  12882   Fri Mar 10 19:48:56 2017 johannesUpdateComputer Scripts / Programsloss script

Loss script running again, on Pianosa this time. Due to an oversight in the code the beam wasn't actually moved across ETMY last night. This time I confirmed that the correct offset value is written as a demodulation parameter to the correct mirror degree of freedom. Script will probably run through the night. Yarm is currently misaligned but previous alignment was saved.

  12883   Sat Mar 11 20:11:58 2017 johannesUpdateComputer Scripts / Programsloss script

Yarm script running on Pianosa. Still working on visualization of the ETMX lossmap.

Quote:

Loss script running again, on Pianosa this time. Due to an oversight in the code the beam wasn't actually moved across ETMY last night. This time I confirmed that the correct offset value is written as a demodulation parameter to the correct mirror degree of freedom. Script will probably run through the night. Yarm is currently misaligned but previous alignment was saved.

 

  12923   Sun Apr 2 23:14:30 2017 ranaUpdateComputer Scripts / Programsnodus update/upgrade/reboot

I just did remote apt-get update, apt-get upgrade, and then reboot on nodus. ELOG started up by itself.

  12992   Mon May 15 19:21:04 2017 KojiUpdateComputer Scripts / ProgramsFSSslow / MCautolocker restarted

It seems that FSS slow servo stopped working.

I found that megatron was restarted (by Rana, to finish an apt-get upgrade) on ~18:47 PDT today.

controls@megatron|~> last -5
controls pts/0        192.168.113.216  Mon May 15 19:15   still logged in   
controls pts/0        192.168.113.216  Mon May 15 19:14 - 19:15  (00:01)    
reboot   system boot  3.2.0-126-generi Mon May 15 18:50 - 19:19  (00:29)    
controls pts/0        192.168.113.200  Mon May 15 18:43 - down   (00:04)    
controls pts/0        192.168.113.200  Mon May 15 15:25 - 17:38  (02:12)


FSSslow / MCautolocker were restarted on megatron.

  13013   Thu May 25 16:42:41 2017 jigyasaUpdateComputer Scripts / ProgramsMaking pylon installation on shared directory

I have been working on interfacing with the GigE’s. I went through Joe Be’s paper and the previous elogs and verified that the code files are installed.

I then downloaded and extracted a copy of the Pylon software onto my home directory on Allegra. Gautam helped me find installation instructions on Johannes’ directory so that I could make the installation on the shared directory.

So far , according to instructions, these commands need to be executed so that the installation takes place and the rules for camera permissions are set up.

sudo tar –C /opt/rtcds/caltech/c1/scripts/GigE –xzf pylon SDK*.tar.gz

followed by ./setup-usb.sh

The Pylon viewer can then be accessed with /scripts/GigE/pylon5/bin/PylonViewerApp 

Should I go ahead with the installation in the shared directory?

  13014   Thu May 25 18:37:11 2017 jigyasaUpdateComputer Scripts / ProgramsMaking pylon installation on shared directory

Gautam helped me execute the commands mentioned above and Pylon has now been installed on the shared directory. We extracted the pylon installation from Johannes's directory to the shared drive and executing the command tar –C /opt/rtcds/caltech/c1/scripts/GigE –xzf pylon SDK*.tar.gz created an unzipped pylon5 folder in /scripts. The ./setup-usb.sh set up the udev rules for the GigE.

The installation took place without any errors.

The Pylon viewer app can now be accessed at /opt/rtcds/caltech/c1/scripts/GigE/pylon5/bin followed by ./PylonViewerApp 

Quote:

Should I go ahead with the installation in the shared directory?

 

  13023   Wed May 31 14:23:42 2017 jigyasaUpdateComputer Scripts / ProgramsEstablishing the EPICs channels for the GigE

To set up the EPICs channels for the GigE, Gautam and I followed the steps in the elog by him  8957 .

We copied the 11 required channels from scripts/GigE/SnapPy/example_camera.db to c1cam.db that we created, however due to conflicts with the existing CAM-AS_PORT channels, the channels could not be accessed.

We later changed the database file to Video.db and on restarting the slow machine, it was verified that the channels indeed could be written to and read from.

11 channels were added

C1: CAM-MC1_X (X centroid position)
C1: CAM-MC1_Y (Y centroid position)
C1: CAM-MC1_WX (Gaussian width in the X direction)
C1: CAM-MC1_WY (Gaussian width in the Y direction)
C1: CAM-MC1_XY (Gaussian width along XY line)
C1: CAM-MC1_SUM (Pixel sum)
C1: CAM-MC1_EXP (Exposure time in microseconds)
C1: CAM-MC1_SNAP (Control signal for taking snapshots)
C1: CAM-MC1_FILE(File name for image to saved to - time stamp automatically appended)
C1: CAM-MC1_RELOAD (Reloads configuration file)
C1: CAM-MC1_AUTO (1 means autoexposure on, 0 means autoexposure off)

The procedure followed –

  • Add the channel names to the file C0EDCU.ini (path = /opt/rtcds/caltech/c1/chans/daq/C0EDCU.ini).
  • Make a database (.db) file so that these channels are actually recorded (path = /cvs/cds/caltech/target/c1aux/Video.db).
  • Restarted the slow machine. FB
  • Verify that the channels indeed exist and can be read and written to using ezcaread and ezcawrite.

GV: Initially, I made a new directory called c1cam in /cvs/cds/caltech/target/ and made a .db file in there. However, the channels were not accessible after re-starting FB (attempting to read these channels threw up the "Channel does not exist" error). On digging a little further, I saw that there were already some "C1:CAM-AS_PORT" channels in C0EDCU.ini. The corresponding database records were defined inside /cvs/cds/caltech/target/c1aux/Video.db. So I just added the new records there. I also had to uncomment out the dummy channel in C0EDCU.ini to keep an even number of channels. Restarting FB still did not allow read/write access to the channels. Looking through the files in /cvs/cds/caltech/target/c1aux, I suspected that the epics database records are loaded when the machine is first booted up - so on a hunch I re-started c1aux by keying the crate, and this did the trick. The channels can now be read / written to (tested using Python cdsutils).

  13056   Fri Jun 9 16:37:29 2017 jigyasaUpdateComputer Scripts / ProgramsOpenCV installation

OpenCV 3.1.0 has been installed by following the commands locally on Donatella

git clone https://github.com/Itseez/opencv.git
cd opencv
git checkout 3.1.0
git clone https://github.com/Itseez/opencv_contrib.git
cd opencv_contrib
git checkout 3.0.0
cd ~/opencv
mkdir release
cd release
cmake -D CMAKE_BUILD_TYPE=RELEASE -D CMAKE_INSTALL_PREFIX=/usr/local -D OPENCV_EXTRA_MODULES_PATH=/~/opencv_contrib/modules/ ~/opencv/

In ~/opencv/release, make and sudo make install were executed.

This completed the installation. The version of the installation was verified pkg-config --modversion opencv which showed 3.1.0. Also verified the import of cv2 module in python and it seems to work fine. 

 

  13066   Thu Jun 15 18:56:31 2017 jigyasaUpdateComputer Scripts / ProgramsMC2 Pitch-Yaw offset

A python script to randomly vary the MC2 pitch and yaw offset and correspondingly record the value of MC transmission has been started on Donatella in the control room and should run for a couple of hours overnight.

The script is named MC_TRANS_1.py and is located in my user directory at /users/jigyasa

Apologies for any inconvenience.
Data analysis will follow.

  13072   Mon Jun 19 18:32:18 2017 jigyasaUpdateComputer Scripts / ProgramsSoftware Installation for image analysis

The IRAF software from the National Optical Astronomy Observatory has been installed locally on Donatella(for testing) following the instructions listed here at http://www.astronomy.ohio-state.edu/~khan/iraf/iraf_step_by_step_installation_64bit
This is a step towards "aperture photometry" and would help identify point scatterers in the images of the test masses.

I will be testing this software, in particular, the use of DAOPHOT and if it seems to work out, we may install it on the shared directory.
Hope this isn't an inconvenience.

