40m QIL Cryo_Lab CTN SUS_Lab CAML OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m elog, Page 48 of 357  Not logged in ELOG logo
ID Dateup Author Type Category Subject
  2369   Wed Dec 9 00:23:28 2009 KojiUpdateSUSfree swinging spectra of ETMY and ITMX

Where is the plot for the trend?
It can be either something very important or just a daydream of you.
We can't say anything before we see the data.

We like to see it if you think this is interesting.

... Just a naive guess: Is it just because the seismic level got quiet in the night?

 

P.S.

You looks consistently confused some words like damping, Q, and peak height.

  • Q is defined by the transfer function of the system (= pendulum).
     
  • Damping (either active or passive) makes the Q lower.
     
  • The peak height of the resonance in the spectrum dy is determined by the disturbance dx and the transfer function H (=y/x).

dy = H dx

As the damping makes the Q lower, the peak height also gets lowered by the damping.
But if the disturbance gets smaller, the peak height can become small even without any change of the damping and the Q.

Quote:

By the way I found a trend, which can be seen in all of the data taken today and yesterday.

The resonances of pitch and yaw around 0.5Hz look like being damped, because their height from the floor become lower than the past.

I don't know what goes on, but it is interesting because you can see the trend in all of the data.

 

  2370   Wed Dec 9 09:07:32 2009 steveUpdatePEMhigh seismic activity

The construction activity is shaking the tables in the control room.  The compactor- large remote controlled jackhammer is in the bottom of the 16-17 ft deep hole 15 ft east of ITMY in CES bay. The suspensions  are holding OK. PRM, MC1 and MC3 are effected mostly.

Attachment 1: seis24d.png
seis24d.png
Attachment 2: seis4h.png
seis4h.png
Attachment 3: P1050833.JPG
P1050833.JPG
Attachment 4: P1050836.JPG
P1050836.JPG
  2371   Wed Dec 9 10:53:41 2009 josephbUpdateCamerasCamera client wasn't able to talk to server on port 5010, reboot fixed it.

I finally got around to taking a look at the digital camera setup today.  Rob had complained the client had stopped working on Rosalba.

After looking at the code start up and not complain, yet not produce any window output, it looks like it was a network problem.  I tried rebooting Rosalba, but that didn't fix anything.

Using netstat -an, I looked for the port 5010 on both rosalba and ottavia, since that is the port that was being used by the camera.  Ottavia was saying there were 6 established connections after Rosalba had rebooted (rosalba is 131.215.113.103).  I can only presume 6 instances of the camera code had somehow shutdown in such a way they had not closed the connection.

[root@ottavia controls]#netstat -an | grep 5010
tcp        0      0 0.0.0.0:5010                0.0.0.0:*                   LISTEN     
tcp        0      0 131.215.113.97:5010         131.215.113.103:57366       ESTABLISHED
tcp        0      0 131.215.113.97:5010         131.215.113.103:58417       ESTABLISHED
tcp        1      0 131.215.113.97:46459        131.215.113.97:5010         CLOSE_WAIT 
tcp        0      0 131.215.113.97:5010         131.215.113.103:57211       ESTABLISHED
tcp        0      0 131.215.113.97:5010         131.215.113.103:57300       ESTABLISHED
tcp        0      0 131.215.113.97:5010         131.215.113.103:57299       ESTABLISHED
tcp        0      0 131.215.113.97:5010         131.215.113.103:57315       ESTABLISHED

 

I switched the code to use port 5022 which worked fine.  However, I'm not sure what would have caused the original connection closure failures, as I test several close methods (including the kill command on the server end used by the medm screen), and none seemed to generate this broken connection state.  I rebooted Ottavia, and this seemed to fix the connections, and allowed port 5010 to work.  I also tried creating 10 connections, which all seem to run fine simultaneously.  So its not someone overloading that port with too many connections which caused the problem.  Its like the the port stopped working somehow, which froze the connection status, but how or why I don't know at this point.

  2372   Wed Dec 9 17:51:03 2009 kiwamuUpdateSUSwatchdogs

Please do not touch the watchdogs for all SUSs except for MCs,

because I am going to measure the free swinging spectra for ITMs, ETMs, BS, PRM, SRM tonight.

Today, it is good chance to summarize those data under atmospheric pressure.

thank you.

 

  2373   Wed Dec 9 18:01:06 2009 KojiUpdateCOCWiping finished

[Kiwamu, Jenne, Alberto, Steve, Bob, Koji]

We finished wiping of four test masses without any trouble. ITMY looked little bit dusty, but not as much as ITMX did.
We confirmed the surface of the ITMX again as we worked at vertex a lot today. It still looked clean.

We closed the light doors. The suspensions are left free tonight in order to check their behavior.
Tomorrow morning from 9AM, we will replace the door to the heavy ones.

  2374   Wed Dec 9 21:09:28 2009 kiwamuUpdateSUSRe: free swinging spectra of ETMY and ITMX

Okay, now the data are attached. At that time I just wanted to say like the follower.

- - -

In the free-swinging spectra around ~0.5Hz, you can see the two resonances, which come from pitch and yaw mode of the pendulum.

Note that, the vertical and the horizontal axis are adjusted to be the same for the two plots in the figure .

And I found that

* the floor levels are almost the same (the factor of about 1.5 or something like that) compared to the past.

* however the peak heights for two resonances are several 10 times smaller than the past.

* this tendency are shown in all of the data (ITMX, ETMX, ETMY).

