Thu Jun 25 12:03:42 2015 |
Max Isi | Update | General | Summary page status | The summary pages have been down due to incompatibilities with a software update and problems with the LDAS cluster. I'm working at the moment to fix the former and the LDAS admins are looking into the latter. Overall, we can expect the pages will be fully functional again by Monday. |
11374
|
Wed Jun 24 17:30:45 2015 |
ericq | Update | LSC | DFD Delay length | This afternoon, I had a fruitful conversation with Rich Abbott about various kinds of cables.
I've sent an email to Steve to ask him to buy 2 x 50m LMR-195 cables, with male SMA connectors. Rich highly recommends these for their polyethylene insulation, which makes them less microphonic and less susceptible to thermal expansion, low loss, and multi-ply bonded foil shielding.
50m means that the peak to peak mixer output swing corresponds to about a MHz. 1nV of mixer output noise looks like ~6mHz frequency noise, for a Level 3 mixer appropriately driven. As a comparison, the lowest our in-loop green PDH error signals get is 0.1Hz/rtHz.
The cable attenuation should be around 4.2dB at 50MHz, and 7.3dB at 150MHz, according to the data sheet. Thus, we should not be in the regieme where we are losing sensitivity to the attenuation.
By my rough geometric estimation, these two should fit in the 2U box I got ahold of today fine. Jessica is designing the front panel.
We currently have ~30m of RG-408 and RG-142 as our delay lines. |
11373
|
Wed Jun 24 08:05:43 2015 |
Steve | Update | PSL | laser turned on | The laser went off around 11am yesterday. It was turned on |
Attachment 1: laserWasOff.png
|
|
11372
|
Tue Jun 23 11:18:20 2015 |
Steve | Update | PEM | cleaning around chambers | Keven is our regular janitor is out for a few weeks.
The sub is careful- gentel Mario. We wiped arouind the vertex cambers north side on the floor and east arm racks.
|
11371
|
Mon Jun 22 20:59:19 2015 |
ericq | Update | LSC | DFD Delay length | I've been thinking a bit about what the ideal cable length / delay time for the upgraded ALS beatbox should be. Here are some thoughts, but no conclusions yet.
If you're not running your beatbox mixer in compression, there are two competing effects when you change the cable length. At first, more delay gives better sensitivity, but this does not go on to infinity, because cable attenuation eventually kills your signal. It turns out that the ideal length can be derived to be whatever length gives you 20/ln(10) = -8.7dB of attenuation. Frank found this out in PSL ELOG 825, and I found an HP document that derives this (and other useful DFD math) to the wiki, here.
In PSL ELOG 826, Frank calculated this ideal length for a 160MHz carrier in various kinds of cables.
However, this is not the end of the story. In the case of the DFD, we actually benefit from operating the mixer in compression, as makes our sensitivity less sensitive to flucuations in the beat amplitude. In this situation, the HP doc states "For maximum sensitivity, more delay can be added until the signal level out of the delay line is 8.7dB below the phase detector (mixer) compression point." I'm not sure I really understand the logic behind this statement, though.
Lastly, Koji mentioned the fact that the splitter in the demod board does not split at exactly 90 degrees, making the trajectory in the IQ plane an ellipse. This means that if the beat signal is moving around the ellipse a lot, or even wrapping around it, we can suffer from some nonlinear signal conversion. Also, if the raw DFD sensitivity is very high, the free swinging mirrors will cause the signal to swing around faster than the phase tracker can keep up. This should be easy to avoid, however; I doubt we will use so much cable that the beat would move by so much.
I intend to take all of this into account when picking a cable length! Jessica is going to help us make a nice box for them, too. |
11370
|
Mon Jun 22 14:53:37 2015 |
rana | Summary | LSC | X/Y green beat mode overlap measurement redone |
- Why is there a factor of 20 power difference? Some of it is the IR laser power difference, but I thought that was just a factor of 4 in green.
- Why is the mode overlap only 50% and not more like 75%?
- IF we have enough PSL green power, we could do the Y-beat with a 80/20 instead of a 50/50 and get better SNR.
- The FFD-100 response is more like 0.33 A/W at 532 nm, not 0.25 A/W.
In any case, this signal difference is not big, so we should not need a different amplifier chain for the two signals. The 20 dB of amplification in the BeatBox was a fine way, but not great in circuit layout.
The BBPD has an input referred current noise of 10 pA/rHz and a transimpedance of 2 kOhm, so an output voltage noise of 20 nV/rHz (into 50 Ohms). This would be matched by an Amp with NF = 26 dB, which is way worse than anything we could bur from mini-circuits, so we should definitely NOT use anything like the low-noise, low output power amps used currently (e.g. ZFL-1000LN....never, ever use these for anything). We should use a single ZHL-3A-S (G = 25 dB, NF < 6 dB, Max Out = 30 dBm) for each channel (and nothing else) before driving the cables over to the LSC rack into the aLIGO demod board. I just ordered two of these now. |
11369
|
Mon Jun 22 14:21:42 2015 |
Steve | Metaphysics | Treasure | Jenne and Den graduated | Last supper before departing at "Grazie" El Portal. All the best on your journey! |
Attachment 1: ls.jpg
|
|
11368
|
Mon Jun 22 12:57:09 2015 |
ericq | Summary | LSC | X/Y green beat mode overlap measurement redone | I took measurements at the green beat setup on the PSL table, and found that our power / mode overlap situation is still consistent with what Koji and Manasa measured last September [ELOG 10492]. I also measured the powers at the BBPDs with the Ophir power meter.
Both mode overlaps are around 50%, which is fine.
The beatnote amplitudes at the BBPD outputs at a frequency of about 50MHz are -20.0 and -27.5 dBm for the X and Y beats, respectively. This is consistent with the measured optical power levels and a PD response of ~0.25 A/W at 532nm. The main reason for the disparity is that there is much more X green light than Y green light on the table (factor of ~20), and the greater amount of green PSL light on the Y BBPD (factor of ~3) does not quite make up for it.
One way to punch up the Y beat a little might be to adjust the pickoff optics. Of 25uW of Y arm transmitted green light incident on the polarizing beamsplitter that seperates the X and Y beams, only 13uW makes it to the Y BBPD, but this would only win us a couple dBms at most.
In any case, with the beat setup as it exists, it looks like we should design the next beatbox iteration to accept RF inputs of around -20 to -30 dBm.
In the style of the referenced ELOG, here are today's numbers.