 

 

  13073   Mon Jun 19 18:41:12 2017 jigyasaUpdateComputer Scripts / ProgramsMC2 Pitch-Yaw offset

The previous run of the script had produced some dubious results!

The script has been modified and now scans the transmission sum for a longer duration to provide a better estimate on the average transmission. The pitch and yaw offsets have been set to the values that were randomly generated in the previous run as this would enable comparison with the current data.

I am starting it on Donatella and it should run for a couple of hours.

Apologies for the inconvenience.

Quote:

A python script to randomly vary the MC2 pitch and yaw offset and correspondingly record the value of MC transmission has been started on Donatella in the control room and should run for a couple of hours overnight.

The script is named MC_TRANS_1.py and is located in my user directory at /users/jigyasa

  13076   Tue Jun 20 17:44:12 2017 jigyasaUpdateComputer Scripts / ProgramsMC2 Pitch-Yaw offset

The script didn't run properly last night, due to an oversight of variable names! It's been started again and has been running for half an hour now.

Quote:

I am starting it on Donatella and it should run for a couple of hours.

Apologies for the inconvenience.

Quote:

A python script to randomly vary the MC2 pitch and yaw offset and correspondingly record the value of MC transmission has been started on Donatella in the control room and should run for a couple of hours overnight.

The script is named MC_TRANS_1.py and is located in my user directory at /users/jigyasa

 

  13077   Fri Jun 23 02:43:43 2017 KaustubhHowToComputer Scripts / ProgramsTaking Measurements From AG4395A

Summary:

I have written a code(a basic one which needs a lot of improvements, but still does the job) for taking multiple measurements from the AG4395A. I have also written a separate code for plotting the data taken from the previoius code along with the error bars upto 1 standard deviation.

 

Details on How To Operate AG4395A:

  1. Under 'Measurement' tab, press the 'Meas' button and select the Analyzer Type (Network Analyzer or Spectrum Analyzer).
  2. Then under the same options select which 'ratio' needs to be measured (A/R, B/R or A/B).
  3. Then press the 'Format' button to select what needs to be measured (Eg - Log|Mag|, Phase, etc.).
  4. In order to measure and see two channels at the same time (Eg - Log|Mag| and Phase), press the 'Display' button and select 'Dual Channel'.
  5. Using the 'Scale' button we can set the scale/div or use autoscale and also set the attenuator values of the different channels.
  6. The 'Bw/Avg' option gives us an averaging option which averages few sets of data to produce the result. In doing this we lose quiet a lot of data and the resulting plot isn't able to give us the information on the statistical errors.
  7. This option also allows us to set the 'Intermediate Frequency' Bandwidth. This basically dictates the sampling rate of the Analyzer. The lower the IF bw, the higher is lesser is the noise (due to less uncertainty in Frequency).
  8. The 'Cal' button helps us calibrate the Analyzer to the current connections and signals. This is done because there is usually a difference in the 'cable lengths' for the two channels which introduces an extra phase term depnding upon the rf frequency. The calibration can be simply done by removing the Device Under Test (DUT) and diectly connecting the coaxial cables to the channels. After this the 'Calibrate Menu' allows us to calibrate the response using the short, open and thru methods.
  9. Now, under the 'Sweep' tab, the 'Sweep' button allows us to select various sweep options such as 'Sweep Time' (Auto, or set a time), 'Number of Points' (b/w 201-801) and 'Sweep Type' (Linear, Log, List Freq. etc.).
  10. Using the 'Source' button we can set the source power in dBm units (Usually kept as -20 to -10 dBm).
  11. The Scan Range can be set in a few ways such as using the start and end points or using the center and span range/width.
  12. After setting up all of the above, we can take the measurement either from the analyzer itself or using one of the control PCs. The command to download the data from AG4395A is netgpibdata -i 192.168.113.105 -d AG4395A -a 10 -f [filename].

 

Brief Details on How the 'AGmeasure' command works:

AGmeasure is a python script developed by some of the people who work at 40m. It is set as a global command and can be used from within any directory. The source code is in the scripts folder on the network, or else it can also be found in Eric Quintero's git repository. This command accepts at the very least a parameter file. This is supposed to be a .yml file. A template (TFAG4395Atemplate.yml) can be found in the scripts folder or in Eric's repo. There are some other options that can be passed to this command, see the help for more details.

 

The Multi_Measurement Script:

This script calls the 'AGmeasure' command repetitively and keeps storing the data files in a folder. Right now, the script needs to be fed in th template file manually at prompt.

 

The Test_Plotting Script:

This script plots the a set of data files obtained from the above mentioned script and produces a plot along with the errors bands upto 1 standard deviation of the data. The format (names) and total number of text files need to be explicitly known, for now at least.

 

Attachments:

  1. The output test files and the two scripts.
  2. This is the 'Bode Plot' for a data set made using the above two scripts.

 

To Do:

  • Improve upon the two scripts to be as compatible as the AGmeasure function itself.
  • Try and incorporate the whole script into AGmeasure itself along with improving upon the templates.
  • The above details, with some edits perhaps, can go into the 40m wiki too(?).

 

Update: Increased the font size in the plot. Added a few comments to the two scripts

To Do: Need to consider the transfer function as a single physical quantity (both the magnitude and phase) and then take the averages and calculate the standard deviation and then plot these results. 

 

EDIT:

The attachment with the test files and the code now also contains a pdf with all the relations/equations I have used to calculate the averages and errors.

Attachment 1: Test_Files_and_Code.zip
Attachment 2: Bode_Plot_with_Error_Bands.pdf
Bode_Plot_with_Error_Bands.pdf
  13078   Fri Jun 23 02:55:18 2017 KaustubhUpdateComputer Scripts / ProgramsScript Running

I am leaving a script running on the Pianoso for the night. For this purpose, even the AG4395A is kept on. I'll see the result of the script in the morning (it should be complete by then). Just check so before fiddling with the Analyzer.

Thank you.

  13084   Tue Jun 27 18:47:49 2017 jigyasaUpdateComputer Scripts / ProgramsMC2 Pitch-Yaw offset

The values generated from the script were analyzed and a 3D scatter plot in addition to a 2D map were plotted. 

Yesterday, Rana pointed me to another method of collecting and analyzing the data. So I worked on the code today and have left a script (MC2rerun.py) running on Ottavia which should run overnight.

Quote:

The script didn't run properly last night, due to an oversight of variable names! It's been started again and has been running for half an hour now.

 

 

  13086   Thu Jun 29 00:13:08 2017 KaustubhUpdateComputer Scripts / ProgramsTransfer Function Testing

In continuation to my previous posts, I have been working on evaluating the data on transfer function. Recently, I have calculated the correlation values between the real and imaginary part of the transfer function. Also I have written the code for plotting the transfer function data stream at each frequency in the argand plane just for referring to. Also I have done a few calculations and found the errors in magnitude and phase using those in the real and imaginary parts of the transfer function. More details for the process are in this git repository.

The following attachments have been added:

  1. The correlation plot at different frequencies. This data is for a 100 data files.
  2. The Test files used to produce the abover plot along with the code for the plotting it as well as the text file containing the correlation values. (Most of the code is commented as that part wasn't needed fo rhte recent changes.)

 

Conclusion:

Seeing the correlation values, it sounds reasonable that the gaussian in real and imaginary parts approximation is actually holding. This is because the correlation values are mostly quite small. This can be seen by studying the distribution of the transfer function on the argand plane. The entire distribution can be seen to be somewhat, if not entirely, circular. Even when the ellipticity of the curve seems to be high, the curve still appears to be elliptical along the real and imaginary axes, i.e., correlation in them is still low.

 

To Do:

  1. Use a better way to estimate the errors in magnitude and phase as the method used right now is a only valid with the liner approximation and gives insane values which are totally out of bounds when the magnitude is extrmely small and the phase is varying as mad.
  2. Use the errors in the transfer function to estimate the coherence in the data for each frequency point. That is basically plot a cohernece Vs frequency plot showing how the coherence of the measurements vary as the frequency is varied.