If such variation of the peak heights is cased by the seismic activity, it means the seismic level change by several 10 times. It sounds large to me.
 

Quote:

Where is the plot for the trend?
It can be either something very important or just a daydream of you.
We can't say anything before we see the data.

Quote:

By the way I found a trend, which can be seen in all of the data taken today and yesterday.

The resonances of pitch and yaw around 0.5Hz look like being damped, because their height from the floor become lower than the past.

I don't know what goes on, but it is interesting because you can see the trend in all of the data.

 

 

Attachment 1: Pitch-Yaw_modes.png
Pitch-Yaw_modes.png
  2375   Thu Dec 10 00:46:15 2009 KojiUpdateSUSRe: free swinging spectra of ETMY and ITMX

Well, I get the point now. It could be either seismic or change in the suspension Q.

The pendulum memorizes its own state for a period of ~ Q T_pend. (T_pend is the period of the pendulum)
If the pendulum Q is very high (>104), once the pendulum is excited, the effect of the excitation can last many hours.

On the other hand, in our current case, we turned on the damping once, and then turned off the damping.
Again it takes ~Q T_pend to be excited. 

In those cases, the peak height is not yet before in equilibrium, and can be higher or lower than expected. 

So, my suggestion is:
Track the peak height along the long time scale (~10hrs) and compare between the previous one and the current one.
This may indicate whether it is equilibrium or not, and where the equilibrium is.

Quote:

If such variation of the peak heights is cased by the seismic activity, it means the seismic level change by several 10 times. It sounds large to me.

 

  2376   Thu Dec 10 08:40:12 2009 AlbertoUpdateComputersFronte-ends down

I found all the front-ends, except for C1SUSVME1 and C0DCU1 down this morning. DAQAWG shows up green on the C0DAQ_DETAIL screen but it is on a "bad" satus.

I'll go for a big boot fest.

  2377   Thu Dec 10 08:43:25 2009 steveFrogsEnvironmentdiesel fumes are less

Quote:

 Instead of doing RCG stuff, I went to Millikan to work on data analysis as I couldn't stand the fumes from the construction.  (this morning, 8am) 

 Diesel fumes are pumped away from control room AC intakes  with the help of newly installed  reflector boxes on the CES wall  fans.........see # 2272

Attachment 1: P1050817.JPG
P1050817.JPG
  2378   Thu Dec 10 08:50:33 2009 AlbertoUpdateComputersFronte-ends down

Quote:

I found all the front-ends, except for C1SUSVME1 and C0DCU1 down this morning. DAQAWG shows up green on the C0DAQ_DETAIL screen but it is on a "bad" satus.

I'll go for a big boot fest.

Since I wanted to single out the faulting system when these situations occur, I tried to reboot the computers one by one.

1) I reset the RFM Network by pushing the reset button on the bypass switch on the 1Y7 rack. Then I tried to bring C1SOSVME up by power-cycling and restarting it as in the procedure in the wiki. I repeated a second time but it didn't work. At some point of the restarting process I get the error message "No response from EPICS".
2) I also tried rebooting only C1DCUEPICS but it didn't work: I kept having the same response when restarting C1SOSVME
3) I tried to reboot C0DAQCTRL and C1DCU1 by power cycling their crate; power-cycled and restarted C1SOSVME. Nada. Same response from C1SOSVME.
4) I restarted the framebuilder;  power-cycled and restarted C1SOSVME. Nothing. Same response from C1SOSVME.
5) I restarted the framebuilder, then rebooted C0DAQCTRL and C1DCU, then power-cycled and restarted C1SOSVME. Niente. Same response from C1SOSVME.
 
Then I did the so called "Nuclear Option", the only solution that so far has proven to work in these circumstances. I executed the steps in the order they are listed, waiting for each step to be completed before passing to the next one.
0) Switch off: the frame builder, the C0DAQCTRL and C1DCU crate, C1DCUEPICS
1) turn on the frame builder
2) reset of the RFM Network switch on 1Y7 (although, it's not sure whether this step is really necessary; but it's costless)
3) turn on C1DCUEPICS
4) turn on the C0DAQCTRL and C1DCU crate
5) power-cycle and restart the single front-ends
6) burt-restore all the snapshots
 
When I tried to restart C1SOSVME by power-cycling it I still got the same response: "No response from EPICS". But I then reset C1SUSVME1 and C1SUSVME2 I was able to restart C1SOSVME.
 
It turned out that while I was checking the efficacy of the steps of the Grand Reboot to single out the crucial one, I was getting fooled by C1SOSVME's status. C1SOSVME was stuck, hanging on C1SUSVME1 and C1SUSVME2.
 
So the Nuclear option is still unproven as the only working procedure. It might be not necessary.
 
Maybe restating BOTH RFM switches, the one in 1Y7 and the one in 1Y6, would be sufficient. Or maybe just power-cycling the C0DAQCTRL and C1DCU1 is sufficient. This has to be confirmed next time we incur on the same problem.
  2379   Thu Dec 10 09:51:06 2009 robUpdatePSLRCPID settings not saved

Koji, Jenne, Rob

 

We found that the RCPID servo "setpoint" was not in the relevant saverestore.req file, and so when c1psl got rebooted earlier this week, this setting was left at zero.  Thus, the RC got a bit chilly over the last few days.  This channel has been added. 

 

Also, RCPID channels have been added (manually) to conlog_channels. 