XARM YARM
o BBPD DC output (mV)
V_DARK: + 1.0 + 2.2
V_PSL: + 7.1 +21.3
V_ARM: +165.0 + 8.2
o BBPD DC photocurrent (uA)
I_DC = V_DC / R_DC ... R_DC: DC transimpedance (2kOhm)
I_PSL: 3.6 10.7
I_ARM: 82.5 4.1
o Expected beat note amplitude
I_beat_full = I1 + I2 + 2 sqrt(e I1 I2) cos(w t) ... e: mode overwrap (in power)
I_beat_RF = 2 sqrt(e I1 I2)
V_RF = 2 R sqrt(e I1 I2) ... R: RF transimpedance (2kOhm)
P_RF = V_RF^2/2/50 [Watt]
= 10 log10(V_RF^2/2/50*1000) [dBm]
= 10 log10(e I1 I2) + 82.0412 [dBm]
= 10 log10(e) +10 log10(I1 I2) + 82.0412 [dBm]
for e=1, the expected RF power at the PDs [dBm]
P_RF: -13.2 -21.5
o Measured beat note power (no alignment done)
P_RF: -20.0 -27.5 [dBm] (53.0MHz and 46.5MHz)
e: 45.7 50.1 [%]
|
11367
|
Sun Jun 21 13:21:03 2015 |
rana | Update | PEM | Wilcoxon Accelerometer Huddle Test | To improve our sensor noise estimate, the ACC cables should be clamped using a rubber pad and a big dog clamp on the SP tabletop before exiting the table. Also the sensors themselves should be covered with a foam box or a double cardboard box. The high-frequency acoustic noise may limit the 10 Hz performance since piezos are not very linear.
Quote: |
I've set up the Wilcoxen accelerometers on the SP table for a huddle test, to estimate their noise levels.
They're clamped down to the table along the same axis, and their housings are in good contact, in hopes to make them as correlated as possible.
|
Wilcoxon. Not Wilconox or Wilco or Wilcoxen. Have some pity on future keyword searchers. |
11366
|
Fri Jun 19 16:54:20 2015 |
Jenne | Update | Computer Scripts / Programs | Wiener scripts in scripts directory | I have put the Wiener filter scripts into /opt/rtcds/caltech/c1/scripts/Wiener/ . They are under version control.
The idea is that you should copy ParameterFile_Example.m into your own directory, and modify parameters at the top of the file, and then when you run that script, it will output fitted filters ready to go into Foton. (Obviously you must check before actually implementing them that you're happy with the efficacy and fits of the filters).
Things to be edited in the ParameterFile include:
- Channel names for the witness sensors (which should each have a corresponding .txt file with the raw data)
- Channel name for the target
- Folder where this raw data is saved
- Folder to save results
- 1 or 0 to determine if need to load and downsample the raw data, or if can use pre-downsampled data
- This should probably be changed to just look to see if the pre-downsampled data already exists, and if not, do the downsampling
- 1 or 0 to determine if should use actuator pre-weighting
- Data folder for measured actuator TFs (only if using actuator pre-weighting)
- Actuator TFs can be many different exported text files from DTT, and they will be stitched together to make one set of measurements, where all points have coherence above some quantity (that you set in the ParameterFile)
- Coherence threshold for actuator data (only use data points with coherence above this amount)
- Fit order for actuator transfer function's vectfit
- 1 or 0 to decide if should use preweighting filter
- zeros and poles for preweighting filters
- 1 or 0 to decide if should use lowpass after Wiener filters (will be provided corresponding SOS coefficients for this filter, if you say yes)
- Lowpass filter parameters: cuttoff freq, order and ripple for the Cheby filter
- New sample rate for the data
- Number of Wiener filter taps
- Decide if use brute force matrix inversion or Levinson method
- Calibrations for witnesses and target
- Fit order for each of the Wiener filters
I think that's everything that is required.
|
11365
|
Fri Jun 19 03:00:56 2015 |
ericq | Update | CDS | Library Model Parts examined | All simulink diagrams being used at the 40m are now under version control. I have compiled, installed, and restarted all current models to make sure that the files are all in a working state, which they seem to be. I have checked the latest version of the userapps svn repository to /opt/rtcds/userapps2.9 , to compare the files therein with our current state.
Surprisingly, only two files in the userapps svn have been changed since they were checked out here, and only one of these is a real change of any kind.
LSC_TRIGGER.mdl was edited at some point simply to align some drawn lines; no functionality was changed.
SCHMITTTRIGGER.mdl was edited to change the "INVERT" epics channel from an arbitrary EPICS input, to binary (true/false) input. This does not change the connectivity diagram, and in fact, I don't think we use this option in any of our scripts, nor is it exposed on our medm screens.
Thus, I think that the only place for block changes can bite us is changes in the fundamental blocks in CDS_PARTS that are used in our custom 40m library parts.
For posterity, these are the files used in compiling all of our running models. (Path base: /opt/rtcds/userapps/release )
/isc/common/models/CALIBRATION.mdl
/isc/common/models/FILTBANK_TRIGGER.mdl
/isc/common/models/LSC_TRIGGER.mdl
/isc/common/models/QPD.mdl
/cds/common/models/FILTBANK_MASK.mdl
/cds/common/models/lockin.mdl
/cds/common/models/rtbitget.mdl
/cds/common/models/rtdemod.mdl
/cds/common/models/SCHMITTTRIGGER.mdl
/cds/common/models/SQRT_SWITCH.mdl
/cds/c1/models/c1rfm.mdl
/cds/c1/models/c1tst.mdl
/cds/c1/models/c1x01.mdl
/cds/c1/models/c1x02.mdl
/cds/c1/models/c1x03.mdl
/cds/c1/models/c1x04.mdl
/cds/c1/models/c1x05.mdl
/isc/c1/models/ALS_END.mdl
/isc/c1/models/BLRMS_40M.mdl
/isc/c1/models/BLRMS_HIGHFREQ.mdl
/isc/c1/models/blrms.mdl
/isc/c1/models/c1als.mdl
/isc/c1/models/c1ass.mdl
/isc/c1/models/c1asx.mdl
/isc/c1/models/c1cal.mdl
/isc/c1/models/c1ioo.mdl
/isc/c1/models/c1lsc.mdl
/isc/c1/models/c1oaf.mdl
/isc/c1/models/c1pem.mdl
/isc/c1/models/IQLOCK.mdl
/isc/c1/models/IQ_TO_MAGPHASE.mdl
/isc/c1/models/PHASEROT.mdl
/isc/c1/models/RF_PD_WITH_WHITENING_TRIGGERING.mdl
/isc/c1/models/SENSMAT_LOCKINS.mdl
/isc/c1/models/TT_CONTROL.mdl
/isc/c1/models/UGF_SERVO_40m.mdl
/sus/c1/models/c1mcs.mdl
/sus/c1/models/c1scx.mdl
/sus/c1/models/c1scy.mdl
/sus/c1/models/c1sus.mdl
/sus/c1/models/lib/sus_single_control.mdl
/sus/c1/models/QPD_WHITE_CTRL_MUX.mdl
|
11364
|
Fri Jun 19 01:55:35 2015 |
ericq | Update | Green Locking | BeatBox Assay: not looking good |
Quote: |
But, it has no whitening. Can we do the whitening part externally? Perhaps we can run the RF signals from the output of the beat RF Amps over to the LSC rack and then put the outputs into the LSC Whitening board and acquire the signals in the LSC ?
|
I like this idea; it gives us more control over the whitening, and saves the IPC delay. We could use the currently vacant AS165 and POP55 channels.
