  12871   Mon Mar 6 16:32:36 2017 SteveUpdateGeneralold NPRO

16 years old Lightwave NPRO M126-1064-700, sn 415 power output is tripping continously to zero.

The Lightwave Controller 125/126-OPN-POS sn516 was used in this test. Settings were lowered to close to nominal values without any success.

One can not determine what is broken: head or controller. This NPRO head was under Manasa's desk.
 

  12872   Tue Mar 7 15:17:19 2017 SteveBureaucracyGeneralproperty tag

Property tag found.

Attachment 1: property_tag.jpg
property_tag.jpg
  12875   Thu Mar 9 15:25:12 2017 KojiUpdateGeneralIMC/XYarms aligned/locked

As per Steve's request, I've checked the alignment of the IMC and the arms. These three cavities are locked and aligned.

  12876   Thu Mar 9 17:26:43 2017 SteveUpdateGeneralattempted ETMY picture taking

I removed the video monitoring can and replaced it with Olympus SP-570UZ camera. It has no IR blocker. The OSEM light are dominant because I can not zoom in more.

I left the camera in place so you can try it. Leave the LEXAN plate on the glass window so no accident can happen. The illuminator is on and you can turn it off-on with the manual switch, close to the camera. Camera manual is on my desk.

 

  12877   Thu Mar 9 20:11:04 2017 KojiUpdateGeneralattempted ETMY picture taking

The attached is the ETMY image with the single arm locked. This was the best I could do. Here is the recipe

  • Turn on SP570UZ
  • Switch to "M" mode (Manual aperture and exposure)
  • Set the aperture to be the widest (smallest F number) and the exposure to be maximum (15 second).
  • Switch to AF mode by the lens side switch
  • Use the lens dial to adjust the zoom until the OSEMs fill the central 1/3 box (i.e. 1/9 area of the field of view). If you zoom more, you can't focus the spot later.
  • Use menu button to switch to ISO1600 (You are now capable to see the beam spot)
  • Switch to MF mode by the lens side switch
  • Use the lens dial to adjust the focus to have the sharpest image of the spot. This can be achieved at the focal distance of ~1m
  • Use menu button to switch back to ISO64
  • Push the shutter (I didn't use it, but you should be able to use 2sec timer)
Attachment 1: P3090032.JPG
P3090032.JPG
  12881   Fri Mar 10 18:00:22 2017 SteveUpdateGeneralattempted ETMY picture taking

Your technique works Koji

Attachment 1: P3100044.JPG
P3100044.JPG
Attachment 2: P3090024.JPG
P3090024.JPG
  12929   Wed Apr 5 16:05:47 2017 gautamUpdateGeneralNB code checkout

[evan, gautam]

We spent some time trying to get the noise-budgeting code running today. I guess eventually we want this to be usable on the workstations so we cloned the git repo into /ligo/svncommon. The main objective was to see if we had all the dependencies for getting this code running already installed. The way Evan has set the code up is with a bunch of dictionaries for each of the noise curves we are interested in - so we just commented out everything that required real IFO data. We also commented out all the gwpy stuff, since (if I remember right) we want to be using nds2 to get the data. 

Running the code with just the gwinc curves produces the plots it is supposed to, so it looks like we have all the dependencies required. It now remains to integrate actual IFO data, I will try and set up the infrastructure for this using the archived frame data from the 2016 DRFPMI locks..

  12948   Wed Apr 19 15:46:24 2017 gautamUpdateGeneral1611/1811 inventory check

I looked through the lab area to do a fast photodiode inventory check, as we may need to buy some for the higher order mode spectroscopy SURF project. I looked on the following optical tables: ETMY, ITMY, BS, AS, PSL, SP, ITMX, Jenne laser table, and ETMX, as well as the photodiode cabinet, and could only find two 1611s. Here is a summary of the inventory: 

  • Power supply 0901: 2x in photodiode cabinet (E6 along the Y arm), 1x on Jenne laser table
  • Newfocus 1611 S/N 7284-WX, labelled "REF DET" on ITMY optical table, currently unused
  • Newfocus 1611 S/N 57109 on Jenne laser table

I have not yet checked if these photodiodes are in working order.

 

  12950   Tue Apr 25 19:35:41 2017 gautamUpdateGeneralIPCS -q

Dataviewer wouldn't launch on pianosa - it seemed to work fine on Donatella though. Rana suggested using the ipcs -q command. The complete fix can be found in this elog. This did the trick, dataviewer runs fine on Pianosa now...

  12951   Wed Apr 26 01:00:23 2017 gautamUpdateGeneralDRMI locking

Since we'd like to get back to DRSE locking, I tried locking the DRMI tonight. I did the following:

  • First, I aligned the arms, and ran the dither alignment scripts to maximize the arm transmission
  • Next, I misaligned the ETMs, and tried to lock the PRC resonant for the carrier (i.e. PRCL on REFL11I, MICH on AS55Q). I got brief lock stretches of a few seconds but not longer. Turns out the AS55 beam was barely hitting the photodiode. I guess this wasn't looked at since Johannes modified the AS path for the loss measurements. Anyways, it just required a minor tweak to center the beam on the AS55 photodiode.
  • Once the PRC was locked, I ran the PRC and MICH dither align scripts. The way these are set up right now, the error signals to these servos are REFLDC and ASDC respectively (demodulated at the respective dither frequencies). But looking at the spots on the ITM cameras with the PRC resonant, the spots seem shifted (in both PIT and YAW) relative to the spots when the arm cavity is resonant. Shouldn't they be the same mode? Or maybe I am missing something.
       
  • Next, I tried to lock the DRMI with the 1f error signals: i.e. PRCL on REFL11 I, SRCL on REFL55 I, and MICH on AS55 Q. After some demod phase tweaking, I was able to get some locks going. Turning on the PRC angular feedforward seemed to help the locking, but I have no idea if the installed filters are still the correct ones. I believe the POP QPD channels are the witnesses used to train this filter, I will look at the predicted vs achieved subtraction.
  • At this point, I was able to get locks lasting a few minutes - see the attachment. I ran the UGF servos and tweaked the loop gains a little, but before I could start a loop measurement, I lost the lock. I am calling it for the night.

GV 26 April 2017, 3pm: Forgot to note yesterday that I re-connected the suspect Satellite box, which has been connected to the SRM signal chain, back to the SRM suspension. I did not see any instances of glitching during my work last night. Also added pictures showing shifted spots on ITMs when PRC is locked relative to when arms are locked...

  12954   Fri Apr 28 02:04:36 2017 gautamUpdateGeneralDRMI locking

I got a couple of ~30min long DRMI lock stretches today. The settings I used are essentially the same as what I had back in November. Though we have since made some changes to the IMC RF signal chain, I guess it is not unreasonable that the LSC Demod phases that worked then work now as well. 

In the lock stretches, I did the following:

  • Took loop measurements for MICH, PRCL, SRCL
  • Turned on the sensing oscillator lines for error signal calibration
  • Tried turning on the analog whitening on AS55, REFL11 and REFL55. The latter two worked fine, but everytime I turned the REFL55 whitening on, I broke the lock. I'm also unable to acquire lock if I leave the whitening turned on all the time. The ADC overflow indicators also indicate frequent overflows when I turn the whitening on. Oddly, this seems to happen even if I turn the analog whitening gain to 0dB - the signals look well within the ADC range on dataviewer and DTT timeseries mode. Not sure what's going on here, I will investigate further tomorrow.
  • We should have some stretches where we can look at the possibility of seismic feedforward for some DRMI length DOFs.

On the side, I'm also looking at whether the PRC angular feedforward filters, last trained in October 2016, remain valid. Even post midnight, I am unable to lock the DRMI without turning on the FF, and looking at the POP QPD PIT and YAW signal spectra with the FF on vs FF off, there is definitely some improvement in the 1-4Hz band (plot to follow), question is whether we can do better and hence improve the DRMI duty cycle/ make the lock acquisition easier. To this end, I centered the beam on the POP QPD after locking and dither aligning the PRC on carrier, and have taken some data to look at.

So, much data analysis to follow - the idea is to put together a DRMI noise budget with Evan's NB code. For now, here are the uncalibrated control signal spectra.

Attachment 1: 20170428_DRMI.pdf
20170428_DRMI.pdf
  12955   Fri Apr 28 13:56:26 2017 ranaUpdateGeneralDRMI locking

one of these signals does not look like the others: explanation?

  12957   Fri Apr 28 19:32:06 2017 gautamUpdateGeneralDRMI locking - PRCL angular FF

I took a closer look at the POP QPD/ PRC angular feedforward situation yesterday. I thought it would be useful to have a POP QPD MEDM screen. Looking at the PIT and YAW channel filter modules, the anti-whitening filters seemed different from what we have for other channels that are connected to the Pentek interface board (e.g. MCL). So I copied over the 150:15 (z:p) filter, and also turned on a 60Hz comb. The LSC offsets script does not set the dark offsets for this QPD, so I manually put in the dark offsets for the PIT, YAW and SUM channels as well. For the locking, I first locked the arms on IR an dither aligned them. Then I locked the PRMI on carrier, ran the PRC dither alignment, and went over to the ITMX pickoff table and centered the beam on the QPD by making the PIT and YAW channel timeseries oscillate around approximately zero. 

After these tweaks, I collected ~40mins of data with the angular FF OFF/ON. I did not DC couple the ITM Oplev servos, but Eric tells me that this did not make a difference to the achievable subtraction in the past. Here is the frequency domain multicoherence analysis - I used the BS_X and BS_Y seismometer channels as witnesses. I've also put a plot with what the raw FF filter coefficients look like (no fitting yet). 

