40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log, Page 87 of 339  Not logged in ELOG logo
ID Date Author Typeup Category Subject
  1318   Wed Feb 18 03:25:25 2009 YoichiUpdateComputersmedm directory back
I restored the medm directory from the backup on the tape.
The directory had an svn property svn:ignore set and the value of the property included *.snap and *.req.
This resulted in the exclusion of those files from the repository.
I fixed this problem by changing the property of all the directories under /cvs/cds/caltech/medm.
After fixing several other svn problems, the current medm directory contents were checked in to the repository.
  1321   Wed Feb 18 21:03:22 2009 ranaUpdateCamerasETMY Camera work not elogged!

The control room video is showing us a false ETMY image. Who worked on the ETMY camera or video today??!!

  1322   Wed Feb 18 21:10:21 2009 ranaUpdateIOOMC Drumhead mode lost again
In early December, Caryn and I noticed that the MC Drumhead mode was visible at the Qmon point of
the MC demod board using a spectrum analyzer and no external excitation of the MC mirrors. We then
started tracking the MC Drumhead modes.

Today I found that it is gone again. It also wasn't there when I looked for it in 2007. Mad

I looked at the MC error point spectrum and it seemed reasonable. Changing the gains in the MZ, ISS, PMC, & FSS
had no good effect on the noise spectrum.

The voltage noise above 10 kHz in the MC error point is increasing like ~f. I think that this means that
the leftover is the noise from the FSS. Below 10 kHz it is the noise of the VCO (10 mHz/rHz).

One possibility is that the high frequency noise changes with the mood of the NPRO. There should be no
frequency noise induced by the decay of the PA diode power. We can do an NPRO SLOW scan to see if there
is some kind of mode hop noise happening.
  1323   Thu Feb 19 04:16:17 2009 YoichiUpdateLSCLocking status
Rob, Yoichi

We checked the CM-MC cross over just before turning off the moving zero.
There was a slight bump in the gain of the MC_L loop at (I believe) the optical spring freq. (~400Hz) just below 0 dB. The phase margin there was very thin.
Removing the moving zero will increase the bump more and make the loop unstable.
Rob suggested to increase the AO gain a bit more.

To see if the AO path is really working, I connected the OUT2 of the MC board to a spare DAQ channel (C1:PEM-OSA_APTEMP).
I confirmed that the PO_DC signal is actually coming to the AO path input of the MC board.
I also hooked up the SR785 to the A excitation channel of the common mode board, so that we can measure the loop gain of the AO path.
After these preparation, the lock acquisition process became somewhat unstable. The ifo loses lock randomly at various places in the lock acquisition steps.
So, as of 4:00 am, I have not gotten a chance to try Rob's suggestion nor the TF measurement with SR785 yet.
I will continue the work tomorrow (i.e. tonight ??).

  1324   Thu Feb 19 11:51:56 2009 steveUpdateMOPAHTEMP variation is too much
The C1:PSL-MOPA_HTEMP variation is more than 0.5 C daily
Normally this temp stays well within 0.1 C
This 80 days plots shows that we have just entered this unstable region some days ago.
The control room temp set unchanged at 70 F, actual temp at ac-mon 69-70 with occasional peaks at 74 F
 
Water temp at chiller repeatedly around 20.6 C at 8 am
This should be rock solid at 20.00C +- 0.02C
 

 

Attachment 1: 80dhtemp.jpg
80dhtemp.jpg
  1325   Thu Feb 19 16:29:43 2009 YoichiUpdateComputersMartian wireless router bad
The Martian wireless router is dead.
I rebooted it several times, but it hangs up in a minute.
I will ask steve to buy a new one.
  1326   Thu Feb 19 22:40:33 2009 KiwamuUpdateElectronicsPSL angle QPD

I checked a broken QPD, which was placed for PSL angle monitor, and finally I cocluded one segment of the quadrant diode was broken.

The broken segment has a offset voltage of -0.7V after 1st I-V amplifier. It means the diode segment has a current offset without any injection of light.

Tomorrow I will check a new QPD for replacement.

Kiwamu IZUMI

 

  1327   Thu Feb 19 23:50:31 2009 peteUpdateLockingaligned pd's on AP table

Yoichi, Peter

While continuing our efforts to lock, we noticed the procedure failed at a point it had gotten past last night:  turning on the bounce/roll filters in MICH, PRC, and SRC.  We checked the MICH transfer function and noticed that the unity gain point was ~10 Hz, well below the bounce modes.   We tried increasing the gain but found saturation, and Rob suggested that there could be misalignment on the AP table, which Steve worked on today.  We went out and found two of the PDs (ASDD133 and AS166) to be badly misaligned probably due to a bumped optic upstream.  We re-aligned.

 

 

  1328   Fri Feb 20 01:54:18 2009 KakeruUpdateComputer Scripts / Programstdsdata might have a bug

I found a strange jump of value in my data taken with tdsdata.
I couldn't find same jump in a playback of DataViewer, so I think this is a problem of tdsdata.
Be careful when you use tdsdata!

The attached file is an example of jumped data.
I try to get data with allegra and op440m, and both has same kind of jump.
(A downsampling or interpolation may be wrong.)

Rana said there is a fixed version of tdsdata in some PC, but 64bit linux may not have.
I try it tomorrow.

Attachment 1: jumped_data.png
jumped_data.png
  1329   Fri Feb 20 03:52:23 2009 YoichiUpdateLockingLocking Tonight
Yoichi, Peter

Tonight, we had a problem with the DD hand off.
It failed when the RG filters of MICH for the bounce-roll modes are engaged.
The reason for the failure was that the MICH UGF was too low (~10Hz).
As in the Peter's elog entry, we found that the AS PDs are mis-centered.
Even after we fixed the centering, the MICH UGF was still too low. So we increased the MICH feedback gain by a factor of 10.
The reason for the gain decrease is unknown. It seems almost like the BS coils get weaker.
I checked the UGF of the BS OL loops. These are around 4Hz, so fine. We should check the HWP on the AP table tomorrow.