  2381   Thu Dec 10 09:56:32 2009 KojiUpdatePSLRCPID settings not saved

Note: The set point C1:PSL-FSS_RCPID_SETPOINT is 37.0 on C1PSL_FSS_RCPID.adl.

Now the temp is recovering with its full speed. At some point we have to restore the value of the FSS SLOW DC as the temp change drag it up.

Quote:

Koji, Jenne, Rob

We found that the RCPID servo "setpoint" was not in the relevant saverestore.req file, and so when c1psl got rebooted earlier this week, this setting was left at zero.  Thus, the RC got a bit chilly over the last few days.  This channel has been added. 

Also, RCPID channels have been added (manually) to conlog_channels. 

 

Attachment 1: RC_TEMP.png
RC_TEMP.png
  2382   Thu Dec 10 10:01:16 2009 JenneUpdateComputersFronte-ends down

All the front ends are back up.  

Quote:

Quote:

I found all the front-ends, except for C1SUSVME1 and C0DCU1 down this morning. DAQAWG shows up green on the C0DAQ_DETAIL screen but it is on a "bad" satus.

I'll go for a big boot fest.

Since I wanted to understand once for all what's the faulting system when these situations occur, I tried to reboot the computers one by one.

1) I reset the RFM Network by pushing the reset button on the bypass switch on the 1Y7 rack. Then I tried to bring C1SOSVME up by power-cycling and restarting it as in the procedure in the wiki. I repeated a second time but it didn't work. At some point of the restarting process I get the error message "No response from EPICS".
2) I also tried rebooting only C1DCUEPICS but it didn't work: I kept having the same response when restarting C1SOSVME
3) I tried to reboot C0DAQCTRL and C1DCU1 by power cycling their crate; power-cycled and restarted C1SOSVME. Nada. Same response from C1SOSVME.
4) I restarted the framebuilder;  power-cycled and restarted C1SOSVME. Nothing. Same response from C1SOSVME.
5) I restarted the framebuilder, then rebooted C0DAQCTRL and C1DCU, then power-cycled and restarted C1SOSVME. Niente. Same response from C1SOSVME.
 
The following is the so called "Nuclear Option", the only solution that so far has proven to work in these circumstances. Execute the following steps in the order they are listed, waiting for each step to be completed before passing to the next one.
0) Switch off: the frame builder, the C0DAQCTRL and C1DCU crate, C1DCUEPICS
1) turn on the frame builder
2) reset of the RFM Network switch on 1Y7 (although, it's not sure whether this step is really necessary; but it's costless)
3) turn on C1DCUEPICS
4) turn on the C0DAQCTRL and C1DCU crate
 
One other possibility remains to be explored to avoid the Nuclear Option. And that is to just try to reset both RFM Network switches: the one in 1Y7 and the one in 1Y6.

 

  2383   Thu Dec 10 10:31:18 2009 JenneUpdateComputersFronte-ends down

Quote:

All the front ends are back up.  

Quote:

Quote:

I found all the front-ends, except for C1SUSVME1 and C0DCU1 down this morning. DAQAWG shows up green on the C0DAQ_DETAIL screen but it is on a "bad" satus.

I'll go for a big boot fest.

Since I wanted to understand once for all what's the faulting system when these situations occur, I tried to reboot the computers one by one.

1) I reset the RFM Network by pushing the reset button on the bypass switch on the 1Y7 rack. Then I tried to bring C1SOSVME up by power-cycling and restarting it as in the procedure in the wiki. I repeated a second time but it didn't work. At some point of the restarting process I get the error message "No response from EPICS".
2) I also tried rebooting only C1DCUEPICS but it didn't work: I kept having the same response when restarting C1SOSVME
3) I tried to reboot C0DAQCTRL and C1DCU1 by power cycling their crate; power-cycled and restarted C1SOSVME. Nada. Same response from C1SOSVME.
4) I restarted the framebuilder;  power-cycled and restarted C1SOSVME. Nothing. Same response from C1SOSVME.
5) I restarted the framebuilder, then rebooted C0DAQCTRL and C1DCU, then power-cycled and restarted C1SOSVME. Niente. Same response from C1SOSVME.
 
The following is the so called "Nuclear Option", the only solution that so far has proven to work in these circumstances. Execute the following steps in the order they are listed, waiting for each step to be completed before passing to the next one.
0) Switch off: the frame builder, the C0DAQCTRL and C1DCU crate, C1DCUEPICS
1) turn on the frame builder
2) reset of the RFM Network switch on 1Y7 (although, it's not sure whether this step is really necessary; but it's costless)
3) turn on C1DCUEPICS
4) turn on the C0DAQCTRL and C1DCU crate
 
One other possibility remains to be explored to avoid the Nuclear Option. And that is to just try to reset both RFM Network switches: the one in 1Y7 and the one in 1Y6.

 

 I burtrestored all the snapshots to Dec 9 2009 at 18:00.

  2384   Thu Dec 10 13:10:25 2009 AlbertoConfigurationLSC166 LO Disconnected

I temporarily disconnected the Heliax cable that brings the 166MHz LO to the LSC rack.

I'm doing a couple of measurement and I'll put it back in as soon as I'm done.

  2385   Thu Dec 10 13:13:08 2009 JenneUpdateComputersfb40m backup restarted

The frame builder was power cycled during the morning bootfest.  I have restarted the backup script once more.

  2386   Thu Dec 10 13:50:02 2009 JenneUpdateVACAll doors on, ready to pump

[Everybody:  Alberto, Kiwamu, Joe, Koji, Steve, Bob, Jenne]

The last heavy door was put on after lunch.  We're now ready to pump.

