ID |
Date |
Author |
Type |
Category |
Subject |
12430
|
Mon Aug 22 18:04:24 2016 |
gautam | Update | SUS | ETMY OSEMs inserted | [Johannes, gautam]
We worked on trying to insert the OSEMs in the optimal positions such that the coupling of the bounce mode into the OSEM sensor signals was minimised.
First, I gave the barrel of the optic a wipe with some optical tissue + acetone in order to remove what looked like some thin fibres of dried first contact. It may be that while I was applying the F.C., the HEPA air flow deposited these on the barrel. In any case, they came off easily enough. There is still a few specks of dust on various parts of the barrel, but it is likely that these can just be removed with the ionized air jet, which we can do after putting the optic in the chamber.
We then did the usual OSEM insertion till the magnets neutral position was such that the sensor output was ~50% of the fully open value (turned the HEPA off for the remainder of this work). I tweaked the bottom OSEM plate a little in order to center the magnets relative to the coil as best as possible. Once this was done, we attempted to look at spectra of the sensor outputs, with 0.05 Hz bandwidth - however, we were unable to identify any peak at 16.4 Hz, which is what a Jan 2015 measured value wiki page claims the bounce mode frequency is (although this was an in vacuum measurement). There were a couple of peaks at ~15.7 Hz and ~16.7 Hz, but I can't think of any reason why the bounce mode resonance should have changed so much - after all, this is ETMY for which no standoff regluing was done. The only difference is that there is some first contact + peek mesh on the HR face now, but I doubt this can modify the bounce resonance frequency so much (this is just my guess, I will have to back this up with a calculation).
Anyways we decided to take this up again tomorrow. Things are progressing fairly well now, I hope to be able to put in ETMY back into the chamber at some point tomorrow and commence re-alignment of the interferometer. I've left the OSEMs in for today, with the EQ stops not engaged but close by. HEPA has been turned back on. |
12429
|
Mon Aug 22 16:33:32 2016 |
Steve | Update | PEM | the lab temp is up | The IFO room temp is up a bit and it is coming down. The out side temp is not really high. |
Attachment 1: temp.png
|
|
12428
|
Mon Aug 22 13:06:11 2016 |
gautam | Update | SUS | ETMY suspended | Today morning, I suspended ETMY and made the same checks dscribed below. The clamping went smoothly, 5 in. lb. of torque seems sufficient, in the limited observation time, there has been no evidence of wire sag. Today afternoon, we will go about putting the OSEM coils in, setting their equilibrium points etc. This may need to be re-done once the optic is in the chamber and the first contact has come off, but at least we can coarsely place them in the relative convenience of the cleanroom.
GV EDIT 9.15pm 22 Aug: Eric had a look at both towers and pointed out that I had neglected to use washers on the wire stops. After consultation with Steve, I decided that it is not worth it to remove the clamp and re-suspend the optic - it is likely that the current suspension process will have caused new grooves in the suspension block, which will have to be removed, and the sanding process did not work so well last time. In any case, the net effect of this will be that the actual torque with which the clamp is tightened will be slightly different from 5 in. lb., but since there is no evidence that the clamp isn't tight enough / is too tight, I think it is okay to push ahead.
Quote: |
- ETMX has been successfully suspended
- I've used one of the new wire clamps, and also the new suspension wire
- Because the HR face has first contact, pitch balancing cannot be checked at this point. But since the pitch balance was checked after the standoff was glued, there is no reason to believe it would have changed
- Heights of the two scribe lines were checked with the microscope and verified to be at 5.5" above the tabletop. Also checked the position of the scribe line on the bottom of the optic to make sure the optic wasn't somehow rotated
- Checked that wire was in the groove in the standoff on both sides, and that the optic was freely hanging with no EQ stops engaged. I also verified that there are no obvious kinks/other funny features where the wire is in contact with the optic barrel below the standoffs.
- Wire clamps were tightened with the new torque wrench and 5 in. lb. (0.56 N m) of torque. Primary clamp was successfully tightened. However, the wire snapped between the primary and secondary clamps on one side. It is unclear to me how or why this happened. But since the primary wire clamp is the important one, I don't think it is worth re-suspending ETMX all over again
- I've left the cage on the flow bench for now, with EQ stops engaged. OSEM coils have yet to be inserted, but I suppose we want to do this in the vacuum chamber now to do the fine rotation to minimize the bounce mode in the OSEM signals
- I've prepared ETMY and its cage for suspension, will work on it tomorrow
|
|
Attachment 1: IMG_3019.JPG
|
|
12427
|
Sun Aug 21 17:21:22 2016 |
Praful | Update | Electronics | Problems with PCB Circuit | For the past week, I've been trying to make a soldered amplifier circuit to use in a prototype box, However, I've been running into this same issue. The circuit, pictured below, works fine on a solderless breadboard.

simple_amp.png
When I amplify a sine wave, I get a clean looking result at the output on the solderless breadboard:

However, on my soldered circuit, if I turn up the negative voltage supply from the power supply past about -12.5V (the target is -15V), I get a strange signal that Gautam suggested looks like some kind of discharging.
At -12.3 V (soldered breadboard):

At -15.0 V (soldered breadboard):

The signal is much noisier. Zooming in on this second signal, this pattern appears:


This pattern is also showing up even when there is no input from the function generator and the circuit is just given a voltage supply of +/- 15V:

I have tried switching out both the positive and negative voltage regulators, the opamp, and remaking and resoldering the entire circuit but I'm still getting the same signal, which is absent from the solderless circuit. This output was produced with a function generator, so I have also ruled out the microphone as a source of this extra noise. The voltage dependence of this problem made me think it was the voltage regulator, but I've switched out the voltage regulator multiple times and it's still showing up. I'm not sure why this signal appears only as the negative voltage supply is increased- there is no problem with increasing the positive input voltage. Please let me know if you have any ideas as to what component or issue could be causing this. |
Attachment 2: simple_amp.png
|
|
Attachment 4: clean.jpg
|
|
Attachment 5: -12.jpg
|
|
Attachment 6: -15.jpg
|
|
Attachment 7: pat1.jpg
|
|
Attachment 8: pat2.jpg
|
|
Attachment 10: bad.jpg
|
|
Attachment 11: pattern.jpg
|
|
Attachment 12: pattern2.jpg
|
|
Attachment 13: pat2.jpg
|
|
Attachment 16: patternzoomed.jpg
|
|
12426
|
Sun Aug 21 16:23:05 2016 |
gautam | Update | SUS | ETMX suspended |
- ETMX has been successfully suspended
- I've used one of the new wire clamps, and also the new suspension wire
- Because the HR face has first contact, pitch balancing cannot be checked at this point. But since the pitch balance was checked after the standoff was glued, there is no reason to believe it would have changed
- Heights of the two scribe lines were checked with the microscope and verified to be at 5.5" above the tabletop. Also checked the position of the scribe line on the bottom of the optic to make sure the optic wasn't somehow rotated
- Checked that wire was in the groove in the standoff on both sides, and that the optic was freely hanging with no EQ stops engaged. I also verified that there are no obvious kinks/other funny features where the wire is in contact with the optic barrel below the standoffs.
- Wire clamps were tightened with the new torque wrench and 5 in. lb. (0.56 N m) of torque. Primary clamp was successfully tightened. However, the wire snapped between the primary and secondary clamps on one side. It is unclear to me how or why this happened. But since the primary wire clamp is the important one, I don't think it is worth re-suspending ETMX all over again
- I've left the cage on the flow bench for now, with EQ stops engaged. OSEM coils have yet to be inserted, but I suppose we want to do this in the vacuum chamber now to do the fine rotation to minimize the bounce mode in the OSEM signals
- I've prepared ETMY and its cage for suspension, will work on it tomorrow
|
Attachment 1: IMG_3018.JPG
|
|
12424
|
Fri Aug 19 22:51:12 2016 |
gautam | Update | SUS | ETMs first-contacted | I've applied first contact to both the ETMs. They're now ready to be suspended. I've also cut up some lengths of the new wire and put them in the oven for a 12 hour 70C bake.
- For both ETMs, I first applied first contact to the bulk of the HR and AR surfaces (all the way out to the edge for the HR, for the AR as large an area as possible without getting too close to the magnets). Calum recommended pouring first contact onto the horizontal optic, but since I had no practise with this method, I opted not to try it out for the first time on our ETMs
- After allowing this to dry for 24 hours, I peeled this layer off. Visual inspection suggests that the whole film came off cleanly.
- I then applied first contact to a smaller area around the center of the optic for only the HR surface. This will only be peeled off once the suspended optic is back in the vacuum chamber. This way, we keep the HR face protected for as long as possible.
- Even though we applied F.C to both faces of the ITMs, I don't think its so important to keep a film on the AR side of the ETMs till we take it in. So I didnt re-coat the AR side with a smaller area of F.C. This way, if we want, we can do the OSEM assembly in the cleanroom without having to worry about peeling the F.C off with limited access to the rear of the optic.
- I also opted to bake some lengths of the newly arrived steel wire for suspension. Not sure how important/useful this bake will be.
Unless we want the AR surface to also have a small F.C coat until the optic is in the vacuum chamber, I think I will proceed with re-suspending the ETMs.. |
12423
|
Thu Aug 18 15:16:09 2016 |
Steve | Update | SUS | SOS sus wire is in |
Stress Relieved 0.0017" Music Wire CFW P/N: CFW2035025, Made 08-17-2016
|
Old 2003 |
New 2016 |
|
GBL |
358.9 |
240.610 |
grams |
UTS |
357,061 |
229,603 |
PSI |
YTS |
343,211 |
177,371 |
PSI |
ELONG |
2.38 |
0.8 |
% |
HEAT |
10622 |
10622 |
|
GBL (grams breaking load )
UTS (ultimate tensile strength)
YTS (yield tensile strength)
ELONG (elongation)
Quote: |
0.0017" OD., 500ft steel music wire ordered. Pictures of the existing roll are below. It will be on 8" OD. spool too.
|
|
Attachment 1: 0.0017new.jpg
|
|
12422
|
Thu Aug 18 14:14:20 2016 |
gautam | Update | SUS | Air-bake of towers - finished | I took the two cages, wires and wire clamps out this morning, back into the cleanroom after their 12 hour 70C bake.
I've also applied first contact to the AR face of the optics. Steve is preparing a jig which will allow us to apply first contact on the HR side with the optic horizontal. The idea is to apply a large coating first, to clean the bulk of the HR surface, and peel it off before re-suspending the optic. Then we can paint on a smaller area, suspend the optic (and hope the pitch balancing is alright) before taking the whole assembly into the chamber where it will be peeled off.
Calum recommended that we buy a new ionizing gun + electrometer assembly (apparently our current set up is woefully obsolete) but I don't know if we can have these in time for the first contact peeling... |
12421
|
Thu Aug 18 08:17:16 2016 |
Steve | Update | SUS | wire clamping preparation | The wire inprints were removed by 800P grain paper [Norton 73568] The SS bridge block now has an undesireble vally in the wire location.
The sus bridges were soaked in acetone over night and sonicated to remove residual sand paper.
Quote: |
I just put in the following into the air bake oven for a 12 hour, 70C bake:
- ETMX and ETMY cages (with sanded suspension blocks loosely tightened for now, we will tighten them after the bake)
- 13 new wire clamps that were recently made by the shop
- 7 lengths of suspension wire (since the new wire is unlikely to arrive for another 2 weeks). This should be sufficient in case we overtighten the wire clamps a couple of times and the wire snaps.
I put these in at 10.30pm. So the oven will be turned off at 10.30am tomorrow morning. The oven temperature seems stable in the region 70-80 C (there is no temperature control except for the in built oven control, I just adjusted the dial till I found the oven remains at ~70C.
Tomorrow, we will look to put on first contact onto the ETMs, and then get about to re-suspending them.
|
|
Attachment 1: 800P.jpg
|
|
12420
|
Wed Aug 17 23:00:57 2016 |
gautam | Update | SUS | Air-bake of towers | I just put in the following into the air bake oven for a 12 hour, 70C bake:
- ETMX and ETMY cages (with sanded suspension blocks loosely tightened for now, we will tighten them after the bake)
- 13 new wire clamps that were recently made by the shop
- 7 lengths of suspension wire (since the new wire is unlikely to arrive for another 2 weeks). This should be sufficient in case we overtighten the wire clamps a couple of times and the wire snaps.
I put these in at 10.30pm. So the oven will be turned off at 10.30am tomorrow morning. The oven temperature seems stable in the region 70-80 C (there is no temperature control except for the in built oven control, I just adjusted the dial till I found the oven remains at ~70C.
Tomorrow, we will look to put on first contact onto the ETMs, and then get about to re-suspending them. |
Attachment 1: IMG_3006.JPG
|
|
12419
|
Wed Aug 17 22:09:04 2016 |
rana | Update | SUS | SOS sus wire ordered | Not really true that it passed. That's just an arbitrary margin. Best to throw away all the old wire. We have no quantitative estimate of what the real torque should be. Its just feelings.
Quote: |
The wire will arrive in 1-2 weeks. It is a new production. Brad Snook of Ca Fine Wire was suprised that we are still using the 13 years old wire. Oxidation is an issue with iron contained steel wire.
He would not give me a shelf life time on it. He recommended to check the strenght of it before usage. It passed with safety factor of 2 just recently.
In the future we'll store the new spool in oxigen free nitrogen environment..
|
|
12418
|
Wed Aug 17 16:28:46 2016 |
Koji | Update | COC | RC folding mirrors - Numerical review | For the given range of the PR3/SR3 RoCs for both cases, all the resulting numbers such as TMSs/mode matching ratios look reasonable to me. |
12417
|
Wed Aug 17 14:37:36 2016 |
gautam | Update | COC | RC folding mirrors - Numerical review |
Quote: |
Cavity |
One-way Gouy phase [rad] |
TMS [MHz] |
PRX |
0.244 |
1.730 |
PRY |
0.243 |
1.716 |
SRX |
0.197 |
1.743 |
SRY |
0.194 |
1.717 |
So, there are regions in parameter space for both options (i.e. keep current G&H mirrors, or order two new sets of folding mirrors) that get us close to the design numbers...
|
Keeping these design numbers in mind, here are a few possible scenarios. The "designed" TMS numbers from my previous elog are above for quick reference.
Case 1: Keep existing G&H mirror, flip it back the right way, and order new PR3/SR3.
- Spec PR3 to be concave with RoC 600 +/- 50m
- This means the TMS in the PRC is in the range 1.4 MHz - 1.6 MHz [see this plot]
- The mode matching efficiency for the PRC is > 98.5% [see this plot]
- The TMS in the SRC is in the range 1.6 MHz - 1.8 MHz [see this plot]
- Mode matching efficiency for SRC is > 98.5% [see this plot]
- PRG between 34-38, depending on uncertainty in measurement of RoC of existing G&H mirror [see Attachment #1, added Nov 11 2016]
Case 2: Order two new sets of folding mirrors
- Spec PR3/SR3 to be flat - for purposes of simulation, let's make it concave with RoC 10 +/- 5 km
- Spec PR2/SR2 to be concave with RoC 1500 +/- 500m
- The TMS in the PRC is between 1.7 MHz and 1.85 MHz [see this plot]
- Mode matching efficiency is >98.5% in the PRC [see this plot]
- TMS in the SRC is between 1.7 MHz and 2 MHz [see this plot]
- Mode matching efficiency >99.0% in the SRC [see this plot]
At first glance, it looks like the tolerances are much larger for Case 2, but we also have to keep in mind that for such large RoCs in the km range, it may be impractical to specify as tight tolerances as in the 100s of metres range. So these are a set of numbers to keep in mind, that we can re-iterate once we hear back from vendors as to what they can do.
For consolidation purposes, here are the aLIGO requirements for the coatings on the RC folding mirrors: PR2, PR3, SR2, SR3 |
Attachment 1: PRG.pdf
|
|
12416
|
Wed Aug 17 08:47:43 2016 |
ericq | Update | SUS | Air-bake finished | I turned off the air bake oven at 8:45AM. I'll leave the optics alone for a bit while it cools. |
12415
|
Tue Aug 16 21:54:27 2016 |
gautam | Update | SUS | Air-bake - IN PROGRESS | I put in both ETMX and ETMY into the air-bake oven at approximately 8.45pm tonight. They can be removed at 8.45am tomorrow morning.
- Given that we had previously melted a thermocouple in this oven, and there have been no high temperature bakes in it since, we ran the oven at 100C for about 3 hours in the afternoon
- After that, I left the oven door open for an hour for the interior to return to room temperature