After the DD hand-off goes ok, the switching of DARM signal from DC to RF failed.
I found that the gain and the polarity of the RF signal were wrong.
AS166 is one of the PDs we found mis-centered (and re-centered). But how can you flip the sign of the signal ?

After this, the cm_step script goes until the activation of the moving zero, but fails when the arm power is increased to 0.7.
Also the ontoMCL script succeeds only 50% of the time.
  1330   Fri Feb 20 19:31:16 2009 YoichiUpdateLSCMICH low gain problem
Last night, we found that MICH UGF was too low. Even after re-aligning the PDs, it was still too low.
Today, I compared the UGFs of MICH and PRC when in the DRMI configuration locked with the single demod. signals.
In this configuration, MICH signal comes from REFL33Q and the PRC signal comes from REFL33I (the same PD).
The PRC UGF was about 100Hz whereas MICH was only ~10Hz.
Since they uses the same PD, the low gain is not caused by the PD.
I checked conlog history and confirmed there is no change in the MICH->BS path in the last few days.
I also checked the svn history of chans directory for changes in filters. Nothing problematic found.

Then I noticed that the susvme computers were overloaded.
This time, I rebooted all the FE computers just in case.

Then the MICH gain was somewhat recovered (by a factor of 3 or so). Don't know why.

I adjusted the DD_handoff script to set the MICH gain to 0.7 before the bounce-roll filter is engaged.
  1334   Tue Feb 24 02:23:40 2009 YoichiUpdateLockingLocking - MC board bad
Rob, Yoichi, Alberto, Kiwamu, Kakeru

We found that the OMC alignment feedback was on for the POS X loop even though the OMC was not locked.
This caused the PZT mirror to be tilted in yaw a lot. This was probably the reason for the mysterious shift in the AS beam last week, because the AS RF beam is picked up after this PZT mirror.
Rob aligned the OMC and we re-centered the AS PDs and the CCD.
This changed the DARM RF gain, so we changed it from 3 to 1. This gain used to be -1. It is still not understood why the polarity was changed.
The MC length was changed ? We should check the sideband transmission.

After this, we reached to the arm power 4. But the IFO loses lock immediately after the moving zero is turned off.
At this stage, the CARM loop bandwidth is supposed to be high enough that the moving zero is no longer necessary.
However, when we measured the MCL loop gain with several different AO path gains, the loop shape did not change at all.
This led us to suspect the AO path may not be connected. The cabling from the common mode board to the MC board seemed ok.
We tested the signal flow in the MC board using a signal generator and an oscilloscope.
Then we found that a signal injected to the IN2 (AO path) does not reach to the TP1A (right after the boost stages), though the signal is visible in the OUT2 (monitor BNC right after the initial amplifier (B-amp) for the AO path). The signal from IN1 (MC REFL) can be observed at TP1A. This means something is broken between the B-amp and the sum-amp in the AO path.
We will check the MC board tomorrow.
  1335   Tue Feb 24 18:42:15 2009 peteUpdateLockingmc board repair
Peter, Yoichi
Last night:


Quote:
However, when we measured the MCL loop gain with several different AO path gains, the loop shape did not change at all. This led us to suspect the AO path may not be connected. The cabling from the common mode board to the MC board seemed ok. We tested the signal flow in the MC board using a signal generator and an oscilloscope. Then we found that a signal injected to the IN2 (AO path) does not reach to the TP1A (right after the boost stages), though the signal is visible in the OUT2 (monitor BNC right after the initial amplifier (B-amp) for the AO path). The signal from IN1 (MC REFL) can be observed at TP1A. This means something is broken between the B-amp and the sum-amp in the AO path. We will check the MC board tomorrow.


Today we examined the MC board. With the extension board in place everything seemed fine. Without the extension board we could replicate the problem. Jiggling the IN2 jack caused the injected signal observed at TP1A to come and go. These jacks are unfortunately mounted directly on the board. We traced the problem to a resistor in this path (R30) which looked fishy. We soldered on a new 2K resistor with OWC and it fixed the problem.
  1336   Wed Feb 25 03:10:24 2009 YoichiUpdateLockingLocking status
Rob, Yoichi, Kakeru, Kiwamu

Tonight, CARM -> MCL hand off was not stable. The MCF signal monotonically went up to +2V after CARM and MCL gain was turned down to zero.
This was repeatable and it only goes up (not down).
After a while, we found that putting sleep (~5sec) between the zeroing of CARM gain and MCL gain prevents this problem.

Handing off of CARM error signal from TR to PODC was also not robust.
It seems that the suitable gain changes every time.

tdsavg started to exit with errors. We rebooted fb40m.
When tdsavg returns an error, the cm_step script tries to write NaN into SPOB DC offset.
To prevent this, I put the tdsavg in a while loop which runs until tdsavg returns something other than NaN.

I was able to hand off to PODC several times, but could not proceed further because the IFO lost lock soon.
  1337   Wed Feb 25 11:48:02 2009 JenneUpdatePEMWiener filtering update - work on filtering some S5 DARM_CTRL data

Quick update on my wiener filtering status:

Joe has been helping me get on the GRID, so I now have  a grid certificate, and accounts on most/all of the clusters.

Joe also helped me get menkar to get S5 data so that I can do wiener filtering to the back-data. 

 

I've been running the wiener filtering algorithm, and right now, it doesn't do anything to improve the DARM_CTRL data.  I am confident that this is because something is funky in the wiener filtering algorithm somewhere.  The indicator of this is that the wiener filtering calculation takes the same amount of time (~95 seconds) to calculate a filter for 64 seconds of data as for 1 hour of data (both for N = 2000 taps). 