 

In order to test the above again, with an even larger data set, I am leaving a script running on Ottavia. It should take more than just the night(I estimate around 10-11 hours) if there are no problems.

Attachment 1: Correlation_Plot.pdf
Correlation_Plot.pdf
Attachment 2: 2x100_Test_Files_and_Code_and_Correlation_Files.zip
  13087   Thu Jun 29 10:04:18 2017 jigyasaUpdateComputer Scripts / ProgramsMC2 Pitch-Yaw offset

The script is being executed again, now.

Quote:

I worked on the code today and have left a script (MC2rerun.py) running on Ottavia which should run overnight.

 

 

  13105   Mon Jul 10 17:13:21 2017 jigyasaUpdateComputer Scripts / ProgramsCapture image without pylon GUI

Over the day, I have been working on a C++ program to interface with Pylon to capture images and reduce dependence on the Pylon GUI. The program uses the Pylon header files along with opencv headers. While ultimately a wrapper in python may be developed for the program, the current C++ program at, 

/users/jigyasa/GigEcode/Grab/Grab.cpp when compiled as

g++ -Wl,--enable-new-dtags -Wl,-rpath,/opt/pylon5/lib64 -o Grab Grab.o -L/opt/pylon5/lib64 -Wl,-E -lpylonbase -lpylonutility -lGenApi_gcc_v3_0_Basler_pylon_v5_0 -lGCBase_gcc_v3_0_Basler_pylon_v5_0 `pkg-config opencv --cflags --libs`

returns an executable file named Grab which can be executed as ./Grab

This captures one image from the camera and displays it, additionally it also displays the gray value of the first pixel.

I am working on adding more utility to the program such as manually adjusting exposure, gain and also on the python wrapper (Cython has been installed locally on Ottavia for the purpose)!

  13109   Mon Jul 10 21:31:15 2017 KaustubhHowToComputer Scripts / ProgramsDetails on Cavity Scan Analysis

Summary:

The following elog describes the procedure followed for generating a sample simulation for a cavity scan, fitting an actual cavity scan and calculating the relevant paramaters using the cavity scan and fit data.

 

1. Cavity Scan Simulation:

  1. First, we define the sample cavity parameters, i.e., the reflectivitie,transmissivities of the mirrors, the RoCs of the mirrors and the absolute cavity length.
  2. We then define a frequency range using numpy.linspace function for which we want to take a scan.
  3. We then define a function that returns the tranmission power output of a Fabry-Perot cavity using the cavity equations as follows: P_{t} = \frac{t_{1}t_{2}}{1-r_{1}r_{2}\exp({\frac{4\pi Lf}{c}+(n+m+1)\phi_G)}} where Pt is the transmission power ratio of the output power to input power, t1,t2,r1,r2 are the transmissivities and reflectivities of the two mirrors, L is the absolute cavity length, f is the frequency of the input laser, c is the speed of light, \phi_G = \arccos{g_{1}g_{2}} is the gouy phase shift with g1,g2 being the g-factors for the two cavity mirrors(g=1-L/R). 'n' and 'm' correspond to the TEMnm higher order mode.
  4. We now obtain a cavity scan by giving the above defined function the cavity parameters and by adding the outputs for different higher order mode('n', 'm' values). Appropriate factors for the HOMs need to be chosen. The above function with appropriate coefficients can be used ti also add the modulated sidebands to the total transmission power.
  5. To this obtained total power we can add some random noise using numpy modules random.normal function. We need to normalise the data with respect to the max. power transmission ratio.
  6. We can now perform fitting on the above data using the procedure stated in the next section and then plot the two data sets using matplotlib module.
  7. A similar code to do the above is given here.

 

2. Fitting a Cavity Scan:

  1. The actual data for a cavity scan can be found in this elog entry or attached below in the zip folder.
  2. We read this data and separate the frequency data and the transmission data.
  3. Using the peakutils module's indexes function, we find the indices of the various peaks in the data set.
  4. These peaks are from the fundamental resonances, sideband resonances(both 11MHz and 55MHz) as well as a few HOMs.
  5. Each of these resonances follows the cavity equations and hence can be modelled as Lorentzian within small intervals around the peak frequencies. A detailed description of how this is possible is given here and is in the atached zip folder('Functionsused.pdf').
  6. We define a Lorentzian function which returns the fo\frac{a}{1+(\frac{\nu - \nu_0}{b})^{2}}llows:  where 'a' is the peak transmission value, 'b' is the 'linewidth' of the Lorentzian and \nu_0 is the peak frequency  about which the cavity equations behave like a lorentzian.
  7. We now, using the Lorentzian function, fit the various identified peaks using the curve_fit function of the scipy module. Remember to turn the 'absolute_sigma' parameter to 'True'.
  8. The parameters now obtained can be evaluated using the procedure given in the next section.
  9. The total transmission power is evaluated by feeding in the above obtained parameters back into the Lorentzian function and adding it for each peak.
  10. We can plot the actual data set and the data obtained using the fit of different peaks in a plot using matplotlib module. We can also plot the residuals for a better depiction of the fit quality.  
  11. The code to analyse the above mentioned cavity scan data is given here and attached below in the zip folder.

 

3. Calculating Physically Relevant Parameters:

  1. The data obtained from the fitting the peaks in the previous section now needs to be analysed in order to obtain some physically relevant information such as the FSR value, the TMS value, the modulation depths of the sidebands and perhaps even the linear caliberation of the frequency.
  2. First we need to identify the fundamental, TEM00 resonances among all the peaks. This we do by using the numpy.where function. We find the peaks with transmission values more than 0.9(or any suitable value).
  3. Using these indices we will now calculate the FSR and the Finesse of the peaks. A description of the correlation between the Fit Parameters and the FSR and Finesse is given here.
  4. We define a Linear fitting function for fitting the frequency values of the fundamental resonances against the ith fundamental resonance. The slope of this line gives us the value of FSR and the error in it.
  5. The Finesse can be calculated by fitting the linewidth with a constant function.
  6. The cavity length can be calculated using the FSR values as follows: L = \frac{c}{2\nu_{FSR}}.
  7. Now, the approximate positions of the sideband frequncies is given by 11*106%FSR and 55*106%FSR away from the fundamental, carrier resonances.
  8. The modulation depth, 'm', is given as \sqrt{\frac{P_{c}}{P_{s}}} = \frac{J_{0}(m)}{J_{1}(m)} where Pc is the carrier transmission power, Ps is the transmission power of the sideband and Jv is the Bessel Function of order 'v'.
  9. We define a function 'Bessel Ratio' using which we'll fit the transmission power ratio of the carrier to the sideband for the multiple sideband resonances.
  10. We also check for the Linearity in frequency data by plotting Fitting the frequencies corresponding to peaks in the actual data to ones obtained after fitting.
  11. After this we attempt to identify the other HOMs. For this we first determine a rough estimate for the value of TMS using the already known parameters of the mirrors,i.e., the RoC. We then look in small intervals (0.5 MHz) around frequencies where we would expect the HOMs to be, i.e., 1*TMS, 2*TMS, 3*TMS... away from the fundamental resonances. These positions are all modulo FSR.
  12. After identifying the HOMs, we take the difference from the fundamental resonance and then study these modulo the FSR.
  13. We perform a Linear Fit between these obtained values and (n+m).  As 'n','m' are degenerate, we can simply perform the fit against some variable 'k' and obtain the value of TMS as the slope of the linear fit.
  14. The code to do the above stated analysis is given here.

 

Most of the above info and some smaller details can be found in the markdown readme file in this git repo.

Attachment 1: Attachments.zip
  13116   Thu Jul 13 16:10:34 2017 KaustubhSummaryComputer Scripts / ProgramsCavity Scan Simulation Code

The code to produce a cavity scan simulation and then fitting the data and re-evaluating the initiallt set parameters can be found in this git repo.