- I then re-connected the controller (which doesn't seem very precise, it pulses the AC power to the oven in order to control the temperature), and dialled the oven back down to heating level 4, which is what Bob had it set at. I then waited for a couple of hours for the oven to reach ~34C
- Before putting the optics in, I gave the inside of the oven a quick wipe with a clean wipe, and palced a layer of Al foil on the bottom of the oven
- The optics are sitting on their donuts (see Attachment #1) - the copper wire elevates the optic+donut slightly and provides a path for air flow
- ETMY was drag wiped with acetone+isopropanol prior to baking (to remove acetone stains from soaking to remove epoxy residue
- We will of course be cleaning the optics with first contact prior to re-installation in the vacuum chambers
- I am not sure what the extra cylindrical piece in there is, but Bob advised me to leave it in there so that's what I did
- I've observed the temperature over ~2hours since I first put it in, and the oven/controller isn't going bonkers, so I'm trusting the controller and leaving for the night
|
Attachment 1: IMG_3005.JPG
|
|
12414
|
Tue Aug 16 16:38:00 2016 |
gautam | Update | COC | RC folding mirrors - Numerical review | Here are the results for case 2: (flat PR3/SR3, for purpose of simulation, I've used a concave mirror with RoC in the range 5-15km, and concave PR2/SR2 - I've looked at the RoC range 300m-4km).
- This is where we order two new sets of mirrors, one for use as PR2/SR2, and the other for use as PR3/SR3.
- RoC of flat PR3/SR3 in simulation explored in the range 5km-15km (concave)
- RoC of concave PR2/SR2 in simulation explored in the range 300m-4km (concave)
Attachment #1: Mode matching between PRC cavities and arm cavities with some contour plots
Attachment #2: Mode matching between SRC cavities and arm cavities with some contour plots
Attachment #3: Gouy phase and TMS for the PRC. I've plotted two sets of curves, one for a PR3 with RoC 5km, and the other for a PR3 with RoC 15km
Attachment #4: Gouy phase and TMS for the SRC. Two sets of curves plotted, as above.
Hopefully EricG will have some information with regards to what is practical to spec at tomorrow's meeting.
EDIT: Added 9pm, 16 Aug 2016
A useful number to have is the designed one-way Gouy phase and TMS for the various cavities. To calculate these, I assume flat folding mirrors, and that the PRM has an RoC of 115.5m, SRM has an RoC of 148m (numbers taken from the wiki). The results may be summarized as:
Cavity |
One-way Gouy phase [rad] |
TMS [MHz] |
PRX |
0.244 |
1.730 |
PRY |
0.243 |
1.716 |
SRX |
0.197 |
1.743 |
SRY |
0.194 |
1.717 |
So, there are regions in parameter space for both options (i.e. keep current G&H mirrors, or order two new sets of folding mirrors) that get us close to the design numbers... |
Attachment 1: PRC_consolidated.pdf
|
|
Attachment 2: SRC_consolidated.pdf
|
|
Attachment 3: GouyPRC.pdf
|
|
Attachment 4: GouySRC.pdf
|
|
12413
|
Tue Aug 16 11:51:43 2016 |
gautam | Update | COC | RC folding mirrors - Numerical review | Summary of roundtable meeting yesterday between EricG, EricQ, Koji and Gautam:
We identified two possible courses of action.
- Flip the G&H mirror (PR2/SR2) back such that the (convex) HR face is the right way round. We want to investigate what are the requirements on a new PR3/SR3 optic that will guarantee cavity stability and also give good mode matching.
- Order two new sets of mirrors (i.e. replace all 4 folding mirrors). In this case, we want to spec a flat (how flat is reasonable to specify? EricG will update us) PR3/SR3, and design a PR2/SR2 with some concavity that will guarantee cavity stability in the event PR3/SR3 deviates from flatness (but still within what we spec). The choice to make PR3 as close to flat as possible is because the angle of incidence in our arrangement means that any curvature on PR3 dominates astigmatism.
I have done some calculations to evaluate the first alternative.
- Based on yesterday's preliminary discussion, we felt it is not reasonable to spec mirrors with RoC > 4km (sag of ~80nm). So I restrict my analyses to the range 300m-4km
- Koji has a measurement of the phase maps for the G&H mirrors. The measured curvature is ~-500m. In my simulations, I've tried to allow for error in this measurement, so I look at the range -450m to -700m for the G&H mirror.
- The Gouy phase analysis suggests we should look for an RoC of +500m (concave) for the new PR3/SR3 to have a TMS of ~1.5 MHz. Anything flatter (but still concave) means the TMS gets smaller.
- The mode-matching in this region also looks pretty good, between 98% and 99%
- I will post results of the analysis for the second alternative here for comparison
Something else that came up in yesterdays meeting was if we should go in for 1" optics rather than 2", seeing as the beam spot is only ~3mm on these. It is not clear what (if any) advantages this will offer us (indeed, for the same RoC, the sag is smaller for a 1" optic than a 2").
Attachments:
Attachment #1: Mode-matching maps between PRX and Xarm cavities, PRY and Yarm cavities with some contours overlaid.
Attachment #2: Mode-matching maps between SRX and Xarm cavities, SRY and Yarm cavities with some contours overlaid.
Attachment #3: Gouy phase calculations for the PRC
Attachment #3: Gouy phase calculations for the SRC
|
Attachment 1: PRC_consolidated.pdf
|
|
Attachment 2: SRC_consolidated.pdf
|
|
Attachment 3: GouyPRC.pdf
|
|
Attachment 4: GouySRC.pdf
|
|
12412
|
Tue Aug 16 08:04:49 2016 |
Steve | Update | SUS | ETMY had Aluminum bridge | The SOS ETMY tower had and Aluminum bridge. How is it possible that this was true? Is SS better than Al for some quantitative scientifc reason ???
Their weight ratio as measured = 922 / 307 g = 3
Destinies: SS 304 / Al 6061 > 0.289 / 0.098 [ lb /in3 ] = 2.94 |
Attachment 1: AlBridge.jpg
|
|
12411
|
Mon Aug 15 18:28:15 2016 |
gautam | Update | SUS | Air-bake preparation | I assume that we are prepared to live with the pitch bias situation of ETMY (i.e. we can achieve a configuration in which there is some pitch bias to the coils, and the OSEMs are inserted such that the PD outputs are half their maximum value). Or at least that we don't want to go through the whole standoff-regluing procedure for ETMY as well.
So today I took the optic out, and began to make some preparations for the air bake.
- Both optics are now sitting in their respective metal donuts.
- How do we want to bake the optics? Bob has said he has prepared the oven for this bake, and that he has configured the temperature controller to a setpoint of 34C, and a ramp time of 2 hours to reach that temperature from lab temperature (we should check this before putting the optics in there with our independent temperature sensor - also, he is away for the week now so we can't get his input on any of these). But what about the actual logistics of how the optics are going to be housed? Specifically:
- Do we want the donut to sit on some sort of tray? Presumably it is not ideal to have the HR surface in close proximity to the oven floor?
- Does the oven need any special cleaning?
- Do we cover the donut+optic setup with a glass jar? If we do, any particles we eject off the optic can't escape the confines of the bowl, and if we don't, detritus from elsewhere may settle on the optic?
- How long do we want this bake to last? 24hours? 48 hours? Bob didn't have an answer when I asked him earlier in the afternoon...
- I also removed the suspension block from the top of the towers of both ETMX and ETMY, so that Steve could work on sanding them before we acetone-wipe and bake the towers themselves.
- It was very apparent that the weights of the two pieces were largely different (ETMY suspension block ~350g, ETMX suspension block ~960g), even though they have the same physical dimensions.
- Investigation into why this was yielded nothing conclusive. But Steve and I think that the ETMY suspension block is made out of Aluminum rather than SS, which would explain why the wire grooves seem deeper in the ETMY piece than the ETMX piece. It is worth noting that the specification calls for SS and not aluminum. But the top piece of the ETMY suspension (and indeed the old ETMX suspension) looks different from the specification, in that they don't have tapped holes for the secondary wire clamps (see Attachment #1).
- I'm not sure if this is important, but it is worth noting. Steve and I also checked the remaining suspension towers. We think that ITMY, BS, SRM and PRM have the correct (to specification) suspension block. We couldn't get a look at ITMX and didn't want to take the door off. So ETMY (and possibly ITMX) will be the only suspension(s?) with a different suspension block.
- Steve's sanding efforts did not go ideally.
- He was successful in removing the wire grooves.
- But the sharp edge which is supposed to clamp the wire seems to have been rounded a little bit (see Attachment #1).
- Overall, the section that we was sanded looks lower (i.e. its like we've dug a small channel into the plane of the suspension block)
- Given that we suspect the ETMY suspension block is Aluminum, it is likely that attempting to sand it will yield an even deeper channel.
- Do we want to bake the suspension towers in the large baking oven? Presumably we don't want to bake the optics with anything else. But does the large oven need any special cleaning before we stick the towers in there?
- ETMY has some acetone marks on it. I will try and have this removed by drag wiping with more acetone and isopropanol prior to the bake tomorrow. Anyways we will first-contact clean the HR (and AR) sides after the bake before installing the optic.
In summary, the questions that remain (to me) are:
- Are we okay using an Al suspension block?
- How perfectly do we want wire grooves from prior suspensions removed? It looks like sanding doesn't work well, do we want to consider sending this into the shop?
- Baking logistics, as described above.
I think we can start the baking of the optics tomorrow. The timeline for the suspension towers is unclear, depends on how we want to deal with the sanding dilemma. |
Attachment 1: IMG_6816.JPG
|
|
12410
|
Mon Aug 15 14:34:33 2016 |
ericq | Update | SUS | SOS sus wire ordered | We have indeed seen numerous tarnished/rusty points along the wires, and just tried to choose lengths free of any of these. I wonder if this can explain the brittleness/ease with which we've been breaking it. My feeling is that we should use the newer wire if feasible. |
12409
|
Mon Aug 15 14:29:32 2016 |
Steve | Update | SUS | SOS sus wire ordered | The wire will arrive in 1-2 weeks. It is a new production. Brad Snook of Ca Fine Wire was suprised that we are still using the 13 years old wire. Oxidation is an issue with iron contained steel wire.
He would not give me a shelf life time on it. He recommended to check the strenght of it before usage. It passed with safety factor of 2 just recently.
In the future we'll store the new spool in oxigen free nitrogen environment..
Quote: |
0.0017" OD., 500ft steel music wire ordered. Pictures of the existing roll are below. It will on 9" OD. spool too.
|
|
12408
|
Mon Aug 15 12:23:56 2016 |
Praful | Update | PEM | Mic Self Noise | I didn't have a separate training set and data set, so I think that's why the graphs came out looking too good. The units on the graphs are also incorrect, I was interpreting PSD as ASD. I haven't been able to get my Wiener filtering code working well- I get unreasonable subtractions like the noise being larger than the unfiltered signal, so Eric showed me this frequency-dependent calculation described here: https://dcc.ligo.org/LIGO-P990002
This seems to be working well so far:
freq1.pdf