 

For reference, attached are my plots for the wiener filtering result for (1) 64 seconds of S5 data, and for (2) 3600 seconds of S5 data.

These plots were made using H1:DARM_CTRL as the signal to minimize, with 4 seismometers as the witness channels (EX_SEISX, EY_SEISY, LVEA_SEISX, LVEA_SEISY)

 

I'm working on figuring out what's going on with the filtering algorithm, and why it does work for C1:MC_L minimization, but does not work for H1:DARM_CTRL minimization.

 

 

 

Attachment 1: h1_DARM_64s_4seis_25Feb09.png
h1_DARM_64s_4seis_25Feb09.png
Attachment 2: h1_DARM_3600s_4seis_25Feb09.png
h1_DARM_3600s_4seis_25Feb09.png
  1339   Thu Feb 26 01:24:44 2009 YoichiUpdateComputersMartian wireless is back
Today, a new wireless router arrived.
I configured and installed it. Now the martian wireless network is back.
I updated the wiki page about the wireless network.
http://lhocds.ligo-wa.caltech.edu:8000/40m/Network
  1341   Thu Feb 26 19:59:23 2009 YoichiUpdateLockingDaytime locking
Osamu, Yoichi

We tried locking today from about 2PM.
It took about 1000sec on average to acquire the initial lock.
After the initial lock is achieved, the hand-off/ramp-up steps were reasonably robust, although the AS beam sometimes fluctuates a lot (not good for mental health).

Like last night, the IFO loses lock at around arm-power=8.
We measured the CARM AO path loop gain at arm-power=4. We used the SR785 connected to the A-excitation channel of the common mode board through my TFSR785.py script.

The first attachment is the transfer function measured right after the arm power was ramped up to 4.
The overall bandwidth of the CM servo is only 400Hz. Note that since this is the loop gain of only the AO path, the low frequency gain is eaten by the MCL path.
The second attachment is the same transfer function measured after the AO path gain was increased by 6dB.
It is evident that the AO path is working.
We increased both the AO path and MCL gain by 18dB. The third attachment is the AO path TF in this state.
We then increased the arm power but lost lock at arm-power=6. We should have checked the DARM loop too.
BTW, these plots are automatically generated when you use TFSR785.py for transfer function measurements.


I added -notickle option to c1_watch_dr_bang, since tickling seems to be not necessary during the daytime (actually the initial lock was easier with no tickling).

As the construction work in the next building is now calmed down, I think it is ok to do locking during the day time, though I still plan to come at night.
The improvement of my brain efficiency during the day time may compensate for the longer wait time for initial lock.
Attachment 1: CM1.png
CM1.png
Attachment 2: CM2.png
CM2.png
Attachment 3: CM3.png
CM3.png
  1343   Fri Feb 27 13:49:19 2009 robUpdateLockingthurs night

Could not get past arm power of ~11 or so.  I was suspicious of the transmon high-gain/low-gain PD handover, so I ran the matchTransMon scripts, but that did not help.  I also removed the line in the cm_step script that increased the CM gain by 18dB at an arm power of 4.  The gain of the CM servo will increase naturally as the power in the IFO builds up, so it may not be good to crank it right away.  I tried several other CM gains, and watched the DARM loop, but still could not get past an arm power of ~10-11.  I'm not sure what's wrong, but it may be that mysterious CM-servo/McWFS conspiracy, so we can try turning down the McWFS gain next time.

  1344   Mon Mar 2 03:57:44 2009 YoichiUpdateLockingSunday night locking
Tonight's locking started with a boot fest of the FE computers which were all red when I came in.
It also took me sometime to realize that C1:IOO-MC_F was returning always zero to tdsavg, causing the offloadMCF script to do nothing.
I fixed this by rebooting c1iovme and c1iool0.

Like Rob on the thursday night, I was only able to reach arm power around 10.
This time, I turned down the MC WFS gain to 0.02 (from 0.3).
I also checked gains of most of the loops (MICH, PRC, SRC, DARM, CARM-MCL, CARM-AO).
All the loops looked fine until the lock was lost suddenly. Also the spectrum of MC_F did not change as the arm power was ramped up.
Actually, I was able to reach arm power=10 only once because I spent a long time checking the loop gains and spectrum at fine steps of the arm power.
So it is quite possible that this loss of lock was just caused by a seismic kick.
  1346   Mon Mar 2 21:16:32 2009 YoichiUpdateLockingLow-gain High-gain PD switching may not be working well
Osamu, Yoichi

This afternoon, I run the locking script while doing calculations for the upgrade.
The IFO lost lock even at lower arm powers (around 6) if it was operated for a while (~ 5min).
It seemed as if there were some intermittent glitches (seismic? laser?) causing the lock losses.
We also saw once the TRX and TRY signals saturated at around arm power = 11 when there was a large fluctuation in the arm power.
Osamu suggested that it looked like the high-gain to low-gain PD switching was not working.

I won't come tonight as I may have caught a cold, but if someone comes tonight, it is worth checking the PD switching.
  1347   Tue Mar 3 08:44:31 2009 steveUpdatePEMair cond. maintenance today
IFO room 104 air conditions will be shut down for maintenance today.
This should be finished by noon.
The temperature and particle count variation can be more than usual.
  1348   Tue Mar 3 10:39:07 2009 AlbertoUpdateLSCc1lsc discontinued functioning

The c1lsc has been unstable since last night. Its status on the DAQ screen was oscillating from green to red every minute.

Yesterday, I power recycled it. That brought it back but the MC got unclocked and the autolocker could not get engaged. I think it's because the power recycling also turned c1iscaux2 off which occupies the same rack crate.