  2387   Thu Dec 10 15:18:55 2009 JenneUpdateVACseisBLRMS

last 20 days - including the pounding from next door

Attachment 1: Untitled.png
Untitled.png
  2388   Thu Dec 10 16:51:35 2009 steveUpdateVACpumpdown will start tomorrow morning

The vacuum system is at 760 torr  All chambers with doors on and their annuloses are pumped down.

PSL output shutter is still closed. We are fully prepared to star slow pump down tomorrow.

The plan is to reach 1 torr ~ in 6 hrs  without a creating a sand storm.

  2389   Thu Dec 10 17:05:21 2009 AlbertoConfigurationLSC166 MHz LO SMA-to-Heliax connection repaired

I replaced the SMA end connector for the 166 MHZ Local Oscillator signal that goes to the back of the flange in the 1Y2 rack. The connector had got damaged after it twisted when I was tigheting the N connector of the Heliax cable on the front panel.

  2391   Thu Dec 10 17:13:36 2009 kiwamuUpdateSUSFree swiging spectra under the atmospheric pressure

The free swinging spectra of ITMs, ETMs, BS, PRM and SRM were measured last night in order to make sure that nothing wrong have happened by the wiping.

I think there are nothing wrong with ITMs, ETMs, BS, PRM and SRM successfully.

For the comparison, Yoichi's figure in his elog entry of Aug.7 2008 is good, but in his figure somehow PRM spectrum doesn't look correct.

Anyway, compared with his past data, there are no significant changes in the spectra. For PRM which has no counterpart to compare with, its shape of spectra looks similar to any other spectra. So I think PRM is also OK. The measured spectra are attached below.

Indeed the current data are still showing smaller peak height for their pitch and yaw modes (roughly factor of 10 ).
 
Attachment 1: summary_freeswing.pdf
summary_freeswing.pdf summary_freeswing.pdf summary_freeswing.pdf summary_freeswing.pdf summary_freeswing.pdf summary_freeswing.pdf summary_freeswing.pdf
  2392   Thu Dec 10 18:27:48 2009 MottUpdateGeneralUpdated R & T Measurements

Attached are updated plots of the T&R Measurements for a variety of mirrors, and diagrams for the setup used to make the measurements.

T is plotted for the 1064 nm measurement, since these mirrors are highly reflective at 1064, and either R or R&T are plotted for the 532 nm measurement, depending on how larger the R signal is.

As with the previous set of plots, the error bars here are purely statistical, and there are certainly other sources of error not accounted for in these plots.  In general, the T measurement was quite stable, and the additional errors
are probably not enormous, perhaps a few percent.

The mirrors are:

Y1-1037-00.50CC

Y1-2037-45S

Y1-2037-45P

Y1S-1025-0

Y1S-1025-45

 

Attachment 1: Y1S-0.png
Y1S-0.png
Attachment 2: Y1S-45.png
Y1S-45.png
Attachment 3: Y45P.png
Y45P.png
Attachment 4: Y45S.png
Y45S.png
Attachment 5: Y150CC.png
Y150CC.png
Attachment 6: Setup.png
Setup.png
  2393   Thu Dec 10 18:31:44 2009 AlbertoConfigurationLSC166 LO Disconnected

Quote:

I temporarily disconnected the Heliax cable that brings the 166MHz LO to the LSC rack.

I'm doing a couple of measurement and I'll put it back in as soon as I'm done.

 These are the losses I measured on a RG-174 cable for the two frequencies that we're planning to use in the Upgrade:

@11MHz Loss=0.22dBm/meter

@55MHz Loss=0.41dBm/meter

(The cable was 2.07m long. The input signal was +10dBm and the output voltages at the oscilloscope where: Vpk-pk(11MHz)=1.90V, Vpk-pk(11MHz)=1.82V )

  2394   Fri Dec 11 08:35:04 2009 steveConfigurationVACpumpdown has started

Oplev positions before and after drag wiped arm TMs as of yesterday. Slow-mode pumpdown has started with 3/4 turn opened  RV1 valve at 8am today.

Attachment 1: wiping.jpg
wiping.jpg
  2395   Fri Dec 11 09:30:09 2009 KojiConfigurationLSC166 LO Disconnected

They must not be dBm, must be dB

Quote:

Quote:

I temporarily disconnected the Heliax cable that brings the 166MHz LO to the LSC rack.

I'm doing a couple of measurement and I'll put it back in as soon as I'm done.

 These are the losses I measured on a RG-174 cable for the two frequencies that we're planning to use in the Upgrade:

@11MHz Loss=0.22dBm/meter

@55MHz Loss=0.41dBm/meter

(The cable was 2.07m long. The input signal was +10dBm and the output voltages at the oscilloscope where: Vpk-pk(11MHz)=1.90V, Vpk-pk(11MHz)=1.82V )

 

  2396   Fri Dec 11 11:42:26 2009 AlbertoConfigurationLSC166 LO Disconnected

Quote:

They must not be dBm, must be dB

Quote:

Quote:

I temporarily disconnected the Heliax cable that brings the 166MHz LO to the LSC rack.

I'm doing a couple of measurement and I'll put it back in as soon as I'm done.