freq2.pdf

freq3.pdf

Here's all the plots on one figure:
frequency_dependent.pdf

Let me know if this looks believable.
Quote: |
Seems to good to be true. Maybe you're over fitting? Please put all the traces on one plot and let us know how you do the parameter setting. You should use half the data for training the filter and the second half for doing the subtraction.
|
|
Attachment 1: freq1.png
|
|
Attachment 2: freq1.pdf
|
|
Attachment 4: freq2.pdf
|
|
Attachment 6: freq3.pdf
|
|
Attachment 8: frequency_dependent.pdf
|
|
12407
|
Sat Aug 13 18:25:22 2016 |
gautam | Update | COC | RC folding mirrors - Numerical review | This elog is meant to summarize my numerical simulations for looking into the effects of curvature on the RC mirrors. I've tried to go through my reasoning (which may or may not be correct) and once this gets a bit more refined, I will put all of this into a technical note.
Motivation:
- Both the G&H (PR2, SR2) and Laseroptik (PR3 SR3) are convex on the HR side with RoCs of approximately -600m and -700m (though as stated in the linked elog, I'm not actually sure if there are measurements of this number) EDIT AUG15: There are measurements for the Laseroptik mirrors here
GV April 8 2017: This elog by Jenne suggests that the installed PR2 has an RoC of approximately -700m. Koji has uploaded the phase map data for the RC TT mirrors to /users/public_html/40m_phasemap/40m_TT and /users/public_html/40m_phasemap/40m_TT2. The G&H mirror data seems to be in the former folder, and it looks like there are two mirrors, one with RoC of ~ -700m and the other with RoC of ~ -500m. Does this mean PR2 has RoC -700m and SR2 has RoC -500m?
- As a result, both the PRC and SRC were close to instability
- By flipping the folding mirrors, the instability has been mitigated, but at the expense of the non-ideal situation where the AR coated side and the substrate are now inside the recycling cavity
- We would like to order some new folding mirrors. In order to avoid receiving convex mirrors from the vendor, we want to specify a concave curvature for the HR side
- The aim of this investigation is to look at how concave we should make these mirrors, because although the cavity stability improves with concavity of the HR side, possible disadvantages of having too convex mirrors are:
- Mode-mismatch between the recycling cavities and the arms
- Astigmatism
The study:
- I've built a Finesse model for the 40m, which has been used for all the numerical studies quoted here
- In constructing this Finesse model, I've used the following sources to specify various paramaters:
- RoCs, R, T and physical dimensions of 4 test-masses, PRM, SRM and BS: Core optics wiki page
- Losses - arm losses from Yutaro's measurements in elog11857 and elog11818 (distributed equally between ITM and ETM). For other optics, a generic value of 25ppm was used
- "Ideal" lengths for our current modulation frequency were used for the various cavities (37.795m for the arms, 6.753m for PRC, 5.399 for SRC)
- The folding mirrors (PR2, PR3, SR2, SR3) are initialized as flat in the model
- I performed some low-level checks (e.g. arm linewidth, PRC FSR etc) to check that the model was sensible
- I then proceeded to investigate the effects of curvature on the folding mirrors. Specifically, I investigated the following:
- What is the mode mismatch between the recycling cavity mode and the arm as a function of the RoC of the folding mirror?
- What is the effect of the RoC of the folding mirrors on the round-trip gouy phase accumulated (and hence the transverse mode spacing) in the recycling cavities?
- For now, the parameter space explored is from 300m concave to 1000m concave. An RoC of 1km for a 2" optic corresponds to a sag of ~0.3 microns. I will explore the 1km-10km concave space and update the results shortly
Results:
- Attachments #1 and #2 show the mode mismatch between the recycling cavity and the arm for various curvatures. The colorbars have been normalized to span the same range in all the plots
- For both the PRC and the SRC, if we have folding mirrors with an RoC of 1000m concave, we will have a mode mismatch of 2-3%. The number gets worse the more convex the mirror
- Attachments #3 and #4 show the one-way accumulated Gouy phase. Here, I have varied the curvature of the folding mirrors along a specific axis at a time (i.e. I've assumed that the folding mirrors are identical). I've also added the transverse mode spacing as a second y-axis. I have yet to check how these numbers compare with the linewidth of the 00-mode for the various fields, but for 1km concave folding mirrors, the TMS is in the region of 2MHz
To do:
- I will extend the range of RoCs explored to 10km concave and post results - but I will have to check with EricG to make sure that it is feasible for us to specify curvatures in this range
- I was trying to use the RT gouy phase as calcluated by my Finesse simulations to plug into some analytical expressions to try and generate plots like this for various RoCs of the folding mirrors, but if the TMS calculations suffice, I will abandon these efforts
- What are the other specifications we need to worry about before placing an order? Some thoughts from Rana's earlier elog:
- The coatings need to be dichroic to allow extraction of the green beam (but only PR3/SR3 is currently dichroic?)
- Wedge angle on the AR side?
- Are there any other obvious sanity checks I should carry out?
|
Attachment 1: PRX_consolidated.pdf
|
|
Attachment 2: SRX_consolidated.pdf
|
|
Attachment 3: Gouy_PRC.pdf
|
|
Attachment 4: Gouy_SRC.pdf
|
|
12406
|
Fri Aug 12 21:26:28 2016 |
rana | Update | PEM | Mic Self Noise | Seems to good to be true. Maybe you're over fitting? Please put all the traces on one plot and let us know how you do the parameter setting. You should use half the data for training the filter and the second half for doing the subtraction. |
12405
|
Fri Aug 12 19:13:25 2016 |
Praful | Update | Electronics | Mic Self Noise | I used the Wiener filtering method described by Ignacio and Jessica (https://dcc.ligo.org/DocDB/0119/T1500195/002/SURF_Final.pdf and https://dcc.ligo.org/public/0119/T1500194/001/Final_Report.pdf) and got the following results:
mic1_wiener.pdf

mic2_wiener.pdf

mic3_wiener.pdf

The channel readout has a gain of 0.0005 and the ADC is 16-bit and operates are 20V. The channel also reads the data out in Pa. I therefore had to multiply the timeseries by 1/0.0005=2000 to get it in units of counts and then by (20 Volts)/(2^16 counts) to get back to the original signal in volts. The PSDs were generated after doing this calibration. I also squared, integrated, and square rooted the PSDs to get an RMS voltage for each microphone as a sanity check:
Mic 1: 0.00036 V
Mic 2: 0.00023 V
Mic 3: 0.00028 V
These values seem reasonable given that the timeseries look like this:
timeseries_elog.pdf

|
Attachment 4: mic1_wiener.pdf
|
|
Attachment 5: mic2_wiener.pdf
|
|
Attachment 6: mic3_wiener.pdf
|
|
Attachment 7: timeseries_elog.pdf
|
|
12404
|
Fri Aug 12 14:37:34 2016 |
Steve | Update | SEI | working seismometers as they are | 2.1 mag earth quake in Norhten Ca
Our seimometers need professorial centering. Related electronics must be checked too.
Quote: |
The saga has started here We have to give credit to the Boss who fixed it. The seismometers themself are not labeled yet.
Atm6 added on 8-12-2016 EX needed to be centered
Thanks to Max for the nice plost at summery pages
|
|
Attachment 1: eq_2.1m_Geysers_CA.png
|
|
Attachment 2: eq2.1m_GeysersCA.png
|
|
Attachment 3: oscillations.png
|
|
12402
|
Thu Aug 11 17:30:05 2016 |
Praful | Update | Electronics | Mic Amplifier | The results of my first huddle test were not so good- one of the signals did not match the other two very well- so I changed the setup so that the mics would be better oriented to receive the same signal. Pictures of the new setup are attached.


