little graph with a dot indicating the COM of mass location. Currently this screen is associated a GC750 camera looking at pickoff of the AS port. I'm having some issues with getting shell scripts to run from it, as well as a slider having limits other than 0 and 0. |
3880
|
Mon Nov 8 14:30:15 2010 |
josephb, yuta | Update | CDS | Added LIGONDSIP setting to cshrc.40m |
We added the following line to the cshrc.40m file in the 64-bit linux and 32-bit linux sections:
setenv LIGONDSIP fb
This allows codes like tdsdmd to work properly on the linux machines (seemed to already work fine on the solaris op440m without this change). |
6831
|
Mon Jun 18 23:38:39 2012 |
Jenne | Update | LSC | Added LSC channels to frames |
Since the .ini files get overwritten every time a model is compiled now, we need to put all channels we want saved to frames in the DAQ Channels list inside the model.
I added the _ERR channels for all RFPDs (I and Q for each), as well as the _OUT channels for the DCPDs. I also added the _OUT channels for the DoF servos (ex. C1:LSC-DARM_OUT). I don't remember off the top of my head what else we used to save from the LSC model, but those all seemed like ones we'll possibly want access to later.
We need to go through and do this to all the models we use regularly.
Since SUS hasn't been recompiled in a while, all those channels are saved (until such time as someone does a recompile). Den has gone through and edited the PEM and OAF .ini files by hand each time he recompiles, so we have that data, although we need to put it into the model (which is the new proper way to acquire channels). |
1661
|
Mon Jun 8 18:22:27 2009 |
Alberto | DAQ | LSC | Added PD11 I amd Q slow channels |
I just added two slow channels to C0EDCUEPICS to monitor the input of PD11. The names are:
[C1:LSC-PD11_I_INMON]
[C1:LSC-PD11_Q_INMON] |
3634
|
Fri Oct 1 11:53:42 2010 |
josephb | Configuration | CDS | Added RCG simlink files to the 40m svn |
I've added a new directory in /opt/rtcds/caltech/c1/core called rts_simlink. This directory is now in the 40m svn. Unfortunately, the simlink files used to generate the front end c codes live in a directory controlled by the CDS svn. So I've copied the .mdl files from /opt/rtcds/caltech/c1/core/advLigoRTS/src/epics/simLink/ into this new directory and added them into the 40m svn. When making changes to the simlink files, please copy them to this new directory and check them in so we can a useful history of the models.
|
3178
|
Thu Jul 8 15:19:27 2010 |
josephb, koji | Configuration | Computers | Added Zonet camera to IP table on linux1 |
We gave the Zonet camera the IP 192.168.113.26 and the name Zonet1.
We did this by modifying the /var/named/chroot/var/named/113.168.192.in-addr.arpa.zone and martian.zone files on linux1 as root. |
4460
|
Wed Mar 30 16:32:29 2011 |
Aidan | Configuration | Computer Scripts / Programs | Added a sitemap alias |
I added an alias to the sitemap MEDM screen in /cvs/cds/caltech/target/cshrc.40m
Now you can enjoy launching sitemap from a terminal.
alias sitemap 'medm -x /cvs/cds/rtcds/caltech/c1/medm/sitemap.adl'
|
4463
|
Wed Mar 30 18:50:57 2011 |
Koji | Configuration | Computer Scripts / Programs | Added a sitemap alias |
I thought that "m40m" was the traditional alias for the sitemap...
rossa:~>alias
...
m40m ${medm_base} ${medm_newtail} &
...
sitemap medm -x /cvs/cds/rtcds/caltech/c1/medm/sitemap.adl
rossa:~>set|grep medm
medm_base medm
medm_newtail -x /opt/rtcds/caltech/c1/medm/sitemap.adl
medm_tail -x /cvs/cds/caltech/medm/sitemap.adl
Quote: |
I added an alias to the sitemap MEDM screen in /cvs/cds/caltech/target/cshrc.40m
Now you can enjoy launching sitemap from a terminal.