We'd only have to move the phase trackers to c1lsc, which means 12 more FMs total. This is really the only part of the c1als model our current system uses, the rest is from before the ALS->LSC integration. |
11363
|
Fri Jun 19 01:24:26 2015 |
rana, koji | Update | Green Locking | BeatBox Assay: not looking good | We had decided a few days ago, to bypass the IF part of the BeatBox board and put some of the Haixing Maglev generic filter boards in there so that we could get more whitening and also have it be low noise.
Tonight we wondered if we can ditch the whole BeatBox and just use the quad aLIGO demod box (D0902745) that Rich gave us a few years ago. Seems like it can.
But, it has no whitening. Can we do the whitening part externally? Perhaps we can run the RF signals from the output of the beat RF Amps over to the LSC rack and then put the outputs into the LSC Whitening board and acquire the signals in the LSC ? |
11362
|
Wed Jun 17 15:31:50 2015 |
ericq | Update | PEM | Accelerometers fully installed | MC1 accelerometer has been plugged in. The modecleaner locking has been intermittent today, but I looked at a 15 minute lock in DTT, looking at the STS1 seismometer and both accelerometer triplets. Plot and DTT xml attached.
For the sake of not cluttering up everything with legends, the coherence plots are organized by direction (x, y, z), and include the coherence of each of the three sensors (sts, acc1, acc2) with the IMC control signal and the IMC transmitted RIN.
Some remarks:
- The 1 Hz pendulum motion is about equal amounts of X and Y, which makes sense, as MC1 and 3 are at an angle
- The ~3 Hz stack motion seems to be entirely in the X direction. Why?
- The bounce/roll bands are strongly coherent with Z motion at MC2.
- The STS does not appear to have any low frequency advantage over the accelerometers, in terms of coherence, contrary to what I would expect even without a thermal enclosure.
- The control signal and RIN RMSs are clearly dominated by noise in the 1-3Hz band, where we have reasonable coherence. Good prospects for noise subtraction...

|
Attachment 1: IMCcoherence_Jun172015.xml.zip
|
Attachment 2: IMCcoherence.png
|
|
11361
|
Mon Jun 15 22:36:40 2015 |
rana, koji | Update | Green Locking | BeatBox Assay: not looking good | Because the ALS beatbox schematic is out-of-date and misleading, we pulled the box to photograph the current implementation and figure out how to proceed. The box is out on the EE bench right now. Schematic Doc added to 40m Document tree: https://dcc.ligo.org/LIGO-D1102241. Some notes:
- The soldering on this board is pretty messy and there are a lot of flying wire and flying component hacks. I wouldn't trust all of the connections.
- The GV-81 RF amps in the front end are both stuffed. The 1 dB compression point is 19 dBm, so we want to use them below 10 dBm output. They have a gain of +10.5 dB, so that means they should not be used with and input to the beatbox of more than -10 dBm. Otherwise there will be nonlinear noise generation.
- Not stuffed: U1-Comparator, A1-attenuator, U2-splitter.
- Why is the filter after the mixer only 2nd order?? That's not a valid filter choice in any RF world. How much do we want to cut off the 2f mixer output before sending into our low noise, audio frequency (and prone to downconversion) amplifier? The Mini-Circuits amplifiers would have given us >60 dB attenuation in the stop band. This one is only going to give us 20-30 dB when the beat frequency is low. Get rid of diplexer. The schematic claims that its just one pole?? Seems like a 2nd order LP filter to me.
- The modified schematic (see Koji elog 8855) shows that an OP27 is used for the whitening stage. The current noise of the OP27 with the 3k resistor makes the OP27 current noise dominate below 1 Hz. And what is going on with that filter capacitor choice? We never want to use these tiny things for sensitive filter applications. (cf. Sigg doc on resistor and capacitor choice, the noise reduction book by Ott, H&H, etc.). That's why we have the larger metal-poly, paper, mylar, etc. caps sitting around.
Probably we ought to install a little daughter board to avoid having to keep hacking this dead horse. Koji has some of Haixing'g maglev filter boards. Meanwhile Koji is going to make us a new beatbox circuit in Altium and we can start fresh later this summer.
Interesting link on new SMD cap technology.
Photos of circuit as found |
11360
|
Mon Jun 15 20:36:48 2015 |
rana | Update | IOO | IOO QPDs centred | after re-aligning the beam into the PMC, I touched up the steering mirros into the IOO QPDs so that the beams are now centered again. Please don't adjust these references without prior authorization and training.
This plot shows the 10-minute trends for these QPDs over the last 400 days. |
Attachment 1: Untitled.png
|
|
11359
|
Mon Jun 15 16:55:39 2015 |
ericq | Update | PEM | Accelerometers installed | The accelerometers have been installed at MC1 and MC2. MC2 data is live, I haven't yet run the cables from the MC1 set to the preamp yet, though.
 

|
Attachment 1: MC2.jpg
|
|
Attachment 2: MC1.jpg
|
|
Attachment 3: mc2accspectra.png
|
|
11358
|
Mon Jun 15 11:54:44 2015 |
ericq | Update | CDS | Parts not in SVN | I ran the following command to find which models/parts are not under version control, or have modifications not commited:
grep "mdl" $(cat models.txt) | awk '{print $NF}' | sort | uniq | xargs svn status
models.txt includes lines like "/opt/rtcds/caltech/c1/rtbuild/c1ass.log" for each running model. These are the build logs that detail every file being sourced.
The resultant list is:
? /opt/rtcds/userapps/release/cds/common/models/BLRMS_HIGHFREQ.mdl
C /opt/rtcds/userapps/release/cds/common/models/rtdemod.mdl
M /opt/rtcds/userapps/release/cds/common/models/SCHMITTTRIGGER.mdl
? /opt/rtcds/userapps/release/isc/c1/models/blrms.mdl
? /opt/rtcds/userapps/release/isc/c1/models/IQLOCK.mdl
? /opt/rtcds/userapps/release/isc/c1/models/PHASEROT.mdl
? /opt/rtcds/userapps/release/sus/c1/models/QPD_WHITE_CTRL_MUX.mdl
I will commit the uncommited c1 parts, and think about what to do about the rtdemod and SCHMITTRIGGER parts.
Once I check out the latest revision of the userapps repo (in a seperate location), I will do something similar to look for models that have changed since the svn revision that is checked out in our running system. |
11357
|
Sat Jun 13 23:52:14 2015 |
ericq | Update | LSC | Beatbox needs whitening gain | Nice find!
We ought to use our noise model of the ALS signal chain to determine what the right gains are, rather than hunt and peck. More likely we'll start from the right gains.