 These are the losses I measured on a RG-174 cable for the two frequencies that we're planning to use in the Upgrade:

@11MHz Loss=0.22dBm/meter

@55MHz Loss=0.41dBm/meter

(The cable was 2.07m long. The input signal was +10dBm and the output voltages at the oscilloscope where: Vpk-pk(11MHz)=1.90V, Vpk-pk(11MHz)=1.82V )

 

I apologize for the lack of correctness on the units in yesterday's elog entry, but I wasn't very sharp last night.

I repeated the measurement today, this time also making sure that I had a 50ohm input impedance set in the scope. These the results for the losses.

RG-174 Cable 0.053 dB/m @ 11MHz  0.155 dB/m @ 55 MHz

 I also measured the losses in the Heliax cable going from the 166 MHz LO to the LSC rack:

166MHz LO Heliax 0.378 dB @ 11MHz  1.084 dB @ 55 MHz

 

  2397   Fri Dec 11 13:18:17 2009 AlbertoConfigurationVACpumpdown has started

Quote:

Oplev positions before and after drag wiped arm TMs as of yesterday. Slow-mode pumpdown has started with 3/4 turn opened  RV1 valve at 8am today.

 I'm leaving the lab now for less than 2 hours. I should be back in time for when the pumping is finished so that I can measure the finesse again.

  2398   Fri Dec 11 14:12:32 2009 ranaConfigurationLSC166 LO Disconnected

 

 Seems like very strange cable loss numbers. The Heliax is lossier than the RG-174? I wonder how these compare with the specs in the cable catalog?

  2399   Fri Dec 11 14:19:22 2009 KojiConfigurationVACpumpdown has started

Wait, Wait, Wait. You are moving too fast. Go one by one.

Check the PZTs, the MC, initial pointings, IFO mirrors, some of the partial locks, and maybe some momentary full locks?
Once the recover of the IFO is declared, you can proceed to the measurements.

I hope the grad students can take this precious opportunity to have their fun time for restoring everything by themselves.

Quote:

 I'm leaving the lab now for less than 2 hours. I should be back in time for when the pumping is finished so that I can measure the finesse again.

 

  2400   Fri Dec 11 15:21:40 2009 steveConfigurationVACpumpdown#67 is completed

Quote:

Oplev positions before and after drag wiped arm TMs as of yesterday. Slow-mode pumpdown has started with 3/4 turn opened  RV1 valve at 8am today.

 Pump down is completed. Valve configuration is VACUUM NORMAL. CC1 pressure is in the ~8 e-5 torr   PSL output shutter is opened.

Attachment 1: pd7h.jpg
pd7h.jpg
Attachment 2: oplevpd8h.jpg
oplevpd8h.jpg
  2401   Fri Dec 11 17:36:37 2009 kiwamuUpdateGeneralIFO restoring plan

Alberto, Jenne, Kiwamu

 

We together will lead the IFO restoring and the following is our plan.

- - - - -|- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

#.0     |  measuring the free swinging spectra                     (weekend by kiwamu)   DONE

#.1     |  turn ON the PZTs for steering mirror and so on.         (Dec.14 Mon.) DONE

#. 1            |    lock around PSL  DONE

#.2     |  deal with mechanical shutter                            (Dec.14 Mon.)DONE

#.3     |  lock MCs                                                (Dec.14 Mon.)DONE

#.4     |  align the IFO                                           (Dec.15 Tue.)DONE

#.5     |  lock full IFO                                           (Dec.15 Tue.)DONE

- - - - -|- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

 

Thank you.

  2402   Fri Dec 11 19:51:13 2009 AlbertoConfigurationLSC166 LO Disconnected

Quote:

 

 Seems like very strange cable loss numbers. The Heliax is lossier than the RG-174? I wonder how these compare with the specs in the cable catalog?

In my last entry there was a typo for the measurement of the Heliax at 55 MHz. I corrected it. It was 1.084 dB instead of 1.084 dB/m.
For the Heliax I don't have the measurement of the loss per meter since I don't know the cable actual length.
 
Except for that, I checked the values I found on the Internet.
My measurements for the RG-174 seem comparable to the loss specified in the catalog (here): 6.6dB in 100ft @ 55 MHz, that is 0.22 dB/m, that compare with 0.155 dB/m that I measured.

I did the measurement on a 4.33 meter long cable with SMA connectors at the ends.

  2403   Sat Dec 12 07:36:56 2009 ranaHowToElectronicsHow to Measure the Length of a Cable: Interferometry

Need to measure the length of the cable, but too lazy to use a measuring tape?

Then you too can become an expert cable length measurer by just using an RF signal generator and a scope:

  1. Disconnect or short (not 50 Ohm term) the far side of the cable.
  2. Put a T on the near side of the cable.
  3. Drive the input of the T with your signal source.
  4. Look at the output of the T with the scope while sweeping the signal source's frequency knob.

The T is kind of acting like a beamsplitter in an asymmetric length Michelson in this case. Just as we can use the RF phase shift between the arms to measure the Schnupp asymmetry, we can also use a T to measure the cable length. The speed of light in the cable is documented in the cable catalog, but in most cases its just 66% of the speed of light in the vacuum.

  2405   Sun Dec 13 17:43:10 2009 kiwamuUpdateSUSfree swinging spectra (vacuum)

The free swinging spectra of ITMs, ETMs, BS, PRM and SRM were measured under the vacuum-condition. The attachment are measured spectra.

It looks there are nothing wrong because no significant difference appear from the past data and the current data (under atmosperic pressure).

So everything is going well.