      

Looks like we can do better for both DOFs - it even seems like we are injecting noise with the current FF filters in some bands, perhaps we can do a better job of rolling off the filters outside the band of interest. Eric and I were discussing MATLAB's "reduce" routine for this purpose, I will play around with it and see if I get a better fit.

Unfortunately, I encountered a strange error when trying to pull data with nds2 today, it kept complaining RuntimeError: Too many channels or too much data requestedeven though I have pulled longer stretches of data for more channels with 16k sampling rate as recently as last week. Shorter duration requests (<600 seconds) seemed to work fine though... So I had to use cds.getdata to pull the data, and they're much too large to attach. Has anyone else encountered a similar error?


The mystery of the spots on the ITMs when the PRC is locked on carrier remains - after talking this over with Koji, we figured that even with the carrier resonant, the spot will be much dimmer than the spots when the arms are locked, but what I see on the cameras is still a pretty beefy spot. The real cavity mode is actually visible where it should be (I marked the locations of the spots with arms well-aligned with a marker on the monitors), as given away by some twinkling that is visible only when the cavity is locked. But what ghost beam is so intense it looks almost as bright as when the arm is locked?

GV 10pm 28 April 2017: Turns out this is the spot from the single bounce off the ETM transmitting back through the ITM and hitting the suspension cage (hence the bright spot). Johannes and I confirmed by moving the ETM, the spot moved with it. I just never paid attention to this spot before.

Attachment 1: PRC_angularFF.pdf
PRC_angularFF.pdf
Attachment 2: PRC_TFs.pdf
PRC_TFs.pdf
  12960   Mon May 1 16:29:51 2017 gautamUpdateGeneralDRMI locking

For the traces I posted, I had not turned on the whitening for the SRCL sensing PD (REFL55). However, I took a spectrum on a subsequent lock, with the analog whitening + digital dewhitening turned on for all 3 PDs (AS55, REFL11 and REFL55), and the HF part of the SRCL spectrum still looked anomalous. I'm putting together the detailed NB, but here's a comparison between the signals from the 3 RFPDs with the PSL shutter closed (but whitening engaged, and with the analog gains at the same values as used during the locking).

 

To convert the y-axis into m/rtHz, I used data from a sensing matrix measurement I took yesterday night during a DRMI lock - I turned on lines between 300 Hz and 325 Hz for the 3DOFs for ~5 minutes, downloaded the RFPD error signal data and did the demodulation. I used numbers from this elog to convert the actuator drive from cts to m. The final numbers I used were:

MICH (AS55_Q):   8.706 * 10^11 cts/m

PRCL (REFL11_I): 2.757 * 10^12 cts/m

SRCL (REFL55_I): 1.995 * 10^10 cts/m

So it looks like there may be something weird going on with the REFL55 signal chain. Looking at the LSC rack (and also suggested by an elog search), it looks like the demodulation is done by a demod board labelled "POP55" - moreover, the demodulated outputs are taken not from the regular output ports on this board, but from the "MON" ports on the front panel. 

Quote:

one of these signals does not look like the others: explanation?

 

Attachment 1: LSC_sensingNoise.pdf
LSC_sensingNoise.pdf
  12962   Mon May 1 21:45:54 2017 ericqUpdateGeneralDRMI locking
Comparing counts doesn't get you anywhere; each PD has different whitening gain which may vary from measurement to measurement. The better thing to compare is Volts coming out of the demod board, since this (hopefully) only changes when we touch the PD or analog signal chain; this is what I used for the most recent DRMI sensing measurements. (ELOG 11589) We have calibrated actuator channels in the CAL model, which will give you the control signal in m for the DRMI lengths. Perhaps you can convert your sensing matrix measurement to demod board output volts per meter to compare with the last measurement.

Also, the monitor ports are the LEMO ports to the left; the SMA ports where the signal is coming from are from a daughter board that has a better output opamp that the nominal output; we're using the same output on the REFL11 and AS55 demod boards.
  12963   Wed May 3 16:00:00 2017 gautamSummaryGeneralNetwork Topology Check

[johannes, gautam]

I forgot we had done this last year already, but we updated the control room network switch labels and double checked all the connections. Here is the status of the connections and labels as of today:

There are a few minor changes w.r.t. labeling and port numbers compared to the Dec 2015 entry. But it looks like there was no IP clash between Rossa and anything (which was one of the motivations behind embarking on this cleanup). We confirmed by detatching the cable at the PC end of Rossa, and noticed the break in the ping signals. Plugging the cable back in returned the pings. Because Rossa is currently un-bootable, I couldn't check the MAC address.

We also confirmed all of this by using the web browser interface for the switch (IP = 192.168.113.249).

Attachment 1: Network_topology_3May2017.pdf
Network_topology_3May2017.pdf
  12964   Wed May 3 16:02:36 2017 SteveUpdateGeneralPI pzt inventory check

One is broken, two are ready to steer green and 3 available in un known condition

 

Attachment 1: IMG_3678.JPG
IMG_3678.JPG
Attachment 2: PIpztETMYgreen.jpg
PIpztETMYgreen.jpg
  12967   Wed May 3 16:47:45 2017 KojiUpdateGeneralPI pzt inventory check

I also have a functional one on my desk, which has one of the wires repaired.

Quote:

One is broken, two are ready to steer green and 3 available in un known condition

 

 

  12969   Wed May 3 18:45:45 2017 ranaUpdateGeneralDRMI locking

Quote:
Comparing counts doesn't get you anywhere; each PD has different whitening gain which may vary from measurement to measurement. The better thing to compare is Volts coming out of the demod board, since this (hopefully) only changes when we touch the PD or analog signal chain; this is what I used for the most recent DRMI sensing measurements. (ELOG 11589) We have calibrated actuator channels in the CAL model, which will give you the control signal in m for the DRMI lengths. Perhaps you can convert your sensing matrix measurement to demod board output volts per meter to compare with the last measurement.

Also, the monitor ports are the LEMO ports to the left; the SMA ports where the signal is coming from are from a daughter board that has a better output opamp that the nominal output; we're using the same output on the REFL11 and AS55 demod boards.


Wrong! RTFS.

SMA outputs are the bare, passive outputs of the mixer/lowpass.
TNC outputs are the low-noise, acti amplified outputs via the daughter board.
LEMO outputs are the high noise, G=2, LT1125 buffered outputs
  12972   Thu May 4 19:03:15 2017 gautamUpdateGeneralDRMI locking - preliminary MICH NB

Summary:

I've been playing around with Evan's NB code trying to put together a noise budget for the data collected during the DRMI locks last week. Here is what I have so far.

Attachment #1: Sensing matrix measurement.

  • This is basically to show that the MICH error signal is mostly in AS55Q.
  • The whitening gain used was 0dB, and the demod phase was -82 degrees.
  • The MICH sensing response was 5.31*10^8 V/m, where V is the demod board output. The 40m wiki RFPD page for AS55 says the RF transimpedance is ~550ohms, and I measured the Demod Board puts out 5.1V of IF signal (measured at after the Preamp, which is what goes to the ADC) for 1V of RF signal at the PD input. Using these numbers, and assuming a PD responsivity of 0.8 A/W at 1064nm, the sensing response is 2.37*10^5 W/m. I don't have a feeling yet for whether this is a reasonable number, but it would be a number to compare to what my Finesse model tells me to expect, for example.
  • Actuator calibration used to arrive at these numbers was taken from this elog

Attachment #2: MICH OLTF measurement vs model

  • In order to build the MICH OLTF model, I used MATLAB to put together the following transfer functions:
    • BS pendulum
    • Digital servo filters from LSC_MICH
    • Violin mode filters 
    • Analog/Digital AA and AI filters. For the digital AA/AI filters, I took the coefficients from /opt/rtcds/rtscore/release/src/fe/controller.c
  • The loop measurement was taken with digital filter modules FM1, FM2, FM3, FM7, FM9 engaged. 
  • In order to fit the model to the measurement, I tried finding the best-fit values for an overall loop gain and delay. 
  • The agreement between model and measurement isn't stellar, but I decided to push ahead for a first attempt. This loop TF was used to convert various noises into displacement noise for plotting.

Attachment #3: Noise budget

  • It took me a while to get Evan's code going, the main changes I made were to use nds2 to grab data instead of GWPy, and also to replace reading in .txt files with importing .mat files. This is a work in progress.
  • Noises plotted:
    • Measured - I took the in loop error signal and estimated the free-running displacement noise with the model OLTF, and calibrated it into metres using the sensing response measurement. This looks consistent with what was measured back in Dec 2015.
    • Shot noise - I used the measured DC power incident on the PD, 13mW, RF transimpedance of 550 V/A, and the V/m calibration factor mentioned above, to calculate this (labelled "Quantum Noise").
    • Dark noise - measured with PSL shutter closed.
    • Seismic noise, thermal noise, gas noise - calculated with GWINC

I think I did the various conversions/calibrations/loop algebra correctly, but I may have overlooked something. Now that the framework for doing this is somewhat set up, I will try and put together analogous NBs for PRCL and SRCL. 

GV 22 August 2017: Attachment #4 is the summary of my demod board efficiency investigations, useful for converting sensing measurement numbers from cts/m to W/m.

Attachment 1: DRMI_noArms_April30.pdf
DRMI_noArms_April30.pdf
Attachment 2: MICH_OLTF.pdf
MICH_OLTF.pdf
Attachment 3: C1NB_disp_40m_MICH_NB_30_April_2017.pdf
C1NB_disp_40m_MICH_NB_30_April_2017.pdf
Attachment 4: 40m_REFL_RFPDs_efficiency.pdf
40m_REFL_RFPDs_efficiency.pdf
  12974   Fri May 5 10:13:02 2017 ericqUpdateGeneralMICH NB questions
Is suspension thermal noise missing? I take it "Thermal" refers just to thermal things going on in the optic, since I don't see any peaks at the bounce/roll modes as I would expect from suspension thermal noise.