The 'CavitScanSimulation' python script now produces a cavity scan with custom parameters which can be easily modified. It also introduces the first TEMs(n+m=0,1,2,3,4) to the laser with power going as (1/(2(n+m)+1))^2 {Selected carefully}. The only care that needs to be taken is that the frequency span should be somewhere near an integral multiple of the FSR so that there are equal number of resonances for all modes and sidebands. This code, as of now also calls the 'FitCavityScan' script which performs the fitting procedure on the data generated above{This data is actually written in a '.mat' file} and generates the Fit parameter data files. The Simulation code then calls the 'CalculatingPhysicalParameters' script which evaluates the data based on the Fit parameters and outputs some physically relevant results like the FSR, Finesse, Modulation Depths, TMS{Current Output is the Estimated RoCs of the two mirrors which isn't something we want directly, so it can be modified a bit to output TMS based on the HOMs}. The scripts do some 'Linearity' checks which might not really be of much significance but can be seen as a reference. Also, the ipython notebook will show all intermediate plots for the actual data and data with custom noise, fit data, FSR fitting, linearity checks, Bessel Ratio plot with mod_depths.

 

Note: The scripts should be run using either an IDE like 'spyder'{for .py files}{Comes with Anaconda} or using an ipython notebook{for .ipynb files}.
  13131   Fri Jul 21 19:44:58 2017 NaomiSummaryComputer Scripts / ProgramsUsing PyKat to run Finesse

I have been working on using PyKat to run Finesse. There appear to be several ways to run an equivalent simulation using Finesse:

1: .kat only

Run a .kat file directly from the terminal. For example, if in the directory containing the Finesse kat.ini file, run the command ‘./kat file.kat’. This method does not use PyKat.

To edit the simulation using this method, one must directly edit the .kat file. This is not ideal, as all parameters must be hard-coded, and there is no looping method for duplicate commands.


Both of the following methods use PyKat in some manner. To run Finesse using PyKat from a .py file, the command ‘from pykat import finesse’ should be included. In addition, two environment variables must be defined:

  • FINESSE_DIR': directory containing ‘kat’ executable
  • KATINI’: location and name of kat.ini file

Within a .py file running PyKat, the kat object contains all of the optical components and their states. To create a kat object, we use the command:

kat = finesse.kat()


2: .kat + .py

To load Finesse commands from a .kat file, we can use the command loadKatFile(). For example, using the kat object as defined above:

kat.loadKatFile(‘file.kat’)

The kat object now contains any components defined in the .kat file. The states of these components can be altered using PyKat. For example, if in the .kat file, we defined a mirror named ‘ITM’, with R = 0.9, T = 0.1, phi = 0, and with nodes 1 and 2 to its left and right, respectively, using the Finesse command

m ITM 0.9 0.1 0 n1 n2

we can now alter the state of the mirror using a PyKat command such as

kat.ITM.phi = 30

which changes the ‘phi’ property of the mirror to 30 degrees. Once all alterations to objects are made, we can run Finesse using the command

out = kat.run()

which stores the output of the Finesse simulation in the variable out.


3: .py only

We can also run a Finesse simulation without any .kat file. There are two ways to define Finesse objects within a .py file.

- Parse a string containing Finesse commands, as would be found in a .kat file, using the command parseCommands(). For example,

            kat.parseCommands(‘m ITM 0.9 0.1 0 n1 n2’)

defines the same mirror as above. This object can now be altered using pyKat in the same manner as above.

- Define an object using the classes defined in PyKat. For example, to define the same ITM mirror, we can use:

ITM = mirror(‘ITM’, ‘n1’, ‘n2’, 0.9, 0.1, 0)

kat.add(ITM)

The syntax for these classes can be found in the files included in the PyKat package named ‘commands.py’, ‘detectors.py’, and ‘components.py’.

We can also run Finesse commands (rather than just defining components) using their respective classes. These must also be added to the kat object. For example:

x = xaxis(‘lin’, [‘-4M’, ‘4M’], ‘f’, 1000, ‘laser’)

kat.add(x)

This runs the command ‘xaxis’, which sets the x-axis of the output data to run from freq = -4 MHz to 4 MHz, in 1000 steps. This is equivalent to the following Finesse command:

xaxis laser f lin -4M 4M 1000

In theory, we should be able to use PyKat to run any Finesse command. However, not all Finesse commands appear to be defined in PyKat; one example is the Finesse command ‘yaxis’, which I cannot locate in PyKat. In addition, I have had difficulty running some commands such as ‘cav’ and ‘pd’, despite following their class definitions in the PyKat files. However, these commands can still be easily run in PyKat using parseCommands().

  13299   Wed Sep 6 01:09:11 2017 johannesUpdateComputer Scripts / ProgramsNew set of loss measurements

I stumbled upon a faster way to stream data from the TDS3014 oscilloscopes to disk, which speeds the loss measurements up by a lot:  ftp://sprite.ssl.berkeley.edu/pub/sharris/MAVEN_LPW_Preamp/109_TDS3014B_control/tds3014b.py
This convenient(!) set of scripts contains a function that parses the scope's native binary format, for which the acquisition of 1 screenful of data takes <1s as opposed to ~20s, into readable data. I tested it for a bit and concluded that it does what it actually claims to do, but there's one weirdness: It get's the channel offset wrong. However this doesn't matter in our measurement because we're subtracting the dark level, which sees the same (wrong) offset. Other than that it seems okay.

So I started a new set of armloss measurements, and since the data acquisition is now much faster, I was able to squeeze a set of 20 individual measurements for each arm into ~30 minutes. This is the procedure I follow when I take these measurements for the XARM (symmetric under XARM <-> YARM):

  1. Dither-align the interferometer with both arms locked. Freeze outputs when done.
  2. Misalign ETMY + ITMY.
  3. ITMY needs to be misaligned further. Moving the slider by at least +0.2 is plentiful to not have the other beam interfere with the measurement.
  4. Start the script, which does the following:
    1. Resume dithering of the XARM
    2. Check XARM dither error signal rms with CDS. If they're calm enough, proceed.
    3. Freeze dithering
    4. Start a new set of averages on the scope, wait T_WAIT (5 seconds)
    5. Read data (= ASDC power and MC2 trans) from scope and save
    6. Misalign ETMX and wait 5s
    7. Read data from scope and save
    8. Repeat desired amount of times
  5. Close the PSL shutter and measure the PD dark levels

I will write a more comprehensive post describing the data acquisition and processing, let's just look at the results for now: The "uncertainties" reported by the individual measurements are on the order of 1-2 ppm (~1.9 for the XARM, ~1.3 for the YARM). This accounts for fluctuations of the data read from the scope and uncertainties in mode-matching and modulation depths in the EOM. I made histograms for the 20 datapoints taken for each arm: the standard deviation of the spread is a little over 2ppm. We end up with something like:

XARM: 49.3 +/- 2.1 ppm
YARM: 20.3 +/- 2.3 ppm

 

    

Attachment 1: XARM_20170905.pdf
XARM_20170905.pdf
Attachment 2: YARM_20170905.pdf
YARM_20170905.pdf
  13307   Mon Sep 11 12:56:40 2017 johannesUpdateComputer Scripts / Programslossmap attempts

I was trying to get a lossmap measurement over the weekend but had some trouble first with the IMC and then with the PMC.

For the IMC: It was a bit too misaligned to catch and maintain lock, but I had a hard time improving the alignment by hand. Fortunately, turning on the WFS quickly once it was locked restored the transmission to nominal levels and made it maintain the lock for longer, but only for several minutes, not enough for a lossmap scan (can take up to an hour). Using the WFS information I manually realigned the IMC, which made locking easier but wouldn't help with staying locked.

For the PMC: The PZT feedback signal had railed and the PMC had been unlocked for 8+ hours. The PMC medm screen controls were generally responsive (I could see the modes on the CCDs changing) but I just couldn't get it locked. c1psl was responding to ping but refusing telnet so I keyed the crate, followed by a burt restore and finally it worked.

After the PMC came back the IMC has already maintained lock for more than an hour, so I'm now running the first lossmap measurements.

  13316   Mon Sep 18 15:00:15 2017 rana, gautamFrogsComputer Scripts / Programsgateway PWD change

We implemented the post-SURF-season nodus password change today.

New password can be found at the usual location.