Once the gains and/or whitening filters make sense, maybe we'll see some effect from fixing the green PDH loops. |
11356
|
Sat Jun 13 18:37:46 2015 |
ericq | Update | LSC | Beatbox needs whitening gain | Here are the promised details!
I was worried about the lack of whitening gain, as I saw that the DC phase tracker Q output (which is the magnitude of the signal in the beatbox's I-Q plane) was no more than 200 ADC counts for X (~120mV), and 800 for Y (~500mV). I.e. this is the largest DC value that either the I or Q ADC channels can see, and the RMS fluctuations are on the order of mV, meaning we're wasting our entire ADC range. 
However, I also had doubts about this since, even in the nominal state, the ASD of the ADC signals before dewhitening was higher than the expected ADC noise level. However, because of the non-linearity of the conversion of the BEAT_I and Q signals into the phase tracker output, evaluating the contribution of the beatbox output and ADC input voltage noises takes a few more steps.
So, I hooked up a Marconi as the signal source for the beatbox's X channel , with no modulation (presumably the phase noise of the Marconi output is significantly lower than the sensitivity of the beatbox). For all of these measurements, the beat frequency was kept around 50MHz, with an amplitude of -30dBm on the control room analyzer, which is a typical X ALS operating point.
At this point, the beatbox output noise was below the ADC noise (as measured by an SR785). Nevertheless, I found that the beat spectrum driven by the Marconi lined up to be very close to the ALS beat spectrum across a wide band, explaining much of the noise.

At this point, I inserted SR560s in between the beatbox I and Q outputs, and the AA chassis leading to the ADC. A gain of 10 reduced the resultant phase tracker noise by that same factor at nearly all frequencies. A further increase in gain did not lead to the noise changing appreciably, probably because the real beatbox noise was now contributing, as is suggested by some common peaks in the direct beatbox output phase tracker spectra.
Going back to the real green beat signal with the SR560s still at G=10, I obtained the result shown in ELOG 11355. I will soon repeat this process with the Y ALS.
As I mentioned in the previous ELOG, we may be further helped by more whitening gain than can be provided by the SR560s (and we obviously need a robust long term circuit for this gain). If it then turns out we are limited by beatbox noise to a degree we are not happy with, we could perhaps look into reintroducing some RF gain into the X beat. As Koji mentions in ELOG 8855, the beatbox operates best at an RF input of around -4dBm. |
Attachment 1: ADCdiagnosis.png
|
|
11355
|
Fri Jun 12 17:09:58 2015 |
ericq | Update | LSC | Beatbox needs whitening gain | Short entry just to preview a new development; more detail about this investigation will soon follow.
The beatbox I and Q signals are too small at the ADC! I was able to reduce the RMS out of loop ALS sensitivity (arm locked on POX) by 300Hz with G=10 SR560s between the beatbox output and the ADC whitening chassis input. Increasing the beat note amplitude via RF amplifier had no positive effect.

There is still a reasonable gap between this and the beatbox's noise levels, as measured with a marconi. There may be additional headroom for whitening gain; the SR560 maximum output range is smaller than the ADC input range.
The high frequency noise has >0.5 coherence with the PDH error signal above a kHz or so, but not much below that.
We should probably either modify the output whitening of the beatbox, or introduce some (variable?) whitening gain in a seperate circuit. |
Attachment 1: beatGain.png
|
|
11354
|
Fri Jun 12 08:40:17 2015 |
Steve | Update | VAC | Vacuum comp. rebooted | Koji and Steve,
One computer expert and one vacuum expert required.
Quote: |
The serial connections to the vacuum gauges were recovered by rebooting c1vac1 and c1vac2.
Steve claimed that the vacuum screen had showed "NO COMM" at the vacuum pressure values.
The epics connection to c1vac was fine. We could logged in to c1vac1 with telnet too although c1vac2 had no response.
After some inspection, we decided to reboot the slow machines. Steve manually XXXed YYY valves (to be described)
to prepare for any possible unwanted switching. Initially Koji thought only c1vac2 can be rebooted. But it was wrong.
If the reset button is pushed, all of the modules on the same crate is reset. So everything was reset. After ~3min we still
don't have the connection to c1vac1 restored. We decided to another reboot. This time I pushed c1vac1 reset button.
After waiting about two minutes, the ADCs started to show green lights and the switch box started scanning.
We recovered the telnet connection to c1vac1 and epics functions. c1vac2 is still note responding to telnet, and
the values associated with c1vac2 are still blank.
Steve restored the valves and everything was back to normal.
|
Atm 1, problem condition: gauges are not reading for a week, error message "NO COMM" and all computer LEDs are green
Atm 2, prepare to safe reboot:
a, close V1, disconnect it's power cable and turn off Maglev, wait till rotation stops
b, close PSL shutter ( take adrenaline if needed )
c, close V4, V5, VA6 valves and disconnect their cables. "Moving" error message indicating this condition.
V1 is not showing "Moving" because its power cable disconnected only! It will show it if its position indicator cable is disconnected too. There is no need for that.
These valves closed and disabled will not allow accidental venting of main volume.
d, push reset, reseting c1vac2 will reset c1vac1 also, wait ~ 6 minutes
"Vacuum Normal" valve configuration was restored after succesful reboot as follows:
a, reconnect cable and open V4 and V5 at P2 & P3 <1e-1 Torr
b, observe that P2 < 1e-3 Torr and retsart Maglev
c, wait till Maglev reaches full speed of 560 Hz and reconnect-open V1
d, reconnect-open VA6 at P3 <1e-3 Torr
NOTE: VM1 valve was locked in open position and it was not responding before and after reboot
Error message on Atm2 is indicating this locked condition: "opening VM1 will vent IFO"
This is a fauls message. The valve is frozen in open position. We need a softwear expert help.
|
Attachment 1: vacMonNoGauges.png
|
|
Attachment 2: prepReboot.png
|
|
11353
|
Thu Jun 11 19:40:59 2015 |
Koji | Update | VAC | Vacuum comp. rebooted | The serial connections to the vacuum gauges were recovered by rebooting c1vac1 and c1vac2.
Steve claimed that the vacuum screen had showed "NO COMM" at the vacuum pressure values.
The epics connection to c1vac was fine. We could logged in to c1vac1 with telnet too although c1vac2 had no response.
After some inspection, we decided to reboot the slow machines. Steve manually XXXed YYY valves (to be described)
to prepare for any possible unwanted switching. Initially Koji thought only c1vac2 can be rebooted. But it was wrong.
If the reset button is pushed, all of the modules on the same crate is reset. So everything was reset. After ~3min we still
don't have the connection to c1vac1 restored. We decided to another reboot. This time I pushed c1vac1 reset button.
After waiting about two minutes, the ADCs started to show green lights and the switch box started scanning.
We recovered the telnet connection to c1vac1 and epics functions. c1vac2 is still note responding to telnet, and
the values associated with c1vac2 are still blank.