What goes into the GWINC calculation of seismic noise? Does it include real 40m ground motion data and our seismic stacks?

I'm surprised to see such a sharp corner in the "Dark Noise" trace, did you apply the OLG correction to a measured dark noise ASD? (The OLG correction only needs to be applied to the in-lock error signals to recover open loop behavior, there is no closed loop when you're measuring the dark noise so nothing to correct for.)
  12975   Fri May 5 12:10:53 2017 gautamUpdateGeneralMICH NB questions

Quote:
Is suspension thermal noise missing? I take it &quot;Thermal&quot; refers just to thermal things going on in the optic, since I don&#39;t see any peaks at the bounce/roll modes as I would expect from suspension thermal noise. What goes into the GWINC calculation of seismic noise? Does it include real 40m ground motion data and our seismic stacks? I&#39;m surprised to see such a sharp corner in the &quot;Dark Noise&quot; trace, did you apply the OLG correction to a measured dark noise ASD? (The OLG correction only needs to be applied to the in-lock error signals to recover open loop behavior, there is no closed loop when you&#39;re measuring the dark noise so nothing to correct for.)


I've included the suspension thermal noise in the "Thermal" trace, but I guess the GWINC file I've been using to generate this trace only computes the thermal noise for the displacement DoF. I think this paper has the formulas to account for them, I will look into including these.

For the seismic noise, I've just been using the seis40.mat file from the 40m SVN. I think it includes a model of our stacks, but I did not re-calculate anything with current seismometer spectra. In the NB I updated yesterday, however, I think I was off by a factor of sqrt(3) as I had only included the seismic noise from 1 suspended optic. I've corrected this in the attached plot.

For the dark noise, you are right, I had it grouped in the wrong dictionary in the code so it was applying the OLG inversion. I've fixed this in the attached plot.
Attachment 1: C1NB_disp_40m_MICH_NB_30_April_2017.pdf
C1NB_disp_40m_MICH_NB_30_April_2017.pdf
  12976   Sat May 6 21:52:11 2017 ranaUpdateGeneralMICH NB questions

I think the most important next two items to budget are the optical lever noise, and the coil driver noise. The coil driver noise is dominated at the moment by the DAC noise since we're operating with the dewhitening filters turned off.

  12979   Wed May 10 01:56:06 2017 gautamUpdateGeneralMICH NB - OL coupling

Last night, I tried to estimate the contribution of OL feedback signal to the MICH length error signal.

In order to do so, I took a swept sine measurement with a few points between 50 Hz and 500 Hz. The transfer function between C1:LSC-MICH_OUT_DQ and the Oplev Servo Output point (e.g. C1:SUS-BS_OL_PIT_OUT etc) was measured. I played around with the excitation amplitude till I got coherence > 0.9 for the TF measurement, while making sure I wasn't driving the Oplev error point too hard that side-lobes began to show up in the MICH control signal spectrum.

The Oplev control signal is not DQ-ed. So I locked the DRMI again and downloaded the 16k data "live" for ~5min stretch using cdsutils.getdata on the workstation. The Oplev error point is DQ-ed at 2k, but I found that the excitation amplitude needed for good SNR at the error point drove the servo to the limiter value of 2000cts - so I decided to use the control signal instead. Knowing the transfer function from the Oplev *_OUT* channel to C1:LSC-MICH_IN1_DQ, I backed out the coupling - the transfer function was only measured between 50 Hz and 500 Hz, and no extrapolation is done, so the estimation is only really valid in this range, which looks like where it is important anyways (see Attachment #2, contributions from ITMX, ITMY and BS PIT and YAW servos added in quadrature).

I was also looking at the Oplev servo shapes and noticed that they are different for the ITMs and the BS (Attachment #1). Specifically, for the ITM Oplevs, an "ELP15" is used to do the roll-off while an "ELP35" is employed in the BS servo (though an ELP35 also exists in the ITM Oplev filter banks). I got lost in an elog search for when these were tuned, but I guess the principles outlined in this elog still hold and can serve as a guideline for Oplev loop tweaking.

Coil driver noise estimation to follow

Quote:

I think the most important next two items to budget are the optical lever noise, and the coil driver noise. The coil driver noise is dominated at the moment by the DAC noise since we're operating with the dewhitening filters turned off.

GV 10 May 12:30pm: I've uploaded another copy of the NB (Attachment #3) with the contributions from the ITMs and BS separated. Looks like below 100Hz, the BS coupling dominates, while the hump/plateau around 350Hz is coming from ITMX.

Attachment 1: OL_BS_ITM_comp.pdf
OL_BS_ITM_comp.pdf
Attachment 2: C1NB_disp_40m_MICH_NB_8_May_2017.pdf
C1NB_disp_40m_MICH_NB_8_May_2017.pdf
Attachment 3: C1NB_disp_40m_MICH_NB_10_May_2017.pdf
C1NB_disp_40m_MICH_NB_10_May_2017.pdf
  12981   Wed May 10 16:53:38 2017 ranaUpdateGeneralMICH NB - OL coupling

That's a good find.

  1. The OL control signal can be gotten from the DQ error signal. You just need to multiply it by the digital filters and the gain. The state of the filters and the gain can be gotten using matlab tools like getFotonFilt.m. For python ChrisW wrote a tool called foton.py which is in the GDS SVN. You should ask him for it. It requires access to some ROOT libraries to run.
  2. We should have sub budgets for everything like OL and thermal, etc. They should be automatically produced each time you run the main budget and should be separate pages in the same PDF file. Jamie / Chris may have something going along these lines so check to see if they are already on it.
  12983   Wed May 10 17:17:05 2017 gautamUpdateGeneralDAC / Coil Driver noise

Suspension Actuator noise:

There are 3 main sources of electronics noise which come in through the coil driver:

  1. Voltage noise of the coil driver.
    1. The input referred noise is ~5 nV/rHz, so not a big issue.
    2. The Johnson noise of the output resistor which is in series with the coil is sqrt(4*k*T*R) ~ 3 nV/rHz. We probably want to increase this resistor from 200 to 1000 Ohms once Gautam convinces us that we don't need that range for lock acquisition.
  2. Voltage noise of the dewhitening board.
    1. In order to reduce DAC noise, we have a "dewhitening" filter which provides some low passing. There is an "antiDW" filter in the digital part which is the inverse of this, so that when they are both turned on, the result is that the main signal path has a flat transfer function, but the DAC noise gets attenuated.
    2. In particular, ours have 2 second order filters (each with 2 poles at 15 Hz and 2 zeros at 100 Hz).
    3. We also have a passive pole:zero network at the output which has z=130, 530 Hz and p = 14, 3185 Hz.
    4. The dewhitening board has an overall gain of 3 at DC to account for our old DACs having a range of +/-5 V and our coil drivers having +/- 15 V power supplies. We should get rid of this gain of 3.
    5. The dewhitening board (and probably the coil driver) use thick film resistors and so their noise is much worse than expected at low frequencies.
  3. DAC voltage noise. 
    1. The General Standards 16-bit DACs have a noise of ~5 uV/rHz.
  4. the satellite box is passive and not a significant source of noise; its just a flaky construction and so its problematic.
Attachment 1: actuation.jpg
actuation.jpg
  12984   Wed May 10 17:46:44 2017 gautamUpdateGeneralDAC / Coil Driver noise - SRM coil driver + dewhite board removed

I've removed the SOS coil driver (D010001-B, S/N B151, labelled "SRM") + Universal Dewhitening Board (D000183 Rev C, S/N B5172, labelled "B5") combo for SRM from 1X4, for photo taking + inspection.

I first shutdown the SRM watchdog, noted cabling between these boards and also the AI board as well as output to Sat. Box. I also needed to shutdown the MC2 watchdog as I had to remove the DAC output to MC2 in order to remove the SRM Dewhitening board from the rack. This connection has been restored, MC locks fine now.

 

  12985   Thu May 11 09:45:46 2017 ranaUpdateGeneralDAC / Coil Driver noise - SRM coil driver + dewhite board removed

I believe the ETMs and ITMs are different from the others.

  12986   Thu May 11 18:59:22 2017 gautamUpdateGeneralSRM coil driver + dewhite board initial survey

I've added marked-up schematics + high-res photographs of the SRM coil driver board and dewhitening board to the 40m DCC Document tree (D1700217 and D1700218). 

In the attached marked-up schematics, I've also added the proposed changes which Rana and I discussed earlier today. For the thick-film -> thin-film resistor switching, I will try and make a quick LISO model to see if we can get away with replacing just a few rather than re-stuff the whole board.

Since I have the board out, should I implement some of these changes (like AD797 removal) before sticking it back in and pulling out one of the ITM boards? I need to look at the locking transients and current digital limit-values for the various DoFs before deciding on what is an appropriate value for the output resistance in series with the coil.

Another change I think should be made, but I forgot to include on the markups: On the dewhitening board, we should probably replace the decoupling capacitors C41 and C52 with equivalent value electrolytic caps (they are currently tantalum caps which I think are susceptible to fail by shorting input to output).

Attachment 1: D010001-B_40m.pdf
D010001-B_40m.pdf D010001-B_40m.pdf D010001-B_40m.pdf D010001-B_40m.pdf
Attachment 2: D000183-C8_40m.pdf
D000183-C8_40m.pdf D000183-C8_40m.pdf D000183-C8_40m.pdf
  12987   Fri May 12 01:36:04 2017 gautamUpdateGeneralSRM coil driver + dewhite board LISO modeling

I've made the LISO models for the dewhitening board and coil driver boards I pulled out.