I also noticed some problems with one of my microphones so I soldered a new mic to bnc and switched it out. Just judging from Dataviewer, the signals seem to be more similar now. I'll be taking data for another few hours to confirm. |
12401
|
Thu Aug 11 11:56:40 2016 |
gautam | Update | SUS | ETMY re-suspended |
Quote: |
How much pitch bias do you need in order to correct this pitch misalignment?
That may give you the idea how bad this misalignment is.
|
I needed to move the pitch slider on the IFO align screen to -2.10 (V?) from 0 to get the HeNe spot to the center of the iris. The slider runs from -10V to 10V, so this is something like 10% of its range. I am not sure if it means anything, but the last saved backup value of this pitch slider was -3.70. Of course, application of the bias will affect all the coils, and when the optic is pitch balanced, the lower magnets are a little too far out and the upper magnets are a little too far in (see Attachment #1), as we expect for a downward pitch misalignment to be corrected. I suppose we can iteratively play with the coil positions and the bias such that the coils are centered and we are well balanced (maybe this explains the old value of -3.70).
I also checked that the side magnet can completely occlude its PD. With the damping on, by pushing the coil all the way in, the output of the side PD went down to 0. |
Attachment 1: pitchBalancingWithBias.PDF
|
|
12400
|
Thu Aug 11 11:51:38 2016 |
Praful | Update | Computer Scripts / Programs | Summary Pages | The summary pages have been updated with the new naming seismometer channel naming conventions. Here's a link to them working on my own page: https://ldas-jobs.ligo.caltech.edu/~praful.vasireddy/1154908817-1154909717/pem/seismic/
Let me know if the actual pages aren't working when they come back online or if there's something that needs to be changed. |
12399
|
Thu Aug 11 11:09:52 2016 |
Max Isi | Update | General | Summary pages status | This problem has been fixed.
> Summary pages are currently empty due to a problem with the code responsible for locating frame files in the cluster. This should be fixed soon and the
> pages should go back to normal automatically at that point. See Dan Kozak's email below for details.
>
>
> Date: Wed, 10 Aug 2016 13:28:50 -0700
> From: Dan Kozak <dkozak@ligo.caltech.edu>
>
>
> > Dan, maybe it's a gw_data_find problem?
>
> Almost certainly that's the problem. The diskcache program that finds
> new data died on Saturday and no one noticed. I couldn't restart it,
> but fortunately it's author just returned from several weeks vacation
> today. He's working on it and I'll let you know when it's back up.
>
> --
> Dan Kozak
> dkozak@ligo.caltech.edu |
12398
|
Thu Aug 11 00:20:41 2016 |
Koji | Update | SUS | ETMY re-suspended | How much pitch bias do you need in order to correct this pitch misalignment?
That may give you the idea how bad this misalignment is. |
12397
|
Wed Aug 10 23:45:03 2016 |
gautam | Update | SUS | ETMY re-suspended | Summary:
- ETMY has been re-suspended
- Reglued magnets (and also those that weren't knocked off) quite well with OSEM coils (see attachments)
- Pitch balance is off by ~2.8mrad (8mm over 1.5m lever arm) after inserting and centering OSEMs
- The same damping scheme used during the ETMX re-suspension process works reasonably well with ETMY as well
Details:
- I suspected that I had not tightened the wire clamp enough yesterday, and that the wire had slipped once the winches were removed
- Steve and I looked into the torque wrench situation today, and I realised that I had not been using the torque wrench correctly. What I thought were clicks indicating that the set torque has been reached was in fact just the sound the piece makes when going the opposite way relative to the direction set by the clip on the torque wrench. Anyways, the point is that while I thought I was tightening the screws with ~1.3Nm of torque, what was actually being applied was much less (although I don't have a good way to quantify how much less)
- So today I put the winches back on top of the tower, and winched the optic back up to the correct height using the ususal scribe line + microscope prescription
- I then tightened the wire clamp by hand. This is obviously not very repeatable, but it will have to do until we get a torque wrench with the correct range
- This seems to have done the trick - I did the tightening shortly after lunch, and after ~10 hours, there is no evidence of any wire sag
- I then proceeded to insert the OSEMs, first not all the way in to check the clearance available to the magnet, and once I was satisfied there was no danger of knocking anything off, went ahead and inserted the coils till the PD readouts were approximately half of the maximum (i.e. fully un-occluded) values. I used the OSEM coils originally on the ETMY tower, but all the other readout and drive electronics in the signal chain (satellite box included) belong to the ETMX setup (so as to avoid any cable routing over 80m from the Y end to the cleanroom). After some adjustment of the OSEM holding plates, I was able to center the magnets relative to the coils
- The tower only allows for a side OSEM to be inserted on one side. The other side does not have a threaded hole for a set screw. So we are forced to use the reglued magnet and not the side magnet that was not knocked off. By eye, it looks like the magnet may never completely occlude the LED, but the Striptool trace I was using to monitor the output of the PD did not yield any conclusive evidence. The optic was moving around a lot and I did not perform this check after turning the damping on
- I was able to damp the optic as well as we were able to damp ETMX on the clean bench (with the HEPA turned OFF). I had to turn the YAW gain down from 100-->75 to avoid some oscillations
- I then proceeded to check the pitch balance with the HeNe. The spot is low on an iris 1.43m away by ~8mm, which corresponds to a pitch misalignment of ~2.8mrad. I am not sure what to make of this - but perhaps its not unreasonable that we see this? Is there any record of what fine pitch balancing was achieved when the optic was put together back in 2010? This is also very sensitive to how far in/out the OSEM coils are, and though I've tried to center the coils as best as I can, I obviously have not done a perfect job...
What's next?
- Is the observed pitch imbalance a deal breaker? If so, I guess we need to re-glue a standoff?
- Are we willing to accept the side OSEM situation? (Tomorrow, I need to do a check to see what, if any, dynamic range we lose, with the damping on)
- If both the above are not problems we need to worry about, then:
- ETMY + ETMX -------> Vacuum bake on 22nd August (? - Bob also told me earlier today that he will try and put in some old turbo pump next week, and if that works, we could possibly get in the queue even before the 22nd)
- ETMY tower -------> Steve for sanding and removing wire grooves -------> Air bake
- ETMX tower -------> Air bake (provided the latest round of wire tightening has not left any grooves in the top piece of the tower, if it has, this needs to be cleaned up too)
- Some lengths of SOS wire (for re-suspending optics after bake) -------> Air bake
Attachments:
Attachment #1: Striptool trace showing all OSEM coils have been pushed in till the PD readout is approximately half the fully open value
Attachment #2: Pitch balance is off by ~2.8mrad (the Iris center is 5.5" above the table)
Attachment #3: UR magnet
Attachment #4: UL magnet
Attachment #5: LR magnet
Attachment #6: LR magnet
Attachment #7: SD magnet |
Attachment 1: ETMY_OSEMStrip.PDF
|
|
Attachment 2: IMG_2998.JPG
|
|
Attachment 3: IMG_3000.JPG
|
|
Attachment 4: IMG_3001.JPG
|
|
Attachment 5: IMG_3002.JPG
|
|
Attachment 6: IMG_3003.JPG
|
|
Attachment 7: IMG_3004.JPG
|
|
12396
|
Wed Aug 10 19:37:08 2016 |
gautam | Update | Electronics | Mic Amplifier | In order to help Praful do his huddle test, I have temporarily arranged for the outputs of the 3 channels he wants to monitor to be acquired as DQ channels at 2048 Hz by editing the C1PEM model. No prior DQ channels were set up for the microphones. Data collected overnight should be sufficient for Praful's analysis, so we can remove these DQ channels from C1PEM before committing the updated model to the svn. There is in fact a filter that is enabled for these microphone channels that claims to convert the amplified microphone output to Pascals, but it is just a gain of 0.0005.
In the long term, once we install microphones around the IFO, we can update C1PEM to reflect the naming conventions for the microphones as is appropriate. |
12395
|
Wed Aug 10 18:10:26 2016 |
Praful | Update | Electronics | Mic Amplifier | I set up 3 of my circuits in the interferometer near MC2 to do a huddle test. I have the signals from my microphones going into C1:PEM-MIC_1_IN1, C1:PEM-MIC_2_IN1, and C1:PEM-MIC_3_IN1. These are channels C17-C19. Here are some pictures of my setup:



I'll likely be collecting data from this for a couple of hours. Please don't touch it for now- it should be gone soon. There are some wires running along the floor near MC2 as well. |
12394
|
Wed Aug 10 17:30:26 2016 |
Max Isi | Update | General | Summary pages status | Summary pages are currently empty due to a problem with the code responsible for locating frame files in the cluster. This should be fixed soon and the
pages should go back to normal automatically at that point. See Dan Kozak's email below for details.
Date: Wed, 10 Aug 2016 13:28:50 -0700
From: Dan Kozak <dkozak@ligo.caltech.edu>
> Dan, maybe it's a gw_data_find problem?
Almost certainly that's the problem. The diskcache program that finds
new data died on Saturday and no one noticed. I couldn't restart it,
but fortunately it's author just returned from several weeks vacation
today. He's working on it and I'll let you know when it's back up.
--
Dan Kozak
dkozak@ligo.caltech.edu |
12393
|
Wed Aug 10 17:20:16 2016 |
Lydia | Update | General | Seismometer channel names changed | [ericq, Lydia]
We changed the seismometer channel names from, e.g. C1:PEM-SEIS_GUR1_X to C1:PEM-SEIS_EY_X.
- GUR1 referred to the seismometer at the Y end, GUR2 referred to the seismometer at the X end, and STS1 referred to the seismometer at the vertex. These have been renamed EY, EX, and BS respectively in all channel and filter names.
- The models for c1pem and c1oaf were changed in Simulink. The DAQ boxes were also updated with the new names. The script which forcibly renames channels saved to disk was edited to no longer refer to the GUR1, GUR2 etc channels. Going along with what Rana suggested we decided not to change the names of the renamed channels this way when saving, so the data saved from the seismometers can be found under e.g. C1:PEM-SEIS_EY_X_OUT_DQ.
- The filter files generated by Foton were changed to reflect the new channel names.
- We compiled the changes to the models and restarted the models on the relevant machines (c1lsc for the c1oaf model and c1sus for the c1pem model). c1sus_aux was down so we manually restarted it to turn off the watchdogs as a precaution before putting too much strain on c1sus.
- The MEDM screens now show the correct information, relabeled with the new names under PEM-RMS.
- The striptool display projected on the wall now shows the appropriate C1:PEM-RMS_BS channels and has been renamed to "SeismicRainbowBS.strip"
- We verivied that the new channels can be accessed live and the data from the DQ channels is saved to disk.
- After the changes were complete, we attempted to commiting to svn (the commit also included bringing the MEDM screen files into version control.) However the svn server was taking a long time to respond, so we will try again tomorrow to commit the file changes.
- There are still some lefotver unused channels with name sinculding STS_2 and STS_3 that refer to seismomters we no longer use. We left these alone.
Summary of new channel names:
C1:PEM-RMS_{BS, EX, or EY}_{X, Y, or Z} followed by the same filtering options as before, e.g. _BP_1_3_OUT
C1:PEM-SEIS_{BS, EX, or EY}_{X, Y, or Z}_{EXC, IN1, IN2, OUT, or OUT_DQ}
C1:OAF-WIT_{BS, EX, or EY}_{X, Y, or Z}_{EXC, IN1, IN2, or OUT}
|
12392
|
Wed Aug 10 15:34:24 2016 |
Steve | Update | SUS | 6 in-lbs torque driver for wire clamp screw | The 7.5 in-lb of Wiha seems at the upper end of torque range for a 4-40 SS screw
Wiha 28502 ordered with range 5 -10 in-lb for silver plated 4-40 screws
Do not trust the Venzo torque wrench under 2 Nm ! It miss lead me.
Recommended torque values for silver-plated fasteners are here. For aLIGO we use the guidelines in T1100066-v6, This doc is posted in 40m wiki under Mechanics also.
So, we'll use 6 in-lbs on silver plated 18-8 stainless steel socket head cap screw 4-40 x 3/8 into SS tower bridge.
Please replace these clamp screws every time if they were tightened without a torque wrench.
Quote: |
New Wiha 28504 torque wrench for SOS wire clamping. It's range 7.5 - 20 in-lb in 0.5 steps [ 0.9 - 2.2 Nm ] Audible and perceptible click when the pre-set torque has been attained at ±6% accuracy.
The new ETMX sus wire torqued to ~ 11.5 in-lb [1.3 Nm ]
Quote: |
Gautam and Steve,
The clamp's left side was jammed onto the left guide pin. It was installed slit facing left. Gautam had to use force to remove it. The clamp should move freely seating on the guide rods till torque aplied. Do not move on with the hanging of optic with a jammed clamp. Fix it.
Never use force as you are hanging - aligning optic. The clamp is in the shop for resurfacing and slit opening.
|
|
|
12391
|
Wed Aug 10 10:32:44 2016 |
Steve | Update | SEI | working Guralps as they are | The saga has started here We have to give credit to the Boss who fixed it. The seismometers themself are not labeled yet.
Atm6 added on 8-12-2016 EX needed to be centered
Thanks to Max for the nice plost at summery pages
|
Attachment 1: InterfboxOutpFixed.png
|
|
Attachment 2: GuralpsCoherences.png
|
|
Attachment 3: GurInterfBox.jpg
|
|
Attachment 4: EX.jpg
|
|
Attachment 5: EY.jpg
|
|
Attachment 6: EX_EY.png
|
|
12390
|
Wed Aug 10 03:08:03 2016 |
gautam | Update | SUS | ETMY patch-up | [lydia, gautam]
Rana felt it was alright to use the wire clamp and suspension cage in its existing condition for checking the ETMY magnet-OSEM coil alignment. So we set about trying to re-suspend ETMY. The summary of our attempts:
- Transferred optic from magnet gluing rig to the suspension cage
- Adjusted bottom EQ stops till the scribe lines on both sides were at 5.5" as verified with the microscope
- Looped cleaned length of wire around optic, attached free ends to winches, placed the wires under light tension by finger-pulling the slack out
- Lowered the bottom EQ stops
- Winched the optic to the right height
- Clamped the wire with the only wire clamp on this variant of the suspension cage. We used the same torque wrench at the same torque setting as was successful for ETMX. But after removing the winches, and releasing the face EQ stops, the optic seems to have sagged a lot - it now touches all the bottom EQ stops, and the more I lower it, the more it seems to come down. Perhaps it is the effect of the wire grooves in the cage, or that the wire-clamp itself is slightly different from the piece used on the ETMX cage, but 1.3Nm of torque doesn't seem to have tightened the wire clamp sufficiently
- We can still probably salvage the situation by re-attaching the winches to the top of the cage, setting the optic to the right height again, and clamping the wire clamp with more torque (as this is just a check to see that the reglued magnet configuration is compatible with the OSEM coil positions on the cage). Before air baking the cage, we will have the old wire grooves removed, and then suspend the optic with a fresh loop of wire after the bake
- We could not check the magnet-OSEM alignment because of the slipping of the wire through the clamp. We decided against pushing on tonight
- Optic is currently in the cage, resting on the bottom EQ stops and with all face EQ stops within 1mm of the optic. The OSEM coils have not been inserted into the holders
Regarding the vacuum bake of the optics: why do we want to do this again? Koji mentioned that the EP30-2 curing process does not require a bake, and there is also no mention of requiring a vacuum bake in the EP30-2 gluing guide. Is there any other reason for us to vacuum bake the optic? |
12389
|
Tue Aug 9 19:35:49 2016 |
rana | Update | SEI | both Guralp seismometers are functioning and being acquired | After some cable swapping, we now have both Guralp seismometers running and the times series and spectra look similar to each other and motley healthy.
Bean and I took a look at the whole situation today. Ben had nicely fixed the Dsub end of the EX cable (the EY one is still just a sad joke), After installing this newly fixed cable, we still saw no signals. There was some confusion in the control room about using the MED displays to diagnose seismometers: flickering MEDM values cannot be used for this. It would be like checking a pizza box temperature to determine if the pizza is any good.
- Although the +/- 12V LEDs on the front panel are dim, we confirmed that the produce 11.94V even when loaded with a seismometer. So its a LED circuit problem not a power problem.
- We were able to inject signals into the front panel with a breakout board and see them in DV for Input 1, but not Input 2.
- After Ben left, I kept poking around and found that the Guralp chassis output gets broken out into 3x3 BNC cables before going to the PEM BNC panel (and then on to the PEM ADC). This is where the problem was.
- The Input #3 BNC cables were connected to the long cables going to the 'GUR2' channels of the PEM. The Input#2 BNC cables were connected to some short BNC cables that were just hanging from the rack. So, somewhere during the debugging of the past N months, someone plugged this in wrong and didn't notice or forgot to switch it back. So all of the tests using DV or DTT or MEDM since that time have been invalid.
Tomorrow, Lydia is going to change all of the labels and channel names. The new names will be EX & EY to prevent this kind of huge waste of time with channel name swapping. That means no more illegal names with the label maker, Steve.
From the spectrum you can see that the EX seismometer (GUR2) is still not centered or at least its oscillating at 245 Hz for some reason. This should go away after some power cycling or recentering using the magic wand.
I noticed some anomalies in the mechanical setups at the ends:
- Some junk has been stored on top of the EX seismometer. Please never, even temporarily, store your power supplies, tools, or donuts
on top of the vibration sensitive sensors. Just put it on the floor and improve your carma.
- The EY seismometer has some fishy wires being fished between the can and the rubber seal. This is verboten. That seal must be flush to prevent pressure fluctuations and wires in there will ruin the smooth contact permanently. Temperature sensor wires must go through the grantie block feed-through or else its pointless.
- The flimsy insulation on the EY seismo is waay toooo mickey mouse. Real thermal insulation should be done using the yellow foam that Jenne used for the seismo huddle test. This flimsy silvery stuff is OK for making hats and mittens and beer cozy's, but its not research grade foam.