alias sitemap 'medm -x /cvs/cds/rtcds/caltech/c1/medm/sitemap.adl'
|
|
3226
|
Thu Jul 15 11:58:50 2010 |
josephb | Update | Computers | Added channel to ADCU_PEM (C0DCU1) |
I modified the C1ADCU_PEM.ini file in /cvs/cds/caltech/chans/daq/ (after making a backup), and added a temporary channel called C1:PEM-TEMP_9, the 9 corresponding to the labeled 9 channel on the front of the BNC breakout in the 1Y7 rack. The chnnum it was set to is 15008 (it was commented out and called C1:PEM-PETER_FE). I also set the data rate to 2048.
I then did telnet fb40m 8087, and shutdown, and also hit the blue reconfig button on the DAQ status screen for the C0DCU1 machine. The framebuilder came back up. I confirmed the temporary channel, as well as the Guralp channels were still working from C0DCU1.
We have strung a cable in the cable trays from the SP table to the 1Y7 rack, which has been labeled as "Phasecam PD". This will be used to record the output of an additional photodiode.
|
7521
|
Wed Oct 10 19:22:03 2012 |
jamie | Update | IOO | Added control for input tip-tilts to c1ass |
I have added some control logic and appropriate output DAC channels for the input tip-tilts (TT1 and TT2) to the c1ass model.
The plan is for all the tip-tilt drive electronics to live in a Eurocrate in 1Y2. They will then interface with a DAC in c1lsc.
c1ass runs on the c1lsc front-end machine, and therefore seemed like an appropriate place for the control logic to go.
I added and interface to DAC0, and a top_named IOO block, to c1ass:

The IOO block includes two TT_CONTROL library parts, one for each of TT1 and TT2:

This is just a start so that I can start testing the DAC output.
I have not recompiled c1ass yet. I will do that tomorrow. |
4396
|
Thu Mar 10 13:44:56 2011 |
josephb | Update | CDS | Added digitization noise to the c1spy model for simulated ADCs/DACs |
To simulate digitization noise, the easiest way I found was to use the MathFunction block, found in the CDS_PARTS model, under simLinkParts.
The MathFunction block supports square of input value, square root of input value, reciprocal of input value, and modulo of two input values.
The last is useful because it casts the input values as integers before taking the modulo.By placing this block after the saturation block (set to +/- 32768), adding 32768.5, choosing the 2nd input to be larger than 2 * 32768 (100,000 in this case), and then subtracting 32768, we wind up with a rounding function. 
The above method has been applied to the c1spy model in the CI and SO out sub-blocks. |
16269
|
Wed Aug 4 18:19:26 2021 |
paco | Update | General | Added infrasensing temperature unit to martian network |
[ian, anchal, paco]
We hooked up the infrasensing unit to power and changed its default IP address from 192.168.11.160 (factory default) to 192.168.113.240 in the martian network. The sensor is online with user controls and the usual password for most workstations in that IP address. |
6869
|
Mon Jun 25 15:19:07 2012 |
Yaakov | Update | PEM | Added microphone channels, moved accelorometer channels |
Jenne and I renamed the mic channels Den created (elog 6664) to MIC_1, MIC_2, etc from the original accelerometer names to keep things clear. We then added 6 new channels (22-27) for the accelerometers, named ACC_MC1_X, Y, Z, ACC_MC2_X, Y, Z, etc. (See the screenshot below). We also added a DAQ channel block and listed out the IN1 channel of all the sensors. We compiled and started the model, and checked that all the channels were there in DataViewer.

|
2477
|
Tue Jan 5 10:26:32 2010 |
Alberto, Steve | Omnistructure | Environment | Added new wall cable-racks |
we hung two new WALL cable racks. One is on the pillar next to the Sp table, the other is next to the PSL computer rack.
To do that we had to drill holes in the wall since the simple screws weren't strong enough to keep them up.
One of the racks, the yellow, is dedicated to 4-pin lemos and other thick cables.