  13329   Sun Sep 24 20:47:15 2017 ranaUpdateComputer Scripts / ProgramsRF TF Uncertainties

I have made several changes to Craig's script for better pythonism. Its more robust with different libraries and syntaxes and makes a tarball by default (w/o a command line flag). These kinds of general util scripts will be going into a general use folder in the git.ligo.org/40m/ team area so that it can be used throughout the LSC.

I don't think we need/want a coherence calculation, so I have not included it. Usually, we use coherence to estimate the uncertainty, and here we are just plotting it directly from the dist of the sweeps so coherence seems superfluous.

Attachment 1: TFAG4395A_21-09-2017_115547_FourSquare.pdf
TFAG4395A_21-09-2017_115547_FourSquare.pdf
Attachment 2: TFAG4395A_21-09-2017_115547.tgz
  13330   Mon Sep 25 17:56:33 2017 johannesUpdateComputer Scripts / Programstransmitted power during lossmap

I had to do a reboot + burt restore of c1psl today. It was unresponsive and I couldn't get the PMC to lock. I also had to slightly realign the PMC, and the IMC was too misaligned for the autolocker to catch lock. Adjusting it manually, it was predominantly MC1 PIT that was off. The YARM locked on a 10 mode and had to be aligned manually as well.

I left a script running on Donnatella that tilts ETMX and thus moves the beam on ITMX. I'm monitoring the transmitted power to evaluate sane thresholds for the demodulation offsets in a lossmap measurement. The script will return the IFO to normal after it is done and will take <2 hours to complete (no real clue, but there's no way it takes longer than that for ~50 datapoints).

  13464   Thu Dec 7 11:14:37 2017 johannesHowToComputer Scripts / ProgramsLots of red on the FE status screen

Since we're getting ready to put the replacement slow DAQ for c1auxex in I wanted to bring the IFO back to operating condition after the PMC hasn't been locked for days. Something seems wrong with the CDS system though, many of the frontent models have red background and don't seem to be responsive. I followed the instructions laid out in https://wiki-40m.ligo.caltech.edu/Computer_Restart_Procedures.

In the attached screenshot, initially all c1ioo models were red, and on c1iscex only c1x01 was blue, the other ones red. I was able to ssh into both machines and tried to restart indivitual models, which didn't work and instead turned their background white. Still following the wiki page, I restarted both machines but they don't respond to pinging anymore and thus I cannot use ssh to reach them. Not sure what to do, I also rebooted fb over telnet.

So far I couldn't find any records of how to fix this situation.

Attachment 1: 22.png
22.png
  13465   Thu Dec 7 15:02:37 2017 KojiHowToComputer Scripts / ProgramsLots of red on the FE status screen

Once a realtime machine was rebooted, it did not come back. I suspect that the diskless hosts have a difficulty to boot up.

Attachment 1: DSC_0552.JPG
DSC_0552.JPG
  13466   Thu Dec 7 15:46:31 2017 johannesHowToComputer Scripts / ProgramsLots of red on the FE status screen

[Koji, Johannes]

The issue was partially fixed and the interferometer is in workable condition now.

What -probably- fixed it was restarting the dhcp server on chiara

sudo service isc-dhcp-server restart

Afterwards the frontends were restarted one by one. SSH access was possible and the essential models for IFO operation were started.

c1iscex reported initially that no DAQ card was found, and inside the IO chassis the LED indicator strip was red. Turning off the machine, checking the cables and rebooting fixed this.

Attachment 1: 04.png
04.png
  13524   Wed Jan 10 14:17:57 2018 johannesConfigurationComputer Scripts / Programsautoburt no longer making backups

I was looking into setting up autoburt for the new c1auxex2 and found that it stopped making automatic backups for all machines after the beginning of the new year. There is no 2018 folder (it was the only one missing) in /opt/rtcds/caltech/c1/burt/autoburt/snapshots and the /latest/ link in /opt/rtcds/caltech/c1/burt/autoburt/ leads to the last backup of 2017 on 12/31/17 at 23:19.

The autoburt log file shows that the back script last ran today 01/10/18 at 14:19, as it should have, but doesn't show any errors and ends with "You are at the 40m".

I'm not familiar with the autoburt scheduling using cronjobs. If I'm not mistaken the relevant cronjob file is /cvs/cds/rtcds/caltech/c1/scripts/autoburt/autoburt.cron which executes /cvs/cds/rtcds/caltech/c1/scripts/autoburt/autoburt.pl

I've never used perl but there's the following statement when establishing the directory for the new backup:

  $yearpath = $autoburtpath."/snapshots/".$thisyear;
  # print "scanning for path $yearpath\n";
  if (!-e $yearpath) {
    die "ERROR: Year directory $yearpath does not exist\n";
  }

I manually created the /opt/rtcds/caltech/c1/burt/autoburt/snapshots/2018/ directory. Maybe this fixes the hickup? Gotta wait about 30 minutes.

  13525   Wed Jan 10 15:25:43 2018 johannesConfigurationComputer Scripts / Programsautoburt making backups again
Quote:

I manually created the /opt/rtcds/caltech/c1/burt/autoburt/snapshots/2018/ directory. Maybe this fixes the hickup? Gotta wait about 30 minutes.

It worked. The first backup of the year is now from Wednesday, 01/10/18 at 15:19. Ten days of automatic backups are missing. Up until 2204 the year folders had been pre-emptively created so why was 2018 missing?

gautam: this is a bit suspect still - the snapshot file for c1auxex at least seemed to be too light on channels recorded. this was before any c1auxex switching. to be investigated.

  13603   Fri Feb 2 23:28:13 2018 KojiUpdateComputer Scripts / Programsnetgpib data missing / PROLOGIX yellow box (crocetta) not working

I could not understand why 'netgpibdata' scripts are missing in "scripts/general" folder on pianosa... Where did they go???

Also, I found that the PROLOGIX GPIB-LAN controller for crocetta (192.168.113.108) is no longer working. I need to reconfigure it with "telnet"...

  13604   Sat Feb 3 13:03:45 2018 gautamUpdateComputer Scripts / Programsnetgpib data missing / PROLOGIX yellow box (crocetta) not working

The netgpibdata scripts are now under git version control at /opt/rtcds/caltech/c1/scripts/general/labutils/netgpibdata. I think the idea was to make this directory a collection of useful utilities that we could then pull at various labs / at the sites.

Quote:

I could not understand why 'netgpibdata' scripts are missing in "scripts/general" folder on pianosa... Where did they go???

  13607   Mon Feb 5 18:04:35 2018 KojiUpdateComputer Scripts / Programsnetgpib data missing / PROLOGIX yellow box (crocetta) not working

crochetta was reconfigured to have 192.168.113.108. It was confirmed that it can be used with netgpibdata.py

Configuration: I have connected my mac with the unit using an Apple USB-Ethernet adapter. The adapter was configured to have a manual IP of 192.168.113.222/255.255.255.0. "netfinder.exe" was run to assign the IP addr to the unit. It seemed that NVRAM of the unit evaporated as it had the IP of 0.0.0.0. Once it was configrued, it could be run with netgpibdata as usual.

  13783   Tue Apr 24 10:10:43 2018 gautamUpdateComputer Scripts / ProgramsParticle swarm hyper parameter optimization

I'm copying and pasting Nikhil's email here as he was unable to login to the elog (but should now be able to in order to reply to any comments, and add more details about this test, motivation, methodology etc).

I did some post-processing after running the grid search. The following steps were carried out:

1) Selected those sets whose cost fun were less than a specific threshold (here 10000)

2) Next task was to see if the parameters of these good solutions had some pattern

3) I used a dimensionality reduction technique called t-SNE to project the 6 dimensional parameter space to 2 dim (for better visualization )

4) Made a scatter plot of these (see fig )

5) Used K-Means to find the clusters in this data

6) MarkerSize & Color reflect the cost fun. Bigger the marker size means better the solution.

7) Visual inspection implied cluster 5 had the best ranking points & more than any other cluster

8) These points had the following Parameter set: Workers {20,40}, SwarmSize {500}, MaxIter {500}, Self Adjustment {1}, Social Adjustment {1}, Tolerance {1e-3,1e-8} 

     See fig: for the box plot 

9) It looks like is a particular set of values rather than individual values that gives the best results.