|
Attachment 1: 1goodday.png
|
|
12388
|
Tue Aug 9 16:19:27 2016 |
Steve | Update | SUS | torque driver for wire clamp | New Wiha 28504 torque wrench for SOS wire clamping. It's range 7.5 - 20 in-lb in 0.5 steps [ 0.9 - 2.2 Nm ] Audible and perceptible click when the pre-set torque has been attained at ±6% accuracy.
The new ETMX sus wire torqued to ~ 11.5 in-lb [1.3 Nm ]
Quote: |
Gautam and Steve,
The clamp's left side was jammed onto the left guide pin. It was installed slit facing left. Gautam had to use force to remove it. The clamp should move freely seating on the guide rods till torque aplied. Do not move on with the hanging of optic with a jammed clamp. Fix it.
Never use force as you are hanging - aligning optic. The clamp is in the shop for resurfacing and slit opening.
|
|
12387
|
Tue Aug 9 15:50:30 2016 |
Praful | Update | General | Guralp Cable | The Guralp cable has been reconnected and powered after having the connector changed out.
|
12386
|
Tue Aug 9 15:27:57 2016 |
gautam | Update | SUS | ETMY patch-up | The pickle pickers came off nicely and both magnets seem to be glued on okay. The alignment of the face magnets look pretty good, but we will only really know once we suspend the mirror, check the pitch balance, and put in the OSEM coils.
I brought the ETMY suspension tower + OSEM coils out of the vacuum chamber into the cleanroom. Given that the old wire had a pretty sharp kink in it, I removed it with the intention of suspending the optic with a new length of wire. I noticed a few potential problems:
Attachment #1 - ETMY tower is different from ETMX tower:
- The ETMY suspension seems to be of an older generation - it does not have the the two secondary wire clamps.
- The top piece was attached to the body of the tower using non-silver-plated screws. Steve tells me this is the wrong type, and we can switch these out when we put it back together.
- The wire clamp itself doesn't have much of a groove from the wire. But the wires have made asymmetric grooves in the tower itself (the left groove is deeper than the right as seen in Attachment #1), that are clearly visible. Should we get these grooves removed before attempting re-suspension? How do we want to remove it? Steve thinks the best option is to send it to the shop for milling, as there is hardly any room to rub sandpaper along the piece because of the pins, and these pins don't come out.
- Or do we just not care about these grooves for now, if we are planning to use new wire anyways after air-baking the towers?
- Steve thinks we should have a few spares of these top blocks handy (the latest version, with the secondary clamps), he wants to know if we should place an order for these (we already have 10 spare wire clamp pieces available for if/when we need them)
Attachment #2 - the base of the tower is significantly rusty:
- A few wipes with an acetone soaked rag yielded quite a lot of rust
- Steve thinks this is because the wrong type of stainless steel was used
- Does this have to do with the cage being of an older variety? After a few vigorous wipes, no more rust came off, but the rusting process will presumably keep generating new rust? Is this a concern? Do we want to change this piece before putting the tower back in?
I am holding off on attempting to re-suspend the optic for now, until we decide if the old wire grooves need to be removed or not. If we are okay with re-using the same piece as is, or if we are okay with using sandpaper and not the machine shop to remove the grooves, I will resume the re-suspension process.
Eric suggested another alternative, which is to use the old ETMX tower. I don't recall it being rusted, but this has to be checked again. The other problem of the wire-grooves would possibly still be an issue.
Regarding the vacuum bake of the ETMs, Bob tells us that the best case scenario we are looking at is September.
|
Attachment 1: IMG_2996.JPG
|
|
Attachment 2: IMG_2997.JPG
|
|
12385
|
Tue Aug 9 13:53:57 2016 |
babbott | Update | SEI | long Guralp EX cable repaired on the D-sub side | I checked out the cable that I took from you, and all of the connections looked right. The only thing I did notice was that some of the soldered wires on the 37-pin connector had gotten hot enough to melt their insulation, and potentially short together. I cut off that connector, and left it on your desk to check out. I put on a new connector, and checked the pinout. If the Guralps still doesn't work, we'll have to check out other possibilities. |
12384
|
Tue Aug 9 00:44:43 2016 |
gautam | Update | SUS | ETMY patch-up | Summary:
Given that ETMX looks to be in good shape and the optic and suspension tower are ready for vacuum and air bakes respectively, I set about re-gluing the knocked off magnet of ETMY. In my previous elog, I had identified the knocked off magnet as the UL magnet. But in fact, it was the LR magnet that broke off. This is actually one of the magnets that was knocked off when Johannes was removing the optic from the vacuum chamber. I have edited the old elog accordingly.
Step 1: Removing epoxy residue
- I used the teflon+glass rig Steve put together for this purpose
- After soaking for ~2 hours in acetone, I was able to remove approximately half of the ring residue by lightly pushing with a wipe.
- The other half wouldn't budge so I let it soak for another 4 hours
- After 6 hours of soaking, I was able to get all of the epoxy residue off - it doesn't simply dissolve in the acetone, I had to push a little with one of the cotton-tipped paddles in the cleanroom
- I gave the portion exposed to acetone a quick drag wipe with isopropanol. I didn't spend too much time trying to clean the AR side given that we will be using first contact anyways.
- I have not touched the HR side for now, even though a small portion of it was exposed to acetone. While cleaning the HR face with first contact, this portion can be inspected and cleaned if necessary
Step 2: Putting the optic in the magnet gluing jig
- I transferred the optic to the magnet gluing jig
- Given that we weren't touching any side magnets, I reasoned I did not have to go through the elaborate shimming routine to account for the wedge of the optic that we had to do in the recent past
- However, I did not think to put a thicker teflon spacer on the lower side of the wedge, and as a result, I knocked off the UR magnet as well as the jig did not have sufficient clearance
- Fortunately, the UR magnet came off cleanly, there was hardly any epoxy residue left on the optic. The UR magnet was NOT one of the magnets knocked off by Johannes while removing the optic from the vacuum chamber
- I gave the area formerly occupied by the UL magnet 3-4 wipes with acetone and then 1-2 wipes with isopropanol
- At this stage, I proceeded to re-insert the magnet-gluing jig. I used the two scribe lines on the outer side of the jig to fix the rotation of the jig, and used the remaining two attached face magnets to fix the overall position of the jig (by centering these magnets relative to the apertures on the jig). In order to center well, I had to unscrew the stuck silver plated screw on the jig by 1 turn
- Having arranged the jig satisfactorily, I proceeded to remove epoxy residue off the dumbbell of the recently knocked off UL magnet using first a razor blade, then sandpaper and finally made some new grooves with a razor blade. I then cleaned the surface of the dumbbell to be in contact with the optic with isopropanol. All of this was done for the LR magnet two weeks ago right after it was knocked off
Step 3: Gluing the magnets
- I prepared the magnets in the pickle pickers
- I discarded 1 full squeeze of the epoxy after it reached the tip of the mixing fixture, and then extracted another full squeeze of the gun for mixing and gluing the magnets
- I mixed the epoxy in an Al foil vessel for 3-4 minutes, and then placed a few drops on a piece of Al foil for a test bake at 200F for ~15 minutes
- The test bake went well, so I proceeded to apply glue to the dumbbells and re-glue the magnets to the optic
- The gluing was done around midnight, so we should be able to have a look at this post lunch tomorrow.
Provided the gluing goes well, the plan for tomorrow is:
- Bring ETMY suspension tower from the vacuum chamber to the cleanroom along with its OSEMs
- Suspend ETMY with a new length of wire (this should be much more straightforward than our ETMX exploits as both standoffs are already glued)
- Insert OSEMs, check that all 4 face magnets are well centered w.r.t. their coils and also that at least one side magnet is well aligned relative to its coil and can be used
- If step 3 goes well, then ETMY is also ready for a vacuum bake. I guess we can also air bake the ETMY suspension tower, there's plenty of room in the oven
|
12383
|
Mon Aug 8 10:35:12 2016 |
Steve | Update | PEM | vent 78 day 38 |
Quote: |
RGA background scan
Quote: |
Vacuum Status: Chamber Open
All chamber annuloses are vented. Vac Monitor screen is not communicating with gauges. The valve position indicator are working.
RGA is pumped by Maglev through VM2
|
|
|
Attachment 1: Vent38days.png
|
|
Attachment 2: vent78d40RgaBg.png
|
|
12382
|
Sun Aug 7 14:53:39 2016 |
ericq | Update | SUS | ETMX Standoff gluing was successful | I came in to check on ETMX. I freed the earthquake stops, and found that the OSEMS were reasonably, but not perfectly, centered. Turning on the damping, I found that the pitch balance is biased slightly downwards at about ~0.5mrad, which is acceptable.
As another check for how much we moved the standoff while gluing, we can look at the spectra of the OSEMS while the mirror is free swinging, and see if/how the resonance frequencies have moved around. As Gautam previously mentioned, the pitch frequency is even softer than we expected from the thicker ruby standoff alone. This is due to the excess glue around the guide rod forcing us to position the standoff even lower to have good contact with the optic's barrel. In the plot below, the design yaw/pit/pos frequencies are the dashed lines, and the measured frequencies are the solid lines.
[The plot is not in spectral density units, so that the peak heights reflect real units of motion at each resonance frequency. Data and code used to generate the plot is attached]