Attachment 1: summary_FreeSwinging_vacuum.pdf
summary_FreeSwinging_vacuum.pdf summary_FreeSwinging_vacuum.pdf summary_FreeSwinging_vacuum.pdf summary_FreeSwinging_vacuum.pdf summary_FreeSwinging_vacuum.pdf summary_FreeSwinging_vacuum.pdf summary_FreeSwinging_vacuum.pdf
  2406   Sun Dec 13 20:50:45 2009 ranaSummaryIOOMach Zender Calibration

I ramped the MZ PZT (with the loop disabled on the input switch) to calibrate it. Since the transmission has been blocked, I used the so-called "REFL" port of the MZ to do this.

The dark-to-dark distance for the MZ corresponds to 2 consecutive destructive interferences. Therefore, that's 2 pi in phase or 1 full wavelength of length change in the arm with the moving mirror.

Eyeballing it on the DTT plot (after lowpassing at 0.1 Hz) and using its cursors, I find that the dark-to-dark distance corresponds to 47.4 +/- 5 seconds.

So the calibration of the MZ PZT is 88 +/- 8 Volts/micron.

Inversely, that's a mean of 12 nm / V.

why am I calibrating the MZ? Maybe because Rob may want it later, but mainly because Koji won't let me lock the IFO.

Apparently, we haven't had a fast channel for any of the MZ board. So I have temporarily hooked it up to MC_DRUM at 21:13 and also turned down the HEPA. Now, let's see how stable the MZ and PMC really are overnight.

EDIT: it railed the +/- 2V ADCwe have so I put in a 1:4 attenuator via Pomona box. The calibration of MC_DRUM in terms of MZ_PZT volts is 31.8 cts/V.

So the calibration of MC_DRUM1 in meters is: 0.38 nm / count


Attachment 1: Untitled.png
Untitled.png
  2407   Sun Dec 13 23:18:09 2009 ranaSummaryIOODisplacement noise on the PSL table

For the Laser Gyro, I wondered how much mechanical noise we might get with a non-suspended cavity. My guess is that the PMC is better than we could do with a large ring and that the MZ is much worse than we could do.

Below 5 Hz, I think the MZ is "wind noise" limited. Above 10 Hz, its just ADC noise in the readout of the PZT voltage.

Attachment 1: mz.pdf
mz.pdf
  2408   Mon Dec 14 00:37:28 2009 KojiOmnistructureEnvironmentAnts in the coffee maker

I made a short stop at the 40m on Sunday night and found that hundreds ants are in the coffee maker.
I removed ants around the sink and washed the coffee maker.

It looked the ants were everywhere in the lab tonight. They seemed to prefer warm places like in the coffee maker and below the coffee mill.
So, I recommend that Steve should confirm there is no ants in the coffee maker again before the first coffee of the week is made.
Othewise they will add some more acidity to your cup.

 

  2409   Mon Dec 14 11:21:23 2009 steveOmnistructureEnvironmentAnts in the coffee maker

 

 We still had some ants visiting the sink area this morning. These ants seem to be addicted to our our Peet's coffe

Spectracide: Bug Stop insect killer was sprayed. Please wash your eating dishes well ! and keep area clean.

 

  2410   Mon Dec 14 12:13:52 2009 JenneUpdateTreasureWe are *ROCKSTARS* ! IFO is back up

[Jenne, Kiwamu, Koji]

We got the IFO back up and running!  After all of our aligning, we even managed to get both arms locked simultaneously.  Basically, we are awesome. 

 This morning, we did the following:

*  Turned on the PZT High voltages for both the steering mirrors and the OMC.  (For the steering mirrors, turn on the power, then hit "close loop" on each.  For the OMC, hit Output ON/OFF).

*  Looked at the PZT strain gauges, to confirm that the PZTs came back to where they had been.  (Look at the snapshot of C1ASC_PZT_Al)

*  Locked all components of the PSL (This had already been done.)

*  Removed beam dump which was blocking the PSL, and opened the PSL mechanical shutter.  Light into the IFO!

*  Locked the Mode Cleaner.  The auto-locker handled this with no problem.

*  Confirm that light is going through the Faraday.  (Look at the TV sitting on top of MC13 tank...it shows the Faraday, and we're hitting the input of the Faraday pretty much dead-on).

*  Look at IP_ANG and IP_POS.  Adjust the steering mirrors slightly to zero the X&Y readings on IP_ANG.  This did not change the PZTs by very much, so that's good.

*  Align all of the Core Optics to their OpLev positions.

*  On the IFO_Align screen, save these positions.

*  Run the IFO_Configure scripts, in the usual order.  (Xarm, Yarm, PRM, DRM).  Save the appropriate optics' positions after running the alignment scripts.  We ended up running each alignment script twice, because there was some residual misalignment after the first iteration, which we could see in the signal as viewed on DataViewer (Either TRX, TRY, or SPOB, for those respective DoFs).

*  Restore Full IFO.

*  Watch the beauty of both arms and the central cavity snapping together all by themselves!  In the attached screenshot, notice that TRX and TRY are both ~0.5, and SPOB and AS166Q are high.  Yay!

Conclusions: 

*  The wiping may have helped.  While aligning X and Y separately, TRX got as high as ~1.08, and TRY got as high as 0.98  This seems to be a little bit higher than it was previously.

*  Since everything locked up in pretty short order, and the free swinging spectra (as measured by Kiwamu in elog 2405) looks good, we didn't break anything while we were in the chambers last week.  Excellent.

*  We are now ready for a finesse measurement to tell us more quantitatively how we did with the wiping last week.