Killing the autolocker on op340 e restarting didn't work. So I rebooted also c1dcuepis and burt-restored almost all snapshot files. To do that, as usual, I had to edit the snapshot files of c1dcuepics to move the quotes from the last line.

After that I restarted the autolocker that time it worked.

This morning c1lsc was again in the same unstable status as yesterday. This time I just reset it (no power recycling) and then I restarted it. It worked and now everything seems to be fine.

  1350   Tue Mar 3 19:26:44 2009 YoichiUpdateLockingLow-gain High-gain PD switching may not be working well
I checked the switching of the QPDX from high gain to low gain.
Switching happens as expected, but the low gain QPDX output was very low compared to QPDY.
Also the digital gain for the high gain TRX was not matched with the low gain one. So when the switching happens, there is a large jump in the TRX.

I also found that the offset values for the low gain QPDX were totally wrong. I adjusted it.
Then I removed a beam splitter in front of the QPDX to increase the power falling on it.
But still the low gain QPDX output is four times lower than the low gain QPDY.

I'm still working on it. So don't expect the switching to work correctly at this moment.
I'm planning to be back after the dinner.


Quote:
Osamu, Yoichi

This afternoon, I run the locking script while doing calculations for the upgrade.
The IFO lost lock even at lower arm powers (around 6) if it was operated for a while (~ 5min).
It seemed as if there were some intermittent glitches (seismic? laser?) causing the lock losses.
We also saw once the TRX and TRY signals saturated at around arm power = 11 when there was a large fluctuation in the arm power.
Osamu suggested that it looked like the high-gain to low-gain PD switching was not working.

I won't come tonight as I may have caught a cold, but if someone comes tonight, it is worth checking the PD switching.
  1351   Tue Mar 3 23:59:26 2009 YoichiUpdateLockingLow-gain High-gain PD switching may not be working well

Quote:
I checked the switching of the QPDX from high gain to low gain.
Switching happens as expected, but the low gain QPDX output was very low compared to QPDY.
Also the digital gain for the high gain TRX was not matched with the low gain one. So when the switching happens, there is a large jump in the TRX.

I also found that the offset values for the low gain QPDX were totally wrong. I adjusted it.
Then I removed a beam splitter in front of the QPDX to increase the power falling on it.
But still the low gain QPDX output is four times lower than the low gain QPDY.

I'm still working on it. So don't expect the switching to work correctly at this moment.
I'm planning to be back after the dinner.



This sounds like the QPD whitening gain sliders may be stuck. The slider twiddling script should be run, or the sliders should be twiddled by hand.
  1352   Wed Mar 4 03:37:51 2009 YoichiUpdateLockingLow-gain High-gain PD switching may not be working well

Quote:

This sounds like the QPD whitening gain sliders may be stuck. The slider twiddling script should be run, or the sliders should be twiddled by hand.


Yes, it was indeed the whitening gain slider problem.
I moved them and the QPDX gain went up suddenly. I reinstalled the BS in front of the QPDX and adjusted the offsets, gains accordingly.
Now the high-gain to low-gain switching is fine.
  1353   Wed Mar 4 03:50:17 2009 YoichiUpdateLockingStill won't go above arm power 10
Yoichi, Kentaro

The IFO still loses lock at around arm power 10.
I attached time series of various error signals when losing lock.
In all of the three cases, both of the arm powers go up rapidly just before losing lock.
(The first attachment was taken before I fixed the QPDX switching, so you can see saturation in TRX.)
But PD1_DC (the DC power in the PRC) did not go up in the third case.

I should also check correlations with laser power, CARM length (OSEM signals), seismic noise etc.
Attachment 1: lockLoss1.png
lockLoss1.png
Attachment 2: lockLoss2.png
lockLoss2.png
Attachment 3: lockLoss3.png
lockLoss3.png
  1354   Wed Mar 4 12:38:07 2009 AlbertoUpdateComputer Scripts / Programs3f locking simulations
I simulated the REFL signals demodulated at the differential frequencies of the sidebands (f2-f1), at their summed frequencies (f2+f1). I also simulated their combination as in the Double Demodulation.
 
I repeated the simulation for:
- Old (current) 40m
- 40m Upgrade
- AdvLIGO
 
I'm attaching the results to this elog entry.
 
The plots show how the signal varies exploring the two-dimensional space of the demodulation frequencies (differential and sum).

 Both the Upgrade and the Old40m's signals look anomalous since the zero-crossing point does not change with the demodulation phases.

I suspect there's is a problem with the optickle model of the 40m.

Attachment 1: 23_3f_40mOld_DDplots.pdf
23_3f_40mOld_DDplots.pdf 23_3f_40mOld_DDplots.pdf 23_3f_40mOld_DDplots.pdf
Attachment 2: 59_3f40m_Upgrade_DDplots.pdf
59_3f40m_Upgrade_DDplots.pdf 59_3f40m_Upgrade_DDplots.pdf 59_3f40m_Upgrade_DDplots.pdf
Attachment 3: 20_3f_AdvL_DDplots.pdf
20_3f_AdvL_DDplots.pdf 20_3f_AdvL_DDplots.pdf 20_3f_AdvL_DDplots.pdf
  1355   Wed Mar 4 17:20:04 2009 josephbUpdateCamerasCamera code upgrades

I've updated the digital camera python code as well as changed the network topology.

At the moment, both cameras are connected to a small gigabit switch which only talks to Ottavia.  This means all camera servers must be run on Ottavia, allow camera output is still UDP multicast so any machine capable of running gstreamer can pick up the images.