 

Attachment 1: ClusterFminScaled.png
ClusterFminScaled.png
Attachment 2: ClusterID_5.png
ClusterID_5.png
  13801   Mon Apr 30 23:13:12 2018 KevinUpdateComputer Scripts / ProgramsDataViewer leapseconds

I was trying to plot trends (min, 10 min, and hour) in DataViewer and got the following error message

Connecting.... done
 mjd = 58235
leapsecs_read()
  Opening leapsecs.dat
  Open of leapsecs.dat failed
leapsecs_read() returning 0
frameMemRead - gpstimest = 1208844718

 

thoough the plots showed up fine after. Do we need to fix something with the leapsecs.dat file?

  13929   Thu Jun 7 20:21:15 2018 KojiUpdateComputer Scripts / Programs/cvs/cds Backup in danger

Local backup on chiara seems not working since Nov 19, 2017.
/opt/rtcds/caltech/c1/scripts/backup/localbackup.log

2017-11-18 07:00:01,504 INFO       Updating backup image of /cvs/cds
2017-11-18 07:03:00,113 INFO       Backup rsync job ran successfully, transferred 1954 files.
2017-11-19 07:00:02,564 INFO       Updating backup image of /cvs/cds
2017-11-19 07:00:02,592 ERROR      External drive not mounted!!!

  13963   Thu Jun 14 15:21:58 2018 gautamUpdateComputer Scripts / Programs/cvs/cds Backup in danger

I think this is because /cvs/cds is getting too big. lsblk reveals:

controls@chiara|~> lsblk
NAME   MAJ:MIN RM   SIZE RO TYPE MOUNTPOINT
sda      8:0    0 465.8G  0 disk 
├─sda1   8:1    0 446.9G  0 part /
├─sda2   8:2    0     1K  0 part 
└─sda5   8:5    0  18.9G  0 part [SWAP]
sdb      8:16   0   2.7T  0 disk 
└─sdb1   8:17   0     2T  0 part /home/cds
sr0     11:0    1  1024M  0 rom  
sdc      8:32   0   1.8T  0 disk 
└─sdc1   8:33   0   1.8T  0 part /media/40mBackup
sdd      8:48   0   1.8T  0 disk 
└─sdd1   8:49   0   1.8T  0 part 

I believe one of sdc or sdd is connected via SATA while the other is an external USB drive. Maybe we have to get bigger backup disks, but this may be a huge pain to setup as it will involve taking chiara down. Actually, now that I check the backup log, seems like backup is executing successfully - not sure if this is due to my unelogged mounting of sdc (using sudo mount /dev/sdc1 /media/40mBackup) last week, or if this is some LDAS backup. But in any case, seems undesirable that sdb1 is larger than sdc1 or sdd1.

2018-06-06 07:00:01,086 INFO       Updating backup image of /cvs/cds
2018-06-06 07:00:01,086 ERROR      External drive not mounted!!!
2018-06-07 07:00:01,147 INFO       Updating backup image of /cvs/cds
2018-06-07 07:00:01,147 ERROR      External drive not mounted!!!
2018-06-08 07:00:01,244 INFO       Updating backup image of /cvs/cds
2018-06-08 08:23:32,939 INFO       Backup rsync job ran successfully, transferred 316870 files.
2018-06-09 07:00:01,465 INFO       Updating backup image of /cvs/cds
2018-06-09 07:12:11,865 INFO       Backup rsync job ran successfully, transferred 1926 files.
2018-06-10 07:00:01,842 INFO       Updating backup image of /cvs/cds
2018-06-10 07:12:28,931 INFO       Backup rsync job ran successfully, transferred 1656 files.
2018-06-11 07:00:01,294 INFO       Updating backup image of /cvs/cds
2018-06-11 07:06:14,748 INFO       Backup rsync job ran successfully, transferred 1664 files.
2018-06-12 07:00:02,081 INFO       Updating backup image of /cvs/cds
2018-06-12 07:07:36,775 INFO       Backup rsync job ran successfully, transferred 1870 files.
2018-06-13 07:00:02,194 INFO       Updating backup image of /cvs/cds
2018-06-13 07:08:37,356 INFO       Backup rsync job ran successfully, transferred 1818 files.
2018-06-14 07:00:01,753 INFO       Updating backup image of /cvs/cds
2018-06-14 07:01:43,270 INFO       Backup rsync job ran successfully, transferred 1744 files.
Quote:

Local backup on chiara seems not working since Nov 19, 2017.
/opt/rtcds/caltech/c1/scripts/backup/localbackup.log

2017-11-18 07:00:01,504 INFO       Updating backup image of /cvs/cds
2017-11-18 07:03:00,113 INFO       Backup rsync job ran successfully, transferred 1954 files.
2017-11-19 07:00:02,564 INFO       Updating backup image of /cvs/cds
2017-11-19 07:00:02,592 ERROR      External drive not mounted!!!

 

  13978   Mon Jun 18 10:34:45 2018 johannesUpdateComputer Scripts / Programsrunning comsol job on optimus

I'm running a comsol job on optimus in a tmux session named cryocavs. Should be done in less than 24 hours, judging by past durations.

  14049   Tue Jul 10 16:59:12 2018 Izabella PastranaHowToComputer Scripts / ProgramsTaking Remote TF Measurements with the Agilent 4395A

I copied the netgpibdata folder onto rossa (under the directory ~/Agilent/), which contains all the necessary scripts and templates you'll need to remotely set up, run, and download the results of measurements taken on the AG4395A network analyzer. The computer will communicate with the network analyzer through the GPIB device (plugged into the back of the Agilent, and whose communication protocol is found in the AG4395A.py file in the directory ~/Agilent/netgpibdata/).

The parameter template file you'll be concerned with is TFAG4395Atemplate.yml (again, under ~/Agilent/netgpibdata/), which you can edit to fit your measurement needs. (The parameters you can change are all helpfully commented, so it's pretty straightforward to use! Note: this template file should remain in the same directory as AGmeasure, which is the executable python script you'll be using). Then, to actually set up, run, and download your measurement, you'll want to navigate to the ~/Agilent/netgpibdata/ directory, where you can run on the command line the following: python AGmeasure TFAG4395Atemplate.yml

The above command will run the measurement defined in your template file and then save a .txt file of your measured data points to the directory specified in your parameters. If you set up the template file such that the data is also plotted and saved after the measurement, a .pdf of the plot will be saved along with your .txt file.

Now if you want to just download the data currently on the instrument display, you can run: python AGmeasure -i 192.168.113.105 -a 10 --getdata

Those are the big points, but you can also run python AGmeasure --help to learn about all the other functions of AGmeasure (alternatively, you can read through the actual python script).

Happy remote measuring! :)

 

 

 

 

  14157   Mon Aug 13 11:44:32 2018 gautamUpdateComputer Scripts / ProgramsPatch updates on nodus

Larry W said that some security issues were flagged on nodus. So I ran

sudo yum upgrade --exclude=elog-3.1.3-2.el7.x86_64

on nodus. The exclude flag is because there were some conflicts related to that particular package. Hopefully this has fixed the problem. It's been a while since the last update, which was in January of this year.