 
|
2478
|
Tue Jan 5 11:00:04 2010 |
rana | Omnistructure | Environment | Added new wall cable-racks |
Quote: |
we hung two new WALL cable racks. One is on the pillar next to the Sp table, the other is next to the PSL computer rack.
|
awesome - I have ordered 5 blue racks so that we can hang power cables. The fat BLUE ones are for fat cables and the orange ones for the coax cables. |
15275
|
Fri Mar 13 14:28:39 2020 |
gautam | Update | PSL | Added tees to PMC Trans and REFL |
I want to monitor the PMC TRANS and REFL levels on the PSL table - previously there were some cables going to the oscilloscope on the shelf but someone had removed these. I re-installed them just now. While there, I disconnected the drive to the AOM - there must've been some DC signal going to it because when I removed the cable, the PMC and IMC transmission were recovered to their nominal levels. |
16330
|
Tue Sep 14 17:22:21 2021 |
Anchal | Update | CDS | Added temp sensor channels to DAQ list |
[Tega, Paco, Anchal]
We attempted to reboot fb1 daqd today to get the new temperature sensor channels recording. However, the FE models got stuck, apparantely due to reasons explaine din 40m/16325. Jamie cleared the /var/logs in fb1 so that FE can reboot. We were able to reboot the FE machines after this work successfully and get the models running too. During the day, the FE machines were shut down manually and brought back on manually, a couple of times on the c1iscex machine. Only change in fb1 is in the /opt/rtcds/caltech/c1/chans/daq/C0EDCU.ini where the new channels were added, and some hacking was done by Jamie in gpstime module (See 40m/16327). |
16270
|
Thu Aug 5 14:59:31 2021 |
Anchal | Update | General | Added temperature sensors at Yend and Vertex too |
I've added the other two temperature sensor modules on Y end (on 1Y4, IP: 192.168.113.241) and in the vertex on (1X2, IP: 192.168.113.242). I've updated the martian host table accordingly. From inside martian network, one can go to the browser and go to the IP address to see the temperature sensor status . These sensors can be set to trigger alarm and send emails/sms etc if temperature goes out of a defined range.
I feel something is off though. The vertex sensor shows temperature of ~28 degrees C, Xend says 20 degrees C and Yend says 26 degrees C. I believe these sensors might need calibration.
Remaining tasks are following:
- Modbus TCP solution:
- If we get it right, this will be easiest solution.
- We just need to add these sensors as streaming devices in some slow EPICS machine in there .cmd file and add the temperature sensing channels in a corresponding database file.
- Python workaround:
- Might be faster but dirty.
- We run a python script on megatron which requests temperature values every second or so from the IP addresses and write them on a soft EPICs channel.
- We still would need to create a soft EPICs channel fro this and add it to framebuilder data acquisition list.
- Even shorted workaround for near future could be to just write temperature every 30 min to a log file in some location.
[anchal, paco]
We made a script under scripts/PEM/temp_logger.py and ran it on megatron. The script uses the requests package to query the latest sensor data from the three sensors every 10 minutes as a json file and outputs accordingly. This is not a permanent solution. |
16319
|
Mon Sep 13 04:12:01 2021 |
Tega | Update | General | Added temperature sensors at Yend and Vertex too |
I finally got the modbus part working on chiara, so we can now view the temperature data on any machine on the martian network, see Attachment 1.
I also updated the entries on /opt/rtcds/caltech/c1/chans/daq/C0EDCU.ini, as suggested by Koji, to include the SensorGatway temperature channels, but I still don't see their EPICs channels on https://ldvw.ligo.caltech.edu/ldvw/view. This means the channels are not available via nds so I think the temperature data is not being to be written to frame files on framebuilder but I am not sure what this entails, since I assumed C0EDCU.ini is the framebuilder daq channel list.
When the EPICs channels are available via nds, we should be able to display the temperature data on the summary pages.
Quote: |
I've added the other two temperature sensor modules on Y end (on 1Y4, IP: 192.168.113.241) and in the vertex on (1X2, IP: 192.168.113.242). I've updated the martian host table accordingly. From inside martian network, one can go to the browser and go to the IP address to see the temperature sensor status . These sensors can be set to trigger alarm and send emails/sms etc if temperature goes out of a defined range.