Steve restored the valves and everything was back to normal. |
11352
|
Wed Jun 10 15:54:14 2015 |
Steve | Update | VAC | Vacuum comp. rebooted | Koji and Steve succeded rebooting C1vac1, C1vac2 and pressure reading is working now
More tomorrow .........
|
Attachment 1: afterReb061015.png
|
|
11351
|
Wed Jun 10 03:19:58 2015 |
ericq | Update | LSC | AUX PDH error measured in CDS | Looking over the old noise budget in the green locking paper, it seems the main technical noise sources were the AUX PDH error and DFD noise. I'm working on quantifying the current state of these noises.
Rather than lugging out the analyzers every time I wanted to make a measurement of the AUX PDH error signals, I set out to make the existing digitized channels (ALS-[X/Y]_ERR_MON) usable for easier, and continuous, monitoring. Sadly, up until now the signals were poorly conditioned, and drowning in ADC noise. (When locked, the Y error signal was only +-10 counts!)
Of course, given the bandwidth of the green servos (10kHz), this won't tell us the full story of the what the green PDH lock is doing, but does indicate how much residual frequency noise exists in the ALS control band.
I'm currently using SR560s at G=20 at each end to amplify the ERR MON outputs of the uPDH boards before sending them to the ADCs; now that I've found a gain that works, I'll modify the error point monitor buffer opamps inside the uPDH boards themselves during the daytime.
The AUX Y error signal was going into an AA board with some funky filtering going on that I didn't want to mess with. Instead, I've moved the signal to the pentek generic board whose first four channels are used for the oplev segments, and the second quartet are unused, save for the TST channel I hooked up yesterday.
On the pentek board, I changed the 4th order 800Hz lowpass to a 4th order 8kHz lowpass on the last three channels through some resistor swapping. (At first it was just the last two, but I found I was getting weird signals in the 32nd channel; and if I recall correctly from my cymac work, the 32 ADC channel is used for some timing signal or something...). I also turned off the 1:10 whitening filters on the last four channels via PCB jumper.
I then unhooked the PZT drive and let the PDH error signals swing around, to calibrate the ADC counts into HZ. Now, the ALS-[X/Y]_ERR_MON_OUT_DQ are calibrated in green Hz! Here are the spectra.

As we've seen in the past (ELOG 10464), the X green is limited by the dark noise of the PD from 10-100Hz. This isn't so great. The RMS noise from 300Hz downwards (which is maybe the band where the ALS control would inject noise into the mirror motion) is about Y:10Hz X:40Hz.
During this time, the test masses were not under any longnitudinal control, so I'm not sure why there is such a difference in the height of the suspension resonance peaks, unless there's some differene in the low frequency PDH TFs that I've forgotten about.
Now, with these references, we can easily check if the PDH loops change qualitatively over longer time periods.
I'll be including the effect of these noises in the upcoming revised ALS noise budget. |
Attachment 1: AUXspectra.png
|
|
11350
|
Wed Jun 10 02:50:39 2015 |
ericq | Update | PEM | Wilcoxon Accelerometer Huddle Test | Here are some results for the 3-corner hat subtraction for the six accelerometers, from ~1 hour of data that didn't look to have any sharp features/glitches from human activity in the lab.
I used the python uncertainties package to try and get an estimate of the uncertainty in the subtracted noise floor, by taking into account every possible possible combination of 3 sensors and the fluctuations in the spectrograms of the subtracted signals. I've attached the python code if anyone is interested in the implementation.
I pulled out the accelerometer data sheets to read their slightly varying V/g calibration (which differs on the order of a few percent from unit to unit). This improved the subtraction factor from ~20 to over 100 at some frequencies. I've edited the filter modules such that the OUT_DQ channels are already calibrated into m/s^2.

|
Attachment 1: hats2Acc.png
|
|
Attachment 2: 3hatcode.zip
|
11349
|
Tue Jun 9 10:57:12 2015 |
Steve | Update | SUS | BS oplev laser tuned | Lenses removed from oplev beam path at elog entry 11246
|
Attachment 1: 100dBStrend0425.png
|
|
Attachment 2: 18dBSdrift.png
|
|
11347
|
Mon Jun 8 15:51:31 2015 |
ericq | Update | CDS | RCG Diagnostics | I've started making some model changes for RCG diagnostic tests.
I put some blocks down in C1TST and C1RFM to test the delays of all-digital loops and one loop with a direct DAC->ADC (which currently uses a janky 1-pin lemo -> BNC -> 2-pin lemo situation (which will be improved)).
Here's what C1TST looks like now.

I've taken TFs of all three loops. The all digital loops are flat on the order of microdBs.
The delay in loop A (single loop, one model) is consistent with one 16k cycle, plus or minus 0.25 nsec.
The delay in loop C (single loop, two models connected via RFM) is consistent with two 16k cycles, plus or minus 0.5 nsec.
I haven't yet grabbed the whitening and AA/AI shapes for loopB, to calibrate the real delay.
All of these files currently live in /users/ericq/2015-06-CDSdiag, but I'll make somewhere outside of the user directory to collect all of these tests soon. |
Attachment 1: newTST.png
|
|
11346
|
Fri Jun 5 11:59:59 2015 |
ericq | Bureaucracy | General | Maintenance Tasks, IFO upgrades | At wednesday's meeting, Rana, Koji, Steve, and I started making a list of maintenence tasks that should be done/checked on a regular basis. The actual scheduling of these has not yet been considered. They include:
- N2 Tank pressure / cylinder replacement
- Headlamp, walkie talkie battery recharge
- Workstation software updates
- Coffee bean and filters
- Multimeter battery levels
- Sorensen DC power supply voltage settings and current draws
- UPS' status (Vacuum, NFS host, workstations)
- SR560s, battery powered scopes plugged in
- Rack Fuses intact
- Take pictures of electronics racks, optical tables
- Replace PSL HEPA filter
Next, we brainstormed work that can be done to improve the interferometer performance, and what order/precedence they should take.
In the end, it was decided that the plan for the next few weeks was to focus on improving the ALS noise levels, and, more importantly, seeking to make the performance more consistent. We need to know what is limiting us, and what we extent we can expect to improve things. To this end, I am working on reviving a ALS noise budget; using the noise budget from the green locking paper to inform a simulinkNB kind of thing.
Here are all of the items we listed during this brainstorming session.
Some near-term/priority tasks are:
- Installing the accelerometers near MC1 and 2
- Installing green steering PZT mirrors at the Y end table, commission dither alignment
- Improving the X end green mode matching
- ALS noise budgeting
- Upgrade the realtime system to RCG v2.9
More down the line, other things we thought about were:
- Cleanup of bench power supplies (FSS box has one, where else?)