The server and client programs now have the ability to read a configuration file for the setup of the cameras.  They default to pcameraSettings.ini, but this can default can be changed with a -c or --config option

For example, "serverV3.py --config pcam1.ini" will run the server using the pcam1.ini settings file.  Similarly, "client.py --config pcam1.ini" will also take the IP settings from the config file so that it knows at which port and IP to listen.

These programs and .ini files have been placed in /cvs/cds/caltech/apps/linux64/python/pcamera/

I've updated the cshrc.40m aliases so that it uses the new configuration file options, so now pcam1 calls "client.py -c pcam1.ini" in the above directory.

So to start a client use pcam1 or pcam2 (for the 32223 camera in PSL looking at MC trans or 44026 looking at an analog moniter in the control room respectively).  These can be run on Allegra, Rosalba or Ottavia at the moment.

To start a server, use pserv1 or pserv2.  These *must* be run on Ottavia.

I've also added a -n or --no-gui option at Yoichi's request, one which just starts up and plays, with no graphical gui.

Lastly, I've made some changes to the base pcamerasrc.py file, which should make display more robust.  After a failed transmission of an image from the camera to Ottavia, it should re-attempt up to 10 times before giving up. I'm hoping this will make it more robust against packet loss.  The change in network topology has also helped this, allowing 640x480 to be transmitted on both cameras before tens of minutes before a packet loss causes a stop.

  1358   Thu Mar 5 00:06:32 2009 Kakeru, RanaUpdateIOOWFS centering
We found that the MC REFL image was no longer round and that the MCWFS DC quadrant spots were mostly
in one quadrant. So we re-centered the MCWFS beams in the following way:

1) We unlocked the MZ and adjusted the PZT voltage to keep the beam on the WFS from saturating.
2) Re-aligned the black hole beam dump to center its beam in its aperture.
3) centered the beam on the MCWFS optics and MCWFS QPD displays.
4) Relocked MC.

Below is the image of the IOO Strip tool. You can see that the MC REFL DC is now more flat. The
MC pointing has also been changed (see the MC TRANS HOR & VERT channels). The MC transmitted
light is also now more stable and higher.

We tried to center the QPD, and we found that there were a few hundred mV of dark offset for each 
quadrant of QPD. We adjusted them with this scripts:
/cvs/cds/caltech/scripts/MC/WFS/McWFS_dc_offsets
Attachment 1: IOO_graph.jpg
IOO_graph.jpg
  1359   Thu Mar 5 01:09:29 2009 rana, albertoUpdateDMFstill not working
We tried to run DMF on mafalda, but it didn't work. I tried to compile it using Rob's elog instructions.
On mafalda, I started matlab and ran the mdv_config.m to set up mDV. I tested that the seisBLRMS.m
script ran and correctly produced changes in the seisBLRMS strip tool. but when I tried to compile it I got:
>> mcc -v -m -R -nojvm seisBLRMS.m
Warning: Duplicate directory name: /cvs/cds/caltech/apps/linux/matlab/toolbox/local.
Compiler version: 4.6 (R2007a)
Processing /cvs/cds/caltech/apps/linux/matlab/toolbox/matlab/mcc.enc
Processing /cvs/cds/caltech/apps/linux/matlab/toolbox/signal/mcc.enc
Processing /cvs/cds/caltech/apps/linux/matlab/toolbox/control/mcc.enc
Processing /cvs/cds/caltech/apps/linux/matlab/toolbox/filterdesign/mcc.enc
Processing /cvs/cds/caltech/apps/linux/matlab/toolbox/shared/controllib/mcc.enc
Processing /cvs/cds/caltech/apps/linux/matlab/toolbox/ident/mcc.enc
Warning: an error occurred while parsing class FilterDesignDialog.AbstractEditor:
Undefined function or variable 'DAStudio.Object'.

> In /cvs/cds/caltech/apps/linux/matlab/toolbox/shared/filterdesignlib/@FilterDesignDialog/@CoeffEditor/schema.p>schema at 9
Warning: an error occurred while parsing class FilterDesignDialog.CoeffEditor:
Invalid superclass handle.

Processing /cvs/cds/caltech/apps/linux/matlab/toolbox/fixedpoint/mcc.enc
terminate called after throwing an instance of 'ApplicationRedefinedException*'
Abort (core dumped)
"/cvs/cds/caltech/apps/linux/matlab/bin/mcc" -E "/tmp/fileRnU5Qj_31324": Aborted
??? Error executing mcc, return status = 134.

In the meantime, I've started up a green terminal on allegra which is ssh'd into megatron.
On megatron there is a regular matlab session which is running seisBLRMS.m as a matlab script
and the seis DMF channels are getting updated.
  1361   Thu Mar 5 05:07:09 2009 YoichiUpdateLockingWednesday night locking
Tonight, I was having a problem with the PO_DC hand-off.
It fails most of the time.
I increased the averaging time for the PD1_DC offset measurement.
I also wrote a script to match the gain of the transmission DC and the PO_DC signals.
This script (/cvs/cds/caltech/scripts/CM/matchPODCGain ) measures the gains of the old (TRX+TRY) and new (PO_DC) signals at 150Hz and returns the optimal value to be put into the input matrix.
cm_step script calls matchPODCGain to determine the matrix element value for the PO_DC signal.

Even with this script, the hand-off was still unreliable.
I checked the AO path loop gain just before the hand off. It looked normal.
Then I realized that the oscilloscope I hooked up to the PO_DC signal using a T-BNC may be introducing some noise into the channel.
So I removed it. Then the PO_DC hand off went well at least once.
The IFO still loses lock at around arm power 10.