Attached is a plot of the current noise in the current configuration (i.e. dewhitening board just has a gain x3 stage, and then propagated through the coil driver path), with the top 3 noise contributions: The op-amps (op3 and op5) are the LT1125s on the coil driver board in the bias path, while "R12" is the Johnson noise from the 1k input resistace to the OP27 in the signal path.

Assuming the OSEMs have an actuation gain of 0.016 N/A (so 0.064 N/A for 4 OSEMs), the current noise of ~1e-10 A/rtHz translates to a displacement noise of ~3e-15m/rtHz at ~100Hz (assuming a mirror mass of 0.25kg). 

I have NOT included the noise from the LM6321 current buffers as I couldn't find anything about their noise characteristics in the datasheet. LISO files used to generate this plot are attached.

Quote:

I've added marked-up schematics + high-res photographs of the SRM coil driver board and dewhitening board to the 40m DCC Document tree (D1700217 and D1700218). 

In the attached marked-up schematics, I've also added the proposed changes which Rana and I discussed earlier today. For the thick-film -> thin-film resistor switching, I will try and make a quick LISO model to see if we can get away with replacing just a few rather than re-stuff the whole board.

Since I have the board out, should I implement some of these changes (like AD797 removal) before sticking it back in and pulling out one of the ITM boards? I need to look at the locking transients and current digital limit-values for the various DoFs before deciding on what is an appropriate value for the output resistance in series with the coil.

Another change I think should be made, but I forgot to include on the markups: On the dewhitening board, we should probably replace the decoupling capacitors C41 and C52 with equivalent value electrolytic caps (they are currently tantalum caps which I think are susceptible to fail by shorting input to output).

 

Attachment 1: SRM_bypass_plus_CoilDriver.pdf
SRM_bypass_plus_CoilDriver.pdf
Attachment 2: liso.zip
  12988   Fri May 12 12:34:55 2017 gautamUpdateGeneralITM and BS coil driver + dewhite board pulled out

I first set the bias sliders to 0 on the MEDM screen (after checking that the nominal values were stored), then shut down the watchdogs, and then pulled out the boards for inspection + photo-taking.

  12990   Fri May 12 18:50:08 2017 gautamUpdateGeneralITM and BS coil driver + dewhite board pulled out

I've uploaded high-res photos + marked up schematics to the same DCC page linked in the previous page. I've noted the S/Ns of the ITM, BS and SRM boards on the page, I think it makes sense to collect everything on one page, and I guess eventually we will unify everything to a one or two versions.

To take the photos, I tried to reproduce the "LED light painting" technique reported here. I mounted the Canon EOS Rebel T3i on a tripod, and used some A3 sheets of paper to make a white background against which the board to be photographed was placed. I also used the new Macro lens we recently got. I then played around with the aperture and exposure time till I got what I judged to be good photos. The room lights were turned off, and I used the LED on my phone to do the "painting", from ~a metre away. I think the photos have turned out pretty well, the component values are readable.

Quote:

I first set the bias sliders to 0 on the MEDM screen (after checking that the nominal values were stored), then shut down the watchdogs, and then pulled out the boards for inspection + photo-taking.

 

  12999   Fri May 19 19:18:53 2017 KaustubhSummaryGeneralTesting of the new Photo Detectors ET-3010 and ET-3040

Motivation:

I got some hands-on-experience on using RF photodetectors and the Network Analyzer from Koji. There were newly purchased RF photodetectors from Electro-Optics Technology, Inc.. These were InGaAs Photodetectors with model no.: 120-10050-0001(ET-3010) and 120-10056-0001(ET-3040). The User Guide for the two detectors can be found here. This is the first time we bought the ET-3010 model PD for the 40m lab. It has an operation bandwith >1.5GHz(not tested yet), much higher than other PDs of its kind. This can be used for detecting the output as we 'sweep' the laser frequency for getting data on the optical cavities and the resonating modes inside the cavity. We just tested out the ET-3040 model today but will test out the ET-3010 next week.

Tools and Machines Used:

We worked on the optical bench right in front of the main entrance to the lab. We put the cables, power chords, etc. to their respective places. We used screws, poles, T's, I's, multimeter, Network/Spectrum Analyzer(along with the moving table), a lab computer, Oscilloscope, power supply and the aforementioned PDs for our testing. We took these items from the stack of tools at the Y-arm and the boxes of various different labelled palced near the X-arm. We moved the Network Analyzer(along with the bench) from near the Y-arm to our workplace.

Procedure:

I will include a rough schematic of the setup later.

We alligned the reference PD(High Speed Photoreceiver model 1611) and the test PD(ET-3040 in this case) to get optimal power output. We had set the pump current for the laser at 19.5mA which produced a power of 1.00mW at the output of the fiber couple. At the reference detector the measured voltage was about 1.8V and at the DUT it was about 15mV. The DC transimpedance for the reference detector is 10kOhm and its responsivity to 1064 nm is around 0.75A/W. Using this we calculate the power at the reference detector to be 0.24mW. The DC transimpedance for the DUT is 50Ohm and the responsivity of about 0.9A/W. This amounts to a power of about 0.33mW. After measuring the DC voltages, we connected the laser input to the Network Analyzer and gave in an RF signal with -10dBm and frequency modulation from 100 kHz to 500 MHz. The RF output from the Analyzer is coupled to the Reference Channel(CHR) of the analyzer via a 20dB directional coupler. The AC output of the reference detector is given at Channel A(CHA) and the output from the DUT is given to Channel B(CHB). We got plots of the ratios between the reference detector, DUT and the coupled refernce for the Transfer Function and the Phase. We found that the cut-off frequency for the ET3040 model was at arounf 55 MHz(stated as >50MHz in the data sheet). We have stored the data using the lab PC in the directory .../scripts/general/netgpibdata/data.

Result:

The bandwidth of the ET-3040 PD is as stated in the data sheet, >50 MHz.

Precaution:

These PDs have an internal power supply of 3V for ET-3040 and 6V for ET-3010. Do not leave these connected to any instruments after the experiments have been performed or else the batteries will get drained if there is any photocurrent on the PDs.

To Do:

A similar procedure has to be followed in order to test the ET-3010 PD. I will be doing this tentatively on Monday.

Attachment 1: IMG_20170519_173247922.jpg
IMG_20170519_173247922.jpg
Attachment 2: IMG_20170519_173253252.jpg
IMG_20170519_173253252.jpg
Attachment 3: IMG_20170519_173300174.jpg
IMG_20170519_173300174.jpg
Attachment 4: PD_test_setup.png
PD_test_setup.png
  13003   Mon May 22 13:37:01 2017 gautamUpdateGeneralDAC noise estimate

Summary:

I've spent the last week investigating various parts of the DAC -> OSEM coil signal chain in order to add these noises to the MICH NB. Here is what I have thus far.

Current situation:

  • Coils are operated with no DAC whitening
  • So we expect the DAC noise will dominate any contribution from the electronics noise of the analog De-Whitening and Coil Driver boards
  • There is a factor of 3 gain in the analog De-Whitening board

DAC noise measurement:

  • I essentially followed the prescription in G1401335 and G1401399
  • So far, I only measured one DAC channel (ITMX UL)
  • The noise shaping filter in the above documents was adapted for this measurement. The noise used was uniform between DC and 1kHz for this test.
  • For the >50Hz bandstops, I used 1 complex pole pair at 5Hz, and 1 compelx zero pair at 50Hz to level off the noise.
  • For <50Hz bandstops, I used 1 compelx pole pair at 1Hz and 1 complex zero pair at 5Hz to push the RMS to lower frequencies
  • I set the amplitude ("gain" = 10,000 in awggui) to roughly match the Vpp when the ITM local damping loops are on - this is ~300mVpp (measured with a scope). 
  • The elliptic bandstops were 6th order, with 50dB stopband attenuation.
  • The SR785 input auto-ranging was disabled to allow a fair comparison of the various bandstops - this was fixed to -20 dBVpk for all measurements, and the SR785 noise floor shown is also for this value of the input range. Input was also AC coupled, and since I was using the front-panel LEMO for this test, the signal was effectively single-ended (but the ground of the SR785 was set to "floating" in order to get the differential signal from the DAC) 
  • Attachment #1 shows the results of this measurement - I've subtracted the SR785 noise from the other curves. The noise model was motivated by G1401399, but I use an f^-1/2 model rather than an f^-1 model. It seems to fit the measurement alright (though the "fit" is just done by eye and not by systematic optimization of the parameters of the model function).

Noise budget:

  • I then tried to translate this result into the noise budget
  • The noises for the 4 face coils are added in quadrature, and then the contribution from 3 optics (2 ITMs and BS) are added in quadrature
  • To calibrate into metres, I converted the DAC noise spectral density into cts/rtHz, and used the numbers from this elog. I thought I had missed out on the factor of 3 gain in the de-white board, but the cts-to-meters number from the referenced elog already takes into account this factor.
  • Just to be clear, the black line for DAC noise in Attachment #2 is computed from the single-channel measurement of Attachment #1 according to the following relation: \script{n}_{\mathrm{DAC}} ~ (m/\sqrt{Hz}) = n_{1-ch} (V/\sqrt{Hz}) \times (2^{15}/20) (cts/V) \times G_{act} \times 2 \times \sqrt{6}, where G_act is the coil transfer function from the referenced elog, taken as 5nm/f^2 on average for the 2 ITMs and BS, the factor of 2 comes from adding the noise from 4 coils in quadrature, and the factor of sqrt(6) comes from adding the noise from 3 optics in quadrature (and since the BS has 4 times the noise of the ITMs)
  • Using the 0.016N/A number for each coil gave me an answer than was off by more than an order of magnitude - I am not sure what to make of this. But since the other curves in the NB are made using numbers from the referenced elog, I think the answer I get isn't too crazy...
  • Attachment #2 shows the noise budget in its current form, with DAC noise added. Except for the 30-70Hz region, it looks like the measured noise is accounted for.