 

Attachment 1: Jenne14Dec09_IFOlocked.png
Jenne14Dec09_IFOlocked.png
  2411   Mon Dec 14 13:08:33 2009 KojiUpdateTreasureLOCKSTARS

Good job guys. What I did was saying "I don't know", "Maybe", and "Ants...".

Now you can proceed to measurements for the visibility and the cavity pole! 

Quote:

[Jenne, Kiwamu, Koji]

We got the IFO back up and running!  After all of our aligning, we even managed to get both arms locked simultaneously.  

  2412   Mon Dec 14 13:17:33 2009 robUpdateTreasureWe are *ROCKSTARS* ! IFO is back up

 

 

Attachment 1: two-thumbs-up.jpeg
two-thumbs-up.jpeg
  2413   Mon Dec 14 14:16:14 2009 AlbertoUpdateTreasureLOCKSTARS

Quote:

Good job guys. What I did was saying "I don't know", "Maybe", and "Ants...".

Now you can proceed to measurements for the visibility and the cavity pole! 

Quote:

[Jenne, Kiwamu, Koji]

We got the IFO back up and running!  After all of our aligning, we even managed to get both arms locked simultaneously.  

 I'm going to do it right now.

  2414   Mon Dec 14 15:18:18 2009 JenneUpdatelorearmLoss script ran....results confidential

I ran the armLoss script for both Xarm and Yarm.  The results are confidential, pending the completion of Alberto's cavity pole/finesse measurement due to the 'bet' as to what the new losses are after the drag wiping.

If you're the kind of person who likes to look at their Chrismas presents, the log files with the results are in the usual place for this script: /scripts/LSC/loss-ARM-GPStime.log  (loss-Y-944865071.log and loss-X-944865946.log)

  2415   Mon Dec 14 19:33:04 2009 AlbertoUpdateGeneralArm Cavity Poles measured again after cleaning the optics last week

Last week we vented and we cleaned the main optics of the arm cavities.

I measured the frequency of the cavity poles for both the arm cavities to see how they changed (see previous elog entry 2226). These the results:

fp_X = 1616 +/- 14 Hz

fp_Y = 1590 +/- 4 Hz

The error is the statistical error that I got with the Matlab NonLinearLeastSquare fitting function.
 
I calculated the cavity pole frequencies by measuring the transfer function between a photodiode located at the end of the arms (either X or Y) and another photodiode placed after the mode cleaner. Both diodes where Thorlabs PDA255.
(Last time, I had measured that the pair of diode had a flat calibration).
 
With the SR785 I measured the transfer function by exciting the OMC_ISS_EXC input cable.
For both arms I set to 1V the excitation amplitude. I repeated the measurements for different excitation amplitudes without observing any changes.
I then fitted the data with the NonLinearLeastSquare function of matlab. The script I wrote to do that is attached to this entry in a compressed file.
The files also contains the PDF with the output plots and the data from both set of measurements performed before and after the cleaning.
The data is commented in a file called measurements.log.
 
In the end I disabled again the test switch on the ISS MEDM screen.
I disconnected the excitation cable from the OMC_ISS_EXC input cable.
I removed the photodiode that measured the Mode Cleaner transmission from the PSL clearing the way for the beam to get back to its path to the RFAM photodiode.
Attachment 1: XarmTF01_fit.pdf
XarmTF01_fit.pdf
Attachment 2: YarmTF01_fit.pdf
YarmTF01_fit.pdf
Attachment 3: ArmCavityFinesseMeasurment.tar.gz
  2416   Mon Dec 14 22:32:56 2009 KojiUpdateGeneralArm cavity loss ~ result

I like to ask someone to review the calculation on the wiki.

In the calculation, the round trip loss and the front mirror T are the unknown variables.
The end mirror T of 10ppm was assumed. (End mirror T)+(Round trip loss) is almost invariant, and T_end does not change the other results much.

Arm cavity loss measurement (Dec. 14, 2009)

X Arm:
  • Arm visibility (given): 0.897 +/- 0.005 (20 pts) (2.5%UP!!)
  • Cut off freq (given): 1616 +/- 14 [Hz] (2.1%UP!!)
  • Finesse (derived): 1206 +/- 10 (2.1%UP!!)
  • Round Trip loss (estimated): 127 +/- 6 [ppm] (28%DOWN!!)
  • Front Mirror T (estimated): 0.00506 +/- 0.00004
Y Arm:
  • Arm visibility (given): 0.893 +/- 0.004 (20 pts) (2.1%UP!!)
  • Cut off freq (given): 1590 +/- 4 [Hz] (8.2%UP!!)
  • Finesse (derived): 1220 +/- 3 (8.2%UP!!)
  • Round Trip loss (estimated): 131 +/- 6 [ppm] (37%DOWN!!)
  • Front Mirror T (estimated): 0.00500 +/- 0.00001 

Previous measurement (Oct 07, 2009 & Nov 10, 2009)

X Arm:  

  • Arm visibility (given): 0.875 +/- 0.005 (34 pts)
  • Cut off freq (given): 1650 +/- 70 [Hz]
     
  • Finesse (derived): 1181 +/- 50
  • Round Trip loss (estimated): 162 +/- 10 [ppm]
  • Front Mirror T (estimated): 0.0051 +/- 0.0002

Y Arm: 

  • Arm visibility (given): 0.869 +/- 0.006 (26 pts)
  • Cut off freq (given): 1720 +/- 70 [Hz]
     
  • Finesse (derived): 1128 +/- 46
  • Round Trip loss (estimated): 179 +/- 12 [ppm]
  • Front Mirror T (estimated): 0.0054 +/- 0.0002
  2417   Tue Dec 15 00:02:10 2009 ranaUpdatePEMNoise of the Ranger SS-1 Seismometer

I wanted to see what the noise of the Ranger seismometer should be. I used LISO and file ranger.fil (in our LISO SVN) to calculate the voltage noise referred to the input. In this model, we represent the EMF from the moving magnet in the coil as a voltage source at 'nin' which drives the coil impedance. This is the same approach that Brian Lantz uses in his noise modeling of the GS-13 (PDF is on our Ranger wiki page).