I attached time series of the latest lock loss. The second attachment is a zoom of the first one.
This time, there is a glitch in the ETM feedback signals, which is also present in the DARM and CARM and error signals.
I saw this kind of glitches several times today.
Attachment 1: lockLoss5.png
lockLoss5.png
Attachment 2: lockLoss5-zoom.png
lockLoss5-zoom.png
  1364   Fri Mar 6 05:44:14 2009 YoichiUpdateLockingLocking distracted by the QPD whitenning problem again
Tonight, I was able to ramp up the arm power to around 20. Then the DARM loop started to oscillate and the IFO lost lock in a few seconds.
I repeated this several times, then realized that the transmission QPDs were not working properly again due to the well known sticky slider problem.
I should have run slider_twiddle script. Since the DARM RF signal is normalized by the sqrt(TRX+TRY), it is reasonable that the DARM loop got unstable.

The fact that I was able to go up to arm power = 20 means there is nothing saturating below this power level.
  1365   Fri Mar 6 15:23:39 2009 YoichiUpdateLockingLocking distracted by the QPD whitenning problem again
By looking at the time series of DARM signal at the time of a lock loss, the oscillation frequency was about 3.5kHz (see the attm1 and its zoomed version attm2).
I will measure the DARM loop gain around this frequency next.


Quote:
Tonight, I was able to ramp up the arm power to around 20. Then the DARM loop started to oscillate and the IFO lost lock in a few seconds.
I repeated this several times, then realized that the transmission QPDs were not working properly again due to the well known sticky slider problem.
I should have run slider_twiddle script. Since the DARM RF signal is normalized by the sqrt(TRX+TRY), it is reasonable that the DARM loop got unstable.

The fact that I was able to go up to arm power = 20 means there is nothing saturating below this power level.
Attachment 1: lockLoss3.pdf
lockLoss3.pdf
Attachment 2: lockLoss3-zoom.pdf
lockLoss3-zoom.pdf
  1366   Fri Mar 6 18:14:58 2009 YoichiUpdateComputersawg not working
Starting from this afternoon, the awg is not working.
I rebooted FE computers, c0daqawg as well as tpman and daqd processes on fb40m several times.
But the problem is still there.
I sent an email to Alex.
  1369   Sat Mar 7 16:50:25 2009 YoichiUpdateComputersNot even data retrieval working
Now our digital system is really in trouble.
We can't even get data from tp channels.

I did another round of computer reboots, this time including the RFM bypass switch, c0daqctrl, c0dcu1 and fb40m itself.
But the problem still persists.

I guess there is nothing I can do until Alex comes in.
  1370   Sun Mar 8 23:09:26 2009 ranaUpdateComputersNot even data retrieval working
Although getting the regular DAQ data works, we can't get any testpoints.

I tried restarting tpman several times; there's no inittab on fb40m for this so we should get Alex to set one up when he comes.
I also tried various power cycles and reboots: daqawg, daqctrl, etc. I also notice that Osamu's setup of new stuff is connected to
the same rack and power strips as all of our sensitive DAQ machines. We should find out if there was any hardware installed in the
last couple days; it would be easy to accidentally unplug or damage on of our fibers.

I moved the old tpman.log over to tpman.log.090308. It starts out with a header and then just lists when each TP is requested.

When restarting tpman it puts the following into the terminal:
fb:controls>./tpman &
[1] 1037
fb:controls>VMIC RFM 5565 (0) found, mapped at 0x2868c90
VMIC RFM 5579 (1) found, mapped at 0x2868c90
Could not open 5565 reflective memory in /dev/daqd-rfm1
16 kHz system
Spawn testpoint manager
Channel list length for node 0 is 4168
Test point manager (31001001 / 1): node 0
which is OK?; its the same startup outputs that are in the old log file. It would be nice if there was not and error message about the RFM.
Requesting new testpoints via tdsdata, dtt, or the diag command line doesn't seem to work. tpman doesn't spit anything out although 'tp show 0'
does show that the TP is selected.

Once Alex fixes the 'tpman' issue, we should make sure to put an inittab or startup script in there so that tpman writes a log
file and also archives its old log files upon a restart.
  1371   Sun Mar 8 23:14:52 2009 ranaUpdateComputer Scripts / Programstdsdata doesn't work

Matt logged in and rebuilt the TDS stuff for us on Mafalda in /cvs/cds/caltech/apps/linux/tds_090304.

He says that he can't build his stuff on 64-bit because there's not a sanctioned 64-bit build of GDS yet.

This should have all the latest fixes in it. I tried using both the old and new code from allegra and they both are fine:

./tdsdata 16384 2 C1:IOO-MC_F > /users/rana/test.txt

I loaded the data I got with the above command and there were no data dropouts. Possibly the dropout problem is only

associated with testpoints and so we have to wait for the TP fix.

  1373   Mon Mar 9 11:09:33 2009 AlbertoUpdateComputersRe: Not even data retrieval working

Quote:
Although getting the regular DAQ data works, we can't get any testpoints.

I tried restarting tpman several times; there's no inittab on fb40m for this so we should get Alex to set one up when he comes.
I also tried various power cycles and reboots: daqawg, daqctrl, etc. I also notice that Osamu's setup of new stuff is connected to
the same rack and power strips as all of our sensitive DAQ machines. We should find out if there was any hardware installed in the
last couple days; it would be easy to accidentally unplug or damage on of our fibers.

I moved the old tpman.log over to tpman.log.090308. It starts out with a header and then just lists when each TP is requested.

When restarting tpman it puts the following into the terminal:
fb:controls>./tpman &
[1] 1037
fb:controls>VMIC RFM 5565 (0) found, mapped at 0x2868c90
VMIC RFM 5579 (1) found, mapped at 0x2868c90
Could not open 5565 reflective memory in /dev/daqd-rfm1
16 kHz system
Spawn testpoint manager
Channel list length for node 0 is 4168
Test point manager (31001001 / 1): node 0
which is OK?; its the same startup outputs that are in the old log file. It would be nice if there was not and error message about the RFM.
Requesting new testpoints via tdsdata, dtt, or the diag command line doesn't seem to work. tpman doesn't spit anything out although 'tp show 0'
does show that the TP is selected.