Comments:

  • I have made a number of assumptions:
    • All DAC channels have similar noise levels
    • Tried to account for asymmetry between BS and ITMs (BS has 100 ohm resistance in series with the coil driver while the ITMs have 400 ohms) but the individual noises haven't been measured yet
    • This noise estimate holds for the BS, which is the MICH actuator (I didn't attempt to simulate the in-lock MICH control signal and then measure the DAC noise)
  • But this seems sensible as a first estimate
  • The dmesg logs for C1SUS don't tell me what DACs we are using, but I believe they are 16-bit DACs (I'll have to restart the machine to make sure)
  • In the NB, the flattening out of some curves beyond 1kHz is just an artefact of the fact that I don't have data to interpolate in that region, and isn't physical.
  • I had a brief chat with ChrisW who told me that the modified EEPROM/Auto-Cal procedure was only required for 18-bit DACs. So if it is true that our DACs are 16-bit, then he advised that apart from the DAC noise measurement above, the next most important thing to be characterized is the quantization noise (by subtracting the calculated digital control signal from the actual analog signal sent to the coils in lock)
  • More details of my coil driver electronics investigations to follow...
Attachment 1: DAC_noise_model.pdf
DAC_noise_model.pdf
Attachment 2: C1NB_disp_40m_MICH_NB_22_May_2017.pdf
C1NB_disp_40m_MICH_NB_22_May_2017.pdf
  13005   Mon May 22 18:20:27 2017 KaustubhSummaryGeneralTesting of the new Photo Detectors ET-3010 and ET-3040

I am adding the text files with the data readings and paramater settings along with the Bode Plot of the data. I plotted these graphs using matplotlib module with python 2.7.

Quote:

Motivation:

I got some hands-on-experience on using RF photodetectors and the Network Analyzer from Koji. There were newly purchased RF photodetectors from Electro-Optics Technology, Inc.. These were InGaAs Photodetectors with model no.: 120-10050-0001(ET-3010) and 120-10056-0001(ET-3040). The User Guide for the two detectors can be found here. This is the first time we bought the ET-3010 model PD for the 40m lab. It has an operation bandwith >1.5GHz(not tested yet), much higher than other PDs of its kind. This can be used for detecting the output as we 'sweep' the laser frequency for getting data on the optical cavities and the resonating modes inside the cavity. We just tested out the ET-3040 model today but will test out the ET-3010 next week...

 

Attachment 1: ET-3040_test.zip
Attachment 2: ET-3040_test.pdf
ET-3040_test.pdf
  13009   Tue May 23 18:09:18 2017 KaustubhConfigurationGeneralTesting ET-3010 PD

In continuation with the previous(ET-3040 PD) test.

The ET-3010 PD requires to be fiber coupled for optimal use. I will try to test this model without the fiber couple tomorrow and see whether it works or not.

  13010   Tue May 23 22:58:23 2017 gautamUpdateGeneralDe-Whitening board noises

Summary:

I wanted to match a noise model to noise measurement for the coil-driver de-whitening boards. The main objectives were:

  1. Make sure the various poles/zeros of the Bi-Quad stages and the output stage were as expected from the schematics
  2. Figure out which components are dominating the noise contribution, so that these can be prioritized while swapping out the existing thick-film resistors on the board for lower noise thin-film ones
  3. Compare the noise performance of the existing configuration, which uses an LT1128 op-amp (max output current ~20mA) to drive the input of the coil-driver board, with that when we use a TLE2027 (max output current ~50mA) instead. This last change is motivated by the fact that an earlier noise-simulation suggested that the Johnson noise of the 1kohm input resistor on the coil driver board was one of the major noise contributors in the de-whitening board + coil driver board signal chain. Since the TLE2027 can drive an output current of up to 300mA, we could reduce the input impedance of the coil-driver board to mitigate this noise source to some extent. 

Measurement:

  • The back-plane pin controlling the MAX333A that determines whether de-whitening is engaged or not (P1A) was pulled to ground (by means of one of the new extender boards given to us by Ben Abbott). So two de-whitening stages were engaged for subsequent tests.
  • I first measured the transfer function of the signal path with whitening engaged, and then fit my LISO model to the measurement to tweak the values of the various components. This fitted file is what I used for subsequent noise analysis. 
  • ​For the noise measurement, I shorted the input of the de-whitening board (10-pin IDE connector) directly to ground.
  • I then measured the voltage noise at the front-panel SMA connector with the SR785
  • The measurements were only done for 1 channel (CH1, which is the UL coil) for 4 de-whitening boards (2 ITMs, BS, and SRM). The 2 ITM boards are basically identical, and the BS and SRM boards are similar. Here, only results for the board labelled "ITMX" are presented.
  • For this board, I also measured the output voltage noise when the LT1128 was replaced with a TLE2027 (SOIC package, soldered onto a SOIC-to-DIP adaptor). Steve has found (ordered?) some DIP variants of this IC, so we can compare its noise performance when we get it.

Results:

  • Attachment #1 shows the modeled and measured noises, which are in fairly good agreement.
  • The transfer function measurement/fitting (not attached) also suggests that the poles/zeros in the signal path are where we expect as per the schematic. I had already verified the various resistances, but now we can be confident that the capacitance values on the schematic are also correct. 
  • The LT1128 and TLE2027 show pretty much identical noise performance.
  • The SR785 noise floor was low enough to allow this measurement without any pre-amp in between. 
  • I have identified 3 resistors from the LISO model that dominate the noise (all 3 are in the Bi-Quad stages), which should be the first to be replaced. 
  • There are some pretty large 60 Hz harmonics visible. I thought I was careful enough avoiding any ground loops in the measurement, and I have gotten some more tips from Koji about how to better set up the measurement. This was a real problem when trying to characterize the Coil Driver noise.

Next steps:

  • I have data from the other 3 boards I pulled out, to be updated shortly.
  • The last piece (?) in this puzzle is the coil driver noise - this needs to be modeled and measured.
  • Once the coil driver board has been characterized, we need to decide what changes to make to these boards. Some things that come to mind at the moment:
    • Replace critical resistors (from noise-performance point of view) with low noise thin film ones.
    • Remove the "fast analog" path on the coil driver boards - these have potentiometers in series with the coil, which we should remove since we are not using this path anyways.
    • Remove all AD797s from both de-whitening and coil driver boards - these are mostly employed as monitor points that go to the backplane connector, which we don't use, and so can be removed.
    • Increase the series resistor at the output of the coil driver (currently, these are either 100ohm or 400ohm depending on the optic/channel). I need to double check the limits on the various LSC servos to make sure we can live with the reduced range we will have if we up these resistances to 1 kohm (which serves to reduce the current noise to the coils, which is ultimately what matters).
Attachment 1: ITMX_deWhite_ch1_noise.pdf
ITMX_deWhite_ch1_noise.pdf
  13011   Wed May 24 18:19:15 2017 KaustubhUpdateGeneralET-3010 PD Test

Summary:

In continuation to the previous test conducted on the ET-3040 PD,  I performed a similar test on the ET-3010 model. This model requires a fiber couple input for proper testing, but I tested it in free space without a fiber couple as the laser power was only 1.00 mW and there was not much danger of scattering of the laser beam. The Data Sheet can be found here

Procedure:

The schematic(attached below) and the procedure are the same as the previous time. The pump current was set to 19.5 mA giving us a laser beam of power 1.00mW at the fiber couple output. The measured voltage for the reference detector was 1.8V. For the DUT, the voltage is amplified using a low noise amplifier(model SR-560) with a gain of 100. Without any laser incidence on the DUT, the multimeter reads 120.6 mV. After alligning the laser with the DUT, the multimeter reads 348.5 mV, i.e. the voltage for the DUT is 227.9/100 ~ 2.28 mV. The DC transimpedance of the reference detector is 10kOhm and its responsivity to 1064 nm is around 0.75 A/W. Using this we calculate the power at the reference detector to be 0.24 mW. The DC transimpedance for the DUT is 50Ohm and the responsivity is around 0.85 A/W. Using this we calculate the power at the DUT to be 0.054 mW. After this we connect the the laser input to the Netwrok Analyzer(AG4395A) and give an RF signal with -10dBm and frequency modulation from 100 kHz to 500 MHz.The RF output from the Analyzer is coupled to the Reference Channel(CHR) of the analyzer via a 20dB directional coupler. The AC output of the reference detector is given at Channel A(CHA) and the output from the DUT is given to Channel B(CHB). We got plots of the ratios between the reference detector, DUT and the coupled refernce for the Transfer Function and the Phase. I stored the data under the directory.../scripts/general/netgpibdata/data. The Bode Plot has been attached below and seeing it we observe that the cut-off frequency for the ET-3010 model is atleast over 500 MHz(stated as >1.5 GHz in the data sheet).

Result:

The bandwidth of the ET-3010 PD is atleast 500MHz, stated in the data sheet as >1.5GHz.

Precaution:

The ET-3010 PD has an internal power supply of 6V. Don't leave the PD connected to any instrument after the experimentation is done or else the batteries will get drained if there is any photocurrent on the PDs.

To Do:

Caliberate the vertical axis in the Bode Plot with transimpedance(Ohms) for the two PDs. Automate the procedure by making a Python script for taking multiple set of readings from the Netwrok Analyzer and aslo plot the error bands.