I feel something is off though. The vertex sensor shows temperature of ~28 degrees C, Xend says 20 degrees C and Yend says 26 degrees C. I believe these sensors might need calibration.
Remaining tasks are following:
- Modbus TCP solution:
- If we get it right, this will be easiest solution.
- We just need to add these sensors as streaming devices in some slow EPICS machine in there .cmd file and add the temperature sensing channels in a corresponding database file.
- Python workaround:
- Might be faster but dirty.
- We run a python script on megatron which requests temperature values every second or so from the IP addresses and write them on a soft EPICs channel.
- We still would need to create a soft EPICs channel fro this and add it to framebuilder data acquisition list.
- Even shorted workaround for near future could be to just write temperature every 30 min to a log file in some location.
[anchal, paco]
We made a script under scripts/PEM/temp_logger.py and ran it on megatron. The script uses the requests package to query the latest sensor data from the three sensors every 10 minutes as a json file and outputs accordingly. This is not a permanent solution.
|
|
Attachment 1: Screen_Shot_2021-09-13_at_4.16.22_AM.png
|
|
4243
|
Thu Feb 3 04:43:58 2011 |
Suresh | Update | Electronics | Added two new DAQ channels |
[Suresh, Joe]
We added the following two new DAQ channels into the c1:GCV model. The daq:analog input channels are on card ADC0 and correspond to channels 3 and 4 on the card.
c1:GCV-EXT_REF_OUT_DAQ Sampling rate=2kHz acquiring a 1Hz sine wave from the SRS Function Generator DS345. This is using the Rb 10MHz signal as an external frequency reference.
c1:GCV-PLL_OUT_DAQ Sa.rate=2kHz acquiring the demodulated signal from the PLL servo.
This work is connected to the study of VCO PLL loop noise at frequencies below 0.1Hz. We are trying to measure phase noise in the VCO PLL servo at low frequencies as this noise would result in arm length fluctuations in the green-locking scheme.
|
16802
|
Fri Apr 22 07:01:58 2022 |
Jc | Update | Coil Drivers | Adding Resistors and Reinstalling |
[Koji, JC]
Coil Drivers LO2, SR2, AS4, and AS1 have been updated a reinstalled into the system.
LO2 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3 Unit: S2100008
LO2 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3 Unit: S2100530
SR2 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3 Unit: S2100614
SR2 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3 Unit: S2100615
AS1 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3 Unit: S2100610
AS1 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3 Unit: S2100611
AS4 Coil Driver 1 (UL/LL/UR)now has R=100 // 1.2k ~ 92Ohm for CH1/2/3 Unit: S2100612
AS4 Coil Driver 2 (LR/SD)now has R=100 // 1.2k ~ 92Ohm for CH3 Unit: S2100613 |
Attachment 1: IMG_0536.jpeg
|
|
Attachment 2: IMG_0539.jpeg
|
|
Attachment 3: IMG_0542.jpeg
|
|
Attachment 4: IMG_0545.jpeg
|
|
Attachment 5: IMG_0548.jpeg
|
|
Attachment 6: IMG_0551.jpeg
|
|
Attachment 7: IMG_0554.jpeg
|
|
Attachment 8: IMG_0557.jpeg
|
|
10130
|
Sat Jul 5 04:18:45 2014 |
Andres | Update | 40m Xend Table upgrade | Adding Two Lenses After the Second Steering Mirror in Order Two Increase the Gouy Phase Difference Between the Sterring Mirrors |
I had been working on the Xend table optical layout update. Since the two steering mirrors in the Xend green are too close to each, there is a very small Gouy Phase different between these two mirrors. It was suggested to place two lenses so that we can increase the Gouy Phase. I have been working with Nick on this problem, and we had found a solution by using a la mode. We had written an a la mode code that optimize the Gouy Phase and the Mode Matching at the same time. After trying different lenses, we found the following results: a mode matching of 0.9939 as it is show in