Once Alex fixes the 'tpman' issue, we should make sure to put an inittab or startup script in there so that tpman writes a log
file and also archives its old log files upon a restart.


Alex fixed the problem. It was caused by the awgtpman running on kami1.martian which conflicted with the tpman in fb0.

Killing awgtpman on kami1 allowed for the tpman on tp0 to work properly again.

If more test points are needed, Alex suggested to tune the GDS settings accordingly.
What this actually means, I still have to understand it.
  1374   Mon Mar 9 12:04:18 2009 YoichiUpdateComputersTPs and AWG are back
I had to do one more reboot of tpman and daqd to get the TPs working.
I confirmed the alignment scripts run fine.

Now the oplevs of some optics are largely mis-centered. Alberto and I will center them after lunch.
  1375   Mon Mar 9 14:57:30 2009 KakeruUpdateComputer Scripts / Programstdsdata doesn't work

I tested new tdsdata and found it was working well.

I excited C1:SUS-ITMY_SUSPIT_EXC with tdssine, and get data from C1:LSC-TRY_OUT (testpoint) and C1:SUS- ITMY_OPLEV_PERROR (recorded point) with new and old tdsdata.

With old tdsdata (/cvs/cds/caltech/apps/linux/tds/bin/tdsdata), I found some jumps of datapoint, which is a same problem with before (Attachment 1).

With new tdsdata (/cvs/cds/caltech/apps/linux/tds_090304/bin/tdsdata), there looks to be no jumps (Attachment 2; taken about 10 minutes after Attachment 1).

The problem of old tdsdata looks to be remaining even for recordedpoints.

You should use /cvs/cds/caltech/apps/linux/tds_090304/bin/tdsdata.

Quote:

Matt logged in and rebuilt the TDS stuff for us on Mafalda in /cvs/cds/caltech/apps/linux/tds_090304.

He says that he can't build his stuff on 64-bit because there's not a sanctioned 64-bit build of GDS yet.

This should have all the latest fixes in it. I tried using both the old and new code from allegra and they both are fine:

./tdsdata 16384 2 C1:IOO-MC_F > /users/rana/test.txt

I loaded the data I got with the above command and there were no data dropouts. Possibly the dropout problem is only

associated with testpoints and so we have to wait for the TP fix.

 

Attachment 1: oldtds.png
oldtds.png
Attachment 2: newtds.png
newtds.png
  1376   Mon Mar 9 16:54:52 2009 Kakeru, RanaUpdateComputer Scripts / Programstdsdata doesn't work

We confirmed that new tds(/cvs/cds/caltech/apps/linux/tds_090304/) works well on linux 64, and replaced it to /cvs/cds/caltech/apps/linux/tds/

The old /cvs/cds/caltech/apps/linux/tds is put in /cvs/cds/caltech/apps/linux/tds.bak

  1379   Mon Mar 9 19:33:10 2009 ranaUpdateDMFseisBLRMS in temp condition

The seisBLRMS has been running on megatron via an open terminal ssh'd into there from allegra with matlab running. This

is because I couldn't get the compiled matlab functionality to work.

Even so, this running script has been dying lately because of some bogus 'NDS' error. So for today I

have set the NDS server for mDV on megatron to be fb40m:8088 instead of nodus.ligo.caltech.edu. If this seems to fix the problem

I will make this permanent by putting in a case statement to check whether or not the mDV'ing machine is a 40m-martian or not.

Attachment 1: Untitled.png
Untitled.png
  1380   Mon Mar 9 23:13:22 2009 YoichiUpdateComputer Scripts / Programstdsdata doesn't work

Quote:

We confirmed that new tds(/cvs/cds/caltech/apps/linux/tds_090304/) works well on linux 64, and replaced it to /cvs/cds/caltech/apps/linux/tds/

The old /cvs/cds/caltech/apps/linux/tds is put in /cvs/cds/caltech/apps/linux/tds.bak

 The tdscntr.pl in the new tds was probably the one from LLO, which is actually the version I sent to Tobin. It had paths and channel names defined for the LLO. So I copied back my original 40m version.

  1382   Tue Mar 10 04:55:41 2009 YoichiUpdateLockingLocking: 3.7kHz large oscillation
Yoichi, Jenne, Alberto,

As I reported on the last Thursday, there is a large oscillation in CARM and DARM error signals (attm1).
I put notch filters (3.75kHz, Q=10, 30dB) in the CARM and DARM loops. This let us go up to the arm power of more than 20 and stay there for a while.
The dashed curves in the attm1 are the spectra when the notches are off, and the solid curves are when the notches are used.
We could somewhat suppress the DARM peaks but not CARM.
Of course this is clearly not a good solution. We should find the cause of the oscillation and kill it.

Attm2 is the spectrum of the PO_DC signal flowing in the CM board measured by the SR785. More specifically, CH1 is TP1A and CH2 is TP2A of the CM board.
This was taken right after the AO path was engaged. At this stage, the AO path gain is very low. But you can already see a seed of the oscillation in the spectrum.

Attm3 shows the same spectra taken after the arm power is increased to 4 but before the PO_DC hand off. You can see large peaks around 3.75kHz.
After this, the peaks grow as the power goes up.

Attm4 is the loop gain of the AO path after the PO_DC hand off (arm power = 4).
Attm5 is the zoom of the same TF around 3.7kHz. Clearly there is something wrong at this frequency. We should check the CM board and the MC board as well as the SPOB PD.