- Fixing the ETMX suspension issues
- Upgrade SOS suspension code to the appropriate aLIGO block
- Upgrade to the green PDH electronics
- Understanding / tuning the FSS servo laser PZT vs. PC crossover
- Undestanding / tuning the 11MHz vs 55MHz modulation phase
- Replacing the slow vxworks machines with the Acromag setup Aiden has set up for the CTN lab
- QPD upgrade
- New/better green beatbox
- Finalize the manifestation of the IR beat control (Freq counter vs. fast DFD)
- Explore the idea of using an analog output of the ALS beatbox as fast input to the CM board
- Replace triple resonant EOM driving circuit with double resonant one
- X end table layout/enclosure upgrade
|
11345
|
Wed Jun 3 17:04:08 2015 |
ericq | Update | PEM | Wilcoxon Accelerometer Huddle Test | I've set up the Wilcoxon accelerometers on the SP table for a huddle test, to estimate their noise levels.
They're clamped down to the table along the same axis, and their housings are in good contact, in hopes to make them as correlated as possible.

Steve helped me move the DAQ cables from under the BS/PRM oplev table, to dropping from the cable tray above the POX table, so I could get them plugged in at the SP table. This also helps reduce the rats nest by the access connector, and is a fine location for when the accelerometers are attached at MC1 & MC2.
A quick glance at DTT shows coherence of >0.9 from about 2-100Hz for all six. A three-corner-hat deal will provide an easy estimate of the noise floor of each one. If we feel like being fancy / accounting for possible gain differences, we could try some MISO wiener action, but that is probably overkill. |
Attachment 1: AccHuddle.jpg
|
|
11344
|
Wed Jun 3 11:55:52 2015 |
Steve | Update | Optical Levers | PRM-BS oplev | I'm getting ready change the Newport Ultima U100-AC to SS-Polaris-K1 LOW DRIFT MIRROR MOUNTS
Note: there is only one lens in the PRM lunching path ( only realized later ) , so the spots are large ~ 3 mm at PRM qpd and ~4.5 mm at BS qpd
The spots are well centered.
Atm3, the spots were well centered yesterday ( the PRM is misaligned in pitch and retsore does not work today )
|
Attachment 1: PRM-BS_oplEr_noLenses.png
|
|
Attachment 2: PRM-BSopl_3jun2015.jpg
|
|
Attachment 3: opServos.png
|
|
Attachment 4: 10d_trend_opl.png
|
|
Attachment 5: opl_trend_40d.png
|
|
11343
|
Tue Jun 2 21:22:07 2015 |
rana, koji | Configuration | IOO | AOM inserted in beam and aligned | We spent an hour today to put the AOM back in the beam before the PMC and verified that the diffraction is working.
- The fuse holder was missing from the rack. We inserted a 5A fuse. We expect that the quiesscent draw is < 0.5 A. The power is from the +24V Sorensen supply.
- The alignment was tricky, but we optimized it as well as we could in translation and the RZ direction. Its a fixed mount still.
- We noticed that according to the datasheet, the polarization is wrong! It wants S-Pol light and we're giving it P-Pol. How come no one noticed this? We expect that the efficiency is reduced because of this. We (Steve) need to brainstorm what kind of mount we can use there to mount it at 90 deg to the plane of the table.
- The lens after the AOM has f = +400 mm. The distance from the AOM to the lens is ~800-900 mm so its not so terrible. However, if someone were to put the AOM halfway between the turning mirrors there, the beam diffraction would be canceled.
- The AOM input impedance seems to be 50 Ohm as advertised. The previous Koji entry claim of 25 Ohm is mysterious. We checked the Ohmage by sending a signal into the AM input of the AOM using the DS345 which as a 50 Ohm output. 1 Vpp from the DS345 made 1 Vpp on the input of the AM input as measured by Oscope connected by T with high impedance setting.
- With 0.5 V offset and a 1 Vpp signal, we get ~20-25% modulation of the power.

- We have left it running with a 4444.4 Hz modulation and a small amplitude. This is to see if we can use this to measure the cavity poles of the MC and the arms.
- We noticed some hash on the Teed input monitor. It was backstreaming of the RF drive. Whoever uses this thing in an ISS feedback ought to make sure to put an RF choke between the servo and this AOM driver.
We also removed a 50/50 pickoff mirror which was used to take one of the NPRO -> EOM polarizer reject beams and send it across the table into a floppy dump. Its now hitting a closer floppy dump. Let's stop using these crappy anodized aluminum flappers anywhere, Steve.
We also noticed that the PMC REFL path uses a W1 from CVI to send the PMC reflection to the REFL RFPD. The dim beam from the AR coated surface is being used rather than the bright beam from the uncoated surface. Ooops. Steve, can you please order another W1 for 1064 from CVI, but get it with a 2-3 deg wedge angle? This one has a wedge which is too small. |
11342
|
Mon Jun 1 20:05:36 2015 |
rana | Update | CDS | RCG Upgrade Imminent | 
Quote: |
Additionally, we want to have some set of tests and diagnostics, to make sure we have not introduced unwanted behavior.
To this end I will create some test models and DTT templates, where a series of measurements can be run, like
- OLTF/delay measurement of a single all-digital loop within one model
- OLTF/delay measurements of a few all-digital loops split across two models, using IPC communcation, RFM, dolphin
- Hook up DAC -> resistor/amplifier/??? -> ADC, to check things like DAC output, ADC noise levels, IOP delays.
I'll run these test before touching anything, and make sure I understand all of the results, so that an apples-to-apples comparison can be made after the upgrade is complete.
|
I got goosebumps just imagining this. |
11341
|
Mon Jun 1 15:22:19 2015 |
Steve | Update | PEM | Guralps X- short cable is bad |
Quote: |
Koji and Steve,
The result: bad Guralp x-arm cable.
I will swap the short cables tomorrow at the base.
|
Short 46" long cables at the base plates were swapped. Their solderings looked horrible.
This cable actually worked at 5-5-2015
Bad cable at ETMY station now. The new cable should be a little bit longer ~52" |
Attachment 1: seismGur1-2.png
|
|
11340
|
Mon Jun 1 10:26:53 2015 |
ericq | Update | CDS | RCG Upgrade Imminent | We are planning on upgrading the 40 CDS system to the latest version of the LIGO RCG software, in three weeks when Jamie is back in town.
Jamie and I spoke last week, to hash out a general plan for upgrade and preperations I can make in the meantime.
Preparation of our models for the upgrade includes;
- Check if any of the default RCG parts (filter modules, etc.) have substantially different default behavior / ports
- Cleaning up unterminated/hanging connections in the diagrams (Jamie tells me RCG is more strict about this now)
- Going through all of the build logs for our models to find what custom blocks are being pulled in from the userapps svn
- Confirm all of our running blocks and models are committed to svn
- In a new, isolated, folder, checkout the latest version of the userapps repo
- See what blocks have changed, and what model changes might be neccesary.