the first attachment below, and we found a Gouy Phase different between the two mirrors of about 60 degrees. I took photos of the Xend Table. The first photo is the Xend table as we had it right now. In the second photo, I moved the 2nd lens, and I placed the two more lenses that we need it, with more or lenses the correct position where they will be placed. The three old lenses will be replaced by three lenses of different focal length as it can be seen in the first attachment below. The first lens and third lens will stay in the same position where the old first lens and old third lens are, and the second lens will be moved by about half of an inch. We might have one or two of the lenses that we need, but we will have to order the rest of the lenses that need. My plan is to verify the lenses that we already have. Then, I need to let Nick know with lenses we need to order. Hopefully, we will be able to update the table by the end of this week if everything turn out fine. |
Attachment 1: OverlapAndComponentsOfTheSolution.png
|
|
Attachment 2: CloseLookToTheGouyPhaseBetweenMirr1AndMirr2.jpg
|
|
Attachment 3: EntireRangeOfBeamPath.jpg
|
|
Attachment 4: XendTableWithTwoNeedLensesAdding.JPG
|
|
Attachment 5: SchematicOfSolutionForTheLensesGouyPhase.jpeg
|
|
Attachment 6: XendGreenModeMatchingAndGouyPhaseOptimization.m
|
clear all
% In this code we are using a la mode to optimatize the mode matching and
% to optimatize the Gouy phase between mirror 1 and mirror 2. All the units
% are in meter
w0=2.943*1e-5; % The Waist of the laser measured before the faraday
z0_laser=-0.039; % position measured where the waist is located
lamb= 532*10^-9; % wavelength of green light in mm
lFaraday=.0638; % Length of the faraday
... 148 more lines ...
|
Attachment 7: BeforeIncludingLensesORMovingLenses.JPG
|
|
14074
|
Mon Jul 16 18:12:00 2018 |
Koji | Update | VAC | Adding a manual gate valve between TP1 and V1/VM2 |
[Steve Koji]
We are in the process of adding a manual gate valve between TP1 (Osaka Maglev) and the other gate valves (I suppose V1 and VM2).
The work is still on going and we will continue to work on this tomorrow. Because this section is isolated from the main volume, this work does not hold off the possible rough pumping tomorrow morning.
The motivation of this work is as follows:
- Since TP2 failed, the main vacuum volume has been pumped down by TP1 and TP3. However TP3 is not capable to handle the large pressure difference at the early stage of the turbo pumping. This cause TP3 to have excessive heating or even thermal shutdown.
- The remedy is to put a gate valve between TPs and the main vacuum to limit the amount of gas flowing into the TPs. This indeed slows down the pumping speed of turbo, but this is not the dominant part of the pumping time.
Actual work:
- Comfirmed TP1 is isolated.
- Unscrewed the flange of TP1.
- Remove TP1. This required to lift up TP1 with some shim as the nuts interferes with the TP1 body. (Attachment1, 2, 3)
- Now remove 10inch flange adapter. (Attachment4)
-Attach 10"-8" adapter and 8" rotational sleeve. (Attachment5) |
Attachment 1: P_20180716_155413.jpg
|
|
Attachment 2: P_20180716_155645.jpg
|
|
Attachment 3: P_20180716_155738.jpg
|
|
Attachment 4: P_20180716_162307.jpg
|
|
Attachment 5: P_20180716_172000.jpg
|
|
4271
|
Thu Feb 10 14:52:36 2011 |
Aidan | HowTo | Computers | Adding filenames in MATLAB plots |
The following code is incredibly useful when creating MATLAB plots as it adds the filename of the script to the plot itself. I think it should be used for all MATLAB plots that go on the elog.
For example, I have no idea where the data/script is that was used to generate these plots.
orient landscape
xposn = 0.0;
yposn = -0.13; % you sometimes have to tweak this value depending on the page size and the number of subplots
text(xposn,yposn,[filename('fullpath'), '.m'], ...