In the simulation, I used the OP27 as a placeholder for the SR560 that we use (I don't know the current noise of the SR560). To do this, I use the new 'inputnoise' feature in LISO (its in the README, but not in the manual).

You can see that we would be limited by the input current noise of the OP27. So we would do a lot better if we used an FET based readout amp like the AD743 (or equivalent) or even better using the new multi-FET readout circuit that Rich Abbott has developed. Clearly, its also silly to have a load resistance in there - I put it in because the manual says to do it, but all it does is damp the mass and reduce the size of the signal.

#  Noise sim for the Ranger SS-1 seismometer
#
#                            \
#                           | \            
#           n2- - - ----  - |  \          
#            |    |         | op1>-- n4 - r4 -- no
#           Rg   RL     n3- |  /     |           
#       n1 - |    |     |   | /      |                
#           Lg    |     |    /       |                        
#            |    |     | - - - R2 - -                    
#           nin  gnd   R1 
#                       |
#                      gnd

We previously measured the Ranger's self noise by locking it down.

The 1/f^3 noise that we see below 1 Hz is roughly consistent with the noise model: to get from my plot into meters you have to multiply by:

(1 + f)^2

----------

340 * f^2

P.S. Secret PDF handshake: You can make your non-compliant applications like LISO or DTT produce a thumbnailing PDF by using Acrobat to open the file and export it as PDF/A.

In the second attachment, I have used an OPA827 (new low-noise FET input amp from TI) as the readout amplifier. This seems like a good choice - main drawback is that Digikey backordered my OPA827s by 19 weeks!

Attachment 1: rangerx.pdf
rangerx.pdf
Attachment 2: ranger827.pdf
ranger827.pdf
  2418   Tue Dec 15 05:29:31 2009 AlbertoUpdateGeneralArm Cavity Poles measured again after cleaning the optics last week

 

 The Y arm cavity pole moved down by 130 Hz, whereas the X arm moved by only 34 Hz. I wonder if that is because Kiwamu spent much more time on cleaning ITMY than on any other optic.

  2419   Tue Dec 15 17:16:22 2009 KojiUpdateGeneralTable distance measurements

During the vent we have tried to measure the distances of the optical tables for BS-ITMX and BS-ITMY.
We need to take into account the difference between the AutoCAD drawing and the reality.

X direction distance of the table center for BS and ITMX:
84.086" (= 2135.8mm)
(This is 84.0000" in AutoCAD drawing)

Y direction distance of the table center for BS and ITMX:
83.9685" (= 2132.8mm)
(This is 83.5397" in AutoCAD drawing)

We used two scales attached each other in order to measure the distance of the certain holes on the tables.

We got more numbers that were estimated from several separated measurements.
I think they were not so accurate, but just as a record, I also put the figure as an attachment 2.

Attachment 1: Table_distance_by_metal_scale.pdf
Table_distance_by_metal_scale.pdf
Attachment 2: Table_distance_by_chambers.pdf
Table_distance_by_chambers.pdf
  2420   Tue Dec 15 21:39:34 2009 AlbertoUpdateABSLbrief summary of this afternoon's measurements
I took measurements of the open loop gain of the AbsL PLL with the old Universal PDH Box.
I Also measured the filter shape of both the new and the old PDH box.
I'm going to plot the results in a nice form tomorrow morning.
For who's interested, the PLL UGF was at 10KHz.
 
I can't lock the PLL with the new PDH box. Measuring its filter's shape, as suggested by Koji, I found out that it differs from the old one. That despite the fact that the two boxes should share the same circuit schematic. O,r at least, that is what it looks like from the schematics in the DCC.
I need to understand whether that is intentional and, if that was the case, what kind of use  Rich Abbott designed it for.
 
Tomorrow I'm going to post in the elog the filter's transfer functions too.
 
Before leaving the lab I closed the auxiliary laser's mechanical shutter.
  2421   Wed Dec 16 11:21:20 2009 AlbertoUpdateABSLUniversal PDH Box Servo Filters
Yesterday I measured the shape of the servo filter of both the old and the new Universal PDH boxes.
Here they are compared.

NewandOlfFilterTF.png

The way the filter's transfer function has been measured is by a swept sine between the "SERVO INPUT" and the "PIEZO DRIVE OUTPUT" connection on the box front panel. The spectrum analyzer used for the measurement is the SR785 and the source amplitude is set at 0.1V.

The two transfer functions are clearly different. In particular the old one looks like a simple integrator, whereas the new one already includes some sort of boost.

That probably explains why the new one is unable to lock the PLL. Indeed what the PLL needs, at least to acquire lock, is an 1/f filter.

I thought the two boxes were almost identical, at least in the filter shapes. Also the two schematics available in the DCC coincide.

Attachment 1: NewandOlfFilterTF.png
NewandOlfFilterTF.png
ELOG V3.1.3-