Attachment 1: PD_test_setup.png
PD_test_setup.png
Attachment 2: ET-3010_test.pdf
ET-3010_test.pdf
Attachment 3: ET-3010_test.zip
  13015   Thu May 25 19:27:29 2017 gautamUpdateGeneralCoil driver board noises

[Koji, Gautam]

Summary: 

  • Attachment #1 shows the measured/modeled noise of the coil driver board (labelled ITMX). 
  • Measurement was made with "TEST" input (which is what the DAC drives) is connected to ground via 50ohm terminator, and "BIAS" input grounded.
  • The model tells us to expect a noise of the order of 5nV/rtHz - this is comparable to (or below) the input noise of the SR785, and even the SR560. So this measurement only serves to place an upper bound on the coil driver board noise.
  • There is some excess noise below 40Hz, would be interesting to see if this disappears with swapping out thick-film resistors for thin film ones.
  • The LISO model says that the dominant contribution is from the voltage and input current noise of the two op-amps (LT1125) in the bias LP filter path. 
  • But if we can indeed realize this noise level of ~10-20nV/rtHz, we are already at the ~10^-17m/rtHz displacement noise for MICH at about 200Hz. I suspect there are other noises that will prevent us from realizing this performance in displacement noise.

Details:

This measurement has been troublesome - I was plagued by large 60Hz harmonics (see Attachment #1), the cause of which was unknown. I powered all electronics used in the measurement set up from the same power strip (one of the new surge-protecting ones Steve recently acquired for us), but these remained present. Yesterday, Koji helped me troubleshoot this issue. We did the various things, I try to put them here in the order we did them:

  1. Double check that all electronics were indeed being powered from the same power strip - OK, but harmonics remained present.
  2. Tried using a different DC power supply - no effect.
  3. Checked the signal with an oscilloscope - got no additional insight.
  4. I was using a DB25 breakout board + pomona minigrabbers to measure the output signal and pipe it to the SR785. Koji suggested using twisted ribbon wire + soldered BNC connector (recycled from some used ones lying around the lab). The idea was to minimize stray radiation pickup. We also disconnected the WiFi extender and GPIB box from the analyzer and also disconnected these from the power - this finaly had the desired effect, the large harmonics vanished. 

Today, I tried to repeat the measurement, with the newly made twisted ribbon cable, but the large 60Hz harmonics were back. Then I realized we had also disconnected the WiFi extender and GPIB box yesterday.

Turns out that connecting the Prologix box to the SR785 (even with no power) is the culprit! Disconnecting the Prologix box makes these harmonics go away. I was using the box labelled "Santuzza.martian" (192.168.113.109), but I double-checked with the box labelled "vanna.martian" (192.168.113.105, also a different DC power supply adapter for the box), the effect is the same. I checked various combinations like 

  • GPIB box connected but not powered
  • GPIB box connected with no network cable

but it looks like connecting the GPIB box to the analyzer is what causes the problem. This was reproducible on both SR785s in the lab. So to make this measurement, I had to do things the painful way - acquire the spectrum by manually pushing buttons with the GPIB box disconnected, then re-connect the box and download the data using SRmeasure --getdata. I don't fully understand what is going on, especially since if the input connector is directly terminated using a 50ohm BNC terminator, there are no harmonics, regardless of whether the GPIB box is connected or not. But it is worth keeping this problem in mind for future low-noise measurements. My elog searches did not reveal past reports of similar problems, has anyone seen something like this before?

It also looks like my previous measurement of the de-whitening board noises was plagued by the same problem (I took all those spectra with the GPIB boxes connected). I will repeat this measurement.

Next steps:

At the meeting this week, it was decided that

  • All AD797s would be removed from de-whitening boards and also coil-driver boards (as they are unused).
  • Thick film resistors with the most dominant noise contributions to be replaced with thin-film ones.
  • Gain of 3 on de-whitening board to be changed to gain of 1.

I also think it would be a good idea to up the 100-ohm resistors in the bias path on the ITM coil driver boards to 1kohm wire-wound. Since the dominant noise on the coil-driver boards is from the voltage noise of the Op-Amps in the bias path, this would definitely be an improvement. Looking at the current values of the bias MEDM sliders, a 10x increase in the resistance for ITMX will not be possible (the yaw bias is ~-1.5V), but perhaps we can go for a 4x increase?

The plan is to then re-install the boards, and see if we can 

  1. Turn on the whitening successfully (I checked with an extender board that the switching of the whitening stages works - turning OFF the "simDW" filter in the coil driver filter banks enables the analog de-whitening).
  2. Relize the promised improvement in MICH displacement noise with the existing whitening configuration.

We can then take a call on how much to up the series resistance in the DAC signal path. 

Now that I have figured out the cause of the harmonics, I will also try and measure the combined electronics noise of de-whitening board + coil driver board and compare it to the model.

Quote:
  • The last piece (?) in this puzzle is the coil driver noise - this needs to be modeled and measured.

 

Attachment 1: coilDriverNoises.pdf
coilDriverNoises.pdf
  13016   Sat May 27 10:26:28 2017 KaustubhUpdateGeneralTransimpedance Calibration

Using Alberto's paper LIGO-T10002-09-R titled "40m RF PDs Upgrade", I calibrated the vertical axis in the bode plots I had obtained for the two PDs ET-3010 and ET-3040.

I am not sure whether the values I have obtained are correct or not(i.e. whether the calibration is correct or not). Kindly review them.

EDIT: Attached the formula used to calculate transimpedance for each data point and the values of other paramaters.

EDIT 2: Updated the plots by changing the conversion for gettin ghte ratio of the transfer functions from 10^(y/10) to 10^(y/20).

Attachment 1: ET-3040_test_transimpedance.pdf
ET-3040_test_transimpedance.pdf
Attachment 2: ET-3010_test_transimpedance.pdf
ET-3010_test_transimpedance.pdf
Attachment 3: Formula_for_Transimpedance.pdf
Formula_for_Transimpedance.pdf
  13017   Mon May 29 16:47:38 2017 gautamUpdateGeneralCoil driver boards reinstalled

Yesterday, I reinstalled the de-whitening boards + coil driver boards into their respective Eurocrate slots, and reconnected the cabling. I then roughly re-aligned the ITMs using the green beams. 

I've given Steve a list of the thin-film resistors we need to implement the changes discussed in the preceeding elogs - but I figured it would be good to see if we can realize the projected improvement in MICH displacement noise just by fixing the BS Oplev loop shape and turning the existing whitening on. Before re-installing them however, I did make a few changes:

  • Removed the gain of x3 on all the signal paths on the De-Whitening boards, and made them gain x1. For the De-Whitened path, this was done by changing the feedback resistor in the final op-amp (OP27) from 7.5kohm to 2.49kOhm, while for the bypass path, the feedback resistor in the LT1125 stages were changed from 3.01kohm to 1kohm. 
  • To recap - this gain of x3 was originally implemented because the DACs were +/- 5V, while the coil driver electronics had supply voltage of +/- 15V. Now, our DACs are +/- 10V, and even though the supply voltage to the coil driver boards is +/- 15V, in reality, the op-amps saturate at around 12V, so we aren't really losing much in terms of range.
  • I also modified the de-whitening path in the BS de-whitening board to mimic the configuration on the ITM de-whitening boards. Mainly, this involved replacing the final stage AD797 with an OP27, and also implementing the passive pole-zero network at the output of the de-whitened path. I couldn't find capacitors similar to those used on the ITM de-whitening boards, so I used WIMA capacitors.
  • The SRM de-whitening path was not touched for now.
  • On all the boards, I replaced any AD797s that were being used with OP27s, and simply removed AD797s that were in DAQ paths.
  • I removed all the potentiometers on all the boards (FAST analog path on the coil driver boards, and some offset trim Pots on the BS and SRM de-whitening boards for the AD797s, which were also removed).
  • For one signal path on the coil driver board (ITMX ch1), I replaced all of the resistors with thin-film ones and re-measured the noise. However, the excess noise in the measurement below ~40Hz (relative to the model) remained.

Photos of all the boards were taken prior to re-installation, and have been uploaded to the 40m Google Photos page - I will update schematics + photos on the DCC page once other planned changes are implemented.

I also measured the transfer functions on the de-whitened signal paths on all the boards before re-installing them. I then fit everything using LISO, and updated the filter banks in Foton to match these measurements - the original filters were copied over from FM9 and FM10 to FM7 and FM8. The new filters are appended with the suffix "_0517", and live in FM9 and FM10 of the coil output filter banks. The measured TFs (for ITMs and BS) are summarized in Attachment #1, while Attachment #2 contains the data and LISO file used to do the fits (path to the .bod files in the .fil file will have to be changed appropriately). I used 2 complex pole pairs at ~10 Hz, two complex zero pairs at ~100Hz, real poles at ~15Hz and ~3kHz, and real zeros at ~100Hz and ~550Hz for the fits. The fits line up well with the measured data, and are close enough to the "expected" values (as calculated from component values) to be explained by tolerances on the installed components - I omit the plots here. 

After re-installing the boards in the Eurocrate, restoring rough alignment, and updating the filter banks with the most recent measured values, I wanted to see if I could turn the whitening on for one of the optics (ITMY) smoothly before trying to do so in the full DRMI - switching off the "SimDW_0517" filter (FM9) should switch the signal path on the de-whitening board from bypass to de-whitened, and I had confirmed last week with an extender board that the voltage at the appropriate backplane connector pin does change as expected when the FM9 MEDM button is toggled (for both ITMs, BS and SRM). But today I was not able to engage this transition smoothly, the optic seems to be getting kicked around when I engage the whitening. I will need to investigate this further. 