Once we think we know what needs to be changed in our models, we can check out the latest version of the RCG source, without linking it as the active version, and create a new build directory, without touching the old one, and create new copies of the 40m models with the neccesary modifications. This way, we can work on getting all of the 40m models compiled, without touching any of the live, running, systems.
Once our models are compiling successfully, we can work on building daqd, nds, mxstream, etc.
Additionally, we want to have some set of tests and diagnostics, to make sure we have not introduced unwanted behavior.
To this end I will create some test models and DTT templates, where a series of measurements can be run, like
- OLTF/delay measurement of a single all-digital loop within one model
- OLTF/delay measurements of a few all-digital loops split across two models, using IPC communcation, RFM, dolphin
- Hook up DAC -> resistor/amplifier/??? -> ADC, to check things like DAC output, ADC noise levels, IOP delays.
I'll run these test before touching anything, and make sure I understand all of the results, so that an apples-to-apples comparison can be made after the upgrade is complete.
Updates will be posted as I hash things out. I'm sure we have not yet thought of everything to think about and test, so ideas and feedback are very welcome. |
11339
|
Mon Jun 1 08:32:14 2015 |
Steve | Update | SUS | PRM damping restored | Local earthquake 3.8 Mag tripped only PRM
Vac monitor is not communicating.
PSL HEPA turned on
|
Attachment 1: indio3.8Meq.png
|
|
11338
|
Fri May 29 15:12:39 2015 |
Koji | Update | Computer Scripts / Programs | Chiara Backup Hiccup | Actual data |
Attachment 1: backup_hours.pdf
|
|
11337
|
Fri May 29 12:49:53 2015 |
Koji | Update | Computer Scripts / Programs | Chiara Backup Hiccup | In fact, the file access is supposed to be WAY faster now than in the RAID case.
As noted in ELOG 9511, it was SCSI-2(or 3?) that had ~6MB/s thruput. Previously the backup took ~2hours.
This was improved to 30min by SATA HDD on llinux1.
I am looking at /opt/rtcds/caltech/c1/scripts/backup/rsync.backup.cumlog
In fact, this "30-min backup" was true until the end of March. After that the backup is taking 1h~1.5h.
This could be related to the recent NFS issue? |
11336
|
Fri May 29 11:28:42 2015 |
ericq | Update | Computer Scripts / Programs | Chiara Backup Hiccup | I've changed the chiara local backup script to read a folder exclusion file, and excluded /users/public_html/detcharsummary, and things are working again.
This was neccesary because the summary pages are being updated every half hour, which is faster than the time it takes for the backup script to run, so the file index that it builds at the start becomes invalid later on in the process.
Thinking about chiara's disk, it strikes me that when we went from the linux1 RAID to a single HDD on chiara, we may have tightened a bottleneck on our NFS latency, i.e. we are limited to that single hard drive's IO rates. This of course isn't the culprit for the more recent dramatic slowdowns, but in addition to fixing whatever has happened more recently, we may want to consider some kind of setup with higher IO capability for the NFS filesystem. |
11335
|
Fri May 29 02:05:08 2015 |
ericq | Update | LSC | End Laser temperatures set | Both green beatnotes have been found with nominal amplitudes. (X: -30dBm Y: -20dBm), at temperatures which don't seem to be prone to mode hopping.
X = 42.64, Y ~40.15
Both arms can lock on ALS, but as Koji mentioned in ELOG 11334, the ALS noise seems anomalously high.
Details
The temperatures I posted in the previous log ended up not being so useful. To find the right end laser temperatures, I looked at the IR beatnotes on the control room analyzer out to 1GHz, and swept around the SLOW_SERVO2_OFFSET channels to change the laser temperature. During this time, the end green shutters were closed, the PSL shutter was closed, the FSS slow servo was off, and the FSS_SLOWDC was set to 0.0. (The green PSL shutter had to be open, because the IR beat fiber is coupled after it.)
For each arm, I found three temperatures where an IR beat could be observed; as Koji mentioned on Wednesday, we should use the middle mode. For each of the arms, scanning around from the middle to move the beat by +-1GHz did not cause a mode hop - the beat stayed visible on the scope. Once I found a real IR beat for the X arm, I took at look at the RF output of the X BBPD, and found my alignment from the other night was actually pretty good; I made a minor touch up to maximize the green beat.
For the AUX X innolight, I was able to find the actual temperature of the laser crystal when at the correct point, remove the digital offset, and return to the same temperature with the controller front panel dial. This temperature is 42.64 degrees Celcius. There is no diode temperature control on the unit, as far as I could tell, but the maximum green transmitted power through the X arm is about the same as it's ever been, around GTRX=0.5.
My motivation for doing this was that I always have problems remembering the historical typical values for the digital temperature offset. It seems much cleaner to me to set things up such that a beat is visible "at the origin" (i.e. FSS_SLOWDC=0 and SLOW_SERVO2_OFFSET=0) I suppose that this will also depend on what mode of the PMC we're locked to. During my time working today, its been locked near the middle of its range, currently sitting at 145V.
However, for the AUX Y lightwave, I was a little perplexed to find that moving the the laser crystal setpoint around does not apear to change the real laser temperature at all. Thus I could not offload the digital offset in the same way. Aditionally, the lightwave controller does not have the same temperature measuring accuracy as the innolight, even with the back panel voltage readout that is hooked up to the multimeter that lives under the optical table. The best I can tell is around 40.1-40.2 degrees C. Y_SLOW_SERVO2_OFFSET of -10690 counts gives a beat <100MHz at FSS_SLOWDC=0. This is actually very close to the previous operating point, so the same mode seems to still work.
The arms now easily lock on ALS, albeit with higher noise. With arms locked on ALS, POX and POY show >1kHz rms noise.
I gave the PRFPMI locking script a few tries, but it's having problems keeping the PRMI locked. The ALS is a few times noisier than usual, and I haven't revisited the validity of the PRC angular feedforward, so I'm not so surprised. |
11334
|
Thu May 28 21:10:46 2015 |
Koji | Update | Green Locking | ALS-X noise hunting | I have been looking at the X-end ALS setup.
I was playing with the control bandwidth to see the effect to the phase tracker output (i.e. ALS err).
For this test the arm was locked with the IR and the green beat note was used as the monitor.
From the shape of the error signal, the UGF of the green PDH was ~10kHz. When I increased the gain
to make the servo peaky, actually the floor level of the ALS err became WORSE. I did not see any improvement
anywhere. So, high residual error RMS cause some broadband noise in the ALS??? This should be checked.
Then when the UGF was lowered to 3kHz, I could see some bump at 3kHz showed up in the ALS error.
I didn't see the change of the PSD below 1kHz. So, more supression of the green PDH does not help
to improve the ALS error?
Then, I started to play with the phase tracker. It seems that someone already added the LF booster
to the phase tracker servo. I checked the phase tracker error and confirmed it is well supressed.
Further integrator does not help to reduce the phase tracker error.