'Units', 'normalized', ...
'Interpreter', 'none', ...
'FontSize', 6)
print('-dpdf', [filename('fullpath'), '.pdf'])
|
9430
|
Wed Nov 27 18:31:26 2013 |
Koji | Summary | LSC | Adittion of the ALS error signals to the LSC input matrix |
The Phase tracker outputs (= ALS X/Y error signals) are now conveyed to the LSC model.
Their entry points at the LSC model are C1:LSC-ALSX_IN1 and C1:LSC-ALSY_IN1.
They are connected to the signal matrix (28th and 29th signals) via signal conditioning filters (C1:LSC-ALSX and C1:LSC-ALSY).
The main LSC screen has not been updated. The conventional ALS servos are still remains as they were.
This renovation required the recompilation of c1als, c1rfm, and c1lsc. Two PCIe-RFM bridge paths were added resulting in
increase of the c1rfm timing budget from 38 to 44. |
6409
|
Wed Mar 14 03:34:44 2012 |
kiwamu | Update | SUS | Adjustment of BS suspension output matrix : coupling from SIDE to POS |
[Rana / Kiwamu]
We put some elements in the BS output matrix to mitigate the actuator coupling from SIDE to POS.
As a result the degree of the coupling reduced by a factor of 2 or so.
Rana did the "Q of 5" test on the SIDE damping servo after putting the elements and set the gain to be 40.
The attached screen shot is the new elements that we put in the suspension output matrix.

(How to)
- Excite the SIDE motion by AWG at 3 Hz.
- Monitor the POS signal in DTT
- Try some numbers in the matrix elements until the peak at 3 Hz in the POS signal is minimized
Quote from #6369 |
The BS SIDE damping gain seemed too low. The gain had been 5 while the rest of the suspensions had gains of 90-500.
I increased the gain and set it to be 80.
I did the "Q of 5" test by kicking the BS SIDE motion to find the right gain value.
However there was a big cross coupling, which was most likely a coupling from the SIDE actuator to the POS motion.
Due to the cross coupling, the Q of 5 test didn't really show a nice ring down time series. I just put a gain of 80 to let the Q value sort of 5.
I think we should diagonalize the out matrices for all the suspensions at some point.
|
|
5288
|
Tue Aug 23 14:49:14 2011 |
jamie, jenne, kiwamu, suresh, keiko | Update | SUS | Adjustment of ETMY, issue with ITMY whitening |
Before lunch we took a closer look at two of the suspensions that were most problematic: ITMY and ETMY. Over lunch we took new free swinging data. Results below:
- For ITMY we discovered that the whitening on the UL sensor was not switching. This was causing the UL sensor to have a different response, with a steeper roll of, which was causing all of the transfer function estimates to the other sensors to have large imaginary components. We took new free swing data with all of the whitening turned OFF. The result is a much improved matrix and diagnalization. The input matrix elements are mostly the same, but the coupling is basically gone. We'll fix the whitening after the pump down.
ITMY |
 |
pit yaw pos side butt
UL 0.157 1.311 1.213 -0.090 0.956
UR 1.749 -0.490 0.886 -0.038 -1.042
LR -0.251 -2.000 0.787 -0.007 1.066
LL -1.843 -0.199 1.114 -0.059 -0.936
SD -0.973 -0.205 1.428 1.000 0.239 |
4.34779 |
- ETMY has a very problematic SIDE OSEM. The magnet does not line up with the OSEM axis, and since there is no lateral adjustment in the side OSEMs, there's not much we can do about this. We're using aluminum foil to wedge the OSEM over as far as possible, but it's not quite enough. With the OSEM plates horizontal there is a lot of POS->SIDE coupling. With the OSEM plates vertical, the magnetic sits a little too close to the rear face, which can cause the magnet to get stuck to the LED plate. We're trying to decide where to leave it now, but the new diagnalization with the OSEM plates vertical is definitely better:
ETMX |
 |
pit yaw pos side butt
UL -0.138 1.224 1.463 -0.086 0.944
UR 0.867 -0.776 1.501 -0.072 -1.051
LR -0.995 -0.896 0.537 -0.045 0.754
LL -2.000 1.104 0.499 -0.059 -1.251
SD 0.011 0.220 1.917 1.000 0.224 |
4.42482 |
|
7901
|
Tue Jan 15 19:26:35 2013 |
jamie | Update | Alignment | Adjustment of active TTs and input alignment |
[Jamie, Manasa, Jenne]
We started by verifying that the tip-tilts were getting the correct signals at the correct coils, and were hanging properly without touching.