controls@nodus|~> sudo yum history
Loaded plugins: langpacks
ID     | Command line             | Date and time    | Action(s)      | Altered
-------------------------------------------------------------------------------
    29 | upgrade --exclude=elog-3 | 2018-08-13 11:36 | E, I, U        |  136 EE
    28 | install yum-utils        | 2018-08-13 11:31 | Update         |    1   
    27 | install nmap             | 2018-06-29 01:57 | Install        |    2   
    26 | install grace            | 2018-05-31 16:52 | Install        |   11   
    25 | install https://dl.fedor | 2018-05-31 16:51 | Install        |    1   
    24 | install perl-Digest-SHA1 | 2018-05-31 15:34 | Install        |    1   
    23 | install python-devel     | 2018-01-13 15:33 | Install        |    1   
    22 | install gcc              | 2018-01-13 15:32 | Install        |    6   
    21 | install git              | 2018-01-12 18:11 | Install        |    4   
    20 | update                   | 2018-01-12 18:01 | I, U           |   39   
    19 | install motif            | 2018-01-05 17:35 | Install        |    3   
    18 | install sendmail sendmai | 2017-12-03 05:11 | Install        |    6   
    17 | install vim              | 2017-11-21 18:12 | Install        |    3   
    16 | reinstall mod_dav_svn    | 2017-11-21 17:40 | Reinstall      |    1   
    15 | install mod_dav_svn      | 2017-11-21 17:39 | Install        |    1   
    14 | install subversion       | 2017-11-21 15:36 | Install        |    2   
    13 | -y install php           | 2017-11-20 22:15 | Install        |    4   
    12 | install links            | 2017-11-20 19:10 | Install        |    2   
    11 | install openssl098e.i686 | 2017-11-18 18:28 | Install        |    1   
    10 | install openssl-libs.i68 | 2017-11-18 18:26 | Install        |   11   
history list
  14243   Thu Oct 11 13:40:51 2018 yukiUpdateComputer Scripts / Programsloss measurements
Quote:

This is the procedure I follow when I take these measurements for the XARM (symmetric under XARM <-> YARM):

  1. Dither-align the interferometer with both arms locked. Freeze outputs when done.
  2. Misalign ETMY + ITMY.
  3. ITMY needs to be misaligned further. Moving the slider by at least +0.2 is plentiful to not have the other beam interfere with the measurement.
  4. Start the script, which does the following:
    1. Resume dithering of the XARM
    2. Check XARM dither error signal rms with CDS. If they're calm enough, proceed.
    3. Freeze dithering
    4. Start a new set of averages on the scope, wait T_WAIT (5 seconds)
    5. Read data (= ASDC power and MC2 trans) from scope and save
    6. Misalign ETMX and wait 5s
    7. Read data from scope and save
    8. Repeat desired amount of times
  5. Close the PSL shutter and measure the PD dark levels

Information for the armloss measurement:

  • Script which gets the data:  /users/johannes/40m/armloss/scripts/armloss_scope/armloss_dcrefl_asdcpd_scope.py
  • Script which calculates the loss: /users/johannes/40m/armloss/scripts/misc/armloss_AS_calc.py
  • Before doing the procedure Johannes wrote you have to prepare as follows:
    • put a PD in anti-symmetric beam path to get ASDC signal.
    • put a PD in MC2 box to get tranmitted light of IMC. It is used to normalize the beam power.
    • connect those 2 PDs to oscilloscope and insert an internet cable to it.
  • Usage: python2 armloss_dcrefl_asdcpd_scope.py [IP address of Scope] [ScopeCH for AS] [ScopeCH for MC] [Num of iteration] [ArmMode]

Note: The scripts uses httplib2 module. You have to install it if you don't have.

The locked arms are needed to calculate armloss but the alignment of PMC is deadly bad now. So at first I will make it aligned. (Gautam aligned it and PMC is locked now.) 

gautam: The PMC alignment was fine, the problem was that the c1psl slow machine had become unresponsive, which prevented the PMC length servo from functioning correctly. I rebooted the machine and undid the alignment changes Yuki had made on the PSL table.

  14245   Fri Oct 12 12:29:34 2018 yukiUpdateComputer Scripts / Programsloss measurements

With Gautam's help, Y-arm was locked. Then I ran the script "armloss_dcrefl_asdcpd_scope.py" which gets the signals from oscilloscope. It ran and got data, but I found some problems.

  1. It seemed that a process which makes arm cavity mislaigned in the script didn't work.
  2. The script "armloss_dcrefl_asdcpd_scope.py" gets the signal and the another script "armloss_AS_calc.py" calculates the arm loss. But output file the former makes doesn't match with the type the latter requires. A script converts format is needed.

Anyway, I got the data needed so I will calculate the loss after converting the format.

  14248   Fri Oct 12 20:20:29 2018 yukiUpdateComputer Scripts / Programsloss measurements

I ran the script for measuring arm-loss and calculated rough Y-arm round trip loss temporally. The result was 89.6ppm. (The error should be considered later.)

The measurement was done as follows:

  1. install hardware
    1. Put a PD (PDA520) in anti-symmetric beam path to get ASDC signal.
    2. Use a PD (PDA255) in MC2 box to get tranmitted light of IMC. It is used to normalize the beam power.
    3. Connect those 2 PDs to oscilloscope (IP: 192.168.113.25) and insert an internet cable to it.
  2. measure DARK noise
    1. Block beam going into PDs with dampers and turn off the room light.
    2. Run the script "armloss_dcrefl_acdcpd_scope.py" using "DARK" mode.
  3. measure the ASDC power when Y-arm locked and misaligned
    1. Remove dampers and turn off the room light.
    2. Dither-align the interferometer with both arms locked. Freeze outputs when done. (Click C1ASS.adl>!MoreScripts>ON and click C1ASS.adl>!MoreScripts>FreezeOutputs.)
    3. Misalign ETMX + ITMX. (Just click "Misalign" button.)
    4. Further misalign ITMX with the slider. (see previous study: ITMX needs to be misaligned further. Moving the slider by at least +0.2 is plentiful to not have the other beam interfere with the measurement.)
    5. Start the script "armloss_dcrefl_acdcpd_scope.py" using "ETMY" mode, which does the following:
      1. Resume dithering of the YARM.
      2. Check YARM dither error signal rms with CDS. If they're calm enough, proceed. (In the previous study the rms threshold was 0.7. Now "ETM_YAW_L_DEMOD_I" signal was 15 (noisy), then the threshold was set 17.)
      3. Freeze dithering.
      4. Start a new set of averages on the scope, wait T_WAIT (5 seconds).
      5. Read data (= ASDC power and MC2 trans) from scope and save.
      6. Misalign ETMY and wait 5s. (I added a code which switchs LSC mode ON and OFF.)
      7. Read data from scope and save.
      8. Repeat desired amount of times.
  4. calculate the arm loss
    1. Start the script "armloss_AS_calc.py", whose content is follows:
      • requires given parameters: Mode-Matching effeciency, modulation depth, transmissivity. I used the same value as Johannes did last year, which are (huga)
      • reads datafile of beam power at ASDC and MC2 trans, which file is created by "armloss_dcrefl_acdcpd_scope.py".
      • calculates arm loss from the equation (see 12528 and 12854).

Result:

YARM
('AS_DARK =', '0.0019517200000000003') #dark noise at ASDC 
('MC_DARK =', '0.02792') #dark noise at MC2 trans
('AS_LOCKED =', '2.04293') #beam power at ASDC when the cavity was locked 
('MC_LOCKED =', '2.6951620000000003')
('AS_MISALIGNED =', '2.0445439999999997') #beam power at ASDC when the cavity was misaligned
('MC_MISALIGNED =', '2.665312')

\hat{P} = \frac{P_{AS}-P_{AS}^{DARK}}{P_{MC}-P_{MC}^{DARK}} #normalized beam power 

\hat{P}^{LOCKED}=0.765,\ \hat{P}^{MISALIGNED}=0.775,\ \mathcal{L}=89.6\ \mathrm{ppm}

Comments:

  • "ETM_YAW_L_DEMOD_I_OUTPUT" was little noisy even when the arm was locked.
  • The reflected beam power when locked was higher than when misaligned. It seemed strange for me at first. Johannes suggested that it was caused by over-coupling cavity. It is possible when r_{ETMY}>>r1_{ITMY}.
  • My first (wrong) measurement said the arm loss was negative(!). That was caused by lack of enough misalignment of another arm mirrors. If you don't misalign ITMX enough then the beam or scattered light from X-arm would bring bad. The calculated negative loss would be appeared only when \frac{\hat{P}^{LOCKED}}{\hat{P}^{MISALIGNED}} > 1 + T_{ITM}
  • Error should be considered.
  • Parameters given this time should be measured again. 
  14251   Sat Oct 13 20:11:10 2018 yukiUpdateComputer Scripts / Programsloss measurements
Quote:

the script "armloss_AS_calc.py",

  • "ETM_YAW_L_DEMOD_I_OUTPUT" was little noisy even when the arm was locked.
  • The reflected beam power when locked was higher than when misaligned. It seemed strange for me at first. Johannes suggested that it was caused by over-coupling cavity. It is possible when r_{ETMY}>>r1_{ITMY}.