|
Yaw |
Pitch |
Pos |
Side |
Design frequencies from T000134: |
0.773 Hz |
0.856 Hz |
1.001 Hz |
|
ETMX Measurement in-air 2010 |
0.828 Hz |
1.04 Hz |
0.908 Hz |
0.949 Hz |
Pre-gluing |
0.785 Hz |
0.709 Hz |
0.949 Hz |
0.975 Hz |
Post-gluing |
0.789 Hz |
0.705 Hz |
0.953 Hz |
0.984 Hz |
According to the calculations from ELOG 12316, this pitch frequency implies the support point is 0.317mm lower than the design value of 0.985mm. (However, this is just an approximation and does not include the fact that each standoff is at a different height.)
Nevertheless, this difference is frequency is not so large that the dynamics of the suspension will be qualitatively changed in some important way; really, the pitch frequency is just ~1.5dB lower. So, I deemed our standoff gluing a success, removed the optic from the suspension, and placed it in an optic holding ring after giving the top of the barrel a gentle drap wipe with some iso. At this point, I used the microscope to look at the ruby standoff groove. As far as I can tell, no glue has invaded the groove - it looks sharp as ever. (whew)
I also wiped the wire with acetone and easily removed the glue droplets. However, I noted that (as is the case for ETMY) the wire is deformed at the points where it was in contact with the standoffs. I wonder if we should re-suspend with new wire, or accept the current deformed wires.
In any case, we can now move on to air baking the ETMX tower and gluing the stray magnet back onto ETMY. |
Attachment 1: ETMX_resonances.pdf
|
|
Attachment 2: ETMX_SUSspectra.zip
|
12381
|
Fri Aug 5 18:21:28 2016 |
gautam | Update | SUS | ETMX Ruby Wire Standoff GLUED |
- The latest twist in this apparently never-ending saga was that even though fine pitch balancing was achieved, the wire was out of the groove on both sides!
- I rectified this situation in the morning, did the fine pitch balancing in the afternoon
- Koji's suggestion of adjusting the OSEM holding plate totally did the trick, all four magnets are reasonably well centered relative to the vertical now...
- After the latest round of fine pitch balancing, we are now tilted in pitch backwards (i.e. towards the AR face) by <0.7mrad.
- Prior to gluing, I visually inspected the optic to check that (see attachments):
- Wires are in grooves on both sides
- Unglued ruby standoff has the correct "rotation", i.e. that the wire contacts the standoff after the groove has started, and leaves it before the groove ends, since the groove doesn't go all the way around the standoff
- Section of wire around the bottom half of the optic has no obvious kinks/other funny features
- Unglued standoff is in contact with the barrel
- All magnets are well clear of teflon in OSEM coils on both sides
- Eric also checked the frequencies of the various modes (PIT, YAW, POS and SIDE) by looking at the power spectrum of the free-swinging error signals on the coils. The pitch mode is now softer than before, at ~710mHz
- We then proceeded to glue the optic, using a needle to apply the glue (optic was clamped using face EQ stops, bottom EQ stops were not engaged as we felt this would affect the fine pitch balancing
- During the process, it looks like we may have inadvertently gotten some glue onto the wire (see attachments) - it doesn't look like any has seeped into the groove itself, but there is definitely some on the wire. We can possibly try cleaning this once the optic is out. In the worst case scenario, we will have to loop another section of wire, but the fine pitch balancing should be unaffected provided we did not perturb the optic too much
- Bob has said the large oven will be available to bake the cages on Tuesday, August 9th. By this time, we should have ETMY suspended as well (we were unable to glue the knocked off magnet on ETMY as the glass bowl we had for soaking the edge of the optic in acetone to remove the epoxy residue broke while I was assembling the various pieces of Teflon inside it. Steve is procuring a new one on Monday). It is still unclear when we can vacuum bake the two ETMs...
Attachments:
Attachment #1: Wire is in the groove in the unglued wire-standoff, groove rotation looks pretty good.
Attachment #2: Ruby standoff is sitting on the barrel of the optic (if you zoom in)
Attachment #3: Side magnet is well centered w.r.t OSEM coil
Attachment #4: UR magnet is well centered w.r.t OSEM coil
Attachment #5: UL magnet is well centered w.r.t OSEM coil
Attachment #6: LL magnet is well centered w.r.t OSEM coil
Attachment #7: LR magnet is well centered w.r.t OSEM coil
Attachment #8: Wire is in the groove in the glued Ruby standoff
Attachment #9: Standoff after gluing. 3-4 drops of epoxy are visible on the wire, but none looks to have seeped into the groove itself
Attachment #10: Side view of newly glued Ruby standoff
Attachment #11: Before and After gluing shots.
|
Attachment 1: IMG_2966.JPG
|
|
Attachment 2: IMG_2969.JPG
|
|
Attachment 3: IMG_2970.JPG
|
|
Attachment 4: IMG_2972.JPG
|
|
Attachment 5: IMG_2972.JPG
|
|
Attachment 6: IMG_2973.JPG
|
|
Attachment 7: IMG_2974.JPG
|
|
Attachment 8: IMG_2967.JPG
|
|
Attachment 9: IMG_2980.JPG
|
|
Attachment 10: IMG_2985.JPG
|
|
Attachment 11: BeforeNAfter.PDF
|
|
12380
|
Fri Aug 5 16:25:08 2016 |
Praful | Update | Electronics | Mic Amplifier | I took the spectrum of an EM172 connected to my amplifier inside and outside a large box filled with foam layers:

I also made a diagram with my plan for the microphone amplifier boxes. This is a bottom view:

The dimensions I got from this box: http://www.digikey.com/product-detail/en/bud-industries/CU-4472/377-1476-ND/696705
This seemed like the size I was looking for and it has a mounting flange that could make suspending it easier. Let me know if you have any suggestions.
I'll be doing a Huddle test next week to get a better idea of the noise floor and well as starting construction of the circuits to go inside the boxes and the boxes themselves.
|
12379
|
Fri Aug 5 09:38:12 2016 |
Steve | Update | SUS | SOS sus wire ordered | 0.0017" OD., 500ft steel music wire ordered. Pictures of the existing roll are below. It will be on 8" OD. spool too.
|
Attachment 1: 0.0017.jpg
|
|
Attachment 2: 0.0017spec.jpg
|
|
|