One time I was able to go up to arm power = 27 or so. At this power level, the DARM loop started to oscillate, probably, around the UGF.
However because of the 3.7kHz problem, we can't stay at this power level long enough to make diagnostic measurements (like open loop TF).
We should tackle the 3.7kHz issue first.
Attachment 1: CARM_DARM_Spectra.pdf
CARM_DARM_Spectra.pdf
Attachment 2: PODC_Spe_AOPath_Engaged.png
PODC_Spe_AOPath_Engaged.png
Attachment 3: PODC_Spe_before_PODC_handoff.png
PODC_Spe_before_PODC_handoff.png
Attachment 4: AOGain3.png
AOGain3.png
Attachment 5: AOGain2.png
AOGain2.png
  1386   Wed Mar 11 14:51:01 2009 Kakeru, Joe, RobUpdateIOOMC alignment

This morning, MC alignment was gone and MC wasn't lock.

We checked old value of pitch, yaw, and position offset of each MC mirror, and found they were jumped.

We don't know the reason of this jump, but we restore each offset value and MC backed to lock.

  1387   Wed Mar 11 16:41:22 2009 steveUpdateMOPAspare NPRO power

The spare M126N-1064-700,  sn 5519 of Dec 2006 rebuilt  NPRO's power output

 measured   750mW at DC2.06A with Ohpir meter.

Alberto's controller  unit 125/126-OPN-PS,  sn516m was disconnected from lenght measurment NPRO on the AP table.

5519 NPRO  was clamp to the optical table  without heatsink and it was on for 15 minutes.

  1388   Wed Mar 11 16:53:48 2009 YoichiUpdateLockingJunks in around kHz
Rana, Yoichi

Last night, we tried to find out the source of the kHz region peaks in the DARM and CARM error signals.
These peaks are also present in the error signal of the single arm locking by RF (both X and Y).
The attachment 1 shows spectra of MC_F and XARM error signal when XARM is locked by the POX PDH signal.
There is a sharp peak at 3.8kHz in MC_F. This peak was there in a reference spectrum taken on June 24 2008.

In the XARM error signal, there is also a broad peak around 3.8kHz. This peak moves between 3.75kHz and 3.8kHz from time to time.
(the brown curve was taken when the peak moved to 3.75kHz).
Also there is a notch like structure at 3.8kHz in the XARM error spectrum. Looks like the peak in the MC_F is creating a notch here, but
no idea why.

We tapped on the PSL table, the end chambers and the SPOB table and looked at the spectra to see if there is any change.
Rana also developed a cool Walkie-Talkie excitation technique, where he put one of the walkie-talkies on the PSL table by the MZ and yelled at the other one while looking at a DTT screen in the control room.
None of these had any effect on the XARM error, while MC_F responded to the disturbances.

We also turned on and off the steering mirror PZT closed loop buttons, moved the PMC, MZ and the ISS gain sliders and changed the MC gain, offset.
Nothing affected the XARM error.

Osamu found old spectra of the XARM signal (attm2). The legends say DARM but these are XARM signals.
Almost the same structures can be seen including the notch at 3.8kHz. Seems like it's been like this for long time.

We should check, RF-AM, MC coil dirivers, Piezo-Jena noise etc.
Attachment 1: MC_F-XARM.pdf
MC_F-XARM.pdf
Attachment 2: old-xarm.pdf
old-xarm.pdf
  1389   Wed Mar 11 21:03:51 2009 Kakeru and KiwamuUpdateIOOPSL angle QPD

Kakeru and Kiwamu

We placed a QPD on the PSL bench for PSL angle monitor.

 

Quote:

I checked a broken QPD, which was placed for PSL angle monitor, and finally I cocluded one segment of the quadrant diode was broken.

The broken segment has a offset voltage of -0.7V after 1st I-V amplifier. It means the diode segment has a current offset without any injection of light.

Tomorrow I will check a new QPD for replacement.

Kiwamu IZUMI

 

 As we mentioned before, old QPD which used to be placed is broken.

 And we put broken QPD into the "photodiodes" box under the soldering table.

 

 

  1390   Wed Mar 11 22:57:48 2009 YoichiUpdateLockingCalibrated XARM error signal spectrum
I did a rough calibration of the XARM error spectrum.
See the attached calibrated spectrum.

I started from this Rana's elog entry.
http://www.ldas-sw.ligo.caltech.edu/ilog/pub/ilog.cgi?group=40m&task=view&date_to_view=04/07/2005&anchor_to_scroll_to=2005:04:07:20:28:36-rana

I first injected a 20Hz sin signal into C1:SUS-ETMX_LSC_EXC and measured the response to the ETMX SUSPOS.
Using the calibration of the SUSPOS given in the above entry, I calibrated the ETMX coil actuation efficiency.
It was 3.4e-12 m/cnt @20Hz for C1:SUS-ETMX_LSC_EXC.

Then I locked the X-arm and injected a calibration peak at 20Hz.
From the ratio of the peaks in C1:SUS-ETMX_LSC_IN2 and C1:LSC-XARM_IN1, I calibrated the X-arm error signal to be 4.2e-13 m/cnt.
We have to also take into account the cavity pole of the arm, 1525Hz (the design value, may not be actual).
So I used the following calibration in the DTT:

G: 4.2e-13
P: 1525
Z:

Note that the attached spectrum shows the actual motion of the X-arm (or equivalent frequency noise) after suppressed by the feedback servo,
unlike conventional noise spectra showing "virtual" displacement which would have been induced in the absence of servos.
Attachment 1: XarmErrorSpeCalibrated.pdf
XarmErrorSpeCalibrated.pdf
  1391   Wed Mar 11 23:41:33 2009 Kakeru, YoichiUpdateIOOWFS centering

We found the MC reflection was distorted . And WFC beam went to upward of QPD

We recentered WFC beam and these problems were fixed

ELOG V3.1.3-