Unrelated to this work: the ETMY Oplev HeNe is dead (see Attachment #3). I thought we had just replaced this laser a couple of months ago - what is the expected lifetime of these? Perhaps the power supply at the Y-end is wonky and somehow damaging the HeNe heads?

Attachment 1: deWhitening_consolidated.pdf
deWhitening_consolidated.pdf deWhitening_consolidated.pdf deWhitening_consolidated.pdf
Attachment 2: deWhitening_measurements.zip
Attachment 3: ETMY_OL.png
ETMY_OL.png
  13019   Tue May 30 16:02:59 2017 gautamUpdateGeneralCoil driver boards reinstalled

I think the reason I am unable to engage the de-whitening is that the OL loop is injecting a ton of control noise - see Attachment #1. With the OL loop off (i.e. just local damping loops engaged for the ITMs), the RMS control signal at 100Hz is ~6 orders of magnitude (!) lower than with the OL loop on. So turning on the whitening was just railing the DAC I guess (since the whitening has something like 60dB gain at 100Hz).

The Oplev loops for the ITMs use an "Ellip15" low-pass filter to do the roll-off (2nd order Elliptic low pass filter with 15dB stopband atten and 2dB ripple). I confirmed that if I disable the OL loops, I was able to turn on the whitening for ITMY smoothly.

Now that the ETMY OL HeNe has been replaced, I restored alignment of the IFO. Both arms lock fine (I was also able to engage the ITMY Coil Driver whitening smoothly with the arm locked). However, something funny is going on with ASS - running the dither seems to inject huge offsets into the ITMY pit and yaw such that it almost immediately breaks the lock. This probably has to do with some EPICS values not being reset correctly since the recent slow-machine restarts (for instance, the c1iscaux restart caused all the LSC RFPD whitening gains to be reset to random values, I had to burt-restore the POX11 and POY11 values before I could get the arms to lock), I will have to investigate further.

GV edit 2pm 31 May: After talking to Koji at the meeting, I realized I did not specify what channel the attached spectra are for - it is  C1:SUS-ITMY_ULCOIL_OUT.

Quote:
 

But today I was not able to engage this transition smoothly, the optic seems to be getting kicked around when I engage the whitening. I will need to investigate this further. 


Unrelated to this work: the ETMY Oplev HeNe is dead (see Attachment #3). I thought we had just replaced this laser a couple of months ago - what is the expected lifetime of these? Perhaps the power supply at the Y-end is wonky and somehow damaging the HeNe heads?

 

Attachment 1: OL_noiseInjection.pdf
OL_noiseInjection.pdf
  13026   Thu Jun 1 00:10:15 2017 gautamUpdateGeneralCoil driver boards reinstalled

[Koji, Gautam]

We tried to debug the mysterious sudden failure of ASS - here is a summary of what we did tonight. These are just notes for now, so I don't forget tomorrow.

What are the problems/symptoms?

  • After re-installing the coil driver electronics, the ASS loops do not appear to converge - one or more loops seem to run away to the point we lose the lock.
  • For the Y-arm dithers, the previously nominal ITM PIT and YAW oscillator amplitudes (of ~1000cts each) now appears far too large (the fuzz on the Y arm transmission increases by x3 as viewed on StripTool).
  • The convergence problem exists for the X arm alignment servos too.

What are the (known) changes since the servos were last working?

  • Gain of x3 on the de-whitening boards for ITMX, ITMY, BS and SRM have been replaced with gain x1. But I had measurements for all transfer functions (De-White board input to De-White Board outputs) before and after this change, so I compensated by adding a filter of gain ~x3 to all the coil filter banks for these optics (the exact value was the ratio of the DC gain of the transfer functions before/after).
  • The ETMY Oplev has been replaced. I walked over to the endtable and there doesn't seem to be any obvious clipping of either the Oplev beam or the IR transmission.

Hypotheses plus checks (indented bullets) to test them:

  1. The actuation on the ITMs are ~x10 times stronger now (for reasons unknown).
    • I locked the Y-arm and drove a line in the channels C1:SUS-ETMY_LSC_EXC and C1:SUS-ITMY_LSC_EXC at ~100Hz and ~30Hz, (one optic at one frequency at a time), and looked at the response in the LSC control signal. The peaks at both frequencies for the ITMs and ETMs were within a factor of ~2. Seems reasonable.
    • We further checked by driving lines in C1:SUS-ETMY_ASCPIT_EXC and C1:SUS-ITMY_ASCPIT_EXC (and also the corresponding YAW channels), and looked at peak heights at the drive frequencies in the OL control signal spectra - the peak heights matched up well in both the ITM and ETM spectra (the drive was in the same number of counts).

      So it doesn't look like there is any strange actuation imbalance between the ITM and ETM as a result of the recent electronics work, which makes sense as the other control loops acting on the suspensions (local damping, Oplevs etc seem to work fine). 
  2. The way the dither servo is set up for the Y-arm, the tip-tilts are used to set the input axis to the cavity axis, while actuation to the ITM and ETM takes care of the spot centering. The problem lies with one of these subsystems.
    • We tried disabling the ASS servo inputs to all the spot-centering loops - but even with just actuation on the TTs, the arm transmission isn't maximized.
    • We tried the other combination - disable actuation path to TTs, leave those to ITM and ETM on - same result, but the divergence is much faster (lock lost within a couple of seconds, large offsets appear in the ETM_PIT_L / ETM_YAW_L error signals.
    • Tried turning on loops one at a time - but still the arm transmission isn't maximized.
  3. Something is funny with the IR transmon QPD / ETMY Oplev.
    • I quickly measured Oplev PIT and YAW OLTFs, they seem normal with upper UGFs around 5Hz and phase margins of ~30 degrees.
    • We had no success using either of the two available Transmon QPDs
    • Looking at the QPD quadrants, the alignment isn't stellar but we get roughly the same number of counts on all quadrants, and the spot isn't drastically misaligned in either PIT or YAW.

For whatever reasons, it appears that dithering the cavity mirrors at frequencies with amplitudes that worked ~3 weeks ago is no longer giving us the correct error signals for dither alignment. We are out of ideas for tonight, TBC tomorrow...

 

  13034   Fri Jun 2 12:32:16 2017 gautamUpdateGeneralPower glitch

Looks like there was a power glitch at around 10am today.

All frontends, FB, Megatron, Optimus were offline. Chiara reports an uptime of 666 days so looks like its UPS works fine. PSL was tripped, probably the end lasers too (yet to check). Slow machines seem alright (Responds to ping, and I can also telnet into them).

Since all the frontends have to be re-started manually, I am taking this opportunity to investigate some cds issues like the lack of a dmesg log file on some of the frontends. So the IFO will be offline for sometime.

  13035   Fri Jun 2 16:02:34 2017 gautamUpdateGeneralPower glitch

Today's recovery seems to be a lot more complicated than usual.

  • The vertex area of the lab is pretty warm - I think the ACs are not running. The wall switch-box (see Attachment #1) shows some red lights which I'm pretty sure are usually green. I pressed the push-buttons above the red light, hopefully this fixed the AC and the lab cools down soon.
  • Related to the above - C1IOO has a bunch of warning orange indicator lights ON that suggest it is feeling the heat. Not sure if that is why, but I am unable to bring any of the C1IOO models back online - the rtcds compilation just fails, after which I am unable to ssh back into the machine as well.
  • C1SUS was problematic as well. I found that the expansion chassis was not powered. Fortunately, this was fixed by simply switching to the one free socket on the power strip that powers a bunch of stuff on 1X4 - this brought the expansion chassis back alive, and after a soft reboot of c1sus, I was able to get these models up and running. Fortunately, none of the electronics seem to have been damaged. Perhaps it is time for surge-protecting power strips inside the lab area as well (if they aren't already)? 
  • I was unable to successfully resolve the dmesg problem alluded to earlier. Looking through some forums, I gather that the output of dmesg should be written to a file in /var/log/. But no such file exists on any of our 5 front-ends (but it does on Megatron, for example). So is this way of setting up the front end machines deliberate? Why does this matter? Because it seems that the buffer which we see when we simply run "dmesg" on the console gets preiodically cleared. So sometime back, when I was trying to verify that the installed DACs are indeed 16-bit DACs by looking at dmesg, running "dmesg | head" showed a first line that was written to well after the last reboot of the machine. Anyway, this probably isn't a big deal, and I also verified during the model recompilation that all our DACs are indeed 16-bit.
  • I was also trying to set up the Upstart processes on megatron such that the MC autolocker and FSS slow control scripts start up automatically when the machine is rebooted. But since C1IOO isn't co-operating, I wasn't able to get very far on this front either...

So current status is that all front-end models except those hosted on C1IOO are back up and running. Further recovery efforts in progress.  

GV Jun 5 6pm: From my discussion with jamie, I gather that the fact that the dmesg output is not written to file is because our front-ends are diskless (this is also why the ring buffer, which is what we are reading from when running "dmesg", gets cleared periodically)

 

Quote:

Looks like there was a power glitch at around 10am today.

All frontends, FB, Megatron, Optimus were offline. Chiara reports an uptime of 666 days so looks like its UPS works fine. PSL was tripped, probably the end lasers too (yet to check). Slow machines seem alright (Responds to ping, and I can also telnet into them).

Since all the frontends have to be re-started manually, I am taking this opportunity to investigate some cds issues like the lack of a dmesg log file on some of the frontends. So the IFO will be offline for sometime.

 

Attachment 1: IMG_7399.JPG
IMG_7399.JPG
  13036   Fri Jun 2 22:01:52 2017 gautamUpdateGeneralPower glitch - recovery

[Koji, Rana, Gautam]

Attachment #1 - CDS status at the end of todays efforts. There is one red indicator light showing an RFM error which couldn't be fixed by running "global diag reset" or "mxstream restart" scripts, but getting to this point was a journey so we decided to call it for today.