Some changes were made in the script for getting the signals of beam power:

  • The script sees "C1:ASS-X(Y)ARM_ETM_PIT/YAW_L_DEMOD_I_OUTPUT" and stops running until the signals become small, however some offset could be on the signal. So I changed it into waiting until (DEMOD - OFFSET) becomes small. (Yesterday I wrote ETM_YAW_L_DEMOD_I_OUTPUT was about 15 and was little noisy. I was wrong. That was just a offset value.)
  • I added a code which stops running the script when the power of transmitted IR beam is low. You can set this threshold. The nominal value of "C1:LSC-TRX(Y)_OUT16" is 1.2 (1.0), so the threshold is set 0.8 now.  

In the yesterday measurement the beam power of ASDC is higher when locked than when misaligned and I wrote it maybe caused by over-coupled cavity. Then I did a calculation as following to explain this:

  • assume power transmissivity of ITM and ETM are 1.4e-2 and 1.4e-5.
  • assume loss-less mirror, you can calculate amplitude reflectivity of ITM and ETM.
  • consider a cavity which consists two mirrors and is loss-less, then \frac{E_{r}}{E_{in}} = \frac{-r_1+r_2e^{i\phi}}{1-r_1r_2e^{i\phi}} holds. r1 and r2 are amplitude reflectivity of ITM and ETM, and E is electric filed.
  • Then you can calculate the power of reflected beam when resonated and when anti-resonated. The fraction of these value is \frac{P_{RESONANT}}{P_{ANTI-RESO}} = 0.996, which is smaller than 1.
  • I found this calculation was wrong! Above calculatation only holds when cavity is aligned, not when misaligned. 99.04% of incident beam power reflects when locked, and (100-1.4)% reflects when misaligned. The proportion is P(locked)/P(misaligned)=1.004, higher than 1.

 

  14254   Mon Oct 15 10:32:13 2018 yukiUpdateComputer Scripts / Programsloss measurements

I used these values for measuring armloss:

  • Transmissivitity of ITM = 1.384e-2 * (1 +/- 1e-2) 
  • Transmissivitity of ETM = 13.7e-6 * (1 +/- 5e-2)
  • Mode-Matching efficiency of XARM = 0.912 * (1 +/- 2e-2)
  • Mode-Matching efficiency of YARM = 0.867 * (1 +/- 2e-2)
  • modulation depth m1 (11MHz) = 0.179 * (1 +/- 2e-2)
  • modulation depth m2 = 0.226 * (1 +/- 2e-2),

then the uncertainties reported by the individual measurements are on the order of 6 ppm (~6.2 for the XARM, ~6.3 for the YARM). This accounts for fluctuations of the data read from the scope and uncertainties in mode-matching and modulation depths in the EOM. I made histograms for the 20 datapoints taken for each arm: the standard deviation of the spread is over 6ppm. We end up with something like:

XARM: 123 +/- 50 ppm
YARM: 152+/- 50 ppm

This result has about 40% of uncertaintities in XARM and 33% in YARM (so big... no).

In the previous measurement, the fluctuation of each power was 0.1% and the fluctuation of P(Locked)/P(misaligned) was also 0.1%. Then the uncertainty was small. On the other hand in my measurement, the fluctuation of power is about 2% and the fluctuation of P(Locked)/P(misaligned) is 2%. That's why the uncertainty became big.

We want to measure tiny value of loss (~100ppm). So the fluctuation of P(Locked)/P(misaligned) must be smaller than 1.6%.

(Edit on 10/23)
I think the error is dominated by systematic error in scope. The data of beam power had only 3 degits. If P(Locked) and P(misaligned) have 2% error, then
\frac{P_L}{P_M}\frac{1}{1+T_{\mathrm{ITM}}} = 0.99(3).
You have to check the configuration of scope.

Attachment 1: XARM_20181015_1500.pdf
XARM_20181015_1500.pdf
Attachment 2: YARM_20181015_1500.pdf
YARM_20181015_1500.pdf
  14255   Mon Oct 15 12:52:54 2018 yukiUpdateComputer Scripts / Programsadditional comments
Quote:

but there's one weirdness: It get's the channel offset wrong. However this doesn't matter in our measurement because we're subtracting the dark level, which sees the same (wrong) offset.

When you do this measurement with oscilloscope, take care two things:

  1. set y-range of scope as to every signal fits in display: otherwise the data sent from scope would be saturated.
  2. set y-position of scope to the center and don't change it; otherwise some offset would be on the data.
  14258   Tue Oct 16 00:44:29 2018 yukiUpdateComputer Scripts / Programsloss measurements

The scripts for measuring armloss are in the directory "/opt/rtcds/caltech/c1/scripts/lossmap_scripts/armloss_scope".

  • armloss_derefl_asdcpd_scope.py: gets data and makes ascii file.
  • armloss_AS_calc.py: calculates armloss from selected a set of files.
  • armloss_calc_histogram.py: calculates armloss from selected files and makes histogram.
  14268   Fri Nov 2 16:42:31 2018 aaronUpdateComputer Scripts / Programsarm loss measuremenents

I'm continuing the arm loss measurements Yuki was making. I'm first familiarizing myself with the procedures for the measurement Johannes describes.

I'm not very familiar with the medm screens, so I'm just kind of poking around and checking with Gautam. I do the following:

  1. Turned Xarm ASS dither on, then off.
  2. Turned X and Y ALS on, then off shortly after
    1. Realizing I needed some guidance, I found this page on lock acquisition on the wiki
    2. Gautam showed me how to align/lock the IFO so I could take some notes, and we locked the Y arm, misaligned X.
  3. I put the PD back in the AS beam path to get the ASDC signal, and approximately centered the beam. This PD is on channel 1 of the scope, which is at 192.168.113.24.
  4. I centered the beam onto the MC2 PD that Yuki had installed. This PD is on channel 2 of the scope.
    1. Both scope channels are set to 1V scale (I also had tried 500mV, and it didn't seem to make a difference) and 10s time axis spacing (maximum integration time, since we're looking for a DC effect. Is this what we want?)
    2. The impedance for both channels is 1MOhm.
  5. I ran the script to start the loss measurement on the Y arm.
    1. python2 armloss_dcrefl_asdcpd_scope.py 192.168.113.24 1 2 5 YARM
    2. I'm reading ~15 (au?) for the MC channel and ~5% of that out the AS, which seems to make sense to me and looked to be about what Yuki the ratios when I checked the log files. However, I'm a bit confused by the normalization, because the maximum output of the MC PD is 10V, and indeed the scope's display is reading under 10V.

I've left the script running.

  14269   Fri Nov 2 19:25:16 2018 gautamUpdateComputer Scripts / Programsloss measurements

Some facts which should be considered when doing this measurement and the associated uncertainty:

  1. When Johannes did the measurement, there was no light from the AS port diverted to the OMC. This represents ~70% loss in the absolute amount of power available for this measurement. I estimate ~1W*Tprm * Ritm * Tbs * Rbs * Tsrm * OMCsplit ~ 300uW which should still be plenty, but the real parameter of interest is the difference in reflected power between locked/no cavity situations, and how that compares to the RMS of the scope readout. For comparison, the POX DC light level is expected to be ~20uW, assuming a 600ppm AR coating on the ITMs.
  2. Even though the reflection from the arm not being measured may look like it's completely misaligned looking at the AS camera, the PDA520 which is used at the AS port has a large active area and so one must check on the oscilloscope that the other arm is truly misaligned and not hitting the photodiode to avoid interference effects artifically bloating the uncertainty.
  3. The PDA255 monitoring the MC transmission has a tiny active area. I'm not sure the beam has been centered on it anytime recently. If the beam is not well centered on that PD, and you normalize the measurements by "MC Transmission", you're likely to end up with larger error.