For the next thing I started to change the offset of the phase tracker. This actually changes
the ALS error level! The attached plot shows the dependence of the ALS error PSD on the phase tracker
output. At the time of this measurement, the offset of -10 exhibited the best noise level.
This was, indeed, factor of 3~5 improvement compared to the zero offset case below 100Hz.
I'm afraid that this offset changes the beat frequency as I had the best noise level at the offset of -5
with a different lock streatch. We should look at this more carefully. If the beat freq changes the offset,
this give us another reason to fix the beat frequency (i.e. we need the frequency control loop.
= Today's ALSX error would have not been the usual low noise state.
We should recover the nominal state of the ALS and make the same test = |
Attachment 1: 150528_ALS.pdf
|
|
11333
|
Thu May 28 17:12:32 2015 |
ericq | Update | IOO | FSS SLOW not engaged: is this intentional? | Yes, I had turned it off while looking for the PSL/X AUX beat, and forgot to turn it back on.
I will post an elog with more detail this evening, but I found a temperature which restored the X green beatnote at its nominal amplitude (-30dBm) with no mode hops within +-1 IR beat GHz, and offloaded the slow offset slider to the X-end laser crystal dial. I will look for the Y beatnote after dinner.
Currently the control room analyzer is hooked up to recieve the Y IR and green beats; no X signals. |
11332
|
Thu May 28 17:00:04 2015 |
Koji | Update | IOO | FSS SLOW not engaged: is this intentional? | I found that FSS SLOW servo is not engaged. Is this intentional test to keep the NPRO temp constant?
This is making the FSS Fast unhappy (~ -7.5V right now). |
11331
|
Thu May 28 16:43:52 2015 |
Steve | Update | PEM | Guralps swapped | Koji and Steve,
The result: bad Guralp x-arm cable.
I will swap the short cables tomorrow at the base.
|
Attachment 1: GursSwapped.png
|
|
11330
|
Thu May 28 15:10:44 2015 |
rana | Update | PEM | Seismic Confusion | You are plotting the STS channels, not Guralp. These are for the Trillium 240 seismometer.
Also, you cannot tell if the seismometer is working by plotting the MEAN trend. That just gives the average and we need the fluctuations. Better off looking at the spectrum like I did last time.
And....its not good enough to just do the bubble. You have to do the mass centering procedure that you and I did last time with the breakout paddle. |
11329
|
Thu May 28 10:42:19 2015 |
Steve | Update | PEM | bad Guralp X-cable |
Quote: |
Tried swapping cables at the Guralp interface box side. It seems that all of our seismic signal problems have to do with the GUR2 cable being flaky (not surprising since it looks like it was patched with Orange Electrical tape!! rather than proper mechanical strain relief).
After swapping the cables today, the GUR2 DAQ channels all look fine: i.e. GUR1 (the one at the Y end) is fine, as is its cable and the GUR2 analog channels inside the interface box.
OTOH, the GUR1 DAQ channels (which have GUR2 (EX) connected into it) are too small by a factor of ~1000. Seems like that end of the cable will need to be remade. Luckily Jenne is still around this week and can point us to the pinout / instructions. Looks like there could be some shorting inside the backshell, so I've left it disconnected rather than risk damaging the seismometer. We should get a GUR1 style backshell to remake this cable. It might also be possible that the end at the seismometer is bad - Steve was supposed to swap the screws on the granite-aluminum plate on Thursday; I'll double check.
|
The Guralps were swapped.
What I did: turned DC power off at 1X1, hand carried them, centered each leveling bubbles, gently locked the jack screws and turned power back on.
ETMY at the east end now has CMG-T40-0008, sn T4157 as channel C1:PEM-SEIS_STS_1_Y_OUT_DQ.........
ETMX at south end now has CMG-T40-0053, sn T4Q17 as channel C1:PEM-SEIS_STS_2_Y_OUT_DQ.........
Conclusion: Guralps are fine. X cable is bad. It was bad 6 months ago when it was made.
We can still swap the 3ft short cables at the granite base if Rana has not done it.
|
Attachment 1: Gurs180dXbad.png
|
|
Attachment 2: swappedGurs.png
|
|
11328
|
Wed May 27 17:14:08 2015 |
ericq | Update | LSC | X Aux Laser crystal temperature changed | Rana suspects that the lack of X beatnote is related to the PSL laser temperature change (ELOG 11294).
I used the information on the wiki and old elogs (wiki-40m, ELOG 6732), to deduce that the new end laser temperatures should be:
- X end-> 38.98 C
- Y end-> 35.80 C
I went out to the X end and found the laser crystal temperature set to 40.87, which is not what the measurements I linked to suggest would be the ideal temperature for the previous NPRO laser temperature of 30.89, which would be 37.02. I could not find any elog describing the choice of this setpoint.
I've changed the X end laser crystal temperature to the value above. I've hooked up the X IR and green beatnotes to go the control room analzyer, and have been looking for the beatnote as I adjust the digital temperature offset, but haven't found it yet...
If this proves totally fruitless, we can just put the lasers back to their original temperatures, since it's unclear if it helped the PC drive noise levels. |
11327
|
Wed May 27 15:20:54 2015 |
ericq | Update | Computer Scripts / Programs | Chiara Backup Hiccup | The local chiara backups are still failing due to vanished source files. I've emailed Max about the summary page jobs, since I think they're running remotely. |
11326
|
Wed May 27 02:53:57 2015 |
ericq | Update | General | ifo recovery log | Given my suspicion of fault with the X Green BBPD, Koji generously provided me with another one that he had confirmed to be working.
However, I turns out I was mistaken. With the existing BBPD, I did indeed witness a beat in the RF output, but it is somehow something like 20dBm smaller than it should be. This is why I missed it the other night. Here's a video of a RF output on a scope, wherein the beat is only barely visible because I've set the trigger level very low. I could not make the beatnote any larger through any alignment motions; I had gotten to this point by doing near/far field overlap on the PSL table.
I'm not sure what could have caused this. Mode mismatch? By eye, the beam spots looked about the same in the near and far fields, and we haven't had to touch the mode matching in quite some time... I've given up on trying to solve this for tonight.
Just for kicks, I hooked up the fiber PD IR beatnotes as inputs to the ALS DFD. The X beat is too small to even really see above the control room analyzer's noise floor, but the Y beat looked big enough. With the arms locked on IR, the phase tracker output RMS was a few kHz, so not even worth thinking about any further. Not so surprising.
Finally, I put back / hooked up everything in its nominal state. |
11325
|
Tue May 26 19:57:11 2015 |
rana | Update | Computer Scripts / Programs | ifoCoupling | Looks like a very handy code, especially with the real statistical tests.
I would make sure to use much smaller excitation amplitudes. Since the coupling is nonlinear, we expect that its only a good noise budget estimator when the excitation amplitude is less than a factor of 3 above the quiesscent excitation. |
|