The state this work was started in was as indicated in the previous elog - c1ioo wasn't ssh-able, but was responding to ping. We then did the following:

  1. Killed all models on all four other front ends other than c1ioo. 
  2. Hard reboot for c1ioo - at this point, we could ssh into c1ioo. With all other models killed, we restarted the c1ioo models one by one. They all came online smoothly.
  3. We then set about restarting the models on the other machines.
    • We started with the IOP models, and then restarted the others one by one
    • We then tried running "global diag reset", "mxstream restart" and "telnet fb 8087 -> shutdown" to get rid of all the red indicator fields on the CDS overview screen.
    • All models came back online, but the models on c1sus indicated a DC (data concentrator?) error. 
  4. After a few minutes, I noticed that all the models on c1iscex had stalled
    • dmesg pointed to a synchronization error when trying to initialize the ADC
    • The field that normally pulses at ~1pps on the CDS overview MEDM screen when the models are running normally was stuck
    • Repeated attempts to restart the models kept throwing up the same error in dmesg 
    • We even tried killing all models on all other frontends and restarting just those on c1iscex as detailed earlier in this elog for c1ioo - to no avail.
    • A walk to the end station to do a hard reboot of c1iscex revealed that both green indicator lights on the slave timing card in the expansion chassis were OFF.
    • The corresponding lights on the Master Timing Sequencer (which supplies the synchronization signal to all the front ends via optical fiber) were also off.
    • Sometime ago, Eric and I had noticed a similar problem. Back then, we simply switched the connection on the Master Timing Sequencer to the one unused available port, this fixed the problem. This time, switching the fiber connection on the Master Timing Sequencer had no effect.
    • Power cycling the Master Timing Sequencer had no effect
    • However, switching the optical fiber connections going to the X and Y ends lead to the green LED on the suspect port on the Master Timing Sequencer (originally the X end fiber was plugged in here) turning back ON when the Y end fiber was plugged in.
    • This suggested a problem with the slave timing card, and not the master. 
  5. Koji and I then did the following at the X-end electronics rack:
    • Shutdown c1iscex, toggled the switches in the front and back of the expansion chassis
    • Disconnect AC power from rear of c1iscex as well as the expansion chassis. This meant all LEDs in the expansion chassis went off, except a single one labelled "+5AUX" on the PCB - to make this go off, we had to disconnect a jumper on the PCB (see Attachment #2), and then toggle the power switches on the front and back of the expansion chassis (with the AC power still disconnected). Finally all lights were off.
    • Confident we had completely cut all power to the board, we then started re-connecting AC power. First we re-started the expansion chassis, and then re-booted c1iscex.
    • The lights on the slave timing card came on (including the one that pulses at ~1pps, which indicates normal operation)!
  6. Then we went back to the control room, and essentially repeated bullet points 2 and 3, but starting with c1iscex instead of c1ioo.
  7. The last twist in this tale was that though all the models came back online, the DC errors on c1sus models persisted. No amount of "mxstream restart", "global diag reset", or restarting fb would make these go away.
  8. Eventually, Koji noticed that there was a large discrepancy in the gpstimes indicated in c1x02 (the IOP model on c1sus), compared to all the other IOP models (even though the PDT displayed was correct). There were also a large number or IRIG-B errors indicated on the same c1x02 status screen, and the "TIM" indicator in the status word was red.
  9. Turns out, running ntpdate before restarting all the models somehow doesn't sync the gps time - so this was what was causing the DC errors. 
  10. So we did a hard reboot of c1sus (and for good measure, repeated the bullet points of 5 above on c1sus and its expansion chassis). Then, we tried starting the c1x02 model without running ntpdate first (on startup, there is an 8 hour mismatch between the actual time in Pasadena and the system time - but system time is 8 hours behind, so it isn't even somehow syncing to UTC or any other real timezone?)
    • Model started up smoothly
    • But there was still a 1 second discrepancy between the gpstime on c1x02 and all the other IOPs (and the 8 hour discrepancy between displayed PDT and actual time in Pasadena)
    • So we tried running ntpdate after starting c1x02 - this finally fixed the problem, gpstime and PDT on c1x02 agreed with the other frontends and the actual time in Pasadena.
    • However, the models on c1lsc and c1ioo crashed
    • So we restarted the IOPs on both these machines, and then the rest of the models.
  11. Finally, we ran "mxstream restart", "global diag reset", and restarted fb, to make the CDS overview screen look like it does now.

Why does ntpdate behave this way? And only on one of the frontends? And what is the remaining RFM error? 

Koji then restarted the IMC autolocker and FSS slow processes on megatron. The IMC locked almost immediately. The MC2 transmon indicated a large shift in the spot position, and also the PMC transmission is pretty low (while the lab temperature equilibriates after the AC being off during peak daytime heat). So the MC transmission is ~14500 counts, while we are used to more like 16,500 counts nowadays.

Re-alignment of the IFO remains to be done. I also did not restart the end lasers, or set up the Marconi with nominal params. 

Attachment #3 - Status of the Master Timing Sequencer after various reboots and power cycling of front ends and associated electronics.

Attachment #4 - Warning lights on C1IOO

Quote:

Today's recovery seems to be a lot more complicated than usual.

So current status is that all front-end models except those hosted on C1IOO are back up and running. Further recovery efforts in progress.  

 

Attachment 1: power_glitch_recovery.png
power_glitch_recovery.png
Attachment 2: IMG_7406.JPG
IMG_7406.JPG
Attachment 3: IMG_7407.JPG
IMG_7407.JPG
Attachment 4: IMG_7400.JPG
IMG_7400.JPG
  13038   Sun Jun 4 15:59:50 2017 gautamUpdateGeneralPower glitch - recovery

I think the CDS status is back to normal.

  • Bit 2 of the C1RFM status word was red, indicating something was wrong with "GE FANUC RFM Card 0".
  • You would think the RFM errors occur in pairs, in C1RFM and in some other model - but in this case, the only red light was on c1rfm.
  • While trying to re-align the IFO, I noticed that the TRY time series flatlined at 0 even though I could see flashes on the TRANSMON camera.
  • Quick trip to the Y-End with an oscilloscope confirmed that there was nothing wrong with the PD.
  • I crawled through some elogs, but didn't really find any instructions on how to fix this problem - the couple of references I did find to similar problems reported red indicator lights occurring in pairs on two or more models, and the problem was then fixed by restarting said models.
  • So on a hunch, I restarted all models on c1iscey (no hard or soft reboot of the FE was required)
  • This fixed the problem
  • I also had to start the monit process manually on some of the FEs like c1sus. 

Now IFO work like fixing ASS can continue...

Attachment 1: powerGlitchRecovery.png
powerGlitchRecovery.png
  13079   Sun Jun 25 22:30:57 2017 gautamUpdateGeneralc1iscex timing troubles

I saw that the CDS overview screen indicated problems with c1iscex (also ETMX was erratic). I took a closer look and thought it might be a timing issue - a walk to the X-end confirmed this, the 1pps status light on the timing slave card was no longer blinking. 

I tried all versions of power cycling and debugging this problem known to me, including those suggested in this thread and from a more recent time. I am leaving things as it for the night, will look into this more tomorrow. I've also shutdown the ETMX watchdog for the time being. Looks like this has been down since 24Jun 8am UTC.

Attachment 1: c1iscex_status.png
c1iscex_status.png
  13080   Mon Jun 26 09:39:15 2017 NaomiSummaryGeneralMeasure transfer functions of Mini-Circuits filters

I have spent my first few days as a SURF getting experience working with the Network/Spectrum Analyzer (AG 4395A). After an introduction to the 40m by Koji, I was tasked with using the AG4395A to measure the transfer function of several filters (for example, Mini-Circuits Low Pass Filter SLP-30). I am now familiar with configuring the AG 4395A, taking a single set of data using a command from one of the control computers, and plotting the dataset as a Bode plot (separate plots for magnitude and phase) using Python.

To Do:

  • Use AGmeasure to take multiple datasets with a single command.
  • Plot multiple datasets for each filter on a single Bode plot and perform some statistical analysis. 

To experiment with plotting multiple datasets on a single Bode plot, I used a single dataset from the Network Analyzer using the SLP-30 filter and added random noise to create ten datasets to plot. I am attaching the resulting Bode plot, which has the ten generated sets of data plotted along with their average.

We discussed with Rana and Koji how to interpret this type of dataset from the Network Analyzer. Instead of considering the magnitude and phase as separate quantities, we should consider them together as a single complex number in the form H(f) = M exp(iπP/180), where M is the magnitude and P is the phase in degrees. We can then find the average value of the measured quantity in its complex number form (x + iy), as opposed to just taking the average of the magnitude and phase separately.  

Attachment 1: Bode_Plot.png
Bode_Plot.png
  13081   Mon Jun 26 22:01:08 2017 KojiUpdateGeneralc1iscex timing troubles

I tried a couple of things, but no fundamental improvement of the missing LED light on the timing board.

- The power supply cable to the timing board at c1iscex indicated +12.3V

- I swapped the timing fiber to the new one (orange) in the digital cabinet. It didn't help.

- I swapped the opto-electronic I/F for the timing fiber with the Y-end one. The X-end one worked at Y-end, and Y-end one didn't work at X-end.

- I suspected the timing board itself -> I brought a "spare" timing board from the digital cabinet and tried to swap the board. This didn't help.

 

Some ideas:

- Bring the X-end fiber to C1SUS or C1IOO to see if the fiber is OK or not.

- We checked the opto-electronic I/F is OK

- Try to swap the IO chassis with the Y-end one.

- If this helps, swap the timing board only to see this is the problem or not.