We started with TT2. It was not hanging freely. One of the coils was in much further than the others, and the mirror frame was basically sitting on the back side yaw dampers. I backed out the coil to match the others, and backed off all of the dampers, both in back and the corner dampers on the front.
Once the mirror was freely suspended, we borrowed the BS oplev to verify that the mirror was hanging vertically. I adjusted the adjustment screw on the bottom of the frame to make it level. Once that was done, we verified our EPICS control. We finally figured out that some of the coils have polarity flipped relative to the others, which is why we were seeing pitch as yaw and vice-versa. At that point we were satisfied with how TT2 was hanging, and went back to TT1.
Given how hard it is to look at TT1, I just made sure all the dampers were backed out and touched the mirror frame to verify that it was freely swinging. I leveled TT1 with the lower frame adjustment screw by looking at the spot position on MMT1. Once it was level, we adjusted the EPICS biases in yaw to get it centered in yaw on MMT1.
I then adjusted the screws on MMT1 to get the beam centered at MMT2, and did the same at MMT2 to get the beam centered vertically at TT2.
I put the target at PRM and the double target at BS. I loosened TT2 from it's base so that I could push it around a bit. Once I had it in a reasonable position, with a beam coming out at PR3, I adjusted MMT1 to get the beam centered through the PRM target. I went back and checked that we were still centered at MMT1. We then adjusted the pitch and yaw of TT2 to get the transmitted beam through the BS targets as clear as possible.
At this point we stopped and closed up. Tomorrow first thing AM we'll get our beams at the ETMs, try to finalize the input alignment, and see if we can do some in-air locking.
The plan is still to close up at the end of the week. |
7902
|
Tue Jan 15 20:00:42 2013 |
Manasa | Update | Alignment | Adjustment of active TTs and input alignment |
Just for reference! The changes made to the TT matrix in order to fix the polarity problem:
The old matrix values are mentioned in elog!
PIT YAW New
Pit slider | -100 -100 | UL
0 | -100 100 | LL
Yaw slider | 100 -100 | UR
0 | 100 100 | LR
|
4580
|
Thu Apr 28 10:53:50 2011 |
josephb | Update | CDS | Adventures in Hyper-threading |
What was done:
1) Turn off MC1, MC2, MC3, BS, ITMX, ITMY, PRM, SRM watchdogs.
2) Turn c1sus computer off (sudo shutdown now)
3) Go connect monitor and keyboard to c1sus. Turn c1sus on.
4) Hit "del" key at the right time to go to setup (BIOS).
5) Go to BIOS advanced tab, CPU options, enable Multi-threading.
6) Hit F10 to save and let the computer continue booting.
What went wrong:
Once c1sus was up, I noticed several red lights and dead keep alives for the c1sus models.
Typing dmesg on c1sus revealed many messages like:
[ 107.583420] c1x02: cycle 33737 time 20; adcWait 10; write1 0; write2 0; longest write2 0
[ 107.583771] c1x02: cycle 33760 time 19; adcWait 11; write1 0; write2 0; longest write2 0
This indicates the Input/Output Processor (IOP) is not completing its duties within the 15 microseconds (1/64 kHz) that it has. These lines indicate its take 20 or 19 microseconds. (I saw messages ranging from 16 to 22 microseconds).
So this seems to agree with Rolf's observations that hyperthreading can cause a 5-10 microsecond increase in computation time.
So the next thing to do is modify which core the codes are running on, and try to get them paired up on the same physical core. |